AUTOMATION OF LC/SPME-COUPLING IN

96-WELLPLATE FORMAT

Dietmar Hein PAS Technology Deutschland GmbH

Page 2: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

Plunger

Barrel

Z-slot

Plunger Retaining Screw

Hub-Viewing Window

Commercial Design of the SPME Device by Supelco

11/15/2011 2

Adjustable NeedleGuide/Depth Gauge

Hub-Viewing Window

Tensioning Spring

Sealing Septum

Septum Piercing Needle

Fiber Attachment Tubing

Coated Fused Silica Fiber

Page 3: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

Schematic of Manual Interface

Commercial SPME fibre assembly

Valco tee (SS) (0.75 mm ID)

fingertight nut

Inner needle

Outerneedle

Inner needle

11/15/2011 3

ferrule (VG) 0.4 mm ID

sorbent

From pump

To column

Six-port valve

1/16” tubing (0.03” ID)

Page 4: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

LC-SPME coupling

SPME-LC INTERFACES AND AUTOMATION IN THE PAST

SPME-LC INTERFACE AUTOMATIONSAMPLE

THROUGHPUT

11/15/2011 4

THROUGHPUT

MANUAL INTERFACE LOW LOW

IN-TUBE SPME HIGH LOW

OFF-LINE DESORPTION LOW HIGH

Page 5: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

LC-SPME coupling

ACHIEVE HIGH DEGREE OF AUTOMATION AND HIGH SAMPLE-THROUGHPUT

11/15/2011 5

PERFORM EXTRACTION AND DESORPTION OF MANY SIMILAR

SAMPLES IN PARALLEL

Page 6: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

Fibre Multiwell System

a)

Multi-fibre top plate

a) For simplification one row and one column of SPME fibres are shown to be inserted into the top plate.

11/15/2011 6

PDMS-coated flexible wire

96-well plate

b)

b) The multi-fibre top plate can then be placed into a commercial multi-well plates for extraction and desorption.

Page 7: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

Well filled with sample

Insert SPME fibre for extraction

Agitate well until equilibrium is reached

11/15/2011 7

Remove fibre from well

Desorb in well filled with solvent

N2

Evaporate solventReconstitute and inject into LC

Page 8: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

Advantages of SPME

0CVKn ffs=

no further sample preparation!!!

11/15/2011 8

increasing sensitivity by increasing Vf!!!

Vf = d f x A f

increased area improves sensitivity and extraction rate!!!

Page 9: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

The “Brush”

11/15/2011 9

Page 10: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

• Pre-washing/conditioning of fiber/membrane in well-plate

• Extraction (variable time setting) of analytes using a PAN/C18 coated blade device

• Washing of fiber/membrane prior to desorption

CONCEPT 96

11/15/2011 10

Washing of fiber/membrane prior to desorption

• Desorption (variable time setting) of analytes from coated blade device into different solvent system

• Evaporation of solvent – pre-concentration/re-constitution step (depends on type of solvent been used for desorption)

• (Injection)

Page 11: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

CONCEPT 96

11/15/2011 11

Page 12: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

• 4 stepper-motor driven axis (X, Y, Z, Z1)Axis max. range step widthMain axis Y 560 mm 0.250 mmArm axis X 170 mm 0.030 mmInjector axis Z 160 mm 0.150 mmSyringe piston Z1 70 mm 0.047 mm

• 2 pneumatically driven axis

CONCEPT 96

11/15/2011 12

• 2 pneumatically driven axisEXTRACTION-Tool (96 fiber „Brush“)DRYING-Tool („Shower-Head“)

• Up to 4 agitators (Conditioning, Extraction, Washin g,Desorption/Evaporation), optionally heatable, (0 – 1 .700 rpm)

Page 13: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

Available coatings:

• C18-polyacrylonitrile (C18-PAN)

• Polystyryne-divynyl benzene- polyacrylonitrile (DVB-PAN )

• Phenylboronic acid -polyacrylonitrile (PBA-PAN)

CONCEPT 96

11/15/2011 13

• Phenylboronic acid -polyacrylonitrile (PBA-PAN)

• Methacryloyloxy ethyl phosphorylcholine (MPC)

• Chromabond Easy/PAN

• LC Diol/PAN

• C18SCX/PAN

Page 14: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

CONCEPT 96

Method for determination of reusability and reprodu cibility

PRECONDITIONINGPAN-C18/ -PS-DVB30 min MeOH/water (1/1)

EXTRACTION60 min, 1000 rpm

Wash15 seconds

Purified water

11/15/2011 14

MeOH/water (1/1)Purified water

DESORPTION40 min120 rpm

Acetonitrile/water (1/1)

Sample Prep. Time: 130 minutes!!

Page 15: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

Reusability and reproducibility

of C18-PAN coating for

extraction from PBS

and plasma

40

80

120

%E

xtra

ctio

n re

cove

ry

Reusability and reproducibility of C18-PAN: extract ion from PBS

Diazepam

Lorazepam

0

1 5 15 25 35 45 55 63 73 85 95 100%E

xtra

ctio

n re

cove

ry

Extraction #

Lorazepam

0

2.5

5

7.5

10

% A

bsol

ute

reco

very

Extraction

Reusability of C18-PAN coating: extraction from pla sma

Diazepam

Page 16: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

Reusability and reproducibility of DVB -PAN coating in PBS

Reusability of DVB-PAN for diazepam in PBS

20

30

40

50

60

70

80

90

100

110

120

1 5 10 15 20 25 30 40

% A

bso

lute

reco

very

DIAZEPAM

20

40

60

80

100

120

7 12 20 25 30 40

% A

bsol

ute

reco

very

Extraction

Reusability of DVB-PAN for caffeine and riboflavin in PBS

Caffeine

Riboflavin

1 5 10 15 20 25 30 40

Extraction

Page 17: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

Efficiency of DVB-PAN for extraction of diazepam fro m plasma

Reusability and reproducibility of DVB -PAN coating in Plasma

0

10

20

30

40

50

1 10 25

ng e

xtra

cted

Extraction #

Diazepam

Page 18: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

Comparison of absolute recovery of the PAN -PS-DVB coating with C18-PAN

Comparison of absolute recovery of the PAN-PS-DVB coating with C18-PAN (extraction from PBS and n=12) Analyte Log P Pka %recovery

for PAN-PS-

%recovery for PAN-C18 coating PAN-PS-

DVB coating

C18 coating

Diazepam 2.82 3.4 96 ±4 97±3 Oxazepam 2.24 12.4 97±3 80±4 Caffeine -0.07 10.4 99±5 40±6 Riboflavin -1.46 10.2 75±4 60±5 Sucrose -3.7 12.6 3±0.2 -

Page 19: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

Limits of detection and quantitation

PAN-PS-DVB SPME-LC-MS/MS detection and quantitation

LOD and LOQ for extraction of benzodiazepines from plasma LOD for extraction from plasma (3×S/N)

LOQ for extraction from plasma (10×S/N)

0.1- 0.3 ng/mL 0.5-1 ng/mL

PAN-PS-DVB SPME-LC-MS/MS detection and quantitationAnalyte LOD (extraction from

plasma) ng/mL LOQ (extraction from plasma) ng/mL

Oxazepam 0.1 0.5 Diazepam 0.3 1 Caffeine 0.3 1 Riboflavin 0.5 1.5 Sucrose 10 25

Page 20: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

CONCEPT 96 Application

ANALYSIS OF BENZODIAZEPINES IN WHOLE BLOOD

PRECONDITIONINGRPA coating

30 minMethanol/water (1/1)

EXTRACTION30 min, 850 rpm

Whole blood or plasma (0.8 mL) + IS

RINSE30 seconds

Purified water

11/15/2011 20

Methanol/water (1/1) (0.8 mL) + ISPurified water

DESORPTION30 min, 850 rpm

Acetonitrile/water (1/1)

Sample Prep. Time: 90 minutes!!

Page 21: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

CONCEPT 96 Application

ANALYSIS OF BENZODIAZEPINES IN WHOLE BLOOD

5 ng/mL std in whole blood

11/15/2011 21

5 ng/mL std in whole blood

blank whole blood

Page 22: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

CONCEPT 96 Application

ANALYSIS OF BENZODIAZEPINES IN WHOLE BLOOD

Validation Parameter Diazepam Oxazepam Nordiazepam Lorazepam

LLOQ (ng/mL) 4 4 4 4

LLOQ Accuracy and 102 111 105 102

11/15/2011 22

LLOQ Accuracy and Precision (%) (11) (17) (9) (14)

LLOQ S/N RATIO 5 19 16 16

Linear Range (ng/mL) 4-1000 4-500 4-1000 4-500

Accuracy (%) 94-103 91-98 98-106 97-106

Intra-batch Precision (%) 2-8 8-20 5-6 7-11

Inter-batch Precision (%) 3-6 7-12 2-4 7-12

Page 23: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

AUTOMATED DRUG-PROTEIN BINDING STUDY

CONCEPT 96 Application

• Determine time required to reach equilibrium between ligand and receptor

• Obtain fibre constant (fC) by calibration using standard solutions containing no receptor

11/15/2011 23

• Prepare standard solutions containing different amounts of ligand and constant amount of receptor

• Perform SPME-LC to determine the amount of analyte extracted by the fibre (m)

• Calculate free concentration of ligand (Cf)

Page 24: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

AUTOMATED DRUG-PROTEIN BINDING STUDY

CONCEPT 96 Application

BINDING OF DIAZEPAM AND

HUMAN SERUM ALBUMIN

Diazepam - Human Serum Albumin Binding Curve

0.10

0.12

0.14

11/15/2011 24

ALBUMIN

1-SITE BINDING MODEL

regression coefficient = 0.991

Free Concentration of Diazepam (M)

0.0 2.0e-7 4.0e-7 6.0e-7 8.0e-7 1.0e-6 1.2e-6 1.4e-6 1.6e-6

B

0.00

0.02

0.04

0.06

0.08

Experimental1-site binding model

Page 25: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

AUTOMATED DRUG-PROTEIN BINDING STUDY

CONCEPT 96 Application

ADVANTAGES OF SPME TO STUDY LIGAND-RECEPTOR BINDING

•Ability to study binding under any conditions and for any

11/15/2011 25

•Ability to study binding under any conditions and for any concentrations of receptors and ligands

•Fast

•Works well for highly bound drugs

•Automation and increased sample throughput

Page 26: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

Ochratoxin A in Human Urine

CONCEPT 96 Application

Ochratoxins:

•structurally related secondarymetabolites, produced by Penicillium verrucosum

11/15/2011 26

Penicillium verrucosum

•a common storage fungus

•OTA has nephrotoxic, carcinogenic and immunosuppres sive properties

Page 27: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

Ochratoxin A in Human Urine

CONCEPT 96 Application

•Sample volume: 1 mL urine

•Extraction Time: 1 h @ 850 rpm

•Extraction Fiber: 12 mm of Carbon Tape on 0.061 SS -wire

11/15/2011 27

•Extraction Fiber: 12 mm of Carbon Tape on 0.061 SS -wire

•Desorption Solvent: 0.9 ml MeOH

•Desorption Time @ 15 min

•Injection Volume LC-MS/MS: 20 µl

Sample Prep. Time: 75 minutes!!

Page 28: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

Ochratoxin A in Human Urine

CONCEPT 96 Application

Validation Parameter

Ochratoxin A – Summary of Validation results

LOD (ng/mL) 0.3

LLOQ (ng/mL) 0.7

11/15/2011 28

LLOQ (ng/mL) 0.7

Linear Range (ng/mL) 0.7-50

1 ng/mL 10 ng/mL 50 ng/mL

Accuracy and Intra-batch

Precision (%)

106(12)

114(2)

93(4)

Accuracy and Inter-batch

Precision (%)

91 (14)

100(5)

109(4)

Page 29: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

References

• Erasmus Cudjoe, Janusz Pawliszyn, PBA-6864, 2008, „ A new approach to the application of solid phase extraction disks with LC– MS/MS for the analysis of drugs on a 96-well plate format“

• R. Vatinnoa, D. Vuckovic, C.G. Zambonin, J. Pawliszy n, Journal ofChromatography, 2008, „Automated high-throughput me thod using solid-phase microextraction–liquid chromatography–tandem mass sp ectrometry for the determination of ochratoxin A in human urine“

• Dajana Vuckovic, Erasmus Cudjoe, Dietmar Hein, Janusz Pawliszyn, Anal. Chem. 2008, „Automation of Solid-Phase Microextraction in H igh-Throughput Format and Applications to Drug Analysis“

• Dajana Vuckovic , Erasmus Cudjoe, Florin Marcel Musteata , Janusz Pawliszyn , Vol.

11/15/2011 29

• Dajana Vuckovic , Erasmus Cudjoe, Florin Marcel Musteata , Janusz Pawliszyn , Vol. 5, No. 1, 2010 Nature Protocols, „Automated solid-p hase microextraction and thin-film microextraction for high-throughput analysis of biological fluids and ligand-receptor binding studies“

• Fatemeh Mirnaghi, Janusz Pawliszyn, Yong Chen, Leon ard Sidisky, Dietmar Hein, ASMS 2011 Poster, „Biocompatible and reusable octad ecyl-polyacrylonitrilecoating for high throughput automated 96-thin-film solid phase microextractionsystem coupled with LC-MS/MS”

• Fatemeh Mirnaghi, Janusz Pawliszyn, ASMS 2011 Poste r „Modified PAN-PS-DVB 96-Thin-FilmSPME System, Capable of Extracting wide polarity range of Analytes from Biological Fluids”

Page 30: AUTOMATION OF LC/SPME-COUPLING IN 96-WELLPLATE FORMAT · ASMS 2011 Poster, „Biocompatible and reusable octadecyl-polyacrylonitrile coating for high throughput automated 96-thin-film

Acknowledgements• Prof. Pawliszyn and his group members, specifically

Fatemeh Mirnaghi, Erasmus Cudjoe and Dajana Vuckovic , who supplied most of the data

11/15/2011 30

Vuckovic , who supplied most of the data http://www.spme.uwaterloo.ca/

• Supelco http://www.sigmaaldrich.com/Brands/Supelco_Home/Spotlights/SPME_central.html