Upload
dinhcong
View
218
Download
0
Embed Size (px)
Citation preview
BIOSENSOR PROJECT
Roi Leibowitch
Shahar Kvatinsky
Idith Farkas
Instructors: Prof. Aharon J. Agranat, Harel Ilan
Picture of the system
2
Contents
Abstract 3
Chapter 1 - Previous work 4
Chapter 2 - Biological Background 6
Chapter 3 - The system 10
Chapter 4 - Experiments Summary 16
Chapter 5 – future work 28
Summary 31
References 32
Appendix 1 – Data sheets 33
Appendix 2 – Labview control program 37
Appendix 3 – NI CompactRio Spec 40
Appendix 4 – Minimal medium MOPS glucose 43
Appendix 5 –An experiment protocol 44
3
Abstract Early warning of chemical hazards that occur in water reservoirs and supply systems is
becoming a key issue in maintaining water quality and safety in urban and rural
environments.
We present an integrated system for realtime inline monitoring of pollutants in water,
based on E-Coli bacteria.
The core of the system is genetically engineered E-coli microbes that respond to the
presence of the pollutants by producing a dye material. The amount of the dye produced
is in turn measured in an opto-electronic detection device which consists of a LED and a
detector. The system includes an automatic sampling apparatus consisting of 10 cuvettes1
in which the bacteria, dried, lie until a sample is taken into a cuvette from the reservoir.
Up to two hours from the time a request is made to sample the water, a result stating
whether the water is toxic or not is available then transmitted via WiFi to a base station.
1 A type of square test tube designed to let light pass through it for optical testing
4
Chapter 1 – Previous work In this section we will present the previous work done in the field of water toxicity
analysis (WTA).
It is common to divide WTA to two separate fields, chemical and biological. The
difference is in the discovery mechanism. In the chemical mechanism, the response to a
certain chemical is measured, while when dealing with biological mechanism, a reaction
of a biological element is measured.
As for today, both of the options above usually require manual sampling of the water
source, and transferring the sample to a central lab, where the process of examining the
water might take up to 24 hours. Obviously, this time is too long when we deal with
water pollution. In addition, the cost is several hundreds of dollars for each sample.
The chemical analysis of the tested sample is a more mature technology, and is used to
identify specific toxic chemicals or biological elements, from a closed list of materials (as
opposed to biological detection, that can identify a general problem, Meaning to react to
unknown toxins in a sample, as can be done with bacteria, more on this subject will be
explained in the next chapters).
The biological mechanisms can be divided into a number of categories, based on the
reaction of the bacteria2:
• Luminescence - the bacteria are emitting light.
• Fluorescence – the bacteria are creating fluorescence material.
• Colorimetric – the bacteria cause a change in the color of the environment.
A great number of academic and industrial researches were performed in this field [not
only in regards to water toxicity but for general detection as well] here are some of the
main actions:
• TAU water toxicity research group - “Tel-Aviv University – Toxicity
Measurement System” (TAU-TMS) was a solution for working with fresh whole
cell biosensors, however wasn’t sensitive enough to work with rehydrated
biosensors. This system used fluorescence bacteria. Another TAU system was
based on luminescence bacteria [ 8, 2].
2 Elaboration on this terms is in the Biological Background section
5
• Toshiba Corporation has developed essential technology for an advanced
biosensor chip capable of wide-ranging applications. Such system can measure
glucose in blood with up to 500 times the sensitivity of current biosensors. That
new biosensor comprises of a sensor chip that interacts with the sample and a
detection section that records the result [ 2].
• PointSource Inc. has developed a cost-efficient testing apparatus that will allow
for continuous, real time monitoring of the water supply for the presence of
dangerous pathogens. The patented system relies on a laser beam to identify
microorganisms in water. A side stream of water, which assumed to be
representative of the water in general, passes through a laser beam. As it passes
through the beam, light is scattered. The system gathers the light and, using a
mathematical technique, can determine what kind of particle it is by its shape, size
and internal composition [ 2].
• Cyanide detection: Colorimeters test kit includes reagents that when added to a
water sample, react with the available cyanide ion to form a colored solution. A
vial containing this solution is inserted into the hand-held colorimeter that
measures the intensity of the sample's color and reports the cyanide concentration
[ 2].
• South Korea Advanced Environmental Monitoring Research Center - multi-
channel continuous water toxicity monitoring system was implemented to sample
water discharged from power plants in order to detect and classify their toxicity,
using several recombinant bioluminescent bacteria. Each channel of the system is
composed of a series of two mini-bioreactors to enable a continuous operation,
i.e. without system interruption due to highly toxic samples [ 5].
As for now, there is no system which is based on colorimetric biological detection,
although this field is well known for biological measurement. There is a lot of biological
data on colorimetric detection, but not from engineering perspective.
Another issue that can be noticed is that there is no biological water sampling system
which is continuous and stand alone.
6
Chapter 2 - Biological Background The basic concept is to use bacterial reporter strains that are genetically engineered to
react to different types of selected toxins. The reaction can have optical characteristics
and is dependent on the bacterial strain and the level of the toxin concentration.
The concept behind the engineering of the reporter strains is the fusion of the DNA
sequence of a gene promoter, known to be activated by the presence of genotoxic chemicals,
to a gene or a group of genes the presence or activity of which can be monitored
quantitatively, preferably in real time. After that, whenever the bacteria are exposed to this
material, the promoter will initiate transcription and protein formation.
We compared several types of bacteria's reactions; each reaction is involved with
different type of protein:
1) Fluorescence bacteria – The bacteria’s reaction creates a fluorescence protein, that
is excited (absorbed) at a unique wave length and emit at longer wave length.
In the world of Biosensors the drawback of the fluorescent proteins is the fact that
there is no amplification of the signal- one protein=one excitation, not like with
enzymatic reactions in which one protein can create 100 products that produce a
signal.
Common proteins - Red Fluorescence Protein (DsRed), green fluorescence protein
(GFP)
Advantages - very stable, lasts for days.
Disadvantages – initial reaction time is slow (reaction starts after 60-100
minutes), not sensitive and has long time intervals for a substantial reaction.
2) Luminescence bacteria – The bacteria’s reaction is creating a luminescent protein,
the bacteria are producing a proteins complex that enables the enzymatic reaction
in which photons are released. The creation of the reaction consists of two
processes – 1. The synthesis of the enzymatic complex that produces the substrate
and creates the luminescent reaction. 2. The enzymatic reactions in which the
substrate is produced and photon is released.
Common proteins - luxCDABE
Advantages - fast detection (after 30-60 minutes), sensitive signal detection.
Disadvantages – fade after ~3 hours, low intensity.
7
3) Colorimetric bacteria – The bacteria’s reaction is creating an enzyme, that when it
reacts with colorimetric substrate it creates a colorimetric product. The product
absorbs light at a specific wave length.
Common colorimetric product - p-nitrophenol (PNP), ortho-nitrophenol (ONP).
Advantages – fast initial reaction and fast detection (after ~30 minutes),
continuous enzymatic activity (the process is continuous and therefore keeps
producing the protein).
Disadvantages – no relevant data for engineering purposes (never been used for
such purposes), there is a spontaneous reaction even without a toxin (slower
reaction), toxins with color may influence the readings.
Figure 1 – Colorimetric detection
In figure 1 there is a schematic of the detection process, as explained above.
We decided to use colorimetric bacteria, due to its advantages.
Substrate
There are several types of colorimetric substrates:
1) para-nitrophnyl phosphate (PNPP) – The reaction of the substrate with alkaline
phosphatase (type of an enzyme) produces the product p-nitrophenol (PNP), a
yellow material that its absorption spectrum is as seen at figure 2 (center of
absorption is 405nm). More soluble than ONPG.
8
Figure 2 -– PNP absorption (different PH) [1]
2) Ortho-Nitrophenyl-β-galactoside (ONPG) - The reaction of the substrate with β-
galactosidase (type of an enzyme) produces the chromogenic molecule ortho-
nitrophenol (PNPG), center of absorption 420 nm.
We characterized the substrates (see experiments chapter) and decide to use PNPP.
Drying of the bacteria:
Different types of storage of the bacteria:
1) Liquid – a solution of substrate and bacteria. The fastest way for reaction (the
bacteria is alive). In order to be preserved for longer time (several weeks) and to
prevent bacteria reaction (until sampling) it needs to be cooled (4°C).
2) Solid firmly fixed – the bacteria is fixed on an agar (complex sugar), which lets
the liquid to penetrate into the bacteria, and also requires cooling.
3) Dried – drying the bacteria, the substrate and the medium into powder. Can be
stored for a long time (1-2 years). Needs vacuum or a very dry environment. It is
better to store the powder at low temperature. Takes more time to “rehydrate” the
bacteria (another ~20 minutes). 50% of the initial live bacterial cells die during
the process.
9
We decided to use dried bacteria in order to get an independent machine that can be
placed near a source of water for a long time, without the need of maintenance.
The drying process:
We mixed together the correct concentration of bacteria and PNPP, including MOPS (the
medium, water with sugar, salts and amino acid, see appendix 4) and threalose (a sugar
that helps the bacteria to survive the drying process).
The drying process is actually lyophilization process.
The drying stages:
1) Freezing – in order to stop the biological processes cooling the solvent to -40°C.
There is a faster process which is to dip the solvent in liquid Nitrogen. We didn’t
use this process.
2) Vacuum – in order to take out the water vapor.
3) Temperature rising – as the water vapor is taken out, the pressure becomes lower.
Therefore we raise the temperature in order to keep the process of taking out the
water vapor – the process is finished at 30°C and lasts about 2 days.
Since it was the first time that Professor Belkin’s lab dried bacteria in cuvettes, we had to
deal with several aspects:
1) Cuvettes temperature proof at low temperature – we cooled the plastic cuvette to
-80°C and saw that it can withstand such low temperature.
2) Find a way to seal the cuvette - we ordered special taps, which allow us to keep
the dried powder under vacuum.
3) Mixing substrate and bacteria together in the drying process.
4) The cuvettes are made from plastic and therefore have worse heat conductivity;
we cope with it by using Copper small plates that we coupled to the cuvettes.
10
Chapter 3 - The system In this section we will present the system, its various parts and their way of work.
The system we built is capable of detecting the presence of genotoxicants in water
sampled from a reservoir and transmitting its findings to an off location base station. We
used the substance Naladixic acid (NA)3 as a model toxin. In our system, as stated in the
Biological Background section of this document, genetically engineered E. coli reporter
strain respond to the presence of the NA by altering the color of a substrate material.
NA was used as a model toxin to characterize our system and calibrate it, but any of a
large number of substances can be detected using our system with the need only to
recalibrate it. In such a case the use of a different strain of bacteria, engineered to respond
to another toxin should be used.4
The biological reaction takes place inside cuvettes5. The bacteria are dried in a process,
also described in the Biological Background section of this document. The dried bacteria
stay in the cuvettes under vacuum untouched for up to months until the water is sampled.
The cuvettes are part of a motorized stack, in the case of our prototype containing 10
cuvettes. When all the samples in the stack are finished, the stack needs to be replenished.
The conclusion stating whether there is NA present or not is made according to a
predetermined toxin level under which we state the water isn’t toxic and above it that it
is. It takes up to two hours to get a clear reaction from the bacteria and reach the
conclusion. The reference level of toxin was determined based on the series of
experiments we conducted on both dried and non dried bacteria using different levels of
NA. These experiments are presented in the Experiments Summary section of this
document.
3 Type of antibiotics 4 See Future Work section for further details on the possibilities of using such bacteria 5 Type of testing tube used to put light through the sample inside it
11
Figure 3: Scheme of the system6
The biological reaction
Originally the substrate used in the biological reaction is PNPP – a colorless material.
Once the bacteria are mixed in with the NA and the PNPP, the latter turns into PNP
which is yellow. As the concentration of NA is higher, so is the concentration of yellow
6 Whenever a connection is marked by D17 for example, it is connected to port 17 of the DAQ
12
material in the mixture. It is the amount of color produced that our system, in fact,
detects.
Still, we take into account that this reaction occurs also when there is no NA present,
since the enzyme responsible for it is produced by the bacteria at all times, in basal levels.
For this reason all reaction is measured against the reaction seen with no NA at all,
referenced in this document as a concentration of 0 PPM NA.
The opto-electronic system
A LED of 400 nm7 was picked so that its spectrum matches that absorbed by the yellow
product – PNP. Figure 4 and figure 5 show that match.
Figure 4: Emission spectrum of our LED
The light of the LED is concentrated using an objective lens and then illuminates a
cuvette containing bacteria, PNPP substrate turning into PNP and a sample of water from
the reservoir. At the other side of the cuvette, a visible light photodiode detector8 collects
the light that was not absorbed by the sample and changes its reading accordingly.
The detector's reading in volts is received by a DAQ card and processed in a computer.
7 See appendix1 for the data sheet of the LED 8 See appendix 1 for the detector’s photodiode’s datasheet
Figure 5: PNP absorption (different PH)[1]
13
The sampling system
10 cuvettes are aligned in a motorized stack. Each cuvette is held in place by a copper
mount and separated by an insulating material from the others. The copper mount eases
heat flow from a Thermo-electric cooler (TEC) placed under the cuvette used for the
current sample. The TEC is used to heat the sample to 37ºC – an ideal temperature for the
bacteria's reaction.
A new cuvette comes to its place in the opto-electronic system whenever a new sample is
requested. The TEC moving under them is placed on a spring so that it is pressed against
the copper mount for maximum contact and heat flow. The mounts are curved at the
edges to facilitate the TEC's movement across them.
A pump with adjustable speed pours water into the current cuvette. At the end of the pipe
there is a syringe and a needle. A stepping motor connected to the syringe moves it down
to puncture the stopper of the cuvette, then, the water flows through the syringe into the
cuvette.
Before each sample is taken, 0.5 liter of water from the reservoir is pumped through into
a sink, which is part of the cuvette stack, to remove any leftovers of the latest sample in
the pumping system.
To stir the sample, mini magnetic stirring rods placed inside each cuvette and a magnet
outside the cuvette are used.
Figure 6: A picture of the opto-electronic system
14
Controlling and processing
All instruments of the system are controlled via computer using a DAQ card and a
Labview program shown in appendix 2.
The result, stating if there is toxin present or not, is transmitted via e-mail to an off
location base station.
When a request is made from the base station to take a sample the following commands
are given to the system:
Figure 7 : A sketch of the cuvette stack used in its planning
Figure 8: A picture of the cuvette stack
Figure 9: A cross section of the cuvette stack showing the TEC under the current cuvette
mounted on a spring for maximum contact
15
0. Move to sink
1. Puncture the cuvette stopper
2. Set pump speed to 4.35 V (speed 200 RPM)
3. Activate pump
4. Wait 4 minutes to clean the pipes
5. Deactivate pump
6. Get out of stopper
7. Move to sample location, set TEC to 37°
8. Puncture the stopper
9. Set pump speed to 1.407 V (speed 10 RPM)
10. Activate pump
11. Wait 4 seconds to pump 0.6 ml of water
12. Deactivate pump
13. Get out of stopper
14. Stir
15. Activate LED
16. Wait 2 hours and collect data
17. Compare to reference toxin level and send message
18. Deactivate LED and turn off TEC
16
Chapter 4 - Experiments Summary All the experiments were made on bacteria that were genetically engineered to react to
Nalidixic acid (NA) as the toxin. The genetically engineering was done in Prof. Belkin
Lab.
The experiments’ steps:
1. Optimization test – substrate – 12/3/09.
2. Optimization test – toxin concentration - 25/3/09.
3. Optimization test – bacteria concentration – 24/6/09.
4. Kinetic test (reaction over time) – 1/4/09, 13/5/09, 20/5/09, 1/6/09.
5. Dried Bacteria characterization – 29-31/7/09, 25-27/8/09, 1/9/09-3/9/09,
6. Heating characterization – TEC vs. incubator – 31/7/09, 4/8/09.
7. Full system characterization – 8/9/09, 10/9/09, 17/9/09.
The optimization tests were done on each component in the biological part (substrate,
bacteria and toxin). We made those tests using liquid bacteria and characterized the
optimal substrate (PNPP vs. ONPG) concentration, the optimal bacteria concentration
and the reaction of the bacteria to different toxin concentration. From those experiments
we determined the substrate and bacteria concentration and the reference voltage for
the bio sensor system.
In the kinetic assay we monitored the reaction over time, this way we could determine the
sampling time of our system.
The assay using the dried bacteria was made in order to find the differences between
dried and liquid bacteria and to get a more accurate determination of the bacterial
concentration, the reference voltage and the sampling time.
In the heating characterization we determined the correct TEC temperature in order to
get 37°C inside the cuvette. We also compared the reaction inside an incubator and in a
cuvette coupled to a TEC.
In the full system characterization we performed a full running of our system. We
determined from those experiments the time and speed of the pump, considered the
steering issues and checked the full integration of the system.
17
Note – different experiments were made in different dates and different conditions (new
bacteria, different equipment etc.), therefore we can't compare the absolute results of one
experiment to another and can only compare the trend. The parameters for the final
system are determined in the last experiments on the system in its final setup.
The experiments:
1. Optimization test – substrate:
Date: 12/3/09.
Goal: decide which type of substrate to use (PNPP vs. ONPG).
Description: we compared the colorimetric product of PNPP (PNP) vs. the
colorimetric product of ONPG (PNPG). We wanted to see the concentration of
the product in the water and the values that were measured in this concentration
(the measured voltage, which implies on the absorption of light).
We determined the height of the water in the cuvette (different volumes in similar
cuvettes) and how the measured voltage depends on it, in order to determine the
optimal water volume for sampling (minimum volume that can collect all the light
from the LED).
We validated that presence of PNPP and ONPG keeps the water clear (no change
in the measured voltage).
Results: we saw that PNPP is much more soluble than ONPG and therefore made
more tests on it.
The results of the measured voltage in different concentration of PNP are shown
in graph 1.
18
PNP concetration test
0
1
2
3
4
5
6
7
0.01 0.1 1 10 100
PNP concetration [mg/ml]
Vo
ltag
e [V
]
Graph 1 – PNP concentration test
Conclusions:
a. Both substrates (PNPP and ONPG) keep the water clear.
b. PNPP is much more soluble and therefore is preferred.
c. Even a small amount of PNP is noticeable; there is a significant difference
between 0 and 0.05 mg/ml.
d. The optimal volume of water in the used cuvettes is 0.6 ml.
2. Optimization test – toxin concentration:
Date: 25/3/09.
Goal: characterization of the bacteria's reaction to different toxin concentration.
Description: we mixed the bacteria with the medium, the substrate and various
concentrations of the toxin (NA) and incubated it for one hour and a half. Then
we measured the reading in our detector 20 and 40 minutes after the incubation
(left in room temperature).
We put in every cuvette 0.3ml MOPS G (the medium), 0.3ml distilled water and
0.08% PNPP.
The bacteria OD (optical density) were 0.25.
We added 2 sets of cuvettes with different NA concentrations (0.15, 0.3, 0.6, 1.25,
2.5, 5 and 10 ppm) and also 2 control cuvettes (0 ppm of NA) and a blank cuvette
(a cuvette with NA but without bacteria).
The biological experiment protocol is in Appendix 5
Results: the results relative to the control (no NA) and in absolute value in volts
minus reading of control after 20 minutes are presented in graph 2 and graph 3.
19
20 minutes afer incubation, relative to 0 PPM
0
0.5
1
1.5
2
2.5
3
0 2 4 6 8 10 12
NA concentration [PPM]
Rea
din
g r
elat
ive
to 0
PP
M
Maximum value
Minimum value
20 minutes afer incubation, subtracted from 0 PPM
0
1
2
3
4
5
6
7
8
0 2 4 6 8 10 12
NA concentration [PPM]
Rea
din
g s
ub
trac
ted
fro
m 0
PP
M [
V]
Maximum value
Minimum value
We tested two samples of each NA concentration, which produce 4 data points of
results relative to 0PPM NA since there are two 0PPM samples as well. In the
graphs, maximum and minimum values refer to the maximum and minimum of
those 4 data points.
Graph 2 – Different NA concentration relative to 0 ppm
Graph 3 – Different NA concentration subtracted from 0 ppm
20
Voltage reading vs. bacteria concentration with close detector
0
5
10
15
20
25
0 0.2 0.4 0.6 0.8 1 1.2
bacteria concentration [OD]
det
ecto
r re
adin
g [
volt
]
10 ppm, A
2.5 ppm, A
0 ppm, A
10 ppm, B
2.5 ppm, B
0 ppm, B
Conclusions:
a. There is a significant difference between no pollution (0 ppm) and
presence of toxin (at least 1.5 V difference was measured).
b. The optimal reaction was measured at 2.5 ppm; there is a decline in
the reaction in a more concentrated toxin (due to bacteria's death
from the toxin).
3. Optimization test – bacteria concentration:
Date: 24/6/09.
Goal: determine the optimal bacteria concentration.
Description: we measured the response after an hour and a half in an incubator
with 4 different concentrations of bacteria (OD of 0.1, 0.25, 0.5 and 1), each with
3 different concentrations of NA (0, 2.5 and 10 ppm).
We also made two separate measurements with our detector placed at different
distances from the cuvette, each more suitable for a different range of voltages
corresponding to different concentration of bacteria.
Results: the results for the closer detector present in graph 4.
Graph 4 – The measured voltage for different bacteria concentration
21
Conclusions:
a. For an OD of 0.25 the results are best, with greatest separation
between various NA concentrations.
b. The detector placed closer, fits bacteria concentration of 0.25 OD
better, giving a voltage range of 15 volts between 0 ppm and 10 ppm
of NA.
c. For greater OD we get great dispersion of the light giving less
reliable results, since they can vary with changes in the cuvette
angle, and anyway the difference of voltages is not better than in the
0.25 OD case.
4. Kinetic test (reaction over time):
Date: 1/4/09, 13/5/09, 20/5/09, 10/6/09.
Goal: testing the reaction upon time and approximate the time for optimal
measurement of the level of toxicity.
Description: we prepared the cuvettes as mentioned in Appendix 5 (2 samples for
each NA concentration – 0, 0.3, 0.6, 1.25, 2.5, 5, 10 ppm), we put all cuvettes in
an incubator and tested it every 20 minutes.
We used OD 0.14 on 13/5, OD 0.22 on 20/5 and OD 0.238 on 10/6.
Results: we had difficulties to get good results for those experiments. The
experiments on 1/4/09 and 13/5/09 were a failure. There are several options for
the reasons of the failures, but we couldn't define the exact reason:
• Contamination of the samples
• Temperature changes
• Low OD
The result on 10/6 for all concentrations is presented in graph 5.
The result on 10/6 after 2.5 hours is presented in graph 6.
22
Kinetics Experiment Results
0
2
4
6
8
10
12
14
16
18
0:43:12 0:57:36 1:12:00 1:26:24 1:40:48 1:55:12 2:09:36 2:24:00 2:38:24 2:52:48
Time [Hours]
De
tect
or
rea
din
g [
V]
0א 0ב
0.3א 0.3ב
0.6א 0.6ב
1.25א 1.25ב
2.5א 2.5ב
5א 5ב
10א 10ב
NA concentration [PPM]
Detector reading for various NA concentration after about 2.5 hours
00.5
11.5
22.5
33.5
0 2 4 6 8 10 12
NA concentration [PPM]
Det
ecto
r re
adin
g [
V]
Reading a
Reading b
Graph 5– Measured voltage depend on time, for all NA concentration
Graph 6– Measured voltage depend on NA concentration, after 2.5 hours
Conclusions:
a. There is great importance to the bacteria OD (must be at least 0.2 OD). This is
probably the reason for the failure experiments in our first experiments.
b. The optimal time for sampling is about 1.5 hours (biggest difference in
voltage between different concentrations.
c. With respect to previous experiments ,in this experiment we didn't have a
decline for high NA concentration.
23
5. Dried Bacteria characterization:
Date: 29-31/7/09, 25-27/8/09, 1/9/09-3/9/09
Goal: find the differences between dried and liquid bacteria; prove the feasibility
of drying the bacteria in cuvettes.
Description:
First, we dried bacteria in small glass bottles (the orthodox way), as described in
the biological part of this document. We dried 3 bacteria concentrations (0.25,
0.35, 0.5 OD).
The technique for this is to grow and fresh the bacteria up to 0.25 OD, then
decrease its volume 20 times to 5 OD and by adding threalose9 we diluted the
bacteria and made it a suspension (2.5, 3.5, 5 OD).
We dried 4 bottles of each bacteria concentration (2 with copper stands in order to
improve heat conductance and 2 without it), each one contained 0.15 ml of
bacteria, 0.15 ml MOPS G (being condensed 10 times) and 0.015 ml PNPP (40
mg/ml concentration).
The drying process had taken 2 days, and then the dried bacteria were diluted with
1.5 ml of distilled water (in order to get 0.25, 0.35 and 0.5 OD).
We checked the reaction of the rehydrated bacteria by transferring 0.6 ml of it to 2
cuvettes. We add 24 cuvettes, for each bacteria concentration we had 2 cuvettes of
each NA concentration (0, 2.5 and 10 ppm).
After making the cuvettes we inserted it to an incubator and check the reaction
after 1.5 hours, 2.5 hours and 3 hours.
The next drying experiments were in order to check the drying process in
cuvettes. We had to find special stoppers and to use coupling for the cuvettes for
better heat conductance.
We dried the bacteria in the same way as in bottles (in each cuvette 0.06 ml of
bacteria, 0.06 ml of MOPS G condensed 10 times, 0.006 ml of PNPP of 40 mg/ml
concentration). We also used UV cuvettes in order to check their impact on drying
(they have different stoppers).
9 Substance used to facilitate the rehydration of the bacteria
24
After three hours
0
1
2
3
4
5
6
7
0 2 4 6 8 10 12
NA concentration [PPM]
De
tect
or
rea
din
g [
Vo
lt]
0.5 OD
0.35 OD
0.25 OD
After one hour and a half
0
2
4
6
8
10
12
14
16
0 2 4 6 8 10 12
NA concentration [PPM]
De
tect
or
rea
din
g [
Vo
lt]
0.5 OD
0.35 OD
0.25 OD
The drying process had taken 2 days, and then the dried bacteria were diluted with
0.6 ml of distilled water (in order to get 0.35 OD).
After making the cuvettes we inserted them into an incubator and checked the
reaction after 2 hours, 3 hours and 4.25 hours.
The last drying experiment was the same as the second one, only with 0.25 OD.
Results:
The results after 1.5 hours and 3 hours for the first experiment (drying in bottles)
are presented in graph 7 and graph 8.
Graph 7– Measured voltage vs. NA concentration, after 1.5 hours
Graph 8– Measured voltage vs. NA concentration, after 3 hours
25
The results after 3 hours for the second experiment (drying in cuvettes, OD –
0.35) are presented in graph 9.
With copper stands - after 3 hours
0
2
4
6
8
10
12
14
16
0 2 4 6 8 10 12
NA concentration [PPM]
De
tect
or
rea
qd
ing [
Vo
lt]
three hours
Graph 9– Measured voltage vs. NA concentration, after 3 hours
Conclusions:
a. We can see from graph 7 that the optimal OD for drying bacteria is 0.25 (same
as in liquid bacteria).
b. There is great importance during the drying process to heat conductivity (the
cuvettes are made of plastic) as well as to symmetry between the cuvettes.
c. There was a technical problem in the lyophilization machine, causing the
drying process to be unreliable.
d. For each set of drying bacteria, our system needs to be characterized.
6. Heating characterization – TEC vs. incubator:
Date: 31/7/09, 4/8/09.
Goal: determine the correct TEC temperature in order to get 37°C inside the
cuvette and comparing between the reaction inside an incubator and in a cuvette
coupled to a TEC.
Description: first we determine the temperature in the TEC in order to get to a
temperature of 37°C for the water inside the cuvette.
Then we took the bottles of dried bacteria, we transferred the rehydrated bacteria
to cuvettes with NA concentration of 2.5 ppm. We measured the reaction for
26
different bacteria concentration (0.25, 0.35, 0.5 OD) and put the cuvette inside the
incubator and similar cuvette coupled to a TEC.
We compared the measured voltage between the coupled to TEC cuvette and the
incubator inserted cuvette.
Results: in order to get 37°C for the water inside the cuvette, we need to set the
TEC up to 41°C.
The results for 0.25 OD is presented in table 1.
Time
[hours]
Incubator
[Volt]
TEC
[Volt]
1.5 3.47 3.61
2.5 2.16 2.17
Table 1 – 0.25 OD, 2.5 ppm NA concentration, comparing the measured voltage between TEC and incubator
Conclusions:
a. In order to get 37°C for the water inside the cuvette, we need to set the TEC
up to 41°C .
b. We see that there is no significant difference between TEC coupling
cuvettes and incubator inserted cuvettes.
7. Full system characterization:
Date: 8/9/09, 10/9/09, 17/9/09.
Goal: show the functionality of the system, determine the time and speed of the
pump, consider the steering issues and check the full integration of the system.
Description: We took the complete system and tested it on several NA
concentrations. We characterized the time and speed needed for 0.5L of water and
0.6ml of water to being pumped and checked the system reliability.
Results: as mentioned in the drying experiments, we had technical problem in the
lyophilization machine and therefore we had a problem to characterize the full
system reference voltage.
We made several sampling and saw the time needed for proper work of the
system and the parameters needed for the10.
10 For more details, see the system part of this document
27
Conclusions:
a. The system proof of concept is done – the system is working properly.
b. There is a need to characterize the reference voltage for each set of drying
bacteria and each toxin.
c. The system seems reliable, although it is needs to be improved when it comes
to puncturing the taps of the cuvettes and the reliability of the stack.
28
Chapter 5 – future work
The Bio-sensor system we presented in previous chapters has various options for
improvements and enhancements which we discovered during the ongoing work.
In this chapter we will present briefly the most important findings we have on the future
implementation of the system:
• Miniaturization and power saving: currently, the system is defined as a
prototype and it is based on several power supply and desktop computers. At
future deployment the system will have to be deployed in the field, which will
make it impossible to use desktops, and even notebooks might present some
problems. An optimum solution might be the NI CompactRio system.
CompactRio includes main processor and various other hardware modules such
as: Wireless module, Satellite communications and more. All the modules are in
rigid a case which is suitable for field deployment. One of the most important
benefits of this system is that the VI controlling the system can be easily exported
to the CompactRio, and save the development time.
Figure 10 -NI CompactRio system
• WSN – Wireless Sensor Networks is a relatively new discipline in the academy
and industry. It's main interest is creating sensor networks which :
o Share information between the sensors in minimum payload.
o Perform various calculations on the information shares.
o Change deployment and information sharing on the fly.
o Raise issues to the network manager.
Bio-sensor network which will work in WSN mode will enable a lot of vital
information that can enhance even further the system's capabilities. For example:
o Detecting malicious attacks on the Bio-sensor system.
29
o Monitor pollution propagating in water sources.
o Cross checking of results from a certain system.
A good example of WSN deployment is of systems that are on the shores of a
water source. These systems communicate with each other and send the
aggregated information to a base station.
• Detecting a large number of toxics: as presented in the previous chapters,
during the stage of genetically engineering the E.coli reporter strains, it is possible
to design strains that will be activated by a different spectrum of toxic substances,
more over, it is possible that each strain will have a different reporting gene which
would activate different substrates, so the signal produced will be unique in
wavelength and/or intensity. We can use these options to discover a few toxins in
one system in one of the following methods :
o Create a few cuvettes in one system; each one contains a different type of
bacteria. In this way, in every sample we can discover a different type of
toxin.
o Create a cuvette which contains few bacteria, each one can sense a
different toxin. The reaction for each kind of toxin will change the
substrate color to a different one. Using advanced technique, it is possible
to discover which toxin or combination of toxins caused the specific
reactions.
• Complex data analysis - The system prototype can only discover the existence of
a toxin or lack of it, as described in the previous chapters. More complex data
analysis that includes performing various mathematics calculations on the data
from the system might obtain the level of toxicity as well. For this analysis to take
place there is a need to characterize further the reaction of the bacteria to all level
of toxins.
• Enabling continuous sampling – As for now, the system is limited to 10
cuvettes, which enables limited number of samples. There is a need to allow
more cuvettes to be part of the system, and yet allowing easy replacements of the
cuvette stack.
30
• Adding controls: - As for now, the system lacks internal controls. The purpose of
the controls is to know that the system is working properly. There are 3 different
controls :
o Positive controls: In these controls, we will use cuvettes with a known
amount of toxin, and add clean water, not from the water source. The
purpose of this control is to know that our system is working properly, and
discover the known amount of toxin.
o Negative control: In this control, we will use cuvettes with bacteria and a
clean water sample. This control will help us to calibrate the system and
know we don’t get false alarms.
o Constructive control – in this control, we will use cuvettes with a know
amount of toxin, and water from the water source. If we will get a low
read in this control as well as in a sample from the reservoir, this means
that the water sample has a cytotoxic effect on the bacteria and that the
bacteria died before reacting with the substrate. If we get high reading in
the control this means that the water in the water source had little amount
of toxin, plus the toxin we added and the reading is reliable.
In order to add this control, some changes will have to be done in the controlling
VI. As part of the prototype, we inserted some basic infrastructure to enable easy
addition of the control. Also, the system should have access to a clean water
source or alternatively, add clean water to the cuvettes in an enclosed cell. When
the needle goes down on these special cuvettes, then this cell will be breached and
clean water will get into the cuvette.
Memebrane
Clean water cell
Needle arrive into the
cuvette and allows
clean water to flow to
the cuvette
Figure 11 -Clean water cell sample
The needle arrives inside the cuvette and allows clear water to flow into it
31
Summary We managed to characterize and build a system capable of detecting pollutants in water
in a fast, automated and simple manner. Our system makes use of bacteria producing a
colorimetric reaction never before used in a biosensor which includes a complete
sampling and detecting system. The use of such bacteria enables minimizing the time
needed to reach a conclusion about the pollution in comparison with other types of
bacteria usually used in such systems – mainly the fluorescent bacteria.
While the ways of detecting pollutant in water reservoirs used today take up to 24 hours
and require reaching the reservoir whenever a sample is taken, our system enables fast
detection and monitoring from a distance with the need for maintenance only once every
few months, thanks to the use of freeze-dried bacteria.
We present a prototype showing the feasibility of such an all included biosensor using
colorimetric bacteria. In order to improve the system in the future one would have to
examine ways of minimization and cost lowering as well as conduct further biological
experiments and better process the bacteria's reaction to enable more complex data to be
extracted from it. The drying process is yet to mature and is still under investigation in
several groups around the world. It needs to be fitted specifically to a system such as ours
in order to make the best of it when taking our need under consideration.
32
References 1. Ali Niazi and Ateesa Yazdanipour - Spectrophotometric simultaneous determination of
nitrophenol isomers by orthogonal signal correction and partial least squares 2. Arthur Rabner MA thesis – TAU.
3. Baselt, D.R., Lee, G.U., Hansen, K.M., Chrisey, L.A., Colton, R.J. (1997) A high-
sensitivity micromachined biosensor. Proc. IEEE 85: 672-680.–
4. Belkin, S.(2003) Microbial whole-cell sensing systems of environmental pollutants Curr.
Opin. Microbiol. 6: 206–212.
5. Kim, B.C., Gu, M.B. (2005) A Multi-Channel Continuous Water Toxicity Monitoring
System: Its Evaluation and Application to Water Discharged from a Power Plant.
Environ. Monit. Assess. 109:,123-133
6. Frederick C. Neidhardt, Philip L. Bloch and David F. Smith - Culture Medium for
Enterobacteria J.bacteriology 1974 119/3 p.736-747
7. A luxCDABE-based bioluminescent bioreporter for detection of phenol – Abd-El-
Haleem D, Ripp S, Scott C, Sayler GS - J Ind Microbiol Biotechnol. 2002
Nov:29(5):233-7.
8. Towards toxicity detection using a lab-on-chip based on the integration of MOEMS and
whole-cell sensors - Noel M. Elman, Hadar Ben-Yoav , Marek Sternheim, Rachel Rosen,
Slava Krylov, Yosi Shacham-Diamand
9. Recombinant Bacterial Reporter Systems - Shimshon Belkin
10. A Dual-Color Bacterial Reporter Strain for the Detection of Toxic and Genotoxic Effects
- N. Hever, S. Belkin
11. Bioreporters: gfp versus lux revisited and single-cell response - Stefanie Kohlmeier,
Matthew Mancuso, Robin Tecon, Hauke Harms, Jan Roelof van der Meer, Mona Wells.
12. Fluorescence and bioluminescence reporter functions in genetically modified bacterial
sensor strains - Eran Sagi, Navit Hever, Rachel Rosen, Amelita J. Bartolome, J. Rajan
Premkumar, Roland Ulber, Ovadia Lev, Thomas Scheper, Shimshon Belkin.
13. http://www.nuncbrand.com/NAG/DP0010.htm
14. Biran, A. (2009) Genetically engineered bacteria for electrochemical detection of
genotoxicants. MSc Thesis. Plant and Environmental Sciences, the Alexander Silberman
Institute for Life Sciences, Faculty of Science, the Hebrew University of Jerusalem, Israel
33
Appendix 1 – Data sheets Data sheet of the LED used in our system.
From: http://docs-asia.origin.electrocomponents.com/webdocs/02a0/0900766b802a04e3.pdf
34
Data sheet of the pump used in our system
From: http://mrceng2.b-smart.co.il/Media/Uploads/BT00300TSPEC(1).pdf
35
Data sheet of the photodiode used in the detector in our system.
36
From: http://jp.hamamatsu.com/resources/products/ssd/pdf/s1223_series_kpin1050e01.pdf
37
Appendix 2 – Labview control program
And inside the stacked sequence:
38
39
40
Appendix 3 – NI CompactRio Spec
41
42
Appendix 4: Minimal medium MOPS glucose MOPSX10 – after preparing and filtering with 0.22µm filter, it is preferred to divide it to 50ml sterile tubes and to keep it in -20OC. Mixing the following solutions in the given order Conc. MW ml g coments
MOPS PH=8.2 with KOH
1M 209.26 400 83.7 Freshly prepared
Tricine pH=8.2 with KOH
1M 179.2 40 7.168 Freshly prepared
FeSO4 0.01M 278 10 0.0278 Freshly prepared
NH4Cl 1.9M 53.49 50 5.1
K2SO4 0.276M 174.27 10 0.48
CaCl2 5*10^-4M
111 10 7.35mg/100ml
MgCl2*6H2O 0.528M 203.3 10 1.073424 NaCl 5M 58.44 100 29.22 micronutrients 10 H2O 360
Total 1000 You prepare the micronutrients separately and add 10 µl to the medium. You can keep it after filtration in 4°c.
Micronutrients solution Conc. MW
(NH4)6(Mo7)24 3*10-6M
H3BO3 4*10-4M 61.83
CoCl2 3*10-5 237.9
CuSO4 10-5 249.68
MnCl2 8*10-5 161.88
ZnSO4 10-5 287.54
MOPS glucose+0.1% Yeast extract- 0.5l MOPS*10 50ml 0.0132M K2HPO4 5 ml D-glucose 1g YE 0.5g Total 500ml (445ml H2O) After the medium is ready, filtration with 0.22µm mc.
43
Appendix 5: an experiment protocol . rfaE/2TTSsulA::phoA∆של O/Nתרבית .1
.OD=0.25גידול שעתיים עד ). כלי זכוכית(בארלנמייר , MOPS - ב 1:100רענון .2 ):קיווטות יש ליד הספקטרופוטומטר(הכנת סטוקים וקיווטות , בזמן הרענון .3אליהם מוסיפים , mg pNPP 8שוקלים ( mg/ml 0.8 - בריכוז MOPS -מומס ב pNPPסטוק -
10ml MOPS( ,צלזיוס 20- בטמפרטורה של , הסובסטראט נמצא במקפיא .לכל קיווטה MOPS+pNPP( ,300µl(מחלקים את המדיום שכבר מכיל את הסובסטראט - . במבחנות אפנדורף )Nalidixic acid )NAמכינים סטוק שריכוז הרצוי של המשרן -
1000ppm NA= 90µl DDW+10µl NA 10,000ppmהכנת סטוק של • 100ppm NA= 90µl DDW+10µl NA 1000 ppmק של הכנת סטו •מכל ריכוז מכינים דופליקאט - הוספת הנפח הרצוי לכל אחת מהקווטות מהסטוק הנכון •
ניכנס לתוך נפח הנוזל בגלל הנפחים הקטנים NA -חשוב לוודא שהטיפ המכיל את ה( ...)המוספים
נפח הוספה ריכוז סופי בקיווטה סטוק1000 ppm 10 ppm 6 µl 1000 ppm 5 ppm 3 µl 1000 ppm 2.5 ppm 1.5 µl 100 ppm 1.25 ppm 7.5 µl 100 ppm 0.6 ppm 3.75 µl 100 ppm 0.3 ppm 1.9 µl
0
חיידקים מרועננים µl 300מוסיפים לכל קיווטה , לאחר הכנת כל הקיווטות עם הריכוזים הנכונים - . הרצוי OD -שהגיעו ל
)'דק 60 - סטנדרטי(ומכניסים לאינקובאטור לזמן הרצוי , ם כיסוי פלאטותסוגרים את הקיווטות ע -