27
APEX III Site Guide version August 2000 Page 1 of 27 Bruker Daltonics Support Center Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected] APEX TM III MASS SPECTROMETER SITE GUIDE AND PRE-INSTALLATION INSTRUCTIONS NOTE: Important Facilities and Installation Data Within Please Forward to Cognizant Engineering or Using Department All measurements in millimeters

APEX III Site Guide 08-00 Orig

Embed Size (px)

DESCRIPTION

Pre installation instructions for Apex III

Citation preview

APEX III Site Guide version August 2000

Page 1 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

APEXTM III

MASS SPECTROMETER

SITE GUIDE AND PRE-INSTALLATION INSTRUCTIONS

NOTE:

Important Facilities and Installation Data Within

Please Forward to Cognizant Engineering or Using Department

All measurements in millimeters

APEX III Site Guide version August 2000

Page 2 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

APEX III Site Guide version August 2000

Page 3 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

1.0.0 INTRODUCTION

It is advantageous for the user to make certain preliminary preparations prior to the arrival of the Bruker

field service engineer. These arrangements vary according to the support equipment purchased.

However, the basic installation is identical for all units.

The basic installation preparations include:

a. The selection of the location for the APEXTM III Mass Spectrometer.

b. The utility installation necessary for its operation.

c. Floor loading determination.

d. A method of handling the magnet during installation.

e. The purchase of the required cryogens (nitrogen and helium)

The following are recommendations for the preparation of your site for the installation of a

superconducting magnet FTMS spectrometer system. If any of these recommendations cannot be

achieved, please contact your local Bruker service office as soon as possible so that alternate plans can be

made prior to the beginning of the installation.

2.0.0 SELECTION OF THE LOCATION FOR THE MASS SPECTROMETER

2.1.0 SAFETY CONSIDERATIONS

Superconducting magnets can cause potential safety hazards due to their extended magnetic stray field,

their large attractive forces on ferromagnetic objects, and their large content of cryogenic liquids. It is the

sole responsibility of our customers to ensure safety in their FTMS laboratory and to comply with local

safety regulations. Bruker Daltonics Inc. (BDI) is not responsible for any injuries or damages due to an

improper room layout or due to improper operation routines.

2.1.1 MAGNETIC ATTRACTIVE FORCES

Strong attraction of ferromagnetic objects may occur at close distance to the magnet, where the magnetic

field is above 50 to 100 Gauss. The attractive force depends on the mass of the object and on the distance

from the magnet. It increases rapidly when the distance is reduced (by the 7th power). Hence, it may

change from barely noticeable to uncontrollable in a very short distance. It is therefore very important

that a safety space is established and marked around the magnet, and that all pathways in the FTMS room

stay clear of that area. Massive iron objects such as pressurized gas cylinders used for GCMS, QEA, etc.

are extremely dangerous in the vicinity of a superconducting magnet. They should be mounted very close

to the door and away from the magnet, or even better, outside the magnet room. In the minimum they

should remain outside the 5 gauss line as indicated by the field plots given in the Appendix in this

document. Inside the room with the magnet, a wall mounted gas distribution system is recommended.

APEX III Site Guide version August 2000

Page 4 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

2.1.2 MAGNETIC STRAY FIELD

It is generally accepted that stray fields are harmless below 3 to 5 Gauss (ten times the earth's magnetic

field). Stronger stray fields closer to the magnet may disturb some heart pacemakers, erase magnetic

cards and storage devices, and adversely affect watches and micro mechanical devices.

It is therefore recommended to mark the 5 Gauss line around the magnet, display warning signs and to

limit access to areas with more than 10 to 20 Gauss field to FTMS staff only. A safe location for

mechanical watches, wallets, and software disks at the entrance to the FTMS laboratory or in a separate

room is recommended for storage of these items by the FTMS staff. Be aware that the magnetic stray

field extends in all three dimensions and is not affected by walls, floors, or ceilings. For FTMS magnets

the horizontal extension is larger than the vertical one. High fields may also affect the rooms above and

below the magnet.

Local shielding of sensitive devices such as computer displays is possible in stray fields below 10 Gauss.

Global shielding of the entire magnet stray field is usually very costly. However, stray fields can be

greatly reduced by using a shielded magnet. Bruker strongly recommends the use of shielded magnets,

especially in multi-user installations and facilities where access to the magnet cannot be easily restricted.

Stray field plots and warning signs will be included with each magnet. Full stray field plots are also

included in this document and extra copies of field plots or warning signs are also available from our

customer service representative. A summary of the 5 and 10 gauss field lines for each system follows in

table 1 as well as a summary of the effects of magnetic stray fields contained in table 2.

Table 1: Stray field lines

Magnet z - Axis (Field

Axis)

z – Axis (Field

Axis)

x, y - Axis

x, y - Axis

5 Gauss 10 Gauss 5 Gauss 10 Gauss

APEXTM II-30

Passively Shielded

1.0 m 0.8 m ** **

APEXTM II-47

Actively Shielded

1.6 m 1.4 m 0.7m 0.8m

APEXTM II-70

Actively Shielded

2.0 m 1.7 m 1.1m 0.9m

APEXTM II-94

Actively Shielded

2.2m 2m 1.5m 1.4m

** Does not extend beyond magnet dewar.

APEX III Site Guide version August 2000

Page 5 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

Table 2: Summary of Magnetic Field Effects (level where damage can occur)

Effects on electron microscopes 1 Gauss

Disturbance of color computer displays* 1-3 Gauss

Disturbance of monochrome computer displays 3-5 Gauss

Erasing of credit cards and bank cards 10 Gauss

Effects on watches and micro mechanical devices 10 Gauss

Lowest known field effect on pace makers 1-7 Gauss

Erasing of magnetic tapes 20 Gauss

Saturation of transformers and amplifiers 50 Gauss

Erasing of floppy diskettes or magnetic tape 350 Gauss

*Disturbances of computer displays can be effectively reduced by shielding the display, as long as the

field is below 10 Gauss.

2.1.3 RECOMMENDATIONS ON SAFETY

These recommendations are adequate for industries and research institutes with limited public access.

The general public should not be allowed access to areas with fields above 5 Gauss. No inexperienced

persons (such as visitors, cleaning staff, electricians, plumbers, painters, etc.) should have access to the

magnet without supervision by the FTMS staff.

Table 3: Recommended Area Restrictions

Warning signs (safety limit for medical devices) 5 Gauss

Access limited to FTMS staff only (safety limit for heavy iron objects) 10 Gauss

Floor markings (safety limit for attraction of iron) 5 Gauss

Pressurized gas cylinders (wall mounted) 10 Gauss

2.1.4 CRYOGENIC LIQUIDS

Superconducting magnets use liquid Nitrogen and Helium as cooling agents. These liquids expand their

volume by a factor of 700 during evaporation and warm-up. The gases are nontoxic and completely

harmless as long as adequate ventilation is provided to avoid suffocation. During normal operation only

100 to 150 cubic feet per day of nitrogen are evaporated, but during a quench 1500 to 3500 cubic feet of

helium gas are produced within a short time. Windows and doors are sufficient for ventilation even after

a quench, but the magnet should never be in an airtight room. Room layout, room clearance and magnet

height should be such that an easy transfer of liquid nitrogen and helium is possible. This will reduce the

risks of accidents considerably.

APEX III Site Guide version August 2000

Page 6 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

2.2.0 OPERATIONAL CONSIDERATIONS

The ease or difficulty of day-to-day operation depends strongly on the room dimensions and layout.

Easy access to the magnet for refilling the cryogenic liquids and adequately located gas and power outlets

are important. Room dimensions should be sufficient to allow convenient pathways around the magnet.

The addition of auxiliary inlets and equipment such as GC, LC, ethernet connections, etc. should be

taken into account at this time also.

The minimum room height is given by the space needed for refilling helium. However some extra space

makes handling much easier. Special short helium transfer lines are available, but they do have higher

helium transfer losses. Large magnets require rooms with oversized doors to enable installation. Magnet

dewars are shipped with support fixtures (shipping bungs), which must be removed at the installation

site. This requires a suitable hoist with sufficient room height, which is generally more than the minimum

required for helium transfer. If this cannot be done in the lab, the magnet system may be moved after

assembling, however this will be done at the customers risk. There is no danger of damage to the magnet

or dewar, but the very tight tolerances of alignment may be disturbed by moving a dewar after assembly.

If so, it only shows up during the cool down or after energization. This requires the warming up and

realignment of the dewar with a corresponding waste of much time and helium.

The vacuum cart and electronics consoles require sufficient clearance around them to provide for

adequate cooling. This space also allows for installation and service of the system which includes

convenient changing of the optional sources, changing pump oils and general system maintenance.

Please refer to Chapter 5.1.1 for system specific recommended room layouts.

2.2.1 FLOOR LOADING

The FTMS magnet system is quite heavy. Depending on field strength, bore diameter and specially the

shielding, the weight may be 1,500 to 13,200 kg. The floor of the room must be strong enough to support

the weight of the instrument allowing for a certain amount of mild vibration. In general, for field

strengths of < 7.0 Tesla, the dead weight floor loading for the magnet is approximately 0.40 kg/cm2 for

an unshielded magnet and 1.4 kg/cm2 for a shielded magnet. Please refer to Chapter 5.1.1 for the exact

footprint of each instrument model.

2.2.2 ELECTRIC POWER

An important consideration when selecting the area for the Mass Spectrometer is availability of adequate

power. The magnet system needs electric power only during charging and discharging and for the

operation of the optional helium recovery system. Regardless of what accessory equipment is purchased

for a particular installation, sufficient power should be brought into the area to provide for future needs.

There are recommendations for power distribution located in Chapter 5.1.1, along with the floor layout.

A complete APEXTM III Mass Spectrometer requires up to 20 KVA (please refer to Chapter 5.1.1 for a

explanation of the power consumptions of the various instrument configurations.) Bruker recommends

APEX III Site Guide version August 2000

Page 7 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

that one 3 phase five wire 380 V AC phase to phase, 30 Amp, 50 or 60 Hertz circuit (220 Volt to neutral)

be available before the instrument arrives. Power distribution within the mass spectrometer consists of

three 220V, 30A circuits-split off from the three phase 220 volt to neutral circuit

If the only available power is 208 V AC (common in the United States) then a boost transformer

(optionally available) must be installed to raise the output to 380 V AC. If this option is to be used, the

transformer should be installed prior to the delivery of the instrument. For additional information on

using the boost transformer please contact your sales/service office.

If the line voltage fluctuations exceed +10% or -5%, a voltage regulator must be installed.

A quick disconnect, in addition to the emergency kill switch on the system, may be required by safety

codes in many areas.

The customer is responsible for appropriate connection of the instrument to the outlets. The instrument is

delivered with a long enough power cable without connector.

2.2.3 WATER

The ultra high vacuum pumps in the mass spectrometer are water-cooled and require a continuous stream

of water. Turbo-pumped systems require 150 Liter/hour of filtered, soft water that is not to exceed 20C.

Cryopump systems require 200 Liter/hour of filtered, soft water that is not to exceed 20C. If filtered soft

"house" water is used, a drain for the water is necessary. A constant pressure regulator on the water line

is recommended if the water pressure is likely to change a large amount over a 24-hour period.

Many communities prohibit the use of "house" water for cooling and require a closed loop heat exchange

system. Bruker, due to environmental considerations, encourages the use of closed loop water chiller

systems. Turbo pump systems require water chillers with a 500 W cooling capacity and 40psi supply

pressure while cryo pump systems require water chiller with a 3 KW (or better) cooling capacity. Please

note that cooling capacity for heat exchangers may depend on the room temperature. These specifications

should be examined closely and possibly forwarded to your heat exchanger vendor, for a

recommendation based on the proper requirements.

2.2.4 VENTILATION AND AIR CONDITIONING

For most efficient operation, the mass spectrometer must be in an air-conditioned room where both

temperature and humidity are controlled. It is important to note, a complete operating mass spectrometer

will produce between 15 and 20,000 BTU/hr exclusive of other sources of heat. Room temperature must

not exceed 72deg. F and should be maintained within ±10deg F for best performance of the FTMS

spectrometer. Relative humidity should remain in the 40 to 60% range. If the air in the lab gets too damp

high voltage arcing can be a problem as well as excessive condensation on the cryostats and cooling

water lines. If it is too dry there are problems with static electricity. Ventilation should be adequate to

avoid depletion of oxygen. Constant air temperature and pressure is very important for high performance

operation. Rapid ambient pressure and temperature fluctuations reduce the performance of the FTMS

spectrometer. If an air conditioning unit is installed, it should be far away from the magnet. The air in the

vicinity of the magnet dewar should not be disturbed by drafts from an inlet or an outlet of the ventilation

APEX III Site Guide version August 2000

Page 8 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

system. If the optional water chiller is used as the source of cooling water for the turbo or cryopumps, the

heat load due to these units must be taken into account when calculating the air conditioning capacity.

2.2.5 GASES

Dry, oil and dust free, clean regulated pressurized air at 60 to 100 psi with a dew point of -40deg. F or

lower is required for operation of the pneumatic valves on the vacuum system. For refilling the magnet

dewar with cryogenic liquids both pressurized nitrogen and helium gas are also required. The

electrospray ion source on the APEXTM III requires a source of dry nitrogen 30-100 psi and 2 scfm and

dry oxygen (for negative ions) at 30-100 psi and 2 scfm. If no fixed gas supplies are available for refilling

or the electrospray source, transportable cylinders may be used. However, if ferro-magnetic cylinders are

used in the presence of strong magnetic fields, they may present a possible source of serious accidents.

Non-magnetic liquid nitrogen cylinders with a pressure build-up system are convenient sources of dry

nitrogen gas. Gas requirements for the optional GC inlet depend on the actual experiment planned.

MS/MS experiments require a small cylinder of argon or another suitable collision gas.

2.2.6 LIQUID NITROGEN AND HELIUM

A constant and dependable source of the liquid cryogens is essential to maintaining the magnetic field.

Therefore, it will be necessary to arrange for the delivery of liquid nitrogen and helium if the facility

does not have it readily available. These arrangements may be made through a local supplier of liquefied

gases.

Typical requirements for installation are:

approx. 300 Liters (for 3 Tesla) to 1400 Liters (for the 9.4Tesla) of liquid Helium

approx. 500 Liters (for 3 Tesla) to 1500 Liters (for the 9.4Tesla) of liquid Nitrogen

Typical requirements for routine fills are:

approx. 100 L of liquid He for refills every approx. 90 days

approx. 75 L of liquid N2 for refills once per week

Please refer to the individual Magnet specifications for exact numbers.

2.2.7 RADIO FREQUENCY INTERFERENCE

RF interference can cause spikes in the spectra at any m/z ratio whose resonance frequency is close to the

frequency of the interference. The most common source of RF interference is TV and AM radio stations.

They broadcast at frequencies in the range of interest for FTMS (10 MHz to 10 KHz). Other possible

sources can be paging systems, ham radio transmitters, arc welders, and other RF spectrometers. Due to

the relationship of transmission power and distance radiated and since a common mode rejection of 80-

APEX III Site Guide version August 2000

Page 9 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

100 dB is attainable in the receiver chain, these interferents are rarely a major problem RF interferents

are often introduced on a poor power grounding coming into the FTMS laboratory.

2.2.8 TELEPHONE AND ETHERNET

It is highly recommended that a telephone and ethernet be installed at the console site. It should be

accessible to the operator of the instrument while seated at the console. This would be used for

troubleshooting assistance as well as applications support. In addition, APEXTM III spectrometers with

the UNIX or Windows NT based acquisition computer use an ethernet connection for remote diagnostics

and remote software upgrades or correction of corrupt data, an ethernet line within reach of the computer

is required for this function. For facilities where outside ethernet or telephone connection to computers is

not permitted for security reasons, an outgoing only telephone line and a voice connection are

recommended for remote diagnosis

2.2.9 RECOMMENDATIONS FOR ROOM LAYOUT

For safety reasons, the magnet should be located away from doors and in a location that the stray field of

10 to 20 Gauss does not extend beyond the room. Refilling of cryogenic liquids is more comfortable if

the magnet system is easily accessible for the transport dewars and if adequately located gas outlets are

available. The rate of nitrogen boil-off from the system depends on the dewar surface temperature; hence

the dewar should not be located close to a heat source or in direct sunlight. The Computer workstation

and the electronics console should be placed at sufficient distance from the magnet and oriented so that

the display monitor, disk and other magnetic storage devices are on the far end from the magnet.

Normally the magnet is located on the right side of the console. If a left hand system is required, this

should be specified at the time of order. An installation and access space to the rear of the console and

the right side of the vacuum system is recommended with a minimum clearance of 2 feet to the walls.

Recommended room layouts are provided at the end of this guide.

2.2.10 OTHER CONSIDERATIONS

In addition to the above, the following should be considered.

• Dust and corrosive gases or vapors are to be avoided if possible. The same consideration should be

given to the mass spectrometer as would be given to any precision electrical equipment containing

relays, meters, insulators, etc. The instrument presents no problems in this connection which are not

encountered in other electrical and electronic equipment.

• Most biological samples require refrigeration for long and short-term storage. It is recommended that

a freezer be located near the instrument for storage of biological samples and calibration compounds.

However, when placing the refrigerator in the same vicinity as the magnet system, remember to

maintain a distance from the 5 gauss stray field so the compressor of the refrigerator is not strained.

• Sufficient space for sample preparation, short-term storage of LC solvents, and storage of optional

ion sources and inlets, spare parts and tools is desirable for day-to-day operation.

APEX III Site Guide version August 2000

Page 10 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

APEX III Site Guide version August 2000

Page 11 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

3.0.0 PERFORMANCE CONSIDERATIONS

Almost any site is suitable for routine FTMS operations and good performance may be achieved even

with moderate environmental conditions of the FTMS room and its close vicinity. In general, higher

performance is possible with FTMS in a given set of environmental conditions than is possible with other

high resolution MS systems. Even so, ultra-high performance operation requires a very stable and

constant environment. Top performance will only be achieved if you select an adequate magnet location.

It is therefore recommended for large FTMS laboratories to plan ahead and influence the building

construction in the planning phase. It should be considered that the trend will always be towards larger

magnetic fields with inherently higher demands on the conditions of the installation site in order to

maintain and optimal operation.

3.1.0 STATIC IRON DISTRIBUTION

Static iron distribution in the vicinity of the magnet will exert a force on the magnet and disturb the field

homogeneity in the FTMS cell. The force may disturb the very tight tolerances of alignment, if so, it

shows up during the energizing of the magnet. In that case the magnet must be moved away from the

iron. An increase in the distance of 25% is usually sufficient. Extended areas of iron with near constant

density distributions generate rather harmless field disturbances. Their effects can easily be removed with

the superconducting X, Y, and Z shims that are standard on Bruker systems. Concentrated iron objects,

however, can produce rather strong local field variations, which are difficult to correct for.

Since these may be corrected by shimming during installation, it is recommended that the location of any

large iron objects such as gas cylinders be decided before installation so that they may be present at the

installation. This aspect becomes more important for high field and wide bore magnets with extended

stray fields.

If the magnet must be located in the vicinity of iron or steel support beams, a proper alignment is

important. Horizontal iron concentrations at heights close to the magnetic center and iron objects

terminating near the magnet should be avoided. They lead to very asymmetric field disturbances.

Since these are, in most cases, heating or plumbing devices rather than structural support beams, they

may need to be removed before installation. Section 3.3 gives some examples of minimum distances,

which should be considered for magnet positioning.

3.2.0 INTERFERENCE FROM VARIABLE MAGNETIC FIELDS

Variable field effects are very difficult to eliminate. Therefore they should be considered early in the

planning stage. Variations can originate from moving iron objects such as metal chairs, doors, handcarts,

elevators, and cars. They also can be caused by electromagnetic equipment, from power lines carrying

DC currents or from floor vibrations or oscillations of the entire building. The most disturbing

fluctuations are step-like field changes produced by large switching DC currents from subways, trolley

cars, etc. or from field ramping mass spectrometers during fly-back of the field. If you have any of these

near your site contact Bruker for a magnetic interference survey of the lab as early in the planning stage

APEX III Site Guide version August 2000

Page 12 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

as possible. In general, mass calibration of the FTMS spectrometer is not affected if the variation of a

moving iron object occurs outside of the 5 gauss line for the magnet used in the system.

3.3.0 SUMMARY OF MAGNETIC FIELD DISTURBANCES:

Table 4: Stray field or distance from the magnet at which effects may occur

Force from large iron objects of over 2000 lbs. 20 Gauss

Force from iron objects over 400 Lbs. 50 Gauss

Field from localized static iron of approx. 200 Lbs. 50 Gauss

Field from moving iron objects of approx. 200 Lbs 3 Gauss

Metal chairs and doors 6 to 10 Feet

Hand-carts and heavy metal doors 10 to 15 Feet

Large moving objects (elevators and cars) 20 to 40 Feet

Maximum floor vibrations below 1 Hz Approx. 0.1 mm

Table 5: Stray field or distance from the magnet which may lead to noticeable

instabilities

Minimum field fluctuations (steps) 1 mGauss

Maximum ac fields (5 to 10 Hz range) 5 to 10 mGauss

Sweeping fields from mass spectrometers (with slow

ramping up and no fly-back)

25 to 30 feet

Large switching DC currents (subway or trolley) 500 to 1500 feet

3.4.0 RECOMMENDATIONS FOR ROOM SELECTION

We recommend locating the APEXTM III system at sufficient distance from large moving objects such as

cars and elevators. Optimal locations are far away from internal and external traffic areas. For installation

sites of high field magnets (7.0 and 9.4T) the rooms above and below the magnet must also be rather

quiescent. We do not recommend locations in high rise buildings above the 5th floor because of

vibration, sway, and atmospheric pressure fluctuations. The optimum site is in the basement or ground

floor directly on the concrete slab. We also advise against locating high field systems in buildings near

subway or train lines because of the high magnetic field instabilities caused by the passing of trains and

DC currents.

Floor support should be rigid in order to avoid vibrations. Windows should be located in such a way that

sunlight does not shine directly on the dewar, and constructed such that no sudden pressure fluctuations

are produced by winds. For optimal performance the room should be air conditioned with constant

temperature control.

APEX III Site Guide version August 2000

Page 13 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

Bruker Daltonics Inc. can provide site information and site inspections for your specific lab.

Please contact your sales or service representative for more information and help in preparing the best

possible site for your instrument.

4.0.0 MASS SPECTROMETER INSTALLATION

4.1.0 RECEIVING

Purchasers are notified at least seven days in advance of the proposed arrival date of the instrument, so

that arrangements may be made to move it to the mass spectrometer room. In advance of shipment,

contact the local Bruker office with any questions concerning installation of the instrument. It is assumed

by Bruker that the instrument can be unloaded on a loading dock and pallet jacks are available at the

customer site. If there is no standard height loading dock, Bruker has to be notified prior to shipment to

make necessary arrangements. The console, vacuum system, magnet, and accessories will be separately

crated allowing the use of a forklift truck to move the system to its final location. The customer has to

make sure all hallways and doors from the loading area to the final position of the instrument are wide

and high enough to move the magnet and vacuum cart through. The vacuum cart is 110cm (44”) long and

127cm (50”) wide, so a minimum door with of 130cm (51”) is required on all doors. Some passively

shielded magnets may require even wider doors; please refer to the magnet guide for dimensions. The

minimum door height for most magnets is 7 feet (2.1m).

Shipping crate dimensions: Vacuum cart ~1200lbs (140 x 175 x 175cm)

Console: ~400lbs (85 x 95 x 140cm for U.S. or

155 x 95 x 160cm for international freight)

Accessories: 300 .. 500lbs (155 x 95 x 90cm) one or two boxes

Magnet: refer to Magnet Guide

Upon request, a Bruker representative will visit the customer site for an inspection.

APEX III Site Guide version August 2000

Page 14 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

4.2.0 GENERAL INSTALLATION PROCEDURE

It is advisable to wait for arrival of the Bruker representative before any equipment is removed from the

shipping crates. However, the crates may be placed near the system’s final position. The utilities and

cryogens should have been obtained. No attempt should be made to assemble or cool the magnet prior to

the arrival of the field service engineer.

It is advantageous to have at least one scientist or engineer who can spend most of his/her time working

with the Bruker field service engineering during the installation. He/she will acquire valuable experience

during this period as well as reduce the time of installation.

During the installation, the personnel who are to be responsible for the operation and maintenance of the

instrument should also be familiarizing themselves with the data system, circuits, manuals and as much

of the instrument as possible.

4.2.1 LIFTING EQUIPMENT

Assembly of the magnet system may be required. In all cases the facility must supply the necessary

equipment to remove the magnet and vacuum cart from their shipping crates. The magnet system may

require assembly onto a stand requiring lifting equipment, which can raise the magnet to the minimum

ceiling height for your magnet. Other magnets (shielded Bruker-Magnex) may require assembly of the

shielding. In this case the shields will have to be raised into position alongside the magnet. This usually

requires a mobile hoist and the proper slings.

For specific requirements for your system, please contact your local service office.

APEX III Site Guide version August 2000

Page 15 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

4.2.2 CRYOGENIC LIQUIDS

The customer is responsible for up to the amounts of cryogen in the table in the following table. Total

volume used may be less. Bruker offers a liquid helium plan, which provides all of the liquid helium

needed for the cool down for a fixed price. Please contact your local sales representative or the magnet

department for details and availability.

Table 6: Cryogen Requirements

Magnet LHe for

installation

(liters)

LHe hold

time

(days)

LHe refill

volume

(liters)

N2 initial

cooldown

(liters)

N2 hold

time

(days)

N2 refill

volume

(liters)

Bruker/Magnex

3 T 160mm

300

>140

135

500

>14

135

Bruker/Magnex

4.7T 160mm

Actively Shielded

500

>90

115

700

>14

145

Bruker/Magnex

7T 160mm

Actively Shielded

800

>150

200

800

>14

190

Bruker/Magnex

9.4 T 160mm

Actively shielded

1400

>100

200

1400

>14

190

Note: All liquid nitrogen must be in low-pressure liquid withdrawal dewars (LX type). Converted

gas packs cannot be used.

4.3.0 COOL DOWN EQUIPMENT

The following items are also needed for the cool down of the magnet and must be supplied by the

customer:

• One T cylinder of (industrial grade) nitrogen gas.

• One T cylinder of (industrial grade) helium gas.

• One low pressure gas regulator for inert gas (0-25 psi)

• Twenty feet of 3/8 inch Tygon tubing.

APEX III Site Guide version August 2000

Page 16 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

• One Heat gun for de-icing and drying.

APEX III Site Guide version August 2000

Page 17 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

5.0.0 Appendix

5.1.0 PRE-INSTALLATION CHECKLIST

APEXTM

II MASS SPECTROMETERS

Space Requirements Installation

___Clearance around instrument ___Receiving

___Auxiliary equipment ___Handling the magnet

___Floor loading ___Personnel Required

___Ways from loading dock to Instrument location ___Electrician

Utilities Environmental

___Filtered Soft Water ___Air conditioning

___Electrical power (380V 3-phase) ___Electrical interference

___Liquid nitrogen and helium ___Magnetic Interference

___Gas and Air ___Vibration

___Freezer for sample storage ___Dust and Corrosion

_____________________ _______________________________

Customer Signature / Date Bruker Representative Signature / Date

APEX III Site Guide version August 2000

Page 18 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

5.1.1 RECOMMENDED SITE LAYOUT for the APEXTM III

The following figure shows the minimal room requirements for an APEXTM III FTMS system:

All dimensions in mm

*: Recommended place for all utility outlets (power, water, gases)

For dimensions of other magnets please refer to the individual magnet specifications.

Power Requirements:

Three phase 380/220 V AC to neutral (+10%, -5%), 30 A, with ground (vacuum system and consoles)

Single phase 110 or 220 V AC, 30/15A, (optional closed loop water chiller / LC / GC).

Single phase 110 or 220 V AC, 20/10 A, quad outlet for computer and accessories

Gas Requirements

Dry, filtered dust and oil free; dew point below -40deg. F; 60-100 psi regulated. Drying gas for the ESI

source and Air for the pneumatic valves. Connected to a ¼” tubing each.

Cooling Water

Closed loop water chiller with at least 500W cooling capacity and 40psi pressure or 75 Liter/hour of

filtered, soft house water that is not to exceed 20C.

APEX III Site Guide version August 2000

Page 19 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

Environmental

Room temperature not to exceed 72deg. F.

Total heat output of the system is around 4 kW (15,000-20,000 BTU/hr), depending on options

minimum ceiling height: Passively shielded magnets: 3085mm, Actively shielded magnets: 3130mm ,

9.4T AS magnet: 3300mm

Magnet must be located as to be accessible for filling with cryogens.

APEX III Site Guide version August 2000

Page 20 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

5.1.2 Magnet Stray Fields

3T Passively Shielded Magnet Stray Field

APEX III Site Guide version August 2000

Page 21 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

4.7 T Actively Shielded Magnet Stray Field

APEX III Site Guide version August 2000

Page 22 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

APEX III Site Guide version August 2000

Page 23 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

APEX III Site Guide version August 2000

Page 24 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

7 T Actively Shielded Magnet Stray Field

APEX III Site Guide version August 2000

Page 25 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

APEX III Site Guide version August 2000

Page 26 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]

9.4 T Actively Shielded Magnet Stray Field

APEX III Site Guide version August 2000

Page 27 of 27 Bruker Daltonics

Support Center

Tel.: (978) 663-3660 x 1445 -- Fax.: (978) 663 2962 - e-mail : [email protected]