Upload
dwain-cecil-hamilton
View
214
Download
0
Embed Size (px)
Citation preview
AV Blocks AV Blocks Artificial PacemakersArtificial Pacemakers
Terry White, RN, EMT-PTerry White, RN, EMT-P
AV BlocksAV Blocks
Disorders of conduction at AV Disorders of conduction at AV JunctionJunction
CategoriesCategories– First degree (1° AV Block)First degree (1° AV Block)– Second degree (2° AV Block)Second degree (2° AV Block)
Type IType I Type IIType II
– Third degree (3° AV Block or Complete Third degree (3° AV Block or Complete AV Block)AV Block)
AV BlocksAV Blocks
First DegreeFirst Degree– Prolonged AV conduction timeProlonged AV conduction time– PR interval > 0.20 secondsPR interval > 0.20 seconds– Characteristics of that of any other Characteristics of that of any other
rhythm with a SINGLE sinus or atrial rhythm with a SINGLE sinus or atrial pacemaker sitepacemaker site
– Associated with an underlying sinus Associated with an underlying sinus or atrial rhythm!!!or atrial rhythm!!!
AV BlocksAV Blocks
First DegreeFirst Degree– CausesCauses
AV node ischemia/hypoxiaAV node ischemia/hypoxia Increased vagal or decreased sympathetic Increased vagal or decreased sympathetic
tone tone Drug effectsDrug effects
– DigitalisDigitalis– Beta blockersBeta blockers– Calcium channel blockersCalcium channel blockers– QuinidineQuinidine– PronestylPronestyl
AV BlocksAV Blocks
First DegreeFirst Degree– ManagementManagement
Usually requires no specific treatmentUsually requires no specific treatment Treat the patient!!!Treat the patient!!! Monitor for progression to higher degree Monitor for progression to higher degree
blockblock
AV BlocksAV Blocks
Second DegreeSecond Degree– DefinitionDefinition
More Ps than QRSsMore Ps than QRSs Every QRS caused by a PEvery QRS caused by a P
– The pattern determines the type of 2° The pattern determines the type of 2° AV blockAV block
– Since requires presence of P waves, it Since requires presence of P waves, it also requires an underlying sinus or also requires an underlying sinus or atrial rhythmatrial rhythm
AV BlocksAV Blocks
Second DegreeSecond Degree– TypesTypes
Type IType I– VariableVariable– Wenckebach phenomenonWenckebach phenomenon
Type IIType II– FixedFixed– ClassicalClassical
AV BlocksAV Blocks
Second DegreeSecond Degree– Type IType I
DefinitionDefinition– PR interval lengthensPR interval lengthens– Beat dropsBeat drops
PathophysiologyPathophysiology– Usually physiologicUsually physiologic
Increased vagal tone (Acute inferior MI, RVI)Increased vagal tone (Acute inferior MI, RVI) Drug effects (digitalis, beta blockers, CCBs)Drug effects (digitalis, beta blockers, CCBs)
– Frequently resolvesFrequently resolves
AV BlocksAV Blocks
Second DegreeSecond Degree– Type IType I
Good prognosisGood prognosis Specific therapy usually not necessarySpecific therapy usually not necessary
– therapy, if indicated, most likely targeted therapy, if indicated, most likely targeted towards bradycardiatowards bradycardia
Treat the patient!!!Treat the patient!!!
AV BlocksAV Blocks
Second DegreeSecond Degree– Type IIType II
DefinitionDefinition– P waves fail to conduct without warningP waves fail to conduct without warning– PR interval does not lengthenPR interval does not lengthen
CharacteristicsCharacteristics– Atrial rate > Ventricular rateAtrial rate > Ventricular rate– QRS QRS usuallyusually longer than 0.12 sec longer than 0.12 sec– Usually 4:3 or 3:2 conduction ratio (P:QRS Usually 4:3 or 3:2 conduction ratio (P:QRS
ratio)ratio)
AV BlocksAV Blocks
Second DegreeSecond Degree– Type IIType II
PathophysiologyPathophysiology– Organic lesions in bundle branchesOrganic lesions in bundle branches
Usually occurs below bundle of His in the Usually occurs below bundle of His in the bundle branches (infranodal AV block)bundle branches (infranodal AV block)
Intermittent block of conduction through Intermittent block of conduction through one bundle and complete block in otherone bundle and complete block in other
Usually caused by Acute anterior or Usually caused by Acute anterior or anteroseptal MIanteroseptal MI
AV BlocksAV Blocks
Second DegreeSecond Degree– Type IIType II
OutlookOutlook– Not goodNot good– Usually associated with anterior or Usually associated with anterior or
anteroseptal MIanteroseptal MI– Frequent progression to complete AV blockFrequent progression to complete AV block
Requires pacemakerRequires pacemaker Worsened by digitalis, procainamide, Worsened by digitalis, procainamide,
lidocaine, propranolol, TCAslidocaine, propranolol, TCAs
AV BlocksAV Blocks CompleteComplete
– DefinitionDefinition No conduction through AV nodeNo conduction through AV node Independent atrial and ventricular rhythmsIndependent atrial and ventricular rhythms Ventricular depolarization dependent on Ventricular depolarization dependent on
automaticity of ventricular pacemaker sitesautomaticity of ventricular pacemaker sites
– PathophysiologyPathophysiology AV node hypoxia/ischemiaAV node hypoxia/ischemia Myocardial infarctionMyocardial infarction Increased vagal or decreased sympathetic tone Increased vagal or decreased sympathetic tone
AV BlocksAV Blocks CompleteComplete
– CharacteristicsCharacteristics Atrioventricular dissociationAtrioventricular dissociation Regular P-P and R-R but without Regular P-P and R-R but without
association between the twoassociation between the two Atrial rate > Ventricular rateAtrial rate > Ventricular rate QRS > 0.12 secQRS > 0.12 sec
AV BlocksAV Blocks
CompleteComplete– OutlookOutlook
Junctional escape rhythm: goodJunctional escape rhythm: good Ventricular escape rhythm: badVentricular escape rhythm: bad
– WarningWarning Do NOT give lidocaine or other ventricular Do NOT give lidocaine or other ventricular
antidysrhythmics!!!antidysrhythmics!!!
AV BlocksAV Blocks ManagementManagement
– Treatment based on Sx/SxTreatment based on Sx/Sx– Most common complication = BradycardiaMost common complication = Bradycardia
IV/OIV/O22/ECG Monitor/12 lead ECG/ECG Monitor/12 lead ECG Atropine (not useful in 2° Type II or 3° AV Block)Atropine (not useful in 2° Type II or 3° AV Block) TCP (bridge to transvenous pacer)TCP (bridge to transvenous pacer) Catecholamine dripCatecholamine drip
– Prophylactic pacer application (standby)Prophylactic pacer application (standby) 2° Type II AV block2° Type II AV block 3° AV Block3° AV Block
Cardiac PacemakersCardiac Pacemakers
DefinitionDefinition– Delivers artificial stimulus to heartDelivers artificial stimulus to heart– Causes depolarization and contractionCauses depolarization and contraction
UsesUses– BradyarrhythmiasBradyarrhythmias– AsystoleAsystole– Tachyarrhythmias (overdrive pacing)Tachyarrhythmias (overdrive pacing)
Cardiac PacemakersCardiac Pacemakers TypesTypes
– FixedFixed Fires at constant rateFires at constant rate Can discharge on T-waveCan discharge on T-wave Very rareVery rare
– DemandDemand Senses patient’s rhythmSenses patient’s rhythm Fires only if no activity sensed after preset Fires only if no activity sensed after preset
interval (escape interval)interval (escape interval)
– Transcutaneous vs Transvenous vs Transcutaneous vs Transvenous vs ImplantedImplanted
Cardiac PacemakersCardiac Pacemakers
Demand Pacemaker TypesDemand Pacemaker Types– VentricularVentricular
Fires ventriclesFires ventricles
– AtrialAtrial Fires atriaFires atria Atria fire ventriclesAtria fire ventricles Requires intact AV conductionRequires intact AV conduction
Cardiac PacemakersCardiac Pacemakers
Demand Pacemaker TypesDemand Pacemaker Types– Atrial SynchronousAtrial Synchronous
Senses atriaSenses atria Fires ventriclesFires ventricles
– AV SequentialAV Sequential Two electrodesTwo electrodes Fires atria/ventricles in sequenceFires atria/ventricles in sequence
Cardiac PacemakersCardiac Pacemakers
ProblemsProblems– Failure to captureFailure to capture
No response to pacemaker artifactNo response to pacemaker artifact Bradycardia may resultBradycardia may result Cause: high “threshold”Cause: high “threshold” ManagementManagement
– Increase amps on temporary pacemakerIncrease amps on temporary pacemaker– Treat as symptomatic bradycardiaTreat as symptomatic bradycardia
Cardiac PacemakersCardiac Pacemakers
ProblemsProblems– Failure to senseFailure to sense
Spike follows QRS within escape intervalSpike follows QRS within escape interval May cause R-on-T phenomenonMay cause R-on-T phenomenon ManagementManagement
– Increase sensitivityIncrease sensitivity– Attempt to override permanent pacer with Attempt to override permanent pacer with
temporarytemporary– Be prepared to manage VFBe prepared to manage VF
Cardiac PacemakersCardiac Pacemakers
ProblemsProblems– Inappropriate absence of pacer artifactInappropriate absence of pacer artifact
CausesCauses– Depleted batteryDepleted battery– Circuit malfunctionCircuit malfunction– OversenseOversense
ManagementManagement– Decrease sensitivityDecrease sensitivity– Treat bradycardiaTreat bradycardia– Replace pacemakerReplace pacemaker
Cardiac PacemakersCardiac Pacemakers ProblemsProblems
– Runaway pacemakerRunaway pacemaker Rates of up to 400/minuteRates of up to 400/minute Increasing rate = EmergencyIncreasing rate = Emergency CausesCauses
– Component failureComponent failure– Battery depletionBattery depletion
ManagementManagement– TransportTransport– Enter site surgically, cut leadEnter site surgically, cut lead– Some may be turned “off” by donut-shaped magnetSome may be turned “off” by donut-shaped magnet
Cardiac PacemakersCardiac Pacemakers
Special ConsiderationsSpecial Considerations– Pacemaker does NOT affect treatment Pacemaker does NOT affect treatment
of cardiac arrestof cardiac arrest– Do NOT fire defibrillator directly over Do NOT fire defibrillator directly over
pacemaker generatorpacemaker generator– Pacemakers may keep AEDs from Pacemakers may keep AEDs from
advising shockadvising shock
Cardiac PacemakersCardiac Pacemakers Transcutaneous PacingTranscutaneous Pacing
– Electrical vs Mechanical captureElectrical vs Mechanical capture– Tips for improving electrical captureTips for improving electrical capture
Ensure adequate conductanceEnsure adequate conductance Increase current (mA)Increase current (mA)
– Tips for improving mechanical captureTips for improving mechanical capture Ensure the “tank” is topped offEnsure the “tank” is topped off Increase the electrical rateIncrease the electrical rate Increase peripheral vascular resistanceIncrease peripheral vascular resistance
Cardiac PacemakersCardiac Pacemakers
Transcutaneous Pacing RealizationsTranscutaneous Pacing Realizations
It is much easier to increase the electrical rate of depolarization than it is to increase the
mechanical rate of contraction!
You can’t create mechanical capture in dead muscle!
Implanted DefibrillatorsImplanted Defibrillators
AICDAICD– Automated Automated
Implanted Cardio-Implanted Cardio-DefibrillatorDefibrillator
UsesUses– TachyarrhythmiasTachyarrhythmias– Malignant Malignant
arrhythmiasarrhythmias VTVT VFVF
Implanted DefibrillatorsImplanted Defibrillators
Programmed at insertion to deliver Programmed at insertion to deliver predetermined therapies with a set order predetermined therapies with a set order and number of therapies including:and number of therapies including:– pacingpacing– overdrive pacingoverdrive pacing– cardioversion with increasing energiescardioversion with increasing energies– defibrillation with increasing energiesdefibrillation with increasing energies– standby modestandby mode
Effect of standby mode on Paramedic Effect of standby mode on Paramedic treatmentstreatments
Implanted DefibrillatorsImplanted Defibrillators Potential ComplicationsPotential Complications
– Fails to deliver therapies as intendedFails to deliver therapies as intended worst complicationworst complication requires Paramedic interventionrequires Paramedic intervention
– Delivers therapies when NOT appropriateDelivers therapies when NOT appropriate broken or malfunctioning leadbroken or malfunctioning lead parameters for delivery are not specific enoughparameters for delivery are not specific enough
– Continues to deliver shocksContinues to deliver shocks parameters for delivery are not specific enough and parameters for delivery are not specific enough and
device senses a resetdevice senses a reset may be shut off (not standby mode) with donut-may be shut off (not standby mode) with donut-
magnetmagnet