Analytical method validation Presented by Debbie Parker 4 July, 2016

Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Embed Size (px)

Citation preview

Page 1: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Analytical method validation

Presented by Debbie Parker

4 July, 2016

Page 2: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 2 © PharmOut 2015

Introduction

This session will cover:

Guidance and references

The types of test methods

Validation requirements

Summary

Page 3: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 3 © PharmOut 2015

References

• ICH Q2(R1) – Validation of Analytical procedures: Text and Methodology

• FDA Guidance for Industry – Analytical Procedures and Methods Validation for Drugs and Biologics July 2015

• FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology

• FDA CDER reviewer guideline for validation of chromatographic methods (1994)

• WHO TRS 937 Appendix 4 “Analytical Method Validation 2006

• ICH Q2(R1) 2005

• Compendia General Chapters

Page 4: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 4 © PharmOut 2015

Recent changes

FDA has just released a new guidance document:

• Analytical Procedures and Methods Validation for Drugs and Biologics, Guidance for Industry, July 2015

• Supersedes the draft document (same name) published Feb 2014

Replaces:

• Analytical Procedures and Methods Validation, 2000

Page 5: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 5 © PharmOut 2015

New guidance – Method content

• Principle/ Scope

• Apparatus/Equipment

• Operating parameters

• Reagents/Standards

• Sample preparation

• Standards/Control Solution preparation

• Procedure

• System Suitability

• Calculations

• Data reporting

Page 6: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 6 © PharmOut 2015

New Guidance - Reagents / Standards

• Grade of chemical (e.g., USP/NF, American Chemical Society, High 157 Performance or Pressure Liquid Chromatography, or Gas 158 Chromatography and preservative-free)

• Source (e.g., USP reference standard, qualified in-house reference material, 160 WHO International Standard/Reference Material, CBER standard)

• Purity (for pure chemicals only), State (e.g., dried, undried), and concentration

• Potencies (where required by CFR, USP)

• Storage conditions

• Directions for safe use (as per current Safety Data Sheet)

• Validated or documented shelf life

Page 7: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 7 © PharmOut 2015

Goal

The aim of validating an analytical procedure is to demonstrate that it is ‘fit for purpose’.

Page 8: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 8 © PharmOut 2015

Types of methods

ICH splits into four :

Identification tests

Impurities - Quantitative (content of)

Impurities - Limit (control of)

Quantitative tests (API or component(s) in drug substance / product)

Page 9: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 10 © PharmOut 2015

Validation requirements for analytical assays: ICH Q2(R1)

Page 10: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 11 © PharmOut 2015

Validation requirements for analytical assays: ICH Q2(R1)

Page 11: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 12 © PharmOut 2015

Methods to be validated -Identification

Identification tests:

• identify the analyte

• compare test sample to a standard

• E.g. Spectrum, chemical reactivity, chromatograph etc.

Page 12: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 13 © PharmOut 2015

Methods to be validated - Impurities

Impurities:

• ensure impurities are as low as required

• quantitative test or limit test

• accurate reflection of the purity characteristics

• validation requirements differ between the two impurity tests

Page 13: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 14 © PharmOut 2015

Methods to be validated - assays

Assay:

• Quantitatively measure the major component(s) in the drug product

• Validation requirements similar for:

› Active ingredients

› Other selected component(s)

› Other analytical procedures (e.g. dissolution)

Page 14: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 15 © PharmOut 2015

Validation tests

• The objective of the analytical procedure must be understood to determine which of the validation characteristics will be evaluated.

• Consider the following characteristics:

Accuracy

Precision

Repeatability

Intermediate precision

Specificity

Detection Limit

Quantitation Limit

Linearity

Range

Page 15: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 16 © PharmOut 2015

Accuracy

• closeness of the value obtained to the reference (‘true’) value

• determined across the range of the analytical procedure

Page 16: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 17 © PharmOut 2015

Accuracy

There are several methods to determine accuracy. Examples:

Apply the procedure to an analyte of known purity (reference)

Apply the procedure to synthetic mixtures of the drug product components to which known quantities of the drug substance to be analysed have been added

Page 17: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 18 © PharmOut 2015

Precision

• The precision of a procedure is the closeness of a series of measurements from different sampling of the same homogenous sample.

• If a homogenous sample cannot be obtained, use artificially prepared samples or a sample solution.

Page 18: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 19 © PharmOut 2015

Accurate and precise

Page 19: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 20 © PharmOut 2015

Precision

There are three levels of precision:

RepeatabilityIntermediate

precisionReproducibility

Page 20: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 21 © PharmOut 2015

Repeatability

Repeatability uses the same operating conditions over a short interval of time, also called ‘intra-assay’ precision.

Same laboratory

One analyst

Same daySame

equipment

Page 21: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 22 © PharmOut 2015

Repeatability

Repeatability should be assessed using either:

• A minimum of 9 determinations covering the specified range (e.g. 3 concentrations/3 replicates)

• A minimum of 6 determinations at 100% of the test concentration

Page 22: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 23 © PharmOut 2015

Intermediate precision

Same laboratory

Multiple days

Multiple analysts

Different equipment

Intermediate precision assesses the variation withinlaboratories:

Page 23: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 24 © PharmOut 2015

Reproducibility

• assesses the precision between laboratories; it is an inter-laboratory trial

• should be considered when standardising methodology e.g. for inclusion in pharmacopoeias or collaborative studies

Page 24: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 25 © PharmOut 2015

Specificity

Specificity should be investigated for the validation of:

Ensure the identity of an analyteIdentification

tests

Demonstrate accurate content of impurities of an analytePurity tests

Provide accurate content or potency of the analyte in a given sample

Assay (content or potency)

Page 25: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 26 © PharmOut 2015

Specificity

• Identity, impurities and assay tests

• ability to detect the analyte in a sample in the presence of other components which may also be present.

• may not always be able to demonstrate that an analytical procedure is specific for a particular analyte (complete discrimination).

› Use two or more procedures to achieve the required discrimination.

Page 26: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 27 © PharmOut 2015

Specificity

• Should be able to discriminate between compounds of related structures which may be present.

• Use scientific judgement to choose potential interfering materials.

• Typically would confirm discrimination by obtaining:

• positive results from samples with the analyte

• negative results from samples without the analyte

Identification test

Page 27: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 28 © PharmOut 2015

Detection limit

• The lowest amount of analyte in a sample that can be detected is the ‘detection limit’.

• Not necessarily quantified as an exact value.

Page 28: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 29 © PharmOut 2015

Quantitation limit

• The quantitation limit is the lowest amount of analyte that can be quantitatively determined with acceptable precision and accuracy.

• The quantitation limit is typically required for impurities and/or degradation product determinations.

Page 29: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 30 © PharmOut 2015

Linearity

• Ability to obtain results that are directly proportional to the concentration of an analyte in the sample.

• A minimum of 5 concentrations should be used to establish linearity.

• May demonstrate linearity directly on the drug substance (dilution of a standard stock solution).

Page 30: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 31 © PharmOut 2015

Range

• determined by confirming that the procedure provides an acceptable degree of linearity, accuracy and precision.

• use amounts of analyte within or at the extremes of the specified range of the procedure being validated.

Page 31: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 32 © PharmOut 2015

Range

Consider the following minimum specified ranges:

80 – 120 percent of the test concentration

Assay for drug substance or

finished product

70 – 130 percent of the test concentration

Content uniformity

+/- 20 percent over the specified range

Dissolution testing

Page 32: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 33 © PharmOut 2015

Robustness

• ability to remain unaffected by small, deliberate variations in method parameters

• provides an indication of the reliability of the method during routine use

• usually only for method development phase; if not done then, it should be done during method validation

Page 33: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 34 © PharmOut 2015

Robustness

Examples of typical variations:

Stability of solutions

Temperature / humidity

MP flow rate, pH, composition

Different columns for chromatography tests (lots/suppliers)

Page 34: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 35 © PharmOut 2015

Validation requirements for analytical assays: ICH Q2(R1)

Page 35: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 36 © PharmOut 2015

System suitability testing (SST)

• based on the concept that the equipment, electronics, analytical operations and samples to be analysed constitute an integral system

• SST parameters for a particular procedure depend on the type of procedure being validated.

• Refer to Pharmacopoeias for more information.

Reference: ICH Q2(R1)

Page 36: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 38 © PharmOut 2015

Re-validation of methods

What situations will require re-validation of an analytical

test?

Page 37: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 39 © PharmOut 2015

Re-validation of methods

May be required in the following situations:

• Changes in the synthesis of the drug substance

• Finished product composition has changed

• Testing procedure has changed

Page 38: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 45 © PharmOut 2015

Method validation report

Complete a validation report once validation protocol has been executed.

The report should have enough detail to be reviewed without the need for raw data.

Must be linked to the protocol (reference protocol number).

Report should have clear conclusions supported by data.

Deviations should be reported and explained.

Page 39: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 46 © PharmOut 2015

Sample verification report

Page 40: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 47 © PharmOut 2015

Summary

The aim of validating an analytical procedure is to demonstrate that it is ‘fit for purpose’.

The objective of the analytical procedure must be understood to determine which of the validation characteristics need to be evaluated.

Method validation must be documented, reviewed and approved.

Page 41: Analytical method validation · • FDA Guidance for Industry - Q2B Validation of Analytical Procedures: Methodology • FDA CDER reviewer guideline for validation of chromatographic

Slide 48 © PharmOut 2015

Thank you for your time.Questions?

Debbie Parker

[email protected]

Senior Consultant

www.pharmout.net