AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY. CHAPTER 13, 14. PROBLEM SET. Chapter 13 13 -5, -7, -8, -10, -11, -12, - 22 Chapter 14 14-1, 14-6, 14-7, 14-11 Due Tues., May 1, 2012. FIRST LONG EXAMINATION Ch 135 A/B/C. May 3, Thursday. 1:30 – 4 :00 PM, C114 - PowerPoint PPT Presentation

Citation preview

Page 1: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION

SPECTROSCOPY

CHAPTER 13, 14

1

Page 2: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

2

Chapter 1313-5, -7, -8, -10, -11, -12, -22Chapter 1414-1, 14-6, 14-7, 14-11

Due Tues., May 1, 2012

PROBLEM SET

Page 3: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

3

May 3, Thursday. 1:30 – 4:00 PM, C114Coverage: Chapters 1, 6-10, 13-17 (5th Ed., Skoog)

FIRST LONG EXAMINATIONCh 135 A/B/C

Page 4: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

4

In this chapter, absorption by molecules, rather than atoms, is considered.

Absorption in the ultraviolet and visible regions occurs due to electronic transitions from the ground state to excited state.

Broad band spectra are obtained since molecules have vibrational and rotational energy levels associated with electronic energy levels.

The signal is either absorbance or percent transmittance of the analyte solution.

Page 5: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

5

Absorption measurements based upon ultraviolet and visible radiation find widespread application for the quantitative determination of a large variety species.

Beer’s Law:A = -log T = logP0/P = bcA = absorbance = molar absorptivity [M-1 cm-1]c = concentration [M]P0 = incident powerP = transmitted power (after passing through sample)

AN INTRODUCTION TO ULTRAVIOLET/VISIBLEMOLECULAR ABSORPTION SPECTROMETRY

Page 6: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

6

Page 7: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

7

Page 8: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

8

Page 9: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

MEASUREMENT OF TRANSMITTANCE AND ABSORBANCE:The power of the beam transmitted by the analyte solution is usually compared with the power of the beam transmitted by an identical cell containing only solvent. An experimental transmittance and absorbance are then obtained with the equations.

P0 and P refers to the power of radiation after it has passed through the solvent and the analyte.

9

TPP

PP

so lu tion

so lven t

0A

PP

PP

so lven t

so lu tion log log

0

Page 10: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

10

BEER’S LAW AND MIXTURES

Each analyte present in the solution absorbs light! The magnitude of the absorption depends on its

A total = A1+A2+…+An A total = 1bc1+2bc2+…+nbcn

If 1 = 2 = n then simultaneous determination is impossible

Need nl’s where ’s are different to solve the mixture

Page 11: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

11

Real limitations Chemical deviations Instrumental deviations

LIMITATIONS TO BEER’S LAW

Page 12: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

12

Beer’s law is good for dilute analyte solutions only. High concentrations (>0.01M) will cause a negative error since as the distance between molecules become smaller the charge distribution will be affected which alter the molecules ability to absorb a specific wavelength.

The same phenomenon is also observed for solutions with high electrolyte concentration, even at low analyte concentration. The molar absorptivity is altered due to electrostatic interactions.

REAL LIMITATIONS

Page 13: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

13

In the derivation of Beer’s law we have introduced a constant (). However, e is dependent on the refractive index and the refractive index is a function of concentration.

Therefore, will be concentration dependent. However, the refractive index changes very slightly for dilute solutions and thus we can practically assume that is constant.

REAL LIMITATIONS

Page 14: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

14

In rare cases, the molar absorptivity changes widely with concentration, even at dilute solutions. Therefore, Beer’s law is never a linear relation for such compounds, like methylene blue.

REAL LIMITATIONS

Page 15: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

15

This factor is an important one which largely affects linearity in Beer’s law.

It originates when an analyte dissociates, associates, or reacts in the solvent.

For example, an acid base indicator when dissolved in water will partially dissociate according to its acid dissociation constant:

CHEMICAL DEVIATIONS

Page 16: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

16

Page 17: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

17

Page 18: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

18

Page 19: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

19

Page 20: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

20

Chemical deviations from Beer’s law for unbuffered solutions of the indicator Hln. Note that there are positive deviations at 430 nm and negative deviations at 570 nm. At 430 nm, the absorbance is primarily due to the ionized In- form of the indicator and is proportional to the fraction ionized, which varies nonlinearly with the total indicator concentration. At 570 nm, the absorbance is due principally to the undissociated acid Hln, which increases nonlinearly with the total concentration.

Page 21: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

21

CALCULATED ABSORBANCE DATA FOR VARIOUS INDICATOR CONCENTRATIONS

Page 22: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

22

Page 23: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

23

Beer’s law is good for monochromatic light only since is wavelength dependent. It is enough to assume a dichromatic beam passing through a sample to appreciate the need for a monochromatic light. Assume that the radiant power of incident radiation is Po and Po’ while transmitted power is P and P’. The absorbance of solution can be written as:

INSTRUMENTAL DEVIATIONS

Page 24: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

24

Page 25: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

25

Page 26: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

26

The effect of polychromatic radiation on Beer’s law. In the spectrum at the top, the absorptivity of the analyte is nearly constant over Band A from the source. Note in the Beer’s law plot at the bottom that using Band A gives a linear relationship. In the spectrum, Band B corresponds to a region where the absorptivity shows substantial changes. In the lower plot, note the dramatic deviation from Beer’s law that results.

Page 27: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

27

Page 28: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

28

Therefore, the linearity between absorbance and concentration breaks down if incident radiation was polychromatic.

In most cases with UV-Vis spectroscopy, the effect small changes in wavelengths is insignificant since differs only slightly; especially at the wavelength maximum.

Page 29: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

29

STRAY RADIATION Stray radiation resulting from scattering or various

reflections in the instrument will reach the detector without passing through the sample. The problem can be severe in cases of high absorbance or when the wavelengths of stray radiation is in such a range where the detector is highly sensitive as well as at wavelengths extremes of an instrument. The absorbance recorded can be represented by the relation:A = log (Po + Ps)/(P + Ps)

Where; Ps is the radiant power of stray radiation.

Page 30: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

30

Page 31: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

31

INSTRUMENTAL NOISE AS A FUNCTION IN TRANSMITTANCE

Page 32: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

32

Page 33: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

33

Page 34: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

34

Therefore, an absorbance between 0.2-0.7 may be advantageous in terms of a lower uncertainty in concentration measurements. At higher or lower absorbances, an increase in uncertainty is encountered. It is therefore advised that the test solution be in the concentration range which gives an absorbance value in the range from 0.2-0.7 for best precision.

Page 35: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

35

Effect of bandwidth on spectral detail for a sample of benzene vapor. Note that as the spectral bandwidth increases, the fine structure in the spectrum is lost. At a bandwidth of 10 nm, only a broad absorption band is observed.

EFFECT OF BANDWIDTH

Page 36: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

36

Effect of slit width (spectral bandwidth) on peak heights. Here, the sample was s solution of praseodymium chloride. Note that as the spectral bandwidth decreases by decreasing the slit width from 1.0 mm to 0.1 mm, the peak heights increase.

Page 37: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

37

Wavelength extremes of an instrument are dependent on type of source, detector and optical components used in the manufacture of the instrument. Outside the working range of the instrument, it is not possible to use it for accurate determinations. However, the extremes of the instrument are very close to the region of invalid instrumental performance and would thus be not very accurate. An example may be a visible photometer which, in principle, can be used in the range from 340-780 nm. It may be obvious that glass windows, cells and prism will start to absorb significantly below 380 nm and thus a decrease in the incident radiant power is significant.

EFFECT OF SCATTERED RADIATION AT WAVELENGTH EXTREMES OF AN

INSTRUMENT

Page 38: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

38

Spectrum of cerium (IV) obtained with a spectrophotometer having glass optics (A) and quartz optics (B). The false peak in A arises from transmission of stray radiation of longer wavelengths.

B: UV-VIS spectrophotometerA: VIS spectrophotometer

EFFECT OF SCATTERED RADIATION

Page 39: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

39

The output from the source at the low wavelength range is minimal. Also, the detector has best sensitivities around 550 nm which means that away up and down this value, the sensitivity significantly decrease. However, scattered radiation, and stray radiation in general, will reach the detector without passing through these surfaces as well as these radiation are constituted from wavelengths for which the detector is highly sensitive. In some cases, stray and scattered radiation reaching the detector can be far more intense than the monochromatic beam from the source. False peaks may appear in such cases and one should be aware of this cause of such peaks.

Page 40: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

40

INSTRUMENTATION Light source l - selectionSample container DetectorSignal processing Light Sources (commercial instruments)D2 lamp (UV: 160 – 375 nm)W lamp (vis: 350 – 2500 nm)

Page 41: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

41

SOURCESDeuterium and hydrogen lamps (160 – 375 nm)

D2 + Ee → D2* → D’ + D’’ + h

Excited deuterium molecule with fixed quantized energy

Dissociated into two deuterium atoms with different kinetic energies

Ee = ED2* = ED’ + ED’’ + hvEe is the electrical energy absorbed by the molecule. ED2* is the fixed quantized energy of D2*, ED’ and ED’’ are kinetic energy of the two deuterium atoms.

Page 42: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

42

Page 43: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

43

(a) A deuterium lamp of the type used in spectrophotometers and (b) its spectrum. The plot is of irradiance Eλ (proportional to radiant power)

versus wavelength. Note that the maximum intensity occurs at ~225 nm.Typically, instruments switch from deuterium to tungsten at ~350 nm.

Deuterium lampUV region

SOURCES

Page 44: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

44

(a) A tungsten lamp of the type used in spectroscopy and its spectrum (b). Intensity of the tungsten source is usually quite low at wavelengths shorter

than about 350 nm. Note that the intensity reaches a maximum in the near-IR

region of the spectrum (~1200 nm in this case).

VISIBLE AND NEAR-IR REGION

Page 45: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

45

The tungsten lamp is by far the most common source in the visible and near IR region with a continuum output wavelength in the range from 350-2500 nm. The lamp is formed from a tungsten filament heated to about 3000 oC housed in a glass envelope. The output of the lamp approaches a black body radiation where it is observed that the energy of a tungsten lamp varies as the fourth power of the operating voltage.

Page 46: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

46

Tungsten halogen lamps are currently more popular than just tungsten lamps since they have longer lifetime. Tungsten halogen lamps contain small quantities of iodine in a quartz envelope. The quartz envelope is necessary due to the higher temperature of the tungsten halogen lamps (3500 oC). The longer lifetime of tungsten halogen lamps stems from the fact that sublimed tungsten forms volatile WI2 which redeposits on the filament thus increasing its lifetime. The output of tungsten halogen lamps are more efficient and extend well into the UV.

Page 47: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

47

SOURCESTungsten lamps (350-2500 nm)

Why add I2 in the lamps?W + I2 → WI2

Low limit: 350 nm

Low intensityGlass envelope

Page 48: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

48

Xenon Arc Lamps Passage of current through an atmosphere of high

pressured xenon excites xenon and produces a continuum in the range from 200-1000 nm with maximum output at about 500 nm. Although the output of the xenon arc lamp covers the whole UV and visible regions, it is seldom used as a conventional source in the UV-Vis. The radiant power of the lamp is very high as to preclude the use of the lamp in UV-Vis instruments. However, an important application of this source will be discussed in luminescence spectroscopy which will be discussed later

Page 49: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

49

Page 50: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

50

Sample containers are called cells or cuvettes and are made of either glass or quartz depending on the region of the electromagnetic spectrum.

The path length of the cell varies between 0.1 and 10 cm but the most common path length is 1.0 cm. Rectangular cells or cylindrical cells are routinely used.

In addition, disposable polypropylene cells are used in the visible region. The quality of the absorbance signal is dependent on the quality of the cells used in terms of matching, cleaning as well as freedom from scratches.

SAMPLE CONTAINERS

Page 51: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

51

INSTRUMENTAL COMPONENTS Source l - selection (monochromators) Sample holders Cuvettes (b = 1 cm typically)

1. Glass (Vis)2. Fused silica (UV+Vis)

Detectors Photodiodes PMTs

Page 52: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

52

SINGLE BEAM

Place cuvette with blank (i.e., solvent) in instrument and take a reading 100% T

Replace cuvette with sample and take reading % T for analyte (from which absorbance is calc’d)

Page 53: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

53

Most common spectrophotometer: Spectronic 20.INSTRUMENTATION

1. On/Off switch and zero transmission adjustment knob

2. Wavelength selector/Readout

3. Sample chamber

4. Blank adjustment knob

5. Absorbance/Transmittance scale

Page 54: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

54

Page 55: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

55

End view of the exit slit of the Spectronic 20 spectrophotometer pictured earlier

Page 56: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

56

SINGLE-BEAM INSTRUMENTS FOR THE ULTRAVIOLET/VISIBLE REGION

Page 57: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

57

SINGLE-BEAM COMPUTERIZED SPECTROPHOTOMETERS

Inside of a single-beam spectrophotometer connected to a computer.

Page 58: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

58

Instrumental designs for UV-visible photometers or spectrophotometers. In (a), a single-beam instrument is shown. Radiation from the filter or monochromator passes through either the reference cell or the sample cell before striking the photodetector.

TYPES OF INSTRUMENTS

Page 59: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

59

Page 60: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

60

DOUBLE BEAM (most commercial instruments)o Light is split and directed towards both reference

cell (blank) and sample cello Two detectors; electronics measure ratio (i.e.,

measure/calculate absorbance)o Advantages:

Compensates for fluctuations in source intensity and drift in detector

Better design for continuous recording of spectra

Page 61: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

61

GENERAL INSTRUMENT DESIGNSDOUBLE BEAM: IN-SPACE

Needs two detectors

Page 62: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

62

GENERAL INSTRUMENT DESIGNSDOUBLE BEAM: IN-TIME

Page 63: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

63

Page 64: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

64

MERITS OF DOUBLE BEAM INSTRUMENTS

• Compensate for all but the most short term fluctuation in radiant output of the source

• Compensate drift in transducer and amplifier

• Compensate for wide variations in source intensity with wavelength

Page 65: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

65

In all photometers and scanning spectrophotpmeters described above, the cell has been positioned after the monochromators. This is important to decrease the possibility of sample photodecomposition due to prolonged exposure to all frequencies coming from the source. However, the sample is positioned before the monochromator in multichannel instruments like a photodiode array spectrophotometer. This can be done without fear of photodecomposition since the sample exposure time is usually less than 1 s. Therefore, it is now clear that in UV-Vis where photodecomposition of samples can take place, the sample is placed after the monochromators in scanning instruments while positioning of the sample before the monochromators is advised in multichannel instruments.

LOCATION OF SAMPLE CELL

Page 66: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

66

MULTICHANNEL INSTRUMENTS Photodiode array detectors used (multichannel

detector, can measure all wavelengths dispersed by grating simultaneously).

Advantage: scan spectrum very quickly “snapshot” < 1 sec.

Powerful tool for studies of transient intermediates in moderately fast reactions.

Useful for kinetic studies. Useful for qualitative and quantitative

determination of the components exiting from a liquid chromatographic column.

Page 67: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

67

MULTI-CHANNEL DESIGN

Page 68: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

68

Page 69: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

69

A multichannel diode-array spectrophotometer, the Agilent Technologies 8453.

Page 70: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

70

These are the same as conventional single beam instruments but the beam from the monochromators is guided through a bifurcated optical fiber to the sample container where absorption takes place. The attenuation in reflected beam at the specified wavelength is thus measured and related to concentration of analyte in the sample.

A fiber optic cable can be referred to as a light pipe where light can be transmitted by the fiber without loss in intensity (when light hits the internal surface of the fiber at an angle larger than a critical angle). Therefore, fiber optics can be used to transmit light for very long distances without losses. A group of fibers can be combined together to form a fiber optic cable or bundle. A bifurcated fiber optic cable has three terminals where fibers from two separate cables are combined at one end to form the new configuration.

PROBE TYPE INSTRUMENTS

Page 71: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

71

Page 72: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

72

FIBER OPTIC PROBE

Page 73: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

73

Page 74: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

74

The instrument in this case has two gratings where the light beam leaving the first monochromators at a specified wavelength is directed to the second grating. This procedure results in better spectral resolution as well as decreased scattered radiation. However, double dispersing instruments are expensive and seem to offer limited advantages as compared to cost; especially in the UV-Vis region where exact wavelength may not be crucial.

DOUBLE DISPERSING INSTRUMENTS

Page 75: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

75

Optical diagram of the Varian Cary 300 double-dispersing spectrophotometer. A second monochromator is added immediately after the source.

Page 76: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

76

Page 77: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

77

MOLAR ABSORPTIVITIES

= 8.7 x 10 19 P A A: cross section of molecule in cm2 (~10-

15) P: Probability of the electronic transition

(0-1) P>0.1-1 allowable transitions P<0.01 forbidden transitions

Page 78: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

78

MOLECULAR ABSORPTION M + h M* (absorption 10-8 sec) M* M + heat (relaxation process) M* A+B+C (photochemical decomposition) M* M + h (emission)

Page 79: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

79

VISIBLE ABSORPTION SPECTRA

Page 80: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

80

The absorption of UV-visible radiation generally results from excitation of bonding electrons.

can be used for quantitative and qualitative analysis

Page 81: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

81

Molecular orbital is the nonlocalized fields between atoms that are occupied by bonding electrons. (when two atom orbitals combine, either a low-energy bonding molecular orbital or a high energy antibonding molecular orbital results.)

Sigma () orbitalThe molecular orbital associated with single bonds in organic compounds

Pi () orbitalThe molecular orbital associated with parallel overlap of atomic P orbital.

n electronsNo bonding electrons

Page 82: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

82

What electrons can we use for these transitions?

MOLECULAR TRANSITIONS FOR UV-VISIBLE ABSORPTIONS

Page 83: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

83

MO Diagram for Formaldehyde (CH2O)

HC

HO

= =n =

Page 84: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

84

SINGLET VS. TRIPLET

In these diagrams, one electron has been excited (promoted) from the n to * energy levels (non-bonding to anti-bonding).

One is a Singlet excited state, the other is a Triplet.

Page 85: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

85

TYPE OF TRANSITIONS σ → σ*

High energy required, vacuum UV rangeCH4: l = 125 nm

n → σ*Saturated compounds, CH3OH etc (l = 150 - 250

nm) n → * and → *

Mostly used! l = 200 - 700 nm

Page 86: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

86

EXAMPLES OF UV-VISIBLE ABSORPTIONS

LOW!

Page 87: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

87

UV-VISIBLE ABSORPTION CHROMOPHORES

Page 88: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

88

EFFECTS OF SOLVENTS Blue shift (n- *) (Hypsocromic shift)

Increasing polarity of solvent better solvation of electron pairs (n level has lower E)

peak shifts to the blue (more energetic) 30 nm (hydrogen bond energy)

Red shift (n- * and –*) (Bathochromic shift) Increasing polarity of solvent, then increase the attractive

polarization forces between solvent and absorber, thus decreases the energy of the unexcited and excited states with the later greater

peaks shift to the red 5 nm

Page 89: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

89

UV-VISIBLE ABSORPTION CHROMOPHORES

Page 90: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

90

TYPICAL UV ABSORPTION SPECTRA

Chromophores?

Page 91: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

91

THE EFFECTS OF SUBSTITUTION

Auxochrome function group

Auxochrome is a functional group that does not absorb in UV region but has the effect of shifting chromophore peaks to longer wavelength as well as increasing their intensity.

Page 92: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

92

solvents are your “container”They need to be transparent and do not erase the

fine structure arising from the vibrational effectsPolar solvents generally tend to cause this problem

Same solvent must beUsed when comparingabsorption spectra foridentification purpose.

Page 93: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

93

Page 94: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

94

SUMMARY OF TRANSITIONS FOR ORGANIC MOLECULES

* transition in vacuum UV (single bonds)n * saturated compounds with non-bonding electrons

l ~ 150-250 nm ~ 100-3000 ( not strong)

n *, * requires unsaturated functional groups (eq. double bonds) most commonly used, energy good range for UV/Vis

l ~ 200 - 700 nmn * : ~ 10-100 *: ~ 1000 – 10,000

Page 95: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

95

List of common chromophores and their transitions

Page 96: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

96

ORGANIC COMPOUNDS

Most organic spectra are complex Electronic and vibration transitions superimposed Absorption bands usually broad Detailed theoretical analysis not possible, but

semi-quantitative or qualitative analysis of types of bonds is possible.

Effects of solvent & molecular details complicate comparison

Page 97: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

97

If greater then one single bond apart - are relatively additive (hyperchromic

shift) - l constant

CH3CH2CH2CH=CH2 lmax= 184 max = ~10,000

CH2=CHCH2CH2CH=CH2 lmax=185 max = ~20,000

If conjugated- shifts to higher l’s (red shift)

H2C=CHCH=CH2 lmax=217 max = ~21,000

RULE OF THUMB FOR CONJUGATION

Page 98: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

98

SPECTRAL NOMENCLATURE OF SHIFTS

Page 99: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

99

WHAT ABOUT INORGANICS? Common anions n * nitrate (313 nm), carbonate (217 nm) Most transition-metal ions absorb in the UV/Vis region. In the lanthanide and actinide series the absorption process

results from electronic transitions of 4f and 5f electrons. For the first and second transition metal series the absorption

process results from transitions of 3d and 4d electrons. The bands are often broad. The position of the maxima are strongly influenced by the

chemical environment. The metal forms a complex with other stuff, called ligands. The

presence of the ligands splits the d-orbital energies.

Page 100: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

100

TRANSITION METAL IONS

Page 101: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

101

CHARGE-TRANSFER-ABSORPTIONA charge-transfer complex consists of an electron-donor group bonded to an electron acceptor. When this product absorbs radiation, an electron from the donor is transferred to an orbital that is largely associated with the acceptor.

• Large molar absorptivity (εmax >10,000)• Many organic and inorganic complexes

Page 102: AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY

102