Upload
yoshihiko-hayakawa
View
217
Download
1
Embed Size (px)
Citation preview
Techn ica l R e p o r t
Absorbed Doses Reduced by the Use of Rare-Earth
Intensifying Screens in Rotational Panoramic
Radiography with Constant Potential Generator
Yoshihiko H A Y A K A W A , B.Sc., M.Sc., Nor io K O B A Y A S H I , R.T.,
Yuji KOUSUGE, R.T., Hisao F U J I M O R I , R.T.,
Kouichi H A R A D A , D.D.S. and Kinya K U R O Y A N A G I , D.D.S., Ph.D.
Department of Oral and Maxillofacial Radiology, Tokyo Dental CoIlege, Chiba, Japan
(Received : March 1, 1993, Revision received : May 7, 1993, Accepted : May 14, 1993)
Key Words : Radiography,dental ; Radiography,panoramic ; Radiation dosage ;Screen-film
Absorbed doses modified by the use of rare-earth intensifying screens were measured with rotational
panoramic radiography. Equipments with the constant potential, Veraview (J.Morita) and PM 2002 CC
(Planmeca) was utilized. The optimum exposure settings at each tube voltage were determined by
radiography of a Rando | phantom. Absorbed doses to parotid and thyroid glands were measured under
the optimum exposure settings by using the Rando | phantom and thermoluminescence dosimeters.
The sensitivity of the rare-earth intesifying screen system increased as the tube voltage became
higher. Absorbed doses decreased steeply as the tube voltage became higher. The highly sensitive
system of the LANEX REGULAR and T-Mat G (Eastman Kodak) reduced the thyroid gland dose to a
half at 70 kV in comparison with the regular system of the PX-III (Kasei Optonics) and X-Omat S
(Eastman Kodak). This high sensitivity, however, made it impossible to set the optimum tube current
at above 70 kV. The another rare-earth system, the TRIMAX 16 and TRIMAX XUD (3M) had a lower
sensitivity than the regular system of the PX-III and X-Omat S at below 80 kV. But they had equal
sensitivity at the highest tube voltage of 80 kV.
I n t r o d u c t i o n
T h e use of r a r e - e a r t h in tens i fy ing
screens and o r thochromat i c films is an
effect ive method for dose reduct ion in
ro ta t iona l panoramic rad iography ~-1~ How-
Oral Radiol. Vol,9 No.1 1993(49~55)
ever, some problems have been ra ised on this
application.
The sensi t iv i ty of ra re -ea r th intensifying
screens falls down at low x- ray energy 2'1').
Keur 2) repor ted tha t its sensi t ivi ty at 60 kVp
49
was lower than t ha t of CaWO~ intensifying
screens and regu la r - type films. Therefore ,
r a r e -ea r th screens should be used with h igher
tube vo l tage for dose reduct ion.
High sens i t iv i ty requi res a reduced x - r a y
output. However , the e q u i p m e n t used by
Keur 2> or Hur lbu r t and Coggins 4) had a fixed
tube current which made it imposs ib le to
reduce x - r ay output when the higher x - r a y
energy was used .The mer i t of high sens i t iv i ty
to the higher x - r a y energy 12) could not be put
to good use. Tynda l l and W a s h b u r n 5) or
Ponce et al. 9''~ a t t e m p t e d to resolve this
p rob lem by using a th ick added filter. Ponce
et a12 ) reduced absorbed doses by 60 percent
when a thin foil of r a r e - e a r t h sc reen or alumi-
num of 4 m m th ickness was added. The
n a r r o w beam width which was ob ta ined by
the thinner sli t is the o ther me thod for output
reduct ion ~,7~. T h i s me thod s imul taneous ly
p roduced a th i cke r image layer .
Fur ther , the a l t e r a t i on of i m a g e qua l i ty
is also a p rob lem TM. But in observer perfor-
mance examina t i ons a'4'~'8-1~ resu l ted tha t the
r a r e - ea r t h screen produced the d iagnos t ic
i m a g e qual i ty compa t ib l e wi th the CAW04
screens. McDavid et al. 8> desc r ibed tha t the
high sens i t iv i ty of the imag ing sys t em did not
d i r ec t ly re la ted to the low sharpness .
On the o ther hand, the use of a cons t an t
po ten t ia l of x - r a y gene ra to r was thought to
be the other m e t h o d for dose reduction13L
This m a k e s i t poss ib le to genera te a h igher
energy of x-ray.
The purpose of this s tudy was to ob ta in
the op t imum e x p o s u r e se t t ings when rare-
e a r t h i n t e n s i f y i n g s c r e e n s and o r tho -
ch roma t i c films were used for ro t a t i ona l
p a n o r a m i c r a d i o g r a p h y with cons t an t poten-
tial. The m e a s u r e m e n t of absorbed doses to
both pa ro t id and thy ro id glands was perfor-
med. I t was inves t iga ted how the abso rbed
dose changed by depending on the exposu re
se t t ings and wha t ex t en t of dose reduct ion
was caused by using r a r e ea r th screens.
Materials and Methods
X-ray equipment, intensifying screen and film
T w o x - r ay mach ines for ro t a t iona l pano-
r amic r ad iography , Verav iew (J. Mor i t a
Corp., Kyoto, Japan) and PM 2002 CC (Plan-
meca Oy, Hels inki , Finland) were used. The
specif icat ions a r e l is ted in T a b l e 1.
T w o types of r a r e - ea r th intensifying
screens and o r t h o c h r o m a t i c films combina-
tion, such as the L A N E X R E G U L A R screen
and T - M a t G film (Eas tman K o d a k Company,
Rochester , N e w York, U.S.A.), and the
Table 1 Specifications for Veraview and PM 2002 CC
X-ray generator Tube voltage
Tube current
Total filtration Exposure time Focal spot Focus-film distance
Veraview PM 2002 CC
constant potential 60-80 kV
(5 kV intervals) 5-10 mA
(1 mA intervals) 2.1 mmA1
15 sec 0.5 x 0.5 mm
510 mm
constant potential 60-80 kV
(2 kV intervals) 4-12 mA
(1 mA intervals) 2.5 mmAl 3.1-18 sec*
0.6 x 0.6 mm 480 mm
*Exposure time is set at 18 sec in this study.
50
TRIMAX 16 screen and T R I M A X XUD film
( 3M company, Germany) were used. These
rare-earth intensifying screens were com-
posed of Gd202S:Tb. The combination of PX-
III (Kasei Optonix Corp., Kanagawa, Japan)
and X-Omat S film (Eastman Kodak) as a
CaWO4 intensifying screen and regular-type
film was used. All films were processed by an
automatic image processor New RN (Fuji
Medical System Corp., Tokyo, Japan) at the
processing t ime of 210 sec and a temperature
of 30~ according to the manufacturer ' s
recommended method.
Exposure settings An operator can choose the settings of
both the tube voltage and current for the
optimum exposure. The optimum tube cur-
rent at each tube voltage was determined by
radiography of a Rando | anthropomorphic
phantom (Alderson Research Laboratories,
Stamford, CN, USA). A Rando | phantom
which was sectioned horizontally at 2.54 em
intervals was used. Sections 6 and 7 corre-
spond to the upper and lower jaws. The four
radiographs obtained by different tube cur-
rents in turn were examined by four
observers, two oral radiologists and two
radiological technologists. The best radio-
graph in four with every tube voltage was
determined by them.
Dosimetry Absorbed doses to the parotid and thy-
roid glands were measured with the optimum
exposure settings of tube voltage and current.
The absorbed doses to the parotid gland and
thyroid gland were measured by the use of
fifty thermoluminescence dosimeters (TLD-
100, Harshaw Chemical Co.,Solon, OH, USA)
in the form of LiF rods 1 • 1• 6 mm.
Since the method of thermolumineseent
dosimetry which was utilized in this study
was the same as was described in a previous
article 14~, the method in detail is abbreviated
in this article. Four TLDs were placed at the
appropriate sites within the sections between
6 and 10. Two were placed in both sides of
section 6 at the location of the parotid gland,
and two were used for the thyroid gland in
section 10. Each dosimeter received ten x-ray
exposures. Each measurement was repeated
three times. The results of the absorbed dose
per single exposure are presented.
Results
The selected opt imum exposure settings
with every tube voltage are indicated in
Figure 1. Observers more than three in four
agreed with all opt imum exposure settings.
The alteration in the sensitivity made a
difference in the optimum setting. The opti-
mum tube current decreased as the tube
voltage increased at all three combinations
examined.
The absorbed doses to the parotid and
thyroid glaads at each optimum setting are
indicated in Figures 2 and 3. Absorbed doses
obtained from using the orthochromatic sys-
tems decreased steeply as the tube voltage
became higher. The highly sensitive ortho-
chromatic system of the LANEX REGULAR
intensifying screen and T-Mat G film reduced
the thyroid gland dose to a half at 70 kV in
comparison with the regular system of the
PX-III intensifying screen and the X-Omat S
film. This high sensitivity, however, made it
impossible to set the optimum tube current at
the tube voltage above 70 kV. The sensitivity
of the or thochromat ic sys tem of the
T R I M A X 16 intensi fying screen and
TRIMAX XUD film was lower than the sys-
tem of the PX-III and X-Omat S at below 80
kV. But, they had equal sensitivity at the
highest tube voltage of 80 kV.
5/
Fig. 1 Opt imum e x p o s u r e set t ing of tube cur- rent with every tribe voltage. (A : Ver- aview, B : PM 2002 ~ 10
CO) ~ 9 �9 : LANEX REGU- --
c 8 LAR and T-Mat G �9 : T R I M A X 16 and k 7
D T RI M AX XUD C) 6 �9 : PX-III and X-
_o 5 O r n a t S
I--
A
Fig. ~ Parot id gland dose A (mean+_S.D.) at each
o p t i m u m e x p o s u r e 1600 set t ing wi th every tube voltage. >, (A : Veraview, B : PM (.9 2002 CC) ~ 1 2 0 0 �9 : LANEX REGU- r LAR and T-Mat G o
�9 : T R I M A X 16 and -o 8 0 0 T R I M A X XUD e-
ra �9 : PX-[II and X- Omat S -o
�9 . = 400 s s
Fig. 3 Thyroid gland dose (mean_+S.D.) at each o p t i m u m e x p o s u r e 40 set t ing wi th every ~., tube voltage. C~ (A : Veraview, B : PM :::1.
v 30 2002 CC) r �9 : LANEX REGU- tn
o LAR and T Mat G
�9 : T R I M A X 16 and "o 2 0 c T R I M A X XUD r
�9 : PX-III and X- Omat S "o
�9 1o (-
I.-.-
0
1 I I I I 6O 70 8O
12 11 lC 9 8' 71
f ,1 I I, I . I I I I I I I 60 70 80
Tube Voltage (kV)
B
1600
1200
800
400
I I I . 60 70
I 0 " I i I I I I I I I I I 80 60 70 80
Tube Voltage ( kV )
I I I I I 6 0 70 8 0
40
30
20
10
0 ,, I I I t I I 1 I I I I 60 70 80
52 Tube Voltage ( kV )
Discuss ion
Optimum exposure settings and absorbed
doses
The optimum exposure settings were
determined as the tube current with every
tube voltage. The high sensitivity of rare-
earth intensifying screens and ortho-
chromatic films need a reduced x-ray output.
The higher sensitivity of the LANEX REGU-
LAR screen and T-Mat G film made it impos-
sible to have an adequate image quality
above 70 kV, because of the excess x-ray
output even if the lowest tube current was
used. The merit of the high sensitivity to the
higher x-ray energy 12) could not be put to
good use. Accordingly, the reduction of x-ray
output by some methods, such as the setting
of lower tube current, use of narrower slit
width 1'7~ or installation of thicker filter 5,9,1~
was needed.
Parotid and thyroid glands doses when
using rare-earth intensifying screens and
orthochromatic films decreased steeply as the
tube voltage became higher. On the other
hand, doses from PX-III CaWO4 intensifying
screens and X-Omat S regular-type films
showed to be nearly constant, although the
tube voltage changed. The rare-earth screen
is more sensitive to the higher x-ray energy in
comparison with the CaWO4 screen, because
the high energy x-ray above the k-shell
absorption energy of rare-earth metal is
absorbed efficiently by the rare-earth screen.
The use of the high energy x-ray with the
rare-earth screen is effective for dose reduc-
tion. The constant potential also generate a
highly effective energy of x-ray. The combi-
nation of rare-earth screens and constant
potential generator result in dose reduction.
Parotid gland and thyroid gland doses
Absorbed doses measured in this study
are compared with previous studies 1~-23).
Some researchers gave attention to absorbed
doses to the parotid and thyroid glands which
had stochastic radiological risks.
The parotid gland doses were much
lower than the doses reported by previous
studies which were shown in Table 2. The
parotid gland doses measured in this study
were distributed in the range of 0.4-1.6 mGy
which are thought to be comparatively low.
This decrease of the parotid gland doses was
caused by the difference with the position of
the rotational axis of the x-ray beam. The
rotational axis at the start and end of projec-
tion in the PM 2002 CC x-ray machine were
17 cm apart from each other and located
outside and behind the mandibular ramus 24).
Namely, the position of the lateral rotational
axis where the absorbed dose increased was
not close to the parotid gland in this equip-
ment.
The thyroid gland doses which were
reported previously were shown in Table 3.
The thyroid gland doses measured in this
study were distributed in the range of 0.01-
0.04 mGy which are similar to previously
reported absorbed doses.
When rare-earth intensifying screens
were used and the tube voltage became
higher, the parotid gland dose showed a stee-
Table 2 The parotid gland doses in the literatures and in this study
Authors Parotid gland dose (mGy)
Stenstr6m et al. TM
Stenstrtim et al. 21) Manson-Hing and Greer ~6) Wall et al. ~71 Bartolotta et aI. ~9) Nilsson et al. TM
Underhill et al. 22) this study
0.30-1.42 0.26-0.36 0.07-1.9 0.45-2.8 0.84-4.17 1.0-3.2 0.1-1.2 0.4-1.6
53
Table 3 The thyroid doses in the literatures and in this study
Authors Thyroid gland dose (mGy)
Antoku et al. TM
Manson-Hing and Greer 1~ WalIet al. In
Stenstr0m et al. TM
Bartolotta et al. TM
Nilsson et al. TM
Stenstr0m et al. TM
Underhill et al. TM
Kassebaum et al. TM
this study
0.01-2.69
0.04-0.51
0.01-0.30
0.03-0.09
0.13-0.37
0.01-0.09
0.03
0.03-0.06
0.05
0.01-0.04
per d e c r e a s e t h a n the t h y r o i d g land . T h i s
was caused by the s c a t t e r i n g r a d i a t i o n w h i c h
was e f f ic ien t ly de l ive red by the h ighe r t ube
vo l t age . T h e r a t e of s c a t t e r i n g r a d i a t i o n
which i n c r e a s e d the abso rbed doses was sup-
posed to be h igher in the t h y r o i d g l a n d t h a n in
the p a r o t i d gland.
S t o c h a s t i c r i sk f r o m r a d i a t i o n e x p o s u r e
a r e in p r o p o r t i o n to e f fec t ive dose. E f f e c t i v e
dose is g iven by the s u m of t he p r o d u c t of t h e
dose e q u i v a l e n t to each t i s sue and i ts we igh t -
ing f ac to r . A b s o r b e d doses to t he t h y r o i d
g l a n d and s a l i v a r y g lands inc lud ing p a r o t i d
g l a n d occup ied the m o s t of e f f e c t i v e dose
w i t h r o t a t i o n a l p a n o r a m i c radiography25,2%
C o n c l u s i o n
W e r e a c h e d the fo l l owing conc lu s ions "
1 . T h e sens i t iv i ty of the r a r e - e a r t h in tes i fy-
ing s c r een s y s t e m irwin:eased as t he t u b e
v o l t a g e b e c a m e higher. A b s o r b e d doses
to the pa ro t i d and t h y r o i d g l ands de-
c r e a s e d s t eep ly as the t u b e v o l t a g e
b e c a m e higher .
2 . T h e h igh sens i t ive s y s t e m of t he L A N E X
R E G U L A R and T - M a t G r e d u c e d the
t h y r o i d g l and dose to a ha l f a t 70 k V in
c o m p a r i s o n wi th the r e g u l a r s y s t e m of
t he PX-III and X - O m a t S . T h i s h igh
sens i t iv i ty , h o w e v e r , m a d e i t imposs ib le
to set t he o p t i m u m tube c u r r e n t a t a b o v e
70kV.
�9 T h e r a r e - e a r t h s y s t e m of t he T R I M A X
16 and T R I M A X X U D had the l o w e r
s ens i t i v i t y t h a n t h e r e g u l a r s y s t e m of the
PX-III and X - O m a t S a t b e l o w 80 kV.
But t h e y had the equa l s ens i t i v i t y a t the
h ighes t t ube v o l t a g e 80 kV.
References 1 ) F0rsgren,L. and Julin,P. : Radiation dose reduction in
panoramic radiography : Orthopantomograph Model OP 3 modified for rare earth intensifying screens. Swed. Dent. f 6 : 225-231, 1982
2 ) Keur,J.J. : A rare earth screen-film system for dental panoramic radiography. Australian Dental Journal 28 : 105-108, 1983
3 )Gratt, B.M., White, S.C., Packard, F.L. and Petersson, A.R. : An evaluation of rare-earth imaging systems in panoramic radiography. Oral Surg. 58 : 475-482, 1984
4 ) Hurlburt, C.E. and Coggins, L.J. : Rare earth screens for panoramic radiography Oral Surg. 57 : 451-454, 1984
5) Tyndall,D.A. and Washburn,D.B. : Rare-earth filters in panoramic radiography:A means of reducing patient dose without compromising image quality Dentomaxillofac. Radiol. 15 : 19-25, 1986
6 ) D'Ambrosio,J.A., Schiff,T.G., McDavid,W.D. and Lan- gland,O.E. : Diagnostic quality versus patient expo- sure with five panoramic screen-film combinations. Oral Surg. 61 : 409 411, 1986
7 ) Aagaard,A. and Sewerin I. : Reduction of body doses in rotational panoramic radiography by means of reduced beam width in combination with rare earth intensifying screens. Scand. J. dent. Res. 94 : 530-535, 1986
8 ) McDavid,W.D., Morris,C.R., Tronje,G. and Welander, U. : Resolution of several screen-film combinations in rotational panoramic radiography. Oral Surg. 61 : 629-634, 1986 Ponce,A.Z., McDavid,W.D., Lundeen,R.C. and Morris, C.R. : Adaptation of the Panorex II for use with rare earth screen-film combinations. Oral Surg. 61 : 645-648, 1986 Ponce,A.Z., McDavid,W.D., Lundeen,R.C. and Morris, C.R. : Kodak T-Mat G film in rotational panoramic radiography. Oral Surg. 61 : 649-652, 1986 Reynolds,J., Skucas,J. and Gorski,J. : An evaluation of screen-film speed characteristics. Radiology 118 : 711- 713, 1976
9)
lO)
11)
5 4
12) Thunthy,K.H., Boozer,C.H. and Weinberg,R.: Sen-
sitometric evaluation of rare earth intensifying screen
systems. Oral Surg. 59 : 102-106, 1985
13) Forest,D., Deschamps,M., Normandeau,L. : A compar-
ative study of radiation doses received from constant
direct current and conventional self rectified pano-
ramic dental x-ray equipment, jr. Dent. Que. 18 : 9-17,
1981
14) Hayakawa, Y., Sakoh, T., Fujimori, H. and Kur-
oyanagi, K. : Dose reduction to the thyroid gland in
intraoral source radiography. Dentomaxillofac. Radiol. 22 : 21-24, 1993
15) Antoku,S., Kihara,T., RusselI,W.J. and Beach,D.R. :
Doses to critical organs from dental radiography.
Oral Surg. 41 : 251-60, 1976
16) Manson-Hing,L.R., Greer,D.F. : Radiation exposure
and distribution measurements for three panoramic
x-ray machines. Oral Surg. 44 : 313-321, 1977
17) WalI,B.F., Fisher,E.S., Paynter,R., Hudson,A. and
Bird,P.D. : Doses to patients from pantomographic
and conventional dental radiography. Brit. J. Radiol. 52 : 727-34, 1979
18) Stenstr0m, B., Julin, P. Richter, S. : Comparison
between panoramic radiographic techniques: Part III:
Radiation absorbed doses with Status-X |
Orthopantomograph| Model OPS, and Panelete |
Dentomaxillofac. Radiol. i1 : 107-116, 1982
19) Bartolotta, A., Calenda, E., Caliccbia, A. and In-
dovina, P.L. : Dental orthopantomography: survey of
patient dose. Radiology 146 : 821-823, 1983
20) Nilsson, L., Rohlin, M. and Tapper, K. : Exposure
distribution, absorbed doses, and energy imparted for
panoramic radiography using Orthopanotomograph
model OP 5. Oral Surg. 59 : 212-219, 1985
21) Stenstr0m, B., Julin, P. and Karlsson,L. : Comparison
between panoramic radiographic techniques: Part IV:
Absorbed doses and energy imparted from the Orth-
opantomograph, model OP 10. Dentomaxillofac. Radiol. 16 : 11-15, 1987
22) Underhill,T.E, Chilvarquer,I., Kimura,K., Langlais,R.
P., McDavid, W.D., Preece, J.W. and Barnwell, G. :
Radiobiologic risk estimation from dental radiology.
Part 1. Absorbed doses to critical organs. Oral Surg. 66 : 111-120, 1988
23) Kassbaum,D.K., Stoller,N.E., McDavid,W.D., Goshorn,
B. and Ahrens,C.R. : Absorbed dose determination for
tomographic implant assessment techniques. Oral Surg. 73 : 502-509, 1992
24) Yamamoto, K., Hayakawa, Y. and Sakoh,T. : Reduc-
tion of redundant shadows using new rotational pano-
ramic radiograph "PM 2002 CC". Bull. Tokyo Dent. Coll. 30 : 175-184, 1989
25) Wall, B.F. and Kendall, G.M. : Collective doses and
risks from dental radiology in Great Britain. Brit. J. Radiol. 56 : 511-516, 1983
26) Stenstr~m, B., Henrikson ,C.O., Karlsson, L. and
Sarby,B. : Effective dose equivalent from intraoral
radiography. Swed. Denl. jr. 10 : 95-101, 1986
Reprint requests to : Yoshihiko HAYAKAWA, B. SC., M, Sc. Department of 0ral and Maxillofacial Radiology,
Tokyo Dental College i-2-2 Masago, Mihama-ku, Chiba 261, Japan
55