View
577
Download
4
Tags:
Embed Size (px)
Citation preview
1
MLIBA PHARMACY COLLEGE,BARDOLI.
1
A Seminar On
Contents
Introduction to Validation
Stages of qualifications
Validation of Autoclave
Validation Protocol of Autoclave
Validation of Dry Heat Sterilizers And
Tunnel
Validation
Validation may be defined as ” Establishing documented
evidence which provides a high degree of assurance that a
specific process will consistently produce a product
meeting its pre-determined specifications and quality
attributes.”
It has been made mandatory by the regulatory bodies to
prove the safety efficacy, Purity & effectiveness of the drug
product, medical devices & biologics in the marketplace &
health system.
Why Validation of Equipment?
Equipment validation is Vital forSafetyFewer interruptions of workLower repair costsElimination of premature replacementLess standby equipmentIdentification of high maintenance costReduction of variation in resultsGreater confidence in the reliability of results
Who should do Equipment Validation?
The vendor or the userThe user has the ultimate responsibility for the accuracy of
the analysis results and also for equipment qualification.
DQ should always be done by the user. While IQ for a small and low cost instrument is usually done
by the user, IQ for large, complex and high cost instruments should be done by the vendor.
OQ can be done by either the user or the vendor. PQ should always be done by the user because it is very
application specific, and the vendor may not be familiar with these. As PQ should be done on a daily basis, this practically limits this task to the user.
Validation
Part 1. Overview on qualification and validation
Part 2. Qualification of HVAC and water systems
Part 3. Cleaning validation
Part 4. Analytical method validation
Part 5. Computerized system validation
Part 6. Qualification of systems and equipment
Part 7. Non sterile product process validation
ValidationStages of qualification
Design qualification
Installation qualification
Operational qualification
Performance qualification Change control
Validation
Requalification
Defined schedule
Frequency based onFactors
Results of calibrationmaintenance,
verification
Periodic
After change
Part ofChange control procedure
Extent based onRisk assessment
Equipment qualification
Equipment qualification / validation includes
following things:
Design qualification (DQ)
Installation qualification (IQ)
Operational qualification (OQ)
Performance qualification (PQ)
Design Qualification (DQ)"Design qualification (DQ) defines the
functional and operational specifications of the instrument and details for the conscious decisions in the selection of the supplier".
List below recommends steps that should be considered for inclusion in a design qualification. Description of the analysis problemDescription of the intended use of the
equipmentDescription of the intended environment
Preliminary selection of the functional and
performance specifications
Preliminary selection of the supplier
Instrument tests (if the technique is new)
Final selection of the equipment
Final selection of the supplier and equipment
Development and documentation of final
functional and operational specifications
Installation Qualification(IQ)“Installation qualification establishes that the
instrument is received as designed and specified, that it is properly installed in the selected environment, and that this environment is suitable for the operation and use of the instrument.”
The qualification involves the coordinated efforts of –
The vendor
The operating department
The project team (which provide input into the
purchase, installation, operation and maintenance
of the equipment).
Operational Qualification (OQ)
"Operational qualification (OQ) is the process
of demonstrating that an instrument will
function according to its operational
specification in the selected environment."
The proper operation of equipment is verified
by performing the test functions specified in
the protocol.
A conclusion is drawn regarding the operation
of equipment after the test functions are
checked and all data has been analyzed.
Following are the contents of equipment
operation qualification
1.Application S.O.P’s
2.Utilization List
3.Process Description
4.Test Instrument Utilized To Conduct Test
5.Test Instrument Calibration
6.Critical Parameters
7.Test Function (List)
8.Test Function Summaries
Performance Qualification(PQ)"Performance Qualification (PQ) is the
process of demonstrating that an instrument consistently performs according to a specification appropriate for its routine use ".
PQ should always be performed under conditions that are similar to routine sample analysis.
PQ should be performed on a daily basis or whenever the equipment is being used.
In practice, PQ can mean system suitability testing, where critical key system performance characteristics are measured and compared with documented.
A. IntroductionSterile products have several unique dosage
form properties, such asFreedom from micro-organisms, Freedom from pyrogens, Freedom from particulates, Extremely high standards of purity and
quality; However, the ultimate goal in the
manufacture of a sterile product is absolute absence of microbial contamination.
Introduction(Con..)
Three principles are involved in the validation
process for sterile product.
1. To build sterility into a product
2. To demonstrate to a certain maximum level of
probability that the processing and sterilization
methods have established sterility to all units of a
product batch
3. To provide greater assurance and support of the
results of the end product sterility test
D value“It is time required for a 90% reduction in
microbial population. Quantitative expression of rate of killing of micro organism.”
In other words, the D value will be affected by
The type of microorganism used as BI,The formulation components and
characteristics The surface on which the micro-organism is
exposed The temperature, gas concentration, or
radiation dose of sterilization process.
D value found by 2 methods, 1) Survivor curve method (log number of surviving
organism versus time/gas concentration/radiation dose)2) Fraction negative method
Z valueUsed exclusively in validation of heat sterilization
process. Z value is reciprocal of slope of plot of log D verses T at which D value is found i.e. increase in temperature required to reduce D value of organism by 90 % (1 log reduction)
F valueUsed exclusively in validation of heat sterilization
process. It is time in min required to kill all spores in suspension at 121oC. Measures equivalent time
Methods of Sterilization of Products
1.Heat
Moist heat (autoclave)
Dry heat oven or tunnel
2.Gas
Ethylene oxide
Peracetic acid
Vapor phase hydrogen peroxide
Chlorine dioxide
3.Radiation
Gamma
Beta
Ultraviolet
B. Qualification and Calibration
1) Mechanically Checking, Upgrading, and Qualifying the
Sterilizer Unit
The main concern with steam sterilization is the
complete removal of air from the chamber and
replacement with saturated steam.
Autoclaves can also involve air–steam mixtures for
Sterilizing flexible packaging systems and syringes.
When autoclave system is used, the unit must be
installed properly and all operations qualified through
installation qualification and operation qualification
(IQ/OQ).
2) Selection and Calibration of Thermocouples
Thermocouples must be durable for repeated use as
temperature indicators in steam sterilization validation and
monitoring.
Copper constantan wires coated with Teflon are a popular
choice as thermocouple monitors.
Accuracy of thermocouples should be 0.5°C. Temperature
accuracy is especially important in steam sterilization
validation.
Thermocouple accuracy is determined using National
Bureau of Standards (NBS).
3) Selection and Calibration of BISr. No
Sterilization process Biological Indicator(BI)
1. Autoclave B. steriothermophillus sporesB. subtilis var. niger sporesB. subtilis, 5230 sporesB. coagulance sporesClostridium sporogenes spores
2. Dry heat B. subtilis var. niger sporesB. subtilis, 5230 spores
3. Ethylene Oxide B. subtilis var. niger spores
4. Radiation B. pumilus sporesMicrococcus radiodurans vegetative cells
C. Heat-Distribution Studies
Heat-distribution studies include two phases:
1) Heat distribution in an empty autoclave
chamber
2) Heat distribution in a loaded autoclave
chamber.
The trips where the wires are soldered should not
make contact with the autoclave interior walls or
any metal surface.
Cont..
Heat-distribution studies may employ thermocouples as
the cool spot in the chamber.
The principle is the location of the cool spot and the
effect of the load size and/or configuration on the cool
spot location.
The difference in temperature between the coolest spot
and the mean chamber temperature should be not greater
than 2.5°C .
Greater temperature differences may be indicative of
equipment malfunction.
D. Heat-Penetration Studies
This is the most critical component of the entire
validation process.
The main purpose is to determine the F0 value of
the cold spot inside the commodity.
The container cold spot for containers ≥100 ml is
determined using container-mapping studies.
Thermocouple probes are inserted within a
container and repeat cycles are run to establish the
point inside the container.
Cont..
Thermocouples will be placed both inside and
outside the container at the cool spot location(s), in
the steam exhaust line, and in constant-
temperature baths outside the chamber.
F0 value will be calculated based on the
temperature recorded by the thermocouple inside
the container at the coolest area of the load.
F0 value will indicate whether the cycle is
adequate or alterations are needed.
Heat-Penetration Studies(Con..) Three critical parameter associated with all
wet heat sterilization Processes:1.A minimum F value2.A design F value3.A sterilization process time Any changes in the load size, load
configuration, or container characteristics must be accompanied;
To prove that the cool spot location has not changed or,
If it has, that it receives the design F0 time exposure from the sterilization cycle used.
E. Equipment Qualification
Prior to the initiation of process, it is important that the
sterilizer be suitably qualified to perform its function.
Typical critical requirements that are considered to
affect the sterilization process (e.g.“quality”
requirements) are:
Accurate temperature and pressure measurement
Air removal to some predefined level of vacuum
Temperature distribution and uniformity in the
chamber.
The qualification of a sterilizer should include the
following :
1.Calibration of temperature and pressure sensors
(traceable to national or international standard)
2.Air removal (usually measured by vacuum level
achieved vs. defined requirement)
3.Demonstration of the sequence of operations,
4.Confirmation of alarms and interlocks
5.Precision of temperature control
6.Temperature distribution and uniformity
F. Microbiological Challenge Studies
Microbiological challenges studies are employed to
provide additional necessary assurance that adequate
lethality has been delivered to all parts of the load.
Calibrated BIs used as bioburden models providing
data that can be employed to calculate Fo.
The microorganisms used to challenge moist heat
sterilization cycles are G. stearothermophilus and
Clostridium sporogenes.
After the sterilization cycle is complete, the
inoculated items or spore strips are recovered
and subjected to microbiological test
procedures.
Strips are immersed in a suitable growth
medium (soybean casein digest medium is
typical) and incubated for up to seven days.
G. Sterilizer Filter Evaluation
Microbial filters are employed on most parts of
sterilizers to ensure that loads are not contaminated
by air used to vent the chamber as it cools or dries.
Product loads are protected from such
contamination by their primary containers (vials,
bags) and many nonproduct loads are protected by
wraps to provide a microbial barrier.
For filters, two issues are of concern:
Sterility and Integrity.
If the load will undergo a bioburden cycle, it may
be necessary to sterilize the filter in a separate
phase of the cycle.
To ensure that filters will remain functional under
all expected conditions, the integrity tests should
be done following the maximum cycle time and
temperature.
Triplicate studies are recommended.
A. Introduction
Mainly three types of dry-heat sterilization
systems are utilized in the pharmaceutical
industry today.
I. Batch Sterilizer Ovens
II.Tunnel Sterilizers
III.Microwave Sterilizers
PRINCIPLES OF HEAT TRANSFER AND
CIRCULATION:
The dry heat process must effectively heat the
article, and air surrounding the article, to achieve
sterilization or depyrogenation.
In moist heat, the condensation of the steam
sterilizer releases large amounts of heat energy that
serves to heat the items in the sterilizer.
In dry heat processes the hot air carries
significantly less heat energy than an equivalent
volume of saturated steam.
Key Process Features to Control Prior to Validating Dry-Heat Sterilizer
Batch(Oven) Tunnel Steriliser
Intake air system Positive pressure to entrance
Exhaust air system Even distribution of heat
Internal air circulation Belt speed recorder
Exhaust HEPA filter HEPA-filtered cooling air
Static pressure gauge Exhaust HEPA filter
Heater current Particulate control
The four main mechanism through which Heat
transfer occurs are:
Convection
Circulation
Conduction
Radiation
B. Batch Oven Validation1. Air balance determination: In an empty oven, data are obtained on the flow
rates of both intake and exhaust air. Air should be balanced so that positive pressure
is exerted to the nonsterile side when the door is opened
2. Heat distribution of an empty chamber: Thermocouples should be situated according to
a specific predetermined pattern. Repeatability of temperature attainment and
identification of the cold spot can be achieved if the temperature range is 15°C at all monitored locations.
3. Heat-penetration studies:
These studies should be designed to determine
the location of the slowest heating point within a
commodity at various locations of a test load in
the sterilizer.
Thermocouples are placed in the commodities
located in the areas likely to present the greatest
resistance to reaching the desired temperature.
Minimum and maximum temperatures as
defined in the process specifications should be
studied.
4. Mechanical repeatability:
During all these studies, mechanical
repeatability in terms of air velocity,
temperature consistency, and reliability and
sensitivity of all the oven and instrumental
controls must be verified.
C. Tunnel Sterilizer Validation 1. Air Balance Determination:
In this study items being sterilized are moving exposed to
different air systems (e.g., heating zone and cooling zone).
Air flow must be balanced in order to provide a gradual
decrease in air temperature as items move along the
conveyor.
In the absence of a critical balance of air dynamics, either
the items will not be cooled or they will be cooled too
quickly, causing contamination of the entire tunnel area.
2. Heat-Distribution Studies:
Thermocouples used in tunnel sterilizer validation
must be sufficiently durable to withstand the extremely
high (≥300°C) temperatures in the heating zone area of
the tunnel.
Heat-distribution studies should determine where the
cold spots are located as a function of the width of the
belt and height of the tunnel chamber.
Peak temperature readings should remain within
10°C across the belt for at least three replicate runs.
3. Heat-Penetration Studies:
Prior to microbial challenge testing of the tunnel
sterilization, heat-penetration studies must be
completed in order to identify the coolest container in
the entire load.
Three to five replicate runs for each commodity size
and every loading configuration should be done
using 10 to 20 thermocouples distributed throughout
the load.
Careful analysis of the temperature data after each
run will be invaluable in the determination of the
cool spot
4. Mechanical Repeatability:
Tunnel sterilizers must demonstrate
mechanical repeatability in the same manner
as batch ovens.
Air velocity, air particulates, temperature
consistency and reliability of all the tunnel
controls (heat zone temperatures, belt speed)
must be proved during the physical validation
studies.
D. Biological Process Validation of Dry D. Biological Process Validation of Dry Heat Sterilization CyclesHeat Sterilization Cycles
If the dry-heat process is claimed to produce both
sterile and pyrogen-free commodities, validation
studies must be done using both micro-organisms
and microbial endotoxins.
The goal is to validate a heating cycle that can
produce a 12-log reduction in the biological
indicator population.
The most widely used biological indicators for
dry heat have been spores of B. Subtilis.
Procedures for the validation of a tunnel sterilization:
The overkill approach is selected for the validation
study.
Select the type of biological indicator to be used.
Run a complete cycle using the desired loading
pattern.
Determine the number of survivors by plate-counting
or fraction negative Methods.
Determine the number of spore log reductions (SLRs)
E. Endotoxin challenge in Dry Heat Sterilization
Inoculate commodity samples with a known
amount of endotoxin. (e.g., 10–100 ng Escherichia
coli lipopolysaccharide)
Thermocouples should be placed in commodities
adjacent to those containing endotoxin for
temperature monitoring and correlation with LAL
test results.
Endotoxin destruction should be ascertained at the
coolest location of the load.
Several endotoxin challenge samples should be
done per cycle, and the studies must be adequately
replicated.
Following the dry-heat cycle, aseptically transfer
the units containing endotoxin to an aseptic area
for extraction procedures.
F values required for endotoxin destruction at
various temperatures and/or cycle time–
temperature variations can be determined using a
Z value of 54°C.
VALIDATION OF TEST EQUIPMENT
Equipment required to conduct the IQ, OQ
and PQ are discussed here.
All temperature equipment employed to
perform the validation studies must be
traceable and calibrated to the International
Temperature Scale
The equipments used for validation testing of dry heat
processes are discussed here:
Resistance Temperature Detectors
Thermocouples
Data Loggers
Wireless Temperature Logger
Infrared Thermometer
Constant Temperature Baths
Stopwatch
Voltmeter or Ammeter
Optical Tachometer
INSTALLATION QUALIFICATION
The IQ is designed to compare the system against the
manufacturer’s specifications for proper installation.
All equipment, utilities, and connections must be
checked against the manufacturer’s
recommendations.
A. Structural:
Check dimensions, presence of identification plates,
correct leveling, proper insulation, presence of seals,
and inspect for structural damage.
B. Filters:All filters used within the system must be
recorded, such as those used with air (supply, re-circulating) or in other utilities (e.g., steam, water).
Some HEPA filters may need to be checked periodically by performing an integrity test or DOP.
C. Electrical:Ensure conformance to National Electrical Code
StandardsD. HVAC:Ensure the system provides the RH, temperature,
and pressure differential required.
E. Air Supply:Identify source (direct from the HVAC system
or room air), duct size, duct material of construction, and air classification.
F. Ventilation:Check that the ventilation exhaust duct
exhausts to an appropriate area (not to an aseptic environment), and identify the method used to prevent back-flow.
G. Door Gaskets:Check integrity of gaskets and materials of
construction.
H. Heaters:Record the manufacturer’s model number, the
number of heating elements, and the voltage, amperage, and wattage of the elements for the heaters.
I. Lubricants:Make certain that any lubricants used cannot
contaminate the material being sterilized or depyrogenated.
J. Blowers:The blower must be mechanically sound, the
volute in place and correctly balanced, and that the blades rotate in the correct direction.
OPERATIONAL QUALIFICATIONA. Temperature Monitors:The temperature controllers, recorders, and
sensors on the process equipment must be calibrated before the unit can be operated reliably.
B. Cycle Timer:The accuracy of the timer must be determined,
so that assurance is provided for cycle length.C. Door Interlocks:If a unit is equipped with double doors, the
interlocks must operate such that the door leading to the aseptic area cannot be opened.
D. Heaters:All of the heating elements must be functional. It
is preferable to have them monitored continuously with ammeters in order that burned-out elements can be immediately detected.
E. Cooling Coils:To enable a faster cool-down cycle, the air is
often circulated across coolant coils.F. Belts:The belt speed is a critical operating parameter in
both continuous hot-air tunnels and flame sterilizers.
Recorders for charting the belt speed are recommended for units with adjustable speed settings.
G. Particulate Counts:
Particulate counts should be checked within
the containers before and after sterilization to
quantitate the particle load.
H. Chamber Leaks:
The perimeter of the doors for batch sterilizers
should be checked for air leakage while
operating.
QUALIFICATION TESTING
Upon completion of IQ and OQ efforts and
approval of the protocol, testing may begin.
The testing will include empty-chamber testing
for:
Heat distribution studies,
Loaded-chamber testing consisting of heat
distribution and heat penetration studies.
1) Component Mapping StudiesBefore conducting the loaded-chamber heat
penetration studies, component mapping should be conducted.
The studies help to determine the coolest point within a specific load and item.
2) Empty-Chamber TestingThe initial testing is performed on an empty
oven or tunnel to establish the uniformity of temperature distribution.
The thermodynamic characteristics of the empty unit are depicted in a temperature distribution profile.
3) Loaded-Chamber Studies
For validation purposes, the loads tested must be
representative of standard items and quantities.
Ideally, each size and type of material should be
tested by penetration studies.
For ovens, the time and temperature set points
should be reduced. For tunnels, the temperature set
point should be reduced and the belt speed
increased if possible.
4) Bio-Challenge/Pyro-Challenge Studies
The challenge should demonstrate the lethality
delivered by the cycle with either microorganisms
or endotoxin.
The challenge can be accomplished using
commercial strips or suspensions of B. subtilis
spores for sterilization or E. Coli endotoxin for
depyrogenation.
The concentration of the challenge for overkill
processes must demonstrate adequate sterility
assurance.
QUALIFICATION REPORT After the empty and loaded-chamber studies and bio-
challenge studies have been completed, the data must be analyzed to ascertain that all testing requirements have been achieved.
The results of the biochallenge studies and F value computation must demonstrate the required degree of lethality according to the protocol.
The following information should be provided in the process qualification validation report:
1. Protocol achievement
2. Summary of data
3. Deviations
4. Diagram
68