3
Automatic Optimizaiton of MS and MS/MS Assays with Dilute Samples or Weak Transitions Bennett S. Kalafut Thermo Fisher Scientific, San Jose, California, USA Poster Note 64440 Overview Purpose: To optimally tune an MS/MS or single ion monitoring (SIM) assay using a dilute sample or weak signal. Methods: Kernel intensity function estimation is used to recover the relationship between intensity and tuning from a noisy raw signal. Results: Optimal collision energies can be recovered from full range collision energy scans even if the signal consists of as few as ten to thirty single-ion events registered at the detector. Introduction SIM and MS/MS assays must typically be optimized for each compound to be tested. Optimal tunings of ion optics devices like the S-lens (SRIG or ion funnel) can vary from compound to compound. CID collision energy tuning, especially, must be tuned for each MS/MS assay since it is particular to each transition. When the concentration of the compound to be optimized is limited due to cost or scarcity or the informative CID transitions are weak, generation of smooth breakdown or tuning curves with clear optima can be costly in time and resources. The method presented here uses kernel intensity function estimation, a statistical algorithm related to kernel density estimation, to optimize collision energies or other assay parameters using signals that are very sparse, resulting from the detection of as few as tens of ions even in the neighborhood of the optimal instrument tuning. Methods Sample Preparation Testosterone and estradiol samples were prepared by serial dilution in 50% methanol, 50% water, 0.1% formic acid from 1 mg/mL stock solutions. Mass Spectrometry Samples were infused to the HESI source of a Thermo Scientific™ TSQ Enduramass spectrometer at a rate of 5 μL per minute. Ionization voltage was set to 3500 V. Sheath and auxiliary gas flow were set using the mass spectrometer’s source optimization routines. Data Analysis Kernel intensity function estimation was performed on-instrument using algorithms coded for use in compound optimization in version 2.0 of the control software of the TSQ Endura MS and Thermo Scientific™ TSQ Quantiva™ mass spectrometer. Further analysis of the results was performed offline using the SciPy statistics library (www.scipy.org). FIGURE 1. At low signal intensities, observed tuning curves become trains of discrete single- or few-ion arrival events. The problem (kernal intensity function estimation): Given steady production of ions at the source and stable filtration of ions by quadrupole mass analyzers, arrival of ions at a triple quadrupole’s mass detector is approximately a Poisson process and the detector signal itself obeys Polya statistics. The relative error varies with intensity and when the ion current is weak the observed signal is often 0: the signal is a series of discrete single- or few-ion arrivals (Fig. 1). Complicating matters further, the relationship between collision energy or instrument tuning voltage (“tuning curve”) and signal intensity is not a simple function consistent from compound to compound. We cannot simply fit a regression curve to the data even taking into account the non-homoscedastic noise of the signal; we do not know a priori what kind of curve to fit. FIGURE 2. A sample kernel intensity estimate. For each ion arrival event, represented by the green X, a kernel (blue curves) is added to the reconstructed tuning curve (red curve.) Here, the kernel bandwidth is an arbitrary 10 units wide. In practice, the kernel bandwidth is set by considering the statistical structure of the data.

Automatic Optimizaiton of MS and MS/MS Assays with Dilute ...tools.thermofisher.com/content/sfs/posters/PN... · Automatic Optimizaiton of MS and MS/MS Assays with Dilute Samples

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Automatic Optimizaiton of MS and MS/MS Assays with Dilute ...tools.thermofisher.com/content/sfs/posters/PN... · Automatic Optimizaiton of MS and MS/MS Assays with Dilute Samples

Automatic Optimizaiton of MS and MS/MS Assays with Dilute Samples or Weak Transitions Bennett S. Kalafut Thermo Fisher Scientifi c, San Jose, California, USA

Po

ster No

te 64

44

0

Automatic optimization of MS and MS/MS assays with dilute samples or weak transitions. Bennett S. Kalafut Thermo Fisher Scientific, San Jose, California, USA

Conclusion Using kernel intensity function estimation, it is possible to tune MS/MS collision

energies using a weak signal consisting of several hundred to as few as (on average) 35 single-ion arrival events.

More application-specific methods of kernel bandwidth selection (requiring novel mathematical development) will allow even greater accuracy in recovery of tuning and breakdown curves. This is an area of ongoing research.

Version 2.0 of the TSQ Endura and TSQ Quantiva MS instrument control software uses the optimization method used to generate the data in this presentation when running compound optimization or QuickQuan™.

A preliminary version of this method was used for data cleanup during compound optimization as implemented in version 1.1 of the TSQ Endura and TSQ Quantiva MS control software.

References 1. M.P. Wand and M.C. Jones. Kernel Smoothing. CRC Press, 1994.

Acknowledgements The author thanks Mark Dreyer (Thermo Fisher Scientific, San Jose, CA) for providing samples for testing this method, including the stock solutions used to prepare reagents for this poster.

Overview Purpose: To optimally tune an MS/MS or single ion monitoring (SIM) assay using a dilute sample or weak signal.

Methods: Kernel intensity function estimation is used to recover the relationship between intensity and tuning from a noisy raw signal.

Results: Optimal collision energies can be recovered from full range collision energy scans even if the signal consists of as few as ten to thirty single-ion events registered at the detector.

Introduction SIM and MS/MS assays must typically be optimized for each compound to be tested. Optimal tunings of ion optics devices like the S-lens (SRIG or ion funnel) can vary from compound to compound. CID collision energy tuning, especially, must be tuned for each MS/MS assay since it is particular to each transition.

When the concentration of the compound to be optimized is limited due to cost or scarcity or the informative CID transitions are weak, generation of smooth breakdown or tuning curves with clear optima can be costly in time and resources.

The method presented here uses kernel intensity function estimation, a statistical algorithm related to kernel density estimation, to optimize collision energies or other assay parameters using signals that are very sparse, resulting from the detection of as few as tens of ions even in the neighborhood of the optimal instrument tuning.

Methods Sample Preparation

Testosterone and estradiol samples were prepared by serial dilution in 50% methanol, 50% water, 0.1% formic acid from 1 mg/mL stock solutions.

Mass Spectrometry

Samples were infused to the HESI source of a Thermo Scientific™ TSQ Endura™ mass spectrometer at a rate of 5 µL per minute. Ionization voltage was set to 3500 V. Sheath and auxiliary gas flow were set using the mass spectrometer’s source optimization routines.

Data Analysis

Kernel intensity function estimation was performed on-instrument using algorithms coded for use in compound optimization in version 2.0 of the control software of the TSQ Endura MS and Thermo Scientific™ TSQ Quantiva™ mass spectrometer. Further analysis of the results was performed offline using the SciPy statistics library (www.scipy.org).

FIGURE 1. At low signal intensities, observed tuning curves become trains of discrete single- or few-ion arrival events.

FIGURE 6. Breakdown curve for 289 to 97 amu CID transition of testosterone.

Figure 7. Breakdown curve for 289 to 109 amu CID transition of testosterone.

At high signal strengths, kernel intensity function estimation becomes a kind of nonparametric regression, similar to LOWESS, kernel-weighted local regression, or spline regression: a way to fit a smooth curve to the noisy data without making strong assumptions about shape or the source of the noise. For consistency, we use kernel estimation across the range of signal intensities.

© 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

FIGURE 2. A sample kernel intensity estimate. For each ion arrival event, represented by the green X, a kernel (blue curves) is added to the reconstructed tuning curve (red curve.)

Here, the kernel bandwidth is an arbitrary 10 units wide. In practice, the kernel bandwidth is set by considering the statistical structure of the data.

The problem (kernal intensity function estimation):

Given steady production of ions at the source and stable filtration of ions by quadrupole mass analyzers, arrival of ions at a triple quadrupole’s mass detector is approximately a Poisson process and the detector signal itself obeys Polya statistics. The relative error varies with intensity and when the ion current is weak the observed signal is often 0: the signal is a series of discrete single- or few-ion arrivals (Fig. 1).

Complicating matters further, the relationship between collision energy or instrument tuning voltage (“tuning curve”) and signal intensity is not a simple function consistent from compound to compound. We cannot simply fit a regression curve to the data even taking into account the non-homoscedastic noise of the signal; we do not know a priori what kind of curve to fit.

Results

FIGURE 3. Kernel intensity estimation applied to a strong signal (here, 1 µg/ml estradiol, in positive mode) is a kind of nonparametric regression or smoothing. technique.

FIGURE 5. Summary of the result of 250 collision energy optimizations for the 97 amu and 109 amu CID transitions of testosterone. (1 pg/ml solution, S-Lens detuned to produce low, spiky signal.)

Optimal collision energy was determined by analysis of a 1 µg/ml solution with a well-tuned S-lens.

Mean number of events per trial: 97 amu: 54.4 109 amu: 34.1

Analysis of the raw detector signal results in a systematic bias away from the optimal collision energy, resulting from the asymmetric shape of the breakdown curve. (Fig. 6-7) Kernel-based optimization removes this bias.

FIGURE 4. Reference breakdown curve for 289 to 109 amu CID transition of testosterone (dark blue) and three breakdown curves estimated from approximately 500 detected ions apiece (red, indigo, and green).

As seen in figure 4, kernel estimation based on several hundred detected ions across a full-range collision energy scan can recover the coarse features of a breakdown curve. Differences in the recovered optimum, at their extreme (purple curve), correspond to 10% weaker signal than the true optimum.

The relatively broad shape of the breakdown curve for the 289 to 109 amu transition (and 289 to 97 amu transition) of testosterone makes it a somewhat challenging case due to the high probability of detecting ions far away from the peak.

Below (figure 5) we see that this method allows some degree of optimization with tenfold fewer detected ions.

Automatic optimization of MS and MS/MS assays with dilute samples or weak transitions. Bennett S. Kalafut Thermo Fisher Scientific, San Jose, California, USA

Conclusion Using kernel intensity function estimation, it is possible to tune MS/MS collision

energies using a weak signal consisting of several hundred to as few as (on average) 35 single-ion arrival events.

More application-specific methods of kernel bandwidth selection (requiring novel mathematical development) will allow even greater accuracy in recovery of tuning and breakdown curves. This is an area of ongoing research.

Version 2.0 of the TSQ Endura and TSQ Quantiva MS instrument control software uses the optimization method used to generate the data in this presentation when running compound optimization or QuickQuan™.

A preliminary version of this method was used for data cleanup during compound optimization as implemented in version 1.1 of the TSQ Endura and TSQ Quantiva MS control software.

References 1. M.P. Wand and M.C. Jones. Kernel Smoothing. CRC Press, 1994.

Acknowledgements The author thanks Mark Dreyer (Thermo Fisher Scientific, San Jose, CA) for providing samples for testing this method, including the stock solutions used to prepare reagents for this poster.

Overview Purpose: To optimally tune an MS/MS or single ion monitoring (SIM) assay using a dilute sample or weak signal.

Methods: Kernel intensity function estimation is used to recover the relationship between intensity and tuning from a noisy raw signal.

Results: Optimal collision energies can be recovered from full range collision energy scans even if the signal consists of as few as ten to thirty single-ion events registered at the detector.

Introduction SIM and MS/MS assays must typically be optimized for each compound to be tested. Optimal tunings of ion optics devices like the S-lens (SRIG or ion funnel) can vary from compound to compound. CID collision energy tuning, especially, must be tuned for each MS/MS assay since it is particular to each transition.

When the concentration of the compound to be optimized is limited due to cost or scarcity or the informative CID transitions are weak, generation of smooth breakdown or tuning curves with clear optima can be costly in time and resources.

The method presented here uses kernel intensity function estimation, a statistical algorithm related to kernel density estimation, to optimize collision energies or other assay parameters using signals that are very sparse, resulting from the detection of as few as tens of ions even in the neighborhood of the optimal instrument tuning.

Methods Sample Preparation

Testosterone and estradiol samples were prepared by serial dilution in 50% methanol, 50% water, 0.1% formic acid from 1 mg/mL stock solutions.

Mass Spectrometry

Samples were infused to the HESI source of a Thermo Scientific™ TSQ Endura™ mass spectrometer at a rate of 5 µL per minute. Ionization voltage was set to 3500 V. Sheath and auxiliary gas flow were set using the mass spectrometer’s source optimization routines.

Data Analysis

Kernel intensity function estimation was performed on-instrument using algorithms coded for use in compound optimization in version 2.0 of the control software of the TSQ Endura MS and Thermo Scientific™ TSQ Quantiva™ mass spectrometer. Further analysis of the results was performed offline using the SciPy statistics library (www.scipy.org).

FIGURE 1. At low signal intensities, observed tuning curves become trains of discrete single- or few-ion arrival events.

FIGURE 6. Breakdown curve for 289 to 97 amu CID transition of testosterone.

Figure 7. Breakdown curve for 289 to 109 amu CID transition of testosterone.

At high signal strengths, kernel intensity function estimation becomes a kind of nonparametric regression, similar to LOWESS, kernel-weighted local regression, or spline regression: a way to fit a smooth curve to the noisy data without making strong assumptions about shape or the source of the noise. For consistency, we use kernel estimation across the range of signal intensities.

© 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

FIGURE 2. A sample kernel intensity estimate. For each ion arrival event, represented by the green X, a kernel (blue curves) is added to the reconstructed tuning curve (red curve.)

Here, the kernel bandwidth is an arbitrary 10 units wide. In practice, the kernel bandwidth is set by considering the statistical structure of the data.

The problem (kernal intensity function estimation):

Given steady production of ions at the source and stable filtration of ions by quadrupole mass analyzers, arrival of ions at a triple quadrupole’s mass detector is approximately a Poisson process and the detector signal itself obeys Polya statistics. The relative error varies with intensity and when the ion current is weak the observed signal is often 0: the signal is a series of discrete single- or few-ion arrivals (Fig. 1).

Complicating matters further, the relationship between collision energy or instrument tuning voltage (“tuning curve”) and signal intensity is not a simple function consistent from compound to compound. We cannot simply fit a regression curve to the data even taking into account the non-homoscedastic noise of the signal; we do not know a priori what kind of curve to fit.

Results

FIGURE 3. Kernel intensity estimation applied to a strong signal (here, 1 µg/ml estradiol, in positive mode) is a kind of nonparametric regression or smoothing. technique.

FIGURE 5. Summary of the result of 250 collision energy optimizations for the 97 amu and 109 amu CID transitions of testosterone. (1 pg/ml solution, S-Lens detuned to produce low, spiky signal.)

Optimal collision energy was determined by analysis of a 1 µg/ml solution with a well-tuned S-lens.

Mean number of events per trial: 97 amu: 54.4 109 amu: 34.1

Analysis of the raw detector signal results in a systematic bias away from the optimal collision energy, resulting from the asymmetric shape of the breakdown curve. (Fig. 6-7) Kernel-based optimization removes this bias.

FIGURE 4. Reference breakdown curve for 289 to 109 amu CID transition of testosterone (dark blue) and three breakdown curves estimated from approximately 500 detected ions apiece (red, indigo, and green).

As seen in figure 4, kernel estimation based on several hundred detected ions across a full-range collision energy scan can recover the coarse features of a breakdown curve. Differences in the recovered optimum, at their extreme (purple curve), correspond to 10% weaker signal than the true optimum.

The relatively broad shape of the breakdown curve for the 289 to 109 amu transition (and 289 to 97 amu transition) of testosterone makes it a somewhat challenging case due to the high probability of detecting ions far away from the peak.

Below (figure 5) we see that this method allows some degree of optimization with tenfold fewer detected ions.

Automatic optimization of MS and MS/MS assays with dilute samples or weak transitions. Bennett S. Kalafut Thermo Fisher Scientific, San Jose, California, USA

Conclusion Using kernel intensity function estimation, it is possible to tune MS/MS collision

energies using a weak signal consisting of several hundred to as few as (on average) 35 single-ion arrival events.

More application-specific methods of kernel bandwidth selection (requiring novel mathematical development) will allow even greater accuracy in recovery of tuning and breakdown curves. This is an area of ongoing research.

Version 2.0 of the TSQ Endura and TSQ Quantiva MS instrument control software uses the optimization method used to generate the data in this presentation when running compound optimization or QuickQuan™.

A preliminary version of this method was used for data cleanup during compound optimization as implemented in version 1.1 of the TSQ Endura and TSQ Quantiva MS control software.

References 1. M.P. Wand and M.C. Jones. Kernel Smoothing. CRC Press, 1994.

Acknowledgements The author thanks Mark Dreyer (Thermo Fisher Scientific, San Jose, CA) for providing samples for testing this method, including the stock solutions used to prepare reagents for this poster.

Overview Purpose: To optimally tune an MS/MS or single ion monitoring (SIM) assay using a dilute sample or weak signal.

Methods: Kernel intensity function estimation is used to recover the relationship between intensity and tuning from a noisy raw signal.

Results: Optimal collision energies can be recovered from full range collision energy scans even if the signal consists of as few as ten to thirty single-ion events registered at the detector.

Introduction SIM and MS/MS assays must typically be optimized for each compound to be tested. Optimal tunings of ion optics devices like the S-lens (SRIG or ion funnel) can vary from compound to compound. CID collision energy tuning, especially, must be tuned for each MS/MS assay since it is particular to each transition.

When the concentration of the compound to be optimized is limited due to cost or scarcity or the informative CID transitions are weak, generation of smooth breakdown or tuning curves with clear optima can be costly in time and resources.

The method presented here uses kernel intensity function estimation, a statistical algorithm related to kernel density estimation, to optimize collision energies or other assay parameters using signals that are very sparse, resulting from the detection of as few as tens of ions even in the neighborhood of the optimal instrument tuning.

Methods Sample Preparation

Testosterone and estradiol samples were prepared by serial dilution in 50% methanol, 50% water, 0.1% formic acid from 1 mg/mL stock solutions.

Mass Spectrometry

Samples were infused to the HESI source of a Thermo Scientific™ TSQ Endura™ mass spectrometer at a rate of 5 µL per minute. Ionization voltage was set to 3500 V. Sheath and auxiliary gas flow were set using the mass spectrometer’s source optimization routines.

Data Analysis

Kernel intensity function estimation was performed on-instrument using algorithms coded for use in compound optimization in version 2.0 of the control software of the TSQ Endura MS and Thermo Scientific™ TSQ Quantiva™ mass spectrometer. Further analysis of the results was performed offline using the SciPy statistics library (www.scipy.org).

FIGURE 1. At low signal intensities, observed tuning curves become trains of discrete single- or few-ion arrival events.

FIGURE 6. Breakdown curve for 289 to 97 amu CID transition of testosterone.

Figure 7. Breakdown curve for 289 to 109 amu CID transition of testosterone.

At high signal strengths, kernel intensity function estimation becomes a kind of nonparametric regression, similar to LOWESS, kernel-weighted local regression, or spline regression: a way to fit a smooth curve to the noisy data without making strong assumptions about shape or the source of the noise. For consistency, we use kernel estimation across the range of signal intensities.

© 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

FIGURE 2. A sample kernel intensity estimate. For each ion arrival event, represented by the green X, a kernel (blue curves) is added to the reconstructed tuning curve (red curve.)

Here, the kernel bandwidth is an arbitrary 10 units wide. In practice, the kernel bandwidth is set by considering the statistical structure of the data.

The problem (kernal intensity function estimation):

Given steady production of ions at the source and stable filtration of ions by quadrupole mass analyzers, arrival of ions at a triple quadrupole’s mass detector is approximately a Poisson process and the detector signal itself obeys Polya statistics. The relative error varies with intensity and when the ion current is weak the observed signal is often 0: the signal is a series of discrete single- or few-ion arrivals (Fig. 1).

Complicating matters further, the relationship between collision energy or instrument tuning voltage (“tuning curve”) and signal intensity is not a simple function consistent from compound to compound. We cannot simply fit a regression curve to the data even taking into account the non-homoscedastic noise of the signal; we do not know a priori what kind of curve to fit.

Results

FIGURE 3. Kernel intensity estimation applied to a strong signal (here, 1 µg/ml estradiol, in positive mode) is a kind of nonparametric regression or smoothing. technique.

FIGURE 5. Summary of the result of 250 collision energy optimizations for the 97 amu and 109 amu CID transitions of testosterone. (1 pg/ml solution, S-Lens detuned to produce low, spiky signal.)

Optimal collision energy was determined by analysis of a 1 µg/ml solution with a well-tuned S-lens.

Mean number of events per trial: 97 amu: 54.4 109 amu: 34.1

Analysis of the raw detector signal results in a systematic bias away from the optimal collision energy, resulting from the asymmetric shape of the breakdown curve. (Fig. 6-7) Kernel-based optimization removes this bias.

FIGURE 4. Reference breakdown curve for 289 to 109 amu CID transition of testosterone (dark blue) and three breakdown curves estimated from approximately 500 detected ions apiece (red, indigo, and green).

As seen in figure 4, kernel estimation based on several hundred detected ions across a full-range collision energy scan can recover the coarse features of a breakdown curve. Differences in the recovered optimum, at their extreme (purple curve), correspond to 10% weaker signal than the true optimum.

The relatively broad shape of the breakdown curve for the 289 to 109 amu transition (and 289 to 97 amu transition) of testosterone makes it a somewhat challenging case due to the high probability of detecting ions far away from the peak.

Below (figure 5) we see that this method allows some degree of optimization with tenfold fewer detected ions.

Page 2: Automatic Optimizaiton of MS and MS/MS Assays with Dilute ...tools.thermofisher.com/content/sfs/posters/PN... · Automatic Optimizaiton of MS and MS/MS Assays with Dilute Samples

2 Automatic Optimizaiton of MS and MS/MS Assays with Dilute Samples or Weak Transitions

Automatic optimization of MS and MS/MS assays with dilute samples or weak transitions. Bennett S. Kalafut Thermo Fisher Scientific, San Jose, California, USA

Conclusion Using kernel intensity function estimation, it is possible to tune MS/MS collision

energies using a weak signal consisting of several hundred to as few as (on average) 35 single-ion arrival events.

More application-specific methods of kernel bandwidth selection (requiring novel mathematical development) will allow even greater accuracy in recovery of tuning and breakdown curves. This is an area of ongoing research.

Version 2.0 of the TSQ Endura and TSQ Quantiva MS instrument control software uses the optimization method used to generate the data in this presentation when running compound optimization or QuickQuan™.

A preliminary version of this method was used for data cleanup during compound optimization as implemented in version 1.1 of the TSQ Endura and TSQ Quantiva MS control software.

References 1. M.P. Wand and M.C. Jones. Kernel Smoothing. CRC Press, 1994.

Acknowledgements The author thanks Mark Dreyer (Thermo Fisher Scientific, San Jose, CA) for providing samples for testing this method, including the stock solutions used to prepare reagents for this poster.

Overview Purpose: To optimally tune an MS/MS or single ion monitoring (SIM) assay using a dilute sample or weak signal.

Methods: Kernel intensity function estimation is used to recover the relationship between intensity and tuning from a noisy raw signal.

Results: Optimal collision energies can be recovered from full range collision energy scans even if the signal consists of as few as ten to thirty single-ion events registered at the detector.

Introduction SIM and MS/MS assays must typically be optimized for each compound to be tested. Optimal tunings of ion optics devices like the S-lens (SRIG or ion funnel) can vary from compound to compound. CID collision energy tuning, especially, must be tuned for each MS/MS assay since it is particular to each transition.

When the concentration of the compound to be optimized is limited due to cost or scarcity or the informative CID transitions are weak, generation of smooth breakdown or tuning curves with clear optima can be costly in time and resources.

The method presented here uses kernel intensity function estimation, a statistical algorithm related to kernel density estimation, to optimize collision energies or other assay parameters using signals that are very sparse, resulting from the detection of as few as tens of ions even in the neighborhood of the optimal instrument tuning.

Methods Sample Preparation

Testosterone and estradiol samples were prepared by serial dilution in 50% methanol, 50% water, 0.1% formic acid from 1 mg/mL stock solutions.

Mass Spectrometry

Samples were infused to the HESI source of a Thermo Scientific™ TSQ Endura™ mass spectrometer at a rate of 5 µL per minute. Ionization voltage was set to 3500 V. Sheath and auxiliary gas flow were set using the mass spectrometer’s source optimization routines.

Data Analysis

Kernel intensity function estimation was performed on-instrument using algorithms coded for use in compound optimization in version 2.0 of the control software of the TSQ Endura MS and Thermo Scientific™ TSQ Quantiva™ mass spectrometer. Further analysis of the results was performed offline using the SciPy statistics library (www.scipy.org).

FIGURE 1. At low signal intensities, observed tuning curves become trains of discrete single- or few-ion arrival events.

FIGURE 6. Breakdown curve for 289 to 97 amu CID transition of testosterone.

Figure 7. Breakdown curve for 289 to 109 amu CID transition of testosterone.

At high signal strengths, kernel intensity function estimation becomes a kind of nonparametric regression, similar to LOWESS, kernel-weighted local regression, or spline regression: a way to fit a smooth curve to the noisy data without making strong assumptions about shape or the source of the noise. For consistency, we use kernel estimation across the range of signal intensities.

© 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

FIGURE 2. A sample kernel intensity estimate. For each ion arrival event, represented by the green X, a kernel (blue curves) is added to the reconstructed tuning curve (red curve.)

Here, the kernel bandwidth is an arbitrary 10 units wide. In practice, the kernel bandwidth is set by considering the statistical structure of the data.

The problem (kernal intensity function estimation):

Given steady production of ions at the source and stable filtration of ions by quadrupole mass analyzers, arrival of ions at a triple quadrupole’s mass detector is approximately a Poisson process and the detector signal itself obeys Polya statistics. The relative error varies with intensity and when the ion current is weak the observed signal is often 0: the signal is a series of discrete single- or few-ion arrivals (Fig. 1).

Complicating matters further, the relationship between collision energy or instrument tuning voltage (“tuning curve”) and signal intensity is not a simple function consistent from compound to compound. We cannot simply fit a regression curve to the data even taking into account the non-homoscedastic noise of the signal; we do not know a priori what kind of curve to fit.

Results

FIGURE 3. Kernel intensity estimation applied to a strong signal (here, 1 µg/ml estradiol, in positive mode) is a kind of nonparametric regression or smoothing. technique.

FIGURE 5. Summary of the result of 250 collision energy optimizations for the 97 amu and 109 amu CID transitions of testosterone. (1 pg/ml solution, S-Lens detuned to produce low, spiky signal.)

Optimal collision energy was determined by analysis of a 1 µg/ml solution with a well-tuned S-lens.

Mean number of events per trial: 97 amu: 54.4 109 amu: 34.1

Analysis of the raw detector signal results in a systematic bias away from the optimal collision energy, resulting from the asymmetric shape of the breakdown curve. (Fig. 6-7) Kernel-based optimization removes this bias.

FIGURE 4. Reference breakdown curve for 289 to 109 amu CID transition of testosterone (dark blue) and three breakdown curves estimated from approximately 500 detected ions apiece (red, indigo, and green).

As seen in figure 4, kernel estimation based on several hundred detected ions across a full-range collision energy scan can recover the coarse features of a breakdown curve. Differences in the recovered optimum, at their extreme (purple curve), correspond to 10% weaker signal than the true optimum.

The relatively broad shape of the breakdown curve for the 289 to 109 amu transition (and 289 to 97 amu transition) of testosterone makes it a somewhat challenging case due to the high probability of detecting ions far away from the peak.

Below (figure 5) we see that this method allows some degree of optimization with tenfold fewer detected ions.

Automatic optimization of MS and MS/MS assays with dilute samples or weak transitions. Bennett S. Kalafut Thermo Fisher Scientific, San Jose, California, USA

Conclusion Using kernel intensity function estimation, it is possible to tune MS/MS collision

energies using a weak signal consisting of several hundred to as few as (on average) 35 single-ion arrival events.

More application-specific methods of kernel bandwidth selection (requiring novel mathematical development) will allow even greater accuracy in recovery of tuning and breakdown curves. This is an area of ongoing research.

Version 2.0 of the TSQ Endura and TSQ Quantiva MS instrument control software uses the optimization method used to generate the data in this presentation when running compound optimization or QuickQuan™.

A preliminary version of this method was used for data cleanup during compound optimization as implemented in version 1.1 of the TSQ Endura and TSQ Quantiva MS control software.

References 1. M.P. Wand and M.C. Jones. Kernel Smoothing. CRC Press, 1994.

Acknowledgements The author thanks Mark Dreyer (Thermo Fisher Scientific, San Jose, CA) for providing samples for testing this method, including the stock solutions used to prepare reagents for this poster.

Overview Purpose: To optimally tune an MS/MS or single ion monitoring (SIM) assay using a dilute sample or weak signal.

Methods: Kernel intensity function estimation is used to recover the relationship between intensity and tuning from a noisy raw signal.

Results: Optimal collision energies can be recovered from full range collision energy scans even if the signal consists of as few as ten to thirty single-ion events registered at the detector.

Introduction SIM and MS/MS assays must typically be optimized for each compound to be tested. Optimal tunings of ion optics devices like the S-lens (SRIG or ion funnel) can vary from compound to compound. CID collision energy tuning, especially, must be tuned for each MS/MS assay since it is particular to each transition.

When the concentration of the compound to be optimized is limited due to cost or scarcity or the informative CID transitions are weak, generation of smooth breakdown or tuning curves with clear optima can be costly in time and resources.

The method presented here uses kernel intensity function estimation, a statistical algorithm related to kernel density estimation, to optimize collision energies or other assay parameters using signals that are very sparse, resulting from the detection of as few as tens of ions even in the neighborhood of the optimal instrument tuning.

Methods Sample Preparation

Testosterone and estradiol samples were prepared by serial dilution in 50% methanol, 50% water, 0.1% formic acid from 1 mg/mL stock solutions.

Mass Spectrometry

Samples were infused to the HESI source of a Thermo Scientific™ TSQ Endura™ mass spectrometer at a rate of 5 µL per minute. Ionization voltage was set to 3500 V. Sheath and auxiliary gas flow were set using the mass spectrometer’s source optimization routines.

Data Analysis

Kernel intensity function estimation was performed on-instrument using algorithms coded for use in compound optimization in version 2.0 of the control software of the TSQ Endura MS and Thermo Scientific™ TSQ Quantiva™ mass spectrometer. Further analysis of the results was performed offline using the SciPy statistics library (www.scipy.org).

FIGURE 1. At low signal intensities, observed tuning curves become trains of discrete single- or few-ion arrival events.

FIGURE 6. Breakdown curve for 289 to 97 amu CID transition of testosterone.

Figure 7. Breakdown curve for 289 to 109 amu CID transition of testosterone.

At high signal strengths, kernel intensity function estimation becomes a kind of nonparametric regression, similar to LOWESS, kernel-weighted local regression, or spline regression: a way to fit a smooth curve to the noisy data without making strong assumptions about shape or the source of the noise. For consistency, we use kernel estimation across the range of signal intensities.

© 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

FIGURE 2. A sample kernel intensity estimate. For each ion arrival event, represented by the green X, a kernel (blue curves) is added to the reconstructed tuning curve (red curve.)

Here, the kernel bandwidth is an arbitrary 10 units wide. In practice, the kernel bandwidth is set by considering the statistical structure of the data.

The problem (kernal intensity function estimation):

Given steady production of ions at the source and stable filtration of ions by quadrupole mass analyzers, arrival of ions at a triple quadrupole’s mass detector is approximately a Poisson process and the detector signal itself obeys Polya statistics. The relative error varies with intensity and when the ion current is weak the observed signal is often 0: the signal is a series of discrete single- or few-ion arrivals (Fig. 1).

Complicating matters further, the relationship between collision energy or instrument tuning voltage (“tuning curve”) and signal intensity is not a simple function consistent from compound to compound. We cannot simply fit a regression curve to the data even taking into account the non-homoscedastic noise of the signal; we do not know a priori what kind of curve to fit.

Results

FIGURE 3. Kernel intensity estimation applied to a strong signal (here, 1 µg/ml estradiol, in positive mode) is a kind of nonparametric regression or smoothing. technique.

FIGURE 5. Summary of the result of 250 collision energy optimizations for the 97 amu and 109 amu CID transitions of testosterone. (1 pg/ml solution, S-Lens detuned to produce low, spiky signal.)

Optimal collision energy was determined by analysis of a 1 µg/ml solution with a well-tuned S-lens.

Mean number of events per trial: 97 amu: 54.4 109 amu: 34.1

Analysis of the raw detector signal results in a systematic bias away from the optimal collision energy, resulting from the asymmetric shape of the breakdown curve. (Fig. 6-7) Kernel-based optimization removes this bias.

FIGURE 4. Reference breakdown curve for 289 to 109 amu CID transition of testosterone (dark blue) and three breakdown curves estimated from approximately 500 detected ions apiece (red, indigo, and green).

As seen in figure 4, kernel estimation based on several hundred detected ions across a full-range collision energy scan can recover the coarse features of a breakdown curve. Differences in the recovered optimum, at their extreme (purple curve), correspond to 10% weaker signal than the true optimum.

The relatively broad shape of the breakdown curve for the 289 to 109 amu transition (and 289 to 97 amu transition) of testosterone makes it a somewhat challenging case due to the high probability of detecting ions far away from the peak.

Below (figure 5) we see that this method allows some degree of optimization with tenfold fewer detected ions.

Page 3: Automatic Optimizaiton of MS and MS/MS Assays with Dilute ...tools.thermofisher.com/content/sfs/posters/PN... · Automatic Optimizaiton of MS and MS/MS Assays with Dilute Samples

PN64440-EN 0615S

Automatic optimization of MS and MS/MS assays with dilute samples or weak transitions. Bennett S. Kalafut Thermo Fisher Scientific, San Jose, California, USA

Conclusion Using kernel intensity function estimation, it is possible to tune MS/MS collision

energies using a weak signal consisting of several hundred to as few as (on average) 35 single-ion arrival events.

More application-specific methods of kernel bandwidth selection (requiring novel mathematical development) will allow even greater accuracy in recovery of tuning and breakdown curves. This is an area of ongoing research.

Version 2.0 of the TSQ Endura and TSQ Quantiva MS instrument control software uses the optimization method used to generate the data in this presentation when running compound optimization or QuickQuan™.

A preliminary version of this method was used for data cleanup during compound optimization as implemented in version 1.1 of the TSQ Endura and TSQ Quantiva MS control software.

References 1. M.P. Wand and M.C. Jones. Kernel Smoothing. CRC Press, 1994.

Acknowledgements The author thanks Mark Dreyer (Thermo Fisher Scientific, San Jose, CA) for providing samples for testing this method, including the stock solutions used to prepare reagents for this poster.

Overview Purpose: To optimally tune an MS/MS or single ion monitoring (SIM) assay using a dilute sample or weak signal.

Methods: Kernel intensity function estimation is used to recover the relationship between intensity and tuning from a noisy raw signal.

Results: Optimal collision energies can be recovered from full range collision energy scans even if the signal consists of as few as ten to thirty single-ion events registered at the detector.

Introduction SIM and MS/MS assays must typically be optimized for each compound to be tested. Optimal tunings of ion optics devices like the S-lens (SRIG or ion funnel) can vary from compound to compound. CID collision energy tuning, especially, must be tuned for each MS/MS assay since it is particular to each transition.

When the concentration of the compound to be optimized is limited due to cost or scarcity or the informative CID transitions are weak, generation of smooth breakdown or tuning curves with clear optima can be costly in time and resources.

The method presented here uses kernel intensity function estimation, a statistical algorithm related to kernel density estimation, to optimize collision energies or other assay parameters using signals that are very sparse, resulting from the detection of as few as tens of ions even in the neighborhood of the optimal instrument tuning.

Methods Sample Preparation

Testosterone and estradiol samples were prepared by serial dilution in 50% methanol, 50% water, 0.1% formic acid from 1 mg/mL stock solutions.

Mass Spectrometry

Samples were infused to the HESI source of a Thermo Scientific™ TSQ Endura™ mass spectrometer at a rate of 5 µL per minute. Ionization voltage was set to 3500 V. Sheath and auxiliary gas flow were set using the mass spectrometer’s source optimization routines.

Data Analysis

Kernel intensity function estimation was performed on-instrument using algorithms coded for use in compound optimization in version 2.0 of the control software of the TSQ Endura MS and Thermo Scientific™ TSQ Quantiva™ mass spectrometer. Further analysis of the results was performed offline using the SciPy statistics library (www.scipy.org).

FIGURE 1. At low signal intensities, observed tuning curves become trains of discrete single- or few-ion arrival events.

FIGURE 6. Breakdown curve for 289 to 97 amu CID transition of testosterone.

Figure 7. Breakdown curve for 289 to 109 amu CID transition of testosterone.

At high signal strengths, kernel intensity function estimation becomes a kind of nonparametric regression, similar to LOWESS, kernel-weighted local regression, or spline regression: a way to fit a smooth curve to the noisy data without making strong assumptions about shape or the source of the noise. For consistency, we use kernel estimation across the range of signal intensities.

© 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

FIGURE 2. A sample kernel intensity estimate. For each ion arrival event, represented by the green X, a kernel (blue curves) is added to the reconstructed tuning curve (red curve.)

Here, the kernel bandwidth is an arbitrary 10 units wide. In practice, the kernel bandwidth is set by considering the statistical structure of the data.

The problem (kernal intensity function estimation):

Given steady production of ions at the source and stable filtration of ions by quadrupole mass analyzers, arrival of ions at a triple quadrupole’s mass detector is approximately a Poisson process and the detector signal itself obeys Polya statistics. The relative error varies with intensity and when the ion current is weak the observed signal is often 0: the signal is a series of discrete single- or few-ion arrivals (Fig. 1).

Complicating matters further, the relationship between collision energy or instrument tuning voltage (“tuning curve”) and signal intensity is not a simple function consistent from compound to compound. We cannot simply fit a regression curve to the data even taking into account the non-homoscedastic noise of the signal; we do not know a priori what kind of curve to fit.

Results

FIGURE 3. Kernel intensity estimation applied to a strong signal (here, 1 µg/ml estradiol, in positive mode) is a kind of nonparametric regression or smoothing. technique.

FIGURE 5. Summary of the result of 250 collision energy optimizations for the 97 amu and 109 amu CID transitions of testosterone. (1 pg/ml solution, S-Lens detuned to produce low, spiky signal.)

Optimal collision energy was determined by analysis of a 1 µg/ml solution with a well-tuned S-lens.

Mean number of events per trial: 97 amu: 54.4 109 amu: 34.1

Analysis of the raw detector signal results in a systematic bias away from the optimal collision energy, resulting from the asymmetric shape of the breakdown curve. (Fig. 6-7) Kernel-based optimization removes this bias.

FIGURE 4. Reference breakdown curve for 289 to 109 amu CID transition of testosterone (dark blue) and three breakdown curves estimated from approximately 500 detected ions apiece (red, indigo, and green).

As seen in figure 4, kernel estimation based on several hundred detected ions across a full-range collision energy scan can recover the coarse features of a breakdown curve. Differences in the recovered optimum, at their extreme (purple curve), correspond to 10% weaker signal than the true optimum.

The relatively broad shape of the breakdown curve for the 289 to 109 amu transition (and 289 to 97 amu transition) of testosterone makes it a somewhat challenging case due to the high probability of detecting ions far away from the peak.

Below (figure 5) we see that this method allows some degree of optimization with tenfold fewer detected ions.

Africa +43 1 333 50 34 0Australia +61 3 9757 4300Austria +43 810 282 206Belgium +32 53 73 42 41Canada +1 800 530 8447China 800 810 5118 (free call domestic)

400 650 5118

Denmark +45 70 23 62 60Europe-Other +43 1 333 50 34 0Finland +358 10 3292 200France +33 1 60 92 48 00Germany +49 6103 408 1014India +91 22 6742 9494Italy +39 02 950 591

Japan +81 45 453 9100Korea +82 2 3420 8600Latin America +1 561 688 8700Middle East +43 1 333 50 34 0Netherlands +31 76 579 55 55New Zealand +64 9 980 6700Norway +46 8 556 468 00

Russia/CIS +43 1 333 50 34 0Singapore +65 6289 1190Spain +34 914 845 965Sweden +46 8 556 468 00Switzerland +41 61 716 77 00UK +44 1442 233555USA +1 800 532 4752

www.thermoscientific.com©2015 Thermo Fisher Scientifi c Inc. All rights reserved. ISO is a trademark of the International Standards Organization. All other trademarks are the property of Thermo Fisher Scientifi c and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientifi c products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifi cations, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.