Aqua Glo Inspections

Presented by

Howard GammonGammon Technical Products

Inc.

Page 2: Aqua Glo Inspections Presented by Howard Gammon Gammon Technical Products Inc

Aqua Glo Inspections

There are two standards required for every Aqua Glo instrument.Fluorescing

StandardCalibrating

Standard

Page 3: Aqua Glo Inspections Presented by Howard Gammon Gammon Technical Products Inc

Aqua Glo Inspections

Remove the Fluorescing Standard held by 4 screws under the hinged flap.

Page 4: Aqua Glo Inspections Presented by Howard Gammon Gammon Technical Products Inc

Aqua Glo Inspections

Read the Set Point on the back of the Fluorescing Standard

Page 5: Aqua Glo Inspections Presented by Howard Gammon Gammon Technical Products Inc

Aqua Glo Inspections

Set the adjusting lever to be perpendicular to the side of the comparator chamber.

Page 6: Aqua Glo Inspections Presented by Howard Gammon Gammon Technical Products Inc

Aqua Glo Inspections Read the position of the

lever on the scale.

If the lever reading is the same as the Fluorescing Standard set point, this confirms that the correct Standard is being used and that nobody has switched with another instrument.

Page 7: Aqua Glo Inspections Presented by Howard Gammon Gammon Technical Products Inc

Aqua Glo Inspections

Compare the Code Number on the Fluorescing Standard with the code on the Calibrating Standard. They must be the same. However, the Set Points are usually different. This is acceptable.

Page 8: Aqua Glo Inspections Presented by Howard Gammon Gammon Technical Products Inc

Aqua Glo Inspections

Newer Aqua Glo’s have the code number on a decal on the hinged cover. If yours does not have a decal, we recommend that you make one or scribe the code into the metal.

Page 9: Aqua Glo Inspections Presented by Howard Gammon Gammon Technical Products Inc

Aqua Glo Inspections

If the lever reading is different than the Fluorescing Standard Set Point, remove the Calibrated Scale and install a new one, positioned correctly.

Order P/N GTP-834-153 and you will receive a packet of 3, in case you do not install correctly.

Page 10: Aqua Glo Inspections Presented by Howard Gammon Gammon Technical Products Inc

Aqua Glo Inspections

If the code on the Calibrating Standard is different than the code on the Fluorescing Standard, order a new set of standards, GTP-835.

A new calibrated scale will be included.

Page 11: Aqua Glo Inspections Presented by Howard Gammon Gammon Technical Products Inc

Aqua Glo Inspections

Somebody is sure to ask why the Standards are not all the same.

Answer: Each one is manually painted in ultraviolet light and we have never mastered a technique to make them the same. There are too many variables.

Page 12: Aqua Glo Inspections Presented by Howard Gammon Gammon Technical Products Inc

Return-Repair Procedure

Call 1-800-524-0287 or 732-223-4600.

Say that you need an RFR number so that you can return an Aqua Glo for repair.

State your name, telephone number, unit number, and location unless it is classified.

We can ship to an APO address.

Page 13: Aqua Glo Inspections Presented by Howard Gammon Gammon Technical Products Inc

Return-Repair Procedure

Package the instrument carefully, preferably in it’s original case, and mail it to us with the RFR number on the OUTSIDE.

Gammon Technical Products2300 Hwy 34 (P.O. Box 400)

Manasquan, NJ 08736

Page 14: Aqua Glo Inspections Presented by Howard Gammon Gammon Technical Products Inc

Return-repair Procedure

Be sure to return the Calibrating and Fluorescing Standards and the instrument pack that mounts on the side of the chamber.

We will telephone you within 1 or 2 days, reporting the damage and the cost of repairs.

Page 15: Aqua Glo Inspections Presented by Howard Gammon Gammon Technical Products Inc

Return-Repair Procedure

If you elect to have us make the repair, give us an IMPAC card number. If you do not want the repair made, we will return it to you at our expense.

Repairs are normally completed in a day or two.

Page 16: Aqua Glo Inspections Presented by Howard Gammon Gammon Technical Products Inc

Other Services

We have an ongoing program to service the:

U.S. ArmyAviation Fuel

Contamination Test Kit: Petroleum

NSN: 6630-01-008-5524

Page 17: Aqua Glo Inspections Presented by Howard Gammon Gammon Technical Products Inc

Other Services

You can ship the entire box to us. We will quote the cost to repair or replace all missing items.

Alternatively, rather than shipping the box to us, we will quote the cost of any component in response to your inquiry.