2
Back-to-back calibration of accelerometers, using FFT analysis for sensitivity comparison at 800 frequencies illlf-Tnis: application note lie- ) scribes a baeii-to-back compare 11 !-!v. VA-■■■-■-■-■:■ ■■ w ■* ; >* •:-^-:o >',;■: o:o:-;^-:- Si ls:on-, method, losing the Dual- I Chanm! Analyzer Type 2032 -ou 112034j for accelerorneter calibralll tion simultaneously at 800 fre- ^uencies. By using broadband random noise excitation, . ^tntfei technique can be used to projl duce sensitivity and phase measurements witn an accura- cy similar to that produced by | traditional sine calibration is ^techniques, i iiiiiiiliiiiiiiiiii|i«ili|i The calibration system pre- Bented can be expanded to a ^computer- based systemi|ii tor &SSg: . - ■»:■;■■< . : :■-;-;■- '<:**■■■*< ' "x-'-^v-' ? semi-automatic transducer call! ibratiom suitable for large caliiii i bration laboratories. Kill H y The 2032- or 2034-based system for back-to-back calibration of accelerome- ters at 800 frequencies. Introduction Back-to-back calibration using FFT analysis The back-to-back calibration tech- The advent of the dual-channel an- tio of their outputs. By feeding the nique is based on the principle illus- alyzer (with built-in broadband ran- output from the reference accelerome- trated in Figure 1. An accelerometer, dom noise generator) enables relative- ter to the channel A input, and the the sensitivity of which is unknown, is ly fast frequency response and phase output from the unknown accelerome- mounted in a back-to-back manner measurement, at 800 frequencies. ter to the channel B input, the un- with a standard reference accelerome- Consequently, the sensitivities of two known sensitivity can be presented on ter, and the combination is mounted accelerometers can be compared by the analyzer's display screen as fre- on a suitable vibration source. The in- using a Dual-Channel Signal Analyzer quency response and phase response put acceleration to each accelerometer Type 2032 or 2034 to measure the ra- functions. is identical; and consequently, the ra- tio of their sensitivities is simply the ratio of their outputs. Traditionally, the accelerometers are excited at a single frequency, and their outputs are measured (after suitable preamplifica- tion) by using a high-quality electron- . ic voltmeter, the accuracy of which is known. This method produces good results. However, it produces a mea- sure of the sensitivity at a single fre- quency. Therefore, attaining a com- prehensive knowledge of an acceler- ometer's characteristics can be rather time consuming. Fig. 1. Principle of the traditional back-to-back calibration technique. \ BO 0237-11 1

Application notes - Back-to-back calibration of ... · Back-to-back calibration of acceler-ometers, using FFT analysis for sensi-Fig. 2. Principle of the back-to-back calibration

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Application notes - Back-to-back calibration of ... · Back-to-back calibration of acceler-ometers, using FFT analysis for sensi-Fig. 2. Principle of the back-to-back calibration

Back-to-back calibration of accelerometers, using FFT analysis for sensitivity comparison

at 800 frequencies —

illlf-Tnis: application note l ie- ) scribes a baeii-to-back compare 11

!-!v. VA-■■■-■-■-■:■ ■■ w ■ * ; >* • : - ^ - : o >',;■: o:o:-;^-:-Sils:on-, method, losing the Dual-I C h a n m ! Analyzer Type 2032 - o u 112034j for accelerorneter calibralll

tion simultaneously at 800 fre-^uencies. By using broadband random noise excitation,. ^tntfei technique can be used to p r o j l duce sensitivity and phase measurements witn an accura-cy similar to tha t produced by | traditional sine calibration is

^techniques, i iiiiiiiliiiiiiiiiii|i«ili|i

The calibration system pre-Bented can be expanded to a

^computer- based systemi|ii tor &SSg: . - ■»:■;■■< . : :■-;-;■- '<:**■■■*< ' " x - ' - ^ v - '

? semi-automatic transducer c a l l ! ibratiom suitable for large caliiii

i bration laboratories. K i l l H y The 2032- or 2034-based system for back-to-back calibration of accelerome­ters at 800 frequencies.

Introduction Back-to-back calibration using FFT analysis The back-to-back calibration tech- The advent of the dual-channel an- tio of their outputs. By feeding the

nique is based on the principle illus- alyzer (with built-in broadband ran- output from the reference accelerome-trated in Figure 1. An accelerometer, dom noise generator) enables relative- ter to the channel A input, and the the sensitivity of which is unknown, is ly fast frequency response and phase output from the unknown accelerome-mounted in a back-to-back manner measurement, at 800 frequencies. ter to the channel B input, the un-with a standard reference accelerome- Consequently, the sensitivities of two known sensitivity can be presented on ter, and the combination is mounted accelerometers can be compared by the analyzer's display screen as fre-on a suitable vibration source. The in- using a Dual-Channel Signal Analyzer quency response and phase response put acceleration to each accelerometer Type 2032 or 2034 to measure the ra- functions. is identical; and consequently, the ra­tio of their sensitivities is simply the ratio of their outputs. Traditionally, the accelerometers are excited at a single frequency, and their outputs are measured (after suitable preamplifica-tion) by using a high-quality electron- . ic voltmeter, the accuracy of which is known. This method produces good results. However, it produces a mea­sure of the sensitivity at a single fre­quency. Therefore, attaining a com­prehensive knowledge of an acceler-ometer's characteristics can be rather time consuming. Fig. 1. Principle of the traditional back-to-back calibration technique.

\

BO 0237-11 1

Page 2: Application notes - Back-to-back calibration of ... · Back-to-back calibration of acceler-ometers, using FFT analysis for sensi-Fig. 2. Principle of the back-to-back calibration

f'y~

A basic calibration system, centered around the Type 2032 or Type 2034, is presented in the opening figure. The system features the Vibration Exciter Type 4805 and Head Type 4815. This head is especially designed for calibra­tion work, and as such incorporates the Standard Reference Accelerome-ter Type 8305. An automated calibra­tion system providing hard copy of each accelerometer's sensitivity can be formed by adding a computer and a plotter to the instrumentation.

—'

Calibration by substitution In practice, it has proved advanta­

geous, with regard to accuracy, to em­ploy the FFT-based back-to-back cali­bration method, using a self-correct­ing calibration by substitution technique.

The calibration by substitution technique, which is based on the prin­ciple shown in Figure 2, involves mak­ing two measurements. The reference measurement, in which one reference accelerometer is calibrated against an­other, is stored before calibration of the unknown accelerometer is made. An accelerometer is used as a refer­ence, and remains fixed to the exciter head. The standard reference acceler­ometer and the accelerometer of un­known sensitivity are individually compared to the reference. The charge sensitivity of the unknown accelerom­eter is then calculated as follows:

SJf) = Sr(f) x | ^

The ratio Hu(f)/Hr(f) can be found directly from the equalized frequency response function. This is a post-pro­cessing function of the analyzer, which calculates the complex ratio between the measured and stored frequency re­sponse functions.

Advantages of using a random-noise based sys­tem

Back-to-back calibration of acceler-ometers, using FFT analysis for sensi-

Fig. 2. Principle of the back-to-back calibration method using the correction by substitution method.

tivity comparison, is especially suit- curate calibration results are attain-able for making frequency response able. checks against a standard reference accelerometer. If the Calibration Ex­citer Type 4290 is used, then acceler­ometer frequency responses up to C o n c l u s i o n s 26 kHz can be checked.

, . The Dual-Channel Signal Analyzers The accuracy of the presented FFT- Type 2032 and 2034 offer an alterna-

calibration technique, using the sub- tive approach to accelerometer cali-stitution method, is comparable to the bration. By using broadband random accuracy attained by dedicated com- n o iSe excitation with the FFT calibra-parison systems. The calibration by tion technique, accelerometers can be substitution method produces results easily calibrated at 800 frequencies, having uncertainties in the order of with an accuracy comparable to tradi-1% with 99% confidence limits, for tional methods. This provides the op-low to medium frequency measure- portunity for fast frequency response ments. checking against a reference acceler­

ometer, particularly when the analyzer If your calibration standard requires is used in a computer based system. .

the use of sine excitation, then the sine generator of the Type 2032 or Type 2034 can be used to make a point by point calibration. By making the cali­bration at frequencies corresponding to the actual analysis lines, and by using rectangular weighting, very ac-