Atmospheric and Climate Sciences, 2014, 4, 636-641 Published Online October 2014 in SciRes. http://www.scirp.org/journal/acs http://dx.doi.org/10.4236/acs.2014.44056 How to cite this paper: Nicosia, A., Belosi, F. and Vazquez, B. (2014) Application of the Soft X-Ray TSI Advanced Aerosol Neutralizer to Aerosol Measurements Made by a Grimm SMPS. Atmospheric and Climate Sciences, 4, 636-641. http://dx.doi.org/10.4236/acs.2014.44056 Application of the Soft X-Ray TSI Advanced Aerosol Neutralizer to Aerosol Measurements Made by a Grimm SMPS Alessia Nicosia, Franco Belosi, Brais Vazquez Institute of Atmospheric Sciences and Climate (ISAC), Italian National Research Council, Bologna, Italy Email: [email protected] Received 21 July 2014; revised 18 August 2014; accepted 6 September 2014 Copyright © 2014 by authors and Scientific Research Publishing Inc. This work is licensed under the Creative Commons Attribution International License (CC BY). http://creativecommons.org/licenses/by/4.0/ Abstract Scanning Mobility Particle Sizer (SMPS) spectrometers are subject to several restrictions when a radioactive source is employed to bring the aerosol to a steady-state charge distribution. An al- ternative solution, commercially available, is represented by the soft X-ray neutralizer. The present study investigates the outcome of a combination of a Grimm SMPS, which employs a 241 Am radioactive source, with the soft X-ray advanced aerosol neutralizer (TSI model 3087). To date, the latter device has been interfaced only with TSI’s Electrostatic Classifiers. Particle size distribution of sodium chloride aerosol was measured with both neutralizers and it was found that the particle number concentration agreed to be within 9% for the 10 - 700 nm range. This difference mainly corresponds to the bias reported by TSI, when the X-ray device is mounted on a TSI spectrometer. It was concluded that the X-ray neutralizer could be usefully employed, as a standalone device, in combination with Grimm Electrostatic Classifiers. Keywords Aerosol Neutralizer, SMPS, Particle Size Distribution 1. Introduction Scanning Mobility Particle Sizers (SMPS) are widely used to classify submicron particulate matter according to their electrical mobility. They employ radioactive sources such as 85 Kr, 210 Po or 241 Am to apply a well-known charge distribution to aerosols. However, these aerosol neutralizers are subject to several restrictions, especially in Europe, which limited their transport and handling. Commercially available alternatives are the aerosol neutralizers based on a low energy (<9.5 KeV) soft X-ray

Application of the Soft X-Ray TSI Advanced Aerosol ...file.scirp.org/pdf/ACS_2014093010362943.pdf · 5400 Vienna type) and a condensation particle counter ... compares well with the

Embed Size (px)

Citation preview

Page 1: Application of the Soft X-Ray TSI Advanced Aerosol ...file.scirp.org/pdf/ACS_2014093010362943.pdf · 5400 Vienna type) and a condensation particle counter ... compares well with the

Atmospheric and Climate Sciences, 2014, 4, 636-641 Published Online October 2014 in SciRes. http://www.scirp.org/journal/acs http://dx.doi.org/10.4236/acs.2014.44056

How to cite this paper: Nicosia, A., Belosi, F. and Vazquez, B. (2014) Application of the Soft X-Ray TSI Advanced Aerosol Neutralizer to Aerosol Measurements Made by a Grimm SMPS. Atmospheric and Climate Sciences, 4, 636-641. http://dx.doi.org/10.4236/acs.2014.44056

Application of the Soft X-Ray TSI Advanced Aerosol Neutralizer to Aerosol Measurements Made by a Grimm SMPS Alessia Nicosia, Franco Belosi, Brais Vazquez Institute of Atmospheric Sciences and Climate (ISAC), Italian National Research Council, Bologna, Italy Email: [email protected] Received 21 July 2014; revised 18 August 2014; accepted 6 September 2014

Copyright © 2014 by authors and Scientific Research Publishing Inc. This work is licensed under the Creative Commons Attribution International License (CC BY). http://creativecommons.org/licenses/by/4.0/

Abstract Scanning Mobility Particle Sizer (SMPS) spectrometers are subject to several restrictions when a radioactive source is employed to bring the aerosol to a steady-state charge distribution. An al-ternative solution, commercially available, is represented by the soft X-ray neutralizer. The present study investigates the outcome of a combination of a Grimm SMPS, which employs a 241Am radioactive source, with the soft X-ray advanced aerosol neutralizer (TSI model 3087). To date, the latter device has been interfaced only with TSI’s Electrostatic Classifiers. Particle size distribution of sodium chloride aerosol was measured with both neutralizers and it was found that the particle number concentration agreed to be within 9% for the 10 - 700 nm range. This difference mainly corresponds to the bias reported by TSI, when the X-ray device is mounted on a TSI spectrometer. It was concluded that the X-ray neutralizer could be usefully employed, as a standalone device, in combination with Grimm Electrostatic Classifiers.

Keywords Aerosol Neutralizer, SMPS, Particle Size Distribution

1. Introduction Scanning Mobility Particle Sizers (SMPS) are widely used to classify submicron particulate matter according to their electrical mobility. They employ radioactive sources such as 85Kr, 210Po or 241Am to apply a well-known charge distribution to aerosols. However, these aerosol neutralizers are subject to several restrictions, especially in Europe, which limited their transport and handling.

Commercially available alternatives are the aerosol neutralizers based on a low energy (<9.5 KeV) soft X-ray

Page 2: Application of the Soft X-Ray TSI Advanced Aerosol ...file.scirp.org/pdf/ACS_2014093010362943.pdf · 5400 Vienna type) and a condensation particle counter ... compares well with the

A. Nicosia et al.

637

source. Such devices can be easily turned on and off, thereby increasing both safety and overall operating life-time, and the respective regulations are more flexible.

Therefore, it is important to compare instruments working with radioactive and soft X-ray aerosol neutralizers. The novelty of this study was to evaluate the accuracy of the measurements obtained with a soft X-ray neutra-lizer (TSI, mod. 3087) in combination with a Grimm SMPS (L-DMA) under controlled laboratory conditions. TSI provides information on the neutralizer 3087 compared with the neutralizer 3077A (based on 85Kr). Howev-er, all data and specifications apply to TSI instruments, since the X-ray neutralizer 3087 was specifically de-signed to interface with TSI’s Electrostatic Classifier 3082.

Nevertheless, the aptness of this X-ray device assembled on a SMPS spectrometer other than TSI has not been reported in literature yet.

2. Experimental Set-Up and Procedures The experimental set-up is schematically represented in Figure 1. Sodium Chloride particles were generated from a solution concentration of 1% w/w, by means of a Collison nebulizer working at an operating pressure of 1 bar. The aerosol produced was slightly charged [1] and so, once dried inside a silica gel column, the particles were passed through a parallel plate condenser. The voltage applied to the condenser was previously adjusted in order to remove the charged portion of the aerosol, as verified by means of an aerosol electrometer. Experiments were carried out both with and without this voltage applied.

The resulting uncharged/charged polydisperse particles are hereinafter called “neutral/charged aerosol”, re-spectively. The aerosol was then diluted with purified air and carried into the 3087 TSI advanced aerosol neutra-lizer, followed by the Grimm SMPS spectrometer. Since the X-ray device was permanently fixed to the Grimm SMPS, the particles losses by Brownian diffusion inside the X-ray tube were the same for all the experiments. The 241Am neutralizer was removed every time the X-ray device was turned on. Conversely, when the 241Am neutralizer was inserted the X-ray device was switched off. This procedure was feasible as the aerosol particle number concentration was constant (±4% throughout 30 minutes of measurement controlled by a condensation particle counter, TSI mod. 3775). The Grimm SMPS spectrometer used in this study (SMPS + C Grimm Aerosol Technik GmbH & Co. KG, Ainring, Germany) consists of a long differential mobility analyser (L-DMA model 5400 Vienna type) and a condensation particle counter (model 5.403). The standard neutralizer is a 241Am source with an activity of 3.7 MBq. The particle size distribution was measured by neutralizing the aerosol alternatively with the 241Am source or the X-ray neutralizer.

The data acquisition software (Grimm Universal Nano Software Version 1.2.3) was used for all measure-ments, based on the hypothesis that the charging ratio was the same for both neutralizers. This assumption was validated by Shimada et al. [2], who observed that the particle charging probabilities were equal whether using a

Figure 1. Experimental set-up.

Page 3: Application of the Soft X-Ray TSI Advanced Aerosol ...file.scirp.org/pdf/ACS_2014093010362943.pdf · 5400 Vienna type) and a condensation particle counter ... compares well with the

A. Nicosia et al.

638

soft X-ray of <9.5 keV or a 241Am foil source of 3.7 MBq at two different particle residence times in the charg-ers (0.5 s and 3 s).

3. Results and Discussion The average particle size distributions obtained with the two neutralizers are compared in Figure 2.

It was found that there is a slight decrease in the measured particle number concentration obtained with X-ray when compared to 241Am. For neutral sodium chloride particles, the difference of the total concentration was about 9%, while for sodium chloride charged particles, it was about 3% (see Table 1).

Figure 3 reports the paired absolute difference percentage (PADP), according to Watson et al. [3], defined as

( )100* i i iPADP abs y x x= − (1)

where xi and yi refer to the particle concentration of the i-th bin for the Grimm SMPS with 241Am neutralizer and with X-ray neutralizer, respectively. In the 10 - 700 nm size interval the average paired absolute difference is

Figure 2. Aerosol size distribution for 241Am and X-ray neutralizers (charged and neutral particles).

Figure 3. Paired absolute difference percentage.

Page 4: Application of the Soft X-Ray TSI Advanced Aerosol ...file.scirp.org/pdf/ACS_2014093010362943.pdf · 5400 Vienna type) and a condensation particle counter ... compares well with the

A. Nicosia et al.

639

Table 1. Statistical parameters of the particle size distributions.

Median Geom. St. Dev. Total Conc.

NEUTRAL AEROSOL

241Am neutralizer 35.3 ± 0.2 nm 2.09 ± 0.01 (2.10 ± 0.06)·105 cm−3

X-ray TSI 3087 34.6 ± 0.2 nm 2.09 ± 0.01 (1.90 ± 0.02)·105 cm−3

Relative difference (respect to 241Am) (1.88 ± 0.01)% (0.400 ± 0.002)% (9.4 ± 0.3)%

CHARGED AEROSOL

241Am neutralizer 47.0 ± 0.5 nm 2.01 ± 0.02 (8.5 ± 0.2)·105 cm−3

X-ray TSI 3087 45.2 ± 0.2 nm 2.01 ± 0.01 (8.2 ± 0.2)·105 cm−3

Relative difference (respect to 241Am) (3.80 ± 0.04)% (0.166 ± 0.002)% (3.2 ± 0.1)%

INDOOR

241Am neutralizer 69 ± 6 nm 2.35 ± 0.06 (2.04 ± 0.18)·103 cm−3

X-ray TSI 3087 67 ± 6 nm 2.32 ± 0.06 (1.84 ± 0.12)·103 cm−3

Relative difference (respect to 241Am) (2.76 ± 0.37)% (0.92 ± 0.03)% (9.8 ± 1.1)%

about 9% for neutral aerosol and about 7% for charged aerosol. The 700 - 1090 nm size range was not consi-dered due to the low particle number concentration.

The mean statistical parameters were compared and are reported in Table 1. For both test aerosols, it was found that the peak of the distribution is slightly shifted backward for X-ray, and all statistical parameters are within 4%.

These results should be compared with the values provided by TSI in the X-ray neutralizer datasheet. Highly charged particles of different type, size and concentration were investigated, after passing through a unipolar charger upstream of the SMPS spectrometer. TSI reports that the geometric means and geometric standard devi-ations for the entire test matrix were within 5%, and the concentrations within 10% - 20%. TSI excluded that these differences were due to an incomplete charge neutralization of the X-ray device. The accuracy limit pro-vided in the datasheet is greater than the differences observed in the present study. To extend our comparison, we used the Grimm SMPS to measure the distribution of indoor air at 25% relative humidity, with both the X-ray and the 241Am neutralizer. Four series were taken through each device and the average particle size dis-tributions are illustrated in Figure 4.

Apart the 90 - 120 nm interval, the two particle size distributions are fully included inside the measurement variability. The statistical parameters, as well as the median and standard deviation show good agreement, while the total particle number concentration is 10% lower with X-ray than with 241Am.

Figure 5 shows the correlation of the average particle number concentration in each size bin (for indoor, un-charged and charged aerosol measurements) between the SMPS with 241Am or X-ray neutralizers.

The correlation coefficient turned out to be 0.998, while the one reported in the TSI datasheet was 0.994, and compares well with the measurements reported in the literature for instruments of different design. Watson et al., [3] reported an average correlation of 0.95 in the 5 - 200 nm size intervals while comparing four different SMPS spectrometers (TSI nano, TSI standard, Grimm SMPS + C and MSP WPS) at Fresno Supersite.

In addition, Wiedensohler et al. [4] reported a discrepancy of 10% in the 20 - 200 nm particle size range, and 30% for particles larger than 200 nm, under controlled laboratory conditions, by comparing distributions meas-ured with spectrometers of different design.

4. Conclusions The applicability of the X-ray neutralizer (TSI model 3087) in combination with a Grimm SMPS (L-DMA) has been evaluated under controlled laboratory conditions. The comparison between the aerosol size distributions obtained with the standard Grimm neutralizer (241Am radioactive source) and with the soft X-ray neutralizer shows:

Page 5: Application of the Soft X-Ray TSI Advanced Aerosol ...file.scirp.org/pdf/ACS_2014093010362943.pdf · 5400 Vienna type) and a condensation particle counter ... compares well with the

A. Nicosia et al.

640

Figure 4. Indoor aerosol size distribution for 241Am and X-ray neutralizers.

Figure 5. Particle concentration correlation between SMPS spectrometer data taken with the X-ray model 3087 TSI and the 241Am neutralizers.

a 0.998 correlation in the particle number concentration; differences in the distribution parameters (median and geometric standard deviation) below 4%; in the 10 - 700 nm size range the PADP is about 9% for neutral aerosol and 7% for charged aerosol.

In conclusion our results show the applicability of the soft X-ray neutralizer as a standalone device.

Acknowledgements This work was partially supported by the European Project “Smart Nano-structured Devices Hierarchically As-sembled by Mineralization Processes” (SMILEY, NMP.2012.1.4-2 FP7 SMALL-6-310637).

We are particularly grateful to Dr. Gianni Santachiara for his constructive comments and corrections.

References [1] Hoppel, W.A. and Frick, G.M. (1990) The Nonequilibrium Character of the Aerosol Charge. Distributions Produced by

Neutralizers. Aerosol Science and Technology, 12, 471-496. http://dx.doi.org/10.1080/02786829008959363

Page 6: Application of the Soft X-Ray TSI Advanced Aerosol ...file.scirp.org/pdf/ACS_2014093010362943.pdf · 5400 Vienna type) and a condensation particle counter ... compares well with the

A. Nicosia et al.

641

[2] Shimada, M., Han, B., Okuyama, K. and Otani, Y. (2002) Bipolar Charging of Aerosol Nanoparticles by a Soft X-Ray Photoionizer. Journal of Chemical Engineering of Japan, 35, 786-793. http://dx.doi.org/10.1252/jcej.35.786

[3] Watson, J.G., Chow, J.C., Sodeman, D.A., Lowenthal, D.H., Chang, O.M.C., Park, K. and Wang, X. (2011) Compari-son of four Scanning Mobility Particle Sizers at the Fresno Supersite. Particuology, 9, 204-209. http://dx.doi.org/10.1016/j.partic.2011.03.002

[4] Wiedensohler, A., Birmili, W., Nowak, A., Sonntag, A., Weinhold, K., Merkel, M., Wehner, B., Tuch, T., Pfeifer, S., Fiebig, M., Fjäraa, A.M., Asmi, E., Sellegri, K., Depuy, R., Venzac, H., Villani, P., Laj, P., Aalto, P., Ogren, J.A., Swietlicki, E., Williams, P., Roldin, P., Quincey, P., Hüglin, C., Fierz-Schmidhauser, R., Gysel, M., Weingartner, E., Riccobono, F., Santos, S., Grüning, C., Faloon, K., Beddows, D., Harrison, R., Monahan, C., Jennings, S.G., O’Dowd, C.D., Marinoni, A., Horn, H.G., Keck, L., Jiang, J., Scheckman, J., McMurry, P.H., Deng, Z., Zhao, C.S., Moerman, M., Henzing, B., de Leeuw, G., Löschau, G. and Bastian, S. (2012) Mobility Particle Size Spectrometers: Harmoniza-tion of Technical Standards and Data Structure to Facilitate High Quality Long-Term Observations of Atmospheric Particle Number Size Distributions. Atmospheric Measurement Techniques, 5, 657-685. http://dx.doi.org/10.5194/amt-5-657-2012

Page 7: Application of the Soft X-Ray TSI Advanced Aerosol ...file.scirp.org/pdf/ACS_2014093010362943.pdf · 5400 Vienna type) and a condensation particle counter ... compares well with the