BLOOD IRRADIATOR 2000 (BI 2000) INSTRUCTION MANUAL FOR OPERATION BOARD OF RADIATION AND ISOTOPE TECHNOLOGY (BRIT) DEPARTMENT OF ATOMIC ENERGY GOVERNMENT OF INDIA TURBHE, SECTOR – 20, VASHI, NAVI MUMBAI MAHARASTRA, PIN – 400 705, INDIA TEL Nos. 00-91-22-2788 7304, 00-91-22- 27887311 FAX No. 00-91-22-2784 0022

BI 2000 for Touch Screen Messung

Embed Size (px)

Citation preview

Page 1: BI 2000 for Touch Screen Messung

BLOOD IRRADIATOR 2000(BI 2000)

INSTRUCTION MANUALFOR

OPERATION

BOARD OF RADIATION AND ISOTOPE TECHNOLOGY(BRIT)DEPARTMENT OF ATOMIC ENERGYGOVERNMENT OF INDIATURBHE, SECTOR – 20,VASHI, NAVI MUMBAIMAHARASTRA,PIN – 400 705,INDIATEL Nos. 00-91-22-2788 7304, 00-91-22-27887311FAX No. 00-91-22-2784 0022

Page 2: BI 2000 for Touch Screen Messung

CONTENTSSr.No TITLE PAGE No.

1. INTRODUCTION 3

2. SPECIFICATION OF THE UNIT: 3

3. MAIN FEATURES OF BI 2000 4

4. DESCRIPTION OF UNIT: 5

5. BRIEF DESCRIPTION OF THE PARTS 5

6. OPERATORS QUALIFICATIONS 7

7. INSTRUCTIONS FOR OPERATION 7

8. EVERYDAY INSPECTION OF THE UNIT 8

9. ROUTINE MAINTENANCE 9

10. GENERAL RECOMMENDATIONS 9

11. PROCEDURE FOR LIMIT SWITCH ADJUSTMENT AND REPLACEMENT

10

12. DECAY TABLE 11

13. SKETCH FOR THERMOCOUPLE SENSOR 12

14. CONTROL SYSTEM 13

15. OPERATION MODES 14

16. MANUAL MODE 15

17. MANUAL TIMER MODE 15

18. AUTO MODE 16

19. PROXIMITY SENSOR 18

20. PAUSE AND CONTINUE 19

21. DATA OF PREVIOUS SAMPLES 20

22. PLC I/O STATUS 22

23. REAL TIME CLOCK 23

24. DOSE RATE PROFILE 23

25. SET DOSE RATE 25

26. SET TIMER 26

2

Page 3: BI 2000 for Touch Screen Messung

27. SET OTHER PARAMETERS 27

28. OPERATIONS SEQUENCE 29

29. ANNEXURE – A (CONTROL CIRCUIT DIAGRAMS)

32

30. ANNEXURE - B (RADIOMETRY TEST RESULTS)

33

31. ANNEXURE - C (DOSIMETRY REPORT)

34

32. ANNEXURE – D (COMMON CHECK LIST FOR THE PACKAGE)

35

33. INSTALLATION CERTIFICATE 36

BLOOD IRRADIATOR 2000

INTRODUCTIONBlood Irradiator 2000 is a compact, portable, self, - shielded type of a Co-60 Gamma Irradiator. Its design conforms to American National Standards ANSI- N 433.1, 1977 forSafe design and use of self contained Dry Source Storage Gamma Irradiator (category - 1). Atomic Energy Regulatory Board (AERB), India, approves it as Type B (U) package for safety in transport of radioactive material. It also meets the IAEA regulations for the safe transport of radioactive material as per safety series - 6,1985(as amended in 1990). The Unit is designed to house 30 TBq (810 Ci) of Cobalt-60 source and it provides An Irradiation volume of about 2000 c.c. approximately. The doubly Encapsulated sealed radioactive sources are used in cylindrical form, which are completely contained in a dry container called flask unit. The sealed sources are shielded at all times and human access to it and the volume undergoing irradiation is not physically possible in its designed configuration. The irradiation chamber is located in a vertical shaft (Drawer). The shaft moves up and down with the help of a drive system, which enables exact positioning of the irradiation chamber in the center of the radiation field. Access holes of 8 mm diameter are provided in the vertical shaft for introduction of service sleeves for gases, RTE) censer etc. for temperature measurement inside irradiation zone.

Specification of the Unit:

Maximum Co-60 source capacity : 30 TBq (810 Ci)

Dose rate at maximum capacity : 11 Gy/Min (in air)

9.5 Gy/Min (in water)

3

Page 4: BI 2000 for Touch Screen Messung

Dose rate uniformity : Radial + 25% or better

And axial -25% or better

Irradiation volume : Sample Chamber -3000 liters

Rotating Chamber -2000 liters

Size of sample chamber : Sample Chamber-17.2 cm (dia) x 20.5cm(h)

Shielding material : Lead & stainless steel

Weight of the unit : ~2600 kg approx.

Size of the unit : ~112cm (L) x 95 cm(W) x270 cm (H)

Power requirement : 220/230 V, 50 Hz, 15 Amps, Single phase,

A.C supply.

Timer range : 6 sec onward

Minimum Irradiation Time: 6 Seconds

MAIN FEATURES OF BI 2000

1. Low external leakage radiation level typically 5 mr/hr (50 micro Sv/hr) or less at the surface with maximum source loading.

2. Comfortable Sample loading height a platform of convenient height has been provided for easy loading/unloading of products.

3. Mechanism of rotating/stirring samples at about 60 rpm inside the sample chamber during irradiation to give better uniformity of dose.

4. Dose rate at maximum capacity of source loading is 11 Gy/Min. approximately in air at center of irradiation volume and 9.5 Gy/Min in water approx. at the center of irradiation volume and it gives Dose rate uniformity is +25% or better radically and -25% or better axially.

5. Control system the operation of the unit is controlled by a PLC based control system consisting of main control unit having a provision of automatic termination of the irradiation time from 6 seconds onwards. 220/230 volts operate it, 50 Hz, 15 Amperes, single phase of A.C. electrical power supply. The control system is PLC base with user-friendly program having touch sensitive screen and printer for data recording.

6. Manual operation In the event of power failure, the unit can be operated

4

Page 5: BI 2000 for Touch Screen Messung

mechanically by using a hand crank.

7. Temperature control Service sleeves in the drawer are provided for manual controlling of temperature in irradiation zone inside the sample chamber by passing the cool or hot air/gas through it. Thermocouple sensor within 0 to 200 degrees Celsius senses irradiation temperature and the same is displayed in the main control panel.

8. Radioactive decay correction. Radioactive decay correction in the dose rate on day-to-day basis is automatically done and displayed in the Auto mode of operation.

9. Sensing of Sample chamber Removable half. Sensing have closed or opened condition of sample chamber removable half (door) is done by a proximity switch. This ensures further safety in operation of the unit by not allowing the Drawer to go down when sample chamber door is open.

DESCRIPTION OF UNIT:

Blood Irradiator 2000 unit is designed to house up to 30 TBq (810 Ci) Cobalt-60 source and the lead shielding provided is adequate enough to bring down the radiation leakage level on the exterior of the unit well below the accepted standards. It offers an irradiation volume of about 2000CC. The unit mainly consists of the following parts

a. Source and cylindrical source cageb. Lead flask as Biological shieldc. Central drawer with sample chamberd. Mechanical drive systeme. Electrical/Solid state control systemf. External cabinet

BRIEF DESCRIPTION OF THE PARTS

a) Source and cylindrical source cage:

The source cage holds the radiation source pencils vertically and symmetrically along its cylindrical periphery. The cage can hold a maximum of 24 pencils and is designed to provide better uniformity of dose rate in the sample chamber. Each pencil contains cobalt-60 source in the form of pellet of 1 mm (dia.) x l mm (h) or 6 mm (dia) x 25 mm (h) slug and is doubly encapsulated in stainless steel capsules (pencils). The sources are

5

Page 6: BI 2000 for Touch Screen Messung

further tested for any leak and made free from any radioactive contaminations.

b) Lead flask (biological shield):

This is a lead filled stainless steel container around the source and consists of two parts, (i) main outer body and (ii) inner removable shielding plug (lead plug). The radiation source is housed in the main outer body which is designed to reduce the radiation leakage level on the outer surface well bellow the permissible level for maximum source loading capacity of 30 TBq (810 Ci) of cobalt-60 source. A minimum of 251 mm lead as shielding encased in stainless steel shells has been provided in the lead flask.

Lead plug is a cylindrical plug constructed of lead encased in stainless steel shells, which forms an integral part of flask. This plug is provided to close the top cavity of the flask through which the source cage with source are inserted or withdrawn from the flask at the time of source loading or unloading. This operation is done inside a hot cell with remote handling equipment. The lead plug is firmly bolted down with the main body of the flask with 8 nos. of M-16 bolts, which will ensure that the lead plug remains firmly fixed to the flask during all conditions of transportation. At no time, lead plug should be attempted to be removed without the use of a hot cell under an expertise guidance to avoid any situation of radiation hazards.

c). Central drawer assembly:

The central drawer consists of three parts namely (I) Top drawer, (ii) sample chamber and (iii). Bottom drawer. These parts are aligned, assembled, and secured firmly with the Allen screws. The central Drawer runs vertically up and down through the cavity of flask unit with the help of a drive system to bring the sample chamber either in irradiation position or out-side the unit in sample loading position. The precise positioning of Drawer in irradiation position or sample loading position is done with the help of a control system.

The sample chamber is a hollow thin walled stainless steel cylinder in two halves, the removable half (door) is held in position by two arrestors. The internal size of sample chamber is about 13 cm (dia) x 21.7 cm (h) providing an irradiation volume of 3000CC approx. and that the rotating chamber is 11.2 cm (dia) x 17.9 cm (h) providing an irradiation volume of 2000 CC approx.

CAUTION

THE ASSEMBLY OF DRAWER AND THE INNER REMOVABLE SHIELDING PLUG OF FLASK MUST NEVER BE REMOVED EXCEPT BY REMOTE MEANS IN A SHIELDED ROOM (HOT CELL) UNDER EXPERTISE GUIDANCE.

6

Page 7: BI 2000 for Touch Screen Messung

d) Mechanical drive system:

The drawer is raised or lowered by a pair of stainless steel wire ropes, which run through a series of pulleys mounted on the bass frame of the unit. One end of each wire rope is anchored to the wire rope winding drum, while the other end is attached to a turnbuckle. An electric motor is coupled to a reduction gear through a V- belt drive and also an electro magnetic clutch, which rotates the rope drum at a fixed rpm. The precise locations of the sample chamber to "irradiation" or "sample loading" position are facilitated with the help of two limit switches mounted on the top of the flask and its associated control system. In case of power failure, Drawer with sample chamber can be raised or lowered mechanically with the use of a hand crank.

e) Electrical/Solid state control system:

The unit is designed to operate on a 220/230 V, 10A, and 50 Hz single phases, AC power supply.

The control system mainly consists of the following:1. Main Control panel assembly2. Contractor Panel3. Electrical motor4. Limit switches etc.

The movement of the drawer is controlled from the main control panel fixed on the front side of the unit. The features of the control system have been described in complete details with circuit diagram separately in this instruction manual for the operation of the unit, which should be strictly followed for the smooth and trouble free functioning of the unit.

f) External cabinet:

The cabinet is a housing for the flask assembly of the unit and it is made out of carbon steel sheets. The right and back side doors of the cabinet provide an access to the mechanical drive system as well as contactor box and other wirings. Removal of top panel gives an access to the limit switches and proximity switch. The control panel is mounted at the front face of the cabinet.

OPERATORS QUALIFICATIONS

The unit is simple to operate; no special training is required to operate it. However, for proper operation and preventive maintenance of the unit, it is desirable to have persons from mechanical and electrical engineering background with some knowledge of radiation protection as well Normally two persons of the users department are associated with the installation of the unit who would be trained by BRIT installation team regarding the operation and maintenance aspects of the unit.

INSTRUCTIONS FOR OPERATION:

7

Page 8: BI 2000 for Touch Screen Messung

The Gamma chamber is simple to operate. The following sequence of operation is recommended:

1. Connect the two three pin plugs of the main power supply and control panel of the unit to 230V, 50Hz single phase AC Supply with proper grounding connection and switch on them. A study power supply of about 220 Volt must be made available for the proper functioning of the unit

2. When the sample chamber is up, remove the sample chamber removable half (door) by turning the top and the bottom arrestors.

3. Place the product to be irradiated in the sample chamber. Any flexible tubing or electrical leads if required could be introduced through the top access tube into the sample chamber. Replace the removable half (door) of sample chamber to its position, and lock it by turning both the arrestors before power is switched on. Ensure that the hand crank has been detached from the unit

4. Switch on the power by actuating Mains key and computer key on the control panel, this will initialize the unit in ''MANUAL MODE".

5. MANUAL MODE: Press ‘LOWER’ switch. The Drawer will start moving downwards and will stop in irradiation zone. At the end of the desired time, the 'RAISE' switch should be pressed to bring the sample chamber to unloading position. The product now can be removed from the sample chamber by opening the door.

6. SOLID STATE TIMER IN MANUAL MODE: In the manual mode, there is an option to use a Solid State Timer for Irradiation ranging from 6 seconds to 99hrs. This timer is located in the control cabinet rack, adjacent to contactor panel. This timer operation is available only in the manual mode. If the switch (located adjacent to the timer in the control cabinet rack) in the “OFF” position, then the sample chamber can be raised and lowered only manually as indicated above in b & c. If the timer switch is in “ON” position then the sample chamber has to be lowered manually and the timer will start functioning as soon as the sample chamber reaches the Irradiation zone. At the end of the set time, the sample chamber will come up automatically.

7. AUTO MODE: Switch 'ON' the mains key as above and wait for few seconds till it displays the MENU Screen. Then press the operation mode on the touch screen display. This will set the unit in the auto mode. For the details of operation refer the instructions given on control system of BI 2000

CAUTION

8

Page 9: BI 2000 for Touch Screen Messung

ALWAYS ENSURE THAT THE HAND CRANK IS DETACHED FROM THE UNIT BEFORE SWITCHING THE POWER ON.

EVERYDAY INSPECTION OF THE UNIT BY THE OPERATOR

a) Voltage : It should be around 220/230 ±10v if the voltage is less than 210 it may not operate in that case. Use hand crank only for drawer movement after disconnecting the unit electrically.

b) Wire Rope Winding: Before every operation for taking "UP" see the wire rope winding on the drum. If it is O.K. then only operates otherwise first set it O.K.

c) Arrestor Screws Of Sample Chamber: 2 numbers of arrestor screws are keeping the arrestor pieces (locking pieces) of sample chamber in their position and also allow them to rotate to facilitate removing the removable half of the sample chamber it should be seen that the screws head do not come out or project out of the drawer surface. In case it is projecting out, arrestor screws should be screwed in by a screwdriver ensuring that locking pieces still rotates smoothly.

d) For Other Major/Unseen Problems: Contact Electrical / Electronics / Mechanical persons.

ROUTINE MAINTENANCE

a. MOTOR: Grease the motor bearings once in six months.

b. REDUCTION GEAR: Check the level of oil in the gearbox every six months by opening the oil level plug. If there is any reduction in oil level, make it up with SAE 90 or any other equivalent good quality gear oil.

c. V-BELT DRIVES: Check the V-belt drive periodically. The belt may get slackened due to wear. If tightening of the belt is required loosen the four corresponding mounting bolts and move to suit the tension.

d. LIMIT SWITCHES: There are two limit switches. These are of roller head actuator type, If any of these become defective replace by a similar one taking care to connect as per circuit diagram as well as to provide a normal gap (approx 1 mm) between the limit switch roller (press condition) and Drawer.

e. WIRE ROPES: Grease the wire ropes once in six months. Replace with the correct size and type when necessary taking care of the direction of rope wining on the drum. The tension of the both the wire ropes should be always same and to be checked periodically every three months. Adjustment of tension can be done through the turnbuckles attached to the wire rope system.

9

Page 10: BI 2000 for Touch Screen Messung

GENERAL RECOMMENDATIONS:

1. Whenever the unit is not in use, allow the drawer Sample-loading position.

2. For very long manual exposure, power should be switched off after taking the drawer to irradiation position. This will eliminate the possibility of someone disturbing the sample chamber during irradiation.

3. The various switch keys and the band crank should be kept under the safe custody of the operator to avoid unauthorized operation of the Unit.

4. Prior to operating the unit the supply voltage should be checked, it should be around 220V. In the event of voltage, dropping to less than 210V the unit may not function properly. Where power fluctuations are generally occurring a suitable power stabilizer (3 KVA, 1 Phase, 230Volts) to be used.

5. In case of any emergency or for any advice regarding the unit, please contact

General Manager, Engg, BRITBARC-Vashi complex,Thurbhe, Navi Mumbai 400 705.Phone No. 00-91-22-27887300, 00-91-22-27887311Fax No. 00-91-22-2784 00 22

OR General Manager,

CSSC, BRITBARC-Vashi complex,Thurbhe, Navi Mumbai 400 705.Phone No. 00-91-22-27887204, 00-91-22-27887311Fax No. 00-91-22-2784 00 22

PROCEDURE FOR LIMIT SWITCH ADJUSTMENT AND REPLACEMENT

Top and bottom limit switches are mounted on the top of lead Flask on the backside. These two limit switches are sensing the extreme top or bottom position of the drawer. For the proper functioning of the limit switches, it should be firmly fixed to its bracket as well as it should have a clearance of about 1-2 mm. between its roller and the drawer Surface. Therefore its resetting is required if the clearance has Changed over a period of time (say 6 months or so). The following steps are needed to adjust it or replace it.

STEPS1. First mark the present location of the bottom plate of the bracket by a pen /

marker.

10

Page 11: BI 2000 for Touch Screen Messung

2. Remove the bolts of the bottom bracket and take out the limit switch bracket.

3. Now move the bottom bracket of the limit switch a little backward or forward so that clearance of 1-2 mm. is obtained and then fixed the bolts as before.

4. For changing a new limit switch, remove the old one after Noting down its wiring connection.

5. Then fix the new one and reconnect the electrical wiring, as before help of the circuit diagram should also be taken while Disconnecting and reconnecting the wiring.

(Decay Table—Cobalt-60)

RADIATIONS : HALF LIFE - 5.2 YEARS

BETA - 0.31 MeV GAMMA — 1.17 MeV, 1 .33 MeV 1 CURIE —1.3 rhm*

11

Page 12: BI 2000 for Touch Screen Messung

*rhm = roentgen per hour at one meter

YEAR MONTHS

12

Page 13: BI 2000 for Touch Screen Messung

CONTROL SYSTEM

DESCRIPTION The control system is built around a state of the art programmable logic controller consisting of a central processing unit, in built 8 digital inputs and 8 digital outputs, a touch screen for input the data, a power supply module. The system is cable of operation under MANUAL as well as AUTO modes. Keys provided on touch screen display can do the Selection of the Auto and Manual mode. When the unit is switched on the unit is activated in the manual mode as default mode. Also, whenever the PLC is not functional the system can be operated in the manual mode. Solid State Timer In Manual Mode operating in range 6 seconds to 99 Hr. has also been provided as addition feature. Its setting can be done in seconds, minutes or hours. It can be used with the control system in the MANUAL mode with a toggle switch. This timer is providing inside the control system cabinet, which can be accessed through the side door.

CONTROL UNIT

OPERATION AND PROGRAMMING

The rightmost key switch on the control panel can switch on the unit. When switched on BRIT logo is displayed on the screen.

13

BOARD OF RADIATION & ISOTOPE TECHNOLOGY

BLOOD IRRADIATOR 2000

V 1 . 2

Page 14: BI 2000 for Touch Screen Messung

This screen displays the unit name (BI 2000) at the top along with a calendar and real time clock. The status position of the sample chamber as UP/DOWN and the rotation chamber status as ON/OFF is also displayed.The user can invoke any of the available screens by choosing from the option. For activating any of the screens, he has to touch the desired region.

Operation modes key when touched, another screen appears as shown, which asks the user to select the mode of operation as MANUAL or AUTO.

14

BI 2000 24/11/2004 16:19:49

Chamber UP Rotation OFF

Operation Modes

Set Dose Rate

Set Timer Preset Values

Set Other Parameters

Data of Previous Samples

PLC I/C Status

Real-Time Clock

Dose Rate Profiles

Chamber UP Rotation OFF

Select mode of operation

EFFECTIVE DOSE RATE : 10.4 Gy/Min.CHAMBER TEMPERATURE : 45 oC

MENU

MANUAL AUTO

LOG IN

LOG OUT

BI 2000 24/11/2004 16:19:49

Page 15: BI 2000 for Touch Screen Messung

If MANUAL is selected a screen appears as: There are two options in manual mode.

MANUAL MODE (WITHOUT TIMER) After positioning the material in the sample chamber, press the LOWER push button to bring the sample in the radiation field. When the required time for irradiation is over, press the RAISE push button.

MANUAL MODE (WITH TIMER.)The required TIME has to be set using the TIMER, which is situated inside the cabinet of the unit. In this, the time can be set from zero seconds to 300 hours. After setting the time the TIMER, toggle switch has to be turned ON. Then the screen appears as shown:

Then, press the LOWER push button on the control panel. When the set Irradiation Time is over the sample chamber will be raised automatically.

AUTO MODE

15

BI 2000 24/11/2004 16:19:49

Chamber UP Rotation OFF

SYSTEM IS CURRENTLY IN MANUAL MODEMANUAL TIMER IS TIMER OFF

EFFECTIVE DOSE RATE : 10.4 Gy/Min.CHAMBER TEMPERATURE : 45 oCIRRADIATION TM : 00:00:00EFFECTIVE DOSE : 00 GyATTENUATION : 0 %

MENU

BI 2000 24/11/2004 16:19:49

Chamber UP Rotation OFF

RAISE LOWER ATTN

SYSTEM IS CURRENTLY IN MANUAL MODEMANUAL TIMER IS TIMER ON

EFFECTIVE DOSE RATE : 10.4 Gy/Min.CHAMBER TEMPERATURE : 35 oCIRRADIATION TM : 00:00:00EFFECTIVE DOSE : 00 GyATTENUATION : 0 %

MENURAISE LOWER ATTN

Page 16: BI 2000 for Touch Screen Messung

Press AUTO key to select the AUTO mode screen. Further, in auto mode dose and time modes are possible. Select Dose mode. This frame indicates that EFFECTIVE DOSE RATE this is the current dose rate in Gy/Min at the center of the sample chamber. The value of dose rate used for calculating the irradiation time by the system. SELECTED DOSE This is the Dose to be delivered to the material. After pressing against the item and using the alphabetical/numerical pad, user can enter required dose in Gy. ATTENUATION The dose rate can be changed in steps of 25% with the help of the attenuator. IRRADIATION TIME this is the time required delivering the selected dose to the material. Irradiation time is calculated by PCL automatically after entering dose.

Press key to go to next screen, a next screen appears as follow

16

1

4

7

ENTER -

0

CLR

3

5

9

2

6

8

BI 2000 24/11/2004 16:19:49

Chamber UP

System is Currently in Auto ModeSample No. 00000001

DELAY TIME 00:00:00

Eff. Dose Rate : 10.4 Gy/Min

Selected Dose : 10.4 Gy

Irradiation TM : 01:00:00

DOSE MODE

TIME MODE

MENUATTN DELAYTIME

Page 17: BI 2000 for Touch Screen Messung

If required user can enter sample data, by pressing a next screen appears. Pressing against the item and using the alphabetical/numerical pad, which then appears, can enter the Sample data. (Shown below)

After typing, the data press the ENT key and the CLOSE key. The data entered has to be confirmed by pressing the Confirm key. When confirmed, previous screen appears this screen reflects the dose and the irradiation time selected by the user. If this acceptable to

17

BI 2000 24/11/2004 16:19:49

Chamber Down Rotation OFF

System is Currently in Auto ModeSample No. 00000001

Eff. Dose Rate : Gy/Min

Selected Dose : Gy

Attenuation :

Irradiation Time :

10.4

3.78

0 %

00:23:17

Start MENU

BI 2000 29/11/2000 16:19:49

Chamber Down Rotation OFF

Customer Data For Sample No.

Name

Address

0000000001

Board Of Radiation & Isotope Technology

/BRIT

X Y Z

Ref BI - 0001

Desc. I/2000/1

Confirm MENU

Page 18: BI 2000 for Touch Screen Messung

user, user can press START function key to begin the irradiation. Then the chamber will descend automatically to take its position in the irradiation zone as the chamber reaches the irradiation zone the irradiation begins. The following screen is displayed.

PROXIMITY SENSOR A proximity sensor has been providing for preventing movement of the sample

chamber to the irradiation zone with the sample chamber door removed. In such a case, if the “LOWER push Button is depressed by mistake, the drawer containing the sample chamber after going down a few centimeters will be brought to a halt immediately by the control system which receives a signal from the proximity sensor indicating the sample chamber is uncover or the door is not closed. The sample chamber can then be brought up by the depressing of the “RAISE” push button, the door placed in position and the irradiation carried out as mentioned above

ABORT The key will abort the ongoing irradiation process, by automatically raising the

sample chamber from irradiation zone to raise position.

This can be used to ABORT the processing

18

System Working in AUTO mode

Eff.Dose Rate : 10.4 Gy/MinSelected Dose : 3.78 GyAttenuation : 0 %Irradiation Time : 00:23:17Balance Time : 00:00:25

Chamber Temp: 30.0 ºC

BI 2000 24/11/2004 16:19:49

Chamber Down Rotation OFF

PAUSE ABORT LOG IN

ABORT

System Working in AUTO mode

Eff.Dose Rate : 10.4 Gy/MinSelected Dose : 3.78 GyAttenuation : 0 %Irradiation Time : 00:23:17Balance Time : 00:00:25

Chamber Temp: 30.0 ºC

BI 2000 24/11/2004 16:19:49

Chamber Down Rotation OFF

PAUSE ABORTLOG IN

Page 19: BI 2000 for Touch Screen Messung

PAUSE AND CONTINUE

During irradiation if the user wants either to examine/modify the sample, he can do so by using he pause key, pressing pause key will interrupt the ongoing irradiation process, by automatically raising the sample chamber from irradiation zone to raise position. The accompanying screen indicates that the irradiation process is paused. The balanced time is frozen at the time the sample chamber started its upward motion. After examining the sample user can continue the irradiation process from where it is paused. When the balance time equals 00:00:00, the sample chamber is raised and brought to RAISE position.

This can be used to PAUSE the processing.

This can be used to CONTINUE the processing

19

IRRADIATION PAUSED

Eff.Dose Rate : 10.4 Gy/MinSelected Dose : 3.78 GyAttenuation : 0 %Irradiation Time : 00:23:17Balance Time : 00:00:45

PAUSE

CONT.

PRINTER ENABLE

Page 20: BI 2000 for Touch Screen Messung

When the balance time becomes 00:00:00, the irradiation process ends and the sample chamber is raised automatically from the irradiation zone to the UP position. The accompanying screen indicates the termination of the irradiation process indicating the upward movement of the sample chamber. As the sample chamber comes to rest in the UP position.

PREVIOUS SAMPLES DATA (To be activated from MENU screen) When previous Sample data key on the MENU screen is pressed a screen appears as which can show record of 50 nos. of previous sample data.

20

BI 2000 24/11/2004 16:19:49

Chamber UP Rotation OFF

Sample Offset : +1 Sample Number : 000000001Eff. Dose Rate : 0.00 Gy/MinSelected Dose : 0.00 GyAttenuation : 0 %Irradiation Time : 00:00:00 Irradiation over Time : 00:00:00

System Working in AUTO mode

Eff.Dose Rate : 10.4 Gy/MinSelected Dose : 3.78 GyAttenuation : 0 %Irradiation Time : 00:23:17Balance Time : 00:00:00

BI 2000 24/11/2004 16:19:49

Chamber IN TRANSIT RotationOFF

GC5000 29/11/2000 16:19:49

Chamber IN TRANSIT RotationOFF

CONT.

PAUSE ABORTLOG IN

PRINTER ENABLE

Page 21: BI 2000 for Touch Screen Messung

Press key for viewing the first sample data

Press key for viewing data in steps of 10

Press key and for sample data in steps of one

Backward or forward.

PLC I/O STATUS (To be activated from MENU screen) When this key on the MENU screen is invoked, a screen appears as shown below:

21

-10 MENU

BI 2000 24/11/2004 16:19:49

-10

Page 22: BI 2000 for Touch Screen Messung

By invoking D I the Digital Inputs status can be viewed as shown below:

By invoking D O , the Dgital Outputs can be viewed as shown:

By invoking T C , the Temperature Controller status can be viewed as shown:

22

Chamber UP Rotation OFF

oo0D

CHANGE PLC MEMORY BACK UPBATTERY ON OR BEFORE DD/MM/YY

INPUTS

TemperatureController

OUTPUT

MENU

BI 2000 24/11/2004 16:19:49

Chamber UP Rotation OFF

X00 CHAMBER DOWN X04 STIRRER ON/OFFX01 CHAMBER UP X05 PROX. BY PASSEDX02 DOOR OPEN SW. X06 TIMER ONX03 SPARE X07 SPARE

BI 2000 24/11/2004 16:19:49

Chamber UP Rotation OFF

Y00 AUTO MODE Y04 CHAMBER DOWNY01 SPARE Y05 CHAMBER UPY02 SPARE Y06 SPAREY03 CHAMBER ROTATION Y07 BUZZER

BI 2000 24/11/2004 16:19:49

Chamber UP Rotation OFF

Chamber temp. : º CCorrection Factor:

Digital Inputs NEXT PLC MENU

PREV NEXT PLC MENU

Digital Outputs

Temperature Controller

NEXT

Page 23: BI 2000 for Touch Screen Messung

REAL TIME CLOCK ( To be activated from the MENUscreen)When this on the MENU screen is invoked a screen appears as shown. The Date and Time can be set using the calculator pad, which will appear, when pressed against the item to be changed. Then press SET to set the data entered.

DOSE RATE PROFILE (To be activated from the MENU screen) When Dose Rate Profile key from the MENU screen is pressed, a screen appears as shown: This displays the dose rate in Gy/Min at 5 locations in the Sample chamber.

23

BI 2000 24/11/2004 16:19:49

Chamber UP Rotation OFF

SET PLC REAL TIME CLOCKDate: 5 / 12 / 00

Time: 13: 11: 14

SET MENU

BI 2000 24/11/2004 16:19:49

Chamber UP Rotation OFF

Dose Rate Pattern of Sample Chamber in Gy/Min.

Size Of Sample Chamber 17.2 cm (dia)20.5 cm (H)

8.51

9.56 9.01 9.69

8.42

45

PREV PLC MENU

Page 24: BI 2000 for Touch Screen Messung

By pressing, the key the next screen can be displayed (shown below) which shows

The dose rates in Gy/Min at 5 locations in the Rotation Chamber.

SET OTHER PARAMETERSThe Enable key (lock) switch on the Control panel can enable these six touch buttons on the MENU screen. When enabled, the lock will be opened as shown:

24

MENU

BI 2000 24/11/2004 16:19:49

Chamber UP Rotation OFF

Dose Rate Pattern in Rotating Chamber in Gy/Min. 8.51

9.56 9.01 9.69 8.42

MENU

BI 2000 24/11/2004 16:19:49

Chamber UP Rotation OFF

Page 25: BI 2000 for Touch Screen Messung

SET DOSE RATEThis when pressed a screen appears as shown below: The dose rates in Sample Chamber (S C) at 9 locations and in the Rotary Chamber (R C) at 5 locations can be Set by pressing on the displayed number and using the numerical pad which then Appears. Press SET to set the entered data.

25

Operation Modes

Set Dose Rate

Set Timer Preset Values

Set Other Parameters

Data of Previous Samples PLC I/C Status

Real_Time Clock

Dose Rate Profiles

BI 2000 24/11/2004 16:19:49

Chamber UP Rotation OFF

BI 2000 24/11/2004 16:19:49

Chamber UP Rotation OFF

Date: 01 / 08 /04 Dose Rate of Sample Chamber in Gy/Min.

MENU

Dose Rate of Rotating Chamber in Gy/Min.

8.61

9.56 9.31 9.69

8.42

8.61

9.56 9.31 9.69

8.42

Page 26: BI 2000 for Touch Screen Messung

SET TIMER This when pressed a screen appears as shown:

The time values can be modified, if required, by pressing on the displayed number and using the calculator pad, which then appears.

Set Other Parameters When Set Other Parameters key is pressed, a screen appears as shown:

The Base Date, Battery Change Date, Sample Number can be set on this by pressing against the required item and using the calculator pad which then appears.

26

BI 2000 24/11/2004 16:19:49

Chamber UP Rotation OFF

BI 2000 24/11/2004 16:19:49

Chamber UP Rotation OFF

MENU

Down Fault, Time 22.0 secUp Fault, Time 22.0 secTop LS Release Time 5.0 secBTM LS Release Time 5.0 sec

Buzzer ON Time Door Open Fault Time 5.0 sec 10.0 sec

Min Irradiation Time 6.0 sec

MENU

Page 27: BI 2000 for Touch Screen Messung

By the key the next screen comes and pressing against the item and the alphabetical/numerical pad can change the client information, which then appears:

By the key the previous screen comes, Previous Sample Data can be reset by pressing

Reset Data key

When pressed a screen appears as:

27

BI 2000 24/11/2004 16:19:49

Chamber UP Rotation OFF

BatteryChangeDate : 15 / 10/ 2008

Sample Number : 00000001

Reset Prev

MENU

BI 2000 24/11/2004 16:19:49

Chamber UP Rotation OFF

BRIT Client Information 1

2

Model No. MENU

ABC PVT. LTD.

Mumbai

BI 2000

Page 28: BI 2000 for Touch Screen Messung

If YES is selected, all the previous data will be lost. If NO is selected, all the previous data will be preserved.

OPERATIONS SEQUENCE

1. MANUAL

28

Data of previous samples Will be permanently lost Do you want to continue?

MENU

YES NO

OPERATION MODE

Page 29: BI 2000 for Touch Screen Messung

2. AUTO – MODE WITH PREVIOUS DATA

3. AUTO-DOSE MODE

29

MANUAL

LOWER

RAISE

OPERATION MODE

AUTO

CONFIRM

START

Page 30: BI 2000 for Touch Screen Messung

4. AUTO-TIME MODE

30

LOG INPassword (Master)

OPERATION MODE

AUTO

CONFIRM

START

ENTER SELECTED DOSE

DOSE MODE

LOG INPassword (Master)

OPERATION MODE

AUTOCONFIRMSTARTENTER IRRADIATION TIME

TIME MODE

Page 31: BI 2000 for Touch Screen Messung

31

Page 32: BI 2000 for Touch Screen Messung

ANNEXURE – A(CONTROL CIRCUIT DIAGRAMS)

32

Page 33: BI 2000 for Touch Screen Messung

ANNEXURE – B(RADIOMETRY TEST RESULTS)

33

Page 34: BI 2000 for Touch Screen Messung

ANNEXURE – C(DOSIMETRY REPORT)

34

Page 35: BI 2000 for Touch Screen Messung

ANNEXURE – D(COMMON CHECK LIST FOR THE PACKAGE)

INSTALLATION CERTIFICATE(Gamma Chamber 5000)

Date ____________

35

Page 36: BI 2000 for Touch Screen Messung

It is certified that the unit is installed and commissioned by BRIT engineers. The

unit is working satisfactorily. The operating instruction manual with dosimetry report and

radiometry test results report has been supplied. The users personnel have been

familiarized with operation and preventive maintenance aspects of the unit. The

following are the details of the unit.

Type of the unit / Sr. No

Users institution

Activity and Date

Central Dose rate and Date

Date of installation

Contact persons/users

BRIT Engineers Users Institution

1. _________________ 1._________________ ____________________ _________________

2. __________________ 2._________________ _____________________ __________________

Place ____________________

Date _________________

36