Upload
others
View
14
Download
0
Embed Size (px)
Citation preview
Moyers
PTCOG 2010
Basic Interactions for Light Ion Beam Treatments: Penetration
Michael F. Moyers
Moyers
PTCOG 2010
Objectives
1. Describe the underlying physical processes that shape the depth dose curves for different particles.
2. Describe various parameters associated with energy and range.
3. Provide example sources and magnitudes of uncertainties for light ion penetration into a patient.
Moyers
PTCOG 2010
Assumptions
This presentation assumes the audience is familiar with:» ionization and stopping powers» multiple Coulombic scattering» bremstrahlung» nuclear reactions
The purpose of the presentation is to give the audience a "feel" for the magnitudes of the effects and how they influence light ion treatments.
Moyers
PTCOG 2010
Material for Presentation
Much of the material used in this presentation was taken from the forthcoming book:» Moyers, M. F. Vatnitsky, S. M. "Practical
Implementation of Light Ion Beam Treatments" (Medical Physics Publishing, Wisconsin, 2011)
Moyers
PTCOG 2010
Rationale for Light Ion Treatments
1. The dose delivered to non-target tissues relative to the dose delivered to target tissues is lower than for other radiation beams due to the depth dose distribution.
2. The lateral and distal dose gradients are higher than for other radiation beams enabling better splitting of the target and normal tissues.
3. For ions heavier than helium, a differential RBE with depth results in a higher effective dose in target tissues compared to surrounding normal tissues.
Moyers
PTCOG 2010
Depth Dose Distributions - Non-modulated[all distributions normalized to maximum dose]
Moyers
PTCOG 2010
Processes that Determine theShape of the Depth Dose Distribution
++++++++++otail fromsecondary particles
++++++++++buildup fromsecondary particles
++++++onuclearattenuation
+++++++radiation head scatter
ooooo+bremstrahlung
++++++++++++++straggling from multiple heterogeneity paths
++++++++++straggling from multiple scattering paths
++++++straggling from stochasticenergy losses
ooooo+++increasing obliquityfluence buildup
++++++++++++++++increasing stopping power with decreasing energy
Fe-56C-12He-4anti-HH-1eParticle
Moyers
PTCOG 2010
Electronic Stopping PowerVersus Energy and Particle
rule of thumb -
Stopping power of highest energy (clinically used) proton beam is
twice that of electrons; i.e.
4 MeV/cm instead of 2 MeV/cm.
Moyers
PTCOG 2010
Speed of Particles
1070.7350924.240.58216.62.191.0088.830.0
1540.5826216.560.45115.12.051.0021.810.0
2470.44131410.90.3458.61.901.005.903.0
3950.3471117.80.2531.81.820.982.041.0
6770.2536430.50.1816.31.920.910.740.3
11190.1819649.70.148.832.090.800.340.1
stoppingpower
[MeV/cm]
relativespeed [v/c]
energy[MeV]
stoppingpower
[MeV/cm]
relativespeed [v/c]
energy[MeV]
stoppingpower
[MeV/cm]
relativespeed [v/c]
energy[MeV]
range[cm]
carbon-12protonelectron
Moyers
PTCOG 2010
Proton Depth Dose Curve -Ionization Loss Only Along Path
max S / ent S≈ 185:1
CSDApeak/entrance
≈ 29:1
measuredpeak/entrance
≈ 4:1
Moyers
PTCOG 2010
Electron Depth Dose Curve -Ionization Loss Only Along Path
max S / ent S≈ 27:1
CSDApeak/entrance
≈ 2.5:1
measuredpeak/entrance
≈ 1.1:1
Moyers
PTCOG 2010
Straggling from StochasticEnergy Losses
ICRU 36 / Paretzke 1980
Moyers
PTCOG 2010
Straggling from MultipleParticle Paths
10 MeV electrons50 histories
80 MeV protons50 histories
150 MeV/n carbon ions500 histories
MCNPX simulations
Moyers
PTCOG 2010
Straggling from MultiplePaths Through Heterogeneities
ions traversing different paths encounter different water equivalent materialsions scattered toward the CAXhave traversed equivalent water equivalent pathlengths but stop at different physical depths; i.e. the dose deposition with respect to depth is not coherent
Moyers
PTCOG 2010
Straggling from MultiplePaths Through Heterogeneities
Urie et al., 1983
Moyers
PTCOG 2010
Electron Depth Dose Curve -Multiple Processes
Moyers
PTCOG 2010
Depth Number Distribution /Depth Dose Distribution
Moyers et al., 2007
Moyers
PTCOG 2010
End of Range / Peak Region Magnified
Moyers et al., 2007
Moyers
PTCOG 2010
Range of Protons in Non-modulatedBeam and Number of Protons Lost
rule of thumb -
for every cm of depth, ≈ 1% of
protons undergo nuclear reaction
Moyers et al., 2007
Moyers
PTCOG 2010
Range of Carbon Ions in Non-modulated Beam and Number of Carbon Ions Lost
rule of thumb -
for every cm of depth, ≈ 4% of
carbon ions undergo nuclear
reactions
Moyers
PTCOG 2010
Example Range Modulation Scheme
energy stacking from acceleratorenergy stacking via rangeshiftingrotating stepped propellorrotating stepped propellor with flux modulationstationary ridge filterstationary cones
Moyers
PTCOG 2010
Effect of Energy Spread on Depth Dose Distribution
increased energy spread →» increased entrance dose» rounding at proximal mesa» rounding at distal mesa» increased distal penumbra
Hsi et al., 2009
Moyers
PTCOG 2010
Distal Penumbra
250 MeVsynchrotron with direct energy extraction205 MeV cyclotron with rangeshifterand slit
Hsi et al., 2009
Moyers
PTCOG 2010
Various Range Parameters
Moyers
PTCOG 2010
Uncertainties in Penetration
patient information
beam delivery
mitigated by planning techniques
Moyers
PTCOG 2010
CT Number Dependencies
scanner calibrationkVp, filter, and FOVvolume and configuration of patientposition in scanartifactsimplants
Moyers
PTCOG 2010
CT Number Scaling
XCTNmat - XCTNwat SXCTN = + 1 x 1000
XCTNwat - XCTNair
Wrightstone et al., 1989
Moyers
PTCOG 2010
Uncertainties in Stopping Powers
range in water» ICRU 49 (1993) versus Janni (1982)
≈ 4 mm different at 250 MeV (≈1%)» several studies have shown the I-value is closer to the 81.77 eV
used by Janni than the 75.0 eV used by ICRU 49
Relative Linear Stopping Power for protons» Moyers et al. (1992): tissue substitutes and ancillary equipment -
calculated/literature versus measured 0.4 - 3.3%» Schneider et al. (1996): tissues - calculated/literature versus
measured 1.6%» small energy dependence (ignored in most TPSs)
≤ ±1% for z<13 between 30 and 250 MeV» almost no dependence on ion species
Moyers
PTCOG 2010
Uncertainties in Converting CT # to RLSP
for std. tissues, CT # to RLSP≤ 1.6%
for some other materials, CT # to RLSP>> 1.6%
Moyers
PTCOG 2010
Uncertainties in Penetration -Patient Information Related
± 2.2± 3.5--Total(soft tissues only)
± 0.5MC± 1.2-± 1.2%energy dependence ofRLSP for low Z
± 1.6± 1.6-± 1.6%WEQ vs RLSP(soft tissues only)
± 1.0%± 1.0%contour and substitute
± 0.0 to 3.0%RLSP of tissues anddevices
± 0.5%± 1.0%-± 1.0%stopping powerof water
± 5.0% metal*± 5.0% metal*z ≤ 22 - MVXCTz > 22 - substitution
100%metal implants
± 0.5% waterDE*± 0.8% tissueDE
> ± 1.0% boneDE*
± 1.5% water*± 2.5% tissue
> ± 3.0% bone*
-± 1.5% water± 2.5% tissue> ± 3.0% bone
position in scan
± 0.0%± 0.0%patient specificscaling
± 2.5%volume andconfiguration scanned
± 0.0%± 0.0%use only calibratedconditions
± 2.0% PMMA, PC> ± 2.0% bone
kVp, filter, andFOV selection
± 0.0%± 0.0%patient specificscaling
± 0.3% day-to-dayscanner calibration forstandard conditions
possible future uncertainty
uncertaintyafter mitigation
mitigationuncertaintybefore mitigation
cause
* ≡ not considered in totalDE ≡ dual energy CTMC ≡ Monte Carlo calc.
Moyers et al., 2009
Moyers
PTCOG 2010
Uncertainties in MeasuringStandard Range in Water
• water tank window ± 0.008 mm• parallel plate chamber waterproof lid ± 0.036 mm• parallel plate chamber front wall ± 0.044 mm• setting of reference depth with spacer ± 0.3 mm• calibration of water tank scanning
mechanism ± 0.23 mm• backlash in water tank scanning
mechanism ± 0.3 mm• rigidity and tilt of ion chamber mount on
scanning mechanism ± 0.3 mm
• total range uncertainty ± 0.571 mm
Moyers
PTCOG 2010
Uncertainties in Bolus WaterEquivalent Thickness
bolus material density (with lot sampling) ± 0.5% → 0.4 mmbolus manufacturing ± 0.3 mmbolus voids (with CT sampling) and/or contamination ± 0.5 mm» will not be everywhere
total bolus: ((0.4 + 0.3)2 + 0.52)0.5 ± 0.9 mm
Moyers
PTCOG 2010
Alignment Devices
face masks» unlikely that holes and net will cover same anatomical
location as when CT scan and plan were performed » ± 1 mm
table tops and pods» materials often do not lie on the CT number to RLSP
conversion curve» ± 1 mm
Moyers
PTCOG 2010
Energy/Range Accuracy Requirements
absolute accuracy» Monte Carlo or analytical based planning
reproducibility» measured depth dose curve used in planning» TPS planned versus treatment delivered» adequate tumor coverage» distal blocking of normal tissues» patch field combinations
relative accuracy» energy stacking
Moyers
PTCOG 2010
Simulated Treatment IncludingVarious Errors
selection of average beam flux levelflux fluctuations during deliverynoise on flux monitor electronicsspot positionsbeam range
Yang et al., 2003
Moyers
PTCOG 2010
Sample Simulation Results[all curves for random energy error selected over 0.2 MeV]
Yang et al., 2003
Moyers
PTCOG 2010
Energy Accuracy Requirements
energy stacking for proton range modulation» uniform mesa requires relative accuracy of 0.1 MeV at 70
MeV and 0.25 MeV at 250 MeV (i.e. at 70 MeV a 2.3 MeVjump must actually be a 2.2 to 2.4 MeV jump)
Moyers
PTCOG 2010
Uncertainties in Delivering CorrectEnergy or Range During Treatment
• synchrotron under normal operation, small deviations (microns)
• when a component fails, can have large deviations (need interlocks)
Moyers et al., 2008
Moyers
PTCOG 2010
Uncertainties in Delivering CorrectEnergy or Range Each Treatment
• accelerator beam energy• variable scattering foils• green area represents
± 0.1 MeV (± 0.18 mm)
Moyers et al., 2008
Moyers
PTCOG 2010
Bolus Position Relative to Patient(Lateral Set-up Uncertainty Effect on Range)
nominal patient/bolus alignment 3 mm patient/bolus miss-alignment
Moyers
PTCOG 2010
Heterogeneous Material Straggling
use of multiple beam entry angles to reduce uncertainty of straggling and relocate localized regions of under and over dose
Moyers
PTCOG 2010
Patient Motion
bolus expansiontarget expansionRLSP assignmentsbreath holdgatingtrackingrescanning with synchronized delayapneic oxygenation
Moyers
PTCOG 2010
Summary of TypicalPenetration Uncertainties
standard energy (or range) ± 0.6 mmenergy (or range) reproducibility ± 0.1 to 2.0 mmbolus WET ± 0.9 mmalignment devices ± 1.0 mm
CT# accuracy (after scaling) ± 2.5%RLSP of tissues and devices ± 1.6%energy dependence of RLSP ± 1.0%CT# to RLSP (soft tissues only) ± 1.6%
bolus position relative to patient variableheterogeneity straggling variablepatient motion variable
Beam Delivery2.6 to 4.5 mm(linearly)
Patient Information3.5%(quadratically)
Planning Mitigationbolus expansionmultiple anglesbreathing control
Moyers
PTCOG 2010
Summary
The shape of light ion beam depth dose distributions is determined by multiple processes.There are multiple range parameters.The depth of penetration of a light ion beam in a patient is uncertain due to several factors. Allowances must be made to account for these factors.Delivery systems must be designed to detect/correct abnormal energy/range delivery.