8
Analysis of High- and Low-Dosed Vitamins in a Single Run Using the Agilent 1200 Infinity Series High Dynamic Range Diode Array Detector Solution Application Note Food Testing & Agriculture Abstract The Agilent 1200 Infinity Series High Dynamic Range Diode Array Detector (HDR DAD) solution expands the linear dynamic range by a factor of 30. By combining the signals from two diode array detectors with different path length Agilent Max-Light flow cells, the 1200 Infinity Series HDR DAD solution facilitates detection and quantification of main and trace components in a single run without exceeding the linear range of the HDR DAD solution. This Application Note analyzed water soluble vitamins from vitamin tablets. All vitamins were separated from each other and from matrix peaks. High-dosed vitamin C and low-dosed vitamin B12 were quantified in a single run. Author A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany min 10 10.2 10.4 10.6 10.8 11 11.2 11.4 mAU/cm 0 0.025 0.05 0.075 0.1 0.125 0.15 0.175 0.2

Analysis of High- and Low-Dosed Vitamins in a Single Run ... · 220 240 260 280 300 320 340 360 380 nm mAU 0 50 100 150 200 250 220 240 260 280 300 320 340 360 380 nm mAU 0 50 100

Embed Size (px)

Citation preview

Page 1: Analysis of High- and Low-Dosed Vitamins in a Single Run ... · 220 240 260 280 300 320 340 360 380 nm mAU 0 50 100 150 200 250 220 240 260 280 300 320 340 360 380 nm mAU 0 50 100

Analysis of High- and Low-Dosed Vitamins in a Single Run Using the Agilent 1200 Infi nity Series High Dynamic Range Diode Array Detector SolutionApplication Note

Food Testing & Agriculture

AbstractThe Agilent 1200 Infi nity Series High Dynamic Range Diode Array Detector (HDR DAD) solution expands the linear dynamic range by a factor of 30. By combining the signals from two diode array detectors with different path length Agilent Max-Light fl ow cells, the 1200 Infi nity Series HDR DAD solution facilitates detection and quantifi cation of main and trace components in a single run without exceeding the linear range of the HDR DAD solution. This Application Note analyzed water soluble vitamins from vitamin tablets. All vitamins were separated from each other and from matrix peaks. High-dosed vitamin C and low-dosed vitamin B12 were quantifi ed in a single run.

AuthorA.G.HuesgenAgilent Technologies, Inc.Waldbronn, Germany

min10 10.2 10.4 10.6 10.8 11 11.2 11.4

mAU/cm

00.0250.05

0.0750.1

0.1250.15

0.1750.2

Page 2: Analysis of High- and Low-Dosed Vitamins in a Single Run ... · 220 240 260 280 300 320 340 360 380 nm mAU 0 50 100 150 200 250 220 240 260 280 300 320 340 360 380 nm mAU 0 50 100

2

IntroductionWhen using conventional diode array detectors in HPLC, the analysis of main and trace components often requires two separate analyses to be able to quantify all components. This challenge can be solved by using the 1200 Infi nity HDR DAD solution. The detector design facilitates the analysis of main and trace components simultaneously in a single run. This is possible due to the enhanced dynamic range for main compounds and the improved signal-to-noise (S/N) ratios for measured trace components using the HDR DAD solution.

In vitamin tablets, some vitamins are present at high concentrations, whereas, for example, vitamin B12 is present in very low concentrations. In the literature for vitamin B12, different sample preparation procedures and different chromatographic methods are used to be able to detect and quantify vitamin B12 in vitamin preparations1,2. Until now, it was not possible, by applying one sample preparation and one chromatographic method, to detect and quantify all water soluble vitamins present in a vitamin tablet. The combination of an Agilent 1200 Infi nity LC with the 1200 Infi nity Series HDR DAD solution offers a perfect solution.

ExperimentalThe following instruments, conditions, and compounds were used.

Instrument Part no.Agilent 1260 Infi nity Diode Array Detector 2 G4212B with a 3.7-mm cellAgilent 1260 Infi nity Diode Array Detector 1 G4212B with a 60-mm cell or G4212B with a 10-mm cellAgilent 1260 Infi nity Column Compartment G1316C Agilent 1260 Infi nity Standard Autosampler G1329B Agilent 1260 Infi nity ALS Cooler G1330B Agilent 1260 Infi nity Binary Pump G1312B

Instrumentation

Compounds analyzed

Vitamin C Pantothenic acid Vitamin B6

Vitamin B12

R = 5’deoxyadenosyl, ME, OH, CN

Vitamin B2 Vitamin B1Niacinamid

HOHO

HO OH

HO

OH

OH

OH

H

O

O

CH3

CH3

OH

HO

HOO

O

O

NH2

OH

+

NH2

H3C H

3C

OH

H

O

O

CH2

H3C

H3C

OHHOHHOHH

O

H

R

Co+

HO

HO

O

O–

H2NOC

H2NOC

H2NOC

CONH2

CONH2

CONH2

Page 3: Analysis of High- and Low-Dosed Vitamins in a Single Run ... · 220 240 260 280 300 320 340 360 380 nm mAU 0 50 100 150 200 250 220 240 260 280 300 320 340 360 380 nm mAU 0 50 100

3

Sample preparation• Two vitamin effervescent tablets

were placed in 20 mL of water and 20 mL of ethanol. Ethanol is the optimum solvent for niacinamid and vitamin B12.

• The solution was stirred for 10 minutes in the dark.

• A 20-mL aliquot was fi ltered using two Agilent Captiva Premium Syringe Filters, regenerated cellulose, p/n 5190-5111.

• 1.5 mL was transferred into an autosampler vial and the sequence was started immediately.

Concentration and injected amountCompound Concentration Injected amountVitamin C 60 mg/tablet 6,000 ngNiacinamid 18 mg/tablet 1,800 ngPantothenic acid 6 mg/tablet 600 ngVitamin B6 2 mg/tablet 200 ngVitamin B2 1.6 mg/tablet 160 ngVitamin B1 1.4 mg/tablet 140 ngVitamin B12 2 µg/tablet 0.2 ng

Parameter ValueColumn Agilent ZORBAX RRHT Eclipse plus C18, 4.6 × 150 mm, 1.8 µm

(p/n 959994-902)Mobile phases A) water + 4 g/L K2HPO4 + 0.5 g/L hexanesulfonate, pH 3 with phosphoric acid,

B) methanolGradient at 0 minutes 3 % B,

at 1 minute 3 % B, at 10 minutes 34 % B, at 11 minutes 50 %

Flow 1.2 mL/minStop time 14 minutesPost time 5 minutes Injection volume 2 µL, sample temperature was kept at 4 °CUV 220/10 nm for pantothenic acid and vitamin B6,

265/10 nm for vitamin C, niacinamid, and vitamin B1, 360/10 nm for vitamin B12 and B2, Ref: off, 20 Hz

Column temperature 40 °C

Chromatographic conditions

Software Agilent OpenLAB CDS ChemStation Edition, revision C.01.05

Page 4: Analysis of High- and Low-Dosed Vitamins in a Single Run ... · 220 240 260 280 300 320 340 360 380 nm mAU 0 50 100 150 200 250 220 240 260 280 300 320 340 360 380 nm mAU 0 50 100

4

Principle and functionality of the 1200 Infi nity Series HDR DAD solutionThe 1200 Infi nity Series HDR DAD solution expands the linear dynamic range by a factor of more than 30. By combining the signals from two diode array detectors with different pathlength Max-Light fl ow cells, the HDR DAD solution enables detection and quantifi cation of components with signifi cantly different concentrations in a single run. The HDR DAD solution clusters two 1260 Infi nity or 1290 Infi nity DADs together, see Figure 2. Detector 1 is equipped with a 60-mm path length cell for analyzing low concentration components, and Detector 2 is equipped with a 3.7-mm path length cell for analyzing high concentration compounds. The 60-mm cell must be installed in the fi rst detector and the 3.7-mm cell in the second detector. The resulting HDR DAD signal is one combined signal, normalized to 10-mm path length. The HDR DAD linear range is typically as wide as 0.6 × 10–6 to 6.7 AU/cm. A conventional 1200 Infi nity Series Diode Array Detector has a maximum linear range of 7 × 10–6 to 2 AU/cm. For detection of trace components, the HDR DAD signal is based on the signal acquired by the 60-mm cell. The 3.7-mm cell is used to provide the HDR DAD signal for the main component, which is typically out of the linear range of the 60-mm cell. For peaks between trace and main component absorbance range, a combination of both signals is used combined by a weighting function.

Figure 2. Agilent 1200 Infi nity Series HDR DAD solution with two clustered detectors.

Detector 1 with a 60-mm path length flow cell

Detector 2 with a 3.7-mm path length flow cell

Figure 1. Compound spectra.

Vitamin C Niacinamid

Pantothenic acidVitamin B6

Vitamin B1

Vitamin B12Vitamin B2

nm220 240 260 280 300 320 340 360 380

mAU

0

50

100

150

200

250

nm220 240 260 280 300 320 340 360 380

mAU

050

100150200250300

nm220 240 260 280 300 320 340 360 380

mAU

020406080

100120140

nm220 240 260 280 300 320 340 360 380

mAU

0100200300400500600

nm220 240 260 280 300 320 340 360 380

mAU

0102030405060

nm220 240 260 280 300 320 340 360 380

mAU

05

1015202530

nm220 240 260 280 300 320 340 360 380

mAU

050

100150200250300

very low UV response

265 nmselected

265 nmselected

265 nmselected

220 nm selected

360 nm selected

220 nm selected

360 nm selected

Page 5: Analysis of High- and Low-Dosed Vitamins in a Single Run ... · 220 240 260 280 300 320 340 360 380 nm mAU 0 50 100 150 200 250 220 240 260 280 300 320 340 360 380 nm mAU 0 50 100

5

The high dynamic range (HDR) tool is confi gured during instrument confi guration. Both detectors are clustered and the delay volume of the capillary connecting both detectors is fi lled in. In the user interface, both detectors appear as one detector, see Figure 3.

Results and Discussion Optimizing the separationIn vitamin tablets, the concentration of vitamin B12 is very low, approximately 2 to 5 µg/tablet. Therefore, selective detection is mandatory. Vitamin B12 is detected with high selectivity at 360 nm. Further coelution of vitamin B12 with matrix compounds must be avoided. In addition, vitamin B12 tends to decompose if exposed to light or oxygen, which causes additional problems during data evaluation.

Figure 3. Confi guring the Agilent 1200 Infi nity Series HDR DAD tool in the Agilent ChemStation.

Clustered detectors

Other vitamins, for example vitamin C, are present in 30,000 times higher concentrations than vitamin B12. Vitamin C is a very polar compound and tends to elute with low retention on reversed phase material. This creates a need to start the gradient at a very low organic percentage.

Pantothenic acid shows very low UV absorbance, see Figure 1. Even though it was present at 18 mg/tablet, the peak height only showed approximately 15 mAU at 220 nm. Good separation from other vitamins and matrix peaks was also mandatory.

Several experiments with different columns, mobile phases, gradients, and different pHs were evaluated. The best separation from each other and from matrix peaks was achieved by using the column, mobile phase, and gradient as described under Chromatographic conditions.

The ion pairing reagent was needed to obtain good peak shape and more retention on the reversed phase material. Methanol gave better separation than acetonitrile. To elute vitamin B12 earlier, the organic percentage was increased to 50 % after 11 minutes.

Page 6: Analysis of High- and Low-Dosed Vitamins in a Single Run ... · 220 240 260 280 300 320 340 360 380 nm mAU 0 50 100 150 200 250 220 240 260 280 300 320 340 360 380 nm mAU 0 50 100

6

Analysis of vitamins with different detector confi gurationsTo prove that using the Infi nity HDR-DAD, vitamin B12, and, for example, vitamin C can be quantifi ed in one run, three different detector confi gurations were used.

Analysis of standard and sample with:

• Agilent 1260 Infi nity DAD with 10-mm cell

• Agilent 1260 Infi nity DAD with 60-mm cell

• Agilent 1260 Infi nity HDR DAD with 3.7 and 60-mm cell

Analysis on a 1260 Infi nity DAD with a 10-mm path length cellQuantitation of pantothenic acid and B6 at 220 nm, quantitation of vitamin C, niacinamid, and B1 at 265 nm, and quantitation of B2 at 360 nm was possible with excellent precision for retention times and areas. The precision of retention times was between 0.0112 % RSD for vitamin C, and 0.033 % RSD for vitamin B1. The area precision was between 0.29 % RSD and 0.46 % RSD.

No quantitation of B12 at 360 nm was feasible. With an S/N ratio of 4.4, which is very close to the detection limit with S/N = 3, reliable quantitation is very diffi cult. The S/N ratio should be at least 10 for quantitation. For better overview, the results are summarized in Table 1.

mAU mAU0.60.50.40.30.20.1

0

> 2,000 mAU linearity limit

Peak height = 0.145998 mAUPeak-to-peak noise = 0.03282 mAUS/N = 4.4

Red: 220 nmBlue: 265 nmGreen: 360 nm2,000

Vita

min

C

Niac

inam

id

Pant

othe

nic a

cid

Vita

min

B6

Vita

min

B1

Vita

min

B2

Vita

min

B12

1,500

1,000

500

02

9.5 10 10.5 11 11.5 12

4 6 8 10 12 min

min

Figure 4. Analysis of water soluble vitamins with an Agilent 1260 Infi nity DAD with a 10-mm path length cell.

Figure 5. Analysis of vitamins with an Agilent 1260 Infi nity DAD with a 60-mm path length cell.

Vita

min

B12

min2 4 6 8 10 12

mAU

0

500

1,000

1,500

2,000

2,500

3,000

3,500

4,000

min10 10.2 10.4 10.6 10.8 11

mAU

0

0.2

0.4

0.6

1.0

0.8 Peak height = 0.768691 mAUPeak-to-peak noise = 0.04776 mAUS/N = 16.1

> 350 mAU linearity limitRed: 220 nmBlue: 265 nmGreen: 360 nm

Vita

min

C

Niac

inam

id

Pant

othe

nic a

cid

Vita

min

B6

Vita

min

B1

Vita

min

B2

Table 1. Summary of results for Figure 4 ( Quantifi cation possible; Quantifi cation not possible).

Compound Quantifi cation with an Agilent 1260 Infi nity DAD with a 10-mm cellVitamin CNiacinamidPantothenic acidVitamin B6Vitamin B1Vitamin B12Vitamin B2

Page 7: Analysis of High- and Low-Dosed Vitamins in a Single Run ... · 220 240 260 280 300 320 340 360 380 nm mAU 0 50 100 150 200 250 220 240 260 280 300 320 340 360 380 nm mAU 0 50 100

7

Analysis on a 1260 Infi nity DAD with a 60-mm path length cellQuantitation of vitamin B12 at 360 nm with a S/N = 16.1 was feasible as well as the quantitation of pantothenic acid, B6, B1, and B2, see Table 2.

No quantitation of vitamin C and niacinamid was possible. Both were out of the linear range of the DAD with the 60-mm cell.

Analysis with a 1260 Infi nity HDR DADQuantitation of all vitamins was possible due to the enhanced linear range of the 1260 Infi nity HDR DAD. One injection was suffi cient to analyze and evaluate low-dosed and high-dosed vitamins, see Table 3.

One remaining challenge was the decomposition of vitamin B12 after dissolution, see Figure 7. Each run took approximately 20 minutes, and after three injections approximately 50 % was decomposed even though the sample temperature was kept at 4 °C, and the vials were completely fi lled to exclude oxygen as much as possible.

We assume that decomposition of vitamin B12 started immediately after dissolution during sample preparation. Further decomposition was observed in the fi rst and second injections. After three injections, the area counts became more stable. Preparing the sample in the dark, and applying a protective gas might result in less decomposition and better area precision. The precision of retention times was 0.055 % RSD over seven runs. After stabilization of the area counts, the area precision was approximately 11.5 % RSD evaluating runs 4 to 7. The study on degradation of vitamin B12 in dietary supplements3 notes that if vitamin C is present in the preparation, it infl uences the decomposition of vitamin B12.

Figure 6. Analysis of vitamins with an Agilent 1260 Infi nity HDR DAD.

Vita

min

B12

Peak height = 0.126319 mAU/cmPeak-to-peak noise = 0.007922 mAU/cmS/N = 16

> 5,400 mAU linearity limittypically 6,700 mAU Red: 220 nm

Blue: 265 nmGreen: 360 nm

Vita

min

C

Niac

inam

id

Pant

othe

nic a

cid

Vita

min

B6

Vita

min

B1

Vita

min

B2

min0 2 4 6 8 10 12

mAU/cm

mAU/cm

0

250

500

750

1,000

1,250

1,500

1,750

2,000

min9.5 9.75 10.00 10.25 10.5 10.75 11.00 11.25 11.50

0.020.040.060.080.1

0.120.140.160.18

Table 3. Summary of results for Figures 4, 5, and 6 ( Quantifi cation possible; Quantifi cation not possible).

Compound

Quantifi cation with an Agilent 1260 Infi nity DAD with a 10-mm cell

Quantifi cation with an Agilent 1260 Infi nity DAD with a 60-mm cell

Quantifi cation with an Agilent 1260 Infi nity DAD HDR Solution

Vitamin CNiacinamidPantothenic acidVitamin B6Vitamin B1Vitamin B12Vitamin B2

Compound Quantifi cation with an Agilent 1260 Infi nity DAD with a 10-mm cell

Quantitation with an Agilent 1260 Infi nity DAD with a 60-mm cell

Vitamin CNiacinamidPantothenic acidVitamin B6Vitamin B1Vitamin B12Vitamin B2

Table 2. Summary of results for Figures 4 and 5 ( Quantifi cation possible; Quantifi cation not possible).

Page 8: Analysis of High- and Low-Dosed Vitamins in a Single Run ... · 220 240 260 280 300 320 340 360 380 nm mAU 0 50 100 150 200 250 220 240 260 280 300 320 340 360 380 nm mAU 0 50 100

ConclusionThe Agilent 1260 Infi nity HDR DAD allowed the analysis of main and trace compounds in one single run due to an increased linear dynamic range by a factor of > 30. The analysis of all water soluble vitamins in vitamin tablets was possible by using the 1260 Infi nity HDR DAD. Vitamin B12 with a concentration of 2 µg/tablet was quantifi ed as well as vitamin C with a concentration of 60 mg/tablet in one run.

References1. U.Kirchner, K.Degenhardt, G.Raffl er,

“Determination of Vitamin B12 in infant formula and adult nutritionals using HPLC after purifi cation on an Immunoaffi nity column: First Action 2011.09“, Journal of AOAC International Vol.95, No4, (2012).

2. Rada Amidzic et al, “RP-HPLC Determination of vitamins B1,B3,B6, and B12 in multivitamin tablets“, J.Serb.Chem.Soc. 70 (10) 1229-1235 (2005).

3. Yamada K, Shimodaira M, Chida S, Yamada N, Matsushima N, Fukuda M, Yamada S., Department of Liberal Arts and Sciences, School of Health Sciences, Sapporo Medical University, Sapporo, Japan. [email protected], “Degradation of vitamin B12 in dietary supplements.” Int J Vitam Nutr Res. Jul-Sep; 78(4-5):195-203. doi: 10.1024/0300-9831.78.45.195. (2008).

www.agilent.com/chem/hdr

© Agilent Technologies, Inc., 2013-2014Published in the USA, March 1, 20145991-3876EN

Figure 7. Decomposition of vitamin B12 over time.

min10 10.2 10.4 10.6 10.8 11 11.2 11.4

mAU/cm

0

0.025

0.05

0.075

0.1

0.125

0.15

0.175

0.2