Upload
transilvanpop
View
258
Download
0
Embed Size (px)
Citation preview
7/30/2019 Analizor Gaze - Manual
1/106
GA3000PLUS Gas Analyser
Operating Manual
Geotechnical Instruments (UK) Ltd
Sovereign House
QueenswayLeamington Spa
Warwickshire
CV31 3JR
England
Tel: +44 (0)1926 338111
Fax: +44 (0)1926 3381110
Email:[email protected]
Website:www.geotechuk.com
mailto:[email protected]:[email protected]:[email protected]://www.geotech.co.uk/http://www.geotech.co.uk/http://www.geotech.co.uk/mailto:[email protected]7/30/2019 Analizor Gaze - Manual
2/106
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
Table of Contents
1 Manual Guidelines ..........................................................................................7
1.1 Document History .......................................................................................... 7
1.2 Safety Symbol ................................................................................................ 7
1.3 Notes ............................................................................................................. 7
2 Introduction ....................................................................................................8
2.1 GA3000 PLUS Model Type Definitions ........................................................... 9
2.1.1 GA3KPLUS1........................................................................................... 9
2.1.2 GA3KPLUS2........................................................................................... 9
2.1.3 GA3KPLUS3......................................................................................... 10
2.1.4 GA3KPLUSC ........................................................................................ 10
2.1.5 GA3KPLUSD ........................................................................................ 11
2.2 Safety Instructions ....................................................................................... 12
2.3 ATEX Certification ........................................................................................ 13
2.4 Range of Environmental Conditions............................................................. 13
2.5 Safety Symbols Used on the GA3000 PLUS .................................................. 14
3 GA3000 PLUS Gas Analyser ............................................................................ 15
3.1 The GA3000 PLUS ........................................................................................ 15
3.1.1 Features ............................................................................................. 15
3.1.2 Benefits .............................................................................................. 16
3.1.3 Main Applications ............................................................................... 16
3.2 Instrument Components .............................................................................. 17
4 GA3000 PLUS Instrument Features ................................................................ 19
4.1 Physical Characteristics of the Gas Analyser ................................................ 19
4.2 Gas Analyser Definitions .............................................................................. 20
4.2.1 Front View .......................................................................................... 20
4.2.2 Rear View ........................................................................................... 21
7/30/2019 Analizor Gaze - Manual
3/106
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
4.3 Instrument Connection Points ..................................................................... 22
4.3.1 Right Side View ................................................................................... 23
4.3.2 Left Side View ..................................................................................... 23
5 Installation .................................................................................................... 24
5.1 Pre-Installation Requirements ..................................................................... 24
5.1.1 General ............................................................................................... 24
5.1.2 Storage of the Gas Analyser................................................................ 24
5.1.3 Ventilation Requirements ................................................................... 24
5.2 Mounting the Cabinet .................................................................................. 25
5.3 External Customer Connections ................................................................... 26
5.4 Terminal Conductor Sizes ............................................................................ 28
5.5 Mains Wiring ............................................................................................... 28
5.5.1 Protective Earthing ............................................................................. 28
5.5.2 How to Wire the Mains Supply ........................................................... 29
5.6 Alarm, Fault and Sample Point Relays .......................................................... 30
5.6.1 Alarm Relays ....................................................................................... 30
5.6.2 Fault Relay .......................................................................................... 30
5.6.3 Sample Point Relays ........................................................................... 31
5.7 4-20mA Outputs .......................................................................................... 31
5.7.1 Wiring the GA3000 PLUS to Current Source Inputs........................... 32
5.7.2 Wiring the GA3000 PLUS to Current Sink Inputs ............................... 32
5.7.3 4-20mA Scaling ................................................................................... 36
5.8 Modbus Digital Output ................................................................................ 36
5.8.1 Wiring the GA3000 PLUS Modbus Outputs ......................................... 36
5.8.2 Configuration of the GA3000 PLUS Modbus Port ............................... 37
5.8.3 Readable Parameters of the GA3000 PLUS PLC .................................. 38
5.9 Connecting the Gas Lines to the GA3000 PLUS ............................................ 41
5.9.1 Sample Lines ....................................................................................... 41
7/30/2019 Analizor Gaze - Manual
4/106
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
5.9.2 Air Purge Inlet..................................................................................... 42
5.9.3 Landfill Return .................................................................................... 42
5.9.4 Vent Atmosphere ............................................................................... 42
5.9.5 Drain ................................................................................................... 43
5.10 Final Checks ................................................................................................. 43
6 General Operational Instructions .................................................................. 45
6.1 Switching the Gas Analyser On .................................................................... 45
6.2 Power On Self Test (POST) ........................................................................... 45
6.3 First Time Run Set-up .................................................................................. 46
6.4 Instrument Main Read Screen ..................................................................... 50
6.4.1 General ............................................................................................... 50
6.4.2 Alarm Indications ............................................................................... 52
6.4.3 Fault Indication ................................................................................... 53
6.4.4 Service Overdue Indication ................................................................. 53
6.4.5 View Alarms ....................................................................................... 54
6.4.6 Pause Sampling .................................................................................. 55
6.5 Main Menu .................................................................................................. 58
6.5.1 Exit Menu and Resume Sampling ....................................................... 58
6.5.2 Alarm Set-up ...................................................................................... 58
6.5.3 User Calibration .................................................................................. 60
6.5.4 Set Instrument Time and Date ............................................................ 60
6.5.5 Sampling Options ............................................................................... 62
6.5.6 View Event Log ................................................................................... 63
6.5.7 Display Contrast ................................................................................. 64
6.5.8 System Information ............................................................................ 64
6.5.9 Administrator Passcode ...................................................................... 65
6.6 GA3000 PLUS Sampling Process ................................................................... 67
6.6.1 GA3KPLUS1......................................................................................... 67
7/30/2019 Analizor Gaze - Manual
5/106
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
6.6.2 GA3KPLUS2 and GA3KPLUSD .............................................................. 67
6.6.3 GA3KPLUS3......................................................................................... 68
6.6.4 GA3KPLUSC ........................................................................................ 69
6.7 Switching the Gas Analyser Off .................................................................... 69
7 Calibration .................................................................................................... 70
7.1 Calibration Introduction .............................................................................. 70
7.2 Required Calibration Equipment .................................................................. 70
7.2.1 Gas ..................................................................................................... 70
7.2.2 Flow Regulator ................................................................................... 71
7.2.3 Set-up ................................................................................................. 72
7.3 Calibration Mixture Processes Best Practice ............................................. 73
7.3.1 Internal (CH4/CO2/O2/Internal Sensor) Calibration ............................. 73
7.3.2 External Sensor Calibration ................................................................ 75
7.4 Last Field Calibration ................................................................................... 77
8 Maintenance ................................................................................................. 78
8.1 GA3000 PLUS Consumable Products ............................................................ 79
8.2 Replacing the Coalescing (Catchpot) Filter ................................................... 81
8.3 Replacing the Water Trap Filter ................................................................... 82
8.4 Replacing the External Sensor ..................................................................... 83
8.5 Replacing the Inlet Port Filter ...................................................................... 84
8.6 Cleaning and Decontamination .................................................................... 85
9 Service .......................................................................................................... 86
9.1 General ........................................................................................................ 86
9.2 Replacement of Gas Analyser for Service Hot Swap .................................. 86
9.3 Sample Certificate of Calibration Page 1 ................................................... 88
9.4 Sample Certificate of Calibration Page 2 ................................................... 89
10 Problem Solving ............................................................................................ 90
10.1 Fault Detection ............................................................................................ 90
7/30/2019 Analizor Gaze - Manual
6/106
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
10.1.1 Error, Warning and Fault Display ........................................................ 90
10.1.2 Types of Faults .................................................................................... 90
10.1.3 Fault Screens ...................................................................................... 92
10.1.4 Under and Over Range Codes ............................................................. 93
10.2 Instrument Lock-Up ..................................................................................... 93
10.3 Analogue Outputs not Working ................................................................... 94
10.4 Low Flow ..................................................................................................... 94
10.5 Fuses ........................................................................................................... 95
10.6 Cold Start ..................................................................................................... 96
10.7 Event Log ..................................................................................................... 97
11 Warranty Information ................................................................................. 101
12 WEEE Compliance ........................................................................................ 102
13 Declaration of Conformity English Language ............................................. 103
13.1 ATEX Directive 94/9/EC References ........................................................... 103
13.2 EMC Directive 89/336/EEC References ...................................................... 103
13.3 Low Voltage Directive References ............................................................. 103
14 Glossary of Terms ........................................................................................ 104
7/30/2019 Analizor Gaze - Manual
7/106
7 GA3000 PLUS Gas Analyser Operating Manual
Manual Guidelines
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
1 Manual Guidelines1.1 Document HistoryIssued by Issue date
Change
control IDIssue No. Reason for change
C. Millar 06/09/2012 N/A 01 New document
1.2 Safety SymbolInformation in this manual that may affect the safety of users and others is
preceded by the following symbol:
Warning
Failure to follow this information may result in physical injury which in some
cases could be fatal.
1.3 NotesImportant /useful information and instructions are shown clearly throughout
the manual in a note format.
For example:
Note: For further information please contact technical support atGeotech on +44(0)1926 338111 or [email protected].
mailto:[email protected]:[email protected]:[email protected]:[email protected]7/30/2019 Analizor Gaze - Manual
8/106
GA3000 PLUS Gas Analyser Operating Manual 8
Introduction
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
2 IntroductionThis manual explains how to install and use the GA3000 PLUS range of gas
analysers for the following model types listed below:
GA3KPLUS1 Measures CH4, CO2 and O2 as standard from one gassample point.
GA3KPLUS2 Measures CH4, CO2 and O2 as standard from two gassample points.
GA3KPLUS3 Measures CH4, CO2 and O2 as standard from three gassample points.
Note: Each of the above variants can be fitted with up to two additionalgas sensors. For further information contact sales at Geotech on
+44(0)1926 338111 or [email protected].
GA3KPLUSC Measures CH4, CO2 and O2 continuously as standard fromone gas sample point.
Note: The above variant can be fitted with up to one additional gassensor. For further information contact sales at Geotech on +44(0)1926
338111 or [email protected].
GA3KPLUSD Measures CH4, CO2, O2 and high & low levels of H2S fromtwo gas sample points.
Note: The above variant is fitted with two additional H2S gas sensors. Forfurther information contact sales at Geotech on +44(0)1926 338111 or
mailto:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]7/30/2019 Analizor Gaze - Manual
9/106
9 GA3000 PLUS Gas Analyser Operating Manual
Introduction
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
2.1 GA3000 PLUS Model Type Definitions2.1.1 GA3KPLUS1
The GA3KPLUS1 measures CH4, CO2 and O2 as standard with a choice ofup to two additional gases (one internal to the GA and one external) and
measures from one sample point.
It has user selectable sample and air purge intervals which are exhaustedback to the process and atmosphere respectively.
The 4-20mA and Modbus outputs are updated at the end of eachsample.
Following is a simple block diagram of the system including the optionalexternal gas sensor:
2.1.2 GA3KPLUS2 The GA3KPLUS2 measures CH4, CO2 and O2 as standard with a choice of
up to two additional gases (one internal to the GA and one external) and
measures from two sample points.
It has user selectable sample and air purge intervals which are exhaustedback to the process and atmosphere respectively.
The 4-20mA and Modbus outputs are updated at the end of eachsample.
Following is a simple block diagram of the system including the optionalexternal gas sensor:
7/30/2019 Analizor Gaze - Manual
10/106
GA3000 PLUS Gas Analyser Operating Manual 10
Introduction
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
2.1.3 GA3KPLUS3 The GA3KPLUS3 measures CH4, CO2 and O2 as standard with a choice of
up to two additional gases (one internal to the GA and one external)from three sample points.
It has user selectable sample and air purge intervals which are exhaustedback to the process and atmosphere respectively.
The 4-20mA and Modbus outputs are updated at the end of eachsample.
Following is a simple block diagram of the system including the optionalexternal gas sensor:
2.1.4 GA3KPLUSC The GA3KPLUSC measures CH4, CO2 and O2 as standard with a choice of
up to one additional gas (external to GA only) from one sample point.
It measures CH4, CO2 and O2 continuously which is exhausted back tothe process.
The external sensor has user selectable sample intervals which areexhausted back to the process. The external sensor cannot be subjected
to the gas stream continuously.
After each defined cycle the external sensor is subjected to a userdefinable air purge (minimum three minutes) which is exhausted to the
atmosphere.
The 4-20mA and Modbus outputs are updated continuously, with theexternal sensor data being refreshed at the end of each sample.
Following is two simple block diagrams of the system with one includingthe optional external gas sensor:
7/30/2019 Analizor Gaze - Manual
11/106
11 GA3000 PLUS Gas Analyser Operating Manual
Introduction
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
2.1.5 GA3KPLUSD The GA3KPLUSD measures CH4, CO2, O2 and high & low range H2S as
standard (high range internal to the GA and low range external) from
two sample points.
It has user selectable sample and air purge intervals which are exhaustedback to the process and atmosphere respectively.
The system has a bypass solenoid in order to divert high levels of H2Sfrom the low range sensor, which will prolong its life.
The 4-20mA and Modbus outputs are updated at the end of eachsample.
Following is a simple block diagram of the system:
Note: Letters P, A, and R indicate tubing connection points to thesolenoids. These letters can be found on each solenoid within the
system.
7/30/2019 Analizor Gaze - Manual
12/106
GA3000 PLUS Gas Analyser Operating Manual 12
Introduction
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
2.2 Safety InstructionsWarning The GA3000 PLUS range of gas analysers can be used
for measuring gases in a wide range of environmentalapplications as described in this manual.
Inhaling toxic gases may be harmful to health and in
some cases may be fatal. It is the responsibility of the
user to ensure that he/she is adequately trained in the
safety aspects of the gases being used and appropriate
procedures are followed. In particular, where
hazardous gases are being used, the gas exhausted
from the analyser must be piped to an area where it issafe to discharge the gas, or returned to the process.
Hazardous gas can be expelled from the instrument,
including when purging with clean air.
The amount of sample gas that could be expelled
during an air purge is typically 50ml. If in any doubt
consult an appropriately qualified person about the
dangers of the gas you are measuring.
The equipment is only for use in ambient temperatures in the range 0Cto +50C and should not be used outside this range, unless the heater
option is purchased when the lower ambient temperature can be
extended to -20C.
The equipment is only for use in ambient pressures in the range 500 to1500 mbar and should not be used outside of this range.
The maximum surface temperature recorded during the ATEXassessment was 180C; this is from a component within the pumpassembly.
Installation should be carried out in accordance with the applicablecode of practice by suitably trained personnel.
Repair of this equipment should be carried out in accordance with theapplicable code of practice.
If the equipment is likely to come into contact with aggressivesubstances, e.g. acidic liquids or gases that may attack metals, or
7/30/2019 Analizor Gaze - Manual
13/106
13 GA3000 PLUS Gas Analyser Operating Manual
Introduction
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
solvents that may affect polymeric materials, then it is the
responsibility of the user to take suitable precautions, e.g. regular
checks as part of routine inspections or establishing from the materials
datasheet that it is resistant to specific chemicals that prevent it frombeing adversely affected, thus ensuring that the type of protection is
not compromised.
Warning When opening the cabinet great care must be taken by
the operator as mains voltages are present. It is the
responsibility of the user of the equipment to ensure
that all personnel are adequately trained.
Note: It is vital that the instructions in this operating manual arefollowed closely. Failure to comply could invalidate the ATEX
certification or cause injury to the user.
2.3 ATEX CertificationReference European ATEX Directive 94/9/EC
The GA3000 PLUS gas analyser has been certified to hazardous area
classification and is ATEX certified according to the following designation:
II 3 G Ex nA IIA T1 Gc (-20C Ta +50C)
It is the responsibility of the user to determine the protection concept and
classification for a particular application.
Note: For further information please contact technical support atGeotech on +44(0)1926 338111 or [email protected].
Warning When opening the cabinet great care must be taken by
the operator as mains voltages are present. It is the
responsibility of the user of the equipment to ensure
that all personnel are adequately trained.
2.4 Range of Environmental ConditionsThe GA3000 PLUS is designed for use outdoors and has an IP65 rating. The
operating temperature range is 0C to +50C without the use of the Geotechapproved ATEX certified heater, and -20C to +50C with the use of the
mailto:[email protected]:[email protected]:[email protected]:[email protected]7/30/2019 Analizor Gaze - Manual
14/106
GA3000 PLUS Gas Analyser Operating Manual 14
Introduction
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
Geotech approved ATEX certified heater. The mains voltage (110-240V) can
fluctuate up to 10% of the nominal voltage.
2.5 Safety Symbols Used on the GA3000 PLUSThe following safety symbols are used on the GA3000 PLUS range of gas
analysers:
Earth (ground) Terminal
Protective Conductor Terminal
Caution, risk of electric shock
Caution, risk of danger
7/30/2019 Analizor Gaze - Manual
15/106
15 GA3000 PLUS Gas Analyser Operating Manual
GA3000 PLUS Gas Analyser
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
3 GA3000 PLUS Gas Analyser3.1 The GA3000 PLUS
The GA3000 PLUS range of gas analysers builds on field-proven gas analysis
technology to offer effective online monitoring with local data outputs.
3.1.1 Features CH4, CO2 and O2 standard measurement Optional H2S, H2 and CO measurement (choice of up to two additional
gases)
Multi-point sampling (up to 3 gas sample points) H2S ranges from 0-50ppm to 0-10,000ppm Monitor before and after desulphurisation Continuous monitoring option Clean air purge Large and clear back-lit display Four 4-20mA outputs Modbus communication Gas return to process as standard IP65 rated stainless steel enclosure
7/30/2019 Analizor Gaze - Manual
16/106
GA3000 PLUS Gas Analyser Operating Manual 16
GA3000 PLUS Gas Analyser
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
ATEX certified for use in Zone 2 areas3.1.2 Benefits
No training required Low cost of ownership Calibration accredited to ISO 17025 (applies to CH4, CO2, and non-
continuous O2 only)
Quick and easy self-installation Compact self-contained system Moisture removal included as standard Zero service downtime Clear servicing schedules Field-proven equipment Start-up configuration wizard Simple user calibration
3.1.3 Main Applications Landfill and biogas Anaerobic digestion projects
o Waste water treatmento Food processing plantso Food and animal waste
Clean development mechanism (CDM)
7/30/2019 Analizor Gaze - Manual
17/106
17 GA3000 PLUS Gas Analyser Operating Manual
GA3000 PLUS Gas Analyser
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
3.2 Instrument Components
Diagram 1 GA3KPLUS instrument components
A Gas analyser H Heater thermostat
B Solenoids I Mains input
C Pump J Modbus terminals
D External sensor K 4-20mA terminals
E Fuses L Gas in/out and drain valvesF Catchpot M Inline coupling
G Alarm/fault/solenoid relays N Calibration valve
B
G
C
H
KJI
E
D
A
F
L
M
N
7/30/2019 Analizor Gaze - Manual
18/106
GA3000 PLUS Gas Analyser Operating Manual 18
GA3000 PLUS Gas Analyser
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
Diagram 2 - GA3KPLUS3 inside of door
O Viewing window
P Insulation (when heater installed, also fitted to main enclosure)
Q Heater
O
P
Q
7/30/2019 Analizor Gaze - Manual
19/106
19 GA3000 PLUS Gas Analyser Operating Manual
GA3000 PLUS Instrument Features
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
4 GA3000 PLUS Instrument Features4.1 Physical Characteristics of the Gas AnalyserFront view: Reference:
A Main gas read screen
B Key 2, scroll up
C Key 4, scroll left
D Key 8, scroll down
E Key 0
F Key 6, scroll right
G On/off key
H Enter key
I General operational keys
J Backspace/delete key
K Instrument label
Rear view: Reference:
L Inlet port filter
M Serial number
N Product option number
A
B
C
D
E
F
G
H
I
J
K
L
M N
7/30/2019 Analizor Gaze - Manual
20/106
GA3000 PLUS Gas Analyser Operating Manual 20
GA3000 PLUS Instrument Features
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
4.2 Gas Analyser Definitions4.2.1 Front View
Reference DefinitionA Main read screen Start and end screen when using the
instrument
B Key 2, scroll up Press scroll up to display more information.
Also key 2.
C Key 4, scroll left Press scroll left to display more information.
Also key 4.
D Key 8, scroll down Press scroll down to display more
information. Also key 8.
E Key 0 Key 0. (Note backlight is always on and
cannot be turned off).
F Key 6, scroll right Press scroll right to display more information.
Also key 6.
G On/off key Hold the on/off key for two seconds to turn
the gas analyser on and off
H Enter key The enter key accepts/confirms choices
made by the operator for various functions
and operations. Also required to confirm
numeric data entry and enable edit mode in
certain menu options.
I General operational keys Keys 1, 2, 3, 4, 5, 6, 7, 8, 9, 0
J Backspace/delete key Can be used to delete an entry. (Note the
pump cannot be controlled by this button).
K Instrument label Instrument product label
7/30/2019 Analizor Gaze - Manual
21/106
21 GA3000 PLUS Gas Analyser Operating Manual
GA3000 PLUS Instrument Features
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
4.2.2 Rear ViewReference Definition
L Inlet port filter The inlet port filter acts as a particulate filter.
The grey cover houses and protects the inlet
port filter.
M Serial number Unique identification number for the gas
analyser.
N Product option number Records gas analyser internal configuration at
the time of manufacture, e.g. FG3KPLUS0
7/30/2019 Analizor Gaze - Manual
22/106
GA3000 PLUS Gas Analyser Operating Manual 22
GA3000 PLUS Instrument Features
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
4.3 Instrument Connection PointsRight side view: Reference:
A Inlet port
B Unused port
Left side view: Reference:
C Gas outlet port
D Connector D
E Connector E
A
B
C
D
E
7/30/2019 Analizor Gaze - Manual
23/106
23 GA3000 PLUS Gas Analyser Operating Manual
GA3000 PLUS Instrument Features
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
4.3.1 Right Side ViewReference Definition
A Inlet port The sample tube is attached to this port for
the gas analyser to take a reading.
B Unused port This port is not used on the GA3000 PLUS.
4.3.2 Left Side ViewReference Definition
C Gas outlet port The sample tube is attached to this port to
allow the gas analyser to take a gas reading.
D Connector D
(Half-moon)
Communications port used to connect the
gas analyser to the PLC.
E Connector E Used to connect the gas analyser to the 9V
external supply.
7/30/2019 Analizor Gaze - Manual
24/106
GA3000 PLUS Gas Analyser Operating Manual 24
Installation
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
5 Installation5.1 Pre-Installation Requirements5.1.1 GeneralIn order to effectively install the GA3000 PLUS system it is important that the
site is ready and in a fit state. In particular, the following points should be
noted:
A suitable mains supply as detailed in this manual is installed. A suitable location is determined for the installation of the
instrumentation. There are no health and safety problems on site a risk assessment may
be required.
All required gas lines are installed. The GA3000 PLUS system has been received on site, unpacked and
checked for obvious damage.
4 20mA/Modbus cable has been installed to the instrument location. Note: Failure to comply with any of the above may result in additional
time on site and additional costs.
Warning Power should NOT be applied before all plumbing and
wiring has been completed and tested.
5.1.2 Storage of the Gas AnalyserThe analyser should not be exposed to extremes of temperature. It is the
users responsibility to ensure the instrument is kept within its ambient
operating temperature range.
5.1.3 Ventilation RequirementsThere is a breather fitted to the GA3000 PLUS situated at the bottom of the
enclosure (see annotation B onDiagram 4 - GA3000 PLUS connection points).
It is the users responsibility to ensure there is a free circulation of air around
the cabinet.
7/30/2019 Analizor Gaze - Manual
25/106
25 GA3000 PLUS Gas Analyser Operating Manual
Installation
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
5.2 Mounting the CabinetThe system is contained in one cabinet. The cabinet is weatherproof with a
rating to IP65. The cabinet must be fixed to a suitable wall or frameworkcapable of holding the weight of the system.
Although the cabinet is weatherproof, consideration should be given to
mounting it in an area that is protected from the worst of the weather. For
example, maintenance of the analyser will be made easier if it is not exposed
to driving rain. Positioning the cabinet in direct sunlight in hot countries should
be avoided (where possible) as this may increase the internal temperature of
the cabinet.
The weight of the cabinet and contents will depend on the options that are
fitted, but will be approximately 36 kg; it is therefore recommended that the
installation is undertaken by a minimum of two people.
Four mounting fixings are supplied and fitted to each corner of the GA3000
PLUS. Suitable nut & bolt or raw bolt arrangements will have to be defined by
the operator for fixing to the wall or framework (it is recommended that these
are stainless steel). Dimensions of the cabinet are provided in diagram 3
below:
7/30/2019 Analizor Gaze - Manual
26/106
GA3000 PLUS Gas Analyser Operating Manual 26
Installation
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
Diagram 3 - Enclosure wall mounting bracket dimensions
5.3 External Customer ConnectionsPart of the installation will require the operator to connect a mains cable,
output cables and gas pipes to the equipment. Diagram 4 below identifies the
available connection points on the GA3000 PLUS:
Diagram 4 - GA3000 PLUS connection points
A Customer cable entry E Landfill return
B Breather F Gas sample in
C Air purge in G Vent atmosphere
D Drain
7/30/2019 Analizor Gaze - Manual
27/106
27 GA3000 PLUS Gas Analyser Operating Manual
Installation
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
Note: The user must remove an M20 stopping plug from annotation Awhere they plan to place a cable gland. The remaining M20 stopping
plugs must remain in place if the holes are unused in order to maintain
the IP rating of the enclosure.
Warning The mains and output cables must enter the cabinet by
armoured cable via a metal Ex e cable gland and the
mains supply should be isolated.
7/30/2019 Analizor Gaze - Manual
28/106
GA3000 PLUS Gas Analyser Operating Manual 28
Installation
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
5.4 Terminal Conductor SizesFor the mains, 4 20mA, Modbus and relay terminals the conductor size must
be as in the table below:
Conductor type Allowed conductor size
(mm2)
Solid H07V-U mm2
0.5 4
Stranded H07V-R mm2 1.5 4
Flexible H07V-K mm2 0.5 4
Flexible H07V-K and AEH DIN 46 228/1 mm2 0.5 2.5
Flexible H07V-K and AEH with plastic
insulation collar mm2
0.5 2.5
Warning The use of conductors outside of the specification will
affect the safety and ATEX certification of the
apparatus
5.5 Mains Wiring5.5.1 Protective EarthingThe safety of the equipment depends on there being effective earthing of the
apparatus via a mains supply.
The mains requirement for the system is:
Type AC
Voltage 100 250VAC
Frequency 50/60Hz
Current 3A
The fuse rating is 3.15A.
The mains cable must enter the cabinet via a metal Ex e cable gland and
armoured cable through one of the three available customer cable entry
points (refer to annotation A onDiagram 4 - GA3000 PLUS connection points).
Warning Use of a non EX e gland will invalidate the ATEX
certification.
7/30/2019 Analizor Gaze - Manual
29/106
29 GA3000 PLUS Gas Analyser Operating Manual
Installation
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
Failure to use armoured cable will invalidate the ATEX
certification.
This equipment must be provided with a switched andfused mains supply. The switch must be mounted close
to the equipment and clearly identified as the
disconnecting device for the equipment.
Failure to connect a suitable earth to the analyser could
result in serious injury.
5.5.2 How to Wire the Mains SupplyRefer to annotation I onDiagram 1 GA3KPLUS instrument componentsfor
the location of the mains wiring terminals, and diagram 5 below for the mains
wiring diagram:
Diagram 5 - Mains wiring diagram
Note: For terminal conductor sizes please refer to sectionTerminalConductor Sizes
7/30/2019 Analizor Gaze - Manual
30/106
GA3000 PLUS Gas Analyser Operating Manual 30
Installation
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
5.6 Alarm, Fault and Sample Point RelaysDependent upon the model chosen at point of manufacture, the GA3000 PLUS
will have the following available relays (refer to annotation G onDiagram 1GA3KPLUS instrument componentsfor relay location):
Model Available Relays
GA3KPLUS1 Sample 1
Alarm 1
Alarm 2
Fault
GA3KPLUS2 Sample 1
Sample 2Air
Fault
GA3KPLUS3 Sample 1
Sample 2
Sample 3
Air
Fault
GA3KPLUSC Alarm 1
Alarm 2
Fault
GA3KPLUSD As per GA3KPLUS2
5.6.1 Alarm RelaysAlarm relays are only available on a single sample point system. The relays are
volt free change-over contacts. The maximum rated voltage is 240V.
The relays operate in a fail safe mode, i.e. normally energised. This means
wiring across connections 11 and 12 during an alarm condition will complete
the circuit (de-energising the relay). Refer toDiagram 6 - Relay wiring diagram
in non-energised statefor a relay wiring diagram.
5.6.2 Fault RelayThe fault relay is available on all systems. The relays are volt free change-over
contacts. The maximum rated voltage is 240V.
7/30/2019 Analizor Gaze - Manual
31/106
31 GA3000 PLUS Gas Analyser Operating Manual
Installation
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
The relays operate in a fail safe mode, i.e. normally energised. This means
wiring across connections 11 and 12 during a fault alarm condition will
complete the circuit (de-energising the relay). Refer toDiagram 6 - Relay wiring
diagram in non-energised statefor a relay wiring diagram.
5.6.3 Sample Point RelaysThe sample point relays are available on all systems except for the GA3KPLUSC
(this is because it is monitoring all of the time). The sample point relays should
be used as an indication as to which sample point the GA3000 PLUS is
monitoring.
The relays are volt free change-over contacts. The maximum rated voltage is240V. The relays operate by energising when the particular sample point is
being monitored and thus the 4-20mA and Modbus outputs should be read
when the solenoid changes state from energised to de-energised. Refer to
Diagram 6 - Relay wiring diagram in non-energised statefor a relay wiring
diagram:
Diagram 6 - Relay wiring diagram in non-energised state
Note: For terminal conductor sizes please refer to sectionTerminalConductor Sizes
5.7 4-20mA OutputsIf analogue outputs are being used the cable must enter the cabinet via a
metal Ex e cable gland and armoured cable through one of the three available
customer cable entry points (refer to annotation A onDiagram 1 GA3KPLUS
instrument components).
The outputs current sink in to the GA3000 PLUS. Labels identify the
appropriate outputs namely from left to right (1, 2, 3 and 4). Refer to
7/30/2019 Analizor Gaze - Manual
32/106
GA3000 PLUS Gas Analyser Operating Manual 32
Installation
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
annotation K onDiagram 1 GA3KPLUS instrument componentsfor the 4-
20mA terminal locations.
It is important for the user to understand the analogue inputs on their loggingsystem There are two typescurrent sink and current source.
5.7.1 Wiring the GA3000 PLUS to Current Source InputsWire the outputs in accordance withWiring Diagram 1 Wiring to current sink
inputs, for optimum performance it is recommended that screened twisted
pair cable is used.
Warning For this method the operating voltage input in to the
GA3000 PLUS must not exceed 30V. Operating outside
of this value will affect the safety and ATEX certification
of the apparatus.
Note: For terminal conductor sizes please refer to sectionTerminalConductor Sizes
5.7.2 Wiring the GA3000 PLUS to Current Sink InputsWhen wiring the GA3000 PLUS to current sink inputs, loop powered isolators
must be used to convert the signal. Wire the outputs in accordance with
Wiring Diagram 2 Wiring to current sink inputsusing the recommended loop
isolator. For optimum performance it is recommended that screened twisted
pair cable is used.
Warning For this method there are two recommended loop
powered isolators:
1) Lutze WPAA 7-05262) SEM1000
Using loop powered isolators that have not been
recommended by the manufacturer will affect the
safety and ATEX certification of the apparatus.
Note: If isolation is important and the current source input method isrequired the loop powered isolators can still be used, but in an
alternative wiring arrangement. For further information please contact
7/30/2019 Analizor Gaze - Manual
33/106
33 GA3000 PLUS Gas Analyser Operating Manual
Installation
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
technical support at Geotech on +44(0)1926 338111 or email
Note: For terminal conductor sizes please refer to sectionTerminalConductor Sizes
mailto:[email protected]:[email protected]:[email protected]7/30/2019 Analizor Gaze - Manual
34/106
GA3000 PLUS Gas Analyser Operating Manual 34
Installation
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
Wiring Diagram 1 Wiring to current sink inputs
7/30/2019 Analizor Gaze - Manual
35/106
35 GA3000 PLUS Gas Analyser Operating Manual
Installation
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
Wiring Diagram 2 Wiring to current sink inputs
7/30/2019 Analizor Gaze - Manual
36/106
GA3000 PLUS Gas Analyser Operating Manual 36
Installation
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
5.7.3 4-20mA ScalingThe following table details the scaling on the 4-20mA channels:
Gas 4mA Reading 20mA ReadingCH4 0.0% 100.0%
CO2 0.0% 100.0%
O2 0.0% 25.0%
H2S 0-50ppm 0ppm 50ppm
H2S 0-200ppm 0ppm 200ppm
H2S 0-500pm 0ppm 500ppm
H2S 0-1000ppm 0ppm 1000ppm
H2S 0-5000ppm 0ppm 5000ppm
H2S 0-10000ppm 0ppm 10000ppm
H2 0-1000ppm 0ppm 1000ppm
CO 0-1000ppm 0ppm 1000ppm
5.8 Modbus Digital OutputIf Modbus digital outputs are being used, the cable must enter the cabinet via
a metal Ex e cable gland and armoured cable through one of the three
available customer cable entry points (refer to annotation A onDiagram 4 -
GA3000 PLUS connection points).
Labels identify the appropriate outputs, namely from left to right +, - and LG.
Refer to annotation J onDiagram 1 GA3KPLUS instrument componentsfor
the Modbus terminal locations.
Warning The GA3000 PLUS Modbus terminals must only be used
for standard Modbus communications; no other
connections must be made. Connections outside of this
could affect the safety and ATEX certification of the
apparatus.
5.8.1 Wiring the GA3000 PLUS Modbus OutputsWire the outputs in accordance withWiring Diagram 3 Wiring to Modbus
inputs. For optimum performance it is recommended that screened twisted
pair cable is used.
7/30/2019 Analizor Gaze - Manual
37/106
37 GA3000 PLUS Gas Analyser Operating Manual
Installation
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
Note: For terminal conductor sizes please refer to sectionTerminalConductor Sizes
When wiring the outputs the twisted pairs must be as follows:
Terminal Colour Wiring Information
Orange Signal A (RS485) +
Blue Signal B (RS485) -
White Logic Ground (0V) LG
Warning The input voltage range to the GA3000 PLUS Modbus
terminals must not exceed -7 to +12V. Operating
outside of this range could affect the safety and ATEX
certification of the apparatus.
5.8.2 Configuration of the GA3000 PLUS Modbus PortThe GA3000 PLUS Modbus port has been configured as follows:
Node Address 1
Baud Rate 38400
Parity Odd
Stop Bits 1
The PLC acts as a slave.
The protocol is MODBUS RTU.
Note: A termination resistance of 200 ohms is fitted between thepositive and negative terminals of the GA3000 PLUS and thus must be
the last connection on the bus. Similarly the first device on the bus
should have a termination resistor.
7/30/2019 Analizor Gaze - Manual
38/106
GA3000 PLUS Gas Analyser Operating Manual 38
Installation
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
5.8.3 Readable Parameters of the GA3000 PLUS PLCBelow is a table of addresses that can be read from the GA3000 PLUS PLC:
Single 16-bit Registers:
PLC Internal
Name
Type Modbus
(Hex)
Modbus
(984)
Parameter
DS100 INT 63H 400100 Last reading: YearDS101 INT 64H 400101 MonthDS102 INT 65H 400102 DayDS103 INT 66H 400103 HourDS104 INT 67H 400104 MinuteDS105 INT 68H 400105 SecondDS107 INT 6AH 400107 Sample Point 1: CH4 Reading x 10DS108 INT 6BH 400108 CO2 x 10DS109 INT 6CH 400109 O2 x 10DS110 INT 6DH 400110 INT CELL 1 (not currently supported)DS111 INT 6EH 400111 INT CELL 2DS112 INT 6FH 400112 INT CELL 3 (not currently supported)DS113 INT 70H 400113 EXT CELL 1DS114 INT 71H 400114 EXT CELL 2 (not currently supported)DS115 INT 72H 400115 Sample Point 2: CH4 Reading x 10
Repeats as per sample point 1DS123 INT 7aH 400123 Sample Point 3: CH4 Reading x 10
Repeats as per sample point 1Readings stored in two 16-bit registers as a single precision real/floating point
number:
PLC Internal
Name
Type Modbus
(Hex)
Modbus
(984)
Parameter
DF50 FLOAT 7062H 428771 Sample Point 1: CH4 readingDF51 FLOAT 7064H 428773 CO2DF52 FLOAT 7066H 428775 O2
7/30/2019 Analizor Gaze - Manual
39/106
39 GA3000 PLUS Gas Analyser Operating Manual
Installation
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
DF53 FLOAT 7068H 428777 INT CELL 1 (not currentlysupported)
DF54 FLOAT 706aH 428779 INT CELL 2DF55 FLOAT 706cH 428781 INT CELL 3 (not currently
supported)DF56 FLOAT 706eH 428783 EXT CELL 1DF57 FLOAT 7070H 428785 EXT CELL 2 (not currently
supported)DF58 FLOAT 7072H 428787 Sample Point 2
Repeats as per sample point 1DF66 FLOAT 7082H 428803 Sample Point 3
Repeats as per sample point 1
Note: Floating point numbers, consist of two 16 bit words to give a 32 bitsingle precision floating point number. The first word (e.g. 428771) holds
the sign in bit 15, the exponent in bits 14-7, part of the mantissa in bits
6-0. The remaining part of the mantissa is in the next register (e.g.
428772) bits 15 0.
7/30/2019 Analizor Gaze - Manual
40/106
GA3000 PLUS Gas Analyser Operating Manual 40
Installation
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
Wiring Diagram 3 Wiring to Modbus inputs
7/30/2019 Analizor Gaze - Manual
41/106
41 GA3000 PLUS Gas Analyser Operating Manual
Installation
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
5.9 Connecting the Gas Lines to the GA3000 PLUS5.9.1 Sample Lines
The sample should be taken and returned to suitable locations at therequired monitoring point. It is recommended that a valve is
incorporated in the assembly so that the gas can be shut off if the
sample line is removed. SeeImage 1 - Inline ball valve on sample pipe:
Image 1 - Inline ball valve on sample pipe
The line should connect to the upper surface of a horizontal pipe or on avertical standing pipe. This will prevent excessive amounts of water
entering the sample line.
The gas connections on the analyser (refer to annotation F onDiagram 4- GA3000 PLUS connection points) are 6mm stainless steel bulkhead
connectors, suitable for 6mm o/d stainless steel tubing.
The sample tube should be connected to the bulkhead connectorslabelled sample gas 1, sample gas 2 and sample gas 3, whereapplicable.
Ensure all tube connections are tight and free from leaks.5.9.1.1 Sample Gas ConditioningCare should be taken in routing the sample line, especially in cold
environments. The sample line may need insulating or even trace heat to
prevent freezing of water within the pipe. This is not part of the GA3000 PLUS
range and is the responsibility of the operator.
7/30/2019 Analizor Gaze - Manual
42/106
GA3000 PLUS Gas Analyser Operating Manual 42
Installation
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
The system incorporates a catchpot for removal of liquid and filters to help
prevent water from entering the analyser. However, additional water filtering
may be required where the sample is heavily contaminated with water.
5.9.2 Air Purge Inlet The air purge inlet requires uncontaminated air for the purge line. If uncontaminated air cannot be guaranteed at the location of the
analyser, a pipe should be connected to the air purge inlet (refer to
annotation C onDiagram 4 - GA3000 PLUS connection points) and
terminated at a point where uncontaminated air is present.
The connection on the cabinet is a BSP female bulkhead.
It is recommended that the supplied particulate filter (Geotech partnumber 2008277, see sectionGA3000 PLUS Consumable Products) is
fitted to the air purge inlet bulkhead prior to use of the system.
Ensure all tube connections are tight and free from leaks.5.9.3 Landfill Return
This line is where the measured gas from the gas analyser is exhausted.It can be returned to the process (recommended) or vented toatmosphere.
The sample gas returning to the process line should be connected to thebulkhead connector labelled landfill return.
If the sample gas is being vented to atmosphere it should be run to aposition where it is safe to do so.
The gas connection on the analyser (refer to annotation E onDiagram 4 -GA3000 PLUS connection points) is a 6mm stainless steel bulkhead
connector, suitable for 6mm o/d stainless steel tubing.
5.9.4 Vent Atmosphere This line is only used on systems with an air purge. During an air purge air is vented from this line, meaning it can be left
open to atmosphere.
If the GA3000 PLUS is being installed inside a building it is recommendedthat this line is vented to a safe area away from users.
7/30/2019 Analizor Gaze - Manual
43/106
43 GA3000 PLUS Gas Analyser Operating Manual
Installation
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
The connection on the analyser (refer to annotation G onDiagram 4 -GA3000 PLUS connection points) is a 6mm stainless steel bulkhead
connector, suitable for 6mm o/d stainless steel tubing.
Warning When performing an air purge the gas that was
previously sampled will be vented to atmosphere for a
short period of time, typically 10 seconds. This equates
to approximately 50ml of gas per air purge.
5.9.5 Drain The drain line should be run to a position where it is safe to discharge
the small amount of water that is removed from the sample gas.
Periodic inspection of the catchpot is required to check if water has beencollected. To drain the water, pause the system (please refer to section
Pause Sampling), close the sample valve(s) (refer to annotation L on
Diagram 1 GA3KPLUS instrument components) to the horizontal
position and open the drain valve (refer to annotation L onDiagram 1
GA3KPLUS instrument components) to the vertical position.
After the liquid has drained close the drain valve, open the samplevalve(s) and resume monitoring.
The connection is a BSP female bulkhead.Warning Dependent upon the application the water/liquid that
is removed may be contaminated.
Note: This line may also vent sample gas for a brief period during eachdraining operation if the sample inlet valves are not closed.
5.10 Final ChecksEnsuring the sample valves are open (refer to annotation L onDiagram 1
GA3KPLUS instrument components, these must be in the vertical
position) and the catchpot drain closed (refer to annotation L on
Diagram 1 GA3KPLUS instrument components, these must be in the
horizontal position), power up the system.
Five seconds after power is applied the gas analyser should start-up,displaying the logo and power-on self-test message.
7/30/2019 Analizor Gaze - Manual
44/106
GA3000 PLUS Gas Analyser Operating Manual 44
Installation
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
The first time the gas analyser is started after the self-test hassuccessfully completed, the first time run set-up wizard will begin
please refer to sectionFirst Time Run Set-upin this manual for details of
this process.
It is strongly recommended that a known concentration of gas is passedthrough the gas analyser to ensure that the system is still reading
accurately following installation.
For further information please contact technical support at Geotech on+44(0)1926 338111 or [email protected].
mailto:[email protected]:[email protected]:[email protected]:[email protected]7/30/2019 Analizor Gaze - Manual
45/106
45 GA3000 PLUS Gas Analyser Operating Manual
General Operational Instructions
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
6 General Operational Instructions6.1 Switching the Gas Analyser On
1) The instrument will automatically turn on five seconds after power isconnected. If not, check the connections and press the on/off key. A
long beep will be emitted followed by the GA3000 PLUS logo display.
2) The power on selftest will then commence. Refer to the messageperforming power-on self-test at the bottom of the analyser read
screen, seeScreen 1- Power on .
Screen 1- Power on self test
6.2 Power On Self Test (POST)When switched on the gas analyser will perform a pre-determined self-test
sequence taking approximately sixty seconds. The logo remains on screen until
the POST has finished. Performing Power-On Self Test is displayed on the
bottom line.
During this time many of the gas analysers functions are tested, including:
Testing of the CH4, CO2, O2, reference, barometer, internal cell andtransducers is performed continuously over a short period of time to
check for faults and instability.
The pump is switched on, and the system checked for blockages. The time and date is checked. The next service due date is checked. Valid communications to the PLCs is checked. Test whether the first time run set-up is required.
7/30/2019 Analizor Gaze - Manual
46/106
GA3000 PLUS Gas Analyser Operating Manual 46
General Operational Instructions
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
Note: After completion the logo screen is removed. If any failures occurthen the self test summary screen is shown. The self-test will
automatically reattempt every five minutes up to three times. If no
faults are found then the first time set-up or monitoring will begin. After
three failed attempts user intervention is required, however the self test
will retest every 24 hours for a further three times.
Note: If the fault is service overdue or invalid time, then the user cancontinue to the next stage by pressing key 1. The GA3000 PLUS will
continue automatically after thirty seconds if key 1 is not pressed.
6.3 First Time Run Set-up1) When switching on the gas analyser for the first time the instrument will
detect the first run conditions and run set-up mode. The GA3000 PLUS is
designed to be fully configurable by the end-user without Geotech
support or configuration.
2) The following screens are shown in this order:
Screen 2 - First time set-up introduction
3) Press the Enter key to continue.4) Set the instrument time and date. For more information about this
screen refer to section
5) Set Instrument Time and Datein this operating manual.
7/30/2019 Analizor Gaze - Manual
47/106
47 GA3000 PLUS Gas Analyser Operating Manual
General Operational Instructions
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
Screen 3 - Instrument time and date set-up
6) Configure your sample intervals for the various sample points. For moreinformation about this screen refer to section
7/30/2019 Analizor Gaze - Manual
48/106
GA3000 PLUS Gas Analyser Operating Manual 48
General Operational Instructions
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
8) Sampling Options in this operating manual.
Screen 4 - Sampling options set-up
Note: On the GA3KPLUSC without an external sensor this screen is notpresent. On the GA3KPLUSC with an external sensor this screen is for the
external sensor sampling option only as the gas analyser measures
continuously.
9) Configure alarm levels (this option can be skipped). For moreinformation about this screen refer to sectionAlarm Set-upin this
operating manual.
Screen 5 - Alarm set-up
Note: Alarms are only available on single sample point systems. For twoand three sample point systems this screen will not be present.
10)Configure 4-20mA output selection (allows you to choose which gas isoutput on which available channel) and safe values (outputs a fixed 4-
20mA and Modbus signal for each available channel so maintenance can
7/30/2019 Analizor Gaze - Manual
49/106
49 GA3000 PLUS Gas Analyser Operating Manual
General Operational Instructions
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
be performed on the system without disturbing the process). For more
information about this screen refer to sectionSafe Value and Analogue
Set-upin this operating manual.
Screen 6 - Analogue output and safe value set-up
Note: There are only four analogue output channels available. Thereforeon a five gas system one gas must be compromised for the 4-20mA
channels or alternatively the Modbus outputs used.
11)Define whether an administrator passcode is required on the system.For more information on this screen refer to sectionAdministrator
Passcodein this operating manual.
Screen 7 - Administrator passcode set-up
12)Once the first time run set-up is complete, monitoring will begin after awarm-up period of approximately sixty seconds.
7/30/2019 Analizor Gaze - Manual
50/106
GA3000 PLUS Gas Analyser Operating Manual 50
General Operational Instructions
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
6.4 Instrument Main Read Screen6.4.1 GeneralThe main gas read screen is considered to be the normal operating screenand all options are carried out from this starting point. An example screen is
shown in the following screenshot.
Screen 8 - Main read screen
Note: The actual data shown on this display will depend on the versionof the system and the option(s) that were selected at the point of sale.
A
B
C
D
E
F
G
H
K
J
I
7/30/2019 Analizor Gaze - Manual
51/106
51 GA3000 PLUS Gas Analyser Operating Manual
General Operational Instructions
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
Reference Definition
A Time and date Displays the time and date which is updated
continuously.
B Sampling Defines the current status of the overall
sampling process.
C This SP Defines the current sample point being
monitored.
D Duration Is the time remaining for the current
sampling process (see reference B). If the
system is a GA3KPLUSC this is relating to theexternal sensor.
E Cycle Is the time remaining for the overall sample
process. If the system is a GA3KPLUSC this is
relating to the external sensor.
F Flow Is the flow rate of the system pump in
ml/min.
G Use arrow keys to view
sample point
Using the scroll keys allows you to view the
previous sample points data.
H Menu Pressing key 1 brings up the menu.
Monitoring will continue whilst the menu is
displayed.
I Pause sampling Pauses the sampling process so maintenanceand can be performed.
J Gases Displays the gas reading for the available
channels. This is dependent upon the sample
point (annotation K) setting. In live mode
these readings are updated every 10-15
seconds.
K Sample point Defines the current display mode for the
7/30/2019 Analizor Gaze - Manual
52/106
GA3000 PLUS Gas Analyser Operating Manual 52
General Operational Instructions
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
gases:
Live = gas readings currently beingtaken from the current sample point
(annotation C). This is refreshed every10-15 seconds.
1, 2, 3 = the last data that was outputon the 4-20mA & Modbus channels for
the particular sample point selected.
6.4.1.1 External Sensor If fitted, the external sensor is the bottom gas displayed on the main
read screen.
If fitted, the external sensor will be sampled when the gas analyser issampling gas.
On a GA3KPLUSC system the external sensor is sampled periodicallybecause the sensor needs to be purged with air.
On a GA3KPLUSC system the duration and cycle is referring to theexternal sensor timings as the gas analyser is operating continuously.
Note: The lifespan of the sensor can be prolonged by subjecting it toshorter sample durations and longer air purges. The typical lifespan forelectrochemical sensors is 18 months in air, and this can be greatly
reduced if it is subjected to high concentrations of gases for long periods
of times.
6.4.2 Alarm Indications Alarms are only available on single sample point systems. A flashing bell next a channel indicates it is exceeding the alarm
condition.
An audible warning from the gas analyser will accompany the alarm,however this will not be audible with the main cabinet door closed.
A static bell symbol indicates that the channel is between the recoveryvalue and the trigger value.
L indicates that the alarm is, or was, active and is latched until the userclears/acknowledges it.
7/30/2019 Analizor Gaze - Manual
53/106
53 GA3000 PLUS Gas Analyser Operating Manual
General Operational Instructions
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
D indicates that the alarm has been disabled by the user. A disabledalarm causes the associated relay to be re-energised.
When an alarm is active the option for view alarms can be accessed bypressing key 3. SeeScreen 9 - Alarm indication:
Screen 9 - Alarm indication
6.4.3 Fault IndicationRefer to theFault Detectionsection in this operating manual.
6.4.4 Service Overdue IndicationWhen the gas analyser is overdue a service, the following message will appear
on the analysers screen:
Screen 10 - Service overdue message
During this time the gas analyser will continue to monitor as normal.
It is recommended that the user returns the gas analyser to the manufacturer
for a full service and calibration when this screen appears, refer to theService
section in this operating manual for further information.
7/30/2019 Analizor Gaze - Manual
54/106
GA3000 PLUS Gas Analyser Operating Manual 54
General Operational Instructions
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
Alternatively, the user can acknowledge this message simply by clicking key 1
and the message will re-appear in 28 days. This allows the user to return the
gas analyser for service at a more convenient time.
The service message cannot be turned off.6.4.5 View Alarms
As mentioned in theAlarm Indicationssection, when the GA3000 PLUS isalarming the user can view the alarm by pressing key 3.
Pressing this will take them to a new screen which can be seen inScreen11 - View alarms.
Selecting key 1 will exit the screen and return to the previous screen. If there is more than one alarm pressing key 3 will move to the next
alarm.
Selecting key 5 will disable the alarm until it is manually enabled by theuser or the alarm condition clears.
To permanently disable the alarm features refer to theAlarm Set-upsection in this operating manual.
For latched alarms that are no longer active, key 7 will be available toclear the alarm. This will re-energise the alarm relay.
Note: If during an alarm condition there is an instrument fault, the viewfault will take precedence over view alarms.
Screen 11 - View alarms
7/30/2019 Analizor Gaze - Manual
55/106
55 GA3000 PLUS Gas Analyser Operating Manual
General Operational Instructions
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
Note: The view alarm function is only available in single sample pointsystems.
6.4.6 Pause Sampling Selecting key 5 from the main read screen will pause the system
sampling process.
In addition, the 4-20mA and Modbus outputs are frozen to their lastknown value.
This option can be used when maintenance is required on the plantequipment and will stop unnecessary and false alarms from occurring.
Selecting this option causes the sampling to stop until the user manuallyresumes the operation (pressing key 1 to resume).
Pressing key 3 in this screen will set the 4-20mA and Modbus outputsto the pre-set safe values (refer to theSafe Value and Analogue Set-up
section in this operating manual for further information).
When sampling is paused the following screen is displayed:
Screen 12 - Pause sampling screen
6.4.6.1 Safe Values Safe values allow the user to set the 4-20mA and Modbus outputs to
pre-set values.
This is suitable to prevent erroneous errors or alarm conditions in theusers remote system(s) whilst maintenance is being performed on the
GA3000 PLUS or any other plant equipment.
7/30/2019 Analizor Gaze - Manual
56/106
GA3000 PLUS Gas Analyser Operating Manual 56
General Operational Instructions
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
The safe values currently pre-set within the system can be seen in thebottom section of the pause sampling screen, seeScreen 12 - Pause
sampling screen.
The safe values can be edited from the pause screen by selecting key 5.This takes you to the safe value and analogue setup screen. Refer to
theSafe Value and Analogue Set-upsection in this operating manual for
information on how to edit the values.
6.4.6.1.1 Safe Value and Analogue Set-upThe safe value and analogue set-up allows the user to define the analogue
outputs for the measured gas and the safe values to be output when sampling
is paused. The screen can be seen in the following image:
Screen 13 - Safe value and analogue output set-up
1) From the pause sampling screen press key 5 to enter the safe value andanalogue set-up screen.
2) Select key 7 to edit settings the CH4 output will become highlighted.3) Using the scroll keys choose the required field that needs editing and
select using the enter key.
Edit 4-20mA Output Channels
4) A list of five options will be displayed. The user can select which outputchannel they would like the selected gas to be assigned to. See following
Screen 14 - Edit gas 4-20mA output channel.
7/30/2019 Analizor Gaze - Manual
57/106
57 GA3000 PLUS Gas Analyser Operating Manual
General Operational Instructions
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
Screen 14 - Edit gas 4-20mA output channel
5) Using the scroll keys choose the required output channel for the gas andselect using the enter key.
Note: There are only four available analogue output channels. Thereforeon a five gas system, one gas must be set to not from the options list.
6) Once complete, press key 7 to save changes.Edit Safe Reading
7) When attempting to edit the safe reading, a ? will become available onthe selected gas channel. See the following screen:
Screen 15 - Edit safe reading
8) Type in the required value to be output on the 4-20mA and Modbuschannel when in fault or pause sampling mode and store with the enter
key.
Note: The value entered here is the channels reading in percent or ppmas appropriate.
7/30/2019 Analizor Gaze - Manual
58/106
GA3000 PLUS Gas Analyser Operating Manual 58
General Operational Instructions
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
9) Once complete, press key 7 to save changes.6.5 Main MenuThe main menu enables the operator to select options to set-up specific
parameters and perform operational tasks. The main menu can be seen in
Screen 16 - Main menu:
Screen 16 - Main menu
The main menu can be accessed from the main read screen by pressingkey 1.
Options can be selected by pressing the relevant option number on thegas analyser keypad.
Note: The keyboard is scanned continuously. The menu is displayed forthirty seconds and if no options are chosen the main read screen will
reappear.
6.5.1
Exit Menu and Resume SamplingBy pressing key 1 the user will exit from the main menu and return the main
read screen.
6.5.2 Alarm Set-up Note: Alarm set-up is only available on single sample point systems.
The system has two relay outputs available for alarm purposes.
7/30/2019 Analizor Gaze - Manual
59/106
59 GA3000 PLUS Gas Analyser Operating Manual
General Operational Instructions
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
1) From the main menu press key 2 to enter the alarm set-up screen, asshown inScreen 5 - Alarm set-up.
2) Select key 7 to edit settings the channel 1 gas will becomehighlighted.
3) Once in edit mode the following keys become available:a. Key 3 becomes available to enable or disable the alarm and relay.b. Key 7 becomes available to save and exit.c. Key 1 becomes available to cancel.
4) Press the enter key to edit the channel, a list of gases will be providedas displayed inScreen 17 - Edit alarm channel:
Screen 17 - Edit alarm channel
5) Choose the required gas (e.g. H2S) by using the enter key.6) Each alarm relay can be configured independently. For each relay the
following parameters can be specified by using the enter key to edit:
a. Gas channelb. Trigger direction (above or below the specified value)c. Trigger valued. Mode (latching or normal)e. Recovery value
7) Pressing key 3 whilst in edit mode will enable or disable an alarm. AnX symbol to the left of the channel indicates that a channel is disabled,
seeScreen 18 - Alarm channel disabledfor an example:
7/30/2019 Analizor Gaze - Manual
60/106
GA3000 PLUS Gas Analyser Operating Manual 60
General Operational Instructions
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
Screen 18 - Alarm channel disabled
Note: Alarm mode definitions: with a latching alarm the relay outputremains de-energised until the user selects the option to cancel thealarm. With a normal alarm the relay output remains de-energised until
the value reaches the recovery value.
6.5.3 User CalibrationThe ability has been provided to perform a calibration on each available gas
channel. This calibration ensures the accuracy of the analyser in its current
operating condition. To ensure optimum performance please ensure your
analyser is returned for service and calibration on time.
1) From the main menu press key 3 to select user calibration. Note: For more information on user calibration, please refer to section
Calibrationin this operating manual.
6.5.4 Set Instrument Time and DateThis option enables the user to set the time and date on the gas analyser.
1) From the main menu press key 4 for set time and date, the followingscreen is displayed:
7/30/2019 Analizor Gaze - Manual
61/106
61 GA3000 PLUS Gas Analyser Operating Manual
General Operational Instructions
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
Screen 19 - Set time and date
2) Press key 1 to exit without saving any changes.3) Press key 3 to toggle the date format between dd/MM/yy and
MM/dd/yy.
4) Press the enter key to edit the time and date. A ? will becomeavailable in the time and date box.
Screen 20 - Time and date edit
5) Enter the appropriate time and date and once complete press the enterkey to store. The gas analyser will return to the main read screen.
Note: When returning to the main read screen the sampling process willbegin from the start. This is because the timing intervals are dictated
from the time of the gas analyser.
7/30/2019 Analizor Gaze - Manual
62/106
GA3000 PLUS Gas Analyser Operating Manual 62
General Operational Instructions
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
6.5.5 Sampling OptionsThis option allows the user to define the sample timings for each sample point
and the time between a full set of readings.
1) From the main menu press key 5 for sampling options, the followingscreen is displayed:
Screen 21 - Sampling options
2) Select key 1 to exit without saving any changes.3) Select key 7 to edit the settings and the duration on SP1 will become
highlighted.
4) Use the scroll keys to navigate to the required field that needs changingand select using the enter key. A ? will become present in the selected
field.
Note: Fields available for editing are:1. The duration that each sample point is sampled for.2. The duration that the system is air purged for after each sample point.3.
The interval between the start of sample point 1 being monitored on aset of sample points.
Note: On the GA3KPLUSC without an external sensor this screen is notavailable. On the GA3KPLUSC with an external sensor this screen is for
the external sensor sampling option only as the gas analyser measures
continuously.
5) Key in the new value for the required field and press enter whencomplete.
7/30/2019 Analizor Gaze - Manual
63/106
63 GA3000 PLUS Gas Analyser Operating Manual
General Operational Instructions
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
6) Press key 7 to save the changes made. The gas analyser will return tothe main read screen.
Note: When returning to the main read screen the sampling process willbegin from the start; this is because the timing intervals are dictated
from the time of the gas analyser.
6.5.6 View Event LogThis option allows the user to view the gas analyser event log to assist in
problem solving and fault detection.
1) From the main menu press key 6 to view the event log, the followingscreen will be displayed:
Screen 22 - View event log
Note: Date values are displayed in the format yy/MM/dd.2) Press key 1 to exit and return to the main read screen.3) Press key 3 to display the oldest record first.4) Press key 5 to delete all the records in the event log.5) Press key 7 to provide more data on the events. This information will
likely be used when contacting the technical support department.
6) After pressing key 7, pressing key 4 and 6 will allow the user to cycleround the data values associated with the event.
Note: For more information please refer to theEvent Logsection in thisoperating manual.
7/30/2019 Analizor Gaze - Manual
64/106
GA3000 PLUS Gas Analyser Operating Manual 64
General Operational Instructions
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
6.5.7 Display ContrastThe GA3000 PLUS automatically adjusts the screen contrast to maintain a
normal viewing contrast according to the current read-out temperature. This
can also be adjusted by the user.
1) From the main menu press key 7 to adjust the display contrast, thefollowing screen will be displayed:
Screen 23 - Adjust display contrast
2) Press key 1 to exit and return to the main read screen without savingchanges.
3) Use the left and right scroll keys (keys 4 and 6 respectively) to adjustthe contrast.
4) Press the enter key to store the new contrast.6.5.8 System InformationThis screen provides information regarding the gas analyser, including the
serial number and model number.
1) From the main menu press key 8 to view the system information, thefollowing screen will be displayed:
7/30/2019 Analizor Gaze - Manual
65/106
65 GA3000 PLUS Gas Analyser Operating Manual
General Operational Instructions
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
Screen 24 - System information
2) Press key 1 to exit.6.5.9 Administrator PasscodeAn administrator passcode helps prevent unauthorised users modifying system
settings. The default setting at point of manufacture is for this to be disabled.
The following options are administrator passcode protected:
Time and date setting Alarm set-up Sampling options 4-20mA configuration and safe values Disabling or changing the administrator passcode
1) From the main menu press key 9 to enter the administrator passcodeoptions, the following screen will appear:
Screen 25 - Administrator passcode set-up
7/30/2019 Analizor Gaze - Manual
66/106
GA3000 PLUS Gas Analyser Operating Manual 66
General Operational Instructions
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
2) Press key 1 to exit and return to the main menu without savingchanges.
3) Press key 5 to change the passcode and a ? will appear in the currentpasscode box.
4) Key in the new code and press enter to confirm Note: The length of the passcode must be 1-6 digits in length.
5) Press the enter key to save the changes you have made and return tothe main menu.
Warning Please ensure you do not lose the passcode as it cannot
be recovered.
When trying to change a setting that is administrator passcode protected the
following screen will appear:
Screen 26 - Administrator passcode required
Key in the required passcode and press the enter key when complete allows
the user to then make changes to the required setting.
7/30/2019 Analizor Gaze - Manual
67/106
67 GA3000 PLUS Gas Analyser Operating Manual
General Operational Instructions
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
6.6 GA3000 PLUS Sampling Process6.6.1 GA3KPLUS1
Start
Complete first time
run set-up
Sample point1 monitored
4-20mA and
Modbus
outputsupdated
Wait
interval
duration
Sample point
1 air purged
6.6.2 GA3KPLUS2 and GA3KPLUSDStart
Complete first time
run set-up
Sample point
1 monitored
Wait
interval
duration
Sample point
2 monitored
Sample point
1 air purged
Sample point
2 air purged
4-20mA and
Modbus
outputs
updated
4-20mA and
Modbus
outputs
updated
7/30/2019 Analizor Gaze - Manual
68/106
GA3000 PLUS Gas Analyser Operating Manual 68
General Operational Instructions
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
6.6.3 GA3KPLUS3Start
Complete first time
run set-up
Sample point
1 monitored
Wait
interval
duration
Sample point
2 monitored
Sample point
3 monitored
Sample point
1 air purged
Sample point
2 air purged
Sample point
3 air purged
4-20mA and
Modbus
outputsupdated
4-20mA and
Modbusoutputs
updated
4-20mA and
Modbus
outputs
updated
7/30/2019 Analizor Gaze - Manual
69/106
69 GA3000 PLUS Gas Analyser Operating Manual
General Operational Instructions
OMGA3KPLUS1.00 Copyright 2012 Geotechnical Instruments (UK) Limited
6.6.4 GA3KPLUSCStart
Complete first time
run set-up
Sample point1 monitored
continuously
4-20mA and
Modbus
outputs
updated
Waitinterval
duration
CH4, CO2, O2
Waitinterval
duration
External
Sensor
Externalsensor
monitored
4-20mA and
Modbus
outputs
updated
Externalsensor air
purged
6.7 Switching the Gas Analyser OffTo switch off the gas analyser, press and hold down the On/Off key for
approximately five seconds. This will close all solenoids and switch off the
pumps. Power remains on to the rest of the system. The system fault signal is
de-energised when the analyser is switched off.
The instrument can also be turned off at the mains power supply.
7/30/2019 Analizor Gaze - Manual
70/106
GA3000 PLUS Gas Analyser Operating Manual 70
Calibration
Copyright 2012 Geotechnical Instruments (UK) Limited OMGA3KPLUS1.00
7 Calibration7.1 Calibration IntroductionThe GA3000 PLUS range of gas analysers are carefully calibrated at
manufacture and when returned for service. However, it is sometimes
desirable to carry out a calibration process between services.
This section outlines the correct procedures to enable the site engineer to field
calibrate the gas analyser.
Note: This does not replace the factory service and calibration. If thiscalibration is completed incorrectly it may decrease the accuracy of the
gas analyser.
CH4, CO2 and O2 can be measured by all GA3000 PLUS range analysers as
standard; these channels can be user calibrated. There are also optional
additional gases to these which are specified at manufacture; these too can be
calibrated. This section will describe in detail how to calibrate the three
standard gas channels plus the additional internal and external sensors.
Two important terms that are used within this section are Zero and Span.
Zero: The point at which the gas analyser is calibrated when there is none of
the target gas present.
Span: The point at which the gas analyser is calibrated when a known quantity
of the target gas is present.
7.2 Required Calibration Equipment7.2.1 GasUser calibration of a gas analyser will greatly improve the data accuracy in the
range of the calibration gases used, and the environmental conditions for
which the gas analyser is calibrated in. This may cause less accurate readings of
concentrations outside this calibrated range. Users should select the correct
calibrati