Upload
catherine-m
View
216
Download
5
Embed Size (px)
Citation preview
Analysis of caramels by capillaryelectrophoresis and ultrafiltrationLouise Royle* and Catherine M RadcliffeDepartment of Food Science and Technology, The University of Reading, Whiteknights, PO Box 226, Reading, RG6 6AP, UK
Abstract: Class I, III and IV caramels are distinguished from each other by capillary electrophoresis at
pH 2.5 and 9.5. The majority of the colour is shown to be in the high molecular weight (MW) fraction of
all the caramels. The high MW Class I and Class IV caramel peaks migrate with a negative charge at
both pHs but the Class IV caramel also has several sharp peaks from low MW components compared to
only a small neutral peak from the Class I caramels. The Class III caramels have a high MW peak which
is positively charged at pH 2.5. The migration time, and hence charge, of the high MW peak of Class III
caramels is shown to be related to the caramel's nitrogen content.
# Crown copyright 1999
Keywords: caramel; Class I caramel; Class III caramel; Class IV caramel; capillary electrophoresis
INTRODUCTIONThe International Technical Caramel Association
divides caramels into four different categories accord-
ing to the reactants added to the sugar during
caramelisation.1 Class I caramels are produced by
heating sugar at high temperature without the addition
of nitrogenous `catalysts'; they are used for colouring
spirits such as whisky. Class II caramels have small
amounts of sulphite added as reactant, and are used in
aperitifs. Class III caramels have ammonia as reactant,
producing highly coloured caramels which are widely
used in beers and foodstuffs. Ammonium sulphites are
used as reactants for production of Class IV caramels;
these highly coloured, negatively charged caramels are
used in cola type beverages. By varying the exact
processing conditions, a wide range of products within
each of these classes can be obtained. The relative
consumption of caramels in the UK and world-wide
differs considerably, being respectively: 5% and 2% for
Class I; <1% and 1% for Class II; 70% and 28% for
Class III and; 25% and 70% for Class IV.1,2
Spectrophotometric measurement of colour is cur-
rently used for the determination of caramel concen-
tration. The different classes are distinguished by the
addition of positively charged DEAE cellulose or
negatively charged phosphoryl cellulose to an acidi®ed
solution of the caramel, which causes precipitation of
Class IV or Class III caramels, respectively.3 These
methods are fairly crude and cannot be used reliably
with foodstuffs where other food components can
interfere with both the colour measurement and the
precipitation reactions. Therefore, more speci®c ana-
lytical methods for caramels are needed. Coffey et al4
used ion-pair HPLC of a low-molecular-weight
marker compound for Class III caramel to estimate
the amounts of this caramel in various foods. Coffey etal5 also used capillary electrophoresis (CE) to analyse
Class III caramels and identi®ed some peaks which
were present only in this class of caramel. We have
found CE more successful for the analysis of high MW
melanoidins which are similar to caramels, than
HPLC, as these compounds adhere to the HPLC
columns.6 Also, we7 have shown that CE can be used
to both identify and quantify Class IV caramels in soft
drinks.
This study aims to identify, and distinguish
between, the major caramel classes (I, III and IV)
using CE.
MATERIALS AND METHODSMaterialsA range of Class I, III and IV caramels was obtained
from British manufacturers. Caramel solutions were
made up at 5mgmlÿ1 in high purity water which was
prepared in-house with a Purite Labwater RO50 unit
(Purite Ltd, High Wycombe, UK) and used through-
out. Combustion elemental analysis for nitrogen and
sulphur content was performed by either Medac Ltd
(Egham, UK) or The University of Kent (Canterbury,
UK) on samples dried at 105°C.
Sodium hydroxide solutions for CE (1.0 and 0.1M)
were obtained from Fluka (Gillingham, UK). Phos-
phate buffer was prepared from 30mM sodium
Journal of the Science of Food and Agriculture J Sci Food Agric 79:1709±1714 (1999)
* Correspondence to: Louise Royle, Department of Food Science and Technology, The University of Reading, Whiteknights, PO Box 226,Reading RG6 6AP, UKContract/grant sponsor: Ministry of Agriculture, Fisheries and Food, UK(Received 27 February 1998; revised version received 22 March 1999; accepted 13 May 1999)Neither Her Majesty’s Stationery Office nor the Department of the Ministry of Agriculture, Fisheries and Food, UK accept any responsibility forthe accuracy of any recipe, formula or instruction published in this Article or in the Journal.
# Crown copyright 1999. Reproduced with the permission of the Controller of Her Majesty's Stationery Of®ce.J Sci Food Agric 0022±5142/99/$17.50 1709
dihydrogen orthophosphate monohydrate (AnalaR
grade, Aldrich, Gillingham, UK) adjusted to pH 2.5
with hydrochloric acid (AnalaR grade, Aldrich).
Carbonate buffer was made by mixing 50mM solu-
tions of disodium carbonate and sodium hydrogen
carbonate (GPR grade, BDH) to give pH 9.5. All
solutions were ®ltered before use with a 0.2-mm PVDF
®lter (Whatman, Maidstone, UK).
Uncoated fused-silica capillaries were obtained
from Hewlett Packard (Bracknell, UK).
MethodsCapillary electrophoresis
A Hewlett Packard HP3D CE with diode-array
detection and an HP3D CE ChemStation for instru-
ment control, data acquisition and data analysis was
used throughout.
At the beginning of each day, the capillary was
preconditioned by ¯ushing successively with 1M
NaOH for 10min, 0.1M NaOH for 5min and water
for 5min. CE running conditions were: buffer: pH 9.5,
50mM carbonate buffer (changed every four runs) or
pH 2.5, 30mM phosphate buffer (changed every run);
capillary: 40cm to detector, 48.5cm total length,
50mm ID, X3 bubble; voltage: 20kV with pH 9.5
buffer, 30kV with pH 2.5 buffer, ramped up over
0.3min; temperature: 25°C; injection: 50mbar, 5s;
preconditioning: 3min 0.1M NaOH ¯ush, 3min
buffer ¯ush; detection: 200, 280, 360 and 460nm
with a band width of 10nm, plus optional full spectra
collection, 195±600nm.
Ultra®ltration
Caramel solutions were prepared at 5mgmlÿ1 in
water. This `total' caramel solution was separated into
high and low MW fractions using Microcon centrifu-
gal ®ltration devices with low-protein binding, regen-
erated cellulose ultra®ltration membranes of 5000
Dalton nominal molecular weight (NMW) cut-off
(Millipore, Watford, UK). A 200ml sample of the
caramel solution was ultra®ltered by centrifugation
and the LMW fraction collected. The retained HMW
fraction was then washed with 2�200ml of water,
reconstituted to 200ml, then 0.22mm ®ltered using an
Amicon micropure separator (Millipore, Watford,
UK).
RESULTS AND DISCUSSIONThe 11 Class III caramels analysed had nitrogen
contents of 9.08, 8.53, 8.26, 7.65, 6.63, 6.38, 5.83,
5.76, 4.84, 4.29 and 4.00%. The four Class IV
caramels analysed had nitrogen contents of 5.35,
0.86, 0.52 and 1.02% and sulphur contents of 8.90,
2.59, 2.07 and 1.79%, respectively. The ®ve Class I
caramels were assumed to have little or no nitrogen or
sulphur content, as neither ammonium nor sulphite is
added during their manufacture.
During the caramelisation reaction, some low MW
byproducts are also produced4 and ultra®ltration
membranes with a nominal MW cutoff of 5000
daltons were used as a means of separating the high
MW caramels from these components. The MW range
of caramels produced is expected to be wide, so,
although the high MW fraction should only contain
caramel, some lower MW caramels may be present in
the low MW fractions.
All the electopherograms (e-grams) shown are taken
at 200nm, because this is the most sensitive wave-
length for detection of all the components in the
caramel mixtures. The high MW caramels absorb over
the full wavelength range of the instrument (195±
600nm) with very similar absorbance spectra (Fig 1).
As there are no distinguishing differences between the
spectra from the different caramels, this peak is
referred to as the `coloured' caramel peak throughout.
The other, sharp peaks detected at 200nm show
Figure 1. Absorption spectrum of the coloured caramel peak from a ClassIV caramel.
Figure 2. E-grams obtained in carbonate buffer at pH 9.5, detection at200nm of Class I caramel as (a) total caramel; and ultrafiltration fractions(b) >5000 NMW and (c) <5000 NMW.
1710 J Sci Food Agric 79:1709±1714 (1999)
L Royle, CM Radcliffe
absorbance only in the UV and therefore are not
coloured.
Analysis in carbonate buffer at pH 9.5At pH 9.5, the electro-osmotic ¯ow (EOF) is fast
(2.9min), as all the silanol groups on the capillary are
ionised. This ¯ow is fast enough to carry negatively
charged compounds past the detector, which is located
near the cathodic end of the capillary, even though
they are migrating in the opposite direction. At pH 9.5,
all the caramel classes would be expected to be
negatively charged.
All the Class I caramels ran as a single, broad,
coloured peak without any signi®cant spikes from low
molecular weight (MW) compounds, except for a
small neutral peak migrating with the EOF. Figure 2a
shows an e-gram typical of this class of caramel. When
the ultra®ltration fractions are examined (Fig 2b above
and Fig 2c below 5000 NMW), it is clear that most of
the broad coloured peak is of high MW, although there
are some components within this peak which are of a
lower MW.
Figure 3 shows e-grams of a Class III caramel with
5.83% nitrogen content. At pH 9.5, all the Class III
caramels analysed gave similar e-grams with a large
neutral peak, and a coloured peak with a mean
migration time of between 3.4 and 4.0min. The broad
coloured peak consists of high MW caramels (Fig 3b).
The majority of the low MW compounds migrated as
though neutral, with only a small number of positively
and negatively charged peaks detectable (Fig 3c).
Figure 4 shows e-grams of a high-nitrogen high-
sulphur Class IV caramel. Previous work7 has shown
that the migration time of Class IV caramels in
carbonate buffer at pH 9.5 is related to the caramel's
sulphur content; the higher the sulphur content, the
longer the migration time of the caramel peak.
Ultra®ltration gives a clear separation between the
high MW coloured caramel peak (Fig 4b) and the low
MW components (Fig 4c). The Class IV caramels
contain the highest number of peaks from low MW
components of all the caramels analysed at this pH.
Analysis in phosphate buffer at pH 2.5At pH 2.5, the EOF is very slow (about 25min), as
most of the silanol groups on the capillary are not
charged. Compounds which are negatively charged at
this pH are not carried past the detector, as they are
migrating in the opposite direction faster than the
EOF.
The e-grams of the Class I caramel (Fig 5) at this pH
show only one peak, which is neutral and runs with the
EOF. This peak is from low MW components (Fig
5c), but its total area is greater than that seen from
neutral components at pH 9.5. This could be due to
compounds which are negatively charged at pH 9.5
becoming neutral at pH 2.5 and/or absorbance
changes due to different buffer interactions. However,
no peak is seen for the high MW coloured caramels,
Figure 3. E-grams obtained in carbonate buffer at pH 9.5, detection at200nm of Class III (5.83% nitrogen) caramel as (a) total caramel; andultrafiltration fractions (b) >5000 NMW and (c) <5000 NMW.
Figure 4. E-grams obtained in carbonate buffer at pH 9.5, detection at200nm of Class IV (5.35% nitrogen, 8.90% sulphur) caramel as (a) totalcaramel; and ultrafiltration fractions (b) >5000 NMW and (c) <5000 NMW.
J Sci Food Agric 79:1709±1714 (1999) 1711
Analysis of caramels
indicating that the Class I caramels are still negatively
charged and thus have isoelectric points of less than
pH 2.5.
Figure 6 shows e-grams typical of the Class III
caramels. This is the only class of caramel where the
high MW coloured caramel component is positively
charged at pH 2.5. Thus the Class III caramels have
isoelectric points between pH 2.5 and 9.5. Figure 6b
shows that the majority of the broad peak is made up of
high MW coloured caramel. Figure 6c shows that
there are low MW compounds present, some posi-
tively charged and some neutral. It also shows that
some of the coloured peak is of a NMW lower than
5000. Figure 6a is similar to the e-gram obtained by
Coffey et al5 of Class III caramel analysed in 30mM
phosphate buffer at pH 1.9, suggesting that the large
peaks picked up by their HPLC analysis were low MW
neutral compounds.
In Fig 7, the e-grams of the Class IV caramel show
sharp peaks for the low MW compounds, some of
which are positively charged and some neutral at this
low pH (Fig 7c). The coloured, high MW caramel
peak is not seen (Fig 7b), indicating that, like the Class
I caramels, the Class IV caramels are still negatively
charged and thus also have isoelectric points of less
than pH 2.5. All the different Class IV caramels gave
e-grams similar to Fig 7, but with progressively lessFigure 5. E-grams obtained in phosphate buffer at pH 2.5, detection at200nm of Class I caramel as (a) total caramel; and ultrafiltration fractions(b) >5000 NMW and (c) <5000 NMW.
Figure 6. E-grams obtained in phosphate buffer at pH 2.5, detection at200nm of Class III (5.83% nitrogen) caramel as (a) total caramel; andultrafiltration fractions (b) >5000 NMW and (c) <5000 NMW.
Figure 7. E-grams obtained in phosphate buffer at pH 2.5, detection at200nm of Class IV (5.35% nitrogen, 8.90% sulphur) caramel as (a) totalcaramel; and ultrafiltration fractions (b) >5000 NMW and (c) <5000 NMW.
1712 J Sci Food Agric 79:1709±1714 (1999)
L Royle, CM Radcliffe
marked low MW peaks as the caramel sulphur content
decreased.
Figure 8 compares four of the Class III caramels in
the early region of the e-gram where the positive,
largely high MW, components appear. When the
migration times of this broad peak (taken at the apex
of the parent peak) from all 11 Class III caramels are
examined, a plot of mean migration time (from three
runs) vs % nitrogen gives a moderately good ®t to a
straight line (Fig 9, r2=0.858), whereas a plot of mean
peak width at half height (from three runs) vs %
nitrogen gives a somewhat lower ®t (Fig 10,
r2=0.706). This shows that at pH 2.5 the mean
migration time and peak width at half height are
related to the nitrogen content of the Class III caramel
in a more or less linear fashion. This relationship can
be explained by the nitrogen-containing chemical
groups carrying a positive charge at this pH; therefore,
the higher the nitrogen content, the higher the positive
charge on the caramel and the shorter its migration
time. The decrease in peak width with increasing
nitrogen content could be due to a narrowing of the
range of size-to-charge ratio of caramel, as the
likelihood of an even nitrogen distribution increases,
and/or there could be less band broadening with
decreasing migration time due to capillary-wall inter-
actions.
CONCLUSIONSThe high MW component of Class I caramels migrates
as a broad coloured peak at pH 9.5, but this peak is not
detected at pH 2.5, suggesting an isoelectric point of
less than 2.5. There is only one small neutral peak of
low MW detected at both pH 9.5 and 2.5, presumably
due to the sugar solution undergoing a less complex
caramelisation reaction in the absence of added
chemicals than in their presence, as with Class III
and IV caramels.
Class III caramels all have similar e-grams at pH 9.5,
a broad coloured peak of negatively charged high MW
caramel and a large sharp peak from neutral low
MW components. At pH 2.5 the coloured high MW
caramel migrates as a positively charged broad band.
The mean migration time and width at half height of
this peak are shown to be related to the caramel's
nitrogen content. There are also several positively
charged and neutral low MW components detectable
at pH 2.5. The Class III caramels have isoelectric
points between 2.5 and 9.5.
For the Class IV caramels, the high MW coloured
caramel migrates as a broad band at pH 9.5, the mean
migration time being related to the caramel's sulphur
content.7 This peak is not detected at pH 2.5,
suggesting isoelectric points of less than 2.5. There
are also many low MW components detected at both
pH 9.5 and 2.5.
The major caramel classes can now be distinguished
by CE in simple aqueous solutions. Class III and Class
Figure 8. E-grams obtained in phosphate buffer at pH 2.5, detection at200nm of Class III caramels with nitrogen contents of (a) 9.08%, (b) 7.65%,(c) 5.83% and (d) 4.00%.
Figure 9. Plot of migration time at 200nm (mean of three runs, relativestandard deviations <1%) vs % nitrogen, for CE analysis of Class IIIcaramels in phosphate buffer at pH 2.5.
Figure 10. Plot of peak width at half height at 200nm (standard deviationsshown) vs % nitrogen, for CE analysis of Class III caramels in phosphatebuffer at pH 2.5.
J Sci Food Agric 79:1709±1714 (1999) 1713
Analysis of caramels
IV caramels can be further categorised according to
their nitrogen or sulphur content, respectively. Since
the type of caramel present can thus be identi®ed, the
quanti®cation of caramels in food has become a
realistic possibility. Future work will investigate the
degree to which other food components affect these
analyses.
ACKNOWLEDGEMENTSThe authors thank Professor HE Nursten and Dr JM
Ames (The University of Reading, Department of
Food Science and Technology) and Dr L Castle (CSL
Food Science Laboratory, Norwich) for their helpful
comments and discussions, the caramel producers
who kindly donated samples for analysis, and MAFF
for funding the work.
REFERENCES1 International Technical Caramel Association, Guide to the Speci-
®cations of Caramel Colour (unpublished). Atlanta, Ga, USA
(1984).
2 MAFF (Ministry of Agriculture, Fisheries and Food), Dietary
intake of food-additives in the UK: initial surveillance. Food
Surveillance Paper No 37, HMSO, London (1993).
3 Licht BH, Shaw K, Smith C, Mendoza M, Orr J and Myers DV,
Development of speci®cations for caramel colours. Food &
Chemical Toxicology 30:383±387 (1992).
4 Coffey JS, Nursten HE, Ames JM and Castle L, A liquid
chromatographic method for the estimation of Class III caramel
added to foods. Food Chem 58:259±267 (1997).
5 Coffey JS and Castle L, Analysis of caramel colour (Class III).
Food Chem 51:413±416 (1994).
6 Royle L, Bailey RG and Ames JM, Separation of Maillard reaction
products from xylose-glycine and glucose-glycine model sys-
tems by capillary electrophoresis and comparison to reverse
phase HPLC. Food Chem 62:425±430 (1998).
7 Royle L, Ames JM, Castle L, Nursten HE and Radcliffe CM,
Identi®cation and quanti®cation of Class IV caramels using
capillary electrophoresis and its application to soft drinks. J Sci
Food Agri 76:579±587 (1998).
1714 J Sci Food Agric 79:1709±1714 (1999)
L Royle, CM Radcliffe