Agilent Vac Elut 12 and 20 Position Manifold Instructions The Agilent 12- and 20-position extraction manifold features: 12- or 20-place positions for luer tip SPE cartridges Simple tip design allows for easy use of stopcocks or disposable needle tips Vacuum gauge with control and release valves for monitoring and adjustment of vacuum level A choice of four racks to accommodate a variety of collection vessels Three-point interlocking system to ensure proper orientation of racks, keying cartridge position to corresponding collection vessel Chemically resistant, non-contaminating construction of glass, polyethylene, polypropylene, and Delrin components Individually controlled, needle-tipped valves (stopcocks) permit flow control of each position Identifying Parts Before assembling the extraction manifold, ensure that you have these parts, pictured in Figure 1. Glass chamber with exit valve installed White Delrin cover* with polyethylene gasket, vacuum gauge, control and release valves installed Disposable needle tips, long (polypropylene) Stopcock valves Luer plugs, male Ejector tool Rack, 12- or 20- position, as specified *The three posts on the underside of the cover serve as elevated feet so the cover can be set down without damaging the needle tips. They also maintain specific alignment of sample cartridge and collection vessel. Using the Extraction Manifold A typical solid phase extraction (SPE) procedure consists of the following steps: 1. Condition / equilibrate the SPE column bed. 2. Load the sample. 3. Wash off interferences. 4. Elute analyte(s) of interest. The Agilent extraction manifold permits these steps to be carried out on up to twenty cartridges in parallel. When sending all the eluents to waste, a rack is not necessary. This is usually the case for a typical SPE protocol when performing steps 1 through 3 above. In these cases waste is typically drawn into a liquid trap through the manifold’s exit valve located at the base of the glass chamber. When optimizing or troubleshooting an SPE protocol, the waste from these steps can be collected in tubes located in the racks for individual analysis. Figure 1. Vac Elut 20-position unit. Vacuum Release (white plastic ball) Luer Plugs Disposable Needle Tips (protruding through the cover and into the glass chamber) Vacuum Gauge Control Valve Alignment Pegs Glass Chamber Rack Exit Valve

Agilent Vac Elut 12 and 20 Position Manifold Instructions · PDF fileAgilent Vac Elut 12 and 20 Position Manifold Instructions ... The Agilent extraction manifold permits these steps

Embed Size (px)

Citation preview

Page 1: Agilent Vac Elut 12 and 20 Position Manifold Instructions · PDF fileAgilent Vac Elut 12 and 20 Position Manifold Instructions ... The Agilent extraction manifold permits these steps

Agilent Vac Elut 12 and 20 Position Manifold Instructions

The Agilent 12- and 20-position extraction manifold features:

• 12-or20-placepositionsforluertipSPEcartridges

• Simpletipdesignallowsforeasyuseofstopcocksordisposableneedletips

• Vacuumgaugewithcontrolandreleasevalvesformonitoringandadjustmentofvacuumlevel

• Achoiceoffourrackstoaccommodateavarietyofcollectionvessels

• Three-pointinterlockingsystemtoensureproperorientationofracks,keyingcartridgepositiontocorrespondingcollectionvessel

• Chemicallyresistant,non-contaminatingconstructionofglass,polyethylene,polypropylene,andDelrincomponents

• Individuallycontrolled,needle-tippedvalves(stopcocks)permitflowcontrolofeachposition

Identifying Parts

Beforeassemblingtheextractionmanifold,ensurethatyouhavetheseparts,picturedinFigure1.

• Glasschamberwithexitvalveinstalled

• WhiteDelrincover*withpolyethylenegasket,vacuumgauge,controlandreleasevalvesinstalled

• Disposableneedletips,long(polypropylene)

• Stopcockvalves

• Luerplugs,male

• Ejectortool

• Rack,12-or20-position,asspecified

*Thethreepostsontheundersideofthecoverserveaselevatedfeetsothecovercanbesetdownwithoutdamagingtheneedletips.Theyalsomaintainspecificalignmentofsamplecartridgeandcollectionvessel.

Using the Extraction Manifold

Atypicalsolidphaseextraction(SPE)procedureconsistsofthefollowingsteps:

1. Condition/equilibratetheSPEcolumnbed.

2. Loadthesample.

3. Washoffinterferences.

4. Eluteanalyte(s)ofinterest.

TheAgilentextractionmanifoldpermitsthesestepstobecarriedoutonuptotwentycartridgesinparallel.

Whensendingalltheeluentstowaste,arackisnotnecessary.ThisisusuallythecaseforatypicalSPEprotocolwhenperformingsteps1through3above.Inthesecaseswasteistypicallydrawnintoaliquidtrapthroughthemanifold’sexitvalvelocatedatthebaseoftheglasschamber.WhenoptimizingortroubleshootinganSPEprotocol,thewastefromthesestepscanbecollectedintubeslocatedintheracksforindividualanalysis.

Figure 1. Vac Elut 20-position unit.

VacuumRelease

(whiteplasticball)

LuerPlugs

DisposableNeedleTips

(protrudingthroughthecoverandintotheglasschamber)

VacuumGauge

ControlValve

AlignmentPegs

GlassChamber

Rack

ExitValve

Page 2: Agilent Vac Elut 12 and 20 Position Manifold Instructions · PDF fileAgilent Vac Elut 12 and 20 Position Manifold Instructions ... The Agilent extraction manifold permits these steps

2

Set up a vacuum system

1. Connecttheoutletsideofaliquidtraptoavacuumsourceusingvacuumtubing.

CAUTION: Makecertainthattheliquidtrapiscompatiblewith,andhassufficientcapacitytocollectallthewastethatwillbegeneratedinyourSPEprocedure.Ifthetrapcapacityistoosmall,wastemaycontaminateand/ordamagethevacuumpump.

2. Connectthemanifold’sexitvalveatthebaseoftheglasschambertothetrapinletusing3/8inidvacuumtubing.

3. Putthecovertopontheglasschamber.Donotinstallarackatthistimeunlesstheneedletipsandstopcocksrequiredforyourprotocolaresecurelymountedinthetopcover.

Installing needle tips and stopcock valves

InstallastopcockwiththevalvehandlepointingoutwardoradisposableneedletipineachofthetopcoverportsasdetailedinthisprocedureandasshowninFigure1.Thismakesthevalveaccessibleforoperation.Thetipsoftheneedlesandstopcockswillextendintocollectiontubesmountedinarackinsidethemanifold.

CAUTION: Youmustfollowthisproceduretoobtainpropersealingofcomponentsmountedinthetopcover.Failuretofollowthisprocedurewillcausejointleaksresultinginvacuumloss.

1. Ifinstalled,removetherackfromthemanifold.

2. Placethetopcoveronthemanifoldandbesureitisproperlyseatedontheedgeoftheglasschamber.

CAUTION: Damagetothetopcoverorrackmayresultiftherackisleftinthemanifoldandthethreealignmentpostsarerestingontherackwhenpressureisappliedtothetopcover.Iftherackmustbeleftinthemanifold,verifythetopcoverisrestingontheglasschamberandnotontherackbeforeproceeding.

3. Insertaneedletipintoahexagonalholeandlightlywiggletheneedlewhilepushingitdowninsidethehexagonalhole.Usingyourthumb,pushtheneedleasfarasitcango.Repeatthisstepfortherestoftheneedletips.Usethissamemethodtoinstallstopcocks.

4. Placeanejectorhandlehorizontallyontopoftwoneedlesandpressdownhardonit.Repeatthisprocedureonalloftheneedles.Alsousethisproceduretosealthestopcockjointtothetopcover.

5. Forneedletipsthatarenotinuse,installaluerplugbypushingitdownintheneedletip’sinlethole.Theluerplugisusedonlywithneedletips.DO NOT INSTALL LUER PLUGS DIRECTLY ON THE COVER!

Removing items from the top cover

1. Removeallliquidfromthecolumns.

2. Toremovealuerplugorcolumn,firsttwistittobreakthefrictionjoint.Whenthepartturnsfreely,liftitout.

CAUTION: Donotpullaplugorcolumnwithatightconnectionstraightupward.Thisforcecancausethejointbetweenthetopcoverandtheneedletiporstopcocktoloosen,resultinginalossofvacuum.

Remove or install a rack of collection tubes

1. Closetheexitvalveor,alternatively,turnoffthevacuumpump.

2. Breakthevacuuminthemanifoldchamberbyslightlyoffsettingthewhiteballfromitsseat(seeFigure2).

3. Removethecoverwhenthevacuumgaugereadsatmosphericpressure.

4. Installorremovearackoftubes.

5. Replacethetopcover.Ifarackisinstalled,carefullyalignthethreepostsonthebottomofthecoverwiththecorrespondingnotchesonthetopplateoftherack.

Manifold setup for protocol

1. Installtheneedletipsandstopcocksrequiredbytheprotocol.

2. Ifrequired,removeorinstallarackoftubes.

3. Putanextractioncartridgeintoeachvalvehubandseatitfirmly.

4. Closeallvalvesbyturningthehandlesuntileachindicatorshowsthatthevalveisintheclosedposition.

5. Setupthevacuumsystemandturnonthevacuumpump.Adjustthecheckvalvetosetthedesiredvacuumlevel.

6. FollowtheSPEprotocolrecommendedforthetypeofcolumnusedandtheapplication.Addfluidintothereservoirsbeforeopeningthestopcockstostartorresumeflow.

7. Openeachcolumn’sstopcockvalvetoevacuatetheliquidasrequiredbytheprotocol.

Figure 2. Breaking the vacuum.

Ballring,unseattobreakvacuum

3. Toremoveaneedletipand/orstopcockvalve,pryitoutcarefullywiththeejectortool.

CAUTION: Pushtheejectortooldownslowlytopreventthelaunchingoftheneedletip.

Page 3: Agilent Vac Elut 12 and 20 Position Manifold Instructions · PDF fileAgilent Vac Elut 12 and 20 Position Manifold Instructions ... The Agilent extraction manifold permits these steps

3

Note:Picturesshownareforillustrationpurposesonly

CORRECTLY SEATED NEEDLEShowing aligned serrations with corners of hexagonal hole

Needle serrated area

NEEDLE INSERTIONHexagon shaped hole on the cover

Figure 3.

Figure 4.

Figure 5.

Figure 6.

Page 4: Agilent Vac Elut 12 and 20 Position Manifold Instructions · PDF fileAgilent Vac Elut 12 and 20 Position Manifold Instructions ... The Agilent extraction manifold permits these steps

4

Ordering Information

Description Part No.

Standard Vac Elut 20 Manifold

VacElutmanifoldwithcollectionrackfor10x75mmtesttubes

12234105

VacElutmanifoldwithcollectionrackfor13x75mmtesttubes

12234100

VacElutmanifoldwithcollectionrackfor13x100mmtesttubes

12234101

VacElutmanifoldwithcollectionrackfor16x75mmtesttubes

12234102

VacElutmanifoldwithcollectionrackfor16x100mmtesttubes

12234103

Racks and replacement parts for Standard Vac Elut 20 Manifold

Standardglassbasin 12234505

20-portcollectionrackfor13x75mmtesttubes 12234507

20-portcollectionrackfor13x100mmtesttubes 12234508

20-portcollectionrackfor16x75mmtesttubes 12234509

20-portcollectionrackfor16x100mmtesttubes 12234510

Replacementlidgasket 12234502

VacElut20lidcover 12234501

Tall Vac Elut 20 Manifold

VacElut20manifoldwithtallglassbasinandcollectionrackfor16x150mmtesttubes,completesystem

12234104

20-portCollectionrackfor10x75mmtesttubes 12234517

Tallglassbasin 12234512

Description Part No.

Vac Elut 12 Manifold

12-portvacuummanifoldprocessingstationandcollectionrackfor16x100mmtesttubes

5982-9110

Racks and replacement parts for Vac Elut 12 Manifolds

VacElut12lidcover 5982-9111

VacElut12sealinggasket 5982-9112

VacElut12glasschamber 5982-9113

12-portcollectionrackfor13x75mmtesttubes 5982-9114

12-portcollectionrackfor13x100mmtesttubes 5982-9115

12-portcollectionrackfor16x75mmtesttubes 5982-9116

12-portcollectionrackfor16x100mmtesttubes 5982-9117

Description Unit Part No.

Common parts and disposables for all Manifolds

Manifoldneedletipejectortool 5982-9105

Polypropylenedeliveryneedles 25/pk 12234511

ManifoldSSneedlewithpolypropylenecoating

25/pk 12234042

Stainlesssteelneedletip 25/pk 12234038

Replacementexitvalveforglassbasin 12234506

Manifoldexitvalvereplacementkit 12234005

Vacuumgaugeassembly 12234504

Manifoldballring/vacuumquickrelease 12234013

Manifoldshortvalvestopcock 15/pk 12131005

Manifoldlongvalvestopcock 20/pk 12234520

Manifoldmaleluerplugs 25/pk 12234518

Page 5: Agilent Vac Elut 12 and 20 Position Manifold Instructions · PDF fileAgilent Vac Elut 12 and 20 Position Manifold Instructions ... The Agilent extraction manifold permits these steps

Learn more

www.agilent.com/chem/sampleprep

Buy online

www.agilent.com/chem/store

Find an Agilent office or authorized distributor

www.agilent.com/chem/contactus

U.S. and Canada

1-800-227-9770,[email protected]

Europe

[email protected]

Asia Pacific

[email protected]

India

[email protected]

Agilentshallnotbeliableforerrorscontainedhereinorforincidentalorconsequentialdamagesinconnectionwiththefurnishing,

performance,oruseofthismaterial.

Information,descriptions,andspecificationsinthispublicationaresubjecttochangewithoutnotice.

©AgilentTechnologies,Inc.,2012PublishedintheUSA,January24,2012

EditionCode04100-06/11PartNumber7704100

A portfolio of solutions designed to give you ultimate confidence in your results.

Agilentsamplepreparationproductsimprovethequalityofyoursamples,soyouimprovethequalityofyouranalysis.Fromsolidphaseextraction...toindustryleadinginstruments...toqualitycolumns,Agilenthasallthesolutionstogiveyougreaterconfidenceinyourresults.