Upload
others
View
2
Download
0
Embed Size (px)
Citation preview
Environmental Chemistry and Analysis Prof. M.S.Subramanian
Indian Institute of Technology Madras
MODULE 2.3
Analysis of trace pollutants in water BioConcentration 1 Accumulation in sediments 2 Biomagnification 2 Degradation 3 Gas liquid Chromatography (GC) 7 Mobile phase 8
Sample injection 8
Column and stationary phase 9
Detectors 11 Extraction procedures or sample preparation 12 Solvent extraction 13
Head space analysis 14
Purge and trap 14
Solid phase extraction (SPE) 14
Analysis of Individual Trace Pollutants 15 Analysis of DDT 16
Alumina-silver nitrate column 16
Silica gel column 17
Technique for measuring total concentration of group of Compounds
19
Phenols 19
Surfactants 20
Total hydrocarbons 20
Environmental Chemistry and Analysis Prof. M.S.Subramanian
Indian Institute of Technology Madras
MODULE 2.3
Analysis of trace pollutants in water This chapter gives the outline of the analysis of organic compounds and
inorganic metal ions present in water at trace levels (i.e at µg l-1). The need to
analyse these species at such a low concentration of µg l-1 studies from the fact
that they tend to bioaccumulate (especially the neutral organic compounds and
metal ions) and are found in organisms at concentration exceeding background
levels by many factors of ten. Analysis of species at such low concentration
needs a preconcentration step so that they are amenable for instrumental
analysis.
BioConcentration:
Bioconcentration is a widely recognised phenomenon whereby
organisms accumulate particular component chemicals in their tissues. The
uptake of these chemicals occurs via a variety of mechanisms depending on the
organism and the chemical. Micro organisms, aquatic plants, fish, and other
animals living in water are capable of accumulating soluble metals and organic
chemicals. Organic compoundswhich do not contain polar groups such as OH or
NH2 , or which are not ionic, will have low solubility in water. Among the same
group of compounds the solubility generally decreases with increasing molecular
mass.As the solubility in water decreases, the solubility in organic solvent
increases.
Since fatty tissues are organophilic, the lower solubility organic
compounds tend to accumulate in these tissues, particularly that is found in
organs in closest contact with aqueous fluids e.g. kidneys. The tendency to
accumulate in fatty tissues is greater, the greater the molecular mass of
1
Environmental Chemistry and Analysis Prof. M.S.Subramanian
Indian Institute of Technology Madras
compound. The ability to accumulate in an organism is measured by the term
bioconcentration factor which is given by the relation:
concentration of a compound in an organismbioconcentration factorconcentration in surrounding water
=
Accumulation in sediments:
As discussed earlier the species which have low solubility in water tend to
adhere to any available solid; the larger the surface area greater will be its ability
to adsorb the compound. Such kind of material is found in sediments which are in
abundnce in estuaries where there are often discharges from major industries.
Organisms such as mussels and scallops which feed by filtration of sediments
ingest these organics.
Biomagnification:
The tendency for a pollutant which is present at very low concentrations in
a water body to get accumulated to an unacceptable level in a food chain is
explained in fig.1 with pesticide as an example.
As we go up the food chain it can be seen that the concentration of
pesticide from lake water which was only at 0.02 mg.kg-1 has been increased to
1600 mg kg-1 in birds where one species is dependent for survival on the
consumption of previous species.
If a pollutant is present in the first organism, then as we proceed down the
food chain there will be increase in concentration in each subsequent species.
This is illustrated in fig 1
2
Environmental Chemistry and Analysis Prof. M.S.Subramanian
Indian Institute of Technology Madras
Lake water ................................. 0.02
Concentration of pesticide (DDT) (mg kg-1)
Plankton ................................... 5.0
Non-predatory fish ................................... 40-100 (fatty tissue)
Predatory fish
..................... 80-2500 (fatty tissue)
Birds feedingon fish
..................................... 1600 (fatty tissue)
Fig.1 A typical Food chain
(Redrawn by permission of John Wiley & Sons Ltd, Chichester, England, 779777 from
Environmental analysis by Roger N. Reeve and John D. Barnes, p.25,1994)
Degradation:
The accumulation of the compound in the fatty tissue also depends on the
rate at which it is metabolised to produce a water soluble product which can be
excreted. The rate of metabolism of a compound depends on the structure of the
molecule and it has been found that the rate is slow especially with chlorinated
compounds.
3
Environmental Chemistry and Analysis Prof. M.S.Subramanian
Indian Institute of Technology Madras
If we take as an example p, p'-DDT, the metabolism of this compound
occurs in two stages as shown in fig.2.
H
ccl2
cl cl clclc
ccl3
c
p,p'-DDEp-p' DDT
slow
cl c
H
cl
cooH
p,p'-DDA
(water solubility increased by the presence of cooH group)
Fast
Fig 2 Metabolism of p,p’-DDT
(Redrawn by permission of John Wiley & Sons Ltd, Chichester, England, 779777 from
Environmental analysis by Roger N. Reeve and John D. Barnes, p.26,1994)
The first stage is rapid and normally takes only a few days for completion
but the second stage is extremely slow, often taking many months in some
species. It is because of this reason ,the first degradation product is often found
as the predominant species in environmental samples.
4
Environmental Chemistry and Analysis Prof. M.S.Subramanian
Indian Institute of Technology Madras
The o,p'-isomer of DDT is present only as a minor component in
commercial DDT sample which however is metabolised rapidly by the reaction
shown in Fig.3.
H
cl cl clOHc
ccl3
c
o,p' DDT
Fast
ccl3
Hcl
(Water solubility increasedby the presence of OH group)
Fig 3. Metabolism of o, p’-DDT
(Redrawn by permission of John Wiley & Sons Ltd, Chichester, England, 779777 from
Environmental analysis by Roger N. Reeve and John D. Barnes, p.26,1994)
In addition to DDT, other chlorine containing pesticides, polychlorinated
biphenyls and dioxins which are bioaccumulative compounds, there are a
number of non-bioaccumulative organic compounds which are suspected to be
carcinogens. A typical example is chloroform, which can be produced in trace
quantities during the disinfection of water by chlorination and which is thought to
be harmful at µg l -1concentrations.
Analysis of such compounds at such low levels requires pretreatment
procedure. These pretreatment procedures are necessary to remove potential
interferences and in many instances to increase the analyte concentration to
within the instrument sensitivity.
5
Environmental Chemistry and Analysis Prof. M.S.Subramanian
Indian Institute of Technology Madras
The preconcentrations methods will be followed by chromatographic
separation of the components. As most organic compounds have significant
volatility even at room temperature, gas chromatography would be expected to
be a useful technique. The alternative of high performance liquid chromatography
is used only where there are advantages over established gas chromatographic
methods.
In the determination of groups of compounds such as phenols, and also of
classes of detergents, where the total concentration of the group of substances is
required rather than the concentration of individual compounds
nonchromatographic methods are used. Ultraviolet / visible absorption
spectrometric techniques can be used to determine the total concentration of
related classes of organic compounds.
A detailed description of gas chromatophic method is given in this chapter.
The liquid chromatographic methods, flame spectroscopic methods, ICP AES,
anodic stripping voltammetry amd chelation ion chromatography are discussed in
the next chapter.
(i) Gas liquid chromatographic methods:
The following precautions have to be taken in the storage and the subsequent
analysis of the organic trace pollutants:
a) To minimise the volalitisation of the organic sample (e.g pesticide) the sample
should be stored at 4oC, if the analysis is not performed immediately.
(b) Since the organic compounds are tend to undergo microbial degradation, it is
6
Environmental Chemistry and Analysis Prof. M.S.Subramanian
Indian Institute of Technology Madras
better to store these compounds below 0oC in deep freezer to minimise the
microbial activity.
(c) Plastic sample containers should not be used for storing these compounds
because of the possibility of leaching of potential monomers and additives into
the samples and hence only glass containers should be used.
(d) The samples should be analysed as soon as possible after collection to
minimise the loss of the analyte by adsorption on the walls of the container.
The volume of the sample required for injection into a chromatograph is of the
order of a few microlitres and that for spectrophotometric detection a few mililitres
and this solution might have been obtained through a preconcentration step from
several litres of the original sample. Precaution has to be taken to avoid
contamination or loss of material at these low concentration during the
subsequent analysis.
Gas liquid Chromatography (GC):
Gas chromatography(GC) is a separation technique where volatile, thermally
stable solutes migrate through the column containing a stationary phase (which
is a liquid adsorbed on or chemically bonded to a solid) at rates dependent on
their distribution ratios. These are inversely proportional to their volatilities, which
in turn are determined by their partial vapour pressures and hence their boiling
points. Solutes are therefore generally eluted in order of increasing boiling
point,except where there are specific interactions with the stationary phase.
Usually an elevated temperature in the range of 50-350oC is normally employed
to ensure that the solutes have adequate volatility. Gas liquid chromatography
has high separation efficiencies and can be interfaced with highly specific
7
Environmental Chemistry and Analysis Prof. M.S.Subramanian
Indian Institute of Technology Madras
sensitive detectors. The main components of a gas liquid chromatograph is
shown in fig.4.
In gas liquid chromatography, the mobile phase is a gas and the stationary
phase is a liquid adsorbed on or chemically bonded to a solid. In the liquid
stationary phase the solutes can dissolve and the sorption process is called
partition. Specific interactions of the solute in the stationary phase may alter the
order of elution from that of increasing boiling points. Gas liquid chromatography
has high separation efficiencies and can be interfaced with highly specific and
sensitive detectors.The main components of a gas chromatograph is shown in
fig.4.
Computer or recorder
Column oven
Detectoroven
column
Exit
Carrier gas
Injectoroven
Siliconerubberseptum
Injectorport
Detector
Fig 4 Components of gas liquid chromotograph
8
Environmental Chemistry and Analysis Prof. M.S.Subramanian
Indian Institute of Technology Madras
Mobile phase:
The mobile phase is known as the carrier-gas because its sole purpose is to
transport solutes through the column, thus not contributing to chromatographic
selectivity. Commonly used carrier gases are nitrogen for packed columns and
hydrogen or helium for capillary columns.
Sample injection:
Small volumes of liquids or solutions (0.1-10 µl) are generally injected into a
heated injection port, through which the carrier gas continuously flows, from a
calibrated microsyringe used to pierce a self-sealing silicone-rubber septum. Split
and splitless modes of injection are used. In the latter the whole injection passes
straight to the column whereas in the former the sample is split in the ratios of
50:1 and upto 500:1 with a large part vented waste. This avoids overloading the
column and is especially an important facility where capillary column are used.
Column and stationary phase:
The column is the central component of a gas chromatograph since it is
where the separation occurs. There are two types of GC columns and they are
capillary (open tubular) and packed columns. Capillary columns consist of long
narrow bore, high-purity quartz tabing with a thin layer of liquid or stationary
phase coated or chemically bonded to the surface of inner wall. Very long lengths
(up to 100 m) can be used resulting in high efficiencies for the separation of
complex mixtures. Capillary columns are available with a range of internal
diameters and thicknesses of the stationary phase. The narrower the bore and
the thinner the coating, the greater the efficiency, but the lower the sample
capacity. Packed columns are much shorter than the capillary columns, rarely
9
Environmental Chemistry and Analysis Prof. M.S.Subramanian
Indian Institute of Technology Madras
exceeding 2m in length. The stainless steel or glass tubing has an internal
diameter of 2 or 3mm and is filled with a granular material that acts a support for
a thin coating of a liquid stationary phase for GLC . Alternatively the liquid
stationary phase can be bound to the solid support through a chemical reaction.
Solid supports are inert, porous silaceous materials with a large surface area.
The particle sizes vary between 0.125 mm to 0.25 mm. The smaller the particle
size and the thinner the coating of stationary phase, the lesser is the band
spread.
Given below are the three types of columns with decreasing separation
efficiency (fig.5)
• Narrow-bore capillary columns:
Typical dimensions 30 - 60m length, 0.2 mm i.d, flow rate
0.4 ml min-1 of He
• Wide -bore capillary columns:
Typical dimensions 15 - 30m length, 0.53mm i.d., flow rate
2.5ml min-1 of He
• Packed columns:
Typical dimensions 2m length, 2mm i.d., flow rate
20ml min-1of He
Fig 5. Typical dimensions of capillary and packed columns
10
Environmental Chemistry and Analysis Prof. M.S.Subramanian
Indian Institute of Technology Madras
Many of the organic compounds of environmental interest are of high
relative molecular mass and have low volatility. High oven temperatures are
necessary for these and consequently silicone polymers are often the favored
stationary phases. Best separation efficiencies are achieved when the stationary
phase has a similar polarity to the components of analyte. Fuel oils are separated
on non-polar columns (e.g .dimethylsilicone), pesticides and chlorinated solvents
are often separated on medium-polarity columns
(e.g. diphenyl / dimethylsilicone), whereas 2, 3, 7, 8- tetrachlorodibenzo-p-dioxin
can be separated from its isomers using highly polar columns (e.g.
cyanopropylsilicone).
In order to confirm whether peak observed in a chromatogram is due to a
single component rather than two, the chromatogram should be produced on two
columns with different polarities. Under such circumstance, it would be unlikely
that the peaks would remain unresolved on both column. Capillary columns
because of their greater separation efficiency decreases the probability
unresolved peaks.
Detectors:
Depending upon the characteristics of the analyte and the circumstances of
its determination different detectors are used in GC. The one with the most
comprehensive response is based on the change in thermal conductivity in the
mobile phase as a component is eluted. Principally this results from the mass
difference between the molecules eluted and those of the carrier gas. The
thermal conductivity detector is robust and reliable, but has only moderate
sensitivity and a limited dynamic range. The preferred option for general purpose
GC use, is flame ionisation detector (FID). The eluate stream is mixed with air
and hydrogen allowing combustion of most components. Ionic species produced
11
Environmental Chemistry and Analysis Prof. M.S.Subramanian
Indian Institute of Technology Madras
in the combustion facilitate the passage of an electronic current between two
electrodes in contact with flame. The magnitude of current reflects the number of
ions and thus the amount of the eluted compound present. There are some
selective detectors which can detect only certain compounds, yielding a simple
chromatogram. The ideal situation for the determination is to have a detector that
can detect only this type of analyte. One such, is the electron capture detector
(ECD).
High energy electrons (b-particles) from a radioactive source such as 63 Ni or 3H, are used to ionise the carrier gas stream. As a result, in the absence
of an eluted component, a steady electronic current will flow between the
electrodes. An eluting component which is electronegative will 'capture' electrons
and reduce the current flowing, leading to a peak in the chromatogram. ECDs
are very sensitive detectors and have been extensively employed in the
determination of chlorinated pesticides. Another selective detector mentioned is
flame photometric detectors (FPD). It selectively monitors radiation emitted by
the species such as S2 and HPO produced in the combustion of sulphur or
phosphorous containing compounds.
Nowadays the interfacing of GC with dedicated mass spectrometers is
quite common. Not only the mass spectrometer can function as aid to peak
identification but also be adjusted for selective response to a single type of ion
originating in one component of the mixture.
Extraction procedures or sample preparation:
Although gas chromatography is primarily a technique for analysing
mixtures of volatile solutes, there are many compounds and materials that are
either nonvolatile, have volatile components in a non-volatile matrix or thermally
12
Environmental Chemistry and Analysis Prof. M.S.Subramanian
Indian Institute of Technology Madras
labile. Direct injection of the sample would deposit nonvolatile solids on the
column which would cause blockage and shorter the column lifetime. In some
instances, samples may contain solutes at such low levels that they must be
preconcentrated prior to analysis.
Special extraction methods have been developed for handling such
samples and they are
(a) solvent extraction
(b) head space analysis
(c) purge and trap technique
(d) solid phase extraction
(a) Solvent extraction:
Extraction of a sample component into a small volume of organic solvent
can be highly selective and may serve to both concentrate and enhance
technique sensitivity. The water sample is shaken with an immiscible organic
solvent in which the analyte species are soluble. The organic layer is separated,
and after drying, is injected into the chromatograph. The extractions can be made
selective by adjusting the pH of the aqueous layer. For extracting selectively
acidic components the sample solution should be made acidic. Similarly for
extractive selectively basic components, the sample solution should be made
basic.
While choosing the solvent for extraction, the compatibility with the
detector also should be taken into account. For example, chlorinated solvents
13
Environmental Chemistry and Analysis Prof. M.S.Subramanian
Indian Institute of Technology Madras
should not be used for extraction if electron-capture detector is used. For that
hexane or light petroleum will be suitable.
(b) Head space analysis: This involves chromatographing the vapours
derived from a sample by warming it in a partially filled vial sealed with septum
cap. After equilibration under controlled conditions, the proportions of volatile
sample components in the head space above the sample are representative of
those in the bulk sample. The head space vapours, which are under slight
positive pressure, are sampled by a modified automated injection system or gas
syringe, and injected on to the column. The procedure is useful for mixtures of
volatile and nonvolatile components, such as residual monomers in polymers,
alcohol or solvents in blood samples, and flavours and perfumes in manufactured
products, as it simplifies the chromatograms and protects the column from
contamination by non volatile substances.
(c)Purge and trap:
This is a procedure where volatile solutes can be collected on a solid
sorbent in a preconcentration step using a purge gas stream. The organics are
collected in a short tube of absorbent material such as activated charcoal, or a
porous polymer (e.g. Tenax). After the collection period, the tube is flash heated
to release the organics into the chromatograph.
(d) Solid phase extraction (SPE):
It is a relatively new techique and is used as a primary means of
sample pretreatment. The removal of interfering matrix components in general
and the preconcentration of trace and ultratrace levels of analytes are other
important uses of SPE. It is versatile, rapid and requires only small volumes of
14
Environmental Chemistry and Analysis Prof. M.S.Subramanian
Indian Institute of Technology Madras
solvents, or none at all in the case of solid phase micro extraction. SPE has been
applied in the determination of pesticides and herbicides in polluted surface
water and soils, polycyclic aromatic hydrocarbons (PAHs) in drinking water,
polluted industrial and urban atmospheres, and drugs in biological fluids.
Most of the SPE is carried out using a small bed of sorbent
(25-500 mg) contained in a cartridge made from a polypropylene syringe barrel,
the sorbent being retained in position by polyethylene fritted disks. A typical
cartridge procedure consists of four distinct steps. They are
(1) Solvent conditioning which involves wetting the surface of the sorbent and to
create the same pH and solvent composition as those of the sample. (2) sample
loading in which the sample solution is passed through the cartridge with the
object of either retaining the analytes of interest whilst the matrix components
pass through or retaining the matrix components whilst the analytes pass
through.(3) Rinsing is necessary to remove all the components which are not
retained by the sorbent. (4) Elution is the final step to recover retained analytes.
A further development is the use of extraction discs where the
absorbent material is held with in a filter disc. The sample is passed through the
disc by suction and the solutes get extracted into the disc. By using a suitable
solvent the extracted components are eluted.
Analysis of Individual Trace Pollutants
In the analysis of individual components in a mixture which are present at
low concentrations, simple extraction of the analyte and injection of the extract
into the chromatograph is not sufficient. Further treatment of the sample is
necessary and it involves (1)extraction (2) clean-up to remove interfering
components (3) concentration of the extract.
15
Environmental Chemistry and Analysis Prof. M.S.Subramanian
Indian Institute of Technology Madras
To illustrate the method the analytical scheme for the commercial
pesticide DDT is discussed in detail as an example.
Analysis of DDT:
DDT is now considered as a contaminant. Commercially available DDT
insecticide is not a single compound, the major active component (p-p'-DDT) only
being 70 - 80% of the total content. In addition it also contains (15-20%) o,p'-DDT
one of the minor components, p, p'-DDD (similar in structure to p,p'-DDT with a
CHCl2 side chain rather than CCl3) (1-4%) is, in fact more toxic to insects than
p,p'-DDT. Hence three peaks are expected in a chromatograph.
If environmental samples are considered a number of decomposition and
metabolic products will also be present. In fact for many samples, the highest
concentration component is not p, p' DDT but its primary metabolic product,
DDE. Thus after considering the environmental DDT samples, a total of atleast 5
peaks are expected including the peaks for the two decomposition products p,p'
DDE(aerobic decomposition) and p,p' DDD (anaerobic decomposition). In
addition the interfering components in the sample such as other pesticides and
polychlorinated biphenyls may be present.
For the analysis of individual DDT components, in such a sample the
following pretreatment is given before injecting the sample into GLC.
1. The organic components are extracted into hexane (A 40 fold increase in
concentration is achieved by extracting 2litre of the sample into 50ml solvent).
2. The hexane extract is dried using 5g sodium sulphate.
3. The sample is further concentrated to 1ml by partial evaporation on a steam
bath followed by bubling nitrogen.
16
Environmental Chemistry and Analysis Prof. M.S.Subramanian
Indian Institute of Technology Madras
The dried sample is placed on alumina-silver nitrate column.
(a) Alumina-silver nitrate column
In this column polar components in the extract are retained. The silver
nitrate helps to retain compounds containing unsaturated carbon-carbon bonds.
Hence non polar materials including DDT components are eluted using 30ml of
hexane. The hexane extract is again concentrated to 1ml and placed on the top
of silica gel column.
(b) Silica gel column
This is a less polar column and therefore can be used to separate
potential non-polar interferences from the sample. By passing hexane (10ml)
through the column, polychlorinated biphenyls (PCBS) are eluted out, but the
DDT Components are retained in the column. DDT is then eluted with a more
polar solvent mixture (12ml of 10% diethyl ether in hexane).
17
Environmental Chemistry and Analysis Prof. M.S.Subramanian
Indian Institute of Technology Madras
0 30 40 50
p,p'-DDE
o,p'-DDT
p,p'-DDT
Time (min)
25m x 0.32mm id methyl silicone capillary column with a temperature gradient to 220oC.
Fig 6 Chromatographic separation of DDT components.
(Redrawn by permission of John Wiley & Sons Ltd, Chichester, England, 779777 from
Environmental analysis by Roger N. Reeve and John D. Barnes, p.108,1994)
The eluates are again concentrated to 1ml before injection into the
chromatograph. A typical chromatogram is shown in fig 6. The detection limit for
each component is approximately 10ngl-1.
18
Environmental Chemistry and Analysis Prof. M.S.Subramanian
Indian Institute of Technology Madras
Technique for measuring total concentration of group of compounds:
This technique includes the measurement of total phenols, surfactants
(total, anionic, cationic and non-ionic surfactants), and total hydrocarbons. Visible
spectrophotometry is often used for phenols and surfactant analysis after the
formation of derivatives. IR absorption technique is used for the total
hydrocarbon concentration.
Phenols:
Phenol analysis is important in water pollution studies because of the
objectionable taste of chlorinated phenols in chlorinated drinking water containing
as little as 1 ppb. The effluents from coke-oven and coal distillation plants are the
major source for phenol pollution. The determination first involves the distillation
of phenols and other phenolic compounds from waste water. Then the distilled
phenol is allowed to react with 4-aminoantipyrene at pH10 in the presence of
K3Fe(CN)6. A red dye is formed which is extractable into chloroform. The
absorbance is measured at 460nm.
N + OH NOH 3C OH
The method is quite sensitive and has a detection limit of about 1ppb of phenol.
N 3C N
NH 2 H 3 C H N O3C
Red dye
19
Environmental Chemistry and Analysis Prof. M.S.Subramanian
Indian Institute of Technology Madras
Surfactants:
Similarly anionic surfactants can be determined by spectrophotometry
using methylene blue method. Under basic conditions a salt is formed between
methylene blue and the surfactant and this salt can be extracted into chloroform.
The absorbance of the extract can be measured at 652 nm and the concentration
can be determined by comparison with a calibration graph.
Total hydrocarbons:
For the determination of total hydrocarbon content, the hydrocarbons are
extracted from acidified water using a non-hydrocarbon solvent such as
carbontetrachloride and the (IR) absorption is measured at 2920 cm-1,
corresponding to C-H stretching frequency.
20