Upload
others
View
2
Download
0
Embed Size (px)
Citation preview
ASSESSING LEAKAGE WORKLOADS OF MEDICAL LINEAR
ACCELERATORS FOR IMRT AND TBI TECHNIQUES
A Thesis
submitted to the Faculty of the
Graduate School of Arts and Sciences
of Georgetown University
in partial fulfillment of the requirements for the
degree of
Master of Science
in Health Physics
By
James R. Jordan, B.S.
Washington, DC
December 10, 2007
ii
Thanks to all people and institutions that provided data for this thesis specifically:
Brad Murray of Cross Cancer Institute
Ralph Young of Martin Memorial Cancer Center
Kathryn Wall and Cara Sullivan of Rock Hill Radiation Therapy Centre
Richard Emery and Dr. Anthony Berson of St. Vincent’s Comprehensive Cancer Center
Julius V. Turian of University of Illinois Medical Center
Allan Caggiano of Holy Name Hospital
Marc S. Miner of Hughes Cancer Center of Pocono Medical Center
St. Joseph Hospital of Orange, California
Wolfgang Tomé and Bhudatt Paliwal of University of Wisconsin Hospital
And all other people and institutions that wished to remain nameless.
Many thanks,
James R. Jordan
iii
ASSESSING LEAKAGE WORKLOADS OF MEDICAL LINEAR
ACCELERATORS FOR IMRT AND TBI TECHNIQUES
James R. Jordan, B.S.
Thesis Advisor: James E. Rodgers, Ph.D.
ABSTRACT
Current estimates for leakage workloads are not well quantified for Intensity Modulated
Radiation Therapy (IMRT) and Total Body Irradiation (TBI) treatments. When analyzed
on a large scale, using a large sample, a well defined leakage workload may be
established. A database of 17 cancer treatment centers and 73 linear accelerators were
used to make the assessment. There were 374,003 total treatments, with 213,757 low
energy (4 and 6 MV) treatments, 106,343 of which were IMRT and 149,730 high energy
(15, 16, 18, and 20 MV) treatments, 64,138 of which were IMRT. There were 184 TBI
treatments, with 98 being low energy and 86 being high energy. The numbers are too
scant to make any statements about the contribution TBI makes to workload, there is
more than enough data to estimate the IMRT contribution to leakage workload. An
IMRT factor, CI, of 5.1 may be used for low energy photon calculations, and a CI of 4.4
may be used for high energy photon calculations, where CI is:
conv
IMRT
IMU
MUC =
.
iv
TABLE OF CONTENTS
Introduction …………………………………………………………………………...... 1
Chapter I: Material and Methods ..……………………………………………………... 4
Chapter II: Results …………………………………………………………………..…. 9
Chapter III: Discussion ……………………………………………………………….. 15
Appendix ………………………………………………………………………………. 26
Bibliography ………………………………………………………………………… 122
1
INTRODUCTION
Due to its smaller field sizes and the number of segments used per field, Intensity
Modulated Radiation Therapy (IMRT) uses many more monitor units (MU) per delivered
dose (usually measured in centigray abbreviated cGy) than does conventional external
beam radiation treatment. This does not have an effect on the IMRT workload for the
primary barrier or for scatter as demonstrated by Rodgers (8), but the increased beam-on
time can greatly increase the leakage of radiation from the head of the medical linear
accelerator and secondary barrier thickness requirements. Thus, the primary barrier
IMRT Workload (WIMRT) alone, in dose per week, is not sufficient to determine the
contribution of IMRT to the total leakage-radiation workload. WIMRT must be multiplied
by some factor in order to account for its increased MU per cGy in order to calculate its
full contribution to the leakage workload (WL). To find this factor, which the National
Council on Radiation Protection and Measurements (NCRP) in NCRP Report Number
151 calls the IMRT factor or CI, a ratio is taken of the of the MU per cGy of prescribed
dose for IMRT and the MU per cGy for conventional treatments (6):
∑=i
ipre
i
IMRTD
MUMU
)(
so,
conv
IMRT
IMU
MUC =
.
2
Where MUIMRT is measured in MU/cGy, and CI is a dimensionless quantity. Since
MUconv is conservatively defined as 1 MU/cGy, in most cases CI is the same value as
MUIMRT.
Just as IMRT disproportionately affects WL, Total Body Irradiation (TBI) does as
well, due to the inverse square law. Conventional workloads (Wconv) are calculated at a
distance of one meter from x-ray target to the patient, and TBI is treated at much greater
distances, up to five meters, with the TBI workload (WTBI) normalized to one meter (6).
This will greatly increase the leakage radiation contribution of TBI, since the amount of
MU per delivered dose is increased by a power of two for any increase in distance
between the x-ray target and patient.
Followill et al. calculated an IMRT factor of 3.4 MU/cGy for Varian Multi-leaf
Collimators (MLC) at 6 Megavolts (MV) and 2.8 MU/cGy at 18 and 25 MV. For Nomos
Tomotherapy an IMRT factor of 9.7 MU/cGy at 6 MV and 8.1 MU/cGy at 18 and 25 MV
was calculated. These calculations fall in line with the NCRP value of 2 to 10 (1, 6);
however, their study was limited to four patients using conventional unwedged
treatments, who were also planned using the MLC and tomotherapy IMRT techniques.
Mechalakos et al. also examined the workload for IMRT. They found an increase
in MU per week on their Varian 2100C machine with photon energies of 6 and 18 MV
using IMRT compared to two other machines (a Siemens Mevatron KD with photon
energies of 6 and 15 MV and a Varian 600C with photon energy of 6 MV), which were
limited to conventional treatments only. Their IMRT factor was between 2.2 to 2.5 (4).
3
This study was conducted with one year of data, but was limited in scope with only a
single machine performing IMRT.
Since an increased WL may translate into an increased amount of secondary
shielding needed to protect radiation workers and the public, it is very important to
carefully ascertain the CI and find the increase in leakage due to IMRT. Other studies
have found increases in workload from IMRT and TBI, but the scope of these studies was
not very broad. In this study, with a wide range of accelerators and treatment types, I
hope to more firmly establish the contributions of IMRT and TBI to the total leakage
workload.
This paper will set values for the CI contribution to the WL for the increased
leakage radiation in IMRT and TBI. I hypothesize that current estimates for leakage
workloads are not well quantified for IMRT and TBI treatments, but when analyzed on a
large scale, using a large sample, a well defined CI may be established.
4
CHAPTER I: MATERIALS AND METHODS
In this study, data was collected from seventeen radiation treatment centers,
which included 71 Varian accelerators (which included CL 6/100, 600C, 600C/D, CL
1800, 2100C, 2100C/D, 21EX, 2300C/D, and 23EX) and almost 375,000 treatment fields
of a combination of conventional unwedged, conventional wedged, IMRT, and TBI
treatments at energies of 4, 6, 15, 16, 18, and 20MV. The data was organized by Varian
using an InfoMaker report to collect the data from the treatment planning systems and
entered into Excel spreadsheets where it could be analyzed. The data that was collected
included the field, treatment time, gantry angle, energy mode, MU, dose, and accessory
used (wedge, cone, etc.).
The collected data was separated by hospital system and by individual machine.
Individual treatments that were IMRT or TBI were identified. It was then analyzed by
energy mode for the conventional MU per cGy, IMRT MU per cGy, TBI MU per cGy,
and the fraction of treatments that were IMRT was noted.
The data was analyzed by multiple methods. First, for all categories a simple
mean was computed by:
n
n
i i∑ == 1µ
µ
with the standard deviation being:
5
( )
n
n
i i
2
1∑ =−
=µµ
σ.
For an unbiased estimator of the standard deviation for a small sample n-1 is used in the
denominator. However, with the large sample sizes involved in this study, n-1 is not
significantly different than n.
The mean MU/cGy was calculated for each gantry angle used in the treatment for
conventional treatments with and without wedges, IMRT, and TBI at each individual
energy for all 73 machines in which the treatment applied. This was done simply by
summing all MU/cGy and dividing by the total number of occurrences. The IMRT, TBI,
and conventional (which includes wedges) means are displayed in Tables 2.46 through
2.58 in the appendix for machines that conducted IMRT or TBI treatments. Machines
that were not used for IMRT or TBI had means calculated for their individual energies
treated, but they are not represented in this report.
Second, the MU/cGy data for IMRT treatments was placed into value bins in
order to be able to graph the results. To do this, bins of different MU/cGy values were
created. The total number of these occurrences were created and then normalized to 100.
The width of each bin was charted for its relative weight, and the mid-point of each bin
was noted. An example of the collected data is shown in Table 1.1.
6
Delta 0.5
Bin Unit 1 6X IMRT
Unit 1 6X IMRT
0 n Ni wi b'i Ni*wi*b'i Ni*wi Ni*wi*(b'i-mean(0-15))^2
0.5 0 0.0 1.00 0.25 0.00 0.00 0.00
1 0 0.0 1.00 0.75 0.00 0.00 0.00
1.5 0 0.0 1.00 1.25 0.00 0.00 0.00
2 20 2.4 1.00 1.75 4.16 2.38 14.62
2.5 174 20.7 1.00 2.25 46.50 20.67 81.10
3 124 14.7 1.00 2.75 40.50 14.73 32.30
3.5 74 8.8 1.00 3.25 28.56 8.79 8.46
4 116 13.8 1.00 3.75 51.66 13.78 3.19
4.5 75 8.9 1.00 4.25 37.86 8.91 0.00
5 24 2.9 1.00 4.75 13.54 2.85 0.77
Table 1.1. Chart of collected data.
The first bin is all occurrences that are greater than or equal to 0 MU/cGy through
all occurrences less than 0.5 MU/cGy. The second bin is from 0.5 to less than 1.0. Delta
is the smallest bin size. The number of occurrences is denoted by n. Ni is the number of
occurrences normalized to 100 total occurences. The bin weight calculated by
subtracting the bottom of the bin from the top of the bin and dividing by the smallest bin
width is wi. The midpoint of each bin is b’i. The unit number (Unit 1 in Table 1.1) is the
tracking number for the machine at the facility being analyzed. Each facility has a unique
hospital serial number (HSN) and each machine analyzed in a facility has a number
assigned as a resource serial number (RSN) starting at one for each facility, e.g., HSN 1,
RSN 1. A table on page 12 shows each machine that conducted IMRT treatments, the
energies at which the machine conducted IMRTs, and the page where the data is located.
These results were then graphed. The abscissa is the value of mid-point of the
bin, and the ordinate is the number of occurrences for that bin. This can be seen in Figure
7
1.1.
RSN 1 6X IMRT
0.0
5.0
10.0
15.0
20.0
25.0
0 1 2 3 4 5
bins of MU/cGy
Nu
mb
er
of
Occu
rren
ces,
N
RSN 1 6X IMRT
Figure 1.1. Frequency distribution of MU/cGy plotted versus bin mid-point value.
A table was also created of means and standard deviations with differing
maximum cutoffs. Since the means and standard deviations are greatly affected by
outliers, this will give the ability to disregard higher numbers for anomalous high values.
An example of this table is shown in Table 1.2.
The mean becomes:
∑
∑
=
=′
=n
i
n
i
wiNi
ibwiNi
1
1
*
**µ
and the standard deviation becomes:
8
( )
∑
∑
=
=−′
=n
i
n
i
wiNi
ibwiNi
1
2
1
*
** µσ
.
mean for indicated range RSN 1 6X IMRT
mean(0-15)= 4.23
mean(0-25)= 4.23
mean(0-50)= 4.94
mean(0-200)= 4.94
mean(0-1500)= 4.94
stdev(0-15)= 2.33
stdev(0-25)= 2.33
stdev(0-50)= 4.76
stdev(0-200)= 4.76
stdev(0-1500)= 4.76
Table 1.2. Table of means and standard deviations at different cutoffs of MU/cGy.
As mentioned in the introduction, I set out to analyze CI on a large sample group. I
believe I have achieved this by using a database of 17 cancer treatment centers and 73
linear accelerators. There were 374,003 total treatments, with 213,757 low energy (4 and
6 MV) treatments, 106,343 of which were IMRT and 149,730 high energy (15, 16, 18,
and 20 MV) treatments, 64,138 of which were IMRT. There were 184 TBI treatments,
with 98 being low energy and 86 being high energy. All 374,003 treatments have several
data recorded besides MU and dose, these include machine model, accessories used
(wedge, cone, et cetera), field identification and name, reference identification and name,
gantry angle, energy mode, patient and session serial number, treatment technique, and
treatment date and time.
9
CHAPTER II: RESULTS
Figures 2.1 through 2.44 in the appendix show the graphs of all machines that
performed IMRTs. Each figure has a corresponding table of means and standard
deviations at MU/cGy cutoffs of 15, 25, 50, 200, and 1500 MU/cGy. These cutoffs are
useful for determining the most useful data to use. For example, if IMRT set-up fields
were included with the data, then the extremely low dose (nominally zero assessed to
these fields) causes the MU/cGy value to be inordinately high and these numbers may be
discarded since they are fictitiously high. The data in the tables below represent the
number of MU/cGy of IMRT treatments which represents the MUIMRT. The IMRT factor
or CI may be found by dividing the MUIMRT by the MUconv. A MUconv of 1 may be used
for conservative values of CI which also makes MUIMRT equal to CI for conservative
estimates.
The overwhelming majority of data, before being analyzed whether there is a
need to restrict the ranges to lower cutoffs, show an IMRT factor within the NCRP
suggested value range of between 2 and 10. All values above the range noted in NCRP
151 will be discussed here.
The first elevated values, as seen in Tables 2.4 and 2.5, are from HSN 2, a
hospital cancer center, RSN 4 at 15 MV and 7 at 6 MV. The computed mean with a
cutoff of 1500 MU/cGy was 36.52 and 9.17 with a standard deviation 77.39 and 7.71 for
RSN 4 and 7 respectively. For RSN 4 these numbers are representative of only three
patients, so the numbers are easily skewed, and for RSN 7 the IMRT data was for a single
10
head and neck patient. Two of the three patients for RSN 4 are eight field head and neck
patients. The last is an eight field plan with an unrecorded reference point. All contain
split fields. Since the dose calculation point is taken from isocenter, if isocenter is
blocked off by MLCs then the dose becomes very small, making the MU/cGy become
extremely high and really meaningless. In this case it causes RSN 7 to have an outlier at
25.81, which has a dose which is an order of magnitude less than any other field, when all
the rest of the field doses are below 10. Using the cutoff of 25 MU/cGy the mean and
standard deviation become more acceptable values of 5.85 and 2.29. As for RSN 4, its
values range from 46.15 to a little over 323 MU/cGy. If the data is analyzed for
treatments of less than 46 MU/cGy then the mean becomes 3.14 and the standard
deviation 1.10. In both cases the numbers are now well within the NCRP recommended
value range.
HSN 3, another hospital center, RSN 5 at 6 MV (Table 2.8) has a mean and
standard deviation of 7.36 and 6.72. This number is a decent value since the IMRT
treatments performed on the machine are primarily head and necks which call for many
small fields with more MU/cGy. As mentioned in the introduction MUconv is usually
defined to be 1 MU/cGy, this creates a conservative value for CI, however, vaults are
currently shielded for a conventional MU/cGy which exceeds this number due to
differing techniques and wedge use. It is then practical, to examine higher MUIMRT in
this light. Therefore, when the value of the MUIMRT mean is divided by the mean of the
machine’s conventional treatments (Table 2.46) the CI becomes 4.74.
11
HSN 6, a stand alone cancer center, RSN 2 at 6 MV (Table 2.12) has a mean and
standard deviation of 6.29 and 7.12. Just like the previous machine if the MUIMRT mean
is divided by the conventional mean, the CI become 5.11, but also like HSN 2 it contains
a few fields which contain doses of an order of magnitude less than other fields in a
patient’s treatment. If these fields are thrown out then a cutoff of 25 MU/cGy may be
used to yield a mean and standard deviation of 5.04 and 3.25.
HSN 8, a hospital center, RSN 1 (Table 2.14) has a mean and standard deviation
of 8.33 and 17.96 and RSN 2 has 10.15 and 18.85, both at 6 MV. When fields are
removed that again are less in dose by an order of magnitude than other fields in a patient
treatment then values of MU/cGy between 55 and 440 drop out and the 25 MU/cGy
cutoff may be used, which brings the mean and standard deviation down to 7.15 and 2.80
for RSN 1 and 7.35 and 2.96 for RSN 2.
Table 2.17 for HSN 10, a hospital center, RSN 3 shows its 6 MV mode has a
mean and standard deviation of 11.52 and 7.62 and its 15 MV mode has 14.14 and 8.34.
The field doses are uniform throughout and there is no reason to use a cutoff. Head and
necks and brains are treated with 6MVs driving up the MU/cGy. When both treatment
modes are compared to the conventional means, then the CI drops to 5.35 for 6MV and
9.48 for 15MV. The 15 MV is still a little high in comparison to the NCRP
recommended value.
A lower cutoff may be used for HSN 12, a hospital center, RSN 1 (Table 2.19) at
6 MV which has a mean and standard deviation of 5.87 and 6.85. When values that are
derived from field doses that are an order of magnitude less than the other field doses in a
12
patient treatment, then the 25 MU/cGy cutoff may be used and the mean and standard
deviation becomes 4.93 and 2.88.
For HSN 13, a hospital cancer center, RSN 1 (Table 2.21) in 6MV mode some
MU/cGy are almost as high as 800 MU/cGy due to the low doses assigned to fields where
the isocenter is blocked. If these high values are disallowed then the 50 MU/cGy cutoff
may be used which brings the mean to 10.76. When divided by the conventional mean
the CI becomes 6.15. Its 18MV treatment mode has a mean and standard deviation of
7.37 and 6.91. Since 2% of the treatments are above 25 MU/cGy and these are split
fields around critical structures, i.e., with blocked isocenter, a cutoff of 25 MU/cGy may
be used for the 18 MV mode. This brings the mean and standard deviation down to 6.49
and 3.40.
HSN 15, a university hospital center, RSN 3 (Table 2.24) at 4 MV has a mean and
standard deviation of 8.01 and 2.89. Again, when divided by the conventional mean, the
CI drops to 4.67.
When the 600 MU/cGy IMRT setup fields are removed from HSN 16, a
university hospital center, RSN 3 (Table 2.28) at 6MV, the mean and standard deviation
fall from 4.25 and 13.26 to 3.96 and 1.48. This is the same for HSN 17, another
university hospital center, RSN 4 (Table 2.31) at 6 MV which sees its mean and standard
deviation drop from 6.44 and 32.48 to 6.43 and 3.05 when the setup fields that are greater
than 50 MU/cGy are removed.
Unrestricted values for IMRT MU/cGy means can be seen in tables 2.46 to 2.51
of the appendix for different photon energies. For these tables the IMRT MU/cGy is a
13
simple mean. The value, C, is the MU/cGy factor for conventional treatments, which do
not include any IMRT or TBI treatments, but does include wedges. The value C differs
from MUconv in that C is a mean of all conventional treatments, and MUconv is defined as
the MU required to deliver the same dose as MUIMRT to a phantom at a 10 cm depth at
100 source-to-axis distance with a 10cm by 10cm field (6). FI is the fraction of
treatments that are IMRT.
A conservative value of the IMRT factor can be ascertained by simply using the
IMRT MU/cGy. To get a number that more closely resembles the reality of clinical
operation, the IMRT MU/cGy can be divided by conventional MU/cGy (C). When tables
2.46 and 2.47 are examined, it can be seen that three out of 44 machines using low energy
photons have an IMRT MU/cGy value that is above the conservative CI of 2 to 10. When
the IMRT MU/cGy is divided by the conventional MU/cGy this falls to one out of 44. In
four cases the IMRT MU/cGy is lower than the conventional MU/cGy.
For machines performing IMRT at energies greater than 6 MV, only two out of 24
had a value greater than 10 MU/cGy. When IMRT MU/cGy was divided by the
conventional MU/cGy, one outlier disappeared and the other will disappear when a cutoff
is used. The tables were also recreated for only IMRT MU/cGy using cutoffs and can be
seen in the appendix in Tables 2.52 to Table 2.57.
TBI has been presented the same as IMRT. Even with the volume of fields in this
study. There were only a total of nine patients divided between four centers using four
treatment machines and three different energies. This makes it difficult to draw any
14
definitive numbers for a TBI Workload. The treatments may be seen table 2.58 in the
appendix.
15
CHAPTER III: DISCUSSION AND CONCLUSION
The NCRP estimate for the IMRT factor of 2-10 is sufficient, based on the data
from the reporting centers. These results were independent of energy. For low energy
beams, 4 or 6 MV, there were only three deviations from the range suggested in NCRP
151 for 44 different machines. For high energy beam, 10 MV or greater, there were two
values of CI above the NCRP range for 24 machines.
All of the above mentioned deviations disappear when a reasonable cutoff is
applied or the IMRT MU/cGy is divided by the conventional MU/cGy. Also, the
deviations were almost exclusively from head and neck treatments which usually contain
many small fields which cause IMRT treatments to have a larger amount of monitor units
per centigray.
If we look first strictly at only removing outliers at a reasonable cutoff, the only
outliers that remain with a combined mean plus one standard deviation over 10 are seen
in table 3.1. The cutoffs were established by either removing extraneous high numbers
caused either by the insertion of IMRT set-up fields or by removing fields with
uncharacteristically low dose. The low dose numbers are caused by isocenter being
blocked for the majority of the segments delivered in the treatment field.
HSN RSN Energy Mean Standard Deviation
3 5 6X 7.36 6.72
8 2 6X 7.35 2.96
10 3 6X 11.52 7.62
10 3 15X 14.14 8.34
13 1 6X 10.76 9.99
15 3 4X 8.01 2.89
20 2 6X 8.57 1.87
Table 3.1. Mean and standard deviations above NCRP suggested range after cutoff.
16
If we take these remaining deviations and divide their means by their respective
conventional mean, then we see that all of the outliers now disappear. Although using a
strict IMRT mean gives a more conservative answer, dividing the IMRT mean by the
conventional mean yields a number that is closer to reality, since secondary shielding in
vaults are already designed to handle this conventional workload. The results of
performing this operation can be seen in table 3.2. Only HSN 10, RSN 3 still hovers near
10. This machine represents 526 of 34,059 or 1.5% of IMRT treatments at 15 MV.
HSN RSN Energy IMRT Mean Conventional Mean IMRT/Conventional
3 5 6X 7.36 1.55 4.75
8 2 6X 7.35 1.20 6.13
10 3 6X 11.52 2.15 5.36
10 3 15X 14.14 1.49 9.49
13 1 6X 10.76 1.75 6.15
15 3 4X 8.01 1.72 4.66
20 2 6X 8.57 1.68 5.10
Table 3.2. IMRT mean divided by Conventional mean.
A chart of the mean values of CI for each MV, in the above table, gives:
CI
0
1
2
3
4
5
6
7
8
9
10
0 5 10 15 20 25
MV
CI
IMRT/C
Linear (IMRT/C)
Figure 3.1. Mean values of CI for each MV.
17
If charts are created for all IMRT machines the trend appears first without a cutoff as:
Total IMRT Treatments without Cutoff
0
1
2
3
4
5
6
7
8
9
10
0 5 10 15 20 25
MV
CI
Total IMRT Treatments without
Cutoff
Linear (Total IMRT Treatments
without Cutoff)
Figure 3.2. Mean values of CI for each MV for total treatments without cutoff.
Energy Sample Size Mean Standard Deviation
4 1 8.01 0.00
6 43 5.44 3.07
15 15 7.04 8.56
16 2 3.85 0.06
18 6 3.77 2.03
20 1 3.59 0.00
Low Energy 44 5.50 3.06
High Energy 24 5.81 6.96
Table 3.3. IMRT mean and standard deviation for total treatments without cutoff.
Low energy is defined as 4 or 6 MV x-rays and high energy treatments are any treatments
above 10 MV.
Next the trend is plotted with the cutoffs discussed in section two:
18
Total IMRT Treatments
0
1
2
3
4
5
6
7
8
9
10
0 5 10 15 20 25
MV
CI
Total
Linear (Total)
Figure 3.3. Mean values of CI for each MV for total treatments with cutoff.
Energy Sample Size Mean Standard Deviation
4 1 8.01 0.00
6 43 5.02 2.14
15 15 4.81 2.75
16 2 3.85 0.06
18 6 3.62 1.72
20 1 3.59 0.00
Low Energy 44 5.09 2.16
High Energy 24 4.38 2.36
Table 3.4. IMRT mean and standard deviation for total treatments with cutoff.
It is interesting to note that with or without a cutoff both low and high energy means are
very similar. The low energy mean is 5.50 with a standard deviation of 3.06 without a
cutoff and 5.09 and 2.16 with a cutoff. The high energy mean and standard deviation
without a cutoff are greatly affected by the 36.52 MU/cGy outlier. Without a cutoff the
19
high energy mean and standard deviation are 5.81 and 6.96, with a cutoff they are 4.38
and 2.36.
The remainder of the plots uses the data with the cutoffs. The next couple is for
mono and dual energy machines. A mono energy machine is defined as a machine that is
used for only one photon energy. Likewise, a dual energy machine uses two photon
energies.
Mono Energy Machines
0
1
2
3
4
5
6
7
8
9
10
0 5 10 15 20 25
MV
CI
Mono Energy Machines
Linear (Mono Energy Machines)
Figure 3.4. Mean values of CI for each MV for mono energy machines.
Energy Sample Size Mean Standard Deviation
4 1 8.01 0
6 4 4.885 1.824582144
15 1 5.23 0
16 0 0 0
18 0 0 0
20 0 0 0
Low Energy 5 5.51 5.92
High Energy 1 5.23 0.00
Table 3.5. IMRT mean and standard deviation for mono energy machines.
20
Dual Energy Machines
0
1
2
3
4
5
6
7
8
9
10
0 5 10 15 20 25
MV
CI
Dual Energy Machines
Linear (Dual Energy Machines)
Figure 3.5. Mean values of CI for each MV for dual energy machines.
Energy Sample Size Mean Standard Deviation
4 0 0.00 0.00
6 39 5.04 2.19
15 14 4.78 2.86
16 2 3.85 0.06
18 6 3.62 1.72
20 1 3.59 0.00
Low Energy 39 5.04 2.19
High Energy 23 4.35 2.41
Table 3.6. IMRT mean and standard deviation for dual energy machines.
With a small sample for mono energy machines, five low energy machines and one high
energy machine the numbers are easily skewed, and are very similar to the numbers
without a cutoff. The dual energy machine results are very close to the mean and
standard deviation of the total with cutoff.
21
The next pair is for mono and dual use machines. Use refers to how the machine
treats IMRT. A mono use machine only uses one energy for IMRT, and a dual use
machine treats IMRT patients at both of its photon energies.
Mono Use Machines
0
1
2
3
4
5
6
7
8
9
10
0 5 10 15 20 25
MV
CI
Mono Use Machines
Linear (Mono Use Machines)
Figure 3.6. Mean values of CI for each MV for mono use machines.
Energy Sample Size Mean Standard Deviation
4 1 8.01 0.00
6 23 5.11 1.25
15 1 5.23 0.00
16 1 3.80 0.00
18 1 3.37 0.00
20 0 0.00 0.00
Low Energy 24 5.24 2.65
High Energy 3 4.13 0.97
Table 3.7. IMRT mean and standard deviation for mono use machines.
22
Dual Use Machines
0
1
2
3
4
5
6
7
8
9
10
0 5 10 15 20 25
MV
CI
Dual Use Machines
Linear (Dual Use Machines)
Figure 3.7. Mean values of CI for each MV for dual use machines.
Energy Sample Size Mean Standard Deviation
4 0 0.00 0.00
6 20 4.92 2.88
15 14 4.78 2.86
16 1 3.89 0.00
18 5 3.67 1.92
20 1 3.59 0.00
Low Energy 20 4.92 2.88
High Energy 21 4.42 2.51
Table 3.8. IMRT mean and standard deviation for dual use machines.
In this case the low energy sample was almost evenly split with 24 mono use and 20 dual
use and it can be seen that their numbers are very close and well within a standard
deviation of the total. The high energy sample has only three for mono energy and 21 for
dual energy, yet the results are still within a standard deviation of the total.
23
The next two plots divide the data between large and small facilities. A small
facility is defined as only having one or two machines. A large facility has three or more
and would thus require a larger staff with more than a single doctor.
Large Facilities
0
1
2
3
4
5
6
7
8
9
10
0 5 10 15 20 25
MV
CI
Large Facilities
Linear (Large Facilities)
Figure 3.8. Mean values of CI for each MV for large facilities.
Energy Sample Size Mean Standard Deviation
4 1 8.01 0.00
6 34 4.80 1.49
15 14 4.81 2.86
16 2 3.85 0.06
18 3 3.82 0.42
20 1 3.59 0.00
Low Energy 35 4.90 2.10
High Energy 20 4.51 2.42
Table 3.9. IMRT mean and standard deviation for large facilities.
24
Small Facilities
0
1
2
3
4
5
6
7
8
9
10
0 5 10 15 20 25
MV
CI
Small Facilities
Linear (Small Facilities)
Figure 3.9. Mean values of CI for each MV for small facilities.
Energy Sample Size Mean Standard Deviation
4 0 0.00 0.00
6 9 5.85 2.55
15 1 4.82 0.00
16 0 0.00 0.00
18 3 3.43 2.67
20 0 0.00 0.00
Low Energy 9 5.85 2.55
High Energy 4 3.78 2.29
Table 3.10. IMRT mean and standard deviation for small facilities.
The large facility sample group is much larger than the small facility group, with 35 low
energy and 20 high energy samples. The small facility group has 9 low energy and 4
high energy samples.
The trend of interest is that in all cases, besides the group computed without a
cutoff where a single point was greatly skewing the results, is that the numbers are
25
consistently lower, in MU/cGy, for high energy treatments, but still close (within a
standard deviation). Also there is no significant difference for machine type or use or
facility size. For low energy calculations for WL a CI of 5.1 may be used, this number
may be increased by its standard deviation if the facility engages in procedures such as
head and necks which use many more MU/cGy than other procedures. For high energy
calculations of WL a CI of 4.4 may be used. Again this number may be increased if a
machine is specialized in procedures that use more MU/cGy. It is interesting to note that
in both cases the mean plus two standard deviations for the calculated value of the CI, are
less than the maximum value of ten for the range discussed in NCRP Report 151.
The low amount of TBI treatments, a total of nine patients for 184 out of 374,003
total exposures, makes any conclusions drawn tenuous at best. In all cases the workload
in MU/cGy was much higher than conventional treatments, as would be expected. For 6
and 15 MV treatments the mean MU/cGy hovered around 20, and for 18 MV TBI
treatments the number increased to a little over 30.
In conclusion, although the numbers are too scant to make any statements about
the contribution TBI makes to workload, there is more than enough data to estimate the
IMRT contribution to leakage workload. An IMRT factor, CI, of 5.1 may be used for low
energy photon calculations, and a CI of 4.4 may be used for high energy photon
calculations. These numbers are higher than the studies of Followill and Mechalakos, but
except in the case of low energy IMRT for Mechalakos are within a standard deviation.
26
APPENDIX: FIGURES AND TABLES FROM RESULTS
HSN RSN MV of IMRT Table Number page
1 1 6 and 18 2.2 29
1 2 6 2.3 31
2 4 15 2.4 33
2 7 6 and 15 2.5 35
3 2 6 2.6 37
3 4 6 2.7 39
3 5 6 2.8 41
3 6 6 2.9 43
5 2 6 and 18 2.10 45
5 3 6 and 20 2.11 47
6 2 6 2.12 49
7 1 6 and 15 2.13 51
8 1 6 2.14 53
8 2 6 2.14 53
9 1 6 2.15 55
10 2 15 2.16 57
10 3 6 and 15 2.17 59
11 1 6 and 18 2.18 61
12 1 6 2.19 63
12 2 6 and 18 2.20 65
13 1 6 and 18 2.21 67
14 1 6 2.22 69
15 2 6 and 15 2.23 71
15 3 4 2.24 73
15 4 6 and 15 2.25 75
16 1 6 2.26 77
16 2 6 2.27 79
16 3 6 2.28 81
17 1 6 2.29 83
17 2 6 2.30 85
17 4 6 2.31 87
17 5 6 2.32 89
17 6 6 2.32 91
17 7 6 2.32 91
17 8 6 2.32 91
27
18 1 6 2.33 91
18 2 6 and 15 2.34 93
18 3 6 and 15 2.35 95
18 4 6 and 15 2.36 97
18 5 6 and 15 2.37 99
18 6 6 2.38 101
19 1 6 and 15 2.39 103
19 2 6 and 15 2.40 105
20 1 16 2.41 107
20 2 6 and 16 2.42 109
21 1 18 2.43 111
22 1 6 and 15 2.44 113
22 2 6 and 15 2.45 115
Table 2.1. Page location of machines by HSN, RSN, and MV combination.
28
0.0
10.0
20.0
30.0
40.0
50.0
60.0
70.0
80.0
90.0
0 2 4 6 8 10 12
MU/cGy
N
RSN 1 6X IMRT
RSN 1 6X
RSN 1 18X IMRT
RSN 1 18X
RSN 1 All IMRT
RSN 1 All Energies
Figure 2.1. HSN 1, RSN 1 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
29
RSN 1 6X IMRT
RSN 1 6X
RSN 1 18X IMRT
RSN 1 18X
RSN 1 All IMRT
RSN 1 All Energies
mean(0-15)= 4.23 2.66 1.58 1.35 4.08 2.18
mean(0-25)= 4.23 2.66 1.58 1.35 4.08 2.18
mean(0-50)= 4.94 2.97 1.58 1.35 4.76 2.38
mean(0-200)= 4.94 2.97 1.58 1.35 4.76 2.38
mean(0-1500)= 4.94 2.97 1.58 1.35 4.76 2.38
stdev(0-15)= 2.33 1.98 0.24 0.28 2.34 1.71
stdev(0-25)= 2.33 1.98 0.24 0.28 2.34 1.71
stdev(0-50)= 4.76 3.47 0.24 0.28 4.69 2.88
stdev(0-200)= 4.76 3.47 0.24 0.28 4.69 2.88
stdev(0-1500)= 4.76 3.47 0.24 0.28 4.69 2.88
Table 2.2. HSN 1, RSN 1 mean and standard deviation at different cutoffs of MU/cGy.
30
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
40.0
0 5 10 15 20
MU/cGy
N
RSN 2 6X IMRT
RSN 2 6X
Figure 2.2. HSN 1, RSN 2 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
31
RSN 2 6X IMRT RSN 2 6X
mean(0-15)= 3.17 2.06
mean(0-25)= 4.95 2.58
mean(0-50)= 4.95 2.58
mean(0-200)= 4.95 2.58
mean(0-1500)= 4.95 2.58
stdev(0-15)= 0.77 0.93
stdev(0-25)= 5.11 3.10
stdev(0-50)= 5.11 3.10
stdev(0-200)= 5.11 3.10
stdev(0-1500)= 5.11 3.10
Table 2.3. HSN 1, RSN 2 mean and standard deviation at different cutoffs of MU/cGy.
32
0.0
10.0
20.0
30.0
40.0
50.0
60.0
0 2 4 6 8 10
MU/cGy
N
RSN 4 15X IMRT
RSN 4 15X
Figure 2.3. HSN 2, RSN 4 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
33
RSN 4 15X IMRT RSN 4 15X
mean(0-15)= 3.14 1.72
mean(0-25)= 3.14 1.72
mean(0-50)= 5.63 2.20
mean(0-200)= 17.68 4.83
mean(0-1500)= 36.52 9.24
stdev(0-15)= 1.10 0.85
stdev(0-25)= 1.10 0.85
stdev(0-50)= 10.11 4.66
stdev(0-200)= 31.11 15.73
stdev(0-1500)= 77.39 38.58
Table 2.4. HSN 2, RSN 4 mean and standard deviation at different cutoffs of MU/cGy.
34
0.0
10.0
20.0
30.0
40.0
50.0
60.0
70.0
0 2 4 6 8 10
MU/cGy
N
RSN 7 6X IMRT
RSN 7 6X
RSN 7 15X IMRT
RSN 7 15X
RSN 7 All IMRT
RSN 7 All Energies
Figure 2.4. HSN 2, RSN 7 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
35
RSN 7 6X IMRT
RSN 7 6X
RSN 7 15X IMRT
RSN 7 15X
RSN 7 All IMRT
RSN 7 All Energies
mean(0-15)= 5.85 1.89 3.00 1.53 5.04 1.67
mean(0-25)= 5.85 1.89 3.00 1.53 5.04 1.67
mean(0-50)= 9.17 2.46 3.00 1.53 7.63 1.91
mean(0-200)= 9.17 2.46 3.00 1.53 7.63 1.91
mean(0-1500)= 9.17 2.46 3.00 1.53 7.63 1.91
stdev(0-15)= 2.29 1.75 0.75 0.53 2.36 1.20
stdev(0-25)= 2.29 1.75 0.75 0.53 2.36 1.20
stdev(0-50)= 7.71 4.04 0.75 0.53 7.20 2.68
stdev(0-200)= 7.71 4.04 0.75 0.53 7.20 2.68
stdev(0-1500)= 7.71 4.04 0.75 0.53 7.20 2.68
Table 2.5. HSN 2, RSN 7 mean and standard deviation at different cutoffs of MU/cGy.
36
0.0
2.0
4.0
6.0
8.0
10.0
12.0
14.0
16.0
0 2 4 6 8 10 12 14
MU/cGy
N
RSN 2 6X IMRT
RSN 2 6X
Figure 2.5. HSN 3, RSN 2 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
37
RSN 2 6X IMRT RSN 2 6X
mean(0-15)= 4.88 4.61
mean(0-25)= 4.89 4.63
mean(0-50)= 4.89 4.63
mean(0-200)= 4.89 4.87
mean(0-1500)= 4.89 4.87
stdev(0-15)= 1.85 2.02
stdev(0-25)= 1.9 2.06
stdev(0-50)= 1.9 2.06
stdev(0-200)= 1.9 5.7
stdev(0-1500)= 1.9 5.7
Table 2.6. HSN 3, RSN 2 mean and standard deviation at different cutoffs of MU/cGy.
38
0.0
5.0
10.0
15.0
20.0
25.0
30.0
0 2 4 6 8 10 12 14
MU/cGy
N
RSN 4 6X IMRT
RSN 4 6X
Figure 2.6. HSN 3, RSN 4 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
39
RSN 4 6X IMRT RSN 4 6X
mean(0-15)= 4.54 2.77
mean(0-25)= 4.54 2.77
mean(0-50)= 4.54 2.77
mean(0-200)= 4.54 2.77
mean(0-1500)= 4.54 2.77
stdev(0-15)= 1.18 1.65
stdev(0-25)= 1.18 1.65
stdev(0-50)= 1.18 1.65
stdev(0-200)= 1.18 1.65
stdev(0-1500)= 1.18 1.65
Table 2.7. HSN 3, RSN 4 mean and standard deviation at different cutoffs of MU/cGy.
40
0.0
2.0
4.0
6.0
8.0
10.0
12.0
14.0
16.0
18.0
0 2 4 6 8 10 12 14
MU/cGy
N
RSN 5 6X IMRT
RSN 5 6X
Figure 2.7. HSN 3, RSN 5 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
41
RSN 5 6X IMRT RSN 5 6X
mean(0-15)= 5.78 4.49
mean(0-25)= 6.04 4.69
mean(0-50)= 7.36 5.67
mean(0-200)= 7.36 5.67
mean(0-1500)= 7.36 5.67
stdev(0-15)= 3.00 3.18
stdev(0-25)= 3.69 3.72
stdev(0-50)= 6.72 6.25
stdev(0-200)= 6.72 6.25
stdev(0-1500)= 6.72 6.25
Table 2.8. HSN 3, RSN 5 mean and standard deviation at different cutoffs of MU/cGy.
42
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
0 2 4 6 8 10 12 14
MU/cGy
N
RSN 6 6X IMRT
RSN 6 6X
Figure 2.8. HSN 3, RSN 6 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
43
RSN 6 6X IMRT RSN 6 6X
mean(0-15)= 4.53 2.76
mean(0-25)= 4.53 2.76
mean(0-50)= 4.53 2.76
mean(0-200)= 4.53 4.75
mean(0-1500)= 4.53 4.75
stdev(0-15)= 1.43 1.52
stdev(0-25)= 1.43 1.52
stdev(0-50)= 1.43 1.52
stdev(0-200)= 1.43 16.19
stdev(0-1500)= 1.43 16.19
Table 2.9. HSN 3, RSN 6 mean and standard deviation at different cutoffs of MU/cGy.
44
0.0
10.0
20.0
30.0
40.0
50.0
60.0
70.0
0 1 2 3 4 5 6 7
MU/cGy
N
RSN 2 6X IMRT
RSN 2 6X
RSN 2 18X IMRT
RSN 2 18X
RSN 2 All IMRT
RSN 2 All Energies
Figure 2.9. HSN 5, RSN 2 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
45
RSN 2 6X IMRT
RSN 2 6X
RSN 2 18X IMRT
RSN 2 18X
RSN 2 All IMRT
RSN 2 All Energies
mean(0-15)= 3.19 1.44 3.87 3.20 3.86 2.73
mean(0-25)= 3.19 1.44 3.87 3.20 3.86 2.73
mean(0-50)= 3.19 1.44 3.87 3.20 3.86 2.73
mean(0-200)= 3.19 27.81 3.87 13.83 3.86 17.86
mean(0-1500)= 3.19 27.81 3.87 13.83 3.86 17.86
stdev(0-15)= 0.77 0.47 1.82 1.95 1.81 1.86
stdev(0-25)= 0.77 0.47 1.82 1.95 1.81 1.86
stdev(0-50)= 0.77 0.47 1.82 1.95 1.81 1.86
stdev(0-200)= 0.77 55.99 1.82 37.75 1.81 44.25
stdev(0-1500)= 0.77 55.99 1.82 37.75 1.81 44.25
Table 2.10. HSN 5, RSN 2 mean and standard deviation at different cutoffs of MU/cGy.
46
0.0
10.0
20.0
30.0
40.0
50.0
60.0
70.0
0 2 4 6 8
MU/cGy
N
RSN 3 6X IMRT
RSN 3 6X
RSN 3 20X IMRT
RSN 3 20X
RSN 3 All IMRT
RSN 3 All Energies
Figure 2.10. HSN 5, RSN 3 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
47
RSN 3 6X IMRT
RSN 3 6X
RSN 3 20X IMRT
RSN 3 20X
RSN 3 All IMRT
RSN 3 All Energies
mean(0-15)= 3.55 1.58 3.59 2.95 3.59 2.52
mean(0-25)= 3.55 1.58 3.59 2.95 3.59 2.52
mean(0-50)= 3.55 1.58 3.59 2.95 3.59 2.52
mean(0-200)= 3.55 25.65 3.59 13.72 3.59 17.67
mean(0-1500)= 3.55 25.65 3.59 13.72 3.59 17.67
stdev(0-15)= 0.40 0.62 0.99 1.39 0.97 1.36
stdev(0-25)= 0.40 0.62 0.99 1.39 0.97 1.36
stdev(0-50)= 0.40 0.62 0.99 1.39 0.97 1.36
stdev(0-200)= 0.40 54.23 0.99 37.98 0.97 44.40
stdev(0-1500)= 0.40 54.23 0.99 37.98 0.97 44.40
Table 2.11. HSN 5, RSN 3 mean and standard deviation at different cutoffs of MU/cGy.
48
0.0
2.0
4.0
6.0
8.0
10.0
12.0
14.0
0 10 20 30 40 50
MU/cGy
N
RSN 2 6X IMRT
RSN 2 6X
Figure 2.11. HSN 6, RSN 2 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
49
RSN 2 6X IMRT RSN 2 6X
mean(0-15)= 4.64 4.45
mean(0-25)= 5.04 4.85
mean(0-50)= 6.29 6.03
mean(0-200)= 6.29 6.03
mean(0-1500)= 6.29 6.03
stdev(0-15)= 2.27 2.34
stdev(0-25)= 3.25 3.28
stdev(0-50)= 7.12 7.03
stdev(0-200)= 7.12 7.03
stdev(0-1500)= 7.12 7.03
Table 2.12. HSN 6, RSN 2 mean and standard deviation at different cutoffs of MU/cGy.
50
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
0 2 4 6 8 10 12 14
MU/cGy
N
RSN 1 6X IMRT
RSN 1 6X
RSN 1 15X IMRT
RSN 1 15X
RSN 1 All IMRT
RSN 1 All Energies
Figure 2.12. HSN 7, RSN 1 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
51
RSN 1 6X IMRT
RSN 1 6X
RSN 1 15X IMRT
RSN 1 15X
RSN 1 All IMRT
RSN 1 All Energies
mean(0-15)= 6.51 4.21 4.77 4.07 5.28 4.12
mean(0-25)= 6.73 4.37 4.82 4.11 5.39 4.21
mean(0-50)= 6.77 4.39 4.82 4.15 5.40 4.24
mean(0-200)= 6.77 4.39 4.82 4.47 5.40 4.44
mean(0-1500)= 6.77 4.39 4.82 191.22 5.40 126.72
stdev(0-15)= 2.26 2.99 2.64 2.74 2.65 2.84
stdev(0-25)= 2.74 3.31 2.74 2.83 2.88 3.03
stdev(0-50)= 2.87 3.38 2.74 3.08 2.92 3.20
stdev(0-200)= 2.87 3.38 2.74 5.79 2.92 5.01
stdev(0-1500)= 2.87 3.38 2.74 623.21 2.92 512.06
Table 2.13. HSN 7, RSN 1 mean and standard deviation at different cutoffs of MU/cGy.
52
0.0
2.0
4.0
6.0
8.0
10.0
12.0
0 5 10 15 20
MU/cGy
N
RSN 1 6X IMRT
RSN 1 6X
RSN 2 6X IMRT
RSN 2 6X
Figure 2.13. HSN 8, RSN 1 and 2 Frequency distribution of MU/cGy, plotted versus bin
mid-point value. The different x-ray beams and combinations are indicated on the graph.
53
RSN 1 6X IMRT
RSN 1 6X
RSN 2 6X IMRT
RSN 2 6X
mean(0-15)= 6.95 6.40 7.15 6.57
mean(0-25)= 7.15 6.58 7.35 6.76
mean(0-50)= 7.22 6.65 7.82 7.19
mean(0-200)= 7.39 6.81 9.87 9.05
mean(0-1500)= 8.33 7.66 10.15 9.31
stdev(0-15)= 2.34 2.79 2.51 2.97
stdev(0-25)= 2.80 3.18 2.96 3.34
stdev(0-50)= 3.12 3.44 4.89 5.04
stdev(0-200)= 5.02 5.11 17.19 16.55
stdev(0-1500)= 17.96 17.22 18.85 18.13
Table 2.14. HSN 8, RSN 1 and 2 mean and standard deviation at different cutoffs of
MU/cGy.
54
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
0 2 4 6 8 10
MU/cGy
N
RSN 1 6X IMRT
RSN 1 6X
Figure 2.14. HSN 9, RSN 1 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
55
RSN 1 6X IMRT RSN 1 6X
mean(0-15)= 3.86 2.61
mean(0-25)= 3.86 2.61
mean(0-50)= 3.86 2.61
mean(0-200)= 3.86 2.61
mean(0-1500)= 3.86 2.61
stdev(0-15)= 1.57 1.60
stdev(0-25)= 1.57 1.60
stdev(0-50)= 1.57 1.60
stdev(0-200)= 1.57 1.60
stdev(0-1500)= 1.57 1.60
Table 2.15. HSN 9, RSN 1 mean and standard deviation at different cutoffs of MU/cGy.
56
0.0
2.0
4.0
6.0
8.0
10.0
12.0
14.0
0 5 10 15 20
MU/cGy
N
RSN 2 15X IMRT
RSN 2 15X
Figure 2.15. HSN 10, RSN 2 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
57
RSN 2 15X IMRT RSN 2 15X
mean(0-15)= 5.23 5.23
mean(0-25)= 5.23 5.23
mean(0-50)= 5.23 5.23
mean(0-200)= 5.23 5.23
mean(0-1500)= 5.23 5.23
stdev(0-15)= 1.78 1.78
stdev(0-25)= 1.78 1.78
stdev(0-50)= 1.78 1.78
stdev(0-200)= 1.78 1.78
stdev(0-1500)= 1.78 1.78
Table 2.16. HSN 10, RSN 2 mean and standard deviation at different cutoffs of MU/cGy.
58
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
40.0
0 5 10 15 20 25 30 35 40
MU/cGy
N
RSN 3 6X IMRT
RSN 3 6X
RSN 3 15X IMRT
RSN 3 All IMRT
RSN 3 All Energies
RSN 3 15X
Figure 2.16. HSN 10, RSN 3 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
59
RSN 3 6X IMRT
RSN 3 6X
RSN 3 15X IMRT
RSN 3 15X
RSN 3 All IMRT
RSN 3 All Energies
mean(0-15)= 7.82 5.22 10.69 5.62 8.82 5.38
mean(0-25)= 10.91 7.66 11.37 6.15 11.04 7.11
mean(0-50)= 11.52 8.18 14.14 7.95 12.37 8.10
mean(0-200)= 11.52 8.18 14.14 7.95 12.37 8.10
mean(0-1500)= 11.52 8.18 14.14 7.95 12.37 8.10
stdev(0-15)= 3.26 3.81 2.45 4.86 3.30 4.26
stdev(0-25)= 6.58 6.78 3.30 5.44 5.65 6.30
stdev(0-50)= 7.62 7.76 8.34 8.65 7.92 8.09
stdev(0-200)= 7.62 7.76 8.34 8.65 7.92 8.09
stdev(0-1500)= 7.62 7.76 8.34 8.65 7.92 8.09
Table 2.17. HSN 10, RSN 3 mean and standard deviation at different cutoffs of MU/cGy.
60
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
40.0
45.0
50.0
0 2 4 6 8 10 12 14
MU/cGy
N
RSN 1 6X IMRT
RSN 1 6X
RSN 1 18X IMRT
RSN 1 18X
RSN 1 All IMRT
RSN 1 All Energies
Figure 2.17. HSN 11, RSN 1 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
61
Table 2.18. HSN 11, RSN 1 mean and standard deviation at different cutoffs of MU/cGy.
RSN 1 6X IMRT
RSN 1 6X
RSN 1 18X IMRT
RSN 1 18X
RSN 1 All IMRT
RSN 1 All Energies
mean(0-15)= 1.70 1.90 2.22 1.68 2.14 1.79
mean(0-25)= 1.70 1.90 2.22 1.68 2.14 1.79
mean(0-50)= 1.70 1.90 2.22 1.68 2.14 1.79
mean(0-200)= 1.70 1.90 2.22 17.93 2.14 10.49
mean(0-1500)= 1.70 1.90 2.22 17.93 2.14 10.49
stdev(0-15)= 0.56 0.65 1.19 1.38 1.13 1.09
stdev(0-25)= 0.56 0.65 1.19 1.38 1.13 1.09
stdev(0-50)= 0.56 0.65 1.19 1.38 1.13 1.09
stdev(0-200)= 0.56 0.65 1.19 45.91 1.13 34.56
stdev(0-1500)= 0.56 0.65 1.19 45.91 1.13 34.56
62
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
40.0
0 2 4 6 8 10
MU/cGy
N
RSN 1 6X IMRT
RSN 1 6X
Figure 2.18. HSN 12, RSN 1 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
63
RSN 1 6X IMRT RSN 1 6X
mean(0-15)= 4.72 2.23
mean(0-25)= 4.93 2.29
mean(0-50)= 5.41 2.41
mean(0-200)= 5.87 2.48
mean(0-1500)= 5.87 2.48
stdev(0-15)= 2.15 1.49
stdev(0-25)= 2.88 1.81
stdev(0-50)= 4.66 2.67
stdev(0-200)= 6.85 3.30
stdev(0-1500)= 6.85 3.30
Table 2.18. HSN 12, RSN 1 mean and standard deviation at different cutoffs of MU/cGy.
64
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
40.0
45.0
50.0
0 2 4 6 8 10
MU/cGy
N
RSN 2 6X IMRT
RSN 2 6X
RSN 2 18X IMRT
RSN 2 18X
RSN 2 All IMRT
RSN 2 All Energies
Figure 2.19. HSN 12, RSN 2 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
65
Table 2.20. HSN 12, RSN 2 mean and standard deviation at different cutoffs of MU/cGy.
RSN 2 6X IMRT
RSN 2 6X
RSN 2 18X IMRT
RSN 2 18X
RSN 2 All IMRT
RSN 2 All Energies
mean(0-15)= 4.90 2.02 3.87 2.65 3.88 2.58
mean(0-25)= 4.90 2.02 4.00 2.73 4.01 2.66
mean(0-50)= 4.90 2.08 4.21 2.85 4.22 2.77
mean(0-200)= 4.90 2.08 4.21 2.85 4.22 2.77
mean(0-1500)= 4.90 2.08 4.21 2.85 4.22 2.77
stdev(0-15)= 1.18 1.24 2.10 2.05 2.10 1.99
stdev(0-25)= 1.18 1.24 2.56 2.36 2.55 2.28
stdev(0-50)= 1.18 2.02 3.46 2.97 3.44 2.90
stdev(0-200)= 1.18 2.02 3.46 2.97 3.44 2.90
stdev(0-1500)= 1.18 2.02 3.46 2.97 3.44 2.90
66
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
0 5 10 15 20 25
MU/cGy
N
RSN 1 6X IMRT
RSN 1 6X
RSN 1 18X IMRT
RSN 1 18X
RSN 1 All IMRT
RSN 1 All Energies
Figure 2.20. HSN 13, RSN 1 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
67
Table 2.21. HSN 13, RSN 1 mean and standard deviation at different cutoffs of MU/cGy.
RSN 1 6X IMRT
RSN 1 6X
RSN 1 18X IMRT
RSN 1 18X
RSN 1 All IMRT
RSN 1 All Energies
mean(0-15)= 6.37 5.56 6.00 3.81 6.22 4.63
mean(0-25)= 8.21 7.22 6.49 4.23 7.59 5.70
mean(0-50)= 10.76 9.48 7.10 4.76 9.49 7.16
mean(0-200)= 13.60 11.97 7.37 4.89 11.48 8.53
mean(0-1500)= 19.08 16.72 7.37 4.89 15.14 11.01
stdev(0-15)= 3.38 3.57 2.37 2.97 3.03 3.38
stdev(0-25)= 5.54 5.62 3.40 3.84 4.94 5.03
stdev(0-50)= 9.99 9.78 6.00 5.98 8.99 8.47
stdev(0-200)= 19.81 18.85 6.91 6.47 16.85 14.69
stdev(0-1500)= 60.42 56.46 6.91 6.47 49.67 41.28
68
0.0
10.0
20.0
30.0
40.0
50.0
60.0
0 5 10 15 20 25
MU/cGy
N
RSN 1 6X IMRT
RSN 1 6X
Figure 2.21. HSN 14, RSN 1 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
69
RSN 1 6X IMRT RSN 1 6X
mean(0-15)= 3.70 2.19
mean(0-25)= 3.95 3.06
mean(0-50)= 3.95 3.06
mean(0-200)= 3.95 3.06
mean(0-1500)= 3.95 3.06
stdev(0-15)= 1.36 1.45
stdev(0-25)= 2.46 4.11
stdev(0-50)= 2.46 4.11
stdev(0-200)= 2.46 4.11
stdev(0-1500)= 2.46 4.11
Table 2.22. HSN 14, RSN 1 mean and standard deviation at different cutoffs of MU/cGy.
70
0.0
10.0
20.0
30.0
40.0
50.0
60.0
0 2 4 6 8 10
MU/cGy
N
RSN 2 6X IMRT
RSN 2 6X
RSN 2 15X IMRT
RSN 2 15X
RSN 2 All IMRT
RSN 2 All Energies
Figure 2.22. HSN 15, RSN 2 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
71
Table 2.23. HSN 15, RSN 2 mean and standard deviation at different cutoffs of MU/cGy.
RSN 2 6X IMRT
RSN 2 6X
RSN 2 15X IMRT
RSN 2 15X
RSN 2 All IMRT
RSN 2 All Energies
mean(0-15)= 5.66 2.86 4.40 3.57 4.53 3.43
mean(0-25)= 5.66 2.86 4.40 3.57 4.53 3.43
mean(0-50)= 5.66 2.86 4.40 3.57 4.53 3.43
mean(0-200)= 5.66 2.86 4.40 3.57 4.53 3.43
mean(0-1500)= 5.66 2.86 4.40 3.57 4.53 3.43
stdev(0-15)= 1.08 2.12 1.79 2.06 1.78 2.10
stdev(0-25)= 1.08 2.12 1.79 2.06 1.78 2.10
stdev(0-50)= 1.08 2.12 1.79 2.06 1.78 2.10
stdev(0-200)= 1.08 2.12 1.79 2.06 1.78 2.10
stdev(0-1500)= 1.08 2.12 1.79 2.06 1.78 2.10
72
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
40.0
45.0
0 5 10 15 20
MU/cGy
N
RSN 3 4X IMRT
RSN 3 4X
Figure 2.23. HSN 15, RSN 3 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
73
RSN 3 4X IMRT RSN 3 4X
mean(0-15)= 7.91 3.97
mean(0-25)= 8.01 4.09
mean(0-50)= 8.01 4.09
mean(0-200)= 8.01 4.09
mean(0-1500)= 8.01 4.09
stdev(0-15)= 2.76 3.53
stdev(0-25)= 2.89 3.72
stdev(0-50)= 2.89 3.72
stdev(0-200)= 2.89 3.72
stdev(0-1500)= 2.89 3.72
Table 2.24. HSN 15, RSN 3 mean and standard deviation at different cutoffs of MU/cGy.
74
0.0
10.0
20.0
30.0
40.0
50.0
60.0
0 2 4 6 8 10 12 14
MU/cGy
N
RSN 4 6X IMRT
RSN 4 6X
RSN 4 15X IMRT
RSN 4 15X
RSN 4 All IMRT
RSN 4 All Energies
Figure 2.24. HSN 15, RSN 4 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
75
Table 2.25. HSN 15, RSN 4 mean and standard deviation at different cutoffs of MU/cGy.
RSN 4 6X IMRT
RSN 4 6X
RSN 4 15X IMRT
RSN 4 15X
RSN 4 All IMRT
RSN 4 All Energies
mean(0-15)= 6.57 2.64 6.67 4.22 6.66 3.80
mean(0-25)= 6.57 2.64 6.67 4.22 6.66 3.80
mean(0-50)= 6.57 2.64 6.67 4.22 6.66 3.80
mean(0-200)= 6.57 2.64 6.67 4.22 6.66 3.80
mean(0-1500)= 6.57 2.64 6.67 4.22 6.66 3.80
stdev(0-15)= 2.35 2.48 2.75 3.33 2.70 3.20
stdev(0-25)= 2.35 2.48 2.75 3.33 2.70 3.20
stdev(0-50)= 2.35 2.48 2.75 3.33 2.70 3.20
stdev(0-200)= 2.35 2.48 2.75 3.33 2.70 3.20
stdev(0-1500)= 2.35 2.48 2.75 3.33 2.70 3.20
76
0.0
2.0
4.0
6.0
8.0
10.0
12.0
14.0
16.0
18.0
20.0
0 2 4 6 8 10
MU/cGy
N
RSN 1 6X IMRT
RSN 1 6X
Figure 2.25. HSN 16, RSN 1 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
77
RSN 1 6X IMRT RSN 1 6X
mean(0-15)= 3.82 3.28
mean(0-25)= 3.89 3.33
mean(0-50)= 3.96 3.39
mean(0-200)= 3.96 3.39
mean(0-1500)= 3.96 3.39
stdev(0-15)= 1.58 1.73
stdev(0-25)= 1.85 1.95
stdev(0-50)= 2.25 2.26
stdev(0-200)= 2.25 2.26
stdev(0-1500)= 2.25 2.26
Table 2.26. HSN 16, RSN 1 mean and standard deviation at different cutoffs of MU/cGy.
78
0.0
2.0
4.0
6.0
8.0
10.0
12.0
14.0
16.0
18.0
20.0
0 2 4 6 8 10
MU/cGy
N
RSN 2 6X IMRT
RSN 2 6X
Figure 2.26. HSN 16, RSN 2 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
79
RSN 2 6X IMRT RSN 2 6X
mean(0-15)= 3.81 3.20
mean(0-25)= 3.81 3.20
mean(0-50)= 3.81 3.20
mean(0-200)= 3.81 3.20
mean(0-1500)= 3.81 3.20
stdev(0-15)= 1.38 1.61
stdev(0-25)= 1.38 1.61
stdev(0-50)= 1.38 1.61
stdev(0-200)= 1.38 1.61
stdev(0-1500)= 1.38 1.61
Table 2.27. HSN 16, RSN 2 mean and standard deviation at different cutoffs of MU/cGy.
80
0.0
5.0
10.0
15.0
20.0
25.0
30.0
0 2 4 6 8 10
MU/cGy
N
RSN 3 6X IMRT
RSN 3 6X
Figure 2.27. HSN 16, RSN 3 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
81
RSN 3 6X IMRT RSN 3 6X
mean(0-15)= 3.96 2.92
mean(0-25)= 3.96 2.92
mean(0-50)= 3.96 2.92
mean(0-200)= 3.96 2.92
mean(0-1500)= 4.25 3.10
stdev(0-15)= 1.48 1.73
stdev(0-25)= 1.48 1.73
stdev(0-50)= 1.48 1.73
stdev(0-200)= 1.48 1.73
stdev(0-1500)= 13.26 10.38
Table 2.28. HSN 16, RSN 3 mean and standard deviation at different cutoffs of MU/cGy.
82
0.0
5.0
10.0
15.0
20.0
25.0
0 2 4 6 8 10 12 14
MU/cGy
N
RSN 1 6X IMRT
RSN 1 6X
Figure 2.28. HSN 17, RSN 1 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
83
RSN 1 6X IMRT RSN 1 6X
mean(0-15)= 4.79 3.35
mean(0-25)= 5.17 3.58
mean(0-50)= 5.44 3.74
mean(0-200)= 5.71 3.89
mean(0-1500)= 5.71 3.89
stdev(0-15)= 2.25 2.27
stdev(0-25)= 3.32 2.99
stdev(0-50)= 4.19 3.60
stdev(0-200)= 5.51 4.52
stdev(0-1500)= 5.51 4.52
Table 2.29. HSN 17, RSN 1 mean and standard deviation at different cutoffs of MU/cGy.
84
0.0
2.0
4.0
6.0
8.0
10.0
12.0
14.0
0 2 4 6 8 10 12 14
MU/cGy
N
RSN 2 6X IMRT
RSN 2 6X
Figure 2.29. HSN 17, RSN 2 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
85
RSN 2 6X IMRT RSN 2 6X
mean(0-15)= 7.08 6.01
mean(0-25)= 7.18 6.1
mean(0-50)= 7.25 6.16
mean(0-200)= 7.25 6.16
mean(0-1500)= 7.25 6.16
stdev(0-15)= 2.34 3.01
stdev(0-25)= 2.58 3.18
stdev(0-50)= 2.97 3.46
stdev(0-200)= 2.97 3.46
stdev(0-1500)= 2.97 3.46
Table 2.30. HSN 17, RSN 2 mean and standard deviation at different cutoffs of MU/cGy.
86
0.0
2.0
4.0
6.0
8.0
10.0
12.0
14.0
16.0
0 5 10 15 20
MU/cGy
N
RSN 4 6X IMRT
RSN 4 6X
Figure 2.30. HSN 17, RSN 4 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
87
RSN 4 6X IMRT RSN 4 6X
mean(0-15)= 6.32 4.21
mean(0-25)= 6.37 4.32
mean(0-50)= 6.43 4.39
mean(0-200)= 6.43 4.42
mean(0-1500)= 6.44 4.43
stdev(0-15)= 2.11 2.77
stdev(0-25)= 2.31 3.1
stdev(0-50)= 3.05 3.71
stdev(0-200)= 4.95 8.34
stdev(0-1500)= 32.48 24.18
Table 2.31. HSN 17, RSN 4 mean and standard deviation at different cutoffs of
MU/cGy.
88
0.0
2.0
4.0
6.0
8.0
10.0
12.0
14.0
16.0
18.0
0 5 10 15 20
MU/cGy
N
RSN 5 6X IMRT
RSN 6 6X IMRT
RSN 7 6X IMRT
RSN 8 6X IMRT
Figure 2.31. HSN 17, RSN 5, 6, 7, and 8 Frequency distribution of MU/cGy, plotted
versus bin mid-point value. The different x-ray beams and combinations are indicated on
the graph.
89
RSN 5 6X IMRT
RSN 6 6X IMRT
RSN 7 6X IMRT
RSN 8 6X IMRT
mean(0-15)= 4.54 5.44 5.27 5.26
mean(0-25)= 4.55 5.46 5.27 5.38
mean(0-50)= 4.60 5.46 5.27 5.38
mean(0-200)= 4.60 5.46 5.27 5.38
mean(0-1500)= 4.60 5.46 5.27 5.38
stdev(0-15)= 2.13 2.23 2.40 2.30
stdev(0-25)= 2.24 2.33 2.40 2.70
stdev(0-50)= 2.86 2.33 2.40 2.70
stdev(0-200)= 2.86 2.33 2.40 2.70
stdev(0-1500)= 2.86 2.33 2.40 2.70
Table 2.32. HSN 17, RSN 5, 6, 7, and 8 mean and standard deviation at different cutoffs
of MU/cGy.
90
0.0
2.0
4.0
6.0
8.0
10.0
12.0
14.0
16.0
18.0
20.0
0 2 4 6 8 10 12
MU/cGy
N
RSN 1 6X IMRT
RSN 1 6X
Figure 2.32. HSN 18, RSN 1 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
91
RSN 1 6X IMRT RSN 1 6X
mean(0-15)= 4.21 3.83
mean(0-25)= 4.21 3.83
mean(0-50)= 4.21 3.83
mean(0-200)= 4.21 3.83
mean(0-1500)= 4.21 3.83
stdev(0-15)= 1.96 2.06
stdev(0-25)= 1.96 2.06
stdev(0-50)= 1.96 2.06
stdev(0-200)= 1.96 2.06
stdev(0-1500)= 1.96 2.06
Table 2.33. HSN 18, RSN 1 mean and standard deviation at different cutoffs of MU/cGy.
92
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
40.0
0 3 6 9 12 15
MU/cGy
NRSN 2 6X IMRT
RSN 2 6X
RSN 2 15X IMRT
RSN 2 15X
RSN 2 All IMRT
RSN 2 All Energies
Figure 2.33. HSN 18, RSN 2 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
93
RSN 2 6X IMRT
RSN 2 6X
RSN 2 15X IMRT
RSN 2 15X
RSN 2 All IMRT
RSN 2 All Energies
mean(0-15)= 3.68 2.75 3.53 3.02 3.55 2.89
mean(0-25)= 3.68 2.75 3.53 3.02 3.55 2.89
mean(0-50)= 3.68 2.75 3.53 3.02 3.55 2.89
mean(0-200)= 3.68 2.75 3.53 3.02 3.55 2.89
mean(0-1500)= 3.68 2.75 3.53 3.02 3.55 2.89
stdev(0-15)= 1.76 1.76 0.84 1.13 0.99 1.27
stdev(0-25)= 1.76 1.76 0.84 1.13 0.99 1.27
stdev(0-50)= 1.76 1.76 0.84 1.13 0.99 1.27
stdev(0-200)= 1.76 1.76 0.84 1.13 0.99 1.27
stdev(0-1500)= 1.76 1.76 0.84 1.13 0.99 1.27
Table 2.34. HSN 18, RSN 2 mean and standard deviation at different cutoffs of MU/cGy.
94
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
0 2 4 6 8
MU/cGy
NRSN 3 6X IMRT
RSN 3 6X
RSN 3 15X IMRT
RSN 3 15X
RSN 3 All IMRT
RSN 3 All Energies
Figure 2.34. HSN 18, RSN 3 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
95
RSN 3 6X IMRT
RSN 3 6X
RSN 3 15X IMRT
RSN 3 15X
RSN 3 All IMRT
RSN 3 All Energies
mean(0-15)= 6.18 3.90 3.54 3.10 3.61 3.10
mean(0-25)= 6.18 3.90 3.54 3.10 3.61 3.10
mean(0-50)= 6.18 3.90 3.54 3.10 3.61 3.10
mean(0-200)= 6.18 3.90 3.54 3.10 3.61 3.10
mean(0-1500)= 6.18 3.90 3.54 3.10 3.61 3.10
stdev(0-15)= 0.98 2.40 0.60 1.06 0.73 1.16
stdev(0-25)= 0.98 2.40 0.60 1.06 0.73 1.16
stdev(0-50)= 0.98 2.40 0.60 1.06 0.73 1.16
stdev(0-200)= 0.98 2.40 0.60 1.06 0.73 1.16
stdev(0-1500)= 0.98 2.40 0.60 1.06 0.73 1.16
Table 2.35. HSN 18, RSN 3 mean and standard deviation at different cutoffs of MU/cGy.
96
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
0 2 4 6 8 10
MU/cGy
N
RSN 4 6X IMRT
RSN 4 6X
RSN 4 15X IMRT
RSN 4 15X
RSN 4 All IMRT
RSN 4 All Energies
Figure 2.35. HSN 18, RSN 4 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
97
RSN 4 6X IMRT
RSN 4 6X
RSN 4 15X IMRT
RSN 4 15X
RSN 4 All IMRT
RSN 4 All Energies
mean(0-15)= 3.48 2.64 3.81 3.10 3.74 2.90
mean(0-25)= 3.48 2.64 3.81 3.10 3.74 2.90
mean(0-50)= 3.48 2.64 3.81 3.10 3.74 2.90
mean(0-200)= 3.48 2.64 3.81 3.10 3.74 2.90
mean(0-1500)= 3.48 2.64 3.81 3.10 3.74 2.90
stdev(0-15)= 0.78 1.25 0.84 3.31 0.84 1.33
stdev(0-25)= 0.78 1.25 0.84 3.31 0.84 1.33
stdev(0-50)= 0.78 1.25 0.84 3.31 0.84 1.33
stdev(0-200)= 0.78 1.25 0.84 3.31 0.84 1.33
stdev(0-1500)= 0.78 1.25 0.84 3.31 0.84 1.33
Table 2.36. HSN 18, RSN 4 mean and standard deviation at different cutoffs of MU/cGy.
98
0.0
10.0
20.0
30.0
40.0
50.0
60.0
70.0
80.0
90.0
100.0
0 1 2 3 4 5 6 7 8 9 10
MU/cGy
N
RSN 5 6X IMRT
RSN 5 6X
RSN 5 15X IMRT
RSN 5 15X
RSN 5 All IMRT
RSN 5 All Energies
Figure 2.36. HSN 18, RSN 5 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
99
RSN 5 6X IMRT
RSN 5 6X
RSN 5 15X IMRT
RSN 5 15X
RSN 5 All IMRT
RSN 5 All Energies
mean(0-15)= 1.75 1.51 3.36 2.74 3.26 2.40
mean(0-25)= 1.75 1.51 3.36 3.26 3.26 2.80
mean(0-50)= 1.75 1.51 3.36 3.26 3.26 2.80
mean(0-200)= 1.75 1.51 3.36 3.26 3.26 2.80
mean(0-1500)= 1.75 1.51 3.36 3.26 3.26 2.80
stdev(0-15)= 0.00 0.40 1.07 1.29 1.10 1.27
stdev(0-25)= 0.00 0.40 1.07 1.29 1.10 1.27
stdev(0-50)= 0.00 0.40 1.07 1.29 1.10 1.27
stdev(0-200)= 0.00 0.40 1.07 1.29 1.10 1.27
stdev(0-1500)= 0.00 0.40 1.07 1.29 1.10 1.27
Table 2.37. HSN 18, RSN 5 mean and standard deviation at different cutoffs of MU/cGy.
100
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
40.0
45.0
0 2 4 6 8 10
MU/cGy
N
RSN 6 6X IMRT
RSN 6 6X
Figure 2.37. HSN 18, RSN 6 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
101
Table 2.38. HSN 18, RSN 6 mean and standard deviation at different cutoffs of MU/cGy.
RSN 6 6X IMRT RSN 6 6X
mean(0-15)= 3.04 2.78
mean(0-25)= 3.04 2.78
mean(0-50)= 3.04 2.78
mean(0-200)= 3.04 2.78
mean(0-1500)= 3.04 2.78
stdev(0-15)= 1.53 1.54
stdev(0-25)= 1.53 1.54
stdev(0-50)= 1.53 1.54
stdev(0-200)= 1.53 1.54
stdev(0-1500)= 1.53 1.54
102
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
40.0
0 2 4 6 8 10
MU/cGy
N
RSN 1 6X IMRT
RSN 1 6X
RSN 1 15X IMRT
RSN 1 15X
RSN 1 All IMRT
RSN 1 All Energies
Figure 2.38. HSN 19, RSN 1 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
103
RSN 1 6X IMRT
RSN 1 6X
RSN 1 15X IMRT
RSN 1 15X
RSN 1 All IMRT
RSN 1 All Energies
mean(0-15)= 3.02 2.44 3.95 3.59 3.35 2.64
mean(0-25)= 3.02 2.44 3.95 3.59 3.35 2.64
mean(0-50)= 3.02 2.44 3.95 3.59 3.35 2.64
mean(0-200)= 3.02 2.44 3.95 3.59 3.35 2.64
mean(0-1500)= 3.02 2.44 3.95 3.59 3.35 2.64
stdev(0-15)= 1.47 1.40 0.74 1.11 1.34 1.41
stdev(0-25)= 1.47 1.40 0.74 1.11 1.34 1.41
stdev(0-50)= 1.47 1.40 0.74 1.11 1.34 1.41
stdev(0-200)= 1.47 1.40 0.74 1.11 1.34 1.41
stdev(0-1500)= 1.47 1.40 0.74 1.11 1.34 1.41
Table 2.39. HSN 19, RSN 1 mean and standard deviation at different cutoffs of MU/cGy.
104
0.0
10.0
20.0
30.0
40.0
50.0
60.0
0 2 4 6 8 10
MU/cGy
N
RSN 2 6X IMRT
RSN 2 6X
RSN 2 15X IMRT
RSN 2 15X
RSN 2 All IMRT
RSN 2 All Energies
Figure 2.39. HSN 19, RSN 2 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
105
RSN 2 6X IMRT
RSN 2 6X
RSN 2 15X IMRT
RSN 2 15X
RSN 2 All IMRT
RSN 2 All Energies
mean(0-15)= 2.50 1.83 5.01 3.91 4.76 3.44
mean(0-25)= 2.50 1.83 5.01 3.91 4.76 3.44
mean(0-50)= 2.50 1.83 5.01 3.91 4.76 3.44
mean(0-200)= 2.50 1.83 5.01 3.91 4.76 3.44
mean(0-1500)= 2.50 1.83 5.01 3.91 4.76 3.44
stdev(0-15)= 0.25 0.52 1.96 2.28 2.00 2.22
stdev(0-25)= 0.25 0.52 1.96 2.28 2.00 2.22
stdev(0-50)= 0.25 0.52 1.96 2.28 2.00 2.22
stdev(0-200)= 0.25 0.52 1.96 2.28 2.00 2.22
stdev(0-1500)= 0.25 0.52 1.96 2.28 2.00 2.22
Table 2.40. HSN 19, RSN 2 mean and standard deviation at different cutoffs of MU/cGy.
106
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
40.0
0 2 4 6 8 10
MU/cGy
N
RSN 1 16X IMRT
RSN 1 16X
Figure 2.40. HSN 20, RSN 1 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
107
RSN 1 16X IMRT RSN 1 16X
mean(0-15)= 3.80 2.60
mean(0-25)= 3.80 2.60
mean(0-50)= 3.80 2.60
mean(0-200)= 3.80 2.60
mean(0-1500)= 3.80 2.60
stdev(0-15)= 0.82 1.33
stdev(0-25)= 0.82 1.33
stdev(0-50)= 0.82 1.33
stdev(0-200)= 0.82 1.33
stdev(0-1500)= 0.82 1.33
Table 2.41. HSN 20, RSN 1 mean and standard deviation at different cutoffs of MU/cGy.
108
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
40.0
0 2 4 6 8 10 12 14
MU/cGy
N
RSN 2 6X IMRT
RSN 2 6X
RSN 2 16X IMRT
RSN 2 16X
RSN 2 All IMRT
RSN 2 All Energies
Figure 2.41. HSN 20, RSN 2 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
109
RSN 2 6X IMRT
RSN 2 6X
RSN 2 16X IMRT
RSN 2 16X
RSN 2 All IMRT
RSN 2 All Energies
mean(0-15)= 8.57 4.73 3.89 3.18 5.82 3.91
mean(0-25)= 8.57 4.73 3.89 3.18 5.82 3.91
mean(0-50)= 8.57 4.73 3.89 3.18 5.82 3.91
mean(0-200)= 8.57 4.73 3.89 3.18 5.82 3.91
mean(0-1500)= 8.57 4.73 3.89 3.18 5.82 3.91
stdev(0-15)= 1.87 3.68 1.62 1.77 2.88 18.92
stdev(0-25)= 1.87 3.68 1.62 1.77 2.88 18.92
stdev(0-50)= 1.87 3.68 1.62 1.77 2.88 18.92
stdev(0-200)= 1.87 3.68 1.62 1.77 2.88 18.92
stdev(0-1500)= 1.87 3.68 1.62 1.77 2.88 18.92
Table 2.42. HSN 20, RSN 2 mean and standard deviation at different cutoffs of MU/cGy.
110
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
0 1 2 3 4 5 6 7 8
MU/cGy
N
RSN 1 18X IMRT
RSN 1 18X
Figure 2.42. HSN 21, RSN 1 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
111
RSN 1 18X IMRT RSN 1 18X
mean(0-15)= 3.37 3.08
mean(0-25)= 3.37 3.08
mean(0-50)= 3.37 3.08
mean(0-200)= 3.37 3.08
mean(0-1500)= 3.37 3.08
stdev(0-15)= 0.68 0.95
stdev(0-25)= 0.68 0.95
stdev(0-50)= 0.68 0.95
stdev(0-200)= 0.68 0.95
stdev(0-1500)= 0.68 0.95
Table 2.43. HSN 21, RSN 1 mean and standard deviation at different cutoffs of MU/cGy.
112
0.0
10.0
20.0
30.0
40.0
50.0
60.0
70.0
80.0
90.0
100.0
0 1 2 3 4 5
MU/cGy
N
RSN 1 6X IMRT
RSN 1 6X
RSN 1 15X IMRT
RSN 1 15X
RSN 1 All IMRT
RSN 1 All Energies
Figure 2.43. HSN 22, RSN 1 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
113
Table 2.44. HSN 22, RSN 1 mean and standard deviation at different cutoffs of MU/cGy.
RSN 1 6X IMRT
RSN 1 6X
RSN 1 15X IMRT
RSN 1 15X
RSN 1 All IMRT
RSN 1 All Energies
mean(0-15)= 1.96 1.64 3.53 2.93 2.83 1.99
mean(0-25)= 1.96 1.64 3.53 2.93 2.83 1.99
mean(0-50)= 1.96 1.64 3.53 2.93 2.83 1.99
mean(0-200)= 1.96 1.64 3.53 2.93 2.83 1.99
mean(0-1500)= 1.96 1.64 3.53 2.93 2.83 1.99
stdev(0-15)= 0.63 0.54 0.40 0.92 0.94 0.89
stdev(0-25)= 0.63 0.54 0.40 0.92 0.94 0.89
stdev(0-50)= 0.63 0.54 0.40 0.92 0.94 0.89
stdev(0-200)= 0.63 0.54 0.40 0.92 0.94 0.89
stdev(0-1500)= 0.63 0.54 0.40 0.92 0.94 0.89
114
0.0
10.0
20.0
30.0
40.0
50.0
60.0
70.0
80.0
90.0
100.0
0 1 2 3 4 5 6 7 8
MU/cGy
N
RSN 2 6X IMRT
RSN 2 6X
RSN 2 15X IMRT
RSN 2 15X
RSN 2 All IMRT
RSN 2 All Energies
Figure 2.44. HSN 22, RSN 2 Frequency distribution of MU/cGy, plotted versus bin mid-
point value. The different x-ray beams and combinations are indicated on the graph.
115
RSN 2 6X IMRT
RSN 2 6X
RSN 2 15X IMRT
RSN 2 15X
RSN 2 All IMRT
RSN 2 All Energies
mean(0-15)= 1.75 1.70 4.08 3.53 3.66 2.96
mean(0-25)= 1.75 1.70 4.08 3.53 3.66 2.96
mean(0-50)= 1.75 1.70 4.08 3.53 3.66 2.96
mean(0-200)= 1.75 1.70 4.08 3.53 3.66 2.96
mean(0-1500)= 1.75 1.70 4.08 3.53 3.66 2.96
stdev(0-15)= 0.00 0.72 1.35 1.57 1.51 1.61
stdev(0-25)= 0.00 0.72 1.35 1.57 1.51 1.61
stdev(0-50)= 0.00 0.72 1.35 1.57 1.51 1.61
stdev(0-200)= 0.00 0.72 1.35 1.57 1.51 1.61
stdev(0-1500)= 0.00 0.72 1.35 1.57 1.51 1.61
Table 2.45. HSN 22, RSN 2 mean and standard deviation at different cutoffs of MU/cGy.
116
4X
HSN RSN IMRT MU/cGy C IMRT/C FI
Total Treatments
IMRT Treatments
15 3 8.03 1.72 4.66 0.38 2364 892
Table 2.46. 4 MV MU/cGy with unrestricted values.
6X IMRT Total IMRT
HSN RSN MU/cGy C IMRT/C FI Treatments Treatments
1 1 4.96 1.57 3.15 0.41 2062 842
1 2 5.04 1.65 3.05 0.28 1963 540
2 7 9.22 1.28 7.21 0.14 790 114
3 2 4.91 1.72 2.86 0.92 6174 5702
3 4 4.59 1.56 2.94 0.40 1705 679
3 5 7.37 1.55 4.74 0.71 1678 1193
3 6 4.51 1.94 2.32 0.31 577 178
5 2 3.11 1.47 2.11 0.03 1883 56
5 3 3.58 1.42 2.53 0.06 2381 141
6 2 6.33 1.24 5.11 0.95 1215 1154
7 1 6.78 1.62 4.19 0.53 2943 1567
8 1 8.34 1.25 6.67 0.90 6934 6271
8 2 10.17 1.20 8.47 0.91 4925 4459
9 1 3.88 1.75 2.22 0.41 2615 1067
10 3 11.52 2.15 5.35 0.64 1720 1104
11 1 1.68 1.90 0.88 0.08 2871 226
12 1 5.67 1.76 3.23 0.13 12608 1679
12 2 4.51 1.65 2.73 0.06 2672 169
13 1 19.11 1.75 10.93 0.86 10117 8738
14 1 3.96 1.21 3.28 0.36 2236 816
15 2 5.65 1.46 3.87 0.34 618 208
15 4 6.56 1.42 4.62 0.23 630 148
16 1 3.99 1.36 2.93 0.78 9035 7036
16 2 3.84 1.32 2.90 0.75 8780 6574
16 3 4.27 1.37 3.11 0.60 6797 4089
17 1 5.72 1.68 3.41 0.54 5669 3056
17 2 5.43 1.74 3.12 0.80 8260 6625
17 4 7.58 2.74 2.77 0.52 15330 7915
17 5 4.70 1.77 2.65 0.75 8871 6632
17 6 5.48 3.61 1.52 0.61 8005 4867
17 7 5.29 6.16 0.86 0.82 6731 5517
17 8 5.44 5.78 0.94 0.87 10120 8822
18 1 4.22 1.45 2.91 0.86 3603 3106
18 2 3.67 1.42 2.58 0.56 662 372
117
18 3 6.27 1.58 3.96 0.50 168 84
18 4 3.47 1.37 2.53 0.61 751 455
18 5 1.67 1.33 1.25 0.27 286 78
18 6 3.01 1.37 2.20 0.85 2416 2063
19 1 2.87 1.28 2.25 0.39 458 178
19 2 2.43 1.52 1.61 0.29 304 88
20 2 8.59 1.68 5.11 0.44 1160 514
22 1 1.85 1.52 1.22 0.22 653 143
22 2 1.65 1.71 0.96 0.42 448 186
Table 2.47. 6 MV MU/cGy with unrestricted values.
15X
HSN RSN IMRT MU/cGy C IMRT/C FI Total Treatments IMRT Treatments
2 4 36.58 1.38 26.56 0.22 1759 392
2 7 2.91 1.45 2.01 0.03 1107 38
7 1 4.84 1.34 3.62 0.77 4746 3658
10 2 5.24 1 5.24 1.00 14943 14943
10 3 14.14 1.49 9.48 0.51 1030 526
15 2 4.43 1.30 3.41 0.73 2487 1814
15 4 6.69 1.34 4.98 0.54 1711 918
18 2 3.54 1.57 2.26 0.73 3769 2769
18 3 3.55 1.38 2.58 0.80 4317 3453
18 4 3.83 1.54 2.49 0.70 2604 1830
18 5 3.39 1.38 2.46 0.67 1693 1140
19 1 4.08 1.75 2.34 0.79 925 733
19 2 4.98 1.59 3.14 0.68 1201 816
22 1 3.46 1.88 1.85 0.64 274 176
22 2 4.12 1.71 2.41 0.77 1111 853
Table 2.48. 15 MV MU/cGy with unrestricted values.
118
16X
HSN RSN IMRT MU/cGy C IMRT/C FI
Total Treatments
IMRT Treatments
20 1 3.84 1.43 2.68 0.49 973 475
20 2 3.91 1.47 2.66 0.71 1038 733
Table 2.49. 16 MV MU/cGy with unrestricted values.
18X
HSN RSN IMRT
MU/cGy C IMRT/C FI Total
Treatments IMRT
Treatments
1 1 1.65 1.33 1.24 0.04 1185 50
5 2 3.88 1.18 3.30 0.74 5195 3836
11 1 2.21 1.17 1.90 0.41 2987 1232
12 2 4.23 1.21 3.50 0.54 23604 12859
13 1 7.40 2.67 2.77 0.47 9440 4439
21 1 3.39 1.49 2.28 0.85 3070 2605
Table 2.50. 18 MV MU/cGy with unrestricted values.
20X
HSN RSN IMRT
MU/cGy C IMRT/C FI Total
Treatments IMRT
Treatments
5 3 3.57 1.42 2.52 0.73 5259 3850
Table 2.51. 20 MV MU/cGy with unrestricted values.
119
4X
HSN RSN IMRT MU/cGy
15 3 8.01
Table 2.52. 4 MV MU/cGy with cutoffs
6X
HSN RSN IMRT MU/cGy HSN RSN IMRT MU/cGy
1 1 4.94 16 1 3.96
1 2 4.95 16 2 3.81
2 7 5.85 16 3 3.96
3 2 4.89 17 1 5.71
3 4 4.54 17 2 7.25
3 5 7.36 17 4 6.43
3 6 4.53 17 5 4.60
5 2 3.19 17 6 5.46
5 3 3.55 17 7 5.27
6 2 5.04 17 8 5.38
7 1 6.77 18 1 4.21
8 1 7.15 18 2 3.68
8 2 7.35 18 3 6.18
9 1 3.86 18 4 3.48
10 3 11.52 18 5 1.75
11 1 1.7 18 6 3.04
12 1 4.93 19 1 3.02
120
12 2 4.90 19 2 2.50
13 1 10.76 20 2 8.57
14 1 3.95 22 1 1.96
15 2 5.66 22 2 1.75
15 4 6.57
Table 2.53. 6 MV MU/cGy with cutoffs.
15X
HSN RSN IMRT MU/cGy HSN RSN IMRT MU/cGy
2 4 3.14 18 3 3.54
2 7 3.00 18 4 3.81
7 1 4.82 18 5 3.36
10 2 5.23 19 1 3,95
10 3 14.14 19 2 5.01
15 2 4.40 22 1 3.53
15 4 6.67 22 2 4.08
18 2 3.53
Table 2.54. 15 MV MU/cGy with cutoffs.
121
16X
HSN RSN IMRT MU/cGy HSN RSN IMRT MU/cGy
20 1 3.80 20 2 3.89
Table 2.55. 16 MV MU/cGy with cutoffs.
18X
HSN RSN IMRT MU/cGy HSN RSN IMRT MU/cGy
1 1 1.58 12 2 4.21
5 2 3.87 13 1 6.49
11 1 2.22 21 1 3.37
Table 2.56. 18 MV MU/cGy with cutoffs.
20X
HSN RSN IMRT MU/cGy
5 3 3.59
Table 2.57. 20 MV MU/cGy with cutoffs.
TBI
HSN RSN TBI MU/cGy C TBI/C FI Total
Treatments TBI Treatments
9 1 35.65 1.39 25.57 0.02 1368 33
18X
14 1 19.89 1.21 16.48 0.04 2236 98
6X
16 2 23.71 1.31 18.16 0.01 2182 28
18X
18 5 20.89 1.38 15.13 0.01 1693 25
15X
Table 2.58. TBI MU/cGy with unrestricted values and different energies.
122
BIBLIOGRAPHY
1. FOLLOWILL, D., GEIS, P., and BOYER, A. (1997). “Estimates of whole-body dose
equivalent produced by beam intensity modulated conformal therapy,” Int. J.
Radiat. Oncol. Biol. Phys. 38(3), 667-672. (Errata 38, 783).
2. MCGINLEY, P.H. (2002). Shielding Techniques for Radiation Oncology Facilities,
2nd
ed. (Medical Physics Publishing, Madison, Wisconsin).
3. MECHALAKOS, J. and ST. GERMAIN, J. (2002). “Estimation of linear accelerator
workloads and use factors,” Health Physics 83(11 Suppl), S65-67.
4. MECHALAKOS, J., ST. GERMAIN, J., and BURMAN, C.M. (2004). “Results of a
one year survey of output for linear accelerators using IMRT and non-IMRT
techniques,” J. Appl. Clin. Med. Phys. 5(1), 64-72.
5. MUTIC, S. and LOW, D.A. (1998). “Whole-body dose from tomotherapy delivery,”
Int. J. Radiat. Oncol. Biol. Phys. 42(1), 229-232.
6. NCRP (2005) National Council on Radiation Protection and Measurements.
Structural Shielding Design and Evaluation for Megavoltage X- and Gamma-Ray
Radiotherapy Facilities, NCRP Report No. 151 (National Council on Radiation
Protection and Measurements, Bethesda, Maryland).
7. PURDY, J.A., BOYER, A.L., BUTLER, E.B., DIPETRILLO, T.A., ENGLER, M.J.,
FRAASS, B., GRANT, W., III., LING, C.C., LOW, D.A., MACKIE, T.R.,
MOHAN, R., ROACH, M., ROSENMAN, J.G., VERHEY, L.J., WONG, J.W.,
CUMBERLIN, R.L., STONE, H., and PALTA, J.R. (2001). “Intensity-modulated
radiotherapy: Current status and issues of interest,” Int. J. Radiat. Oncol. Biol.
Phys. 51(4), 880-914.
8. RODGERS, J.E. (2001). “Radiation therapy vault shielding calculational methods
when IMRT and TBI procedures contribute,” J. Appl. Clin. Med. Phys. 2(3), 157-
164.