Rx Imola C1 wash procedure
The following Procedure should be carried out at least twice a year at 1000 hour and 2000
hour Preventative aintenance visits or as re!uired to remove "acterial or #ungal build$
up from the R% Imola &nalyser'
C1 (ash solution is available from Randox order number R%)*+)'
1. Switch off the analyser and the PC.
2. Remove maintenance covers L1 and R1 by loosening 2xM4x screws of each side.
!. P"ll o"t the L#s"b tan$ as shown on the %ict"re below.
4. Loosen bottle to%s and remove all t"bes and the floating sensor.&. 'm%ty the s"b tan$ and fill it with C1 sol"tion "% to the high level mar$.
(. )nsert all t"bes bac$ into the s"b *an$.
+. Re%eat ste%s from ! to ( for the R#s"b tan$.
. ,isconnect water s"%%ly t"bing.
-. Switch on the analyser and the PC.
-. Press cancel on the %o%#"% screen a%%ears. See on %ict"re below
1. /se engineering %assword to login.
Com%osed 0
Chec$ed R,
May 2&th2( ver. 1.
7/25/2019 C1 Wash Procedure Ver1.0
2/3
11. Connect inector syringe as shown on the %ict"re below.
12. 3%en ainte ,#10-./e!uencescreen on the "ser interface software.
1!. *ic$ S"b#*an$ R '0. Set *ime to 2 secs.
14. Press xecto o%en the valve. hile valve is o%en withdraw a f"ll syringe of the li5"id.
1&. ,isconnect syringe t"bing and em%ty the syringe.
1(. Re#connect the syringe.
1+. *ic$ S"b#*an$ L '0 box. Set *ime to 2secs.
1. Press 'xec to o%en the valve. hile valve is o%en withdraw a f"ll syringe of the li5"id.
Leave syringe connected to analyser for 1 min to allow C1 dissolve stains inside the valves.
Com%osed 0
Chec$ed R,
May 2&th2( ver. 1.
7/25/2019 C1 Wash Procedure Ver1.0
3/3
1-. 6ee% the syringe connected and %roceed as follows.
2. Prime analyser once. 7fter %rime is finished to% "% the s"b tan$s with C1 sol"tion.
21. R"n c"vette wash.
22. hen c"vette wash is finished8 r"n c"vette water %lacement and leave analyser for 1 mins to allow
internals stains to be dissolved.
2!. 'm%ty s"b tan$s of all remaining C1 sol"tion.
24. Connect system water s"%%ly.
2&. Perform a C"vette ater ,is%lacement
2(. Prime analyser ! times to remove remaining C1 sol"tion.
2+. P"t maintenance covers bac$ on8 the C1 wash is now com%lete.2. R"n a Rabs and Sabs test to ens"re o%eration is normal.