28

Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed
Page 2: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

Bredikhin Ivan

Gammatech LLC

SOLUTIONS FOR

RADIATION MONITORING

IN PULSED FIELDS

Page 3: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

Detectors

Gamma Sphere and GASP detectors

Clover detectors

Double sided Ge Strip detectors

Special point contact technology (Dark Matter)

PopTop/Encapsulated

Discrete element and Segmented Array detectors

Page 4: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

CAEN Spa: HV&R.O.

Detectors

Beam Line Users

Machine Staff

CAEN SpA (Costruzioni Apparecchiature

Elettroniche Nucleari) was founded in 1979 as an

important industrial spin-off of the INFN.

Core business&Primary Market: Electronic

Instrumentation for particle accelerator physics

experiments (world leader)

Page 5: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

CAEN for the Synchrotrons Labs

Core business&Primary Market: Bipolar Power

Supplies and Beam Profile Monitors for synchrotron

accelerators

CAENEls:

PS&BPM Beam Magnets

Machine Staff

Page 6: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

CAENEls:

PS&BPM

Detectors

Beam Magnets

Beam Line Users

Machine Staff

Radiation Monitors

Environmental fixed monitors (gamma and neutrons)

Mobile monitoring stations, including

Ultra-sensitive pressurised ionisation chamber

Innovative neutron rem counter

Air monitoring system

Page 7: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

Detection of & :

Could it be a problem at all?

Page 8: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

t

Dose rate

x

STEADY

DETECTION

Radiation field

Page 9: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

t

Dose rate Dose rate

t

y

tOFF

tON

x

STEADY

DETECTION

IN PULSED FIELDS Radiation field

PULSED

Same averaged dose rate but

different instantaneous dose rates

Page 10: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

t

Dose rate Dose rate

t

y

tOFF

tON

x

STEADY

DETECTION

IN PULSED FIELDS Radiation field

PULSED

Small DUTY FACTORS (=> high instantaneous dose rates)

impose severe limitations on the survey meters to be employed

Same averaged dose rate but

different instantaneous dose rates

ON

ON OFF

tx y

t t

DUTY FACTOR

Page 11: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

Fundamental property for a detector working in pulsed fields

Two response models

Typical values

GM: τ = 100 μs

Rem counter: τ = 1-10 μs

DEAD TIME

Page 12: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

Fundamental property for a detector working in pulsed fields

Two response models

Typical values

GM: τ = 100 μs

Rem counter: τ = 1-10 μs

DEAD TIME

Page 13: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

0

0,1

0,2

0,3

0,4

0,5

0,6

0,7

0 0,5 1 1,5 2 2,5 3 3,5

Me

as

ure

d d

os

e r

ate

[S

v/h

]

True dose rate [Sv/h]

EXAMPLE

Rem counter with dead time = 5 μs, sensitivity = 1 nSv/count

Page 14: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

0

0,1

0,2

0,3

0,4

0,5

0,6

0,7

0 0,5 1 1,5 2 2,5 3 3,5

Me

as

ure

d d

os

e r

ate

[S

v/h

]

True dose rate [Sv/h]

SAME VALUE DISPLAYED ON THE DETECTOR

BUT FACTOR OF 4 DIFFERENCE!

Correction equations work, but…

• Valid only for relatively low dead time losses

• Valid under the assumption that the interactions are uniformly distributed (=>

This is not the case, by definition, for pulsed fields)

EXAMPLE

Rem counter with dead time = 5 μs, sensitivity = 1 nSv/count

Page 15: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

Damage caused by a complete beam loss

to an accelerator magnet (synchrotron)

BEAM LOSS

CONSEQUENCES

Page 16: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

CASE STUDY

Paul Scherrer Institute, Villigen, Switzerland

Page 17: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

SwissFEL (Free Electron Laser) accelerator

Deliver short (1-60 fs) and intense flashes of X-ray radiation of

laser quality

Enable new investigations into the structure and dynamics of the

illuminated matter

CASE STUDY

Page 18: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

GROUND LEVEL

0 m

Fitness area, where radiation monitoring must be ensured

but…

Installation of active radiation detectors is forbidden!

THE PROBLEM

(only passive systems)

Page 19: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

ACTIVATION DETECTORS (Ag or 12C)

DIAMOND DETECTORS (12C)

INSTRUMENTS WORKING IN CURRENT MODE

(Ionisation chambers,

innovative rem counter (LUPIN))

CORRECTION TECHNIQUES

1. Capability to withstand very high fluxes with little saturation

2. High sensitivity

3. Capability to measure correctly the intensity of a single burst

4. Capability to reject the photon contribution

OPTIONS

(STATE-OF-THE-ART

DETECTORS)

IDEAL DETECTOR SHOULD

Page 20: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

Ion chamber

16 atm pressure (6 atm Ar + 10 atm N2):

higher pressure, higher sensitivity, but…

Ulltra-sensitive electrometer (fA) range from 1 fA to 10 µA

NAUSICAA

GAMMA MONITORING

Page 21: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

Ion chamber

16 atm pressure (6 atm Ar + 10 atm N2):

higher pressure, higher sensitivity, but…

Ulltra-sensitive electrometer (fA) range from 1 fA to 10 µA

NAUSICAA

GAMMA MONITORING

Argon is sensitive

to neutron as

well!

(n,p), (n,d),

(n,α)

Page 22: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

LUPIN BF3 or He3

Extended dynamic range (> 20 MeV)

Specifically conceived for pulsed fields

Excellent neutron/gamma discrimination

Fast alarm response (50 ms)

Ultra short pulses (1-60 fs!)

High energies (5 GeV primary beam)

Need of having a dual-use

(RP monitor + Beam Loss Monitor)

Dose rate from 10 nSv/h to 100 mSv/h

Excellent gamma rejection 105 rejection factor

NEUTRON MONITORING

REM COUNTER

Page 23: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

Detection of pulsed neutron fields shows an advantage, if compared to photons

NEUTRONS

γ

γ

γ

Neutron detection mechanism:

1) They reach the moderator surface

2) They are thermalized (scattering events)

3) Once thermalized they diffuse

4) They reach the detector (BF3 or 3He)

Photons do not need

thermalization in order to

be detected

DETECTION PROCESS

Page 24: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

OPERATION AT SWISSFEL

PROMPT GAMMA PEAK

DELAYED NEUTRON SIGNAL

Page 25: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

NO Single Channel Analyser => NO dead time losses

Logarithmic amplifier => Wide dynamic range

I

t

PRE-

AMPLIFIER AMPLIFIER

SINGLE

CHANNEL

ANALYZER COUNTER

PRE-

AMPLIFIER

SINGLE

CHANNEL

ANALYZER

LOG

AMPLIFIER ADC + PC

t

I

t

I

t

I +1

t

I I n + 10B → 7Li +

α

n

t

LUPIN WORKING PRINCIPLE (I)

Page 26: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

Completely customisable

Cable up to 30 m for remote measurements with both probes

Conceived for 100% reliable use also in rough environments

Mobile stations

Open protocols

Can handle multiple

detectors

Customisable reports

Page 27: Bredikhin Ivan Gammatech LLCisinn.jinr.ru/past-isinns/isinn-27/presentations/11/Bredikhin.pdf · LUPIN BF3 or He3 Extended dynamic range (> 20 MeV) Specifically conceived for pulsed

LUPIN

1. BF3 or He3 with energies up to 5GeV

2. Dose rate range: 10 nSv/h to 100 mSv/h

3. Excellent gamma rejection

105 rejection factor

4. Ultra short pulses (1-60 fs!)

Best instrument on the market capable of

distinguishing the single neutron burst!