Burning Through the Confusion:See How Combustion IC Provides Superior Halide Analysis

Jay ShefferMetrohm USAWednesday, October 17, 2018

Gulf Coast ConferenceAbstract # 187

Page 2: Burning Through the Confusion: See How Combustion IC ......Burning Through the Confusion: See How Combustion IC Provides Superior Halide Analysis Jay Sheffer Metrohm USA. Wednesday,

Organic Halides in Crude

• Rarely occur naturally in crude oils• Introduced at production sites, pipelines, or tanks in

cleaning operations• Produce HCl, HF, etc., in reforming and hydro-

treating processes• Leads to corrosion and catalyst poisoning

Speciation of halides is important to trace the contamination source. Current specifications expect < 2ppm org Cl in

crude oil

Detection at low levels (< 5ppm) is of most importance to industry

Page 3: Burning Through the Confusion: See How Combustion IC ......Burning Through the Confusion: See How Combustion IC Provides Superior Halide Analysis Jay Sheffer Metrohm USA. Wednesday,

Recent ASTM Development

• New methodology was developed in ASTM D02 for additional testing methodologies for organic chloride in crude:

1) Updated ASTM D4929: organic chloride measurement in crude oil by distillation and XRF detection (new Procedure C)

2) New ASTM D8150: organic chloride measurement in crude oil by distillation and Combustion IC detection

Page 4: Burning Through the Confusion: See How Combustion IC ......Burning Through the Confusion: See How Combustion IC Provides Superior Halide Analysis Jay Sheffer Metrohm USA. Wednesday,

Development of ASTM D4929 and D8150

Test methods consist of:• Distillation to 400ºF• Water wash (removes inorganic chloride)• Caustic wash (removes sulfur)• Detection by:

1) D4929-Test Method A: Sodium biphenyl reduction and potentiometry, or

2) D4929-Test Method B: Combustion and microcoulometry, or

3) D4929-Test Method C: X-ray fluorescence spectroscopy (including MWDXRF, MEDXRF, and EDXRF)

4) D8150-Test Method: Combustion Ion Chromatography

Page 5: Burning Through the Confusion: See How Combustion IC ......Burning Through the Confusion: See How Combustion IC Provides Superior Halide Analysis Jay Sheffer Metrohm USA. Wednesday,

Method Confusion???

Combustion IC

Page 6: Burning Through the Confusion: See How Combustion IC ......Burning Through the Confusion: See How Combustion IC Provides Superior Halide Analysis Jay Sheffer Metrohm USA. Wednesday,

Precision: Original D4929 - Methods A & B

• Test Method A(Sodium biphenyl reduction, potentiometry)

r = 0.32 ( X + 0.33)0.644

R = 0.7 ( X + 0.33)0.644

• Test Method B(Combustion, microcoulometry)

r = 1.01 ( X – 0.17)0.467

R = 1.32 ( X – 0.17)0.4670

1

2

3

4

5

6

0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20

Std

Dev

Cl in Crude Conc

Method A and B: repeatability and Reproducibility

Method A r

Method A R

Method B r

Method B R

Data from ILS reports for D4929 and D8150 development and available from ASTM

Page 7: Burning Through the Confusion: See How Combustion IC ......Burning Through the Confusion: See How Combustion IC Provides Superior Halide Analysis Jay Sheffer Metrohm USA. Wednesday,

0

1

2

3

0 1 2 3 4 5

Std

Dev

Cl in Crude Conc

D4929 - Methods A and B: repeatability (r) and Reproducibility (R)

Method A r

Method A R

Method B r

Method B R

• Concentrations of chloride < 5ppm is of most importance to industry

• Specifications often dictate organic chloride < 2ppm

Precision: Original D4929 - Methods A & B

Data from ILS reports for D4929 and D8150 development and available from ASTM

Page 8: Burning Through the Confusion: See How Combustion IC ......Burning Through the Confusion: See How Combustion IC Provides Superior Halide Analysis Jay Sheffer Metrohm USA. Wednesday,

Precision: D4929 Method C - XRF

0

0.5

1

1.5

2

2.5

3

0 1 2 3 4 5

Std

Dev

Cl in Crude Conc

Method C: repeatability and Reproducibility

MWDXRF r

MWDXRF R

MEDXRF r

MEDXRF R

EDXRF r

EDXRF R

Data from ILS reports for D4929 and D8150 development and available from ASTM

Page 9: Burning Through the Confusion: See How Combustion IC ......Burning Through the Confusion: See How Combustion IC Provides Superior Halide Analysis Jay Sheffer Metrohm USA. Wednesday,

Precision: D8150 - CIC

0

0.5

1

1.5

2

2.5

3

0 1 2 3 4 5

Std

Dev

Cl in Crude Conc

CIC + XRF + Method A and B: repeatability and Reproducibility

CIC r

CIC R

Data from ILS reports for D4929 and D8150 development and available from ASTM

Page 10: Burning Through the Confusion: See How Combustion IC ......Burning Through the Confusion: See How Combustion IC Provides Superior Halide Analysis Jay Sheffer Metrohm USA. Wednesday,

Comparing Precision of D4929 and D8150

0

0.5

1

1.5

2

2.5

3

0 1 2 3 4 5

Std

Dev

Cl in Crude Conc

CIC + XRF + Method A and B: repeatability and Reproducibility

CIC r

CIC R

MWDXRF r

MWDXRF R

MEDXRF r

MEDXRF R

EDXRF r

EDXRF R

Method A r

Method A R

Method B r

Method B R

Data from ILS reports for D4929 and D8150 development and available from ASTM

Page 11: Burning Through the Confusion: See How Combustion IC ......Burning Through the Confusion: See How Combustion IC Provides Superior Halide Analysis Jay Sheffer Metrohm USA. Wednesday,

Spike Recovery: D8150 - CIC

Sample ID Expected Concentration(Approximate)(mg/kg)

CIC Mean Concentration(mg/kg)

SA 1 1.024

SB 12 11.771

SC 10 8.435

SD 3 3.156

SE 0 (no spike) 0.185

SF 1 1.007

SG 5 4.946

SH 0 (no spike) 0.154

SI 6 4.069

SJ 5 5.076 0

1

2

3

4

5

6

7

0 1 1 2 2 3 3 4 4 5 5

Reco

vere

d Co

ncen

tratio

n, m

g/kg

Cl

Spiked Concentration, mg/kg Cl

Average Recovery of Org Cl Spike by CIC and XRF(multiple samples spiked with 0,1,5ppm were averaged as a single point)

CIC Averaged

Ideal Recovery

13Data from ILS reports for D4929 and D8150 development and available from ASTM

Page 12: Burning Through the Confusion: See How Combustion IC ......Burning Through the Confusion: See How Combustion IC Provides Superior Halide Analysis Jay Sheffer Metrohm USA. Wednesday,

Challenges with D4929 Test Methods A and B

• Sodium biphenyl reduction and potentiometry• Combustion and microcoulometry

• Interference from other titratable halides including HBr and HI• Interference with oxyhalides HOBr and HOI• Detection limit of 1ppm Cl

Page 13: Burning Through the Confusion: See How Combustion IC ......Burning Through the Confusion: See How Combustion IC Provides Superior Halide Analysis Jay Sheffer Metrohm USA. Wednesday,

Challenges with D4929 Test Method C

• EDXRF, MEDXRF, MWDXRF

• Interference with sulfurS Ka = 2.3 keVS KB = 2.45 keVCl Ka = 2.55 keV• When using Ag anode x-ray tube

(less overlap than Rh or Pd anode tubes)

100ppm Cl + 5,000ppm S interferentHigh Sulfur Crudes = 1ppm Cl + 10,000ppm S interferent

Page 14: Burning Through the Confusion: See How Combustion IC ......Burning Through the Confusion: See How Combustion IC Provides Superior Halide Analysis Jay Sheffer Metrohm USA. Wednesday,

Detection Limits Comparison

• D4929 Test Methods A, B:Cl < 1ppm

• D4929 Test Method C:• EDXRF, MEDXRFCl < 1ppm• MWDXRFCl < 0.13 to 0.3ppm

• D8150 Combustion IC• Cl < 0.1ppm• Can be pushed lower with good techniques

Page 15: Burning Through the Confusion: See How Combustion IC ......Burning Through the Confusion: See How Combustion IC Provides Superior Halide Analysis Jay Sheffer Metrohm USA. Wednesday,

Combustion IC Advantages

• Determination of chloride without interference from other species

• Determination of individual halides and sulfur in one sample run

• No need to remove sulfur (eliminating caustic wash)

• Determine total, individual halides including organic and inorganic sources (without water wash)

• Determine organic, individual halides (with water wash)

Allows more flexibility with your analysis

Page 16: Burning Through the Confusion: See How Combustion IC ......Burning Through the Confusion: See How Combustion IC Provides Superior Halide Analysis Jay Sheffer Metrohm USA. Wednesday,

• Automated pyrohydrolytic combustion• Heat + Oxygen - Breaks covalent bonds to remove halogens from

organic molecules

Combustion Ion Chromatography (CIC)

Page 17: Burning Through the Confusion: See How Combustion IC ......Burning Through the Confusion: See How Combustion IC Provides Superior Halide Analysis Jay Sheffer Metrohm USA. Wednesday,

• Capable of performing simultaneous determination of sulfur and individual halides in a single run• F, Cl, Br• S (as SO4)

Combustion IC Advantages

Page 18: Burning Through the Confusion: See How Combustion IC ......Burning Through the Confusion: See How Combustion IC Provides Superior Halide Analysis Jay Sheffer Metrohm USA. Wednesday,

Combustion IC Advantages

Column Metrosep A Supp 5 150 x 4.0mm

Eluent3.2mmol/L Na2CO3+1.0 mmol/L NaHCO3

Eluent flow 0.7mL/min

Combustion temperature 1050°C

Combustion volume 100µL

Absorption solution Ultra-pure water

Injection volume 200µL

Detector Suppressed Conductivity

Cl

SO4-2

ASTM D8150 - CIC

Component Name

Retention Time[min]

Organic Chloride Concentration Naphtha[mg/kg]

OrganicChlorideConcentration Crude[mg/kg]

Chloride 6.40 12.037 2.923

Unwashed Naphtha Distillate

Page 19: Burning Through the Confusion: See How Combustion IC ......Burning Through the Confusion: See How Combustion IC Provides Superior Halide Analysis Jay Sheffer Metrohm USA. Wednesday,

Summary

Combustion Ion Chromatography demonstrates superior performance, selectivity, sensitivity, and lack of interferences.• Able to determine individual

concentrations of halides and sulfur in a variety of matrices

• Permits refineries to control corrosion and associated costs by predicting, locating, and preventing it

• Becoming widely accepted with several new methods under development in ASTM

Page 20: Burning Through the Confusion: See How Combustion IC ......Burning Through the Confusion: See How Combustion IC Provides Superior Halide Analysis Jay Sheffer Metrohm USA. Wednesday,

Metrohm at GCC 2018

Abstract # Title Time

251 Simple, Efficient Petrochemical Analysis with NIR and Raman Spectroscopy

10:30am – 11:00am

193 Better Surfactant Analysis and a New Approach for Cloud Point

1:30pm – 2:00pm

194 Robotic Titration for Consolidating Petrochemical Titrations 2:00pm – 2:30pm

Technical Presentations: Wednesday 10/17, Wysteria

Page 21: Burning Through the Confusion: See How Combustion IC ......Burning Through the Confusion: See How Combustion IC Provides Superior Halide Analysis Jay Sheffer Metrohm USA. Wednesday,

Abstract # Title Time

196 OMNIS: Titration for Faster, Safer and Easier Analysis 10:00am – 12:30pm

1976 Acid Number of Crude Oils and Petroleum Products by Catalytic Thermometric Titration using ASTM D8045

1:30pm – 4:00pm

Technical Posters: Wednesday 10/17, South Lobby

Metrohm at GCC 2018