Additives & Instrument s A member of Measure what you see. BYK-mac BYK-mac COLOR Manual

BYK-mac BYK-mac · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

Embed Size (px)

Citation preview

Page 1: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

Additives & InstrumentsA member of

Measure what you see.

BYK-mac BYK-mac COLOR

Manual

Page 2: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

1

BYK-macBYK-mac COLOR

Manual

Patent pending 269 021 053 E 1012

BYK-Gardner GmbHLausitzer Str. 8D-82538 GeretsriedGermanyTel. 0-800-gardner

(0-800-4273637)+49-8171-3493-0

Fax +49-8171-3493-140

BYK - Gardner USA9104 Guilford RoadColumbia, MD 21046USAPhone800-343-7721

301-483-6500Fax 800-394-8215

301-483-6555

www.byk.com/instruments

Page 3: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

2

Dear customer,thank you for having decided for a BYK-Gardnerproduct. BYK-Gardner is committed to providing youwith quality products and services. We offercomplete system solutions to solve your problemsin areas of color, appearance and physicalproperties. As the basis of our worldwide business,we strongly believe in total customer satisfaction.Therefore, in addition to our products, we offer manyVALUE-ADDED services:- Technical Sales Force- Technical & Application Support- Application and Technical Seminars- Repair & Certification Service

BYK-Gardner is part of the Additives and InstrumentDivision of ALTANA AG, a leading supplier ofadditives for coatings and plastics. Together, weoffer complete and unique solutions for you, ourcustomer.

Thank you for your trust and confidence. If there isanything we can do better to serve your needs, donot hesitate to let us know.

Your BYK-Gardner Team

Page 4: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

3

Table of content

Inhaltsverzeichnis

1. Safety instructions ............................................................................. 5

2. System description and Delivery notes ...........................................10

3. Power supply ....................................................................................12

4. Controls .............................................................................................15

5. Getting started ..................................................................................16

6. Calibrate ...........................................................................................186.1 Cyan Check ................................................................................................ 206.2 Calibrate ..................................................................................................... 216.3 Effect Check ............................................................................................... 23

7. Measurement techniques .................................................................24

8. Measure ............................................................................................258.1 Continue last series .................................................................................. 258.2 Absolute ...................................................................................................... 268.3 Difference ................................................................................................... 268.4 Organizer .................................................................................................... 30

9. Delete ................................................................................................339.1 Last Measurement .................................................................................... 339.2 Checkzone .................................................................................................. 349.3 Test series .................................................................................................. 349.4 Standard ..................................................................................................... 35

10. Data View .........................................................................................3610.1 Test series .................................................................................................. 3610.2 Standard ..................................................................................................... 38

11. Setup .................................................................................................3911.1 Measurement parameters ....................................................................... 3911.2 Instrument .................................................................................................. 4411.3 Calibrate ..................................................................................................... 48

Page 5: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

4

12. Interface ............................................................................................ 49

13. Standards.......................................................................................... 50

14. Technical Data .................................................................................. 51

15. Errors and Warning Messages ......................................................... 53

16. Cleaning and Maintenance .............................................................. 55

17. Service and Certification ................................................................ 57

18. Copyright .......................................................................................... 59

Page 6: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

5

Safety instructions

1. Safety instructions • Before operating the instrument the first time,

please read the operating instructions and takeparticular notice of the safety instructions.

• If you use the unit and accessories properly,there are no hazards to fear.

• This product is equipped with safety features.Nevertheless, read the safety warnings carefullyand use the product only as described in theseinstructions to avoid accidental injury or damage.

• No claims of product liability or warranty can behonored if the device is not operated inaccordance with the operating instructions.

• Keep these instructions for future reference.

• If you pass this instrument to somebody else,make sure to include these instructions.

Page 7: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

6

Safety instructions

The following symbols and terms are used.

This symbol warns of the danger of injury.

This symbol warns of the danger of injury caused byelectricity.

This sign points out additional information.

DANGERThe term DANGER warns of possible severeinjuries and danger to life.

WARNINGThe term WARNING warns of injuries and severematerial damage.

CAUTIONThe term CAUTION warns of slight injuries ordamage.

Page 8: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

7

DANGER injuries possible

• Defects and extraordinary loadsIf safe operation can no longer be presumed,shut down the device and secure it againstunintended operation.

The device must be presumed unsafe to operate:• if visible damage is evident• if the instrument is no longer working• if it has been stored for long periods under

adverse conditions• after harsh treatment during shipping.• Safety advices for batteries and rechargeable

battery pack: Do not crush or dismantle, do notheat or incinerate, do not immerse in any liquid.This may cause explosion or release harmfulsubstances.

• For operation with the external power supply, careshould be taken to ensure the nominal voltage of thepower supply unit (see the manufacturer’s plate onthe power supply unit) matches the voltage suppliedby the power outlet.

• Do not perform any repairs on the unit yourself.The unit must be opened by trained professionalsonly. Please contact our customer servicedepartment in such cases.

• The measurement device and accessories maybe disconnected from the power supply asfollows:Instrument:

a) by removing the battery compartment orb) by removing the rechargeable battery pack orc) by removing the measurement unit from the

docking stationDocking station:

a) by disconnecting the plug from the power supplyunit on the docking station

Safety instructions

Page 9: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

8

Safety instructions

b) by disconnecting the power supply unit plug fromthe socket

The power supply unit can be disconnect from thepower supply by disconnecting the powerconnection line plug from the socket. Make certainthat the power supply unit plug is easily accessible.Use only the power supply connection line includedwith delivery.• When working with the batteries and the

rechargeable battery pack, make certain there isno short circuit on the contacts. Metallic objectsmust not come in contact with the barecontacts.

WARNING severe material damage

• The measurement unit consists of sensitiveoptical and electronic precision parts. Prevent itfrom being dropped, bumped or shaken!

• Avoid exposure to continuous humidity andcondensation. Avoid splashing with water,chemicals or other liquids.

• Please use only accessories that are availablefor the unit.

• Only devices that meet the requirements for low-voltage safety may be connected to the USBinterface.

CAUTION material damage

• Do not allow any foreign objects to get into themeasurement opening.

• Do not expose the unit to direct sunlight forextended periods of time. Do not store it in a hotor dusty environment. Use the instrument casefor storage.

• Align the protection cap when the instrument isnot in use.

Page 10: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

9

• Avoid prolonged high relative humidity and do notallow condensation water.

• Do not use any acetone for cleaning theunit! The unit housing is resistant to manysolvents. For cleaning you should use a soft,moist cloth. Excessive dirt and dust can beremoved with ethanol or cleaning alcohol.

• In case you intend not to use the instrument fora longer period of time, take out the rechargeablebatteries.

• Never use different types of batteries.

Additonal information on use:

• You will find the technical data for all systemcomponents such as the measurement unit andbattery compartment on the respectivemanufacturer’s plates and in the sectionTechnical Data

• Batteries and rechargeable batteries are specialwaste and must therefore not be disposed of withhousehold trash. Make certain to observe thedisposal instructions of the battery orrechargeable battery manufacturer.

Safety instructions

Page 11: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

10

2. System description and Delivery notesBYK-mac COLOR measures color traditionally atfive aspecular angles (15°/25°/45°/75°/110°) with a45° illumination. Additional color measurement”behind” the gloss for color travel of interferencepigments is implemented at -15°. Besides color, theBYK-mac uses additional illuminations to simulatesparkle under 15° /45° and 75°. Diffused illuminationfor graininess evaluation is created by using aBaSO4 coated sphere. Detection is performed by ahigh resolution camera situated perpendicular to thesurface.The instrument is operated by the operate buttonand the mode -scroll wheel. Both, the operatebutton and the scroll - wheel are used to switch onthe instrument. Functions described in this manualin regards to effect measurement are only availablewith BYK-mac.The instrument complies with the followingstandards: DIN 5033, 5036, 6174, 6175-2; ISO7724; ASTM D2244, E308, E2194, E 1164,SAE J1545.

BYK - mac* 23mm Ø aperture CM-6362BYK - mac* 12mm Ø aperture CM-6397BYK - mac Sensor 23mm Ø aperture CM-6363BYK - mac Sensor 12mm Ø aperture CM-6398

Comes complete with:

Multi-angle spectrophotometer, black calibrationstandard, white calibration standard with certificate,cyan and effect checking reference, protective cap,cleaning set for bottom plate, two light protectioncovers, seal replacement kit, *BYKWARE smart-chart software, docking station with USB cable formemory transfer, instrument interface cable foronline data transfer, 2 rechargeable Li-ion batterypacks, battery holder, 4 AA batteries, shortinstructions, operating manual on CD, carryingcase, training.

System description and Delivery notes

Page 12: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

11

BYK - mac COLOR* CM-6395BYK - mac COLOR Sensor CM-6396

Comes complete with:

Multi-angle spectrophotometer, black calibrationstandard, white calibration standard with certificate,cyan checking reference, protective cap, cleaningset for bottom plate, two light protection covers, sealreplacement kit, *BYKWARE smart-chart software,docking station with USB cable for memory transfer,instrument interface cable for online data transfer,2 rechargeable Li-ion battery packs, battery holder,4 AA batteries, short instructions, operating manualon CD, carrying case, training.

Accessories and spare partsBlack calibration standard CM-6332Protective cap 23mm Ø aperture CM-6336Protective cap 12mm Ø aperture CM-6399Docking station CM-6360USB interface cable CM-6337Instrument interface cable, online CM-6413Battery pack CM-6359Cleaning set for Bottom Plate CM-6364Seal Replacement Set CM-6348Light Protection Cover CM-6414BYKWARE smart-chart AW-4831

System description and Delivery notes

Page 13: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

12

3. Power supplyBefore operating the instrument for the first time,please read the operating manual and takeparticular notice of the Safety Instructions.

Powering the instrumentThe measurement unit can be operated either with arechargeable battery pack or with AA (LR6) alkalinebatteries.

Battery:To operate the instrument using batteries, thebattery compartment must be fitted with four 1.5-VAA(LR6) batteries

Ensure that the batteries are correctlyoriented in the compartment according to the

(+) and (-) marks.Use only alkaline batteries (AA /LR6)!

When the battery voltage falls below the requiredoperating voltage, the following message appears onthe display:

“Battery Low !”

Insert the battery box into the battery compartmentand press it down so that the strap moves into thecut-out of the housing.

Set the battery compartment cover in the slotagain. Lock the cover by sliding it downwards until itsnaps.

Power supply

Rechargeablebatterypack

Batterycompartment

Page 14: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

13

Power supply

Rechargeable battery packTo operate the instrument with the rechargeablebattery pack, make sure it is inserted until it locksin place.When inserting the rechargeable battery pack,ensure that its contacts are aligned with those ofthe instrument. See adjacent figure.The capacity of the rechargeable battery packincluded with delivery is sufficient for about 1000measurements. When the voltage of therechargeable battery pack falls below the requiredoperating voltage the following message appears onthe display:

“Battery Low !”

Set the battery compartment cover in the slotagain. Lock the cover by sliding it downwards until itsnaps.Note: To ensure uniform utilization, the rechargeblebattery packs should be exchanged regularlybetween instrument and docking station (weeklyrecommended).

Docking station power supplyPower is supplied to the docking station through theexternal power supply unit. Connect the externalpower supply unit to the docking station. Connectthe appropriate end of the power connection line tothe power supply unit and the plug end of the powerconnection line to the power outlet. Please verifythat the specifications of the power supply unitmatch the power source in terms of current andvoltage.

Page 15: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

14

Charging the rechargeable battery packThe rechargeable battery pack is charged in thedocking station. Charging time is approx. 2 hours.1. Battery pack in the instrument:The lithium ion rechargeable battery will begincharging immediately upon insertion of theinstrument into the docking station. To do this,power must be supplied to the docking stationthrough the corresponding power supply unit.Insert the measurement unit into the docking stationas shown in the illustration.

2. Battery pack in docking station:A second charging shaft is located below the shaftfor the instrument. Insert the second battery packhere for charging, so it will be handy at any time toreplace the other battery when it is discharged.

The compartment for AAbatteries may not be inserted into the

charging shaft.

Charging indicator for:1: additional battery pack2: instrumentIndication light:green: readyred: charging.

Power supply

12

Page 16: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

15

123 4

6

7

5

Controls

4. Controls

Measurement unit

1 Mode scroll wheel: switch on and menu selection

2 esc button: changes between main menu and measurement mode

3 std button: short-cut to difference mode

4 Display for user guidance and measurement values

5 Signal lamps

6 Operate button (measurement button): switch on and measure

7 Docking station and online cable connector

8 Pins to detect correct placement on sample surface

9 Protective cap

89

Page 17: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

16

Getting started

5. Getting startedTurning on the unit

To turn on the unit, press the mode scroll wheel orthe operate button.

The main menu appears.

For the first steps select ”Absolute” from the”Measure” menu.

Place the instrument on the sample and press”operate”. While the reading is in progress, the lightdiodes are green and ”Measuring” is displayed.Keep the instrument stable as long as the greenlights are on. According to the preselected numberof measurements to be averaged, the instrument iswaiting for additional readings.Having finished all readings the result is displayed.

To see all measurement data you can switchbetween the different displays with the scroll wheel.The small black rectangle in the lower left indicatesthe number of pages.

Sparkle and Graininess data is only available whenusing the BYK-mac.

Page 18: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

17

The following error messages might appear duringmeasurement.A warning signal is heard and the light diodes flashred. At the same time, a message appears in thedisplay indicating the type of error.

”Pins not pressed”

Not all of the pins have been pressed thoroughly onthe sample surface (e.g. curvature too high,instrument has been moved during measurement).Confirm the message by pressing the scroll wheelor the operate button and repeat the reading.

”Ambient light”

Appears when ambient light enters themeasurement aperture. Confirm the message bypressing the scroll wheel or the operate button andrepeat the reading.This warning can also appear when measuring brightcolors (white and light silver) in direct sunlight. Verybright light adds to the internal reflection of thecoating.In that case please use the light protection coverwhich is supplied with the instrument. It is a black,flexible foam attachment that snaps on the bottomplate of the instrument. You will find it in the carryingcase under the black trap.

Getting started

Page 19: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

18

6. Calibrate

Calibration Information

StandardsThe instrument comes with 2 calibration standards.The black trap is used to perform the zeroreflectance calibration (black calibration). Alwaysstore the black trap with the protection cover inplace when not in use.

The white standard is fixed in a case that has ahinged cover. It is used to perform the 100%reflectance calibration and to calibrate the camera.Additionally two checking standards are supplied toaudit instrument performance:A cyan standard for color and an effect standard forsparkle and graininess (only with BYK-mac).When calibration is required, the device will displaya message.

The cyan check must be accomplished every 7days. Black and white calibration must beaccomplished every 3 months.It is recommended to perform the effect check onceper month but it can also be performed every 7 daystogether with the cyan check.

Black trap

Protection

Measurement

White Standard

Cyan standard, effectstandard

Calibrate

Page 20: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

19

Calibrate

Calibration Notes

• Dirt or dust in the optics area will cause aninaccurate calibration reading. Therefore, alwaysuse the protective cap, if no readings are taken.

• The standards and black trap should be cleanedperiodically. For cleaning procedure see section"Cleaning and Maintenance“. Please make surethat the standards are not scratched.

• Do not move the instrument while taking acalibration measurement. If motion is detected,an error message will be displayed andcalibration is aborted.

• When moving from cold to warm environment,there is a danger of condensation. For thisreason, you should wait for an appropriateamount of time to allow the optical componentsto adjust before calibrating and using the unit.

Page 21: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

20

6.1 Cyan CheckChoose "Cyan Check" from the "Calibrate" menu.Open the cover of the cyan checking standard andposition the instrument on the standard. Theinstrument snaps into the indentation of thestandard and can only be positioned in onedirection.Choose "OK" with the scroll wheel.

When the cyan check is completed, an affirmationis shown in the display.Confirm by pressing the operate button or the wheelto return to the main menu.

If the reading of the cyan standard was out oftolerance, a black and white calibration isnecessary. The instrument leads you directly to thecalibration display.

Calibrate

Page 22: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

21

6.2 Calibrate

Turn the unit on by pressing either the menu scrollwheel or the operate button.Choose the menu item Calibrate from the Calibratemenu.

Remove the protective cap. Place the instrument onthe black trap. Make sure that the instrument fitsinto the indentation of the black trap.

Choose OK with the wheel and press the operatebutton to perform the black trap reading.

The reading process is indicated on the display.

Choosing Cancel will abortthe calibration procedure and calibration will

not be updated.

Calibrate

Page 23: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

22

Calibrate

When black trap calibration is completed, theinstrument must be placed on the white standard.

Position the instrument as thoroughly on the whitestandard as it is described for the black trap andpress the operate button.

The reading process is indicated on the display.When calibration is complete, an affirmation isshown in the display. Confirm by pressing theOperate button or the wheel to proceed to the Cyancheck.

Choose "OK" with the scroll wheel.

When the cyan check is completed, an affirmationis shown in the display. Confirm by pressing theoperate button or the wheel to return to the mainmenu.

If an error occurs during calibration, try toread the plaque again. If the error message

still occurs, refer to section "Errors and warningmessages“.

Page 24: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

23

6.3 Effect CheckChoose "Absolute" from the "Measure" menu. Thebasic configuration is the average of threemeasurements.Open the cover of the effect checking standard andposition the instrument on the standard. Theinstrument snaps into the indentation of thestandard and can only be positioned in onedirection.

Perform three measurements.

Compare the average of the measurements with therange indicated on the lid.

If the values measured on the effect standard arewithin the printed tolerance range, the requirementsare met.If the mean value is not within the desired tolerancerange, try carefully cleaning the effect standard. Ifthis produces no improvement, please contact ourCustomer Service department.

Calibrate

Page 25: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

24

7. Measurement techniques

In order to obtain accurate and repeatablemeasurements, the bottom of the instrument mustbe aligned flat on the sample surface. Do not movethe instrument while measuring.To perform a measurement, press the operate keyor set-up the instrument to automatically perform ameasurement when the four pins are pressed down(see chapter ”Setup/Instrument”).You can adjust the number of readings in “Setup/Measurement parameters/Sample/Statistic/Numberof measurements”. When using an Organizer, thenumber of readings is predefined.You can adjust the sensitivity of the pins to thesample curvature. Refer to the paragraph “Samplecurvature” in chapter “Setup/Instrument”.

Pressing the mode wheel while being inmeasurement mode opens a submenu:Main menuInterupt statisticDelete last measurementDelete checkzone.

By pressing the ”esc” button you can switchbetween measurement mode and the main menu.

Measurement techniques

Page 26: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

25

8. MeasureSelect the menu item Measure by pressing theoperate button/scroll wheel.A selection menu appears.

”Continue last series”: To continue the last testseries. If no series is stored, it will be created.

”Absolute”: To measure absolute data.

”Difference”: To measure differences. Select anexisting standard, create a new one or use thefunction „Auto Standard“.

”car_01”: To take readings according to a predefinedmeasurement sequence (organizer).

8.1 Continue last series

”Continue last series” leads you directly to themeasuring sequence of the last series. Pressoperate to start the reading.

Measure

Page 27: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

26

Measure

8.2 AbsoluteYou are asked if you like to continue the test series.

Confirm by pressing the scroll wheel leads you tothe last test series.You can continue to take readings by pressing”operate”.

Choosing ”No” opens a new test series, starting withSample 001. Press operate to continue.

8.3 DifferenceChoosing ”Difference” leads you to a selectionmenu. You can choose an existing standard, createa new standard or use the function „Auto Standard“.

std - ButtonYou can easily access the ”Difference” menu bypressing the std button on top of the instrument.You can choose an existing standard or you cancreate a new standard as described before.

Page 28: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

27

New Standard

You are asked to input a name. You can accept thepredefined suggestion or change the character byturning the scroll wheel. When you press the wheel,the highlighted position moves to the next character.If you did a wrong input, you can move backwardwith the esc button.

After inputting the name, you can input the numberof readings by using the scroll wheel.

Then measure the standard by pressing ”operate”.

After the readings you are asked to copy settings.You can choose among several predefined settings:Solid: Displays only 45° angle and no effect parameters.Metallic: Displays all 6 color angles and effectparameters.

The instrument is now ready to take samplereadings.

Measure

Page 29: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

28

Measure

Existing standard

Choose the appropriate standard from the list.

You are asked to continue the last series.

If you choose ”Yes”, you can directly continue theseries by pressing the ”operate” button.

In case you choose ”No”, a new series is started.

Page 30: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

29

Measure

Auto standardThe function „Auto Standard“ is displayed if at leastone standard is stored in the memory. Whenchoosing this function the closest standardcompared to the sample is automatically recalled.

Choose „Auto Standard“ from the menu and pressoperate to start the sample reading.

If only one standard is close to the sample, thisstandard is automatically selected and the initialsample reading is compared to this standard.

If more than one standard is close to the sample,you have the choice to select from the list ofstandards.

If no standard is close, the message „No goodmatch“ is displayed and you can either measure anew standard or select from the list of all standardsstored in the instrument.

Page 31: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

30

8.4 OrganizerAn organizer file defines a measurement sequencefor user guidance, e. g. measuring a car body withseveral checkzones. These files can be generatedwith the smart-chart software and transferred to theinstrument.

Once an Organizer is selected, a menu appears forParameter 2 (Color).

After you have selected the appropriate color, aselection menu appears for Parameter 3(e.g.automotive paint line).

If ”Input Comment” is activated in the organizer, youare prompted to enter additional information.If “Input ID” is activated, you are prompted to enter acode, e.g. the vehicle ID.After definition of all parameters you are asked tostart a new test series.

If the answer is ”No”, the instrument returns to themain menu.If the answer is ”Yes”, the instrument goes to themeasurement mode.In case the standard (color) has not been stored inthe instrument, you can measure it now. Pleaseproceed as described at ”Measure/New standard”.

Measure

Page 32: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

31

Measure

After the standard has been measured, or when it isalready stored in the instrument, the name of thecheckzone to be measured first, appears on the leftside of the display.The number of performed and predefinedmeasurements (0 of 2) appears in the upper rightcorner.Press the operate button to start the reading.

Once the number of measurements for thecheckzone is reached, a double audio signal isheard.The display shows the results of the measurementand indicates that the measurement of thecheckzone is complete (e.g.2/2).

Pressing briefly the operate button allows the nextcheckzone to appear in the display.Once all checkzones have been measured, theinstrument returns to the Measure menu.

Pass/Fail Mode

In the software smart-chart tolerances can be set foreach color. If Pass/Fail mode is selected in theorganizer and the measurements are within thetolerance range, PASS is indicated in the display. Ifnot, FAIL.

To see which measurement value is out of tolerance,you can scroll through the display pages with thescroll wheel.The appropriate value will be highlighted.

Page 33: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

32

Measure

During the measurement you have access toadditional functions by pressing the mode wheel.In case you took a false reading, you can delete thelast measurement or a complete checkzone.

”Interrupt statistic” must be activated when theorganizer is created. If not, you will get aninformation on the display that it is not allowed.

Page 34: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

33

Delete

9. DeleteSelect the menu item Delete by pressing theoperate button/scroll wheel.A selection menu appears.”Last Measurement”: To delete the lastmeasurement of a checkzone.”Checkzone: To delete the whole checkzone.”Test series”: To delete a test series.”Standard”: To delete a standard.

9.1 Last MeasurementIn case you took a false reading you can deletemeasurements.Select the menu item ”Last Measurement” bypressing the scroll wheel or the operate button.

The last measurement is deleted and the number ofthe performed measurements in the upper rightcorner of the display is diminuished by one.

Page 35: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

34

9.2 CheckzoneTo delete a whole checkzone, choose the menuitem ”Delete checkzone” and press the mode wheel.

.

The instrument is now ready again to measure thecheckzone.

Delete

9.3 Test seriesWhen you choose ”Test series” from the menu, asubmenu opens.You can delete all test series or individual testseries taken in absolute, difference or organizermode.

To delete the test series in absolute mode, choose”Absolute” from the menu.A list of the stored test series is displayed.Choose the appropriate test series and press theoperate button or the mode wheel.

To confirm choose ”Delete” with the mode wheel andpress the wheel or the operate button.If you choose ”Cancel” the instrument returns to theselection menu.

Page 36: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

35

Delete

To delete test series in difference mode, you have tochoose the appropriate standard first and then thetest series to be deleted. Test series in Organizermode can be deleted accordingly. The organizeritself can only be deleted with smart-chart.

9.4 Standard

With this function you can delete standards.Choose the appropriate standard from the list andpress the mode wheel or the operate button.

To confirm, choose ”Delete” with the mode wheeland press the wheel or the operate button.If you choose ”Cancel” the instrument returns to theselection menu.

If a test series is stored in the instrument, that usesthe standard, it can not be deleted. You have todelete the test series first.

Page 37: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

36

10. Data ViewYou can use this function to display the readings ofmeasurements stored in the instrument.

10.1 Test seriesAbsoluteTo view the data of readings taken in absolute mode,choose ”Absolute” from the menu.A list of all test series taken in absolute mode isdisplayed.Choose the appropriate test series and press theoperate button or the mode wheel.

The data of the first reading is displayed.Use the operate button to switch to the nextreading.

DifferenceTo view the data taken in difference mode, choose”Difference” from the menu.Then select the standard and the appropriate testseries by pressing the operate button or the modewheel.

Data View

Page 38: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

37

Data View

Organizer

To view the data of an organizer, choose theappropriate organizer from the list.

Then select parameter 2 (color) and the appropriatetest series by pressing the operate button or themode wheel.

The data of the first checkzone is displayed. Youcan move through the different displays with thescroll wheel to see all measurement data. The smallblack rectangle in the lower left corner indicates thenumber of pages.

With the operate button you can move to the nextcheckzone.

Page 39: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

38

Data View

10.2 Standard

To view the data of a standard, choose theappropriate standard from the menu.The data of the standard is displayed.

Page 40: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

39

11. SetupIn the Setup menu you find functions to adjustgeneral settings for:Measurement parametersInstrumentCalibrate

11.1 Measurement parameters

Measurement parameters can be adjusted forabsolute mode or any individual standard. Select thestandard you want to change and a submenuopens:StatisticColorEffect (only with BYK-mac)Sample Temp. Alarm

11.1.1 Statistic

The function ”Statistic” lets you set the number foraverage readings and switch ”SMC” (StatisticalMeasurement Control) on or off.

11.1.1.1 Number of measurements

This function allows selection of the number ofreadings to be taken per sample. To change thecharacter, turn the scroll wheel. When you press thewheel, the highlighted position moves to the nextcharacter. Pressing operate completes the processand the instrument returns to the selection menu. Ifyou entered a wrong input, you can move backwardwith the esc button.

Setup

Page 41: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

40

11.1.1.2 SMC - Statistical Measurement ControlHighlight the function with the scroll wheel. Pressingthe wheel activates or deactivates the function. IfSMC is activated a checkmark appears.

For SMC mode you must preset a minimum numberof measurements ( n > 3 ). To keep the requirednumber of measurements as low as possible, adefault standard error is stored in the instrument. Incase of uniform surfaces, the standard error is oftenreached after the minimum number ofmeasurements. In case of non-uniform surfaces,e.g. problem areas, additional measurements will berequired by the instrument.

11.1.2 Color

This function lets you choose the measurementangles, scales and illuminant/observer.

Setup

Page 42: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

41

Setup

11.1.2.1 Angles

You can choose individual angles or ”All”.

11.1.2.2 Scales

Here you can decide according to which color scaleresults are calculated and how they are displayed.

In the sub menu you may find the following:• Solid:

displays only 45° angle• Metallic:

displays all 6 color angles

Page 43: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

42

Setup

11.1.2.3 Illuminant/Observer

You can choose among the illuminants/observersshown on the left side.

11.1.3 Effect (only available with BYK-mac)

This function lets you choose the effect parametersto be displayed.

You can choose individual effect parameters or ”All”.

Page 44: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

43

Setup

11.1.4 Sample Temp Alarm

The device has a built in temperature sensor. If thealarm is activated, a checkmark appears.You can set minimum and maximum values.

Min:Choose the appropriate sample temperature and analarm is set to alert when the temperature is below.Max:Choose the appropriate sample temperature and analarm is set to alert when the temperature isexceeded.

Page 45: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

44

Setup

11.2 Instrument

In the Setup menu you find functions to adjustgeneral settings of the instrument:

11.2.1 Sample curvature

You can adapt the sensitivity of the pins dependenton the sample curvature.Flat: 4 pins < 0.1mmLow: 3 pins < 0.1mm, 1 pin < 0.3mmMedium: 3 pins < 0.3mm, 1 pin < 0.9mmHigh: 3 pins < 0.6mm, 1 pin offOff: pins are deactivated, but it is ensured

that no ambient light will enter theaperture.

11.2.2 Automatic measurement

The instrument can be set-up to execute a readingby just pressing the four pins.When the automatic measurement is activated, acheckmark appears.

Page 46: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

45

Setup

11.2.3 Beeper

This menu option turns the beeper on or off. Use thescroll wheel to move the cursor to Beeper and pressthe wheel.When the beeper is activated, a checkmarkappears.

11.2.4 Language

You can use this menu to select the displaylanguage.If a foreign language is activated accidently, you canfind the language settings by following the * symbolin the menus.Use the scroll wheel to move the cursor to thedesired language and press the wheel.

11.2.5 Info

This menu displays the following information aboutthe device:

• Serial No.• Catalog No.• Version• Date of last calibration• Free Memory• Battery Capacity

Page 47: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

46

Setup

11.2.6 Date/Time

The unit contains an integrated clock. This makesdate and time of the measurement available for datatransfer to a PC. Date and time are not lost evenwhen the battery is changed. If necessary, adjustthe data by using the scroll wheel.

11.2.7 Display Time

To save energy, the unit automatically turns off aftera certain amount of time. This time period can beset between 15 and 99 seconds.

If you entered a wrong input, you can movebackward with the esc button.

11.2.8 Advanced

You can use this menu item to activate:• Interrupt statistic• Input Parameter 3• Input Comment• Input ID• Input Checkzone

Page 48: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

47

11.2.8.1 Interrupt statisticIf interrupt statistic is activated you can interrupt acheckzone (sample) before reaching the presetnumber of measurements.

11.2.8.2,3 Input Parameter 3 and CommentWhen active, these parameters allow to assignindividual names for identification of a new testseries.

11.2.8.4 Input IDWhen this function is active, you can enter anidentification code e.g. vehicle ID number for eachnew test series.

11.2.8.5 Input CheckzoneWhen active, you can enter a name for each sampleduring a test series. When deactivated, theinstrument automatically assigns sample namesincrementally, beginning with SAMPLE 01.

Setup

Page 49: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

48

Setup

11.3 Calibrate

You can perform a cyan check and a completecalibration.Please refer to the chapter ”Calibration”.

Page 50: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

49

12. InterfaceTransfer of data stored in the memory of theinstrument takes place through the docking station.It contains the USB interface connecting theinstrument with a PC.

To transfer data, the connection cable must beconnected to a USB port on the computer. For theposition and assignment of the socket, please referto your computer manual.

The connection point for the USB cable is located atthe back of the docking station. Plug in the cableincluded with delivery.To transfer data, the instrument must be insertedinto the docking station.

To transfer data online from the instrument to thePC, the instrument interface cable can be used. Itmust be pluged directly into the cable connector onthe instrument.

Interface

Page 51: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

50

13. Standards

DIN 5033 Colorimetry; basic concepts.

DIN 5036 Radiometric and photometric properties of materials;definitions characteristic.

DIN 6174 Colorimetric evaluation of colour differences ofsurface colours according to the CIELAB formula.

DIN 6175-2 Tolerances for automotive paintsPart 2: Goniochromatic paints

ISO 7724 Paints and varnishes - Colorimetry

ASTM D 2244 Standard Test Method for Calculation of ColorDifferences From Instrumentally MeasuredColor Coordinates.

ASTM E 308 Standard Practice for Computing the Colors ofObjects by Using the CIE System.

ASTM E 1164 Standard Practice for Obtaining SpectrophotometricData for Object-Color Evaluation.

ASTM E 2194 Standard Practice for Multiangle Color Measurementof Metal Flake Pigmented Materials

SAE J 1545 Instrumental Color Difference Measurement forExterior Finishes, Textiles and Colored Trim

Standards

Page 52: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

51

14. Technical Data

General technical data:

Temperature range 10 °C to 42 °C (50 °F to 110 °F)for operation0 °C to +60 °C (32 °F to 140 °F )for storage

Rel. humidity Up to 85% non-condensing/35°C (95 °F)

Measurement unit:

ColorMeasuring Geometry Illumination 45°,

-15°,15°, 25°, 45°, 75°, 110° aspecular viewingMeasuring Area Ø23mm (0.91 in.) or 12mm (0.47 in.)Spectral Range 400 - 700 nm, 10 nm resolutionMeasurement Range 0 to 400 % reflectanceMeasuring timeBYK-mac < 6 secondsBYK-mac COLOR < 4 secondsRepeatability 0.02 dE*, 1

10 consecutive readings on white tileReproducibility 0.2 dE*, 1

average on 12 BCRA II tilesColor scales dE*, dE CMC, dE 94,dE DIN 6175Illuminants A; C; D50; D65; F2; F7; F11; F12Observer 2°; 10°Effect only with BYK-macMeasurement Geometry 15°, 45°, 75° and diffused illumination

perpendicular viewingEffect parameters dS, dS_a, dS_i, dGRepeatability S_a / S_i: 5% or > 0.50 / G = ± 0.05, 1Reproducibility S_a / S_i: 10% or > 1.00 / G = ±0.15, 1Memory 1000 Standards/SamplesPower supply Rechargeable battery pack or 4 mignon AA batteries

(alkaline or rechargeable)Dimensions (LxWxH) 218 x 81 x 147mm (8.6 x 3.2 x 5.8 in.)Weight approx. 1300g/ 2.86 lbsInterface USB-B

Technical Data

Page 53: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

52

External power supply:

External power supply Docking stationInput 90 - 264 VAC, 50 - 60 Hz max. 0,8 AOutput 12 VDC, 2,5 A

Technical Data

Page 54: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

53

Errors and Warning Messages

15. Errors and Warning Messages

Info

WarningThe surface The sample temperature is lower than thetemperature is under ..°C sample temperature alarm.

WarningThe surface The sample temperature is higher than thetemperature is over.°C sample temperature alarm.

Single measurements can You can delete measurements only within theno longer be deleted. active checkzone.

Calibration interval expired: Appears when the instrument is switched onEffect, Color, Cyan and the calibration interval has expired.

Perform calibration.

Error

Memory full The maximum number of measurements hasbeen stored. Delete data which is no longerneeded.

Pins not pressed! Appears when sensor pins are notpressed during a reading. Confirm and repeatthe reading.

Wrong Standard The wrong standard was used for blackUse Black Trap calibration.

Wrong Standard The wrong standard was used for whiteUse White Standard calibration.

Page 55: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

54

Error

Instrument temperature Temperature is outside the operatingout of range temperature range.

Ambient light Appears when there is too much ambientlight during measurement or when measuringbright colors (white and light silver) in directsunlight. Please refer to chapter 5“Getting started/Ambient light“

Cyan Check Appears when check on cyan tile failed.

Please calibrate Appears when the instrument is notcalibrated and the user wants to take areading.

Humidity too high Humidity is outside the operatinghumidity range.

Battery Low Replace the batteries or recharge therechargeable battery pack.

Input required Data must be input.

Errors and Warning Messages

Page 56: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

55

Cleaning and Maintenance

16. Cleaning and Maintenance• Do not insert any objects into the measurement

aperture for cleaning. The instrument could getdamaged.

• Do not use any acetone for cleaning theunit! The instrument housing is resistant to anumber of solvents, but cannot be guaranteed towithstand all chemicals. You should thereforeuse a soft, moist cloth for cleaning. For cleaningexcessive dirt, use ethanol or cleaning alcohol.

• Do not attempt to make any repairs yourself! If amalfunction occurs on your measuring device,our Customer Service department will be happyto help you as quickly as possible.

Cleaning standards

Do not use any acetone!The accuracy of the measurement can besignificantly impacted by using dirty or damagedstandards.Since the surfaces of the standards are highlysensitive, cleaning must be undertaken withgreat care.To clean standards, use a new lint-free cloth,dust-free lens paper or an optical cloth.Apply only slight pressure as you clean andmake certain there are no large particles stuck inthe cloth that could damage the surface.For dirt that is difficult to remove, use an opticalcloth dipped in liquid. Then wipe the surface witha dry optical cloth.It is highly recommended to handle thestandards with great care. They should alwaysbe stored enclosed.

Page 57: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

56

Cleaning the Black TrapFrom time to time the black trap should be cleanedwith clean, dry air to remove any dust orcontamination.The black trap is taken apart by compressing thelower part on both sides of the case with yourfingers, and then separating the two sections.The black glass should be cleaned using a clean,lint-free cloth. In the second step, wipe off the glasswith alcohol. You must let the glass dry completelybefore taking a black trap reading.

Exact calibration is not possible unless thestandard is in perfect condition. If the condition ofthe standard seems doubtful because of itsappearance or measurement errors, we will behappy to check it for you.

Cleaning the Bottom PlatePlease use only the cleaning set included withthe instrument.Place the instrument on the white paper pad.Repeat placing the instrument several times tomake the pins grease-free.Then place the instrument on the blue rubber mat.Repeat placing the instrument several times tomake the pins lint-free.Do not drag the instrument over the surface!Please make sure that all rubber pins are stillattached after cleaning.Clean the blue rubber mat with Isopropanol and amicrofiber cloth.

Cleaning and Maintenance

White paper pad

Blue rubber mat

Page 58: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

57

17. Service and Certification

ServiceBesides the repair of your instrument we offer thefollowing additional services:First diagnosis on the telephone or by e-mailCall us or send us an e-mail and we will try to solveyour problem. If this is not successful, please sendus the instrument for repair.Preventive maintenance, calibration, andrecertificationFor precautionary reasons we recommend regularpreventive maintenance. We carry out this preventivemaintenance automatically when you send us yourinstrument for maintenance and recertification. Weclean the optics, check all functions, test and, ifrequired, adjust the measured values by usingreference standards. You will receive a certificate,which includes the retraceability to internationalstandards.LoanersDuring the period of repair we furnish you with aloaner on request and availability.Maintenance agreementIn case you want to make sure that the necessarymaintenance is being done on a regular basis andon time, we recommend a maintenance agreement.Extended warranty contractsFurthermore, you can request an extended warrantycontract for additional 12 months.Ordering information:SP-6362 Calibration service BYK-macSP-6395 Calibration service BYK-mac COLORSE-6340 Extended warranty BYK-macSE-6395 Extended warranty BYK-mac COLOR

Service and Certification

Page 59: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

58

Service and Certification

Service Centers for BYK-Gardner products

GermanyBYK-Gardner GmbHLausitzer Strasse 882538 GeretsriedGermanyPhone:+49-8171-3493-0Fax: +49-8171-3493-166USABYK-Gardner USA9104 Guilford RoadColumbia, MD 21046USAPhone:+1-301-483-6500Fax: +1-301-483-6555ChinaBYK-Gardner ChinaInstruments ServiceBYK (Tongling) Co., Ltd.Shanghai Branch3/F, Bldg A, No8, Lane 1305Huajing Road, Xuhui DistrictShanghai 200231P.R. ChinaTel.: +86(021)6496-7931Fax: +86(021)6496-7932BrazilBYK-Gardner Latin AmericaRua Itaporanga, 340Bairro Paraíso -Santo André-SPCEP 09190-640BrazilTel.: +55-11-2147-1199Fax: +55-11-2147-1168

Page 60: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

59

18. Copyright

This instruction manual is an important part of this instrument. It contains es-sential information about setting up, placing in service and use. If you passthe device on to another user, please ensure that the instruction manual isincluded with the instrument. The manual must be studied carefully beforeworking with the equipment. Please contact your regional service office ifyou have any questions or require additional information about the device.

The technology and fittings are based on state-of-the art optic andelectronic technology. New developments and innovations are constantlybeing integrated into the equipment. Thus, the diagrams, dimensions, andtechnical data used in this manual may have changed as a result of adaptingthe device to new information and improvements.

© Copyright 2010 BYK-Gardner GmbHAll rights reserved

No portion of the software, documentation or other accompanying materialsmay be translated, modified, reproduced, copied or otherwise duplicated(with the exception of a backup copy), or distributed to a third party, withoutprior written authorization from BYK-Gardner GmbH. In any case, thisrequires the prior written consent of BYK-Gardner.

BYK-Gardner GmbH offers no guarantee that the software will functionwithout error or that the functions incorporated therein can be executed in allapplications and combinations selected by you.

No liability other than as provided by law is assumed for direct or indirectdamage sustained in association with the use of the instrument, the softwareor documentation.

BYK-Gardner GmbH reserves the right to update the software and writtendocumentation without prior notice.

Copyright

Page 61: BYK-mac BYK-mac  · PDF fileBYK-mac BYK-mac COLOR Manual ... complete system solutions to solve your problems ... • Do not use any acetone for cleaning the unit!

269 021 053 E 1012