94
Biophysics Biophysics Understanding Radiation Understanding Radiation Units Units Dr. Yousif Mohamed Y. Abdallah Dr. Yousif Mohamed Y. Abdallah

Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Embed Size (px)

DESCRIPTION

Understanding radiation units Answer True or False 1.The same amount of radiation falling on the person at level of breast, head or gonads will have the same biological effects. 2.Effective dose can be easily measured. 3.Diagnostic reference levels are not applicable to paediatric radiology. 3

Citation preview

Page 1: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Biophysics Biophysics

Understanding Radiation UnitsUnderstanding Radiation Units

Dr. Yousif Mohamed Y. AbdallahDr. Yousif Mohamed Y. Abdallah

Page 2: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Educational Objectives

At the end of the lecture, the students should become familiar with the following:

• Why is it important to measure radiation dose ?• How radiation dose can and should be expressed?• Understand the radiation quantities and units used

in diagnostic radiology.

2

Page 3: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Answer True or False

1. The same amount of radiation falling on the person at level of breast, head or gonads will have the same biological effects.

2. Effective dose can be easily measured.3. Diagnostic reference levels are not applicable to

paediatric radiology.

3

Page 4: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Contents• Dose descriptors outside the patient’s body.• Dose descriptors for effects that have threshold

(deterministic effects)• Dose descriptors to estimate stochastic risks• Diagnostic reference levels• Dose descriptors and units for staff dose

assessment

Understanding radiation units

Page 5: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Introduction

• Several quantities and units are used in the field of diagnostic radiology to measure and describe radiation dose

• Some can be measured directly while others can only be mathematically estimated

5

Page 6: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

6

Two types of radiation effectsStochastic effects

– Where the severity of the result is the same but the probability of occurrence increases with radiation dose, e.g., development of cancer

– There is no threshold for stochastic effects– Examples: cancer, hereditary effects

Deterministic effects– Where the severity depends upon the radiation dose, e.g.,

skin burns– The higher the dose, the greater the effect– There is a threshold for deterministic effects– Examples: skin burns, cataract

Radiation Protection in Paediatric Radiology L02. Understanding radiation unitsUnderstanding radiation units

Page 7: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Hot Coffee – Energy contained in a sip

Excess Temperature = 60º - 37 = 23º1 sip = 3ml3x 23 = 69 calories

Understanding radiation units

Page 8: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Radiation Dose

Lethal Dose= 4GyLD 50/60 = 4 GyFor man of 70 kg

Energy absorbed = 4 x 70 = 280 J= 280/418= 67 calories= 1 sip

Energy content of a sip of coffee if derived in the form of X-rays can be lethal

Understanding radiation units

X-rays

Page 9: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Dose of Radiation

• Radiation energy absorbed by a body per unit mass.

Understanding radiation units 9

Page 10: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Dose Quantities and Radiation units

- Dose quantities external to the patient’s body.

- Dose quantities to estimate risks of skin injuries and effects that have threshold.

- Dose quantities to estimate stochastic risks.

Understanding radiation units

Page 11: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Why so many quantities?Radiation dose is a complex topic

Understanding radiation units

• 1000 Watt heater giving off heat (IR radiation) - unit is of power which is related with emission intensity

• Heat perceived by the person will vary with so many factors: distance, clothing, room temperature

• As can be seen with the example of heat, the energy transformation is a highly complicated issue

• This is the case with X-rays - radiation can’t be perceived

Page 12: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Basic Radiation Quantities

• Used to quantify a beam of X or γ-rays

• There are:– Quantities to express

total amount of radiation.

– Quantities to express radiation at a specific point

12

Radiation at a specific point•Photon fluence•Absorbed dose•Kerma•Dose equivalent

Total radiation•Total photons•Integral dose

Page 13: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Exposure: X• Exposure is a dosimetric quantity for measuring ionizing

electromagnetic radiation (X-rays & Ɣ-rays), based on the ability of the radiation to produce ionization in air.

Units: coulomb/kg (C/kg)

or roentgen (R)

1 R = 0.000258 C/kg

Page 14: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

KERMAKERMA (Kinetic Energy Released in a Material):

– Is the sum of the initial kinetic energies of all charged ionizing particles liberated by uncharged ionizing particles in a material of unit mass

– For medical imaging use, KERMA is usually expressed in air

SI unit = joule per kilogram (J/kg) or gray (Gy)

1 J/kg = 1 Gy

Page 15: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Absorbed dose, D, is the mean energy imparted by ionizing radiation to matter per unit mass.

SI unit = joule per kg (J/kg) or gray (Gy).

In diagnostic radiology, KERMA and D are equal.

Absorbed dose: D

Harold Gray

Page 16: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Mean absorbed dose in a tissue or organ

The mean absorbed dose in a tissue or organ DT is the energy deposited in the organ divided by the mass of that organ.

Page 17: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Now things get a little more complicated !

17

Page 18: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Radiation Dose Quantities

• Primary physical quantities are not used directly for dose limitation

• The International Council on Radiation Protection (ICRP) has defined values for dose limits in occupational exposure

Radiation Protection in Paediatric Radiology L02. Understanding radiation unitsUnderstanding radiation units

Page 19: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Radiation Dose Quantities

Equivalent Dose:

• Accounts for the type of radiation• Different radiation types have different level of

biologic damage per unit absorbed dose

Radiation Protection in Paediatric Radiology L02. Understanding radiation unitsUnderstanding radiation units

Page 20: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Radiation Weighting Factors, wR

Radiation type Radiation weighting factor, wR

Photons 1Electrons and muons 1Protons and charged pions 2Alpha particle, fission fragments, heavy ions

20

Neutrons A continuous curveas a function ofneutron energy

(Source: ICRP 103) Radiation Protection in Paediatric Radiology L02. Understanding radiation unitsUnderstanding radiation units

Page 21: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Equivalent Dose : HT,R

The absorbed dose in an organ or tissue multiplied by the relevant radiation weighting factor :

where DT,R is the average absorbed dose in the organ or tissue T, and wR is the radiation weighting factor for radiation R.

RTRRT DwH ,,

Page 22: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Radiation Quantities and Units

Equivalent dose (Unit = sievert, Sv )– Compares the biological effects for

different types of radiation, X-rays, Ɣ-rays, electrons, neutrons, protons, α-particles etc.

– For X-rays, Ɣ-rays, electrons : absorbed dose and equivalent dose have the same value Gy = Sv.

22

Rolph Sievert

Page 23: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Detriment

• Radiation exposure to different organs and tissues in the body results in different probabilities of harm and different levels of severity.

• The combination of probability and severity of harm is called “detriment”.

• Effective dose reflects the combined detriment from stochastic effects due to the equivalent doses in all the organs and tissues of the body.

Page 24: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Effective Dose: ET

• Effective dose takes into account the organ specific radio-sensitivity to develop cancer and hereditary effects from radiation

• Unit = sievert, Sv

24

Page 25: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Effective Dose: ET

A summation of the tissue equivalent doses, each multiplied by the appropriate tissue weighting factor:

where HT is the equivalent dose in tissue T and wT is the tissue weighting factor for tissue T.

T

TTHwE

Page 26: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

• The organs have different weighting factors, wT.

• These factors are published in ICRP 103 (2007) and have been changed over the years due to increased knowledge.

Understanding radiation units 26

Tissue Weighting Factors, wT

Page 27: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Radiation Protection in Paediatric Radiology L02. Understanding radiation units 2727

• The weighting factors sum up to 1.0.• They are relative and compares one organ with the

other.• They are the same for children and adults!

Tissue Weighting Factors

Understanding radiation units

Page 28: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

• Data is primarily taken from knowledge derived from studying the Japanese population exposed to atomic bombs in Hiroshima and Nagasaki

• On going research has changed the weighting factors from 1990 (ICRP 60) to 2007 (ICRP 103).

28

Tissue Weighting Factors

Page 29: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Multipliers of the equivalent dose to an organ or tissue to account for the different sensitivities to the induction of stochastic effects of radiation.

Tissue Weighting Factors

Tissue weighting

factorwT

*

∑ wT

Bone-marrow (red), Colon, Lung, Stomach, Breast, Remainder Tissues**(nominal weighting factor applied to the average dose to 14 tissues)

0.12 0.72

Gonads 0.08 0.08Bladder, Esophagus, Liver, Thyroid 0.04 0.16Bone surface, Brain, Salivary glands, Skin 0.01 0.04*ICRP 103

**Remainder Tissues (14 in total): Adrenals, Extrathoracic (ET) region, Gall bladder,Heart, Kidneys, Lymphatic nodes, Muscle, Oral mucosa, Pancreas, Prostate, Small intestine, Spleen, Thymus, Uterus/cervix..

Page 30: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units 3030

Dose to lungs times their weighting factor; DL x wL

+Dose (mean absorbed dose)

to gastrointestinal tract times their weighting factor;

DGI x wGI+

....(summation over organ after organ)

=Effective dose

Effective Dose (E)

where T stands for tissuewhere T stands for tissue

T

TTHwE

Page 31: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Effective Dose (E)

We can compare different paediatric imaging procedures through their different effective doses, E.

31

Page 32: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Radiation Quantities and Units used in Diagnostic Radiology

– Incident air kerma– Entrance surface air kerma– Air kerma-area product– Air kerma-length product– Dosimetric quantities for CT – Dosimetric quantities for interventional radiology

32

Page 33: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Incident Air Kerma

Measured Free in Air on the central beam axis at the focal spot to surface distance.

Only primary beam is considered, that is, no scatter contribution.

Unit: joule/kg or gray (Gy)

Page 34: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Entrance Surface Air Kerma (ESAK)

• ESAK measured on the surface of the patient or phantom where X-ray beam enters the patient or phantom.

• Includes a contribution from photons scattered back from deeper tissues, which is not included in free in air measurements.

Page 35: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Entrance Surface Air Kerma (ESAK)

• If measurements are made at other distances than the true focus - to - skin distance, doses must be corrected by the inverse square law and backscatter factor incorporated into the calculation.

References:– Dosimetry in Diagnostic Radiology: An International code of

practice, TRS 457, IAEA, 2007– Phys. Med. Biol. 43 (1998) 2237-2250.

Page 36: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Kerma in X-ray field can be measured using calibrated: • Ionization chamber

• Semiconductor dosimeter

• Thermoluminescent dosimeter (TLD)

Dose Measurement

36

Page 37: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Kerma-Area Product: KAP

• The kerma - area product (KAP) is defined as the kerma in air in a plane perpendicular to the incident beam axis, integrated over the area of interest.

• This is the dose related quantity measured and displayed on all modern X-ray equipment excluding CT.

KAP meter

Page 38: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Kerma-Area Product: KAP• The KAP (Gy·cm2) is constant

with distance since the cross section of the beam is a quadratic function which cancels the inverse quadratic dependence on dose .

• KAP remains constant along the beam axis as long as it is not measured close to the patient/phantom surface which introduces backscatter.

Page 39: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

KAP = K x Area

the SI unit of KAP is the Gy·cm2

Kerma-Area Product: KAP

39

Area = 1Dose = 1

Area = 4Dose = 1/4

d1=1

d2=2

Page 40: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Kerma-Area Product: KAP

40

KAP is independent of distance from the X-ray source, as:

Air Kerma decreases with the inverse square law.

Area increase with the square distance

KAP is usually measured at the level of the tube diaphragms

Area = 1Dose = 1

Area = 4Dose = 1/4

d1=1

d2=2

Page 41: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

KAPKAP (kerma(kerma--area product)area product)

UnitUnit: : GyGy··cmcm22This is a picture of a KAP meter which measures the kerma area product

Page 42: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units 42

Example of a dose display during fluoroscopy or cine

runs with dose rate as shown

Page 43: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

In paediatric radiology KAP may be used for:

– Diagnostic reference levels (DRLs)– By use of conversion factors, it can be converted to

skin dose and/or effective dose

43

Kerma-Area Product

Page 44: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Dosimetric Quantities for CT• Computed Tomography Dose

Index (CTDI) - determined using scan protocol

parameters.-useful for comparison of different scanners.

• Dose-Length Product (DLP)- measure of dose to patient- used to estimate effective dose

Page 45: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

CT and Risk

Understanding radiation units

KERMA (in phantom)

CTDI (dose in phantom per slice)

Length of scan and pitch

DLP

Effective dose

Risk

Page 46: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Measurement of Dosimetric Quantities in CT

• Pencil ionisation chamber with active length of 100 mm.

• • Measurements free-in-air or in

standard dosimetry phantom.

• Alternatives: TLD, solid state detectors.

• CTDIVOLshould be displayed on the console, reflecting the conditions of operation selected (IEC, 2003)

Page 47: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Dose Indicators in Interventional Radiology

• For quality assurance purposes• To estimate the probability of occurrence of

stochastic effects use:

Kerma-air product rate (KAP, PKA)

Understanding radiation units

Page 48: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Dose Indicators in Interventional Radiology

• For quantifying the threshold and severity of deterministic effects use:

• Maximum skin dose (MSD)• Cumulative dose (CD) to Interventional

Reference Point (IRP)

• In a complex procedure skin dose is highly variable

Understanding radiation units

Page 49: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

• In some procedures, patient skin doses approach those used in radiotherapy fractions

• Maximum skin dose (MSD) or peak skin dose is the maximum dose received by a portion of the exposed skin.

Interventional Procedures: Skin Dose

49

Radiodermatitis in the right arm. 7 year-old patient. Photograph taken 4 months after radiofrequency ablation. Surce: ICRP 85

Page 50: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Cumulative Dose to Interventional Reference Point*

• IRP is located 15 cm from the isocentre towards the focal spot

• The air kerma accumulated at a specific point in space relative to the fluoroscopic gantry (IRP) during a procedure

• Cumulative dose does not include tissue backscatter and is measured in Gy.

• Cumulative dose is sometimes referred to as cumulative air kerma

Understanding radiation units

*IRP

Page 51: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Cumulative dose to Interventional Reference Point

Cumulative dose to IRP is measured with a flat ion chamber or calculated by the system and displayed in the angiography room

Understanding radiation units

15 cm

Isocenter

IRP

15 cm

Isocenter

IRP

(IEC-60601-2-43)

Page 52: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

MSD vs. Cumulative dose

• In some procedures, cumulative dose to IRP is well correlated with MSD

• Cumulative dose to IRP can be a good indicator of doses higher than the thresholds for skin injures

• A “trigger value” for cumulative dose can be adopted to alert interventionalists the threshold for skin erythema could be reached.

• A follow-up protocol can be adopted.

Understanding radiation units

Page 53: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Other related dose parametersFluoroscopy time:

• Has a weak correlation with KAP• But, in a quality assurance programme it can be

adopted as a starting unit for– comparison between operators, centres,

procedures– for the evaluation of protocol

optimization, and– to evaluate operator skill

53

Page 54: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Other related dose parametersNumber of acquired images and number of series:

– Patient dose is a function of total acquired images

– But dose/image can have big variations– There is an evidence of large variation in

protocols adopted in different centres

54

Page 55: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Diagnostic reference levels (DRLs)

• ICRP, IAEA, EC: introduced the concept of diagnostic reference levels (DRLs) for patients

• DRLs are a form of investigation level, apply to an easily measured quantity at the surface of a simple standard phantom or a representative patient.

• An optimisation tool, not dose limits

Understanding radiation units

Page 56: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Diagnostic Reference Levels (DRLs)

• DRLs calls for local investigation (often very simple) if constantly exceeded

• DRLs: Management of patient doses must be consistent with the required clinical imaging information

Understanding radiation units

Page 57: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Quantities for Establishment of DRLs

• Incident air kerma and entrance-surface air kerma

• Incident air kerma rate and entrance-surface air kerma rate

• Air kerma–area product• CT Dose index, CT Dose–length product

Understanding radiation units

Page 58: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Quantities and Units for Staff Dose Assessment

• Personal dosimetry services typically provide monthly estimates of Hp(10) (mSv), the dose equivalent in soft tissue at 10 mm depth. This is in most of the cases used to estimate the effective dose.

• Sometimes, Hp(0.07) (mSv) is also reported: the dose equivalent in soft tissue at 0.07 mm depth)

• Personal dosememters (film, thermoluminescent...)

Understanding radiation units

Page 59: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Personal Dosimetry Methods

• Single dosimeter worn– above the apron at neck level

(recommended) or under the apron at waist level

• Two dosimeters worn (recommended in intrevational procedures)– one above the apron at neck

level – another under the lead apron

at waist level

Understanding radiation units

Lens dose, optional

Finger dose, optional

Second dosemeter

outside and above the apronat the neck, optional

Personal dose

dosemeter behind the lead apron

X-ray

tube

Image intensifier

Patient

Radiation

protectionmeasures

Dose limits of occupational exposure

(ICRP 60)

Effective dose 20 mSv in a yearaveraged over a period of 5 years

Anual equivalent dose in the lens of the eye 150 mSv

skin 500 mSvhands and feet 500 mSv

Page 60: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Dose Measurement

Dose due to scatter radiation at a point occupied by the operator can be measured with a portable ionization chamber

Understanding radiation units

Page 61: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Summary

• Dosimetric quantities are useful to know the potential hazard from radiation and to determine radiation protection measures to be taken

• Physical quantities - Directly measurable• Protection quantities - Defined for dose limitation

purposes, but not directly measurable.• Application specific quantities - Measurable in medical

imaging. • Diagnostic Refernce Levels

Page 62: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Answer True or False

1. The same amount of radiation falling on the person at level of breast, head or gonads will have same biological effects.

2. Effective dose can be easily measured. 3. Diagnostic reference levels are not applicable to

paediatric radiology.

62

Page 63: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Radiation Protection in Paediatric Radiology L02. Understanding radiation units 6363

Answer True or False

1. False -Different organs have different radio-sensitivity and tissue weighting factors as given by ICRP.

2. False -It can be only calculated using different methods.

3. False - DRLs apply for paediatric radiology, but these are age-specific.

Understanding radiation units

Page 64: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

References

• INTERNATIONAL COMMISSION ON RADIATION UNITS AND MEASUREMENTS, Patient Dosimetry for X Rays Used in Medical Imaging, ICRU, Rep. 74, ICRU, Bethesda, MD (2006).

• INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION, Radiological Protection in Medicine, Publication 105, Elsevier, Oxford (2008)

• INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION, Recommendations of the ICRP, Publication 103, Elsevier, Oxford (2008)

• EUROPEAN COMMISSION, Guidance on Diagnostic Reference Levels (DRLs) for Medical Exposure, Radiation Protection 109, Office for Official Publications of the European Communities, Luxembourg (1999)

• INTERNATIONAL ATOMIC ENERGY AGENCY, Dosimetry in Diagnostic Radiology: an International Code of Practice, Technical Report Series No 457, IAEA, Vienna (2007)

Page 65: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Additional information

Page 66: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Quantities for radiation measurement

• Physical quantities - Directly measurable

• Protection quantities - Defined for dose limitation purposes, but not directly measurable

• Application specific quantities - Measurable in medical imaging

Page 67: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Radiation quantities and units• Fundamental dosimetric quantities• Protection quantities

– Equivalent dose– Effective dose

• Application specific dosimetric quantities used in DR– Incident air kerma– Entrance surface air kerma– Air kerma area product– Air kerma length product– Dosimetric quantities in CT and mammography

Page 68: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Physical Quantities

Page 69: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Physical quantities

• Fluence

• Exposure

• Kerma

• Absorbed dose

Page 70: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Fluence : f

The fluence, f , is the quotient of dN by da, where dN is the number of particles incident on a sphere of cross section da, thus

f = dN/da

The unit of fluence is m-2

Page 71: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Exposure: X

where dQ is the absolute value of the total charge of ions produced in air when all the electrons liberated in air of mass dm are completely stopped in air.

The SI unit of exposure is the coulomb per kilogram (C/kg)

The special unit of exposure is the röntgen (R).1R = 2.58 x 10-4 C kg-1

dmdQX

Page 72: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

KERMAThe KERMA (Kinetic Energy Released in a MAterial)

where dEtrans is the sum of the initial kinetic

energies of all charged ionizing particles liberatedby uncharged ionizing particles in a material ofmass dm The SI unit of kerma is the joule per kilogram (J/kg),termed gray (Gy).

.

dmdEK trans

Page 73: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Exposure and KERMA

Exposure, X, in units of C kg-1, is related to air kerma as follows:

where W is the average energy spent by an electron to produce an ion pair, g is the fraction of secondary charged particles that is lost to bremsstrahlung radiation production and e is the electronic charge

W

egKX a

1

Page 74: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

The fundamental dosimetric quantity absorbed dose, D, is defined as:

where is the mean energy imparted by ionizing radiation to matter in a volume element and dm is the mass of matter in the volume element.

The SI unit of absorbed dose is the joule per kilogram (J/kg), termed the gray (Gy)

In diagnostic radiology, KERMA and D are equal

Absorbed Dose: D

dmdD

d

Page 75: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Exposure and Absorbed Dose or KERMA

• Exposure can be linked to air dose or kerma by suitable conversion coefficients.

• For example, 100 kV X-rays that produce an exposure of 1 R at a point will also give an air kerma of about 8.7 mGy and a tissue kerma of about 9.5 mGy at that point.

Page 76: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Application Specific Quantities

Page 77: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Imaging modality

Measurement subject

Measured radiation quantity

Remark

Radiography PhantomPatient

Incident air kermaESAK, KAP Calculated from X-ray

tube outputFluoroscopy/ Interventional procedures

PhantomPatient

ESAKKAP/Peak skin dose

CT PhantomPatient

CT air kerma indexCT air kerma- length product

Measured in PMMA head and body phantom

Mammogra-phy

PhantomPatient

Incident air kerma, ESAKIncident air kerma

Calculation of mean glandular dose

Dental radiography

Patient Incident air kermaAir kerma-length product

Page 78: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Application Specific Quantities

Patient thickness

Focal-spot to image receptor distance (FFD)

Focal-spot to patient skin distance (FSD)

Incident air kerma (no backscatter)

Entrance surface air kerma (including backscatter)

X-ray tubefocal spot position

Image receptor

Schematic diagram showing some dosimetric and geometric quantities

Page 79: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Entrance Surface Air Kerma (ESAK)

where Y(kVp, FFD) is tube output for actual kVp used during examination, mAs is actual tube current-time product used during examination and FFD is focus-to-film distance. BSF is the backscatter factor that depends on kVp and total filtration of X-rays

Understanding radiation units

BSFtFFD

FDDmAsFDDkVpYESAKp

2

),(

Page 80: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Backscatter Factors (Water)

HVL Field size (cm x cm)

mmAl 10 x 10 15 x 15 20 x 20 25 x 25 30 x 30

2.0 1.26 1.28 1.29 1.30 1.30

2.5 1.28 1.31 1.32 1.33 1.34

3.0 1.30 1.33 1.35 1.36 1.37

4.0 1.32 1.37 1.39 1.40 1.41

Page 81: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

• If the KAP is calculated by the system, you must know if the user added filtration you use is included or not !

81

Kerma-Area Product: KAP

Page 82: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Kerma-Area Product: KAP

• It is always necessary to calibrate and to check the transmission chamber for the X-ray installation in use

• In some European countries, it is compulsory that new equipment is equipped with an integrated ionization transmission chamber or with automatic calculation methods

Page 83: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Dosimetric Quantities for CT

• Computed Tomography Dose Index (CTDI)

• CT air kerma index

• Dose-Length Product (DLP)• Air kerma-length product

Page 84: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

ICRU 74 / IAEA TRS 457

• CT air kerma index– Free-in-air (Ck)– In phantom (Ck,PMMA)

• Air kerma length product (PKA)

Page 85: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Dosimetric Quantities for CTPrincipal dosimetric quantity in CT is CT air kerma index:

where K(z) is air kerma along a line parallel to the axis of rotation of the scanner over a length of 100 mm.

N = Number of detectors in multi-slice CT T = Individual detector dimension along z-dimension

The product NT defines the nominal scan beam width/collimation for a given protocol.

50

50100, )(1 dzzK

NTCa

Page 86: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Dosimetric Quantities for CT

Weighted CT air kerma index, CW, combines values of CPMMA,100 measured at the centre and periphery of a standard CT dosimetry phantoms

pPMMAcPMMAw CCC ,100,,100, 231

Page 87: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Dosimetric Quantities for CT

Pitch (IEC, 2003):

T= Single detector dimension along z-axis in mm.

N=Number of detectors used in a given scan protocol (N>1 for MDCT), N x T is total detector acquisition width or collimation

I=table travel per rotation

Radiographic, 2002, 22:949-62

NTIp

Page 88: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

• Volume CTDI describes the average dose over the total volume scanned in sequential or helical sequence, taking into account gaps and overlaps of dose profiles (IEC, 2003):

• Average dose over x, y and z direction• Protocol-specific information

Dosimetric Quantities for CT

lNTCC WVOL

Page 89: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Dosimetric Quantities for CT

• Kerma-length product (PKL):

where L is scan length is limited by outer margins of the exposed scan range (irrespective to pitch)

• PKL for different sequences are additive if refer to the same type of phantom (head/body)

LCP VOLKL

Page 90: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Maximum Skin Dose (MSD)

• Measurement/evaluation of MSD• Point or area detectors • Cumulative dose at IRP (interventional radiology point)• Calculation from technical data

• Off line methods• Area detectors: TLD array, slow films, radiochromic

films• From KAP and Cumulative dose measurement

Understanding radiation units

Page 91: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

Method for MSD Evaluation: Radiochromic Large Area Detector

Example: Radiochromic films type Gafchromic XR R 14”x17”• useful dose range: 0.1-15 Gy • minimal photon energy dependence (60 - 120 keV)• acquisition with a flatbed scanner:b/w image, 12-16

bit/pixel or, measure of OD measurement with a reflection densitometer

Page 92: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Benefits of Radiochromic Films• The radiochromic film:

– displays the maximum dose and its location– shows how the total dose is distributed– provides a quantitative record for patient files– provides physician with guidance to enable safe planning of

future fluoroscopically guided procedures– improves fluoroscopic technique and patient safety– possible rapid semi-quantitative evaluation

Understanding radiation units

Example of an exposed radiochromic film in a cardiac interventional procedure

Page 93: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Rapid Semi-Quantitative Evaluation: Example

• For each batch number (lot #) of gafchromic film a Comparison Tablet is provided

• In the reported example we easily can recognise that the darkness area of the film, corresponding to the skin area that has received the maximum local dose, has an Optical Density that correspond at about 4 Gy

Understanding radiation units

Page 94: Biophysics Understanding Radiation Units Dr. Yousif Mohamed Y. Abdallah

Understanding radiation units

DRLs for Complex Procedures

3rd level “Patient risk”

2nd level “Clinical protocol”

1st level“Equipment

performance”

94

Dose rate and dose/image(BSS, CDRH, AAPM)

Level 1 + No. images + fluoroscopy

time

Level 2 + DAP + Peak Skin Dose (MSD)

Reference levels (indicative of the state of the practice): to help operators to conduct optimized procedures with reference to patient exposure

For complex procedures reference levels should include: • more parameters • and, must take into account the complexity of the procedures.(European Dimond Consortium recommendations)