Beam Polarimetry Matthew Musgrave NPDGamma Collaboration Meeting Oak Ridge National Laboratory Oct. 15, 2010

Beam Polarimetry

  • Upload
    cosmo

  • View
    51

  • Download
    5

Embed Size (px)

DESCRIPTION

Beam Polarimetry. Matthew Musgrave NPDGamma Collaboration Meeting Oak Ridge National Laboratory Oct. 15, 2010. Introduction. - PowerPoint PPT Presentation

Citation preview

Page 1: Beam Polarimetry

Beam Polarimetry

Matthew Musgrave

NPDGamma Collaboration MeetingOak Ridge National Laboratory

Oct. 15, 2010

Page 2: Beam Polarimetry

IntroductionThe neutron beam polarization and the spin flip efficiency of the RFSF need to be determined to obtain the parity violating physics ϒ-ray asymmetry from the measured ϒ-ray asymmetry.

Pn is the neutron beam polarization, εrf is the efficiency of the RFSF, and C is a constant representing the rest of the corrections to the measured asymmetry.

The goal of the polarimetry measurement is to obtain the polarization of the neutron beam to within 5% to help achieve a statistical precision of ~10-8 for Aϒ.

CP

AA

rfn

measured

Page 3: Beam Polarimetry

PolarimetryThe neutron beam polarization will be determined from relative neutron flux measurements through a polarized 3He analyzer cell. The neutron spin will be flipped by a RFSF and the 3He spin will be flipped by adiabatic fast passage to provide redundant methods of measuring the signal of opposing spin states of the neutrons.

RFSF 3He analyzer cells

Page 4: Beam Polarimetry

RF Spin FlipperIn the RFSF there is a static field B0=9.7G from the guide field and a RF field Brf produced by the RFSF oscillating in the direction of the neutron beam.

The RF field can be approximated as two counter rotating fields, rotating normal to the guide field B0.

If the RF field oscillates at the Larmor frequency of the neutron magnetic moment in the guide field ωL=ϒB0=29kHz then the frequency of Brf is on resonance and one rotating component of Brf follows the precession of the neutron magnetic moment. The other component is off resonance by twice the Larmor frequency and has a negligible effect on the precession of the neutron magnetic moment. In the rotating frame of reference of the neutron magnetic moment there is only a field in the ẑ direction.

ztByBB rfeff ˆ)cos(ˆ0

)ˆ)sin(ˆ)(cos(2

)ˆ)sin(ˆ)(cos(2

xtztB

xtztB

B rfrfrf

zB

zB

yBB rfrfLrot ˆ

2ˆ0

Page 5: Beam Polarimetry

RF Spin FlipperIn the rotating reference frame on resonance, the neutron magnetic moment precesses about the effective magnetic field at the Larmor frequency for Brot.

To achieve a spin flip of the neutron beam, the neutrons need to remain in the RFSF for the time required to rotate the neutron spins by π radians.

Where n=1, 3, 5… The time spent in the RFSF is dependent on the neutron’s velocity, which for the NPDGamma experiment is characterized by the neutron’s time of flight.

Where L is the length of the RFSF and l is the distance to the neutron moderator. To rotate the neutron spins by π radians, the rf field of the RFSF must be varied with the time of flight.

rotrot B

rfB

nt

2

tofrf tL

lnB

12

toftl

L

v

Lt

Page 6: Beam Polarimetry

Efficiency of RFSF at LANL

The efficiency of the RFSF at LANL was measured to be 98.8±0.5%.

Differences in the design of the beam line and the method of polarizing the neutrons will affect the efficiency of the RFSF and how it is calculated.

Page 7: Beam Polarimetry

Field Uniformity

S. Balascuta

The rf field Brf in the RFSF is not uniform because Brf is produced by a finite solenoid. If the RFSF is tuned to maximize efficiency in the center of the beam, the RFSF efficiency will be less than unity off axis.

dzzyxByxL

L rf ),,(),(2/

2/

The efficiency of the RFSF off axis is determined by the integral of the amplitude of Brf experienced by the neutron over the length of the RFSF.

Page 8: Beam Polarimetry

nn n pp

p n p

The cross-section for neutron capture on 3He with nuclear spin parallel to the neutron spin is nearly zero.

The cross-section for neutrons with spin anti-parallel to the 3He nuclear spin is very large. Twice the cross-section for unpolarized neutrons.

Neutron Polarization by Capture on Polarized 3He

nn n pp

p n p

The strong spin dependence of the neutron capture cross-section of polarized 3He makes it an effective neutron polarizer and analyzer.

Page 9: Beam Polarimetry

3He is polarized through spin exchange with optically pumped alkali metal such as Rb or K.

How do we polarize 3He?

•Only electrons in the S1/2 state with ms=-1/2 can absorb the laser light because the light is circularly polarized with magnetic projection of +1.

•The valence electron in the alkali metal absorbs a photon with angular momentum of +1 and magnetic projection of +1 and is excited to the P1/2 ms=+1/2 state .

•The excited electron will decay back to the S1/2 ground state with either value for the spin state.

•Since only the ground state electron in the ms=-1/2 spin state can absorb a photon but the excited electron can decay to either spin state, the ms=+1/2 spin state will eventually become populated.

•The 3He is then polarized through spin exchange hyperfine interactions with the electrons of the alkali atoms.

Page 10: Beam Polarimetry

Spin Exchange Optical Pumping

Collision Mixing

Page 11: Beam Polarimetry

Optical Pumping StationThe 3He cell is polarized in a static magnetic field of 12-14 Gauss, and its polarization and spin-lattice relaxation time are determined in the lab by NMR.

During optical pumping the Rb’s polarization becomes saturated in under a second, but the 3He polarization takes several hours to saturate and is usually pumped over night.

The spin-lattice relaxation time of a 3He cell will often be over 100 hours, which will provide a stable 3He polarization for polarimetry measurements.

Page 12: Beam Polarimetry

3He Analyzer CellsFive 3He analyzer cells have been produced for the polarimetry measurement and named Maxwell, Nambu, Oppenheimer, Ramsey, and Szilard. The 3He cells are 1in diameter cylindrical cells made of GE180 glass. They are filled with 3He, N2, and trace amounts of Rb for polarization of the 3He. So far three of the 3He cells have been characterized.

0

03 _nl

ThicknessHe

Page 13: Beam Polarimetry

3He Analyzer in the Beamline

Page 14: Beam Polarimetry

Neutron Transmission through a 3He CellNeglecting normalization, the measured neutron flux of an unpolarized neutron beam through an unpolarized and polarized 3He cell are given by:

Where κ is the 3He cell thickness, which is a function of the 3He density n, the length of the cell l, and a normalizing reference cross-section σ0 and neutron wavelength λ0.

The polarization of the neutron beam is given by:

If the wavelength of the neutrons is known, the polarization can be rewritten as a function of neutron flux measurements through the cell.

This is the analyzing power of the 3He cell.

0

0

nl

)cosh()(

)(0

Hepol PeN

eN

)tanh()( HeHen PP

2

0

)(

)(1)(

pol

Hen N

NP

Page 15: Beam Polarimetry

Transmission of a Polarized Neutron Beam through a 3He Cell

Assuming the neutron spin can be flipped with unit efficiency, the transmission of an unpolarized neutron beam can be approximated from a polarized neutron beam by flipping the neutron spins and averaging the flux measurements of the two neutron spin states.

The analyzing power of the 3He cell can then be determined from a polarized neutron beam.

Where R↑=N↑/N0 and R↓=N↓/N0 are ratios of the flux measurements.

)(2

)()(

)(

)(

00

N

NN

N

N pol

RR

RR

NN

NPHen

4)(

)()(

)(21)(

22

0

Page 16: Beam Polarimetry

Polarization of the Neutron Beam

The effective efficiency of the supermirror polarizer and the polarized 3He cell is the product of their polarization efficiencies, PnPn

He. The detected neutron flux through the supermirror and the 3He cell can be determined from the effective efficiency.

The polarization of the neutron beam as a function of neutron wavelength can then be calculated from relative neutron flux measurements.

4)(

21

1)(

2

RR

RR

RR

R

PP

Hen

n

)1(2

)1( Henn

HennunpolPP

PPRR

RPPNN Henn

Page 17: Beam Polarimetry

Neutron Flux Measurements

)))21(tanh()21(1)()21(cosh()()(

)))21(tanh(1)()21(cosh()()(

))tanh()21(1)(cosh()()(

))tanh(1)(cosh()()(

0

0

0

0

Hen

afpnrfHen

afpPFF

Hen

afpnHen

afpPF

HenrfHePF

HenHeP

PPPNN

PPPNN

PPPNN

PPPNN

Heafpsf

Heafp

Hesf

He

Four neutron flux measurements through a polarized 3He cell will be measured corresponding to the RFSF on and off and the two 3He spin states tuned by AFP.

Where N0(λ)=e-κλ is the neutron flux through an unpolarized 3He cell. These measurements will also be done at several 3He polarizations. Since the polarization of the neutron beam is independent of the polarization of the 3He, these additional measurements will provide a redundant check on the beam polarization. The efficiencies of the RFSF and the AFP coils can also be determined from the neutron flux measurements.

Page 18: Beam Polarimetry

RFSF Efficiency

)tanh(

)tanh()21(

0 HesmpPPF

PPF

HesmpsfPFPFF

PFPFF

sf

PPTT

TTR

PPTT

TTR

HeHeafp

HeHeafp

HesfHeafpsf

HesfHeafpsf

0

12

1

R

Rsfsf

The 3He polarization can be flipped with nearly 100% efficiency by adiabatic fast passage. By assuming no loss in polarization of the 3He, the efficiency of the RFSF can be determined by ratios of sums and differences of neutron flux measurements.

Page 19: Beam Polarimetry

Statistical PrecisionTo determine the polarization of the neutron beam to within 5%, the statistical precision of each neutron flux measurement will be measured to better than 1.5%. This equates to about 5000 neutrons detected by the neutron detector. Neutron flux measurements will be taken for about 20 time bins to determine the polarization at different neutron wavelengths .

Assuming a neutron flux of 108 n/cm2s, an aperture of 1 cm2, 20 time bins, 5×102 n/time bin, a relative neutron transmission of 0.01 through a thick 3He cell, and a 10% capture efficiency in the neutron detector. A conservative estimate for the time of a statistically significant neutron flux measurement can be calculated.

The time required to make a neutron flux measurement is small enough that the polarimetry measurement will not be statically limited.

sscmncm

btnbt1

)1.0)(01.0)(/10)(1(

.)./105.)(.20(282

3