69
MEDIFUSION R MP-1000 SERVICE MANUAL Daiwha Corp., Ltd. Copyright 2002, DAIWHA Corporation, Ltd. All rights reserved.

Bomba Infusion Mp 1000

Embed Size (px)

Citation preview

MEDIFUSIONR

MP-1000 SERVICE MANUAL

Daiwha Corp., Ltd.

Copyright 2002, DAIWHA Corporation, Ltd. All rights reserved.

This Manual describes the usage of an infusion pump(model : MP-1000) manufactured by Daiwha Corp., Ltd.

Only the person who is authorized by the company can repair this equipment and change a circuit. The equipment that is dismantled or repaired without prior permission cannot receive a post-management service. Parts and circuits used in this equipment can be changed for better performance

and safety reasons. After reading this manual, put it in a place with easy access so that everybody can

use it. If you have any questions or concerns regarding this equipment and user manual,

feel free to contact the company at the number below.

Address:

Daiwha Corp., Ltd., 733-18 Daiwha Bldg Yeoksam-dong Kangnam-Ku Seoul, Korea TEL: (822) 558 - 1711~6 FAX: (822) 554 - 0317

Safety precautions are divided into the following categories depending on the level of the expected hazard and damage caused by the inappropriate use of the product and the level of urgency that requires actions.

-DANGER: If you do not follow the instruction, it will cause severe injuries or even

death. - WARNING: If you do not follow the instruction, it can cause serious injuries or death.

- CAUTION: If you do not follow the instruction, it can cause minor injuries or damage

on the product.

TABLE OF CONTENTS

1. PRECAUTIONS

2. GENERAL DESCRIPTION

3. PARTS DESCRIPTION

4. CHECKOUT

5. THEORY OF OPERATION

6. TROUBLE SHOOTING

7. DISASSEMBLY AND REASSEMBLY

8. PARTS BREAKDOWN

9. PCB

10. SCHEMATIC

1. PRECAUTIONS The following information is provided to help the user operate the pump safely and accurately and further prevent safety hazard or physical injuries or damage.

DANGER 1. Keep the pump away from inflammable anesthetic.

WARNING

1. Make sure that an inlet that is connected to the external power stays dry all the

time.

2. Do not use a mobile phone, radio set or defibrillator generating high frequency

near the pump.

3. Do not use this pump in MRI rooms, high-pressure rooms or places while high

electromagnetic radiation is generated. 4. Be sure to set the infusion set rightly. If it is set by a reversed manner to a normal

pump operation, the patient’s blood will be drawn up into the tubing.

CAUTION

1. Operate the pump by the designated steps.

2. When the external power sources are used, keep in mind the following matters:

1) Use the power cord for the hospital and plug it into the grounded outlet.

2) Prior to use, check the rated voltage and frequency.

3) Be sure to check whether an extension cord is normal or not.

3. Use only the person educated about the pump or a skilled hand.

4. Attach the pump securely to an infusion stand and ensure its stability. If the pump

is dropped or given any hard shocks, do not use it. Please call the local agencies or

the company. 5. During infusion, frequently check the infusion line such as a rupture in the filter or

a leak in the line. Check both pump and patient’s conditions on a regular basis while drugs are infused. To ensure a good connection between the patients with the infusion line, a male-to-female luer-lock type connection is recommended.

1. PRECAUTIONS CAUTION 6. Pay close attention to the following matters when the infusion set is used.

1) Use a designated infusion set for infusion pump. If any infusion set other than

specified is used, proper flow rate accuracy and alarm function (air and

occlusion) are not guaranteed. If the designated set is not available, contact to

the agent or company.

2) Do not pull on the tubing with excessive force. Then the intended flow rate may

not be achieved.

3) Fit the tubing inside of the door from bottom to the top. Securely set the tubing

along the groove of the PUMPING UNIT 31. If not, free flow or leak can be

occurred.

4) Do not connect an infusion set administrated from an infusion pump to another

infusion line administrated only by the manual roller clamp (gravity infusion)

because this may influence the accuracy of infusion and alarm functions.

5) When drugs are infused for extended time, change the tubing location once every

12 hours to use the tubing within the margin of errors.

6) Do not reuse the infusion set.

7. When the pump is operated under the shown description conditions, the flow rate

accuracy may be incorrect.

1) Using drugs of the higher viscosity such as Dextrose 50%.

2) Operating under an ambient temperature between +5 and 40 , a relative

humidity between 20% and 90%. 8. Pay attention to the following matters when the pump operates on battery. 1) Check the battery use time. (The battery can be used for 2 hours at flow rate

25ml/h if it is fully charged.) 2) Connect the main power to charge the battery if the indicator that displays

LOW BATTERY ALARM INDICATOR 21 on the front is turned on or the alarm sounds.

3) Charge the internal battery to full capacity by connecting the pump to an AC power supply for 15 hours at least.

4) When the external (AC/DC) power is supplied, 3 lamps of the Battery capacity indicator 17 will be turned on without regard to the battery capacity.

1. PRECAUTIONS CAUTION

9. Never connect both AC and DC power sources to the infusion pump at the same time. To use the DC power, be sure to check the polarity of the DC jack.

10. Do not sterilize the pump with autoclave or E.O. gas. Do not wipe the pump with thinner, solvent, benzene, ammonia, acetone, etc. Use a wet cloth to clean.

11. Do not dismantle the pump.

2. GENERAL DESCRIPTION 2.1 INTRODUCTION

This manual provides service information on Volumetric Infusion Pump MEDIFUSIONMP-1000 for qualified engineers and service personnel. Refer to the pump's Instruction for use for detailed operating instructions.

2.2 TECHNICAL SPECIFICATION

+ Product Name Volumetric Infusion Pump

+ Model Name MEDIFUSION MP-1000

+ Infusion Principle Peristaltic Semi-Transit Finger

0.1~999 ml/h

0.1~99.9 ml/h: increase and decrease by 0.1 ml/h

+ Range of Flow Rate

100~999ml/h: increase and decrease by 1 ml/h

+ Accuracy ±5% (Korea Vaccine Type T: Code 243)

+ Range of Volume - Total Volume 1~9999ml or no limit: 1~9999ml: increase by 1 ml - Infused Volume 0.1~9999 ml: 0.0~99.9ml: increase by 0.1ml

100~9999ml: increase by 1ml

+ Purge Rate 700ml/h (subject to change)

+ Purge Volume 3ml (subject to change)

1ml/h: Flow rate 1.0~999 ml/h (subject to change) + K.V.O. Rate

0.1ml/h: Flow rate 0.1~0.9ml/h

+ Display 7 segment (4 digit * 2 line)

3 hours at flow rate 25ml/h + Battery

Charging time – approx. 15 hours

Visible and Audible Alarm

- Door open

- Occlusion (Detection range: 100~950mmHg)

: Down stream Occlusion 3 steps are adjustable

: Maximum infusion pressure: 950 mmHg

- Air Bubble detection/Low battery/

+ Alarm

- Infusion complete / Alarm for Malfunction

100~240 VAC, 50/60Hz (Fuse: 250V, T3. 15A) + Power supply

12VDC

+ Power consumption Less than 30VA

+ Classification Class I, Type BF, IPX 1 (Drip-proof)

+ Dimension (D x W x H) 125 x 166 x 258 mm

+ Weight Approx. 3.2 Kg

3. PARTS DESCRIPTION 3.1 FRONT

< MP-1000 >

Item No. Function

1) Handle 38

2) Door 39 Protects pumping unit, sensor, infusion set etc.,

3) Door handle 40 Opens and closes pump door.

3. PARTS DESCRIPTION 3.2 KEY

ITEM NO FUNCTION

1) F. RATE KEY 11 This key is used to set up the flow rate.

2) T.VOL KEY 9 This key is used to set up the total volume.

3) I.VOL KEY 10 This key is used to view the infused volume.

4) UP KEY 34 This key is used to set up the flow rate or total volume. The number increases by one inch each time when the key is pressed. The number rapidly increases while the key is being held down.

5) DOWN KEY 35 This key is used to set up the flow rate or total volume. The number decreases by one inch each time when the key is pressed. The number rapidly decreases while the key is being held down.

6) SHIFT KEY 12 This key is used to set up the flow rate or total volume. When this key is pressed, the digit shifts to the left.

7) PURGE KEY 8 This key is used for purge. The total purge volume will be infused automatically at the set up purge flow rate each time the key is pressed. (Note: It does not sense air and occlusion.)

8) START KEY 13 This key is used to start the infusion.

9) STOP/CLEAR KEY 14 This key is used to stop infusion. In addition, when the key is pressed for two seconds while setting up the total volume and flow rate (including the infused volume), the set data will be cleared off.

10) ON/OFF KEY 15 This key is used to turn the pump on/off. Press the key for two or three seconds to turn the pump on/off. If external power is supplied, the battery will be charged even if the corresponding indicator is off.

3. PARTS DESCRIPTION 3.2 KEY

< DISPLAY PARTS ON THE FRONT >

3. PARTS DESCRIPTION

3.3 DISPLAY ITEM /NO FUNCTION

F.Rate Display Key / 1 This window displays the infusion flow rate to the patient. It will be blinked when a flow rate set up function is selected (by pressing the Flow rate key ⑪).

Volume Display Window/4 This Window displays the total volume or infused volume to the patient. It will be turned on along with the Total volume indicator when a Total ③

volume set up function is selected (by pressing the Total Volume key ). However, only the Infused ⑨

volume indicator will be turned on when the ②

infused volume is displayed (by pressing the Infused volume key ). ⑩

Total Volume Selection

Indicator / 3

This indicator will be turned on when selecting Total volume selection (by pressing the Total volume key ). At this time, the total volume data ⑨

will be displayed on the Volume display window . ④

Infused Volume Selection

Indicator / 2

This indicator will be turned on when infused volume is selected (by pressing the Infused volume key ). At this time, Infused volume data will be ⑩

displayed on the Volume display windows . ④

Purge Indicator / 5 This indicator will be turned on when purge starts (by pressing the PURGE key ). It will be turned off ⑧

when purge stops.

Stop Indicator / 6 This indicator will be turned on when infusion is stopped. It will be turned off when infusion starts.

Infusion Indicator /7 This indicator displays the drug infusion state in percentage (%)

AC/DC Indicator This indicator will be turned on when external power (AC/DC) is supplied. It will be turned off when operating on batteries.

Battery Capacity Indicator This indicator displays the battery capacity state in three levels. When external (AC/DC) power is supplied, 3 lamps of this indicator is turned on, no concern with battery capacity.

3. PARTS DESCRIPTION

3.3 DISPLAY ITEM / NO FUNCTION

Air Alarm Indicator This indicator will be blinked in addition to the alarm sounding when an air (bubble) sensor detects bubbles during the infusion process.

Occlusion Alarm Indicator This indicator will be blinked in addition to the alarm sounding when an up or down occlusion sensor detects the infusion set being blocked during the infusion process.

Door Open Alarm Indicator This indicator will be blinked in addition to the alarm sounding when the door is open or somebody opens the door forcefully during the infusion process. However, the indicator will be turned on without the alarm sounding when the door is open after the infusion is stopped.

Low Battery Alarm Indicator This indicator will be blinked in addition to the alarm sounding when the battery is exhausted.

Completion Alarm Indicator This indicator will be blinked in addition to the alarm sounding when infusion is completed.

Micro Function Indicator This indicator will be turned on only in micro mode (Flow rate : 0.1˜99.9ml/h).

3. PARTS DESCRIPTION

3.4 REAR

Item No. Function

Nurse call port 26 Connected to nurse call equipment by nurse call cable, this port transmits warning to the nurse call equipment during the infusion.

Drop sensor port (Option)

25 This port is connected to a drop sensor.

Power switch 28 This switch is used to turn power on/off.

DC power port 27 12 - 15 VDC power can be supplied through this port.

AC power port 29 100 - 240 VAC and 50/60 Hz can be supplied through this port.

Clamp 34 Secures the pump to the pole.

Battery cover 35 Easy change to the battery

3. PARTS DESCRIPTION

3.5 DOOR INSIDE

Item No Function

Up occlusion Detector

29 This part senses whether or not the tubing is blocked on the upper part of the pump.

Lever 30 Pull down the lever to open the Pumping unit (31) and push it back to close it. When the door is closed without pushing the lever back, the lever will be pushed back automatically.

Pumping unit 31 This part is used to infuse drugs by pressing the tubing on the infusion set.

Down occlusion Detector

32 This part senses whether or not the tubing is blocked on the lower part of the pump.

Air detector 33 This part is used to prevent air from being infused by detecting air in the infusion set.

4.CHECKOUT 4.1 CHECKOUT OF POWER

1) Connect the infusion pump to AC power. Unload the set into the pump and close the pump door at this time.

2) Verify that AC/DC indicator is on. 3) Turn the power switch on at the rear of the pump. 4) Verify that all of the battery capacity indicators in the front of the pump are on. 5) Verify that AC/DC indicator is off, When the pump disconnect with AC power. And

verify that the battery capacity indicator in the front of the pump is on. 6) Connect the infusion pump to AC power again and Verify that AC/DC indicator is

on.

4.2 SELF TEST

1) Press the on/off key at the front of the pump for 2 seconds and check that the pump performs the following self test.

2) The following displays illuminate momentarily; Alarm indicators, Infusion indicators. Total volume indicator and infused volume indicator, purge indicator, stop

indicator and battery capacity indicator 3) All segments of the flow rate display window and volume display window

illuminate momentarily. 4) Press the on/off key at the front of the pump for 2 seconds and verify that display

turn off.

4.3 ACCURACY TEST

1) Turn the pump on and load an infusion set without air in the tubing 2) Program a flow rate of 100 ml/h and a total volume of 100 ml and verify that

infused volume is 0 ml. 3) Start infusion by pressing the START key. While the pump is running, check for the

normal operation without occlusion and that no bubbles are seen in the tubing. 4) When the pump has delivered the selected volume, the pump activates an audible alarm, illuminates the completion alarm LED, and switches to a KVO (Keep Vein Open) rate.

5) To stop the infusion, press the STOP key and verify infusion time and infused volume.

4.CHECKOUT

4.4 DOOR OPEN ALARM TEST

1) Turn the pump on and load an infusion set without air in the tubing 2) Verify that the door open alarm LED is on when the door is opening. 3) Close the door and press the START key 4) Open the pump door. Verify that the door open alarm indicator flashes and the

audible alarm is activated. 5) To stop the door open alarm, press the STOP. 6) Verify that Door open indicator is on.

4.5 AIR ALARM TEST

1) Turn the pump on and load an infusion set without air in the tubing. 2) Start infusion by pressing the START key. 3) Introduce an air into the infusion set. The air size should be greater than 3 mm

(Approximately 30 ul). 4) When the air sensor detects air bubble, the air alarm indicator flashes and the

audible alarm is activated. 5) To stop the AIR alarm, press the STOP. 6) Verify that the AIR alarm indicator is on.

4.6 DOWN STREAM OCCLUSION ALARM TEST

1) Turn the pump on and load an infusion set without air in the tubing. 2) Start infusion by pressing the START key. 3) Fold and completely pinch the tubing just below the pump. 4) When the occlusion alarm is activated, the occlusion alarm indicator flashes and

the audible alarm are activated. 5) To stop the occlusion alarm, press the STOP. And verify that the occlusion alarm

indicator is on.

4.7 UP STREAM OCCLUSION ALARM TEST

1) Turn the pump on and load an infusion set without air in the tubing. 2) Start infusion by pressing the START key. 3) Fold and completely pinch the tubing just above the pump. 4) When the occlusion alarm is activated, the occlusion alarm indicator flashes and

the audible alarm is activated. 5) To stop the occlusion alarm, press the STOP. And verify that the occlusion alarm

indicator is on.

4.CHECKOUT

4.8. LOW BATTERY ALARM TEST

1) Turn the pump on and load an infusion set without air in the tubing 2) Program a flow rate of 10 ml/h and a total volume of 100 ml and verify that

infused volume is 0 ml 3) Start infusion by pressing the START key. Verify that the micro function indicator

is on and the stop indicator is off and the infusion indicator is blink. 4) Disconnect the infusion pump to AC power. No change in pumping should occur,

but the AC/DC indicator should turn off within a few seconds. 5) IF low battery alarm is activated, plug the pump back in. Operation should

continue without interruption, the AC/DC indicator should turn on and low battery alarm turn off.

4.9. TEST MODE

1) Turn the pump on. 2) Press the MODE Key when infusion is stopped and then Program Version will be

displayed. At this time, press the MODE Key one more time to enter into the <test mode>

3) Check that the voltages displayed in the volume display window are within specification. The voltages displayed show the input voltage of A/D converter.

Description Display Specification

(Hexadecimal)Remark

Above 30H When tubing set Air sensor (VAIR) 28H + 2H When tubing not set

E4H ± 2H When the pump connect with AC power

Low Battery (VLB) 91H ± 2H

When the low battery alarm is activated

Up stream occlusion (VUP) 9AH ± 2H When tubing not set

Down stream occlusion (VDOWN)

9AH ± 2H When tubing not set

Note > Hexadecimal display

0 1 2 3 4 5 6 7 8 9 A B C D E F

4.CHECKOUT

4.10 ELECTRICAL SAFETY TEST

Test the pump's leakage current and ground impedance with IEC 60601-1. Leakage current should not exceed 1mA. Ground impedance should not exceed 0.1Ω-using 25A

4.11 USER MAINTENANCE

4.11.1. REPLACEMENT OF FUSE OF POWER INLET 1) Remove the AC power from the pump 2) Remove fuse box from the power inlet 3) Replace with a new fuse of the same value, type and voltage.(250V, T3.15A) 4) Reassemble fuse box and perform the electrical safety test to verify proper

grounding impedance.

4.11.2. REPLACEMENT OF BATTERY 1) Turn off and unplug the pump. 2) Remove battery cover and disconnect the connector from battery terminals. 3) Reassembly the battery into the pump in reverse order and reassembly battery

cover. 4) Be careful with the polarities of the female connectors.

4.11.3. BATTERY CHARGING

The battery is recharged whenever the pump is plugged in regardless of whether the pump is on or off. To charge the battery, simply plug the pump into a 100-240 VAC outlet. The battery must be stored in a charged condition and should be recharged at least once a month.

4.11.4. CLEANING

The pump should be cleaned as soon as possible after each use to minimize the accumulation and harming of spilled solutions. Case and panel may be cleaned with a soft cloth moistened with cold or warm water. Before cleaning the pump, be sure to turn off the pump by Power switch (26), and disconnect the AC power port (28) or DC power port (27) from the pump. Do not sterilize the pump with autoclave or E.O gas and do not use thinner, solvent, benzene, ammonia or acetone.

5. THEORY OF OPERATION 5.1 BLOCK DIAGRAM

5. THEORY OF OPERATION

5.2 POWER SUPPLY PART Power supply part consists of SMPS and 5V regulator, battery, and battery charger. The AC power inputs into inlet with EMC/EMI filter. SMPS converts from DC power. Whenever the SMPS is activated by AC power, the AC/DC indicator on the front is lit. SMPS (Switching Mode Power Supply) converts AC power (100~240VAC, 50/60Hz) into DC power (15VDC, 2.3A). The SMPS has over-voltage and over-current protection. Two 3.15A slow blow replacement fuses are required for AC. Battery charger will charge the battery as long as the pump is connected to a specified AC power. 5V regulator circuit on the main rear board generates 5V(Vcc) for electronic components from the output of SMPS and internal battery.

5.3 CPU PART

CPU part consists of controller (CPU) and EPROM, EEPROM, address decoder and oscillator, reset circuit. The controller (CPU) controls all pump function and gather data from the key controller, A/D converter, EEPROM and alarm controller. The controller utilizes 16-address line and 8 data line. Program reserved at the EPROM make the system work and main data will be maintained at the serial EEPROM. Oscillator generates 12MHz clock and pulse divider generates a pulse for the A/D converter.

5.4 KEY CONTROL PART

Key control part consists of keypad and key controller. The key is multiplexed 4 x 4 array, which is scanned by key controller. One of 4 select lines determines which four keys are read. All normal key depressions are decoded by this matrix.

5.5 DISPLAY PART

Display part consists of display controller and two 7-segments with 4digits and several LED. Two-display controller manages display and displays are turned on in order. Alarm/operation indicators are managed by the controller, while AC/DC indicator is managed by its own circuit.

5. THEORY OF OPERATION

5.6 MOTOR CONTROL PART Motor control part consists of motor, motor driver and rotation detector. The motor adopts bi-polar stepping motor to precise control and is controlled by 1/8, 1/4, 1/2 micro step control in order to precise than control. The motor is shut down when an alarm or failure occurs. The motor speed range permits infusions from 1 to 999 ml/h. The motor rotation detector assures that pump drive rotation occurs or the motor shuts down. CPU reads the motor rotation signal via the sensor controller.

5.7 SENSOR

1) Air sensor part The air sensor part consists of an ultrasonic transmitter and receiver mounted on opposite sides of the tubing path. The transmitter consists of an oscillator and piezo ceramic that transform electric energy into mechanical energy. The transmitter consists of signal process part and piezo ceramic that transform mechanical energy into electric energy. The transducers operate on the principle that air in the tubing transmits ultrasonic energy much less effectively than fluid. This energy is amplified, rectified, applied to the A/D converter and than converted into a digital signal. The CPU monitors the signal and activates an AIR alarm if it detects the absence of a level energy.

2) Occlusion sensor part The down and up stream occlusion sensor consists of a moving magnet and hall sensor that detects magnet. As the condition of infusion set, up and down of magnet, which is tangled by the end of occlusion is sensed by hall sensor. The moving magnet is spring-loaded against the tubing, when pressure down stream of the pump increase, the magnet moves from its original position, which in turns change the voltage of the sensor. If the occlusion is sufficient to cause a specific voltage change, the CPU activates an alarm. There is a maximum expansion of the tubing beyond which the pump will no longer permit operation.

3) Low battery detector The input voltage is sensed by step-downed in use of resistance. The alarm works when the voltage goes down under base line. When the voltage goes below impracticable figure, the display and motor stop.

4) Door Open Sensor When the door is opened, the drop in voltage that is sensed by the distance between the magnet stuck to the door and hole sensor. After all, the alarm works.

6. TROUBLE SHOOTING

6.1 TROUBLE SHOOTING TABLE 6.1.1 WHEN AC POWER AND BATTERY IS ABNORMAL <TABLE 1. POWER AND BATTERY BLOCK >

No SYMPTOMS CAUSES CORRECTIVE ACTIONS

AC/DC indicator is off 1.Abnormal AC power 1.Check AC power

while the pump is running 2.Abnormal fuse & SMPS (100VAC~240VAC)

on by AC 3.Abnormal AC/DC indicator 2. Check power cable and fuse

4.Malfunction in power circuit 3. Check AC/DC indicator and D

15

1.

4. Check <Chart 1>

AC/DC indicator is on 1. Abnormal D16 1.Check D16

while the pump is running

2.

on by internal battery

Battery capacity indicator 1.Malfunction in power circuit 1. ~5. Check <Chart 17>

is off while the power 2.Malfunction in AD converter

switch is on. 3.Abnormal battery capacity

4.Abnormal operation of

display circuit

5. Abnormal operation of

3.

CPU circuit

Low battery alarm is lit 1. Malfunction in A/D 1.Check <Chart 3>

While the power switch is converter.

4.

On at AC power.

If battery capacity 1.Abnormal on/off key 1. Check on/off key

indicator is on, on/off key 2.Malfunction in key scan 2. Check <Chart 7>

5.

is not accepted. Circuit.

When the on/off key push 1. The pump is infusing 1. ~2. Push the stop key and

the pump is not off. 2.The alarm is activated turn the pump off

3.Error message is displayed 3. Check <Table 9>

6.

4.Abnormal on/off key 4. Check <Table 1~5>

7. When the power switch is 1. Abnormal power switch 1.Check power switch

off, the pump is not off.

When the power switch is 1.The battery is not charged 1. ~2. Check <Table 1~11>

on, the battery capacity 2.Malfunction in battery 3. Check <Table 1~3>

indicators are off on Circuit

8.

internal battery. 3.Same cause as <Table1~3>

6. TROUBLE SHOOTING

6.1 TROUBLE SHOOTING TABLE 6.1.1 WHEN AC POWER AND BATTERY IS ABNORMAL

<TABLE 1. POWER AND BATTERY BLOCK > No SYMPTOMS CAUSES CORRECTIVE ACTIONS

The battery capacity 1.The battery charge time is 1.Charging battery during 15

indicators are off after insufficient. Hours.

charging. 2. The life span of battery is 2. Replace battery

exhausted.

3. Same causes as 3. Check <Table 1~4>

9.

<Table 1~4>

1. Battery operating time 1. Same causes as 1.Check <Table 1~9> 10.

is short. <Table 1~9>

1. Battery is not charged. 1.Abnormal battery 1. Replace battery

2.Malfunction in battery 2. Check <Chart 2>

11.

charge circuit

When supplying DC power 1.Abnormal DC jack 1.Replace DC jack

the pump is not turned on. 2.Abnormal connector J19 2.Check J12 and cable

12.

3.Abnormal D19 3.Replace D9

6.1.2 WHEN DISPLAY IS ABNORMAL <TABLE 2. DISPLAY BLOCK >

No SYMPTOMS CAUSES CORRECTIVE ACTIONS

Central Alarm indicators 1.Connector is not properly 1.Check J19, J21 and alarm

are abnormal. connected. cable.

1.

2.Abnormal alarm indicators 2. Check alarm indicators

One of volume, flow rate 1. Display select controller is 1.Check <Chart 6>

display segment is abnormal.

2.

displayed.

One of more flow rate 1.Flow rate or volume display 1.Check flow rate or volume

and volume display segment is abnormal. Display segment

3.

segment is not displayed.

Flow rate display window 1.Flow rate display segment 1.Check flow rate display

Is not displayed. is abnormal. segment

2.Display data controller is 2. Replace U23

4.

abnormal.

6. TROUBLE SHOOTING

6.1 TROUBLE SHOOTING TABLE 6.1.2 WHEN DISPLAY IS ABNORMAL

<TABLE 2. DISPLAY BLOCK >

No SYMPTOMS CAUSES CORRECTIVE ACTIONS

All flow rate display 1. Display data controller is 1.Replace U23

segment display same abnormal.

5.

character.

One or more flow rate 1.Flow rate display segment is 1.Check flow rate display

display segment is not abnormal. Segment.

displayed. 2.Display select controller is 2. Check <Chart 6>

6.

abnormal.

Volume display window 1.Volume display segment 1. Check volume display 7.

is not displayed. is abnormal. segment.

All volume display 1. Display data controller is 1.Replace U25

Segment and indicator abnormal.

8.

Is not displayed.

All volume display 1.Data display controller is 1.Replace U25

Segment display same abnormal.

9.

character.

One of more volume 1.Volume display segment 1.Check volume display

display segment is not is abnormal. segment

displayed. 2.Display select controller 2.Check <Chart 6>

10.

is abnormal.

Only indicators are not 1.Abnormal indicator 1.Check indicator 11.

displayed.

The brightness of segment 1.Abnormal segment and 1.Check segment and indicator

Is irregular. indicator. (LED).

12.

2. Abnormal RA3-RA6 2.Check RA3-RA6

6. TROUBLE SHOOTING

6.1 TROUBLE SHOOTING TABLE 6.1.3 WHEN KEY IS ABNORMAL

<TABLE 3. KEY PAD BLOCK > No SYMPTOMS CAUSES CORRECTIVE ACTIONS

All front key are disabled. 1.When blinking error 1.Check <Table 9> list

message

2.Malfunction in key scan 2~4. Check <Chart 7>

circuit

3.Abnormal key controller

enable signal

1.

4.Malfunction in CPU circuit

One or more keys are not 1.When infusing 1.Only Stop, Start key is usable

accepted. 2. Alarm is activated 2.Only Stop key is usable

3.When <User Mode> 3. Purge, Select, Start key is not

4.When <Test Mode> usable at <User Mode>

5.Abnormal Key 4. Only slow up/down key, fast

6.Malfucntion in key scan Up/down key is usable at

circuit <Test Mode>

5. Check key

2.

6.Check <Chart 7>

6.1.4 WHEN OCCLUSION SENSOR IS ABNORMAL <TABLE 4. OCCLUSION DETECTOR >

No SYMPTOMS CAUSES CORRECTIVE ACTIONS

Occlusion sensor is not 1.Infusion set other than the 1.Use recommended infusion

detected. specified is used. set

2. Infusion set is misloading 2.Reload infusion set

3. Malfunction in occlusion 3.Check occlusion circuit

Sensor circuit Up OCCL. => Check Chart15

Down OCCL.=> Check Chart16

1.

Occlusion alarm is 1. Infusion set is misloading 1.Reload infusion set

activated without 2. Malfunction in occlusion 2. Check occlusion circuit

occlusion. Sensor circuit Up OCCL. => Check Chart15

2.

Down OCCL.=> Check Chart16

6. TROUBLE SHOOTING

6.1 TROUBLE SHOOTING TABLE 6.1.5 WHEN AIR SENSOR IS ABNORMAL

<TABLE 5. AIR DETECTOR BLOCK> No SYMPTOMS CAUSES CORRECTIVE ACTIONS

Air sensor is not detected 1.Below 2mm air bubble

2.Air sensor is no sensitive 2. ~3. Check <Chart 14>

3.Malfunction in air sensor

circuit

1.

Air alarm is activated 1. Misloading infusion set 1.Reload infusion set

without air bubble 2. Air sensor is sensitive 2. Check <Chart 14>

3.Malfunction in air sensor

2.

circuit

6.1.6 WHEN DOOR SENSOR IS ABNORMAL

<TABLE 6. DOOR BLOCK> No SYMPTOMS CAUSES CORRECTIVE ACTIONS

When opening door, 1.Abnormal door open alarm 1.Check door open alarm

the door open alarm indicator indicator

indicator is off. 2.Malfunction in door circuit 2.Check <Chart 8>

1.

3.Malfunction in display circuit 3. Check <TABLE 2 LIST>

When closing door, 1.Malfunction in door circuit 1. Check <Chart 8>

the door open alarm 2.Malfunction in display circuit 2. Check <TABLE 2 LIST>

2.

indicator is on.

Door open alarm occurs 1.Malfunction in door circuit 1. Check <Chart 8> 3.

while the pump is running.

6. TROUBLE SHOOTING

6.1 TROUBLE SHOOTING TABLE 6.1.7 WHEN DROP SENSOR IS ABNORMAL

<TABLE 7. FREE-DROP DETECTOR BLOCK>

No SYMPTOMS CAUSES CORRECTIVE ACTIONS

“F.Err” alarm is activated 1.Drop sensor is not properly 1.Properly attach the pump.

when free flow is not attached.

occurred. (When using 2.Abnormal drop sensor plug 2.Check plug

drop sensor) 3.Drop sensor is internally 3.Check drop sensor circuit

1.

broken.

“F.Err” alarm is not 1.Option of drop sensor is off. 1.Option of drop sensor sets on

activated when free flow 2.Same cause as <TABLE 7-1> at <User Mode>

occurs (When using drop 2.Check <TABLE 7-1>

2.

Sensor)

6.1.8 WHEN BUZZER IS ABNORMAL <TABLE 8. BUZZER BLOCK>

No SYMPTOMS CAUSES CORRECTIVE ACTIONS

Audible alarm does not 1.Abnormal Buzzer 1.Check buzzer 1.

sound. 2.Malfunction in Buzz circuit 2.Check <Chart 9>

Audible alarm is low or 1.Option of alarm sound is low 1.Reset of alarm sound at

high. or high. <User Mode>

2.

2.Malfunction of buzzer circuit 2.Check R12, R14, R17

3. Audible alarm is abnormal 1.Abnormal buzzer 1.Replace buzzer

Audible alarm is on. 1.Low battery voltage is 1.Recharging battery 4.

(when infusing) detected.

When the stop key press, 1.Error message display 1.Check <TABLE 9> List

audible alarm is not off 2.Abnormal Stop key 2.Check STOP key

5.

(when infusing) 3.Malfunction in key circuit 3.Check Chart 7

Audible alarm is always 1.Malfunction of buzzer circuit 1.Check <Chart 9> 6.

on.

6. TROUBLE SHOOTING

6.1 TROUBLE SHOOTING TABLE 6.1.9. WHEN ERROR CODE IS DISPLAYED

<TABLE 9. ERROR STATUS> No SYMPTOMS CAUSES CORRECTIVE ACTIONS

"Er. 01"is displayed and 1.Two or more key is pressed 1.Check <Chart 7> 1.

the buzzer sounds.

"Er. 02"is displayed and 1.Malfunction in A/D 1.Check <Chart 13> 2.

the buzzer sounds. converter circuit

"Er. 03"is displayed and 1.Malfunction in EEPROM 1.Replace EEPROM 3.

the buzzer sounds.

"Er. 04"is displayed and 1.Rotation of pump is 1.Check pumping unit 4.

the buzzer sounds. abnormal. 2.Check <Chart 11>

"Er. 05"is displayed and 1.When the pump is stop, 1.Check pumping unit 5.

the buzzer sounds. pumping unit rotate. 2.Check <Chart 11>

"Er. 06"is displayed and 1.Backlashing of motor 1.Check Motor and replace 6.

the buzzer sounds.

"Er. 07"is displayed and 1.Initial position of pumping 1.Check checker assembly 7.

the buzzer sounds. unit is abnormal.

6.1.10. ETC

<TABLE 10. OTHERS>

No SYMPTOMS CAUSES CORRECTIVE ACTIONS

The flow rate accuracy 1.The infusion set is not 1.Reloading Infusion set

is low. properly set.

2.Option of accuracy is low or 2.Reset option of accuracy at

high. <User Mode>

1.

3.Abnormal pumping unit 3.Check pumping unit

Nurse call is always on or 1.When the nurse call plus is 1.Check plug

off. abnormal.

2.When the nurse call circuit 2.Check <Chart 10>

2.

is abnormal.

6. TROUBLE SHOOTING

6.2 TROUBLE SHOOTING FLOW CHART 6.2.1 CHART 1 (WHEN POWER IS ABNORMAL)

6. TROUBLE SHOOTING

6.2 TROUBLE SHOOTING FLOW CHART 6.2.2 CHART 2 (WHEN BATTERY IS NOT CHARGED)

6.2.3 CHART3 (WHEN A/D CONVERTER NOT DETECT POWER)

6. TROUBLE SHOOTING

6.2 TROUBLE SHOOTING FLOW CHART 6.2.4 CHART 4 (CHECKOUT OF CPU CIRCUIT)

6.2.5 CHART5 (CHECKOUT OF ENABLE SIGNAL OF CONTROLLERS)

6. TROUBLE SHOOTING

6.2 TROUBLE SHOOTING FLOW CHART 6.2.6 CHART 6 (WHEN DISPLAY IS ABNORMAL)

6. TROUBLE SHOOTING

6.2 TROUBLE SHOOTING FLOW CHART 6.2.7 CHART 7 (WHEN KEY IS ABNORMAL)

6. TROUBLE SHOOTING

6.2 TROUBLE SHOOTING FLOW CHART 6.2.8 CHART8 (CHECKOUT OF DOOR OPEN SENSOR)

6. TROUBLE SHOOTING

6.2 TROUBLE SHOOTING FLOW CHART 6.2.9 CHART 9 (CHECKOUT OF BUZZER)

6.2.10 CHART 10 (CHECKOUT OF NURSE CALL)

6. TROUBLE SHOOTING

6.2 TROUBLE SHOOTING FLOW CHART 6.2.11 CHART 11 (CHECKOUT OF PUMPING UNIT)

6. TROUBLE SHOOTING

6.2 TROUBLE SHOOTING FLOW CHART 6.2.12 CHART 12 (CHECKOUT OF MOTOR DRIVER)

6. TROUBLE SHOOTING

6.2 TROUBLE SHOOTING FLOW CHART 6.2.13 CHART 13 (CHECKOUT OF A/D CONVERTER)

6.2.14 CHART 14 (CHECKOUT OF AIR SENSOR DRIVER)

6. TROUBLE SHOOTING

6.2 TROUBLE SHOOTING FLOW CHART 6.2.15 CHART 15 (CHECKOUT OF UP STREAM OCCLUSION DRIVER)

6. TROUBLE SHOOTING

6.2 TROUBLE SHOOTING FLOW CHART 6.2.16 CHART 16(CHECKOUT OF DOWN STREAM OCCLUSION DRIVER)

6. TROUBLE SHOOTING

6.2 TROUBLE SHOOTING FLOW CHART 6.2.17 CHART 17(WHEN POWER S/W IS ON, BATTERY CAPACITY INDICATOR IS OFF)

6. TROUBLE SHOOTING

6.3 WAVEFORM 6.3.1 WAVEFORM OF CPU

(1) Display Select Controller Enable Signal - U5.1(=U29.5)

(2) PSEN Signal - U5.29

6.3.2 WAVEFORM OF DISPLAY

(1) Display Data Controller Enable Signal - U23.11(=U15.6), U25.11(=U15.8)

(2) Display Select Controller Enable Signal - U29.4, U29.5(=U5.1) : (When flow rate display

window blink)

6. TROUBLE SHOOTING

6.3 WAVEFORM

6.3.2 WAVEFORM OF DISPLAY (3) Display Select Controller Enable Signal - U29.4, U29.5(=U5.1) : (When volume display

window blink)

(4) Display Select Controller Output - U30.11, U29.15(=U30.8)

(5) Display Select Controller Output - U30.12, U29.14(=U30.7)

(6) Display Select Controller Output - U30.13, U29.13(=U30.6)

6. TROUBLE SHOOTING

6.3 WAVEFORM

6.3.2 WAVEFORM OF DISPLAY

(7) Display Select Controller Output - U30.14, U29.12(=U30.5)

(8) Display Select Controller Output - U30.15, U29.11(=U30.4)

(9) Display Select Controller Output - U30.16, U29.10(=U30.3)

(10) Display Select Controller Output - U30.17, U29.9(=U30.2)

6. TROUBLE SHOOTING

6.3 WAVEFORM

6.3.2 WAVEFORM OF DISPLAY (11) Display Select Controller Output - U30.18, U29.7(=U30.1)

6.3.3 WAVEFORM OF KEY (1) Key Controller Enable - U36.1(=U10.11), Key Scan Signal - U36.4

(2) Key Controller Enable - U36.1(=U10.11), Key Scan Signal - U36.5

(3) Key Controller Enable - U36.1(=U10.11), Key Scan Signal - U36.6

6. TROUBLE SHOOTING

6.3 WAVEFORM

6.3.4 WAVEFORM OF BUZZER

(1) Buzzer Controller Output - U16.14 (sound level - Middle): When door open alarm is

activated

※ Output of U16 according to sound level

Low : U16.11

Middle : U16.14

High : U16.1

6.3.5 WAVEFORM OF CHECKER (1) Checker Sensor - J17.3 (SS2), J17.2 (SS1): When flow rate is 100ml/h

※ Typical example

6.3.6 WAVEFORM OF MOTOR

(1) Motor Controller - U28.9, U28.19 : When flow rate is 100ml/h

※ Typical example

6. TROUBLE SHOOTING

6.3 WAVEFORM

6.3. 6 WAVEFORM OF MOTOR (2) Motor Driver - U27.10, U27.15 : When flow rate is 100ml/h

※ Typical example

(3) Motor Driver - U32.10, U32.15 : When flow rate is 100ml/h

※ Typical example

6.3.7 WAVEFORM OF A/D CONVERTER

(1) A/D Converter EOC - U19.7(=U5.14)

(2) A/D Converter - Start Signal : U19.22(U20.1), EOC Signal : U19.7(U5.14)

6. TROUBLE SHOOTING

6.3 WAVEFORM

6.3.8 WAVEFORM OF AIR SENSOR (1) Transmitter - U17.12, Receiver - U19.1:

※ Typical example

6.4 TEST MODE (CALIBRATION) 6.4.1 REFERENCE OF EACH SENSOR

Description Display

Reference

(Hexa-

decimal)

Check

Point

Reference

voltage

Adjust

Component Remark

Air Sensor1, 2 Above

30H

U19.26

U19.3 Above 0.96V

VR4,

VR6 When tubing sets

Up Stream

Occlusion

Sensor 9AH U19.27 3V VR7 When tubing not set

Down Stream

Occlusion sensor 9AH U19.28 3V VR8 When tubing not set

Low battery

(Power

Detector) E4H U19.1 4.46V VR9

When the pump connects

the AC power

[Reference of each sensor at <Test mode>]

6. TROUBLE SHOOTING

6.4 TEST MODE (CALIBRATION) 6.4.2 ADJUST EACH SENSOR (1) Calibration of air Sensor

Adjust waveform of transmitter by VR4, VR6 (See 7.3.8. waveform of air Sensor). and adjust

that the voltage of receiver is 50H when tubing set and it is 10h when tubing not set When

the voltage of air sensor downs 2AH, air alarm is activated. If air sensor is abnormal, check

the air sensor by <7.1. Troubleshooting Table - Table 5>

(2) Calibration of up/down Occlusion Sensor

Adjust that the voltage of A/D converter input is 9AH by VR7 (Up), VR8 (Down).

Down Occlusion adjust sensitivity of 3 levels at <User mode>. When occlusion alarm is

activated, Occlusion alarm indicator is blinks with audible alarm. Up Occlusion not adjust

sensitivity. When occlusion is not detected, check the occlusion sensor by <7.1.

Troubleshooting Table - Table 4>.

6.4.3 CALIBRATION OF LOW BATTERY (POWER DETECTOR)

Voltage

of Battery

Voltage

of U19.1 Display (Hexadecimal) Change of Battery capacity indicator

AC power 4.45V E4H AC/DC indicator LED ON

11.6V(Full) 3.59V B8H All battery capacity indicators are on

9.74V 2.99V 98H Two battery capacity indicators are on

9.56V 2.91V 95H One battery capacity indicator is on

9.33V 2.83V 91H Battery capacity indicators are off

8.8V 2.66V 88H Motor stops

8.33V 2.5V 80H Display is off

7.8V 2.32V 77H System is off (Reset)

[Reference voltage of Low Battery (Power Detector)]

Adjust voltage of low battery is E4H by VR9 when connecting the AC power. If voltage drops above 81H(voltage of U18.1=2.52V), display is turn off regardless of connecting the AC power.

7. DISASSEMBLY AND REASSEMBLY Disassembly of the Infusion pump MP-1000 is limited to the mechanical component and PCB assemblies. The pump that is dismantled or repaired without prior permission cannot receive a post-management service. Always make sure the pump is unplugged and that the pump is turn off before disassemble. Component of Infusion pump MP-1000 is sensitive to electrostatic discharge damage. Always disassemble and reassemble the pump on an anti-static mat.

7.1 SEPARATION OF FRONT CASE AND REAR CASE (SEE 8.5) 1) Turn the pump off and remove screw at rear case. Take care not to lay the pump

face down on components that could damage the front panel. 2) Separate the front and rear case by slowly lifting the rear case. 3) Avoid stressing the pump's internal cabling when separating the front and rear

case. 4) Stand the rear case horizontally 5) Disconnect the connectors, J11, J12 from the main rear PCB ass'y and disconnect

the connector J8 from the main front PCB ass'y and disconnect the ground cable from the power inlet

6) Reassemble case in reverse order. Verify cable route 7) Perform the operational checkout (See 3.)

7.2 REPLACEMENT OF PCB ASS'Y OF FRONT CASE (SEE 8.1)

1) Separate the front case and rear case completely. (See 7.1) 2) Disconnect the connectors, J13, J1, J5, J11, from the main rear PCB ass'y and

disconnect the connector J6 from the main front PCB ass'y 3) Remove screw of the main front PCB ass'y and lift main PCB ass'y. Then

disconnect connector, J22 from bottom of main front PCB ass'y. 4) Remove screw main rear PCB ass'y and separate main rear PCB ass'y and main

front PCB ass'y 5) Disconnect connector, J21 from the alarm PCB ass'y and remove screw of the

display and alarm PCB ass'y. Then separate the display and alarm PCB ass'y. 6) Reassemble PCB ass'y in reverse order.

7.DISASSEMBLY AND REASSEMBLY

7.3 REPLACEMENT OF AIR SENSOR ASS'Y (SEE 7.1)

1) Separate the front case and rear case completely. (See 7.1) 2) Disconnect connector of air sensor ass'y, J6 from the main front PCB ass'y and

remove screw of air sensor ass'y. 3) Separate air sensor ass'y and front case by lifting air sensor ass'y. 4) Reassemble the air sensor ass'y in reverse order. 5) Do not reuse the air sensor ass'y.

7.4 REPLACEMENT OF DOOR AND DOOR HANDLE (SEE 8.5)

1) Open the door and remove screw from door pin. 2) Pull bottom door pin up and pull top door pin down. Then separate door and

front case. 3) Remove pinhole cover and E-ring (82) from the door handle pin and push door

handle pin out of door and separate handle. 4) Reassemble the door and door handle in reverse order. 5) Do not reuse the E-ring

7.5. REPLACEMENT OF UP / DOWN STREAM OCCLUSION SENSOR (SEE 8.4)

1) Separate the front case and rear case completely. (See 7.1) 2) Disconnect the two connectors J10, J11 from Main front PCB ass'y. 3) Remove occlusion sensor PCB ass'y from occlusion sensor ass'y. 4) Separate the door. (See 7.4) 5) Remove screw from the occlusion sensor ass'y inside door, push the occlusion

sensor ass'y out of door 6) Reassemble the occlusion sensor ass'y in reverse order. 7) Do not reuse the PCB ass'y and sensor ass'y.

7.6. REPLACEMENT OF DOOR LATCH (SEE 8.1 AND 8.4)

1) Open the door. Remove screw from the slide knob and separate the knob. 2) Remove screws from the door latch and separate the door latch. 3) Reassemble the door latch in reverse order.

7.DISASSEMBLY AND REASSEMBLY 7.7 REPLACEMENT OF MOTOR (SEE 8.4)

1) Separate the front case and rear case completely. (See 7.1) 2) Disconnect the two connectors J13, J10 from the main front PCB ass'y. 3) Separate the door. (See 7.4) 4) Remove screw from the up occlusion sensor ass'y. 5) Remove screw from door base plate and separate the motor ass'y. 6) Remove wrench screw from the motor ass'y and separate the pulley. Then

separate the motor and buffer jig. 7) Reassemble the motor in reverse order.

7.8 REPLACEMENT OF CHECKER, CHECKER PCB ASS'Y AND DOOR SENSOR PCB ASS'Y (SEE 8.3)

1) Separate the front case and rear case completely. (See 7.1) 2) Disconnect the connectors, J1, J5, J11 from the main front PCB ass'y. 3) Remove screw from the door sensor PCB ass'y and checker PCB ass'y. Separate

the door sensor PCB ass'y and checker PCB ass'y. 4) Separate the door. (See 7.4) 5) Remove screw from the down occlusion sensor ass'y. 6) Remove screw from the checker and separate checker. 7) Reassemble in reverse order.

7.9 REPLACEMENT OF PUMPING UNIT (SEE 8.4)

1) Separate the front case and rear case completely. (See 7.1) 2) Separate the door and separate motor ass'y and checker PCB ass'y, checker. 3) Remove the screw from the pumping unit and separate pumping unit and front

case by pushing the pumping unit out of door. 4) Reassemble the pumping unit in reverse order. 5) Perform the accuracy test.

7.10 REPLACEMENT OF CLAMP (SEE 8.2)

1) Separate the front case and rear case completely. (See 7.1) 2) Remove the screw from the clamp plate and separate the clamp ass'y. 3) Reassemble the clamp in reverse order. 4) Perform the ground impedance test.

7.DISASSEMBLY AND REASSEMBLY 7.11 REPLACEMENT OF EXTERNAL CONNECTOR PCB ASS'Y AND DC JACK PCB

ASS'Y, POWER SWITCH (SEE 8.2)

1) Separate the front case and rear case completely. (See 7.1) 2) Remove the screw from the external PCB ass'y and DC jack PCB ass'y. Separate

the external PCB ass'y and DC jack PCB ass'y. 3) Separate power switch from the rear case by pushing 4) Reassemble the clamp in reverse order.

7.12 REPLACEMENT OF SMPS (SEE 8.2)

1) Separate the front case and rear case completely. (See 7.1) 2) Remove cable from the SMPS 3) Remove label and screw, then separate SMPS. 4) Reassemble the clamp in reverse order. 5) Perform the ground impedance test.

7.13 REPLACEMENT OF POWER INLET (SEE 8.2)

1) Separate the front case and rear case completely. (See 7.1) 2) Separate External PCB ass'y and DC jack PCB ass'y. 3) Remove cable from the power inlet. 4) Separate connector hole cover and DC jack hole cover. 5) Remove inlet label. 6) Remove screw and separate power inlet. 7) Reassemble the clamp in reverse order. 8) Perform the ground impedance test.

8. PARTS BREAKDOWN 8.1 FRONT CASE ASSEMBLY (1)

No. Description Qty No. Description Qty

1 Door base plate 1 10 Flat head screw (M3X8L) 6

2 Case front 1 11 Flat washer (ID=3.2, OD=6) 6

3 Door hinge 2 12 Spring washer (ID=3.2) 22

4 Display PCB ass'y 1 13 Pan head screw (M3X8L) 19

5 Main front PCB ass'y 1 14 Flat washer (ID=3.2, OD=8) 16

6 Main rear PCB ass'y 1 15 Pan head screw (M3X5L) 3

7 Alarm PCB ass'y 1 16 Nut (M3) 4

8 Air sensor ass'y 1 17 Toothed lock washer (ID=3.2) 4

9 Door latch 1 18 PCB supporter (H=17.0) 4

8. PARTS BREAKDOWN 8.2 REAR CASE ASSEMBLY

No. Description Qty No. Description Qty

10 Flat head screw (M3X8L) 2 26 Clamp body 1

12 Spring washer (ID=3.2) 8 27 Clamp Locker 1

13 Pan head screw (M3X8L) 6 28 E-ring (φ5) 1

14 Flat washer (ID=3.2, OD=8) 8 29 Power switch 1

19 SMPS 1 30 Connector hole cover 2

20 DC jack PCB ass'y 1 31 DC jack hole cover 1

21 Ext. connector PCB ass'y 1 32 Inlet label 1

22 Plat head screw (M4x15L) 4 33 Power inlet 1

23 Clamp plate 1 34 Pan head screw (M3X10L) 2

24 Case rear 1 35 Battery pack 1

25 Clamp handle 1 36 Battery cover 1

8. PARTS BREAKDOWN

8.3 PUMPING UNIT ASSEMBLY

No. Description Qty No. Description Qty

10 Flat head screw (M3X8L) 1 53 Finger - A 2

16 Nut (M3) 4 54 Pumping unit case - left 1

39 Magnet(φ5) 2 55 Pan head screw (M3X35L) 4

44 Slide knob 1 56 Buffer pully 1

45 Slide 1 57 Wrench screw (M3X5L) 4

46 Roller 4 58 Bearing 2

47 Presser 1 59 Teflon washer 2

48 Pumping unit cover 1 60 CAM 7

49 Mini Presser spring 2 61 CAM spacer 6

50 Mini presser 2 62 CAM shaft 1

51 Finger shaft 1 63 Pumping unit case - right 1

52 Finger - B 5

8. PARTS BREAKDOWN

8.4 FRONT CASE ASSEMBLY (2)

No. Description Qty No. Description Qty

10 Flat head screw (M3X8L) 4 66 Stepping motor 1

11 Flat washer (ID=3.2, OD=6) 4 67 Buffer jig 1

12 Spring washer (ID=3.2) 17 68 Pulley 1

13 Pan head screw (M3X8L) 13 69 PCB supporter (H=5.0) 2

14 Flat washer (ID=3.2, OD=8) 13 70 Motor checker PCB ass'y 1

15 Pan head screw (M3X5L) 4 71 Checker 1

57 Wrench screw (M3X5L) 2 72 Down Occl. sensor PCB ass'y 1

64 Up Occl. sensor PCB ass'y 1 73 Door sensor PCB ass'y 1

65 Timing belt 1

8. PARTS BREAKDOWN

8.5 CASE ASSEMBLY

No. Description Qty No. Description Qty

11 Flat washer (ID=3.2, OD=6) 2 79 Flat head screw (M3X5L) 2

12 Spring washer (ID=3.2) 6 80 Door 1

14 Flat washer (ID=3.2, OD=8) 4 81 Door handle pin 1

39 Magnet (φ5) 1 82 E-ring (Φ3) 1

74 Overlay 1 83 Handle rubber 1

75 Alarm window 1 84 Rubber foot 4

76 Door Handle 1 85 Pan head screw (M3X75L) 2

77 Pin hole cover 2 86 Pan head screw (M3X45L) 2

78 Door pin 2 87 Pan head screw (M3X15L) 2

9. PCB 9.1 MAIN FRONT PCB

9. PCB 9.2 MAIN REAR PCB

9. PCB 9.3 DISPLAY PCB

9. PCB 9.4 ROTATION CHECKER PCB

9.5 OCCLUSION SENSOR PCB (UP / DOWN STREAM)

9.6 AIR SENSOR PCB

9.7 EXTERNAL CONNECTION PCB

9.8 DC JACK PCB

9. PCB 9.9 DOOR PCB

9.10 ALARM DISPLAY PCB

9.11 TUBE CHECK PCB

9.12 DROP SENSOR PCB

10. SCHEMATIC

INTENTIONAL BLANK

10. SCHEMATIC 10.1 SCHEMATIC OF POWER CONTROL

SMPS

Vcc

GND

AC(L)

ACG

AC(N)

D171N4007

12

D181N4007

5VDrop Sensor 5V

L2330uH

DC Jack123

R539.1

C31470uF

PowerSwitch

L1330uH

R4410K

F1250V T3.15A

Inlet With Noise Filter

U33LM317/TO

1 3

2

VIN VOUT

AD

J

C32104

D191N4007

V1100~240VAC

12

3

R474.7K

C5410uF

C230100uF

D1201N5819

1 2

External Power Signal

F2250V T3.15A

D91N4007

1 2

C231470uF

C5310uF

U134LM2575-5

1 2

3

45

V IN OUTPUT

GN

D

FBON/OFF

SW1Power S/W

1 2

BT19.6V

12

C232C104

C30100uF

D201N5819

1 2

U34LM2575-5

1 2

3

45

V IN OUTPUT

GN

D

FBON/OFF

DC Power

D151N4007

1 2

D161N4007

1 2

R5510k

10.2 SCHEMATIC OF CPU

A12

A0

D2

A3/W/R

A15

D7A7

A15

A1D2

D5

A10

D4

A1

D7

A2

D6

A11

D[0..7]

A3 D3D4

D1

D4A[0..15]

A14

A12

D6

A[0..15]

D5

A6

A10

D[0..7]

D5

D6

A4

D3A4

D0

D2

A0

A14

D7

D0

D4

/R/D

A2

A9

A9

D6

D[0..7]

A5

D1

D3D2

A8

A8

A[0..15]

A6

D0

D0

A13

A11

D3

D5

D1

A13

D1

A5

D7

A7

Data D0

Address A11

R98470

Data D2

Data D4

NF1104/50V

Address A5

5V

C320pF

Z4HF70BTL3.5*9R

1 2

Z1HF70BTL3.5*9R

12

Address A4

Address A7

EOC

Address A14

5V

Drop Sensor 5V

Z5HF70BTL3.5*9R

1 2

5V

Address A0

Address A2

Address A13

Data D7

Address A12

5V

U580C32

31

19

18

9

12131415

12345678

3938373635343332

2122232425262728

171629301110

EA/VP

X1

X2

RESET

INT0INT1T0T1

P1.0/T2P1.1/T2XP1.2P1.3P1.4P1.5P1.6P1.7

P0.0P0.1P0.2P0.3P0.4P0.5P0.6P0.7

P2.0P2.1P2.2P2.3P2.4P2.5P2.6P2.7

RDWR

PSENALE/P

TXDRXD

375 kHz

Data D5

Address A9

J27Drop sensor jack

1

2

3

4

5

6

7

Address A6

U214N35 16

2

5

4

Data D3

Address A10

Read Signal

Address A14

NF2104/50V

R101100

AT93C56U11

123

4

6

CSCLKDI

DO

ORG

R674.7k

C1010uF

R100100

Data D1

U4

74HC4020

10

11

9754613121415123

CLK

RST

Q1Q4Q5Q6Q7Q8Q9

Q10Q11Q12Q13Q14

Motor Control Signal

Address A8

Address A3

U674HC373

111

256912151619

3478

13141718

OELE

1Q2Q3Q4Q5Q6Q7Q8Q

1D2D3D4D5D6D7D8D

Write Signal

D31N4007

12

R1120K

Data D6

C520pF

U13E74HC14

1110

R994.7k

Address A1

C

RA14.7k SIP 9

123456789

X112Mhz

U2C74HC04

56

Z2HF70BTL3.5*9R

12

10. SCHEMATIC

10.3 SCHEMATIC OF ADDRESS DECODER

Read SignalBuzzer Control Signal

Motor Control SignalU20D74HC02

11

1213

U13B74HC14

3 4

U13D74HC14

9 8

FND1 Control Signal

Sensor Selection Signal

ADC Read Signal

Write Signal

Address A14

Address A5

U12A74LS139

23

1

4567

AB

G

Y0Y1Y2Y3

U15B74HC08

4

56

U12B74LS139

15

1211109

AB

G

Y0Y1Y2Y3

Address A15

Address A9

ADC Write Signal

U15A74HC08 1

23

Address A13

U10B74HC32

4

56 Display Selection Signal

U20B74HC02

5

64

FND2 Control Signal

U10D74HC32

12

1311

U20A74HC022

31

U10C74HC32

9

108

Address A4

U13A74HC14

1 2

Key Control Signal

Address A8

U10A74HC32

1

23

U15C74HC089

108

A15

A14

1413A9

A13

/R/D

/W/R

A8

A5

A4

D0

D5

D3

D2

D4

D4

A2

D5

D7

A0

D0

D2

D7

D1

D6

D1

D6

A1

D3

10.4 SCHEMATIC OF DISPLAY CONTROL

LED6Green

Micro

LED21SBD446G

7

1 2

8

3 4

5 64_A

1_A 1_C

4_C

2_A 2_C

3_A 3_C

Data D4

LED13Red

Display Selection Signal

Data D3

LED8Yellow

Data D1

DOOR OPEN

AC/DC

LED10RED

Data D6

INFUSED VOLUME

LED14Red

LED3GREEN

U2574HC373

111

256912151619

3478

13141718

OELE

1Q2Q3Q4Q5Q6Q7Q8Q

1D2D3D4D5D6D7D8D

STOP

LED11Yellow

RA668Y

12345678

12345678

LED9Yellow

External Power Signal

U2374HC373

111

256912151619

3478

13141718

OELE

1Q2Q3Q4Q5Q6Q7Q8Q

1D2D3D4D5D6D7D8D

Address A2

Data D3

Data D6

AIR

TOTAL VOLUME

U26SND540BUG

117421

1053

6

8

9

12

ABCDEFGDP

DIG.4

DIG.3

DIG.2

DIG.1

CHARGE3

Address A0

CHARGE1

COMPLETE

Data D5

LED7Yellow

Data D2

LED26Red

Data D4

LOW BATTERY

CHARGE2Data D7

U30TD62083

12345678

10

18

11121314151617I1

I2I3I4I5I6I7I8

COM

O1

O8O7O6O5O4O3O2

LED20SBD446G

7

1 2

8

3 4

5 64_A

1_A 1_C

4_C

2_A 2_C

3_A 3_C

LED15Green

Volume

Data D5

RA368Y

12345678

12345678

LED12Red

PURGE

U24SND540BUG

117421

1053

6

8

9

12

ABCDEFGDP

DIG.4

DIG.3

DIG.2

DIG.1

FND2 Control Signal

Flow Rate

LED2GREEN

Data D2

FND1 Control Signal

LED1GREEN

Data D0

RA568Y

12345678

12345678

OCCLUSION

5V

RA468Y

12345678

12345678

Data D0

R58120

Data D7

U2974HC237

123

16

15141312111097

654

ABC

VCC

Y0Y1Y2Y3Y4Y5Y6Y7

G1G2GL

Address A1

Display Control Signal

Data D1

10. SCHEMATIC

10.5 SCHEMATIC OF MOTOR CONTROL

5V

R5247K

D121N5817

R453.3K

Data D5

U32A3955SB

12

3

4 5 1312

78

6

149

10

11

15

16

PFDREF

RC

GN

DG

ND

GN

DG

ND

PhaseD2

Vss

D0D1

OUTa

Sense

OUTb

Vbb

Data D7

D131N5817

Motor Control Signal

Data D0

Data D4

R4920K

Data D2

D111N5817

Data D3

R464.7k

5V

R5747K

R511.2

U27A3955SB

12

3

4 5 1312

78

6

149

10

11

15

16

PFDREF

RC

GN

DG

ND

GN

DG

ND

PhaseD2

Vss

D0D1

OUTa

Sense

OUTb

Vbb

5V

D141N5817

Data D1

C29470pF

R561.2

C26470pF

MG1Stepping Motor

1

3

4 6

Data D6

U2874HC373

111

20

256912151619

3478

13141718

OELE

VCC

1Q2Q3Q4Q5Q6Q7Q8Q

1D2D3D4D5D6D7D8D

R4830K

DC Power

D0

D3D4

D6

Motor CTRL

D1

D7

D5

D2

C000H

Up Stream

Air 2

Power

A0

D7

A2

D1Down Stream

D4

D2

A1

D3

D6D5

C005H

D0Air 1

10.6 SCHEMATIC OF AD CONVERTER

ADC Read Signal

C1620pF

+

-

U22BLM29025

67

411

1N4148D2

U2F74HC04

13 12

C20104

R39300K

C410pF

S6Air sensor 2

+-

+-

Address A1

ADC Write Signal

Data D1

C52102

5V

R1351K

EOC

U2A74HC04

1 2

Data D3

R731.8k

R24470k

R161.8k

Data D7

R74470k

VR65021 3

2

R281K

VR45021 3

2

5V

C22104

R41.2K

R4210K

Air 2

5V

300KR43

Address A0

Q92SC1815

R51.2K

5V

Data D6

5V

C24104

R71470k

C19104

R21.2K

R201.8k

C25

10uF

VR710K

13

2

R61.2K

D71N4148

DC Power

R22100k

R75100k

C820pF

R761.8k

C21104

R7920k

C18104 Address A2

Data D4

C1220pF

R8020k

R19100k

R1551K

THS119

VccOUT(+)

GNDOUT(-)

R5010K

5V

5V

C610pF

5V

5V

C23104

R751K

375 kHz

D81N4148

R72100k

C15104

R31300K

U2B74HC04

3 4

U19ADC0809

26272812345

1216

10

97

171415818192021

252423

622

IN0IN1IN2IN3IN4IN5IN6IN7

REF+REF-

CLK

OEEOC

D0D1D2D3D4D5D6D7

A0A1A2

STARTALE

Data D5

+-

U3ALM339

7

61

312 R10

51K

5V

THS119

VccOUT(+)

GNDOUT(-)

R77470k

Data D0

5V

C1120pF

R411K

VR810K

13

2

D11N4148

Air 1

VR910K

13

2

U2E74HC04

11 10 C14102

Data D2

R21470k

Q12SC1815

R951K

R78470k

Q82SC1815

R1851K

Q32SC1815

+-

U3CLM339

9

814

312

+

-

U22DLM290212

1314

411

5V

R3010K

5V

R26300K

S1Air sensor 1

+-

+-

10. SCHEMATIC

10.7 SCHEMATIC OF ADDITIONAL SENSOR

R6420k

R6222

5V

R7022

Data D6

U1D74HC14

98

5V

R61330

U1F74HC14

1312

Data D0

R6620k

SS4UGS3144

Vcc

OUT

GND

Data D7

SS1SG207

Data D3

R664.7K

U1F74HC14

1312

R6720k

Data D5

SS11SG215

SS13SG215

R6820k

74HC244U8

3579

12141618

17151311

8642

191

2Y42Y32Y22Y1

1Y41Y31Y21Y1

2A42A32A22A1

1A41A31A21A1

2OE1OE

Data D4

SS14SG215

R6022

C9104

SS12SG215

SS2SG215

R6320k

Data D1

R69330

5V

U1E74HC14

1110

Sensor Selection Signal

U1A74HC14

12

U1B74HC14

34

R59330

R6520k

Data D2

D0

CHK4

D2

CHK2

D1

D5

D3

D6

CHK1

CHK3

D4

D7

D7

D1D2

D0

D3

D5D6

D4

A0A1

10.8 SCHEMATIC OF KEY CONTROL

SW111103W

12

Data D0

Data D5

SW51103W

12

Data D3

SHIFT

UP

Address A1

PURGE

SW81103W

12

U36A74HC139

23

1

4567

AB

G

Y0Y1Y2Y3

Data D6START

SW41103W

12

SW91103W

12

STOP/CLEAR

SW31103W

12

T.VOL

I.VOL

SW21103W

1 2

Data D4

Address A0

C

RA24.7k SIP 91

2 3 4 5 6 7 8 9

ON/OFF

SW71103W

12

Data D1

SW101103W

12

SW61103W

1274HC244U35

3579

12141618

17151311

8642

191

2Y42Y32Y22Y1

1Y41Y31Y21Y1

2A42A32A22A1

1A41A31A21A1

2OE1OE

DOWN

Data D2

Key Control Signal

Data D7

F.RATE

5V

10. SCHEMATIC 10.9 SCHEMATIC OF BUZZER AND NURSE CALL CONTROL

Q52SC1815

J23Nurse call jack

1

2

3

4

5

6

7

BZ1ALP1205

1

2

Data D2

5V

R14100

Q7TIP41C

D251N4007

12

Data D0

R120

C1104

R231.2k

Data D1

Buzzer Control Signal

U1674HC75

2

3

6

7

134

1611514101198

D1

D2

D3

D4

C12C34

Q1Q1Q2Q2Q3Q3Q4Q4

R17270

Q42SC1815

Data D3

R14.7k

D41N4007

12

Q22SC1815

123

5V

K1RY5W-K

34

5

68

712

Q62SC1815

D0

D3

D1

D2

10.10 SCHEMATIC OF EXTERNAL CONNECTION AND DROP SENSOR

VR2500

13

2

C7103

D24EL-7L

Q10AL-7L

13

C2103

R84.7K

U29C

74LS10

91011

8

R34.7K

Drop Sensor Output

Q11AL-7L

13

5V