11
IAEA/AFRA Training Course on Commissioning of Linear Accelerators used in Radiotherapy Algiers, Algeria 24-28 November 2007 Mehenna ARIB / SSDL Officer IAEA CRNA Workshop on Commissioning of Linear accelerators used in Radiotherapy TABLE OF CONTENTS Introduction Ionization chamber based dosimetry Determination of dose using calibrated chambers Beam quality specification Calibration of MV photon and electron beams Error and uncertainty analysis for ionization chambers IAEA CRNA Workshop on Commissioning of Linear accelerators used in Radiotherapy Modern radiotherapy relies on accurate dose delivery to the prescribed target volume. ICRU recommends an overall accuracy in tumour dose delivery of 5%, based on: An analysis of dose response data. An evaluation of errors in dose delivery in a clinical setting. Considering all uncertainties involved in the dose delivery to the patient, the 5% accuracy is by no means easy to attain. ± ± INTRODUCTION IAEA CRNA Workshop on Commissioning of Linear accelerators used in Radiotherapy Accurate dose delivery to the target with external photon or electron beams is governed by a chain consisting of the following main links: Basic output calibration of the beam Procedures for measuring the relative dose data. Equipment commissioning and quality assurance. Treatment planning Patient set-up on the treatment machine. INTRODUCTION IAEA CRNA Workshop on Commissioning of Linear accelerators used in Radiotherapy The basic output for a clinical beam is usually stated as: Dose rate for a point P in G/min or Gy/MU. At a reference depth z ref (often the depth of dose maximum z max ). In a water phantom for a nominal source to surface distance (SSD) or source to axis distance (SAD). At a reference field size on the phantom surface or the isocentre (usually 10x10 cm 2 ). INTRODUCTION IAEA CRNA Workshop on Commissioning of Linear accelerators used in Radiotherapy Machine basic output is usually given in: Gy/min for kilovoltage x-ray generators and teletherapy units. Gy/MU for clinical linear accelerators. This should be determined At a given distance from the source and For a given nominal collimator or applicator setting. INTRODUCTION

A_Dosimetry Calibration Reference Dosimetry

Embed Size (px)

DESCRIPTION

kalibrasi dosimetry

Citation preview

  • IAEA/AFRA Training Course on Commissioning of Linear Accelerators used in Radiotherapy

    Algiers, Algeria 24-28 November 2007

    Mehenna ARIB / SSDL OfficerIAEACRNA

    Workshop on Commissioning of Linear accelerators used in Radiotherapy

    TABLE OF CONTENTS

    IntroductionIonization chamber based dosimetryDetermination of dose using calibrated chambersBeam quality specificationCalibration of MV photon and electron beamsError and uncertainty analysis for ionization chambers

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Modern radiotherapy relies on accurate dose delivery to the prescribed target volume.

    ICRU recommends an overall accuracy in tumour dose delivery of 5%, based on: An analysis of dose response data. An evaluation of errors in dose delivery in a clinical setting.

    Considering all uncertainties involved in the dose delivery to the patient, the 5% accuracy is by no means easy to attain.

    INTRODUCTION

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Accurate dose delivery to the target with external photon or electron beams is governed by a chain consisting of the following main links: Basic output calibration of the beam Procedures for measuring the relative dose data. Equipment commissioning and quality assurance. Treatment planning Patient set-up on the treatment machine.

    INTRODUCTION

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    The basic output for a clinical beam is usually stated as: Dose rate for a point P in

    G/min or Gy/MU.

    At a reference depth zref(often the depth of dose maximum zmax).

    In a water phantom for a nominal source to surface distance (SSD) or source to axis distance (SAD).

    At a reference field size on the phantom surface or the isocentre (usually 10x10 cm2).

    INTRODUCTION

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Machine basic output is usually given in:

    Gy/min for kilovoltage x-ray generators and teletherapy units.

    Gy/MU for clinical linear accelerators.

    This should be determined At a given distance from the source

    and For a given nominal collimator or applicator setting.

    INTRODUCTION

  • IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    The basic output calibration for photon and electron beams is carried out with: Radiation dosimeters Special dosimetry techniques.

    Radiation dosimetry refers to a determination by measure-ment and/or calculation of Absorbed dose

    at a given point in the medium.

    INTRODUCTION

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Basic output calibration of a clinical radiation beam, by direct determination of dose or dose rate in water under specific reference conditions, is referred to as reference dosimetry.

    Three types of reference dosimetry technique are known: Calorimetry Fricke (chemical, ferrous sulfate) dosimetry Ionization chamber dosimetry

    INTRODUCTION

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Calorimetric dosimetry is the most fundamental of all reference dosimetry techniques, since it relies on basic definition of either electrical energy or temperature. In principle, calorimetric dosimetry (calorimetry) is simple. In practice, calorimetric dosimetry is very complex because of

    the need for measuring very small temperature differences.

    This complexity relegates the calorimetry to sophisticated standards laboratories.

    CalorimetryINTRODUCTION

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Main characteristics of calorimetic dosimetry: Energy imparted to matter by radiation produces an increase in

    temperature which is measured with thermocouples or thermistors.

    T

    CalorimetryINTRODUCTION

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    The following simple relationship holds:

    is the average dose in the sensitive volume is the thermal capacity of the sensitive volume is the thermal defect is the temperature increase

    Note:

    D = dE

    dm=

    CpT1

    D

    Cp T

    T(water, 1 Gy) = 2.4 104 K

    CalorimetryINTRODUCTION

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Ionizing radiation absorbed in certain media produces a chemical change in the media and the amount of this chemical change in the absorbing medium may be used as a measure of absorbed dose.

    The best known chemical radiation dosimeter is the Fricke dosimeter which relies on oxidation of ferrous ions into ferric ions in an irradiated ferrous sulfate FeSO4 solution.

    (Fe2+ ) (Fe3+ )

    Fricke (chemical) dosimetryINTRODUCTION

  • IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Average absorbed dose in a Fricke solution is given as:

    is the change in molar concentration of is the density of the Fricke solution. is the increase in optical density after irradiation. is the extinction coefficient. is the thickness of the solution. is the chemical yield of in mole/J.

    D = M

    G(Fe3+ )=

    (O.D.) l G(Fe3+ )

    = 278(O.D.)

    M

    (O.D.)

    l+3(Fe )G Fe3+

    Fe3+ .

    Fricke (chemical) dosimetryINTRODUCTION

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Recommended G-values in molecule/100 eV Photon beams (ICRU 14)

    Cs-137 15.32 MV 15.4Co-60 15.54 MV 15.55 MV to 10 MV 15.611 MV to 30 MV 15.7

    Electron beams (ICRU 35)1 MeV to 30 MeV 15.7

    Fricke (chemical) dosimetryINTRODUCTION

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Ionization chamber is the most practical and most widely used type of dosimeter for accurate measurement of machine output in radiotherapy.

    It may be used as an absolute or relative dosimeter.

    Its sensitive volume is usually filled with ambient air and: The dose related measured quantity is charge Q, The dose rate related measured quantity is current I,

    produced by radiation in the chamber sensitive volume.

    Ionization chamber dosimetryINTRODUCTION

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Measured charge Q and sensitive air mass mair are related to absorbed dose in air Dair by:

    is the mean energy required to produce an ion pair in air per unit charge e.

    Currently, the value of for dry air is 33.97 eV/ion pair or 33.97 J/C.

    airair

    air

    Q WDm e

    =

    air /W e

    air /W e

    Ionization chamber dosimetryINTRODUCTION

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    The subsequent conversion of the air cavity dose Dair to dose to medium (usually water) Dw is based on: Bragg-Gray cavity theory Spencer-Attix cavity theory

    INTRODUCTIONIonization chamber dosimetry

    Dmed = Dair smed,air

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Reference dosimetry with ionization chambers

    Clinical photon and electron beams are most commonly calibrated with ionization chambers that: Are used as relative dosimeters. Have calibration coefficients determined either in air or in water

    and are traceable to a national primary standards dosimetry laboratory (PSDL).

  • IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Traceability of chamber calibration coefficient to a national PSDL implies that:1. Either the chamber was calibrated directly at the PSDL in terms of:

    Air kerma in air Absorbed dose in water

    2. Or the chamber was calibrated directly at an accredited dosimetry calibration laboratory (ADCL) or at secondary standards dosimetry laboratory (SSDL) that traces its calibration to a PSDL.

    4. Or the chamber calibration coefficient was obtained through a cross-calibration with another ionization chamber, the calibration coefficient of which was measured directly at a PSDL, an ADCL oran SSDL.

    Reference dosimetry with ionization chambers

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Dosimetry protocols or codes of practice state the procedures to be followed when calibrating a clinical photon or electron beam.

    Choice of which protocol to use is left to individual radiotherapy departments or jurisdictions.

    Dosimetry protocols are generally issued by national, regional, or international organizations.

    Reference dosimetry with ionization chambers

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Reference dosimetry with ionization chambers

    19871987

    NK - based IAEA TRS 277

    update ofTRS-277

    IAEATRS-381

    19971997ND,w - based IAEA

    TRS 398

    2000200019971997

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    IONIZATION CHAMBER BASED DOSIMETRY SYSTEMS Ionization chambers

    Examples of typical ionization chambers:(a) Cylindrical chambers

    used for relative dosimetry.(b) Pinpoint mini-chamber

    and Co-60 buildup cap.(c) Farmer type cylindrical

    chamber and cobalt-60buildup cap.

    (d) Parallel-plate Roos typeelectron beam chamber.

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    USE OF CALIBRATED IONIZATION CHAMBERS Air kerma based protocols

    Air kerma based protocols use the air kerma in air cali-bration coefficient NK,Co obtained for a local reference ionization chamber in a cobalt-60 beam at a standards laboratory.

    Two steps are involved in air kerma based protocols for calibration of megavoltage photon and electron beams.

    The cavity air calibration coefficient ND,air is determined from the air kerma in air calibration coefficient NK,Co.

    Absorbed dose to water is determined using the Bragg-Gray relationship in conjunction with the chamber signal MQ and the cavity air calibration coefficient ND,air.

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    USE OF CALIBRATED IONIZATION CHAMBERS Air kerma based protocols

    Calibration in a cobalt-60 beam at standards laboratory:Absorbed dose to air in the cavity Dair,Co is determined from the total air kerma in air (Kair)air as follows:

    is the radiative fraction, i.e., the fraction of the total transferred energy expended in radiation interactions on slowing down of the secondary electrons in air.

    km corrects for the non-air equivalence of the chamber wall and buildup cap needed for an air kerma in air measurement.

    katt corrects for attenuation and scatter in the chamber wall. kcel corrects for non-air equivalence of chamber central electrode.

    air,Co air air m att cel( ) (1 ) D K g k k k=

    g

  • IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    USE OF CALIBRATED IONIZATION CHAMBERS Air kerma based protocols

    Under these special conditions, according to the B-G cavity theory, the dose to the medium Dmed is related to the dose to the cavity Dcav as:

    is the ratio of the average unrestricted mass collision stopping powers medium to cavity.

    The Spencer-Attix (S-A) cavity theory is more general and accounts for creation of secondary (delta) electrons. The dose to medium is given as:

    is the ratio of the average restricted mass collision stopping powers medium to cavity.

    Dmed = Dcav (smed.cav )

    (smed.cav )

    Dmed = Dcav (S / )med,cav

    med,cav( / )S

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    USE OF CALIBRATED IONIZATION CHAMBERS Air kerma based protocols

    With a known value of the cavity air calibration coefficient ND,air for a specific chamber, the chamber signal corrected for influence quantities MQ at a point in phantom allows determination of absorbed dose to water Dw,Q:

    is the ratio of average restricted collision stopping powers of water to air for a radiation beam of quality Q.

    pQ is the perturbation correction factor which corrects for effects that cause deviations from Bragg-Gray behaviour:

    w,Q air,Q w,air Q Q Q D,air w,air Q Q( ) ( ) D D s p M N s p= =

    w,air Q( )s

    w,Q Q K m att cel w,air Q dis wall cel cav Q (1 ) ( ) ( )D M N g k k k s p p p p=

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    USE OF CALIBRATED IONIZATION CHAMBERS Absorbed dose to water based protocols

    Calibration in a cobalt-60 beam at standards laboratory:

    Recent developments have provided support for a change in the quantity used to calibrate ionization chambers and provide calibration coefficients in terms of absorbed dose to water at beam quality Qo .

    At standards laboratory , absorbed dose to water at the reference depth zref in water for a reference beam Qo (usually cobalt-60) is known and used to determine the water dose calibration coefficient .

    Dw,Qo

    ND,w,Qo

    ND,w,Qo

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    DETERMINATION OFABSORBED DOSE TO WATER

    Using ND,w (TRS 298)

    at any other user quality Q:

    at the calibration quality Qo:

    Dw,Qo =MQo ND,w,Qo

    Dw,Q=MQ ND,w,Qo kQ,Qo

    beamqualityfactor

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Q,QoQw,D,QQw,p)airw,airD,QQw,

    p)air,wQ,airQw,

    Qair,D,Qair,D,

    celmatt(Q,K

    Q

    Q,airQ,air,D

    Q

    Q,wQ,w,D

    Q

    Qo,airQ,K

    Q,

    kNM=D(sNM=D

    s(D=D( )5N=N( )4

    kkk)g-1N=

    MD

    =N( )3

    MD

    =NM

    K=N( )2

    oQQ

    QQ

    o

    o

    o

    o

    o

    o

    o

    o

    o

    o

    ( )1 Q,wD oairK o

    6

    Steps in the determination of Dw at the reference point using the NK -ND,air formalism and the ND,w - formalism

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    kkQ,QoQ,Qo is given as a function of a quality index is given as a function of a quality index TPR20,10

    Need to experimentally determine Need to experimentally determine TPR20,10

    Beam Beam qquality uality sspecificationpecificationPhoton beamsPhoton beams

  • IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Beam Beam qquality uality sspecificationpecificationWhy TPR20,10 ?

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Beam Beam qquality factoruality factor

    photon beam quality (TPR20,10)0.55 0.60 0.65 0.70 0.75 0.80 0.85

    k Q

    0.93

    0.94

    0.95

    0.96

    0.97

    0.98

    0.99

    1.00

    1.01

    LPRI (Delaunay)Palm NPL (Sharpe)NRC soft (Seuntjens)(BEL) Palmans ENEA (Guerra)PTB (Shortt) Andreo et al (IAEA CoP 2000)NRC soft (Shortt) NRC soft (Ross)fit to all experimental dataNRC hard (Ross)

    NE 2571

    Co-60

    Andreo Phys Med Biol 45 (2000)

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Beam Beam qquality factoruality factor

    photon beam quality (TPR20,10)0.55 0.60 0.65 0.70 0.75 0.80 0.85

    k Q

    0.93

    0.94

    0.95

    0.96

    0.97

    0.98

    0.99

    1.00

    1.01

    Palm NRC soft (Seuntjens)Vatnitsky Andreo et al (IAEA CoP 2000)fit to all experimental data excl NRC high-E soft

    PTW 30001

    Andreo Phys Med Biol 45 (2000)

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Beam Beam qquality factoruality factor

    photon beam quality (TPR20,10)0.55 0.60 0.65 0.70 0.75 0.80 0.85

    k Q

    0.93

    0.94

    0.95

    0.96

    0.97

    0.98

    0.99

    1.00

    1.01

    PTB (Short) NRC soft (Ross)Palm NRC soft (Seuntjens) Vanitsky NRC soft (Shortt)ENEA (Guerra)Andreo et al (IAEA CoP 2000)NRC hard (Ross)fit to all experimental data

    PR-06C

    Andreo Phys Med Biol 45 (2000)

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Beam Beam qquality uality sspecificationpecification

    photon beams (High energy X-rays)TPR20,10 is given by a ratio of dose (ionization) at two

    depths, for constant SCD, (independent of electron contamination)

    Attenuation property

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Beam Quality: determination of TPR20,10

    10 g/cm2

    20 g/cm2

    SCD100 cm

    10M

    10

    20

    MM

    10

    201020 D

    DTPR =,

    20M

    10 cm x 10 cm

  • IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    SSD100 cm

    Beam Quality: determination of TPR20,10

    10M

    10

    20

    MM

    10

    201020 D

    DPDD =,

    20M

    20 g/cm2

    10 g/cm2

    20TPR

    = 1.2661,10

    1020PDD - 0.0595,

    10 cm x 10 cm

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    M

    MNN field

    refrefD,w

    fieldwD =,

    Cross-calibration of field ionization chambers

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    The beam quality index is the half-value depth in water R50. This is the depth in water (in g/cm2) at which the absorbed dose is 50% of its value at the absorbed dose maximum.

    SSD = 100 cm Field size at the phantom surface of at least 10 cm 10 cm for

    R50 7 g/cm2 and at least 20 cm 20 cm for R50 > 7 g/cm2

    The choice of R50 as the beam quality index is a change from the current practice of specifying beam quality in terms of the meanenergy at the phantom surface Eo.

    As Eo = f(R50), this change in beam quality index is merely a simplification which avoids the need for a conversion to energy.

    Beam Beam qquality uality sspecificationpecificationElectron beamsElectron beams

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Beam Beam qquality uality sspecificationpecificationElectron beamsElectron beams

    R50 = 1.029 R50,ion - 0.06 g/cm2 (R50,ion 10 g/cm2)

    R50 = 1.059 R50,ion - 0.37 g/cm2 (R50,ion 10 g/cm2)

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Beam Beam qquality uality sspecificationpecificationElectron beamsElectron beams

    This depth has been shown to reduce significantly the influence of spectral differences between different accelerators as well as that of electron and photon contamination in clinical electron beams.

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Electron beamsElectron beams

    Plane parallel chamber cylindrical chamber

    Beam Beam qquality factoruality factor

  • IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Cross Calibration of Ionisation Chambers

    Problem: the water absorbed dose has to be determined in a beam with R50 4 g/cm2.

    the Co-60 calibration factor of the plane-parallel chamber is not sufficiently reliable

    Solution: Cross-calibration of the user plane-parallel chamber by direct comparison against a cylindrical reference chamber calibrated in a Co-60 gamma radiation

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Plane parallel chamber

    Zref

    cylindrical chamber

    @ Zref

    refQQ

    xQwDx

    Q

    refQx

    QwD crosscross

    cross

    crosskN

    MM

    N00 ,,,,,

    =

    The highest energy electron beam available should be used; R50 > 7 g/cm2 (Eo 16 MeV) is recommended

    Cross Calibration of Ionisation Chambers

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    xQQ

    xQwD

    xQQw crosscross

    kNMD ,,,, =

    kQ,Qcross(R50) are determined from Table 19 (TRS398)

    xQQ

    xQQx

    QQ

    cross

    cross kk

    kint

    int

    ,

    ,, =

    Cross Calibration of Ionisation Chambers

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Experimental determination DExperimental determination Dw w (TRS 398)(TRS 398)Photon beamsPhoton beams

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Experimental determination DExperimental determination Dw w (TRS 398)(TRS 398)Photon beamsPhoton beams

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Influence quantity Reference value or reference characteristics

    REFERENCE CONDITIONS FOR DE-TERMINATION OF DW IN HIGH-ENERGY PHOTONS

    Phantom material water

    Chamber type cylindrical

    Measurement depth zref for TPR20,10 < 0.7, 10 g/cm2 (or 5 g/cm2)

    for TPR20,10 0.7, 10 g/cm2

    Reference point of chamber on central axis at centre of the cavity volume

    Position of referencepoint of chamber

    at the measurement depth zref

    SSD / SCD 100 cm

    Field size 10 cm x 10 cm

  • IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Reference conditions are the same as with the determination of R50

    Positioning the chamber at reference depth

    zref = 0.6 R50 0.1 g/cm2 (R50 in g/cm2)

    Plane parallel chamber

    Zref

    cylindrical chamber

    Note: zref may be deeper than zmax, cavity perturbation for cyl chamber may then be larger

    REFERENCE CONDITIONS FOR DETERMINATION OF DW IN HIGH-ENERGY ELECTRONS

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Determination of absorbed dose to water

    Absorbed dose at zmaxD(zmax) determined from PDD or TMR distribution

    max )(w refD D PDD Z= max )(w refD D TMR Z=

    SSD SAD

    Dw,Q (Zref) =MQ ND,w,Qo kQ,Qo

    ( ), , , ,cross crossx x x

    w Q ref Q D w Q Q QD Z M N k=

    PHOTONS

    ELECTRONS

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Corrections to the dosimeter reading, MQ

    ( )( )273.2273.2

    oTP

    o

    T PkT P+

    = +

    pol

    M Mk

    2M+ +=

    2

    2

    12

    2

    11

    +

    +=

    MMa

    MMaap os

    Experimental determination DExperimental determination Dw w (TRS 398)(TRS 398)

    Temperature and pressure

    Polarity effect

    Recombination (not all charge produced is collected)

    0

    0

    20 101.325

    T CP kPa= =

    M1 V1 , M2 V2IAEACRNA

    Workshop on Commissioning of Linear accelerators used in Radiotherapy

    PRACTICAL CONSIDERATIONS (1)

    0.994

    0.995

    0.996

    0.997

    0.998

    0.999

    1.000

    1.001

    0 5 10 15 20 25

    Time (min)

    Q/Q

    max

    WDIC70 # 141 WDIC70 # 039 PTW 30001 # 1245 NACP02 # 33.11 NE 2611 # 149

    IAEA TECDOC in press

    Stabilization time for ionization chambers

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    PRACTICAL CONSIDERATIONS (2)

    Stabilization time for a chamber and electrometer

    0.995

    0.997

    0.999

    1.001

    1.003

    1.005

    0 10 20 30 40 50 60

    Time (min)

    Q/Q

    max

    IAEA TECDOC 1455

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    PRACTICAL CONSIDERATIONS (3)

    Temperature measurements in air and in water

    In water ?

    In cavity chamber ? In s

    leeve

    ?

    Free in

    air ?

  • IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    PRACTICAL CONSIDERATIONS (3)

    Temperature measurements in air and in water

    IAEA TECDOC 1455

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    PRACTICAL CONSIDERATIONS (4)

    Phantom window bowing

    IAEA CUBIC PHANTOM

    0

    0.01

    0.02

    0.03

    0.04

    0.05

    0.06

    0.07

    -2 0 2 4 6 8 10 12 14Distance to the center of the window (cm)

    Def

    lect

    ion

    mm

    filling removing

    0

    0.5

    1

    1.5

    2

    2.5

    3

    -14 -9 -4 1 6Distance to the center of the window (cm)

    defle

    ctio

    n (m

    m)

    filling removing

    TG51 PHANTOM

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    PRACTICAL CONSIDERATIONS (5)

    Effect of waterproofing sleeve thickness

    Use the same sleeve used for chamber calibration

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Use of IAEA Worksheet (HE x-rays)

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Use of IAEA Worksheet (HE x-rays)

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Use of IAEA Worksheet (HE x-rays)

  • IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Use of IAEA Worksheet (HE x-rays)

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    These worksheets are available at:http://www-naweb.iaea.org/nahu/dmrp/codeofpractice.shtm

    Use of IAEA Worksheet (HE electrons)

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Estimated uncertainty in Dw (HE X-rays) (Q0 = 60Co)Physical quantity or procedure Rel. std uncertainty (%)

    Step 1: Standards Laboratory ND,w calibration of secondary standard at PSDL 0.5Long term stability of secondary standard 0.1ND,w calibration of the user dosimeter at the standard laboratory 0.4Combined uncertainty of Step 1 0.6

    Step 2: User high-energy photon beamLong-term stability of user dosimeter 0.3Establishment of reference conditions 0.4Dosimeter reading MQ relative to beam monitor 0.6Correction for influence quantities ki 0.4Beam quality correction MQ (calculated values) 1.0Combined uncertainty of Step 2 1.4

    Combined standard uncertainty of Dw,Q (steps 1+2) 1.5

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Estimated uncertainty in Dw (HE electrons) (Q0 = 60Co)

    Physical quantity or procedure User chamber type :

    Relative stand. uncertainty (%) Cyl PP

    Step 1: Standards laboratory

    ND,w calibration of secondary standard at PSDL 0.5 0.5

    Long term stability of secondary standard 0.1 0.1 ND,w calibration of user dosimeter at SSDL 0.4 0.4

    Combined uncertainty of step 1b 0.6 0.6

    Step 2: User electron beamLong term stability of user dosimeter 0.3 0.4 Establishment of reference conditions 0.4 0.6

    Dosimeter reading MQ relative to beam monitor 0.6 0.6

    Correction for influence quantities ki 0.4 0.5 Beam quality correction kQ (calculated values) 1.2 1.7

    Combined uncertainty of step 2 1.5 2.0

    Combined standard uncertainty of Dw,Q (steps 1+2) 1.6 2.1

    IAEACRNAWorkshop on Commissioning of Linear accelerators used in Radiotherapy

    Estimated uncertainty in Dw (HE electrons) (Q0 = electrons)

    Relative standard uncertainty (%)Physical quantity or procedure User chamber type: Cyl PP Step 1: PSDL

    ND,w calibration of user dosimeter at PSDL 0.7 0.7

    Combined uncertainty in step 1 0.7 0.7

    Step 2: User electron beam

    Long term stability of user dosimeter 0.3 0.4

    Establishment of reference conditions 0.4 0.6

    Dosimeter reading MQ relative to beam monitor 0.6 0.6

    Correction for influence quantities ki 0.4 0.5

    Beam quality correction kQ,Qo (calculated values) 0.9 0.6

    Combined uncertainty in step 2 1.3 1.2

    Combined standard uncertainty of Dw,Q (steps 1+2) 1.4 1.4

    IAEACRNA

    THANK YOU FOR YOUR ATTENTION

    Workshop on Commissioning of Linear accelerators used in Radiotherapy