32
An Autonomous Inexpensive Robust CO 2 Analyzer (AIRCOA) Britton Stephens, Andrew Watt, and Gordon Maclean National Center for Atmospheric Research, Boulder, Colorado, USA

An Autonomous Inexpensive Robust CO 2 Analyzer (AIRCOA) Britton Stephens, Andrew Watt, and Gordon Maclean National Center for Atmospheric Research, Boulder,

Embed Size (px)

Citation preview

An Autonomous Inexpensive Robust CO2 Analyzer (AIRCOA)

Britton Stephens, Andrew Watt, and Gordon MacleanNational Center for Atmospheric Research, Boulder, Colorado, USA

Potential source of bias AIRCOA solutionRelating to WMO CO2 Scale Dedicated CO2 and O2 calibration transfer facility

Short-term IRGA noise Average for 2 minutes to get better than 0.1 ppm precision

Drift in IRGA sensitivity 4-hourly 4-point calibrations and 30-minute 1-point calibrations

IRGA pressure sensitivity Automated 4-hourly pressure sensitivity measurements

IRGA temperature sensitivity 30-minute 1-point calibrations, temperature control at some sites

Incomplete drying of air Slow enough flow (100 sccm), two 96” Nafion driers, downstream humidity sensor to verify performance

Drying system altering CO2 Continuous flows and pressures through Nafions and run surveillance gas through entire system

Incomplete flushing of cell and dead volumes

Fast enough flow (100 sccm), alternate calibration sequence low‑to-high / high-to-low to look for effects

Leaks through fittings, solenoid valves, and pumps

Automated 8-hourly positive pressure leak-down and 4-hourly ambient pressure leak-up checks

Pressure broadening without Ar Use calibration gases made with real air

Fossil CO2 in calibration gases and different field and lab 13C sensitivities

Laboratory tests limit current effect to 0.05 ppm, long-term plans to use cylinders with natural CO2

Regulator temperature effects Tests suggest effect is negligible, but could be regulator dependent

Regulator flushing effects Repeat calibration tests suggest the effect is negligible

Whole-system diagnostics and comparability verification

Long-term surveillance tank analyzed every 8 hours, co-location with other programs, rotating cylinders, and laboratory comparisons

Delay in diagnosis of errors Near real-time data acquisition, processing, and dissemination

Potential source of bias AIRCOA solutionRelating to WMO CO2 Scale Dedicated CO2 and O2 calibration transfer facility

Short-term IRGA noise Average for 2 minutes to get better than 0.1 ppm precision

Drift in IRGA sensitivity 4-hourly 4-point calibrations and 30-minute 1-point calibrations

IRGA pressure sensitivity Automated 4-hourly pressure sensitivity measurements

IRGA temperature sensitivity 30-minute 1-point calibrations, temperature control at some sites

Incomplete drying of air Slow enough flow (100 sccm), two 96” Nafion driers, downstream humidity sensor to verify performance

Drying system altering CO2 Continuous flows and pressures through Nafions and run surveillance gas through entire system

Incomplete flushing of cell and dead volumes

Fast enough flow (100 sccm), alternate calibration sequence low‑to-high / high-to-low to look for effects

Leaks through fittings, solenoid valves, and pumps

Automated 8-hourly positive pressure leak-down and 4-hourly ambient pressure leak-up checks

Pressure broadening without Ar Use calibration gases made with real air

Fossil CO2 in calibration gases and different field and lab 13C sensitivities

Laboratory tests limit current effect to 0.05 ppm, long-term plans to use cylinders with natural CO2

Regulator temperature effects Tests suggest effect is negligible, but could be regulator dependent

Regulator flushing effects Repeat calibration tests suggest the effect is negligible

Whole-system diagnostics and comparability verification

Long-term surveillance tank analyzed every 8 hours, co-location with other programs, rotating cylinders, and laboratory comparisons

Delay in diagnosis of errors Near real-time data acquisition, processing, and dissemination

NCAR CO2 and O2/N2 Calibration Facility

Potential source of bias AIRCOA solutionRelating to WMO CO2 Scale Dedicated CO2 and O2 calibration transfer facility

Short-term IRGA noise Average for 2 minutes to get better than 0.1 ppm precision

Drift in IRGA sensitivity 4-hourly 4-point calibrations and 30-minute 1-point calibrations

IRGA pressure sensitivity Automated 4-hourly pressure sensitivity measurements

IRGA temperature sensitivity 30-minute 1-point calibrations, temperature control at some sites

Incomplete drying of air Slow enough flow (100 sccm), two 96” Nafion driers, downstream humidity sensor to verify performance

Drying system altering CO2 Continuous flows and pressures through Nafions and run surveillance gas through entire system

Incomplete flushing of cell and dead volumes

Fast enough flow (100 sccm), alternate calibration sequence low‑to-high / high-to-low to look for effects

Leaks through fittings, solenoid valves, and pumps

Automated 8-hourly positive pressure leak-down and 4-hourly ambient pressure leak-up checks

Pressure broadening without Ar Use calibration gases made with real air

Fossil CO2 in calibration gases and different field and lab 13C sensitivities

Laboratory tests limit current effect to 0.05 ppm, long-term plans to use cylinders with natural CO2

Regulator temperature effects Tests suggest effect is negligible, but could be regulator dependent

Regulator flushing effects Repeat calibration tests suggest the effect is negligible

Whole-system diagnostics and comparability verification

Long-term surveillance tank analyzed every 8 hours, co-location with other programs, rotating cylinders, and laboratory comparisons

Delay in diagnosis of errors Near real-time data acquisition, processing, and dissemination

CO2 signal averaged over 2.5 min. measurement cycle

Potential source of bias AIRCOA solutionRelating to WMO CO2 Scale Dedicated CO2 and O2 calibration transfer facility

Short-term IRGA noise Average for 2 minutes to get better than 0.1 ppm precision

Drift in IRGA sensitivity 4-hourly 4-point calibrations and 30-minute 1-point calibrations

IRGA pressure sensitivity Automated 4-hourly pressure sensitivity measurements

IRGA temperature sensitivity 30-minute 1-point calibrations, temperature control at some sites

Incomplete drying of air Slow enough flow (100 sccm), two 96” Nafion driers, downstream humidity sensor to verify performance

Drying system altering CO2 Continuous flows and pressures through Nafions and run surveillance gas through entire system

Incomplete flushing of cell and dead volumes

Fast enough flow (100 sccm), alternate calibration sequence low‑to-high / high-to-low to look for effects

Leaks through fittings, solenoid valves, and pumps

Automated 8-hourly positive pressure leak-down and 4-hourly ambient pressure leak-up checks

Pressure broadening without Ar Use calibration gases made with real air

Fossil CO2 in calibration gases and different field and lab 13C sensitivities

Laboratory tests limit current effect to 0.05 ppm, long-term plans to use cylinders with natural CO2

Regulator temperature effects Tests suggest effect is negligible, but could be regulator dependent

Regulator flushing effects Repeat calibration tests suggest the effect is negligible

Whole-system diagnostics and comparability verification

Long-term surveillance tank analyzed every 8 hours, co-location with other programs, rotating cylinders, and laboratory comparisons

Delay in diagnosis of errors Near real-time data acquisition, processing, and dissemination

Calibration sequence

Potential source of bias AIRCOA solutionRelating to WMO CO2 Scale Dedicated CO2 and O2 calibration transfer facility

Short-term IRGA noise Average for 2 minutes to get better than 0.1 ppm precision

Drift in IRGA sensitivity 4-hourly 4-point calibrations and 30-minute 1-point calibrations

IRGA pressure sensitivity Automated 4-hourly pressure sensitivity measurements

IRGA temperature sensitivity 30-minute 1-point calibrations, temperature control at some sites

Incomplete drying of air Slow enough flow (100 sccm), two 96” Nafion driers, downstream humidity sensor to verify performance

Drying system altering CO2 Continuous flows and pressures through Nafions and run surveillance gas through entire system

Incomplete flushing of cell and dead volumes

Fast enough flow (100 sccm), alternate calibration sequence low‑to-high / high-to-low to look for effects

Leaks through fittings, solenoid valves, and pumps

Automated 8-hourly positive pressure leak-down and 4-hourly ambient pressure leak-up checks

Pressure broadening without Ar Use calibration gases made with real air

Fossil CO2 in calibration gases and different field and lab 13C sensitivities

Laboratory tests limit current effect to 0.05 ppm, long-term plans to use cylinders with natural CO2

Regulator temperature effects Tests suggest effect is negligible, but could be regulator dependent

Regulator flushing effects Repeat calibration tests suggest the effect is negligible

Whole-system diagnostics and comparability verification

Long-term surveillance tank analyzed every 8 hours, co-location with other programs, rotating cylinders, and laboratory comparisons

Delay in diagnosis of errors Near real-time data acquisition, processing, and dissemination

Empirical pressure correction

Potential source of bias AIRCOA solutionRelating to WMO CO2 Scale Dedicated CO2 and O2 calibration transfer facility

Short-term IRGA noise Average for 2 minutes to get better than 0.1 ppm precision

Drift in IRGA sensitivity 4-hourly 4-point calibrations and 30-minute 1-point calibrations

IRGA pressure sensitivity Automated 4-hourly pressure sensitivity measurements

IRGA temperature sensitivity 30-minute 1-point calibrations, temperature control at some sites

Incomplete drying of air Slow enough flow (100 sccm), two 96” Nafion driers, downstream humidity sensor to verify performance

Drying system altering CO2 Continuous flows and pressures through Nafions and run surveillance gas through entire system

Incomplete flushing of cell and dead volumes

Fast enough flow (100 sccm), alternate calibration sequence low‑to-high / high-to-low to look for effects

Leaks through fittings, solenoid valves, and pumps

Automated 8-hourly positive pressure leak-down and 4-hourly ambient pressure leak-up checks

Pressure broadening without Ar Use calibration gases made with real air

Fossil CO2 in calibration gases and different field and lab 13C sensitivities

Laboratory tests limit current effect to 0.05 ppm, long-term plans to use cylinders with natural CO2

Regulator temperature effects Tests suggest effect is negligible, but could be regulator dependent

Regulator flushing effects Repeat calibration tests suggest the effect is negligible

Whole-system diagnostics and comparability verification

Long-term surveillance tank analyzed every 8 hours, co-location with other programs, rotating cylinders, and laboratory comparisons

Delay in diagnosis of errors Near real-time data acquisition, processing, and dissemination

SPL 9/4-9/18

NWR 9/18

Empirical temperature correction

Potential source of bias AIRCOA solutionRelating to WMO CO2 Scale Dedicated CO2 and O2 calibration transfer facility

Short-term IRGA noise Average for 2 minutes to get better than 0.1 ppm precision

Drift in IRGA sensitivity 4-hourly 4-point calibrations and 30-minute 1-point calibrations

IRGA pressure sensitivity Automated 4-hourly pressure sensitivity measurements

IRGA temperature sensitivity 30-minute 1-point calibrations, temperature control at some sites

Incomplete drying of air Slow enough flow (100 sccm), two 96” Nafion driers, downstream humidity sensor to verify performance

Drying system altering CO2 Continuous flows and pressures through Nafions and run surveillance gas through entire system

Incomplete flushing of cell and dead volumes

Fast enough flow (100 sccm), alternate calibration sequence low‑to-high / high-to-low to look for effects

Leaks through fittings, solenoid valves, and pumps

Automated 8-hourly positive pressure leak-down and 4-hourly ambient pressure leak-up checks

Pressure broadening without Ar Use calibration gases made with real air

Fossil CO2 in calibration gases and different field and lab 13C sensitivities

Laboratory tests limit current effect to 0.05 ppm, long-term plans to use cylinders with natural CO2

Regulator temperature effects Tests suggest effect is negligible, but could be regulator dependent

Regulator flushing effects Repeat calibration tests suggest the effect is negligible

Whole-system diagnostics and comparability verification

Long-term surveillance tank analyzed every 8 hours, co-location with other programs, rotating cylinders, and laboratory comparisons

Delay in diagnosis of errors Near real-time data acquisition, processing, and dissemination

Drying system monitoring

A change of 0.5% RH is approximately 300 ppm H2O, which would cause a dilution error of 0.1 ppm in CO2

Potential source of bias AIRCOA solutionRelating to WMO CO2 Scale Dedicated CO2 and O2 calibration transfer facility

Short-term IRGA noise Average for 2 minutes to get better than 0.1 ppm precision

Drift in IRGA sensitivity 4-hourly 4-point calibrations and 30-minute 1-point calibrations

IRGA pressure sensitivity Automated 4-hourly pressure sensitivity measurements

IRGA temperature sensitivity 30-minute 1-point calibrations, temperature control at some sites

Incomplete drying of air Slow enough flow (100 sccm), two 96” Nafion driers, downstream humidity sensor to verify performance

Drying system altering CO2 Continuous flows and pressures through Nafions and run surveillance gas through entire system

Incomplete flushing of cell and dead volumes

Fast enough flow (100 sccm), alternate calibration sequence low‑to-high / high-to-low to look for effects

Leaks through fittings, solenoid valves, and pumps

Automated 8-hourly positive pressure leak-down and 4-hourly ambient pressure leak-up checks

Pressure broadening without Ar Use calibration gases made with real air

Fossil CO2 in calibration gases and different field and lab 13C sensitivities

Laboratory tests limit current effect to 0.05 ppm, long-term plans to use cylinders with natural CO2

Regulator temperature effects Tests suggest effect is negligible, but could be regulator dependent

Regulator flushing effects Repeat calibration tests suggest the effect is negligible

Whole-system diagnostics and comparability verification

Long-term surveillance tank analyzed every 8 hours, co-location with other programs, rotating cylinders, and laboratory comparisons

Delay in diagnosis of errors Near real-time data acquisition, processing, and dissemination

Nafion absorption effect

Flow pulled through Nafion went from 300 to 50 sccm at t = 30 sec

Potential source of bias AIRCOA solutionRelating to WMO CO2 Scale Dedicated CO2 and O2 calibration transfer facility

Short-term IRGA noise Average for 2 minutes to get better than 0.1 ppm precision

Drift in IRGA sensitivity 4-hourly 4-point calibrations and 30-minute 1-point calibrations

IRGA pressure sensitivity Automated 4-hourly pressure sensitivity measurements

IRGA temperature sensitivity 30-minute 1-point calibrations, temperature control at some sites

Incomplete drying of air Slow enough flow (100 sccm), two 96” Nafion driers, downstream humidity sensor to verify performance

Drying system altering CO2 Continuous flows and pressures through Nafions and run surveillance gas through entire system

Incomplete flushing of cell and dead volumes

Fast enough flow (100 sccm), alternate calibration sequence low‑to-high / high-to-low to look for effects

Leaks through fittings, solenoid valves, and pumps

Automated 8-hourly positive pressure leak-down and 4-hourly ambient pressure leak-up checks

Pressure broadening without Ar Use calibration gases made with real air

Fossil CO2 in calibration gases and different field and lab 13C sensitivities

Laboratory tests limit current effect to 0.05 ppm, long-term plans to use cylinders with natural CO2

Regulator temperature effects Tests suggest effect is negligible, but could be regulator dependent

Regulator flushing effects Repeat calibration tests suggest the effect is negligible

Whole-system diagnostics and comparability verification

Long-term surveillance tank analyzed every 8 hours, co-location with other programs, rotating cylinders, and laboratory comparisons

Delay in diagnosis of errors Near real-time data acquisition, processing, and dissemination

Empirical flushing correction

Potential source of bias AIRCOA solutionRelating to WMO CO2 Scale Dedicated CO2 and O2 calibration transfer facility

Short-term IRGA noise Average for 2 minutes to get better than 0.1 ppm precision

Drift in IRGA sensitivity 4-hourly 4-point calibrations and 30-minute 1-point calibrations

IRGA pressure sensitivity Automated 4-hourly pressure sensitivity measurements

IRGA temperature sensitivity 30-minute 1-point calibrations, temperature control at some sites

Incomplete drying of air Slow enough flow (100 sccm), two 96” Nafion driers, downstream humidity sensor to verify performance

Drying system altering CO2 Continuous flows and pressures through Nafions and run surveillance gas through entire system

Incomplete flushing of cell and dead volumes

Fast enough flow (100 sccm), alternate calibration sequence low‑to-high / high-to-low to look for effects

Leaks through fittings, solenoid valves, and pumps

Automated 8-hourly positive pressure leak-down and 4-hourly ambient pressure leak-up checks

Pressure broadening without Ar Use calibration gases made with real air

Fossil CO2 in calibration gases and different field and lab 13C sensitivities

Laboratory tests limit current effect to 0.05 ppm, long-term plans to use cylinders with natural CO2

Regulator temperature effects Tests suggest effect is negligible, but could be regulator dependent

Regulator flushing effects Repeat calibration tests suggest the effect is negligible

Whole-system diagnostics and comparability verification

Long-term surveillance tank analyzed every 8 hours, co-location with other programs, rotating cylinders, and laboratory comparisons

Delay in diagnosis of errors Near real-time data acquisition, processing, and dissemination

Automated (4- or 8-hourly) leak checks

A positive trend of 0.3 kPa/min would be a leak rate of 0.1 sccm which if 100 ppm different would cause a 0.1 ppm bias

Potential source of bias AIRCOA solutionRelating to WMO CO2 Scale Dedicated CO2 and O2 calibration transfer facility

Short-term IRGA noise Average for 2 minutes to get better than 0.1 ppm precision

Drift in IRGA sensitivity 4-hourly 4-point calibrations and 30-minute 1-point calibrations

IRGA pressure sensitivity Automated 4-hourly pressure sensitivity measurements

IRGA temperature sensitivity 30-minute 1-point calibrations, temperature control at some sites

Incomplete drying of air Slow enough flow (100 sccm), two 96” Nafion driers, downstream humidity sensor to verify performance

Drying system altering CO2 Continuous flows and pressures through Nafions and run surveillance gas through entire system

Incomplete flushing of cell and dead volumes

Fast enough flow (100 sccm), alternate calibration sequence low‑to-high / high-to-low to look for effects

Leaks through fittings, solenoid valves, and pumps

Automated 8-hourly positive pressure leak-down and 4-hourly ambient pressure leak-up checks

Pressure broadening without Ar Use calibration gases made with real air

Fossil CO2 in calibration gases and different field and lab 13C sensitivities

Laboratory tests limit current effect to 0.05 ppm, long-term plans to use cylinders with natural CO2

Regulator temperature effects Tests suggest effect is negligible, but could be regulator dependent

Regulator flushing effects Repeat calibration tests suggest the effect is negligible

Whole-system diagnostics and comparability verification

Long-term surveillance tank analyzed every 8 hours, co-location with other programs, rotating cylinders, and laboratory comparisons

Delay in diagnosis of errors Near real-time data acquisition, processing, and dissemination

Potential source of bias AIRCOA solutionRelating to WMO CO2 Scale Dedicated CO2 and O2 calibration transfer facility

Short-term IRGA noise Average for 2 minutes to get better than 0.1 ppm precision

Drift in IRGA sensitivity 4-hourly 4-point calibrations and 30-minute 1-point calibrations

IRGA pressure sensitivity Automated 4-hourly pressure sensitivity measurements

IRGA temperature sensitivity 30-minute 1-point calibrations, temperature control at some sites

Incomplete drying of air Slow enough flow (100 sccm), two 96” Nafion driers, downstream humidity sensor to verify performance

Drying system altering CO2 Continuous flows and pressures through Nafions and run surveillance gas through entire system

Incomplete flushing of cell and dead volumes

Fast enough flow (100 sccm), alternate calibration sequence low‑to-high / high-to-low to look for effects

Leaks through fittings, solenoid valves, and pumps

Automated 8-hourly positive pressure leak-down and 4-hourly ambient pressure leak-up checks

Pressure broadening without Ar Use calibration gases made with real air

Fossil CO2 in calibration gases and different field and lab 13C sensitivities

Laboratory tests limit current effect to 0.05 ppm, long-term plans to use cylinders with natural CO2

Regulator temperature effects Tests suggest effect is negligible, but could be regulator dependent

Regulator flushing effects Repeat calibration tests suggest the effect is negligible

Whole-system diagnostics and comparability verification

Long-term surveillance tank analyzed every 8 hours, co-location with other programs, rotating cylinders, and laboratory comparisons

Delay in diagnosis of errors Near real-time data acquisition, processing, and dissemination

= Scott Marrin = SIO

JJ8588 JJ13623 CC79567 CC79546 JJ8550 JJ21201 CC161533

Li820-Siemens CO2 0.000 -0.084 -0.028 -0.014 -0.035 -0.051 0.044

ave Scott Marrin -0.043 ave SIO 0.001stdev Scott Marrin 0.035 stdev SIO 0.038stderr Scott Marrin 0.017 stderr SIO 0.022

Scott Marrin cylinders have a small (~ -0.04 ppm) bias on Li820 relative to Siemens

O2 Variability 4.980 5.250 0.600 0.540 4.700 4.030 0.390

ave Scott Marrin 4.740 ave SIO 0.510stdev Scott Marrin 0.524 stdev SIO 0.108stderr Scott Marrin 0.262 stderr SIO 0.062

6/24-6/21 O2 24.060 26.950 1.350 0.850 29.030 25.470 0.480

ave Scott Marrin 26.378 ave SIO 0.893stdev Scott Marrin 2.126 stdev SIO 0.437stderr Scott Marrin 1.063 stderr SIO 0.252

Scott Marrin cylinders have much more short-term and long-term drift for O2

13C bias test

Potential source of bias AIRCOA solutionRelating to WMO CO2 Scale Dedicated CO2 and O2 calibration transfer facility

Short-term IRGA noise Average for 2 minutes to get better than 0.1 ppm precision

Drift in IRGA sensitivity 4-hourly 4-point calibrations and 30-minute 1-point calibrations

IRGA pressure sensitivity Automated 4-hourly pressure sensitivity measurements

IRGA temperature sensitivity 30-minute 1-point calibrations, temperature control at some sites

Incomplete drying of air Slow enough flow (100 sccm), two 96” Nafion driers, downstream humidity sensor to verify performance

Drying system altering CO2 Continuous flows and pressures through Nafions and run surveillance gas through entire system

Incomplete flushing of cell and dead volumes

Fast enough flow (100 sccm), alternate calibration sequence low‑to-high / high-to-low to look for effects

Leaks through fittings, solenoid valves, and pumps

Automated 8-hourly positive pressure leak-down and 4-hourly ambient pressure leak-up checks

Pressure broadening without Ar Use calibration gases made with real air

Fossil CO2 in calibration gases and different field and lab 13C sensitivities

Laboratory tests limit current effect to 0.05 ppm, long-term plans to use cylinders with natural CO2

Regulator temperature effects Tests suggest effect is negligible, but could be regulator dependent

Regulator flushing effects Repeat calibration tests suggest the effect is negligible

Whole-system diagnostics and comparability verification

Long-term surveillance tank analyzed every 8 hours, co-location with other programs, rotating cylinders, and laboratory comparisons

Delay in diagnosis of errors Near real-time data acquisition, processing, and dissemination

Three cylinders were in the oven and one (green dots) was not

Regulator oven tests

Potential source of bias AIRCOA solutionRelating to WMO CO2 Scale Dedicated CO2 and O2 calibration transfer facility

Short-term IRGA noise Average for 2 minutes to get better than 0.1 ppm precision

Drift in IRGA sensitivity 4-hourly 4-point calibrations and 30-minute 1-point calibrations

IRGA pressure sensitivity Automated 4-hourly pressure sensitivity measurements

IRGA temperature sensitivity 30-minute 1-point calibrations, temperature control at some sites

Incomplete drying of air Slow enough flow (100 sccm), two 96” Nafion driers, downstream humidity sensor to verify performance

Drying system altering CO2 Continuous flows and pressures through Nafions and run surveillance gas through entire system

Incomplete flushing of cell and dead volumes

Fast enough flow (100 sccm), alternate calibration sequence low‑to-high / high-to-low to look for effects

Leaks through fittings, solenoid valves, and pumps

Automated 8-hourly positive pressure leak-down and 4-hourly ambient pressure leak-up checks

Pressure broadening without Ar Use calibration gases made with real air

Fossil CO2 in calibration gases and different field and lab 13C sensitivities

Laboratory tests limit current effect to 0.05 ppm, long-term plans to use cylinders with natural CO2

Regulator temperature effects Tests suggest effect is negligible, but could be regulator dependent

Regulator flushing effects Repeat calibration tests suggest the effect is negligible

Whole-system diagnostics and comparability verification

Long-term surveillance tank analyzed every 8 hours, co-location with other programs, rotating cylinders, and laboratory comparisons

Delay in diagnosis of errors Near real-time data acquisition, processing, and dissemination

Regulator flushing tests

Potential source of bias AIRCOA solutionRelating to WMO CO2 Scale Dedicated CO2 and O2 calibration transfer facility

Short-term IRGA noise Average for 2 minutes to get better than 0.1 ppm precision

Drift in IRGA sensitivity 4-hourly 4-point calibrations and 30-minute 1-point calibrations

IRGA pressure sensitivity Automated 4-hourly pressure sensitivity measurements

IRGA temperature sensitivity 30-minute 1-point calibrations, temperature control at some sites

Incomplete drying of air Slow enough flow (100 sccm), two 96” Nafion driers, downstream humidity sensor to verify performance

Drying system altering CO2 Continuous flows and pressures through Nafions and run surveillance gas through entire system

Incomplete flushing of cell and dead volumes

Fast enough flow (100 sccm), alternate calibration sequence low‑to-high / high-to-low to look for effects

Leaks through fittings, solenoid valves, and pumps

Automated 8-hourly positive pressure leak-down and 4-hourly ambient pressure leak-up checks

Pressure broadening without Ar Use calibration gases made with real air

Fossil CO2 in calibration gases and different field and lab 13C sensitivities

Laboratory tests limit current effect to 0.05 ppm, long-term plans to use cylinders with natural CO2

Regulator temperature effects Tests suggest effect is negligible, but could be regulator dependent

Regulator flushing effects Repeat calibration tests suggest the effect is negligible

Whole-system diagnostics and comparability verification

Long-term surveillance tank analyzed every 8 hours, co-location with other programs, rotating cylinders, and laboratory comparisons

Delay in diagnosis of errors Near real-time data acquisition, processing, and dissemination

Laboratory intercomparisons

Field surveillance tanksLaboratory offsets less than 0.06 ppm (1-sigma = 0.12 ppm)

Field differences from assigned values less than 0.15 ppm (1-sigma = 0.15 ppm)

Potential source of bias AIRCOA solutionRelating to WMO CO2 Scale Dedicated CO2 and O2 calibration transfer facility

Short-term IRGA noise Average for 2 minutes to get better than 0.1 ppm precision

Drift in IRGA sensitivity 4-hourly 4-point calibrations and 30-minute 1-point calibrations

IRGA pressure sensitivity Automated 4-hourly pressure sensitivity measurements

IRGA temperature sensitivity 30-minute 1-point calibrations, temperature control at some sites

Incomplete drying of air Slow enough flow (100 sccm), two 96” Nafion driers, downstream humidity sensor to verify performance

Drying system altering CO2 Continuous flows and pressures through Nafions and run surveillance gas through entire system

Incomplete flushing of cell and dead volumes

Fast enough flow (100 sccm), alternate calibration sequence low‑to-high / high-to-low to look for effects

Leaks through fittings, solenoid valves, and pumps

Automated 8-hourly positive pressure leak-down and 4-hourly ambient pressure leak-up checks

Pressure broadening without Ar Use calibration gases made with real air

Fossil CO2 in calibration gases and different field and lab 13C sensitivities

Laboratory tests limit current effect to 0.05 ppm, long-term plans to use cylinders with natural CO2

Regulator temperature effects Tests suggest effect is negligible, but could be regulator dependent

Regulator flushing effects Repeat calibration tests suggest the effect is negligible

Whole-system diagnostics and comparability verification

Long-term surveillance tank analyzed every 8 hours, co-location with other programs, rotating cylinders, and laboratory comparisons

Delay in diagnosis of errors Near real-time data acquisition, processing, and dissemination

Automated web-based output

http://www.eol.ucar.edu/~stephens/RACCOON