B1C116 - Rev 02/12 WHAT YOU NEED FOR THE OPERATION Cemover Cemover Instrument Cement plus cementation system Drill The range of accessories for Cemover includes everything necessary for an effective cement removal procedure. CEMOVER CONSISTS OF: A sterile single-use device A plastic breather tube that allows removing the air in the surgical cavity during bone cement injection. A 02 without 8 mm drill bit Warnings: The device requires the complete specific instrument Bone cement extractor Hip - Knee orthopaedics TECRES S. p. A. Via A. Doria, 6 - Sommacampagna - Verona (Italy) tel. +39 045 92 17 311 - fax +39 045 92 17 330 www.tecres.it - [email protected] distribuited by CodeProduct Reference for orders

Bone cement Drill extractor Hip - Knee of bone cement In cemented prosthesis revisions, the cement may remain adherent to the bone. Cemover, the innovative cement pull extractor allows

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Bone cement Drill extractor Hip - Knee of bone cement In cemented prosthesis revisions, the cement may remain adherent to the bone. Cemover, the innovative cement pull extractor allows

B1C

116

- R

ev 0

2/12

WHAT YOU NEED

FOR THE OPERATION

Cemover

Cemover Instrument

Cement plus cementation system

Drill

The range of accessories

for Cemover includes everything

necessary for an effective cement

removal procedure.

CEMOVER CONSISTS OF:A sterile single-use device

A plastic breather tube that allows removing

the air in the surgical cavity during bone cement

injection.

A

02 without 8 mmdrill bit

War

nin

gs:

The

dev

ice

requir

es t

he

com

ple

te s

pec

ific

inst

rum

ent

Bone cementextractorHip - Knee

o r t h o p a e d i c s

TECRES S. p. A.Via A. Doria, 6 - Sommacampagna - Verona (Italy)tel. +39 045 92 17 311 - fax +39 045 92 17 330www.tecres.it - [email protected]

distribuited by

CodeProduct

Reference for orders

Page 2: Bone cement Drill extractor Hip - Knee of bone cement In cemented prosthesis revisions, the cement may remain adherent to the bone. Cemover, the innovative cement pull extractor allows

Correct patient positionin a hip operation

2

3

5

6

9

10

11

121

2

3

4

5

6

7

8

REMOVING BONE CEMENT

78

9

11

REMOVING THE INTRAMEDULLARY PLUG

10

12

1

It is advisable to use

IDGENTA

The device comes in 3 lengths(120, 150 and 180 mm)

to be chosen basedon the length of the stem

4

Cemoveris a sterile single-usedevice for removalof bone cement

In cemented prosthesis revisions, the

cement may remain adherent to the bone.

Cemover, the innovative cement pull

extractor allows completely and accurately

removing residual cement.

The device consists of metamers

connected to each other that allow

segmental cement removal.

Perforate the plug with an 8 mm Drill using the special centringdevice. It is advisable to cool the 8 mm Drill by spraying thecanal with physiological saline solution

Fully and forcefully screw the Plug Remover Staff down into thehole made with the 8 mm Drill.

Remove the plug using the Plug Remover Staff

Remove the plug.If you have difficulties, connect the instrument used forremoval of the segments to the double-threaded junctionwith the Plug Remover Staff

Remove the prostheses

Mill the femoral canal

Remove and aspirate any fragments and dry thoroughly

Insert the Ventilation tube in the canal and fill it withlow viscosity bone cement

Immediately insert Cemover and wait for the new cement to polymerize

Assemble the instrument in the following order:1) Hammering Mass Staff2) Hammering Mass3) Double-Threaded junction4) Cemover Staff

Screw the instrument onto the first metamer

The metamer is removed by repeatedly acting on the Hammering Mass

screw the instrument onto the next metamer and repeat operation 7for all the metamers until complete removal.

Remove the prostheses

Mill the femoral canal

Remove and aspirate any fragments and dry thoroughly

Insert the Ventilation tube in the canal and fill it withlow viscosity bone cement

Immediately insert Cemover and wait for the new cement to polymerize

Assemble the instrument in the following order:1) Hammering Mass Staff2) Hammering Mass3) Double-Threaded junction4) Cemover Staff

Screw the instrument onto the first metamer

The metamer is removed by repeatedly acting on the Hammering Mass

screw the instrument onto the next metamer and repeat operation 7for all the metamers until complete removal.

BEFORE

AFTER CEMOVER