Upload
duongthuy
View
218
Download
3
Embed Size (px)
Citation preview
Introduction HJ Method 679-2013 describes the determination of acrolein, acrylonitrile, and acetonitrile from soil and sediments by headspace-gas chromatography. It has similarities to EPA methods 8015C (nonhalongenated organics), 8030A (acrolein and acrylonitrile by gas chromatography), and 8031 (acrylonitrile by gas chromatography). Found in the soil as a byproduct of synthetic processes or from previous pesticide use, these three analytes can quickly degrade and evaporate into the air.
ExperimentalThe HJ-679 method specifies a column with an internal diameter of 530 µm, which makes it well suited for an Agilent 7890 GC equipped with an FID and Agilent 7697 headspace sample. The HJ method was followed as closely as possible. Five headspace vials were made at each calibration level by filling each vial with ~2 g of quartz sand, adding 10 mL of a matrix modifier (180 g of NaCl in 500 mL of H2O), and spiking varying amounts of a 2,000 ppm standard to achieve the required levels. Acrolein, acrylonitrile, and acetonitrile were spiked with 2 µg, 5 µg, 20 µg, 40 µg, and 80 µg.
ACROLEIN, ACRYLONITRILE, AND ACETONITRILE BY HS-GC
Demonstration of an HJ Method with an Agilent 7890 GC
For more information, visit:www.agilent.com
Results and DiscussionThe calibration curves determined for acrolein, acrylonitrile, and acetonitrile were found to be within method requirements (correlation coefficient of ≥0.995). The response factor was calculated over the calibration curve range, and an RSD was determined for each analyte. Acrolein was found to have the highest response factor RSD, at 10 %, likely due to its high volatility. Both acrylonitrile and acetonitrile yielded response factor RSDs of 5.3 %.
Instrument conditionsParameter Value
Agilent 7890 GCInlet 150 °C, Split 5:1
Column Agilent J&W DB-WAX UI, 30 m × 0.53 mm, 1.00 µm, 125-7032UI
Column flow 4.5 mL/min
Oven 40 °C (5 minutes) then 5 °C/min to 60 °C then 30 °C/min to 150 °C (5 minutes)
FID 250 °C
Agilent 7697 Headspace SamplerOven 75 °C
Loop 105 °C
Transfer line 150 °C
Vial equilibration 30 minutes
Injection duration 0.1 minutes
Vial 20 mL
Shaking On, level 1
Vial fill flow 50 mL/min
Vial fill pressure 8 psi
Vial pressure equilibration time 2 minutes
Loop fill ramp rate 20 psi/min
Final loop pressure 1.2 psi
Loop equilibration 0.2 minutes
0 4 8 12 16 20 24 28 32 36 40 44 48 52 56 60 64 68 72 76 80 840
0.10.20.30.40.50.60.70.80.91.0 Acrolein
y = 12005.609609*x + 5790.506646R2 = 9967
×106
Resp
onse
Concentration (ng/mL)
Figure 1. The calibration curve for acrolein from 2 µg to 80 µg gave a correlation coefficient of 0.997.
×106
Acrylonitrile
Concentration (µg/L)-4 0 4 8 12 16 20 24 28 32 36 40 44 48 52 56 60 64 68 72 76 80 84
Resp
onse
00.40.81.21.62.02.42.83.2
y = 39797.225910*x – 10068.903978R2 = 0.9983
Figure 2. Calibration curve for acrylonitrile gave a correlation coefficient of 0.998 for five levels, ranging from 2 µg to 80 µg.
×105
Acetonitrile
Concentration (ng/mL)
Resp
onse
012345 y = 6896.947165*x – 1637.622097
R2 = 0.9978
-4 0 4 8 12 16 20 24 28 32 36 40 44 48 52 56 60 64 68 72 76 80 84
Figure 3. Acetonitrile gave a correlation coefficient of 0.998 for five calibration levels, ranging from 2 µg to 80 µg.
1 2 3 4 5 6 7 8 9 10120,000121,000122,000123,000124,000125,000126,000127,000128,000129,000130,000131,000132,000133,000134,000135,000136,000
Time (min)
Resp
onse
1
2
3
Figure 4. Acrolein (1), acrylonitrile (2), and acetonitrile (3) at 1 mg/kg eluted in under 10 minutes.
www.agilent.comInformation, descriptions and specifications in this
publication are subject to change without notice.
© Agilent Technologies, Inc. 2017 Published in USA, November 1, 2017
5991-8096EN
ConclusionThe Agilent 7890 GC configured with an Agilent 7697 headspace sampler achieves the performance specifications of HJ679-2013. Correlation coefficients were found to be 0.995 or better, while response factor RSDs are not specified in the method. Acrolein, acrylonitrile, and acetonitrile yielded RSDs of 10 % or better for five replicate vials. The 10-minute run time is typical of HJ679-2013, but could be shortened easily by implementing a smaller id column.