MIZUHOSI 2009 NW0510 Rev. D ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL Model 6800 This manual is written in five languages in the following order: (en) English (es) Spanish (fr) French (de) German (it) Italian MIZUHO OSI 30031 AHERN AVENUE UNION CITY, CA 94587 Bus: 510-429-1500 Toll Free: 800-777-4674 Fax: 510-429-8500 WWW.MIZUHOSI.COM ·NEWHIPNEWS.COM · NEWSPINENEWS.COM

ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

  • Upload
    lekhanh

  • View
    233

  • Download
    1

Embed Size (px)

Citation preview

Page 1: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D

ALLEGRO™ MOBILE IMAGING TABLE

OWNER’S MANUAL

Model 6800

This manual is written in five

languages in the following order:

• (en) English

• (es) Spanish

• (fr) French

• (de) German

• (it) Italian

MIZUHO OSI

30031 AHERN AVENUE • UNION CITY, CA 94587

Bus: 510-429-1500 • Toll Free: 800-777-4674 Fax: • 510-429-8500

WWW.MIZUHOSI.COM ·• NEWHIPNEWS.COM · •NEWSPINENEWS.COM

Page 2: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 2

IMPORTANT NOTICES

CAUTION: To ensure safe operation of the equipment, please READ THESE INSTRUCTIONS COMPLETELY and keep this manual readily available for future reference.

Carefully observe and comply with all warnings, cautions and instructions placed on the equipment or described in this manual.

In this manual, the WARNING symbol is intended to alert the user to the presence of important operation, maintenance, or safety instructions.

PROTECTION AGAINST ELECTRICAL SHOCK HAZARD:

This symbol indicates this equipment is an applied part, TYPE B in accordance with IEC 60601-1 and is generally suitable for applications involving external or internal contact with the patient, excluding the heart. The patient circuit is connected to protective earth and this equipment should be connected only to outlets with a protective earth ground.

This symbol indicates the external stud that is at the same potential as the equipment chassis. If the AC power cable is not connected to an outlet with a protective earth ground, this external stud should be connected to a protective earth ground.

WARNING: Proper preoperative and intraoperative procedures must be followed to prevent venous stasis and pooling, pressure sore development, neuropathy, improper electro-surgical tissue grounding, hypotension and hypothermia. NOTE: The application techniques outlined in these instructions are the manufacturer’s suggested techniques. The final disposition of each patient’s care as related to the use of this equipment rests with the attending physician.

THIS DEVICE IS TO BE USED BY TRAINED PERSONNEL ONLY.

Page 3: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 3

TABLE OF CONTENTS

1.0 INTRODUCTION ................................................................................................................................. 5 1.1 General Description....................................................................................................................... 5 1.2 Specification .................................................................................................................................. 5 1.2 Specification .................................................................................................................................. 6

2.0 GLOSSARY OF TERMS..................................................................................................................... 7 2.1 Basic Assumptions Of Patient Orientation On The Tabletop ........................................................ 7

3.0 COMPONENT IDENTIFICATION ....................................................................................................... 9 3.1 Table Orientation........................................................................................................................... 9 3.2 Model Number And Serial Number ............................................................................................... 9

4.0 BASIC OPERATION ......................................................................................................................... 10 4.1 Control Operation ........................................................................................................................ 10 4.2 Floor Lock.................................................................................................................................... 10 4.3 Hand Pendant ............................................................................................................................. 11 4.4 Joystick Control – Interactive Proportional Speed Control™ (IPSC) .......................................... 14

5.0 INSPECTION..................................................................................................................................... 16 5.1 Acceptance And Transfer............................................................................................................ 16 5.2 Inspection And Transfer ............................................................................................................. 16

6.0 FUNCTION CHECK .......................................................................................................................... 17 6.1 Function Check ........................................................................................................................... 17

7.0 THE ELECTRICAL SYSTEM............................................................................................................ 20 7.1 Description .................................................................................................................................. 20 7.2 On/Off Circuit Breaker Switch ..................................................................................................... 20 7.3 Component Circuit Breakers ....................................................................................................... 20 7.4 Power Cord ................................................................................................................................. 20 7.5 Floor Lock System....................................................................................................................... 20 7.6 Battery Charging ......................................................................................................................... 21 7.7 Power Supply - AC & DC Operation ........................................................................................... 21 7.8 Battery System ........................................................................................................................... 22

8.0 CLEANING........................................................................................................................................ 24 8.1 Cleaning And Disinfecting ........................................................................................................... 24 1. Table exterior ............................................................................................................................... 24 2. Casters ......................................................................................................................................... 24 4. Table tops..................................................................................................................................... 25 8.2 Lubrication................................................................................................................................... 25 1. Table exterior ............................................................................................................................... 25 2. Table interior ................................................................................................................................ 25 8.3 Table Storage.............................................................................................................................. 25

9.0 TROUBLESHOOTING...................................................................................................................... 26 9.1 Electrical System......................................................................................................................... 26 9.2 Table Functions........................................................................................................................... 27

10.0 REMOVAL AND REPLACEMENT OF COMPONENTS .................................................................. 33 10.1 Floor Lock Foot Actuator ........................................................................................................... 33 10.2 Floor Lock Pad .......................................................................................................................... 34 10.3 Power Supply Tray .................................................................................................................... 35 10.4 Controller Circuit........................................................................................................................ 36 10.5 10-Amp Motor Control PCB ...................................................................................................... 37 10.6 Stepper Controller Tray ............................................................................................................. 38 10.7 Tilt Sensor Actuator................................................................................................................... 40 Left Tilt Cover Removal (The Bell Shaped Housing): ....................................................................... 41

11.0 TECHNICAL DRAWINGS AND PARTS LISTS ............................................................................... 44 11.1 Electrical Interconnect Diagram, 120 VAC................................................................................ 44 11.2 Electrical Interconnect Diagram, 230 VAC................................................................................ 45 11.3 Top Level Table Assembly Exploded View............................................................................... 46 11.4 Table Sub Assembly Parts List ................................................................................................. 47 11.5 Table Assembly Exploded View................................................................................................ 48 11.6 Base Cover Assembly Exploded View ...................................................................................... 49 11.7 Base Assembly Exploded View................................................................................................. 50 11.8 X-Ring Assembly Exploded View.............................................................................................. 51

Page 4: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 4

11.9 Y-Ring Assembly Exploded View.............................................................................................. 52 11.10 Y-Ring Assembly Bottom View ............................................................................................... 52 11.11 X - Y Coupler Assembly Exploded View ................................................................................. 53 11.12 X - Y Coupler Assembly Bottom View..................................................................................... 54 11.13 Replacement/Spare Parts List ................................................................................................ 55

12.0 Mizuho OSI CUSTOMER RESOURCE GROUP ............................................................................. 56 12.1 Contact For Parts And Service.................................................................................................. 56 12.2 Instant Support Value Package................................................................................................. 56 12.3 To Order Replacement Parts (RP)............................................................................................ 56 12.4 Returns To Mizuho OSI (RGA) ................................................................................................. 56 12.5 To Send A Part For Repair (RA) ............................................................................................... 57 12.6 EC Authorized Representative.................................................................................................. 57

13.0 Mizuho OSI ALLEGRO™ 6800 OPTIONAL ACCESSORIES ........................................................ 58

Page 5: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 5

1.0 INTRODUCTION 1.1 General Description

The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or surgical procedures. It is a stand-alone, single-pedestal AC power or battery-operated table designed to position a patient in a prone, lateral or supine position during radiographic or surgical procedures. The table has a 2” (5 cm) MIZUHO OSI® Tempur-Pedic® pad. The TEMPUR® used in the manufacture of MIZUHO OSI® pads has viscoelastic properties and is a temperature sensitive material, becoming softer where the patient’s body is making the most contact with the surface and remaining firm in areas where less body contact is being made. Pressure is distributed evenly over the entire surface area. The pad is radiolucent, MR safe and latex free. The benefits of using MIZUHO OSI® Tempur-Pedic® are improved pressure management, reduced shear forces and enhanced patient comfort. The table has four primary electrically powered functions; up/down, left/right lateral roll (tilt), trendelenburg/reverse trendelenburg and X-Y translation. These motions are performed by electromechanical actuators and electric stepper motors. The table is controlled by means of a low voltage hand pendant which has the functions clearly labeled. In addition to the hand pendant, a joystick controls the X-Y translations of the table top. Positive floor locks are employed to stabilize the table in the operating location.

Figure 1: Allegro™ 6800 Mobile Imaging

Pedestal

end

Cantilever

end

Page 6: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 6

1.2 Specification

• Maximum patient weight 450 lbs (205 kg)

• Height range of table top 30 – 40 inches (76 - 102 cm)

• Trendelenburg range 15°

• Reverse trendelenburg 15°

range

• Powered lateral roll ±20°

• Longitudinal travel 21 inches (53 cm) at speed of 2 to 4 inches (5 cm – 10 cm) per second

• Lateral travel 10 inches (25 cm) at speed of 2 inches (5 cm) per second

• Table top length 86 inches (218 cm)

• Width 21 inches (53 cm)

• Duty cycle 10%

• Battery life 5 hours at 10% duty cycle

• Maximum deflection <1 inches (2.5 cm) at cantilever end

• Radiolucent equivalency <1 mm of aluminum

• Class B medical device

• Power input (configurable) o 120 VAC at 60 Hz o 230 VAC at 50 Hz o internal battery power

• IPX-0 rating

Page 7: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 7

2.0 GLOSSARY OF TERMS

NOTE: The following terms are used throughout this owners manual. A thorough understanding of these terms is necessary to properly operate this table. 2.1 Basic Assumptions Of Patient Orientation On The Tabletop Our glossary of terms and technical procedures assume the patient is oriented with their head at the front of table and their feet at the rear of table as described below. The hand control functions are oriented for this position only. PEDESTAL END, or rear of table, indicates the power cord end of the table, where the control panel is located. CANTILEVER END, or front of table indicates the opposite end of the rear of the table or, the column pedestal. PEDESTAL, refers to the main column structure that supports the tabletop. Left side, of the table refers to the left side of the table as you stand by the pedestal end and look down the table. This also corresponds to the hand pendent button labeled LEFT LATERAL ROLL. Right side, of the table refers to the right side of the table as you stand by the pedestal end and look down the table. This also corresponds to the hand pendant button labeled RIGHT LATERAL ROLL. Raise the table, refers to raising the height of the table. This also corresponds to the hand pendant button labeled HEIGHT UP. Lower the table, refers to lowering the height of the table down toward the floor. This also corresponds to the hand pendant button labeled HEIGHT DOWN. Reverse trendelenburg, refers to moving the cantilever end of the table up. The corresponding hand pendant button is labeled REV. TREN. Trendelenburg, refers to moving the cantilever end of the table down. The corresponding hand pendant button is labeled TREN. Return to level, or level the table, refers to returning the tabletop to a state level regardless of height. The corresponding hand pendant button is labeled return to level. Holding the RETURN TO LEVEL button longer than approximately three seconds will lower the tabletop to its lowest position. Top home, refers to returning the table top to its home position after longitudinal or lateral travel. The corresponding hand pendant button is labeled TOP HOME. The tabletop in the home position is centered over the pedestal pads in its side to side travel and to its extent of travel toward the pedestal end in its longitudinal travel Unlock the floor locks, raise the floor locks, turn off the floor locks, refers to allowing the table to freely roll on its wheels once the floor lock feet next to the wheels, raise off the

Page 8: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 8

floor. The floor locks, when down, prevent the table from moving on its wheels. The corresponding hand pendant buttons labeled FLOOR LOCK. Control panel, is a panel of buttons and lights located on the pedestal end of the table. Floor locks and tabletop home can be controlled by depressing the corresponding button labeled UP TO UNLOCK, DOWN TO LOCK or TABLE TOP HOME. Hospital grade outlet, refers to a 120VAC outlet properly wired with earth ground. Generally this is: neutral to common = 120VAC; neutral to ground = 120VAC; common to ground less than zero point five (0.5)VAC, and will be marked with a green dot.

Page 9: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 9

3.0 COMPONENT IDENTIFICATION

3.1 Table Orientation

The major components of the Allegro™ 6800 and control panel are shown in figure 2. The patient’s head can be located at either end of the table, depending on the procedure. Therefore, the table is described as having a pedestal end and a cantilever end. The on/off switch, floor lock buttons and the power cord socket are located at the pedestal end of the table.

3.2 Model Number And Serial Number

The model number and serial number are printed on the identification label on the pedestal end of the base.

Figure 4: On/ Off switch, serial number/model number/identification label

Component

circuit breakers

Power cord

socket Model No./

Serial No. label

Hand pendant

can be hung here

Figure 3: Component identification of the control panel

Figure 2: Component identification of the pedestal end interface

Floor lock

indicator lights

Hand pendant

can be hung here

On/off

switch

Page 10: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 10

4.0 BASIC OPERATION

4.1 Control Operation

For use with line power: Plug the power cord into a properly grounded receptacle. Refer to the label at the pedestal end of the base for input voltage requirements (see figure 4). Turn on the power switch and observe the green light in the switch indicating that power is applied to the table. This rocker switch is a combination on/off power switch, circuit breaker and pilot light.

WARNING: Ensure that the battery has been properly charged as outlined in section 7.0. Turn on the power switch. The battery status LED on the hand pendant should be green, indicating a charged battery. The expected working life of a fully charged battery is approximately 5 hours at 10% duty cycle. While in storage, it is desirable that the table be plugged in and the power switch turned on so that the batteries remain charged. If this is not possible, the batteries must be charged for at least three hours a week. This can be accomplished when using the table under line power.

This symbol indicates the external stud that is at the same potential as the equipment chassis. If the AC power cable is not connected to an outlet with a protective earth ground, this external stud should be connected to a protective earth ground. 4.2 Floor Lock

To operate the floor locks via the hand pendant, press and hold the FLOOR LOCK button until the floor lock is completely deployed. The green indicator light marked LOCKED WHEN LIT on the control panel and the FLOOR LOCK ON LED located on the hand pendant will be illuminated when it is locked. To stow the floor locks via the hand pendant, press and hold the FLOOR UNLOCK button until the floor lock is completely retracted. The amber indicator light marked STOWED WHEN LIT on the control panel and the FLOOR LOCK, STOWED LED located on the hand pendant will be illuminated when they are stowed. The floor locks can also be operated by the floor lock buttons on the sheet metal housing of the pedestal as follows. Press the button labeled DOWN TO LOCK, and hold it until the floor lock is completely deployed. The green indicator light marked LOCKED WHEN LIT on the control panel (see figure 3) and the FLOOR LOCK ON LED located on the hand pendant (see figure 6) will be illuminated when it is locked. To release the floor lock, press the button labeled UP TO UNLOCK and hold it until the floor lock is completely retracted. The amber indicator light marked STOWED WHEN LIT on the control panel and the FLOOR LOCK STOWED LED on the hand pendant will be illuminated when it is stowed.

Page 11: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 11

NOTE: The table will not function until the floor locks are completely deployed. Floor lock override button:

If floor lock systems fails during table use, press and hold the FLOOR LOCK OVERRIDE button while simultaneously operating the desired hand pendant function. When button is released, no functions will be operable and this button must be pressed again if subsequent table movement is required. NOTE: This OVERRIDE button only powers the height, trendelenburg and lateral roll functions via the hand pendant and is intended for emergency situations only.

To restore floor lock system function:

• Verify power cord is plugged into a live power outlet supplying proper voltage and the power switch is ON and illuminated

• Verify all four floor locks are completely engaged and control panel, green LOCKED WHEN LIT light illuminates and the FLOOR LOCK ON LED on the hand pendant is illuminated. The table should not move on its wheels when the floor locks are engaged.

To manually disengage floor lock for table relocation: a. Use floor lock crank handle (see Replacement / Spare Parts List) stowed on sheet

metal housing behind control panel for each floor lock. b. Remove floor lock access hole-plug and insert crank handle to engage floor lock

shaft.

• For floor lock nearest grounding stud/power cord, turn crank handle clockwise to raise floor lock.

• For all other floor locks, turn crank handle counter-clockwise to raise floor lock.

4.3 Hand Pendant

WARNING: A hand pendant can be plugged in at the pedestal end (back) or the cantilever end (front) of the base, or both locations with two hand pendants. If a hand pendant is removed or added while the table is operational, the table must be shut down for 30 seconds. When turned on, the floor locks must be reengaged. It is best only to plug or unplug hand pendants when power is turned off.

The table is controlled by means of a low voltage hand pendant which has the functions clearly labeled (see figure 6). The hand pendant should be plugged into the receptacle labeled “hand pendant” on either the back or the front of pedestal. The hand pendant is intended to be hung on the side rail independently, placed either with the swivel mount or without. The swivel mount allows the hand pendant to be mounted horizontally on the table. The hand pendant clip slides in the top of the swivel mount and the swivel mount slides on the side rail or auxiliary clamp.

Page 12: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 12

Height: To adjust the height of the table press and hold the appropriate HEIGHT UP or HEIGHT DOWN button until the desired height is achieved. Lateral roll: To adjust the lateral roll press and hold the appropriate LATERAL ROLL button. Note that when lateral roll is within +/- 2° of level the green level indicator will illuminate. Trendelenburg / reverse trendelenburg: To adjust the trendelenburg position of the table, press and hold the appropriate TREN or REV.TREN button. Note that when trendelenburg tilt is within +/- 2° of level the green level indicator will illuminate. Table top translation: The hand pendant is equipped with TOP FORWARD, TOP AFT, TOP LEFT and TOP RIGHT translation button. When one of the translation buttons is depressed the table top will translate in the given direction at constant speed. NOTE: The table can also be translated at variable speeds using the joystick (see section 4.4).

WARNING: Battery status: The hand pendant is equipped with a two color battery status LED. If the indicator is green, the battery is charged and the table is ready for use. If the indicator is red, the battery must be charged for a minimum of 3 hours (see section 7.0 for complete Battery Charging instructions). NOTE: The table can be used on line power if the battery is not fully charged.

Page 13: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 13

The hand pendant is equipped with a TOP HOME button and a RETURN TO LEVEL button. The TOP HOME button is used to return the table top to it’s X-Y home position. The RETURN TO LEVEL button returns all functions to a preset level position. The top home and the return to level buttons must be depressed and held continuously until the function is complete.

Return to level: Press and hold the RETURN TO LEVEL button, this will 1) level lateral roll, 2) level the trendelenburg and 3) bring the table to its lowest height position. A delay of 3 seconds is programmed after the trendelenburg function before the table returns to its lowest height position. The button must be depressed and held continuously for the function to complete.

Figure 5: Hand pendant in the swivel mount

Figure 6: Hand pendant function buttons

FLOOR LOCK

TOP HOME

FLOOR UNLOCK

RETURN TO LEVEL

TOP LEFT TOP RIGHT

TOP FORWARD TOP AFT

LEFT LATERAL ROLL RIGHT LATERAL ROLL

REVERSE TRENDELENBURG TRENDELENBURG

HEIGHT UP HEIGHT DOWN

FLOOR LOCK - ON FLOOR LOCK - STOWED

BATTERY STATUS

Green = OK

Red = CHARGE

TRENDELENBURG - LEVEL

LATERAL ROLL LEVEL

Page 14: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 14

The controller considers the table to be level if it is within +/- 2° of 0° level. If the RETURN TO LEVEL button is pressed while the table is within this range, it will not move. If it is required to adjust the table to a position closer to 0°, simply press the appropriate function button on the hand pendant until the desired position is achieved.

Top home: The X-Y home position of the table top can be achieved by pressing the TABLE TOP HOME button located on the control panel of the pedestal or the TOP HOME button on the hand pendant. This position is defined as the table top being centered in its side to side travel and to its extent of travel toward the pedestal end in its longitudinal travel. 4.4 Joystick Control – Interactive Proportional Speed Control™ (IPSC) The lateral and longitudinal translation movement of the table is controlled by a joystick control mounted on the side rail or auxiliary clamp on either side of the table top. The Allegro™ 6800 has 2 speed ranges of operation for longitudinal travel. Lateral travel is always performed at the normal speed setting. An illuminated green light on the top of the joystick control indicates which speed is engaged. A speed button on the top of the joystick control can be pressed once to switch between speeds. It should be noted that the joystick provides Interactive Proportional Speed Control™ (IPSC). This means that the Allegro™ 6800 travel speed will increase from stop to full speed within the selected range in proportion to how far the joystick is pushed.

Figure 7: Joystick control

To use the joystick control slide it onto the side rail or auxiliary clamp and plug it into the receptacle labeled joystick on either end of the pedestal of the Allegro™ 6800. The locking lever of the joystick control must always be rotated toward the pedestal end of the table to lock it to the side rail or auxiliary clamp. Correct orientation of this lever insures that the direction of travel of the Allegro™ 6800 will be consistent with the direction of the joystick movement.

WARNING: Failure to engage the joystick locking lever in the proper direction will result in travel in the direction opposite of the joystick motion.

Activation

button

Joystick

Speed selection

button

Locking

lever

Normal/high speed

indicator lights

Page 15: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 15

The joystick control includes an activation button (red button) that must be depressed to activate the joystick. This feature prevents unintended motion of the Allegro™ 6800 if the joystick is inadvertently pushed. To translate the Allegro™ 6800, press and hold down the JOYSTICK ACTIVATION button and then push the joystick in the intended direction of travel. Release the joystick when the desired position is achieved. The table will move at a speed that is proportional to the position of the joystick from the starting center position. Pushing the joystick further from center moves the table faster in that direction. Releasing the joystick activation button stops the table in any position.

NOTE: Although the Allegro™ 6800 is designed to travel laterally and longitudinally at any tilt or trendelenburg angle, it is recommended not to translate the table at high speed under full load (i.e. 450 lb. patient) when it is at extreme tilt or trendelenburg angles. NOTE: If the joystick is moved off center before the joystick activation button is pressed the table will start moving at that speed proportional to the position of the joystick. This can result in sudden jerky or rapid movement of the table top. Therefore, do not move the joystick off of center before pressing the joystick activation button.

Figure 8: Patient supine with head toward pedestal end

Page 16: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 16

5.0 INSPECTION

5.1 Acceptance And Transfer

Upon receipt of your Allegro™ 6800 Mobile Imaging Table, remove the table from the shipping crate following the unpacking instructions. Visually inspect all surfaces for freight damage. Check each caster for proper rolling operation. NOTE: Any freight damage must be reported to the freight carrier immediately upon delivery. It is the responsibility of the recipient to make freight damage claims.

• Verify receipt of the completeness and test report provided with each Allegro™ 6800.

• Read the model/serial number identification label found at the pedestal end of the base (see figure 4) to confirm the serial number and the input power requirements.

• Place the Allegro™ 6800 in an area with at least 4 feet of clearance on all sides.

• Perform Function Check; see section 6.0.

Pre-procedure/post-procedure and annually perform functional checks:

• Inspect and test the table as described in Function Check section 6.0.

• Thoroughly clean the table as described in the Cleaning section of this manual. Pay special attention to the cleanliness of the controls, as excessive soil can affect function.

• Inspect the power cord for cuts in the insulation or damage to the connector.

• Check to see that all fasteners are tight.

• Check the operation and security of the floor lock with and without a patient load in place.

5.2 Inspection And Transfer Before use, inspect the device for possible damage, excessive wear or non-functioning parts. Carefully inspect all critical, inaccessible areas, joints, electrical cords and all movable parts for possible damage or non-function. Damaged or defective products should not be used or processed. Contact your local Mizuho OSI sales representative for repair or replacement.

Page 17: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 17

6.0 FUNCTION CHECK

Perform all steps in this procedure. If you have any questions please feel free to contact Mizuho OSI Customer Resource Group at 800-777-4674, 7AM to 5PM PST. For a complete definition of reference terms used in this procedure, please refer to the Glossary of Terms (Section 2) this owners manual. 6.1 Function Check

1. Plug a hand pendant into the desired location. NOTE: A hand pendant can be plugged in at the pedestal end (back) or the cantilever end (front) of the base, or both locations with two hand pendants. If a hand pendant is removed or added while the table is operational, the table must be shut down for 30 seconds. When turned back on, the floor locks must be reengaged. 2. Plug in and tighten the connector for the joystick into the receptacle labeled

JOYSTICK at either end of the pedestal. 3. Attach and correctly orient the locking lever on the joystick. This lever must always

lock in place toward the pedestal end of the table. Correct orientation of this lever ensures the direction of travel of the Allegro™ 6800 will be consistent with the direction of the joystick movement.

4. If the table is to be used under battery power, ensure that the battery has been

properly charged and that the battery status LED on the hand pendant is green when the on/off power switch is on.

5. If the table is to be used under line power, plug in power cord into an appropriate

hospital grade outlet and turn on the main ON/OFF power switch. Note that the switch illuminates indicating that AC power is on.

6. Lockout mode: The Allegro™ 6800 is only operational if the LOCKED WHEN LIT

green indicator located on the control panel is illuminated and the FLOOR LOCK ON LED is illuminated on the hand pendant. Verify that the hand pendant functions do not operate except to activate floor locks.

7. Floor lock check using the control panel:

7.1. Press and hold DOWN TO LOCK button until green locked when lit indicator is illuminated on the control panel and that the FLOOR LOCK ON LED on the hand pendant is illuminated. Table is locked if it cannot be moved on its wheels.

7.2. Press and hold UP TO UNLOCK button until amber indicator is illuminated on the control panel and the FLOOR LOCK STOWED indicator is illuminated on the hand pendant. The table is unlocked if it can move on its wheels.

Page 18: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 18

8. Floor lock override button check: 8.1. With floor locks not engaged, press and hold floor lock override button located

under control panel. While holding the floor lock override button, operate the HEIGHT UP, HEIGHT DOWN, LEFT LATERAL ROLL, RIGHT LATERAL ROLL, TRENDELENBURG/REVERSE TRENDELENBURG. These should function without the floor locks being engaged while holding the override button (see figure 10).

9. Joystick control:

9.1. Move the joystick without pressing the activation button. No movement should occur.

9.2. Verify Interactive Proportional Speed Control™ (IPSC), variable speed operation. Press and hold activation button and move joystick. The tabletop moves in the direction of handle. Small handle movements away from center result in slow speeds while larger movements result in faster tabletop speeds.

9.3. Press the speed select button, illuminating the high speed indicator light and deactivating the normal indicator light. Tabletop can move at twice the normal speed in the long direction.

9.4. Relocate joystick control to other side of table and repeat steps 9.1-9.3 (see step 3 for more information).

10. Hand pendant check:

10.1. Press and hold FLOOR LOCK button. Listen for locking feet to move down and contact the floor. Observe on the control panel LOCKED WHEN LIT green indicator is illuminated and that the FLOOR LOCK ON LED indicator is illuminated on the hand pendant after floor locks engaged. Observe that all four feet have moved down and are in contact with the floor.

10.2. Press and hold FLOOR UNLOCK button. Observe that all four locking feet move up and STOWED WHEN LIT amber indicator is illuminated on the control panel and that the FLOOR LOCKS STOWED LED is illuminated on the hand pendant.

10.3. Press and hold HEIGHT UP button. Observe that the tabletop moves up. 10.4. Press and hold HEIGHT DOWN button. Observe that the tabletop moves

down. Complete range. 10.5. Press and hold REV.TREN button. Cantilever end of tabletop moves up.

Complete range. 10.6. Press and hold TREN button. Cantilever of tabletop moves down. Complete

range. 10.7. Press and hold LEFT LATERAL ROLL button. Tabletop tilts to the left.

Complete range. 10.8. Press and hold RIGHT LATERAL ROLL button. Tilt tabletop to the right.

Complete range. 10.9. Press and hold TOP FORWARD button. Tabletop translates forward. Full

distance. 10.10. Press and hold TOP AFT button. Tabletop translates aft. Full distance. 10.11. Press and hold TOP RIGHT button. Tabletop translates right. Full distance. 10.12. Press and hold TOP LEFT button. Tabletop translates left. Full distance.

Page 19: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 19

10.13. Disorient tabletop by raising height approximately two inches, lateral roll tabletop and performing trendelenburg approximately five degrees. Press and hold RETURN TO LEVEL button. Tabletop first levels roll and trendelenburg and then after a three-second delay, the tabletop moves down. Observe that the table top is level and in its lowest height position when movement stops.

10.14. Use the joystick and move tabletop, forward and laterally. Press and hold TOP HOME button on hand pendant. Tabletop moves to center and furthest position toward the pedestal end. Perform this same step but use the TABLE TOP HOME button on the control panel.

11. Verify the full function of the table system by moving and connecting hand pendant

and joystick to alternate ports. Power should be shut down when changing ports on joystick or hand pendant (see section 6.1 Note).

Page 20: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 20

7.0 THE ELECTRICAL SYSTEM

7.1 Description The electrical system provides control of all the table functions and is comprised of a power cord, an on/off circuit breaker switch, wire harnesses, a power supply, a controller circuit, a hand pendant, a joystick control, floor lock jackscrews, various electromechanical actuators, two stepper motors and two batteries. The electric motor driven leadscrew-type actuators manipulating the height, lateral roll and trendelenburg functions are controlled by the hand pendant. The floor lock buttons raise and lower the table floor lock feet to the floor. The translation motion of the table top is controlled by the joystick control and the hand pendant and driven by the stepper motors. Drawing number 6800-1003 is the Electrical Wiring Diagram for the Allegro™ 6800 Imaging Table. The input power requirement for this table is 120 VAC 60Hz, 5 Amp or 230 VAC 50Hz 2.5 Amps as indicated on the serial number label. Drawing number 6800-1004 is the Electrical Wiring Diagram for the 230 volt model 6800 Imaging Table. 7.2 On/Off Circuit Breaker Switch

An illuminated on/off circuit breaker main switch is located on the back of the pedestal end of the base of the table (see figure 4). When illuminated, it indicates that the table is plugged into a live electrical outlet and the power is on, (when the table is turned on using battery power, this switch will not illuminate). This switch also serves as a circuit breaker. In the event of an overload condition this switch will trip off. To reset, push to the off position and then the on position. NOTE: Determine the source of the overload prior to resetting this switch. 7.3 Component Circuit Breakers The electrical system includes three individual component circuit breakers that provide protection for the major components of the system. These circuit breakers are located in the base at the pedestal end of the table (see figure 4). If a short circuit occurs, the affected circuit breaker will trip as evidenced by a button protruding from the circuit breaker access hole. To reset, press the button in and release.

NOTE: Determine the source of the overload prior to resetting this button. 7.4 Power Cord The table is equipped with a standard IEC power cord with the appropriate hospital grade connector. The power cord is connected to the table at the IEC power entry socket located on the base of the pedestal near the on/off switch. 7.5 Floor Lock System The Allegro™ 6800 Mobile Imaging Table floor lock secures the table by powering 4 jackscrews to lower four legs onto the floor and unloading the swivel casters. It is operated by depressing the DOWN TO LOCK button located on the control panel or the FLOOR LOCK button on the hand pendant. The lock button is pressed and held until the floor lock is completely engaged as evidenced by the illumination of the floor lock

Page 21: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 21

indicator light on the control panel or the hand pendant. This button applies electrical current to each of the four jackscrew motors. Each leg is monitored by the controller for contact with the floor. In the case of an uneven floor, it is possible that one or more legs will continue to extend until firm contact is made with the floor. Complete locking is evidenced by the illumination of the green LOCKED WHEN LIT indicator located on the control panel and the FLOOR LOCK ON LED located on the hand pendant. The table should not move on its wheels when floor locks are engaged. NOTE: The table controller requires that the floor lock system be completely engaged before any other function can be operated. If the hand pendant or the joystick will not operate the table, verify that the floor locks are completely deployed, such that the table does not move. To unlock the floor lock, press and hold the UP TO UNLOCK button until the amber STOWED WHEN LIT indicator on the control panel and the FLOOR LOCK STOWED LED located on the hand pendant are illuminated indicating that the legs are completely retracted.

WARNING: Press and hold DOWN TO LOCK button until green LOCK WHEN LIT light is illuminated on the control panel, and that the FLOOR LOCK ON LED on the hand pendant is illuminated. Table is locked if it cannot move on its wheels. Ensure each floor lock is fully engaged before use. 7.6 Battery Charging The Allegro™ 6800 Table is designed to use two 12 volt batteries in series to provide 24 volts. When battery status LED on the hand pendant is red, batteries need to be charged. To charge the batteries, simply plug the power cord into an appropriate hospital grade outlet and turn the power switch on. While in storage, it is desirable that the table be plugged in and the power switch turned on so that the batteries remained charged. If this is not possible, the batteries must be charged for at least three hours a week. This can be accomplished when using the table under line power. To ensure proper battery function, it is recommended that the batteries be charged for at least 3 hours within the previous 48 hours prior to use. The battery status LED on the hand pendant must be illuminated green when the table is on. The batteries should be replaced every 5 years or when they fail to hold a charge. 7.7 Power Supply - AC & DC Operation

The system power supply converts AC power, 120-volt, 60Hz or 230-volt, 50Hz into 24VDC (nominal) operational power. There are four 24VDC output ports, each with a circuit breaker for thermal and short-circuit protection. Two are unregulated at 10-Amps with voltage between 36VDC and 22VDC depending on load and one at 15-Amps with voltage between 36VDC and 22VDC depending on load. The fourth DC output port is regulated at 24VDC +/- .25 ,and has no circuit protection within the power supply tray. When powered from an AC source, the batteries in series are constantly being charged at a variable rate depending on the charged state. Battery charge status is monitored by

Page 22: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 22

the 10-Amp motor controller PCB via a data port on the power supply PCB. A current-sensing circuit switches to DC battery power when it senses no AC power source is available. The 24VDC battery circuit can power and operate all of the on-board systems, motors and actuators. When running on DC battery power the power switch will not illuminate green but LED’s on the hand pendant will illuminate, indicating the battery circuit powers the table. 7.8 Battery System The 6800 Allegro™ Table uses two 12-volt batteries in series to provide 24-volts. When the battery status LED on the hand pendant is red, the batteries require charging. To charge, plug power cord into an appropriate hospital grade AC outlet and turn the power switch to on.

WARNING: While in storage, the batteries should remain charged for a minimum of three (3) hours a week. Using the table while on AC power will also charge the batteries.

• Normal battery operation is approximately 5 hours at 10% duty cycle. • If recharge is required, charge a minimum of three hours. A full charge should be

available after 18 hours. • If battery fails to hold a charge it should be replaced. Replace only with identical type

and size battery (see Replacement / Spare Parts List). • Replace both batteries at the same time. Also replace foam battery blocks on top of

battery. • Batteries should be replaced once every five (5) years. • Whenever the battery cover is opened, check battery for signs of corrosion. Battery

terminals should be clean and free from corrosion, oil, grease, dirt, or other contaminants.

• Battery ordering number: NV0801; failure to use an approved Mizuho OSI battery, voids warranty and can cause harm to equipment.

1. Turn power switch off, wait 30 seconds for discharge, unplug power cord from wall

outlet. 2. Remove pedestal-end base cover by removing the six screws using 1/8 in hex key.

WARNING: Disconnect batteries only with power cord unplugged and power switch off.

This symbol indicates the external stud that is at the same potential as the equipment chassis. If the AC power cable is not connected to an outlet with a protective earth ground, this external stud should be connected to a protective earth ground.

Page 23: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 23

3. Connect battery terminals as follows (see figure below): a. Red wire: Battery A positive terminal to power supply terminal E16. b. Yellow wire: Battery A negative terminal to battery B positive terminal. c. Black wire: Battery B negative terminal to power supply terminal E17.

Perform the following test to confirm proper battery connection: 1. Verify power cord is unplugged from AC wall outlet. 2. Turn power switch on (utilizing battery power) and verify all hand pendant LED’s

illuminate on and off in sequence. 3. With properly connected batteries, the hand pendant battery status light illuminates

green or red.

• A green battery-status light indicates batteries are charged and installed correctly.

• A red battery-status light indicates batteries are connected correctly but are in need of charging. If a charge is required, plug power cord into wall receptacle; turn power switch on, verify it illuminates green (AC is provided to the charger). Charge for at least three (3) hours.

NOTE: If this does not happen, immediately turn power switch off; verify proper battery connections, verify hand pendant is fully connected to table receptacle. Repeat test again. If problem continues, contact Mizuho OSI Customer Resource Group at 800-777-4674 for further assistance.

Figure 9: Battery orientation and wiring

Page 24: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 24

8.0 CLEANING 8.1 Cleaning And Disinfecting

NOTE: Never pour any liquid directly onto the table. Never subject the Allegro™ 6800 Mobile Imaging Table to any cart or equipment washer.

1. Table exterior

• Exterior surface should be regularly wiped clean with a mild detergent solution and wiped dry with a soft lint-free cloth.

NOTE: Do not wipe the support rails for the translating table top. These should remain with a light coat of oil at all times.

• Care should be taken to avoid exposing the table to excessive moisture. Flooding, fogging or steam cleaning is not recommended.

• Blood or other fluids, etc., if allowed to remain on the table for a long period of time, will require special cleaning to remove. A 5% acetic acid solution or white vinegar and water solution is especially good for this purpose.

• Staining and discoloration of plated or stainless steel surfaces can be corrected by cleaning with a good commercial cleaning compound, such as Stainless Steel Magic or Acme White Finish, and then buffing the surface by hand.

• To disinfect exterior surface use a quaternary ammonium or similar type disinfectant compound according to manufacturer's directions for use. Wipe dry with a soft lint-free cloth.

USE OF IODOPHORS WILL CAUSE STAINING

NOTE: Failure to thoroughly dry surface after cleaning and disinfecting may result in rust.

2. Casters

• Casters should be cleaned and disinfected in the same manner as table exterior. Access to the casters is achieved by removing the sheet metal enclosures at each end of the table base.

NOTE: Before removing any sheet metal covers the table must be turned off and unplugged. Dangerous high voltage may be present in the circuitry under the covers. This task should be performed by trained technicians only.

3. Table Pad MIZUHO OSI® Tempur-med® Pad IMPORTANT: The MIZUHO OSI® Tempur-Pedic® pad should always be stored in a

flat position. It can actually get stiff in cold temperatures and can crack and break if in a rolled position. It is important that you allow the pad to warm to room temperature before attempting to utilize or handle it.

• When handling always grasp by the entire thickness of the pad.

Page 25: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 25

• Do not lift, slide or carry MIZUHO OSI® Tempur-med® pads by grabbing the fabric. The cover may tear or rip.

• The pad is intended to be cleaned in place. It does not need to be rotated or flipped.

• Clean with standard hospital disinfectants labeled for use on table pads. Always dilute and rinse per manufacturer’s label instructions. Wipe dry with a lint free cloth. Do not soak or autoclave pads. The continued use of bleach or highly concentrated chemicals will void the warranty on the cover.

• When cleaning the bottom of the pad or the table top, simply lift one end of the pad, and fold it over onto the other end. Clean the pad or the table top then return the pad to flat on the table top.

4. Table tops

• Disinfect surfaces of table tops according to standard hospital procedure. 8.2 Lubrication 1. Table exterior

No lubrication is necessary. 2. Table interior

The translation and height adjustment sliding mechanisms of the 6800 Imaging Table require periodic lubrication. There are ten sliders in all and each should be lubricated with Thomson Linear Lube grease (or equivalent) once a year. Four sliders are on the lateral translation rails, four are on the longitudinal translation rails and two are on the main table column. Apply the grease through the zerk fittings located on each slide bearing housing. Do not over lubricate. All other components are lubricated for life at the factory and no other lubrication on the table is necessary. 8.3 Table Storage

When not in use the table should be stored in a clean dry environment with temperature between 32 °F & 120 °F (0 °C & 49 °C).

Page 26: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 26

9.0 TROUBLESHOOTING

9.1 Electrical System In the advent of a table malfunction, the first item to investigate is the input power. Be certain that the power cord is plugged into a live electrical outlet and the power switch is on. This is evidenced by an illuminated on/off switch. If none of the functions are operating, the floor lock may not be completely deployed. Press the FLOOR LOCK button until the green LOCKED WHEN LIT indicator is illuminated.

Figure 10: Floor lock override button

WARNING: If the floor lock is deployed as evidenced by visual inspection, all four floor lock feet should be in contact with the floor. Table should not move whether or not the green indicator is illuminated, and the up/down, lateral roll or trendelenburg functions do not operate. The floor lock override button can be used to temporarily bypass the interlock on the floor lock system. This button is located on the underside of the sheet metal housing at the FLOOR LOCK buttons (see figure 10). Press and hold this button while simultaneously operating the desired function on the hand pendant. Once released, no functions will be operable and this button must be pressed again if subsequent table movement is required. This override button only powers the height, trendelenburg and lateral roll functions on the hand pendant and is intended for emergency situations only. If it becomes necessary to use this feature, call Mizuho OSI Customer Resource Group immediately to arrange for table service.

Floor lock override button

(under this surface)

Floor lock button

Floor lock indicator lights

Page 27: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 27

CAUTION: Electrical shock hazard exists with access covers removed; use caution when working in exposed areas. 1. Measure voltage at wall outlet to verify proper AC input is available (120V or 230V). 2. Inspect power supply cord for physical damage, continuity, and adequate resistance. 3. Inspect power switch/circuit breaker:

a. Plastic boot is in good condition. b. Switch locks in the on and off positions. c. Green light illuminates when switched on.

4. Verify proper voltage at output ports: a. Refer to specifications (described above) and interconnect diagram.

5. Verify switchover from AC (mains) to battery operation: a. With charged battery (24 volts or higher with power switch off) turn power switch

on. b. Operate up/down function and pull power cord connector from table power-entry-

socket while motor is running. c. The table should keep moving with only a minor slow-down of motor speed.

6. Repair option for damaged or non-functioning power supply is replacing entire power supply tray assembly with new unit (see Replacement/Spare Parts List). Field repair is not recommended.

9.2 Table Functions Table is “dead”; no functions are operational:

• Verify power cord is plugged into an appropriate AC receptacle and into the power cord socket (located at the back end of pedestal).

• Verify power switch (located on back end of pedestal) is on. If illuminated, table is connected to live AC power. If no illumination, table may still operate on internal battery power (hand pendant control buttons and lights will be illuminated).

• Verify circuit breaker (part of power switch located on table-base panel) is not tripped off (indicating table power overload condition. Find overload source prior to resetting). To reset, switch power switch/circuit breaker off, then on.

• Verify component circuit breakers (located on table-base panel) are not tripped off (indicating component power overload condition. Find overload source prior to resetting). Observe the three component circuit breaker buttons are not protruding from circuit breaker access holes.

• Verify floor locks are deployed (all four locking feet make firm contact with the floor). Observe LOCKED WHEN LIT green light on control panel or FLOOR LOCK ON LED on hand pendant illuminates.

Floor lock button on control panel not functioning:

• Using the control panel, press and hold the DOWN TO LOCK control button. Verify the LOCKED WHEN LIT green indicator on the control panel or FLOOR LOCK ON LED on the hand pendant illuminates.

• Verify the floor locks are deployed (all four locking feet make firm contact with the floor). Using the control panel, press and hold the DOWN TO LOCK button on the control panel. Try pressing the UP TO UNLOCK button and then repeat press and hold the DOWN TO LOCK button.

Page 28: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 28

• Verify power cord is plugged into a properly grounded AC receptacle.

• If on battery power, verify proper charge; observe battery status LED illuminates green.

• Verify table power switch is on.

• Activate floor lock override; press and hold FLOOR LOCK OVERRIDE button (located under control panel). While holding floor lock override control button, press and hold any hand pendant control button for height up, height down, trendelenburg, reverse trendelenburg, left lateral roll or right lateral roll. Verify function.

Floor lock button on hand pendant not functioning:

• Using the hand pendant, press and hold the FLOOR LOCK button. Verify the LOCKED WHEN LIT green indicator on the control panel or the FLOOR LOCK ON LED on the hand pendant illuminates.

• Verify the floor locks are deployed (all four locking feet make firm contact with the floor). Using the hand pendant, press and hold the FLOOR LOCK button. Try pressing the FLOOR UNLOCK button and repeat press and hold the FLOOR LOCK button.

• Verify power cord is plugged into a properly grounded AC receptacle.

• If on battery power, verify proper charge; observe battery status LED illuminates green.

• Verify the table power switch is on.

• Activate floor lock override; press and hold the FLOOR LOCK OVERRIDE control button (located under control panel). While holding the FLOOR LOCK OVERRIDE control button, press and hold any hand pendant control button for height up, height down, trendelenburg, reverse trendelenburg, left lateral roll or right lateral roll. Verify function.

Height up or height down not functioning:

• Press and hold the hand pendant HEIGHT UP or HEIGHT DOWN control button. Observe tabletop movement.

• Verify the hand pendant is properly plugged into a hand pendant receptacle. Observe hand pendant buttons illuminate green.

NOTE: The table power switch must be turned off before disconnecting or connecting the hand pendant. Once reconnected, turn power on and deploy floor locks (all four locking feet make firm contact with the floor).

• With the power switch off, plug the hand pendant into the other hand pendant receptacle. Turn power switch on. Observe hand pendant buttons illuminate green.

• Verify the power cord is plugged into a properly grounded AC receptacle.

• If on battery power, verify proper charge. Observe battery status LED illuminates green.

• Verify power switch is on.

• Verify floor locks are deployed (all four locking feet make firm contact with the floor). Observe LOCKED WHEN LIT illuminates green on the control panel or FLOOR LOCK ON, LED illuminates on the hand pendant.

• Verify if other hand pendant buttons are functional. If not, replace the hand pendant.

Page 29: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 29

• Call Mizuho OSI Customer Resource Group for service or parts (possible hand pendant and/or 10 Amp motor-control PCB).

Trendelenburg or reverse trendelenburg not functioning:

• Press and the hold hand pendant TREN or REV.TREN button. Observe tabletop movement.

• Verify the hand pendant is properly plugged into a hand pendant receptacle. Observe hand pendant buttons illuminate green.

NOTE: The table power switch must be turned off before disconnecting or connecting the hand pendant. Once reconnected, turn power on and deploy floor locks (all four locking feet make firm contact with the floor).

• With the power switch off, plug the hand pendant into the other hand pendant receptacle. Turn power switch on. Observe hand pendant buttons illuminate green.

• Verify power cord is plugged into a properly grounded AC receptacle.

• If on battery power, verify proper charge; observe battery status LED illuminates green on the hand pendant.

• Verify power switch is on.

• Verify floor locks are deployed (all four locking feet make firm contact with the floor); observe LOCKED WHEN LIT indicator illuminates green on the control panel and the FLOOR LOCK ON LED illuminates on the hand pendant.

• Verify if other hand pendant buttons are functional. If not, replace hand pendant.

• Call Mizuho OSI Customer Resource Group for service or parts (possible hand pendant and/or 10 Amp motor-control PCB).

Left lateral roll or right lateral roll not functioning:

• Press and hold hand pendant LEFT LATERAL ROLL or RIGHT LATERAL ROLL control button. Observe tabletop movement.

• Verify hand pendant is properly plugged into a hand pendant receptacle; observe hand pendant control buttons illuminate green.

NOTE: The table power switch must be turned off before disconnecting or connecting the hand pendant. Once reconnected, turn power on and deploy floor locks (all four locking feet make firm contact with the floor).

• With the power switch off, plug the hand pendant into the other hand pendant receptacle. Turn power switch on. Observe hand pendant buttons illuminate green.

• Verify the power cord is plugged into a properly grounded AC receptacle.

• If on battery power, verify proper charge. Observe battery status LED illuminates green on the hand pendant.

• Verify power switch is on.

• Verify floor locks are deployed (all four locking feet make firm contact with the floor); observe LOCKED WHEN LIT indicator illuminates green on the control panel and the FLOOR LOCK ON, LED illuminates on the hand pendant.

• Verify if other hand pendant buttons are functional. If not, replace the hand pendant.

Page 30: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 30

• Call Mizuho OSI Customer Resource Group for service or parts (possible hand pendant and/or 10 Amp motor-control PCB).

Top forward or top aft not functioning:

• Press and hold hand pendant TOP FORWARD or TOP AFT control button. Observe tabletop movement.

• Verify the hand pendant is properly plugged into a hand pendant receptacle. Observe hand pendant buttons illuminate green. NOTE: The table power switch must be turned off before disconnecting or connecting the hand pendant. Once reconnected, turn power on and deploy floor locks (all four locking feet make firm contact with the floor).

• With the power switch off, plug the hand pendant into the other hand pendant receptacle. Turn power switch on. Observe hand pendant buttons illuminate green.

• Verify the power cord is plugged into a properly grounded AC receptacle.

• If on battery power, verify proper charge; observe battery status LED illuminates green on the hand pendant.

• Verify power switch is on.

• Verify floor locks are deployed (all four locking feet make firm contact with the floor); observe locked when lit light illuminates green.

• Verify if other hand pendant buttons are functional. If not, replace hand pendant.

• Call Mizuho OSI Customer Resource Group for service or parts (possible hand pendant and/or 10 Amp motor-control PCB).

Top left or top right not functioning:

• Press and hold hand pendant TOP LEFT or TOP RIGHT control button. Observe tabletop movement.

• Verify hand pendant is properly plugged into a hand pendant receptacle; hand pendant buttons illuminate green.

NOTE: The table power switch must be turned off before disconnecting or connecting the hand pendant. Once reconnected, turn power on and deploy floor locks (all four locking feet make firm contact with the floor).

• With the power switch off, plug the hand pendant into the other hand pendant receptacle. Turn power switch on. Observe hand pendant buttons illuminate green.

• Verify the power cord is plugged into a properly grounded AC receptacle.

• If on battery power, verify proper charge; observe battery status LED illuminates green on the hand pendant.

• Verify power switch is on.

• Verify floor locks are deployed (all four locking feet make firm contact with the floor); observe locked when lit light illuminates green.

• Verify if other hand pendant control buttons are functional. If not, replace the hand pendant.

• Call Mizuho OSI Customer Service Resource Group for service or parts (possible hand pendant and/or 10 Amp motor-control PCB).

Page 31: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 31

Top home not functioning on the hand pendant:

• Press and hold hand pendant TOP HOME button. Observe tabletop movement.

• Call Mizuho OSI Customer Resource Group for service or parts (possible hand pendant and/or 10 Amp motor-control PCB).

Table top home not functioning on the control panel:

• Press and hold TABLE TOP HOME button. Observe tabletop movement.

• Call Mizuho OSI Customer Resource Group for service or parts (possible hand pendant and/or 10 Amp motor-control PCB).

Return to level not functioning:

• Press and hold hand pendant RETURN TO LEVEL button. Observe tabletop movement.

• Disorient tabletop by raising height approximately two inches, lateral roll tabletop and activating trendelenburg approximately five degrees.

• Press and hold hand pendant RETURN TO LEVEL button. Observe roll and trendelenburg functions level first, and after a three-second delay, tabletop moves down to its lowest height position and is level.

• Call Mizuho OSI Customer Resource Group for service or parts (possible hand pendant and/or 10 Amp motor-control PCB).

Battery status LED illuminates red:

• Table battery system requires charging. Charge batteries minimum three (3) hours; plug power cord into appropriate AC outlet.

NOTE: Power switch must be switched on. Battery status LED remains illuminated red after battery charging:

• Call Mizuho OSI Customer Resource Group for service or parts. Joystick control not functioning (allowing longitudinal and lateral tabletop movement):

• Press and hold ACTIVATION button; red button (black on older tables) must be activated for joystick operation. Verify function.

• Verify joystick cable-connector is properly plugged into either joystick receptacle. Observe normal high speed indicator lights illuminate green when corresponding button pushed.

• Verify Joystick locking lever is completely locked toward pedestal-end to move tabletop in same direction.

Joystick control moving tabletop in opposite direction:

• Verify joystick locking lever is locked in proper direction, toward pedestal end.

Page 32: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 32

Joystick variable speed operation not functioning:

• Call Mizuho OSI Customer Resource Group for service or parts. Joystick speed select button not functioning:

• Call Mizuho OSI Customer Resource Group for service or parts.

Joystick: 1. Perform all indicated functions per functional check. 2. If there is no joystick response, inspect for physical damage:

a. Inspect coiled cord for cuts or other damage. b. Inspect the coiled cord connector pins; verify that they are not bent or broken. c. Verify the joystick handle provides firm resistance to motion in all directions and

returns to vertical when pushed to one side and released. d. Inspect the joystick handle support rod (hidden under the rubber boot) for bends. e. Inspect the red activation-button for damage; verify firm resistance to being

pressed with smooth up and down motion; the button should provide an audible click at or near the bottom of its stroke.

3. If there is no physical damage: a. Verify non-function when plugged into either access port. b. Bypass the internal wiring by plugging the joystick directly into the stepper-

controller PCB DB9-type socket marked joystick. c. If the table responds properly, the internal table wiring is at fault.

4. Repair option for damaged or non-functioning joystick is replacing entire unit. Field disassembly is not recommended.

Page 33: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 33

10.0 REMOVAL AND REPLACEMENT OF COMPONENTS 10.1 Floor Lock Foot Actuator

Figure 11: Foot end base: Cover removed, raised with support block and crank handle

Replacement procedure: 1. Turn power switch on. The switch illuminates indicating power is to the table. 2. Lock table to floor; press and hold DOWN TO LOCK button until green LOCKED

WHEN LIT light is on. 3. Turn power switch off. Wait 30 seconds for discharge then unplug power cord from

wall outlet. 4. Remove base-access cover from end of table needing replacement. CAUTION: Electrical shock hazard exists with access covers removed; use caution when working in exposed areas.

a. Pry off two white plastic access plugs for the floor lock jackscrew using a small screwdriver.

b. Remove four screws attaching the cover using a 1/8 inch hex key. c. Remove access cover.

5. Disconnect red wire from battery A positive terminal. 6. Raise end of table using floor lock crank handle (supplied with table). 7. Place 2-1/2 inch high support block under jackscrew floor lock motor mount. 8. Carefully lower table onto support block using floor lock crank handle. CAUTION: Make certain table remains stable. 9. Raise floor lock off the floor to its maximum height, using floor lock crank handle.

Floor lock

(lowered)

Support block

Coupling

and set-screw Floor lock

(raised) Crank handle

Page 34: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 34

10. Loosen floor lock actuator coupling setscrew using 1/8-inch hex key. Slide coupling towards actuator.

11. Remove actuator by removing the two socket-head cap screws using 3/16-inch hex key.

12. Replace floor lock actuator assembly (see Replacement/Spare Parts List section) or pad as described below.

13. Reassemble in reverse order. 14. Verify proper function using pre-op function check.

10.2 Floor Lock Pad If only the floor-lock pad requires replacing, perform the following: 1. Remove brake pad by removing M6-1.00 x 20mm socket-head bolt using 5mm hex

key. 2. If the standard attachment bolt is bent; the first threads may be damaged requiring

replacement with an optional, longer M6-1.00 x 35mm (see Replacement / Spare Parts List).

3. Install replacement floor lock pad (see Replacement / Spare Parts List) with attachment bolt and washer using 5mm hex key.

Figure 12: Floor lock actuator assembly

Page 35: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 35

10.3 Power Supply Tray The power supply tray contains the system power supply PCB, transformer, DC batteries, power switch/circuit breaker, component circuit breakers and power cord receptacle.

Figure 13: Power supply tray, base cover removed

Figure 14: Base cover and power supply tray removal

Battery-A Battery-B

Page 36: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 36

Replacement procedure: 1. Turn power switch off. Wait 30 seconds for discharge, then unplug power cord from

wall outlet. 2. Remove pedestal end base cover by removing the six screws using 1/8 in hex key. CAUTION: Electrical shock hazard exists with access covers removed; use caution when working in exposed areas. 3. Disconnect red wire from battery A positive terminal. 4. Unplug small 3-wire connector P 11. 5. Unplug 3-wire connector P 15. 6. Unplug 2-wire connector P 16. 7. Remove 3/8-inch hex bolt from ground stud using 9/16-inch wrench and remove top

two green / yellow striped wires. 8. Remove three socket head screws on each side of power supply tray and loosen

two socket-head screws at front of tray using 5/32-inch hex key. 9. Remove power supply tray by carefully lifting out of base casting. 10. Install in reverse order. 10.4 Controller Circuit The table functions are controlled by input voltage signals from the hand pendant or control panel to the 10-Amp motor control and stepper controller PCBs located in the cantilever-end of the base assembly. Voltage from the power supply also is fed to these controller boards. Interconnect diagrams are given in section 11 for reference.

Figure 15: 10-Amp motor control PCB and stepper controller tray assemblies

10-Amp motor control

PCB assembly

Stepper control tray

PCB assembly

Page 37: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 37

10.5 10-Amp Motor Control PCB Replacement procedure: 1. Turn the power switch off. Wait 30 seconds for discharge, then unplug power cord

from wall outlet. 2. Remove pedestal end base cover by removing the six screws using 1/8 in hex key. CAUTION: Electrical shock hazard exists with access covers removed; use caution when working in exposed areas. 3. Disconnect red wire from battery A positive terminal. 4. Remove cantilever end base cover by removing the four screws using 1/8 hex key. CAUTION: Electrical shock hazard exists with access covers removed; use caution when working in exposed areas. 5. Remove cables from 10 Amp motor controller PCB.

NOTE: Cable connectors are not marked, mark as needed.

a. J1 - To stepper controller b. P1 - To interrupt - to stepper controller/component Interface. c. P2 - To pc - no connection. d. P3 - Keypad - extension cable to hand pendant. e. P4 - Output - no connection. f. P5 - Power in - power cable to DC power supply. g. P6 - White marked cable to trendelenburg actuators. h. P7 - To tilt sensor assembly. i. P8 - Blue marked cable to trendelenburg actuators. j. P10 - Orange marked cable to tilt actuator. k. P11 - To tilt sensor assembly. l. P12 - Yellow marked cable to vertical actuator. m. P13 - No connection. n. P15 - No connection. o. P17 - No connection. p. P18 - Input - no connection.

6. Remove the two #10 socket head screws attaching the 10 Amp motor controller PCB mounting bracket, located under the PCB.

7. Lift the mid section base cover and remove the two #10 socket head screws attaching the 10-Amp motor controller PCB mounting bracket.

8. Carefully lift out the two #10 socket-head screws attaching the 10 Amp motor controller PCB assembly.

9. Reassemble in reverse order.

Page 38: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 38

10.6 Stepper Controller Tray

Y

X

Figure 16: Stepper control tray assembly

Replacement procedure:

1. Turn the power switch off. Wait 30 seconds for discharge, then unplug power cord from wall outlet.

2. Remove pedestal end base cover by removing the six screws using 1/8 in hex key.

WARNING: Electrical shock hazard exists with access covers removed; use caution when working in exposed areas. 3. Disconnect red wire from battery A positive terminal. 4. Remove cantilever end base cover by removing the four screws using 1/8 hex key.

WARNING: Electrical shock hazard exists with access covers removed; use caution when working in exposed areas. 5. Remove cables from stepper controller PCB.

NOTE: Cable connectors are not marked, mark as needed. a. J1 and J2 - To X & Y stepper drivers. b. P1 - Joystick input. c. P2 - Com port. d. P3 - Power in. e. P4 - Power out.

Stepper control

interface PCB

X-Axis

stepper PCB

Y-Axis

stepper PCB

Page 39: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 39

f. P6 - Senders g. P7 - Analog I/O h. P8 - Digital I/O

6. Remove cables from stepper driver PCBs. NOTE: Y stepper PCB is above and cables marked white, X stepper below marked black. a. J2 - Power b. J3 - Motor

7. Remove the ground wires from the ground stud. 8. Remove the two tray mounting screws closest to the table pedestal using 5/16 in hex

key. 9. Loosen the two, horizontal, tray mounting screws using 5/16 in hex key. 10. Carefully lift out the controller tray. 11. Reassemble in reverse order.

X-Y Axis stepper motor high speed adjustment procedure: With pedestal end base cover removed and using the joystick to control the stepper motors and using a stopwatch to time the total travel time, take the following actions:

1. Position each motor at one end of its travel using the joystick. 2. Switch the joystick to high speed and verify high speed green light is

illuminated. Without activating the button on the top of the joystick, move the joystick in the extreme opposite direction.

3. Press the button and the start time on the stopwatch. a. X-Axis motor (large motor), travel time = 4.7 (four point seven) seconds +/- 1.3

(one point three) seconds. NOTE: If the motor does not meet this requirement, adjust R86 on stepper controller PCB and re-test. b. Y-Axis motor, travel time = 4.7 (four point seven) seconds +/- 1.3 (one point

three) seconds. NOTE: If the motor does not meet this requirement adjust R87 on stepper controller PCB and re-test.

4. Using the hand pendant, joystick and control panel, run all functions through the entire range five times.

Table floor lock system adjustment procedure: Adjust floor lock potentiometers to read .65V (point six five) +/- .02V (point zero two) 1. Using a voltmeter connect ground probe to TP-2 (analog ground). 2. Connect positive probe to TP-3 and observe voltage. 3. Adjust trim pot R11 until voltage reads .65V (point six five) +/- .02V (point zero two). 4. Repeat steps for TP-4 and R19, TP-5 and R27, TP6 and R35.

Page 40: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 40

Table floor lock system test procedure: Using hand pendant and control panel, operate floor lock and unlock buttons through their full range and verify that 1. All four floor lock motors move while the button is pressed and all motors stop when

the button is released. 2. The floor lock LOCKED WHEN LIT light illuminates green when all four motor feet

are extended and all motors stop. 3. The floor lock STOWED WHEN LIT light illuminates amber when all four motor feet

are retracted and motors stop. 4. After the light, amber or green, illuminates all motors stop even when the button is

held. NOTE: If the floor lock motors appear to be out of sync (excessive travel distance between the group), this may indicate problems with the motor, gearbox or the stepper controller PCB.

10.7 Tilt Sensor Actuator

Tilt level sensor assembly: The tilt sensor controls the trendelenburg and tilt table levels. This sensor assembly is protected by the X-Y stage and tilt covers. Removal of the covers requires movement of the table system before, during and after the disassembly process. It is recommended that the power be turned off and the table unplugged, during the removal of the covers. After removal, and/or assembly of the tilt level sensor, a calibration sequence, extreme caution must be taken to ensure safety to the operator and to the table. Following are the steps required for access, removal, replacement and/or calibration:

Figure 17: Tilt sensor assembly

Top (X-Y Stage) cover removal:

WARNING: The following procedure requires removing components with the table on, use extreme caution.

Page 41: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 41

1. Turn power switch on. The green light illuminates. 2. Press and hold DOWN TO LOCK button until green LOCKED WHEN LIT light

illuminates. 3. Move tabletop to highest position and maximum left lateral roll using hand pendant. 4. Move tabletop to maximum left translation position using joystick. 5. Remove left Y-stop bracket (the L-shaped bracket with rubber bumper) by removing

two screws with 5/32 hex key. 6. Move tabletop to maximum right lateral roll using hand pendant. 7. Move tabletop to maximum right-translation position using joystick. 8. Remove right Y-stop bracket (opposite left Y-stop bracket) by removing two screws

with 5/32 hex key. 9. Remove X-Y stage cover by removing four screws, two on each side, using 1/8 hex

key. 10. Turn power off and unplug table. 11. Slide X-Y stage cover off end of tabletop. Left Tilt Cover Removal (The Bell Shaped Housing):

WARNING:

The following procedure requires removing components with the table on, use extreme caution. NOTE: On some older tables, front X-Y stage cover must be removed to access one 4-40 mounting screw. Do not replace this screw during reassembly. 1. Remove seven flat-head screws, using 1/16 inch hex key. 2. Remove three screws located on top surface of left tilt cover using 1/8 inch hex key. 3. Plug table in and turn power on. 4. Move tabletop with joystick to maximum left transitional position. Move tabletop, to

place casting recess (viewed best from below) for best screw clearance and easiest access.

5. Turn power switch off. Wait 30 seconds for discharge, then unplug power cord from wall outlet.

6. Disconnect red wire from battery A positive terminal. 7. Remove left tilt cover.

WARNING: To prevent damage to the sensors use care removing or replacing this cover. Replacement procedure: 1. Cut any cable ties holding the existing tilt sensor wire harness to the tilt sensor

bracket. 2. Gently pull the two sets of wires from the tilt sensor. NOTE: The color coding: White-marked connector goes to white-marked tilt adjust socket. Black-marked connector goes into black-marked trend adjust socket. 3. Remove the tilt sensor by removing two, 1/4-20x1/2 in. cap-head screws using 3/16

hex key.

Page 42: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 42

4. Replace in reverse order. Attach harness to sensor bracket, as needed, with small nylon cable tie. Use care to prevent possible damage by moving parts.

Tilt sensor calibration: NOTE: Use a bubble level resting on the tabletop as a level gauge. The level sensor adjustment screws are accessed from the top of the unit. A two degree maximum dead-band exists for each axis. For return to level to function, the tabletop must be outside of this dead-band. 1. Turn power switch on. The green light illuminates. 2. Press and hold DOWN TO LOCK button until green LOCKED WHEN LIT indicator

illuminates. 3. Raise the tabletop to a comfortable level. 4. Push the LEFT LATERAL ROLL or RIGHT LATERAL ROLL buttons to level

tabletop. 5. Push the LEFT LATERAL ROLL button to tilt the tabletop more than five degrees. 6. Push the RETURN TO LEVEL button to level the tabletop; note the degree of level. 7. Push the RIGHT LATERAL ROLL button to tilt the tabletop more than five degrees. 8. Push and hold the RETURN TO LEVEL button to level the tabletop; note the degree

of level. 9. Turn the tilt adjust screw a partial turn. NOTE: Clockwise will stop the tabletop more to the right, counter-clockwise will stop the tabletop more to the left. The tabletop will overshoot each stopping point, requiring "splitting the difference" to adjust for proper balance.

10. Test the adjustment by:

a. Push the LEFT LATERAL ROLL button to tilt the tabletop to maximum. b. Push the RETURN TO LEVEL button to level the tabletop; note the degree of

level. c. Push the RIGHT LATERAL ROLL button to tilt the tabletop to maximum. d. Push the RETURN TO LEVEL button to level the tabletop; note the degree of

level. 11. Continue adjustments until the tabletop stops from either maximum lateral roll at the

same offset from zero. 12. Trendelenburg / reverse-trendelenburg calibration (TREN/REV.TREN).

a. Repeat steps of left and right lateral roll calibration using TREN/REV.TREN buttons on the hand pendant.

13. Turn power switch off. Wait 30 seconds for discharge, then unplug power cord from wall outlet.

14. Disconnect red wire from battery A positive terminal. 15. Attach left tilt cover.

WARNING:

Use care replacing this cover to prevent damage to the sensors.

16. Place the left tilt cover in its original position. a. Install the three screws attaching left tilt cover, located on top surface, using 1/8

hex key.

Page 43: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 43

b. Move the tabletop with the joystick to maximum left position for best screw clearance and easiest access.

c. Install the seven screws attaching the left and right tilt covers using a 1/16 inch hex key.

17. Attach X-Y stage cover. a. Place the X-Y stage cover onto the tabletop. b. Install the four screws attaching the X-Y stage cover using a 1/8 inch hex key. c. Install the two flat-head screws attaching the left Y-stop bracket using 5/32 inch

hex key. d. Install the two flat-head screws attaching the right Y-stop bracket using 5/32 inch

hex key.

Page 44: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 44

11.0 TECHNICAL DRAWINGS AND PARTS LISTS 11.1 Electrical Interconnect Diagram, 120 VAC

LIM

IT SWITCH ARRAY

6800-203

Y-AXIS

15A

10A

10A

10A

TRENDELENBURG

ACTUATOR

ACTUATOR

TRENDELENBURG

REVERSE-

TILT

ACTUATOR

VERTICAL

6800-247

SENSOR

TILT

ASSY

NV0579

ACTUATOR

LIM

IT SWITCH

6800-80

RIGHT

NV0579

LIM

IT SWITCH

LEFT

CONTROL PCB

5V, 12V , 100mA

MADE PER 6800-320

24V

MADE PER 6800-320

P6

24V

6800-185

MADE PER

6800-199

BLU/BLK

BLK/RED

ACTUATOR

BLK SHRINK W

RAP

WHT SHRINK W

RAP

24V

24V

P1

P2

MADE PER 6800-320

TILT

6800-171

24V

6800-171

6800-171

PCB / CABLE

24V

10 AMP MOTOR

P8

P12

24V

24V

P11

P10

P7

6800-240

P1

NV0564

P3

6800-248

BLU/BRN

6800-274 OR 6800-275

6800-194

PLUS

CAPACITOR

10,000UF

50V DC

(PART OF

6800-196)

+ P5 -

6800-196

J1

2 PLCS

BLACK SHRINK W

RAP

WHITE SHRINK W

RAP

6800-195

X STEPPER M

OTOR

SUPPLY SUPPLY

6800-250

POWER SUPPLY TRAY

6800-204

1 & 2

PINS

JUMPER

6800-193

5A

THRU

6800-200

24VDC

P16

24VDC

P13

PCB

P12

P10

E12

P15

24VDC

P14

GND

E15 24VDC

SW

9600-750

6800-198

60Hz

110V

AC

P8

J1

J2

STEPPER

MOTOR

6800-51

STEPPER

X-AXIS

6800-50

MOTOR

SWITCH

" DOWN TO LOCK"

HAND CONTROL

9 PIN D SUB

FOR JOYSTICK

GREEN LED

"LOCKED W

HEN LIT"

YEL/BLK

"STOWED W

HEN LIT"

GRN/BLK

"UP TO UNLOCK"

SWITCH

GRY/BLK

WHT/BLK

FLOOR LOCK

CONTROL PANEL

SWITCH

AMBER LED

"TABLETOP HOME"

"FLOOR LOCK OVERRIDE"

SWITCH

REAR

BRN/BLK

BLU/BLK

INPUT FOR

Y-AXIS

6800-202

PCB

STEPPER DRIVE

STEPPER-CONTROLLER/DRIVER TRAY ASSEMBLY

GND

J3

J2

X-AXIS

STEPPER DRIVE

J1 GND

J3

J2

PCB

6800-201

6800-201

J1

6800-251

NV0564

9600-450

Y STEPPER M

OTOR

LIM

IT SWITCH ARRAY

24V

6800-238

6800-150

JOYSTICK

HAND PENDANT CONTROL

P3

STEPPER CONTROLLER

INTERFACE BOARD

P2

6800-190

P6

P4

+24V

+24V

+24V

+24V

P5

NV0337

6800-192

P1

GND

6800-197

24V

24V

24V

AUXILIARY PANEL

9650-431A

INPUT FOR

FOR JOYSTICK

9 PIN D SUB

FRONT

HAND CONTROL

CONNECT TO

CONNECT TO

EITHER PANEL

EITHER PANEL

FLOOR

6800-205

6800-206

MOTORS

LOCK

Page 45: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 45

11.2 Electrical Interconnect Diagram, 230 VAC

J1

J2

Page 46: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 46

11.3 Top Level Table Assembly Exploded View

26 6800-290 HAND PENDANT SWIVEL MOUNT, ASSY 1

25 NP1816 WARNING LABEL, NOT FOR STORAGE 3

24 NG0752 #10-32 PEM SELF CLINCHING SS 3

23 NB0331 RUBBER BUMPER #10 FEMALE THREAD 2

22 AAAW075BAC 1/4-20 X .75 SST FLHD SKT SCR 4

21 AAAT075BAC #10-32 UNF X .75 LG, SST FL HD SKT SCR 2

20 6800-175 BRACKET, BUMPER STOP MOUNT, Y AXIS 2

19 NW0367 USER GUIDE, ALLEGRO TABLE 2

18 9600-750 HOSP. GRADE PWR CORD, 10FT, RT ANGLE 1

17 NP1858 LABEL, 6800 ALLEGRO 2

16 NP1625 6800 TABLE SERIAL NUMBER LABEL 1

15 NP1560 LABEL, ETL LISTING 1

14 NP1559 GROUNDING LABEL 1

13 NP1551 LABEL, MEDICAL, TYPE B 1

12 NC0356 CABLE TIE, 4" BLACK 1

11 AAM6030CAA M6-1.00 X 30mm SKT HD CAP SCR, GR12.9 4

10 AAAT037AAC #10-32 UNF X .37 SST BT HD SKT SCR 11

9 6800-280 ASSY, JOYSTICK II, HEAVY DUTY 1

8 6800-279 X-Y STAGE COVER, X-MOTOR COVER (PLASTIC) 1

7 6800-278 X-Y STAGE COVER, FWD (PLASTIC) 1

6 6800-277 BOTTOM PLATE 1

5 6807-3 HANDPENDANT ALLEGRO 1

4 6800-140 ASSEMBLY, TOP PAD, 4 WAY FLOAT TABLE 1

3 6800-66 STOP, TABLE TRAVEL, X-AXIS 2

2 6800-5 ASSY, TABLE TOP, 4 WAY FLOAT TABLE 1

1 6800-4 FINAL BASE ASSEMBLY 1

ITEM PART NO DESCRIPTION QTY

2

4

4X

26

6

12

14

10

32X

114X

9

20

13

15

16

253X

22

1

23

21 17 2X(REF)

202X

23 2X

18

5

224X

212X

10 7X

8

24 3X

7

Page 47: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 47

11.4 Table Sub Assembly Parts List 75 AAAW050CA 1/4-20 X .50 HEX HD CAP SCR 2 150 MZNP1811 LABEL, EQUIPOTENTIAL 1

74 AAAW050BAC 1/4-20 X .50 SST FLHD SKT SCR 4 149 AM062200 .625 DIA X 2.00 SST DOWEL PIN 2

73 AAAT125CAC #10-32 X 1.25 SST CAP HD SKT SCR 5 148 ADAHEC #4 NYLOCK NUT, SST 8

72 AAAT125BAC #10-32 X 1.25 SST FLHD SKT SCR 1 147 6850-1012 INTERNAL HAND PENDANT CABLE EXTENSION 1

71 AAATO75HA #10-32 X .25 SST SKT SET SCR 7 146 NP1837 BATTERY ORIENTATION LABEL 1

70 6800-276 X-RING DOOR, FRONT 1 145 NP1851 BATTERY INSTRUCTION PICTURE LABEL 1

69 AAAT075AAC #10-32 X .75 BTN HD SKT SCREW S.S. 4 144 NP1850 BATTERY INSTRUCTION LABEL 1

68 AAAT050HAC #10-32 X .50 SST SKT SET SCR 1 143 6850-1014 BATTERY CABLE SET 1

67 AAAT050BAC #10-32 X .50 SST FLHD SKT SCR 2 142 NV0711 MOLEX SOCKET CRIMP 28-22 AWG 16

66 AAATO50AAC #10-32 X .50 SST BTN HD SKT SCR 12 141 NV0630 CONNECTOR HOUSING, MALE, 8 CIRCUIT 39-01-2080 4

65 AAAT037CAC #10-32 X .37 SST CAP HD SKT SCR 8 140 6800-76 TILT WORM GEAR SHIM, .005 1

64 AAATO37AAC #10-32 X .37 SST BTN HD SKT SCR 55 139 6800-274 10 A MOTOR CONTROL POWER CABLE 1

63 AAAT031AAC #10-32 X .31 SST BTN HD SKT SCR 34 138 6800-248 CABLE ASSY. ,FLAT RIBBON M-F DB9 1

62 AAAT025HJA #10-32 X .25 SELF LOCKING SKT SET SCR 2 137 6800-246 TILT SENSOR WIRE HARNESS 1

61 AAAT025AAC #10-32 X .25 SST BTN HD SKT SCR 4 136 6800-203 CABLE ASSY, Y-AXIS STEPPER MOTOR CONTROL 1

60 AAAMO37CAC #6-32 X .37 SST CAP HD SKT SCR 1 135 6800-202 CABLE ASSY, X-AXIS STEPPER MOTOR CONTROL 1

59 AAAM018AAC #6-32 X .188 SST BTNHD SKT SCR 2 134 6800-199 DIGITAL I/O WIRE HARNESS 1

58 AAAH025AAC #4-40 X .25 SST BTNHD SKT SCR 8 133 6800-198 SENSOR ARRAY WIRE HARNESS 1

57 AAAH031BAC #4-40 X .31 SST FLHD SKT SCR 21 132 6800-197 FLOOR LOCK HARNESS ASS. 1

56 AAAH018BAC #4-40 X .188 SST FLHD SKT SCR 6 131 6800-193 CABLE ASSY, PWR SPL TO STEPPER CONTROLLER 1

55 9650-231 ½ RUBBER CUSHIONED STEEL LOOP CLAMP 7 130 6800-119 TILT WORM GEAR SHIM .0239 THICK 2

54 9650-227 .375 INCH PLASTIC CABLE CLAMP 10 129 6800-118 TILT WORM GEAR SHIM .007 THICK 2

53 6800-271 BELLOWS MOUNT 2 128 9650-225 .25 INCH PLASTIC CABLE CLAMP 1

52 6800-270 TILT COVER, RIGHT 1 127 NV0649 .75 INCH PLASTIC CABLE CLAMP 3

51 6800-269 TILT COVER, LEFT 16 126 6800-185 ACTUATOR POWER CORD 1

50 6800-268 COLUMN SKIRT, FIXED RIGHT 1 125 NV0653 HORIZONTAL MOUNTING CLIP, CAPACITOR 2

49 6800-267 COLUMN SKIRT, FIXED LEFT 1 124 6800-192 JOYSTICK Y CABLE 1

48 6840-51 X-RING COVER, RIGHT 1 123 NV0650 1 INCH PLASTIC CABLE CLAMP 5

47 6840-50 X-RING COVER, LEFT 1 122 6800-196 CABLE ASSEM, X-AXIS STEPPER MOTOR 1

46 6800-260 BELLOWS SUPPORT 2 121 NV0587 PUSH BUTTON, MOM., SST 4

45 6800-259 BELLOWS 2 120 6850-195 FOAM BATTERY BLOCK 2

44 6800-4250 ASSEMBLY CONTROLLER TRAY (STEPPER) 1 119 NV0801 BATTERY 12V, 12 A H SONNENSCHEIN ES12-12 2

43 6800-250 POWER SUPPLY TRAY ASSEMBLY 1 118 6850-1011 2 OUTPUT INT. HAND PEND. CBL. W/NANABOSHI 1

42 6800-249 TILT SENSOR SPACER 1 117 NV0259 3/8 IN RUBBER CUSHIONED STEEL LOOP CLAMP 7

41 6800-247 TILT SENSOR ASSEMBLY 1 116 NP1549 ELECTRICAL GROUND LABEL 2

40 6800-244 CAPACITOR MOUNT PLATE 1 115 NL1255 ACTUATOR, ELEC., 100MM, LINAK 2

39 6850-1020 ASSY, 10 AMP MOTOR CONTROL 1 114 NG0753 JACKSCREW #4-40 X 187 X .31 THD W/WASH & NUT 6

38 6800-230 ASSEMBLY, MANUAL FLOOR LOCK CRANK 1 113 NG0752 #10-32 PEM SELF CLINCHING SS 8

37 6800-229 X-RING WITH BUSHINGS ASSY 1 112 NB0339 EAGLE PLASTIC NYLON HOLE PLUG ø1.00 4

36 6800-208 FLOOR LOCK LED YELLOW 1 111 6800-195 CABLE ASSEM, Y-AXIS STEPPER MOTOR 1

35 6800-207 FLOOR LOCK LED GREEN 1 110 NA0190 THOMSON RAIL 25MM X 508MM 2

34 6800-206 Y-AXIS SENSOR ARRAY ASSY AND WIRING DIA. 1 109 NA0189 THOMSON 25MM CARRIAGE STD LENGTH 4

33 6800-205 X-AXIS SENSOR ARRAY ASSY AND WIRING DIA. 1 108 NA0186 THOMSON CARRIAGE 25MM LONG 4

32 6800-187 BRACKET, TABLE RETURN TO CENTER FLAG, Y-AXIS 1 107 KA0060400 1/16 X 4 ½ NEOPRENE RUBBER W/ADH 24

31 6800-186 BRACKET TRAVEL STOP FLAG, Y-AXIS 2 106 AJ0114 RETAINING RING EXTERNAL 1” DIA, SST 4

30 6800-171 TILT MOTOR LIMIT SWITCH ASSY 1 105 AE040081A WASHER, FLAT .400 X .81 X .051 2

29 NV0337 10 FT EXTENSION CABLE M-F DB9 1 104 AE037075H WASHER, FLAT HEAVY DUTY .38 ID .75 OD 4

28 6800-174 BRKT, TILT MTR LIM SW MNT 2 103 AE037068BD WASHER, LOCK .38 X .68 HEAVY DUTY 4

27 6800-173 PC BOARD MOUNT, TILT MOTOR 1 102 AE019050KF SHOULDER WASHER #10 NYLON 4

26 6800-148 CABLE TRAY, Y-RING 1 101 AE018037D STAR LOCK WASHER, INTERNAL, .18 X .37 3

25 6800-133 BASE COVER, FRONT 1 100 MZA30108030 8MM HEX NUT 10

24 6800-132 BASE COVER, BACK 1 99 ADATA #10-32 HEX NUT 3

23 6800-130 BASE COVER, MID SECTION 1 98 6850-1010 CABLE BATTERY MONITORING 1

22 6800-129 COLUMN SKIRT FLOATING 1 97 AC012025 POP RIVET, .12 X .25 LG 8

21 6800-127 COLUMN SLIDER COVER 1 96 AAM8035CAN M8 X 35 SOCKET HEAD CAP SCREW CLASS 12.9 8

20 6800-99 PIN, TRENDELENBURG 2 95 AAM8030CA M8 X 30 SOCKET HEAD CAP SCREW CLASS 12.9 14

19 6800-98 PIN, TILT, BACK 1 94 AAM8030BAN M8 X 30 FLAT HEAD SOCKET SCREW CLASS 12.9 3

18 6800-97 PIN, TILT, FRONT 1 93 AAM8025CAN M8 X 25 SOCKET HEAD CAP SCREW CLASS 12.9 7

17 6800-96 SWIVEL BRACKET MOUNT WASHER 2 92 AAM6030CAA M6-1.00 X 30MM SKT HD CAP SCR, CLASS 12.9 18

16 6800-95 PRESSURE PLATE, TILT ACTUATOR MOUNT 1 91 AAM6025CAN M6-1.00 X 25MM SKT HD CAP SCR, CLASS 12.9 16

15 6800-94 SWIVEL BRACKET ADJUSTMENT BRACKET 1 90 AAM10045FA 10MM DIA. X 45MM LG, SKT HD SHLDR SCR 1

14 6800-90 SWIVEL BRACKET WELDMENT, TILT 1 89 AAM10035FA 10MM DIA. X 35MM LG, SKT HD SHLDR SCR 1

13 6800-80 ASSEMBLY, TILT ACTUATOR 1 88 AAM10030FA 10MM DIA. X 30MM LG, SKT HD SHLDR SCR 3

12 6800-79 TILT AXIS WORM GEAR 1 87 AABF125CA ½-13 X 1.25 CAP HD SKT SCR 2

11 6800-77 TILT WORM GEAR SHIM, .010 2 86 AABB075HA 3/8-16 X .75 SCK SET SCR 1

10 6800-75 TILT WORM GEAR SHIM, .002 1 85 AABB175DDA 3/8 – 16 X 1.75 HEX HD CAP SCR, GRADE 8 1

9 6800-73 X-Y CORD COVER PLATE 1 84 AABB175CAN 3/8-16 UNC X 1.75 CAP HD SKT SCR GR 8 3

8 6800-72 GEAR RACK, Y-AXIS, X-Y COUPLER 1 83 AABB125CAC 3/8-16 UNC X 1.25 SST CAP HD SKT SCR 4

7 6800-70 X-Y COUPLER, CASTING MACHINED 1 82 AABB100CAC 3/8-16 X 1.0 SST CAP HD SKT SCR 2

6 6800-54 SHIM, 1.00 DIA PIN 6 81 AABB075CAC 3/8-16 X .75 SST CAP HD SKT SCR 2

5 6800-51 ASSY, STEPPER MOTOR AND GEAR BOX Y-AXIS 1 80 AABB075CAA 3/8-16 X .75 CAP HD SKT SCR GRADE 8 3

4 6800-50 ASSY, STEPPER MOTOR AND GEAR BOX X-AXIS 1 79 AAAY350DD 5/16-18 X 3.5 HEX HEAD CAP SCREW 2

3 6800-38 Y-RING MACHINED 1 78 AAAW125CA ¼-20 X 1.25 CAP HD SKT SCR 2

2 6800-37 TRENDELENBURG ADJUSTMENT LEVER, X-RING 1 77 AAAW075CA ¼-20 X .75 CAP HD SKT SCR 3

1 6800-7 BASE FEET AND COLUMN SUBASSEMBLY 1 76 6800-273 X RING DOOR, REAR 1

ITEM PART NO. DESCRIPTION QTY ITEM PART NO. DESCRIPTION QTY

Page 48: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 48

11.5 Table Assembly Exploded View

NOTE: Numbers correspond to item numbers on parts list in section 11.4.

19

68

18

71

20 2X

106 4X88

10090

286

77

123 4X

634X

1152X

882X

1004X

43 (REF)

123(REF)

63(REF)

115 (REF)

88 (REF)

6

X-Y Coupler assembly

Y-Ring assembly

X-Ring assembly

Stepper-

controller

Base assembly

Power supply

10 Amp controller

Batteries

Page 49: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 49

11.6 Base Cover Assembly Exploded View

NOTE: Numbers correspond to item numbers on parts list in section 11.4.

Base Assembly Exploded View

63 6X

54 5X107 3" X 4.5"

1073" X 4.5"

111

55 2X

59 2X

125

66 2X

40

63 2X

633X127 3X

107 3" X 13"

128

122

64 4X

50

39

642X

25

662X

107 REF

44

23

576X

1

64 4X

107 REF

64 2X

642X

112 4X

49

24

22

692X

21

614X

532X

64 6X

102 4X

45

Page 50: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 50

11.7 Base Assembly Exploded View

NOTE: Numbers correspond to item numbers on parts list in section 11.4.

63 6X

54 5X107 3" X 4.5"

1073" X 4.5"

111

55 2X

59 2X

125

66 2X

40

63 2X

633X127 3X

107 3" X 13"

128

122

Page 51: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 51

11.8 X-Ring Assembly Exploded View

NOTE: Numbers correspond to item numbers on parts list in section 11.4.

47

14

54

55

1173X

123

63

63

45

30

27

28

37

38

462X

13

15

16

17

76

70

48

64

60

78

87

672x

56

57

74

99

79

85

1134X

97

101

Page 52: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 52

11.9 Y-Ring Assembly Exploded View

NOTE: Numbers correspond to item numbers on parts list in section 11.4.

26

41

42

51

52

3

5

10

11

12

54

117

34

63 2X

64 3X

75 2X

77 2X

80 3X

812X

56 3X

57 2X

62 2X

64

63

563X

63

843X

11

11.10 Y-Ring Assembly Bottom View

NOTE: Numbers correspond to item numbers on parts list in section 11.4.

64

63 117

3

4

1094X

116

91 16X

968X

Page 53: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 53

11.11 X - Y Coupler Assembly Exploded View

NOTE: Numbers correspond to item numbers on parts list in section 11.4.

634X

9

4

652X

33

105 2X

82 2X

7

1044X

103 4X

834X

64 3X

116

55

54

4

Page 54: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 54

11.12 X - Y Coupler Assembly Bottom View

NOTE: Numbers correspond to item numbers on parts list in section 11.4.

92

110

8

92

110

954X93

93

943X

95 2X

93

954X

95 4X

65 6X

32

31 2X

7

108 4X

71 6X735X

72

93

93

93

93

1172X

110

Page 55: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 55

11.13 Replacement/Spare Parts List 10-Amp motor controller PCB 6800-1021 Base cover, floating 6800-129 Base cover, mid section 6800-130 Battery, 12V, 12A (2x reqd) NV0801 Column skirt, fixed, left 6800-267 Column skirt, fixed, right 6800-268 Column slider cover 6800-127 Floor lock gearbox assembly 6800-210 Floor lock motor 6800-217 Floor lock pad MZC700004B0 Floor lock pad screw MZA00006200 Floor lock manual crank 6800-230 Hand pendant 6807-3 (14-button) Joystick 6800-280 LED (light assembly), amber 6800-208 LED (light assembly), green 6800-207 On/off power switch NV0586 Power supply tray assembly 6800-250 Power cord, right angle 9600-750 Power cord receptacle NV0585 Stepper controller/driver PCB 6800-4250 Table pad 6800-7140 Tilt sensor assembly 6800-247 X-ring cover, left 6800-262 X-ring cover, right 6800-263

Page 56: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 56

12.0 Mizuho OSI CUSTOMER RESOURCE GROUP 12.1 Contact For Parts And Service For detailed repair information or to order replacement parts, call the Mizuho OSI Customer Resource Group:

1-800-777-4674 A Customer Resource Group line is available from 7AM-5PM PST, Monday through Friday. Please leave a message after business hours. Please state slowly your name, phone number, your facility name and city, and the affected equipment model number. An e-mail message may be left anytime at [email protected] or through the Customer Resource Group icon in our web site: www.mizuhosi.com. 12.2 Instant Support Value Package

Additional technical information and documentation is available via Mizuho OSI’s subscription based instant support. Instant support is a web-based virtual support agent accessible from Mizuho OSI’s web page; www.mizuhosi.com. Instant support provides immediate and actionable answers to problems and questions on a 24/7 basis. That Instant Support Value Package (ISVP) is a comprehensive support package designed to enhance the hospitals internal equipment support efforts. Contact your local Mizuho OSI sales representative for subscription information and additional details of the program. 12.3 To Order Replacement Parts (RP) 1. If unable to identify part please telephone, fax or e-mail to Customer Resource

Group. 2. If part is known please telephone, fax or e-mail part number and description to

Customer Resource Group for price and availability. 3. To place Replacement Parts (RP) order please telephone, fax or e-mail part number,

description, price, customer number and method of shipment to Customer Resource Group with purchase order. Indicate that order is for Replacement Parts (RP).

12.4 Returns To Mizuho OSI (RGA) 1. Identify parts to be returned. 2. Telephone, fax or e-mail part number and description of parts for return to Customer

Resource Group for Return Goods Authorization (RGA) number. 3. Return ship parts with RGA number clearly marked on outside of package. 4. Complete certificate of disinfected parts document #40204-17.

Page 57: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 57

12.5 To Send A Part For Repair (RA) 1. If unable to identify part, please call, fax or e-mail to Customer Resource Group. 2. If part is known please telephone, fax or e-mail part number and description to

Customer Resource Group for Repair Authorization (RA) number. 3. Clean (disinfect) parts prior to shipping; complete certificate of disinfected parts

document #40204-17. 4. Ship part with RA number clearly marked on outside of package. 5. Part will be evaluated and customer will be contacted with the cost of repair. 6. After customer approval of repair cost, part will be repaired and return shipped to

customer.

12.6 EC Authorized Representative Emergo Europe Molenstraat 15 2513 BH The Hague The Netherlands Tel: (31) (0) 345-8570 Fax: (31) (0) 346-7299

Page 58: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 58

13.0 Mizuho OSI ALLEGRO™ 6800 OPTIONAL ACCESSORIES

Figure 18: Stent tray

Stent tray (Part# 6900-21) The stent tray is designed for and intended to support guide wires, stents and catheters. It can be mounted on either the pedestal or cantilever end of the table. NOTE: This accessory is not designed to support a patient.

Figure 19: Radiolucent single arm board

Radiolucent single arm board (Part# 6900-12) The radiolucent arm board is placed between the tabletop and table pad and is held in place by the patient’s weight. This arm board has a carbon fiber surface with a 2 inch (5cm) pad. The single arm board can be used on either side of the table, right or left.

Figure 20: Auxiliary clamp

Auxiliary clamp (Part# 6900-20) The auxiliary clamp mounts where needed on the cantilever of the table to provide 6 inches (15 cm) of side rail.

Page 59: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 59

Figure 21: Radiolucent AV fistula board

Radiolucent AV fistula board (Part# 6900-25) A radiolucent platform for performing upper extremity procedures. The radiolucent AV fistula board has a 2 inch (5 cm) pad and attached auxiliary clamp for securing to the table top.

Figure 22: Rail mounted I.V. pole

Rail mounted I.V. pole (Part# 6900-23) The rail mounted I.V. pole mounts on the side rail. It telescopes 34 – 60 inches (86 – 152 cm).

Figure 23: Joystick

Replacement Parts: Joystick (Part# 6800-280) See section 4.0 Basic Operation, for a complete description.

Page 60: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 60

Figure 24: Hand pendant

Hand pendant (Part# 6807-3) See section 4.0 Basic Operation, for a complete description

Figure 25: Swivel mount for hand pendant

Swivel mount for the hand pendant (Part# 6800-290) The swivel mount allows for mounting the hand pendant horizontally on the side rail or auxiliary clamp. The hand pendant then can be swiveled into the desired position for easy access by the surgeon, anesthesiologist or surgical team.

Figure 26: Drape rod

Drape rod (Part# 6800-350) The drape rod mounts on the side rail or auxiliary clamp and is for holding the surgical drapes in place while the table top is being moved during a procedure.

Page 61: ALLEGRO™ MOBILE IMAGING TABLE OWNER’S MANUAL · The Allegro™ 6800 Mobile Imaging Table is designed to safely hold a patient in the proper position undergoing radiographic or

MIZUHOSI 2009 NW0510 Rev. D 61

Figure 27: Universal foot board with two side rail adaptors

Universal foot board (includes two side rail adaptors) (Part# 6900-51) The universal foot board mounts on the side rail or auxiliary clamps and is designed to provide support in maintaining the patient’s desired position in the table when the table is placed in reverse trendelenburg.