Bridge for Thermistor Thermometers

Embed Size (px)

Citation preview

  • 7/25/2019 Bridge for Thermistor Thermometers

    1/4

    This content has been downloaded from IOPscience. Please scroll down to see the full text.

    Download details:

    IP Address: 128.122.253.228

    This content was downloaded on 10/10/2014 at 08:03

    Please note that terms and conditions apply.

    Constant sensitivity bridge for thermistor thermometers

    View the table of contents for this issue, or go to thejournal homepagefor more

    1962 J. Sci. Instrum. 39 75

    (http://iopscience.iop.org/0950-7671/39/2/316)

    me Search Collections Journals About Contact us My IOPscience

    http://localhost/var/www/apps/conversion/tmp/scratch_3/iopscience.iop.org/page/termshttp://iopscience.iop.org/0950-7671/39/2http://iopscience.iop.org/0950-7671http://iopscience.iop.org/http://iopscience.iop.org/searchhttp://iopscience.iop.org/collectionshttp://iopscience.iop.org/journalshttp://iopscience.iop.org/page/aboutioppublishinghttp://iopscience.iop.org/contacthttp://iopscience.iop.org/myiopsciencehttp://iopscience.iop.org/myiopsciencehttp://iopscience.iop.org/contacthttp://iopscience.iop.org/page/aboutioppublishinghttp://iopscience.iop.org/journalshttp://iopscience.iop.org/collectionshttp://iopscience.iop.org/searchhttp://iopscience.iop.org/http://iopscience.iop.org/0950-7671http://iopscience.iop.org/0950-7671/39/2http://localhost/var/www/apps/conversion/tmp/scratch_3/iopscience.iop.org/page/terms
  • 7/25/2019 Bridge for Thermistor Thermometers

    2/4

    J. SC

    INSTRUM.,

    1962, VOL. 39

    Constant sensitivity bridge for thermistor

    thermometers*

    E.

    PITTS and

    P. T. PRIESTLEY

    Research Laboratories, Kodak Ltd., Wealdstone, Harrow, Middlesex

    MS. received 21st September 1961

    Desp ite the variation of the incremental tempera ture resistance sensitivity of therm istors,

    temperatures c n be measured in a Wheatstone bridge circuit having constant'temperature

    deflection sensitivity if the bridge is fed from a variable voltage supply.

    In

    the circuit

    described the compensation c n be a utom atically set when the bridge is balanced, both

    operations being performed simultaneously on one and the same variable resistor, which

    is in series with a thermistor in one arm of a Wheatstone bridge and

    is

    part of

    a

    fixed

    resistor, the remainder of which is in series with the b ridge voltage supply. s a result,

    the voltage across the bridge changes with the balancing resistor, so that the rate of change

    with temperature of the off-balancecurrent

    is

    independent of the temp erature at which the

    bridge is balanced.

    The resistance values of the other three bridge

    rm

    must be suitably

    chosen to achieve

    this

    automatic compensation, and a method

    is

    described for their

    evaluation. In a typical example, full-scale deflection was obtained for

    a

    change of

    0.31 5 degc , i.e. a sensitivity of 2.45 pA-deg-', which was within 1 of the calculated

    value and constant to 1 over the w orking range of 15 degc.

    1 IntrodIi~on

    Thermometric analysis involves the measurement of small

    temperature changes during a chemical reaction, and the

    thermometer used must therefore accurately measure such

    differences irrespective of the initial temperature. It is very

    desirable to have an instrument more convenient and of a

    much smaller thermal capacity than the B e c k thermo-

    meter, w hich is otherwise suitable for the purpose.

    Direct-reading thermistor thermometers (McLean 1954,

    Beakley 1951) in which the full-scale deflection of a galvano-

    meter corresponds to a tempe rature change

    of

    about 25 degc,

    lack sufficient linearity, require complicated setting up and

    are not immediately adaptable to the abo ve use.

    We here describe a simple bridge circuit in which a

    full-

    scale deflection always corresponds to the same small

    tempe rature difference. The instrum ent has adequate sensi-

    tivity which is

    also

    constant over a working range of about

    15 degc.

    As

    an additiona l advantage, it is very easy to use,

    adjustment being effected by varying a single resistance

    without the need for complicated cyclical procedures. More-

    over, if an accuracy to 1 % in the measurement of the

    temp erature difference s enough, no calibration of the bridge

    is necessary, the theoretically calculated sensitivity being

    sufficiently accurate.

    This

    feature may save

    a

    great amount

    of time.

    2 The

    bridge

    circuit

    Repeated measurements of the thermistor resistance are

    inconvenient, so the thermistor is placed in one

    rm

    of a

    Wheatstone bridge which

    is

    then balanced at a given

    temperature, and subsequent small changes of temperature

    are measured by the c urrent flowing through the galvano -

    meter as the bridge departs from balance. The current

    through he galvanometer is proportional to

    a

    small change

    * Communication No. 2192 H from the Kodak Research

    Laboratories.

    The

    subject matter

    of

    this

    paper

    is

    covered

    by

    U.K.

    Pat. Appl. No. 33420/61.

    in resistance. Since the rate of chang e of therm istor

    resistance with temp erature decreases with rise in tempera-

    ture, the sensitivity of this simple a rrangem ent decreases as

    the working tem perature increases.

    To compensate for

    t i s

    change in ssnsitivity, a resistance

    is included in the circuit outside the bridge, in series with

    the applied voltage.

    This

    resistance is decreased as the

    working temperature increases, thus increasing the voltage

    applied to the bridge.

    At any working temperature, when the bridge is balanced

    the total resistance of the arm containing the thermistor and

    the adjustable resistance must have a k e d value determined

    by the k e d resistances of the other

    three

    arms. If the s u

    of the adjustable resistance and that in the external circuit

    is kept constant, then, as the resistance of the thermistor

    decreases with increasing working temperature so wi l l the

    resistance in the ex ternal circuit.

    By suitably choosing the

    resistances of the other

    three rms

    it is then possible to

    maintain c onstant sensitivity.

    3.

    Theory

    The figure shows the b ridge circuit and th e provision

    of

    a

    resistance

    A-Y

    in the external circuit. The figure indicates

    the symbols used in the following argument.

    Balance

    is

    obtained by varying

    Y.

    The current is through he galva-

    nometer may be found by the usual methods, and diJdT

    may be derived and conditions chosen such that this ate of

    change remains independent of the thermistor resistance R .

    The equation for the current is

    vi?,( R

    Y)

    L ( M + A - Y

    75

  • 7/25/2019 Bridge for Thermistor Thermometers

    3/4

    CONSTANT

    SENSITIVITY B R I D G E

    FOR

    THERMISTOR

    T H E R M O M E T E R S

    Thermistor bridge circuit.

    V voltage of cell;

    ,

    pre-set potentiometer;

    Rb,

    made up

    from a

    high

    stability resistance and a potentiometer;

    R,

    resw

    tance of thermistor;

    G,

    galvanometer resistance; iD cment

    through the galvanometer;Y esistance of potentiometer in

    the bridge circuit;

    A-Y,

    resistance

    of

    potentiometer in extemal

    circuit; A , total resistance of the balancing potentiometer.

    We now consider the expression for

    di dT

    in conditions

    where the bridge isvery nearly balanced.

    There wi l l be small

    changes in Lo and O hen the bridge goes

    off

    balance, but

    it is readily shown that the main change in

    ig

    will be due to

    the change in (Rb R Y).

    good approximation is

    therefore

    where we have

    used

    equation (2) to substitute for Y in the

    denominator. Over a tempera ture range of abo ut 20 degc

    the variation of thermistor resistance with temperature is

    given by

    where a and

    b

    are constants and Tis the absolute temperature.

    Equation

    (5)

    then becomes

    R

    = ebiT, (9

    The requirement of constant sensitivity means that thii

    expression must

    be

    independent of

    R

    throughout the range.

    A way of effectingths is to m ake the sensitivitiesequal at

    each end of the working range; in practice this also sufEces

    to keep sensitivities very nearly constant throughou t.

    Suppose

    R 1

    and

    R2

    are the resistances of the thermistor a t

    the ends of the working range at temperatures

    l

    nd

    T2

    respectively.

    Sensitivitieswill be equal if

    Thevalue of the po tentiometer resistance

    A

    may conveniently

    be

    chosen to

    qual.RI R2,

    and we shall put

    Rb equal

    to

    76

    R 1 .

    Equation

    8)

    may be solved to find O nd hence

    R,,

    which is given by

    From these results we ha ll y obtain two numerically equal

    expressions for the sensitivity at temperatures

    Tl

    nd

    T2

    respectively.

    The simpler equation (11) is used for calculations of the

    sensitivity.

    An

    alternative method is based upon the observation that

    over the working range the change of thermistor resistance

    with temperature may be written approximately as

    d R

    _ -

    T

    (R

    @,

    where a and p are independent of T.

    This

    result can

    be

    easily shown graphically or theoretically.

    If

    this expression

    is used in equation 9, here results

    dip aVRa(R

    k?

    dT

    Lo M0 +

    A

    b +R)'

    This

    expression will be independent of

    R,

    and hence of T,

    provided

    The application of this approximation obviously ensures

    constant sensitivity throughout the working range.

    Its use

    leads in practice to the same value of R, as that found by

    the first method , an d also to th e sam e value of the sensitivity.

    4.

    A practical example

    In

    order to test the validity of the foregoing approximate

    theory, we calibrated an 'aged' thermistor (Stantel

    F23,

    S.T.C. L td.) using an N.P.L.-certiiied 0-50 c thermometer

    and

    a

    Muirhead Wheatstone bridge.

    The values R , and R2

    of the thermistor resistance were 2500 Q a t 1440 c and

    1463 Q at 29.62 c respectively.

    The latter temperature was

    chosen so that R1 R2 was equal to the value of a nominal

    l m 10-turn helipot (Colvem CLR 2301/22) having an

    actual resistance of 1037 Q. These results enable the values

    of

    a

    and

    b in

    equation (6) to be calculated. The result is

    b

    =

    2982 K, = 0-0775Q.

    By use of equation

    (9),

    the

    value of

    R,

    was found to be 749

    Q.

    The bridge circuit was

    then constructed,

    R,

    and

    Rb

    being calibrated against the

    Muirhead Wheatstone bridge.

    The thermistor thermometer was then checked at various

    temperatures against a Beckmann thermometer. The

    experimental values of sensitivity are shown in the table.

    Temp.( c)

    14.8 15.2 18.0

    24.0

    2 7 . 2 2 8 . 2

    dig

    - (p~deg- ' )

    2.45 2 .45 2 4 7 2.46 2 .45 2 .45

    dT

    Substitution of V =

    1-5Ov, G

    = 1308 Cl and other

    quantities in equation 11) gives the theoretical result

    M O

    = Rb A.

    = 2.43 /LA Qeg-'.

    d T

  • 7/25/2019 Bridge for Thermistor Thermometers

    4/4

    CONSTANT

    SENSITIVITY

    BRIDGE

    FOR

    THERMISTOR THERMOMETERS

    From these results it will be seen that the sensitivity varies

    by about 1 % over the working range and

    is

    within about

    1 of the theoretical value obtained from the approximate

    theory. W ith the galvanometer used (Pye Scalamp,

    0-00556

    p 4

    -I) a full-scale deflection corresponded to

    approximately

    3

    degc.

    5. Conclusions

    The approximate theory gives results which differ from

    experiment by about 1 . Greater accuracy cannot be

    expected since the evaluation of the derivative di,ldT is

    only approximate. Nevertheless, for many purposes this

    accuracy is sa ci en t, in which case the theoretical value c n

    be used without the need for time-consuming calibration.

    The sensitivity does remain nearly constant over the whole

    working range .

    It is impo rtant that an 'aged' thermistor is used, otherwise

    results will vary considerably over a period of months see,

    for example, Beck 1956).

    In

    the present experiments, the

    thermistor was immersed altemately in hot and cold water

    until the resistance at

    a

    given tem perature remained constant.

    The theory enables

    a

    suitable choice of galvanometer to

    be

    made if the sensitivity needed is

    known.

    The thermistor

    thermometer is accurate and very convenient to use and

    therefore suitable for many types of work where small

    temperature differences have to be m easured. The balancing

    potentiometer

    A

    may b e ca librated in absolute degrees if

    so

    desired.

    The 10-turn-helipot, used for

    A,

    was fitted wth an

    indicator dial having

    lo00

    divisions so

    that

    in the apparatus

    described lo00 divisions were equivalent to 1 5 . 6 2 ~ r

    1 division to 0.015

    62 c.

    Consequently the bridge sensitivity,

    expressed as degrees

    c

    per full-scale deflection,

    could

    be

    easily evaluated in the following manner, t o tak e into a ccoun t

    any change in the bridge supply voltage due to battery

    deterioration. The thermistor was immersed in a liquid at

    constant temperature, or replaced by a 2.2 kilohm resistor,

    and the bridge balanced. The balance helipot was then

    rotated tw gi ve full-scale deflection of the galvanom eter.

    The number of divisions,

    on

    he indicator

    dial,

    which

    corre-

    sponded to

    ths

    deflection was noted and the temperature

    span of the galvanometer was evaluated.

    The authors have lso made

    a

    thermistor thermometer

    to

    cover the range 56-82'

    c

    for use in the Heitler (1958) ebullio-

    meter. This thermometer has a sensitivity of 1 e94 PA deg-*

    with a galvanometer @ye Scalamp 040556 PA

    mm-')

    and

    its

    full-scale deflection corresponds to 0.4 degc.

    References

    BEAKLEY, . R., 1951,J. Sei. Instrum., 28,176.

    BECK,

    A.,

    1956,

    J.

    Sei. Instrum.,

    33,

    16.

    I-IEITLER

    C., 1958, Analyst, 83,223.

    MCLEAN,.

    A.,

    1954, J.

    Sei.

    Instrum., 31, 455.