Features
Throughput: l00 tests/hour26 positions for samples30 positions for reagentsAuto identification of reagents by RFID technologyOn-board reagent coolingProbe with liquid level detection and collision protection automatic Antigen Excess detectionroad test menu, more than 35 assays available onDesktop PC along with analyzer
Auto IA500Auto IA500
System Description
MethodsOperative ModeThroughputSample PositionSample Containers
Sampling
Minimum Reaction Volume:Reagents
ControlReaction CuvettesReaction TemperatureOptic ModuleOptional AccessoriesPower RequirementsDimensions and Weight
Operating Environment Requirements
Control UnitConnection
Designed for clinical laboratories, Auto IA 500 is a random access bench top, fully automated immuno-Nephelometry SystemFixed time immunonephelometryBatch, random, STATUp to 100 tests/hour26 positions STAT sample availablePrimary tubes (diameter from 8 to 16mm ; height up to 100mm), Cups of different sizesSample volume: 6uL-20uL (1µL inr)Automatic sample probe washing (internal and external)Probe wash cycle is operator-programmable230µL30 positions available for reagent, Auto Reagent Identification , Reagent disk with non-stop cooling functionsAuto drawing of control Levey-Jennings and Westguard QC chart50 disposable , optic grade cuvettesConstantly and precisely monitored (37°C + 0.1)Laser 670 nm 1mWPrinter, Barcode reader100-240VA C, 50/60 Hz60 x 55 x 65cm (H x W x L)Weight: 60kgRoom temperature: 18-32°CRelative humidity: 30-85%Smart information management system with mass data storageCan be connected to LIS by Protocol HL7
Intelligent Software
Intuitive operator interfaceQuick view of real-time running statusLevey-Jennings QC chartsSTAT programming
Reagent tray
30 positions for reagentReagent disk with non-stop cooling functions Reagent can be auto-identified by RFID technology Ready to use, liquid reagent
Carefree Cleaning SystemProbe can be washed by external-inter-nal way
Smart Pipetting System3D transferring armProbe with liquid level detection and collision protectionAuto tracking of reagent residual to ensure the continuity of analysis
Precise Dispensing SystemTeflon piston technology, longer duration Dispensing CV% < 1%
Auto-IA5001
2
34
Sample tray
26 sample positions. Both standard blood collection tube and centrifuge tube can be placed in sample disk STAT samples can be inserted and it's prior to routine samples
Assays availableAssay
Immune Disease
Accute Phase Protein
Linear range
Cardiovascular Disease
Kidney Disease
Diabetes
IgAIgGIgMC3C4ASORF C1 IN KAPLAM
bCRPUsCRPCRPSAAHPTTRFCER AATAAG
0.9-380.0 mg/L0.22-19.00 mg/L1.6-152.0 mg/L3-200 mg/L0.15-3.80 g/L0.25-6.10 g/L0.1-2.5 g/L0.27-2.60 g/L0.28-4.10 g/L
ALBPABApoA1Apo BAT3D-DimerLp(a)
5-124 g/L0.04-1.34 g/L0.24-4.10 g/L0.24-5.20 g/L0.06-0.75 g/L0.28-15.70 mg/L60-1200 mg/L
mALBBMGAMGA1 MIgG U TRFUCys CRBP
9.5-230.0 mg/L0.4-13.0 mg/L0.28-7.20 g/L5-200 mg/L4-100 mg/L1-80 mg/L0.4-7.6 mg/L5-160 mg/L
HbA1c 2.8%-15.5%
0.23-5.10 g/L1.8-26.0 g/L0.18-4.60 g/L0.19-4.70 g/L0.06-2.20 g/L41-607 IU/mL20-306 IU/mL0.05-1.00 g/L0.88-12.20 g/L0.65-7.10 g/L