121
Scholars' Mine Scholars' Mine Masters Theses Student Theses and Dissertations 1965 An apparatus to measure capillary jet thrusts and the evaluation An apparatus to measure capillary jet thrusts and the evaluation of normal stresses in fluids of normal stresses in fluids Charles Donald Green Follow this and additional works at: https://scholarsmine.mst.edu/masters_theses Part of the Mechanical Engineering Commons Department: Department: Recommended Citation Recommended Citation Green, Charles Donald, "An apparatus to measure capillary jet thrusts and the evaluation of normal stresses in fluids" (1965). Masters Theses. 6990. https://scholarsmine.mst.edu/masters_theses/6990 This thesis is brought to you by Scholars' Mine, a service of the Missouri S&T Library and Learning Resources. This work is protected by U. S. Copyright Law. Unauthorized use including reproduction for redistribution requires the permission of the copyright holder. For more information, please contact [email protected].

An apparatus to measure capillary jet thrusts and the

  • Upload
    others

  • View
    4

  • Download
    0

Embed Size (px)

Citation preview

Page 1: An apparatus to measure capillary jet thrusts and the

Scholars' Mine Scholars' Mine

Masters Theses Student Theses and Dissertations

1965

An apparatus to measure capillary jet thrusts and the evaluation An apparatus to measure capillary jet thrusts and the evaluation

of normal stresses in fluids of normal stresses in fluids

Charles Donald Green

Follow this and additional works at: https://scholarsmine.mst.edu/masters_theses

Part of the Mechanical Engineering Commons

Department: Department:

Recommended Citation Recommended Citation Green, Charles Donald, "An apparatus to measure capillary jet thrusts and the evaluation of normal stresses in fluids" (1965). Masters Theses. 6990. https://scholarsmine.mst.edu/masters_theses/6990

This thesis is brought to you by Scholars' Mine, a service of the Missouri S&T Library and Learning Resources. This work is protected by U. S. Copyright Law. Unauthorized use including reproduction for redistribution requires the permission of the copyright holder. For more information, please contact [email protected].

Page 2: An apparatus to measure capillary jet thrusts and the

·'"')·· \

AN. APPARATUS TO MEASURE CAPILLARY JET

THRUSTS AND THE EVALUATION OF NORMAL STRESSES

IN FLUIDS

BY

CHARLES DONALD GREEN; '11-2-

A

THESIS

submitted to the faculty ofthe

UNIVERSITY OF MISSOURI AT ROLlA

in partial fulfillment of the requirements for the

Degree of

MASTER OF SCIENCE IN MECHANICAL ENGINEERING

Rolla, Missouri

1965

11.5236

Approved by

Page 3: An apparatus to measure capillary jet thrusts and the

TABLE OF CONTENTS

LIST OF FIGURES .

LIST OF TABLES

. . . . . . . . . . . .

SUMMARY . . . . . . . . . . . . I. INTRODUCTION

II.

III.

IV.

v. VI.

LITERATURE REVIEW .

EXPERIMENTAL . . . . . . . . . . . . . Purpose of Investigation

Apparatus . . . • • • . . . . . . . . . A. Auxiliary Equipment

B.

c. Thrustometer •

Viscometer • .

Experimental Procedure

A. Capillary Thrust Measurements

B. Viscometer Measurements

RESULTS AND DISCUSSION . . . . . . . . Calibration of Tubes and Entrance Effects Correction . • • . • .

Viscometry Measurements for Toluene and 1.0% polyisobutylene in cyclohexane

. . . .

Thrust Measurement for 1.0% PIB in cyclohexane . . . . . . ... . CONCLUSIONS • • .

RECOMMENDATIONS • .

. . . . . . . . . . . . . . . . . . . . . . .

APPENDIX I . . . . . . . . . . . . . . . . . Sample Calculations . . . . . . .

Page

iv

vi

1

3

4

15

15

15

15

28

34

39

39

43

46

46

52

58

66

67

68

68

ii

Page 4: An apparatus to measure capillary jet thrusts and the

TABLE OF CONTENTS (continued)

APPENDIX II . . . . . . . -Tables of Data . . . . . . . . . . . . . . Materials used in experiment

NOMENCLATURE

BIBLIOGRAPHY

ACKNOWLEDGEMENTS . . . . . . . . . . . . . . . . . . VITA . . . . . . . . . . . . . . . . .

iii

Page

75

76

102

103

105

107

108

Page 5: An apparatus to measure capillary jet thrusts and the

Figures

1.

2.

3-

4.

5-

LIST OF FIGURES

Stresses on incremental volume of fluid when subjected to a continuous deformation

Schematic diagram of process fluid flow system and auxiliary equipment

Page

5

. 16

Auxiliary equipment . . . . . . . . . . . . . 17

Schematic diagram of process fluid reservoir . . . . . • • . . . . .

Close-up view of Zenith pump~ Graham variable speed drive, and Bourdon gauges .

19

20

6. Schematic diagram of constant temperature bath regulating system . . . • . . . . . . . . 22

7. Schematic diagram of viscometer and cooling water flow system . . . • . . . . . . . . . . 23

8. Viscometer jacket connected to constant tem-perature bath and water pump . . . . . 24

9.

10.

Manometer system and manifold board

Schematic drawing of modified analytical balance frame . • . . . • . . . . . . .

. 25

29

11. Side view of analytical balance apparatus with capillary tube in place . . . . . . . 30

12. Back view of analytical balance apparatus with tygon tube disconnected and capillary

13.

14.

15.

16.

17.

18.

tube in place . . . • . . . . . . . . . . . . 31

Capillary tubes with adapters for thrust measurements . . . . . . . . . . . . . . . .

Detailed diagram of viscometer jacket

Viscometer heat exchanger jackets

Capillary tubes with adapters, for viscdmetry measurements . . . . .

. 33

. 35

. 37

. 38

Wall shear stress vs 8V/D for toluene using viscometer tubes . . . . . . . . . . . . . . . 48

Wall shear stress vs 8V/D for toluene using thrust tubes . • . . . • . . . . . . . . . . • 49

iv

Page 6: An apparatus to measure capillary jet thrusts and the

LIST OF FIGURES (continued)

Figures Page

19. Wall shear stress vs 8V/D for 1.0% PIB using viscometer tubes . . . . . . . . . . 53

20. Wall shear stress vs 8V/D for 1.0% PIB

21.

22.

23.

using thrust tubes . . . • . . . . . . . . 54

Wall shear stress vs 8V/D for 1.0% PIB using viscometer and thrust tubes . . .

Thrust vs 8V/D for 1.0% PIB

. 55

. . 59

Normal stress differences vs 8V/D for 1. O% PIB . . . . . . . • • . . . . . . . . . . 62

v

Page 7: An apparatus to measure capillary jet thrusts and the

Tables

I.

II.

III.

IV.

v.

VI.

VII.

VIII.

IX.

LIST OF TABLES

Pressure Indicators . . . . . o

Thrustometer Capillary Tubes

Viscometer Capillary Tubes

Viscometer and Thrust Tube Curve Constants from log~w vs log 8V/D .

Constants for the Curves of the Log Thrust vs Log 8V/D for 1.0% PIB . o

P.Eo in (P11 - P22)w Due to Error in Thrust . . . . . . . • • • . .

Variation of Balance Sensitivity with Pressure Drop . . . . . • • • . . . •

vi

Page

. 26

. • 34

• 0 • • 39

60

. • 65

Comparison of Reynolds Numbers and Thrusts . . 65

Toluene Data for Viscometer and Thrust Capillary Tubes • . . . • • • . . . . . 76

X. Calibration of Tubes and Entrance Effects Coefficient Data . . . • • • • . • . • . • . . 81

XI. Viscometry Data for 1.0% PIB in cyclohexane .. 95

XII.

XIII.

XIV.

Thrust and Normal Stress Data for 1.0% PIB in cyclohexane . • . . • • • . . • . • . . 97

Dimensions and Tolerances of Capillary Tubes. .100

Calibration of Bourdon Tube Pressure Gauges using Dead Weight Tester . o • • • • • • • • .101

Page 8: An apparatus to measure capillary jet thrusts and the

SUMMARY

The purpose of this work was to design~ construct and

calibrate simple~ inexpensive apparatus which could be used

to measure the normal stresses and tangential shear stresses

of a fluid.

For normal stress measurements~ the capillary jet

1

thrust principle was selected. A standard analytical bal­

ance was modified so that capillary tubes could be fitted on

it horizontally. It was designed so that the thrust produced

by a jet of fluid from the tube caused a deviation from the

zero balance position. The weights needed to rezero the

balance gave a measure of the thrust.

Normal stresses were evaluated for a solution of 1.0%

polyisobutylene in cyclohexane and found to be near the

values obtained on a sim~lar solution by others. It was

found that the thrusts measured at high values of 8V/D~ above

4 x 104 sec- 1 ~ did not give reliable normal stress differ­

ences. Thrusts could not be measured accurately at high

values of 8V/D (where the pressure drop across the capillary

tube exceeded 15 psi) because the flexible tygon tubing con­

nector swelled~ stiffened and lost its flexibility~ seriously

reducing the balance sensitivity.

It was noted that in the Generalized Reynolds Number

range useful for these measurements (100-1000)~ the range of

thrust values obtained was the same regardless of tube

diameter. This observation is useful in selecting tubes for

covering desired 8V/D ranges.

Page 9: An apparatus to measure capillary jet thrusts and the

A capillary viscometer was also constructed and used

for obtaining the necessary flow data. It operated satis­

factorily in a continuous range of flow rates of 2 to 560

cc/min and capillary wall shear stresses of 1.0 to 660

dynes/cm2 • With seven capillary tubes, the complete 8V/D

range of 200 to 1.2 x 105 sec- 1 can be covered even for

fluids of low (0.5 cp) viscosity. For more viscous fluids,

fewer tubes are required.

The diameters of the capillary tubes used in the

experiment were obtained by applying an iterative procedure

2

to a standard experimental method. A Newtonian fluid entrance

effects correction coefficient was also checked experimentally

and found to agree with a value proposed in the literature.

Page 10: An apparatus to measure capillary jet thrusts and the

I. INTRODUCTION

It has been increasingly recognized that many fluids

such as slurries and suspensions~ polymer solutions~ poly­

mer melts~ emulsions and others exhibit non-Newtonian

properties. Therefore~ the characterization of their flow

properties has developed into an important field of study

in which stresses developed in these fluids when they are

deformed are evaluated. These stresses are tangential

shearing stresses and normal stresses.

The determination of these stresses~ particularly the

normal stresses at high shear rates~ has been limited by

the methods and instruments available. Therefore~ the

design of the necessary instruments has become an important

problem and will continue to be one until a suitable appa­

ratus is developed which can be used to obtain normal

stresses over a wide range of shear rates for any fluid.

The work described here was therefore concerned pri­

marily with the problem of designing~ building and testing

an instrument for evaluating normal stresses by using the

principle of measuring the thrust jetting from a capillary

tube. To fully describe the flow~ the tangential shearing

stresses must also be evaluated. As part of the problem~ a

capillary viscometer was also developed.

3

Page 11: An apparatus to measure capillary jet thrusts and the

II. LITERATURE REVIEW

The classical theory of elasticity deals with the

mechanical properties of perfectly elastic solids, in which

in accordance with Hooke's law, stress is always directly

proportional to strain3 but independent of the rate of

strain. The theory of hydrodynamics deals with properties

of perfectly viscous liquids for which, in accordance with

Newton's law, the stress is always directly proportional to

the rate of strain, but independent of the strain itself.

These categories are idealizations however. Any real solid

shows deviation from Hooke's law under suitably chosen con­

ditions, and it is probably safe to say that any real liquid

would show deviation from Newtonian flow if subject to

sufficiently precise measurement.

4

There are two important types of deviations. First, the

strain in a solid or the rate of strain in a liquid may not

be directly proportional to the stress, but may depend on

stress in a more complicated manner, such as when the elastic

limit is exceeded for a solid. Second, the st~ess may depend

on both the strain and the rate of. strain together. Mater­

ials demonstrating this type of behavior are partly fluid­

like and partly solid-like. These materials are referred to

as viscoelastic and usually demonstrate what are referred to

as normal stress effects.* *Fredrickson (6) has shown that a normal stress effect can be predicted from a model of a purely viscous fluid (Stokesian) under certain conditions, even though the fluid is nonelastic. However, there are no known examples of such fluids.

Page 12: An apparatus to measure capillary jet thrusts and the

The classical method followed in analyzing the stresses

developed within a fluid~ when it is deformed continuously 3

is to consider an incremental volume of fluid and the

stresses acting on it (Figure 1).

2 "'t-22 )-1

\'-12 ,...32

v J--1-31 111

1-'33

Figure 1. STRESSES ON INCREMENTAL VOLUME OF FLUID

WHEN SUBJECTED TO A CONTINUOUS DEFORMATION

In the stress convention used here~ the first subscript

refers to the direction perpendicular to the plane in which

the stress acts and the second subscript refers to the

direction.

5

The stresses 1 11, '\22 , 133 are called the dtnti:aee~

normal stresses and the terms 1 12, 1 13, '121 , 'T"'23 ~ 1 31, and

1r32 are called the shearing stresses. These stresses can

be represented by the stress tensor:

Page 13: An apparatus to measure capillary jet thrusts and the

6

T 11 \12 .,... 13

""-· . = \21 \22 rr-23 l.J (1)

T 31 '\ 32 .,.....33

In dealing with fluids, it is convenient to divide the

stress tensor into two parts .

..,...,_ . - - p 5 . . + p . . l.J l.J l.J

(2)

where the term p is called hydrostatic or isotropic pressure,

Pij the shear stress tensor, and 5ij the Kronecker delta:

5 .. -l.J

1

0

0

0

1

0

0

0

1

(3)

This stress tensor definition can be made because a fluid

has a fundamental property that, when it is in a state of

rest, the stresses within it become purely isotropic {the

same in all directions). That is:

rr .. = - P 5 .. l.J l.J

where

...... - 0 l.J

i -1 j ; "'f':. • I o i = j l.J

When the velocity field is assumed as:

(4)

Page 14: An apparatus to measure capillary jet thrusts and the

and where the subscripted x refers to the coordinate direc­

tion (see Figure 1), it has been described by Dahler and

Scriven (5) and by Aris (1) that the stress tensor may be

taken as symmetric. That is:

~-. = "-·. ~J J~

(5)

7

Weissenberg (23) noted that in a continuously deforming

fluid with elastic properties, finite strains would be crea­

ted. He reasoned that by analogy to large deformations of

solids these finite strains would cause normal stresses. On

the basis of this analogy he then proposed that for incom­

pressible viscoelastic fluids, the stress tensor could be

written as (24):

pll 1"'12 0

'\"" .. - - p 5 .. + "1""12 p22 0 ~J ~J

(6}

0 0 p33

where p, the hydrostatic pressure, was defined as

1 ("til +~22 +!33)

1tr~ p = - 3 = -3

(7)

and the velocity field was the same as that assumed previ­

ously in this paper. It should be noted that1r13,1r23,~31

andil32 are zero. The reason for this is that, in consider­

ing the rotational equilibrium (22) of an incremental

element of volume, it has been found that in order to have a

Page 15: An apparatus to measure capillary jet thrusts and the

state of rotational equilibrium these stresses must indeed

be zero. It can also be shown that only two of the normal

stress terms are independent. Thus only the shearing stress

1l12 and any two of the normal stresses need be evaluated

independently.

8

A great deal of work has been done in developing methods

with which to measure normal stresses in fluids subjected to

continuous deformation. Methods which have found major use

are: rotating parallel plates (13_,24)_, rotating cones of

small angle and flat plates (13_,24)_, and birefringence

measurements (11_,16_,4).

Measurement of the normal stresses in the cone-plate and

parallel plate rheogoniometer instruments are obtained by

determining the height of rise of fluid in capillary tubes

affixed to the plate_, as a function of radial position. The

force normal to the plate can thus be measured and the normal

stress calculated. These methods are reliable_, but they

have the major disadvantage of being unable to measure normal

stresses at shear rates above 103 sec- 1 . It is of great i~

portance_, in industry and research_, that methods of obtaining

these stresses at shear rates higher than 103 sec-l be found_,

since many practical flow situations fall in this region.

The birefringence method makes use of light ray re­

fraction. Anisotropic crystals split a ray of light into

two separate parts which travel at different speeds and fol­

low different paths. By measuring the responses of the

light rays through the fluid_, it is possible to evaluate the

normal stresses. This procedure has been used for determining

Page 16: An apparatus to measure capillary jet thrusts and the

normal stresses at shear rates slightly greater than 104

sec- 1 (4). \

It has the major disadvantage of being a tedious

procedure limited to only transparent fluids.

It was suggested by Gaskins and Philippoff (7) that

normal stress data at high shear rates might be obtainable

by studying the jet of a fluid issuing from a capillary

tube, as it had been observed (12) that the jet expanded

when emitted from the capillary tube. This expansion was

attributed to the relaxation of elastic stresses developed

in the shear field of the capillary. From an energy bal­

ance across the jet, they suggested that the normal stresses

could be determined by obtaining the ratio of the diameter

of the relaxed jet to the diameter of the capillary tube.

They were able to obtain normal stress data at shear rates

9

up to 105 sec- 1 . However, a mistake was made in their energy

balance. They erroneously omitted, the difficult to evaluate,

dissipation of energy accompanying the change of flow veloc­

ity profile from that within the tube to the flat profile

downstream from the tube exit.

Houghton (10) and Metzner, et. al. (14), using this basic

theory, but with a more refined analysis, were able to obtain

normal stresses at high shear rates. They determined the

momentum balance on the jet between the tube exit and a point

downstream where the velocity profile had completely relaxed.

From this, normal stresses were calculated as a function of

shear rate from data on the ratio of the jet diameter to the

tube diameter, tube dimensions, and n 1 , where n' is the slope

Page 17: An apparatus to measure capillary jet thrusts and the

10

'""'.tk--(;.xJI of the log shear stress}.versus log (8V/D) curve (V =average

velocity in the tube~ D =tube diameter).

The major disadvantage of this procedure was that photo­

graphs of the jet were used in measuring the jet diameter.

Difficulty arose from the inherent parallax in measuring

the photographs accurately. It was also difficult to obtain

photographs of the jet in a completely relaxed state. Thus,

this method had serious deficiencies although high shear

rate data could be obtained.

Harris (8) suggested that the normal stresses might be

evaluated by measuring the reaction, or thrust, of a jet of

viscoelastic fluid. Shertzer and Metzner (19,20) presented

a detailed mathematical analysis of this theory which was

outlined by White and Metzner (25) and a summary of which is

presented here. For a thorough analysis~ the reader is

referred to the original references.

By considering the momentum balance on an element of

fluid at the exit of a capillary tube, the thrust may be

related to the normal stress in the direction of flow by

the equation:

(8)

where T is the thrust exerted by the jet, ~ is the fluid

velocity at radius r in the tube~ p is the fluid density and

(~1) is the total normal stress in the axial direction. By

analyzing this equation they (19~20) obtained the equation:

Page 18: An apparatus to measure capillary jet thrusts and the

/

_ pV2 [3n'~l (n'+l)J _ T [ 1 n* n' rrR2

1 d(log T) J .. + 2n' BV

d(log D) · (9)

where (1r11)w denotes the stress 'r11 evaluated at the wall

shear rate.

11

Using the flow equations~ invoking the Weissenberg

hypothesis that P22 = P33 ~ and assuming that p0 ~L = 0 (where . .

p0~L is the pressure at the tube exit and centerline), they

were able to show that (~11)w - (P11 - P22)w. Tne final

equation thus obtained was: '{ . II

( ) pQ~ [~+l (n0'+.l)J. , Pll - p22 w - rr2R.4 2n *+1

T - rrR2

1 d(tbs T) J + 2n 1 BV

d(log. _D) . . ' Equation (10) can be used to evaluate (P11 - P22)w from

experimental data for n 1 , T and d(log T)/d(log 8V/D).

(10)

The assumptions made in the above derivation were: (1)

the flow was steady~ laminar and isothermal and the tube

from which the fluid was issuing has a sufficiently great

L/D ratio to insure a well-developed velocity profile, (2)

the fluid was !~compressible~ (3) effects of7 interfacial and

surfa~e-terision at the end <:lf the jet were negligible, and

(4) gravitational effects and drag of the air on the jet

could be neglected.

Shertzer (19) constructed a capillary rheometer which

measured the thrust by measuring the force of the jet

Page 19: An apparatus to measure capillary jet thrusts and the

12

transmitted to a series of transducers. He was able to

measure thrusts in the range of 0.223 to 22.3 grams on

dilute polymer solutions up to shear rates of slightly higher

than 105 sec- 1 • The normal stresses were then compared to

those obtained from previous work at lower shear rates and

were found to be in good agreement, thus demonstrating the

validity of the method.

Uebler (21), using the same theory, constructed an

apparatus different in design from Shertzer in that he meas­

ured the thrust by having the jet produce a twisting moment

in a circular bar. Strain gauges were used to determine the

strain developed in the bar, and by using basic mechanics of

materials relationships he was able to evaluate the thrust.

He reported measurements of thrusts in the range of 0.06 to

17.9 grams and he evaluated normal stresses at shear rates

up to 1.3 x 106 sec- 1 • He also compared his data to that

obtained by Shertzer in order to demonstrate the validity of

his design and found good agreement. Uebler also investigated

the assumptions made by Shertzer and Metzner in deriving

equation (10) and found them to be justifiable.

The technique of capillary viscometry is one of the

most common methods for characterizing the flow characteris­

tics of fluids. In this method, a fluid is forced through a

fine-bore tube and the viscosity of the fluid is determined

from the measured volumetric flow rate, applied pressure,

and tube dimensions.

For Newtonian fluids in which the shear stress is

related to the shear rate by the proportionality constant

Page 20: An apparatus to measure capillary jet thrusts and the

~ (the viscosity), the Hagen-Poiseuille law holds for

laminar flow in tubes.

Q = TrR4 6P 8~L

where Q = flow rate, R = internal radius of the tube,

(11)

6P = pressure drop across the tube, and L = length of tube.

The shear rate at the wall is

.6.PR 2T)L

8V D (12)

For non-Newtonian fluids, it can be shown (3,22) that

(-dv) = [3n 1 +1] §:1. Or w 4n 1 D (13)

13

An important consideration in capillary viscometry work

is the determination of the pressure-drop correction factor

C for the capillary end effects correction. This correction

is important, because in all capillary viscometers, a liquid

in a wide reservoir enters the capillary tube in a converging

stream and comes out either into open air or into another

wide reservoir in a divergent stream. This process produces

what is referred to as the pressure drop due to entrance

effects, and is made up of the excess pressure loss downstream

from the contraction plus the increase in the kinetic energy

of the full-developed velocity profile within the tube. This

entrance effect can be expressed in velocity heads as

Page 21: An apparatus to measure capillary jet thrusts and the

14

!::,.PENTRANCE pV.2 - c 2g

c (14)

where V is the average velocity of the fluid within the tube.

Therefore, the actual pressure drop across the tube is the

observed pressure drop minus ~p entrance.

6PACTUAL = ~POBSERVED - ~PENTRANCE (15)

For Newtonian fluids a value of (C) equal to 2.16 was sug­

gested by Bogue (3) and it will be shown later that this

value gave consistent results with the data obtained in this

work. It should be noted here that the end correction

factor should be checked for each tube and type of fluid

in order to obtain a value directly applicable to the

particular experiment being run.

Page 22: An apparatus to measure capillary jet thrusts and the

15

III. EXPERIMENTAL

Purpose of Investigation

The primary object of this investigation was to design

and construct equipment to measure the thrust produced by a

fluid jetting from a capillary tube. The thrust would then

be used to calculate the normal stresses of the fluid. The

secondary object was to construct a capillary viscometer to

obtain viscosity data for the fluids being studied.

Apparatus

The apparatus consisted of two main pieces of equipment:

1. An analytical balance to measure the thrust produced

by a fluid jetting from a capillary tube attached to

it

2. A capillary viscometer.

These two pieces of equipment were used with the follow-

ing auxiliary equipment:

1. Fluid reservoir

2. Pump and variable speed drive

3. Constant temperature bath

4. Manometer and pressure indicator system

A. Auxiliary Equipment

The auxiliary equipment will be described first. A

schematic diagram of the process fluid system and auxiliary

equipment is shown in Figure 2. Figure 3 is a photograph of

the actual equipment. Process fluid flows from the reservoir

to the process fluid pump~ then through coils in the constant

Page 23: An apparatus to measure capillary jet thrusts and the

Figure 2

SCHEMATIC DIAGRAM OF PROCESS FLUID

FLOW SYSTEM AND AUXILIARY EQUIPMENT

1. Mercury manometer

2. Process fluid manometer

3 . Mi.cromanometer

4. Air discharge tube

5. Volume capacity line

6. Bourdon pressure gauge (0-300 psi)

7. Bourdon pressure gauge (0-100 psi)

8. Compensating air pressure line

9. Zenith metering gear pump

10. Graham variable speed transmission Model No. 45R2.8

(0-200 rpm)

11. Fluid reservoir

12. Constant temperature bath

13. Process fluid outlet

14. Mercury manometer well

16

Page 24: An apparatus to measure capillary jet thrusts and the
Page 25: An apparatus to measure capillary jet thrusts and the

17

Page 26: An apparatus to measure capillary jet thrusts and the

18

temperature bath to the thrustometer or capillary viscometer.

A pressure tap which could be connected to any of the five

pressure indicators was teed into the process fluid line

just before it exits from the constant temperature bath.

About 450 cc of process fluid were required to fill the

system.

The fluid reservoir is shown in Figure 4. It consists

of a 2-inch O.D. galvanized pipe coupling with a galvanized

reducer on the bottom to form the outlet port for the fluid.

The top was covered with a sheet of aluminum foil. The

reservoir had a capacity of 15 in3. This small capacity was

used so that the evaporation of the solvent~ which was a

problem~ could be minimized. When flow rate data was not

being taken> the fluid from the thrustometer or viscometer

was directed into the reservoir by way of a curved glass

tube~ producing a recycling system.

Two Zenith metering gear pumps~ with volumetric ranges

of 0-50 and 0-550 cc/min~ were used to deliver the desired

flow rates. A graham variable speed drive (0-190 rpm) pow­

ered by a 1/4 HP explosion-proof motor was used to drive the

pumps. A photograph of the drive and one pump is shown in

Figure 5. The low capacity pump provided consistent and

reproduceable flow rates in the low flow range (as low as

2 cc/min) where the larger pump was unable to reproduce the

desired flow rates. The wide range of flow rates permitted

viscometry data to be obtained in the transition and low

turbulent range as well as the laminar range. The rpm

Page 27: An apparatus to measure capillary jet thrusts and the

Figure 4

SCHEMATIC DIAGRAM OF PROCESS FLUID RESERVOIR

1. Glass tube to catch fluid jet

2. Glass funne 1

3. Aluminum foil cover

4. 2-inch galvanized nipple

5. 2-inch galvanized pipe coupling

6. 2 x 1/4-inch galvanized bushing

7· Process fluid line to gear pump: 1/4-inch O.D. nylon

tubing

8. Aluminum tee: Swage lok A- 400- 3- 4TMT

9. Fluid level indicator tube: 1/4-inch O.D. nylon tubing

19

Page 28: An apparatus to measure capillary jet thrusts and the
Page 29: An apparatus to measure capillary jet thrusts and the

Figure 5 CLOSE-UP VIEW OF ZENITH PUMP, GRAHAM

VARIABLE SPEED DRIVE, AND BOURDON GAUGES 1\) 0

Page 30: An apparatus to measure capillary jet thrusts and the

21

delivered by the variable speed drive could be set to repro­

duce the desired flow rates to at least +1.0 cc/min.

A schematic diagram of the constant temperature water

bath regulating system is shown in Figure 6. The container

was a 5 gallon can. The process fluid passed through an 18-

foot coil of 3/8-inch O.D. stainless steel tubing immersed in

the bath. Heat was supplied by a 1000 watt electric heater.

The bath was cooled by passing cool tap water through a

copper cooling coil. It could be controlled to ±.05°C.

When the viscometer was in use~ mixing was obtained by

recycling water through the bath with a 1/6 HP centrifugal

pump that delivered 1.0 gallon of water per minute. Figure

7 is a schematic diagram of this recirculation system and

Figure 8 is a photograph of the system in operation. When

data were being taken with the thrustometer~ vibration pro­

duced by the water pump was transmitted to the thrust appa­

ratus and affected the thrust measurements. Therefore~ an

air driven stirrer was used for the mixing.

Valves were placed at many points in the fluid and

manometer lines so that the system could be drained and

flushed easily. A valve was also placed at the high point

of the process fluid coil to allow for air to be discharged

from the coil.

The pressure tap in the process fluid coil near the out­

let from the bath was connected to a system of manometers

and pressure gauges. Figure 9 shows the three manometers

and.the valves needed to permit the proper pressure indicator

Page 31: An apparatus to measure capillary jet thrusts and the

Figure 6

SCHKMATIC DIAGRAM OF CONSTANT TEMPERATURE

BATH REGULATING SYSTEM

1. Cooling coil

2. Thermometer: (-5.0 - 50.0°C) precision +0.05°C

3. Thermoregulator: Jumo Model No. 16-50A

0-50°C~ 150 mm long precision +0.01°C

4. Explosion proof box

5. 110-volt electrical recepticle

6. Relay: Fisher~ Model No. 30

7. Circulating-air line

8. Electrical coil (1000 watts)

9. Constant temperature bath

22

Page 32: An apparatus to measure capillary jet thrusts and the
Page 33: An apparatus to measure capillary jet thrusts and the

Figure 7

SCHEMATIC DIAGRAM OF VISCOMETER AND

COOLING WATER FLOW SYSTEM

1. Constant temperature bath

2. Cooling water return line: 1/4-inch galvanized pipe

3. Viscometer heat exchanger jacket: 3/8-inch galvanized

pipe

4. Centrifugal pump and 1/6 HP explosion proof motor

23

Page 34: An apparatus to measure capillary jet thrusts and the
Page 35: An apparatus to measure capillary jet thrusts and the

Figure 8 VISCOMETER JACKET CONNECTED TO CONSTANT

TEMPERATURE BATH AND WATER PUMP 1\) ~

Page 36: An apparatus to measure capillary jet thrusts and the

25

Figure 9

~~NOMETER SYSTEM AND MANIFOLD BOARD

Page 37: An apparatus to measure capillary jet thrusts and the

to be connected to the process fluid pressure tap. The

manometers consisted of a process fluid micromanometer,

a process fluid manometer, and a single-leg, well-type

mercury manometer. All three manometers were constructed

specifically for this equipment. The pressure gauges were

two bourdon gauges which can be seen in Figure 5.

26

The pressure indicators were assigned numbers to identify

them in computer calculations. These numbers and the char-

acteristics of each pressure indicator based on toluene

(density = 0.8564 gm/cm3 ) as the process fluid, are listed

in Table I.

Table I

PRESSURE INDICATORS

Maximum Range Smallest No. Type Reading psi Division

1 Well-Hg 180 em 0-34 1.0 nun Manometer

2 Process fluid 180 em 0-2.25 1.0 rrnn Manometer

3 Process fluid 1.25 inches 0-0.038 0.001 inch Micromanometer fluid

4 Bourdon- tube 100 psi 0-100 1 psi

5 Bourdon- tube 300 psi 0-300 5 psi

For all but No. 3, pressure readings could be estimated to

half the smallest division indicated. No. 3 could be read

to 0.0001 inch.

All manometers were used up to their maximum reading

before the next highest manometer was valved into the system.

Page 38: An apparatus to measure capillary jet thrusts and the

27

By following this procedure~ the initial pressure reading for

each manometer was high enough so that the error in reading

them was minimized. In transferring from manometer No. 3 to

manometer No. 2~ the initial pressure reading was only 3.2 em

of fluid which was actually too small to be reliable. It was

decided that pressure readings lower than 5.0 em on manometer

No. 2 would be taken. However, if that data point deviated

from the other points in the series, that point could justifi­

ably be considered inaccurate due to the error in the

manometer reading.

In order to reduce the amount of process fluid needed

to fill the volume capacity lines to the pressure gauges, a

regulated compensating air pressure of 0-80 psi was fed into

the lines as required. By regulating the pressure in the

lines to approximately the expected reading, the pressure

gauge could attain a constant reading faster. The regulated

air counter-balanced the pressure from the system quickly and

little movement of the process fluid in the lines was neces-

sary. The volume capacity line also served to maintain a

volume of air ahead of the process fluid to insure tMt the

process fluid would not contact the internal parts of the

gauges~ which could have damaged them.

The process fluid and manometer lines were 1/4-inch O.D.

nylon tubing. The volume capacity line of the pressure indi­

cators was a 3/8-inch O.D. nylon tube. The tubing was rated

at 500 psi at room temperature with a safety factor of 4:1.

Page 39: An apparatus to measure capillary jet thrusts and the

B. Thrustometer

While the apparatus designed by Shertzer and Uebler

appear to be useful, it was desired to design a less costly

sensing device to measure capillary jet thrusts. An

analytical balance principle was chosen to measure the

thrust, because accurate and reproduceable weight measure­

ments as small as 0.0002 grams are easily obtained with a

standard analytical balance when it is undamped. It was

believed that if an analytical balance could be modified to

measure thrusts without a substantial loss of sensitivity,

accurate and reproduceable thrusts could be obtained.

Figure 10 shows a schematic diagram of the modified

analytical balance frame used to measure the capillary jet

thrusts. Photographs of a side view and a back view of the

apparatus are shown in Figures 11 and 12.

The balance was fitted with a 4.25 inch length of 3/8-

inch O.D. high pressure sight glass tubing, hung directly

from above the fulcrum of the balance.* A 90° aluminum

28

Swagelok tee was affixed to the top end to produce a length

of 5.50 inches from the fulcrum to the mid-point of the

outlet opening on the elbow.** The 180° section was fitted *A piece of stainless steel tubing was originally used, but it was found that at the time fluid was being pumped into the sys­tem, air would be trapped within the tube. The experimenter was unable to observe when all the air was discharged from the tube. Thus a glass tube was chosen. This also reduced the weight of the modified frame. **Steel ferrules were used as seals in the Swagelok fittings on all fluid lines, manometer lines and pressure indicator lines. Nylon ferrules were used for all glass tubing connections and for seals for the capillary tubes in the thrustometer and vis­cometer. The nylon ferrules were used on the capillary tubes because it was found that this type, while providing a tight seal, could be removed easily and reused when tubes were being changed.

Page 40: An apparatus to measure capillary jet thrusts and the

Figure 10

SCHEMATIC DRAWING OF MODIFIED

ANALYTICAL BALANCE FRAME

1. Capillary tube with adapter

2. Pointer

3. Adjustable lead counter-weights

4. Balance head frame

5. Aluminum tee: Swagelok A-600-3TFT

6. Insert: Swagelok 605-4

7· Front nylon ferrule: Swagelok ZY-603-1

8. Back nylon ferrule: Swagelok ZY-604-1

9. 3/8-inch O.D. tygon tubing

10. 3/8-inch O.D. glass tube

11. Aluminum elbow: Swagelok A-600-9

12. Front nylon ferrule: Swagelok ZY-603-1

13. Back nylon ferrule: Swagelok ZY-604-1

29

Page 41: An apparatus to measure capillary jet thrusts and the
Page 42: An apparatus to measure capillary jet thrusts and the

Figure 11 SIDE VIEW OF ANALYTICAL BALANCE APPARATUS

WITH CAPILLARY TUBE IN PLACE (,.) 0

Page 43: An apparatus to measure capillary jet thrusts and the

Figure 12

BACK VIEW OF ANALYTICAL BALANCE APPARATUS WITH TYGON

TUBE DISCONNECTED AND CAPILLARY TUBE IN PLACE w ......

Page 44: An apparatus to measure capillary jet thrusts and the

with a plug which was used to hang the tee to the balance.

A capillary tube with a 3/8-inch O.D. aluminum adapter

(described below) was fitted to the horizontal section of

32

the 90° elbow and properly aligned. Adjustable counter­

weights were then used to locate the center of gravity of the

balance just below the fulcrum. This produced the greatest

balance sensitivity and was checked whenever the balance was

reassembled.

A 6-inch piece of 3/8-inch O.D. by 5/16-inch I.D. tygon

tubing connected the outlet of the constant temperature bath

to the tee of the balance. Tygon tubing was chosen because of

its resistance to organic solvents~ apparent flexibility and

strength.

The center line of the tygon tubing was located so as

to pass directly through the bottom of the fulcrum of the

balance. This procedure minimized the effect of inherent

torque in the tygon tube.

The distance of the weight pan knife edge from the ful­

crum was 2.75 inches. A jet from the capillary tube would

thus produce a horizontal force which would be counter­

balanced by placing weights on the pan at a ratio of 2:1. In

other words~ the additional weight on the pan needed to re­

balance the thrustometer divided by 2 was the thrust.

Seven stainless steel capillary tubes shown in Figure 13

were prepared for the thrust measurements. The dimensions of

these tubes are listed in !able II and a complete table of

specifications are given in Appendix II~ Table XIII.

Page 45: An apparatus to measure capillary jet thrusts and the

Figure ·13 CAPILLARY TUBES WITH ADAPTERS

FOR THRUST MEASUREMENTS w w

Page 46: An apparatus to measure capillary jet thrusts and the

34

The adapters for the capillary tubes were machined from

1/2-inch aluminum stock to 3/8-inch O.D. and 3/4-inch length.

A hole was drilled in the center to accommodate the proper

capillary tubes which were glued into the adapters using an

epoxy glue.* The capillary tube entrance was made flush

with the end of the adapter. Thus the fluid passed from a

large area through a sharp restriction to a small area.

This procedure was followed so that the entrance effect

was nearly constant for all the tubes.

Table II

THRUSTOMETER CAPILLARY TUBES

I.D.(in} Length {in} L/D

.01022 6.244 611

.01339 6.248 467

.02326 6.240 268

.03283 12.084 368

.04647 12.264 264

.06137 12.016 191

.1045 12.252 116

C. Viscometer

The viscometer arrangement was shown in Figures 7 and

8 and a detailed diagram of the water jacket is shown in

Figure 14. The jacket consisted of a 3/4-inch O.D.

*The epoxy glue~ mixed by Dr. K. Mayhan~ consisted of a mixture of 100 parts by weight CIBA No. 6010 epoxy resin (M. W. 390-410) to 13 parts CIBA hardener No. 951 (Triethylenetetramine); procured from Harry Baumstock Co.~ St. Louis~ Missouri. ·

Page 47: An apparatus to measure capillary jet thrusts and the

Figure V+

DETAILED DIAGRAM OF VISCOMETER JACKET

1. Inlet water line from bath: 1/4-inch galvanized pipe

2. 1/4-inch union

3. 1/2 x 1/4 galvanized bushing

4. 3/8-inch galvanized tee

5· Outlet water line to centrifugal pump: 1/4-inch

galvanized pipe

6. Viscometer heat exchanger jacket: 3/8-inch galvanized

pipe

7. Brass male connector: Swagelok 600-1-4

35

Page 48: An apparatus to measure capillary jet thrusts and the
Page 49: An apparatus to measure capillary jet thrusts and the

galvanized steel pipe with brass Swagelok fittings on each

end.* A capillary tube with 3/8-inch O.D. aluminum adapters

on each end was placed inside the tube and secured at the

ends. One end was connected to the process fluid outlet of

the constant temperature bath. The pipe was then connected

to the water pump and bath so that water would be pumped from

the bath through the galvanized pipe surrounding the capillary

tube and back into the bath. In this way the temperature of

the fluid within the capillary tube could be kept at a con­

stant temperature within the same limits as the bath (±0.059 C).

Temperature control is a factor of great importance in vis­

cometry work, because of the rapid change of viscosity with

temperature.

Four jackets were constructed and are shown in Figure

15. The lengths of these were 48, 24, 12, and 6 inches.

These lengths were used to accommodate the lengths of capil­

lary tubes (see Table III). The viscometer tubes and jackets

could be changed in 10-15 minutes. Seven stainless steel

capillary tubes as shown in Figure 16 were used for the vis­

cometer experimentation. The dimensions for these tubes

are listed in Table III and a complete list of specifications

is given in Appendix II, Table XIII.

*Contact between process fluid and any copper-containing metal was avoided in this apparatus because of the catalytic effect copper has in many organic reactions. The brass Swagelok fittings used here acted only to seal the water in the jackets and help align the tubes, thus they were never in contact with the process fluid.

Page 50: An apparatus to measure capillary jet thrusts and the

Figure 15 VISCOMETER HEAT EXCHANGER JACKETS

UJ -.,'J

Page 51: An apparatus to measure capillary jet thrusts and the

Figure 16 CAPILLARY TUBES WITH ADAPTERS,

FOR VISCOMETRY }ffiASUREMENTS

w 00

Page 52: An apparatus to measure capillary jet thrusts and the

Table III

VISCOMETER CAPILLARY TUBES

I.D.{in) Length (in) L/D

.009639 6.142 637

.01331 6.244 469

.02322 12.087 520

.03254 24.220 744

.04625 24.220 524

.06290 24.220 385

.1033 48.025 465

ExEerimental Procedure

A. Capillary Thrust Measurements

The following procedure was used in obtaining thrust

data. The room temperature was adjusted so as to be close

to the temperature of the process fluid during the experi­

ment. The thermoregulator was turned on and the bath was

regulated to the desired constant temperature. The analy­

tical balance (thrustometer) was placed in position and the

tygon tubing was connected to the outlet of the process

39

fluid coil and to the 90° part of the tee of the thrustometer.

In connecting the tubing, the balance hung fre~~n its ful­

crum, and the tubing was secured with the pointer of the

balance zeroed as closely as possible. This was done to

minimize any twist that may have been put into the tubing in

tightening it. The capillary tube was then inserted into

the lower elbow and secured. A small level was then used

to level the entire thrustometer.

Page 53: An apparatus to measure capillary jet thrusts and the

40

Process fluid was then poured into the reservoir

through the funnel and the gear pump was started. Fluid was

added intermittently to the reservoir as the system began to

fill. With this procedure, a reservoir capacity of only 15

cubic inches was required. Thus, the amount of evaporation

of solvent into the head space was minimized. In filling the

system~ air trapped within the coil was discharged through

the vent valve located at the high point of the coil. The

process fluid manometers were filled to the level of the hori­

zontal capillary tube to maintain a continuous leg of process

fluid.

As soon as the fluid began to jet from the capillary

tube~ it was directed into the reservoir, completing the re­

cycling system. Process fluid was allowed to recirculate

through the system for at least 10 minutes to insure that it

was at the bath temperature.

By placing a finger over the capillary tube exit, pres­

sure could be built up in the process fluid and the process

fluid manometer level was raised. The pump was then stopped

and fluid was allowed to flow from the manometer out through

the capillary tube. This procedure served to flush out air

trapped in the vertical glass tube and the horizontal tube

inside the bath. This was repeated several times until no

air was discharged from the system.

After all the air was out of the system, the pump was

stopped and the pressure was again allowed to go to zero.

The surface tension of the fluid at the end of the capillary

Page 54: An apparatus to measure capillary jet thrusts and the

41

held the fluid in the tube. At this point the level of the

thrustometer and the alignment of the capillary tube were

checked again. The balance was then zeroed by placing the

proper weights in the balance pans. A magnifying glass was

used to reduce parallax in reading the scale. The pump was

then set to deliver the desired flow rate. The proper

manometer was then valved into the system and allowed to

reach equilibrium. The thrustometer was then balanced to

overcome the imbalance caused by the thrust produced by~the­

fluid jetting from the capillary tube. A fluid sample was

then collected over a time period of 1.0 minute to 2.5 min­

utes~ depending upon the flow rate~ and the volume measured.

A stop watch readable to 0.2 seconds was used for the time

measurements. The flow rate in cc/min was thus obtained.

The manometer was checked again and the reading was recorded

as the pressure drop across the capillary tube.*

The temperature of the process fluid~ which was read as

the temperature of the bath3 and that of the manometer fluids

It was found that the pressure drop 3 in the line from the pressure tap inside the bath to the entrance to the capillary tube~ was so small that it could be neglected. Therefore, the manometer or pressure indicator reading recorded could be taken as the actual pressure drop across the capillary tube. Pressure readings were only taken to prove the accuracy of the data during the initial phase of the project. It was found that the manometers required a considerable time (5-10 minutes) to reach equilibrium when data was being taken with one of the smaller diameter tubes. Thus it was decided to valve off the manometer system and wait 2 minutes for the system to reach equilibrium between different flow rate settings and to calculate the pressure drop from viscometry data on the same solution. This procedure proved to be satisfactory and reduced the time required for taking data.

Page 55: An apparatus to measure capillary jet thrusts and the

were then recorded.* The setting of the variable speed

drive was recorded so that the point could be repeated if

desired. Before recording the weight on the weigh pans, a

check was made to see if the pointer had deviated from the

zero position. If it hadJ the point was rerun from the

beginning. It should be noted that deviations generally

did not occur. If no deviation was observedJ the total

weight on the pan was recorded. The pump was stopped and

the pressure drop allowed to reduce to zero, after which a

check was made on the zero thrust point.

42

It was found, after much experimentation, that the zero

thrust reading had a tendency to change from that recorded

before the first run.** It was also noted that after the

first data point was taken, the zero thrust weight usually

remained nearly constant throughout a series of data points.

Apparently after the initial movement, little further movement

occurred. Careful examination of the data and rechecks of

the data showed that the zero thrust reading recorded after

the run produced the most consistent results.

The repeatability of duplicate thrust measurements

obtained in two different series of measurements using the

same tube and flow rates was best using this procedure. *The temperature of the process fluid in this apparatus could not be controlled as well as in the viscometry unit which had jacketed heat exchangers. However, the thrust measurements are not as sensitive to temperature variations as the pres­sure measurements. Viscometry data taken with the thrust tubes is not as reliable as that taken with the capillary tubes. **This is believed to be due to expansion of the tygon tubing which caused a slight shift in the position of the balance knife-edge. This shift was observed to be about 1/32 inch.

Page 56: An apparatus to measure capillary jet thrusts and the

Between each series of data points the capillary and tygon

tubes were dissembled and reassembled, thus insuring no

repeated experimental errors in the assembly of the thrust­

ometer. The thrust was half the difference between the

total thrust weight and the zero thrust weight.

This procedure was followed for all data points. The

time required for equipment set-up and obtaining the first

data point was 45 minutes to an hour. The following data

points in a series required about 15 minutes per point.

After a series of thrust and viscometer runs with a

process fluid was completed, the fluid was drained from the

system. The system was then filled with a good solvent for

the polymer studied and recycled for 10-15 minutes to dis­

solve any residual polymer. The wash solution was then

drained, fresh solvent was added and the procedure was re­

peated. The system was then drained again.

If a solution containing a solvent different from the

wash solvent was to be studied next, the system was flushed

with about 150 cc of the solution to be tested. This was

drained and the system was then filled with test solution.

B. Viscometer Measurements

43

The following procedure was used in obtaining data with

the viscometer. The thrustometer was disengaged from the

bath and a 90°, 3/8-inch steel Swagelok elbow was secured to

the outlet of the bath. A viscometer capillary tube was

placed inside its proper jacket and secured by brass fittings.

The jacket was then connected to the low pressure side of the

Page 57: An apparatus to measure capillary jet thrusts and the

44

centrifugal pump and the fluid entrance end of the capillary

tube was secured in the elbow.

It was then necessary to prime the water pump by filling

the jacket manually with water. Air was bled from the pump

by a capped Swagelok fitting threaded into the pump chamber.

An 18-inch x 3/8-inch O.D. galvanized pipe with an open end

inside the constant temperature bath near the bottom was

connected to the viscometer jacket. A pipe was fitted to

the outlet of the pump with one end placed just under the

water level of the bath. The pump was then turned on and

proceeded to draw water from the bottom of the bath and

discharge it at the top of the bath. The temperature of the

bath was then regulated to the desired temperature.

Process fluid was introduced into the system. The

reservoir was placed in position to receive the fluid dis­

charged from capillary tubes~ completing the recycling sys­

tem. Pressure in the system was built up and released inter­

mittently as described in the previous section in order to

discharge all air that may have been trapped within the sys­

tem. After all the air was out of the system, the gear pump

was set to deliver the desired flow rate and for low pressures

the proper manometer was valved into the system. When the

manometer reached equilibrium, a fluid sample was collected

over a time period of 1.0 to 2.5 minutes depending upon the

flow rate and the volume measured. A stop watch was used

to measure the time. The flow rate in cc/min was thus ob­

tained. The manometer was checked and if it.had not changed~

Page 58: An apparatus to measure capillary jet thrusts and the

the reading was recorded. The bath temperature, manometer

board temperature, and the drive setting were recorded.

It required 45 minutes to an hour to set up the vis­

cometer and record the first data point. The following

data points in a series required about 15 minutes per point.

Page 59: An apparatus to measure capillary jet thrusts and the

46

IV. RESULTS AND DISCUSSION

Calibration of Tubes

and Entrance Effects Correction

The most important variable in capillary viscometry and

capillary thrust measurements is the diameter. (Equations 16~

10). Thus~ accurate calibration of the capillary tubes used

is vital for obtaining ·reliable measurements. The usual pro­

cedure followed in calibrating the capillary tubes is to

use heavy fluids~ such as oils of known viscosity and

density~ and apply the following equation to experimental

data:

(16)

where ~ is the viscosity, Q is the flow rate~ L is the capil­

lary length and ~P is the pressure drop corrected for

entrance effects.

In this work a variation of this procedure was used.

In place of an oil, a light Newtonian fluid (toluene) was

used, because its viscosity was known and it was close to

the viscosities of many of the fluids which would be later

studied. Relatively high velocities could be attained in

each tube with toluene and a sensitive check of the entrance

correction could be obtained.

Using toluene, flow rate data in the laminar region were

obtained for the capillary tubes* to be used in the viscometry *The pressure drops measured for toluene with the largest diameter thrust and viscometer tubes were too small for calibration. Therefore, the procedures described in Appendix I were used.

Page 60: An apparatus to measure capillary jet thrusts and the

47

and in the thrust measurements. The data obtained are tabu-

lated in Appendix II~ Table IX~ and plotted in Figures 17

and 18. The following iterative procedure was then used in

calculating the correct diameters. (See also Appendix I for

a sample calculation.)

The tube diameter listed by the manufacturer or an

estimate based on prelim~nary experiments* was used to

calculate the fluid velocity from the measured flow rate

using the equation:

(17)

where d is the assumed diameter~ to be used in the first

iteration and Q is the flow rate. This velocity was then

used to calculate the pressure drop across the tube using

equation (15)~ reproduced here for convenience.

(15)

The suggested Newtonian value of C equal to 2.16 was assumed

at this point (3). Using these values in equation (16)~ a

new diameter was calculated. The difference between the

newly calculated diameter and the assumed diameter was then

determined. If the difference was more than 5.0 x 10-6

inches, the new diameter was used to recalculate another value

of the diameter, using the same procedure. The resulting *The diameter~ based on preliminary experiments~ was a diameter obtained from data obtained with the apparatus before it was completely calibrated.

Page 61: An apparatus to measure capillary jet thrusts and the

,..... (\) s

103

~ 10 Cll QJ

~ "0 .._,

~ P-f <1 A

10

1

'

10

I I I I I . ..

G> .• 00964 .. • .01331 .. ~ .02322 '·. : ; • 03254

·I. ," .,,

... ..,_..

"--- - ----

ul-

Figure 17 WALL SHEAR STRESS VS 8V/D FOR TOLUENE

USING VISCOMETER TUBES

I I

. & .04625 ... 0629 v .1_033

. / ..

..4 1P .r

!lid

,a-/

tf"

~7 ~

-:r ~

pfC

Jf r

~

v~

lo3 104 8V/D (sec- 1) .

(!

j>l ...

"-~

loS uP .po co

Page 62: An apparatus to measure capillary jet thrusts and the

103

,...... 102 (\J s

~ fll QJ c:l >.

"0 ........,

H

~ Po. <l 10 A

1 10

Figure 18

WALL SHEAR STRESS VS 8V/D FOR TOLUENE USING THRUST TUBES

I I I I

I ~~~~~~-~

•• 03283 _ll 0 .01022 .•.. :OJ.339. ·A.. • 04647 &1 ~~326 A .06137 i

V' ~ .rn s

~~

~ r

.V ~

.F v

~ ~

10 ]03 10

BV/D-(sec- 1)

~-

_."' ~

'*

10

I

I

I

I !

I I

]06

..!=" \0

Page 63: An apparatus to measure capillary jet thrusts and the

50

diameter was then compared to the previously calculated

diameter. This procedure was followed until the difference

between two successively calculated diameters was less than

the specified tolerance. As shown in Table X, two or three

iterations were generally required.

The last value of the diameter thus calculated was

listed as the diameter for the capillary tube for that

particular data point. The same procedure was followed for

each data point in the set of data obtained for each tube.

The average of these calculated diameuers was taken as the

correct diameter for that tube.

Using the same data, but with the newly calculated

diameters, the friction factor {f) for the tube was calc­

ulated using two different approaches. The first value was

obtained from:

f.=_ 16/Re (18)

where Re is the Newtonian Reynolds Number. The second value

using the experimental data was calculated from the equation:

f = :J(@Jgc

(19)

where ~p is found from equation (15)~ p = density and gc is 2 a dimensional constant= 32.17 (lbm- ft./lbf- sec).

Since the fluid used was a Newtonian fluid, the ratio

of the two friction factors calculated should be 1.0. Since

f from equation (19) is dependent on the magnitude of C in

Page 64: An apparatus to measure capillary jet thrusts and the

51

equation (15)~ it was possible to check the value of C from

this ratio. Beginning with an arbitrarily chosen value of

C equal to 2.51~ the friction factors and their ratios were

evaluated for each data point in the set of data for a

particular tube. The average friction factor ratio was then

calculated. Using the same procedure~ an average friction

factor ratio was evaluated for the tube at successive values

of C incremented by -0.05 down to 1.96. It was found that

the average friction factor ratio for each tube was equal

to 1.0000 + .0009 at a value of C equal to 2.16. The small

range of values about 1.0000 indicates that the use of

C = 2.16~ which had been proposed by Bogue (3) for Newtonian

fluids~ gave consistent results.

In order to reduce the error caused by entrance effects

and insure that a proper velocity profile was developed~ all

tubes used had large L/D ratios. (see Tables II and III).

The L/D ratio for the capillary viscometer tubes ranged from

744 to 385. Since the thrust tubes were designed primarily

to obtain thrust data which only requires that a fully

developed velocity profile exist at the exit end of the tube~

lower L/D ratios are tolerable. Nevertheless~ the minimum

L/D ratio of 116 for the thrust tubes permits their use for

viscometry also.

In the process of obtaining the above data~ the manom­

eters~ pressure gauges~ gear pumps and variable speed drive

were calibrated and checked for reproduceability in obtain­

ing data. The calibration data for the Bourdon gauges using

a dead weight tester are shown in Table XIV.

Page 65: An apparatus to measure capillary jet thrusts and the

Viscometry Measurements for Toluene

and 1.0% polyisobutylene in cyclohexane

52

Log-log plots of the shear stress,~w' versus 8V/b for

data obtained from the viscometer and the thrustometer tubes

using toluene and 1.0% polyisobutylene (PIB) in cyclohexane*

are plotted in Figures 17, 18, 19, 20, and 21 and the data

are tabulated in Appendix II, Tables IX, XI, and XII.

For each plot, the data can be described by the equa-

tion:

(20)

The data obtained with the viscometer tubes and the thrust

tubes, using toluene, were fitted separately to first-order

curves by the least squares procedure. The values of n',

log K1 (which is the value of log~w at (8V/D) = 1), the 95%

confidence limits of each, and the standard deviation of the

data from the curves are listed in Table IV. Toluene data

were taken over the range of (8V/D) values of 200-1.2 x 105.

For Newtonian fluids 8V/D is the shear rate at the wall.

Since toluene is a Newtonian fluid, the theoretical

value of n' should be unity. It can be seen from Table IV

that the n's calculated for each set of tubes are very close

to 1.000. The small standard deviation obtained indicated

that the equipment and procedure used are internally con-

sistent.

See Appendix II for a description of the materials used.

Page 66: An apparatus to measure capillary jet thrusts and the

""" (\)

~ Cl) QJ

103

~ l>"l ~ 102

....:! ~ ~ A

10 2 10

Figure 19 WALL SHEAR STRESS VS 8V/D FOR 1.0% PIB USING VISCOMETER TUBES

L

I T I T I

0 .04625 .II

8 _;' .• 06290 fiT A e:.10326 j

. . ;:t:J •

. t .'•' ~

----,.-- • I . ~· ~

j

/ ;

I

/ rdR

/ A, ~'

/ w ~

~,

~

103 8V/D (sec- 1)

10 105 \Jl LV

Page 67: An apparatus to measure capillary jet thrusts and the

Figure 20

WALL SHEAR STRESS VS 8V/D FOR 1.0% PIB USING THRUST TUBES

I -r I I I I

!-

~ ~~~: .

. , '•. ~¢-: :. 03~~.~: : ·:-:.: ( 66l.~·· ~ •,<• •• ~

.:1 .._.to~s·, .::~ ' , ., ' ' " .

lrJ: ; '·· ":::j· "\ '' . :"~i ....

~ ~

r./' /

a /

.J!f ~

~~

) ~

v ~

, / ~

54

I~

v ~ A

Page 68: An apparatus to measure capillary jet thrusts and the

,.... 11 ~

Cl.l Q) c:: >.

'"0 ..._,

103

....:l ~ 102

a

10 2 10

Figure 21 WALL SHEAR STRESS VS 8V/D FOR 1.0% PIB

USING VISCOMETER AND THRUST TUBES I I l :J-.-:J l J I IJ,,.

(i) .03283 • .04625 .• :If .

... . . . . i16 .-~= 'i: . ~: .

•. 10326 . ..-.,.~ -~. -·r""" .,.,_,

, / ~

17 ~

,,.lltl!

J. ll J3

~ v

j/

103

~ b'

• ~ .,

L.l" ,. IJJ -~~~

II ~

10 8V/D (sec- 1)

II ~ • t:l

% ~

( v

-10 106

\Jl \Jl

Page 69: An apparatus to measure capillary jet thrusts and the

Table IV

VISCOMETER AND THRUST TUBE CURVE CONSTANTS FROM log'\ w vs log 8V /D

Solution Toluene**

n' 0.9989

95% limit on n' ±0.0036

n' upper bound 1.0026

n' lower bound .9953

Log K1 -2.282

95% limit on Log K' ±0.015

Standard Deviation 0.0071

*Viscometer tubes only **Thrust tubes only #Viscometer and thrust tubes

Toluene* PIB* PIB** PIB#

1.0047 0.8895 0.8797 0.8882

±0.0026 ±0.0266 ±0.0014 ±0.0111

1.0073 0.9161 0.8801 0.8993

1.0021 0.8629 0.8783 0.8771

-2.305 -0.9174 -0.8676 - .9067

±0.010 ±0.0969 +0.0592 + .0439

0.0075 0.0174 0.0167 0.0176

The 1.0% PIB data obtained, using the viscometer and

thrust tubes, were fitted separately and together to first­

order curves by the least squares procedure.* The constants

obtained from these curves are also given in Table IV.

It can be seen from Table IV that the values of n' ob-

tained for each set of data agreed fairly well.** The value

of n' obtained from the combination of the data was taken to

be the most correct value. This value of n' = .8882 in the

The entrance correction constant for non-Newtonian fluids such as 1.0% PIB in cyclohexane was assumed to be 2.16, the same as for Newtonian fluids (3).

**An analysis of variance check indicated that a second-order least squares curve did not improve the fit of the data.

Page 70: An apparatus to measure capillary jet thrusts and the

57

range of 8V/D values studied (about 300-75,000 sec- 1) is

typical of a mildly pseudoplastic non-Newtonian fluid. The

uncertainty in the value of n' in this fluid is greater than

that for toluene. This is due in part to the fact that 8V/D

above 75,000 could not be attained because of pressure limit­

ations and data were not obtained below 300, so that values

of n 1 were obtained from a narrower range of data. Also,

each set of toluene data had about twice as many measured

points.

It should also be noted from Figure 21 that, when all

the data points for the different diameter tubes are plotted

together, each tube appears to fit the straight line with no

consistent deviations for any tube. Thus it appears that

the tube diameters and entrance corrections obtained with

toluene give satisfactory results with a non-Newtonian fluid.

The tubes were originally selected so that a wide range

of 8V/D values in the laminar range could be attained with­

out too much overlap of data points when low viscosity fluids

such as toluene were used. However, it was found that with

fluids of higher viscosities, a larger range of 8V/D values

could be covered with each tube, in the same range of Rey­

nolds Numbers. Therefore, some intermediate tubes were not

used with fluids of high viscosities, because the attainable

8V/D range could be covered without them. The time required

for taking data was thus reduced considerably.

Page 71: An apparatus to measure capillary jet thrusts and the

Thrust Measurement for

1.0% PIB in cyclohexane

Curves of thrust versus 8V/D are plotted in Figure 22

and the data are tabulated in Appendix II, Table XII. All

thrust measurements were taken at Generalized Reynolds

Numbers* between 100 and 1000. Middleman and Gavis (15)

suggested that acceleration effects caused by the relaxing

velocity profile and surface tension effects at the tube

exit at low Reynolds number could cause a significant normal

stress to be observed independent of elastic stresses. The

minimum Reynolds number above which this effect could be

neglected was 100. Reynolds numbers above 1000 were not

used, because it was reported elsewhere (19,21) that unusual

effects in the data were observed above this value.

An analysis of variance check indicated that the fit of

the data for the .1045 inch diameter tube could be improved

considerably by using a second-order least squares curve as

compared with first-order. However, first-order curves ade­

quately fit the data for the .06137 inch and the .0328 inch

diameter tubes. The constants and their confidence limits

are listed in Table v. The values of d(log T)/d(log 8V/D) for the 0.1045 inch

diameter tube used in calculating normal stress differences

had a range of 2.03 to 2.36· The other two tubes had constant

. n' 2-n' n '- 1 The Generalized Reynolds Number ~s D V p/K1 8 . In the case of Newtonian fluids where n' = 1, this reduces to the conventional Reynolds Number.

Page 72: An apparatus to measure capillary jet thrusts and the

10

-til 51 -+J 1. til :j $-1

eS ':

Figure 22

THRUST VS 8V/D FOR 1.0% PIB

·' :"'

~ ' ~ ..... -l"',

... ~ .• 03283' ' ;:,?~~til • 06137 ·.:·

fA .1045 .

··---- .

' j ~ I

liD I

I( irrf d 1r p. IJ

II J

j lJ

I ~

' I

104 -1) 8V/D (sec

1

' '{

I

1 I

l~

L t

59

I ~

J

v ceJ

ll

~

Page 73: An apparatus to measure capillary jet thrusts and the

Table V

CONSTANTS FOR THE CURVES OF THE LOG THRUST VS LOG 8V/D FOR 1.0% PIB

Tube diameter 0.1045

Order of fit 2nd

Sl d{log T} 7.23- 2(.701) log 8V/D ope d(log 8V/D)

95% limit on slopes 7.23 + 4.800 of coefficients -.701 + .0692

Log T intercept -16.399

95% limit on Log T intercept + 2. 270

Standard deviation 0.0098

0.06137

1st

2.1087

+0.0411

-8.626

+1.696

0.0091

0.0328

1st

2.1445

+0.0370

-9.894

+1.768

0.0021

0'1 0

Page 74: An apparatus to measure capillary jet thrusts and the

values of d(log T)/d(log 8V/D) because their curves were

first-order.

61

Figure 23 is a plot of the log normal stress difference

(Pll - P22)w versus log 8V/D for the two larger tubes. The

data were fitted with a first-order curve of the form

Log (Pll - P22)w = -3-328 + 1.1508 (Log 8V/D)

where the 95% confidence limit on the terms -3.328 and

1.1508 are +0.4158 and +0.1075, respectively. The standard

deviation of the data from the curve was 0.0675.

The values of (P11 - P22)w obtained in this work for

1.0% PIB L-80 in cyclohexane are of the same order of mag­

nitude as those obtained by Uebler (21) for 1.0% PIB L-100

in decalin. His values were slightly higher, which is

explainable since the polymer he used had a slightly higher

molecular weight and was in a different solvent. The slope

of about one is also in agreement with his results.

In order to show why points on this plot, at nearly

equal values of 8V/D taken with the same tubes, had a fairly

large deviation~ the probable error (P.E.) encountered in

evaluating (P11 - P22)w was found following Shertzer (19).

It was assumed that the thrust had a P.E. of 5% and that

the slope of the~w vs 8V/D plot was not in error. From this

assumption, the P.E. of the second term in Equation 10 was

also 5%. But the error introduced into the difference

between the first and second terms (P11 - P22)w was much

larger. This is shown in the last column of Table VI which

Page 75: An apparatus to measure capillary jet thrusts and the

62

Figure 23

NORMAL STRESS DIFFERENCES VS 8V/D FOR l.Oi PIB

*'{_ '+'

' t;;: 0 . 06137

.&.,.1045

f.- .. --·--·· ·---·~· r--- r·-

~ .. {!)L .. L

If" .J

~

10 . ~ ~~

.V •I:¥ ~

L .J'!i

¥ ~ , /

¥ I

1

Page 76: An apparatus to measure capillary jet thrusts and the

shows a few sample points for the .06137 tube. This demon­

strates why the (P11 - P22)w points scatter about the curve.

A small error in the measurement of the thrust results in a

large error in evaluation of (P11 - P22)w, especially at

higher values of 8V/D where the individual terms of equation

(10) increase more rapidly than their difference.

Table VI

P.E. IN (Pll - p22)w DUE TO ERROR IN THRUST

1st 2nd P.E. in P.E. in Term Term pll-P22 P.E. in pll-P22' pll-P22%

{8V/D) (:2sf) (Esf} (Esf} T % {:esf}

19270 463 430 33 s.o 21.5 65.0

14410 363 338 25 s.o 16.9 66.0

7395 92 78 14 s.o 3.9 27-5

It can be noted from the curves of Figure 22 and from

Table XII that in repeated runs, values of thrust varied by

less than about 5%. There is a trend to lower per cent

variations at high thrust values. This demonstrates the

repeatability of the thrust measurements, but as was shown

in the preceding error analysis, is misleading in that small

differences in the thrust measurement will be magnified when

normal stress differences are calculated. It is believed,

however, that the good repeatability obtained with the .03283

inch tube is fortuitous in view of the sensitivity problems

described below.

The normal stress differences for the 0.03283 inch

diameter tube are not plotted in Figure 23 because the normal

Page 77: An apparatus to measure capillary jet thrusts and the

64

stress differences (P11 - P22)w evaluated for this tube were

all small or negative.

It would appear that the values of (P11 - P22 )w for

this tube were small or negative because the measured thrusts

were larger than the true thrusts produced by the jets. The

alternate explanation that negative differences were due t,o

an incorrect diameter requires that the calculated diameter

was too large. This seems unlikely~ particularly in view

of the good agreement of data taken with this tube with the

rest of the data shown in Figure 18. The reason for the

apparent bias in these readings is not understood.

All thrust measurements taken at high pressures were

affected by the tygon tubing connecting the constant tempera­

tube bath to the analytical balance. As the pressure rose~

the tubing had a tendency to swell and elongate. This

reduced the flexibility of the tubing which in turn prevented

the knife edge from pivoting freely on the balance. The

effect of pressure in the tygon tubing on the sensitivity of

the balance is shown in Table VII which lists the weight

required to give the pointer a unit displacement at different

values of pressure drops across the capillary tubes. At the

pressures prevalent in the .03282 inch diameter tube with

this solution (20-60 psi), the sensitivity is inadequate to

obtain meaningful (P11 - P22)w results since a small error

in the thrust term of equation (10) produces a large error

in (P11 - P 22)w. Thus~ the use of the thrustometer is

limited by the flexibility of the tygon tubing at high

Page 78: An apparatus to measure capillary jet thrusts and the

pressures (high shear rates).

Table VII

VARIATION OF BALANCE SENSITIVITY WITH PRESSURE DROP

Tube Diameter (inches)

.1045

.06137

.03283

Weight* (milligram)

(10-20)

(25- 50)

(100-200)

*Thrust equals (Weight/2)

Pressure Drop (psi)

0-5

5-15

20-60

It is of interest to note in Table VIII that the

Generalized Reynolds Numbers are nearly the same for

different tubes at nearly equal values of thrusts.

Tube Diameter {inches)

0.0328

0.06137

0.1045

Table VIII

COMPARISON OF REYNOLDS NUMBERS AND THRUSTS

Generalized Reynolds Thrust

Number (gms)

540 1.255 940 3-755

235 0.340 520 1.420 970 3.810

200 0.260 490 1.475

8V/Dl {sec- }

45530 76416

7395 14400 25032

2337 5288

This fact should be helpful in selecting the proper diameter

capillary tubes when designing an experiment to measure

thrust.

Page 79: An apparatus to measure capillary jet thrusts and the

V. CONCLUSIONS

The viscometer designed and built for this project

operates satisfactorily in a continuous range of flow rates

of 2 to 560 cc/min and capillary wall shear stresses of

1.0 to 660 dynes/cm2 • With seven capillary tubes, the com­

plete 8V/D range of 200 to 1.2 x 105 sec- 1 can be covered

even for fluids of low (0.5 cp) viscosity.

66

The Newtonian entrance correction factor for each capil­

lary tube was checked and a value of 2.16, the same as that

proposed by Bogue (3), appears to be satisfactory.

The capillary jet thrust apparatus using an analytical

balance for thrust measurements, gives reasonable normal

stress results for a 1.0% solution of polyisobutylene in

cyclohexane at values of 8V/D up to 3 x 104 sec- 1 . The

equipment is relatively simple and inexpensive and fairly

sensitive at low pressures. However, at higher shear rates,

where the pressure drop across the capillary tube exceeds 15

psi, reliable thrust measurements could not be obtained.

Under these conditions, the tygon tubing connector to the

balance swelled and stiffened causing the balance to lose

its sensitivity.

It was noted that in the Generalized Reynolds Number

range useful for these measurements (Re ~ 100 to 1000), the

range of thrust values obtained was the same regardless of

tube diameter. This observation is useful in selecting

tubes for covering desired 8V/D ranges.

Page 80: An apparatus to measure capillary jet thrusts and the

VI. RECOMMENDATIONS

No recommendations are offered for the capillary

viscometer as it satisfied all present requirements.

In order to extend the range of the 8V/D values for

the jet thrust apparatus, the following recommendations

are suggested for study:

1. Reduce L/0 ratio of the thrust tubes so that lower

pressure drops will be encountered across them. The minimum

L/b which will permit fully developed velocity profiles to

be achieved will have to be determined.

2. A suitable flexible connector which is less sensi­

tive to pressure than the tygon tubing and still is not

attacked by organic solvents should be sought.

3. It is also recommended that a longer glass tube

lever arm on the balance be used to increase the lever arm

ratio and thus give larger weight measurements for balancing

thrusts.

Page 81: An apparatus to measure capillary jet thrusts and the

68

APPENDIX I

Sample Calculations

I. Calculation of capillary tube diameters.

In order to show the iterative process used in obtaining

the capillary tube diameters~ a review of the calculations

will be presented for the 0.06137049 inch diameter capillary

thrust tube using toluene.

First iteration:

d = Estimated diameter of .06135913 inches

Q = 56. 85 cc/min

L = 12.016 inches

Tj = 0.5179 cp = 1.08268 X 105 lbf sec ft

v = 4Q/ud2

pV2 ~p = ~POBSERVED - V 2g = 0.231 psi

c

C assumed to be 2.16

D = 2(8QLTJ/u~P) 1/4 = 0.0621212 inches

-2

The value of D calculated was then compared to d. If

these two values differred by more than 5.0 x 10- 6 inches~ D was substituted for d in the velocity evaluation and a

different D was calculated. This iterative process was

continued until the difference between two successive

iterations was within the tolerance.

A diameter was calculated in this manner for each point

in the set of data for each tube (Table X). The values

Page 82: An apparatus to measure capillary jet thrusts and the

69

ranged from 0.06094642 to 0.06173789 inches. The average

of these calculated diameters was taken as the true diameter,

which for this tube was

D = DAVERAGE = 0.06137049 inches

Page 83: An apparatus to measure capillary jet thrusts and the

II. Check of the capillary end effects correction

factor C.

The 0.06137 inch diameter tube will be used again to

show a sample of the calculations followed in checking the

correct value of C. C was varied from 2.51 to 1.96 by

increments of 0.05. C equal to 2.16 will be used here.

Q = 56.85 cc/min

f = 16/Re

Re = VDp = 1328 T)

fl - 16/Re = 0.01205

D.P tiPOBSERVED -pV2

c 2g -

L = 12.016 inches

V = ;d~ = 19.55 in/sec

- DD.Pgc - 01 11 f2 - 2LV2 p - • 2

c = 0.231 psi

fl friction factor ratio = I: = 0.9991

2

70

The friction factor ratio was calculated for each data point

with C = 2.16. (Table X).

Page 84: An apparatus to measure capillary jet thrusts and the

71

III. Calculation of Normal Stress Difference

= I - II

An example of the calculations used to obtain (P11 - P22)w

is shown here. The 0.06137 inch diameter tube will be used

for this case. The computer analysis provided the following

relationships from the log T versus log 8V/D and lo~w

versus log 8V/D curves.

log T = -8.627 + 2.1087 log 8V/D

log"l-w = -~ 067 + . 8882 log 8V /D

from which, n' = 0.8882

Q = 249.0 cc/min

8V/D = 7.3953 x lo+3 sec- 1

p~2 = 0.585 psi c

( 3n 1 +l)(n 1 +1) = 2 _810 2n 1 +1 n'

I - 1.645 psi • ¥'_0 i)

T - 0.~ gm

d(log T) - 2.1087 d(log ~v) -

(1 + 1 d(log T) ) = 2.1960 2n • d(log ~V)

T = 0.255 psi 7TR2

Page 85: An apparatus to measure capillary jet thrusts and the

72

II = 0.565 psi 1.

(Pll - P22)w = I - II = 1.080 psi

Page 86: An apparatus to measure capillary jet thrusts and the

73

IV. Procedure to obtain diameter of largest viscometer tube.

Because the viscosity of toluene was too low (~.52 cp),

the pressure drops across the 0.10326 inch diameter viscometer

tube were too small to be reliable in calculating a diameter

for the tube. Therefore a Newtonian fluid of higher vis­

cosity was used to determine the diameter of this tube. The

fluid used was 4.0% Polystyrene in toluene which had a

viscosity of 4. 567 cp. Hershey (9) and Rodriguez (18) found

this fluid to possess Newtonian properties using other tubes~

and obtained flow data for the capillary tube in question.

The data thus obtained was used to calculate the

diameter of the tube by the same iterative procedure followed

for the other tubes. The data are tabulated in Appendix II,

Table X.

Page 87: An apparatus to measure capillary jet thrusts and the

v. Procedure to obtain the diameter of the largest thrust

tube.

74

Because the viscosity of toluene was too low (~.52 cp)~

the pressure drops measured across the largest diameter

thrust tube were too small to be reliable in calculating the

diameter. Therefore~ in order to determine the diameter of

the tube, the data obtained for the 1.0% PIB solution was used.

The log (D6P/4L) vs log (8V/D) data obtained with three

viscometer tubes (D = .1033~ .0629~ and.04625) and two thrust

tubes (D = .06137 and .03283) were fitted with a first-order

curve by the least squares method. This curve could be

represented by equation (20}:

I

D6P/4L = K1 (8V/D)n (20}

where n 1 = 0.8882 and K' = 0.128. (Table IV).

This equation can be rearranged to obtain:

[ r Jl/(3n'+l)

D = (32 Q/rr)n (4LK' /6P)

A reasonable diameter was assumed for the largest thrust tube

in order to correct the 6P term (equation 15)~ and a diameter

(D) was calculated for each measured data point. The numeri­

cal average of the diameters was then determined.and compared

to the assumed diameter. If they were different~ another

diameter was chosen and the same procedure followed. This

was done until the assumed and calculated diameters were

nearly equal. The value thus obtained as the diameter of the

tube was 0.10448 inches which was rounded off to 0.1045 inches.

Page 88: An apparatus to measure capillary jet thrusts and the

75

APPENDIX II

Page 89: An apparatus to measure capillary jet thrusts and the

Table IX

TOLUENE DATA FOR VISCOMETER

Pressure

AND~UST CAPILLARY TUBES

Shear Diameter Drop Velocit) Reynolds Stress 8V/D (inch_es_l . __ (~ (ft/sec Number f(Eq. 19) (dynes/cm2 ) (sec-1)

.00963885 19.10 9.86 1216 .01335 516.645 98215

.00963885 15.75 8.13 1003 .01619 426.035 80978

.00963885 13.21 6.89 851 .01887 357.407 68716

.00963885 11.08 5.81 718 .0~~~4 ~99.696 ~~§64 .00963885 8.83 4.66 576 .02755 238.992

.00963885 6.54 3.46 427 .03710 176.843 34473

.00963885 4.25 2.26 279 .05629 114.900 22558

.01331044 7-34 7.10 1210 .01345 269.920 51223

.01331044 5-65 5-54 944 .01698 207.634 39977

.01331044 4.86 4.80 818 .01945 178.719 34661

.01331044 4.05 4.09 698 .02224 148.692 29565

.01331044 2.83 2.75 470 .03430 103.899 19900

.01331044 1.98 1.93 329 .04886 72.481 13926

.02322385 2.84 4.36 1299 .01236 93.972 18063

.02322385 2.42 3·75 1117 .01425 80.168 15540

.02322385 2.02 3.10 922 .01747 66.882 12819

.02322385 1.63 2.55 759 .02085 54.082 10553

.02322385 1.23 1.90 566 .02830 40.860 7873

*Ratio of f values from Equations 19 and 18.

f(Eq. 18)

.01314

.01594

. 01879

.02227

.02776

. 03746

.05724

.01321

.01693

.01953

.02290

.03402

.04862

.01231

.01431

.01735

.02107

.02825

I

/ I ,J

Drag* Ratio

1.015 1.015 1.004

.998

.992

.990

.983

1.017 1.002

.995

. 971 1.008 1.004

1.004 .996

1.007 .989

1.002

~ 0\

Page 90: An apparatus to measure capillary jet thrusts and the

Table IX (continued)

Pressure Shear Diameter Drop Velocit} Reynolds Stress 8V/D Drag* (inches) __ (psi) (ft/sec Number f(Eq. 19) (dynes/cm2 ) (sec-1) f(Eq. 18) Ratio

.03253884 3.23 4.82 2011 .00806 74.836 14247 .00795 1.014

.03253884 2.87 4.36 1817 .00878 66.521 12873 .00880 .997

.03253884 2.52 3.82 1591 .01007 58.470 11270 .01005 1.001

.03253884 2.21 3.38 1411 .01120 51.219 9998 .01133 .989

.03253884 1.90 2.89 1204 .01322 43.946 8528 .01328 .995

.03253884 1.69 2.56 1069 .01490 39.061 7571 .01496 .996

.03253884 1.47 2.23 932 .01708 34.022 6600 .01716 .995

.03253884 1.26 1.92 800 .01994 29.295 5668 .01999 .997

.03253884 1.05 1.59 664 .02416 24.432 4703 .02409 1.003

.03253884 ·739 1.10 462 .03489 17.103 3274 .03460 1.008

.04624756 .792 2.36 1402 .01169 26.077 4915 .01141 1.024

.04624756 .701 2.11 1250 .01299 23.068 4384 .01279 1 ~JH6

.04624756 .611 1.85 1100 .01461 20.100 3859 .01453 1.005

.04624756 .521 1.62 960 .01638 17.147 3364 .01665 .983

.04624756 .436 1.32 785 .02046 14.337 275 .02036 1.005

.04624756 .347 1.08 640 . 02453 11.409 . 2244 .02499 .981

.04624756 .294 .917 543 .02885 9.670 1905 .02944 .980

.04624756 .241 .749 443 .03550 7.927 1555 .03606 .984

.04624756 .191 .586 347 .04591 6.289 1218 .04605 .997

.04624756 .142 .423 250 .06528 4.665 879 .06375 1.023

*Ratio of f values from Equations 19 and 18

-..l -'1

Page 91: An apparatus to measure capillary jet thrusts and the

Table IX (continued)

Pressure Shear Diameter Drop Velocit) Reynolds Stress 8V/D Drag* _(inches) (psi) (ft/sec Number f(Eq. 19) (dynes/cm2 ) (sec-1) f(Eq. 18) Ratio

.06290381 .243 1.37 fll04 .01451 10.867 2093 .01448 1.002

.06290381 .168 • 973 -:::\7 83 .01999 7·535 1485 .02041 • 979

.06290381 .141 .804 647 .02445 6.292 1227 .024~0 .989

.06290381 .118 .673 542 .02933 5.295 767 .039 8 .984

.06290381 .0639 ·359 289 .05551 2.862 549 .05518 1.005

.06290381 .0489 .274 221 .07307 2.192 419 .07233 1.010

.06290381 .0328 .182 147 .11079 1.471 278 .10873 1.018

.06290381 .0243 .136 109 .14745 1.090 208 .14569 1.012 •

.01021475 23.44 13.31 1740 .00937 661.041 125095 .00919 1.020

.01021475 18.86 10.79 1412 .01146 531.924 101476 .01133 1.012

.01021475 u~.o9 8.33 1090 .01437 397.325 78342 .01467 .979

.01021475 9.47 5·56 727 .02167 267.015 52293 .02198 .985

.01021475 4.71 2.72 355 .04512 132.839 25563 .04497 1.003

.01338756 10.13 10.11 1~33 .00919 374.169 ~2535 .00922 .996

.01338756 8.60 8.70 1 92 .01053 317.618 2432 .01072 .982

.01338756 7.10 7.07 1212 .01316 262.109 50731 .01319 .997

.tl1338756 5-55 5.53 949 .01679 204.977 39722 .01685 .996

.01338756 4.05 4.00 686 .02345 149.603 28712 .02331 1.006

.01338756 2.38 2.33 399 .04078 88.109 16709 .04006 1.018

.01338756 11.14 10.77 1846 .00891 411.587 77242 .00866 1.028

.01338756 8.20 8.24 1413 .01118 302.908 59151 .01131 .988

.01338756 6.40 6.47 1109 .01416 236.142 46414 .01442 .982

.01338756 4.67 4.63 793 .02021 172.428 33202 .02016 1.002

*Ratio of f values from Equations 19 and 18. ~ 00

Page 92: An apparatus to measure capillary jet thrusts and the

Table IX (continued

Pressure Shear Diameter Drop Velocity Reynolds Stress 8V/D (inches) (psi) U:tLsec2 Number f(Eq. 192 (dxnes/cm2 ) (sec- 1)

.02325907 2.14 6.26 1865 .00880 137.376 25852

.02325907 1.68 5·17 1541 .01015 108.222 21365

.02325907 1.36 4.15 1236 .01278 87.680 17133

.02325907 1.05 3.21 956 .01639 67.290 13255

.02325907 .843 2.49 743 .02182 54.167 10308

.02325907 .437 1.29 386 .04195 28.060 5351

.03282976 1.51 4.66 1962 .00813 70.612 13655

.03282976 1.34 4.18 1759 .00900 62.783 12238

.03282976 1.19 3.65 1536 .01045 55·563 10686

.03282976 1.01 3.15 1326 .01197 47.469 9228

.03282976 .853 2.64 1111 .01436 39.936 7729

.03282976 .691 2.15 904 .01755 32.355 6294

.03282976 ·535 1.66 699 .02275 25.035 4863

.03282976 ·378 1.17 492 .03240 17.694 3425

.03282976 .276 .830 348 .04721 12.941 2427

.04646582 .135 .818 486 .03317 8.844 1691

.04646582 .188 1.15 686 .02308 12.245 2385

.04646582 .253 1.52 906 .01787 16.525 3149

.04646582 .301 1.83 1093 .01462 19.654 3796

.04646582 -356 2.17 1292 .01237 23.251 4489

.04646582 .410 2.54 1511 .01042 26.780 5250

*Ratio of f .values from Equations 19 and 18.

f{Eg. 182

.00857

.01037

.01294

.01672

.02151

.04143

.00815

.00909

.01041

.01206

.01440

.01768

.02288

.03249

.04586

.03285

.02329

.01764

.01463

.01237

.01058

Drag* Ratio

1.026 . 978 .988 .980

1.014 1.012

.998

.990 1.003

.993

.997

.992

.994

.997 1.029

1.009 .991

1.013 .999 .999 .984

-...J \0

Page 93: An apparatus to measure capillary jet thrusts and the

Table IX (continued)

Pressure Shear Diameter Drop Ve1ocitl Reynolds Stress 8V/D (!nc_hes) (ps.ll_ Number f(Eq. 19) {dynes(cm2 ) (~~~-l) (ft(sec

.06137049 .146 1.62 1279 .01214 12.817 2547

.06137049 .137 1.52 1201 .01293 12.0~5 2392

.06137049 .123 1.37 1083 .01437 10.8 3 2156

.06137049 .113 1.23 967 .01643 9.917 1926

.06137049 .106 1.13 893 .01812 9.321 1778

.06137049 .0983 1.04 817 .02011 8.657 1626

.06137049 .0843 .902 709 .02291 7.425 1411

.06137049 .0726 .765 601 .02741 6.393 1197

*Ratio of f values from Equations 19 and 18.

f(Eq. 18)

.01250

.01331

.01477

.01653

.01791

.01957

.02256

.02659

Drag* Ratio

. 971

.971

.972

.993 1.012 1.027 1.015 1.030

00 0

Page 94: An apparatus to measure capillary jet thrusts and the

Table X

CALIBRATION OF TUBES AND ENTRANCE

EFFECTS COEFFICIENT DATA

81

Page 95: An apparatus to measure capillary jet thrusts and the

82'

Flow Temp. of Man. Temp. of Diameter Rate Man. Man. Reading Fluid {inches} {cc/min} No. {oc} {em} {oc}

.•

.01021371 12.87 1. 32-50 133-90 30.0

.01021371 10.44 1. 31.90 106.45 30.0

.01021371 8.06 1. 29.80 78.85 30.0

. 01021371 5.38 1 . 33.00 52.65 30.0

. 01021371 2.63 1 . 30.80 26.65 30.0

Calculated Number of Diameter Iterations

.01016798 2

.01018568 2

.01026494 3

.01024923 2

.01020595 2

Average Diameter = .01021475 inches

Average Ratio c

.9906 2.51

.9920 2.46

.9934 2.41

.9947 2.36

.9961 2.31

.9974 2.26

.9988 2.21 1.0002 2.16 1.0015 2.11 1.0029 2.06 1.0042 2.01 1.0056 1.96

Page 96: An apparatus to measure capillary jet thrusts and the

83

Flow Temp. of Man. Temp. of Diameter Rate Man. Man. Reading Fluid {inches} {cc/min} No. {oc} (em} (oc}

.01338529 16.80 1. 33.00 60.07 30.0

.01338529 14.46 1 . 31.60 50.48 30.0

. 01338529 11.75 1 . 30.30 41.09 30.0

. 01338529 9.20 1. 31.30 31.90 30.0

.01338529 6.65 1. 31-30 23.25 30.0

.01338529 3-87 1. 32.50 14.01 30.0

.01338529 17.89 1. 31.20 66.70 30.0

.01338529 13-70 1. 32.30 48.50 30.0

.01338529 10.75 1. 30.00 37-51 30.0

.01338529 7 .. 69 1. 32.40 27.34 30.0

Calculated Number of Diameter Iterations

.01339872 2

.01344058 2

.01339485 2

.01339835 2

.01336781 2

.01332860 2

.01330359 2

.01342148 3

.01344210 2

.01337856 2

Average Diameter = .01338756 inches.

Average Ratio c

.9855 2.51

.9877 2.46

.9898 2.41

.9920 2.36

.9942 2.31

.9963 2.26

.9985 2.21 1.0007 2.16 1.0028 2.11 1.0050 2.06 1.0071 2.01 1.0093 1.96

Page 97: An apparatus to measure capillary jet thrusts and the

84

Flow Temp. of Man. Temp. of Diameter Rate Man. Man. Reading Fluid {inches} {cc/min} No. {oC} {em} {oC}

.02325339 31.40 1. 30.70 15.46 30.0

.02325339 25.95 1. 31.00 12.33 30.0

.02325339 20.81 2. 30.30 143.10 30.0

.02325339 16.10 2. 32.60 110.20 30.0

.02325339 12.52 2. 29.50 88.95 30.0

.02325339 6.50 2. 31.90 51.90 30.0

Calculated Number of Diameter Iterations

. 02313759 4

.02336605 4

.02331817 3

.02336223 3

.02318076 3

.02318964 2

Average Diameter = . 02325907 inches .

Average Ratio c

.9770 2.51

.9803 2.46

.9836 2.41

.9868 2.36

.9901 2.31

.9933 2.26

.9966 2.21

.9999 2.16 1.0031 2.11 1.0064 2 .. 06 1.0097 2·.01. 1.0129 1.96

Page 98: An apparatus to measure capillary jet thrusts and the

Flow Temp. of Man. Temp. of Diameter Rate Man. Man. Reading Fluid (inches) (cc/min) No. (°C) (em) (oc)

.03280583 46.64 2. 31.50 159-75 30.0

.03280583 41.80 2. 34.00 141.95 30.0

.03280583 36.50 2 . 31.40 124.65 30.0

. 03280583 31.52 2. 33.80 107.20 30.0

.03280583 26.40 2 . 32.80 90.80 30.0

. 03280583 21.50 2. 32.00 75-00 30.0

.03280583 16.61 2 . 32.10 60.23 30.0

. 03280583 11.70 2. 33-50 45.95 30.0

.03280583 8.29 2. 32.20 32.20 30.0

Calculated Number of Diameter Iterations

.03283950 3

.03289597 3

.03280003 3

.03287866 3

.03284654 2

.03288563 2

.03287471 2

.03284905 2

.03259778 3

Average Diameter = .03282976 inches.

Average Ratio c

.9829 2.51

.9853 2.46

.9877 2.41

.9901 2.36

.9925 2.31

.9949 2.26

.9973 2.21

.9996 2.16 1.0020 2.11 1.0044 2.06 1.0068 2.01 1.0092 1.96

Page 99: An apparatus to measure capillary jet thrusts and the

86

Flow Temp. of Man. Temp. of Diameter Rate Man. Man. Reading Fluid {inches} {cc/min} No. {oc} {em} {oC}

.04642334 16.38 2. 33-70 25.25 30.1

.04642334 23.10 2. 35.20 30.25 30.0

.04642334 30.50 2. 32.00 36.60 30.0

.04642334 36.77 2. 32.60 41.64 30.0

.04642334 43.48 2 . 32.50 47-55 30.0

. 04642334 50.85 2. 32-50 53-78 30.0

Calculated Number of Diameter Iterations

.04635974 2

.04655883 3

.04632811 3

.04646985 3

.04646488 1

.04661354 4

Average Diameter = .04646582 inches.

Average Ratio c

.9790 2.51

.9819 2.46

.9849 2.41

.9878 2.36

.9908 2.31

.9937 2.26

.. 9967 2.21

.9996 2.16 1.0026 2.11 1.0055 2.06 1.0085 2.01 1.0114 1.96

Page 100: An apparatus to measure capillary jet thrusts and the

87

Flow Temp. of Man. Temp. of Diameter Rate Man. Man. Reading Fluid {inches} (cc/min} No. {oc} {em} {oC}

.06135913 56.85 2 . 32.10 28.10 30.0

. 06135913 53-38 2 . 32.30 27.05 30.0

. 06135913 48.11 2 . 32.00 25.50 30.0

. 06135913 42.99 2. 30.00 24.20 30.0

.06135913 39.68 2. 28.70 23.40 30.0

.06135913 36.30 2 . 28.00 22.56 30.0

. 06135913 31.50 2. 30-30 21.16 30.0

.06135913 26.72 2. 31.20 19.97 30.0

Calculated Number of Diameter Iterations

.06173319 4

.06173789 1

.06172566 2

. 06144971 4

.06121000 4

.06100342 3

.06115765 3

.06094642 3

Average Diameter = .06137049 inches.

Average Ratio c

.9730 2.51

.9768 2.46

.9805 2.41

.9843 2.36

.9881 2.31

.9918 2.26

.9956 2.21

.9994 2.16 1.0031 2.11 1.0069 2.06 1.0107 2.01 1.0144 1.96

Page 101: An apparatus to measure capillary jet thrusts and the

88

Flow Temp. of. Man. Temp. of Diameter Rate Man. Man. Reading Fluid (inches) (cc/min) No. (°C) (em) (° C)

. 01000000 8.49 1 . 28.60 105.90 30.0

.01000000 7.00 1. 29.50 86 ·70 30.0

.01000000 5.94 1. 27.90 72.45 30.0

.01000000 5.01 1. 27-50 60.60 30.0

.01000000 4.02 1. 28.20 48.30 30.0

.01000000 2.98 1. 26.40 35.85 30.1

.01000000 1.95 1. 27-70 23.65 30.0

Calculated Number of Diameter Iterations

.00960317 3

.00960244 1

.00962856 2

.00964206 2

.00965656 2

.00966107 1

.00967807 2

Average Diameter = .00963885 inches.

Average Ratio c

.9938 2.51

.9947 2.46

.9956 2 .. 41

.9965 2.36

.9974 2.31

.9983 2.26

.9991 2.21 1.0000 2.16 1.0009 2.11 1.0018 2.06 1.0027 2.01 1.0036 1.96

Page 102: An apparatus to measure capillary jet thrusts and the

89

Flow Temp. of Man. Temp. of Diameter Rate Man. Man. Reading Fluid (inches) (cc/min) No. (oC) (em) (oC)

.01327935 11.66 1. 24.10 42.35 30.0

.01327935 9.10 1. 23.90 32.38 30.0

.01327935 7-89 1. 23.90 27.85 30.0

.01327935 6.73 1. 23.90 23.25 30.0

.01327935 4.53 1. 23.70 16.40 30.0

.01327935 3-17 1. 23.80 11.76 30.0

Calculated Number of Diameter Iterations

.01325705 3

.01330082 2

.01332377 2

.01340295 2

.01328414 2

.01329381 2

Average Diameter = .01331044 inches.

Average Ratio c

.9913 2.51

.9925 2.46

.9938 2.41

.9950 2.36

.9962 2.31

.9975 2.26

.9987 2.21 1.0000 2.16 1.0012 2.11 1.0024 2.06 1.0037 2.01 1.0049 1.96

Page 103: An apparatus to measure capillary jet thrusts and the

90

Flow Temp. of Man. Temp. of Diameter Rate Man. Man. Reading Fluid (inches) (cc/min) No. CCC) (em) (oc)

.02312708 21.84 1. 28.90 17.21 30.0

.02312708 18.79 1. 29.10 14.75 30.0

.02312708 15-50 1. 30.00 12.40 30.0

.02312708 12.76 2 .. 30.20 154.15 30.0

.02312708 9.52 2. 29.20 118.30 20.0

Calculated Number of Diameter Iterations

.02319894 3

.02324409 2

.02318300 2

.02328145 3

.02321176 2

Average Diameter = .02322385 inches.

Average Ratio c

.9900 2.51

.9914 2.46

.9928 2.41

.9942 2.36

.9956 2.31

.9970 2.26

.9984 2.21

.9998 2.16 1.0012 2.11 1.0026 2.06 1.0040 2.01 1.0054 1.96

Page 104: An apparatus to measure capillary jet thrusts and the

9i

Flow Temp. of Man. Temp. of Diameter Rate Man. Man. Reading Fluid {inches2 {cc/min2 No . {°Cl (em} {oc)

. 03255615 47.38 1. 29.70 19.50 30.0

.03255615 42.81 1. 29.90 17-38 30.0

.03255615 37-48 1. 30.90 15-31 30.0

.03255615 33.25 1. 30.60 13-50 30.0

.03255615 28.36 2 . 28.40 177-45 30.0

. 03255615 25.18 2 . 27.40 158.20 30.0

. 03255615 21.95 2. 27.20 138.75 30.0

.03255615 18.85 2. 26.30 120.60 30.0

.03255615 15.64 2. 25.70 102.20 30.0

.03255615 10.89 2. 25.60 75-00 30.0

Calculated Number of Diameter Iterations

.03243242 3

.03255453 3

.03252473 2

.03262120 3

.03257631 2

.03256797 2

.03257443 2

.03255371 2

.03251307 2

.03247010 2

Average Diameter = .03253884 inches.

Average Ratio c

.9910 2.51

.9923 2.46

.9936 2.41

.9948 2.36

.9961 2.31

.9973 2.26

.9986 2.21

.9998 2.16 1.0011 2.11 1.0023 2.06 1.0036 2.01 1.0049 1.96

Page 105: An apparatus to measure capillary jet thrusts and the

92

Flow Temp. of Man. Temp. of Diameter Rate Man. Man. Reading Fluid {inches} {cc/min} No . {o C) {em} {°C2

. 04630384 46.93 2 . 26.50 83.70 30.0

. 04630384 41.86 2 . 26.10 75-03 30.0

. 04630384 36.85 2 . 25.50 66.60 30.0

. 04630384 32.15 2. 25.90 58.45 30.0

.04630384 26.30 2 . 26.00 50.59 30.0

. 04630384 21.43 2 . 25-50 42.70 30.0

. 04630384 18.19 2 . 25.50 38.05 30.0

. 04630384 14.85 2. 25.10 33.43 30.1

.04630384 11.63 2 . 25.40 29.15 30.0

. 04630384 8.40 2. 25.30 25.30 30.0

Calculated Number of Diameter Iterations

.04599008 3

.04607614 3

.04618531 3

.04642929 3

.04619016 3

.04645264 3

.04647049 2

.04642482 2

.04627933 2

.04597738 2

Average Diameter = . 04624756 inches.

Average Ratio c -.9920 2.51 .9932 2.46 .9943 2.41 .9955 2.36 .9966 2.31 .9978 2.26 .9989 2.21

1.0001 2.16 1.0012 2.11 1.0024 2.06 1.0035 2.01 1.0047 1.96

Page 106: An apparatus to measure capillary jet thrusts and the

93

Flow Temp. of Man. Temp. of Diameter Rate Man. Man. Reading Fluid {inches} (ccLmin} No. (oc} {em} (oc}

.06297296 50.30 2. 26.00 34.95 30.05

.06297296 35.69 2. 26.40 27.92 30.05

.06297296 29.49 2. 26.10 25.34 29.98

.06297296 24.70 2. 26.00 23.32 30.05

.06297296 18.45 2. 26.60 20.60 30.00

.06297296 13.20 2 • 26.30 18.55 30.00

. 06297296 10.07 2 • 25.90 17.27 30.00

. 06297296 6.70 2 . 25-70 15.91 30.05

. 06297296 5.00 2. 25.80 15.20 30.05

Calculated Number of Diameter Iterations

.06286'('82 3

.06320686 3

.06305184 3

.06298433 2

.06314220 3

.06280978 3

.06274526 2

.06261067 2

.06271556 2

Average Diameter = .06290381 inches.

Average Ratio c

.9930 2.51

.9939 2.46 • 9949 2.41 . .9959 2.36 .9968 2.31 .9978 2.26 .9987 2.21 .9997 2.16

1.0007 2.11 1.0016 2.06 1.0026 2.01 1.0035 1.96

Page 107: An apparatus to measure capillary jet thrusts and the

94

Note: 4.0% Polystyrene E in toluene used instead of pure toluene.

Flow Temp. of Man. Temp. of Diameter Rate Man. Man. Reading Fluid (inches) (cc/min) No. (OC) (em) (°C)

.10748413 564.00 1. 27-7 36.55 30.0

.10748413 497.00 1. 28.7 32.60 30.1

.10748413 411.00 1. 27.4 27-35 29.9

.10748413 330.00 1. 28.5 22.20 30.1

.10748413 247.00 1. 27.4 16.95 30.0

.10748413 165.00 2. 27-5 170.40 30.0

.10748413 116.00 2. 27.8 124.70 30.1

.10748413 85.00 2. 29.0 94.45 30.0

.10748413 50.70 2. 27-5 61.53 30.1

.10748413 56.75 2. 27-5 68.45 29.9

.10748413 41.33 2. 27.8 53.45 30.0

.10748413 36.37 2. 27.6 48.65 30.0

.10748413 30-57 2. 28.3 42.95 30.0

.10748413 26.25 2. 28.7 39.05 30.0

.10748413 21.17 2. 27-5 34.15 30.0

.10748413 16.25 2. 27.8 29.35 30.0

.10748413 11.53 2. 29.0 24.70 30.0

.10748413 6.67 2. 27-7 19.87 29.9

Calculated Number of Diameter Iterations

.10423039 4

.10393200 3

.10366417 3

.10367049 2

.10377363 2

.10372214 2

.10338072 2

.10349221 2

.10347936 2

.10293561 2

.10293619 1

.10291640 2

.10299485 2

.10272232 2

.10260762 2

.10258467 2

.10267659 2

.10297701 2

Average Diameter = . 10326088 inches .

Page 108: An apparatus to measure capillary jet thrusts and the

95

Table XI

VISCOMETRY DATA FOR 1.0% PIB IN CYCLOHEXANE

Flow Temp. of · Man. Temp. of Diameter Rate Man. Man. Reading Fluid (inches) (cc/min) No . (OC) (em) (°C)

. 10326088 159.00 1. 28.5 12.35 25.0

.10326088 105.00 2 . 28.5 136.75 25.0

. 10326088 86 .. 50 2 . 28.5 118 .oo 25.1

. 10326088 69.50 2 . 28.5 97-35 25.1

. 10326088 51.15 2 . 28.5 77-15 25.0

. 10326088 33-35 2. 28.5 55-40 24.9

Delta P Velocitl Reynolds Shear Stress 8V/D1 (psi Number (dynes/cm2 ) (sec- ) (ft/sec

2.043 1.609 1121.0 I5-71 1496.0 1.347 1.062 740.2 9.91 987.9 1.145 0.875 609.8 42.44 813.8 0.921 0.703 490.0 34.15 653-9 0.702 0.517 360.6 26.03 481.2 0.465 0.337 235.1 17.24 313-7

Flow Temp. of Man. Temp. of Diameter Rate Man. Man. Reading Fluid (inches) (cc/min) No. ec> (em) coc> .06290381 182.00 1. 28.5 43-55 25.0 .06290381 166 .oo 1 . 28.5 40.10 25.0 . 06290381 141.50 1. 28.5 35.05 25.0 .06290381 118.00 1. 28.5 29.85 25.0 .06290381 76.00 1. 28.5 20.50 25.1 .06290381 50.12 1. 28.5 14.60 25.0 .06290381 32-70 1. 28.5 10.35 25.0 .06290381 15.65 2. 28.5 88.57 25.0

Delta P Ve1ocitl Reynolds Shear Stress 8V/D1 (psi) Number (dynes/cm2 ) (sec- ) (ft/sec

8.013 4.963 2106.5 358.74 7575.0 7-234 4.527 1921.3 324.10 6909.1 6.320 3-859 1637-9 282.95 5889.4 5-360 3 .. 218 1365-7 239.95 4911.3 3.609 2.072 879-5 161.57 3163.2 2.489 1.366 580.1 111.41 2086.0 1.674 0.891 378.4 74.94 1361.0 0.829 0 .. 426 181.1 37-09 651.0

Page 109: An apparatus to measure capillary jet thrusts and the

96

Flow Temp. of Man. Temp. of Diameter Rate Man. Man. Reading Fluid (inches) (cc/min) No . ec) (em) (oC)

. 04624756 38.30 1. 28.1 31.25 25.0

.04624756 26.78 1. 28.1 22.80 25.0

.04624756 14.80 1. 28.1 14.00 25.0

.04624756 114.00 1. 28.2 80.35 25.0

. 04624756 107.00 1. 28.3 75-85 25.0

.04624756 90.00 1. 28.3 66.35 25.0

. 04624756 83.50 1. 28.3 62.15 24.9

. 04624756 48.60 1. 28.4 39.65 24.9

Delta P Velocity Reynolds Shear Stress 8V/D1 (psi) (ft/sec) Number (dynes/cm2 ) (sec- )

5-707 5-707 602.9 181. 8~ 4011.2 4.085 4.085 412.5 13 .4 2804.7 2.387 2.387 232.9 78.56 1550.0

14.929 14.929 1794.6 491.38 11939.4 14.098 14.098 1684.4 464.01 11206.3 12.346 12.346 1416.8 406.34 9425.9 11.561 . 11.561 1314.6 380.51 8745.1 7-316 7-316 765.1 240.78 5089.9

Page 110: An apparatus to measure capillary jet thrusts and the

Table XII

THRUST AND NORMAL STRESS DATA FOR 1.0% PIB IN CYCLOHEXANE

Flow Temp. of Man.* Temp. of Total Zero Diameter Rate Man. Man. Reading Fluid Thrust Thrust (inches) (cc/min) No. (oc) (em) (oC) (gm) (gm)

.03283 260.00 4. 29.30 0.00 25.00 16.090 8.570

.03283 201.00 4. 29.30 0.00 25.00 12.780 8.550

.03283 155·50 4. 29.30 0.00 25.00 11.050 8.540

.03283 104.50 4. 29.40 0.00 25.00 9.640 8.580

. 03283 261.00 4 . 29.40 0.00 25.00 15.480 7. 970

.03283 200.50 . 4. 29.40 0.00 25.00 12.380 7·970

. 03283 156.00 4 . 29.40 0.00 25.00 10.500 8.000

.03283 104.00 4. 29.40 0.00 25.00 9.030 7.980

Pressure Shear Normal Stress Drop Velocitl Stress 8V/D1 Difference Thrust (psi) (ft/sec (dynes/cm2 ) (sec- ) (1b/ft2 ) (gm)

25.3 26.032 2760.56 76123.4 -227.50 3.760 36.8 20.125 2178.52 58849.2 - 27.37 2.115 46.4 15.569 1720.34 45527.6 - 7.48 1.255 59.0 10.463 1193.46 30595.7 26.84 0.530 25.4 26.133 2770.32 76416.1 -201.30 3·755 36.6 20.075 2173.54 58702.8 -109.80 2.205 46.4 15.619 1725.42 45674.0 3.23 1.250 58.9 10.413 1188.21 30449.3 26.54 0.525

*No pressure drops were read during the thrust measurements. Pressure drops were calculated from other viscometry data shown in Tables XI and XII.

1..0 --.J

Page 111: An apparatus to measure capillary jet thrusts and the

Flow Diameter Rate Man. (inches) (cc/min) No.

.06137 430.00 1.

.06137 321.50 1.

.06137 249.50 1.

.06137 165.00 1.

.06137 558.50 1.

.06137 430.00 1.

.06137 320.00 1.

.06137 248.50 1.

.06137 165.00 1.

Pressure Drop (psi)

Velocitl (ft/sec

9.801 12.320 7·321 9.212 5-677 4.1-49 3.804 .728

12.533 16.002 9.644 12.320 7.128 9.169 5.561 4.120 3.745 .728

Table XII (continued)

Temp. of Man. Man. Reading (°C) (em)

28.40 55.20 28 .• ~0 40.65 28.60 31.30 28.90 21.60 29.40 72.05 29.40 54.40 29.40 39.65 29.60 30.70 29.70 21.30

Shear Stress 8V/D

( dynes/cm2 ) (sec- 1)

862.85 19272 ·5 644.39 14409.5 499 ·75 11182.5 334.90 7395.3

11:03.31 25031.8 848.98 19272·5 627 ·52 14342.3 489.57 +1137·7 329.71 7395.3

Temp. of Total Fluid Thrust (oc) (gm)

25.00 18.650 25.00 16.270 25.00 15.070 25.00 14.180 25.00 21.900 25.00 18.420 25.00 16.080 25.00 14.880 25.00 13.920

Normal Stress Difference

(lb/ft2 )

32.99 < 25.64

27 ·41 14. 4 68.01 35-58 21.13

. 23.64 15.62

Zero Thrust

(gm)

13.478 13.440 13.470 13.500 13.280 13.270 13.240 13.260 13.250

Thrust (gm)

2,!580 1.415 0.800 0.340 3.810 2.575 1.420 0.810 0.335

\0 00

Page 112: An apparatus to measure capillary jet thrusts and the

Table XII (continued)

Flow Temp. of Man. Diameter Rate Man. Man. Reading (ill:~he_s) (cc/min) No. (oC) (em) -.1045 246.00 2. 29.70 73·75 .1045 377.00 2. 29.80 118.10 .1045 457.00 2. 29.70 149.50 . 1045 561.00 1 . 29.80 11.20 .1045 248.00 2. 29.70 74.85 .1045 375.00 2. 29.70 119.40 .1045 458.00 2. 29.70 149.65 .1045 559-50 1. 29.70 11.35

Pressure Shear Drop Velocitl Stress 8V/D1 (psi) (ft/sec (dynes/cm2 } (sec- )

0.754 2.43 110.5 2235 1.162 3·73 171.2 3420 1.440 4.52 211.0 4155 1.750 5·55 257.0 5100 0.761 2.45 111.0 2250 1.210 3.71 176.5 3405 1.442 4.53 220.1 4160 1.785 5·54 263.8 5080

Temp. of Total Fluid Thrust ec) (gm)

25.00 27.530 25.00 28.350 25.00 29.000 25.00 29.950 25.00 27.190 25.00 28.100 25.00 28.800 25.00 29.800

Normal Stress Difference

(1b/ft 2 )

3.38 5.64 5·97 8.90 3.09 5.12 6.76 9.60

Zero Thrust

(gm)

26.980 26.990 26.990 27.000 26.670 26.750 26.780 26.830

Thrust (gm)

0.275 0.680 1.005 1.475 0.260 0.675 1.010 1.485

\0 \0

Page 113: An apparatus to measure capillary jet thrusts and the

Table XIII

DIMENSIONS AND TOLERANCES

I .D. (:i.n.) Leng1:h(:i.n.) L/D

. 009639* 6.142 637

.01331* 6.244 469

.02322* 12.087 520

.03254* 24.22 744

.04625* 24.22 524

.06290* 24.22 385

.1033* 48.025 465

.01022** 6.244 611

.01339** 6.248 467

.02326** 6.240 268

.03283** 12.086 368

.04647** 12.264 264

.06137** 12.016 191

.1045** 12.252 116

*Used for viscosity measurements **Used for thrust measurements

OF CAPILLARY TUBES

0 ._R ._(in .J WC]._l_l_(j. n . )

.020 .005

.025 .006

.035 .006

.049 .008

.065 .009

.083 .010

.134 .014

.020 .005

.025 .006

.035 .006

.049 .008

.065 .009

.083 .010

.134 .014

O.D .

+.0005 -.0000

II

+.0005 - rr

II

+.001 - II

+.0005 -.0000

II

+.0005 - II

II

+.001 - II

Tolerance Wall

+.0000 -.0005

" II

+.0005 - II

+.001 - II

+.0000 -.0005

II

II

+.0005 - II

+.001 - II

J-1 0 0

Page 114: An apparatus to measure capillary jet thrusts and the

Table XIV

CALIBRATION OF BOURDON TUBE PRESSURE GAUGES USING DEAD WEIGHT TESTER

Champion Bourdon - Champion Bourdon -

Tube gauge (0-300) psi Tube gauge (0-100) psi

Gauge Tester Gauge Tester Read in~ Force Rea din§ Force (lb/in ) (1b/in2 ) (lb/in ) (lb/in2 )

30.0 30 15.0 15

60.0 60 20.0 20

90.0 90 25.0 25

120.5 120 30.0 30

155.0 155 35.0 35

180.0 180 40.0 40

210.0 210 50.0 50

230.0 230 60.0 60

240.0 240 70.0 70

249.5 250 80.0 80

259.0 260 90.0 90

278.5 280 100.0 100

298.0 300

101

Page 115: An apparatus to measure capillary jet thrusts and the

MATERIALS USED IN EXPERIMENT

Cyclohexane. Purity (wt % cyclohexane) 99.9% minimum;

impurities-non-volatiles~ water~ benzene 0.1% maximum;

maximum boiling range 0.4°C~ including 80.7°C; specific

gravity range 0.780-0.784 at 15.5°C; purchased from

G. S. Robins Co.~ St. Louis~ Missouri.

Toluene. Purity (wt % toluene) 99.5% minimum; impurities­

heptane isomers o.s%·maximum; maximum boiling range l°C~

including 110.6°C; specific gravity between 0.869 and

0.873 at 15.5°C; nitration grade; purchased from G. s. Robins Co.~ St. Louis, Missouri.

102

Polyisobutylene L-80. Enjay MM Vistanex; grade L-80; lot

B40828; code 230; molecular weight approximately 1,000,000,

distribution unknown; produced by a low temperature Friedel­

Craft reaction; catalyst-free with trace amounts of butylated

hydroxytoluene and sodium stearate; color slightly yellow;

donated by Humble Oil and Refining Co., Baton Rouge,

Louisiana.

Page 116: An apparatus to measure capillary jet thrusts and the

NOMENCLATURE

Any consistent units can be used.

c

d

D

(dv/dr)w

f

L

n'

p

Po.,L

p .. l.J

b.P

Q

r

R

Re

T

1-L

v 'J.. .

l.J

'1'--w p

'JT

end correction coefficient

tube diameter used in iteration procedure

tube diameter

shear rate at capillary tube wall

Fanning friction factor

dimensional constant lbm - ft

lbf - sec 2

tube length

8V slope of log~w vs log 1) curve =

hydrostatic pressure

p at tube exit., centerline

deviatoric stress component

pressure drop

flow rate

tube radius at point within tube

tube radius

Reynolds Number

Thrust

local velocity

average velocity

component of total stress tensor iJjJ = 1J2J3

3.1415····

103

Page 117: An apparatus to measure capillary jet thrusts and the

104

~ viscosity

5.. Kronecker delta ~J

Page 118: An apparatus to measure capillary jet thrusts and the

1.

2.

4.

6.

8.

9.

10.

11.

12.

13.

14.

16.

105

BIBLIOGRAPHY

Aris, R., nvectors, Tensors and the Basic Equations of Fluid Mechanics", Prentice-Hall, Inc., Englewood Cliffs, New Jersey (1962)

Bird, R.B., Stewart, W.E., and Lightfoot, E.N., "Transport Phenomena", Johy Wiley and Sons, New York, New York, pp. 67-68, (1960)

Bogue, D.C., Ph.D. Thesis, University of Delaware, Newark, Delaware (1961)

Brodnyan, J.G., Gaskins, F.H. and Philippoff, W., Tran. Soc. Rheol. II 285 (1958)

Dahler, J.S., and Scriven, L.E., Nature, 192, 36 (1961)

Fredricksonfi A.G., T!Principles and Applications of Rheology', Prentice-Hall, Inc., Englewood Cliffs, New Jersey (1964)

Gaskins, F.H., and Philippoff, W., Trans. Soc. Rheol., _3, .181 (1959)

Harris, J., Nature, 190, 993 (1961)

Hershey, H.C., Ph.D. Thesis, University of Missouri at Rolla, Rolla, Missouri (1965)

Houghton, W.T., M. Ch. E. Thesis, University of Delaware, Newark, Delaware (1961)

Lodge, A.S., Nature, 176, 838 (1955)

Merrington, A.C., Nature, 152, 663 (1943)

Metzner, A.B., 11Handbook of Fluid Dynamics n, McGraw-Hi 11 Book Company, New York, New York, p. 7.16, (1961)

Metzner, A.B., Houghton, W.T., Sailor, R.A. and White, J.L., Trans. Soc. Rheo1., 2, 133 (1961)

Middleman, S. and Gavis, J., Physics of Fluids, 4, 355 (1961)

Philippe££, W., J. Appl. Phys., gz, 984 (1956)

17. Philippoff, W., Trans. Soc. Rheol., I, 95 (1957)

18. Rodriguez, J.M., M. Ch. E. Thesis, In Progress, University of Missouri at Rolla, Rolla, Missouri (1966)

Page 119: An apparatus to measure capillary jet thrusts and the

19. Shertzer~ C.R.~ M. Ch. E. Thesis~ University of Delaware~ Newark, Delaware (1964)

20. Shertzer~ C.R.~ and Metzner~ A.B.~ Trans. and Journal of Plastics Institute (Brit)~ ~~ 141 (1963)

106

21. Uebler~ E.A., M. Ch. E. Thesis, University of Delaware~ Newark~ Delaware~ (1964)

22. Van Wazerfi J.R., Lyons~ J.W., Kim, K.Y.~ and Colwell~ R.E.~ 'Viscosity and Flow Measurementn, John Wiley and Sons~ New York, New York, pp. 187-214, (1963)

23. Weissenberg, K.~ Nature, 122, 310 (1947)

24. White, J.L. and Metzner~ A.B., nProgress in Int. Research in Thermodynamic and Transport Props.n A.S.M.E. and Academic Press~ New York; New York, pp. 748-760, (1962)

25. White~ J.L. and Metzner~ A.B., Trans. Soc. Rheol., 1_, 295 (1963)

Page 120: An apparatus to measure capillary jet thrusts and the

ACKNOWLEDGEMENTS

The author wishes to thank Dr. Jacques Zakin for his

patience~ suggestions and assistance throughout the entire

investigation.

107

The author also wishes to thank Harry C. Hershey~ who

helped establish the design of the capillary viscometer and

did a part of its construction~ for his many suggestions

during phases of the project. Mr. Hershey also spent a

great deal of time in helping the author program his data

for the computer.

Thanks are also ·given to Gary K. Patterson who pro­

posed that an analytical balance might be used to measure

the thrusts.

He is grateful for financial assistance provided by

the American Chemical Society Petroleum Research Fund.

Page 121: An apparatus to measure capillary jet thrusts and the

108

VITA

Charles Donald Green~ son of John D. and Mary L. Green~

was born on June 22~ 1942~ in Litchfield, Illinois.

He attended elementary school in Mt. Olive~ Illinois.

He graduated from Mt. Olive Community High School~ Mt. Olive~

Illinois, in May of 1960. He enrolled at the Missouri School

of Mines and Metallurgy~ Rolla, Missouri, in September, 1960,

and received the degree of Bachelor of Science in Mechanical

Engineering in May, 1964.

On August 17, 1963, he married Miss Harriet Jane

Marburger of Mt. Olive, Illinois.

In June, 1964, he entered the Graduate School of the

University of Missouri at Rolla, Rolla, Missouri, to complete

requirements for the degree of Master of Science in Mechan­

ical Engineering. From June, 1964 to February, 1965, he held

the position of American Chemical Society Petroleum Research

Fund Fellow.