PRIMING DRAWING THE CLEARING VOLUME 1 Push the reservoir plunger to the closed and locked position. 2 With the shut-off valve in the open position (parallel to the tubing), hold sampling site above the Edwards VAMP Plus reservoir at 45° angle. 5 Release plunger latch and smoothly draw back on the reservoir plunger until the plunger stops at its 12 cc volume capacity. (Recommended draw rate is 1 second for each cc.) 4 Ensure removal of all air from system. Connect to patient’s catheter. 6 Close the shut-off valve by turning the handle perpendicular to the tubing. 3 Provide flow by pulling the Snap-Tab of the Edwards TruWave disposable pressure transducer. Slowly fill and debubble the reservoir and each sample site in succession. Shut-off valve in open position. Shut-off valve in closed position. VAMP Plus System Inservice Bulletin Venous Arterial blood Management Protection system Safe and Accurate Closed Blood Sampling System Featuring the Edwards TruWave Disposable Pressure Transducer

AR05990 Vamp Plus Inservice - ht.edwards.comht.edwards.com/.../ar05990vampplusinservice.pdfPlus reservoir at 45° angle. 5 Release plunger latch and smoothly draw back on the reservoir

  • Upload
    others

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

Page 1: AR05990 Vamp Plus Inservice - ht.edwards.comht.edwards.com/.../ar05990vampplusinservice.pdfPlus reservoir at 45° angle. 5 Release plunger latch and smoothly draw back on the reservoir

P r i m i n g

D r a w i n g t h e C l e a r i n g V o l u m e

1 Pushthereservoirplungertotheclosedandlocked

position.

2 Withtheshut-offvalveintheopenposition(paralleltothetubing),holdsamplingsiteabovetheEdwardsVAMPPlusreservoirat45°angle.

5 Releaseplungerlatchandsmoothlydrawbackonthereservoirplunger

untiltheplungerstopsatits12ccvolumecapacity.(Recommendeddrawrateis1secondfor

eachcc.)

4 Ensureremovalofallairfromsystem.Connecttopatient’scatheter.

6 Closetheshut-offvalvebyturningthehandle

perpendiculartothetubing.

3 ProvideflowbypullingtheSnap-TaboftheEdwardsTruWavedisposablepressuretransducer.Slowlyfillanddebubblethereservoirand

eachsamplesiteinsuccession.

Shut-off valve in open position.

Shut-off valve in closed position.

VamP Plus System inservice BulletinVenous arterial blood management Protection systemSafe and Accurate Closed Blood Sampling SystemFeaturing the Edwards TruWave Disposable Pressure Transducer

Page 2: AR05990 Vamp Plus Inservice - ht.edwards.comht.edwards.com/.../ar05990vampplusinservice.pdfPlus reservoir at 45° angle. 5 Release plunger latch and smoothly draw back on the reservoir

Edwards Lifesciences Irvine,USAINyon,SwitzerlandITokyo,JapanISingapore,SingaporeISãoPaulo,Braziledwards.com

Drawing BlooD SamPleS from the VamP PluS neeDleleSS SamPling Site

For professional use. CAUTION: Federal (United States) law restricts this device to sale by or on the order of a physician. See instructions for use for full prescribing information, including indications, contraindications, warnings, precautions and adverse events.EdwardsisatrademarkofEdwardsLifesciencesCorporation.EdwardsLifesciences,thestylizedElogo,TruWave,VAMPandVAMPPlusaretrademarksofEdwardsLifesciencesCorporationandareregisteredintheUnitedStatesPatentandTrademarkOffice.©2011EdwardsLifesciencesCorporation.Allrightsreserved.AR05990

No components of this package or the product it contains are made from natural rubber latex or dry natural rubber.

8 Slowlydrawthebloodsample.

10 Opentheshut-offvalvebyturningthehandleparallel

totubing.

9 Whileholdingthecannula,removethesyringe/cannulatogetherfromthesamplingsitebypullingstraightout.

11 Slowly,smoothly,andevenly,reinfusetheclearingvolume.(Recommendedinfusionrateis1secondfor

eachcc.)

12 FlushtheVAMPPlussystembypulling

theSnap-TabontheTruWave

transducerand swabthesampling sitetoremove

any excessblood.

7 Swabsample site.Ensureplunger

isdepressedtothebottomofreservoir.Pushcannulawithsyringeor

Direct-Draw unitinto samplesite.