  • E Instruction Manual

    BOD-System Oxi700

  • Content

    1. Thesystem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51.1 Principleofthemethod. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51.2 Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51.3 Illustrations/Equipmentsupplied. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

    2. Themethod . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62.1 Biochemicaloxygendemand(BOD). . . . . . . . . . . . . . . . . . . . . . . . . . . . 62.2 BODsensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62.3 Principleofmeasurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

    3. Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73.1 Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73.2 Definitions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83.3 Terms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83.4 Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93.5 InitialSetUp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113.6 Switchingon. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113.7 Switchingoff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113.8 Readingstoredvalues . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123.9 Readingactualvalues . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123.10 Startingmeasurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133.11 Parametermenu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143.11.1 Settingdateandtime. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143.11.2 Autostart. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143.11.3 Autoduration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153.11.4 Replacingbatteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153.12 Datatransfer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153.13 Notes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

    4. BODmeasurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164.1 Samplevolume. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164.2 Preparingthewatersample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

    5. Interpretationofresults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

    6. Checkingthemeasurementsystem . . . . . . . . . . . . . . . . . . . . . . . . 19

    7. Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

  • 4

    Content

    8. Inductivestirringsystem. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 208.1 Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 218.1.1 Correctuse. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 218.1.2 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 218.2 Safetyrequirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 228.3 Equipmentdescriptionandfunction. . . . . . . . . . . . . . . . . . . . . . . . . . . 248.4 SetUp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 248.5 Magneticstirringrods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 258.6 Stirring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 258.7 Maintenanceandcleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 268.8 Troubleshouting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 268.8.1 Thestirringrodrotatesunevenly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 268.8.2 Thestirringrodswervesbackandforth . . . . . . . . . . . . . . . . . . . . . . . . 26

    9. Technicaldata . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 279.1 BODsystem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 279.2 Inductivestirringsystem. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

    10. Accessoriesandspareparts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

  • 5

    1. TheSystem

    1.1 Principleofoperation BOD ismeasuredby thepressuredifferencewithinaclosedsystem (respirometric

    BOD).TheintegratedmemorysavesaBODvalueautomaticallyevery24hoursforatestperiodofmorethan3days.

    1.2 Applications DeterminationofBOD5andBOD7,OECD301F/biologicaldegradability.

    Otherbio-technologicalapplications.

    1.3 Illustrations/Equipmentsupplied

    ■ 1xBODmeasuringdevicewithintegratedbottlerack

    ■ 6xBODsensor

    ■ 6xBODbottles

    ■ 6xsealcups(Gasket)

    ■ 6xmagneticstirringrods

    ■ 1xstirringsystemdrive

    ■ 1xstirringsystemcontroller

    ■ 1xnitrificationinhibitor(ATH)

    ■ 1xpotassiumhydroxidesolution(KOHsolution,45%)

    ■ 2xoverflowmeasurementflasks(157mland428ml)

    ■ 1xinstructionmanual

  • 6

    2. TheMethod

    2.1 BiochemicalOxygenDemand(BOD) Thebiochemicaloxygendemand(BOD)ofwastewater,industrialeffluentsandsurface

    waterisanexpressionfortheamountofoxygenconsumedbythedecompositionoforganicmatterinabiochemicalprocess.

    2.2 PrincipleofMeasurement TheBODsystem,consistingofthesamplebottleandtheBODsensor,representsa

    closedsystem.Inthebottle,abovethesampleitself,isadefinedvolumeofair.DuringtheBODmeasurement,thebacteriainthesampleconsumethedissolvedoxygeninthesample.Thisisreplacedbyoxygeninthebottleabovethesample.Thecarbondioxidereleasedatthesametimereactwiththepotassiumhydroxideinthesealgasket.Thisgeneratesadecreaseinpressurewithinthesystem.ThisismeasuredbytheBODsensoranddisplayedasaBODvalueinmg/lO2.

    2.3 PreparingtheSample/BriefSummary• Estimatethemeasurementrangeandselectthevolumeforthesample(4.1).

    • Carryoutnecessarypre-treatmentofthesample(e.g.,settingthepHvalue;filtering;etc.)(4.2).

    • Measurethevolumeofthesampleprecisely,usingtheoverflowmeasurementflaskandtransferittotheBODbottle,withtheaidofafunnel.Ifnecessary,addnitrificationinhibitor(4.1).

    • Insertthemagneticstirringrod.

    • Place3-4dropsofKOHsolutionintothesealgasket.Theninsertthegasketintheneckofthebottle.

    • ScrewtheBODsensorstothesamplebottles.

    • Placethebottleinthebottlerack.

    • Startthemeasurement(3.10).

    • Incubatethesampleinaccordancewiththeinstructions(e.g.BOD5for5daysat20°C)

  • 7

    3. Operation

    3.1 Keys

    Key Functions

    OnOff

    • Switchestheunitonandoff(automaticswitchoffapprox.45secafterlastoperation)

    • Quitssub-menus• Breaksoffoperations

    START • SetUpofmeasurements

    READ • Readingofacurrentvalueforaselectedposition(doesnotstorevalue!)

    + • Increasesparameter/figure

    - • Reducesparameter/figure

    ENTER• Confirmsinput• Switchestoothersectionsinmenus

    • Headkey:selectsthepositionrequired

    • Head LED : indicates the position which has been selected

    1 2

    3 45 6

  • 8

    3.3TechnicalTerms

    Finishedhead Aheadwhichhascompleteditsseriesofmeasurements

    FreeheadAheadwheretheseriesofmeasurementshasbeenfinishedandatleastonefigurehasbeendisplayedinread-outmode(see„Readingoutstoredvalues“).

    Startvalue

    BODmeasurementisbasedonadifferenceinpressure.Whenameasurementisstarted,theinitialpressureisstoredasthestartvalue.BODfiguresarethenobtainedbymeasuringcurrentpressurevaluesandcalculatingtheBODfromthedifferencebetweenthetwo.

    DisplaymodeOperatingstatus,inwhichthe„Oxi700“displaysthelaststoredvalueandtheselectedmeasurementperiodforahead

    Read-outmodeOperatingstatus,inwhichallthestoredvaluesforaheadcanbedisplayed,togetherwiththeassociatedtimeofmeasurement

    Startmode Operatingstatus,inwhichmeasurementscanbestarted

    BackgroundmodeThe„OxiDirect“carriesoutsporadicmeasurementsindependently.Theindicationisalineinthecentreoftheupperdisplay.Donotchangeanybottleswhilethisisinoperation.

    3.2OperatingModes

  • 9

    3.4DisplaysandtheirMeanings

    LEDs Meaning Actionrequired

    LEDsIndicatethepostionwhichhasbeenactivated

    LEDislit Measurementiscomplete

    LEDflashes Measurementisnotyetcomplete

    Dispay Meaning Actionrequired

    LobAt

    Batteryisnearlyempty• Changebatteries(see„Changing

    theBatteries“-3.11.4)

    LoAd„OxiDirect“hasnotfoundaheadatthisposition

    • Checkwhetherthereisaheadattheselectedposition

    • Checkcontactsbetweenheadandrack

    xd

    Theunitisin„displaymode“.Theupperdisplayshowstheselectedmeasurementperiod.Thelowerdisplayshowsthelastvaluewhichhasbeenstored

    dx

    hx

    Theunitisin„read-outmode“andshowsthestoredvalueforday‚d‘/hour‚h‘(x=1...28daysor1...48hrs.)

    UFLThevaluesarebelowthepermittedmeasurementrange(belowzero)

    • Checkthattheflaskisfullysealedwiththemeasurementhead

    • Checkthetemperature

    OFLThevaluesareabovethepermittedmeasurementrange

    • Checkthetemperature

  • 10

    Dispay Meaning Actionrequired

    F1 Nomeasurementfigurepresent

    F2No„start-value”present,sonoactualBODvaluecanbecalculated

    see„Autostart“

    F120 Anerrorhasoccurredon„Start“

    Bottlesmustnotberemovedduringthe„start“process.Also,positionsmustnotbechangedduringthe„start“process

    SetDAteDuringinitialsetuporafterpowersupplyfailsthetimeanddatemustbereset

    See„SettingTimeandDate“(parametersmenu3.11.1)

    SrtundLoAdTheunithasnotlocatedaheadatthepositionwhichhasbeenselected

    • InsertabottlewithheadandpresstheOn/Offkey;or:

    • Selectanotherposition.Theunitwillchangeto„DisplayMode“

    SrtundDELThedatafortheheadattheselectedpositionhavenotyetbeenreadout.

    • Tocontinuethe„start“process,pressthe„Enter“key.(NOTE:Alldataonthisheadwillbedeleted!)

    • Toquitthe„start“process,presstheOn/Offkey.Theunitwillthenchangeto„DisplayMode“

    EndundLiFE

    Theelectronicsontheheadcannotstoreanymorevalues(thisshouldnotoccurinnormaluse).

    Thevaluesmostrecentlystoredcanbereadout.Theheadmustbereplaced.

    3.4DisplaysandtheirMeanings

  • 11

    3.5InitialSetUpThebatteriesarepackedseparatelywhentheunitisdelivered.Ifnecessary,removeallthebottlesfromtherack,turntherackandopenthebatterycompartment.Installthethreealkali-manganesebatteries(C/LR14)inthebatterycompartment(checkingforcorrectpolarity).Theunitshouldthenswitchonautomaticallyandthedisplaywillshow„SetDAte“.Ifnecessary,switchtheuniton.Setthedateandtime(seetheparametersmenu3.11.1).

    3.6SwitchingonTheunitisswitchedonwiththeOn/Offkey.The„OxiDirect“willactivatetheleftpositioninthefront(position1)andwillindicatethisbytheLED,whichwillbepermanentlylit(=measurementiscomplete)orbyflashing(=measurementinprogress).

    Ifthereisnoheadatthisposition,the„OxiDirect“willdisplay„LoAd“.

    Whenitisswitchedon,the„OxiDirect“isin„DisplayMode“,inwhichthelaststoredvalueandtheselectedmeasurementperiodaredisplayed(oradisplayasshowninTable3.4).Forexample:

    Meaning

    Upperdisplay 5d days 5days‘BOD

    Lowerdisplay 178 mg/l 178mg/lO2asthelaststoredvalue

    3.7SwitchingoffTheunitisswitchedoffbypressingtheOn/Offkey.Iftheunitisnotin„DisplayMode“,theOn/Offkeymustbepressedseveraltimesinordertoswitchoff.

    Duringtheswitching-offsequencetheupperdisplaywillshow„OFF“andalinewillthenappearinthemiddleofthedisplay.Duringthisperiod,donotremoveanybottlesfromtherack,orchangeplaces,oraddbottles,orpressanyotherkeys.Theswitching-offsequenceiscompletewhenboththedisplaysareblank.Atthispoint,waitafurther2secondsminimumbeforeswitchingonagain.

  • 12

    3.8ReadingoutstoredvaluesOncetheunitisswitchedon,pressthekeytoselectthepositionrequiredandconfirmbypressingthe„Enter“key.Theunitwillswitchto„DisplayMode“forthisposition.Theupperdisplaywillnowshowthemeasurementtimeandthelowerdisplaywillshowthevaluewhichhasbeenstored.

    Theunitalwaysstartswiththefirstvalue–i.e.,withaBOD-5set-upitwillstartwithday1(d1),forexample:

    Upperdisplay d1 days Valueofthefirstday

    Lowerdisplay 147 mg/l 147mg/lO2asthevalueforthefirstday

    The„+“and„-“keyscanbeusedtodisplayotherstoreddataforthisposition.

    Theheadkeyscanbeusedtoselectanotherposition.Ifthispositionisvacanttheunitremainsin„read-out“modeandshows„LoAd“.Placeaheadinthispositionforreadingoutthestoredvaluesorselectanotherposition.

    PressingtheOn/Offkeywillcausetheunittoquit„read-out“modeandgoto„displaymode“.

    Forameasurementperiodofoneor2days(BOD-1andBOD-2)themeasurementtimeisstatedinhours.Formeasurementperiodsexceeding2days(BOD-3,BOD-4...)itisshownindays.Thehoursareshownintheformof„hxx“sothelatestfigureforaBOD-2,forexample:

    Upperdisplay h1 hrs Valueofthefirsthour

    Lowerdisplay 178 mg/l 178mg/lO2valueforthefirsthour

    3.9ReadingoutthecurrentvaluesOncetheunitisswitchedon,pressthekeyforthepositionrequired.Thenpressthe„READ“key.Thecurrentvaluewillthenanddisplayed.

  • 13

    3.10Startingmeasurement

    Startinganewmeasurementwilldeleteallstoreddatafromthehead!

    Placeapreparedbottleinavacantpositionoftherack.Switchontheunitandpressthecorrespondingheadkeytoactivatethisposition.

    Pressingthe„Start“keywillnowstartthemeasurementforthisbottle.Whenthe„Start“keyispressed,theunitchangesto„StartMode“andwilldisplaythelastselectedrangeusedforthishead,withthesamplevolumerequired(specialcasesaredescribedlaterinthissection).Thedisplaywillflash.Forexample:

    Upperdisplay 157 ml 157ml(samplevolumerequiredforthemeasurementrange)

    Lowerdisplay 400 mg/l Rangeupto400mg/lO2

    Variousactionsarenowpossible:

    • The„+“and„-“keyscanbeusedtochangethesamplevolumeandthemeasurementrange.

    • Pressingthe„Enter“keywillselectthedisplayedsamplevolume/rangeandthe„start“processwillcontinue.

    Theunitcheckswhether„Autoduration“isactive.Ifso,theunitwillautomaticallyselectthespecifiedmeasurementperiodandwillactivatethemeasurement.Themeasurementperiodisnotdisplayed.If„Autoduration“isnotactive,theunitwilldisplaythelastmeasurementperiodwhichwasused(thedisplaywillflash).Forexample:

    Upperdisplay 5d ml 5days‘BOD(BOD-5)

    Lowerdisplay Thelowerdisplayisblank

    Variousactionsarenowpossible:

    • The„+“and„-“keyscanbeusedtochangethemeasurementperiodinone-daysteps.

    • Pressingthe„Enter“keywillselectthedisplayedmeasurementperiodtobeusedandtheunitwillstartthemeasurement.

    Whiletheunitattemptstostartthemeasurement,thedisplayswillshow„0.0.0“and„0.0.0.0“(flashing).Timecanvary.

    Whentheunithassuccessfullystartedthemeasurementthedisplayswillshow„000“and„donE“.

  • 14

    3.11ParametersMenuPressthe„ENTER“keyandholditdepressed.SwitchontheunitwiththeOn/Offkey.Allowthreedecimalpointsinthelowerdisplaytoflashfor3secondsandthenreleasethe„ENTER“key.Theunitisnowintheparametersmenu.The„+“and„-“keyscannowbeusedtoselectbetweentheDate/Time,„Autostart“,„Autoduration“and„ChangeBatteries“menus.Pressingthe„ENTER“keywillgiveaccesstothesub-menuwhichisselected.QuittheparametersmenuwiththeOn/Offkey.Theunitwillreturnto„DisplayMode“.

    3.11.1SettingtheDateandTimeConfirm„Set“„DAtE“bypressingthe„ENTER“key.Theunitwilldisplay„dA“andtheyearwhichiscurrentlyset.Pressthe„ENTER“key.Theunitdisplays„dA“andnowshowsthedatewhichiscurrentlysetintheformof„MM.TT“,where„MM“indicatesthemonthand„TT“indicatestheday.

    Pressthe„ENTER“key.Theunitdisplays„ti“andthetimeintheformof„hh.mm“where„hh“indicatesthehoursand„mm“indicatestheminutes.Pressthe„ENTER“key.Theupperdisplayflashesandshows„JJ“andthelowerdisplayshowstheyearinfigures.Theyearcanbechangedwiththe„+/-“key.Pressthe„ENTER“key.

    Inthiswayitispossibletoresetthefiguresforthemonth(thedisplayshows„Mo“),theday(thedisplayshows„dd“),thehour(thedisplayshows„hh“),thetensofminutes(thedisplayshows„M“)andthesingleminutes(thedisplayshows„m“).Eachentrymustbeconfirmedbypressingthe„ENTER“key.Finally,pressthe„ENTER“keyoncemoretoadoptthefigureswhichhavebeenset.Theunitsetstheinternalclocktothesefiguresand(providednomistakeshavebeenmade)willdisplay„iS“„Set“.Thedate/timemenucanbequittedatanytimebypressingtheOn/Offkey.

    3.11.2AutostartUsingthe„Autostart“functiondoesnotreplacetemperingofthesampletotherequiredtemperaturebeforemeasurementisactivated.

    If„Autostart“isactive,theunitwillwaitforpropertemperatureofthesample,indicatedbyadecreaseinpressure.Ittakesthispressureasthe„start“valueforthefollowingBODcalculations.However,themaximumwaitingtimeis3hours.Afterthistimetheunitwillautomaticallytakethecurrentpressureasthe„start“value,evenifnodecreaseinpressurehasbeendetected.Thiswillbeindicatedlaterbyflashingmeasurementvalues.

    IfthetemperatureoftheBODsamplesissignificantlydifferentfromthespecifiedtemperature,achangeinpressurewillcauseerrorsindeterminingtheBODvalues.

    The„Autostart“functioncanbeswitchedonandoff.Thesettingalwaysappliesfromthetimeitisdoneandappliestoalltheheadsontherack.

    Confirm„Set“„Auto“withthe„ENTER“key.The„+“keyisusedtoswitchonthe„Autostart“function–“On“„Auto“andthe„-“keyisusedtoswitchitoff–„Off“„Auto“.Oncetheselectionhasbeenmade,thesub-menuisquittedwiththeOn/Offkey.

  • 15

    3.11.3AutodurationConfirm„Set““dur“withthe„ENTER“key.The„+“and„-“keyscanbeusedtosetfiguresbetween1and28,aswellas„OFF“.Figuresbetween1and28specifythemeasurementperiodindays,while„OFF“deactivatesthe„Autoduration“.With„Autoduration“deactivated,theunitwillaskfortherequiredmeasurementperiodinthe„StartMode“foreachsampleand/orrackposition.Theunitispresetbythemanufacturerforan„Autoduration“of5–thatis,for5days‘BOD.The„Autoduration“settingisretainedevenifthepowersupplyfailsandvalidalwaysforallpositions.

    3.11.4ChangingtheBatteriesTochangethebatterieswithoutaffectingacurrentmeasurement,makesureyouhavethetoolrequiredtoopenthebatterycompartmentandhavethreereplacementalkali-manganesebatteries(C/LR14)ready.Gototheparametersmenuandselectthebatterychangemenu„Set“„bAt“(„+“key).Pressthe„ENTER“key.Theunitwilldisplay„bAt“.Pressthe„ENTER“keyagain.Theunitdisplaywillnowflash,showing„bAt“„chg“.Waituntilthedisplaydisappears–thismaytakeupto6minutes.Whenameasurementisactive,thebatterychangemustbecarriedoutwithin4minutesandtheunitmustnotbeleftwithoutpowerformorethanoneminutemaximum!Ifnecessary,removeallsamplesfromtheframe,rotatetheframeandopenthebatterycompartment.Takeouttheoldbatteriesandreplacethemimmediatelywiththenewones,checkingforcorrectpolarity.Closethecoverofthebatterycompartment,replaceallthesamples(theirpositionsarenotimportant)andswitchtheunitonagain.Iftheunitswitchesautomaticallytothedate/timemenuafterachangeofbatteries,thismeansthatthebreakinthepowersupplywastoolong.Itwillthenbenecessarytoresetthedateandtime.

    3.12DatatransferSwitchtheOxi700off.ConnecttheOxi700RS232interfacewiththeserialinterfaceofthecomputerusingthecable,whichispartofthedelivery.TransferingstoreddatafromtheOxiDirecttothecomputerrequiresatransferprogram,e.g.Hyperterminal(thisispartofstandardWindowssoftware).Pleasefinddetailedinformationonourhomepageinthedownloadarea.PresstheREADkeyandholditdepressed.SwitchtheunitonwiththeOn/Offkey.Allow3decimalpointsinthelowerdisplaytoflashfor3seconds,thanreleasetheREADkey.Theunitisnowtransferringallstoreddata.

    3.13Notes• Ifabottlehasbeenaddedtotherackwhileameasurementisactive,theunitmustbe

    switchedonagain,sothatthisnewflaskisdetectedandcorrectmeasurementcantakeplace.

    • Ifabottlehasbeenremovedpermanentlyfromtheframe,theunitshouldbeswitchedononceandthenoffagain,sothatthesemissingbottlenolongercauseapowerconsumption.

    • Theunitwillswitchoffautomaticallyifnokeyispressedforapprox.45seconds.However,toachievemaximumbatterylife,theunitshouldalwaysbeswitchedoffimmediately,ifnofurtheruseractionisrequired.Theunitwillswitchinbackgroundmodewhenmeasurementsarerequired,indicatedbyalineinthetheupperdisplay.Oncemeasurementiscomplete,theunitswitchesoffagain.

  • 16

    4. BODMeasurement

    4.1 SampleVolume ThesamplevolumeisrelatedtotheexpectedBODvalue.TheOxi700isdesignedto

    operatewiththefollowingrangesandsamplevolumes,allowingBODmeasurementupto0-4000mg/l,withoutanydilution.

    Note

    Theexpectedresultsshouldbeintheupperhalfoftherange.

    FordomesticwasteitisgenerallyappropriatetoconsideraBOD5valuewhichisap-proximately80%oftheCODvalue.

    RangeBODmg/l Samplevolumeinml DosageATH

    0–40 428 10drops

    0–80 360 10drops

    0–200 244 5drops

    0–400 157 5drops

    0–800 94 3drops

    0–2000 56 3drops

    0–4000 21.7 1drops

    4.2 PreparingtheWaterSample1. Check thepHvalueof theeffluentsample.TheoptimumpHvalue forbiochemical

    oxidationisbetweenpH6.5and7.5.IfthepHvalueofthesampleishigherorlower,itshouldbepre-adjusted.AnysignificantdeviationwillresultinalowerBODvalue.IfthepHvalueistoohigh,itcanbereducedbyaddingdilutehydrochloricacid(1mol/l)ordilutesulphuricacid(1mol/l).IfthepHvalueislow,itcanbeadjustedwithasodiumhydroxidesolution(1mol/l).

    2. Mixthewatersamplewellandallowtosettleforashortwhile.Itmayalsobeadvisabletofilterorhomogenisethesample.

    3. Measurethesamplevolumeprecisely,usingtheappropriateoverflowmeasurementflaskandpourthesampleintothesamplebottle(itmaybehelpfultouseafunnelforthis).Ensurethatthesampleinthebottlecontainsarepresentativeportionofanysolidsinsuspension.Itisrecommendedthateachsampleshouldbetestedtwiceorthreetimes.

  • 17

    4. To inhibit nitrification, we recommend the addition of nitrification inhibitor B(=AllylThiourea,orATH).Thisisparticularlyimportantforthelowrange0-40mg/l(forexample,whencheckingdischargesfromeffluenttreatmentplants).TherightamountofnitrificationinhibitorBisrelatedtothemeasurementrange-see4.1.

    Note Nitrifyingbacteriaalsoconsumeoxygen.Thisconsumptioncanoccurwithinthefirst5

    daysandismorelikelywithsampleswithlowBODlevels.Asageneralrule,theBODmeasurementshouldnotincludetheoxygenconsumptioncausedbynitrifyingbacteriaandthiscanbeinhibitedbyusingnitrificationinhibitorB.Thisinhibitorsuppressestheactivityofthebacteriabyenzymaticinhibition,sothatonlythebreakdownoforganicsubstancesinthesamplewillbemeasuredastheBODvalue.

    If it isdesired tomeasureoxygenconsumption resulting fromnitrification (N-BOD),carryoutmeasurementssamples,onewithandonewithoutnitrificationinhibitorandcomparetheresults.ThedifferencebetweenthetwoBODvalueswillrepresenttheoxygendemandofthenitrifyingbacteria.

    5. Addacleanmagneticstirringrodtoeachsamplebottleandadd3-4dropsof45%potassiumhydroxidesolutiontothesealgasket(thiswillabsorbthecarbondioxide).Theninsertthesealgasketintheneckofthebottle.

    Important The sample must never come into contact with the potassium hydroxide solution.

    Neverusegreaseoranyotherlubricantsasanadditionalsealingagent,fortheBODsensorsorforthesealgasket.Productsofthiskindmaycontainsolventswhichwillattack thesensor, resulting inseveredamage to theplastichousingandeven toafailureofthesensor.Theuseofsealinggreasesandlubricantsisnotcoveredbyourguarantee!

    6. Beforemeasurementbegins, thepreparedsamplemustbebrought to thedesiredtemperature(e.g.BOD5,20°C).Thiscanbeachievedbyplacingthesampleinather-mostaticallycontrolledcabinet,whilestirringthesamplecontinuouslywiththeinductivestirringsystem.

    TheOxi700hasanoptional“Auto-Start”function,whichenablesittostartwithsamplesattemperaturesfrom15°to20°C.Whenthis“Auto-Start”functionisswitchedon,thesystemchecksinspecificintervalswhethertherehasbeenadecreaseofthepressureintheBODbottleandwillnotstartthetimeruntilapressuredecreaseisdetected(latest,thetimerwillstart3hoursaftertheBODsensorhasbeenstarted).

    7. PlacetheBODsensorsonthesamplebottlesandtightencarefully.Thisisextremelyimportant–thesystemmustbecompletelyair-tight.ThenplacetheBODbottle,withthesensorscrewedinposition,intothebottlerack.Thiscanbedoneinthethermo-staticallycontrolledcabinetitself.

    Alternatively,becauseoftheuser-friendlydesignoftheOxi700,youcanremovetheentireBODunit,withitsintegralbottlerack,fromthethermostaticallycontrolledcabinet,whileleavingtheinductivestirringsysteminthecabinet.Thereisnoneedtodisconnectthecabling.OncetheBODbottleshavebeenplacedintherack,thesystemispositionedovertheinductivestirringsystemsothatthe4adjustmentscrewsfitintheassociatedrecessesofthestirringsystem.

    8. Startthemeasurementprocess(seeSection3.10).9. Incubatethesampleinaccordancewiththeinstruction(e.g.BOD5for5daysat20°C).

  • 18

    5. InterpretingtheResults

    Start 1. Day 2. Day 3. Day 4. Day 5. Day

    300

    250

    200

    150

    100

    50

    0

    1. TheBODvalueonanyparticulardaymustbehigherthantheoneofthepreviousday.

    2. BODvaluesdonotincreaseinalinearmanner.Theincreaseisalwayssmallerthantheoneofthepreviousday.

    3. IftheBODresultsincreaseinalinearmanner,thesamplehasahigherBODvaluethanexpected.

    4. IfBODresultsincreasedsuddenlyduringthemeasurement,thismaybetheresultofnitrification.

    5. IfBODresultsdecreasesuddenlyduringthemeasurement,thismayindicatesleakageofthesystem.

    Theinformationgivenreferstonormalsamples(municipalwastes)andthenormalcir-cumstancesofaBODmeasurement.However,deviationarealwayspossible,dependingonthesamples.Forexample,ifthereisareadingofzeroafter5days,thismaybeasignofsevereinhibitionoroxidizingofthesample.

    Inthecaseofindustrialeffluents,samplesmaycontaintoxicsubstances.Thesemustberemoved,orthesamplemustbediluted.Inaddition,nutrients(nitrogen;phosphate)andtraceelements(suchasiron)mustbepresentinsufficientconcentrations;otherwisebacterialgrowthmaybelimitedandartificiallylowBODvaluesmayberecorded.Whereproblemsofthiskindoccur,eachindividualcasemustbeconsidered.

    Forfurtherinformation,pleaseaskforourapplicationreports.

  • 19

    6. CheckingtheMeasurementSystem

    TestSet(Ordercode:L418328) Thetestsetenablesacheckofall thecomponentsof theOxi700.Usingaspecialreagenttablet,adefinedunderpressureisgeneratedinthesealedBODbottle.

    7. Maintenance

    High-qualitymetalalloysareusedforthecontactsbetweentheBODsensorsandthebottlerack.Ifnecessary,thecontactsmaybecleanedcarefullywithasoftcloth.

    TocompensateforanyirregularitiesandtoensureoptimumcontactbetweentheBODsensorandthebottlerackaswellasinsuringproperstirring,theBODmeasurementsystemhas4levellingscrewsonitsbase.

    Ifnecessary,theBODbaseunit(includingthebottlerack)andtheBODsensorscanbecleanedcarefullywithasoftcloth.

    Allpartswhichcomeintocontactwiththesample(BODbottles;sealcups;magneticstirringrod)mustbecleanedthoroughlyaftereachmeasurement.Thebottlesmustbeemptiedcompletelyandwashedseveraltimesinhotwater.

    Whenanycleaningagents,dish-washingliquids,etc.,areused,allpartsmustberinsedthoroughlyafterwards!TheresiduesofsuchagentscaninterferewithBODmeasure-ments.

  • 20

    8. InductiveStirringSystem

    Fig.1:Inductivestirringdrivewithstirringcontroller

    Å Stirringplace

    Ç Stirringdevice

    É Controller

    Ñ Plug

    Ö Cable

    Ñ

    É

    Ö

    Ç

    Å

  • 21

    8.1 Instructions

    8.1.1 Correctuse

    ThisinductivestirringsystemhasbeendesignedforstirringliquidsinBODbottles.

    8.1.2 Symbols

    Thefollowinginstructionsymbolsareusedinthisoperatingmanual:

    DANGER! Thissignreferstodangerousvoltages

    DANGER! Thissignreferstohazardoussituations. Riskstohumanlifeareindicatedbytheterm”LIFEHAZARD”

    y CAUTION! Thissignindicatesriskstoequipmentandmachinery

    i INFORMATION Thissignindicateseasierworkingpractices

  • 22

    8.2 SafetyRequirements Foryourownsafety,youshouldobservethefollowingwarnings.

    Thesafetywarningsindicateapossiblesourceofdanger.Atthesametimetheygiveinformationhowtoavoiddamage.Youwillalwaysfindthesesignswheredangercanoccur.

    DANGER! Magnetism. Magneticormetallicparts(e.g.datacarriers,pacemakers,watches...)canbeaffectedbymagneticfields. Keepsuchpartsawayfromthemagneticstirrer(2)andthemagneticstirringrods

    DANGER! Voltageandfrequencymustbewithintherangespecifiedforthecontrolunit(3)

    y CAUTION! Donotuseinhazardouslocations

    y CAUTION! Generalambientconditions: Stirringdrive(2):-10to+56°Cat100%humidity. Controlunit(3):0to+40°Catmax.80%humidity. Avoidextremetemperaturechanges. Thecontrolunit(3)mustnotbeusedinareasofhighhumidityorinareaswherewatermaysplash!

  • 23

    y CAUTION!

    Donotplacehotsamplesontopofthemagneticstirrer(2). Maximumtemperature:56°C.

    y CAUTION! Incaseofrepair,theequipmentmustonlybeopenedbyanauthorisedservicetechnician.Disconnectfrommainsbeforeopening!

    y CAUTION! Donotallowthemagneticstirringrodtorestinanalternatingmagneticfield.Donotplacethemagneticstirringrodinastronginversemagneticfield.Thismayresultindestructionofthemagneticfieldofthestirringrod.

    i INFORMATION Thelengthofthemagneticstirringrodmustnotexceed40mm. Donotuseellipticalstirringrodswitharoundcross-section.

  • 24

    8.3 EquipmentDescriptionandFunction ThestirringsystemhasbeendesignedforstirringliquidsinstandardBODbottles.The

    ultra-slimstirringdrive(2)has6stirringareas(1)andaseparatecontroller(3).Bottlescanberemovedfromtherackwithoutstoppingthestirringdrive(2).

    The inductivestirringdrivehasnomotorand isthereforewear-free. It isparticularlydesignedforcontinuousstirringinaincubator.Itishermeticallysealed(fullyprotectedagainstsplashingwater)andcanbeusedevenintheroughestenvironments.Thewidespacebetweenthestirringareas(1)ensuresgoodventilationofthebottles,speciallyinanincubator.Thestirringdrive(2)isdrivenbyacontroller(3)withapre-setspeed.Whenturnedon,thesoftstartfeatureensuresauniformstartofthemagneticstirringrodsathalfspeed.

    Every40seconds,thecontrollerreducesthestirringspeedby50%,re-centringthemagneticstirringrodinthemiddleofthebottle.

    Themagneticstirringrodsaredrivensynchronously,preventingpracticallyanyinterfe-rencebetweenthem.

    8.4 SetUp

    DANGER!

    Voltageandfrequencymustbewithintherangespecifiedforthecontrolunit(3)

    DANGER! Magnetism.Magneticormetallicparts(e.g.datacarriers,pacemakers,watches...)canbeaffectedby

    magneticfields.Keepsuchpartsawayfromthemagneticstirrer(2)andthemagneticstirringrods.

    y CAUTION! Donotuseinhazardouslocations

    y CAUTION! Generalambientconditions: Stirringdrive(2):-10to+56°Cat100%humidity. Controlunit(3):0to+40°Catmax.80%humidity. Avoidextremetemperaturechanges. Thecontrolunit(3)mustnotbeusedinareasofhighhumidityorinareaswherewatermaysplash!

    Thecontrolunit(3)doesnothaveanyswitch.Itisreadyforuseassoonasitispluggedin:Connectthecable(5)fromthestirringdrive(2)withthesocket(4)ofthecontrolunit(3)andfixit.Plugthecontrolunit(3)intothemainssocket.Toensureproperstirringthefourscrewsatthebottomoftherackcanbeusedtooptimizethedistancebetweenstirringdriveandmagneticstirringrod.

  • 25

    8.5 Magneticstirringrods WerecommendtheuseofPTFE-coatedmagneticstirringrods.

    i INFORMATION

    Thelengthofthemagneticstirringrod(8)mustnotexceed40mm.

    8.6 Stirring

    y CAUTION! Donotplacehotsamplesonthemagneticstirrer(2). Maximumtemperature:56°C.

    Fillthebottleswiththesample.

    Placeamagneticstirringrodintoeverybottle.

    PlacebottlesintheBODrack.

    Thecontrolunit(3)ispresettooperateat320rpm.Every40secondsthespeedisreducedto200rpm,whichwillre-centerthenagneticstirringrod,ifnecessary.

    Ifyouwanttostopthestirringprocess:

    Disconnectthecontrolunit(3)fromthemainssocket.

  • 26

    8.7 MaintenanceandCleaning Thestirringdrive(2)ismaintenance-free.

    Themagneticdriveisembeddedinasyntheticresinandiscompletelywatertight.Cleanthesurfaceofthestirringdrive(2)inregularintervalsusingacleaningoradisinfectantsolutionsuitableforuseonPVC.Cleanthesurfaceofthestirringcontroller(3)usingdrycloth.

    y CAUTION! Incaseofrepair,theequipmentmustonlybeopenedbyanauthorisedserviceagent.Disconnectfrommainsbeforeopening!

    8.8 Troubleshooting

    8.8.1 Themagneticstirringrodrotatesunevenly:

    Overtime,someageingof thestirringrodcannotbeavoided.Thismayreduce itsmagnetisation.

    y CAUTION! Donotallowthemagneticstirringrodtocometorestinanalternatingmagneticfield(e.g.inanunfilledstirringvesselwhenthestirringormixingmodeisselected).Donotsubjectthemagneticstirringrodtoastronginversemagneticfieldortostrongimpactshocks(e.g.bydroppingit).Thismayresultindestructionofthemagneticstirring

    rod.Replaceitbyanewone.

    8.8.2 Thestirringrodswervesbackandforth:

    Swervingmovementsofthestirringrodmaybecausedbyirregularshapedbottomofthebottle.Replacethebottlebyanewone.

  • 27

    9. TechnicalData

    9.1 BODSystem

    Type Oxi700

    Measurementprinciple Manometric;mercury-free;electronicpressuresensor

    Ranges 0-40,0-80,0-200,0-400,0-800,0-2000,0–4000mg/l[mg/lO2]

    Applications BSB5,BSB7,OECD301F...

    Resultdisplay BSB[mg/l];4-digits;7-segmentLED

    Measurement BOD-range,volume,duration,displaytimeofmeasurementparameterdisplay

    Measurementperiod User-selectable,between1and28days

    Autoresultstorage upto28results,dependingonmeasurementperiod

    Storageinterval –hourly(1day); (formeasurementperiodof...) –every2hours(2days); –daily(3-28days)

    Automaticstartfunction –automaticstartafterpressuredecrease –canbeswitchedoff

    Powersupply 3alkaline-manganesebatteries

    Batterylife 1year(normaluseasBOD5meter–max.onereadingaday); Low-Batt-Indication

    Clock Real-timeclock

    Protectionclass IP54(sensorhead)

    Dimensions 375x195x230mm (LxWxH) includingstirringunit

    Housing ABS

    Approval CE

  • 28

    9.2 InductiveStirringSystem

    Type InductiveStirringSystem

    Stirringvolume 6BODbottles

    Stirringpower 7W

    Rotationspeed 320rpm (reducedto200rpmevery40s)

    Dimensions(WxDxH) 270x180x25mm

    Distancebetween 88mm stirringpoints

    Weight(stirringdrive) 1800g

    Housingmaterial PVC

    GeneralAmbientConditions -10to+56°Cat100%humidity

    Operatingvoltage 20Vmax.

    Controlcable 2m

    Protectionclass IP68(DIN40050)

    Stirringcontroller 230Vversion 220–240V/10V/50-60Hz,protectionclassII, thermalfuse,CE 115Vversion 110–120V/10V/50-60Hz,protectionclassII, thermalfuse,CE Permanentambientconditions 0to+40°Catmax.80%humidity Protectionclass IP20(DIN40050)

  • 29

    10. AccessoriesandSparePartsList

    Article Ordercode

    ReplacementBODsensor L2444430

    ReplacementBODbaseunit L444433

    Stirringcontroller L444413

    Inductivestirringdrive L444411

    TestSet L418328

    Potassiumhydroxidesolution(50ml) L2418634

    NitrificationinhibitorB(50ml) L2418642

    BODsamplebottle L418644

    Sealcup(gasket) L418636

    Diagrampad L418663

    Magneticstirringrod L418637

    Magneticstirringrodremover L418638

    Overflowmeasuringflask,428ml L418660

    Overflowmeasuringflask,360ml L418659

    Overflowmeasuringflask,244ml L418658

    Overflowmeasuringflask,157ml L418657

    Overflowmeasuringflask,94ml L418656

    Overflowmeasuringflask,56ml L418655

    Overflowmeasuringflask,21.7ml L418664

    Completesetoverflowflasks L418654

    Thermostaticallycontrolledincubators seegeneralcatalogue

    Thermostaticallycontrolledincubatorswithglassdoor seegeneralcatalogue

  • Orbeco-HelligeInc.,185MarineStreet,Farmingdale,NY11735

    Phone:631-293-4110,Fax:631-293-8258,[email protected],www.orbeco.com