Upload
others
View
3
Download
4
Embed Size (px)
Citation preview
ANSI N42.48 D4
N42.48
American National Standard Performance Requirements for
Spectroscopic Personal Radiation Detectors (SPRDs) for Homeland
Security
Accredited by the American National Standards Institute
Sponsored by the
National Committee on Radiation Instrumentation, N42
Published by
The Institute of Electrical and Electronics Engineers, Inc.
3 Park Avenue, New York, NY 10016-5997, USA
draft January, 2007
ANSI N42.48
American National Standard Performance Requirements for Spectroscopic
Personal Radiation Detectors (SPRDs) for Homeland Security
Sponsor
National Committee on Radiation Instrumentation, N42
Accredited by the
American National Standards Institute
Secretariat
The Institute of Electrical and Electronics Engineers, Inc.
Approved …..
American National Standards Institute
Abstract:
Keywords: alarm, design criteria, exposure rate, performance specifications, pocket-sized,
personal radiation detectors, radiation, radiation detection, radiation instrumentation,
spectroscopic
TERMS OF USE
Please Read Carefully
The Institute of Electrical and Electronics Engineers, Inc., (“IEEE”) owns the copyright to this
Document in all forms of media. Copyright of the text retrieved, displayed, or output from this
Document is owned by IEEE and is protected by the copyright laws of the United States and by
international treaties. The IEEE reserves all rights not expressly granted to the user.
Except as allowed by the copyright laws of the United States of America, applicable
international treaties, or any license or agreement the user or the user’s employer may have
entered into with the IEEE or one of its authorized distributors, the user may not copy and/or
distribute copies of this Document, nor significant portions of the document, in any form, nor
may the user post this document on any intranet, website, or database without prior written
permission from the IEEE.
Permission and/or requests for licenses should be directed to:
Manager, Standards Licensing and
Contracts
+1-732-562-3804
or in writing at:
IEEE Standards Association
Manager, Standards Licensing and Contracts
445 Hoes Lane
Piscataway, NJ 08854
The Institute of Electrical and Electronics
Engineers, Inc.
3 Park Avenue, New York, NY 10016-5997,
USA
Copyright © 2004 by the Institute of Electrical and Electronic Engineers, Inc.
All rights reserved. Published 30 January 2004. Printed in the United States of America.
Print:
PDF:
No part of this publication may be reproduced in any form, in an electronic retrieval system or
otherwise, without the prior written permission from the publisher.
American National Standard
An American National Standard implies a consensus of those substantially concerned with its
scope and provisions. An American National Standard is intended as a guide to aid the
manufacturer, the consumer, and the general public. The existence of an American National
Standard does not in any respect preclude anyone, whether he has approved the standard or not,
from manufacturing, marketing, purchasing, or using products, processes, or procedures not
conforming to the standard. American National Standards are subject to periodic reviews and
users are cautioned to obtain the latest editions.
CAUTION NOTICE: This American National Standard may be revised or withdrawn at any
time. The procedures of the American National Standards Institute require that action be taken to
affirm, revise, or withdraw this standard no later than five years from the date of publication.
Purchasers of American National Standards may receive current information on all standards by
calling or writing the American National Standards Institute.
Authorization to photocopy portions of any individual standard for internal or personal use is
granted by the Institute of Electrical and Electronics Engineers, Inc., provided that the
appropriate fee is paid to Copyright Clearance Center. To arrange for payment of licensing fee,
please contact Copyright Clearance Center, Customer Service, 222 Rosewood Drive, Danvers,
MA 01923 USA; +1-978-750-8400. Permission to photocopy portions of any individual standard
for educational classroom use can also be obtained through the Copyright Clearance Center.
Copyright © 2004 IEEE. All rights reserved. iv
Introduction
(This introduction is not part of ANSI N42.48, American National Standard Performance
Requirements for Spectroscopic Personal Radiation Detectors (SPRDs) for Homeland Security.)
This standard is the responsibility of the Accredited American Standards Committee on
Radiation Instrumentation, N42. The standard was approved on N42 letter ballot of July–August
2003.
Participants
At the time it approved this standard, the Accredited Standards Committee on Radiation
Instrumentation, N42, had the following members:
Michael P. Unterweger, Chair
Louis Costrell, Vice Chair
Sue Vogel, Administrative Secretary
Organization Represented ..........................................................................Name of Representative
American Conference of Governmental Industrial Hygienists.............................. Jesse Lieberman
Bartlett Services ........................................................................................................... Morgan Cox
Canberra.......................................................................................................................Larry Darken
Chew, M.H. ............................................................................................................... Jack M. Selby
Commerce Dept, U.S., NIST ...................................................................... Michael P. Unterweger
........... ...............................................................................................................Louis Costrell (Alt.)
Consultant ................................................................................................................Frank X. Masse
Entergy-ANO .............................................................................................................Ron Schwartz
Femo-TECH Inc. ..................................................................................................... Richard Straub
Gamma-Metrics .................................................................................................... Ernesto A. Corte
General Activities Inc. .................................................................................................. Karl Reinitz
Health Physics Society ..............................................................................................Joseph Stencel
Institute of Electrical & Electronics Engineers, Inc. ................................................. Louis Costrell
........... ................................................................................................................Julian Forster (Alt.)
......................................................................................................................Anthony Spurgin (Alt.)
............................................................................................................ Michael P. Unterweger (Alt.)
Lawrence Berkeley National Laboratory............................................................. Edward J. Lampo
Lawrence Livermore National Laboratory ................................................................ Gary Johnson
NASA, GFSC...............................................................................................H. Sachidananda Babu
Nuclear Standards Unlimited ................................................................................ Al N. Tschaeche
Oak Ridge National Laboratory ...........................................................................Peter J. Chiaro, Jr
................................................................................................................... .Charles L. Britton (Alt.)
Ortec Corp. ..........................................................................................................Ronald M. Keyser
Overhoff Technology Corp......................................................................................Mario Overhoff
Pacific Northwest National Laboratory……...........................................................Richard Kouzes
Swinth Associates ...............................................................................................Kenneth L. Swinth
Tennessee, University of...................................................................................... William M. Bugg
Thermo-electron...................................................................................................Richard P. Oxford
Translucent..............................................................................................................G. Laurie Miller
Copyright © 2004 IEEE. All rights reserved. v
U. S. Army.........................................................................................................Edward Groeber
U. S. Nuclear Regulatory Commission……………………………………….Cynthia G. Jones
Members-At-Large ........................................................................................ Joseph G. Bellian
........... ..............................................................................................................Edward Fairstein
............................................................................................................................. Paul L. Phelps
............................................................................................................................. Lee J. Wagner
At the time this standard was approved, Subcommittee N42.HSI had the following members:
Morgan Cox, Co-Chair
Jack M. Selby, Co-Chair
Dru Carson
Peter J. Chairo, Jr.
Jack Cooley
Leo Faust
Edward Groeber
Jerry Hiatt
Mark M. Hoover
Ron Keyser
C. McDonald
Robert Murphy
Cheryl Olson
Mario Overhoff
Scott Rogers
Michael P.
Unterweger
Ed Walker
Chuan-Fu Wu
At the time this standard was approved, the ANSI 42.48 Working Group had the following
members:
Peter Chiaro Jr., Chair and
Project Leader
Rolf Arlt Richard Oxford June Wang
David Brown Leticia Pibida
Kevin Carmichael Francis Schulcz
Robert Corsetti Ayman Shourbaji
Andrey Gueorguiev Rick Smith
Cynthia Jones Keith Spero
Stephen Lancaster Dave Trombino
Mike McGuire Michael Unterweger
Copyright © 2004 IEEE. All rights reserved. vi
Contents
1 Overview ................................................................................................................................. 2
1.1 Scope .............................................................................................................................. 2
1.2 Purpose ........................................................................................................................... 2
2 References ............................................................................................................................... 2
3 Definitions ............................................................................................................................... 3
4 General considerations ............................................................................................................ 8
4.1 Standard test conditions.................................................................................................. 8
4.2 Units and uncertainties ................................................................................................... 8
4.3 Special word usage......................................................................................................... 8
5 General requirements .............................................................................................................. 9
5.1 Controls .......................................................................................................................... 9
5.2 Documentation Check .................................................................................................. 10
5.3 Displays ........................................................................................................................ 10
5.4 Effective range of measurement or indication ............................................................. 10
5.5 Audible Alarms ............................................................................................................ 10
5.6 Vibration Alarm ........................................................................................................... 11
5.7 Size ............................................................................................................................... 12
5.8 Mass.............................................................................................................................. 12
5.9 Reference point marking .............................................................................................. 12
5.10 Explosive atmospheres ................................................................................................. 13
5.11 Batteries and Battery Lifetime ..................................................................................... 13
5.12 Data format and communication interface ................................................................... 13
5.13 User interface ............................................................................................................... 14
5.14 Spectral identification................................................................................................... 15
6 Radiological Tests ................................................................................................................. 16
6.1 General test information............................................................................................... 16
6.2 Rate of false alarms ...................................................................................................... 16
6.3 Time to alarm; photons................................................................................................. 17
6.4 Time to alarm; neutrons (if provided) .......................................................................... 17
6.5 Detection of gradually increasing radiation levels ....................................................... 18
6.6 Accuracy....................................................................................................................... 18
6.7 Personal radiation alarm............................................................................................... 18
6.8 Over-range reading....................................................................................................... 19
6.9 Interfering Ionizing Radiation ...................................................................................... 19
6.10 Radionuclide identification .......................................................................................... 19
7 Environmental Performance Requirements........................................................................... 22
7.1 General test practices.................................................................................................... 22
7.2 Temperature.................................................................................................................. 22
7.3 Temperature Shock....................................................................................................... 23
7.4 Humidity....................................................................................................................... 23
7.5 Moisture and dust protection........................................................................................ 24
7.6 Cold temperature start up ............................................................................................. 25
8 Electromagnetic Performance Requirements ........................................................................ 25
8.1 General test practices.................................................................................................... 25
8.2 Electrostatic Discharge (ESD)...................................................................................... 26
8.3 Radio frequency ........................................................................................................... 27
8.4 Magnetic fields ............................................................................................................. 27
Copyright © 2004 IEEE. All rights reserved. vii
8.5 Radiated emissions ....................................................................................................... 27
9 Mechanical Performance Requirements................................................................................ 28
9.1 Vibration....................................................................................................................... 28
9.2 Drop Test ...................................................................................................................... 28
9.3 Impact (Microphonics) ................................................................................................. 29
10 Documentation ...................................................................................................................... 29
10.1 Type test report............................................................................................................. 29
10.2 Certificate ..................................................................................................................... 29
10.3 Operation and maintenance manuals............................................................................ 30
Annex A (informative) Bibliography............................................................................................ 31
Annex B (informative) Detector tests ........................................................................................... 38
Annex C (informative) Sample user interface evaluation technique ............................................ 39
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 2
American National Standard Performance Requirements for
Spectroscopic Personal Radiation Detectors (SPRDs) for
Homeland Security
1 Overview
1.1 Scope
This standard describes design and performance requirements along with testing methods for evaluating
radiation detection instruments that are pocket-sized and worn on the body for the purpose of rapid
detection and identification of radioactive materials.
The performance requirements contained in this standard are meant to provide a means for verifying the
capability of these instruments to reliably detect changes above background levels of radiation, notify the
user to these changes, and provide a means to determine if the alarm was caused by a radionuclide of
interest that may require further evaluation.
These devices are not primarily intended to provide a measurement of dose equivalent rate. However,
their indication can provide an approximate value of exposure rate that should be reasonably accurate.
They are also not meant to provide radionuclide identification at the same level as radionuclide
identification detectors defined in ANSI N42.34.
Successful completion of the tests described in this standard should not be construed as an ability to
identify all radionuclides in all environments.
1.2 Purpose
The purpose of this standard is to specify performance criteria and test methods used to evaluate radiation
detection and radionuclide identification instruments that are pocket-sized and may be worn on the body.
2 References
This standard shall be used in conjunction with the following publications.
[R1] ANSI N42.22-1995 (R2002), American National Standard—Traceability of Radioactive
Sources to the National Institute of Standards and Technology (NIST) and Associated Instrument
Quality Control.2
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 3
[R2] ANSI N42.23-1996, American National Standard Measurement and Associated
Instrumentation Quality Assurance for Radioassay Laboratories.
[R3] International Atomic Energy Agency (IAEA) Safety Guide No. RS-G-1.9, Categorization
of Radioactive Sources, 2005
[R4] IAEA, Code of Conduct on the Safety and Security of Radioactive Sources, 2004
[R5] IEC 61000-4, (2001), Electromagnetic compatibility (EMC), Testing and measurement
techniques
[R6] IEC 61000-4-2 (2001), Electromagnetic compatibility (EMC)- Part 4-2: Testing and
measurement techniques - Electrostatic discharge immunity test
[R7] IEC 61000-4-3 (2002), Electromagnetic compatibility (EMC) - Part 4-3: Testing and
measurement techniques - Radiated, radio-frequency, electromagnetic field immunity test
[R8] IEC 60068-2 (1990) (-1, -2, and other parts as appropriate), Environmental testing - Part 2:
Tests
[R9] IEC 60068-2-18 (2000), Environmental testing - Part 2-18: Tests - Test R and guidance:
Water
[R10] IEC 60068-2-75 (1997), Environmental testing - Part 2-75: Tests - Test Eh: Hammer tests
[R11] IEC 60529 (2001), Degrees of protection provided by enclosures (IP Code), IP53
[R12] ANSI/IEEE N42.42 - American National Standard Data format standard for radiation
detectors used for Homeland Security
[R13] Federal Communication Commission Rules, Code of Federal Regulations, Title 47, Part
15, Radio Frequency Devices
3 Definitions
The following definitions apply to ANSI/IEEE standards that have been developed at the request
of the Department of Homeland Security (DHS) for instruments to be used by DHS and
emergency responders.
3.1 A-weighted sound level: The frequency weighting of an acoustic spectrum according to a
standardized frequency response curve based on the frequency response of the human ear.
3.2 acceptance test: Evaluation or measurement of performance characteristics to verify that
certain stated specifications and contractual requirements are met.
3.3 accepted ambient photon background: The background radiation as measured using a
high-pressure ionization chamber, an energy compensated Geiger-Müeller (GM) tube, an energy
compensated proportional counter, a tissue equivalent plastic scintillator, a scintillator with
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 4
spectral compensation, or any other exposure rate meter having a nearly constant energy
response (±30 % in the energy range from 200 keV to 1.5 MeV).
3.4 accredited testing laboratory: Testing laboratory that has been accredited by an
authoritative body with respect to its qualification to perform verification tests on the type of
instruments covered by this standard.
3.5 accuracy: The degree of agreement between the observed value and the conventionally true
value of the quantity being measured.
3.6 adjust: To alter the reading of an instrument by means of a built-in variable (hardware or
software) control.
3.7 alarm: An audible, visual, or other signal activated when the instrument reading exceeds a
preset value or falls outside of a preset range.
3.8 calibrate: To adjust and/or determine the response or reading of a device relative to a series
of conventionally true values.
3.9 calibration: A set of operations under specified conditions that establishes the relationship
between values indicated by a measuring instrument or measuring system, and the
conventionally true values of the quantity or variable being measured.
3.10 check source: A not-necessarily calibrated source that is used to confirm the continuing
functionality of an instrument.
3.11 coefficient of variation (COV (%): ratio of the standard deviation, s, to the arithmetic mean, x , of a set of n measurements, xi , given by the following formula:
1
)(12
−
−==
∑n
xx
xx
sV
i
3.12 conventionally true value (CTV): The commonly accepted best estimate of the value of
that quantity.
NOTE - This and the associated uncertainty will preferably be determined by a national or
transfer standard, or by a reference instrument which has been calibrated against a national or
transfer standard, or by a measurement quality assurance (MQA) interaction with the National
Institute of Standards and Technology (NIST) or an accredited calibration laboratory. (See ANSI
N42.22 and ANSI N42.23.)
3.13 decade: A range of values for which the upper value is a power of ten above the lower
value.
3.14 detection limits: The extremes of detection or quantification for the radiation of interest.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 5
NOTE - The lower detection limit is the minimum statistically quantifiable instrument response
or reading. The upper detection limit is the maximum level at which the instrument meets the
required accuracy.
3.15 detector: A device or component designed to produce a quantifiable response to ionizing
radiation normally measured electronically.
3.16 effective center: For a given set of irradiation conditions, the point within a detector where
the response is equivalent to that which would be produced if the entire detector were located at
that point.
3.17 effective range of measurement: Range of measurements within which the requirements
of this standard are met.
3.18 energy dependence: Variation in instrument response as a function of radiation energy for
a constant radiation type and exposure rate referenced to air.
3.19 exposure: The measure of ionization produced in air by x or gamma radiation.
NOTE—The special unit of exposure rate is the roentgen per hour, abbreviated in this standard
as R/h
NOTE—In this standard, the International System (SI) units sievert (Sv) or gray (Gy) follow in
parentheses the Roentgen value R, though the two units are not physically equivalent.
3.20 false alarm: Alarm NOT caused by a radioactive source under the specified background
conditions.
3.21 functional check: A frequently used qualitative check to determine that an instrument is
operational and capable of performing its intended function.
NOTE - Such checks may include, for example, battery check, zero setting, or source response
check.
3.22 indicated value: (A) A scale or decade reading. (B) The displayed value of the readout.
See also: reading.
3.23 indication: Displayed signal from the instrument to the user conveying information such as
scale or decade, status, malfunction or other critical information.
3.24 influence quantity: Quantity that may have a bearing on the result of a measurement
without being the subject of the measurement.
3.25 innocent alarm: An alarm resulting from an actual increase in radiation level, but for
reasons that are not due to the detection of illicit radioactive materials.
NOTE—Also known as nuisance alarm.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 6
3.26 instrument: A complete system consisting of one or more assemblies designed to quantify
one or more characteristics of ionizing radiation or radioactive material.
3.27 instrument-hour: The number of operating instruments multiplied by the amount of time
they are operating (e.g. 8 instruments operating for 3.75 hours is equivalent to 30 instrument
hours).
3.28 interdiction: Stopping the illicit or inadvertent movement of radioactive material that has
been discovered as a result of radiation detection or measurement.
3.29 monitoring: Means provided to continuously indicate the state or condition of a system or
assembly.
NOTE – May also be used for the real time measurement of radioactivity or radiation levels.
3.30 overload reading: See: over-range reading
3.31 over-range reading: The reading of an instrument when exposed to radiation intensities
greater than the upper detection limit. Syn: overload reading.
3.32 performance test: An evaluation of the performance of an instrument in response to a
given influence quantity.
3.33 point of measurement: Place at which the conventionally true values are determined and
at which the reference point of the instrument is placed for test purposes.
3.34 precision: Degree of agreement of repeated measurements of the same parameter.
3.35 range: All values lying between the lower and upper detection limits.
3.36 reading: The indicated or displayed value of the readout.
3.37 readout: The portion of the instrument that provides a visual display of the reading of the
instrument or the displayed value, with units, displayed and/or recorded by the instrument as a
result of the instrument’s response to some influence quantity.
3.38 reference point of an instrument: Physical mark, or marks, on the outside of an
instrument used to position it at a point where the conventionally true value of a quantity is to be
measured, unless the position is clearly identifiable from the construction of the instrument.
3.39 percent relative error (εREL(%)
): The difference between instrument’s reading, M, and the
conventionally true value, CTV, of the quantity being measured divided by the conventionally
true value multiplied by 100.
εREL(%) = [(M–CTV) ⁄ (CTV)] × 100
3.40 response: Ratio of the instrument reading to the conventionally true value of the measured
quantity.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 7
3.41 response time: The time interval required for the instrument reading to change from 10
percent to 90 percent of the final reading or vice versa, following a step change in the radiation
field at the detector.
3.42 restricted mode: An advanced operating mode that can be accessed by an expert user (e.g.:
via password) to control the parameters that can affect the result of a measurement (i.e.,
radionuclide library, routine function control, calibration parameters, alarm thresholds, etc.).
NOTE—May be called the “advanced” or “expert” mode.
3.43 routine test: Test that applies to each independent instrument to ascertain compliance with
specified criteria
3.44 standard deviation: The positive square root of the variance.
3.45 standard instrument or source: (A) National standard—a standard determined by a
nationally recognized competent authority to serve as the basis for assigning values to other
standards of the quantity concerned. In the U.S., this is an instrument, source, or other system or
device maintained and promulgated by the National Institute of Standards and Technology
(NIST). (B) Primary standard—a standard that is designated or widely acknowledged as having
the highest metrological qualities and whose value is accepted without reference to other
standards of the same quantity. (C) Secondary standard—a standard whose value is assigned by
comparison with a primary standard of the same quantity. (D) Reference standard—a standard,
generally having the highest metrological quality available at a given location or in a given
organization, from which measurements made there are derived. (E) Working standard—a
standard that is used routinely to calibrate or check material measures, measuring instruments, or
reference materials. A working standard is traceable to NIST (see ANSI N42.22 and ANSI
N42.23).
3.46 standard test conditions: The range of values of a set of influence quantities under which
a calibration or a measurement of response is carried out.
3.47 test: A procedure whereby the instrument, circuit, or component is evaluated.
3.48 type test: Initial test of two or more production instruments made to a specific design to
show that the design meets defined specifications.
3.49 uncertainty: The estimated bounds of the deviation from the conventionally true value,
generally expressed as a percent of the mean, ordinarily taken as the square root of the sum of
the square of two components: 1) Random errors that are evaluated by statistical means; and 2)
systematic errors that are evaluated by other means.
3.50 upper measurement limit (UML): The UML is the maximum level at which the
instrument meets the required accuracy.
3.51 variance (σσσσ 2): A measure of dispersion, which is the sum of the squared deviation of
observations from their mean divided by one less than the number of observations.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 8
( )2
1
2
1
1∑
=
−−
=n
i
i xxn
σ
3.52 confidence indication: An indication provided by the instrument of the reliability assigned
to the determined identification.
NOTE—This definition is applicable to this standard.
4 General considerations
Unless otherwise specified in the individual steps, all tests enumerated in this standard are to be
considered as type tests. Certain tests may be considered as acceptance tests by agreement between the
customer and manufacturer.
All test results shall be documented.
4.1 Standard test conditions
The required standard test conditions for environmental quantities, such as temperature and pressure, as
well as those for other quantities that may influence the performance of instruments, are given in Table 1.
These conditions as given in Table 1 shall be met, except where the effect of the condition or quantity
itself is being tested. Environmental quantities, such as temperature and humidity, are referred to as
influence quantities.
4.2 Units and uncertainties
For the purposes of this standard the radiological unit of exposure rate (R/h) is used when defining
radiation fields used for testing.
Throughout the text radiological quantities will be expressed in conventional units, SI units are given in
parentheses.
If uncertainties are not specified for a measurable quantity, they are set to ± 5 %.
4.3 Special word usage
The following word usage applies:
- The word “shall” signifies a mandatory requirement (where appropriate a qualifying statement is
included to indicate that there may be an allowable exception).
- The word “may” signifies an acceptable method or an example of good practice.
- The word “should” signifies a recommended specification or method.
Table 1—Standard test conditions
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 9
Influence quantity
Standard test conditions
(unless otherwise indicated by the
manufacturer)
Stabilization time As stated by the manufacture, or less than 1 minute
Ambient temperature 18 °C to 25 °C
Relative humidity 20 % to 75 %
Atmospheric pressure 70 kPa to 106.6 kPa (525 to 800 mm of
mercury at 0 °C)
Battery voltage Fresh, or fully charged batteries
Reference Point Effective center of detector as marked
Electromagnetic field of external origin Negligible
Magnetic induction of external origin Negligible
Instrument controls Set up for normal operation; alarm set points
set to default
Radiation background Ambient photon exposure rate of up to
25µR/h (0.25 µGy/h)
Contamination by radioactive elements Negligible
Reference photon radiations 241
Am, 137
Cs, 60
Co
Reference neutron radiation 252Cf (2×10
4 n/s ± 20%) in 1 cm steel plus 1
cm lead
5 General requirements
5.1 Controls
5.1.1 Requirement
Controls shall be clearly identified, easily operable under conditions of expected use, and adequately
protected from accidental operation. The on/off button or any other control that could cause the
instrument not to function as expected by the user shall be protected from accidental operation.
5.1.2 Test method
Place each face of the instrument on a flat, hard surface. The instrument may be supported or braced to
maintain this orientation. Place a 500 g weight on the opposite surface of the instrument and verify that
the instrument does not turn off or change operational mode. Repeat this test for each face of the
instrument.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 10
5.2 Documentation Check
5.2.1 Requirement
Manufacturers shall provide instructions to verify proper operation of the instrument. See documentation
section for a detail list of requirements.
5.2.2 Test method
Review the manufacturer’s provided documentation for adequate instructions to verify proper operation
of the instrument.
5.3 Displays
5.3.1 Requirement
The instrument shall directly display the measured exposure or dose rate with the associated radiological
unit (e.g. µR/h, µGy/h, or µSv/h).
Radionuclide identification results shall be displayed on the instrument. If measurement results can be
viewed via a wireless or network link on a secondary device, the failure of that or any secondary device
shall not affect the operation of the instrument.
5.3.2 Test method
The instrument shall be inspected and the type of display noted. Note whether the display is backlit. If
the measurement results can be viewed by a secondary device, verify that the instrument continues to
function properly when the secondary device is switched off.
5.4 Effective range of measurement or indication
5.4.1 Requirement
The effective range of measurement or indication shall be specified and shall be at least 5 µR/h to 2 mR/h
(0.05 µGy/h to 20 µGy/h).
The instrument response over the effective range specified by the manufacturer shall be tested. When
exposed to radiation fields that are greater than the effective range of measurement, the instrument shall
indicate an over range condition.
5.4.2 Test method
Review the manual and record the stated range. The range will be confirmed during the accuracy test.
The over range indication shall be tested during the over range test.
5.5 Audible Alarms
5.5.1 Requirements
The instrument shall provide an audible alarm to indicate an increase in the radiation level that is greater
than the alarm set point. Different techniques may be used to differentiate between the type of radiation
detected (e.g., gamma, neutron, over range). The frequency of an audible alarm signal shall be from 1000
Hz to at least 4000 Hz. Where an intermittent alarm signal is provided, the interval shall not exceed 2 s.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 11
The A-weighted alarm signal volume at a distance of 30 cm from the instrument shall be at least 80
dB(A) and shall not exceed 100 dB(A).
It shall not be possible to disable both the vibration and audible alarm indications simultaneously, except
through the restricted mode. When both alarm signals are off, an indication shall be provided on the
display to inform the user of this condition.
An earphone connection should be available to enable use of the audible function in a high noise
environment.
5.5.2 Test method
The audible alarm signal of the instrument shall be activated with an appropriate radiation source that
may be placed as close to the instrument as practical. The A-weighted sound level at a distance of 30 cm
shall be measured using a peak measurement device.
Verify that it is not possible to switch off the audible and vibration alarm signals simultaneously without
accessing the restricted mode.
Verify that when both alarm modes are off, an indication is provided on the display.
If an earphone connection is provided, verify that the earphone functions when connected and that the
external audible signal is off.
5.6 Vibration Alarm
5.6.1 Requirements
The instrument shall have a vibration alarm signal capabilty. The vibration alarm shall have sufficient
intensity to inform the user of an alarm condition.
The use of carrying pouches is discouraged. If a holder is used, there should be a rigid connection
between the holder and the instrument such that there is no loss of vibration intensity to the user.
The intensity of the vibration at the surface of the instrument (instrument pouch or holder, when used)
shall be greater than 0.8 g. The vibration motor used by the instrument should rotate between 9000 and
11000 rpm.
5.6.2 Test method
Review the instrument manual and record the vibratory motor rotation frequency stated by the
manufacturer.
Install new or fully charge the batteries, as applicable, in the instrument being tested.
Attach the instrument (in a pouch or attached to a holder, when appropriate) to a flat, hard surface using
non-cushioning double-sided tape if possible. Attach a single axis accelerometer to the side of the
instrument that when worn, is closest to the wearer.
After allowing the vibration measurement system to settle, activate the alarm signal and once the
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 12
measurement is stable, record the results. The measured intensity shall be greater than 0.8 g. After a
period of 10 seconds, reduce the radiation field and allow the instrument to return to normal operation
(non-alarm condition). Repeat the test 2 additional times. Each measured reading shall be greater than
0.8 g to be acceptable.
If the vibration signal is intermittent, measure the intensity during the activated period.
5.7 Size
5.7.1 Requirement
The overall dimensions of the instrument should be similar to that of a personal radiation detector (within
the rectangular solid defined by 20 cm in length, 10 cm in width, and 5 cm in depth).
Means shall be provided to securely fix the instrument to the user (for example, a clip, ring, or lanyard),
with attention given to the necessary orientation of the detector and display.
5.7.2 Test method
Measure the physical dimensions of the instrument. The instrument shall be measured outside of its
holster or carrying case and the measurement shall exclude the clip and/or lanyard.
5.8 Mass
5.8.1 Requirement
The mass of the complete instrument should not exceed 400 g.
5.8.2 Test method
Weigh the instrument including its holster or carrying case, clips, and with batteries installed.
5.9 Reference point marking
5.9.1 Requirement
The instrument shall have reference points on both the front, or back, and side indicating the effective
center of the detector.
The instrument shall have an additional reference point indicating its orientation with respect to the
wearer. The presence of a clip may be used as the reference point to indicate proper orientation.
All reference points shall be described in the instrument manual.
5.9.2 Test method
Inspect the instrument and review the manual to verify compliance.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 13
5.10 Explosive atmospheres
5.10.1 Requirement
The manufacturer shall state whether the instrument is certified for use in explosive atmospheres. If
certification is claimed, documentation shall be provided. Certification should be based on UL-913-2004
[B3].
5.10.2 Test method
Inspect documentation provided by the manufacturer. The documentation shall state whether or not the
instrument is suitable for use in explosive atmospheres. A certificate of compliance shall be provided if
the manufacturer states that the instrument may be used in explosive atmospheres. Compliance should be
based on testing done in accordance with UL-913-2004[B3].
5.11 Batteries and Battery Lifetime
5.11.1 Requirement
If non-rechargeable batteries are used, they shall be widely available, not unique to the instrument, and be
replaceable in the field without the use of special tools. When rechargeable batteries are used, provisions
shall be made to permit recharging from AC or DC (12 Volt) power sources.
The batteries shall be capable of powering the instrument in a non-alarm state for a minimum of 16 hours
in a 50 µR/h (0.5 µGy/h) field. The batteries shall be capable of powering the audible alarm continuously
for 30 minutes.
The instrument shall have a low battery indicator.
5.11.2 Test method
Install fresh batteries or fully charge the rechargeable batteries and increase the radiation field as needed
to activate the audible alarm. Verify that the alarm sounds continuously for 30 minutes.
Install fresh batteries or fully charge the rechargeable batteries, switch the instrument on, and allow it to
stabilize and measure background as specified by the manufacturer. Since the instrument will be exposed
to radiation fields of 50 µR/h (0.5 µGy/h), adjust the alarm threshold as necessary to prevent activation
during test. The instrument shall then be exposed to an exposure rate of 50 µR/h (0.5 µGy/h) using 137
Cs
for 16 hours. The low battery indicator shall not come on during the 16-hour period. At the end of the
16-hour period, verify that each alarm signal as provided on the instrument is operational.
To verify the low battery indication requirement, continue to operate the instrument until the low battery
indication is activated. When low battery is indicated, expose the instrument to a radiation field that
triggers the alarm and verify that alarm functions properly.
5.12 Data format and communication interface
5.12.1 Requirement
The instrument shall have the ability to transfer data to an external device, such as a computer. The
transfer should be based on a bi-directional port that meets the requirements of Ethernet, USB, wireless,
or other electronic means such as a removable media device. The technique used shall conform to
applicable IEEE protocols. The transferred data shall be in the XML format following the format defined
in ANSI N42.42 [R12]. When used, wireless techniques shall have the ability to be encrypted.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 14
Communication protocols shall be described in the technical manual. Proprietary communication formats
shall not be used and any required drivers shall be made available to the user.
Proprietary software should not be required for remote data interpretation. If proprietary software is
needed, the software shall be provided by the manufacturer.
5.12.2 Test method
Data transfer requirement is verified in steps 5.13 and 5.14.
5.13 User interface
5.13.1 Requirement
The instrument shall include:
1. A display that is easily readable over the required temperature range and under different
lighting conditions,
2. Controls that are user-friendly for routine operation,
3. Controls and switches that are designed in a way to minimize accidental operation,
4. A menu structure that is simple and easy to be followed intuitively,
5. Detect, search/localize, and identification functions,
6. The capability to operate if the user is wearing gloves,
7. A method to inform the user of the expected time required to collect a spectrum and a
means to allow the user to extend or reduce the collection time,
8. An automated mode of operation that would automatically start spectrum collection and
attempt to identify the radionuclide, and
9. Provide a status indicator, such as a flashing LED or LCD heartbeat to inform the user
that the instrument is functioning properly including visual indication of an alarm
condition.
5.13.2 Test method
In order to test the above requirements, a minimum of three potential users of this type of instrument shall
review the operating instructions provided by the manufacturer. Following the review, each potential user
shall operate the instrument in the routine operation mode. Specifically, the potential user shall:
1. Turn on the instrument and allow it to operate for a period of 30 min. Verify that it is working properly
(e.g., the battery is charged, the detector is present and working, date and time are correct, memory is
available, self-check passed),
2. Calibrate (if necessary),
3. Go through the menu and compare it to the information provided in the technical manual,
4. Make an exposure rate measurement using 137
Cs,
5. Make an identification using 137
Cs and save the data,
6. Repeat step 5 and verify that the estimated collection time is shown on the display, that the time can be
extended or reduced, and that the result is displayed,
7. Transfer the data to an external device, such as a computer, following the manufacturer-provided
information,
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 15
8. If wireless communication can be used, verify that encryption is available,
9. Turn off the instrument.
Steps 1, 2, and 4 through 6 shall be done in low light levels (<150 lux) and repeated in high light levels
(>10 000 with a range up to 32 000 lux (direct sunlight)). A separate test following steps 1, 2, 4 and
5 shall also be performed with the potential user wearing protective gloves. Gloves worn shall be typical
of those used for thermal protection.
A survey evaluation form (see Annex C) shall be completed by each potential user to assess the usability
of the instrument’s controls, interface, and operation. A report shall be generated based on the survey
results.
5.14 Spectral identification
5.14.1 Requirements
NOTE – The following requirements are verified through the performance of the same ordered test
methods listed in section 5.14.2.
5.14.1.1 The instrument shall have the ability to store and transfer at least 50 complete (unprocessed)
spectra. Each stored spectrum shall contain collection and identification results information in
the ANSI N42.42 format including:
- Time and date,
- Instrument type and serial number,
- Hardware and software version,
- Identified radionuclides and associated confidence indications,
- Spectrum collection time interval,
- Measured gamma-ray exposure rate, and
- Neutron count rate at the time of measurement, if provided.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 16
5.14.1.2 An indication shall be displayed or otherwise provided (e.g., “not identified”, “unknown
radiocnuclide”) if a radionuclide cannot be identified.
5.14.1.3 The manufacturer shall describe the meaning of reliability or confidence indications.
5.14.1.4 The instrument shall indicate if the exposure rate is too high for radionuclide identification.
5.14.2 Test Method
5.14.2.1 Verify the transfer and storage of data by performing 50 radionuclide identifications storing the
results at the end of each identification process. Transfer the stored results to a computer. On
the computer, verify that each individual data set contains the required information in the
required format.
5.14.2.2 Verify through review of the manual and direct observation of the instrument when performing
step 6.10.5.
5.14.2.3 Verify through review of the manual, and record the results.
5.14.2.4 Indication that the exposure rate is too high for identification is verified through the
performance of step 6.10.6, “Over-range characteristics for identification.”
6 Radiological Tests
6.1 General test information
Radiation fields used for testing shall be traceable to the NIST.
The reference point of the instrument shall be placed at the point of measurement with the instrument
oriented with respect to the radiation source as indicated by the manufacturer.
If the instrument requires a background radiation measurement, it will be allowed to acquire the data in a
manner specified by the manufacturer.
6.2 Rate of false alarms
6.2.1 Requirements
The false alarm rate for gamma and neutron (when applicable) shall be less than or equal to 1 alarm
activation in a 10-hour period when operated in a stable background environment.
6.2.2 Test method
For this test, the alarm signal threshold shall be the same as that used for the “time to alarm” test.
Place the instrument in an area with a stable ambient background that is not greater than 25 µR/h (0.25
µGy/h), and monitor the instrument for 30 hours.
The number of alarms during this period shall not exceed 1 alarm per 10-hour interval.
If the alarm threshold is adjusted to meet the requirements of the alarm tests defined in steps 6.3 through
6.5, the rate of false alarm test shall be repeated.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 17
6.3 Time to alarm; photons
6.3.1 Requirement
The alarm shall activate within 2 seconds after exposure to an increase in the ambient radiation level of 50
µR/h (0.5 µGy/h) that occurs over a period of not more than 0.5 seconds.
6.3.2 Test method
The false alarm test shall be completed before this test and the alarm signal threshold shall be the same as
the one used for the false alarm test.
Switch on the instrument and allow it to stabilize. Place the instrument in an area where the ambient
background is not greater than 25 µR/h (0.25 µGy/h). Using 137
Cs, increase the radiation field by 50 µR/h
(0.5 µGy/h) above the background in a period of not more than 0.5 seconds. Verify that the alarm is
activated within 2 seconds. Reduce the field and repeat the test 9 additional times.
Repeat the entire process using 241
Am and 60
Co.
The test result is acceptable if the alarm is activated for each trial for each source.
If the alarm threshold is adjusted to meet the requirements of this test, the rate of false alarm test shall be
repeated.
6.4 Time to alarm; neutrons (if provided)
6.4.1 Requirements
The neutron alarm shall activate within 2 seconds after exposure to an unmoderated neutron field
that occurs over a period of not more than 2 seconds.
6.4.2 Test method
Neutron tests should be made in a low scatter irradiation facility (see ISO 8529-1:2001 [B27]) or in an
area where there is open space on all sides of at least 1 m. The alarm set point shall be set the same as
that used for the false alarm test.
Place the instrument at the center of a 30 cm × 30 cm × 15 cm PMMA phantom facing the radiation
source. Position the instrument and phantom at a location where the reference point of the instrument will
be 25 cm from the radiation source when exposure occurs. Using the 252
Cf source listed in Table 1,
increase the neutron field within a period of not more than 2 seconds. The instrument shall alarm within a
period of less than or equal to 2 seconds after the field increase. Reduce the field and repeat the test 9
additional times.
The test result is acceptable if the alarm occurs in 8 out of 10 trials.
If the alarm threshold is adjusted to meet the requirements of this test, the rate of false alarm test shall be
repeated.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 18
6.5 Detection of gradually increasing radiation levels
6.5.1 Requirement
The instrument’s alarm threshold shall not be affected by slowly increasing radiation levels that may be
caused when a wearer is slowly approaching or is being approached by a radiation source.
6.5.2 Test method
From an ambient background of not more than 25 µR/h (0.25 µGy/h), increase the radiation field by
slowly approaching a 137
Cs source that produces an exposure rate that is approximately 50 µR/h (0.5
µGy/h) above the starting point background at the reference position of the instrument. The approach
speed shall be 0.25 m/s.
Return the instrument to the original position, allow the instrument to stabilize, and repeat the process 9
additional times. The results are acceptable if the instrument alarms as the source is approached or within
2 seconds of being exposed to the 50 µR/h (0.5 µGy/h) above background a radiation field for each trial.
If the instrument has neutron detection capabilities, repeat the test using the 252
Cf source listed in Table 1.
For this test, the instrument shall be configured in the same manner as that used when performing the
neutron time to alarm test (step 6.4). The approach speed shall be 0.25 m/s starting with the reference
position of the instrument positioned at a distance of at least 3m from the source. Return the instrument
to the original position, allow the instrument to stabilize, and repeat the process 9 additional times. The
results are acceptable if the instrument alarms as the source is approached or within 2 seconds of being
fully exposed neutron field (25 cm from the source) for each trial.
6.6 Accuracy
6.6.1 Requirements
Displayed exposure rates shall be within ±30 % of the conventionally true value of the applied exposure
rate using 137
Cs.
6.6.2 Test method
Expose the instrument to radiation fields of 400 µR/h (4 µGy/h), 1 mR/h (10 µGy/h), and a field that is
equivalent to 80 % of the response range of the instrument. The instrument readings shall be within
±30 % for each field.
6.7 Personal radiation alarm
6.7.1 Requirement
The instrument shall provide an alarm that will alert the user to the presence of a relatively high radiation
field. The alarm shall be audible and visible, and be different than those associated with radiation
indication alarm described in step 6.3.
6.7.2 Test method
Using 137
Cs expose the instrument to a 10 mR/h (100 µGy/h) radiation field. The personal alarm shall be
activated within 2 s of the exposure. Reduce the radiation field and repeat the exposure two additional
times for a total of three trials. The alarm shall activate within the time specified for each trial.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 19
6.8 Over-range response
6.8.1 Requirements
When exposed to a radiation field that is two times the maximum exposure rate specified by the
manufacturer, the indication of the instrument shall remain at the maximum of that range and an over-
range indication shall be displayed for the duration of the exposure. The instrument shall respond
normally (recover) within a time period of 5 min after the exposure is reduced.
6.8.2 Test method
Expose the instrument to a 137
Cs field that is twice the maximum range specified by the manufacturer for
1 minute. Verify that an over-range response is indicated. Remove the 137
Cs source and verify that the
instrument responds normally within 5 min.
6.9 Interfering Ionizing Radiation
6.9.1 Requirements
If the instrument has a neutron detector, the neutron detector shall be insensitive to photon radiation at the
tested level.
6.9.2 Test method
The instrument shall be exposed to a 137
Cs radiation field of 10 mR/h (100 µGy/h) for 1 min. The
instrument shall not activate a neutron alarm during exposure.
.
6.10 Radionuclide identification
For the purposes of this standard, when identifying radionuclides, test results are considered acceptable
when an instrument identifies the radionuclide(s) of interest, or that radionuclide(s) and expected
daughter(s). It is considered not acceptable if the instrument identifies unexpected radionuclides or only
the daughter(s) of the radionuclide(s) of interest.
If a library is used as part of the identification process, it shall not be altered during the entire testing
process.
6.10.1 Radionuclide categorization
The radionuclides that are most likely to be encountered by instruments addressed by this standard are
listed in four different categories.
NOTE—This is an informative list and should not be considered as all-inclusive.
- Special Nuclear Materials (SNM): Uranium (used to indicate 233
U, 235
U, 238
U), 237
Np, Pu.
- Medical radionuclides: 18
F, 67
Ga, 51
Cr, 75
Se, 89
Sr, 99
Mo, 99m
Tc, 103
Pd, 111
In, Iodine (123
I, 125
I, 131
I), 153
Sm, 201
Tl, 202
Tl, 133
Xe.
- Naturally occurring radioactive materials (NORM): 40
K, 138
La, 226
Ra, 232
Th and daughters, 238
U and
daughters.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 20
- Industrial radionuclides: 57
Co, 60
Co, 133
Ba, 137
Cs, 192
Ir, 204
Tl, 226
Ra, 152
Eu, and 22
Na.
6.10.1.1 Requirements
The manufacturer shall state the radionuclides that the instrument can identify and their category. The
categories selected should be based on the list shown in 6.10.1.
6.10.2 Single radionuclide
6.10.2.1 Requirements
The instrument shall be able to identify the following radionuclides within the time specified by the
manufacturer with a maximum of 5 minutes. The manufacturer shall provide radionuclide-specific test
results.
- Medical radionuclides: 67
Ga, 99m
Tc, Iodine (123
I, 131
I), 201
Tl.
- NORM: 40
K, 226
Ra, 232
Th.
- Industrial radionuclides: 22
Na, 57
Co, 60
Co, 133
Ba, 137
Cs, 152
Eu, 192
Ir, and 241
Am.
- Special Nuclear Materials: HEU (highly enriched uranium, 235
U > 90%), Pu [Reactor grade plutonium
(> 6% 240
Pu)].
6.10.2.2 Test method
One at a time, expose the instrument to the radionuclides listed in 6.10.2.1. The gamma-ray exposure rate
at the reference point from each source, unshielded or shielded, shall be 100 µR/h (1 µGy/h). The test
shall consist of 10 trials for each radionuclide. The instrument shall be reset between each trial, if
appropriate. The performance is acceptable when the instrument correctly identifies the radionuclide in 8
out of 10 consecutive trials.
The shielded portion of the test shall be done using 5 mm of steel.
6.10.3 Simultaneous radionuclide identification
6.10.3.1 Requirement
The instrument shall be able to identify at least two radionuclides simultaneously.
6.10.3.2 Test method
Expose the instrument to 99m
Tc and 137
Cs simultaneously. Each radionuclide shall produce an exposure
rate of approximately 100 µR/h (1 µGy/h) at the reference point. The test shall consist of 10 trials. The
performance is acceptable when the instrument correctly and simultaneously identifies and categorizes
both of the two test radionuclides in 8 out of 10 consecutive trials.
6.10.4 Masking
6.10.4.1 Requirement
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 21
The instrument shall provide an indication (e.g., the correct identification, “unknown”, “unable to
identify”) when exposed to a radionuclide masked by another radionuclide that is not listed in the library
or that has a much higher radiation intensity than the unmasked radionuclide.
6.10.4.2 Test method
To ensure that an indication is provided when exposed to a masked radionuclide that is not listed in the
library, each radionuclide shall produce an exposure rate of approximately 100 µR/h (1 µGy/h) at the
reference point. The test shall consist of 10 trials. The performance is acceptable when the instrument
indicates that there may be an unidentifiable radionuclide present in 8 out of 10 consecutive trials. The
combination of radionuclides shall include one that is in the library and another that is not. The
combination of 67
Ga and 54
Mn is recommended.
To ensure that an indication is provided when exposed to a radionuclide that is masked by another
radionuclide that has a much higher radiation intensity than the unmasked radionuclide, repeat the test
using a combination of radionuclides where the radionuclide of interest (e.g., 137
Cs) produces an exposure
rate at the reference point of 50 µR/h (0.5 µGy/h) and the masking radionuclide at 100 µR/h (1 µGy/h).
The source combinations for this test shall include 67
Ga as the masking radionuclide with HEU as the
radionuclide of interest, and 137
Cs as the masking radionuclide with Pu (reactor grade) as the radionuclide
of interest.
6.10.5 Response to an “Unknown” radionuclide
6.10.5.1 Requirement
The instrument shall provide an indication (e.g., “unknown”, “unable to identify”, “not in library”) when
exposed to radionuclides that are not in the library.
6.10.5.2 Test method
Expose the instrument to a radionuclide that is not in the library. 166m
Ho is recommended. The source
shall produce an exposure rate of approximately 50 µR/h (0.5 µGy/h) at the reference point. The test shall
consist of 10 trials. The performance is acceptable when the instrument indicates that the result is
unknown in 8 out of 10 consecutive trials.
6.10.6 Over-range characteristics for identification
6.10.6.1 Requirements
The manufacturer shall state the maximum gamma-ray exposure rate (relative to 137
Cs) for identification.
6.10.6.2 Test method
Increase the ambient exposure rate using 137
Cs to 90 % of the maximum exposure rate for radionuclide
identification as stated by the manufacturer and perform a radionuclide identification. The instrument
shall correctly identify 137
Cs in 8 out of 10 trials.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 22
7 Environmental Performance Requirements
7.1 General test practices
To ensure that the instrument functions properly when exposed to the environments stated in section 7,
the following tests shall be performed as required.
7.1.1 Establishing the nominal reading To verify that the gamma exposure rate indication meets the requirements, prior to the test;
establish a nominal reading by collecting 10 independent readings and determining the mean
reading, standard deviation, and COV using 137
Cs. The COV should be ≤12 %. It may be
necessary to increase the alarm threshold to prevent an alarm due to the field used for testing.
7.1.2 Verifying the gamma reading To verify that the instrument meets the gamma exposure rate requirements during or after each
test, expose the instrument to the same radiation field used to determine the nominal reading
using the same technique. Each mean reading shall be within ±15% of the nominal mean reading.
7.1.3 Identification verification To verify identification operation, a series of three radionuclide identifications shall be performed
as required using 133
Ba and 137
Cs positioned to provide an exposure rate of 100 uR/h (0.1 uGy/) at
the reference point. Each identification shall be correct.
7.1.4 Verification of the gamma alarm To verify that the gamma alarm functions as required, using
137Cs increase the exposure rate
above the alarm threshold and verify that the alarm functions based on the requirements in step
6.3.
7.1.5 Verification of the neutron alarm To verify that the neutron alarm functions as required, expose the instrument to the unmoderated 252
Cf source (with the instrument on a phantom, see 6.4) and verify that the alarm functions based
on the requirements in step 6.4
7.2 Temperature
7.2.1 Requirement
The instrument shall function correctly at temperatures from –20 ºC to +50 ºC. If the manufacturer
specifies a broader operating temperature range, the instrument shall be tested at the broader temperature
range as specified by the manufacturer. Relative humidity shall be within the range specified in Table 1,
Standard Test Conditions.
7.2.2 Test method
Switch the instrument on and place it in an environmental chamber at a temperature of 22 ºC ± 2 ºC.
Allow the chamber and instrument to stabilize at 22 °C for a period of 1 hour. During the last 15 minutes
of the stabilization period, establish the nominal reading.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 23
Increase the temperature at a rate of 10 °C per hour to +50 °C. At each 10 °C (30 and 40 °C) increment,
stabilize the temperature for a period of 45 minutes. During the last 15 minutes of each stabilization
period, verify the gamma reading, perform a verification of the identification function, and verify that the
gamma and neutron alarms function as required.
At the high temperature limit of +50 °C, the instrument shall be exposed for a period of 4 hours with the
gamma reading, alarms, and identification functions verified during the last 15 minutes of the 4-hour
period.
This same process shall be performed for temperatures that are less than the reference temperature of
22 °C. The 10 °C intervals are 10, 0, and −10 °C and the lower temperature limit is −20 °C. The test at
−20 °C shall be the same as that performed at +50 °C.
7.3 Temperature Shock
7.3.1 Requirement
The instrument shall be fully functional within 1 hour of exposure to rapid temperature changes from
22°C to –20°C, –20°C to 22 °C, 22°C to 50°C, and 50°C to 22 °C with each change being made in less
than five minutes. Relative humidity shall be within the range specified in Table 1, Standard Test
Conditions.
7.3.2 Test method
Switch the instrument on and place it in an environmental chamber at a temperature of 22 ºC ± 2 ºC.
Allow the chamber and instrument to stabilize at 22 °C for a period of 1 hour. During the last 15 minutes
of the stabilization period, establish the nominal reading.
With the instrument observed continuously, expose it to a temperature of 50 °C with the temperature
change being made in less than 5 minutes. Every 15 minutes, verify the gamma reading, perform a
verification of the identification function, and verify that the gamma and neutron alarms function as
required.
After 30 minutes, the instrument’s mean indicated reading shall be within ±15 % of the mean reading
obtained at 22 °C. No alarms shall occur due to the test. If the instrument is unable to perform properly
after the first 30 minutes, an additional 30 minutes is recommended with the time required for recovery
noted. If the instrument recovers within the first 30 minutes, data does not need to be taken during the
second 30 minutes; however, the instrument should remain in this environment during the period to reach
temperature stabilization.
Following the stabilization period, expose the instrument to a temperature of 22°C ± 2°C. This change
shall be performed in less than 5 minutes and the analysis process stated above repeated.
The entire process shall be repeated for the 22 °C to –20 °C and –20 °C to 22 °C.
7.4 Humidity
7.4.1 Requirement
The instrument shall function correctly over the range of relative humidity up to 93% RH at 35 ºC.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 24
7.4.2 Test method
Switch the instrument on and place it in an environmental chamber at a temperature of 22 ºC ± 2 ºC and
40 % RH. Allow the chamber and instrument to stabilize at 22 °C for a period of 1 hour. During the last
15 minutes of the stabilization period, establish the nominal reading.
The humidity level shall then be increased at a rate not exceeding 10% RH per hour until attaining 93 % ±
3% RH. The temperature shall be increased at a rate not exceeding 10ºC per hour. The humidity and
temperature shall be maintained at this value for 8 hours. The instrument shall be observed continuously
during the ramp and soak.
Following the 8-hour soak, verify the gamma reading, perform a verification of the identification
function, and verify that the gamma and neutron alarms function as required.
The humidity shall then be reduced to 40 % RH at a rate not exceeding 10 % RH per hour while
maintaining the temperature at 35°C ± 2°C. After allowing the instrument to stabilize in those conditions
for a minimum of 1 hour, verify the gamma reading, perform a verification of the identification function,
and verify that the gamma and neutron alarms function as required.
7.5 Moisture and dust protection
7.5.1 Requirements
The instrument case design shall meet the requirements stated for IP code 53 [R11], which means that the
instrument shall be protected from the ingress of dust and spraying water. For IP53, the ingress of dust is
not totally prevented, but dust shall not penetrate in a quantity to interfere with satisfactory operation of
the instrument or to impair safety, and water sprayed at an angle up to 60º on either side of the vertical
shall have no harmful effects.
7.5.2 Test method - Dust
The test shall be conducted in a dust chamber [R11] where the powder circulation pump may be replaced
by other means suitable to maintain the talcum powder in suspension in a closed test chamber. The
amount of powder to be used should be 2 kg/m3. The powder shall not have been used for more than 20
tests.
Prior to the test, establish the nominal reading and verify that the gamma and neutron alarms function as
required. For this test, the alarm threshold shall be the same as that used for the false alarm test.
Remove the radiation field and place the instrument inside the dust chamber. Expose the instrument to the
dust environment for a period of 1 hour. No alarms shall occur during testing.
Following the test, verify the gamma reading and the gamma and neutron alarms. In addition, an
inspection shall be performed to determine the extent of dust ingress. Particular attention shall be made to
the battery compartment and any other easily accessed portions of the instrument. The protection is
satisfactory if, on inspection, powder has not accumulated in a quantity or location such that, as with any
other kind of dust, it could interfere with the correct operation of the instrument.
7.5.3 Test method - Moisture
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 25
The test shall be made using a suitable nozzle [R10] with the water pressure adjusted to give flow rate of
10 L/min ± 5 % which should be kept constant during the test. The water temperature should not differ
by more than 5 ºC from the temperature of the instrument under test. The test duration is 1 min/m2 of the
calculated surface area of the instrument with a minimum duration of 5 min.
Prior to the test, establish the nominal reading and verify that the gamma and neutron alarms function as
required. For this test, the alarm threshold shall be the same as that used for the false alarm test.
Place the instrument inside the test chamber and expose the instrument to the water spray. The spray
nozzle should be located approximately 2 meters from the instrument. The instrument shall be positioned
such that the nozzle is directly pointed at the display. During the exposure, the orientation shall be
changed by +60º and -60º in two orthogonal planes relative to each side of the instrument. The
instrument shall respond to the presence of radiation throughout the test and after the test. No alarms shall
occur during testing.
Following the test, verify the gamma reading and the gamma and neutron alarms. Inspect the instrument,
including the battery compartment, to ensure that moisture did not penetrate into the instrument.
7.6 Cold temperature start up
7.6.1 Requirement
The instrument shall be able to operate when switched on at the cold temperature limit (-20 ºC).
7.6.2 Test method
Switch the instrument on and place it in an environmental chamber at a temperature of 22 ºC ± 2 ºC.
Allow the chamber and instrument to stabilize at 22 °C for a period of 1 hour. During the last 15 minutes
of the stabilization period, establish the nominal reading. Remove the sources, switch the instrument off, and decrease the temperature in the chamber at a rate of
10 °C/h to −20 °C. Allow the temperature to stabilize for a period of 2 hours.
Switch on the instrument, and after the manufacturer’s stated warm-up time, verify the gamma reading,
perform a verification of the identification function, and verify that the gamma and neutron alarms
function as required. Remove the sources, switch off the instrument and return the temperature to 22 °C
at a rate of 10 °C/hr.
8 Electromagnetic Performance Requirements
8.1 General test practices
To ensure that the instrument functions properly when exposed to the environments stated in section 8,
the following tests shall be performed as required.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 26
8.1.1 Establishing the nominal reading To verify that the gamma exposure rate indication meets the requirements, prior to the test;
establish the nominal reading by collecting 10 independent readings and determining the mean
reading, standard deviation, and COV using 137
Cs. The COV should be ≤12 %. The alarm
threshold shall be the same as that used for the false alarm test.
8.1.2 Verifying the gamma reading To verify that the instrument meets the gamma exposure rate requirements during or after each
test, expose the instrument to the same radiation field used to determine the nominal reading
using the same technique. Each mean reading shall be within ±15% of the nominal mean reading.
8.1.3 Identification verification To verify identification operation, a series of three identifications shall be performed at each test
point using 133
Ba and 137
Cs positioned to provide an exposure rate of 100 uR/h (0.1 uGy/) at the
reference point of the instrument. Each identification shall be correct.
8.1.4 Verification of the gamma alarm To verify that the gamma alarm functions, using
137Cs increase the exposure rate above the alarm
threshold and verify that the alarm functions based on the requirements in step 6.3.
8.1.5 Verification of the neutron alarm To verify that the neutron alarm functions, expose the instrument to the unmoderated
252Cf source
(with the instrument on a phantom, see 6.4) and verify that the alarm functions based on the
requirements in step 6.4
8.2 Electrostatic Discharge (ESD)
8.2.1 Requirement
The instrument shall not be affected by exposure to electrostatic discharges at intensities of up to 6 kV
using the contact discharge technique.
8.2.2 Test method
The “contact discharge” technique [R6] for conductive surfaces and coupling planes shall be used.
Discharge points shall be selected based on user accessibility.
Prior to the ESD test, establish the nominal reading and perform an identification verification. Remove the radiation field and expose the instrument to ESD. There shall be ten discharges per discharge
point with a one-second-recovery time between each discharge. It is recommended that tests first be
performed at 2 kV, then if acceptable, 4 kV, followed by 6 kV.
Following the test, verify the gamma reading, alarm and identification functions.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 27
8.3 Radio frequency
8.3.1 Requirement
The instrument shall not be affected by radio frequency (RF) fields over the frequency range of 80 MHz
to 2.5 GHz at an intensity of 50 volts per meter (V/m). When exposed to these RF fields, the instrument
shall function correctly. No alarms shall occur as a result of the RF radiation alone.
8.3.2 Test method
Prior to the RF test, establish the nominal reading.
Place the instrument and source in a RF controlled environment and expose it to a RF field of 50 V/m as
measured without an instrument present in the test cell over a frequency range of 80 MHz to 2.5 GHz that
is 80 % amplitude modulated with a 1-kHz sine wave. The test should be performed using an automated
sweep at a frequency change rate not greater than 1 % of the fundamental. The instrument’s readings
during the test should be within ±15 % of the nominal reading. No other functional changes shall occurr
such as, alarm activation, mode changes, loss of display, etc.
Remove the radiation source and repeat the test. No alarms or substantial changes in the reading of the
instrument shall occur as a result of the RF radiation alone.
8.4 Magnetic fields
8.4.1 Requirements
When exposed to direct current (DC) magnetic fields in all three mutually orthogonal orientations relative
to a 10 gauss (1 mT) magnetic field, the instrument shall function correctly.
8.4.2 Test method
Prior to the magnetic field test, establish the nominal reading.
Expose the instrument to a 10 gauss (1 mT) magnetic field. The instrument’s reading during the test shall
be within ±15 % of the nominal reading. No other functional changes shall occurr such as, alarm
activation, mode changes, loss of display, etc.
Remove the radiation source and repeat the test. No alarms or substantial changes in the reading of the
instrument shall occur as a result of the magnetic field alone.
The test shall be repeated for all three mutually orthogonal orientations of the instrument with respect to
the magnetic field.
8.5 Radiated emissions
8.5.1 Requirement
RF emissions from an instrument shall be less than that which can interfere with other equipment located
in the area of use. RF emissions when measured at three meters shall be less than those shown in Table 2.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 28
Table 2—Radiated RF emission limits
Emission frequency range
(MHz)
Field Strength
(micro volts/meter)
30–88 100
88–216 150
216–960 200
Above 960 500
8.5.2 Method of test
Place the instrument in a shielded room or chamber, as appropriate. Place an antenna three meters from
the instrument. With the instrument off, collect a background radio frequency spectrum using a bandwidth
of 50 kHz.
Switch the instrument on and perform a RF scan. Repeat the test with the instrument performing a
radionuclide identification. RF emissions shall be less than those shown throughout the test.
9 Mechanical Performance Requirements
9.1 Vibration
9.1.1 Requirement
The instrument shall withstand exposure to vibrations associated with the operation of handheld or hand-
carried equipment. The physical condition and functionality of the instrument shall not be affected by
exposure (e.g.: solder joints shall hold, nuts and bolts shall not come loose).
9.1.2 Test method
Prior to the test, establish a nominal reading as stated in step 8.1.1.
Remove the radiation source and subject the instrument to a random vibration at 0.01 g2/Hz (spectral
density) using 5 and 500 Hz for the frequency endpoints for a period of 1 hour in each of three orthogonal
orientations. No alarms shall occur during the test.
Following the test, verify the gamma reading, alarm, and identification functions as stated in steps 8.1.2,
8.1.3, 8.1.4, and 8.1.5. There shall be no visible external damage to the instrument, and all control
functions shall be verified to be operating correctly.
9.2 Drop Test
9.2.1 Requirement
After being subjected to drops on each of its six surfaces from a height of 1.5 m onto a concrete floor, the
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 29
instrument shall function correctly and alarm at a change in the radiation field. It is acceptable if a
transient alarm occurs at the moment of exposure.
9.2.2 Test method
Prior to the test, establish a nominal reading as stated in step 8.1.1.
Remove the radiation source. The instrument shall then be dropped from a height of 1.5 m onto a concrete
surface on each of its six surfaces.
Following the test, verify the gamma reading, alarm, and identification functions as stated in steps 8.1.2,
8.1.3, 8.1.4, and 8.1.5. There shall be no visible external damage to the instrument, and all control
functions shall be verified to be operating correctly.
9.3 Impact (Microphonics)
9.3.1 Requirement
The instrument shall be unaffected by microphonic conditions such as those that may occur from low
intensity impacts from sharp contact with hard surfaces. The physical condition and functionality of the
instrument shall not be affected by exposure (e.g.: solder joints shall hold, nuts and bolts shall not come
loose).
9.3.2 Test method
Prior to the test, establish a nominal reading as stated in step 8.1.1.
Remove the radiation source. Using an appropriate test device (i.e. spring hammer), expose the instrument
case to 3 impacts at an intensity of 0.2 joules (J). 0.2 J is equivalent to a mass of 0.2 kg moving at 1.4 m/s
over a distance of 0.1 m (IEC 60068-2-75 [R11]). The test shall be performed on each side of the
instrument case while observing the readings. No alarms shall occur as a result of the mechanical shock
alone.
Following the test, verify the gamma reading, alarm, and identification functions as stated in steps 8.1.2,
8.1.3, 8.1.4, and 8.1.5. There shall be no visible external damage to the instrument, and all control
functions shall be verified to be operating correctly.
10 Documentation
This section specifies the requirements for documentation.
10.1 Type test report
The manufacturer shall provide a report covering the type tests performed in accordance with the require-
ments of this standard.
10.2 Certificate
The manufacturer shall provide a certificate or other documentation containing at least the following
information:
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 30
— Contacts for the manufacturer including, but not limited to, name, address, telephone number,
fax number, e-mail address, etc.
— Type of instrument, detector, and types of radiation the instrument is designed to measure.
— Range of exposure rates the instrument is designed to measure.
— Reference points and reference orientation for radiation source used for calibration.
— Location and dimensions of the sensitive volumes of the detectors.
— Response of the instrument to different appropriate radiation energies.
— Results of tests for accuracy, linearity, and lower limit of detection.
— Results of calibration tests (isotopes calibration with and date of next calibration due date)
— Weight and dimensions of the instrument.
— Power supply (battery) requirements.
— Results of tests under environmental conditions.
— Results of electrical and mechanical tests.
— List of radionuclides to which the instrument was tested, and
— FWHM and efficiency for 137
Cs.
10.3 Operation and maintenance manuals
The manufacturer shall supply an operational and maintenance manual containing at least the following
information for the user:
— Operating instructions and restrictions. Instructions shall include information regarding alarm
threshold adjustments.
— Troubleshooting guide.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 31
Annex A (informative) Bibliography
A.1 General
[B1] IEC 60068-2, Basic Environmental Testing Procedures—Part 2: Tests. (All Sections.)
[B2] IEEE Std C62.41™-1991, IEEE Recommended Practice on Surge Voltages in Low-Voltage
AC Power Circuits.5, 6
[B3] UL 913– 2002, Intrinsically Safe Apparatus and Associated Apparatus for Use in Class I, II,
and III, Division 1, Hazardous (Classified) Locations.7
A.2 Detectors
[B4] ANSI N42.12-1994, American National Standard for Calibration and Usage of Thallium-
Activated Sodium Iodide Detector Systems for Assay of Radionuclides.
[B5] ANSI N42.13-1986 (R1993), American National Standard for Calibration and Usage of
“Dose Calibrator” Ionization Chambers for the Assay of Radionuclides.
[B6] ANSI N42.14-1999, American National Standard for Calibration and Use of Germanium
Spectrometers for the Measurement of Gamma-Ray Emission Rates of Radionuclides.
[B7] ANSI N42.31-2003 American National Standard – Measurement Procedures for Resolution
and Efficiency of Wide-Bandgap Semiconductor Detectors of Ionizing Radiation.
[B8] IEEE Std 300™-1988, IEEE Standard Test Procedures for Semiconductor Charged-Particle
Detectors.
[B9] IEEE Std 309™-1999/ANSI N42.3-1999, IEEE Standard Test Procedures and Bases for
Geiger-Mueller Counters.
[B10] IEEE Std 325™-1996 (R2002), IEEE Standard Test Procedures for Germanium Gamma-
Ray Detectors
A.3 Detection and identification instruments
[B11] ANSI N42.33-2003, American National Standard for Portable Radiation Detection
Instrumentation for Homeland Security.
5IEEE publications are available from the Institute of Electrical and Electronics Engineers, 445
Hoes Lane, P.O. Box 1331, Piscataway, NJ 08855-1331, USA (http://standards.ieee.org/).
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 32
6The IEEE standards referred to in Annex A are trademarks belonging to the Institute of
Electrical and Electronics Engineers, Inc. 7UL standards are available from Global Engineering Documents, 15 Inverness Way East,
Englewood, Colorado 80112, USA
(http://global.ihs.com/).
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 33
[B12] ANSI N42.34-2003, American National Standard Performance Criteria for Hand-held
Instruments for the Detection and Identification of Radionuclides.
[B13] ANSI PN42.35, Draft American National Standard for Evaluation and Performance of
Radiation Detection Portal Monitors for Use in Homeland Security.8
[B14] IEC WD62327, Radiation Protection Instrumentation—Hand-held Instruments for the
Detection and Identification of Radioactive Isotopes and additionally for the Indication of
Ambient Dose Equivalent Rate from Photon Radiation (Draft).9
[B15] ISO/DIS 22188:2002, Monitoring for Inadvertent Movement and Illicit Trafficking of
Radioactive Material.10
A.4 Radiological protection instruments
[B16] ANSI N42.17A-1989 (R1994), American National Standard Performance Specifications
for Health Physics Instrumentation—Portable Instrumentation for Use in Normal Environmental
Conditions.
[B17] ANSI N42.17B-1989 (R1994), American National Standard Performance Specifications
for Health Physics Instrumentation—Occupational Airborne Radioactivity Monitoring
Instrumentation.
[B18] ANSI N42.17C-1989 (R1994), American National Standard Performance Specifications
for Health Physics Instrumentation—Portable Instrumentation for Use in Extreme Environmental
Conditions.
[B19] ANSI N42.20-2003, American National Standard Performance Criteria for Active
Personnel Radiation Monitors.
[B20] ANSI N323A-1997, American National Standard Radiation Protection Instrumentation
Test and Calibration Portable Survey Instruments.
[B21] ANSI N323B-2003, American National Standard for Radiation Protection Instrumentation
Test and Calibration, Portable Survey Instrumentation for Near Background Operation.11
[B22] IEC 60395 (1972), Portable X or Gamma Radiation Exposure Rate Meters and Monitors
for Use in Radiological Protection.
A.5 Electromagnetic compatibility
[B23] 47 CFR 0-19: 2002, Telecommunication.12, 13
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 34
8This ANSI standards project was not approved at the time this publication went to press. For
information about obtaining a draft, contact the Institute of Electrical and Electronics Engineers,
445 Hoes Lane, P.O. Box 1331, Piscataway, NJ 08855-1331, USA (http://stan-dards.ieee.org/). 9This IEC standards project was not approved at the time this publication went to press. For
information about obtaining a draft, contact the International Electrotechnical Commission, Case
Postale 131, 3, rue de Varembé, CH-1211, Genève 20, Switzerland/Suisse (http:// www.iec.ch/). 10
ISO publications are available from the ISO Central Secretariat, Case Postale 56, 1 rue de
Varembé, CH-1211, Genève 20, Switzer-land/Suisse (http://www.iso.ch/). ISO publications are
also available in the United States from the Sales Department, American National Standards
Institute, 25 West 43rd Street, 4th Floor, New York, NY 10036, USA (http://www.ansi.org/). 11
This approved ANSI standard will be available from the Institute of Electrical and Electronics
Engineers, 445 Hoes Lane, P.O. Box 1331, Piscataway, NJ 08855-1331, USA
(http://standards.ieee.org/), in early 2004. 12
Supersedes FCC P15: 1976, Radio Frequency Devices. 13
CFR publications are available from the Superintendent of Documents, U.S. Government
Printing Office, P.O. Box 37082, Washing
ton, DC 20013-7082, USA (http://www.access.gpo.gov/).
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 35
[B24] IEC 61000-6-2 (1999), Electromagnetic Compatibility (EMC)—Part 6-2: Generic
Standards—Immunity for Industrial Environments.
A.6 Units, quantities, calibrations
[B25] ISO 4037-1:1996, X and Gamma Reference Radiation for Calibrating Dosemeters and
Doserate Meters and for Determining their Response as a Function of Photon Energy—Part 1:
Radiation Characteristics and Production Methods.
[B26] ISO 4037-2:1997, X and Gamma Reference Radiation for Calibrating Dosemeters and
Doserate Meters and for Determining their Response as a Function of Photon Energy—Part 2:
Dosimetry for Radiation Protection over the Energy Ranges from 8 keV to 1,3 MeV and 4 MeV
to 9 MeV.
[B27] ISO 8529-1:2001, Reference Neutron Radiations—Part 1: Characteristics and Methods of
Production.”
[B28] ISO 8529-2:2000, Reference Neutron Radiations—Part 2: Calibration Fundamentals
Related to the Basic Quantities Characterizing the Radiation Field.
[B29] NIST SP 250-98 ED, NIST Calibration Services User’s Guide, 1998 Edition.14
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 36
14Information on NIST Special Publications may be obtained from the National Institute of
Standards and Technology at http:// www.nist.gov/.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 37
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 38
Annex B (informative) Detector tests
This standard and ANSI N42.33-2003 [B11], ANSI N42.34-2003 [B12], and ANSI PN42.35
[B13] utilize some of the following types of detectors:
— Cesium Iodide (CsI) Scintillation detectors: These detectors are used for their high
efficiency of light output per photon incident. They are operated at room temperature and
have moderate energy resolution. Test procedures for systems using scintillation detectors
can be found in ANSI N42.12-1994 [B4].
— Sodium Iodide (NaI) Scintillation detectors: These detectors are available in large sizes
such that they have both high efficiency and moderate energy resolution. They are operated
at room temperature. Test procedures are given in ANSI N42.12-1994 [B4].
— CZT Semiconductor detectors: CZT and other wide-bandgap semiconductor detectors are
semiconductor detectors that can be operated at room temperatures. At this time they are
small physically and therefore have low efficiency. They have good energy resolution
though somewhat poorer than that of Germanium detectors. Standard test procedures for
these detectors are given in ANSI N42.31-2003 [B7].
— Germanium Gamma-ray detectors: These detectors have very high energy resolution and
are currently of sufficient size to have also high efficiency. They must be operated at
cryogenic temperatures. Test procedures for these detectors are given in IEEE Std 325-
1996 [B10].
— Semiconductor charged-particle detectors: These detectors are capable of high resolution
measurements of charged particles. Test procedures for these detectors are given in IEEE
Std 300-1988 [B8].
— Geiger-Mueller Counters: These are widely used for radiation detection and intensity
measurements. They are avalanche detectors, the output signals of which are independent
of the radiation energy. Test procedures for these detectors are given in IEEE Std 309-
1999/ANSI N42.3-1999 [B9].
— Ionization chambers: These are highly accurate detectors for gross measurement of
radiation intensity. They are operated at room temperature. Test procedures for these
detectors are given in ANSI N42.13-1986 [B5].
— Plastic Scintillator detectors: These detectors are particularly useful for portal monitors.
Standards and standard measurement procedures have not yet been developed.
— High-pressure 3
He proportional counters: These are particularly useful for neutron
detection and are commonly used in portal monitors.
ANSI
N42.48-D4 PERFORMANCE REQUIREMENTS FOR SPECTROSCOPIC PERSONAL
RADIATION DETECTORS (SPRDS) FOR HOMELAND SECURITY
Copyright © 2004 IEEE. All rights reserved. 39
Annex C (informative) Sample user interface evaluation technique
Controls
1. Was the on/off switch easy to find? Y/N
2. Were all the controls labeled? Y/N
3. Were all the labeled controls easy to read/interpret? Y/N
4. Were all the controls easy to operate without gloves? Y/N
5. Could all the controls be operated with gloves? Y/N
Interface
6. Was everything readable in low light levels Y/N
7. Was everything readable in high light levels Y/N
8. Did the display contain abbreviations or icons? (If no, skip next
question.) Y/N
9. Were the abbreviations or icons easy to interpret or understand? Y/N
Operation
10. Did the instrument convey it’s state-of-health at start-up (e.g., battery
life, detector present and working, memory available, mode of
operation)?
Y/N
11. Did you have to refer to the instruction manual more than once to
complete the test? Y/N
12. Was the menu structure simple and intuitive? Y/N
13. At any time during the test did the instrument prompt you for action? Y/N
14. Did the instrument issue any cautions or warning? (if no, skip next
question) Y/N
15. Did the instrument provide information on the nature of the cautions or
warning and a corresponding course of action?
Y/N