29
American Society for Radiation Oncology (ASTRO) EMERGING TECHNOLOGY COMMITTEE (ETC) Report on Continuous Localization Technologies for Radiotherapy Delivery January 7, 2010 Eleanor Harris, MD, Co-chairman Evaluation Subcommittee Robert A. Price Jr., PhD, Co-chairman Evaluation Subcommittee John Bayouth, PhD Task Group Leader Indrin Chetty, PhD Candace Correa, MD David D’Ambrosio, MD Thomas J. Dilling, MD Gregory Franklin, MD Ping Xia, PhD

American Society for Radiation Oncology (ASTRO) EMERGING

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: American Society for Radiation Oncology (ASTRO) EMERGING

American Society for Radiation Oncology (ASTRO)

EMERGING TECHNOLOGY COMMITTEE (ETC)

Report on

Continuous Localization Technologies for Radiotherapy Delivery

January 7, 2010

Eleanor Harris, MD, Co-chairman Evaluation Subcommittee Robert A. Price Jr., PhD, Co-chairman Evaluation Subcommittee

John Bayouth, PhD Task Group Leader

Indrin Chetty, PhD Candace Correa, MD

David D’Ambrosio, MD Thomas J. Dilling, MD Gregory Franklin, MD

Ping Xia, PhD

Page 2: American Society for Radiation Oncology (ASTRO) EMERGING

I. INTENDED USE OF TECHNOLOGY

An active arena for technological advancement in radiation oncology treatment

delivery has focused on the motion inherent in target structures and normal organs in

their proximity. These structures are subject to targeting uncertainty due to motion not

adequately represented on planning computer tomography (CT) scans that may occur

both during (intrafraction) and between (interfraction) daily treatments. To address some

of these localization uncertainties, devices that may be implanted in the organ or volume

of interest and tracked during and between treatments have been developed.

For example, several studies have suggested that the prostate gland is subject to

significant intrafraction motion during radiotherapy(1-7). Unlike respiration motion, this

motion is neither periodic nor predictable, with patterns ranging from continuous target

drift, to transient and persistent excursions, to high frequency motion(8). A recent

technology initially developed to address these issues, the Calypso 4D Localization

System has made it possible to monitor intrafraction prostate motion on a real-time basis.

Additionally, with the advances over the last decade (and more so within the last few

years), in Intensity Modulated Radiation Therapy (IMRT), both intra- and extra-cranial

stereotactic radiosurgery (SRS)/radiotherapy(SRT) and stereotactic body radiotherapy

(SBRT), and Image-Guided Radiation Therapy (IGRT), it has become critical to position

patients in the treatment positions precisely and reproducibly. The use of noninvasive and

nonionizing systems presents an attractive solution. The goal of this report is to provide a

review of emerging nonionizing technologies designed to continuously localize the

tumor, patient, or a surrogate, (> 4 Hz). Specifically, this report considers the Calypso

4D Localization System as well as the RadioCameras and the AlignRT system.

Some of the published studies utilize ‘in-house’ hardware and software designs with

differing approaches to acquire both static and real-time patient positions. These studies

have also confirmed that with these systems positioning errors can be reduced and

verified in many treatment sites such as the pelvis, central nervous system, head/neck and

breast/thorax. Similar to angiography where two images are subtracted from one another

2

Page 3: American Society for Radiation Oncology (ASTRO) EMERGING

to create a new image, the optical techniques include the utilization of difference

imaging (video) in a two-dimensional coordinate frame(9, 10), three-dimensional (3-D)

photogrammetry/optical stereoscopic scanning and laser interferometry evaluating

surface morphologic changes(11-16), and infrared optical 3-D tracking with/without tissue

markers both as an independent alignment system or with x-ray alignment(17-25) (e.g.

Novalis Tx/ExacTrac X-ray). For the purposes of this discussion, only the completely

independent (nonionizing) systems were evaluated. Systems integrated with stereoscopic

X-rays will be evaluated in a subsequent report. Furthermore, systems utilizing optical

tracking primarily as input information for motion management and not target

localization (e.g., Varian Real-time Position Management (RPM) system, the Accuray

Synchrony Respiratory and XSIGHT Lung Tracking Systems) will be addressed in a

subsequent report. Presently, we are aware of only three innovative systems which are

both FDA-approved and commercially available; the Calypso system, RadioCameras and

AlignRT. This report summarizes these three systems and their current results.

II. DESCRIPTION OF THE CURRENTLY MARKETED DEVICES A. Calypso 4-D Localization System

The major components of the Calypso 4D Localization System are: (1) a mobile AC

electromagnetic console; (2) a receiver array; (3) three ceiling-mounted infrared optical

cameras and a hub, all located in the treatment room; (4) wireless transponders

(Beacons); and (5) a tracking station located in the control area (Fig. 1)(26). The system

uses an array of AC magnetic coils to generate a resonant response in implanted

transponders that is subsequently detected using a separate array of receiver coils. The

transponders are approximately eight mm in length and two mm in diameter (see Fig. 2),

and are inserted within the prostate gland under ultrasound guidance in a manner

analogous to a needle biopsy(26). Typically, three beacons are implanted, though the

system can use as few as two implanted beacons. The spatial coordinates of the beacons

are identified on a treatment planning CT, and the offset between the beacons’ centroid

and the intended (treatment planning) isocenter is reported at a frequency of 10 Hz (or

every 0.1 second). The location of the array relative to the linac isocenter is defined

3

Page 4: American Society for Radiation Oncology (ASTRO) EMERGING

through a calibration procedure, and the array itself is tracked in the treatment room using

the three ceiling mounted infrared cameras (see Fig. 1).

The Calypso 4D Localization System is used in the clinic as follows: (a) The patient

is first localized using the skin marks and room lasers,(b) The Calypso system is

subsequently used in localization mode by moving the couch in three dimensions until

offsets (between the beacon-defined centroid and the treatment planning isocenter) read

0.0 ± 0.5 mm. This process establishes the patient’s initial treatment position and usually

takes less than two minutes to perform.(27) (c) During treatment, the Calypso system

continuously monitors and reports the offset (in three dimensions) between the actual and

the desired isocenter locations. Individual facilities may choose to use action criteria

(tolerances) to re-localize the patient or interrupt treatment based on the observed

intrafraction motion. For example, during the course of the initial clinical evaluation,

various facilities chose different criteria. These criteria included: no action taken,

regardless of motion observed or treatment interruption during delivery if the motion

exceeded a specified threshold (e.g., a threshold of three mm for a duration of 30 seconds

continuously(6)).

Although pre-clinical studies using the Calypso System for breast and lung cancer

applications have recently begun, this report will focus on the published literature as it

applies to prostate localization since this represents the only currently FDA-approved

indication.

4

Page 5: American Society for Radiation Oncology (ASTRO) EMERGING

Fig.. 1. Diagram of the components of the Calypso 4D Localization System, from Balter

et al.(26).

Fig.. 2. Picture of a transponder (Beacon), from Willoughby et al.

1. Device Operation

When a patient is scheduled to undergo prostate radiation therapy using the Calypso

4D Localization System, the localization process begins when three transponders are

permanently implanted into the prostate under transrectal ultrasound guidance.

Approximately four to fourteen days later the patient undergoes thin-slice CT simulation.

Target volumes are delineated and a radiation treatment plan is created. Transponder

coordinates on the CT scan relative to the treatment plan isocenter are determined and

entered into the Calypso system.

5

Page 6: American Society for Radiation Oncology (ASTRO) EMERGING

The Calypso system can be operated in either localization (interfraction monitoring)

or tracking (intrafraction monitoring) mode. For localization, the patient is placed on a

treatment table and aligned with skin marks and treatment room lasers. Either supine or

prone patient positioning is possible, but immobilization devices (e.g., a belly board) can

not contain carbon fiber as this interferes with the signal from the transponder. Next, the

Calypso receiver coil array is placed over the patient and its position relative to the

isocenter, as well as the transponder position relative to the array, is detected.

Manufacturer-provided software sends an output of the transponder position relative to

the isocenter (the offset) to the tracking station. Radiation therapists at the control panel

can then align the patient to the machine isocenter within desired guidelines. After

alignment is complete, real-time intrafraction tracking of the transponders can be

performed by operating the system in tracking mode. If the target moves beyond pre-

specified motion thresholds, the radiation therapist can take appropriate corrective action.

2. Installation and Calibration

Installation of the Calypso 4D Localization System is performed primarily by the

Calypso engineers with assistance from a radiation physicist and typically requires

approximately three to seven days. Three infrared cameras are permanently mounted in

the treatment room using customized ceiling bolts and camera mounts and the cameras,

console, and tracking station are connected to the system’s network hub(26).

Commissioning of the system requires calibration and acceptance testing. A

manufacturer-provided optical calibration procedure aligns the optical system to the

treatment machine isocenter. Acceptance testing includes system safety tests as well as

performance measurements. Performance testing involves several measurements with a

localizer phantom embedded with three transponders at known distances to determine the

ability of the system to accurately (and precisely) determine the absolute transponder

positions and offsets. Other acceptance tests are performed to verify the functionality and

accuracy of the optical localization system within the room.

Routine quality assurance (QA) of the Calypso system is performed on a daily and

monthly basis. Daily QA is performed before the system is used for patient treatments. It

entails verification of the positions of the transponders within a vendor provided phantom

6

Page 7: American Society for Radiation Oncology (ASTRO) EMERGING

against their calibrated values. The daily QA test is estimated to take 5-15 minutes.

Monthly QA (typically performed by a radiation physicist) is comprised of a camera

calibration procedure and subsequently a measurement of the known transponder

positions within a phantom. The purpose of the monthly QA is to ensure that the camera

systems are able to localize the transponder array in the room coordinate system

accurately, and that the array is able to detect the transponder positions within the patient

coordinate system accurately. In one study, repeated monthly calibrations showed that

the system stability was within one mm(28). The monthly QA procedure requires

approximately 30-45 minutes. Other tests, such as the positional stability of the

transponders over time(29),are also helpful in evaluating the accuracy of the system, and

could provide an indication, if necessary, to acquire additional imaging(30).

B. RadioCameras

Wagner et al performed an excellent review of the RadioCameras system,

commercially available with the trade name RadioCameras. The system is specifically

designed for intracranial radiosurgery with high-precision patient positioning. The system

includes two, two-dimensional charged couple device (CCD) cameras and a rigid array

containing four infrared, passively reflective markers. The cameras are rigidly mounted

in the ceiling of a treatment room and interfaced with a personal computer. The infrared

markers fixed to the array are also connected with the patient by rigid attachment of the

array to a maxillary bite-block. The CCD cameras intercept the infrared light reflected

from the markers, thus localizing the positions of the markers at a frequency of ~15 Hz.

The camera system must be able to identify at least three of the four markers to determine

the array’s position with six degrees of freedom (translations and rotations). The software

application will then display the vector displacement (with 0.1 mm precision) of the

patient, along with the rotations (yaw, pitch, and roll with 0.1 degree precision).

1. Device Operation

7

Page 8: American Society for Radiation Oncology (ASTRO) EMERGING

Clinically, the relationship between the markers on the array and the treatment isocenter

of the patient is established using the treatment planning CT, where the treatment

isocenter is determined and the coordinates of the infrared marker positions in relation to

the treatment isocenter are fixed. It is generally assumed that the treatment isocenter is

rigidly related to the positions of the markers. Therefore, the desired treatment position

for a specific patient in the treatment room is linked to the desired positions of the

infrared markers in the linac coordinate system. A calibration procedure is required for

the RadioCameras system to transform the marker’s position in the camera coordinate

system to the treatment machine’s coordinate system (with its origin at isocenter). When

a patient with a bite-block is set up for treatment, the RadioCameras system applies a

geometry pattern search algorithm to detect the position of the center of each marker and

computes the necessary translation and rotational shifts to co-register the treatment

planning isocenter with the machine isocenter. The system displays the detected marker

positions and the calculated shifts approximately 15 times per second.

For frameless SRS/SBRT of head and neck (H & N) or central nervous system (CNS)

treatments patient flow begins with an initial simulation where a custom bite-block is

attached to the RadioCameras system’s passive IR array. Developing a relationship with

a prosthodontist, who has experience making obturators and other devices to reconstruct

palate and other surgical defects, can help with customization. The bite-block provided

by Varian Medical Systems can then be modified for a very precise, immobile and patient

specific fit. Generally, the patient needs to have the majority of their upper teeth, or very

tight fitting dentures for the system to work. Edentulous patients can be problematic.

After the bite-block is created, the patient undergoes CT simulation. A thermoplastic

mask is created and a hole is pulled open while the thermoplastic is still hot (if the entire

head is to be immobilized). The custom bite-block is placed into the mouth prior to

simulation. After the plan (FastPlan, Varian Medial Systems) is finished the data is sent

to the RadioCameras system. The array balls are localized manually on the computer

screen. Tolerances within 0.4 mm are usually achievable by experienced personnel; this

again is dependent on voxel resolution achieved during acquisition of the CT dataset.

The patient is next brought to the linear accelerator, where he/she is registered in 3-D

space using the passive infrared system. The entire RadioCameras system consists of a

8

Page 9: American Society for Radiation Oncology (ASTRO) EMERGING

ceiling mounted optical position sensor that senses infrared reflections from the optical

reference array attached to the patient. The positions of these reflections are then input to

a workstation that outputs information regarding the patient’s position. The system

generates coordinate shifts: the couch is adjusted for the x, y and z coordinate offsets

(and the stereotactic head-holder can adjust for tilt and spin). With a secure bite-block

and a compliant patient, the system is robust and reproducible.(20) If the patient moves his

mouth or lips or grinds his/her teeth, the system will show an error and the patient will

need to be repositioned. A noncompliant or difficult patient can add significant time to

setup and treatment. During treatment, the patient is monitored continuously and if the

patient moves, he/she needs to be repositioned. Additionally, for each couch move, the

RadioCameras system position must be re-zeroed.

2. Installation and Calibration

A calibration procedure is necessary to transform the coordinates of the inherent

RadioCameras system to the linear accelerator’s coordinate system. A calibration

apparatus consisting of an array of infrared markers is placed at a defined location

relative to the machine isocenter. The positioning of the calibration jig so that its center is

precisely at the radiation isocenter is critical and challenging. The vendor provides a

gantry-mounted and table-top calibration jig; in-house solutions are required to validate

the jig’s positioning with respect to radiation isocenter if ≤ 0.3 mm positioning is desired.

Once positioned, the passively reflective infrared markers of the calibration apparatus are

optically detected by the camera system establishing the transformation between the

camera and the machine coordinate system whose origin is at isocenter. An end-to-end

test is recommended to verify the accuracy of the entire treatment procedure using the

optical tracking system for treatment positioning. An effective end-to-end test includes

imaging and planning treatment on a phantom, which contains a regularly shaped target

(a sphere) at a known location. If the test target is radio-opaque, analysis of portal images

of the radio-opaque target can be used to define the accuracy of treatment positioning.

Since the RadioCameras system utilizes the planning CT to establish the relationship

between the positions of infrared markers and the patient specific treatment isocenter, a

potential error is introduced, referred to as the “mean registration error”. It is the

9

Page 10: American Society for Radiation Oncology (ASTRO) EMERGING

difference between the CT defined geometry of the markers and the known geometry of

the markers, mostly due to the finite CT voxel sizes. The mean registration error is

approximately 0.3 mm in the phantom test, and less than 0.5 mm in most clinical

situations.

Additional sources of uncertainty can exist. Loss of reflectivity of the markers can

result in the center-of-mass of the marker being inaccurately identified. Markers are

attached to the array by metal posts, which when bent cause inaccurate localization of the

array. The camera system’s accuracy diminishes as the radial distance of the array from

isocenter increases. All of these uncertainties should be characterized for a given system

and routine quality assurance of the array is appropriate. Periodic end-to-end testing,

including applying known transformations to a phantom, is also appropriate to verify

maintenance of sub-millimeter positional accuracy.

During patient setup and treatment, the RadioCamera system also displays overall

vector misalignment (the root mean square of three translational misalignments). For

intracranial stereotactic radiotherapy, 0.3 mm vector misalignment and 0.3 degrees of

rotational misalignment about each axis have been accomplished.(31) During treatment

delivery, if the patient moves more than the threshold (<0.5 mm), the treatment can be

interrupted until the patient’s position is corrected. This procedure can be repeated for

each treatment field.

C. AlignRT

The AlignRT system consists of two imaging pods and analysis software running on a

standard personal computer with a two-channel frame-grabber card. Each pod consists of

two charge-coupled device (CCD) stereo cameras, a CCD texture camera, speckle flash

with lens, a clear light flash without lens and a speckle projector as shown in Fig. 3. The

two camera pods are mounted in the ceiling of the treatment room (or simulation room),

and each covers approximately 120 degrees of axial body surface from the midline to

posterior flank of the body. The stereo cameras acquire topological data of interest of the

patient surface. The projected speckle supplies a unique pattern over the smooth patient

surface, allowing a precise match of two stereo images. Data from both pods are merged

to form an integrated surface model of the patient. The six cameras from the two pods are

10

Page 11: American Society for Radiation Oncology (ASTRO) EMERGING

calibrated relative to the linac coordinate system by a procedure using a special

calibration plate with a printed grid. A reference surface can be obtained either during

simulation or the first day of treatment while the patient position is confirmed with a

radiographic technique (e.g., cone beam CT or portal images). If performed daily, this

approach informs the user of the patient’s position relative to the one radiographically

confirmed. Alternatively, the CT-based surface can be used as the reference surface,

which requires transformation from the CT coordinates to the camera coordinates. This

approach provides patient position information relative to that assumed when designing

the treatment planning study. During treatment, the daily-acquired patient surface is

registered to the reference surface using rigid body transformation, minimizing the

distance between the surfaces, which include the user defined region of interest. The

registration process yields treatment position adjustments in six degrees (translation and

rotation) that maximize the congruence of the two surfaces. The AlignRT system

assumes that the user defined surface is rigidly related geometrically to the treatment

isocenter.

Fig. 3. The camera pod is shown at the top. The imaging system includes two camera

pods directed towards the patient (shown middle bottom). The data from the two cameras

are merged creating surface images (shown lower left bottom).

11

Page 12: American Society for Radiation Oncology (ASTRO) EMERGING

1. Device Operation

The system can be used in two principal modes, a single-frame mode and continuous

mode. In the single-frame surface acquisition mode, two flashes are illuminated, one with

the speckle pattern and one without speckle. The latter flash enables acquisition of

textured images. In the continuous mode, the speckle pattern is projected continuously

while acquiring image data at 6.5 frames per second. This mode allows for real-time

monitoring of surface motion due to breathing, or for respiratory-gated image acquisition.

For accelerated partial breast irradiation (APBI) or conventional breast treatments, the

patient flow begins with a standard CT-based simulation with immobilization on a breast

board. The patient setup is initially determined by tattoos (and radio-opaque bb’s) marked

at the time of CT simulation. This allows standard laser and portal film alignment of the

patient to the isocenter on a daily basis. During the time of the first treatment session on

the linac (or in the simulator room with a second imaging system) a reference surface

model is acquired (or acquired from the CT data set). As mentioned above, the imaging

system includes two camera pods suspended from the ceiling which triangulate a given

position; see Fig. 4. The data from the two camera pods are then merged in a 3-D

calculation process to form a single 3-D surface topographic image. An initial region of

interest is chosen (10-12 cm over the tumor bed in APBI but can be as large as needed) to

compare to daily 3-D images acquired after the patient is set up conventionally. The

system then compares the reference to the daily image and provides couch offset values

to align both images to within millimeter resolution.

The AlignRT images are acquired quickly (~10 sec) and the calculation of

displacement coordinates are rapid. A good immobilization device (especially in regards

to arm positions) may be necessary to obtain reproducible breast surface topography.

Variability can be introduced if regions of the patient’s anatomy are visible by only one

of the two cameras (i.e., areas of the breast and supraclavicular region).

12

Page 13: American Society for Radiation Oncology (ASTRO) EMERGING

Fig. 4. The corresponding two points noted in both space and time from each camera and

merged in a 3-D calculation process creating a 3-D topographic image.

2. Installation and Calibration Aligned by the projected light field from the linac, the calibration plate is positioned

on the treatment couch at five specific couch and gantry positions. At each position and

for each camera, four specific points in the grid plate image are identified by the user to

specify the plate origin relative to the machine isocenter. After completion of this

calibration procedure, the calibration software provided by the vendor establishes a

correlation of the system’s camera coordinate system to the linac coordinates. After

initial system calibration, the system is verified daily by a verification procedure that is

similar to the initial system calibration except only one of the four plate positions is

checked. The time required for full system calibration and daily verification is about 10,

and three minutes, respectively. The calibration verification could fail if the adjustable

threshold at the match-line from two stereo images exceeds one mm (personal

communications). The approximate time for registration to generate new couch

13

Page 14: American Society for Radiation Oncology (ASTRO) EMERGING

coordinates is less than 30 seconds, based on initial clinical experience in alignment of 26

breast treatment setup procedures.

Bert et. al(11) conducted a comprehensive evaluation of the AlignRT system. They

reported that the system stability over a period of 57 hours was better than 0.5 mm.

Compared to known couch shifts of a phantom, the maximum standard deviation was

0.75 mm for the three translational degrees of freedom and less than 0.1° for each

rotation. Using the CT-derived surface as a reference surface model, the tested root-

mean-square (rms) of the distance between the surfaces was 0.65 mm. With the known

motion trajectory, the motion trajectories measured by the system were compared to

another established system (Varian Medical System’s RPM™ system). The standard

deviations of the differences to the known trajectory were 0.15 mm and 0.04 mm for the

AlignRT system and the RPM system, respectively. Because of low frame acquisition

rate of 6.5 Hz, rapid motion trajectories can be under sampled. These reported accuracy

and precision values are based on measurements of a rigid phantom. Caution should be

exercised when the alignment region involves great soft tissue deformation because it

may not be approximated with the rigid body assumption.

III. DESCRIPTION OF PATIENTS POTENTIALLY BENEFITING FROM USE

OF TECHNOLOGY It seems intuitive that the use of “continuous localization technologies” would be

implemented hand-in-hand with the use of advanced radiotherapy planning and delivery

techniques (eg., IMRT, SBRT, SRS, etc.). The ability to conform the radiation dose

distribution to the target(s) of interest while sparing surrounding normal tissues has the

potential to reduce the margin for error with respect to localization uncertainty. The

inability to localize appropriately may result in a geographic miss of the intended target

tissues. The use of continuous localization technologies also has the potential to reduce

normal tissue side effects as well as improve outcome due to the inaccurate utilization of

advanced delivery techniques.

Tumors located in the pelvis, abdomen and thorax are subject to motion during

treatment caused by respiration, inherent bowel mobility and peristalsis, and cardiac

motion. This motion is often accounted for by applying a margin to the target of interest

14

Page 15: American Society for Radiation Oncology (ASTRO) EMERGING

to encompass the target’s spatial variability. However, treatment of this margin results in

the delivery of unwanted dose to normal tissues when the target is not occupying a given

position. The use of real-time tracking techniques may allow for a reduction in this

margin and thus has the potential to reduce the morbidity associated with unnecessary

dose to surrounding normal tissues. There is potential for all patients undergoing

radiotherapy to the aforementioned body sites to benefit from real-time tracking

techniques.

IV.EVALUATION/SUMMARY OF RESULTS OF EXISTING STUDIES

A. Calypso 4D Localization System

There have been several articles published on the use of the Calypso system both in

phantom studies as well as in patients. These clinical studies have focused primarily on

prostate cancer. In phantom measurements, Balter et al.(26) observed sub-millimeter

localization and tracking capabilities of the Calypso system, with values that remained

stable over prolonged periods of time. These results have been updated recently by

Litzenberg et al.(29). Willoughby et al.(30), in reporting on the first human use of the

system, evaluated the localization accuracy of the Calypso system relative to radiographic

localization, and assessed its ability to track prostate motion in real-time. Their findings

indicated significant intrafraction prostate motion (greater than 10 mm) in 2 of 11

patients(30). However, the Calypso system demonstrated comparable (within t mm)

isocenter localization accuracy as compared to X-ray localization procedures(30).

Kupelian et al.(6) reported on Calypso-based localization and continuous real-time

monitoring of the prostate gland on a multi-institutional trial consisting of 41 patients

treated at five institutions. They found differences between skin marks versus Calypso

alignment to be greater than 5 mm in vector length in more than 75 percent of all

fractions(6). They also observed that individual patients exhibited displacements of 5 mm

or more, lasting at least 30 seconds, in 56 percent of all fractions(6). Using the criterion

that 90 percent of patients receive 95 percent of the prescribed dose within the PTV,

15

Page 16: American Society for Radiation Oncology (ASTRO) EMERGING

Litzenberg et al.(32) showed that margins required to accommodate intrafraction motion

were approximately 2 mm in all directions, assuming that Calypso-based localization was

performed for each fraction prior to the start of treatment. In the absence of Calypso-

based localization these margins are approximately 10 mm, indicating that a substantial

reduction in margins is possible when daily alignment is performed using the Calypso

system(32). A recent study on dosimetric consequences of intrafraction prostate motion,

showed that significant reductions in treatment planning margins are possible without

compromising target dose coverage, attributed primarily to the significant reduction in

the setup uncertainty (and hence setup margin) when Calypso is used for daily

localization(33)

The efficacy of Calypso localization among patients receiving androgen ablation

therapy is not well-defined. One small comparative study of 41 patients in which 14

received neoadjuvant and concurrent androgen suppression found that the implanted

electromagnetic markers maintained a stable geometry within the prostate gland over

time, both in patients treated with androgen deprivation and in patients treated with

radiation therapy alone(34).

The impact of the electromagnetic detector array on the quality of radiation beams

and portal images is of potential concern, as the array is placed several centimeters above

the patients during treatment to detect signals from the transponders. Preliminary research

demonstrated that the increase in skin dose attributable to the array was “within

acceptable clinical limits(35)” though this was not quantified. Additionally, researchers

found that attenuation of the beams was less than 0.5 percent for radiation incident

normal to the array; no comment was made regarding oblique or tangential beams(35).

Finally, the researchers stated that “portal image quality due to presence of the array in

the beam path was similar to that of patient support devices, such as nylon-strung tennis

racquet table inserts(35).”

Another technical challenge is that of MRI compatibility with the transponder

beacons. Calypso Medical Technologies tested the transponders according to ASTM

F2503-05, and found them to be MRI safe(36). ASTM F2503-05 [American Society for

Testing and Materials: Standard Practice for Marking Medical Devices and Other Items

for Safety in the Magnetic Resonance Environment] defines "MRI Safe" as "an item that

16

Page 17: American Society for Radiation Oncology (ASTRO) EMERGING

poses no known hazards in all MR environments." The Instructions for Use(36) includes a

precaution for planned magnetic resonance imaging (MRI) of the prostate after the

beacon transponders are implanted. The statement reads as follows: "Beacon

transponders are MR safe; however, when imaged with MRI, a local image artifact will

occur in tissue adjacent to the implanted transponders(36)." This radiographic artifact can

extend up to 2 cm from the transponder(36). An example is shown in Fig. 5, where a local

but pronounced radiographic artifact surrounding the transponders is apparent.

Artifact from transponders

Fig. 5. Fast spin echo (FSE) T2-weighted MRI of patient with Calypso transponders implanted in the prostate. Provided by Mark K Buyyounouski, MD, MS, Fox Chase Cancer Center.

Researchers have also performed introductory studies analyzing possible application

of the Calypso localization system to head and neck cancer patients. In one such study, a

dental prosthesis was cast from a volunteer and some of the teeth were filled with dental

amalgam. The prosthesis was placed under the detection array, adjacent to three

transponders, and the resultant measurements were compared to those taken without the

presence of the dental prosthesis. Despite the presence of the amalgam, the system could

localize the transponders up to 20 cm from the array(37). Placement of the transponders

within a mouthpiece does not increase backscatter dose to overlying oral mucosa(38).

17

Page 18: American Society for Radiation Oncology (ASTRO) EMERGING

B. RadioCameras

An optical (infrared) guided bite-plate system was initially developed at the

University of Florida in the 1990s(18). This system decouples the immobilization and

localization functions and has been shown to provide high accuracy. The system has been

utilized mainly for H & N and CNS malignancies and is shown in Fig. 6 and 7. With

active light emitting diodes coupled to an ultrasound transducer, this infrared localization

system has also been utilized for extra-cranial stereotactic localization.(39)

The initial clinical study of this technology focused on stereotactic radiosurgery for

CNS malignancies at University of Florida(40). Sixty patients (both with benign (e.g.,

meningiomas) and malignant (e.g., PNET tumors) received a total of 1426 treatments

using this frameless SRS system. At 11 months follow-up, three patients were found to

have recurrence of their malignant disease and minimal side effects from the treatment.

All low-grade astrocytoma patients had marked involution of the tumor and, of the 36

percent which had post-treatment edema, all had eventual resolution of the edema. The

system proved to be robust with a misalignment vector error of 0.18mm; the tolerance

limit of 0.3mm and 0.3 degrees was achieved in every case (and accomplished within 15

minutes for a daily patient treatment). This accuracy was determined to rival frame-based

systems. Two follow-up studies have further evaluated this technology(31, 41). The most

recent study followed 64 patients who received frameless stereotactic radiosurgery for

intracranial metastatic disease. Some of the patients were treated upfront and others for

progressive disease after initial resection or whole brain radiotherapy. Tumor and dose

characteristics for these treatments confirmed a median of two metastatic sites with two

isocenters and a median dose of 17.5 Gy. With a median actuarial follow-up of 8.2

months, ultimate local control was 88 percent. Median time to progression and overall

survival was 8.1 months and 8.7 months, respectively. The authors noted that these

18

Page 19: American Society for Radiation Oncology (ASTRO) EMERGING

results were similar to outcomes with frame-based techniques. No patient had a serious

(Grade 3 or higher) complication.

The authors also noted some advantages and disadvantages of the system(41). The

advantages included: no pain or discomfort to patient, ability to perform treatment

planning and treatment delivery on different days, minimal monitoring of staff with the

ability to leave facility during treatment delivery and reduction of resource utilization,

personnel, cost and complexity of the stereotactic procedure. The main disadvantage of

the technique was noted to involve the potential for lengthy treatment times. Because of

the use of conventional linac-based machines, treatment times were about 15

minutes/isocenter. Patients requiring more than four isocenters were found to be

technically challenging and otherwise not good candidates because of patient fatigue

during the lengthy treatment times. Highly irregular volumes and patients with poor

Karnofsky Performance Scores also prevented optimal use of the system.

A final publication utilized this optically-based methodology in 10 patients with

advanced head and neck cancers at the University of Wisconsin(42). These patients were

prospectively enrolled to determine the potential impact of traditional daily setup

variations with laser alignment and immobilization mask markings. The passive fiducial

arrays were mounted to the maxillary bite-tray and acted as the ‘gold-standard’. They

found an average composite mean vector error of 6.97mm +/- 3.63 mm with conventional

methods. Based on these findings, the authors noted that the GTV and PTV were under-

dosed and critical normal structures like the parotid or globe were overdosed utilizing

IMRT with conventional daily patient alignment. They concluded that more rigorous

immobilization techniques are required and that routine daily assessment of patient setup

and tracking are important for delivery of high quality IMRT for head and neck cancer

patients.

19

Page 20: American Society for Radiation Oncology (ASTRO) EMERGING

A

B

19. Fig. 6. The patient has an (partial) immobilization mask with the RadioCamera maxillary-bite plate mounted as shown. A) Coordinates of each marker identified on the CT scan, relative to treatment isocenter, are determined from CT. (re-printed from Ryken et al.(31), with permission). B) In the treatment room a workstation coupled to a ceiling mounted optical position sensor computes and display’s the transformation required to reproduce the setup from CT. (re-printed from Hong TS et al.(42), with permission).

20

Page 21: American Society for Radiation Oncology (ASTRO) EMERGING

Fig. 7. A close-up view of the RadioCameras device showing the passive IR array. The

optical reference array has six fiducials that can be localized in a CT scan. Four of these fiducials have a reflective surface that can be used for optical tracking. The reference array is attached to a biteplate.The custom biteplate is fabricated by placing dental impression material in an acrylic dental tray and allowing the impression material to form on the patient’s maxillary dentition. (re-printed from Ryken, et al. (31) with permission).

C. AlignRT AlignRT, as mentioned above, uses two ceiling mounted 3-D camera units to acquire

3-D surface images (photogrammetry) and is shown in Fig. 3. During each treatment

fraction the current patient position (+/- respiratory gated) is acquired repeatedly at the

same point in space (and respiratory cycle) and compared to a reference image providing

21

Page 22: American Society for Radiation Oncology (ASTRO) EMERGING

new couch position coordinates. Most of the data for this system have been provided

during talks or supplemental abstracts and has primarily been related to the setup of

breast cancer patients (11, 43-55).

One clinical publication from Massachusetts General Hospital (MGH) used the

system to assess its utility in patient setup for accelerated partial breast irradiation

(APBI)(44). The accuracy of the system (in nine patients) was compared to traditional

laser and portal image patient setup. Mean 3-D displacements were 7.3 +/- 4.4 mm and

7.6 +/- 4.2 mm for laser and portal image setup, respectively, as compared to 1.0 +/- 1.2

mm for AlignRT. Breathing motion data at isocenter was 1.9 +/- 1.1 mm. As a

comparison, the system was used to evaluate the surface motion of the abdomen and 5.7

+/- 1.3 mm of displacement was noted. Other sites explored in abstracts noted increased

accuracy in positioning the head for use in stereotactic radiosurgery guidance(47, 51).

In theory the system can be used for patient positioning regardless of treatment site

(i.e., extremity for sarcoma/skin treatments). Good correlation has been noted between

surface and bony anatomy(45). The practical utility of the system may be limited by skin-

to-tumor positional correlation, which was investigated in one abstract for the case of

APBI(48) where the registration of lumpectomy-site clip-based imaging (‘gold standard’)

was compared to the skin alignment assessed by AlignRT. Both agreed within 1 mm

suggesting that the surface of the breast may be a reasonable surrogate for the treatment

volume.

V. IDENTIFICATION, ANALYSIS AND EVALUATION OF CONSEQUENCES

OF NON-USE The inability to localize appropriately when using advanced radiation delivery

techniques may result in a geographic miss of the intended target tissues resulting in

uncertainties with respect to tumor control. Additionally, a geographic miss of the

target(s) generally results in the unintended delivery of high dose to healthy tissues and

has the potential to result in undo morbidity.

22

Page 23: American Society for Radiation Oncology (ASTRO) EMERGING

The use of real-time tracking techniques has the potential to limit morbidity by

decreasing dose to normal structures through the reduction of target margins utilized for

spatial uncertainty. Not employing tracking techniques makes it imperative that these

margins be applied in order to maintain treatment outcomes gained to date.

Previous localization and tracking techniques typically utilize ionizing radiation. A

notable exception is the use of ultrasound although to date this usage is limited to a few

tumor sites. None of the technologies evaluated in this study involve the use of ionizing

radiation.

VI. FUTURE PREDICTION BASED ON TECHNOLOGY DEVELOPMENT Given the growing popularity of dose escalation, hypofractionation, and respiratory

gating and the potential improvement of clinical outcomes from each, it can be postulated

that the use of real-time tracking techniques will also gain favor in the radiation oncology

community. Through further research and clinical trials the Calypso system is being

expanded to use in body sites outside the pelvis, which may lead to its use on a routine

basis for a larger population of patients.

The use of technologies that allow the registration of patient topography with planning

data for use with respiratory gating will in all likelihood flourish. Current gating systems

suffer from the uncertainty of correlation of external markers with internal structure

movement. It may be that the increase in the number of registration points (the body

surface) will decrease these uncertainties.

There are two main areas of clinical outcomes improvement that may be expected

as a result of more accurate real-time localization using these technologies. The first

involves target localization, in which the treatment fields are centered on a per-fraction

basis on the center of mass of the target volume itself, as opposed to stable but unrelated

23

Page 24: American Society for Radiation Oncology (ASTRO) EMERGING

bony or other anatomical landmarks. Use of unrelated landmarks requires the use of

wider margins around the target volume, as described in this report, and these margins

must be particularly large in the case of very mobile targets, including tumors located in

the thorax and abdomen. These same tumors generally have poor outcomes overall, and

this is in part due to the inability to escalate dose to large volumes of the surrounding

normal tissues. Therefore the use of real-time continuous localization techniques may

allow for significant reduction in margins, which will then allow for reduced normal

tissue dose-volumes and subsequently dose escalation to the target volume that may

indeed lead to improved tumor control. These are the next phase of clinical trials which

need to be conducted to verify outcomes following the implementation of continuous

localization technologies.

The second important area of potential improvement in clinical outcomes involves

the reduction of normal tissue dose-volumes, which will decrease toxicity. Radiation

toxicity to normal tissue is directly related to the volume of the normal tissue that

receives any given percentage of the prescribed dose. This outcome is important for all

treatment sites, even in the case of tumors that do not exhibit a lot of inherent motion.

For example, in prostate or breast cancer, in which the tumor motion is generally less

extreme than for tumors in the thorax, and in which dose escalation either is already

feasible (prostate) or not indicated (breast), it is still important to minimize the dose to

critical normal tissues that surround the target volumes in order to reduce acute and

especially long term toxicity. Many patients with prostate and breast cancer will enjoy

normal life spans after treatment, therefore the avoidance of late toxicities to the bowel,

24

Page 25: American Society for Radiation Oncology (ASTRO) EMERGING

bladder, lung and heart will contribute to quality of life and reduce the cost of post-

treatment care. In addition, the feasibility of hypofractionation depends upon the ability

to very accurately localize the target volume with minimal margins, as treatment of large

volumes of the surrounding normal tissues would result in a higher likelihood of

developing late toxicities due to the large dose per fraction used in these regimens.

Hypofractionation therefore is highly dependent upon technologies that allow precise and

real-time target localization, in order to reduce normal tissue dose volumes. While the

demonstration of reduced late toxicity may take many years to demonstrate in clinical

trials, these outcomes should also be examined. In the interim, patients may well benefit

from the use of continuous localization techniques as a component of image guided

radiotherapy, and their use should be considered one method for achieving greater

accuracy and precision.

REFERENCES 1. Aubry JF, Beaulieu L, Girouard LM, et al. Measurements of intrafraction motion

and interfraction and intrafraction rotation of prostate by three-dimensional analysis of daily portal imaging with radiopaque markers. Int. J. Radiat. Oncol. Biol. Phys. 2004;60:30-39.

2. Britton KR, Takai Y, Mitsuya M, et al. Evaluation of inter- and intrafraction organ motion during intensity modulated radiation therapy (IMRT) for localized prostate cancer measured by a newly developed on-board image-guided system. Radiat Med 2005;23:14-24.

3. Dawson LA, Litzenberg DW, Brock KK, et al. A comparison of ventilatory prostate movement in four treatment positions. Int. J. Radiat. Oncol. Biol. Phys. 2000;48:319-323.

4. Ghilezan MJ, Jaffray DA, Siewerdsen JH, et al. Prostate gland motion assessed with cine-magnetic resonance imaging (cine-MRI). Int. J. Radiat. Oncol. Biol. Phys. 2005;62:406-417.

5. Kitamura K, Shirato H, Seppenwoolde Y, et al. Three-dimensional intrafractional movement of prostate measured during real-time tumor-tracking radiotherapy in

25

Page 26: American Society for Radiation Oncology (ASTRO) EMERGING

supine and prone treatment positions. Int. J. Radiat. Oncol. Biol. Phys. 2002;53:1117-1123.

6. Kupelian P, Willoughby T, Mahadevan A, et al. Multi-institutional clinical experience with the Calypso System in localization and continuous, real-time monitoring of the prostate gland during external radiotherapy. Int. J. Radiat. Oncol. Biol. Phys. 2007;67:1088-1098.

7. Mah D, Freedman G, Milestone B, et al. Measurement of intrafractional prostate motion using magnetic resonance imaging. Int. J. Radiat. Oncol. Biol. Phys. 2002;54:568-575.

8. Solberg TD, Chetty IJ, Li S, et al. Real-time tracking for radiotherapy of prostate cancer: implication for dose painting. In: Bissonnette J-P, editor. Proceedings of the 15th International Conference on the use of Computers in Radiation Therapy. Oakville, Ontario, Canada: Novel Digital Publishing; 2007. pp. 393-397.

9. Johnson LS, Milliken BD, Hadley SW, et al. Initial clinical experience with a video-based patient positioning system. Int. J. Radiat. Oncol. Biol. Phys. 1999;45:205-213.

10. Milliken B, Rubin SJ, Hamilton RJ. Performance of a video-image-subtraction-based positioning system. Int. J. Radiat. Oncol. Biol. Phys. 1997;38:855-866.

11. Bert C, Methany KG, Doppke K, et al. A phantom evaluation of a stereo-vision surface imaging system for radiotherapy patient setup. . Med. Phys. 2005;32:2753-2762.

12. Li S, Geng J, Djajaputra D. Clinical results for a 3D-surface-guided SRT and IMRT. Int. J. Radiat. Oncol. Biol. Phys. 2004;60:S621.

13. Li S, Walker E, Liu D. Pilot study of a real-time stereovision-based image-guided radiotherapy for breast cancer. Int. J. Radiat. Oncol. Biol. Phys. 2006;66:S228.

14. Mackey RI, Graham PA, Logue JP, et al. Patient positioning using detailed three-dimensional surface data for patients undergoing conformal radiation therapy for carcinoma of the prostate: A feasibility study. Int. J. Radiat. Oncol. Biol. Phys. 2001;49:225-230.

15. Moore C, Lilley F, Sauret V. Opto-electronic sensinsing of body surface topology changes during radiotherapy for rectal cancer. Int. J. Radiat. Oncol. Biol. Phys. 2003;56:248-258.

16. Sohn JW, Kim SS, Monroe JI, et al. Breast image-guided radiation therapy using an optical laser scanner. Int. J. Radiat. Oncol. Biol. Phys. 2005;63:S525.

17. Baroni G, Garibaldi C, Riboldi M, et al. 3D optoelectronic analysis of interfractional patient setup variability in frameless extracranial stereotactic radiotherapy. Int. J. Radiat. Oncol. Biol. Phys. 2006;64:635-642.

18. Bova FJ, Buatti JM, Friedman WA, et al. The University of Florida frameless high-precision stereotactic radiotherapy system. Int. J. Radiat. Oncol. Biol. Phys. 1997;38:875-882.

19. James S, Lyatskaya Y, Soto R. iGPS(Infrared guided patient setup) dramatically improves port films for patient setup of lung cancer patients. Int. J. Radiat. Oncol. Biol. Phys. 2006;66:S497.

20. Meeks SL, Bova FJ, Wagner TH, et al. Image localization for frameless stereotactic radiotherapy. Int. J. Radiat. Oncol. Biol. Phys. 2000;46:1291-1299.

26

Page 27: American Society for Radiation Oncology (ASTRO) EMERGING

21. Sie F, Bootsma D, Moseley D, et al. An efficiency study of an optical approach for routine patient positioning and monitoring. Int. J. Radiat. Oncol. Biol. Phys. 2006;66:S636.

22. Soete G, Van de Steene J, D. V. Int. J. Radiat. Oncol. Biol. Phys. 2002;52:694-698.

23. Spadea M, Riboldi M, Tagaste B. Constrained anatomical surface registration for patient positioning in breast cancer radiotherapy. Int. J. Radiat. Oncol. Biol. Phys. 2004;60:S392.

24. Willoughby T, Meeks S, Wagner T. Gated radiotherapy for lung tumors using infrared detectors, stereoscopic x-rays, and implanted markers. Int. J. Radiat. Oncol. Biol. Phys. 2005;63:S413.

25. Willoughby TR, Forbes AR, Buchholz D, et al. Evaluation of an infrared camera and X-ray system using implanted fiducials in patients with lung tumors for gated radiation therapy. Int. J. Radiat. Oncol. Biol. Phys. 2006;66:568-575.

26. Balter JM, Wright JN, Newell LJ, et al. Accuracy of a wireless localization system for radiotherapy. Int. J. Radiat. Oncol. Biol. Phys. 2005;61:933-937.

27. Beyer D, Liu D, Flores N, et al. Efficiency of a non-ionzing target localization system for radiation therapy. Int. J. Radiat. Oncol. Biol. Phys. 2006;66:S372.

28. Cao D, Ye J, Afghan M, et al. Clinical implementation and initial experience with Calypso 4D Tracking System for prostate cancer treatment Med Phys (abstract) 2007;34:2376.

29. Litzenberg DW, Willoughby JM, Balter HM, et al. Positional stability of electromagnetic transponders used for prostate localization and continuous, real-time tracking. Int. J. Radiat. Oncol. Biol. Phys. 2007;68:1199-1206.

30. Willoughby TR, Kupelian PA, Pouliot J, et al. Target localization and real-time tracking using the Calypso 4D localization system in patients with localized prostate cancer. Int. J. Radiat. Oncol. Biol. Phys. 2006;65:528-534.

31. Ryken TC, Meeks SL, Pennington EC, et al. Initial clinical experience with frameless stereotactic radiosurgery: analysis of accuracy and feasibility. Int. J. Radiat. Oncol. Biol. Phys. 2001;51:1152-1158.

32. Litzenberg DW, Balter JM, Hadley SW, et al. Influence of intrafraction motion on margins for prostate radiotherapy. Int. J. Radiat. Oncol. Biol. Phys. 2006;65:548-553.

33. Li HS, Chetty IJ, Enke CA. Dosimetric consequences of intrafraction prostate motion. Int. J. Radiat. Oncol. Biol. Phys. 2008;71:801.

34. Weinstein G, Jani SK, Kupelian P, et al. Stability of intraprostatic electromagnetic transponders in patients receiving radiation therapy, with and without neoadjuvant and/or concurrent androgen suppression therapy. Int. J. Radiat. Oncol. Biol. Phys. 2006;66 (Suppl. 1):S358-359.

35. Pouliot J, Werner B, Riley J, et al. Demonstration of minimal impact to radiation beam and portal image quality due to the presence of an electromagnetic array. Med. Phys. 2003;30:1473.

36. Instruction Manual: Beacon Care Package for Prostate, Calypso Medical Technologies, Inc. LBL0002-001.

27

Page 28: American Society for Radiation Oncology (ASTRO) EMERGING

37. Mate T, Zeller T, Douglas R, et al. Feasibility of tracking wireless AC electromagnetic transponders in head and neck cancer environment Med Phys (abstract) 2005;32:2112.

38. Ye J, Werner B, Mate T, et al. Assessment of dental amalgam backscatter with a Beacon Transponder Embedded Mouthpiece for real-time tracking during head and neck IMRT Med Phys (abstract) 2006;33:2073.

39. Meeks SL, Buatti JM, Bouchet LG. Ultrasound-guided extracranial radiosurgery: Technique and application. Int. J. Radiat. Oncol. Biol. Phys. 2003;55:1092-1101.

40. Buatti JM, Bova FJ, Friedman WA, et al. Preliminary experience with frameless stereotactic radiotherapy. Int. J. Radiat. Oncol. Biol. Phys. 1998;42:591-599.

41. Kamath R, Ryken TC, Meeks SL. Initial clinical experience with frameless radiosurgery for patients with intracranial metastases. Int. J. Radiat. Oncol. Biol. Phys. 2005;61:1467-1472.

42. Hong TS, Tome WA, Chappell RJ. The impact of daily setup variations on head-and-neck intensity-modulated radiation therapy. Int. J. Radiat. Oncol. Biol. Phys. 2005;61:779-788.

43. Bert C, Metheany KG, Doppke K, et al. Set-Up Uncertainty in Accelerated Partial Breast Irradiation Using 3D-Conformal External Beam Radiotherapy. Presented at the Annual Meeting of the American Society of Therapeutic Radiology and Oncology. Atlanta, Georgia; October 2004.

44. Bert C, Metheany KG, Doppke KP, et al. Clinical experience with a 3D surface patient setup system for alignment of partial-breast irradiation patients. Int. J. Radiat. Oncol. Biol. Phys. 2006;64:1265-1274.

45. Bidmead M, Corsini L, Lindgren-Turner J. Investigating the correlation between surface and bony anatomy using 3D surface and portal imaging. Radiation & Oncology 2004;73:S233.

46. Chen GTY, Riboldi M, Gierga DP, et al. Clinical implementation of IGRT. Radiother. Oncol. 2005;76:S10.

47. Drzymala R, Wood R. Feasibility of tracking head position under an obscuring immobilization mask using a bite block and a 3-D surface imaging system. Med. Phys. 2006;33:1992.

48. Gierga DP, Turcotte JC, Riboldi M, et al. Comparison of target registration errors for multiple modalities in image-guided partial breast irradiation. Int. J. Radiat. Oncol. Biol. Phys. 2006;66:S103-104.

49. Gierga DP, Turcotte JC, Sharp G, et al. Target registration error with three-dimensional surface imaging in setup of image-guided partial breast irradiation. Int. J. Radiat. Oncol. Biol. Phys. 2005;63:S536.

50. Johnson U, Deehan C, Landau D. Real time 3D surface imaging for the analysis of respiratory motion during radiotherapy. Int. J. Radiat. Oncol. Biol. Phys. 2004;60:S603.

51. Lindgren-Turner J, Corsini L, Keane R, et al. Position Verification for Intercranial Stereotactic Radiotherapy Using 3D Surface Imaging. UK Radiation Oncology Conference; April 11-13, 2005.

52. Schöffel PJ, Harms W, Karger CP. Evaluation of repositioning accuracy of patiens with breast cancer using a 3D surface imaging system. Radiother. Oncol. 2006;81:S195.

28

Page 29: American Society for Radiation Oncology (ASTRO) EMERGING

53. Schöffel PJ, Harms W, Sroka-Perez G, et al. Accuracy of a commercial optical 3D surface imaging system for realignment of patients for radiotherapy of the thorax. Phys. Med. Biol. 2007;52:3949-3963.

54. Smith N, Meir I, Hale G, et al. Real-time 3D surface imaging for patient positioning in radiotherapy. Int. J. Radiat. Oncol. Biol. Phys. 2003;57:S187.

55. Tarte S, McClelland J, Hughes S, et al. A non-contact method for the acquisition of breathing signals that enable distinction between abdominal and thoracic breathing. Radiother. Oncol. 2006;81:S209.

29