Instrument Washer Disinfector
AthertonPlatypus Series
Atherton is a dynamic and innovative oganization operating
nationally within Australia, with an office in each capital city, and
internationally through its network of agents throughout Asia and
the Pacific nations.
Established in 1889, Atherton is a privately owned
family business now in its fifth generation.
To find out more aboutAtherton Sterilizers and other infection control productsVisit our web sitewww.atherton.netAlternatively phone or fax any of our offices listed here:
Company Head OfficeMelbourne:A.E.Atherton & Sons Pty Ltd364 Darebin Road, ThornburyVIC 3071 AustraliaPh. 61 03 9497 2500Fax: 61 03 9497 1190
Brisbane:66 Raynham Street, SalisburyQLD 4107 AustraliaPh. 61 07 3277 1321Fax: 61 07 3875 1512
Sydney:23 Lidco Street, Arndell ParkNSW 2148 AustraliaPh. 61 02 9679 8911Fax: 61 02 9679 8922
Adelaide:Unit 22, 11-31 Port Wakefield RoadGepps Cross SA 5012 AustraliaPh. 61 08 8262 9044Fax: 61 08 8262 9344
Perth:Unit 3/1914 Beach RoadMalaga WA 6062 AustraliaPh. 61 08 9249 6166Fax: 61 08 9249 6353
Design and specifications may change without notice.TDS 002 Revision:0 01/09/07.
0805
Instrument Washer Disinfector
Platypus Series
Atherton
890WALL OPENING
REQUIRED FOR PASS THRUUNITS ONLY
25
1400(VENTED UNIT)
880
100
250SERVICES
140 790
1375SINGLE DOOR
1960PASS THRU
700SERVICES
CONTROLS
VENTVENT
1430WALL OPENING
REQUIRED FOR PASSTHRU UNITS ONLY(VENTED MODEL)
585DOOR
CLEARANCE
585DOOR
CLEARANCE
FINISHED FLOOR
HOT WATER
DEMINERALISED WATER
POWER
COLD WATER
CLEAN SIDE(PASS THRU)
DIRTY SIDE(PASS THRU)
CONTROLS
145 270
WASTE 'A' WASTE 'A'
WASTE 'B'WASTE 'B'
125
VENT
190
140
2100
(CO
ND
ENSO
R M
OD
EL)
2130WALL OPENING
REQUIRED FOR PASSTHRU UNITS ONLY
(CONDENSOR MODEL)
OP
TIO
NAL
CO
ND
EN
SOR
The Favourite Cycle function makes this an easy task. Six individual wash / decontaminate cycle formats can be programmed and saved. Pre Rinse, Wash, Disinfecting and Final Rinse temperature and time values can all be set to your personnel requirements.
The Atherton Platypus Series introduces a new era in washer / disinfectors. Utilizing colour touch screen operator interface technology, employing easily understood graphics the Platypus Series provides a powerful yet easy to use resource in the infection control process.
Detergent, neutralising agent, and instrument milk dosing, as well as demineralised water rinse, can all be programmed as required by the operator.
After the washing / disinfection process has been completed, the operator can choose to use the inbuilt drying feature, provided as standard on the Platypus Series, or transfer the load to an Atherton drying cabinet to free the disinfector for additional cleaning cycles. The electrically actuated door on the Atherton Platypus Series washer / disinfector can be opened and closed with the touch of a finger. When the start cycle icon is touched on the control screen, the door will automatically close and the cycle will commence. The door cannot be opened during the cycle process to prevent any injury to operators. At the end of the cycle, the door will open when the operator touches the cycle complete acknowledgment icon. The Platypus Series Washer / Disinfector is manufactured from stainless steel throughout, and can be supplied with single or pass through double doors. Wash, rinse and disinfecting water is circulated by a powerful 500 Lt/min stainless steel pump. The wash chamber and piping system is designed to gravity drain to waste, thereby minimising the possibility of bacterial growth from residual water within the system. Where an external vent from the chamber cannot be supplied, a condenser unit can be mounted on the top of the unit as an optional extra.
The Platypus Series control system provides complete freedom to the operator to choose from 3 user defined preset cycles - Surgical Equipment, Hospital Utensils or Anaesthetic Equipment, or program their individual cycle requirements.
Instrument rack: Capacity – 10 trays Dimension of tray – 485 x 255 x 50 mm Anaesthetic rack: Capacity - Corrugated breathing tubes – 8, maximum length 1500 mm Respiratory bag – 6 Endotracheal / Oropharyngeal tubing – 10 Instrument / miscellaneous rack – 1 (485 x 255 x 50 mm)
Interchangeable racks provide the flexibility to process surgical instruments, anaesthetic tubing and airway equipment, utensils and hollow ware.
An optional transfer trolley allows racks to be loaded and unloaded away from the unit if required, and provides convenient storage for racks not immediately required.
Technical Information. Power: 3 Phase & Neutral 415 V, 20 Amps / Phase Heating elements: 9 kW Total electrical load: 14.5 kW Water connection (hot and cold required): 15mm Waste connection: 40 mm copper Vent: 100mm diam.