Upload
others
View
1
Download
0
Embed Size (px)
Citation preview
Automated Syringe Dosing
Hanson, E.E., Weisshaar, C.L., Wentland, A.L.
BME 402 Department of Biomedical Engineering
University of Wisconsin-Madison May 4, 2005
Advisor
Willis J. Tompkins, Ph.D., Professor Department of Biomedical Engineering
Abstract
Many people, especially elderly individuals, have afflictions along with diabetes, including poor
eyesight, tremors, loss of dexterity, arthritis, multiple sclerosis, and other neuromuscular
disorders. These disabilities make it difficult for patients to measure and administer medications
with a syringe. We have developed a device that electronically drives the plunger on a standard
syringe, reducing the dexterity normally needed with dosing a syringe. We have constructed a
prototype that demonstrates the feasibility of driving the syringe on a screw-based system with a
bimodal stepper motor. We have begun to program a microprocessor for controlling a keypad,
motor, and digital display.
Syringe Dosing 2/40
Table of Contents Page
Abstract 1 Table of Contents 2
Problem Statement 3
National Student Design Competition 3
Schoofs Prize for Creativity 3
Background 4
Current Products 6
Contacts 7
Design Constraints 7
Design Approaches 8
Design Evaluation 10
Our Chosen Design 11
Syringe Dosing Device: Operation 12
Prototype 13
Construction of the Prototype 13
Syringe Dosing Device: Advantages 15
Syringe Dosing Device: Disadvantages 15
Syringe Calibration and Preliminary Testing 16
Schoofs Prize for Creativity: Claims and Marketing Information 17
Smaller Motor 19
Microprocessor 20
Future Work 23
Ethical Considerations 23
Conclusions 24
References 25
Appendix A: PDS 28
Appendix B: Expenses 32
Appendix C: BASIC program for counting zero through nine 33
Appendix D: Master BASIC program for the design 37
Syringe Dosing 3/40
Problem Statement
Our goal is to develop a syringe delivery device that uses standard 1 cc syringes (BD
Ultra-FineTM Needles [1]) and sets the dosage on those syringes within 0.01 cc accuracy. This
device should minimize the dexterity typically needed with dosing a syringe, along with being
easy to use for elderly patients and those with poor eyesight. Ideally, this machine is geared
towards patients who use syringes on a daily basis, such as those diagnosed with diabetes.
Patients afflicted with maladies in addition to diabetes, such as neuromuscular disorders and poor
eyesight, would also benefit greatly from this device.
National Student Design Competition
This project is part of the 2004-2005 National Student Design Competition, an
undergraduate competition funded by the Rehabilitation Engineering Research Center on
Accessible Medical Instrumentation [2]. In conjunction with Marquette University, Professor
John Enderle of the University of Connecticut chairs the competition. The competition is open
to students predominantly in biomedical engineering and industrial design. In the 2004-2005
competition, student teams have a choice of three projects: a weight scale, a syringe dosing
device, or an ergometer, all to assist patients of diabetes, obesity, paralysis, and neuromuscular
disorders. At the University of Wisconsin – Madison, another team is working on an ergometer
[3] while our team is working on a syringe dosing device.
Schoofs Prize for Creativity
The team also chose to participate in the Schoofs competition. Part of this competition
incorporated market research to develop a marketing plan. As part of this research, we
determined the market, insurance costs and coverage, and future prospects for this type of a
device. Marketing research will be provided later in this paper.
Syringe Dosing 4/40
Background
Diabetes is one of the leading causes of disability in the United States, causing dementia,
low testosterone levels in males, excessive thirst, frequent urination, fatigue, changes in vision,
blindness, stroke, nerve damage, and the need for amputation [4, 5]. Diabetes is typically treated
with insulin injections, where the amount of insulin is determined by eating habits, exercise, and
ultimately a patient’s blood glucose level. While the changes in vision make it difficult for
patients to see and set the dosage on the syringe, elderly diabetes patients commonly acquire
neuromuscular diseases, albeit unrelated to diabetes. Nevertheless, these diseases make it
difficult for patients to control syringes.
Type 2 diabetes is the most common form of diabetes, affecting 18.2 million people in
the United States alone. This type is found in 90-
95% of the diabetes patients [6]. Those most often
afflicted with this disease are older people. The
Center for Disease Control [6] has predicted that the
number of American diabetes patients will increase
as Baby Boomers grow older and more sedentary.
Type 2 diabetes is characterized by a high blood
glucose level, a high insulin level, and a resistance
to insulin (Figure 1). Contrary to type 1 diabetes,
where insulin does not get produced in proper quantities, type 2 diabetics cannot use their insulin
effectively [5].
A person with diabetes must take daily precautions to maintain optimal health. These
include healthy eating, exercise, and blood glucose testing. Blood glucose testing will tell if the
Figure 1. Type 2 Diabetes. In this disease, users develop a resistance to insulin, and therefore, blood glucose levels rise [7].
Syringe Dosing 5/40
blood sugar levels are too low (hypoglycemia) or too high (hyperglycemia) [4]. Irregular blood
sugar levels can cause illness, dizziness, nervousness, confusion, fainting, and/or impaired
judgment [4]. Depending on the blood sugar level, insulin may be administered.
The injection procedure should be done as recommended by an endocrinologist and is as
follows [8]:
1. Wash hands and area where injecting (usually a fatty subcutaneous tissue area such as the stomach).
2. Wash medicine bottle top with cotton ball and alcohol.
3. Draw air into the syringe equal to the volume of fluid needed.
4. Insert needle into medicine bottle and depress plunger, pushing all air out of the syringe. This action creates a vacuum and allows for easier and smoother filling and delivery.
5. Invert medicine bottle and fill syringe, making sure the needle is not exposed to air.
Air bubbles in the barrel will lead to an incorrect amount of medication.
6. Once the syringe has been filled to the proper amount, keep the bottle and syringe upside down and flick the syringe barrel. This moves air bubbles that may have formed at the top of the syringe.
7. Push the plunger to move the air bubbles into the vial.
8. Check medication dosage.
9. Insert needle into skin at a 90-degree angle and deliver medicine at a slow, steady rate. All medication should be released within 5 seconds.
10. Dispose of needle properly.
It is evident that this is a complicated and involved procedure, requiring a fair amount of
dexterity in controlling the syringe and perspicacity in detecting bubbles. This process should be
simplified for the elderly and individuals with neuromuscular disorders and/or poor vision.
Syringe Dosing 6/40
Current Products
Research was done on current devices available to diabetes
patients. Most diabetes patients administer insulin using either an
insulin pump or a syringe. An insulin pump, like the Medtronic
MiniMed [9] (Figure 2), provides a constant supply of fast-acting
insulin throughout the day [10, 11]. Additionally, the device must
be programmed to accommodate eating and exercising habits.
While an insulin pump provides many advantages in terms of convenience, flexibility, and
accuracy, the pump also has some drawbacks. Many patients do not enjoy having a pump
attached to their body. Older patients are often wary of the new, computer-driven technology
and don’t want to feel attached to a device. There are still many users of syringes. However,
proper syringe operation requires concentration, dexterity, and good eyesight.
Many currently available devices assist in administering the medication, not setting the
dosage. Owen Mumford offers a product called the Autoject 2 [12]. To use this device, the user
fills the syringe to the proper level and places the syringe inside the Autoject 2 (Figure 3). With
the touch of a button, a spring pushes the syringe into the patient and administers the medication.
The device does not assist the patient in
filling the syringe. Consequently, the device
does not increase the accuracy of the dosage
or compensate for the user’s disabilities to a
significant degree.
Another device that assists the user is the Novo Nordisk NovoPen® [13] (Figure 4). The
device uses cartridges of insulin and disposable needles. The NovoPen® itself is disposable after
Figure 2. Medtronic MiniMed insulin pump. A user programs this device to deliver fast-acting insulin based on the diet and activity of the user. [9]
Figure 3. Autoject2 syringe device by Owen Mumford [12]. This device assists in injection, but does not assist in setting the dosage on the syringe.
Syringe Dosing 7/40
a month or when the cartridge is empty, whichever occurs first. The
device is used by clicking the top for each unit of insulin. While
patients find this easier to use than a syringe [14], it still takes a fair
amount of dexterity to click the device, and particularly to not
overshoot the desired dosage. Furthermore, if the user needs a
large dosage, the device would require a lot of clicks, making it
cumbersome and liable to losing count.
Contacts
We have accumulated a few contacts concerning different aspects of our project. They
are listed below:
1) Missy Midbon [11] – UW student; diabetes patient using a Medtronic MiniMed pump.
2) Kara Yaeger [15] – diabetes educator and nurse.
3) Doug Haist [14] – a 79-year (2004) old diabetes and Parkinson’s disease patient.
4) Bern Jordan – a graduate student in the Trace Center familiar with universal design.
5) John Puccinelli – a biomedical engineering graduate student with diabetes.
The evolution of our design and first prototype is described below.
Design Constraints
On average, the device will be used twice daily with slow-acting insulin and several more
times with fast-acting insulin to accommodate the eating habits of the user. The device should
function for at least five years under normal use. The device will be designed to accommodate
BD UltraFineTM II 1 cc syringes, which have 80% of the market share [16]. Furthermore, the
device should be easy to use for people with poor eyesight, dexterity, and fine motor control.
Figure 4. Novo Nordisk NovoPen®. This device uses insulin cartridges. The dosage is set by clicking the device for each unit desired. [13]
Syringe Dosing 8/40
Creating a syringe dosing device with universal design in mind is a very important aspect
of this project. The device must be small and portable, yet have features accommodating
presbyopia and individuals with poor neuromuscular control. The device should be easier to use
than the NovoPen® and reduce the amount of interaction a patient has with syringes.
Should a user, and not an insurance company, pay for the device, the device should be
inexpensive, especially compared to the price of insulin pumps (>$5,000). We would like the
design to sell for between $300-$500, with manufacturing costs between $150-$200.
Please refer to Appendix A for our Product Design Specifications, a list of requirements
particular to our final design. See Appendix B for a list of this semester’s expenses.
Design Approaches
We brainstormed two major methods in which we could design a syringe dosing device.
One such way was to have a device that not only controls the syringe, but also an insulin bottle at
the same time. A device like this would be prohibitively large to be considered portable and is
therefore described in the Tabletop portion of this section. Another way of designing the device
would be to control the syringe alone and let the user continue to control the insulin bottle. This
device would be portable and is therefore described in the Portable portion of this section.
Tabletop
With a tabletop device, a user inserts an
insulin bottle and a syringe into the device
(Figure 5). A user would then type in the
needed dosage of insulin. This input will trigger
the microprocessor to move the syringe towards
Figure 5. Tabletop syringe dosing device. The device can hold an insulin bottle and syringe, where a user types in a dosage, the syringe itself moves into the insulin bottle, a motor pulls the plunger of the syringe backwards to draw fluid, and the syringe moves back into its starting position. [17]
Syringe Dosing 9/40
the insulin bottle, then to run a motor that pulls the plunger of the syringe backwards, drawing
fluid into the insulin chamber of the syringe. With that process complete, the syringe will move
back to its starting position where it can be taken out of the device. A user would then remove
the syringe and inject the insulin into the body without assistance.
Advantages: Completely eliminates the dexterity needed to dose a syringe and control an
insulin bottle at the same time.
Disadvantages: This design is not portable and provides no assistance in pushing the
plunger while also inserting the needle into the body. Pushing the plunger for injection takes a
fair amount of dexterity because the thumb must be used while also holding the syringe in place.
The authors acknowledge that it would be possible to combine this device with the Owen
Mumford Autoject 2 to assist in the injection.
Portable
A portable device would be useful for all users. While users claim that it is difficult to
hold an insulin bottle at the same time as holding the syringe [14], the difficulty truly arises when
a user needs to hold the bottle but also maneuver the plunger of the syringe at the same time.
Our supposition is that a user would have no trouble holding an insulin bottle and a device at the
same time, because only a single button would need to be pressed on the device. Therefore, one
hand would hold the bottle while the other would hold the device, pressing a button to cause the
plunger to draw fluid into the chamber of the syringe.
Below are two methods of controlling the plunger of a syringe in a portable device.
Hydraulics-driven: In a hydraulics-driven design, a piston would be connected to the
plunger of the syringe (Figure 6). This piston would be controlled by a hydraulic pump, which
in turn is regulated by a control device, e.g. a microcontroller. While this device would provide
Syringe Dosing 10/40
plenty of power for controlling the syringe, it may be difficult to fit all of
the components in a portable device without making it prohibitively
expensive.
Motor/screw-driven: In a motor and screw-driven design, a
bracket would be attached to the plunger. This bracket would translate
along a screw that rotates according to a motor with or without gears.
While this design doesn’t provide as much power as a hydraulics
system, it would be considerably cheaper to build. It would also be
easier to fit this design into a small portable device. (Please see Figure 8
in Our Chosen Design for a representation of the motor/screw-driven system).
Advantages: The authors consider the portability of this design to be very advantageous.
Setting the dosage on a syringe could then be done anywhere and would not rely on the location,
such as the tabletop design.
Disadvantages: The fact that the user still needs to hold the insulin bottle is a noteworthy
flaw in the design, but is considered outweighed by portability.
Design Evaluation
To determine whether to design a tabletop or portable device, we constructed a design
matrix to quantitatively assess the two categories (Table 1). Although this matrix gives equal
weight to all of the criteria, cost and portability are the most important factors. Nevertheless, the
handheld device achieved a lower score than the tabletop design (where a lower score is more
desirable). Qualitatively, a portable device would be more useful, as it provides assistance in any
environment. However, it must be noted that we still suppose that users will not have trouble
holding a device and insulin bottle simultaneously.
Figure 6. A hydraulics-driven plunger. The plunger of the syringe is controlled with a piston. [17]
Syringe Dosing 11/40
Table 1. Design matrix evaluating handheld and tabletop designs, where a lower score is desirable and each criterion is evaluated on a scale of 1-3. Handheld Tabletop Portability 1 3 Cost 1 2 Ease of Use 1 1 Accuracy 1 1 Total 4 7
To determine whether to control a syringe in a handheld device with a hydraulics-driven
system or a motor/screw-driven system, we constructed a design matrix to quantitatively assess
the two designs (Table 2). Although the matrix gives equal weight to all of the criteria, cost and
simplicity are the most important factors. Nevertheless, the motor/screw-driven design achieved
a lower score than the hydraulics-driven system (where a lower score is more desirable).
Qualitatively, a motor/screw-driven device will be much simpler to construct than a hydraulics-
system.
Table 2. Design matrix evaluating hydraulics-driven and motor/screw-driven designs, where a lower score is desirable and each criterion is evaluated on a scale of 1-3. Hydraulics-Driven Motor/Screw-Driven Cost 3 2 Simplicity 2 1 Reliability 2 1 Size 2 1 Strength 1 2 Total 10 7
Our Chosen Design
Based on the criteria we used to evaluate our designs (Tables 1 & 2), we chose to design
a portable device driven with a motor/screw system (Figure 7). A key factor in choosing this
design was that it not only satisfied the requirements of the National Student Design
Competition, but that it will be easy to manufacture. Additionally, the mechanical components
of the device are very cheap. The operating steps of this device are provided in the following
section.
Syringe Dosing 12/40
Syringe Dosing Device: Operation
Taking the normal procedures to sterilize a syringe, a user will insert the syringe into our
device. The syringe is held in place with collapsible brackets (Figures 7 & 8, shown in red); the
plunger of the syringe will fit into a bracket. The user will then use the numeric keypad to type
in the dosage, pressing Enter when complete. The digital display (top of Figure 7) will instruct
the user to insert the needle into a bottle of insulin. Having done so, the user will press a Fill
button located on the right side of the device (not shown in Figures 7 & 8) where a user’s thumb
will typically be located. This button will cause the motor to run and the bracket to pull on the
plunger, therefore drawing fluid into the chamber of the syringe. The digital display will notify
the user when the process is complete. The display will then instruct the user to insert the needle
into the body. Having done so, the user will press the Inject button located adjacent to the Fill
button. This button will cause the motor to run in the opposite direction as before, therefore
pushing on the plunger and expelling the fluid. Again, the user will be notified when the process
is complete.
Figure 7. 3D model of our final design. A syringe can be inserted into the device with the plunger controlled by a motor. The interface includes a numerical keypad for setting the dosage on the syringe. [18]
Figure 8. 3D model of the inner components of our final design. As shown, a screw is attached to a motor. A bracket translates along this screw to move the plunger simultaneously. [18]
Syringe Dosing 13/40
Prototype
The fall semester was devoted to constructing a large-scale model of our design,
mastering its mechanics and feasibility. This prototype was constructed without regard to
dimensions and is therefore much larger than a portable device. As a note: the specific
dimensions of our chosen design (Figure 7) have not been determined. Roughly, the device will
measure the length of two syringes (~10”) and have a diameter that fits in the palm of the hand
(~3”). The exact dimensions of the final prototype will be determined when the device is
dimensioned with CAD.
Construction of the Prototype
Figure 9. Picture of our constructed prototype. Annotations are shown for the description below. This model includes a syringe with its plunger controlled by a bracket. This bracket translates along a screw based on the direction the motor is turning. The motor is controlled by a PC with the shown microcontroller. [19]
The long base platform (Figure 9 – 1) and syringe support block (Figure 9 – 2) both
began as large pieces of kiln-dried carving wood. A table saw and band saw were employed to
cut the pieces to the desired dimensions before they were sanded and stained. General purpose
cement was then used to bond the pieces together. Next, the syringe (Figure 9 – 3) was mounted
to the support block using two brackets (Figure 9 – 4) purchased at a local hardware store. With
the syringe affixed, attention was turned toward the screw and its supports.
Syringe Dosing 14/40
Two screw supports (Figure 9 – 5) were machined from a block of scrap wood using a
drill press, band saw, and an electric powered sander. Wood was chosen instead of other
materials such as steel or aluminum due to the ease at which it could be machined. The supports
were mounted to the base platform using cement. The threaded rod (Figure 9 – 6) was selected
from an assortment of small diameter threaded rods at a hardware store. One end of the rod was
cut to the desired length and filed flat.
The bracket (Figure 9 – 7) that holds the syringe plunger was machined from a small
block of aluminum. This material was chosen because a high degree of precision was required
for this part and a threaded hole needed to be cut into the piece to accept the threaded rod. An
end mill was used to machine most of the piece geometry. A drill press and a tap were used to
create the threaded hole.
The two-phase step motor (1.8°/step) and controller circuit (Figure 9 – 8) were purchased
as a kit from online [20]. The kit also included the software required to control the motor from a
personal computer. The motor was mounted to the base platform using a pair of L-shaped
brackets, screws, and bolts (Figure 9 – 9). A small block of wood (Figure 9 – 10) also was
mounted under the motor to provide additional support. The controller circuit was affixed with
contact cement. Following the instructions that came with the kit, the motor was wired to the
controller.
The last piece of the prototype to be constructed was the sleeve (Figure 9- 11) that would
mechanically link the shaft of the motor to the threaded rod. A small section of aluminum rod
was machined with a lathe. A hole with a diameter slightly larger than the diameter of the motor
shaft was drilled halfway into the rod, along the centerline. Another hole, matching the diameter
of the threaded rod, was drilled through the other end of the sleeve to connect the two holes. A
Syringe Dosing 15/40
pair of set screws, placed in threaded holes, was used to apply pressure to the shafts and ensure
that a snug fit was obtained.
Overall, the construction of the prototype was relatively straightforward. A fair amount
of machining was required but it was accomplished without much difficulty.
Syringe Dosing Device: Advantages
The motor/screw prototype has a number of advantages over other designs. First of all,
using a two-phase step motor with a step size of 1.8° offers a high degree of accuracy. The step
size is sufficiently small enough to reach the level of accuracy required for the design
competition. The second advantage to this design is that the design itself and the underlying
principles of its operation are very straightforward. While sufficient testing has yet to be done,
the design has shown early indications that it will be quite accurate. A final advantage is that the
final prototype will be very user friendly. All of the complexities of the device will be out of
sight. The user will interface with the device solely through a simple keypad with feedback
provided by a series of LEDs. It is our goal to make the device as user friendly as possible.
Syringe Dosing Device: Disadvantages
While the prototype has many advantages, the design has a few drawbacks as well. One
of the major disadvantages is that the motor and other electrical components will require a
significant power supply. Motors in general require relatively high start-up current. Since the
device is to be portable, the power must come from a battery source. This is a concern due to
obvious spatial limitations. Another disadvantage to the device is that the motor creates some
degree of noise and vibration. Proper motor selection should minimize this problem. Finally, the
cost of the device has the potential to be high. The motor and the microprocessor could drive the
Syringe Dosing 16/40
price of the final prototype up, making the device unaffordable to a sufficient group of the
expected users. Again, motor selection will be crucial to ensure that this is not a major issue.
Syringe Calibration and Preliminary Testing
A microprocessor will control
the motor and determine how long it
needs to run for a given dosage. It is
therefore important to determine if the
syringe is approximately linear.
Calipers [21] were obtained to
measure the distance the plunger
moved from a baseline for every ten
units of the syringe. These
measurements were performed along the
length of the syringe, for all 100 units,
and repeated three times. A linear regression was performed showing approximate linearity,
with 0.224 inches between every ten units of the syringe and an error of +/- 0.00049 inches.
Since the motor moves in steps of 1.8°, there are 200 steps/revolution. It was discovered
that approximately 300 steps of the motor correspond to the plunger moving the length of 1.06
units on the syringe. Since we can control the motor to the step, the system will be easily within
the accuracy of 1 unit.
Figure 10. Linear regression of the average distance the plunger moves for every ten units on a 1 cc syringe. These measurements were made along the length of the syringe and repeated three times. The average distance between 10 units was 0.224 +/-0.00049 inches; R2 = 0.99. [22]
Syringe Dosing 17/40
Schoofs Prize for Creativity: Claims and Marketing Information
The first half of the spring 2005 semester was devoted towards market research and
developing a preliminary disclosure paper.
Claims
In disclosing our design for patent submission, we make the following claims:
1. An automatic syringe wherein said apparatus is handheld and portable. 2. The invention as defined in claim 1 wherein said device is powered by stepper motor. 3. The invention as defined in claim 1 wherein said device automatically measures medicine
by pressing numbers on a keypad. 4. The invention as defined in claim 1 wherein said device is more intuitive than insulin
pumps. 5. The invention as defined in claim 1 wherein said device shows a digital display of the
medication to be administered, offering a visual user interface. 6. The feature as defined in claim 5 will not be necessary for operation and said device will
include tactile and auditory clues. 7. The invention as defined in claim 1 wherein said device is accurate to 0.01 milliliter,
offering better accuracy than a person or other marketed devices. 8. The invention as defined in claim 1 wherein said device requires minimal dexterity. 9. The invention as defined in claim 1 where said device assists users with poor vision,
users with poor motor control, and users who need assistance injecting the fluid into the body.
Marketing Information
Our product will be marketed towards persons who have difficulties with fine motor
skills and who also have the need to accurately administer medication via a syringe. These users
will most likely be 65 years or older and have some sort of chronic illness, such as diabetes. In
2002, 12.4% of the U.S. population was represented by the 65 and older age group. This group
represented 40% of the total healthcare costs [23]. The U.S. Census Bureau expects this age
group to expand to approximately 17% of the population by 2020, due to the presence of the
baby boomer generation [24]. Because our product is designed mainly for consumers in this age
group, there is a favorable future for marketing our product. In addition, the World Health
Organization expects the prevalence of chronic disease (such as diabetes) to increase due to
Syringe Dosing 18/40
unhealthy lifestyles [25]. A product that seeks to aid patients suffering with diabetes or other
illnesses that require daily injections would therefore be increasingly necessary.
In addition to an expanding market for our product, there are also pricing trends within
the medical industry that are favorable. Overall pricing in the U.S. medical product industry will
remain competitive in future years [23]. Furthermore, the American population of peoples over
age 65 is 34,974,000 [24]. With 1% of Americans over 65 having Parkinson’s disease [26] and
18.37% of Americans over 60 having diabetes [27], an estimated 0.18%, or 62,247 people, have
both diabetes and Parkinson’s disease, and therefore, desperately need assistance. These
statistics exclude the numerous people with poor eyesight that need assistance with syringes. It
also excludes nurses and doctors who could appreciate the accuracy the device provides.
Although entry into the medical products industry is difficult due to many barriers of
entry and government regulations, our device’s classification as a Class I [28] medical device is
advantageous. Our automated syringe dosing system poses little patient risk and therefore
requires less government approval than Class II or Class III devices. Before manufacturing our
device we would need to register with the FDA, notify the FDA 90 days before any marketing
campaigns, and uphold good manufacturing practices [28]. Endorsement by a major medical
corporation would be necessary for competitiveness, manufacturing, and advertisement.
While the Minimed pump costs $6,195 [29], the Owen Mumford Autoject 2 costs $32.99
[30], and the Novopen costs $31.99 [31], we estimate our device to cost around $300, with $100-
200 being at cost. Initial costs are high due to the need for a small motor, the keypad, and the
circuitry involved. While this price may seem expensive compared to the Autoject 2 and the
Novopen, our device is much more accurate and vastly aids the user over these devices.
Syringe Dosing 19/40
Our device is much less expensive than a pump. Pumps are typically covered by
insurance companies and Medicare [15]. Should our device achieve corporate sponsorship, it
would be a good alternative to any patient wary over using pumps and a strong alternative to
insurance companies who would appreciate the lower cost. Furthermore, our device will be less
prone to wasting medication than a person using a syringe alone. Initial costs may be high for
our device, but saving medication may allow patients to come out ahead in the long term.
Smaller Motor
The team began designing a prototype with a smaller motor. This motor provides
7.5°/step, which has a significantly reduced resolution from the previous motor. However,
granted the excessively high resolution of the larger motor, as shown in Figure 9, resolution can
be sacrificed to get a smaller motor and yet still yield the required accuracy of 0.01 cc.
The bipolar motor purchased came with a controller kit [32], which the team constructed
and soldered. Time ran out in the spring 2005 semester, preventing the team from incorporating
this motor into the prototype with a new fitted sleeve (Figure 9 – 11). The motor and its
controller, along with the electronic connections, are shown in Figures 11 and 12, respectively.
Figure 11. LNS Technologies motor and our soldered controller circuit for this bipolar stepper motor [32].
Syringe Dosing 20/40
Figure 12. LNS Technologies bipolar stepper motor controller circuit diagram with wiring [32].
Microprocessor
In the second half of the 2005 Spring semester, our team began working on the
electronics side of the project. Using a microprocessor kit [33], we taught ourselves how to
program a BASIC Stamp 2 embedded system using PBASIC 2.5. Our first program was a lesson
in using a push button and a seven-segment LED. Each time the button was pressed, the LED
was increment by one. A later version of the code included a sound chip that would beep upon
the LED displaying ‘nine.’
The seven-segment LED button works by using a different pin from the microprocessor
for each segment. In the code, you can light up a segment with ‘HIGH A,’ where A is the pin
number corresponding to a segment on the LED. Subsequently, a segment can be turned off with
Syringe Dosing 21/40
‘LOW A.’ Programming the push button uses a ‘listening’ technique, where a loop cycles every
fourth of a second to see if a signal came through in the microprocessor pin linked to the button.
In the code at hand, a signal from the push button would instruct the seven-segment LED display
to increment one number, where the number to be displayed was determined with a counter. The
actual process of displaying a number worked by using the ‘GOSUB B’ command, which calls a
function B. A function was used for displaying each number, with each
function setting the appropriate pins HIGH and the appropriate pins LOW
from the previous number displayed on the LED. When the number ‘nine’
was reached, the sound chip would beep with the following code,
‘FREQOUT p, t, f,’ where p is the pin number, t is the length of time, in
milliseconds, for the beep to occur, and f is the frequency of the sound.
This code is provided in Appendix C.
The team attempted to incorporate
a numeric keypad [34] and a digital
display, but ran out of time. We ordered a
keypad (Figure 13) and began to
understand the coding, which is described
in Figure 14. Furthermore, we ordered a
10-digit length LCD display, which is
much bigger than desired for a final
device, but was used to begin learning
the programming aspect of the display.
Figure 13. 12-key numeric keypad with ten pins, programmed in a matrix format
Figure 14. 10-pin matrix format of a 12-key keypad [34].
Syringe Dosing 22/40
Further programming of the microprocessor will need to include error handling for the
following situations (Figures 15 and 16).
Figure 15. Microprocessor Error Handling Flow Chart. If the user realizes that he or she entered an incorrect dosage during the filling or injecting stages, the user can stop the device and start over.
Figure 16. Microprocessor Error Handling Flow Chart. If the user realizes that he or she entered an incorrect dosage before filling, the user can press clear and re-enter the correct dosage.
The first flow chart (Figure 15) demonstrates the path followed by a user who realizes an
input error before the filling stage. The device allows the user to clear the entry and restart the
Syringe Dosing 23/40
process. The second flow chart (Figure 16) demonstrates the path followed by a user who
realizes an input error during the filling or injecting stage. The user can stop the syringe’s
current action and view the current volume. The user would then decide the appropriate course
of action.
The team also worked on a master code that would be used in the final product. This code
begins to handle every aspect of the design, including the motor, display, and keypad buttons.
Error-handling is scant at this point. This code is provided in Appendix D. Please note that the
actual code is not meant to work with all components because it doesn’t, for example, include the
proper coding for controlling a motor. This code is meant to outline the true program to be
uploaded onto the microprocessor.
Future Work
Future work includes finishing the microprocessor program, incorporating the coding of
the keypad and digital display (seven-segment LEDs). Further construction is needed to house
the microprocessor and motor in a small compartment. Finally, the team has until 12/3/06 to file
a patent disclosure, which will be done through the Wisconsin Alumni Research Foundation
(WARF).
Ethical Considerations
Delivery of the incorrect amount of insulin or other medication could be very dangerous.
Consequently, there must be a way for the patient to easily check that the correct amount of
medication is being administered, and most importantly, that there are no bubbles in the chamber
of the syringe. Even though the chamber of the syringe will be openly exposed, the patient with
poor eyesight will be unable to check for bubbles or an incorrect dosage. If a mechanism was
Syringe Dosing 24/40
used to detect these errors, such as a laser checking the passage of light along the length of the
syringe, the device would become prohibitively expensive.
Conclusions
Our team has proposed a device that is beneficial to diabetes patients, particularly those
with poor motor skills and/or poor eyesight. The device aids patients in measuring and
administering medication via a syringe. It would be competitive in the current market and more
so in the future with this expanding market. Compared to current products, the device is an
attractive alternative to both patients and insurance companies.
Syringe Dosing 25/40
References
[1] BD Diabetes. BD Ultra-FineTM Insulin Syringes. http://www.bddiabetes.com/us/product/thin wall_consumer.asp. 2004. [2] RERC-AMI. 2004-2005 National Student Design Competition. http://www.rerc-ami.org/rerc ami/r2d2/d22-yr2.htm. 2004. [3] Swift, J., Mehta, A., Millin, J., Poper, R. National Design Competition: Ergometer. http://ww w.cae.wisc.edu/~bmedesgn. 2004. [4] Medline plus: National Library of Medicine. Diabetes tutorial. http://www.nlm.nih.gov/med lineplus/tutorials/diabetesintroduction/id029102.html. 2004. [5] Medline plus: National Library of Medicine. Diabetes. http://www.nlm.nih.gov/medlineplus /diabetes.html. 2004. [6] Center for Disease Control. Diabetes: Frequently Asked Questions. http://www.cdc.gov/diab etes/faq/index.htm. 2004. [7] Health. Diabetes Symptom. http://www.diabetes-immune-system.com/. 2004. [8] NIH Clinical Center. Giving a Subcutaneous Injection. http://www.cc.nih.gov/ccc/patient_ed ucation/pepubs/subq.pdf. 2004. [9] Medtronic. MiniMed. http://www.minimed.com/. 2004. [10] Canadian Diabetes Association. Insulin: Things you Should Know. http://www.diabetes.ca/S ection_About/insulin2.asp. 2004. [11] Midbon, Missy. Diabetes patient interview. 9/10/04. [12] Owen Mumford. Autoject 2: Self injection made simpler…http://www.owenmumford.com /autoject2.html. 2004. [13] Novo Nordisk. NovoPen® 3 – An unsurpassed range. http://www.novonordisk.com/diabete s/public/insulinpens/novopen3/default.asp. 2004. [14] Haist, Doug. Diabetes and Parkinson’s disease patient interview. 10/06/04. [15] Yaeger, Kara. Diabetes educator and nurse interview. 09/14/04. [16] BD. Prefilled Syringes Brochure. http://www.bd.com/pharmaceuticals/products/BDPS_bro chure.pdf. 2004. [17] Wentland, A.L. Drawn in Deneba Canvas 9.0.3. 2004.
Syringe Dosing 26/40
[18] Lienau, W. of Flite Productions, Madison, WI, USA. Drawn and rendered in 3D Studio Max. 2004. [19] Wentland, A.L. Picture taken with a 5.0 MegaPixel EasyShare DX4530 Kodak Camera. Annotations constructed in Deneba Canvas 9.0.3. 2004. [20] Stepper World. MS-1 Microstepping Motor Control Project: Advanced Microstepping Control. http://www.stepperworld.com/. 2004. [21] Enco. Enco Vernier Calipers. 2004. [22] Wentland, A.L. Linear regression using Matlab Student ed. 6.5. 2004. [23] Healthcare Products & Supplies Industry Survey: Industry Trends. Standard & Poor’s Industry Surveys. Sept. 23, 2004. [24] US Census Bureau. Percent Distribution of the Population by Age: 1990 to 2050. http://www.census.gov/prod/1/pop/p25-1130/p251130b.pdf. 2005. [25] World Health Organization. Obesity and Overweight. http://www.who.int/hpr/NPH/docs/ gs_obesity.pdf. 2005. [26] Center for Disease Control. National Center for Health Statistics: Data on Parkinson’s Disease. http://www.cdc.gov/nchs/data/factsheets/Parkinsons.pdf. 2005. [27] American Diabetes Association. National Diabetes Fact Sheet. http://www.diabetes.org /diabetes-statistics/national-diabetes-fact-sheet.jsp. 2005. [28] Food and Drug Administration. Device Classes. http://www.fda.gov/cdrh/devadvice/ 3132.html. 2005. [29] Medtronic Minimed. The Paradigm® Platform FAQ. http://www.minimed.com/patientfam/ pf_ipt_paradigm_faq.shtml. 2005. [30] Diabetic Express. Autoject 2 Fixed Needle Device. http://www.diabeticexpress.com/content /ProductDetail.aspx?CategoryID=0&SubCategoryID=0&ItemID=08-14. 2005. [31] Walgreens. Novopen 3 Insulin Delivery System. http://www.walgreens.com/library/ finddrug/druginfo1.jhtml?id=14415. 2005. [32] LNS Technologies. Part Number: BISTEP-MAN. Bipolar Stepper Motor Controller Kit. 2004. [33] Parallax, Inc. Parallax #27207 BASIC Stamp Discovery Kit. Oct., 2004.
Syringe Dosing 27/40
[34] Ellis, J.R. Embedded Systems & Test Equipment. ElectronicBrains.com. 2005.
Syringe Dosing 28/40
Appendix A PDS Function:
The device will accept a standard 1 cc syringe of ¼” diameter and automate the process of
measuring and delivering medication. After a syringe is inserted into the device, a user can type
in the number of units that should be drawn into the barrel of the syringe. The user will then
insert the needle into a bottle of medication, such as insulin, and press the “Fill” button. A motor
will drive the syringe’s plunger back via a screw mechanism, drawing the medication into the
barrel. Afterwards, the user will insert the needle into the body and press the “Inject” button.
The motor and screw will drive the plunger in and push the medication into the user. Since the
only manipulation required by the user is pressing numbers on a keypad and inserting the needle
into the body, the amount of dexterity is reduced dramatically.
Client Requirements:
The device should accept 1 cc syringes ¼” in diameter.
The device should dose to the nearest 0.01 cc.
The device should be easy to use for people with poor vision, low dexterity, diminished
motor control, Parkinson’s, and/or other neuromuscular disorders.
Design Requirements:
1. Physical and Operational Characteristics
a. Performance requirements
i. On average, the device will be used twice daily with slow-acting insulin and
several more times with fast-acting insulin to accommodate the eating
habits and varying blood sugar levels of the user.
ii. The device will be modeled to fit BD Ultra-Fine II 1 cc syringes (BD has 75 –
80% market share).
iii. The device should be easy to use for people with poor eyesight and/or
dexterity.
b. Safety
Syringe Dosing 29/40
i. Standard syringe safety measures should be taken, including proper sterility
procedures and disposal.
ii. The device will contain batteries, so the batteries should not be left in the
device if they have exceeded their expiration date.
iii. Being electronic, the device should comply with FCC standards.
c. Accuracy and Reliability
i. The device will use 1 cc syringes and dose to the nearest unit, where a unit is
1/100 of a cc.
ii. With the stepper motor in the prototype, the shaft of the motor can move in
steps of 1.8°. In initial testing, 300 steps of the motor corresponded to 1.06 units
being expelled from or drawn in to the plunger. Therefore, the prototype is
accurate to a minimum accuracy of 1/100 of a cc.
iii. This accuracy will be true as long as the syringe dimensions, screw pitch, and
motor increments do not change.
d. Life in Service
i. The device should function for at least five years around five times per day at
five minutes per usage under normal functionality (low-viscosity fluid, such as
insulin, and careful handling).
e. Shelf Life
i. Being electronic, the device should not be left in the sun or in freezing
temperatures. Direct contact of the electrical components with liquids should be
avoided.
ii. The batteries should not be left in the device for more than three years or
beyond their expiration date, whichever comes first.
f. Operating Environment
i. The device should operate and remain accurate at temperatures above freezing
and below 100 °F.
ii. The device should not be affected at different pressures or humidity.
iii. The device will be inoperable if dust, dirt, or water has gotten into the device.
iv. The device can be used by able people of any age above seven.
g. Ergonomics
Syringe Dosing 30/40
i. This device seeks to overcome the typical problems that patients with poor
muscular control, tremors, Parkinson’s, and poor eyesight have operating a
syringe.
ii. Anyone who has some control of the hands should be able to use the device,
because the device only requires the pressing of a few buttons with low resistance.
h. Size
i. The device should fit into a single hand.
ii. The device should not exceed the length of two syringes (~10”) or a width of
3”.
iii. It should be completely portable, such as being able to fit into a purse.
i. Weight
i. Ideally, the device should weigh less than a pound.
j. Materials
i. There are no restrictions on the materials, although non-slippery materials are
best.
ii. Materials that are smooth are ideal, so as to ensure comfort of the user.
k. Aesthetics
i. The color of the device should be mild, so as not to attract attention. A colorful
appearance may bring too much attention to a user that prefers discretion.
ii. The device should have a smooth texture, but not too smooth to slip out of the
hands. Rubber grips in key locations would be best.
iii. The device should contour to the hand in a natural fashion.
2. Production Characteristics
a. Quantity
i. Potential for thousands of units, depending on the number of patients/companies
that desire the device. For the National Student Design Competition, one
prototype is needed.
b. Target Product Cost
i. Suggested retail price: $300 - $500
ii. Manufacturing costs: ~$150-200
iii. Existing devices that don’t assist in dosing:
Syringe Dosing 31/40
- MiniMed pump ~$5000
- Novopen ~$34
- AutoJect ~$44
3. Miscellaneous
a. Standards and Specifications
i. FDA approval required
b. Customer
i. The device should be portable, lightweight, and fit in one hand.
ii. The device should completely eliminate the need to manually set the
dosage on the syringe.
iii. The device should be easy to use, especially for individuals with tremors, poor
muscular control, Parkinson’s, and poor eyesight.
c. Patient-related concerns
i. The device itself will not need to be sterilized, but for each usage, a new syringe
should be used with proper sterilization and swabbing of the skin.
ii. Proper disposal procedures should be taken.
d. Competition
i. Aside from the items listed above, there are no other devices that assist in
dosing syringes to the degree that ours does.
Syringe Dosing 32/40
Appendix B
Expenses Budget: $2,000
Costs to date:
℘ $13.82 – Supplies at Home Depot (screws, brackets, and glues)
℘ $125 – MS-1 Motor Kit [20] from StepperWorld.com
℘ $178 – Parallax BASIC stamp 2 kit [23]
℘ $39.95 – ZiLog microprocessor kit from DigiKey.com
℘ $25.95 – 8-digit LED driver from Parallax
℘ $264.52 – 2x Board of Education (USB) from Parallax
℘ $8.80 – 12-Button Keypad from ElectronicBrains.com
℘ $39.00 – 10 seven-segment LEDs from LC Led Inc.
℘ $4.00 – 10-digit LCD from AllElectronics.com
℘ $5.50 – 5 seven-segment LEDs from Parallax _______________
Total: $704.54
Remaining Budget: $1,295.46
Syringe Dosing 33/40
Appendix C
BASIC program for counting zero through nine
'What's a Microcontroller - ZeroThruNine.bs2 ' {$STAMP BS2} ' {$PBASIC 2.5} counter VAR Nib counter2 VAR Word pulses VAR Word duration VAR Word GOSUB set_Zero DO IF IN3 = 0 THEN counter = counter + 1 IF counter = 1 THEN GOSUB set_One ELSEIF counter = 2 THEN GOSUB set_TWO ELSEIF counter = 3 THEN GOSUB set_Three ELSEIF counter = 4 THEN GOSUB set_Four ELSEIF counter = 5 THEN GOSUB set_Five ELSEIF counter = 6 THEN GOSUB set_Six ELSEIF counter = 7 THEN GOSUB set_Seven ELSEIF counter = 8 THEN GOSUB set_Eight ELSEIF counter = 9 THEN GOSUB set_Nine 'FREQOUT 6, 1000, 2000 FOR counter2 = 1 TO 150 PULSOUT 5, 1000 PAUSE 20 NEXT FOR counter2 = 1 TO 150 PULSOUT 5, 500 PAUSE 20 NEXT FOR counter2 = 1 TO 150 PULSOUT 5, 750 PAUSE 20 NEXT
Syringe Dosing 34/40
ENDIF DEBUG ? counter PAUSE 250 ENDIF LOOP set_Zero: HIGH 15 HIGH 14 HIGH 13 HIGH 10 HIGH 9 HIGH 8 RETURN set_One: HIGH 15 HIGH 10 LOW 14 LOW 13 LOW 9 LOW 8 RETURN set_Two: HIGH 15 HIGH 14 HIGH 12 HIGH 9 HIGH 8 LOW 10 RETURN set_Three: HIGH 15 HIGH 14 HIGH 12 HIGH 10 HIGH 9 LOW 8 RETURN set_Four: HIGH 15 HIGH 13
Syringe Dosing 35/40
HIGH 12 HIGH 10 LOW 14 LOW 9 RETURN set_Five: HIGH 14 HIGH 13 HIGH 12 HIGH 10 HIGH 9 LOW 15 RETURN set_Six: HIGH 14 HIGH 13 HIGH 12 HIGH 10 HIGH 9 HIGH 8 RETURN set_Seven: HIGH 14 HIGH 15 HIGH 10 LOW 13 LOW 12 LOW 9 LOW 8 RETURN set_Eight: HIGH 15 HIGH 14 HIGH 13 HIGH 12 HIGH 10 HIGH 9 HIGH 8 RETURN set_Nine: HIGH 15 HIGH 14
Syringe Dosing 36/40
HIGH 13 HIGH 12 HIGH 10 LOW 9 LOW 8 RETURN
Syringe Dosing 37/40
Appendix D
Master BASIC program for the design
' {$STAMP BS2} ' {$PBASIC 2.5} keypadInput VAR Nib flag VAR Boolean degrees VAR Nib drawn VAR Boolean injected VAR Boolean display VAR String degrees = 295 flag = false drawn = false injected = false DO IF IN3 = 0 THEN IF keypadInput = 1 & flag = false THEN GOSUB set_One ELSEIF keypadInput = 2 & flag = false THEN GOSUB set_TWO ELSEIF keypadInput = 3 & flag = false THEN GOSUB set_Three ELSEIF keypadInput = 4 & flag = false THEN GOSUB set_Four ELSEIF keypadInput = 5 & flag = false THEN GOSUB set_Five ELSEIF keypadInput = 6 & flag = false THEN GOSUB set_Six ELSEIF keypadInput = 7 & flag = false THEN GOSUB set_Seven ELSEIF keypadInput = 8 & flag = false THEN GOSUB set_Eight ELSEIF keypadInput = 9 & flag = false THEN GOSUB set_Nine ELSEIF keypadInput = 0 & flag = false THEN GOSUB set_Zero ELSEIF keypadInput = ENTER THEN GOSUB set_Return ENDIF ENDIF IF IN4 = 0 THEN IF !flag THEN display = 'Press enter.' ELSEIF flag THEN GOSUB draw ENDIF
Syringe Dosing 38/40
ENDIF IF IN1 = 0 THEN IF !drawn THEN display = 'Draw fluid.' ELSEIF GOSUB inject ENDIF ENDIF LOOP set_Zero: HIGH 15 HIGH 14 HIGH 13 HIGH 10 HIGH 9 HIGH 8 RETURN set_One: HIGH 15 HIGH 10 LOW 14 LOW 13 LOW 9 LOW 8 RETURN set_Two: HIGH 15 HIGH 14 HIGH 12 HIGH 9 HIGH 8 LOW 10 RETURN set_Three: HIGH 15 HIGH 14 HIGH 12 HIGH 10 HIGH 9 LOW 8 RETURN set_Four:
Syringe Dosing 39/40
HIGH 15 HIGH 13 HIGH 12 HIGH 10 LOW 14 LOW 9 RETURN set_Five: HIGH 14 HIGH 13 HIGH 12 HIGH 10 HIGH 9 LOW 15 RETURN set_Six: HIGH 14 HIGH 13 HIGH 12 HIGH 10 HIGH 9 HIGH 8 RETURN set_Seven: HIGH 14 HIGH 15 HIGH 10 LOW 13 LOW 12 LOW 9 LOW 8 RETURN set_Eight: HIGH 15 HIGH 14 HIGH 13 HIGH 12 HIGH 10 HIGH 9 HIGH 8 RETURN set_Nine:
Syringe Dosing 40/40
HIGH 15 HIGH 14 HIGH 13 HIGH 12 HIGH 10 LOW 9 LOW 8 RETURN set_Return GOSUB set RETURN set flag = true RETURN draw DONE VAR Nib DONE = keypadInput*degrees FOR counter = 1 TO DONE IN2 = 1 NEXT drawn = true RETURN inject DONE VAR Nib DONE = keypadInput*degrees FOR counter = 1 TO DONE IN0 = 1 NEXT injected = true display = 'Complete' PAUSE 1000 GOSUB set_Zero RETURN