Upload
others
View
2
Download
0
Embed Size (px)
Citation preview
Haemodialysis
Online Clearance MonitoringAssuring the Desired Dose of Dialysis
732
202
1/1
GB (1
,0 V
&B 0
5/07
) ©
Cop
yrig
ht 2
007
Fres
eniu
s M
edic
al C
are
Deut
schl
and
GmbH
Contents
1.Foreword
2.Dialysisdose 2.1 Standardmethodsofdeterminationofthedialysisdose 2.1.1 UreaReductionRatio(URR) 2.1.2 Kt/V ClearanceK Effectivetreatmenttimet UreadistributionvolumeV 2.1.2.1 DeterminationofKt/Vusingbloodsamples Formalureakineticmodelling Single-poolKt/V(Daugirdasformula) Single-poolvariable-volumeKt/V Double-poolKt/V(dpKt/V)anditsapproximation: EquilibratedKt/V(eKt/V) 2.2 Theclinicalrelevanceofdialysisdose 2.3 Recommendationsfordialysisdoseandthefrequency ofmeasurementofdelivereddose
3.OnlineClearanceMonitoring–OCM®
3.1 ThemethodofOCM®operation 3.2 OCM®inONLINEplusHaemodiafiltrationtreatments
4.Optimisationofdialysisefficiency 4.1 Bloodflow 4.2 Dialysistime 4.3 Dialyser(Membranesurfacearea,lowflux/highflux) 4.4 Dialysateflowrate
5.MeasurementofplasmasodiumwithOCM®
6.FrequentlyAskedQuestionsaboutOCM®
7.References
4
5555
7
1011
121217
1818192020
22
24
26
3
1. Foreword
OnlineClearanceMonitoring(OCM®)isastandardfeatureoftheFreseniusMedicalCareTherapySystem5008
thatprovidesanautomaticintradialyticmeasurementoftheeffective in-vivoureaclearanceK,thetotalcleared
bloodwatervolumeKt,thedelivereddoseofdialysisKt/Vandtheplasmasodiumconcentrationofthepatient.
WiththeOCM®itispossibletoeasilymonitortheseessentialparameterswithoutadditionalcostsduringregular
ONLINEHaemodiafiltrationineitherpre-orpost-dilutionmode,andduringhaemodialysistreatments.
TheOCM®helpsphysiciansandnursingstafftoensureanddocumentonaregularbasisthattherenalreplace-
menttherapymeetstherecommendedorrequiredqualitystandardswithouttheneedforadditionallaboratory
testsorexpenses.
ThisuserbrochureprovidesclinicalandtechnicalbackgroundinformationontheOCM®,describesitsoperation
andgivesrecommendationsfortheapplicationoftheOCM®.
4
2. Dialysis dose
Similar to the term “dose”, that defines the specific
quantityofamedicalproductordrugrequiredforthe
treatmentofapatient,thedialysisdosecanbedefined
as the quantity of dialysis treatment delivered for a
given period of time. In general the dialysis dose is
measuredbythecomparisonbetweenthebaselineand
finalconcentrationofadefinedsubstanceintheblood
ofthepatient.Themoreefficientthedialysissession,
thegreaterthereductionofthisgivensubstanceis.
Intheory,itispossibletodeterminehowefficientany
substancewhichispresentinthebloodhasbeenre-
movedduringdialysis.Eventhoughdifferentmethods
have been developed to measure the dialysis dose,
changesinureaconcentrationaremonitoredinorder
todeterminethedialysisdoseinstandardpractice.
2.1Standardmethodsofdeterminationofthedialysisdose
2.1.1UreaReductionRatio(URR)
Asimpleandcommonly-usedmethodofdetermining
thedoseofdialysisisthecalculationoftheUreaReduc-
tionRatio(URR).Thisinvolvesadirectcomparisonof
pre-andpostdialyticureaconcentrationsandshows
the percentage reduction of the urea concentration
duringdialysistreatment(1).
Inviewofthesimplicityofthemethod,URRisfrequently
usedtodeterminethedoseofdialysis,despitethefact
that the method has an in-built analytical weakness
whichmaycauseinaccurateresults.Incontrasttoother
methods,theURRmethoddoesnottakeintoaccount
thatureaisalsoremovedfromthebloodbyultrafiltration.
Cpost:postdialyticureaconcentration/Cpre:predialyticureaconcentration
Ultrafiltration leads toaveryefficient transferofurea
from the blood into the dialysate, but this does not
resultinadirectreductionoftheureaconcentrationin
theblood.Thisremovalofureaisnotaccountedforin
theURRmethod.ThegreatertheUFvolumeremoved
during dialysis, the more inaccurate the results of
dialysisdosecalculationbasedonURR(2,3).
2.1.2Kt/V
41:ThevariablesoftheKt/Vformula
TheformulaKt/V,whichisthemostcommonlyused
index of the adequacy of the dialysis treatment, is
a mathematical representation involving the blood
volume which has been completely cleared of urea
during a specified dialysis and the urea distribution
volumeofthepatientrequiringdetoxification.41
The variables of the Kt/V formula:
Clearance K
Aswell asbeing used as adiagnostic parameter in
renalfunctiontests,theterm“clearance”isalsoused
in renal replacement therapy.Theprimaryparameter
used to determine the volume of blood which has
beenpurifiedduringtreatment is theeffective in-vivo
clearance Keff (mL/min)42. Clearance is defined as
the (hypothetical) volume of blood which has been
totally “cleared” of a given substance each minute;
clearanceisthusexpressedasmL/min.
K x
V
t
Clearance(Manufacturer)
Dialysis time(Nephrologist)
Distribution volume(Patient)
5
URR = 100 1 - Cpost
Cpre ( )
In-vivo must be differentiated from
in-vitro clearance:
In-vitroclearancerepresentsthepurelydiffusiveclearance
achievedbyadialysertestedunderdefinedconditions
(asspecifiedinstandardEN1283)usingstandardised
aqueoussolutions.TheEN1283standardrequiresthat
measurementsbeconductedatapredefineddialysis
fluidflowratewithoutanyultrafiltration(thatis,without
aconvectiveclearancecomponent).Onlyonthebasis
ofthesenon-patient-specificlaboratorymeasurements
isitpossibletocompareclearancevaluesandthusthe
relativeefficiencyofindividualdialysers.In-vitroclear-
ance, i.e. the efficacyof adialyser under laboratory
conditions,isthusprimarilydeterminedbythedesign
featuresofthedialyseraswellasthemembranechar-
acteristics.
In-vitroclearancesofdialysers,ontheotherhand,are
measuredundertheactualdialysistreatmentconditions
involvingthepatient,generallybymeansofanalysisof
thebloodpriortoandafterpassingthedialyser:theterm
“wholebloodclearance”isthususedinthiscontext.
In-vivoclearancesaresubstantiallydifferentfromin-vitro
clearancesas,duringthedialysisofwholeblood,the
effectsof thecorpuscularbloodconstituents (mainly
theerythrocytes),thedissolvedplasmaproteins,andthe
adsorption of plasma proteins on the dialysis mem-
brane(the“secondarymembrane”)considerablyalters
thediffusionpropertiesofthedialysismembraneeven
underotherwisecomparablecircumstances.
In-vivoclearanceisthusinevitablylowerthanthestan-
dardisedin-vitroclearance,butdoesreflecttheactual
conditionsofthetreatmentandisthusofcentralimpor-
tanceforthedeterminationofdialysisefficiency.
Effective dialysis time t
Effectivedialysistimetistheactualdurationofdiffusive
blood detoxification (time of dialysis fluid flow in the
dialyserwiththebloodpumpoperating).Interruptions
to the dialysis treatment for therapeutic or technical
K=clearance(mL/min),QB=effectivebloodflow(mL/min),CBin=concentrationofinflowingblood,CBout=concentrationofoutflowingblood
K = QB CB in – CB out
CB in
6
In-vivo Clearance K influencedby:dialyser,effectiveblood(water)flow(considering
haematocrit),ultrafiltration,recirculation,dialysisfluidflow
Effective dialysis time t sometimesreducedduetoalarmsandbypassconditionsorprematureterminationofthetreatment
Urea distribution volume V dependentonbodysizeandcomposition,weight,ageandgenderofthepatient
42:ThedoseofdialysisKt/V isa resultof theeffective in-vivoureaclearanceK,theeffectivedialysistimetand theureadistributionvolumeV.
reasons are considered in the effective dialysis time
42. The effective dialysis time is, in most cases,
shorterthanthetotaldurationoftreatmentprovided.
Inanobservationalstudy,Seguraetal.havedemon-
stratedthat themeanprescribeddurationofdialysis
is generally not achieved inpractice. Thus, over the
period of 1 year, the loss of time and therefore of
efficiencyperpatientwasequivalenttothatof7whole
dialysissessions(4).
Urea distribution volume V
TheureadistributionvolumeVisequivalenttothetotal
bodyfluid,consistingoftheproportionofwaterinblood
(7%)aswellasthe interstitial (31%)and intracellular
volumecompartments(60%).Uponcommencementof
dialysis,ureaishomogeneouslydistributedthroughout
thebody;hence,more than90%of theureawhich
accumulatesinthehumanbodyisnotpresentinblood,
butintheinterstitialandintracellularvolumecompart-
ments. Thus, only with continuous diffusion of urea
fromthesecompartments into thebloodand further
transport into the extracorporeal circulation is the
largestproportionofureapresent in thebodymade
availablefordialysis.
Thelargerthebloodvolume(highbloodflowrate)which
reaches thedialyserduringa treatment sessionand
thelongerthedurationofthetime-dependentdiffusion
fromtheinterstitialandintercellularcompartmentsinto
theblood(adequatetreatmenttime),themoreefficient
thedialysistreatmentwillbe.
AmajoradvantageoftheKt/Vequationisthefactthat
ittakesintoaccounttheindividualbodyweightofthe
dialysispatient.PatientswiththesameKt/Varedetoxified
tothesameextenteveniftheyhavemarkedlydifferent
bodyweights.Dialysistreatmentscanthusbecompared
intermsoftheeliminationeffectusingtheKt/Vmethod.
The following examples should help explain the
use of the Kt/V concept:
AssumingKtis20litres,thismeansthat20Lofblood
haveleftthedialyserafterbeingclearedofurea.How-
ever,thisinformationdoesnotyettelluswhichdialysis
dosethepatienthasreceived.Inthecaseofapatient
withaureadistribution volumeVof30L, acleared
blood volumeof 20 L hasquite adifferentmeaning
thaninthecaseofapatientwithaureadistributionvol-
umeof40L.IfKt/Visequalto1,thismeansthatthe
volumeofbloodwhichhasbeencompletelycleared
ofureainthedialyserisexactlyequivalenttotheurea
distributionvolumeof thepatient. Inapatientwitha
distributionvolumeof40L,Ktmustbeequalto40L
sothattheresultoftheformulaKt/Visalso1.IftheKt
valueisonly20L,theKt/Vvaluewouldbe0.5.
EvenifKt/Visequalto1,thisdoesnotmeanthatthe
total body fluid or the total urea distribution volume
ofthepatient isfreeofurea,butonlythat,usingthe
exampleabove,40Lbloodhaveleftthedialyserafter
beingfreedofurea.IfKt/Vis1.0,aureaconcentration
of approximately 36 – 37%, relative to the baseline
value,remainsafterdialysis.
2.1.2.1DeterminationofKt/Vusingbloodsamples
Therearevariousmethodswhichcanbeusedtode-
termineKt/V inpractice.Thesediffer in termsof the
underlyingmathematicalmodelsusedtopredictand
determineureakinetics(inotherwords,thechangesin
thebloodureaconcentrationofthepatientovertime).
Somemethodsinvolveverycomplexcalculationsand
requireappropriatesoftwareprogrammes.
7
Formal urea kinetic modelling
Themostcomplex,butatthesametimemostprecise
methodofdeterminationofKt/V is theso-calledfor-
malureakineticmodelling(5,6).This requiresaccurate
measurementof:
•ureaorBUN(BloodUreaNitrogen):concentrations
priortoandafterthefirstdialysissessionoftheweek,
andureaorBUNconcentrationspriortothesecond
dialysissessionoftheweek(assumingthreetreat-
mentsareprovidedperweek)
• thebodyweightofthepatientpriortoandafterthe
firstdialysissessionoftheweek
• theactualeffectivetreatmenttime(nottheprescribed
durationofdialysisandnotthetimefromconnection
toanddisconnectionfromthedialysismachine)
• theeffectiveureaclearanceofthedialyserasmeasured
atthedialysiscentre(notthevalueforin-vitroclearance
quotedbythemanufacturer).
Ureakineticmodellingisareproduciblemethodwhich
notonly allowscalculationof thedialysisdoseKt/V,
butalsoof theparameternPCR (normalisedProtein
CatabolicRate),animportantmarkerofthenutritional
statusofthepatient.Moreover,ureakineticmodelling
allowstheexactcalculationofthepatient-specificurea
distributionvolumesothattheindividualdialysispre-
scriptioncanbegivenonthebasisofmeasureddata.
Themaindisadvantageof theureakineticmodelling
methodisthelogisticalaspect.Someparameters,such
astheeffectiveclearanceachievedbyadialyser,are
difficult tomeasure in clinical practice and the effort
required to obtain and process the necessary data
canbeconsiderableinlargedialysiscentres.
Morerecently,atwopointureakineticmodelhasbeen
validated which is based on only two measurements
ofureaorBUNinsteadofthree,priortoandafterthe
dialysis session and which supplies precise results
withamarkedreductionoflogisticaleffort(7).
Single-pool Kt/V (Daugirdas formula)
The best alternative to the urea kinetic modelling
method, also based on analysis of blood samples,
isprovidedbytheformulaforthecalculationofKt/V
developedbyJ.T.Daugirdas(8).
Single-pool variable volume Kt/V
Mostwidelyusedatpresent is the following formula
forcalculatingKt/V:
Thismathematicalmodel,whichusesthenaturalloga-
rithm tocalculateKt/V,providessufficientlyaccurate
resultsoverthefullrangeofstandardKt/Vvalues(9,10).
Inaddition,theDaugirdasformulatakesintoaccount
thechangeinpatientvolumecausedbyultrafiltration
anditscontributiontoconvectiveureaelimination.
ln:naturallogarithm,R:ratioofpostdialytic÷predialyticBUN,t:effectivedialysistimeinhours,Uf:ultrafiltrationvolumeinlitres,W:weightofthepatientafterdialysisinkg.
43:Bloodureanitrogenconcentrationatthebeginning(1),attheend(2) and60minutesaftertheend(3)ofaHDtreatment
Kt/Vsp = -ln (R – 0.008 x t) + (4 – 3.5 x R) x Uf/W
BUN
conc
entr
atio
n [m
g/dl
] ICV
ICV
ICV
ECV+blood
ICV: Intracellular volume
ECV: Extracellular volume
ECV+blood
ECV+blood
Time [min]
Single pool variablevolume Kt/V (spKt/V)
Equilibrated Kt/V(eKt/V)
“Rebound”
End of treatment
8
Double-pool Kt/V (dpKt/V) and its approximation:
equilibrated Kt/V (eKt/V)
Asinthecaseoftheother(notdiscussedhere)single-
pool models, the Daugirdas formula considers the
body,forthepurposesofmathematicalmodellingas
a single unified fluid pool, and does not take into
account the differences in rates of urea transfer
between the three fluid compartments blood plas-
ma, extracellular space and intracellular space. This
difference inthetransferrateofurea isdescribedas
thedouble-pooleffect.
Withincreasingefficiencyofthedialyser,itispossible
thatureaiseliminatedmorerapidlyfromthebloodthan
itcandiffusefromthecellsintotheblood.Duetothis
relativedelayofureamovementfromcellsintoblood,
the intracellular compartment becomes a non-equilib-
ratedreservoirforurea,andthisaspectisnottakeninto
accountinthesingle-poolmodelsforureakinetics(9,12,13).
Hence,oncompletionofadialysistreatment,theintra-
cellular urea concentration is greater than in blood
plasma,andthediffusionofureafromcellsintoblood
nevertheless continues for approximately 30 – 60
minutesaftertheendoftreatment43.Thisphenom-
enon is called urea rebound(14,15). If this intracellular
ureapoolislargeoncompletionofdialysisandisnot
takenintoaccount,theeffectivedialysisdoseKt/Vwill
beoverestimated.Thiseffect isparticularlymarkedif
high performance dialysers are used in combination
withashortdurationofdialysis(16).
In addition to the relative delay of transfer of urea
betweenthevariousbodycompartments,themarked
differencesinbloodperfusionoforgansalsocontribute
tothereboundeffect.Itshouldbeborneinmindthat
70%oftotalbodywater,andthusalsourea,ispresent
inbodyorganswitha relatively lowbloodperfusion,
suchasskinandmusclesatrest.Astheseurea-rich
organshaveonlyarelativelylowbloodflow,theircon-
tribution to the totalplasmaureawhich reaches the
extracorporeal circulation is less than thatof organs
with lowureaconcentrationsbuthavinghigherrates
ofbloodperfusion.Organswith lowbloodperfusion
thereforealsoconstituteafurtherureareservoir.Inthe
initial 30 minutes after completion of dialysis, the
concentrationequilibrationagainleadstoarebound,
thusanincreaseinbloodureaconcentration.
The extent of total urea rebound can differ greatly
betweenindividualpatients.Inastudyconductedby
Leblancetal.(15), themean rebound, thepercentage
increaseinpostdialyticureaconcentration30minutes
after dialysis, in comparison with that immediately
aftercompletionofdialysis,was17%.
Ifthisreboundeffectistakenintoaccountinthecal-
culationofthedialysisdose,theresultantKt/Vvalueis
referredtoasequilibratedKt/V(abbreviatedtoeKt/V).
Onaverage,equilibratedsingle-poolKt/V(eKt/V)is0.2
lowerthannon-equilibratedsingle-poolKt/V(17,18).
InordertoobtainanexactvalueforequilibratedKt/V,
theBUNconcentrationshouldideallybemeasured30
minutes after completion of dialysis, but this is, un-
fortunately,notpracticableinroutinepracticesofout-
patient haemodialysis centres. In order toovercome
thisproblem,various formulaehavebeendeveloped
which,afterathirdblood(urea)samplehasbeentaken
during the dialysis session (usually after 70 minutes
dialysis),allowthepostdialyticequilibratedureacon-
centration to be approximatively calculated(16,19). The
additionaleffortrequiredtoobtainathirdblood(urea)
sample has meant that this method has not been
widelyaccepted.
Daugirdashasalsodevelopedaformulawhichallows
theresultsofcalculationofsingle-poolKt/Vtobeextra-
polated to equilibrated Kt/V(20). Formula 1 (artKt/V) is
applicable if an arterial postdialytic urea value from
anarterio-venousaccessisavailable,whileformula2
(venKt/V)shouldbeusedifamixedvenousureavalue
fromaveno-venousaccessisavailable:
9
Thekeyadvantageoftheseformulaeisthatonlytwo
bloodsamplesarerequiredforthecalculationofthe
equilibratedKt/V.
The following table shows the relationship between
single-poolKt/V(spKt/V)valuesandequilibratedKt/V
values(eKt/V)relativetothedurationoftreatment.
2.2 The clinical relevance of the dialysis dose
Ifdialysisdoseisdetermined,itispossibletoprescribe
andsubsequentlymonitorthedialysistreatmentonthe
basis of clinical parameters. A significant correlation
betweentheaveragedelivereddialysisdoseandpatient
mortalityrateshasbeendemonstratedinmanyclinical
studies(1,21–29).All these studies reiterate the fact that
thehigherthedelivereddoseofdialysis,thelowerthe
patientmortalityrates.
TheNCDSstudy(30)wasthefirstlong-termstudywhich
investigatedthecorrelationbetweendialysisdoseand
therapeuticoutcome.Theresultsnotonlydemonstrated
astatisticalrelationshipbetweenureaeliminationand
mortality,butalsoallowedGotch&Sargenttodevelop
theureakineticmodellingmethodanddefinetheKt/V.
In2002,Portetal.publishedanobservationalstudy
ofmajorimportanceontheeffectofdelivereddialysis
doseontheriskofmortality(31).
Inthisretrospectivestudy,thedatafor45.967USpa-
tients who had commenced chronic haemodialysis
therapybetweenApril1997andDecember1998were
examined44.
Inordertonotmasktherelationshipbetweenmortality
riskanddialysisdosebytheseparateriskfactor“body
weightofthepatient”,patientswereinitiallysortedinto
threebodyweightgroups–small,mediumandlarge.
The relationship between mortality risk and dialysis
dosewasthenanalysedseparatelyforeachofthese
body weight groups. The result was confirmed over
thewholerangeofdialysisdosesanalysed:thehigher
thedelivereddialysisdose,thelowertherateofpatient
mortality.Thisstudyalsoshowedthatthereisnospecific
maximum dose above which no further decrease in
patientmortalitycanbeachieved.Thismeansthatduring
each dialysis treatment, it is advisable to deliver the
highestdialysisdosepossible.
Otherclinicalaspects,suchastheerythropoietin(EPO)
requirement,canalsobeinfluencedbythedialysisdose
Kt/V. Inastudypublished in2001, theEPOrequire-
mentoftwopatientgroupswithdifferentKt/Vvalues
44:Dialysisdoseandbodymassindexarestronglyassociatedwithsurvival in haemodialysis patients. Retrospective analysis of data from 45.967 US hemodialysispatientsstartingdialysisduringApril1997 throughDecember 1998(adaptedfrom(31)).
Small
MediumLarge
Annu
al a
djus
ted
deat
h ra
tes
[%]
* p<0.05 compared to next lower URR group
Size of circle reflects number of patients per group
Dose Category
*
**
*
*
*
**
*
Small BMI < 23.2 kg/m 2
Medium BMI 23.2 – 27.8 kg/m2
Large BMI > 27.8 kg/m 2
Ratio of single-pool Kt/V (spKt/V) and equilibrated Kt/V (eKt/V)as a function of treatment time (t)
eKt/V
spKTt/V 2.0 hrs 2.5 hrs 3.0 hrs 3.5 hrs 4.0 hrs 4.5 hrs 5.0 hrs
1.0 0.73 0.79 0.83 0.86 0.88 0.90 0.91
1.1 0.80 0.87 0.91 0.94 0.97 0.98 1.00
1.2 0.87 0.94 0.99 1.02 1.05 1.07 1.09
1.3 0.94 1.02 1.07 1.11 1.14 1.16 1.17
1.4 1.01 1.09 1.15 1.19 1.22 1.24 1.26
1.5 1.08 1.17 1.23 1.27 1.31 1.33 1.35
1.6 1.15 1.25 1.31 1.36 1.39 1.42 1.44
Table 1:
10
T:dialysistreatmenttimeinhours,t:dialysistimeinminutes
(1) equilibrated or eKt/V = artKt/Vsp – (0.6 x artKt/Vsp/T) + 0.03 (2) equilibrated or eKt/V = venKt/Vsp – (0.47 x venKt/Vsp/T) + 0.02
wascompared(32).Measuredhaematocritwas identi-
cal,at35±1.5%, inthetwoinvestigatedgroups,A
andB,althoughgroupAreceiveddialysisatalowKt/V
<1.2andgroupBreceiveddialysisatahighKt/V>1.4
45.WhilegroupA(Kt/V<1.2)hadanEPOrequirement
of183±95U/kg/week,patientgroupBwithitshigh
Kt/V (>1.4) required only 86 ± 33 U/kg/week. This
studyshowsthatifasufficientlyhighdialysisdoseis
delivered,thiscanhaveapositiveeffectontheoverall
clinicalconditionof thepatient,which, in turn, is re-
flectedinareductionoftheEPOrequirement.
2.3 Recommendations for dialysis dose and the frequency of measurement of delivered doseAsageneralrule,itisthetaskofthetreatingnephro-
logist,toprescribetheformanddoseofdialysistreat-
ment,taking intoaccountthespecificclinical factors
oftheindividualpatient.
Averycomprehensivecollectionofrecommendations
fortheprescriptionofdialysiswasfirstprovidedbythe
USNationalKidneyFoundation (NKF) in theirguide-
linesonadequatehaemodialysistreatment.TheDOQI
guidelineswerepublished in1997andsubsequently
updated as the K/DOQI guidelines (Kidney Disease
Outcomes Quality Initiative) in 2000(33). The European
BestPracticeGuidelinesforHaemodialysis(EBPGPart1)
were introduced in 2002 and represent the current
consensus on best practices and dialysis adequacy
amongsttheEuropeannephrologicalcommunity(67).
The respective recommendations regarding dia-
lysis dose are specified as follows:
•The delivered dialysis dose should be regularly
measuredandmonitored–atleastmonthlyusinga
standardisedmethod
•Thedelivereddialysisdoseshouldbeexpressedas
equilibratedKt/V(eKt/V)orassingle-poolKt/V(spKt/V)
•Theminimumdosepersessiononathrice-weekly
scheduleshouldbeeKt/V≥1.20(spKt/V~1.4)
TheEBPGforHaemodialysisareconsideredtobethe
referenceguidelines inEuropeandhavebeenadop-
tedortakenasthebasisforthepreparationofnational
guidelines.
Ithasbecomeoneofthemainaimsofhaemodialysis
treatment,particularlyintheUS,todeliverahighKt/V
dialysisdose. IntheUS,theaveragedeliveredKt/V
dialysisdosecontinuouslyincreasedintheperiod1986
to1999from0.9totoday’sfigureofgreaterthan1.4.
TheprescriptionofanadequateKt/Vnotonlyrepre-
sentsthemathematicallystandardisedspecificationof
averifiable therapygoal,but isaqualitycriterion for
dialysistherapyitselfbecauseofthefrequentlydescri-
beddirectcorrelationbetweendialysisdoseandpatient
mortality(1,21–29).Theregularmeasurementofdelivered
Kt/V is thus an important aspect of medical quality
assurance.Todate,however,theneedtoextensively
sample urea concentrations, and the associated
logistical and financial requirements, has made it
difficult toconductclosemonitoringofdoseoreven
tomeasuredelivereddoseineachsession.Asanauto-
matedprocedure,whichisnotbasedonaurea/blood
sampling method, the Online Clearance Monitoring
(OCM®)hasmadeitpossibletomeasuredialysisdose
easily without additional expenses during on-going
dialysistreatmentsateverysession.
45:TheinfluenceofahighKt/VontheweeklyEPO-dose(adaptedfrom(32)).
EPO
dose
[U/k
g/w
eek] Group A
Hct: 35%
Group BHct: 35%
EPO dose
11
The Online Clearance Monitoring (OCM®) is a stan-
dardfeatureintegratedintheFreseniusMedicalCare
5008TherapySystem.Itprovidesautomaticintradia-
lyticmeasurementofthedelivereddialysisdoseKt/V,
theeffective in-vivoureaclearance,theaccumulated
cleared plasma volume Kt and the plasma sodium
concentration of the patient. OCM® can be applied
during regular ONLINEplus haemodiafiltration treat-
mentsinpreorpostdilutionmodeaswellasduring
standardhaemodialysis.
ThedialysisdosedeterminedbytheOnlineClearance
Monitoring(OCM®)isequivalenttoasingle-poolKt/V
(spKt/V).
3.1 The method of OCM® operation
To achieve the objective of developing a low cost
methodofmonitoringclearance,itwasnecessaryto
moveawayfromthecost-intensiveconceptofenzy-
maticureaanalysis. Insearchingforanalternative,a
substancewasconsideredwhich ispresent in large
quantities in the dialysis fluid and where changes in
concentrationscanbemeasuredby the sensors in-
stalled within the dialysis machine – namely ionised
sodium.Sodiumionsrepresentthelargestproportion
offreelymobileelectrolytesinthedialysisfluidandtheir
3. Online Clearance Monitoring – OCM®
OCM®screenduringaHDFtreatment
concentrationessentiallydeterminesthetotalconduc-
tivityofthedialysisfluid.
Althoughthesmall,positively-chargedsodiumiondif-
fersfromthenon-chargedand largerureamolecule,
bothparticlesexhibitcomparable in-vitroand in-vivo
diffusion characteristics across a synthetic dialysis
membrane, i.e. their specific diffusion coefficient is
almost identical at 37°C (Na+: 1.94 x 10 – 5 cm2/s,
Urea:2.20x10–5cm2/s46(34)).Under realdialysis
conditions,thedifferenceinclearanceisevensmaller
than thedifferenceofdiffusioncoefficientssince the
clearanceislimitedbybloodanddialysateflowrates
andnotby thediffusionprocessacross thedialyser
membrane.
Bymeansofindirectdeterminationofionconcentra-
tions in the haemodialysis solution (measurement of
conductivityattheinflowandoutflowofthedialyser)it
istechnicallypossibletodeterminethediffusionprofile
ofsodiumionsacrossthedialysismembraneandthus
calculatethedialysanceorionicclearance(D).Onthe
basisofthedialysanceofsodiumions,the“diffusibi-
lity”ofureathroughthemembrane(permeability)and
thusureaclearancecanbedetermined(35).
For accurate measurement of sodium dialysance,
conductivity sensor cells have been installed in the
inflowandoutflowlinesofthedialysertomeasurethe
conductivityof thedialysisfluid (prior to thedialyser)
andinthedialysate(afterthedialyser)(36).
46:OCM® isbasedon theequivalenceofelectrolyteandureaclearance(adaptedfrom(34))
Elec
trol
yte
clea
ranc
e [m
l/min
]
Urea clearance [ml/min]
aqueous solutions
Coefficients of diffusion at 37°C[cm2/s]
Na+ Urea1.94 x 10-5 2.20 x 10-5
12
In order to achieve adetectablediffusionof sodium
ions across the membrane, the otherwise moderate
diffusiongradientof sodiumbetween thebloodand
dialysisfluidsidesofthedialysermustbetemporarily
increased.
Forthispurpose,thedialysismachineinducesashort
termpulsetoincrease(ordecrease)thesodiumcon-
centration in thedialysisfluid, therebyresulting inan
increase in diffusion of sodium ions into the blood
compartment or in the reverse direction48. Assu-
mingthattheconductivitypulsedoesnotexceedthe
specifiedconductivitylimits,alternatepulsesincrease
anddecreaseconductivitytoensurethatthesodium
balanceremainsasneutralaspossible(37–44).
Thisshort-termincreaseintheconductivityofthedia-
lysisfluidpriortoenteringthedialyserissubsequently
reducedbydiffusionofaportionof thesodium ions
across the dialysis membrane into the blood of the
patientduring thepassage through thedialyser.The
dynamicinputconductivitysignal(pulse)atthedialyser
inflow is continuously monitored by the conductivity
sensorinstalledatthatposition:thesignalattheoutflow
of the dialyser is registered by an equivalent sensor
positionedthere.Therelativeareasunderthecurves
for the two recorded conductivity signals reflect the
diffusionofsodiumionsacrossthedialysermembrane.
Thelowerthevalueofoutflowconductivitycompared
toinflowconductivity,thegreatertheamountofsodium
thathasdiffusedfromthehaemodialysissolutioninto
the blood compartment as a result of the diffusion
gradientor, inotherwords, themorepermeable the
dialysismembraneforsodiumis(45).
Asurea–aspostulatedabove–hasadiffusionprofile
similartothatofthesodiumion,ureaclearancecanbe
determined(usingappropriatecorrectionfactors)irre-
spectiveoftheactualconcentrationofureaintheblood.
Steiletal.wereabletodemonstratein1993thatthe
dialysanceisdirectlyproportionaltoureaclearance(35).
Theyshowedhowcloselymeasured ionicclearance
correlateswithmeasuredureaclearance.Inavalidation
studyconductedin2001,Kuhlmannetal.produced
clinical evidence that dialysance also correlates with
ureaclearancein-vivo,andthatitcanbedetermined
accuratelywithina very lowanalytical error rangeof
only±5%bytheOCM®49(39).Asacomparison:the
analytical error for the determination of urea in the
laboratoryisapprox.±7%.Inbothoftheabove-men-
tioned investigations, itwaspossibletodemonstrate
thatthecorrelationbetweenelectrolyteclearanceand
theclearancevaluesforureadeterminedusingstan-
dardmethodsisalmost ideal.Themajorityofresults
lie,withsufficientprecision,onthelineofidentityand
thecorrelationvaluer isclosetotheidealvalueof1
(if all values for ionic clearance and laboratory data
48:OnlineClearanceMonitoring isoperatingwithashort-termdynamic conductivitypulse
Inle
t/Ou
tlet c
ondu
ctiv
ity
[mS/
cm]
Dialysis Time [secs]
ytivitcudnoc tel
Rec.
tuOytivitcudnoc telnI
Recirculation
13
QDQB
Cd inCb out
Cd outCb in
Dialyser
Blood pumpBlood inflow
Blood outflow
Fistula
Dialysate outflow
Dialysis fluid inflow
Conductivity cell “out”
Conductivity cell “in”
47:Thediagramshowsthelocationofthe2conductivitycellsoftheOCM®
clearancewereexactly identical,all resultswouldbe
ontheidealline).Themajorityofvaluesarewithinan
errorrangeofonly±5–6%.
ThroughfurtherimprovementsoftheOnlineClearance
Monitoring®method,Goldauetal.reportedinafurther
seriesofanalysisin2002animprovedlevelofprecision
ofbelowthe5%errorlimit(45).
Theprecisionoftheconductivitymonitoringisdeter-
mined,amongotherfactors,bythestabilityoftheplasma
sodiumconcentrationduringmeasurement.The less
variation in the plasma sodium concentration of the
patientduringthemeasurementcycle,themorepre-
cisetheresult is.Forthisreason,thetimeintervalof
conductivity variation is kept as short aspossible in
theOCM®technique.
Withthiscomparativelyshort(butstillsufficientlylong)
measurementofdialysateconductivityatthedialyser
outflow,notonlythesodiumdialysanceofthedialyser,
butalsothenegativeeffectsoftotalrecirculation(fis-
tula recirculation plus cardiopulmonary recirculation)
canbedetermined48.Evenaftertheconductivitypulse
inthedialysatehasfadedout,recirculatingbloodtrans-
portsrecirculatingsodiumionstothedialyserwhich,
dependingontheconcentrationintheblood,willdiffuse
backintothedialysate.Consequentlythisresultsina
lessfavourableratiooftheareasundertheconductivity
curves,andthustoareductioninclearanceequivalent
tothecontributionmadebyrecirculation.Althoughitis
notpossibletoproduceactualfiguresforrecirculation
fromtheOCM®,theeffectistakenintoaccountinthe
calculationofeffectivein-vivoclearance.
TheOCM®determinesclearance “online”during the
entire treatment period, i.e. with no time lapse and
atpredefinedmeasurement intervals.Thenumberof
measurementcyclestobeperformedduringthetreat-
mentcanbepre-selectedwithincertainlimits(see5008
Operator’sManual).Inaddition,theOCM®iscapable
ofdetectingadjustmentstobloodand/ordialysateflow
rateswithin1minuteandthenimmediatelyrecalculates
thecorrespondingnewclearancevalues.Thismeans
thattheeffectsofanyalterationtotreatmentparameters
on actual clearance can be continuously monitored
duringanongoingsession.
Inaddition to theureaclearanceK, thedata for the
currenteffectivedialysistimetareobtainedfromthe
FreseniusMedicalCareTherapySystem5008andare
included in the calculation of the dialysis dose. The
effectivedialysistimetisdefinedasthetimeinwhich
physicaldialyticprocessesbetweenthebloodofthe
patient and the dialysis fluid occur. The time of any
interruption of treatment for technical or therapeutic
reasonsandthedurationofpreparationandreinfusion
procedures are not included: only the “true” dialysis
time(diffusiontime)istakenintoaccount.
Onthebasisofthetwomeasuredparameters–urea
clearance and effective dialysis time – the OCM®
calculates the cumulative water volume contained
withinthebloodKxt(inlitres)whichhasbeencleared
ofurea,which isequivalent to thatproportionof the
circulating blood from which all urea has been re-
movedduringdialysis:Kxt=x(L).
14
49: Correlation of electrolyte clearance measured by the OCM® and bloodureaclearance
Electrolyte clearance OCM® [ml/min]
Bloo
d ur
ea c
lear
ance
[ml/m
in]
Error K: ± 5%
It isnotonlypossibletoderivetheVofapatientby
calculation,but thisparametercanalsobeprecisely
measured. In clinical routine, the bioimpedance
spectroscopytechniqueortheureakineticmodelling
method described in section 2 above can be used.
Oncethismeasuredvalueisavailable,itcanbedirectly
entered into the data-entry menu of the OCM®: the
informationnecessary for theanthropometric formulae
(age,gender,etc.)isthen,ofcourse,nolongerrequired.
TheanthropometricformulaeallowVtobesimplyand
quicklyestimated,buttheresultsrepresentonlyastan-
dardised mean value for the population as a whole.
Deviationsattributabletoindividualfactors,suchasan
above-averageproportionoffatormuscletissue,are
notaccountedfor.Insomecases,theremaybemarked
differencesbetween thevalue forVcalculatedusing
theWatsonformulaandthemeasuredvalueforV.
Kloppenburgetal.havedemonstrated,inacompara-
tivestudy,thattheresultsforureadistributionvolume
VcalculatedusingtheanthropometricWatsonformula
regularlyexceedthemeasuredVofthesamepatient
by26%(50).Thus,ifthevalueforVobtainedusingthe
Watsonformulaisusedinsteadofameasuredvalue
forcalculationofKt/V,theachievedKt/Visfrequently
underestimated.
InorderthattheKt/Vvaluecanbedeterminedbydi-
vision of the result for K x t by the urea distribution
volumeV,theoperatormustenteranappropriateva-
luefortheureadistributionvolumeVoftheindividual
dialysispatient.
The easiest method of estimating urea distribution
volumeV is touse theanthropometric formulaede-
veloped by Watson(48), Hume-Weyers(49) and Mellits-
Cheek(70), the latter being especially adapted for
children below the age of 16. All formulae can be
selectedintheOCM®menuofthe5008.TheWatson
formulacalculates theVofapatienton thebasisof
bodyweight,height,ageandgender411.
The Hume-Weyers formula requires information on
bodyweight,heightandthegenderofthepatient;in-
formationonageisnotrequired.Forinformationonthe
Mellits-CheekformulapleaserefertotheFAQsection.
411:AcalculatortoestimatetheureadistributionvolumeVaccordingto anthropometricformulaeisintegratedintheOCM®
15
410: To calculate theKt/V theOCM®measuresKand the5008Therapy Systemmeasurestheexacteffectivedialysistimet;theuserhastoenterthe patient’sV
Effective clearance:accurately knownfrom OCM® (±6%)
Effective dialysis time:precisely known fromthe 5008
Normally unknown
V from anthropometric data (Watson,Hume-Weyers ...),large deviationslead to an approximate Kt/VV from a previous Urea Kinetic Modelling procedure (DCTool ), leads tothe best results for Kt/V (± 0.1 Kt/V)V from other methods, e.g. bioimpedance
K x
V
t
MaleVurea = 2.447 - 0.09516 x age + 0.1074 x height + 0.3362 x weight
FemaleVurea = -2.097 + 0.1069 x height + 0.2466 x weight
ureaprofileofahaemodialysispatientbasedontwo
blood samples (pre andpostdialytic) using the sing-
le-pool variable volume method(7). Important data
relevant to the dialysis and nutritional status of the
patient, such as urea distribution volume, Kt/V and
proteincatabolicrate(PCR)arederivedfromthis.The
DCTool programme differs from the standard urea
kineticprogrammesinthat,insteadofcalculatingtheo-
reticalclearanceonthebasisofbloodanddialysate
flowrates, theeffective in-vivoclearance–averaged
for the treatment as a whole – as measured by the
OCM®isusedformodelling.
Forthedeterminationoftheureadistributionvolume
ofapatientusingtheDCTool,thebasicpatientdata
and information on treatment regimen needs to be
entered.Thenonedialysis treatment isselectedand
the relatedmain treatment andOCM®data (i.e. the
averageeffective in-vivoclearance) togetherwiththe
pre-andpostdialyticureavaluesfrombloodsamples,
areenteredintheDCToolprogramme.Withthisinfor-
mation,theDCToolconductsanureakineticanalysis
anddeterminesaprecisevaluefortheureadistribution
volumeV;apartfromotherveryinterestingdoseand
nutritionalparameters.OncethisassessedvalueforV
isavailable, itcanbeentereddirectly intotheOCM®
template,savedonthePatientCardandishenceavail-
ableforsubsequenttreatments.ThisenablestheOCM®
tocalculateanexactvalueforspKt/Vwithanerror<0.1
foreachroutinedialysissessionconductedoveralon-
InordertodetermineanexactvalueforspKt/Vusing
theOCM®,itisalwayspreferabletoenterthemeasured
ureadistributionvolumeratherthanananthropomet-
rically-derivedV.
With the software “Dose Calculation Tool” (DCTool),
Fresenius Medical Care provides the OCM® user
with an uncomplicated and precise instrument for
the accurate calculation of patient-specific urea dis-
tribution volume V using treatment and laboratory
data. The DCTool is a urea kinetic modelling (UKM)
software programme based on 2 point urea kinetic
modelling. The programme determines the weekly
413:OCM®offersmanyadvantagesincomparisonwiththeconventionalmethodfortheKt/Vmeasurement
16
The conventional procedure today Aspects 5008 with OCM® option
Kt/V ≥ 1.2... 1.8
Frequency
Control
Effort
Accuracy of K
Handling
Quality assurance
Blood samples (expensive)
Once a month/quarterly
Retrospective
Staff, syringes, lab time, cost and energy
6 – 8 %
Inconvenient
Unpractical and uncommon
Dialysate; K and t (no additional costs)
In every session
Continuous, online
None
6 %
Automatic
Standard!
412: DCTool is a urea kinetic modelling software to precisely calculatetheureadistributionvolumeofthepatient
gertimeperiod(upto6monthsforstablepatients)(66).
Itispossibletoconvertthissingle-poolKt/Vdetermined
bytheOCM®toaneKt/V(≈doublepoolKt/V)using
the formulae provided in section 2.1. Alternatively,
eKt/Vvaluesarealsoshowninthetable1onpage10.
In order to ensure that the V value is appropriately
updated,itisadvisabletorepeatureakineticmodel-
lingcalculationswithDCToolwithintheframeworkof
routinelaboratorytestsevery6–12weeks.
Inclinicalstudiesinwhich40treatmentsand151treat-
mentswereanalysed,itwasshownthattheDCTool-
methodprovidedtheoptimumvaluefortheureadistri-
butionvolumeofapatient(51,52,66)andthat theresults
agree well with the total body water measured with
bioimpedancespectroscopy.
The use of the OCM® functionality and DCTool in
combinationmakesitpossibletodeterminetheexact
valueforspKt/Vduringeachroutinedialysis.
Theparametersrelevanttotherapymonitoring,includ-
ing the OCM® measurement results, are displayed
inaclearoverviewwithamulti-coloureddiagramon
the touch screen of the 5008 Therapy System. The
clearance K is displayed during the course of the
treatmentasabluelineinthe“OCM®diagram”graphic,
soarethecurrentandabsolutevaluesforK,Kt/Vor
KtandGoal-Kt/V.Inaddition,thecumulativeKt/V(red
line)orKt(blackline)arevisualisedinrelationtotheset
Kt/Vtarget(greenline).
3.2 OCM® in ONLINEplus Haemodiafiltra- tion treatments
ONLINE Haemodiafiltration (HDF) with its numerous
positiveeffectsonuraemia-relatedcardiovascularrisk
factors,isacknowledgedasthemosteffectivedialysis
treatment modality that comes closest to the elimi-
nationprofileofthenaturalkidney. Inparticular,HDF
showssuperiorclearancesincomparisontostandard
haemodialysisintheremovalofmiddlemolecules.In
selectedEuropeancountries,upto30–40%ofthe
dialysis population is treated with ONLINE HDF, the
averagehoweverbeingaround10%(61).
Several studies have demonstrated clinical advan-
tagesoftheONLINEHDFtreatmentmodalitysuchas
an increased dose of dialysis, a reduced risk of ß2-
microglobulin-associated amyloidosis(62,63), a positive
influence on the treatment of anaemia(64) and im-
proved intradialytic cardiovascular stability(65). The
results of two recent studies even demonstrate that
patients treated with ONLINE HDF had a significant
35%reductionofmortalityriskcomparedtopatients
treatedwithstandardhaemodialysis(68,69).
The precision of the OCM® clearance measurement
under ONLINE HDF conditions has been validated
in-vitro and in clinical experiments by Canaud et al.
(HôpitalLapeyronie,Montpellier,France;publicationin
KIpending(71)). Intheseinvestigations,anerrorrange
of6% for theclearancemeasurementof theOCM®
systemduringONLINEplusHDFwasreported.
WiththeintroductionofOnlineClearanceMonitoringin
ONLINEplusHaemodiafiltrationit ispossibletoeasily
monitor the dose of dialysis during highly efficient,
convectivetreatments.
17
414: The OCM® display: The blue line indicates the urea clearance, the cumulatingKt/Visdisplayedbytheredline,theGoal-Kt/Visvisualisedasa greenline.
4.1 Blood flow
Theextentofureaclearanceachievedduringdialysis
isdeterminedlargelybytheeffectivebloodflowrate.
Increasingthebloodflowrateisthusaveryeffective
method of increasing dialysis efficiency(53). Figure 15
illustrates that in the blood flow rate range 0 – 600
mL/min, the increase inureaclearance isproportional
tothe increaseinbloodflowrate.Underthespecified
conditions,e.g.withtheFreseniusMedicalCareFX60
dialyser,anincreaseinbloodflowratefrom200mL/min
to400mL/minresultsinanincreaseinureaclearance
by 65%, while an increase in blood flow rate from
150mL/minto300mL/minresultsinanimprovement
of82%.Butevenmoderate increases inbloodflow
rateby50mL/min–from250mL/minto300mL/min
Theefficiencyofeachhaemodialysistreatmentisdeterminedbythefollowingfourparameters:
• theeffectiveextracorporealbloodflow
• theactualdialysistime
• theeffectiveureaclearanceachievedbythedialyser
• thedialysateflowrate
Itisessentialtoachievetheappropriatebalanceoftheseparametersinordertoensurethatadequatedialysistreat-
mentisprovidedandalsotoensurethattheresourcesavailableareeconomicallyutilised.
415:UreaclearanceofFreseniusMedicalCareFX60dialyseratdifferentbloodflowrates
4. Optimisation of dialysis efficiency
– under standard conditions of dialysis lead to an
increaseintheureaclearanceofapproximately15%.
When prescribing the required blood flow, and par-
ticularly when increasing blood flow, the individual
vascularaccessofeachpatientmustbeconsidered.
Thefistulabloodflowrate(averageflowrateinradio-
cephalic fistula is approximately 500 – 900 mL/min)
shouldalwaysmarkedlyexceedthatoftheextracor-
porealbloodflowrateinthedialyser,otherwisethere
is the risk that recirculation, which compromises
efficiency,maybeinduced.Recirculationoccurswhen
blood that has already been cleared of toxins flows,
counter to thenaturaldirectionof flow in thefistula,
out of the venous cannula directly into the arterial
cannula and mixes with non-purified blood. Recirc-
ulating blood has only a low toxin load prior to its
repeated passage through the dialyser. As a result,
thetoxinremovalpertimeunitisreducedanddialysis
efficiencydecreases.
Thereare,however,variousdevicesavailable,suchas
theBlood Temperature Monitor (BTM),obtainable
as an optional extra to the 5008 Therapy System,
which can automatically measure recirculation at a
pre-selected blood flow rate. In case of a signifi-
cantdropinthemeasuredclearancevalues(≥20%)
during the same treatment or compared to the
previous one stored on the 5008 PatientCard, a
BTM recirculation measurement can be automati-
callyinitiatedbytheOCM®.
Urea
cle
aran
ce [m
L/m
in]
Blood flow [mL/min]
Dialysate flow 500 mL/min
18
Since the OCM® conducts the clearance measure-
mentonthedialysatesideonly,theaccuracyofKis
notinfluencedbydeviationsinthebloodflowrate.In
thecaseofbloodordialysateflowchangesinbetween
twosubsequentmeasurements,theOCM®isableto
account for thesechangesbyusingamathematical
estimateforKunderthenewconditions.
4.2 Dialysis time
As thedialysis time t is adirect variable in theKt/V
formula,itisimportanttoensurethatasufficientlylong
effectivedialysistimeisdeliveredtoensureprovisionof
anadequatedialysisdose.Itisessentialtoprescribe
a sufficiently long treatment timewhich, by avoiding
all interruptions caused by alarms or other delays,
should,asfaraspossible,alsorepresenttheeffective
dialysistime.
TheDOQIguidelines,forexample,take intoaccount
the discrepancy between prescribed and effective
treatmenttime,whichfrequentlyoccursasaresultof
undesirable interruptions inthedialysistreatment,by
recommendingprescriptionofaslightlyincreasedKt/V
valueof1.3ifaKt/Vof1.2istobeachieved(33). Ina
studypublishedin2001,J.Leonetal.investigatedthe
dialysisprescriptionsfor721dialysispatients(54).They
foundthatin15%ofpatientsthedialysisprescription
wastoolowtoachievetheminimumdoseofKt/V1.2
as recommended in theDOQI guidelines. They also
concluded that in this patient group, a prolongation
of treatment by up to 30 minutes (depending on
patient) would have resulted in the delivery of an
adequatedialysisdosein75%ofthecases.
Thisresultshowsthatevenamoderateprolongation
oftreatmenttimecanbeanessentialfactorinensur-
ing delivery of an adequate dialysis dose. Individual
prolongationof treatment time should thus seriously
betakenintoconsideration,despitethedifficultiesof
incorporatingthisinshiftschedulesandtheunwilling-
nessofpatientstoanincreaseinthedialysistime.
19
4.3 Dialyser (Membrane surface area, low flux/high flux)
Thedialyser itselfplaysan important role in theeffi-
ciencyofdialysistreatment.Theuseofadialyserwith
thehighestpossibleureaclearanceisrequiredforop-
timisationofthedialysistreatment.Asarepresentative
ofotherlowmolecularweightsubstances,ureaisone
ofthemostimportantmarkersubstancesindialysis.
Theureaclearanceisnotonlydeterminedbytheactive
surfaceareaofthemembrane,butalsobythemembrane
permeability(highorlowflux)andtheconstructionof
thedialyseritself.Figure16showstheincreaseinurea
clearanceunderthegivenconditions(bloodflowrate:
300mL/min,dialysateflowrate:500mL/min)inrelation
toanincreaseintheeffectivemembranesurfacearea.
Underidenticalconditions,ureaclearanceisimproved
bychangingfromlowfluxtohighfluxdialysers.Touti-
liseagivendialyserto itsfullcapacity, it is important
toconsidertherelationshipbetweeneffectivesurface
areaand theachievablebloodflow (see instructions
foruse:re-commendedbloodflowrange).
416:UreaclearanceofFreseniusMedicalCare lowandhighfluxdialysers asfunctionoftheeffectivesurface
4.4 Dialysate flow rate
Itistheconcentrationgradientbetweenthebloodand
dialysate which mainly determines the diffusion rate
duringhaemodialysis.Thelargertheflowoffresh,toxin-
free dialysis fluid through the dialyser, the larger the
removaloftoxinsbydiffusionis(55).
In many dialysis centres, a dialysis fluid flow rate of
500mL/minisusedbydefaultinhaemodialysistreat-
ments,althoughoptimalsoluteclearanceswouldal-
readybeachievedatlowerdialysisfluidflowrates.As
showninfigure17,atadialysateflowrateequaltothe
bloodflowrate,approximately90%ofthemaximum
solute clearance is already achieved. Thus, much
higherdialysis fluid flow rateswouldnotsignificantly
contributetoincreasesinclearancesandaretherefore
notnecessaryfromaneconomicalpointofview.
The5008TherapySystemofferstheAutoFlowfunc-
tion,whichautomaticallyadjuststhedialysisfluidflow
rateinordertoderiveanoptimalratiobetweenblood
flowrateanddialysisfluidflowrate.Anypotentialeffects
onlowmolecular(e.g.urea)clearancecanhoweverbe
monitored with OCM®, thus providing another very
meaningfulfunctionforfurthertreatmentoptimisation,
e.g.bymeansofincreasedbloodflowand/orperform-
ingONLINEHDF.
Urea
cle
aran
ce [m
L/m
in]
Effective dialyser surface [m2]
QB: 300 mL/min
1700,6 0,8 1,0 1,2 1,4 1,6 1,8 2,0 2,2 2,4
180
190
200
210
220
230
240
QD: 500 mL/min
FX-class high flux range FX-class low flux range
417:UreaclearanceofaFX60dialyserasfunctionofbloodanddialysate flow.Thehigherthebloodflowthehigheristhepositiveeffectofincreasingthe dialysisfluidflowonureaclearance
In v
itro
urea
cle
aran
ce F
X 60
[mL/
min
]
Dialysate flow [mL/min]
Blood flow
20
QB: 300 mL/minQD: 500 mL/min
Theprerequisite for aneffectivedialysis treatment is
anappropriatebalancebetweenbloodflowrate,dia-
lysateflowrate,treatmenttimeandthedialyserused.
Forthispurpose,FreseniusMedicalCarehasdeveloped
appropriatesoftwareintheformoftheClearanceCal-
culationTool(CCT),whichmakesitpossibletocompare
theefficacyofFreseniusMedicalCaredialysersunder
diversepredefinedtreatmentconditions.
418:TheCCTClearanceCalculationToolisasoftwarefortheplanningof haemodialysistreatments.CCTallowsaclearanceestimationofanyFresenius MedicalCaredialyseroverthefullrangeoftreatmentconditions
21
Inadditiontothedelivereddialysisdose,theOCM®also
measures the levelofplasmasodiumof thepatient.
This information is a useful aid in the appropriate
therapeutic adjustment of the sodium concentration
of the dialysis fluid to the individual plasma sodium
levelofthepatient.
In a clinically stable haemodialysis patient, there is
physiologicalequilibriumbetweenthequantityofsalt
ingested with the nutrition and the quantity of NaCl
removedduringhaemodialysis.
Therearetwomechanismsofintradialyticremovalof
NaCl:ultrafiltration(UF)anddiffusion.
Theelectrolytecompositionoftheultrafiltrateremoved
forthepurposesofdehydrationofthepatientisvery
similartothatofbloodplasma,theNaClconcentration
intheultrafiltratevariesfrompatienttopatient inthe
approximaterange135–140mmol/L.Thequantityof
NaClremovedduringultrafiltrationdependsdirectlyon
thevolumeofUFremoved;thehigherthevolumeofUF,
thegreaterthequantityofNaClremovedfromthepatient.
Diffusion is even more effective than ultrafiltration in
theremovalofNaCl,thoughinthiscasemaintenance
ofaneffectiveconcentrationgradientbetweenblood
anddialysatesidesisrequired.
IftheconcentrationofNaClavailableforfreediffusion
inthebloodplasmaofthepatientishigherthaninthe
dialysisfluid, there isaflowofNaCl fromblood into
dialysis fluid;NaCl is removed from thepatient. The
oppositecan,ofcourse,alsooccur:iftheconcentra-
tionofNaClinthedialysisfluidishigherthaninblood
plasma,itdiffusesfromthedialysisfluidintotheblood
andthereisdialysis-inducedsalt-loading.
Theconcentrationofsodiuminthedialysisfluidthus
determineswhetherthesodiumbalanceispositiveor
negativeduringhaemodialysistreatment.
5. Measurement of plasma sodium with the OCM®
Variousstudieshavedemonstratedthatrestrictionof
sodium intake has a favourable effect on long term
bloodpressurecontrolindialysispatients;lowsodium
content inthedialysate,providedthis istoleratedby
thepatient,canassistindietarycontrolofsodiumin-
take(57–59).Krautzigetal.havedemonstratedthatwith
lowsodiumintake,itispossibletoreduce,orevendis-
continue,theuseofanti-hypertensivemedication(60).
However, if adialysatewithanon-physiological, low
sodium concentration is used, this can frequently
trigger intolerance reactions, such as hypotension,
tirednessandcramps.
Conversely,theuseofadialysatewithanon-physio-
logical, high concentrationof sodiummay improve
blood pressure stability during dialysis, but also in-
creasesthepatient’sthirstandresultsinasubsequent
increase in fluid intake by the patient. This, in its
turn, leads to increased overhydration and long-term
hypertension in the patient. Figure 20 summarises
theadvantagesanddisadvantagesofhighand low
sodiumconcentrationsinthedialysate.
Individualprescriptionofthesodiumconcentrationofthe
dialysisfluidalwaysrequiresachievingacompromise
between long-termbeneficial bloodpressure control
(withlowdialysisfluidsodium)andtheshort-termad-
vantageofbettercardiovascularstabilityduringtreat-
ment (with high dialysis fluid sodium). Hence, when
419:Acuteandchronic effectsofdifferent sodiumconcentrations in the dialysisfluid
Acute effects
cardiovascular stabilitythirst
extracellular dehydrationplasma osmolality
disequilibriumhypotension
Chronic effects
Chronic volume overloadhypertension
blood pressure control
Na145
Na138
Na130
+ ––
–––– +
–
22
decidingontheoptimumconcentrationofsodiumin
thedialysisfluid,thenephrologistshouldconsiderthe
predialyticplasmasodium,thegeneralstatusandthe
cardiovascularstabilityofthepatient.
Figure20showstheeffectsofahaemodialysistreat-
mentontheplasmasodiumconcentrationinindividual
patients(meanvaluesfor10treatmentsarequoted).
Whiletherewaslittlefluctuationinplasmasodiumlevels
inpatients1and2(thedialysisfluidusedforboththese
patientshadasodiumconcentrationof140mmol/L),
plasma sodium levels in patients 3 and 4 (here the
specifieddialysisfluidsodiumconcentrationwas142
mmol/L)increasedby2.5–3mmol/Lduringdialysis.
How can the OCM® be used for measurement of
plasma sodium?
TheOCM®determinessodiumconcentrationsatthe
dialysateinflowandoutflowbymonitoringconductivity.
Thedifferencebetweenthetwovalues,inrelationtothe
dialysateflowrate,showstherateofflowofsodium
intooroutofthepatient.Usingthevalueforclearance
asmeasuredbytheOCM®,theplasmaconcentration
ofsodiumcanbecalculated. In thisway, theOCM®
derivesavalueforplasmasodiumfromeachclearance
measurement.Theresultisadjustedusingcorrection
420: Example of a plasma sodium profile of 4 HD patients measured byOCM®
factorssothatthedisplayedvalueforplasmasodium
isequivalenttothatobtainedbymeansofflamepho-
tometryassay.
Smallabsolutedifferencesbetweentheplasmasodium
concentration determined by the dialysis machine
using this non-invasive method and the results
of laboratory analysis canbeexplainedbydeviations
of the measurement and calibration instruments
used. Of more clinical relevance, however, is the
change inplasmasodiumconcentrationduringdialy-
sis–withitsassociatedeffects–relativetopredialytic
baselineplasmasodium.Usingthedataforchanges
inplasmasodiumovertimeregisteredbythedialysis
machine, it can be determined whether the patient
is being loaded with sodium from the dialysis fluid
(plasma sodium concentration increase) or whether
too much sodium is being removed from the blood
(plasma sodium concentration falls), or whether the
sodiumconcentrationofthehaemodialysissolutionis
appropriateforthepatient(e.g.atendencytocons-
tantplasmasodiumconcentrations).OnlineClearance
Monitoring(OCM®)canassisttoappropriatelyadjust
thesodiumconcentrationof thehaemodialysissolu-
tionduringtheongoingtreatment.
Theplasmasodiumconcentrationovertimeisdisplayed
asaredlineintheUF-Na-Diagramofthe5008.Atlow
flowconditions(QB≤80mL/min;QD<300mL/min),
theplasmasodiumconcentrationisnotdisplayed.The
accuracyofthevaluedisplayedisfurthermorestrongly
affectedbytheexistenceoffistularecirculation.
421:ThekineticsoftheplasmasodiumconcentrationinthecourseofaHDtreatmentcanbedisplayedbytheOCM®(redline)
Plas
ma-
Na [m
mol
/l]
Pat 1 (Dial.-Na: 140 mmol/l)Pat 2 (Dial.-Na: 140 mmol/l)Pat 3 (Dial.-Na: 142 mmol/l)Pat 4 (Dial.-Na: 142 mmol/l)
23
1.IsKt/VasmeasuredbytheOCM®asingle-poolor
double-poolKt/V?
TheKt/VvaluedeterminedbytheOCM®isasingle-pool
Kt/V (spKt/V). It ispossible toconvert thesingle-pool
Kt/VdeterminedbytheOCM®toaneKt/V(≈equilibrated
or double pool Kt/V) using the formulae provided in
Section 2.1.2.1 Alternatively, eKt/V values are also
showninthetableonpage10.
2. In a comparison conducted at a dialysis centre, it
wasobservedthattheOCM®Kt/Vwas10%lowerthan
theKt/Vmeasuredonthebloodside:whatisthecause
ofthisdiscrepancy?
Suchadifferenceisfrequentlyseenifthevalueforurea
distributionvolumeVusedforKt/Vcalculationbythe
OCM®hasbeenestimatedusingtheWatsonformula.
TheanthropometricWatsonformulatendstooveresti-
matetheureadistributionvolumeindialysispatients.As
thepreciselymeasuredresult forKxtdeterminedby
theOCM®isthendividedbyanincorrectlyhighvalue
forV,thisresultsinanunderestimationofthedelivered
dialysisdose(Kt/V).Thisproblemcanbeavoidedifthe
valueforVisdeterminedmorepreciselyusingabioim-
pedancespectroscopy(BIS)techniqueorureakinetic
modelling (e.g.with theaidofDCTool)–seeSection
3–and this valuebeingentered into theOCM®and
storedonthePatientCardforfutureuse.
3.Whatisthesignificanceofhaematocritforthecalcu-
lationofclearanceandplasmasodiumconcentrations?
Whatalternativeisthereifnorecentvalueforhaema-
tocritisavailableforapatient?Couldanestimatedvalue
forhaematocritresultinaninaccurateresultforKt/V?
The haematocrit is required to calculate the effective
flowofwatercontainedwithintheblood(=plasmawa-
ter+intracellularwater)andsubsequentlytheeffective
Kt/V(Comment:HctdoesnothaveanyinfluenceonK,
sinceonlydialysate-sidequantitiesareused todeter-
mineit.Rather,itenterstheestimationformulaincase
6. Frequently Asked Questions about the OCM®
ofabloodflowchangebetween2measurements).Ifan
estimatedHctwhichdeviatesfromtheactualHctofthe
patientisenteredintheOCM®template,thiswillhave
littleeffecton theaccuracyof thevalue forclearance
calculatedby theOCM®. If, for example, theentered
value forHct is 10%higher or lower than the actual
valueforHct,thereisonlya2–3%changeinthere-
sultforclearance.IfnovalueforHctisenteredpriorto
commencementoftreatment,theOCM®automatically
assumesahypotheticalvalueof35%.Incasea5008
Blood Volume Monitor (BVM) isusedsimultaneously
withtheOCM®,thecontinuouslymeasuredHctvaluecan
beautomaticallytransferredfromtheBVMtotheOCM®.
4.WhyistheclearancemeasuredbytheOCM®frequen-
tlylowerthanclearancemeasuredonthebloodside?
This occurs if a simplified method of whole blood
clearance measurement is performed on the blood
side, where the full blood flow value QB is used
for blood side calculation of clearance, while the
OCM® correctly uses the effective flow of water
contained within the blood (= plasma water + intra-
cellularwater). Theeffective flowofwatercontained
within the blood is equivalent to the effective blood
flow x 0.87. The factor 0.87 takes into account the
characteristic composition of human blood. If the
effectiveflowofwatercontainedwithinthebloodisnot
usedforbloodsidecalculationofclearance,therecan
beaninappropriateoverestimationofureaclearance
byupto25%comparedwiththeeffectiveclearance
whichisaccuratelymeasuredbytheOCM®.
5.CouldtheaccuracyofOCM®measurementsbeim-
pairedifequipmentisnotdisinfectedbetweeneachdi-
alysistreatment?
In theory, throughout dialysis deposits may occur on
theconductivitysensorsoftheOCM®justasinanyof
the other components of the hydraulic system of a
dialysismachine.Toavoidthisproblem,itismandatory
todisinfect theequipmentaftereachdialysissession.
24
The risk of an inaccurate clearance measurement is
thusexcluded.
6. Does the accuracy of the OCM® depend on the
membranematerialusedinthedialyser?
Noevidencehasbeenfoundtodatethattheaccuracy
of the clearance measurement by the OCM® is influ-
encedbythetypeofdialysermembraneused.
7.CanOCM®measurementbestartedduringon-going
dialysistreatment?
Inprinciple,theOCM®canbestartedatanytimeduring
on-goingtreatment.If,however,morethan80minutes
havepassedsincetreatmentwasstartedwithoutasuc-
cessfulOCM®measurementperformed,thesystemwill
notcalculateKtandKt/V.Itwouldnotbeworthwhileex-
trapolatingbackwards,astheremayhavebeenmarked
changesinclearancesincetheinitiationoftreatment.
8.Isitpossibletocalculatetheureadistributionvolume
ofchildrenusingtheWatsonformula?
TheMellits-CheekformularatherthantheWatsonfor-
mulashouldbeusedforanthropometricestimationof
Vinchildren:
Forboyswithaheight<132.7cm:
V=-1.927+0.465xbodyweight(kg)
+0.045xheight(cm)
Forboyswithaheight>132.7cm:
V=-21.993+0.465xbodyweight(kg)
+0.209xheight(cm)
Forgirlswithaheight<110.8cm:
V=0.076+0.507xbodyweight(kg)
+0.013xheight(cm)
Forgirlswithaheight>110.8cm:
V=-10.313+0.252xbodyweight(kg)
+0.154xheight(cm)
Inorder toobtainaprecise value forV in children, a
bioimpedancemeasurementorureakineticmodelling
(e.g.withDCTool)ispreferable.
9. How does one take into account an amputation
whencalculatingureadistributionvolume?
Urea distribution volume V in amputated dialysis pa-
tients should not be estimated using anthropometric
formulae(suchastheWatsonformula),butshouldbe
determinedbyabioimpedancemeasurementorurea
kineticmodelling(DCTool).
10.CanOCM®alsobeusedduringONLINEHDF?
OCM®canbeusedduringregularONLINEplusHaemo-
dialfitrationtreatmentsineitherpost-orpredilutionmode
butduetothemissingdialysateflowinthedialysernot
inONLINE-HFnorinSingle-Needlemodes.
25
1. OwenW,LewN,LiuY,LowrieE,LazarusJ:Theureareductionratioandserumalbuminconcentrationaspredictorsofmortalityinpatientsundergoing hemodialysis.NEnglJMed1993;329:1001-1006
2. ShermanR,CodyR,RogersM,SolanchickJ:Accuracyoftheureareductionratioinpredictingdialysisdelivery.KidneyInt1995;47:319-321
3. DepnerT:EstimationofKt/VfromtheURRforvaryinglevelsofdialyticweightloss:Abedsidegraphicaid.SeminDial1993;6:242
4. SeguraL,DolesonM,PastreF,GaffetJ,GranollerasC,OulesR,FourcadeJ,ShaldonS:Dialysistime:whatyouprescribeiswhatyouget?(WYPIWYG). EDTNAERCAJ.1998;24(1):43-4
5. GotchF,SargentJ:AmechanisticanalysisoftheNationalCooperativeDialysisStudy.KidneyInt1985;28:524-534
6. SargentJ,GotchF:Principlesandbiophisicsofdialysis,inMaherJ(ed):Replacementofrenalfunctionbydialysis(ed3).Boston,MA,KluwerAcademic, 1989:87-143
7. DepnerT,CheerA:ModelingureakineticswithtwoversusthreeBUNmeasurements.Acriticalcomparison.ASAIOTrans1989;35:499-502
8. DaugirdasJ:Secondgenerationlogarithmicestimatesofsingle-poolvariablevolumeKt/V:Ananalysisoferror.JAmSocNephrol1993;4:1205-1213
9. FlaninganM,FangmanJ,LimV:Quantitatinghemodialysis:Acomparisonofthreekineticmodels.AmJKidneyDis1991;17:295-302
10. BankheadM,TotoR,StarR:Accuracyofurearemovalestimatedbykineticmodels.KidneyInt1995;48:785-793
11. DaugirdasJ,SmyeS:Effectofatwocompartmentdistributiononapparentureadistributionvolume.KidneyInt1997;51:1270-1273.
12.StarR,HootkinsJ,ThompsonJetal.:Variabilityandstabilityoftwopoolureamasstransfercoefficient.JAmSocNephrol1992;3:395(abstr)
13. CappelloA,GrandiF,LambertiC,SantoroA:Comparativeevaluationofdifferentmethodstoestimateureadistributionvolumeandgenerationrate.IntJ ArtifOrgans1994;17:322-330
14. PedriniL,ZereikS,RasmyS:Causes,kineticsandclinicalimplicationsofpost-hemodialysisurearebound.KidneyInt1988;34:817-824
15. LeblancM,CharbonneauR,LalumiereG,CartierP,DezielC:Postdialysisurearebound:determinantsandinfluenceondialysisdeliveryinchronichemo- dialysispatients.AmJKidneyDis1996;27:253-261
16. SmyeS,DunderdaleS,BrownridgeG,WillE:Estimationoftreatmentdoseinhigh-efficiencyhemodialysis.RevEnferm1994;67:24-29
17. SpiegelD,BakerP,BabcockS,ContigugliaR,KleinM:Hemodialysisurearebound:Theeffectofincreasingdialysisefficiency. AmJKidneyDis1995;25:26-29
18. HEMOStudyGroup,DaugirdasJ,DepnerT,GotchF,GreeneT,KeshaviahP,LevinN,SchulmanG:ComparisonofmethodstopredictequilibratedKt/Vin theHemopilotstudy.KidneyInt1997;52:1395-1405
19. SmyeS,EvansJ,BrocklebankJ:Paediatrichemodialysis:Estimationoftreatmentefficiencyinthepresenceofurearebound.ClinPhysPhysiolMeas 1992;13:51-62
20. DaugirdasJ:EstimationofequilibratedKt/VusingtheunequilibratedpostdialysisBUN.SeminDial1995;8:283-284
21. CollinsA,MaJ,UmenA,KeshaviahP:Ureaindexandotherpredictorsofhemodialysispatientsurvival.AmJKidneyDis1994;23:272-282
22. LowrieE,LairdN,ParkerT,SargentJ:Effectofhemodialysisprescriptiononpatientmorbidity.NEnglJMed1981;305:1176-1180
23. HakimR,BreyerJ,IsmailN,SchulmannG:Effectsofdoseofdialysisonmorbidityandmortality.AmJKidneyDis1994;23:661-669
24. LowrieE:Chronicdialysistreatment:Clinicaloutcomeandrelatedprocessofcare.AmJKidneyDis1994;24:255-266
25. FernandezJ;CarbonellM,MazzuchiN,PetrucelliD:Simultaneousanalysisofmorbidityandmortalityfactorsinchronichemodialysispatients.KidneyInt 1992;41:1029-1034
26. HeldP,PortF,WolfeR,StannardD,CarrollC,DaugirdasJ,GreerJ,HakimR:Thedoseofhemodialysisandpatientmortality.KidneyInt1996;50:550-556
27. ParkerT,HusniL,HuangW,LewN,LowrieE,DallasNephrologyAssociates:SurvivalofhemodialysispatientsintheUSisimprovedwithagreaterquantity ofdialysis.AmJKidneyDis1994;23:670-680
28. GotchF,LevinN,PortF,WolfeR,UehlingerE:Clinicaloutcomerelativetothedoseofdialysisisnotwhatyouthink:thefallacyofthemean.AmJKidney Dis1997;30:1-15
29. OwenW,ChertowG,LazarusJ,LowrieE:Thedoseofhemodialysis:Mortalityresponsesbyraceandgender.JAMA1998;280:1-6
30. GotchF,SargentJ:AmechanisticanalysisoftheNationalCooperativeDialysisStudy(NCDS).KidneyInt1985;28:526-534
31. PortF,AshbyV,DhingraR,RoysE,WolfeR:Dialysisdoseandbodymassindexarestronglyassociatedwithsurvivalinhemodialysispatients.JAmSoc Nephrol2002;13:1061-1066.
32. MovilliE,CancariniG,ZaniR,CameriniC,SandriniM,MaiorcaR:Adequacyofdialysisreducesthedosesofrecombinanterythropoietinindependentlyfrom theuseofbiocompatiblemembranesinhaemodialysispatients.NephrolDialTransplant2001;16:111-114.
33. NKF-K/DOQIClinicalPracticeGuidelinesforHemodialysisAdequacy:UpDate2000.AmJKidneyDis2001;37Suppl.1:7-64.
34.BabbA,MaurerC,PopovichR,McKee:Thedeterminationofmembranepermeabilitiesandsolutediffusivitieswithapplicationstohemodialysis.ChemEng ProgrSympSer1968;84,64:59-68
35. SteilH,KaufmanAM,MorrisAT,LevinNW,PolascheggHD:.InvivoverificationofanautomaticnoninvasivesystemforrealtimeKtevaluation.ASAIOJ 1993;39(3):M348-52
36. PolascheggHD:Automatic,noninvasiveintradialyticclearancemeasurement.IntJArtifOrgans1993;16(4):185-191.
37. GoldauR,KuhlmannU,SamadiN,GrafT,GrossM,OrlandiniG,LangeH:Accuracyofionicclearancemeasurementisdependentonureadistributio volume:Anewapproachtobetterresults.NephrolDialTransplant2000;15(9):A133
38. KuhlmannU,GoldauR,SamadiN,GrafT,GrossM,OrlandiniG,LangeH:Accuracyandsafetyofonlinemonitoringbasedonconductivityvariation.JASN 1998;9:299A
7. References
26
39. KuhlmannU,GoldauR,SamadiN,GrafT,GrossM,OrlandiniG,LangeH:AccuracyandsafetyofOnlineClearanceMonitoringbasedonconductivity variation.NephrolDialTransplant2001;16:1053-1058
40. KuhlmannU,GoldauR,SamadiN,GrafT,GrossM,OrlandiniG,LangeH:OnlineClearanceMonitoringofhighfluxhemodialysisbasedonconductivity variation.JASN1998;9:298A
41. DiFilippoS,GoldauR,ManzoniC,SpickermannR,MarcelliD,PontorieroG,OrlandiniG,LocatelliF:Effectiveureaclearance(Uke)andionicdialysanceby conductivity(ID):Whatistherelationship?JASN1998;9:299A
42. DiFilippoS,ManzoniC,AndrulliS,PontorieroDell’OroC,LaMiliaV,BacchiniG,CrepaldiM,BigiM,LocatelliF:Howtodetermineionicdialysanceforthe onlineassessmentofdelivereddialysisdose.KidneyInt2001;59:774-782.
43. CybullaM,SchwertfegerE,WeinerS,BreitenfeldtK,BöhlerJ:AccuracyofKt/VcalculationbasedonOnline-Clearance-Measurement(OCM)incomparison todialysatesideureakinetics(urea-monitor).NephrolDialTransplant2000;15(9):A134
44. DiFilippoS,GoldauR,ManzoniC,SpickermannR,MarcelliD,PontorieroG,OrlandiniG,LocatelliF:Accurateandinexpensiveon-linemonitoringofhemo dialysisperformance.JASN1998;9:299A
45. GoldauR,KuhlmannU,NaderS,GrossM,GrafT,OrlandiniG,MarcelliD,LangeH:Ionicdialysancemeasurementisureadistributionvolumedependent:A newapproachtobetterresults.ArtifOrgans200226(4):321-32.
46. ElangovanL,ShinabergerC,KrautJ,ShinabergerJ:HEMOequilibratedKt/Vgoalsaredifficulttoachieveinlargemalepatients.ASAIOJ2001;47(3):235-9.
47. HaukM,KuhlmannM,RiegelW,KöhlerH:InvivoeffectsofdialysateflowrateonKt/Vinmaintenancehemodialysispatients.AmJKidneyDis2000;35, No1:105-111.
48. WatsonP,WatsonI,BattR:Totalbodywatervolumesforadultmalesandfemalesestimatedfromsimpleanthropometricmeasurements. AmJClinNutrition1981;33:27-39.
49. HumeR,WeyersE:Relationshipbetweentotalbodywaterandsurfaceareainnormalandobesesubjects.JClinPathol1971,24:234-238.
50. KloppenburgW,StegemanC,DeJongP,HuismanR:Anthropometry-basedequationsoverestimatetheureadistributionvolumeinhemodialysispatients. KidneyInt2001;59:1165-1174.
51. WüpperA,TattersallJ,KrämerM:Theuseofelectrolyticclearancetodetermineureadistributionvolume.ASAIOJ2002;48(2):174
52. GotchF,ManzoniC,PanlilioF.,MorrisT,RahmatiS,RoncoC:ReconciliationofspKt/Vfromureakineticmodellingwithdirectdialysisquantification JASN2000;11:320A
53. HassellD,VanderSandeF,KoomanJ,TordoirJ,LeunissenK:Optimizingdialysisdosebyincreasingbloodflowrateinpatientswithreducedvascular- accessflowrate.AmJKidneyDis2001;38,No5:948-955.
54. LeonJ,SehgalA:Identifyingpatientsatriskforhemodialysisunderprescription.AmJNephrol2001;21:200-207
55. OusephR,WardR:Increasingdialysateflowrateincreasesdialyserureamasstransfer-areacoefficientsduringclinicaluse.AmJKidneyDis2001;37 No.2:316-320.
56. HaukM,KuhlmannM,RiegelW,KöhlerH:InvivoeffectsofdialysateflowrateonKt/Vinmaintenancehemodialysispatients.AmJKidneyDis2000; 35,1:105-111
57. CharraB,CalemardE,RuffetM,ChazotC,TerraJC,VanelT,LaurentG:Survivalasanindexforadequacyinhemodialysis.KidneyInt1992,41: 1286-1291.
58. GoldsmithDJ,CovicA,AckrillP:Interdialysisbloodpressurecontrolbylonghaemodialysis.NephrolDialTransplant1996,11:401-402.
59. OzkahhyaM,OkE,CiritM,AydinS,AkcicekF,BasciA,Dorhout-MeesEJ:Regressionofleftventricularhypertrophyinhaemodialysispatientbyultra-filtra- tionandreducedsaltintakewithoutantihypertensivedrugs.NephrolDialTransplant1998,13:1489-1493.
60. KautzigS,JanssenU,KochK,GranollerasC,ShaldonS:Dietarysaltrestrictionandreductionofdialysatesodiumtocontrolhypertensioninmaintenance haemodialysispatients.NephrolDialTransplant1998;13:552-553.
61. MoellerS:ESRDpatients2005:globaloverviewofpatients,treatmentmodalitiesanddevelopmenttrends.FMEinternalanalysis.
62. LocatelliF,MastrangeloF,RedaelliB,RoncoC,MarcelliD,LaGrecaG,OrlandiniG:Effectsofdifferentmembranesanddialysistechnologiesonpatient treatmenttoleranceandnutritionalparameters.TheItalianCooperativeDialysisStudyGroup.KidneyInt1996;50(4):1293-1302
63. LocatelliF,MarcelliD,ConteF,LimidoA,MalbertiF,SpottiD:ComparisonofmortalityinESRDpatientsonconvectiveanddiffusiveextracorporealtreat- ments.TheRegistroLombardoDialisiETrapianto.KidneyInt1999;55(1):286-293
64.MaduellF,delPozoC,GarciaHetal:Changefromconventionalhaemodiafiltrationtoon-linehaemodiafiltration.NephrolDialTransplant1999;14(5):1202-1207
65.PizzarelliF,CerraiT,DattoloP,TettaC,MaggioreQ:Convectivetreatmentswithon-lineproductionofreplacementfluid:aclinicalexperiencelasting6years. NephrolDialTransplant1998;13:363-36966.WuepperA,TattersallJ,KraemerM,WilkieM,ErwardsL:DeterminationofureadistributionvolumeforKt/V assessedbyconductivitymonitoring.KidneyInt2003;(64):2262-2271
66.WuepperA,TattersallJ,LindleyE,KraemerM,EdwardsL,InglisH:Whichistheoptimalmethodtomeasureureadistributionvolumeforconductivitybased Kt/V.NephrolDialTransplant2003;18(4):173
67. TheEBPGExpertGrouponHaemodialysis:EuropeanBestPractiseGuidelinesforHaemodialysis(Part1).NephrolDialTransplant2002;17:17-2569.
68. Canaudetal:Mortalityriskforpatientsreceivinghemodiafiltrationversushemodialysis:EuropeanresultsfromtheDOPPS.KidneyInt2006;69:2087–2093
69.Jirkaetal:Responseto“MortalityriskforpatientsreceivinghemodiafiltrationversusHemodialysis”.KidneyInt2006;70:1524–1525
70.MellitsED,CheekDB:Theassessmentofbodywaterandfatnessfrominfancytoadulthood.MonographsSocResChildDev1970:140(35):12-26,72.
71. GrossM,MaierhoferA,TettaC,SénecalL,CanaudB:OnlineClearanceMeasurementinHigh-EfficiencyHemodiafiltration,KidneyInt,publicationpending
27
73
2 20
2 1/
1 GB
(1,0
V&B
05/
07)
© C
opyr
ight
200
7 Fr
esen
ius
Med
ical
Car
e De
utsc
hlan
d Gm
bH
Fresenius Medical Care Deutschland GmbH · 61346 Bad Homburg v. d. H. · Germany · Phone: +49 (0) 6172-609-0 · Fax: +49 (0) 6172-609-2191Head office: Else-Kröner-Straße 1 · 61352 Bad Homburg v. d. H.
www.fmc-ag.com