1. Potentiometric measuring principles
2. Amperometric measuring principles
3. Optical measuring principles
4. User-defined corrections
5. Performance specifications
6. Parameters
7. Solutions and gas mixtures
Index
Date of Issue
Reference manual
ABL800 FLEX
System performance
The procedures described in this manual must be observed in order to ensure proper system performance, and to avoid hazards.
Radiometer cannot provide or verify system performance characteristics if the system is not installed, used and maintained in accordance with Radiometer procedures or if accessories not meeting the specifications provided by Radiometer are used.
Radiometer warrants that the data media on which the software included in the system is furnished is free from defects in material and workmanship under normal use for three (3) months from the date of delivery as evidenced by a copy of invoice or receipt.
Third-party software and trademarks
The ABL800 FLEX analyzers comprise the Microsoft® Windows®XP Embedded, VxWorks®, Sybase®, SQL Anywhere®, and Radiometer-developed software.
By using the system, you accept the terms of the Software License Agreement(s) of the provider(s) of the above software as shown in the End User License Agreement(s) in the analyzer start up picture and to the terms of the Microsoft® Windows®XP Embedded End-User Agreement included in this manual. If you cannot accept the terms of the Software License Agreement(s), you should not use the system, but immediately contact your provider for a return of the system and a refund of the purchase price.
Microsoft® and Windows® are trademarks of Microsoft Corporation.
VxWorks® is a registered trademark of WindRiver Systems Incorporated.
Sybase® SQL Anywhere® is a registered trademark of Sybase Incorporated.
Warranties and disclaimer
Radiometer makes no warranties, express or implied, other than expressly stated.
Any warranties expressly stated in this document are conditional upon the system being installed, used and maintained in accordance with Radiometer procedures, including that only accessories meeting the specifications provided by Radiometer are used.
Radiometer disclaims any liability for system performance if the system is not installed, used and maintained in accordance with Radiometer procedures or if accessories not meeting the specifications provided by Radiometer are used.
Further, Radiometer disclaims any liability for loss of data and direct, consequential or other damages, including loss of profit or loss of business, whether such claim for damages is based upon contract, negligence or tort (including strict liability), and even if Radiometer has knowledge of the possibility of the potential damage or loss.
Confidentiality
The contents of this document shall not be reproduced or communicated to any third party without the prior written consent of Radiometer.
Changes
This document is subject to change without notice and you are urged to contact Radiometer to verify whether the document has been changed.
While every effort is made to ensure the correctness of the information provided in this document as changed from time to time, Radiometer disclaims any liability for errors and omissions.
Radiometer, the Radiometer logo, ABL, AQT, TCM, RADIANCE, PICO and CLINITUBES are trademarks of Radiometer Medical ApS.
© 2008 Radiometer Medical ApS. All rights reserved.
Contents
TUContentsUT ................................................................................................................1 1. Potentiometric measuring principles ............................................................. 1-1
Overview........................................................................................................... 1-1 General information .......................................................................................... 1-2 Reference electrode........................................................................................... 1-8 pH electrode ...................................................................................................... 1-9 pCOB2 B electrode ................................................................................................ 1-15 Electrolyte electrodes...................................................................................... 1-23 References....................................................................................................... 1-37
2. Amperometric measuring principles ............................................................. 2-1 Overview........................................................................................................... 2-1 General information .......................................................................................... 2-2 pOB2 B electrode..................................................................................................... 2-4 Glucose and Lactate electrodes....................................................................... 2-13 Crea electrodes................................................................................................ 2-23 References....................................................................................................... 2-31
3. Optical measuring principles.......................................................................... 3-1 Overview........................................................................................................... 3-1 Optical system................................................................................................... 3-2 Correcting for interferences .............................................................................. 3-7 Measurement and corrections ........................................................................... 3-9 References....................................................................................................... 3-15
4. User-defined corrections ................................................................................. 4-1 Overview........................................................................................................... 4-1 General information .......................................................................................... 4-2 Correction factors for oximetry parameters and bilirubin................................. 4-4 Electrolyte and metabolite parameters .............................................................. 4-7
5. Performance characteristics ........................................................................... 5-1 Overview........................................................................................................... 5-1 Definition of terms............................................................................................ 5-2 ABL8x0/8x5 Performance characteristics .................................................... 5-5 Overview........................................................................................................... 5-5 Test conditions .................................................................................................. 5-6 Performance test results – chart description...................................................... 5-7 Performance test results – pH ......................................................................... 5-10
Contents ABL800 FLEX Reference Manual
Performance test results – pCOB2 B ..................................................................... 5-12 Performance test results – pOB2 B ........................................................................ 5-15 Performance test results – cKP
+P ........................................................................ 5-18
Performance test results – cNaP
+P ...................................................................... 5-20
Performance test results – cCl P
–P ....................................................................... 5-22
Performance test results – cCaP
2+P ..................................................................... 5-24
Performance test results – cGlu ...................................................................... 5-26 Performance test results – cLac ...................................................................... 5-28 Performance test results – ctHb ...................................................................... 5-30 Performance test results – oximetry................................................................ 5-32 Performance test results – bilirubin ................................................................ 5-42 Additional information about FLEXMODE................................................... 5-48 ABL8x7 Performance characteristics ......................................................... 5-49 Overview......................................................................................................... 5-49 Test conditions ................................................................................................ 5-50 Performance test results – pH, pCOB2 B, pOB2 B ...................................................... 5-51 Performance test results – electrolytes............................................................ 5-54 Performance test results – cGlu, cLac............................................................. 5-58 Performance test results – ctHb ...................................................................... 5-60 Performance test results – oximetry................................................................ 5-61 Performance test results – bilirubin ................................................................ 5-70 Performance test conditions and results – cCrea............................................. 5-72 Interference tests ............................................................................................. 5-92 References..................................................................................................... 5-103
6. Parameters ....................................................................................................... 6-1 Overview........................................................................................................... 6-1 General information .......................................................................................... 6-2 Measured parameters ........................................................................................ 6-5 Input parameters.............................................................................................. 6-14 Derived parameters ......................................................................................... 6-17 Units of derived parameters ............................................................................ 6-22 List of equations.............................................................................................. 6-28 Oxyhemoglobin dissociation curve (ODC)..................................................... 6-44 Conversion of units ......................................................................................... 6-49 Default values ................................................................................................. 6-51 Altitude correction .......................................................................................... 6-52 References....................................................................................................... 6-53
ABL800 FLEX Reference Manual Contents
7. Solutions and gas mixtures ............................................................................. 7-1 Overview........................................................................................................... 7-1 General information .......................................................................................... 7-2 Calibration solutions ......................................................................................... 7-3 Rinse and Cleaning solutions............................................................................ 7-6 Electrolyte solutions ......................................................................................... 7-8 S5362 Hypochlorite Solution.......................................................................... 7-10 Gas mixtures (Gas 1 and Gas 2)...................................................................... 7-11 Traceability certificates................................................................................... 7-12
TIndex
TDate of Issue
Contents ABL800 FLEX Reference Manual
TTWarnings/Cautions
Throughout the manual, the descriptions may contain operational precautions and warnings.
Notice Definition
TWARNING T Warning alerts users to potential serious outcomes to themselves or the patient (such as death, injury, or serious adverse events).
TPRECAUTION T Precaution alerts users to exercise special care necessary for the safe and effective use of the device. Precaution may include actions to be taken to avoid effects on patients or users that may not be potentially life threatening or result in serious injury, but about which the user should be aware. Precaution may also alert users to adverse effects on the device by use or misuse, and the care necessary to avoid such effects.
TNOTET Notes give practical information.
In this manual a distinction between a warning and a caution is not made. Any notice that alerts the user to possible dangers of any kind is given the title TWARNING/CAUTIONT.
Definitions
TWARNING/ CAUTION T
1. Potentiometric measuring principles
Overview
This chapter describes the potentiometric measuring principles and the pH, pCOB2 B and electrolyte electrodes that are based on this principle.
This chapter contains the following topics.
General information ......................................................................................... X1-2X
Reference electrode.......................................................................................... X1-8X
pH electrode ..................................................................................................... X1-9X
pCOB2 B electrode ................................................................................................. X1-15X
Electrolyte electrodes ....................................................................................... X1-23X
References ........................................................................................................ X1-37X
Introduction
Contents
1. Potentiometric measuring principles ABL800 FLEX reference manual
1-2
General information
The potential of an electrode chain is recorded using a voltmeter, and related to the concentration of the sample (the Nernst equation).
An electrode chain describes an electrical circuit consisting of a sample, electrode, reference electrode, voltmeter, membranes and electrolyte solutions.
Every element in the electrode chain contributes a voltage to the total potential drop through the chain. Thus:
• When immersed in the appropriate electrolyte solution, both electrodes have separate potentials
• The membrane junctions between the sample and electrolyte solutions also have separate potentials
The potentiometric measuring principle is applied to pH, TpTCOB2 B and electrolyte electrodes.
The complete electrode chain potential therefore is the sum of these separate potentials and is the quantity measured by the voltmeter.
TE BtotalB = EBsampleB – EBRef
where the final unknown potential ( TE BsampleTB) can be calculated knowing the total electrode chain potential (TE BtotalTB) and the reference potential (TE BRefTB is constant between two subsequent calibrations).
Having measured the unknown potential (TE BsampleTB), the Nernst equation is then applied to determine the activity (Ta TBxB) of the species under study:
E E Tn
asample = +02 3R
Flog .
x
where: E0 = standard electrode potential R = gas constant (8.3143 Joule × KP
−1 P× molP
−1P)
TT T = absolute temperature (310 K (37 P
oPC ))
Continued on next page
Potentiometric method
SampleElectrolyte
solutionElectrolyte
solutionReferenceelectrode Electrode
V
Membrane Membrane
Voltmeter
Nernst equation
ABL800 FLEX reference manual 1. Potentiometric measuring principles
1-3
General information, TContinuedT
Tn T = charge on the ion F = Faraday constant (96487 coulomb × molP
−1P)
xa = activity of TxT
The Nernst equation is rearranged to express the activity as a function of the potential TE BsampleTB. Having measured TE Bsample B Tthe activity can be calculated since all other quantities are already known. Finally the analyzer converts activity to concentration.
Strictly speaking, the potential of an electrode chain or the magnitude of current flowing through an electrical chain is related to the activity of a substance, and not its concentration.
Activity expresses the "effective concentration" of a species, taking non-ideality of the medium into account.
Activity and concentration are related by the following equation:
Ta TBxB = γ TcTBx B
where: Ta TBxB = the activity of the species x
γ = the activity coefficient of species x under the measurement conditions (for ideal systems γ = 1)
TcTBx B = the concentration of species (mmol/L)
TNOTE: TTTo be exact, activity is related to the molality of species x, i.e., the number of mmoles per kg of solvent. However, molality is converted to concentration (molarity). T
The analyzer automatically converts activities into concentrations [1]. The term concentration is therefore used in explanations of the measuring principles for each of the electrodes further on in this chapter.
The potentiometric measuring principle is applied in the pH, pCOB2B and electrolyte electrodes. It is slightly different for the Tp TCOB2 B electrode, however, since the Nernst equation is not directly applied.
Calibration is an analytical process defining the functional relationship between the obtained readings or analytical responses and the concentration or other quantities present in the calibration material (liquid or gas). Thus, a calibrating solution or a gas mixture (for pCOB2B calibrations) is drawn into the measuring chamber and the analyzer adjusts itself to measure the known value of the liquid or gas.
Continued on next page
Nernst equation T(continued)
Calibration
1. Potentiometric measuring principles ABL800 FLEX reference manual
1-4
General information, TContinuedT
The electrodes are active elements and must be calibrated regularly. Signals from the electrodes change because of, e.g., protein build-up, worn-out membranes, aging electrodes, etc. The responses from the electrodes when measuring on the calibrating solutions are checked to ensure that the amplified signals from the electrodes are converted to accurate values for an unknown sample. The relationship between the electrode amplifiers’ output and the pH/pCOB2 B /electrolyte electrodes are simple mathematical functions. Calibration data can therefore be determined by relating the electrode signals during the calibration process to the values of the calibrating solutions.
The calibration line expresses the relationship between the potential measured at an electrode, and the concentration of the species specific to the electrode. The calibration line forms the basis of the scale used by the analyzer to convert electrode chain potentials to concentrations. Each electrode has a different calibration line.
The pH electrode is used as an example to illustrate how the calibration line is derived from two calibration solutions with known pH.
The calibration solutions give the following two points: T− T64 mV at pH 6.802 (Cal 2) and T− T100 mV at pH 7.398 (Cal 1)
Within the coverage range 6.300 to 8.000 the pH electrode is linear, and the relationship between potential and pH is linear, so a line can be drawn between the two points, as shown below:
Calibration line
pH
Measured potential (mV)
−100
−64
6.802 pH of Cal 2 sol.
7.398 pH of Cal 1 sol.
−97
7.346pH of sample
This is a 2-point calibration. In 1-point calibration, only the position of the calibration line is determined. The slope of the calibration line is maintained from the last 2-point calibration.
The calibration line is stored in the computer and is used during measurement to convert the potential measured at the pH electrode during sample analysis to an actual pH value.
Continued on next page
Calibration T(continued)T
Calibration line
ABL800 FLEX reference manual 1. Potentiometric measuring principles
1-5
General information, TContinuedT
To describe the actual condition of the electrode, its calibration line is compared to the calibration line of the theoretical electrode.
Theoretical calibration line
pH
Measuredpotential(mV)
−112.4
−75.5
6.8 7.4
The theoretical electrode is defined to measure the following: −112.4 mV at pH 7.400, −75.5 mV at pH 6.800.
The position and slope of the calibration line compared to the theoretical calibration line are described by the status and sensitivity.
The electrode sensitivity illustrates the slope of the calibration line compared to the slope of the theoretical electrode.
The sensitivity of the theoretical electrode is 100 % or 1.00.
Theoretical calibration lineSlope= −61.5 mV/pHSensitivity = 100 %
pH
Measuredpotential(mV)
−112.4
−75.5
6.8 7.4
2-point calibration lineSlope = −58.4 mV/pHSensitivity = 95 %
If an electrode has a sensitivity of 95 % or 0.95, its sensitivity is 5 % lower than the theoretical electrode.
The sensitivity of an electrode is calculated as:
(%))802.6398.7(5.61
398.7802.6−×
−=
atPotentialatPotentialySensitivit
where 61.5 = sensitivity of theoretical electrode.
Each electrode has its own sensitivity limits.
Continued on next page
Calibration line T(continued)T
Sensitivity
1. Potentiometric measuring principles ABL800 FLEX reference manual
1-6
General information, TContinuedT
Status reflects the deviation from the theoretical electrode at pH 7.400 and, therefore, indicates the position of the calibration line.
Theoretical calibration line drawn through the 1-point calibration point.
pH
Measured potential (mV)
ECal 1
pHCal 1
Theoretical calibration line for the theoretical pH electrode with known potential of −112.4 mV at a pH = 7.400.
ECal 1 nom
ECal 1 nom(theo) E=−112.4 mV
pHCal 1 nom(7.400)
pHStatus
ΔE
ΔpH
A calibration line with the same slope as the theoretical calibration line (−61.5 mV/ pH) is drawn through this point.
The calibration line for the actual electrode deviates from that of the theoretical electrode. The status value describes this deviation.
Status of the actual pH electrode at pH 7.400 is calculated as:
1005.61400.7.400.7.400.7
×−
−=atelectrltheoreticaofPotentialatpotentialMeasStatus
Each electrode has its own status limits.
Drift of an electrode is a measure of stability obtained by comparing the last accepted calibration with the previous calibration.
The following drift values are used:
• Drift 1 – obtained on Cal 1 and/or Gas 1
• Drift 2 – obtained after a 2-point calibration
The obtained drift values should not exceed the calibration drift tolerances. The drift tolerances can be changed in the Setup program, but Radiometer recommends using the default drift tolerances. Too narrow drift tolerances will cause electrode drift errors even for normal electrode fluctuations. If the drift tolerances are made wider, no warning will be given if the electrodes should become unstable. Significant measurement errors could result.
Continued on next page
Status
Drift
ABL800 FLEX reference manual 1. Potentiometric measuring principles
1-7
General information, TContinuedT
The following calibration materials are used:
TCalibration material T TUsed for...T
Calibration Solutions 1 and 2: the exact composition of the calibration solutions is given in the barcode on the bottle label, which can be read into the analyzer using the barcode reader, or entered manually via the keyboard.
Calibration of the pH, and electrolyte electrodes
Gas 1 and Gas 2: each gas has a precise composition essential for determining the accuracy of the analyzer in each pCOB2 B measurement; the exact composition of the calibration solutions is given in the barcode on the bottle label, which can be read into the analyzer using the barcode reader, or entered manually via the keyboard.
Calibration of the pCOB2 B electrode
The Chemical Reference Laboratory at Radiometer is responsible for the accuracy of the calibrating solutions. Traceability certificates for the individual solutions are enclosed in chapter 7: TSolutions and Gas MixturesT.
The measuring time of the electrode is independent of the electrode type. Electrode signals are registered at 0.982-second intervals during both calibrations and measurements. The registration of each electrode signal begins after the samples, calibration solutions and calibration gases are in position in the measuring modules. The duration of each calibration is predetermined, as is the number of updatings of the electrodes’ signals.
In general, the updatings from an electrode response are numbered from 1 to Tupd. last T, where updating number 1 is the first updating and Tupd.lastT is the last. The diagram below schematically illustrates the electrode response that is calculated on uncorrected electrode updating values.
Calibration materials
Measuring time
Updatings
Signal
Updatings
Upd. 1 Upd. last
1. Potentiometric measuring principles ABL800 FLEX reference manual
1-8
Reference electrode
The reference electrode is used in the measurement of pH and electrolyte parameters and is located in the pH/Blood Gas module.
The reference electrode maintains a stable, fixed potential against which other potential differences can be measured. The potential is not altered by sample composition.
A fixed potential is maintained at the reference electrode by the following equilibrium reactions:
TAgCl ⇔ Ag P
+P + ClP
−TP
TAg P
+P + eP
−P ⇔ Ag T
These reactions are possible because the electrode is made from a Ag rod coated with AgCl to provide the Ag/Ag P
+P equilibrium and determine the reference
potential.
The electrolyte solution acts as a salt-bridge solution that maintains an electrical contact between the coated Ag wire and the sample. The solution is 4 M sodium formate (HCOONa), adjusted to pH 5.5 with hydrochloric acid.
The chloride concentration in the electrolyte solution is adjusted in accordance with the chloride concentration in the Rinse Solution, to reduce Cl P
−P exchange across the
membrane, thereby obtaining a more stable potential.
The electrode is encased in the electrode jacket: The rubber ring seals the electrode in the jacket to prevent evaporation or leakage of the electrolyte solution.
The membrane consists of three separate membranes:
Membrane Function
Inner To limit diffusion through the membrane and stabilize the whole membrane system.
Middle To prevent protein interference.
Outer To reduce the interchange of sample or Rinse Solution and HCOONa solution.
The E1001 reference electrode comes in a box with an insert explaining the preparation of the electrode and its use.
Description
Packaging
Electrode contact
Electrolyte
Ag rod coated with AgCl
3-layer membrane
Electrode jacket
ABL800 FLEX reference manual 1. Potentiometric measuring principles
1-9
pH electrode
The pH electrode (E777) is a pH-sensitive glass electrode. The pH-sensitive glass membrane is located at the tip and seals the inner buffer solution with a constant and known pH.
The air bubble allows for expansion of the inner buffer solution when the electrode is thermostatted to 37 P
oPC.
The potential difference across the glass membrane is due to a change in the charge balance at the membrane.
The glass membrane is sensitive to HP
+P ions. The metal ions in the glass
are exchanged with protons on either side of the membrane, from the inner buffer solution on one side and the sample on the other.
A difference in the ion exchange on either side of the membrane occurs if the HP
+P
concentration (and therefore pH) is unequal on both sides. The number of positive and negative ions is no longer equal, so the potential difference across the membrane changes. If the HP
+P concentrations on either side of the membrane are
equal, the potential difference will theoretically be 0 mV.
The theoretical sensitivity of the pH electrode at 37 P
oPC being equal to −61.5 mV
per pH unit, using pH = −log [HP
+P], and converting concentration to activity, the
Nernst equation can be expressed as:
E E 615 pH mVsample 0= − ×.
The following symbols are used:
−61.5 mV/pH = Theoretical sensitivity of the pH electrode at 37 P
oPC
E(pH,Cal2) = Potential of the pH electrode chain from a calibration measurement on Cal 2 solution
E(pH,Cal1) = Potential of the pH electrode chain from a calibration measurement on Cal 1 solution
E B0B(pH,Cal1) = Standard potential of the pH electrode chain with a nominal pH = 7.4 (the approximate pH of Cal 1 solution)
Continued on next page
Description
Nernst equation
Designation
Electrode contact
Electrode
Inner buffer solution
Glass membrane
1. Potentiometric measuring principles ABL800 FLEX reference manual
1-10
pH electrode, TContinuedT
pH(Cal1,nom) = Nominal pH of Cal 1 solution (pH = 7.4)
pH(Cal1) = pH of Cal 1 solution
E(pH,Cal1prev) = Potential of the pH electrode chain from the previous calibration measurement on Cal 1 solution
Sens(pH,prev) fraction
= Sensitivity of the pH electrode from the previous 2-point calibration
pH(Cal1,prev) = pH of Cal 1 solution in the previous calibration measurement
pH(Cal2) = pH of Cal 2 solution
Sens(pH) = Relative sensitivity of the pH electrode chain.
The sensitivity of the pH electrode (SensBpHB) is obtained from the calibration line obtained from a 2-point calibration on Calibration Solutions 1 and 2 (Cal 1 and Cal 2), and is calculated from the following equation:
[ ]pH(Cal1)pH(Cal2)61.5Cal1)E(pH,Cal2)E(pH,Sens(pH)
−×−−
= (fraction)
The sensitivity of the pH electrode should fall between 0.92-1.03 or 92-103 %.
The status of the pH electrode is calculated from the following equation:
pH(Cal1)nom)pH(Cal1, 261.5-
Cal1)(pH,ECal1)E(pH,Status(pH) 0 −+
−=
The status of the pH electrode should fall between a pH of 6.7 and 8.1.
Drift 1 is calculated from the following equation:
[ ]prev)pH(Cal1,pH(Cal1)prev)Sens(pH,61.5-
Cal1prev)E(pH,Cal1)E(pH,1(pH)Drift −−×
−=
TNOTE: TTUnder normal circumstances, T pH(Cal1) − pH(Cal1,prev) = 0. THowever, in instances where the Cal 1 solution container has been replaced between two consecutive calibrations, TpH(Cal1) − pH(Cal1,prev) ≠ 0T. TT
The default drift tolerances set by Radiometer for Drift 1 are ± 0.020.
Drift 2 is calculated from the following equation:
[ ]Drift 2(pH) E(pH,Cal2) E(pH,Cal1prev)- 61.5 Sens(pH, prev)
pH(Cal2) pH(Cal1, prev)=−
×− −
The default drift tolerances set by Radiometer for Drift 2 are ± 0.020.
Continued on next page
Designation T(continued)
Sensitivity
Status
Drift
ABL800 FLEX reference manual 1. Potentiometric measuring principles
1-11
pH electrode, TContinuedT
The sample pH is calculated as follows:
pH(sample) = E(pH,sample) E(pH,Cal1)61.5 Sens(pH)
pH(Cal1)−− ×
+
The measured pH value is then corrected for systematic deviations from the reference method using the following equation:
TEquation A: T
pH(sample,corr.) = AB0 B × pH(sample) + AB1 B
where:
pH(sample) = uncorrected pH value of the sample
pH(sample,corr.) = corrected pH value of the sample.
AB0 B = instrument-dependent correction factor
AB1 B = instrument-dependent cut-off
TEquation A+:
When an additional correction is needed, equation A is first used together with the constants for the macromode S250 and FLEXMODE (195 and 165 μL, no message) mode. Then the obtained results are put back into equation A as pH(sample) and then treated again, using the constants for the specific sample handling to obtain the corrected value.
Corrections are as follows:
TABL8xx FLEX T
TModeT TAB0 TB TAB1 TB TEquationT
37/27/17 S250 1.000 –0.009 A
C125 1.000 –0.009 A
FLEXMODE (message 903) 1.000 –0.009 A
FLEXMODE (message 902) 1.000 –0.009 A
FLEXMODE (message 901) 1.000 –0.009 A
FLEXMODE (message 900) 1.000 –0.009 A
FLEXMODE (message 899) 1.000 –0.009 A
FLEXMODE (message 898) 1.000 –0.009 A
Continued on next page
Measurement
Corrections
1. Potentiometric measuring principles ABL800 FLEX reference manual
1-12
pH electrode, TContinuedT
ABL8xx FLEX
Mode AB0 B AB1 B Equation
35/25/15 S195 0.9964 0.0150 A
S95 0.9964 0.0150 A
S85 0.9964 0.0150 A
C95 1.007 −0.053 A+
C55 1.025 −0.1880 A+
FLEXMODE (no message) 0,9964 0.0150 A
FLEXMODE (message 897) 1.007 −0.0530 A+
FLEXMODE (message 895) 1.007 −0.0530 A+
FLEXMODE (message 894) 1.007 −0.0530 A+
FLEXMODE (message 873) 1.007 −0.0530 A+
FLEXMODE (message 872) 1.0216 −0.1639 A+
FLEXMODE (message 871) 1.025 −0.1880 A+
FLEXMODE (message 870) 1.030 −0.216 A+
FLEXMODE (message 869) 1.030 −0.216 A+
30/20/10 S85 0,9964 0.0150 A
10 BG only
C55 1.025 −0.1880 A+
FLEX FLEXMODE (no message) 1.0006 −0.0035 A+
FLEXMODE (message 872) 1.0209 −0.1575 A+
FLEXMODE (message 871) 1.025 −0.1880 A+
FLEXMODE (message 870) 1.030 −0.216 A+
FLEXMODE (message 869) 1.030 −0.216 A+
Continued on next page
Corrections T(continued)
ABL800 FLEX reference manual 1. Potentiometric measuring principles
1-13
pH electrode, TContinuedT
ABL8xx FLEX
Mode AB0 B AB1 B Equation
05 S165 0,9964 0.0150 A
S95 0,9964 0.0150 A
S85 0,9964 0.0150 A
C95 1.007 −0.053 A+
C55 1.025 −0.1880 A+
FLEXMODE (no message) 0,9964 0.0150 A
FLEXMODE (message 897) 1.007 −0.053 A+
FLEXMODE (message 895) 1.007 −0.053 A+
FLEXMODE (message 894) 1.007 −0.053 A+
FLEXMODE (message 873) 1.007 −0.053 A+
FLEXMODE (message 872) 1.0216 −0.1639 A+
FLEXMODE (message 871) 1.025 −0.1880 A+
FLEXMODE (message 870) 1.030 −0.216 A+
FLEXMODE (message 869) 1.030 −0.216 A+
00 BASIC S195 0.9964 0.0150 A
S95 0.9964 0.0150 A
S85 0.9964 0.0150 A
C95 1.007 −0.053 A+
C55 1.025 −0.1880 A+
Continued on next page
Corrections T(continued)
1. Potentiometric measuring principles ABL800 FLEX reference manual
1-14
pH electrode, TContinuedT
The following stability criterion must be met to obtain a stable electrode response during 1- and 2-point TcalibrationsT:
pH(limit)i).updpH(sample,last).updpH(sample, ≤−
The following stability criterion must be met to obtain a stable electrode response during TmeasurementT:
pH(limit)i).updpH(sample,last).updpH(sample, ≤−
where:
pH(sample,upd.last) = pH value from the last updating with a measurement on calibration solution or sample. (The last updating is number 30.)
pH(sample,upd.i) = pH value for a given updating with a measurement on calibration solution or sample. (The relationship must be fulfilled for at least one of the updating numbers 20 or 21.)
pH(limit) = pH limiting value for the stability criterion (0.005).
Stability criteria
ABL800 FLEX reference manual 1. Potentiometric measuring principles
1-15
TpTCOB2 B electrode
The pCOB2 B electrode (E788) is a combined pH and Ag/AgCl reference electrode mounted in a plastic jacket, which is filled with a bicarbonate electrolyte.
The jacket is covered by a 20-μm silicone membrane moulded on a 50-μm nylon net. The net both reinforces the silicone membrane and serves as a spacer in order to trap a layer of the electrolyte between the membrane and the glass tip of the electrode. The electrolyte also contains glycerol to prevent collection of air bubbles in the electrode jacket, thus improving electrode stability.
The membrane allows any uncharged molecules of COB2 B, OB2B and NB2B to pass through it. Charged ions such as HP
+P will not pass. Consequently, dissolved COB2 B from the
sample will diffuse into the thin layer of bicarbonate electrolyte until the equilibrium is reached.
This produces carbonic acid: HB2 BO + COB2 B ⇔ HB2BCOB3
Carbonic acid dissociates according to the following equilibrium reaction: H CO H H CO2 3 2 3⇔ ++ −
The release of H P
+P ions changes the HP
+P concentration, and therefore the pH of the
solution on one side of the pH-sensitive glass membrane.
The concentration gradient of HP
+P ions on the other side of the membrane affects
the potential difference across the glass membrane. This change in potential across the glass membrane is measured by the voltmeter.
The Nernst equation is used to convert the potential reading into a pH value: mV)(pH5.610glass ×−=EE
where:
TE TBglassB = potential difference across the glass membrane
TE TB0B = standard electrode potential
61.5 mV/pH = theoretical sensitivity of the pH electrode at 37 P
oPC
Continued on next page
Description
Nernst equation
Electrolyte
Ag/AgCl reference band
Electrode jacket
Electrode contact
Membrane
1. Potentiometric measuring principles ABL800 FLEX reference manual
1-16
TpTCOB2 B electrode, TContinuedT
The pH value is related to the partial pressure of COB2 B in the sample by the following equation:
2CO2
-3
a COHCO
log +pK = pHα×p
c
where: pKa = −log KBa B, the equilibrium constant for the dissociation of carbonic acid in
water
α CO2= solubility coefficient for COB2 B in water
The bicarbonate concentration [ ]HCO3- is so large compared to [ ]H+ that it can be
considered constant. At constant temperatures α CO2 is also constant. So the
equation can be simplified to: 2CO log-K' = pH p where:
K' is a constant incorporating the equilibrium constant for carbonic acid (K Ba B), the bicarbonate concentration and the solubility coefficient α CO2
.
2
3
COHCOH −+ ×
=cc
K a is the equilibrium constant for carbonic acid.
pCOB2 B of the sample is then calculated from the equation above.
The following symbols are used: Tp TCOB2 B(Gas1), Tp TCOB2 B(Gas2)
= Pressure of COB2 B in Gas 1 or Gas 2, respectively
TFTTCOB2 B(Gas1), TTFTTCOB2 B(Gas2) T
T=T TFraction of COB2 B in Gas 1 or Gas 2, respectivelyT
TBTBGas 1 or 2B = Pressure inside the measuring chamber during a measurement on Gas 1 or Gas 2, respectively
pH O2 = Water vapor pressure (6.2751 kPa at 37 P
oPC)
E(COB2 B,Gas1), E(COB2 B,Gas2)
= Potential of the pCOB2B electrode from a measurement on Gas 1 or Gas 2, respectively
Sens(pCOB2 B,theo) = Theoretical (absolute) sensitivity of the pCOB2 B electrode at 37 P
oPC
Sens(pCOB2 B,prev) = Relative sensitivity of the pCOB2 B electrode from the previous 2-point calibration
Continued on next page
Nernst equation T(continued)T
Designation
ABL800 FLEX reference manual 1. Potentiometric measuring principles
1-17
TpTCOB2 B electrode, TContinuedT
E B0B(COB2 B,Gas1) = Standard potential of the pCOB2 B electrode with Gas 1
E(COB2 B,Gas1,prev) = Potential of the pCOB2 B electrode from the previous measurement on Gas 1
δ = Difference between Tp TCOB2 B(sample) from the first and last updatings
predict = Extrapolated value for Tp TCOB2 B
The pCOB2 B electrode is calibrated on two gases with known COB2B contents:
Gas 1: 5.61 % COB2 B and Gas 2: 11.22 % COB2 B. The exact composition of the calibration gases is contained in their barcodes.
The partial pressures of COB2 B in Gas 1 and Gas 2 are calculated from the following equations:
( )p F B pCO CO2 2 Gas 1 2H O kPa( ) ( )Gas Gas1 1= × −
( )p F B pCO CO2 2 Gas 2 2H O kPa( ) ( )Gas Gas2 2= × −
The relative sensitivity of the pCOB2 B electrode is calculated as follows:
Sens( CO E(CO ,Gas2) E(CO ,Gas1)
Sens( CO , theo) log CO (Gas2)CO (Gas1)
22 2
22
2
pp p
p
) =−
×
The sensitivity of the pCOB2B electrode should fall between 0.85-1.00 or 85-100 %.
The status of the pCOB2B electrode is calculated as follows:
kPa10(Gas1)CO)COStatus( theo),COSens(Gas1),(COEGas1),E(CO
222
202
ppp−
×=
The status of the pCOB2B electrode should fall between 6.2-260 mmHg (0.83-34.66 kPa).
Drift 1 is calculated as follows:
kPa)prev(Gas1,CO10(Gas1)CO)CO1(Drift 2theo),COSens(prev),COSens(
prev)Gas1,,E(COGas1),E(CO
2222
22
ppp pp −×= ×−
Drift 2 is calculated as follows:
kPa)prev(Gas2,CO10(Gas2)CO)CO2(Drift 2theo),COSens(prev),COSens(
prev)Gas1,,E(COGas2),E(CO
2222
22
ppp pp −×= ×−
Continued on next page
Designation T(continued)
Sensitivity
Status
Drift
1. Potentiometric measuring principles ABL800 FLEX reference manual
1-18
TpTCOB2 B electrode, TContinuedT
The default drift tolerances set by Radiometer are as follows:
• for Drift 1: ± 0.33 kPa (2.5 mmHg)
• for Drift 2: ± 0.67 kPa (5.0 mmHg)
The Tp TCOB2 B value for a sample is calculated from the following equations:
theo),COSens(prev),COSens(Gas1)E(COupdi)sample,E(CO
2222
22
10)Gas1(CO)updsample,(CO pppip ×−
×=
δ = −p pCO (sample, upd30) CO (sample, upd12 2 )
[ ]predictCO (sample, upd6) CO (sample, upd30 CO (sample, upd18CO (sample,upd6 CO (sample, upd30) CO (sample,upd18)
2 2 2
2 2 2
=× −
+ − ×
p pp p p
) ))
p 2
2
For δ < 1.33 kPa, Tp TCOB2 B(sample) = Tp TCOB2 B(sample,upd30)
ForT T1.33 kPa < δ < 2.66 kPa
p pCO (sample) predict CO (sample, upd30)2
2=× − + × −( . ) ( . )
.δ δ133 2 66
133
ForT Tδ ≥ 2.66 kPa, Tp TCOB2 B(sample) = predict.
The Tp TCOB2 B measured on a sample is then corrected for systematic deviations from the reference method using the following equations:
TEquation A: TTp TCOB2 B(sample,corr) = AB3 B × Tp TCOB2 B(sample)P
3P + AB2 B × Tp TCOB2 B(sample)P
2P +
+ AB0B × Tp TCOB2 B(sample) + AB1 B × (TB T − Tp THB2BO)
where: pCOB2 B(sample) = uncorrected value of Tp TCOB2 B in the sample
pCOB2 B(sample, corr) = corrected value of TpTCOB2 B in the sample
AB0 B = correction factor
AB1 B = correction factor
AB2 B = correction factor
AB3 B = correction factor
Tp THB2BO = partial pressure of saturated water vapor (6.2751 kPa)
and
Continued on next page
Drift T(continued)T
Measurement
Corrections – blood samples
ABL800 FLEX reference manual 1. Potentiometric measuring principles
1-19
TpTCOB2 B electrode, TContinuedT
TEquation B: T Tp TCOB2 B(sample,corr) = BB1 B × Tp TCOB2 B(sample) + BB0 B
where: BB0 B = correction cut-off
BB1 B = correction factor
TABL8xx FLEX T
TModeT TAB0 TB TAB1 TB TAB2 TB TAB3 TB TBB0 TB TBB1 TB TEq. T
37/27/17 S250 1.1233 -0.003303 0.003 -0.00002 A
C125 1.1233 -0.003303 0.003 -0.00002 A
*FM (message 903)
1.1233 -0.003303 0.003 -0.00002 1.05742 -0.158 A, B
*FM (message 902)
1.1233 -0.003303 0.003 -0.00002 1.05742 -0.158 A, B
*FM (message 901)
1.1233 -0.003303 0.003 -0.00002 1.05742 -0.158 A, B
*FM (message 900)
1.1233 -0.003303 0.003 -0.00002 1.05742 -0.158 A, B
*FM (message 899)
1.1233 -0.003303 0.003 -0.00002 1.05742 -0.158 A, B
*FM (message 898)
1.1233 -0.003303 0.003 -0.00002 1.05742 -0.158 A, B
35/25/15 S195 1.1126 -0.003573 0.0051 -0.0000002 A
S95 1.1126 -0.003573 0.0051 -0.0000002 1.000 -0.016 A, B
S85 1.1126 -0.003573 0.0051 -0.0000002 A
C95 1.1126 -0.003573 0.0051 -0.0000002 1.013 0.010 A, B
C55 1.1126 -0.003573 0.0051 -0.0000002 1.12 -0.28 A, B
*FM (no message)
1.1126 -0.003573 0.0051 -0.0000002 A
*FM (message 897)
1.1126 -0.003573 0.0051 -0.0000002 1.013 0.010 A, B
*FM (message 895)
1.1126 -0.003573 0.0051 -0.0000002 1.013 0.010 A, B
*FM (message 894)
1.1126 -0.003573 0.0051 -0.0000002 1.013 0.010 A, B
*FM (message 873)
1.1126 -0.003573 0.0051 -0.0000002 1.013 0.010 A, B
*FM (message 872)
1.1126 -0.003573 0.0051 -0.0000002 1.0884 -0.1619 A, B
*FM (message 871)
1.1126 -0.003573 0.0051 -0.0000002 1.090 -0.150 A, B
Continued on next page
Corrections – blood samples T(continued)T
1. Potentiometric measuring principles ABL800 FLEX reference manual
1-20
TpTCOB2 B electrode, TContinuedT
TABL8xx FLEXT
TModeT TAB0 TB TAB1 TB TAB2 TB TAB3 TB TBB0 TB TBB1 TB TEq.T
30/20/10 S85 1.1126 -0.003573 0.0051 -0.0000002 A
10 BG only
C 55 1.1126 -0.003573 0.0051 -0.0000002 1.12 -0.28 A, B
FLEX *FM (no message)
1.1126 -0.003573 0.0051 -0.0000002 1.013 0.050 A, B
*FM (message 872)
1.1126 -0.003573 0.0051 -0.0000002 1.0819 -0.0495 A, B
*FM (message 871)
1.1126 -0.003573 0.0051 -0.0000002 1.090 -0.150 A, B
05 S165 1.1126 -0.003573 0.0051 -0.0000002 A
S95 1.1126 -0.003573 0.0051 -0.0000002 1.000 -0.016 A, B
S85 1.1126 -0.003573 0.0051 -0.0000002 A
C95 1.1126 -0.003573 0.0051 -0.0000002 1.013 0.010 A, B
C55 1.1126 -0.003573 0.0051 -0.0000002 1.12 -0.28 A, B
*FM (no message)
1.1126 -0.003573 0.0051 -0.0000002 A
*FM (message 897)
1.1126 -0.003573 0.0051 -0.0000002 1.013 0.010 A, B
*FM (message 895)
1.1126 -0.003573 0.0051 -0.0000002 1.013 0.010 A, B
*FM (message 894)
1.1126 -0.003573 0.0051 -0.0000002 1.013 0.010 A, B
*FM (message 873)
1.1126 -0.003573 0.0051 -0.0000002 1.013 0.010 A, B
*FM (message 872)
1.1126 -0.003573 0.0051 -0.0000002 1.0884 -0.1619 A, B
*FM (message 871)
1.1126 -0.003573 0.0051 -0.0000002 1.090 -0.150 A, B
00 S195 -0.003573 1.1126 0.0051 -0.0000002 A
BASIC S95 -0.003573 1.1126 0.0051 -0.0000002 1.000 -0.016 A, B
S85 -0.003573 1.1126 0.0051 -0.0000002 A
C95 -0.003573 1.1126 0.0051 -0.0000002 1.013 0.010 A, B
C55 -0.003573 1.1126 0.0051 -0.0000002 1.12 -0.28 A, B
*FM = FLEXMODE.
Continued on next page
Corrections – blood samples T(continued)
ABL800 FLEX reference manual 1. Potentiometric measuring principles
1-21
TpTCOB2 B electrode, TContinuedT
The pCOB2 B measured from the sample is then corrected for systematic deviations from the reference method using the following equation: )OH(A(sample)COAcorr)(sample,CO 21202 pBpp −×+×= TEquation A: T
where:
Tp TCOB2 B(sample) = uncorrected TpTCOB2 B value of an expired air sample
Tp TCOB2 B(sample,corr) = corrected Tp TCOB2 B value of an expired air sample
AB0 B
= instrument-dependent correction factor
AB1 B
= instrument-dependent correction factor
TB T
= barometric pressure during the measurement
Tp THB2BO = partial pressure of saturated water vapour = 6.2751 kPa
TABL8xx T
TModeT TAB0 TB TAB1 TB TEquationT
All FLEX and BASIC
Expired air 1.0196 –0.00106 A
The following stability criterion must be met to obtain a stable electrode response during calibration:
upd.i)(sample,OCupd.last)(sample,OC 22 pp − ≤ pCOB2 B (limit)
This criterion is valid for calibrations using Gas 1 and Gas 2 where:
Parameter pCOB2 B value from the last updating number...
TABL8x5/8x7 FLEX T TABL8x0 FLEX T
pCOB2 B(sample,upd.last) 92 62
pCOB2 B(sample,upd.i) 86 or 87 56 or 57
(the relationship must be fulfilled for at least one of the updating numbers)
Tp TCOB2 B(limit) value for the stability criterion is 0.40 kPa (3.0 mmHg).
Continued on next page
Corrections – expired air samples
Stability criteria
1. Potentiometric measuring principles ABL800 FLEX reference manual
1-22
TpTCOB2 B electrode, TContinuedT
The following stability criteria must be met to obtain a stable electrode response during measurement:
δ= upd.i)(sample,OC)upd.30(sample,OC 22 pp −
For δ Criterion
≤1.33 kPa 40.0)upd.16(sample,OCupd.30)(sample,OC 22 ≤− pp
>1.33 kPa 5.0)1upd.(sample,OC)upd.16(sample,OC)upd.16(sample,OCupd.30)(sample,OC1.0
22
22 <−
−≤−
pppp
For δ >1.33 kPa: T
if the following criteria are fulfilled, then no result is reported:
0.1)1upd.(sample,OC)upd.16(sample,OC)upd.16(sample,OCupd.30)(sample,OC
22
22 −<−
−pp
pp
or
5.0)1upd.(sample,OC)upd.16(sample,OC)upd.16(sample,OCupd.30)(sample,OC
22
22 ≥−
−pp
pp
TUExpired air samplesUT:
Measurement on an expired air sample is accepted if the following criterion is fulfilled:
⏐pCOB2 B (sample,upd.30) − pCOB2 B (sample,upd.24)⏐≤0.40 kPa (3.0 mmHg)
or
⏐pCOB2 B (sample,upd.30) − pCOB2 B (sample,upd.24)⏐≤0.04 × pCOB2B (sample,upd.30).
Error message "Measurement unstable" (= pCOB2 B response fault during electrode monitoring in Expired air mode) is displayed if the stability criterion is not fulfilled.
Stability criteria T(continued)T
ABL800 FLEX reference manual 1. Potentiometric measuring principles
1-23
Electrolyte electrodes
The K electrode (E722) is an ion- selective electrode whose sensing element is a PVC membrane containing a potassium-neutral ion carrier. The ion-sensitive membrane is covered with a cellophane membrane in order to protect it from the samples.
The electrolyte has a constant and known concentration of potassium ions. When a sample is brought in contact with the electrode, a potential develops across the PVC and cellophane membranes. The potential depends on the difference between the potassium (more precisely, activity) in the electrolyte and the sample. If the cKP
+P in
both solutions is the same, the potential across the electrode tip will be 0 V.
The Na electrode (E755) is an ion-selective electrode whose sensing element is a NaP
+P-sensitive ceramic pin
contained in the tip of the jacket.
The electrolyte has a constant and known concentration of sodium ions. When a sample is brought in contact with the electrode, a potential develops across the ceramic pin. The potential depends on the difference between the sodium (more precisely, activity) in the electrolyte and the sample. If the cNaP
+P
in both solutions is the same, the potential across the electrode tip will be 0 V.
Continued on next page
Description
Cellophane membrane
Electrode jacket
Electrode contact
Electrode jacket
Electrode contact
1. Potentiometric measuring principles ABL800 FLEX reference manual
1-24
Electrolyte electrodes, TContinuedT
The Ca electrode (E733) is an ion- selective electrode whose sensing element is a PVC membrane containing a calcium-neutral ion carrier. The ion-sensitive membrane is covered with a cellophane membrane in order to protect it from the samples.
The electrolyte has a constant and known concentration of calcium ions. When a sample is brought in contact with the electrode, a potential develops across the PVC and cellophane membranes. The potential depends on the difference between the calcium (more precisely, activity) in the electrolyte and the sample. If the cCaP
2+P
in both solutions is the same, the potential across the electrode tip will be 0 V.
The Cl electrode (E744) is an ion- selective electrode whose sensing element is a PVC membrane containing a chloride ion carrier. The ion-sensitive membrane is covered with a cellophane membrane in order to protect it from the samples.
The electrolyte has a constant and known concentration of chloride ions. When a sample is brought in contact with the electrode, a potential develops across the PVC and cellophane membranes. The potential depends on the difference between the chloride (more precisely, activity) in the electrolyte and the sample. If the cClP
−P in
both solutions is the same, the potential across the electrode tip will be 0 V.
Continued on next page
Description T (continued)T
Cellophane membrane
Electrode contact
Cellophane membrane
Electrode jacket
Electrode jacket
Electrode contact
ABL800 FLEX reference manual 1. Potentiometric measuring principles
1-25
Electrolyte electrodes, TContinuedT
The total potential across the electrode chain is a sum of the potential differences at each of the elements in the chain, all but one of which is known and constant.
Element Potential Symbol
Ag/AgCl electrode /electrolyte solution. (Reference electrode)
Known and constant when the Ag/AgCl wire is immersed in the electrolyte solution.
TE TBrefB
Membrane junction between the electrolyte solution in the reference electrode and the sample.
Known and constant, independent of sample composition.
TE TBMJB
Ion-sensitive membrane (or pin) junction separating the sample and the electrode.
TUnknown T, dependent on sample composition.
TE TBSampleB
Ag/AgCl electrode/inner buffer solution. (Electrolyte electrode)
Known and constant when the Ag/AgCl wire is immersed in the electrolyte solution.
TE TBEB
Total potential. Measured by the voltmeter. TE TBtotB
The unknown potential difference across the ion-sensitive membrane or pin is then the difference between the measured total potential and the sum of the known potentials:
( )E E E E ESample tot ref MJ E= mV− + +
The potential difference at the membrane (or pin) in the electrolyte electrodes can be expressed by the Nernst equation:
E E Tn
aSample 0 ion2 3R
Flog mV= + ×
.
where:
E0 = standard electrode potential
R = gas constant (8.3143 J × KP
−1Pmol P
−1P)
TT T = absolute temperature (310.15 K at 37 P
oPC)
Tn T = charge on the ion (n = 1 for KP
+P and NaP
+P, n = −1 for ClP
−P, n = 2 for CaP
2+P)
F = Faraday constant (96487 coulomb × mol P
−1P)
aion = activity of the specific ion
Continued on next page
Electrode chain potential
Nernst equation
1. Potentiometric measuring principles ABL800 FLEX reference manual
1-26
Electrolyte electrodes, TContinuedT
Calibration solutions for the ABL835/30/25/20/15/10/05 FLEX analyzers have the following nominal electrolyte concentrations:
Parameter S1820 Cal 1 S1830 Cal 2
TcTKP
+P 4.0 mmol/L 40.0 mmol/L
TcTNaP
+P 145 mmol/L 20.0 mmol/L
TcTCaP
2+P 1.25 mmol/L 5.0 mmol/L
TcTCl P
−P
102 mmol/L 50 mmol/L
Calibration solutions for the ABL837/27/17 FLEX analyzers have the following nominal electrolyte concentrations:
Parameter S1827 Cal 1 S1837 Cal 2
TcTKP
+P 4.0 mmol/L 10.0 mmol/L
TcTNaP
+P 145 mmol/L 50.0 mmol/L
TcTCaP
2+P 1.25 mmol/L 5.0 mmol/L
TcTCl P
−P
102 mmol/L 50 mmol/L
The precise concentration of each electrolyte ion is contained in the solution’s barcodes.
The following designations are used (X = K/Na/Ca/Cl):
E(X,Cal1) = Potential of the respective electrolyte electrode chain from a calibration on Cal 1 solution
E(X,Cal2) = Potential of the respective electrolyte electrode chain from a calibrration on Cal 2 solution
61.5 = Theoretical sensitivity of the K and Na electrodes at 37 P
oPC
30.75 = Theoretical sensitivity of the Ca electrode at 37 P
oPC
−61.5 = Theoretical sensitivity of the Cl electrode at 37 P
oPC
TcTX(Cal1) = Concentration of the respective electrolyte ion in Cal 1 solution
TcTX(Cal2) = Concentration of the respective electrolyte ion in Cal 2 solution
E B0B(X,Cal1) = Standard potential of the respective electrolyte electrode chain
Continued on next page
Calibration solution values
Designations
ABL800 FLEX reference manual 1. Potentiometric measuring principles
1-27
Electrolyte electrodes, TContinuedT
TcTX(Cal1,nom) = Nominal concentration of the respective electrolyte ion in Cal 1 solution
E(X,Cal1,prev) = Potential of the respective electrolyte electrode chain from the previous calibration on Cal 1 solution
Sens(X,Cal2,prev) = Sensitivity of the respective electrolyte electrode from the last 2-point calibration
cX(Cal1,prev) Concentration of the respective electrolyte ion in Cal 1 solution in the previous calibration
The sensitivity of the electrolyte electrodes is calculated from the following equations:
K electrode
)(Cal2K(Cal1)Klog61.5
Cal2)E(K,Cal1)E(K,Sens(K) +
+×
−=
cc
(fraction)
Na electrode
(fraction)
)(Cal2N(Cal1)Nalog61.5
Cal2)E(Na,Cal1)E(Na,Sens(Na) +
+×
−=
acc
Ca electrode
(fraction)
)(Cal2Ca(Cal1)Calog75.03
Cal2)E(Ca,Cal1)E(Ca,Sens(Ca)
+2
+2
cc
×
−=
Cl electrode
(fraction)
)(Cal2Cl(Cal1)Cllog61.5-
Cal2)E(Cl,Cal1)E(Cl,Sens(Cl)
−
−
×
−=
cc
The sensitivity limits of the electrolyte electrodes are as follows:
Electrode Sensitivity limits
K 92-105 %
Na 90-105 %
Ca 90-105 %
Cl 85-105 %
Continued on next page
Designations T(continued)
Sensitivity
1. Potentiometric measuring principles ABL800 FLEX reference manual
1-28
Electrolyte electrodes, TContinuedT
The status of each of the electrolyte electrodes is calculated from the following equations:
K electrode
(Cal1)Knom)(Cal1,K01Status(K)
261.5Cal1)(K,ECal1)E(K, 0
+
+−
×=
cc mmol/L
Na electrode
(Cal1)Nnom)(Cal1,N01Status(Na)
261.5Cal1)(Na,ECal1)E(Na, 0
+
+−
×=
acac mmol/L
Ca electrode
(Cal1)Cnom)(Cal1,C01Status(Ca) 2
2230.75Cal1)(Ca,ECal1)E(Ca, 0
+
+−
×=
acac mmol/L
Cl electrode
(Cal1)Clnom)(Cal1,Cl01Status(Cl)
261.5-Cal1)(Cl,ECal1)E(Cl, 0
−
−−
×=
cc mmol/L
The status limits of the electrolyte electrodes are as follows:
Electrode Status Limits
K 0.5-12 mmol/L
Na 10-250 mmol/L
Ca 0.1-20 mmol/L
Cl 30-900 mmol/L
Drift equations are given below. K electrode
)Cal1(K)prevCal1,(K101(K)Drift ++prev)Sens(K,61.5prev)Cal1,E(K,Cal1)E(K,
cc −×= ×−
mmol/L
)(Cal2K)prevCal1,(K102(K)Drift +prev)Sens(K,61.5prev)Cal1,E(K,-Cal2)E(K,
cc + −×= × mmol/L
Continued on next page
Status
Drift
ABL800 FLEX reference manual 1. Potentiometric measuring principles
1-29
Electrolyte electrodes, TContinuedT
Na electrode
)Cal1(Na)prevCal1,(Na101(Na)Drift ++prev)Sens(Na,61.5prev)Cal1,E(Na,Cal1)E(Na,
cc −×= ×−
mmol/L
)(Cal2Na)prevCal1,(Na102(Na)Drift +prev)Sens(Na,61.5prev)Cal1,E(Na,-Cal2)E(Na,
cc + −×= × mmol/L Ca electrode
)Cal1(Ca)prevCal1,(Ca101(Ca)Drift +2+2prev)Sens(Ca,75.03prev)Cal1,E(Ca,Cal1)E(Ca,
cc −×= ×−
mmol/L
)(Cal2Ca)prevCal1,(Ca102(Ca)Drift +2prev)Sens(Ca,75.03prev)Cal1,E(Ca,-Cal2)E(Ca,
cc +2 −×= × mmol/L
Cl electrode
)Cal1(Cl)prevCal1,(Cl101(Cl)Drift prev)Sens(Cl,61.5-prev)Cal1,E(Cl,Cal1)E(Cl,
−−×−
−×= cc mmol/L
)(Cal2Cl)prevCal1,(Cl102(Cl)Drift prev)Sens(Cl,61.5-prev)Cal1,E(Cl,-Cal2)E(Cl,
−−× −×= cc mmol/L
NOTE: If Cal 1 solution bottle has not been changed between two consecutive calibrations, the cTX(Cal1,prev T) − cTX(Cal1)T = 0, where X is the respective electrolyte ionT.T
The default drift tolerances set by Radiometer are as follows:
Electrode Drift 1 tolerances Drift 2 tolerances
K ± 0.2 mmol/L ± 1.5 mmol/L
Na ± 3 mmol/L ± 1 mmol/L
Ca ± 0.05 mmol/L ± 0.2 mmol/L
Cl ± 2 mmol/L ± 3 mmol/L
The electrolyte concentration in a sample is calculated from the following equation:
)prevSens(X,Sens(theo)prev)Cal,E(X,-sample)E(X,
101) X(CalX(sample) ××=cc
where: E(X,sample) = potential of the electrolyte electrode chain from a
measurement on the sample. E(X,Cal,prev) = potential of the electrolyte electrode chain from the
previous calibration on Cal 1 solution. TcTX(Cal 1)B B= specific (true) concentration of the electrolyte ion in Cal 1
solution.
Continued on next page
Drift T(continued)T
Measurement
1. Potentiometric measuring principles ABL800 FLEX reference manual
1-30
Electrolyte electrodes, TContinuedT
Sens(theo) = theoretical sensitivity of the electrolyte electrode. Sens(X,prev) = relative sensitivity of the electrolyte electrode chain from
the last 2-point calibration
The measured electrolyte concentration is then corrected for systematic deviations from the reference method by the following equations:
TEquation A:
TcTX(sample, corr) BmacromodeB = AB0,macromodeB × cX(sample)B macromode B+ AB1,macromode B
and
TEquation B: T
TcTX(sample,corr)BmicromodeB = AB0 B, BmicromodeB × TcTX(sample,corr)B macromodeB + AB1,micromodeB
where
TcTX(sample) = uncorrected value of the electrolyte ion in the sample
TcTX(sample,corr) = corrected value of the electrolyte ion in the sample
AB0 B = instrument-dependent correction factor
AB1 B = instrument-dependent correction cut-off
Chloride is corrected for −3HCOc interference. The default value 5.24HCO3 =−c
mmol/L is used in
TEquation C:
cClP
–P(sample,corr)BmacromodeB = AB0,macromodeB × (cCl P
–P(sample) – 0.0956 × cHCOB3 PB
–P) +
+AB1,macromodeB
Note that subscript "macromode" in the equations above is used for the sake of convenience and stands for S250, "FLEXMODE (no message)".
Continued on next page
Measurement T(continued)
Corrections
ABL800 FLEX reference manual 1. Potentiometric measuring principles
1-31
Electrolyte electrodes, TContinuedT
TCorrections for TTcTTNa P
+P:
TABL8xx FLEX T
TModeT TAB0 TB TAB1 TB TEquationT
37/27/17 S250 1.0307 -8.104 A
C125 1.0307 -8.104 A
*FM (message 903) 1.0123 1.3284 A, B
*FM (message 902) 1.0123 1.3284 A, B
*FM (message 901) 1.0123 1.3284 A, B
*FM (message 900) 1.0123 1.3284 A, B
*FM (message 899) 1.0123 1.3284 A, B
*FM (message 898) 1.0123 1.3284 A, B
35/25/15 S195 0.995 -3.00 A
S95 1.01 1.80 A, B
C95 1.03 -1.09 A, B
*FM (no message) 0.995 -3.00 A
*FM (message 897) 1.030 -1.00 A, B
*FM (message 895) 1.030 -1.00 A, B
05 S165 0.995 -3.00 A
S95 1.01 1.80 A, B
C95 1.03 -1.09 A, B
*FM (no message) 0.995 -3.00 A
*FM (message 895) 1.03 -1.00 A, B
*FM (message 897) 1.03 -1.00 A, B
00 BASIC S195 0.995 -3.00 A
S95 1.01 1.80 A, B
C95 1.03 -1.09 A, B
Continued on next page
Corrections T(continued)T
1. Potentiometric measuring principles ABL800 FLEX reference manual
1-32
Electrolyte electrodes, TContinuedT
TCorrections for TTcTTKP
+P:
TABL8xx FLEX T
TModeT TAB0 TB TAB1 TB TEquationT
37/27/17 S250 0.9784 -0.1163 A
C125 0.9784 -0.1163 A
*FM (message 903) 1.0462 -0.1101 A, B
*FM (message 902) 1.0462 -0.1101 A, B
*FM (message 901) 1.0462 -0.1101 A, B
*FM (message 900) 1.0462 -0.1101 A, B
*FM (message 899) 1.0462 -0.1101 A, B
*FM (message 898) 1.0462 -0.1101 A, B
35/25/15 S195 0.985 -0.065 A
S95 1.05 -0.13 A, B
C95 1.11 -0.37 A, B
*FM (no message) 0.985 -0.065 A
*FM (message 897) 1.11 -0.37 A, B
*FM (message 895) 1.11 -0.37 A, B
05 S165 0.985 -0.065 A
S95 1.05 -0.13 A, B
C95 1.11 -0.37 A, B
*FM (no message) 0.985 -0.065 A
*FM (message 897) 1.11 -0.37 A, B
*FM (message 895) 1.11 -0.37 A, B
00 BASIC S195 0.985 -0.065 A
S95 1.05 -0.13 A, B
C95 1.11 -0.37 A, B
*FM = FLEXMODE
TContinued on next page
Corrections T(continued)T
ABL800 FLEX reference manual 1. Potentiometric measuring principles
1-33
Electrolyte electrodes, TContinuedT
TCorrections for TTcTTCa P
2+P:
TABL8xx FLEX T
TModeT TAB0 TB TAB1 TB TEquationT
37/27/17 S250 1.0108 -0.0396 A
C125 1.0108 -0.0396 A
*FM (message 903) 1.0584 -0.001 A, B
*FM (message 902) 1.0584 -0.001 A, B
*FM (message 901) 1.0584 -0.001 A, B
*FM (message 900) 1.0584 -0.001 A, B
*FM (message 899) 1.0584 -0.001 A, B
*FM (message 898) 1.0584 -0.001 A, B
35/25/15 S195 1.004 -0.022 A
S95 1.05 -0.004 A, B
C95 1.08 -0.04 A, B
*FM (no message) 1.004 -0.022 A
*FM (message 897) 1.08 -0.04 A, B
*FM (message 895) 1.08 -0.04 A, B
05 S165 1.004 -0.022 A
S95 1.05 -0.004 A, B
C95 1.08 -0.04 A, B
*FM (no message) 1.004 -0.022 A
*FM (message 897) 1.08 -0.04 A, B
*FM (message 895) 1.08 -0.04 A, B
00 BASIC S195 1.004 -0.022 A
S95 1.05 -0.004 A, B
C95 1.08 -0.04 A, B
Continued on next page
Corrections T(continued)T
1. Potentiometric measuring principles ABL800 FLEX reference manual
1-34
Electrolyte electrodes, TContinuedT
TCorrections for TTcTTCl P
-P:
TABL8xx FLEX T
TModeT TAB0 TB TAB1 TB TEquationT
37/27/17 S250 1.2715 -36.048 C
C125 1.2715 -36.048 C
*FM (message 903) 0.9682 3.4529 C, B
*FM (message 902) 0.9682 3.4529 C, B
*FM (message 901) 0.9682 3.4529 C, B
*FM (message 900) 0.9682 3.4529 C, B
*FM (message 899) 0.9682 3.4529 C, B
*FM (message 898) 0.9682 3.4529 C, B
35/25/15 S195 1.225 -30.7 C
S95 1.000 0.0 C, B
C95 1.01 -1.7 C, B
*FM (no message) 1.225 -30.7 C
*FM (message 897) 1.01 -1.7 C, B
*FM (message 895) 1.01 -1.7 C, B
05 S165 1.225 -30.7 C
S95 1.000 0.0 C, B
C95 1.01 -1.7 C, B
*FM (no message) 1.225 -30.7 C
*FM (message 897) 1.01 -1.7 C, B
*FM (message 895) 1.01 -1.7 C, B
00 BASIC S195 1.225 -30.7 C
S95 1.000 0.0 C, B
C95 1.01 -1.7 C, B
*FM = FLEXMODE
Continued on next page
ABL800 FLEX reference manual 1. Potentiometric measuring principles
1-35
Electrolyte electrodes, TContinuedT
The following stability criterion must be met to obtain a stable electrode response during calibration:
upd.last)X(Cal,K)upd.iX(Cal,upd.last)X(Cal, ccc ×≤−
This criterion is valid for calibrations using Cal 1 and Cal 2 solutions where:
TcTX(Cal,upd.last) = Concentration of the electrolyte ion from the last updating when measuring on calibration solution. (The last updating is number 30.)
TcTX(Cal,upd.i) = Concentration of the electrolyte ion for a given updating when measuring on calibration solution. (The relationship must be fulfilled for at least one of the updating numbers 18 or 19.)
K = Constant for the stability criterion.
TElectrolyte ionT TCal 1 solution T TCal 2 solution T
KP
+P 0.01 0.01
NaP
+P 0.01 0.02
CaP
2+P 0.02 0.02
Cl P
−P
0.022 0.022
The following stability criterion must be met to obtain a stable electrode response during measurement:
( ))()()upd.lastX(sample,K
)upd.iX(sample,upd.last)X(sample,
RinsecXRinsecXc
cc
+−×
≤−
where:
TcTX(sample,upd.last) = Concentration of the electrolyte ion from the median of the last five updatings (for CaP
2+P: Three last
updatings) when measuring on a sample. The last updating number is 30 (or 10 for some micromodes).
TcTX(sample,upd.i) = Concentration of the electrolyte ion for a given updating when measuring on a sample. (The relationship must be fulfilled for at least one of the updating numbers shown below.)
KP
+P NaP
+P CaP
2+P Cl P
−P
22 22 26 22
23 23 27 23
In some micromodes, substract 20 from number above.
Continued on next page
Stability criteria
1. Potentiometric measuring principles ABL800 FLEX reference manual
1-36
Electrolyte electrodes, TContinuedT
K Constant for the stability criterion; it equals to:
KP
+P = 0.012; NaP
+P = 0.012; Ca P
2+P = 0.022; ClP
−P = 0.012
cXRinse Constant used indicates the concentration of the electrolyte ion level in the Rinse Solution:
KP
+P = 4.0; NaP
+P = 130.0; CaP
2+P = 1.25; Cl P
−P = 137.7
Stability criteria T(continued)
ABL800 FLEX reference manual 1. Potentiometric measuring principles
1-37
References
Linnet N. pH measurements in theory and practice. 1st ed. Copenhagen: Radiometer Medical A/S, 1970.
List of references
1. Potentiometric measuring principles ABL800 FLEX reference manual
1-38
2. Amperometric measuring principles
Overview
This chapter describes the amperometric measuring principles and the pOB2B, Glucose, Lactate and Crea electrodes that are based on this principle.
This chapter contains the following topics.
General information ......................................................................................... X2-2X
pOB2B electrode.................................................................................................... X2-4X
Glucose and Lactate electrodes ........................................................................ X2-13X
Crea electrodes ................................................................................................. X2-23X
References ........................................................................................................ X2-31X
Introduction
Contents
2. Amperometric measuring principles ABL800 FLEX reference manual
2-2
General information
The magnitude of an electrical current flowing through an electrode chain, which is in turn proportional to the concentration of the substance being oxidized or reduced at an electrode in the chain
The electrode chain in amperometric measurements consists of the sample, the two electrodes (anode and cathode), an ammeter, a voltage source, the membranes and the electrolyte solutions.
Part Function
Cathode Negative electrode where a reduction reaction occurs and electrons are consumed.
Anode Positive electrode where an oxidation reaction occurs and electrons are released.
Electrolyte solution
Provides electrical contact between the anode and cathode.
Membrane Allows the appropriate molecules to pass through from the sample.
Sample Contacts the membrane.
Applied voltage Applies the necessary potential for the reduction or oxidation reaction under study.
Ammeter Measures the current flowing through the circuit.
To simplify the description of the measuring process in an amperometric electrode, we make the following assumptions:
• there is a species TAT in the sample which is reduced at the cathode to TAP
−TP.
• there is a species TX T in the electrolyte which is oxidized at the anode to TXP
+TP.
Continued on next page
Amperometric method
Cathode
Ammeter
Applied voltage
Anode
Electrolyte solution
Sample
Membrane
ABL800 FLEX reference manual 2. Amperometric measuring principles
2-3
General information, TContinued
The membrane is selective to the species TAT, allowing no other species but it to pass through from the sample into the electrolyte solution.
As an appropriate potential is applied across the electrodes, the species TAT is reduced at the cathode according to the following reaction:
A + eP
−P → AP
−
The reduction of TAT produces a flow of electrons, i.e. an electrical current.
To complete the electrical circuit an oxidation reaction where electrons are released is necessary. Therefore species TXT is oxidized according to the following reaction:
X → XP
+P + eP
−
The magnitude of the current flowing through the circuit is proportional to the concentration of the species being reduced, in this case species TAT. The analyzer thereby automatically calculates the concentration of TAT in the sample.
The amperometric measuring principle is applied in the pOB2B, Glucose, Lactate, Crea A and Crea B electrodes.
The electrodes are active elements and must be calibrated regularly as the signals from the electrodes change with, e.g. age or deposits on the membrane.
Calibration relates the electrode signals during the calibration sequence to the values of the calibrating solutions and must be performed at regular intervals so that the accuracy can be constantly refined after inevitable minor changes in the electrodes’ behavior.
Actual electrode condition is described by zero point and sensitivity and compared with theoretical conditions for an "ideal" electrode. In addition to zero point and sensitivity, an electrode condition is described by drift.
The following calibration materials are used:
Gas 1 and Gas 2: each gas has a precise composition essential for determining the accuracy of the analyzer in each pOB2 B measurement.
Calibration of the pOB2B electrode
Calibration Solution 1 Calibration of the Glucose, Lactate and Crea B electrodes
Calibration Solution 2 Calibration of the Crea A and Crea B electrodes
Amperometric method T(continued)T
Calibration
Calibration material
2. Amperometric measuring principles ABL800 FLEX reference manual
2-4
TpTOB2 B electrode
The pOB2B electrode is an amperometric electrode which consists of a silver anode, platinum cathode and Ag/AgCl reference band, all protected by an electrode jacket which is filled with electrolyte solution. At the tip of the electrode jacket an oxygen-permeable membrane protects the Pt cathode from protein contamination and is covered on the inner side with Pt-black.
The electrode chain is polarized with a constant voltage of -630 mV. Oxygen from the sample diffuses across the membrane into the electrolyte and is reduced on the cathode (electrons are consumed) according to the following equation:
OB2 B + 4HP
+P + 4eP
−P → 2HB2 BO
The H P
+P ions come from the
electrolyte solution. This represents the complete reduction of OB2 B. Some of the OB2 B, however, is only partially reduced according to the following equation:
OB2 B + 2HP
+P + 2eP
−P → HB2BOB2B
In the presence of Pt-black, HB2BOB2 B produced by the incomplete reduction of OB2 B at the cathode is immediately decomposed:
2HB2BOB2 B → 2HB2 BO + OB2 B This oxygen is then also reduced at the cathode. The reduction of oxygen produces a flow of electrons (an electrical current) the size of this current, I, proportional to the amount of oxygen and measured by the ammeter:
TI T = Sens(pOB2B) × pOB2 B + TI TBo BpA where:
Sens(pOB2 B) = sensitivity of the pOB2B electrode
pOB2B = partial pressure of OB2 B in the sample
TI TBoB = zero current, i.e. the current flowing through the circuit when pOB2B = 0 kPa (mmHg)
To complete the electrical circuit, an oxidation reaction where electrons are released is necessary. The reaction at the silver anode is the conversion of Ag to Ag P
+P:
Ag → AgP
+P + eP
−P
In order to maintain a charge balance between the anode and cathode, four atoms of Ag need to be oxidized for one molecule of OB2B to be reduced.
Continued on next page
Description
Electrode jacket
AgCl reference band
Membrane
Electrolyte
Electrode contact
ABL800 FLEX reference manual 2. Amperometric measuring principles
2-5
TpTOB2 B electrode, TContinuedT
The Ag P
+P ions are released into the electrolyte solution where they react with the
Cl P
−P ions present, producing AgCl which is insoluble and forms a layer on the silver
rod:
Ag P
+P + ClP
−P → AgCl
Not all Ag+ ions can be removed from the solution. Some reach the cathode where they are converted back to Ag and form a deposit of silver. This deposit must be periodically removed with the brush provided in the electrode box.
The following designations are used to describe sensitivity, zero point and drift:
I(OB2 B,Gas1) = Current recorded at the pOB2B electrode from a measurement on Gas 1
I(OB2 B,Gas2) = Current recorded at the pOB2B electrode from a measurement on Gas 2
pOB2B(Gas1) = Partial pressure of OB2 B in Gas 1
pOB2B(Gas2) = Partial pressure of OB2 B in Gas 2 TF TOB2 B(Gas1) = Fraction of OB2B in Gas 1 TF TOB2 B(Gas2) = Fraction of OB2B in Gas 2
TB T = Ambient pressure
pH O2 = Water vapor pressure = 6.2571 kPa at 37 P
oPC.
Sens(pOB2 B, prev)
= Sensitivity of the pOB2B electrode measured at the previous 2-point calibration
I(OB2 B,Gas2, prev)
= Current recorded at the pOB2B electrode from the previous measurement on Gas 2
The pOB2B electrode is calibrated on two gases with known OB2 B content.
Gas 1 contains 19.76 % OB2 B and Gas 2 contains 0.0 % OB2 B. The exact composition of the calibration gases is contained in their barcodes.
The sensitivity of the pOB2B electrode, Sens(pOB2 B), is calculated as follows:
pA/kPa(Gas2)O(gas1)OGas2),I(Ogas1),I(O
)OSens(2 2
222 pp
p−−
=
Continued on next page
Description T(continued)
Designations
Sensitivity
2. Amperometric measuring principles ABL800 FLEX reference manual
2-6
TpTOB2 B electrode, TContinuedT
The partial pressures of OB2 B in the gas mixtures Gas 1 and Gas 2 are calculated from the following equation:
[ ] kPaOH)Gas1(O)gas1(O 222 pBFp −×=
[ ] kPaOH)(Gas2O)(gas2O 222 pBFp −×=
The sensitivity of the pOB2B electrode should fall between 5-40 pA/mmHg or 37.5-300 pA/kPa.
The zero point of the pOB2B electrode is the electrode current at pOB2 B= 0. It is calculated from the current measured at the electrode with Gas 2 (0 % OB2 B), and the sensitivity:
kPaprev),OSens(
Gas2),I(O)Opoint Zero
2
22 p
(p =
The zero point value of the pOB2B electrode should be less than 6.0 mmHg or 0.80 kPa.
The zero point current is the current measured at the pOB2 B electrode with Gas 2 in the measuring chamber. When the measurement on Gas 2 begins, a relatively high current is recorded due to residual OB2 B from the Rinse Solution in the measuring chamber. This current falls exponentially with time while Gas 2 is present in the measuring chamber.
Forty seconds into the measurement the current reaches a steady state which is then considered as the zero point current.
Time (secs)
Current (pA)
Drift 1 is a measurement of the difference between two consecutive measurements on Gas 1, and is calculated from the following equation:
kPa)Gas1(O)prev,OSens(
)prevGas1,,I(O)Gas1,I(O)O1(Drift 2
2
)222 p
pp −
−=
Drift 2 reflects the change in sensitivity between 2-point calibrations and is calculated from the following equation:
kPa)Gas2(O)prev,OSens(
)prevGas2,,�I(O)Gas2,I(O)O2(Drift 2
2
)222 p
pp −
−=
The default drift tolerances set by Radiometer are ± 0.80 kPa (6.0 mmHg) for Drift 1 and Drift 2. The drift tolerances can, however, be user-defined in the Setup program.
Continued on next page
Sensitivity T(continued)T
Zero point
Drift
ABL800 FLEX reference manual 2. Amperometric measuring principles
2-7
TpTOB2 B electrode, TContinuedT
The Tp TOB2B value for a sample is calculated from the following equations:
12
222 K
)OSens()prevGas2,,I(Oupd.i)sample,,I(O
upd.i)(sample,O ×−
=p
p
Constant KB1B describes the gas/liquid relationship for the electrode. This constant is defined as:
K 1+0 58370 21712 Sens( O3.662941
2= − + +⎛
⎝⎜
⎞
⎠⎟. . . )01 p
)upd.1(sample,Oupd.30)(sample,O 22 pp −=δ
( ))18upd.sample,(O2)30upd.sample,(O)upd.6sample,(O
)18upd.sample,(O)upd.30sample,(O)upd.6sample,(Opredict
222
2222
pppppp×−+
−×=
where: I(O B2 B,sample,updi) = current recorded at the Tp TOB2B electrode from updating
number Ti T with a measurement on the sample I(OB2 B,Gas2,prev) = current recorded at the Tp TOB2B electrode from the previous
measurement on Gas 2 Sens(pOB2 B) = relative sensitivity of the Tp TOB2 B electrode determined
from the last calibration on Gas 1 and Gas 2 δ = difference between Tp TOB2 B(sample) from the first and last
updatings predict = extrapolated value for Tp TOB2 B
For T Tδ < 2.66 kPa, Tp TOB2B(sample) = Tp TOB2 B(sample, upd.30)
ForT T2.66 kPa < δ < 5.32 kPa
66.2)32.5(upd.30)(sample,O)66.2(predict
(sample)O 22
δδ −×+−×=
pp
For T Tδ≥5.32 kPa Tp TOB2B(sample) = predict
Continued on next page
Measurement
2. Amperometric measuring principles ABL800 FLEX reference manual
2-8
TpTOB2 B electrode, TContinuedT
TGas/liquid relationship T:
KB1 B is a constant that describes the gas/liquid relationship for the electrode. The constant is defined as follows:
⎟⎟⎠
⎞⎜⎜⎝
⎛++−+=
66294.3)O(Sens
712.218370.5100
11K 21
p
The Tp TOB2B measured from the sample is then corrected for systematic deviations from the reference method using the following equation:
TEquation A:
( )p
d d e e eO (sample,corr)
O (sample,v1) O (sample, v1)2
2 2=
− + − × + × + ×1 12
2 3 424 p p
2where:
• Tp TOB2B value of the sample after the first part of correction is as follows:
)398.100()((sample)Ov1)(sample,O4
2 (sample)O2122 Bekkpp pk3 −××−+= ×
• and:
d B1B
= eB0 B × Tp TOB2B(sample, v1) + eB1B
TkTB1 B
= correction constant = 0.02614
TkTB2 B
= correction constant = 0.02107
TkTB3 B = correction constant = −0.00281
eB0 B, eB1B, eB2B, eB3 B, eB4B
= correction constants
B = barometric pressure in kPa
TEquation A+:
When an additional correction is needed, equation A is first used together with the constants for the macoromodes S250 and FLEXMODE (C195 and 165) with no message. Then the obtained results are put back into equation A as pOB2B(sample) and then treated again, using the constants for the specific mode.
Or
TEquation B:
When an additional correction is needed, equation A is first used together with the constants for the macoromodes S250 and FLEXMODE (C195 and 165) with no message. Then the obtained results are put back into equation B as pOB2B(sample) and then treated again, using the constants for the specific mode.
TcTX(sample,corr) = AB0 B × TcTX(sample) + A B1B
Continued on next page
Corrections – blood samples
ABL800 FLEX reference manual 2. Amperometric measuring principles
2-9
TpTOB2 B electrode, TContinuedT
TABL 8xx
FLEX T
TModeT eB0 B eB1 B eB2 B eB3 B eB4 B AB0 B AB1 B Eq.
37/27/ S250 -2.3618 7.05686 0.92629 -7.16718 1.36413 A
17 C125 -2.3618 7.05686 0.92629 -7.16718 1.36413 A
*FM (message 903) 1.0102 0.0431 A, B
*FM (message 902) 1.0102 0.0431 A, B
*FM (message 901) 1.0102 0.0431 A, B
*FM (message 900) 1.0102 0.0431 A, B
*FM (message 899) 1.0102 0.0431 A, B
*FM (message 898) 1.0102 0.0431 A, B
35/25/ S195 -2.30300 5.96942 0.83281 -6.07310 1.30565 A
15 S95 1.020 -0.200 A, B
S85 -2.30300 5.96942 0.83281 -6.07310 1.30565 A
C95 0.9965 -0.0254 A, B
C55 -2.1199 4.7210 -1.0156 -4.2974 1.1035 A, A+
*FM (no message) -2.30300 5.96942 0.83281 -6.07310 1.30565 A
*FM (message 897) 0.9965 -0.0254 A, B
*FM (message 895) 0.9965 -0.0254 A, B
*FM (message 894) 0.9965 -0.0254 A, B
*FM (message 873) 0.9965 -0.0254 A, B
*FM (message 872) -2.20159 5.70807 -0.41342 -5.42718 1.19023 A, A+
*FM (message 871) -2.1199 4.7210 -1.0156 -4.2974 1.1035 A, A+
30/20/ S85 -2.30300 5.96942 0.83281 -6.07310 1.30565 A
10/10 C55 -2.1199 4.7210 -1.0156 -4.2974 1.1035 A, A+
BG *FM (no message) -2.30300 5.96942 0.83281 -6.07310 1.30565 A
only *FM (message 872) -2.19314 5.81012 -0.96320 -5.46921 1.18037 A, A+
*FM (message 871) -2.1199 4.7210 -1.0156 -4.2974 1.1035 A, A+
Continued on next page
Corrections – blood samples T(continued)T
2. Amperometric measuring principles ABL800 FLEX reference manual
2-10
TpTOB2 B electrode, TContinuedT
ABL 8xx
FLEX
Mode eB0B eB1B eB2B eB3 B eB4B AB0B AB1 B Eq.
05 S165 -2.30300 5.96942 0.83281 -6.07310 1.30565 A
S95 1.020 -0.200 A, B
S85 -2.30300 5.96942 0.83281 -6.07310 1.30565 A
C95 0.9965 -0.0254 A, B
C55 -2.1199 4.7210 -1.0156 -4.2974 1.1035 A, A+
*FM (no message) -2.30300 5.96942 0.83281 -6.07310 1.30565 A
*FM (message 897) 0.9965 -0.0254 A, B
*FM (message 895) 0.9965 -0.0254 A, B
*FM (message 894) 0.9965 -0.0254 A, B
*FM (message 873) 0.9965 -0.0254 A, B
*FM (message 872) -2.20159 5.70807 -0.41342 -5.42718 1.19023 A, A+
*FM (message 871) -2.1199 4.7210 -1.0156 -4.2974 1.1035 A, A+
00 S195 -2.30300 5.96942 0.83281 -6.07310 1.30565 A
BASIC S95 1.020 -0.200 B
S85 -2.30300 5.96942 0.83281 -6.07310 1.30565 A
C95 0.9965 -0.0254 B
C55 -2.1199 4.7210 -1.0156 -4.2974 1.1035 A+
*FM = FLEXMODE
Continued on next page
Corrections – blood samples T(continued)T T(continued)
ABL800 FLEX reference manual 2. Amperometric measuring principles
2-11
TpTOB2 B electrode, TContinuedT
The Tp TOB2B measured from the sample is then corrected for systematic deviations from the reference method using the following equation:
)OH(A(sample)OAcorr)(sample,O 21202 pBpp −×+×=
where:
Tp TOB2B(sample) = uncorrected TpTOB2 B value of an expired air sample
Tp TOB2B(sample,corr) = corrected Tp TOB2 B value of an expired air sample
AB0 B
= instrument-dependent correction factor
AB1 B
= instrument-dependent correction factor
TB T
= barometric pressure during the measurement
Tp THB2BO = partial pressure of saturated water vapour = 6.2571 kPa
TABL8xx FLEX T TModeT TAB0 TB TAB1 TB TEquationT
All FLEX and BASIC Expired air 1.016 -0.004 A
When measuring on gas samples, the constant KB1B (describes the gas/liquid relationship for the electrode) is equal to 1.
The following stability criterion must be met to obtain a stable electrode response during TcalibrationT:
)i.upd(sample,Oupd.last)(sample,O 22 pp − ≤ pOB2 B (limit)
This criterion is valid for 1-point calibrations (Gas 2 contains no oxygen) where:
Continued on next page
Corrections – expired air samples
Stability criteria
2. Amperometric measuring principles ABL800 FLEX reference manual
2-12
pOB2 B electrode, TContinuedT
Parameter pOB2 B value from the last updating number...
TABL8x5/8x7 FLEX/00 BASICT TABL8x0 FLEX T
pOB2B(Gas1,upd.last) 92 62
pOB2B(Gas1,upd.i) 86 or 87 56 or 57
(the relationship must be fulfilled for at least one of the updating numbers)
Tp TOB2B(limit) value for the stability criterion is 0.80 kPa/6.0 mmHg.
The following stability criteria must be met in order to obtain a stable electrode response during TmeasurementT:
δ=|pOB2B(sample,upd.30) − pOB2 B(sample,upd.1)|
For δ Criterion
≤ 2.66 kPa 80.0)upd.16(sample,O(sample)O 22 ≤− pp
> 2.66 kPa 6.0upd.6)(sample,O)upd.18(sample,O
)upd.18(sample,O)upd.30(sample,O2.0
22
22 <−−
≤−pppp
For δ > 2.66 kPa: T
if the following criteria are fulfilled, then no result is reported:
0.1upd.6)(sample,O)upd.18(sample,O
)upd.18(sample,O)upd.30(sample,O
22
22 −<−−
pppp
or
6.0upd.6)(sample,O)upd.18(sample,O
)upd.18(sample,O)upd.30(sample,O
22
22 ≥−−
pppp
TUExpired air samplesUT:
Measurement on an expired air sample is accepted if the following criterion is fulfilled:
⏐pOB2 B (sample,upd.30) − pOB2 B (sample,upd.24)⏐≤0.80 kPa/6.0 mmHg,
or
⏐pOB2 B (sample,upd30) − pOB2B (sample,upd.24)⏐≤0.05 × pOB2B (sample,upd.30).
Error message "Measurement unstable" (= pOB2 B response fault during electrode monitoring in Expired air mode) is displayed if the stability criterion is not fulfilled.
Stability criteria T(continued)T
ABL800 FLEX reference manual 2. Amperometric measuring principles
2-13
Glucose and Lactate electrodes
The Glucose electrode (E7066) and the Lactate electrode (E7077) have a similar construction described below.
The electrode consists of a silver cathode and a platinum anode. The electrode is protected by an electrode jacket filled with electrolyte solution and a multilayer membrane mounted at the tip.
The membrane consisting of three layers:
1. outer membrane layer permeable to glucose/lactate.
2. middle enzyme layer.
3. inner membrane layer permeable to HB2 BOB2B.
A polarization voltage of 675 mV is applied to the electrode chain and the current through the chain is measured by an ammeter.
Glucose or lactate molecules are transported across the outer membrane of the multilayer membrane.
The enzyme glucose oxidase or lactate oxidase immobilized between the inner and outer membrane layers converts the glucose or lactate according to the following reactions:
glucose + O B2 B → gluconic acid + H B2 BOB2B
lactate + OB2 B → pyruvate + HB2 BOB2B
OB2 B for this reaction is supplied by the outer membrane layer and also by the oxidation of HB2 BOB2B at the Pt anode.
The HB2 BOB2B produced by the enzyme reaction is transported across the inner membrane to the Pt anode.
Continued on next page
Description
AgCl reference band
Multilayer membrane
Electrolyte
Multilayer membrane
Electrolyte AgCl reference band
Electrode jacket
Electrode contact
Electrode contact
Electrode jacket
2. Amperometric measuring principles ABL800 FLEX reference manual
2-14
Glucose and Lactate electrodes, TContinuedT
HB2 BOB2B → 2HP
+P + OB2 B + 2eP
−P
When a potential is applied to the electrode chain, the oxidation of HB2BOB2 B produces an electrical current proportional to the amount of HB2BOB2 B, which in turn is directly related to the amount of glucose or lactate.
To complete the electrical circuit a reduction reaction (where electrons are consumed) at the cathode converts Ag P
+P
(from AgCl) to Ag:
Ag P
+P + eP
−P → Ag
In order to maintain a charge balance between the anode and the cathode, two AgP
+P
ions need to be reduced for one molecule of HB2BOB2 B to be oxidized.
The zero current is a small background current measured at the electrode when no glucose or lactate is present in a solution. As the Rinse Solution contains no glucose or lactate, a baseline representing the zero current, IB0B as a function of time (IB0B = f(t)), is obtained from continuous measurements on the rinse solution.
Rinse
Time
I (current)
xxxxxx x x x x xxx
Extrapolated baseline
N measurements of I0 on the Rinse Solution
tfinal tmean
I0(t) I0(t)
This IB0B baseline is obtained as follows:
• At the end of a rinse, with the Rinse Solution in the measuring chamber, zero current of the metabolite electrodes is measured periodically (the intervals between these measurements become longer if the analyzer is idle)
• The previous N (N = 8) measurements on the Rinse Solution – before a calibration or a sample measurement starts – are used to obtain a baseline representing the time function of IB0B
Continued on next page
Description T(continued)T
Zero current
ABL800 FLEX reference manual 2. Amperometric measuring principles
2-15
Glucose and Lactate electrodes, TContinuedT
• The baseline is extrapolated thoughout the whole electrode calibration or sample measurement period, and represents the zero current time function
• The I B0B baseline is used to determine the sensitivity of the metabolite electrode
The extrapolated final zero current value at the metabolite electrodes at the last updating (illustrated by the IB0B baseline) is determined as follows:
( ) pA(mean) IttI(final)I 0meanfinal slope10 +−××=A where: TA TB1B = empirical constant dependent on electrode and determined from
tests against the reference method
t BfinalB = time of the last measurement updating on the calibration solution or sample
t Bmean B =
the mean time of the N zero current measurements on the rinse solution:
secN
tt
N
1nn
mean
∑==
where t Bn B is the time of the nP
thP measurement on the rinse solution.
IB0B(mean) =
the zero current at the mean time (tBmean B):
pAN
I(mean)I
N
1nn0,
0
∑==
where I B0,n B is the zero current at the nP
thP measurement on the rinse
solution
IBslopeB = the slope or gradient of the I B0B baseline
( ) ( )
( )pA/second
tt
(mean)IIttI 2N
1=nmeann
0n0,
N
1=nmeann
slope
∑
∑
−
−×−=
If IBslopeB > 0.0, it is set to 0.0
The zero current of the metabolite electrodes should be less than 10,000 pA.
Continued on next page
Zero current T(continued)T
2. Amperometric measuring principles ABL800 FLEX reference manual
2-16
Glucose and Lactate electrodes, TContinuedT
The sensitivities of the metabolite electrodes are calculated by measuring the current on Calibration Solution 1 (Cal 1) and then correcting for the zero current using the extrapolated IB0B baseline.
Cal 1 has a nominal glucose concentration of 10 mmol/L and a nominal lactate concentration of 4 mmol/L. The precise values are batch-individual and contained in the barcodes of the Cal 1 bottles.
The diagram below describes in principle how the sensitivities for the metabolite electrodes are obtained.
I0(final)
t
I
xxx x x x x x x x xxx
Start of Calibration
Extrapolated baseline
N measurements of I0 on rinse solution
End of Calibration
I(Cal 1,upd.2)
I(Cal 1)
∑I(Cal 1,upd.1
I(Cal 1,upd.N) Electrode updatings
The current at the metabolite electrodes with Cal 1 in the measuring chamber, I(Cal B B1), is measured 30 times at regular intervals. The current at the 15th updating is used to determine sensitivity of the glucose electrode, and the current at the 30th updating is used to determine sensitivity of the Lactate electrode.
The current due to the glucose or lactate presence in the sample is then calculated as the difference between the current at the final updating (the 15th for the glucose and the30th for the Lactate electrode) and the zero current at that time point:
I(Cal 1) = I(Cal 1,final) − IB0B(final)T
The sensitivities of the electrodes are calculated as follows:
1)X(Cal1)I(CalSens
c= T T
Continued on next page
Sensitivity
ABL800 FLEX reference manual 2. Amperometric measuring principles
2-17
Glucose and Lactate electrodes, TContinuedT
where:
cX(Cal 1) = actual concentration of glucose/lactate in the Cal 1 solution
IB0B(final) = extrapolated final zero current value of the metabolite electrode at the time of the last updating
TI(Cal 1)T T=T Telectrode current due to presence of glucose/lactateT
The sensitivity limits of the metabolite electrodes are as follows:
Electrode Sensitivity limits
Glucose 100-1800 pA/mM
Lactate 150-2000 pA/mM
The drift in the sensitivity of the metabolite electrodes is calculated from the following equations:
1)X(CalSens
(final)Ifinal)1,I(CalDrift 0 c−
−=
where:
I(Cal 1,final) = current at the final measurement on Cal 1 solution.
Sens = sensitivity of the glucose/lactate electrode from the previous calibration.
cX(Cal 1) = actual concentration of glucose/lactate in the Cal 1 solution.
IB0B(final) = extrapolated final zero current value of the metabolite electrode measured at the time of the last updating
The default drift tolerances set by Radiometer for the metabolite electrodes are:
± 0.5 mM for the Glucose electrode
± 0.2 mM for the Lactate electrode.
Continued on next page
Sensitivity (continued)
Drift
2. Amperometric measuring principles ABL800 FLEX reference manual
2-18
Glucose and Lactate electrodes, TContinuedT
The glucose/lactate concentration in a sample is calculated from the following equation:
Sens(final)II(sample)
X(sample) 0−=c
where:
I(sample) = current of the metabolite electrode measured on the sample
IB0B(final) = extrapolated final zero current value of the metabolite electrode at the time of the last sample updating
Sens = relative sensitivity of the metabolite electrode
The measured metabolite concentration is corrected for systematic deviations from the reference method by the following equations:
TEquation A:
cX(sample, corr)BmacromodeB = AB0,macromodeB × cX(sample) + AB macromodeB
and
TEquation B: T
cX(sample,corr)BmicromodeB = AB0 B, BmicromodeB × cX(sample,corr)B macromodeB + AB1,micromode
where:
cX(sample) = uncorrected measured metabolite concentration from a sample
cX(sample,corr)
= corrected measured metabolite concentration from a sample
AB0 B
= instrument-dependent correction factor
AB1 B
= instrument-dependent cut-off
When an additional correction is needed, equation A is first used together with the constants for the macromode. Then the obtained results are put back into equation B as cX(sample) and then treated again, using the constants for the specific mode.
Note that subscript "macromode" in the equations above is used for the sake of convenience and stands for S250, "FLEXMODE (no message)".
Continued on next page
Measurement
Corrections
ABL800 FLEX reference manual 2. Amperometric measuring principles
2-19
Glucose and Lactate electrodes, TContinued
TCorrections for TTcTTTTGlu:TT
ABL8xx FLEX
Mode AB0 B AB1B Equation
37/27/17 S250 0.937 -0.029 A
C125 0.937 -0.029 A
C35 1.175 -0.05 A, B
*FM (message 903) 1.00 0.00 A, B
*FM (message 902) 1.00 0.00 A, B
*FM (message 901) 1.00 0.00 A, B
*FM (message 900) 1.00 0.00 A, B
*FM (message 899) 1.00 0.00 A, B
*FM (message 898) 1.00 0.00 A, B
35/25/15 S195 0.94 0.1 A
S95 1.00 0.0 A, B
C95 1.06 0.0 A, B
C35 1.16 0.0 A, B
*FM (no message) 0.94 0.1 A
*FM (message 897) 1.06 0.0 A, B
*FM (message 895) 1.06 0.0 A, B
*FM (message 894) 1.06 0.0 A, B
*FM (message 873) 1.06 0.0 A, B
05 S165 0.94 0.1 A
S95 1.00 0.0 A, B
C95 1.06 0.0 A, B
C35 1.16 0.0 A, B
*FM (no message) 0.94 0.1 A
*FM (message 897) 1.06 0.0 A, B
*FM (message 895) 1.06 0.0 A, B
*FM (message 894) 1.06 0.0 A, B
*FM (message 873) 1.06 0.0 A, B
00 BASIC S195 0.94 0.1 A
S95 1.00 0.0 A, B
C95 1.06 0.0 A, B
C35 1.16 0.0 A, B
Continued on next page
Corrections (continued)
2. Amperometric measuring principles ABL800 FLEX reference manual
2-20
Glucose and Lactate electrodes, TContinuedT
Corrections for TcTLac:
ABL8xx FLEX
Mode AB0 B AB1B Equation
37/27/17 S250 0.97 -0.04 A
C125 0.97 -0.04 A
C35 1.202 -0.01 A, B
*FM (message 903) 0.99 0.07 A, B
*FM (message 902) 0.99 0.07 A, B
*FM (message 901) 0.99 0.07 A, B
*FM (message 900) 0.99 0.07 A, B
*FM (message 899) 0.99 0.07 A, B
*FM (message 898) 0.99 0.07 A, B
35/25/15 S195 0.97 -0.04 A
S95 1.03 0.03 A, B
C95 1.03 0.18 A, B
C35 1.13 0.05 A, B
*FM (no message) 0.97 -0.04 A
*FM (message 897) 1.03 0.18 A, B
*FM (message 895) 1.03 0.18 A, B
*FM (message 894) 1.03 0.18 A, B
*FM (message 873) 1.03 0.18 A, B
05 S165 0.97 -0.04 A
S95 1.03 0.03 A, B
C95 1.03 0.18 A, B
C35 1.13 0.05 A, B
*FM (no message) 0.97 -0.04 A
*FM (message 897) 1.03 0.18 A, B
*FM (message 895) 1.03 0.18 A, B
*FM (message 894) 1.03 0.18 A, B
*FM (message 873) 1.03 0.18 A, B
00 BASIC S195 0.97 -0.04 A
S95 1.03 0.03 A, B
C95 1.03 0.18 A, B
C35 1.13 0.05 A, B
*FM = FLEXMODE
Continued on next page
Corrections (continued)
ABL800 FLEX reference manual 2. Amperometric measuring principles
2-21
Glucose and Lactate electrodes, TContinuedT
The following stability criteria must be met to obtain a stable electrode response during calibration:
I(Cal 1,upd.30) − I(Cal 1, upd.21) − 9 × IBslopeB ≤ 0
S Bd,zero B< S Bd,max B
50
upd.21)1,I(Calupd.11)1,I(Calupd.11)1,I(Calupd.1)1,I(Callog
5.9≤
−−
−=τ
All of the three criteria must be fulfilled for a calibration using Cal 1 solution where:
I(Cal 1,upd.30)
I(Cal 1,upd.21)
I(Cal 1,upd.11)
I(Cal 1,upd.1)
= electrode current at the 30P
thP/21P
stP/11P
thP/1 P
stP updating during
measurement on Cal 1 solution, respectively
S Bd,zeroB = spreading of the zero point current updatings around the regression line
S Bd,max B = if Sens > 400 pA/mM, then maxd,S = 0.025 × Sens, otherwise maxd,S = 10.0
τ = Should be less than or equal to 50,
and
upd.21)1, I(Calupd.11)1, I(Calupd.11)1, I(Calupd.1)1, I(Callog
−−
should be negative or equal zero.
The following stability criterion must be met to obtain a stable electrode response during measurement:
S Bd,zero B< S Bd,max B
where:
S Bd,zeroB = spreading of the zero point current updatings around the regression line
S Bd,max B = if Sens > 400 pA/mM, then SBd,max B = 0.025 × Sens,B Botherwise S Bd,max B = 10.0
Continued on next page
Stability criteria
2. Amperometric measuring principles ABL800 FLEX reference manual
2-22
Glucose and Lactate electrodes, TContinuedT
The (glucose or lactate) in the sample is cX(sample,corr).
If the corrected concentration of the metabolite, cX(sample,corr) > 1, the following criteria must be fulfilled:
200.)ero(Iupd.30)I(sample,
I9upd.21)I(sample,upd.30)I(sample,0
0
slope ≤−
×−−≤
z
otherwise
14.0Sens
I9upd.21)I(sample,upd.30)I(sample, slope ≤×−−
where:
I(sample,upd.30)
I(sample,upd.21)
= electrode current at the 30th/21st updating during measurement on sample, respectively
IB0B(zero) = zero current extrapolated to the time of the measurement
If all the criteria below are fulfilled, then the result of the measurement will be marked with an interference error.
1upd.9)I(sample,-upd.16I(sample,upd.23)I(sample,-upd.30)I(sample,
≥
I(sample,upd.16) > I(sample,upd.12)
I(sample,upd.12) > I(sample,upd.9)
cX(sample,corr)> 1.5 mmol/L
where:
I(sample,upd.30) I(sample,upd.23)B BI(sample,upd.16)B BI(sample,upd.12) I(sample,upd.9)
= electrode current at the 30th/23rd/16th/12th/9th updating during measurement on sample, respectively
cX(sample,corr) = corrected concentration of glucose or lactate in the sample
Stability criteria (continued)
ABL800 FLEX reference manual 2. Amperometric measuring principles
2-23
Crea electrodes
Blood contains both creatinine and creatine. In order to measure creatinine, a two-electrode system is used where one electrode (E8089) measures both substances, and the other electrode (E8088) measures creatine only.
The measurement of creatinine is done by subtracting the signals from the two electrodes and by compensating for any interferences in the blood sample. The compensation requires that the electrode jackets have equal properties. To ensure this, the electrode jackets are packed in pairs in the membrane boxes and the electrodes may only be remembraned, using a pair of electrode jackets from the same membrane box.
Both electrodes: Crea A, E8088 (brown with text in ivory), and Crea B, E8089 (ivory with text in brown), have a similar construction: each consists of a silver cathode and a platinum anode. The electrode is protected by an electrode jacket filled with electrolyte solution and a multilayer membrane mounted at the tip.
The membrane consists of three layers:
1. Outer membrane layer permeable to creatinine and creatine.
2. Middle enzyme layer.
3. Inner membrane layer permeable to HB2 BOB2 B.
A polarization voltage of 675 mV is applied to the electrode chain; the current through the chain is measured by an ammeter.
Continued on next page
Introduction
Description
AgCl reference band
Multilayer membrane
Electrolyte
Electrode jacket
Electrode contact
2. Amperometric measuring principles ABL800 FLEX reference manual
2-24
Crea electrodes, TContinuedT
Creatinine and creatine molecules are transported across the outer layer of the multilayer membranes.
The enzymes creatininase (Crea B, E8089 electrode), creatinase and sarcosine oxidase (Crea A, E8088 electrode) immobilized between the inner and outer membrane layers change creatinine/creatine as follows:
Creatinine + HB2 BO ↔ creatine (creatinine amidohydrolase [1])
creatine + HB2 BO → sarcosine + urea (creatine amidohydrolase)
Sarcosine + H B2 BO + OB2 B → glycine + formaldehyde + HB2 BOB2B
(Sarcosin oxidase)
The Crea B electrode (E8089) contains all three enzymes and detects both creatinine and creatine (solid line). The Crea A electrode (E8088) contains only the last two enzymes in the cascade and detects only creatine (dotted line).
OB2 B for these reactions is supplied h the outer membrane layer and also by the oxidation of HB2 BOB2B at the Pt anod
The HB2 BOB2B p by the enzyme reaction is transported across the inne membrane t
In order to mions are red
Description (continued)
roduced
o the Pt anode.aintain a charge baluced for each molecu
througe.
HB2 BOB2B → 2HP
+P + OB2 B + 2eP
−P
When a potential is applied to the echain, the oxidation of HB2BOB2 B produelectrical current proportional to thamount of HB2BOB2 B, which in turn is drelated to the amount of creatininecreatine for the Crea B and the amcreatine for the Crea A electrodes.
To complete the electrical circuit areduction reaction (where electronconsumed) at the cathode converts(from AgCl) to Ag:
Ag P
+P + eP
−P → Ag
ance between the anode and the cathodele of HB2 BOB2B, which is oxidized.
Continued o
r
lectrode ces an e irectly and ount of
s are Ag P
+P
, two AgP
+P
n next page
ABL800 FLEX reference manual 2. Amperometric measuring principles
2-25
Crea electrodes, TContinuedT
S1827 Calibration Solution 1 contains creatinine and is used for calibration of the Crea B electrode.
S1837 Calibration Solution 2 contains creatine and is used for calibration of the Crea A and Crea B electrodes.
The precise start concentrations are contained in the barcodes of the S1827 and S1837 Calibration Solutions that are scanned before the calibrating solutions are installed on the analyzer.
The zero current (IB0B) is a small background current measured without presence of creatinine and creatine (i.e. Rinse Solution). Immediately before the electrode calibration or sample measurement, the measuring chamber is flushed with Rinse solution. The zero current is therefore obtained from continuous updatings during this period. The baseline is then extrapolated (Slope 0) throughout the measurement period for the calibration of the signal, originating from creatinine and/or creatine in the calibration solution or sample.
Slope 1
Slope 2
Electrode updatings
Measurement Zero current
Slope 0
The following areas of the baseline (before measurement) are used:
• Slope 1 = the slope of the baseline between updatings 33 to 57
• Slope 2 = the slope of the baseline between updatings 21 to 45
Slope 0 is calculated, using Slope 1 and Slope 2.
Slope 0 depends on the calculated Slope 1 as follows:
• For Slope 1 > 0.5, Slope 0 = (1+k BohB) × Slope 1 – kBohB Slope 2
• For 0.5 ≤ Slope 1 ≤ -0.5, Slope 0 = 0
• For Slope 1 < -0.5, Slope 0 = (1+k BohB) × Slope 1 – kBohB Slope 2
k BohB is an experimentally determined constant that equals 0.5
Zero current for data point n, IB0 B(n), is then determined, using the following equation:
IB0B(n) = IB0 B(47;57)BmedianB + Slope 0 × (38 - n)
where
IB0B (47;57)Bmedian B= median of the zero current between updatings 47 to 57
The zero current of the creatinine electrodes should be less than 10.000 pA. Continued on next page
Calibration material
Zero current
2. Amperometric measuring principles ABL800 FLEX reference manual
2-26
Crea electrodes, TContinuedT
During calibration the sensitivities of Crea A and Crea B electrodes are calculated.
For creatinine and creatine the following spontaneous reaction occurs in aqueous solution:
creatinine + H B2 BO ↔ creatine.
It means that the nominal concentrations of creatinine and creatine in the calibration solutions vary with time and temperature. The analyzer automatically calculates the actual concentrations from the time the calibration solutions are installed and from the ambient temperature (can be specified by the user). Before the calibration solutions are installed on the analyzer, the substrates are added to ensure that the nominal start concentrations are known.
During a 2-point calibration the sensitivities are calculated as follows:
The Crea B electrode is calibrated, using both calibration solutions.
I BBB(Cal1) – IB0,BB(Cal1) = SensBB,Creatinine B× cCreatinine(Cal1) +SensBB,creatine B× cCreatine(Cal1)
I BBB(Cal2) – IB0,BB(Cal2) = SensBB,Creatinine B× cCreatinine(Cal2) + SensBB,Creatine B× cCreatine(Cal2)
where
IBBB(Cal1) = the Crea B electrode signal obtained on Calibration Solution 1
IB0,BB(Cal1) = zero current of Crea B for measurement on Calibration solution 1
IBBB(Cal2) the Crea B electrode signal obtained on Calibration Solution 2
IB0,BB(Cal2) = zero current of Crea B for measurement on Calibration solution 2
cCreatine (Cal1) = actual concentration of creatine (μmol/L) in Calibration Solution 1
cCreatinine (Cal1) = actual concentration of creatinine (μmol/L) in Calibration Solution 1
cCreatine(Cal2) = actual concentration of creatine (μmol/L) in Calibration Solution 2
cCreatinine(Cal2) = actual concentration of creatinine (μmol/L) in Calibration Solution 2
SensBB,CreatinineB = sensitivity of the Crea B electrode to creatinine (pA(mM))
SensBB,CreatineB = sensitivity of the Crea B electrode to creatine (pA(mM))
Continued on next page
Sensitivity
ABL800 FLEX reference manual 2. Amperometric measuring principles
2-27
Crea electrodes, TContinuedT
For the Crea A electrode (calibrated on Calibration Solution 2):
IBAB(Cal2) – IB0,AB(Cal2) = SensBA,Creatine B× cCreatine(Cal2)
where
IBAB(Cal2) = the Crea A electrode signal obtained on Calibration Solution 2
I B0,AB(Cal2) = zero current of the Crea A for measurement on Calibration solution 2
cCreatine (Cal2) = actual concentration of creatine (μmol/L) in Calibration Solution 2
SensBA,CreatineB = sensitivity of the Crea A electrode to creatine (pA(mM))
Sensitivity of the electrodes is determined from their signals corrected for their zero currents as follows:
(1):
[ ] [ ]1)(CalCreatine2)(CalCreatinine2)(CalCreatine1)(CalCreatinine
(Cal1)Creatine(Cal2)I(Cal2)I(Cal2)Creatine(Cal1)I(Cal1)ISens B0,BB0,B
CreatinineB, cccccc
×−×
×−−×−=
(2):
[ ] [ ]1)(CalCreatine2)(CalCreatinine2)alCreatine(C1)(CalCreatinine(Cal2)Creatinine(Cal1)I-(Cal1)I(Cal1)Creatinine(Cal2)ICal2)(I
Sens B0,BB0,B
CreatineB, cccccc
×−×
×−×−=
(3):
2)alCreatine(C(Cal2)I(Cal2)I
Sens A0,A
CreatineA, c−
=
During the 1-point calibration only SensBB,creatinineB is determined. The sensitivity cannot be determined directly from equation (1). The following equation, which uses the sensitivities determined during the previous 2-point calibration (Cal2, prev), is then used:
(4):
[ ]1) (CalCreatine
prev)(Cal2,Sensprev)(Cal2,Sens
1) (CalCreatinine
(Cal1)ICal1)(ISens
CreatinineB,
CreatineB,
B0,B
CreatinineB,
cc ×+
−=
Continued on next page
Sensitivity (continued)
2. Amperometric measuring principles ABL800 FLEX reference manual
2-28
Crea electrodes, TContinuedT
The following sensitivity limits are used:
Electrode Sensitivity Definition Limits
Crea B Sens 1
Sens 2
SensBB,Creatinine B
%100SensSens
CreatinineB,
CreatineB, ⋅
5-15 pA/μM
65-85 %
Sens 3 %100
SensSens
CreatinineB,
CreatineA, ⋅ 50-200 % Crea A
Sens A SensBA,Creatinine B 3,5-20 pA/μM
S1827 Calibration Solution 1 does not contain creatine. So if the Crea A electrode measures creatine during calibration on the Calibration Solution 1, the reason may be that:
• electrodes were placed in the wrong chambers in the Met II module
• calibration solutions were interchanged
• calibration solutions were not stored as instructed, and some creatinine in the Calibration Solution 1 has been converted to creatine
During each calibration, the performance of the Crea A and Crea B electrodes are tested. If performance of either of the electrodes is not satisfactory it will be indicated.
The drift in the sensitivity of the Crea A and Crea B electrodes are calculated from the following equations:
(5):
[ ]
t)(Cal1,Creatinine1)(tSens
t)(Cal1,Creatine1)(tSenst)(Cal1,It)Cal1,(IDrift
CreatinineB,
CreatineB,B0,B
CreatineB,
c
c−
−
×−−−=
The nominal creatine concentration in the Calibration Solution 1 is used in equation (5).
Drift in the creatine sensitivity of the Crea B electrode is determined only during the 2-point calibration:
(6):
[ ]
t)al2,Creatine(C1)(tSens
t)(Cal2,Creatinine1)(tSenst)(Cal2,It)Cal2,(IDrift
CreatineB,
CreatinineB,B0,B
CreatineB,
c
c−
−
×−−−=
The nominal creatinine concentration in the Calibration Solution 2 is used in equation (6).
Continued on next page
Sensitivity limits
Additional checks during calibration
Drift
ABL800 FLEX reference manual 2. Amperometric measuring principles
2-29
Crea electrodes, TContinuedT
Drift in the creatine sensitivity of the Crea A electrode is determined only during the 2-point calibration:
(7):
[ ]
t)al2,Creatine(C1)(tSens
t)(Cal2,Creatinine1)(tSenst)(Cal2,It)Cal2,(IDrift
CreatineA,
CreatinineA,A0,A
CreatineA,
c
c−
−
×−−−=
The default drift tolerances set by Radiometer are:
Electrode Drift tolerances
Crea B ±15 μM creatinine in Calibration Solution 1
±15 μ M creatine in Calibration Solution 2
Crea A ±15 μ M creatine in Calibration Solution 2
The concentration of creatine in the sample is determined from the electrode response on the Crea A electrode as follows:
(8): CreatineA,
A0,A
uncorr SensI(sample)I
(sample)Creatine−
=c
Since the carry-over from the Rinse Solution for creatine is the same for the Crea A and Crea B electrodes, the uncorrected creatine concentration from equation 8 is then used to obtain the creatinine concentration in the sample:
(9): CreatinineB,
uncorrCreatineB,B0,Buncorr Sens
(sample)CreatineSens)I(sample)(I(sample)Creatinine
cc
×−−=
Continued on next page
Drift (continued)
Measurements and corrections
2. Amperometric measuring principles ABL800 FLEX reference manual
2-30
Crea electrodes, TContinuedT
Then the measured creatinine concentration is corrected for carry-over from the Rinse Solution as follows:
Equation A:
cCrea(sample,corr) B macromodeB = AB0,macromodeB × cCreatinineBuncorrB(sample) + AB1,macromodeB
Equation B:
cCrea(sample,corr) BmicromodeB = AB0,micromodeB × cCrea(sample)BmacromodeB + AB1,micromodeB
where:
cCrea(sample,corr) = corrected measured creatinine concentration from a sample
cCreatinineBuncorrB(sample) = uncorrected measured creatinine concentration from a sample
AB0 B = instrument-dependent correction factor
AB1 B = instrument-dependent cut-off
Analyzer Mode AB0 B AB1 B Equation
37/27/17 S250 1.0262 0.4211 A
C125 1.0262 0.4211 A
*FM (message 903) 1.0819 -3.8515 A, B
*FM (message 902) 1.0819 -3.8515 A, B
*FM (message 901) 1.0819 -3.8515 A, B
*FM (message 900) 1.0819 -3.8515 A, B
*FM (message 899) 1.0819 -3.8515 A, B
*FM (message 898) 1.0819 -3.8515 A, B
*FM = FLEXMODE
The ABL8x7 FLEX analyzers are designed to measure cCrea on whole blood. If plasma, serum or NIST SRM is to be measured on the ABL8x7 FLEX analyzers, cCrea must be corrected as follows:
cCrea(sample,corr) (ABL8x7,plasma/serum) μmol/L = 0.950 × cCrea(ABL8x7,whole blood) – 0.4
cCrea(sample,corr) (ABL8x7,plasma/serum) mg/dL = 0.950 × cCrea(ABL8x7,whole blood) – 0.005
where
cCrea(ABL8x7,whole blood) is the result obtained on the analyzer.
Measurements and corrections (continued)
Whole blood correction
ABL800 FLEX reference manual 2. Amperometric measuring principles
2-31
References
1. Yamamoto K, Oka M, Kikuchi T, Emi S. Cloning of the creatinine amidohydrolase gene from Pseudomonas sp. PS-7; Biosci Biotechnol Biochem 1995;59:1331-32
List of references
2. Amperometric measuring principles ABL800 FLEX reference manual
2-32
3. Optical measuring principles
Overview
This chapter describes the optical system, its construction and the measuring method used.
This chapter contains the following topics.
Optical system.................................................................................................. X3-2X
Correcting for interferences ............................................................................. X3-7X
Measurement and corrections .......................................................................... X3-9X
References ........................................................................................................ X3-15X
Introduction
Contents
3. Optical measuring principles ABL800 FLEX reference manual
3-2
Optical system
The optical system of the ABL800 FLEX analyzer is designed to measure the following parameters:
Parameter Description
TcTtHb Concentration of total hemoglobin
sOB2 B Oxygen saturation
FOB2 BHb Fraction of oxyhemoglobin
FCOHb Fraction of carboxyhemoglobin
FHHb Fraction of deoxyhemoglobin
FMetHb Fraction of methemoglobin
FHbF Fraction of fetal hemoglobin
TcTtBil Concentration of total bilirubin (the sum of unconjugated and conjugated bilirubin) in plasma
TNOTE: TT cTtBil T can be measured on a whole-blood or plasma sample. Plasma samples provide the optimal measurement performance. To obtain optimal accuracy when following a patient trend in cTtBilT, use the same aspiration mode and the same analyzer.
THematocrit (THct T) is also available as a derived parameter. T
The optical system is based on a 128-wavelength spectrophotometer with a measuring range of 478-672 nm. The spectrophotometer is connected via an optical fiber to a combined hemolyzer and measuring chamber.
Sample in
Concave grating Photodiode array
Slit
Optical fiber
Hemolyzer
Lens
Infrared filter
Cuvette
Sample out
Spectrophotometer
Lamp unit
Continued on next page
Measured parameters
Construction
ABL800 FLEX reference manual 3. Optical measuring principles
3-3
Optical system, TContinuedT
The method used in the analyzer's optical system is visible absorption spectroscopy.
Step Description
T1 T The blood sample is transported to the cuvette positioned in the hemolyzer unit. The temperature of the cuvette is regulated to 37 P
oPC.
T2 T 1 μL of the sample is ultrasonically hemolyzed in the cuvette at a frequency of about 30 kHz in order to rupture the walls of the red blood cells so that their content is mixed with the blood plasma, giving an optically clear solution. There is no bilirubin in the red blood cells, so after hemolyzation the red blood cell intracellular fluid dilutes the plasma bilirubin. The calculation discussed in TMeasurement and CorrectionsT corrects for this dilution.
To eliminate air bubbles in the sample and to enhance hemolyzation, an overpressure of one atmosphere is maintained throughout hemolyzation and measurement.
T3 T Light from a 4-Watt halogen lamp is sent to the cuvette via an infra-red filter and a biconvex lens.
The voltage across the halogen lamp is regulated by a thermostatted photodiode so that the amount of light sent to the cuvette has a constant intensity.
T4 T The light transmitted through the cuvette is guided to the spectrophotometer via an optical fiber.
T5 T The light passes through a slit that directs it towards a combined mirror and concave grating.
T6 T The grating separates the light into 128 single wavelengths and the mirror focuses the 128 light signals on a photodiode array.
T7 T The photodiode array has 128 diodes or pixels, one for each wavelength, which convert the monochromatic light signals to currents.
T8 T The currents and therefore the intensity of the light signals are measured at each of the 128 diodes, which form the basis for the absorption spectrum for a particular sample.
T9 T The spectrum is sent to the analyzer’s computer, where the calculations of the oximetry parameter values are made.
Continued on next page
Construction T(continued)
3. Optical measuring principles ABL800 FLEX reference manual
3-4
Optical system, TContinuedT
Absorption spectroscopy is based on Lambert-Beer's law which states that the measured absorbance for a single compound is directly proportional to the concentration of the compound and the length of the light path through the sample [1]:
A c ly yλ λε= × ×y
where: Ay
λ = absorbance of compound y at wavelength λ
ελy = extinction coefficient of compound y at wavelength λ (a constant,
characteristic of the compound)
cy = concentration of compound y in sample
T l T = length of the light path
The absorbance (A) of a compound is defined as the logarithm of the ratio of the light intensity before and after transmission through the compound. In practice it is the logarithm of the ratio of the light intensity transmitted through water to the light intensity transmitted through the compound.
AII
= log 0
where:
I0 = intensity of light transmitted through water (TI B0 TB is measured as the intensity of light transmitted through the Cal 1 or Cal 2 solutions)
I = intensity of light transmitted through the compound
For samples containing more than one optically active compound, the total absorbance (A BtotalB) is the sum of the individual compounds’ absorbance, since absorbance is an additive quantity.
For example, if a sample contains six compounds yB1B, y B2B, ….y B6 B, the total absorbance measured for that sample at wavelength λB1 B is:
A A A A A A Atotal y y y y y y1 2 3 4 5 6
λ λ λ λ λ λ λ1 1 1 1 1 1 1= + + + + +
( )= + + + + +l c c c c c cy y y y y y y y y y y y1 1 2 2 3 3 4 5 5 6 6ε ε ε ε ε ελ λ λ λ λ λ1 1 1 1
4
1 1
If there are Y compounds and measurements are taken at Tn T wavelengths, a general expression can be written for TATBtotalB at the wavelength λBnB:
A c ln ntotal y y
y
Yλ λε= × ×
=∑
1
where: λBn B = the individual wavelengths.
Continued on next page
Lambert-Beer’s law
Absorbance
Total absorbance
ABL800 FLEX reference manual 3. Optical measuring principles
3-5
Optical system, TContinuedT
nAλtotal can be depicted graphically as a function of wavelength, and if the
differences between the wavelengths are small enough, a continuous spectrum is produced.
TEXAMPLES: The figure below shows three spectra; pure O B2 BHb, pure HHb in a low concentration, a spectrum of 92 % oxygenated hemoglobin obtained by adding the spectra of OB2 BHb and HHb. The additivity of absorption and the continuity of the spectra can clearly be seen.
480 500 520 540 560 580 600 620 640 660 680
O2Hb (9.2 mmol/L)HHb (0.8 mmol/L)92 % oxygenated hemoglobin (i.e., 92 % O2Hb + 8 % HHb)
Wavelength/nm
Absorption
Example of the spectrum obtained from unconjugated bilirubin at concentration of 200 μmol/L.
The spectrum of conjugated bilirubin is slightly different.
Continued on next page
Continuous spectrum
200umol/L Unconjugated Bilirubin in Plasma
0
0.02
0.04
0.06
0.08
0.1
470 520 570 620 670
nm
Abs
orba
nce
3. Optical measuring principles ABL800 FLEX reference manual
3-6
Optical system, TContinuedT
In the spectrum taken of a sample, the absorption recorded at each wavelength contains contributions from each of the compounds in the sample. The task then is to determine the magnitude of that contribution and thereby the concentration of each compound in the sample.
The concentrations are determined using the following equation:
c An n
ny y total
=1
128
K=∑ λ λ
where: Ky
nλ
= a constant specific to compound y at wavelength λBnB.
The constants ( Kynλ ) are determined using Multivariate Data Analysis [2] where
the spectra of the calibration compounds were considered together with the reference values of the calibration compounds. The essential interfering substances were also taken into account.
Determining concentrations
Matrix of constants
ABL800 FLEX reference manual 3. Optical measuring principles
3-7
Correcting for interferences
Fetal hemoglobin (HbF) does not have the same spectrum as adult hemoglobin (HbA) due to a slight variation in molecular structure. The presence of HbF in a sample will interfere with the result if it is not corrected for.
It is thus important when measuring hemoglobin levels in premature neonates and neonates aged 0 to 3 months, as well as adults suffering from thalassemia, to take into account this difference [3].
The ABL800 FLEX analyzers automatically correct for HbF.
TNOTE: TT Hb types other than HbA and HbF interfere with haemoglobin measurements and are not compensated for in the ABL800 FLEX analyzers.
The diagram below shows the transition from fetal hemoglobin to adult hemoglobin [4].
This graph is only schematic and cannot be used to determine TFTHbF.
If the difference between the two types of hemoglobin is not accounted for in measurements on samples containing HbF, e.g. from premature neonates and neonates aged 0 to 3 months, then a deviation in the measurement will arise.
The deviation is most important for measurements of oxygen saturation (sOB2B) and the fraction of carboxyhemoglobin (FCOHb), since inaccurate measurements of these parameters can lead to incorrect diagnostic interpretation of the results, and consequent risk of inappropriate treatment.
The presence of HbF in a sample is detected from the difference spectrum between fetal and adult oxyhemoglobin. From the size of the difference spectrum the concentration of fetal oxyhemoglobin, TcTOB2 BHbF, can be measured.
The amount of TcTOB2 BHbF exceeding a certain level indicates HbF interference. The analyzer automatically corrects for this interference by subtracting the difference spectrum of fetal oxyhemoglobin from the measured spectrum. It then makes further calculations, using TcTOB2 BHbF to measure TFTHbF.
Continued on next page
HbF versus HbA
Deviation of results
Detecting HbF
Correcting for HbF
3. Optical measuring principles ABL800 FLEX reference manual
3-8
Correcting for interferences, TContinuedT
Fetal hemoglobin and non-hemoglobin substances present in blood that absorb light within the same wavelength range used to measure the oximetry parameters and bilirubin, will interfere with the true spectra of the blood samples.
The optical system in the ABL800 FLEX analyzers compensates for the most likely interfering substances by repressing their spectra.
The interference from the following substances the analyzer compensates for when measuring the oximetry parameters:
Intralipids (turbidity)
Sulfhemoglobin, SHb
Repressing the spectra of the likely interfering substances is done in two ways depending on the substance:
• TEitherT the substance is taken account of in the calculation of the matrix of constants, K (see the section TMeasuring PrincipleT in this chapter). This applies to Intralipids and Sulfhemoglobin,
• TOrT the substance is detected, and the measured spectrum is corrected accordingly. This applies to HbF.
A measured spectrum is compared to a model spectrum calculated from the determined concentrations. The difference between the two spectra is then called the residual spectrum. If the difference is too high, a warning (Oxi spectrum mismatch) is issued on all the oximetry module parameters ctHb, sOB2 B, FOB2BHb, FCOHb, FMetHb, FHHb, FHbF and TcTtBil.
The same action is taken if one of the following conditions exists and FHb BderivB is defined as one of the parameters sO2, FOB2 BHb, FCOHb, FMetHb, FHHb:
• ctHb<−0.1 mmol/L or ctHb>25 mmol/L.
• FHb(deriv)<–2% or FHb(deriv)>102%.
• Negative fraction of SHb<−2% is detected.
• Value of Turbidity<−0.5%.
Most likely interfering substances
Repressing spectra
Residual spectrum
ABL800 FLEX reference manual 3. Optical measuring principles
3-9
Measurement and corrections
The oximetry parameters are calculated as follows:
Parameter Equation
ctHb(meas) = cOB2 BHb + cCOHb + cHHb + cMetHb
sOB2 B
=c
cO Hb
eHb2
TcTeHb = TcTHHb + TcTOB2 BHb (effective hemoglobin)
FOB2 BHb =
ccO HbtHb
2
FCOHb =
ccCOHb
tHb
FHHb =
ccHHbtHb
FMetHb =
ccMetHb
tHb
FHbF = ccHbFtHb
where:
cOB2 BHb = concentration of oxyhemoglobin in the sample
cCOHb = concentration of carboxyhemoglobin in the sample
cHHb = concentration of deoxyhemoglobin in the sample
cMetHb = concentration of methemoglobin in the sample
TcTHbF = concentration of fetal hemoglobin in the sample
Bilirubin is calculated as follows:
Hct(calc)1tBil(B)tBil(P)
−=
cc
where:
ctBil(P) = concentration of total bilirubin in plasma
ctBil(B) = concentration of diluted plasma bilirubin after sample hemolyzation
Hct(calc) = calculated hematocrit (a fraction).
Continued on next page
Oximetry parameters
Bilirubin
3. Optical measuring principles ABL800 FLEX reference manual
3-10
Measurement and corrections, TContinuedT
Hct(calc)g / dL
tHb= ×0 0301. c
For further details on Hct(calc) please refer to TInterference Tests T and the explanation of MCHC (Mean Corpuscular Hemoglobin Concentration) in chapter 5 T Tin this manual.
The following parameters will not be calculated:
Parameter Is not calculated if…
sOB2 B, FCOHb, FMetHb, FHHb
TcTeHb = cHHb + cOB2 BHb < 0.75 mmol/L;
ctHb < 1 mmol/L
TcTtBil ctHb > 15.5 mmol/L
The following conditions are required to exclude HbF interference:
Parameter or Feature Requirement
TcTeHb > 3 mmol/L
FCOHb < 15 %
FMetHb < 10 %
"HbF correction" has not been activated
If ctHb < 5 mmol/L, TcTOB2 BHbF should be more than 1 mmol/L.
If ctHb > 5 mmol/L, TcTOB2 BHbF/ctHb should be more than 0.2.
"HbF correction" has been activated
No lower limit value for TcTOB2 BHbF is required, i.e. even adult blood samples will be corrected for HbF.
It may be of value when analyzing blood samples from newborns who received adult blood transfusion. In these cases FHbF can be lower than 20 % and significant deviations of oximetry parameters and bilirubin can occur.
HbF suppression has been activated
The FHbF value is displayed by the ABL835/30 FLEX analyzers.
Message “HbF detected” is displayed on the other analyzer versions with the oximetry module installed.
sOB2 B < 50 % or
ctHb < 5 mmol/L
Message “FHbF measurement is not possible” is displayed by the ABL835/30 FLEX analyzers if a HbF suppression has been activated.
Continued on next page
Bilirubin T (continued)T
Restrictions
ABL800 FLEX reference manual 3. Optical measuring principles
3-11
Measurement and corrections, TContinuedT
The uncorrected hemoglobin concentration, ctHb(sample), measured on capillary or syringe samples is corrected as follows: TEquation A:
dilcuv FF)tHb(samplecorr),tHb(sample cc =
where: ctHb(sample,corr) = corrected ctHb
F BcuvB = analyzer-dependent constant determined at tHb calibrations and automatically stored by the analyzer
F BdilB = analyzer-dependent constant determined during tests against the reference method, which corrects for Hb dilution in the different aspiration modes.
TABL8xx FLEX T
TModeT TF BdilTB TEquationT
37/27/17 S250 1.0000 A
C125 1.0110 A
*FM (message 903) 0.9714 A
*FM (message 902) 0.9714 A
*FM (message 901) 0.9714 A
*FM (message 900) 0.9714 A
*FM (message 899) 0.9714 A
*FM (message 898) 0.9714 A
*FM = FLEXMODE
Continued on next page
Corrections for ctHb
3. Optical measuring principles ABL800 FLEX reference manual
3-12
Measurement and corrections, TContinuedT
ABL8xx FLEX
Mode Fdil Equation
35/25/15 S195 1.0000 A
S95 0.9630 A
S85 1.0050 A
C95 0.9630 A
C55 0.9220 A
C35OXI 0.9570 A
*FM (no message) 1.0110 A
*FM (message 897) 0.9630 A
*FM (message 895) 0.9630 A
*FM (message 894) 0.9630 A
*FM (message 873) 0.9630 A
*FM (message 872) 0.9490 A
*FM (message 871) 0.9440 A
*FM (message 870) 0.9230 A
*FM (message 869) 0.9230 A
30/20/10 S85 1.0050 A
10 BG only C55 0.9220 A
FLEX C35 OXI 0.9570 A
*FM (no message) 0.9750 A
*FM (message 872) 0.9490 A
*FM (message 871) 0.9440 A
*FM (message 870) 0.9230 A
*FM (message 869) 0.9230 A
00 BASIC S195 1.0000 A
S95 0.9630 A
S85 1.0050 A
C95 0.9630 A
C35OXI 0.9570 A
*FM = FLEXMODE
Continued on next page
Corrections for TcTtHb T(continued)
ABL800 FLEX reference manual 3. Optical measuring principles
3-13
Measurement and corrections, TContinuedT
The uncorrected total bilirubin concentration, ctBil(sample), measured on capillary or syringe samples is corrected as follows: TEquation A:
dilcuv F Fe)tBil(samplcorr)e,tBil(sampl cc =
where: ctBil(sample,corr) = corrected ctBil
F BcuvB = analyzer-dependent constant determined at tHb calibrations and automatically stored by the analyzer
F BdilB = analyzer-dependent constant determined during tests against the reference method, which corrects for TcTtBil dilution in the different aspiration modes.
TABL8xx T TModeT TF BdilTB TEquationT
37 S250 1.0050 A
C125 0.9900 A
*FM (message 903) 0.9432 A
*FM (message 902) 0.9432 A
*FM (message 901) 0.9432 A
*FM (message 900) 0.9432 A
*FM (message 899) 0.9432 A
*FM (message 898) 0.9432 A
*FM = FLEXMODE
Continued on next page
Corrections for TcTtBil
3. Optical measuring principles ABL800 FLEX reference manual
3-14
Measurement and corrections, TContinuedT
ABL8xx Mode Fdil Equation
35 S195 1.0050 A
S95 0.9320 A
S85 1.0000 A
C95 0.9320 A
C55 0.8640 A
C35 OXI 0.9160 A
*FM (no message) 0.9900 A
*FM (message 897) N/A
*FM (message 895) N/A
*FM (message 894) N/A
*FM (message 873) N/A
*FM (message 872) N/A
*FM (message 871) N/A
*FM (message 870) N/A
*FM (message 869) N/A
30 S85 1.0000 A
C55 0.8640 A
C35 OXI 0.9160 A
*FM (no message) 0.9570 A
*FM (message 872) N/A
*FM (message 871) N/A
*FM (message 870) N/A
*FM (message 869) N/A
*FM = FLEXMODE
Corrections for TcTtBil T(continued)
ABL800 FLEX reference manual 3. Optical measuring principles
3-15
References
The list of the references for Tchapter 3, The Optical Measuring Principles:T
1. Ewing GW. Instrumental methods of chemical analysis. 5th ed. McGraw-Hill, 1985.
2. Martens H. Multivariate calibration: quantitative interpretation of non-selective chemical data. Dr. Techn. Thesis, NTH Univ. of Trondheim, 1986.
3. Krzeminski A. Why correct for fetal hemoglobin in blood oximetry measurements? Radiometer Publication Info. No. 1992-3. Copenhagen: Radiometer Medical A/S, 1992.
4. Huehns ER, Beaven GH. Developmental changes in human hemoglobins. Clin Dev Med 1971; 37: 175-203.
List of references
3. Optical measuring principles ABL800 FLEX reference manual
3-16
4-1
4. User-defined corrections
Overview
This chapter describes the basis of the user-defined corrections available for all the measured parameters.
This chapter contains the following topics.
General information ......................................................................................... X4-2X
Correction factors for oximetry parameters and bilirubin................................ X4-4X
Electrolyte and metabolite parameters ............................................................. X4-7X
Introduction
Contents
4. User-defined corrections ABL800 FLEX reference manual
4-2
General information
User-defined corrections are most commonly implemented in situations where the values measured for a particular parameter by two or more analyzers, deviate consistently from each other.
TNOTE: T Since the performance of all analyzers is tested as described in chapter 5, Performance Characteristics, and each instrument is assumed to operate accurately and optimally, the unnecessary correction of parameter values by the user can lead to inaccurate measurements being reported.
User-defined corrections are based on a linear correlation between the measured values (without user-defined corrections) and the displayed values (with user-defined corrections).
The correction factors for each measured parameter are the slope and the offset of the correction line. With user-defined corrections it is possible to change the values of either one or both of these correction factors, depending on the parameter type.
Corrected value = Slope × Uncorrected value + Offset
The diagram below is a schematic representation of the relationship between correction lines without and with user-defined correction.
Continued on next page
Purpose of use
User-defined corrections
Displayed (corrected) parameter value
Measured (uncorrected) parameter value
Offset
Correction line without user correction
Correction line with user correction
Slope = 1
0.0
ABL800 FLEX reference manual 4. User-defined corrections
4-3
General information, TContinuedT
The slope/offset for each parameter are configured in the General Setup > Parameters Setup. User-corrected values are marked with a “*” after the result.
NOTE: The user-defined corrections will be applied to measurements on QC solution unless the "Apply parameter corrections to QC" option was deactivated in Miscellaneous Setup.
For detailed instructions on how to enter user-defined corrections, refer to the section TParameter Setup T in Tchapter 3 T of the TOperator’s Manual T.
Entering user-defined corrections
4. User-defined corrections ABL800 FLEX reference manual
4-4
Correction factors for oximetry parameters and bilirubin
The following corrections can be user-defined for the oximetry parameters and bilirubin:
Parameter Allowed user-defined corrections
TSlopeT TOffset T
TcTTtHb T Yes No
TsTTOB2 TB Yes Yes
TF TTCOHb T No Yes
TF TTMetHb T No Yes
TF TTOB2 BHb T No No
TF TTHHb T No No
TFHbF T Yes Yes
TcTTtBilT Yes Yes
NOTE: In order to define the corrections accurately, the measurements of the oximetry parameters and bilirubin on the ABL800 FLEX analyzers should be made without any entered corrections. To avoid truncation errors from an enabled “Out of range suppression” function it is important to disable the function.
The following recommendations apply to ctHb:
Item Description
Units g/dL; g/L; mmol/L
Sample Set TcTtHb of a SAT100 sample to ≈15 g/dL (9.3 mmol/L) and pH ≈ 7.4
ctHb, maximum point
Uncorrected or corrected: ≈ 15 g/dL or 9.3 mmol/L
Slope 0.950-1.050
The following recommendations apply to sOB2 B :
Item Description
Units Fraction
Sample Set TcTtHb of gas equilibrated SAT0 and SAT100 samples to ≈ 15 g/dL (9.3 mmol/L) and pH ≈ 7.4
Slope 0.900-1.100
Offset ± 0.050
Continued on next page
Allowed corrections
ctHb
sOB2 B
ABL800 FLEX reference manual 4. User-defined corrections
4-5
Correction factors for oximetry parameters and bilirubin, TContinuedT
The following recommendations apply to FCOHb:
Item Description
Units Fraction
Sample The zero point (TF TCOHb ≈ 0) is saturated to approximately SAT100, and TcTtHb is set to ≈ 15 g/dL (9.3 mmol/L) and pH ≈ 7.4.
Offset ± 0.050
The following recommendations apply to FMetHb:
Item Description
Units Fraction
Sample The zero point (TF TMetHb ≈ 0) is saturated to approximately SAT100, and TcTtHb is set to ≈ 15 g/dL (9.3 mmol/L) and pH ≈ 7.4.
Offset ± 0.050
The following recommendations apply to FHbF:
Item Description
Units Fraction
Sample Radiometer recommends that TcTtHb in the adult samples (with FHbF = 0) and fetal samples (with high FHbF) is set to ≈ 15 g/dL (9.3 mmol/L), sOB2B ≈ 100 % and pH ≈ 7.4.
The “Correction for HbF levels less than 20 %” function should be enabled in order to have the FHbF value displayed for the adult sample.
Averaging repeated measurements on blood from different donors gives an optimized accuracy of the correction. Averaging repeated measurements on blood from the same donor also improves the accuracy.
Slope 0.800-1.200
Offset ± 0.20
Continued on next page
FCOHb
FMetHb
FHbF
4. User-defined corrections ABL800 FLEX reference manual
4-6
Correction factors for oximetry parameters and bilirubin, TContinued
The following recommendations apply to ctBil:
Item Description
Units μmol/L
Sample Radiometer recommends that human plasma or serum is used with pH ≈ 7.4 (the analyzer reading). Zero point sample could be adult sample (ctBil ≈ 0 μmol/L) and maximum point could be an unconjugated bilirubin sample with TcTtBil ≈ 300-400 μmol/L.
Averaging repeated measurements on samples from different donors gives an optimized accuracy of the correction. Averaging repeated measurements on samples from the same donor also improves the accuracy.
Commercial bilirubin standards can interfere with bilirubin measurement because they may have an absorbance spectrum different from that of human plasma.
Slope 0.5-1.5
Offset ± 100
The units for TF TOB2 BHb and TF THHb are [Fraction].
After the user-defined corrections of the parameters TsTOB2 B, TF TCOHb and TF TMetHb have been carried out, FOB2BHb and TF THHb are automatically calculated using the formulae stated below, since the sum of the fractions TF TCOHb, TF TMetHb, FOB2 BHb and TF THHb as defined must be equal to 1.0:
TFOB2 BHb: T
FOB2 BHb = (1 − TF TCOHb − TF TMetHb) × TsTOB2B
TFTTHHb: T
TF THHb = (1 − TF TCOHb − TF TMetHb) × (1 − TsTOB2 B)
TcTtBil
FOB2 BHb and FHHb
ABL800 FLEX reference manual 4. User-defined corrections
4-7
Electrolyte and metabolite parameters
Prior to entering corrections for the electrolyte and metabolite parameters, the user must obtain the reference values for the chosen parameters using the method accepted in his/her laboratory.
It should be noted that in order to define corrections:
• Measurements should be taken on the analyzer without user-defined corrections, and on the reference analyzer
• A series of measurements that cover the entire measuring range should be performed
• The measurements should be made simultaneously on the ABL800 FLEX and reference analyzers, and samples must be handled correctly
• The slope and the offset must be calculated. The user may, for example, make a linear correlation between the values measured on the ABL800 FLEX and the reference analyzers, using the ABL800 FLEX analyzer as an independent variable
• If the measurements are carried out on samples with values within the normal reference range, then the user may change the offset and leave the slope unchanged
• The user must verify the corrections that are entered
Details of these procedures may be found in the section TDefinitions and Test ConditionsT in Tchapter 5T.
The following corrections to the slope are possible within the stated limits:
TParameterT TSlope (mmol/L) T
TcTKP
+P 0.750-1.250
TcTNaP
+P 0.850-1.150
TcTCaP
2+P 0.800-1.200
TcTCl P
−P
0.850-1.150
TcTGlu 0.750-1.250
TcTLac 0.750-1.250
TcTCrea 0.75-1.250
The following corrections to the offset are possible within the stated limits:
TParameter: T TcTKP
+P TcTNaP
+P TcTCaP
2+P
TcTCl P
−P
TcTGlu TcTLac TcTCrea
TOffset (mmol/L): T
± 0.3 ± 5 ± 0.05 ± 5 ± 0.5 ± 0.5 ± 100 μmol/L
Continued on next page
Preparatory actions
Correcting the slope
Correcting the offset
4. User-defined corrections ABL800 FLEX reference manual
4-8
Electrolyte and metabolite parameters, TContinuedT
The Radiometer default values for the electrolyte and metabolite parameters must be reset manually by the user to 1.000 for each parameter on the TParameters Setup T screen.
Resetting corrections to default values
5-1
5. Performance characteristics
Overview
This chapter describes performance characteristics of the ABL8x0/8x5 and ABL8x7 FLEX analyzers for each measured parameter and test conditions to obtain them.
This chapter contains the following topics.
Definition of terms ........................................................................................... X5-2X
ABL8x0/8x5 Performance characteristics ................................................... X5-5X
Test conditions ........................................................................................... X5-6X
Performance test results – chart description............................................... X5-7X
Performance test results –- pH ................................................................... X5-10X
Performance test results – pCOB2 B ................................................................ X5-12X
Performance test results – pOB2B .................................................................. X5-15X
Performance test results – cKP
+P ................................................................... X5-18X
Performance test results – cNaP
+P.................................................................. X5-20X
Performance test results – cCl P
–P................................................................... X5-22X
Performance test results – cCaP
2+P ................................................................ X5-28X
Performance test results – cGlu.................................................................. X5-26X
Performance test results – cLac.................................................................. X5-28X
Performance test results – ctHb.................................................................. X5-30X
Performance test results – oximetry ........................................................... X5-32X
Performance test results – bilirubin............................................................ X5-42X
Additional information about FLEXMODE .............................................. X5-48X
ABL8x7 Performance characteristics........................................................... X5-49X
Test conditions ........................................................................................... X5-50X
Performance test results – pH, pCOB2B, pOB2 B ................................................. X5-51X
Performance test results – electrolytes ....................................................... X5-54X
Performance test results – cGlu, cLac........................................................ X5-58X
Performance test results – ctHb.................................................................. X5-60X
Performance test results – oximetry ........................................................... X5-61X
Performance test results – bilirubin............................................................ X5-70X
Performance test conditions and results – cCrea........................................ X5-72X
Interference tests .............................................................................................. X5-92X
References ........................................................................................................ X5-103X
Introduction
Contents
5. Performance characteristics ABL800 FLEX reference manual
5-2
Definition of terms
Performance specifications are achieved by comparison between the ABL8x0/8x5 FLEX analyzers and the primary reference methods, and by comparison between these analyzers and the ABL735.
Performance specifications of the ABL8x0/8x5 FLEX analyzers are described, using the following:
• BiasBRef B = the mean difference between the ABL8x0/8x5 FLEX analyzer and the primary reference methods.
• Bias BABLB = the mean difference between the ABL8x0/8x5 FLEX analyzer and the ABL735 analyzer.
• Repeatability
• Reproducibility
• Total variation range
• Imprecision
The bias of a quantity is defined as the mean difference between the measured value on a group of test instruments and the estimated true value (as assayed by the reference method). Bias BRef Bis determined as follows:
BiasBRef B = X BABL8x0/8x5 FLEXB − X BReference methodB
BiasBABLB is a relative bias between the ABL835 FLEX analyzer in FLEXMODE and the ABL735 analyzer in macromode (C195 μL mode), and is determined as follows:
BiasBABLB = X BABL8x0/8x5 FLEXB − X BABL735
Bias given in the tables in this chapter have been obtained experimentally.
Samples, assumed to be identical, repeatedly measured on one analyzer will not necessarily yield identical results. The degree of variation in the results is a measure of the repeatability of the analyzer.
The repeatability is obtained from repeated measurements within a short interval of time using:
• The same instrument and location
• The same measurement procedure
• Identical portions of the same sample
• One operator per instrument
The repeatability for each level is pooled for all test instruments and test days.
Continued on next page
General information
Bias
Repeatability
ABL800 FLEX reference manual 5. Performance characteristics
5-3
Definition of terms, TContinuedT
Reproducibility is obtained from repeated measurements over several days using:
• Random instrument
• Random sample
• Random operators
Reproducibility for each level is pooled for all test instruments and test days.
TE TBATB, total analytical error is a quality specification that sets a limit for both the random error (imprecision) and systematic error (inaccuracy) in a single measurement or single test result. In Radiometer reference manual the following expression for total analytical error is either expressed in an absolute number (TEBAB = ±(│Bias│+1,65×SBxB)) or in % (TE BAB = ± (│Bias│+1,65×CV Bx B)). The formula we are using for total error allowable works at 95% probability to allow for 5% error.
Repeated measurements using one analyzer on samples assumed to be identical will not necessarily yield identical results. The degree of variation in the results is a measure of the precision of the analyzer.
The following table describes the parameters used to characterize precision obtained via the performance tests on the ABL8x0/8x5/8x7 FLEX analyzers.
Parameter Description
S B0B TRepeatability
This is a standard deviation obtained from repeated measurements within a short interval of time using:
• The same instrument and location
• The same measurement procedure
• Identical portions of the same sample
• One operator per instrument
S B0B for each level is pooled for all test instruments and test days.
S BDB TDay-to-day variation
This is a standard deviation obtained from repeated measurements over all test days.
Includes contributions from differences in calibration states of the analyzers throughout the test days.
S BABLB TUncertainty of bias on a random instrument
S BABLB is used for repeated determinations on one sample. This standard deviation includes the inter-instrument variations, sample variations, and uncertainties from standard solutions and reference methods.
Continued on next page
Reproducibility
Total analytical error
Imprecision
5. Performance characteristics ABL800 FLEX reference manual
5-4
Definition of terms, TContinuedT
Parameter Description
S BxB Uncertainty of bias on a random instrument for a single measurement S BxB is a standard deviation, which includes SBABLB, B BS BDB and SB0 B.
S B0B, S BDB and SBABLB given in the tables are Radiometer requirements. Experimental data are always lower.
Confidence interval provides a range of values estimated from a study group that is highly likely to include the true, but unknown, value ("confidence interval" applies to the results of a statistical analysis). A 95 % confidence interval means that there is only a 5 % chance that the true value is not included in the interval.
The 95 % prediction interval is much wider than the 95 % confidence interval about, for example, the mean regression line response. For any given value of the independent variable, this interval represents the 95 % probability for the values of the dependent variable.
Imprecision T(continued)T
Confidence intervals T
Prediction interval
ABL800 FLEX reference manual 5. Performance characteristics
5-5
ABL8x0/8x5 Performance characteristics
Overview
This chapter describes performance characteristics of the ABL8x0/8x5 FLEX analyzers for each measured parameter and test conditions to obtain them.
This section contains the following topics.
Test conditions ................................................................................................. X5-6X
Performance test results – chart description..................................................... X5-7X
Performance test results –- pH ......................................................................... X5-10X
Performance test results – pCOB2 B ...................................................................... X5-12X
Performance test results – pOB2B ........................................................................ X5-15X
Performance test results – cKP
+P ......................................................................... X5-18X
Performance test results – cNaP
+P ....................................................................... X5-20X
Performance test results – cCl P
–P......................................................................... X5-22X
Performance test results – cCaP
2+P ...................................................................... X5-28X
Performance test results – cGlu........................................................................ X5-26X
Performance test results – cLac........................................................................ X5-28X
Performance test results – ctHb........................................................................ X5-30X
Performance test results – oximetry ................................................................. X5-32X
Performance test results – bilirubin.................................................................. X5-42X
Additional information about FLEXMODE .................................................... X5-48X
Introduction
Contents
5. Performance characteristics ABL800 FLEX reference manual
5-6
Test conditions
Test conditions to determine biasBABLB, repeatability and total variation for pH, pCOB2 B, pOB2 B, cCaP
2+P, cCl P
–P, cKP
+P, cNaP
+P, cGlu, cLac, ctHb were as follows:
Item Description
Reference analyzers Five ABL735 with AutoCheck module were used as a reference. The C195 mode was used as a reference for all measured parameters.
Primary reference methods
As specified for each parameter further in this chapter.
Analyzers and test modes
Five ABL835, three ABL830 and three ABL805 were tested over 11 days in the following modes:
• Syringe: S195, S165, S95, S85
• Capillary: FLEXMODE, C95, C85, C55, C35 OXI, C35 MET
Blood samples Heparinized blood samples from healthy, voluntary donors.
Eleven blood pools were prepared to cover test ranges for all measured parameters.
Blood measurements The measurements were performed by different operators.
Calibration solution and gases
All calibration solutions and gases used for the tests are traceable to Primary Reference Standards.
Traceability certificates for the ABL800 FLEX calibration solutions and gases are found at the end of chapter 7: TSolutionsT.
Experimental conditions
Ambient temperature: 22-25 °C
Relative humidity: 30-50 %.
• TThe solutions used in the performance tests are those recommended by Radiometer. Performances using other solutions cannot be guaranteed.
• TThe performance tests are performed under conditions where the analyzers are not influenced by electromagnetic fields.T
ABL8x0/8x5
TNOTES:
ABL800 FLEX reference manual 5. Performance characteristics
5-7
Performance test results – chart description
Tests were performed in the following modes:
Mode Syringe Capillary
Macro S195, S165 FLEXMODE ABL835 (no message) FLEXMODE ABL805 (no message)
Micro S95, S85 C95, C85, C55, C35 OXI, C35 MET; FLEXMODE ABL830 (no message) FLEXMODE (message 869) FLEXMODE (message 870) FLEXMODE (message 871) FLEXMODE (message 872) FLEXMODE (message 873) FLEXMODE (message 874)
The legend of the BiasBABLB chart is given below:
Chart Description
"x" axis The ABL735 mean values obtained as follows:
To determine the best possible ABL735 reference value for each parameter of a specific sample, the measurements on five ABL735 analyzers are plotted as a function of time. A regression line is made to represent the best possible mean ABL735 measurement at a given time thus compensating the metabolism of the sample during repeated measurements on it.
"y" axis Bias in %; bias for pH in pH units.
95 % statistical confidence range for bias in macromodes.
95 % statistical confidence range for bias in micromodes.
NBmacro B Number of measurements in macromodes.
NBmicro B Number of measurements in micromodes.
Continued on next page
Modes
BiasBABLB chart description
5. Performance characteristics ABL800 FLEX reference manual
5-8
Performance test results – chart description, TContinuedT
Repeatability is presented as a plot of the coefficient of variation (CV %). Contribution to variation, such as sample matrix and environmental conditions, are not directly included, but compensated for by extending the repeatability values shown in the chart.
Chart Description
"x" axis The ABL800 FLEX mean value.
"y" axis Repeatability in %; repeatability for pH in pH units.
Repeatability in macromodes.
Repeatability in micromodes.
NBmacro B Number of measurements in macromodes.
NBmicro B Number of measurements in micromodes.
Total variation chart is presented as a difference plot against the regression line at 5 ABL735 analyzers. The individual measurements are plotted directly.
Chart Description
"x" axis The ABL735 mean values obtained as follows:
To determine the best possible ABL735 reference value for each parameter of a specific sample, the measurements on five ABL735 analyzers are plotted as a function of time. A regression line is made to represent the best possible mean ABL735 measurement at a given time thus compensating the metabolism of the sample during repeated measurements on it.
"y" axis Total variation in %; total variation for pH in pH units.
At least 95 % statistical confidence range for total variation in macromodes.
At least 95 % statistical confidence range for total variation in micromodes.
o Observations in macromode.
x Observation in micromode.
NBmacroB Number of measurements in macromode – see the next page.
NBmicroB Number of measurements in micromode – see the next page.
Continued on next page
Repeatability chart
Total variation chart
ABL800 FLEX reference manual 5. Performance characteristics
5-9
Performance test results – chart description, TContinuedT
The number of measurements in macro- and micromodes, and the total number of measurements during the test is listed below:
Parameter NBmacroB NBmicroB Total
pH 3334 421 3755
pCOB2 B 2768 397 3165
pOB2B 282 2912 3194
cKP
+P 422 1364 1786
cNaP
+P 423 1362 1785
cCaP
2+P 407 1148 1555
cCl P
–P 426 1360 1786
cGlu 423 1825 2248
cLac 412 1829 2241
ctHb 415 3032 3447
Number of measurements
5. Performance characteristics ABL800 FLEX reference manual
5-10
Performance test results – pH
Capillary-type glass pH electrode with a saturated calomel reference electrode and a liquid junction saturated with KCl (BMS Mk2) [1,2].
The calibration standards are traceable to the Primary Reference Standards for pH.
The FLEXMODE on the ABL805/30/35 FLEX analyzers was tested:
pH Bias BREFB N
7.0 –0.002 90
7.4 –0.002 90
7.7 –0.002 90
N = number of measurements on several analyzers used for the test.
This bias is presented by the following chart:
Continued on next page
Reference method
BiasBREFB
BiasBABLB – blood samples
Bias (pH)
ABL735 (pH)
ABL800 FLEX reference manual 5. Performance characteristics
5-11
Performance test results – pH, TContinuedT
Repeatability is presented by the following chart:
Total variation is presented by the following chart:
Repeatability
Total variation
Repeatability (pH)
ABL800 FLEX
Total variation (pH)
ABL735
5. Performance characteristics ABL800 FLEX reference manual
5-12
Performance test results – TpTCOB2 B
Tonometry [3].
The gases used for tonometry are traceable to NIST-certified Standard Reference Materials.
The FLEXMODE on the ABL805/30/35 FLEX analyzers was tested:
pCOB2 B (mmHg) Bias BREFB N
15 –0.11 60
40 –0.38 60
60 0.29 60
80 –0.20 60
150 –0.21 60
N = number of measurements on several analyzers used for the test.
This bias is presented by the following chart:
Reference method
BiasBREFB
BiasBABLB – blood samples
Bias (%)
ABL735 (mmHg)
Continued on next page
ABL800 FLEX reference manual 5. Performance characteristics
5-13
Performance test results – TpTCOB2 B, TContinuedT
Repeatability is presented by the following chart:
Total variation is presented by the following chart:
Continued on next page
Repeatability
Total variation
Repeatability (%)
ABL800 FLEX (mmHg)
Total variation (%)
ABL735 (mmHg)
5. Performance characteristics ABL800 FLEX reference manual
5-14
Performance test results – TpTCOB2 B, TContinuedT
The bias and imprecision for expired air samples are as follows*:
TpCOB2 B (mmHg)T TBias ABL835/30/25/20/15/10/05T
15 0.2
40 −0.2
60 −0.4
80 −0.2
150 1.6
TpCOB2 B (mmHg)T
TS B0TB TS BDTB TS BABLTB TS BXTB
15 0.25 0.35 0.59 0.73
40 0.40 0.30 0.43 0.66
60 0.50 0.35 0.79 1.00
80 0.70 0.40 1.10 1.44
1.00 1.10 3.07 3.41
Bias and imprecision – expired air samples
150
* The Expired air mode is unchanged in the ABL800 FLEX analyzers compared to the ABL700 Series and, consequently, was not retested for the ABL800 FLEX analyzers.
ABL800 FLEX reference manual 5. Performance characteristics
5-15
Performance test results – TpTO B2 B
Tonometry [3].
The gases used for tonometry are traceable to NIST-certified Standard Reference Materials.
The FLEXMODE on the ABL805/30/35 FLEX analyzers was tested:
pOB2 B (mmHg) Bias BREFB N
15 0.47 60
50 –0.24 60
150 –0.45 60
250 –2.17 60
530 1.01 60
N = number of measurements on several analyzers used for the test.
This bias is presented by the following chart:
Continued on next page
Reference method
BiasBREFB
BiasBABLB – blood samples
Bias (%)
ABL735 (mmHg)
5. Performance characteristics ABL800 FLEX reference manual
5-16
Performance test results – TpTO B2 B, TContinuedT
Repeatability is presented by the following chart:
Repeatability
Total variation
Repeatability (%)
ABL835 (mmHg)
Total variation is presented by the following chart:
Total variation (%)
Continued on next page
ABL735 (mmHg)
ABL800 FLEX reference manual 5. Performance characteristics
Performance test results – TpTO B2 B, TContinuedT
The bias and imprecision for expired air samples are as follows:
TpOB2 B (mmHg)T TBias ABL835/30/25/20/15/10/05T
15 0.8
40 0.4
130 −0.4
230 −0.9
570 4.2
Imprecision:
TpOB2 B (mmHg)T TS B0TB TS BDTB TS BABLTB TS BXTB
15 0.3 0.3 1.2 1.3
40 0.3 0.3 1.0 1.1
130 0.3 0.3 0.7 0.8
230 2 2 3 4
570 5 5 13 15
Bias and imprecision – expired air samples
Fial
5-17
fteen ABL700 Series and ABL800 FLEX analyzers are tested over three days for l levels. Bias is determined against certified gases at sea level.
5. Performance characteristics ABL800 FLEX reference manual
5-18
Performance test results – cK P
+P
NIST-certified Standard Reference Material SRM 909b (human serum).
The FLEXMODE on the ABL805/35 FLEX analyzers was tested using SRM 909b:
cKP
+P (mmol/L) Bias BREFB N
3.424 –0.03 20
6.278 0.23 20
N = number of measurements on several analyzers used for the test.
This bias is presented by the following chart:
Continued on next page
Reference method
BiasBREFB
BiasBABLB – blood samples
Bias (%)
ABL735 (mmol/L)
ABL800 FLEX reference manual 5. Performance characteristics
5-19
Performance test results – TcTK P
+P, TContinuedT
Repeatability is presented by the following chart:
Total variation is presented by the following chart:
Repeatability
Total variation
Repeatability (%)
Total variation (%)
ABL800 FLEX (mmol)L)
ABL735 (mmol/L)
5. Performance characteristics ABL800 FLEX reference manual
5-20
Performance test results – TcTNaP
+P
NIST-certified Standard Reference Material SRM 909b (human serum) and Radiometer specified standard serum material (specified using flame photometry).
The FLEXMODE on the ABL805/35 FLEX analyzers was tested:
cNa P
+P (mmol/L) Bias BREFB N
120.76* –0.25 20
138.5** –0.28 30
N = number of measurements on several analysers used for the test.
(*NIST-certified Standard Reference Material
** Serum (Radiometer-specified).
This bias is presented by the following chart:
Reference method
BiasBREFB
BiasBABLB – blood samples
Bias (%)
Continued on next page
ABL735 (mmol/L)
ABL800 FLEX reference manual 5. Performance characteristics
Performance test results – cNaP
+P, TContinuedT
Repeatability is presented by the following chart:
Repeatability
Total variation
Repeatability (%)
Total variation is presented by the following chart:
Total variation (%)
ABL800 FLEX (mmol/L)
A
5-21
BL735 (mmol/L)
5. Performance characteristics ABL800 FLEX reference manual
5-22
Performance test results – TcTClP
–P
NIST-certified Standard Reference Material SRM 909b (human serum).
The FLEXMODE on the ABL805/35 FLEX analyzers was tested using SRM 909b:
cCl P
–P (mmol/L) Bias BREFB N
89.11 0.6 20
119.43 2.4 20
N = number of measurements on several analyzers used for the test.
This bias is presented by the following chart:
Continued on next page
Reference method
BiasBREFB
BiasBABLB – blood samples
Bias (%)
ABL735 (mmol/L)
ABL800 FLEX reference manual 5. Performance characteristics
Performance test results – cClP
–P, TContinuedT
Repeatability is presented by the following chart:
Repeatability
Total variation
Repeatability (%)
5-23
Total variation is presented by the following chart:
Total variation (%)
ABL800 FLEX (mmol/L)
ABL735 (mmol/L)
5. Performance characteristics ABL800 FLEX reference manual
5-24
Performance test results – cCaP
2+P
The calcium transfer standards were used. These are traceable to NIST SRM915 and have an ionic strength of 160.0 mmol per kg of water and pH 7.40 at 37 °C, using 1 mmol/L (37 °C) HEPES buffer. The standards were produced as indicated in [4].
The FLEXMODE on the ABL805/35 FLEX analyzers was tested:
cCa P
2+P (mmol/L) BiasBREFB N
0.4879 0.038 150
1.2700 0.025 150
2.5657 0.052 150
N = number of measurements on several analyzers used for the test.
This bias is presented by the following chart:
Continued on next page
Reference method
BiasBREFB
BiasBABLB – blood samples
Bias (%)
ABL735 (mmol/L)
ABL800 FLEX reference manual 5. Performance characteristics
5-25
Performance test results – TcTCaP
2+P, TContinuedT
Repeatability is presented by the following chart:
Total variation is presented by the following chart:
Repeatability
Total variation
Repeatability (%)
Total variation (%)
ABL800 FLEX (mmol/L)
ABL735 (mmol/L)
5. Performance characteristics ABL800 FLEX reference manual
5-26
Performance test results – cGlu
Spectrophotometry, using the hexokinase (HK) method recommended by NCCLS [5], measured on serum.
The FLEXMODE on the ABL805/35 FLEX analyzers was tested:
cGlu(mmol/L) BiasBREFB N
0.08 0.03 30
2.09 0.06 30
5.08 0.12 30
14.73 –0.02 30
N = number of measurements on several analyzers used for the test.
This bias is presented by the following chart:
Continued on next page
Reference method
BiasBREFB
BiasBABLB – blood samples
Bias (%)
ABL735 (mmol/L)
ABL800 FLEX reference manual 5. Performance characteristics
5-27
Performance test results – TcTGlu, TContinuedT
Repeatability is presented by the following chart:
Total variation is presented by the following chart:
Repeatability
Total variation
Repeatability (%)
Total variation (%)
ABL800 FLEX (mmol/L)
ABL735 (mmol/L)
5. Performance characteristics ABL800 FLEX reference manual
5-28
Performance test results – cLac
Spectrophotometry using a lactate dehydrogenase (LDH) method, measured on serum [10].
The FLEXMODE on the ABL805/35 FLEX analyzers was tested:
cLac (mmol/L) BiasBREFB N
0.36 -0.08 30
2.06 0.12 30
8.3 –0.81 30
11.3 –0.62 30
N = number of measurements on several analyzers used for the test.
This bias is presented by the following chart:
Continued on next page
Reference method
BiasBREFB
BiasBABLB – blood samples
Bias (%)
ABL735 (mmol/L)
ABL800 FLEX reference manual 5. Performance characteristics
5-29
Performance test results – TcTLac, TContinuedT
Repeatability is presented by the following chart:
Total variation is presented by the following chart:
Repeatability
Total variation
ABL800 FLEX (mmol/L)
Repeatability (%)
Total variation (%)
ABL735 (mmol/L)
5. Performance characteristics ABL800 FLEX reference manual
5-30
Performance test results – TcTtHb
HiCN method recommended by NCCLS [6].
The FLEXMODE on the ABL830/35 FLEX analyzers was tested:
ctHb (mmol/L) BiasBREFB N
15 (SAT0) 0.32 145
7 (SAT100) 0.04 145
15 (SAT100) 0.37 145
25 (SAT100) 0.97 145
N = number of measurements on several analyzers used for the test.
This bias is presented by the following chart:
Continued on next page
Reference method
BiasBREFB
BiasBABLB
Bias (%)
ABL735 (g/dL)
ABL800 FLEX reference manual 5. Performance characteristics
Performance test results – ctHb, TContinuedT
Repeatability is presented by the following chart:
Repeatability
Total variation
ABL800 FLEX (g/dL)
Repeatability (%)
5-31
Total variation is presented by the following chart:
ABL735 (g/dL)
Total variation (%)
5. Performance characteristics ABL800 FLEX reference manual
5-32
Performance test results – oximetry
The optical system is unchanged in the ABL800 FLEX analyzers compared to the ABL700 Series. Test of TcTtHb on the ABL800 FLEX analyzers has been conducted (results given on pages 5-31 to 5-32). The other oximetry parameters have not been retested; the information and results below are from the ABL700 Series.
The reference method established for the oximetry parameters uses modified ABL520 analyzers as the reference instruments. The ABL520 analyzers have been validated and their performance specifications determined according to primary reference methods.
The modified ABL520 analyzers are used in accordance with IFCC’s recommendations for traceability of reference methods.
The reference methods used for the oximetry parameters on the ABL520 analyzers are those presented below.
Parameter Primary reference method
sOB2 B Tonometry: whole blood is tonometered with a gas mixture containing 94.4 % OB2 B and 5.6 % COB2 B.
FHHb Th (TcTtHb = 13-15 g/dL) treated wi
FCOHb Gamopu16
FMetHb Sp
FHbF AlNC
Test
Reference Toacpathe
Eathr
(Fwe
Bia
Explanation
Reference methods
Test conditions for oximetry parameters
e standard is blood
th dithionite.s chromatography. The standards are carbon noxide mixtures with atmospheric air, whose
rity is validated in accordance with NIST SRM 78 (50 ppm CO in NB2 B).
ectrometry, modified Evelyn-Malloy method [7].
kali denaturation method [8]. Corresponds to CLS guideline [9].
Description
verify that the correction constants have been curately determined, 10 analyzers with all rameters available are tested in C195 mode against reference methods.
ch parameter is tested on 3-6 levels over at least ee days, with five repetitions each day.
ive new analyzers with all parameters available re tested against the reference methods for FHbF).
as for each parameter in the C195 measuring mode rence method is determined.
gainst the refeContinued on next page
ABL800 FLEX reference manual 5. Performance characteristics
5-33
Performance test results – oximetry, TContinuedT
Test Description
Verification 6-10 ABL700 Series analyzers are tested over at least two days for all levels. Bias for the given mode is calculated as difference compared to the C195 mode.
Bias against the reference method is determined as follows:
Bias = bias against C195 + C195 bias against reference method.
The following parameters: sOB2 B, FCOHb, FMetHb and FOB2 BHb are measured directly against the reference built into the analyzer, and these parameters are independent of the reference method.
Reduced verification 6-10 new analyzers are used over at least one day for selected levels.
Bias for the tested mode is calculated as follows:
Bias = bias against C195 + C195 bias against reference method.
Modes which are not tested are described as "N/A".
Simple verification 6-10 analyzers are tested at one extreme level over one day. Bias is not determined; bias values for the modes with similar wet section programs are used.
The measuring modes were tested as follows:
Test Analyzer Mode
Reference ABL735/25/15 C195
Verification ABL735/25/15 S195, S95, S85, C95, C55, C35 OXI
Continued on next page
Test conditions for oximetry parameters T(continued)
5. Performance characteristics ABL800 FLEX reference manual
5-34
Performance test results – oximetry, TContinuedT
Bias:
TsTTOB2 B (%)T TABL835/25/15T TABL830/20/10T
TcTTtHb (g/dL) T TsTTOB2 B (%)T TS195T TFM* T TS85T
15 0 0.00 0.05 −0.02
7 100 0.01 0.22 N/A
15 100 0.01 −0.08 0.00
25 100 0.00 −0.29 N/A
* FM = FLEXMODE (no message) corresponding to C195 mode on the ABL735/25/15.
Imprecision:
TcTTtHb (g/dL) T TsOB2 B (%)T TS B0TB TS BDTB TS BABLTB TS BXTB
15 0 0.05 0.05 0.25 0.30
7 100 0.10 0.10 0.25 0.30
15 100 0.05 0.10 0.25 0.30
25 100 0.05 0.10 0.30 0.35
Bias:
TsOB2 TB TABL835/25/15 T TABL830/20/10T
TcTTtHb (g/dL) T
TsOB2 B (%) T
TS95T TC95 T TS85T TC55 T TC35 T TFM*T TC55 T TC35 T
15 0 -0.04 -0.02 -0.02 -0.03 -0.03 N/A -0.03 -0.03
7 100 -0.10 -0.19 N/A -0.22 -0.10 N/A -0.22 -0.10
15 100 -0.10 -0.16 0.00 -0.16 -0.10 -0.05 -0.16 -0.10
25 100 -0.10 -0.17 N/A -0.14 -0.09 N/A -0.14 -0.09
* FM = FLEXMODE (no message) corresponding to C195 mode on the ABL735/25/15.
Imprecision:T
TcTtHb (g/dL) sOB2 B (%) S B0B S BDB S BABLB S BXB
15 0 0.05 0.05 0.25 0.30
7 100 0.10 0.10 0.25 0.30
15 100 0.05 0.10 0.25 0.30
25 100 0.05 0.10 0.30 0.35
Continued on next page
sOB2 B – macromodes
sOB2 B – micromodes
ABL800 FLEX reference manual 5. Performance characteristics
5-35
Performance test results – oximetry, TContinuedT
Bias:
FOB2 BHb ABL835/25 ABL830/20
TcTtHb (g/dL) TFTOB2 BHb (%) S195 FM* S85
15 0 0.00 −0.04 −0.02
7 100 −0.07 N/A N/A
15 100 −0.03 N/A −0.15
25 100 −0.05 N/A N/A
* FM = FLEXMODE (no message)
Imprecision:
TcTtHb (g/dL) FOB2 BHb (%)
S B0B S BDB S BABLB S BXB
15 0 0.05 0.05 0.25 0.30
7 100 0.25 0.20 0.50 0.60
15 100 0.15 0.15 0.45 0.50
25 100 0.10 0.10 0.40 0.45
Bias:
FOB2 BHb (%) ABL835/25
TcTtHb (g/dL) TFTOB2 BHb (%) S95 C95 S85 C55 C35
15 0 −0.04 −0.02 −0.02 −0.03 −0.03
7 100 −0.47 −0.39 N/A −0.48 −0.18
15 100 −0.33 −0.40 −0.15 −0.39 −0.31
25 100 −0.29 −0.46 N/A −0.36 −0.33
FOB2 BHb ABL830/20
TcTtHb (g/dL) TFTOB2 BHb (%) C85 C55 C35
15 0 N/A −0.03 −0.03
7 100 N/A −0.48 −0.18
15 100 −0.16 −0.39 −0.31
25 100 N/A −0.36 −0.33
Continued on next page
FOB2 BHb – macromodes
FOB2 BHb – micromodes
5. Performance characteristics ABL800 FLEX reference manual
5-36
Performance test results – oximetry, TContinuedT
Imprecision:
TcTtHb (g/dL) FOB2 BHb (%) S B0B S BDB S BABLB S BXB
15 0 0.05 0.05 0.25 0.30
7 100 0.25 0.20 0.50 0.60
15 100 0.15 0.15 0.45 0.50
25 100 0.10 0.10 0.40 0.45
Bias:
FCOHb ABL835/25 ABL830/20
TcTtHb (g/dL) sOB2 B (%) TFTCOHb (%)
S195 FM* S85
15 100 0 0.03 0.08 0.12
7 100 20 N/A 0.47 N/A
15 100 20 N/A 0.10 N/A
25 100 20 N/A −0.47 N/A
* FM = FLEXMODE (no message)
Imprecision:
TcTtHb (g/dL) sOB2 B (%) FCOHb (%) S B0B S BDB S BABLB S BXB
15 100 0 0.05 0.10 0.35 0.40
7 100 20 0.10 0.10 0.75 0.80
15 100 20 0.05 0.10 0.70 0.75
25 100 20 0.05 0.10 0.70 0.75
Imprecision (ABL837/27):
TcTtHb (g/dL) sOB2 B (%) FCOHb (%) S B0B S BDB S BABLB S BXB
15 100 0 0.05 0.10 0.35 0.40
7 100 20 0.10 0.10 0.75 0.80
15 100 20 0.05 0.10 0.70 0.75
25 100 20 0.05 0.10 0.70 0.75
Continued on next page
TFTOB2 BHb – micromodes T(continued)T
FCOHb – macromodes
ABL800 FLEX reference manual 5. Performance characteristics
5-37
Performance test results – oximetry, TContinuedT
Bias:
FCOHb ABL835/25
TcTtHb (g/dL)
sOB2 B (%)
TFTCOHb (%)
S95 C95 S85 C55 C35
15 100 0 0.10 0.10 0.12 0.08 0.08
7 100 20 N/A N/A N/A N/A N/A
15 100 20 N/A N/A N/A N/A N/A
25 100 20 N/A N/A N/A N/A N/A
FCOHb ABL830/20
TcTtHb (g/dL) sOB2 B (%) TFTCOHb (%) FM* C55 C35
15 100 0 −0.02 0.08 0.08
7 100 20 N/A N/A N/A
15 100 20 N/A N/A N/A
25 100 20 N/A N/A N/A
* FM = FLEXMODE (no message)
Imprecision:
TcTtHb (g/dL) TsTOB2 B (%) FCOHb (%) S B0B S BDB S BABLB S BXB
15 100 0 0.05 0.10 0.35 0.40
7 100 20 0.10 0.10 0.75 0.80
15 100 20 0.05 0.10 0.70 0.75
25 100 20 0.05 0.10 0.70 0.75
Bias:
FMetHb ABL835/25 ABL830/20
TcTtHb (g/dL) sOB2 B (%) TFTMetHb (%) S195 FM* S85
15 100 0 0.01 −0.03 0.06
15 100 20 N/A 0.10 N/A
* FM = FLEXMODE (no message)
Continued on next page
FCOHb – micromodes
FMetHb – macromodes
5. Performance characteristics ABL800 FLEX reference manual
5-38
Performance test results – oximetry, TContinuedT
Imprecision:
TcTtHb (g/dL) sOB2 B (%) TFTMetHb (%) S B0B S BDB S BABLB S BXB
15 100 0 0.10 0.10 0.25 0.30
15 100 20 0.05 0.10 0.35 0.40
Bias:
FMetHb ABL835/25
TcTtHb (g/dL)
sOB2 B (%) TFTMetHb (%)
S95 C95 S85 C55 C35
15 100 0 0.13 0.14 0.06 0.16 0.14
7 100 20 N/A N/A N/A N/A N/A
15 100 20 N/A N/A N/A N/A N/A
25 100 20 N/A N/A N/A N/A N/A
FMetHb ABL830/20
TcTtHb (g/dL) sOB2 B (%) TFTMetHb (%) FM* C55 C35
15 100 0 0.13 0.16 0.14
7 100 20 N/A N/A N/A
15 100 20 N/A N/A N/A
25 100 20 N/A N/A N/A
* FM = FLEXMODE (no message)
Imprecision:
TcTtHb (g/dL) TsTOB2 B (%) FMetHb(%) SB0B S BDB S BABLB S BXB
15 100 0 0.10 0.10 0.25 0.30
15 100 20 0.05 0.10 0.35 0.40
Continued on next page
TFTMetHb – macromodes T(continued)T
TFTMetHb – micromodes
ABL800 FLEX reference manual 5. Performance characteristics
5-39
Performance test results – oximetry, TContinuedT
Bias:
FHHb ABL835/25 ABL830/20
TFTHHb (%) TcTtHb (g/dL) S195 FM* S85
0 15 –0.01 0.08 −0.05
* FM = FLEXMODE (no message)
Imprecision:
FHHb (%) TcTtHb (g/dL) S B0B S BDB S BABLB S BXB
0 15 0.05 0.10 0.30 0.35
Bias:
FHHb ABL835/25 ABL830/20
TcTtHb (g/dL)
FHHb (%)
S95 C95 S85 C55 C35 FM* C55 C35
15 0 0.09 N/A N/A 0.15 0.10 N/A N/A 0.10
* FM = FLEXMODE (no message)
Imprecision:
TcTtHb (g/dL) FHHb (%) SB0B S BDB S BABLB S BXB
15 0 0.05 0.10 0.30 0.35
Bias (macromodes):
FHbF ABL835 ABL830
FHbF (%) TcTtHb (g/dL) S195 FM* S85
0 10 3.3 3.3 3.3
0 15 5.5 5.5 5.5
0 20 5.6 5.6 5.6
* FM = FLEXMODE (no message)
Imprecision (macromodes):
FHbF (%) TcTtHb (g/dL) sOB2 B (%) S B0B S BDB S BABLB S BXB
0 10 100 4 4 5 8
0 15 100 2 3 7 8
0 20 100 2 2 10 11
Continued on next page
FHHb – macromodes
FHHb – micromodes
FHbF – adult blood
5. Performance characteristics ABL800 FLEX reference manual
5-40
Performance test results – oximetry, TContinuedT
Bias (micromodes):
FHbF ABL835 ABL830
TcTtHb (g/dL)
TFTHbF (%)
S95 C95 S85 C55 C35 FM* C55 C35
10
0 3.3 3.3 3.3 3.3 3.3 3.3 3.3 3.3
15
0 5.5 5.5 5.5 5.5 5.5 5.5 5.5 5.5
20
0 5.6 5.6 5.6 5.6 5.6 5.6 5.6 5.6
* FM = FLEXMODE (no message)
Imprecision (micromodes):
TcTtHb (g/dL) TFTHbF (%) sOB2 B (%) S B0B S BDB S BABLB S BXB
10 4 4 5 8
15 2 3 5 7
20
0 100
2 2 10 11 NOTES: a, b.
Bias (macromodes):
FHbF ABL835 ABL830
FHbF (%) TcTtHb (g/dL) S195 FM*
80 10 5.9 5.9 5.9
80 15 3.3 3.3 3.3
80 20 2.6 2.6 2.6
* FM = FLEXMODE (no message)
Imprecision (macromodes):
FHbF (%) TcTtHb (g/dL) TsTOB2 B (%) S B0B S BDB S BABLB S BXB
80 10 100 4 5 5 9
80 15 100 3 3 6 8
80 20 100 2 3 6 7 NOTES: a, b.
Continued on next page
FHbF – adult blood T(continued)T
FHbF – fetal blood
ABL800 FLEX reference manual 5. Performance characteristics
5-41
Performance test results – oximetry, TContinuedT
Bias (micromodes):
FHbF ABL835 ABL830
TcTtHb (g/dL)
FHbF (%)
S95 C95 S85 C55 C35 FM* C55 C35
10 80 5.9 5.9 5.9 5.9 5.9 5.9 5.9 5.9
15 80 3.3 3.3 3.3 3.3 3.3 3.3 3.3 3.3
20 80 2.6 2.6 2.6 2.6 2.6 2.6 2.6 2.6
* FM = FLEXMODE (no message)
Imprecision (micromodes):
FHbF (%) TcTtHb (g/dL) sOB2 B (%) S B0B S BDB S BABLB S BXB
10 4 5 6 9
15 3 3 6 8
80
20
100
2 3 6 7
NOTES: a, b.
The following corrections should be geometrically added to SBInstB and SBXB for the analyzer's wavelength calibrated with the S7770:
Parameter Mode Level Correction (percentage
point)
Macromode All 0 TcTtHb
Micromode All 0
sOB2 B All sOB2 B (100 %) 0.23
FOB2 BHb All FOB2 BHb (100 %) 0.15
FCOHb All FCOHb (20 % and 0 %) 0.40
FHHb All FHHb (0 %) 0.23
NOTES:
a. pH = 7.4 ± 0.1. FHbF is adjusted with the pH sensitivity to a nominal pH = 7.4. For further details please refer to the TInterference Tests T section for oximetry parameters.
b. Specifications for imprecision are derived from worst-case values found during internal laboratory tests. 40 % relative is then added as a safety factor.
FHbF – fetal blood T(continued)T
Contribution to Imprecision Specifications from S7770
5. Performance characteristics ABL800 FLEX reference manual
5-42
Performance test results – bilirubin
As the optical system is unchanged in the ABL800 FLEX analyzers compared to the ABL700 Series, the specifications for bilirubin have not been re-established.
The ABL735/30 performance specifications for bilirubin were made as a field test the purpose of which was to optimize the bilirubin algorithm for neonatal blood samples.
For neonatal use: The bilirubin method has been evaluated on whole blood and plasma. The allowed analytical error is ± 10 % to satisfy average clinical requirements for bilirubin measurement [1,2,3,4,5]. This requirement is fulfilled for plasma. For whole blood the analytical error is slightly higher. The clinicians and clinical chemists have evaluated bilirubin measurement on whole blood, the conclusion being that the ABL735/30 analyzers have satisfactory performance and can substitute other bilirubin measuring methods.
For adult use: Adult samples within reference range:
The uncertainty in the bilirubin measurement on whole blood can, in some cases, exceed the level required to measure normal bilirubin levels for children older than 3 months and adults (bilirubin reference range 4-22 μmol/L). In these cases it is recommended to measure bilirubin on plasma or serum.
Adult samples with an increased bilirubin level:
Adult field tests were typically performed on samples with 80 % of the total bilirubin in the conjugated form. For these highly conjugated samples the field tests showed a negative bias of 7 % on both plasma and whole blood samples.
The patient samples represented typical variations in TcTtBil, TcTtHb, sOB2 B, pH and MCHC values.
A Hitachi analyzer calibrated with NIST SRM 916a standards was used as a reference. TcTtBil was measured in μmol/L. Each field test place had its own ABL735 analyzer.
Continued on next page
Explanation
Field test results
ABL800 FLEX reference manual 5. Performance characteristics
5-43
Performance test results – bilirubin, TContinuedT
The field test results are given below.
Pos. Field test place
Type N Slope Inter-
cept
μmol/L
R2 Syx
μmol/L
Range
μmol/L
1 A 46 1.026 0.0 0.9914 5.1 18-258
2 B 56 0.986 –1.3 0.9939 5.8 10-334
3 D 4 1.014 –1.4 0.9984 4.5 22-236
4 E
Plasma,
neonatal
47 0.945 1.2 0.9937 5.1 4-253
5 D Plasma, 16 0.950 –0.5 0.9977 5.2 18-313
6 B adult 59 0.924 1.4 0.9981 3.8 2-366
7 F 52
45 (a)
0.904
0.942
5.6
2.6
0.9932
0.9941
12.0
5.3
4-635
4-300
8 A 46 1.075 9.6 0.9661 10.7 18-258
9 B 100 1.057 –1.6 0.9819 12.0 3-297
10 D 32 1.000 –5.6 0.9715 14.4 3-254
11 C 52 0.993 –5.0 0.9790 11.3 6-309
12 E
Blood,
neonatal
47 1.019 –10.2 0.9827 9.5 4-253
13 D Blood, 18 0.950 –6.8 0.9974 5.6 18-313
14 B adult 55 0.909 3.2 0.9974 4.6 2-366
15 F 25 0.939 4.9 0.9967 10.0 21-635
Regression table: Regression results from field tests. N = #samples, SBy/x B is standard deviation about regression line.
NOTE: (a) Datasubset excluding samples above 300 μmol/L.
Continued on next page
Field test results T(continued)
5. Performance characteristics ABL800 FLEX reference manual
5-44
Performance test results – bilirubin, TContinuedT
Data set position 9 from regression table.
y = 1.0572x - 1.6119R2 = 0.9819
0
50
100
150
200
250
300
350
0 50 100 150 200 250 300
Hitachi, NIST
AB
L735
Syx=12.0
Actual field test from a neonatal critical care hospital using whole blood. Values are in μmol/L.
The same data as above but depicted in a Bland-Altman plot below.
-20
-10
0
10
20
30
40
50
0 50 100 150 200 250 300
ctBil
Diff
eren
ce
Lines indicate Mean, Mean+2SD and Mean–2SD. Values are in μmol/L. Difference = ABL835 – Hitachi, NIST.
Continued on next page
Regression and Bland-Altman plot
ABL800 FLEX reference manual 5. Performance characteristics
5-45
Performance test results – bilirubin, TContinuedT
The following parameters are used to describe performance of the ABL835/30 FLEX analyzers for bilirubin measurements.
S B0B: Repeatability. Measurement short time interval variation on the same sample
S BDB: Day-to-day variation
S BTB: Patient-to-patient variation
S BI B: ABL-to-ABL instrumental variation
S BABLB: ABL uncertainty. Variation including SBTB, S BI B and reference uncertainty
S BXB: Reproducibility. Total variation including SB0 B, SBDB and S BABLB
The above field test regression statistics SByx B include variations from SB0B, S BDB and SBT.
Macromodes: 195 μL and 85 μL from syringe and capillary:
TcTtBil (μmol/L)
TcTtHb (g/dL)
TsTOB2 B (%) S B0B S BDB S BTB S BIB S BABLB S BXB
≈0 Plasma 1.1 1.4 2.2 0.4 2.3 2.9
≈0 10 100 1.9 3.1 4.0 3.2 5.1 6.3
≈0 15 100 2.3 2.9 7.4 5.5 9.2 9.9
≈0 20 100 3.4 2.6 10.9 13.0 17.0 17.5
≈200 Plasma 1.3 1.7 3.1 4.7 7.4 7.7
≈200 10 100 2.4 4.4 5.8 6.6 10.1 11.3
≈200 15 100 2.6 3.7 8.5 9.3 13.6 14.4
≈200 20 100 4.2 5.0 12.1 15.4 20.4 21.4
≈400 Plasma 1.7 2.5 4.8 9.3 12.0 12.3
≈400 10 100 3.5 6.8 9.3 12.0 16.5 18.2
≈400 15 100 3.4 5.3 11.4 15.9 20.8 21.7
≈400 20 100 6.0 8.8 15.0 21.0 27.1 29.2
Notes: a, b, c
Continued on next page
Imprecision parameters
Performance test results for bilirubin
5. Performance characteristics ABL800 FLEX reference manual
5-46
Performance test results – bilirubin, TContinuedT
Micromodes: 95 μL – syringe and capillary, 55 μL and 35 μL – capillary:
TcTtBil (μmol/L)
TcTtHb (g/dL)
TsTOB2 B (%)
S B0B S BDB S BTB S BIB S BABLB S BXB
≈0 Plasma 1.1 1.4 2.2 0.4 2.3 2.9
≈0 10 100 1.9 3.1 4.0 3.2 5.1 6.3
≈0 15 100 2.3 2.9 7.4 5.5 9.2 9.9
≈0 20 100 3.4 2.6 10.9 13.0 17.0 17.5
≈200 Plasma 2.0 1.7 2.9 3.9 6.8 7.3
≈200 10 100 3.7 3.9 6.0 5.6 9.6 11.0
≈200 15 100 4.4 4.2 9.3 7.9 13.2 14.6
≈200 20 100 5.6 5.9 13.0 16.3 21.6 23.1
≈400 Plasma 3.5 2.5 4.3 7.8 10.6 11.4
≈400 10 100 6.7 5.7 9.9 9.6 15.2 17.6
≈400 15 100 7.9 6.7 13.5 12.5 19.7 22.3
≈400 20 100 9.5 10.9 17.8 23.6 30.7 33.9
Notes: a, b, c
NOTES:
a. Adult/fetal blood, pH = 7.4 ± 0.1, normal MCHC and albumin variation. Spiked with unconjugated bilirubin.
b. TcTtBil specification at level 200 μmol/L is interpolated from the measured specifications at 0 and 400 μmol/L.
c. The performance specifications apply to measurements performed using CLINITUBES with clot catchers and mixing wire from Radiometer.
1. Fraser CG. The application of theoretical goals based on biological variation data in proficiency testing. Arch Pathol Lab Med 1988; 112: 402-15.
2. Ehrmeyer SS, Laessig RH, Leinweber JE, Oryall JJ. 1990 Medicare/CLIA final rules for proficiency testing: minimum intralaboratory performance characteristics (CV and bias) needed to pass. Clin Chem 1990; 36, 10: 1736-40.
Continued on next page
Performance test results for bilirubin T(continued)T
References
ABL800 FLEX reference manual 5. Performance characteristics
5-47
Performance test results – bilirubin, TContinuedT
3. Fraser CG, Petersen PH, Ricos C, Haeckel R. Proposed quality specifications for the imprecision and inaccuracy of analytical systems for clinical chemistry. Eur J CLin Chem Clin Biochem 1992; 30: 311-17.
4. Westgard JO, Seehafer JJ, Barry PL. Allowable imprecision for laboratory test based on clinical and analytical test outcome criteria. Clin Chem 1994; 40, 10: 1909-14.
5. Vanderline RE, Goodwine J, Koch D, Scheer D, Steindel S, Cembrowski G. Guidelines for providing quality stat laboratory services. 1987 Laboratory Quality Assurance Commitee.
References T(continued)T
5. Performance characteristics ABL800 FLEX reference manual
5-48
Additional information about FLEXMODE
With the FLEXMODE
• Varying sample volumes can be introduced to obtain a given parameter profile, and
• Two different parameter profiles can be reported for the same sample volume as the sample volume intervals overlap one another.
See the TABL800 FLEX Operator’s ManualT, chapter 4, page 4-3, for an overview of sample volume intervals and parameter profiles.
Most of the variation contributed by the difference in sample volume and parameter profile is included in the performance test results (bias, repeatability and imprecision) given in this chapter. Special tests with emphasis on extreme scenarios were conducted. The following scenarios were studied:
• Bias and repeatability for minimum and maximum sample volumes
• Bias and repeatability for parameter profiles with same sample volume
These special studies were conducted for each of the individual parameter profiles. The parameters with the variation exceeding the Performance Characteristics are listed below.
Sample volume < 55 μL:
Parameter Values Bias Repeatability
pH 7.15 7.40
0.015 0.013
0.005 0.005
pCOB2 B (mmHg) 29 80
1.0 -2.9
0.9 2.6
pOB2B (mmHg) 130 230
3.0 -3.7
3.9 3.6
Introduction
Worst-case observations
ABL800 FLEX reference manual 5. Performance characteristics
5-49
ABL8x7 Performance characteristics
Overview
This chapter describes performance characteristics of the ABL8x7 FLEX analyzers for each measured parameter and test conditions to obtain them.
This section contains the following topics.
Test conditions ................................................................................................. X5-50X
Performance test results – pH, pCOB2B, pOB2 B ....................................................... X5-51X
Performance test results – electrolytes ............................................................. X5-54X
Performance test results – cGlu, cLac.............................................................. X5-58X
Performance test results – ctHb........................................................................ X5-60X
Performance test results – oximetry ................................................................. X5-61X
Performance test results – bilirubin.................................................................. X5-70X
Performance test conditions and results – cCrea.............................................. X5-72X
Interference tests .............................................................................................. X5-92X
References ........................................................................................................ X5-103X
Introduction
Contents
5. Performance characteristics ABL800 FLEX reference manual
5-50
Test conditions
Test conditions to determine biasBABLB and imprecision for pH, pCOB2B, pOB2B, cCaP
2+P,
cCl P
–P, cKP
+P, cNaP
+P, cGlu, cLac, ctHb were as follows:
Item Description
Reference analyzers Five ABL735 with AutoCheck module were used as a reference. The C195 mode was used as a reference for all measured parameters.
Primary reference methods
As specified for each parameter further in this chapter.
Analyzers and test modes
Ten ABL837 were tested over 30 days in the following modes:
• Syringe: S250
• Capillary: C125, C35 MET
Blood samples Heparinized blood samples from healthy, voluntary donors.
The blood is prepared to obtain the different concentration levels of each measured parameter.
Blood measurements The measurements were performed by different operators.
Calibration solution and gases
All calibration solutions and gases used for the tests are traceable to Primary Reference Standards.
Traceability certificates for the ABL8x7 FLEX calibration solutions and gases are found at the end of chapter 7: TSolutionsT.
Experimental conditions
Ambient temperature: 22-25 °C
Relative humidity: 30-50 %.
Test conditions for TcTCrea – see pages 5-73 to 5-86 in this chapter.
• TThe solutions used in the performance tests are those recommended by Radiometer. Performances using other solutions cannot be guaranteed.
• TThe performance tests are performed under conditions where the analyzers are not influenced by electromagnetic fields.T
ABL8x7
TNOTES:
ABL800 FLEX reference manual 5. Performance characteristics
5-51
Performance test results – pH, TpTCOB2 B, TpTOB2 B
Parameter Reference method
pH Capillary-type glass pH electrode with a saturated calomel reference electrode and a liquid junction saturated with KCl (BMS Mk2) [1,2].
pCOB2 B and pOB2B Tonometry [3].
The gases used for tonometry are traceable to NIST-certified Standard Reference Material SRM 1701a, 1702a, 1703a.
S250 mode – bias:
pH Bias
7.0 –0.003
7.4 0.002
7.7 0.005
S250 mode – imprecision:
pH S B0B S BDB S BABLB S BXB
7.0 0.0035 0.0024 0.0061 0.0080
7.4 0.0020 0.0019 0.0070 0.0080
7.7 0.0030 0.0024 0.0100 0.0110
C125 mode – bias:
pH Bias
7.0 0.004
7.4 0.001
7.7 –0.001
C125 mode – imprecision:
pH S B0B S BDB S BABLB S BXB
7.0 0.0053 0.0036 0.0070 0.0120
7.4 0.0030 0.0029 0.0079 0.0120
7.7 0.0050 0.0036 0.0108 0.0150
Continued on next page
Reference methods
pH
5. Performance characteristics ABL800 FLEX reference manual
5-52
Performance test results – pH, TpTCOB2 B, TpTOB2 B, TContinuedT
S250 mode – bias:
pCOB2 B (mmHg) Bias
10 –0.26
28 −0.01
40 −0.26
60 0.03
80 –0.12
150 –1.37
S250 mode – imprecision:
pCOB2 B (mmHg)
S B0B S BDB S BABLB S BXB
10 0.25 0.35 0.51 0.90
40 0.40 0.30 0.59 0.70
60 0.60 0.50 1.10 1.80
80 0.80 0.80 1.40 2.40
150 1.50 2.00 3.76 6.10
C125 mode – bias:
pCOB2 B (mmHg) Bias
10 –0.56
40 0.07
60 0.19
80 0.01
150 –2.93
C125 mode – imprecision:
pCOB2 B (mmHg)
S B0B S BDB S BABLB S BXB
10 0.38 0.40 0.80 1.40
40 0.60 0.45 0.85 1.80
60 0.90 0.75 1.60 2.70
80 1.20 1.20 2.01 3.60
150 2.70 1.20 4.83 7.40
Continued on next page
pCOB2 B
ABL800 FLEX reference manual 5. Performance characteristics
5-53
Performance test results – pH, TpTCOB2 B, TpTOB2 B, TContinuedT
S250 mode – bias:
pOB2 B (mmHg) Bias
15 0.42
50 0.33
150 0.11
230 3.82
550 23.4
S250 mode – imprecision:
pOB2 B (mmHg) SB0B S BDB S BABLB S BXB
15 0.40 0.28 0.58 1.00
50 0.45 0.38 0.62 1.10
150 0.90 1.00 1.64 2.80
230 3.00 2.00 3.15 6.00
550 7.00 6.60 9.69 18.00
C125 mode – bias:
pOB2 B (mmHg) Bias
15 0.50
50 0.47
150 1.29
230 –7.17
550 –18.66
C125 mode – imprecision:
pOB2 B (mmHg) SB0B S BDB S BABLB S BXB
15 0.70 0.42 0.86 1.50
50 0.68 0.57 0.88 2.00
150 1.40 1.50 2.34 5.00
230 4.50 3.00 4.52 9.00
550 10.50 9.90 14.25 27.00
pOB2 B
5. Performance characteristics ABL800 FLEX reference manual
5-54
Performance test results – electrolytes
Parameter Reference method
TcTKP
+P, TcTNaP
+P, TcTCl P
−P
The standard sodium and potassium solutions are traceable to NIST-certified Standard Reference Material SRM 909b (human serum)
TcTCaP
2+P Radiometer method used.
The standard calcium solutions used are validated against corresponding NIST standard SRM 915 [4].
S250 mode – bias:
cKP
+P (mmol/L) Bias
2 –0.08
4 –0.06
8 –0.04
S250 mode – imprecision:
cKP
+P (mmol/L) S B0B S BDB S BABLB S BXB
2 0.020 0.010 0.078 0.090
4 0.040 0.020 0.080 0.100
8 0.030 0.038 0.122 0.160
C125 mode – bias:
cKP
+P (mmol/L) Bias
2 0.04
4 –0.01
8 –0.03
C125 mode – imprecision:
cKP
+P (mmol/L) S B0B S BDB S BABLB S BXB
2 0.060 0.025 0.093 0.140
4 0.060 0.048 0.106 0.170
8 0.080 0.070 0.144 0.230
Continued on next page
Reference methods
cKP
+P
ABL800 FLEX reference manual 5. Performance characteristics
5-55
Performance test results – electrolytes, TContinuedT
S250 mode – bias:
cNa P
+P (mmol/L) Bias
120 –0.58
140 –0.69
180 –0.57
S250 mode – imprecision:
cNa P
+P (mmol/L) S B0B S BDB S BABLB S BXB
120 0.40 0.30 0.97 1.30
140 0.40 0.30 1.05 1.30
180 0.50 0.30 1.41 1.65
C125 mode – bias:
cNa P
+P (mmol/L) Bias
120 0.43
140 0.03
180 –0.41
C125 mode – imprecision:
cNa P
+P (mmol/L) S B0B S BDB S BABLB S BXB
120 0.60 0.65 1.28 1.95
140 0.60 0.68 1.28 1.95
180 0.60 0.60 1.55 2.10
Continued on next page
cNa P
+P
5. Performance characteristics ABL800 FLEX reference manual
5-56
Performance test results – electrolytes, TContinuedT
S250 mode – bias:
cCl P
–P (mmol/L) Bias
85 –0.62
105 –0.57
140 –0.63
S250 mode – imprecision:
cCl P
–P (mmol/L) SB0B S BDB S BABLB S BXB
85 0.50 0.40 1.07 1.50
105 0.45 0.40 1.24 1.60
140 0.50 0.45 1.86 2.20
C125 mode – bias:
cCl P
–P (mmol/L) Bias
85 1.22
105 0.38
140 –1.23
C125 mode – imprecision:
cCl P
–P (mmol/L) S B0B S BDB S BABLB S BXB
85 0.75 0.72 1.28 2.25
105 0.75 0.75 1.40 2.40
140 0.75 1.10 2.11 3.30
Continued on next page
cCl P
–P
ABL800 FLEX reference manual 5. Performance characteristics
5-57
Performance test results – electrolytes, TContinuedT
S250 mode – bias:
cCa P
2+P (mmol/L) Bias
0.5 –0.012
1.25 –0.010
2.5 –0.008
S250 mode – imprecision:
cCa P
2+P (mmol/L) S B0B S BDB S BABLB S BXB
0.5 0.008 0.003 0.013 0.019
1.25 0.008 0.005 0.016 0.021
2.5 0.010 0.012 0.030 0.040
C125 mode – bias:
cCa P
2+P (mmol/L) Bias
0.5 0.003
1.25 0.023
2.5 0.046
C125 mode – imprecision:
cCa P
2+P (mmol/L) S B0B S BDB S BABLB S BXB
0.5 0.013 0.020 0.027 0.050
1.25 0.013 0.020 0.028 0.050
2.5 0.015 0.018 0.039 0.059
cCa P
2+P
5. Performance characteristics ABL800 FLEX reference manual
5-58
Performance test results – TcTGlu, TcTLac
Parameter Reference method
TcTGlu Spectrophotometry, using the hexokinase (HK) method recommended by NCCLS [5], measured on serum
TcTLac Spectrophotometry using a lactate dehydrogenase (LDH) method, measured on serum.
S250 mode – bias:
cGlu (mmol/L) Bias
2 0.02
5 –0.12
15 –0.21
S250 mode – imprecision:
cGlu (mmol/L) SB0B S BDB S BABLB S BXB
2 0.10 0.06 0.12 0.20
5 0.10 0.05 0.23 0.30
15 0.35 0.14 0.50 0.70
C125 mode – bias:
cGlu (mmol/L) Bias
C125 C35
2 0.03 0.07
5 0.06 0.04
15 0.12 0.25
C125 mode – imprecision:
cGlu (mmol/L) SB0B S BDB S BABLB S BXB
2 0.15 0.08 0.17 0.30
5 0.15 0.13 0.29 0.50
15 0.60 0.30 0.60 1.10
Continued on next page
Reference methods
cGlu
ABL800 FLEX reference manual 5. Performance characteristics
5-59
Performance test results – TcTGlu, TcTLac, TContinuedT
S250 mode – bias:
cLac (mmol/L) Bias
0.3 0.08
2 0.01
10 –0.42
S250 mode – imprecision:
cLac (mmol/L) S B0B S BDB S BABLB S BXB
0.3 0.10 0.06 0.13 0.23
2 0.10 0.06 0.15 0.23
10 0.20 0.20 0.61 0.85
C125 mode – bias:
cLac (mmol/L) Bias
C125 C35
0.3 0.09 0.03
2 0.15 0.04
10 0.20 0.15
C125 mode – imprecision:
cLac (mmol/L) S B0B S BDB S BABLB S BXB
0.3 0.15 0.12 0.19 0.35
2 0.15 0.10 0.21 0.35
10 0.45 0.30 0.73 1.30
cLac
5. Performance characteristics ABL800 FLEX reference manual
5-60
Performance test results – TcTtHb
Parameter Reference method
TcTtHb HiCN method recommended by NCCLS [6].
S250 – bias:
ctHb (g/dL) sOB2 B (%) Bias
15 0 –0.06
7 100 –0.03
15 100 0.06
25 100 0.48
S250 mode – imprecision:
ctHb (g/dL)
sOB2 B (%) S B0B S BDB S BABLB S BXB
15 0 0.30 0.09 0.27 0.40
7 100 0.30 0.13 0.22 0.40
15 100 0.30 0.10 0.27 0.40
25 100 0.50 0.10 0.48 0.70
C125 – bias:
ctHb (g/dL) sOB2 B (%) Bias
15 0 0.19
7 100 –0.08
15 100 0.12
25 100 –0.11
C125 mode – imprecision:
ctHb (g/dL)
sOB2 B (%) S B0B S BDB S BABLB S BXB
15 0 0.50 0.15 0.36 0.80
7 100 0.50 0.20 0.352 0.80
15 100 0.50 0.17 0.35 0.80
25 100 0.70 0.15 0.66 1.00
Reference method
ctHb
ABL800 FLEX reference manual 5. Performance characteristics
5-61
Performance test results – oximetry
The optical system is unchanged in the ABL800 FLEX analyzers compared to the ABL700 Series. Test of TcTtHb on the ABL800 FLEX analyzers has been conducted (results given on the previous page). The other oximetry parameters have not been retested; the information and results below are from the ABL700 Series.
Parameter Reference method
Oximetry The reference method established for the oximetry parameters uses modified ABL520 analyzers as the field reference instruments. The ABL520 analyzers have been validated and their performance specifications determined according to primary reference methods.
The modified ABL520 analyzers are used in accordance with IFCC’s recommendations for traceability of reference methods.
The reference methods used for the oximetry parameters on the ABL520 analyzers are those presented below.
sOB2 B Tonometry: whole blood is tonometered with a gas mixture containing 94.4 % OB2 B and 5.6 % COB2 B.
FHHb The standard is blood (TcTtHb = 13-15 g/dL) treated with dithionite.
FCOHb Gas chromatography. The standards are carbon monoxide mixtures with atmospheric air, whose purity is validated in accordance with NIST SRM 1678 (50 ppm CO in NB2 B).
FMetHb Spectrometry, modified Evelyn-Malloy method [7].
FHbF Alkali denaturation method [9]. Corresponds to NCCLS guideline [10].
Test Description
Reference To verify that the correction constants have been accurately determined, 10 analyzers with all parameters available are tested in C195 mode against the reference methods. Each parameter is tested on 3-6 levels over at least three days, with five repetitions each day. (Five new analyzers with all parameters available were tested against the reference methods for FHbF.) Bias for each parameter in the C195 measuring mode against the reference method is determined.
Continued on next page
Explanation
Reference methods
Test conditions for oximetry parameters
5. Performance characteristics ABL800 FLEX reference manual
5-62
Performance test results – oximetry, TContinuedT
Test Description
Verification 6-10 ABL700 Series analyzers are tested over at least 2 days for all levels. Bias for the given mode is calculated as difference compared to the C195 mode.
Bias against the reference method is determined as follows:
Bias = bias against C195 + C195 bias against reference method.
The following parameters: sOB2 B, FCOHb, FMetHb and FOB2 BHb are measured directly against the reference built into the analyzer, and these parameters are independent of the reference method.
Reduced verification 6-10 new analyzers are used over at least 1 day for selected levels.
Bias for the tested mode is calculated as follows:
Bias = bias against C195 + C195 bias against reference method.
Untested modes are described as "N/A".
Simple verification 6-10 analyzers are tested at one extreme level over one day. Bias is not determined; bias values for the modes with similar wet section programs are used.
S250 mode – bias:
ctHb (g/dL) sOB2 B (%) Bias
15 0 –0.01
7 100 –0.12
15 100 0.00
25 100 –0.03
S250 mode – imprecision:
ctHb (g/dL)
sOB2 B (%) S B0B S BDB S BABLB S BXB
15 0 0.50 0.12 0.60 1.00
7 100 0.30 0.15 0.42 0.60
15 100 0.30 0.15 0.45 0.60
25 100 0.50 0.16 0.67 1.00
Continued on next page
Test conditions for oximetry parameters T(continued)
sOB2 B
ABL800 FLEX reference manual 5. Performance characteristics
5-63
Performance test results – oximetry, TContinuedT
C125 – bias:
ctHb (g/dL) sOB2 B (%) Bias
15 0 –0.04
7 100 –0.14
15 100 –0.12
25 100 –0.13
C125 mode – imprecision:
ctHb (g/dL)
sOB2 B (%) S B0B S BDB S BABLB S BXB
15 0 0.50 0.15 0.36 1.00
7 100 0.30 0.20 0.41 0.60
15 100 0.30 0.15 0.42 0.60
25 100 0.50 0.12 0.65 1.00
S250 mode – bias:
FOB2 BHb (%) Bias
ctHb (g/dL) sOB2 B (%)
15 0 –0.12
7 100 –0.03
15 100 0.02
25 100 –0.01
S250 mode – imprecision:
ctHb (g/dL)
FOB2 BHb(%) S B0B S BDB S BABLB S BXB
15 0 0.05 0.05 0.25 0.30
7 100 0.25 0.20 0.50 0.60
15 100 0.15 0.15 0.45 0.50
25 100 0.10 0.10 0.40 0.45
Continued on next page
TsTOB2 B T(continued)T
FOB2 BHb
5. Performance characteristics ABL800 FLEX reference manual
5-64
Performance test results – oximetry, TContinuedT
C125 mode – bias:
FOB2 BHb (%) Bias
ctHb (g/dL) sOB2 B (%)
15 0 –0.02
7 100 –0.39
15 100 –0.40
25 100 –0.46
C125 mode – imprecision:
ctHb (g/dL)
FOB2 BHb(%) S B0B S BDB S BABLB S BXB
15 0 0.05 0.05 0.25 0.30
7 100 0.25 0.20 0.50 0.60
15 100 0.15 0.15 0.45 0.50
25 100 0.10 0.10 0.40 0.45
S250 mode – bias:
FCOHb (%) Bias
TcTtHb (g/dL) sOB2 B (%) FCOHb (%)
15 100 0 0.03
7 100 20 N/A
15 100 20 N/A
25 100 20 N/A
S250 mode – imprecision:
ctHb g/dL)
sOB2 B (%) FCOHb (%)
S B0B S BDB S BABLB S BXB
15 100 0 0.05 0.10 0.35 0.40
7 100 20 0.10 0.10 0.75 0.80
15 100 20 0.05 0.10 0.70 0.75
25 100 20 0.05 0.10 0.70 0.75
Continued on next page
TFTOB2 BHb T(continued)
FCOHb
ABL800 FLEX reference manual 5. Performance characteristics
5-65
Performance test results – oximetry, TContinuedT
C125 mode – bias:
FCOHb (%) Bias
ctHb (g/dL) sOB2 B (%) FCOHb (%)
15 100 0 0.10
7 100 20 N/A
15 100 20 N/A
25 100 20 N/A
C125 mode – imprecision:
ctHb g/dL)
sOB2 B (%) FCOHb (%)
S B0B S BDB S BABLB S BXB
15 100 0 0.05 0.10 0.35 0.40
7 100 20 0.10 0.10 0.75 0.80
15 100 20 0.05 0.10 0.70 0.75
25 100 20 0.05 0.10 0.70 0.75
S250 mode – bias:
FMetHb (%) Bias
TcTtHb (g/dL) sOB2 B (%) FMetHb (%)
15 100 0 0.01
7 100 20 N/A
S250 mode – imprecision:
ctHb g/dL)
sOB2 B (%) FMetHb (%)
S B0B S BDB S BABLB S BXB
15 100 0 0.10 0.10 0.25 0.30
7 100 20 0.05 0.10 0.35 0.40
Continued on next page
TFTCOHb T(continued)
FMetHb
5. Performance characteristics ABL800 FLEX reference manual
5-66
Performance test results – oximetry, TContinuedT
C125 mode – bias:
FMetHb (%) Bias
TcTtHb (g/dL) sOB2 B (%) FMetHb (%)
15 100 0 0.14
7 100 20 N/A
15 100 20 N/A
25 100 20 N/A
C125 mode – imprecision:
ctHb g/dL)
sOB2 B (%) FMetHb (%)
S B0B S BDB S BABLB S BXB
15 100 0 0.10 0.10 0.25 0.30
7 100 20 0.05 0. 10 0.35 0.40
S250 mode – bias:
FHHb (%) Bias
TFTHHb (%) ctHb (g/dL)
0 15 –0.01
S250 mode – imprecision:
FHHb (%) tHb (g/dL) SB0B S BDB S BABLB S BXB
0 15 0.05 0.10 0.30 0.35
C125 mode – bias:
FHHb (%) Bias
TFTHHb (%) ctHb (g/dL)
0 15 N/A
C125 mode – imprecision:
FHHb (%) TcTtHb (g/dL) S B0B S BDB S BABLB S BXB
0 15 0.05 0.10 0.30 0.35
Continued on next page
TFTMetHb T(continued)
FHHb
ABL800 FLEX reference manual 5. Performance characteristics
5-67
Performance test results – oximetry, TContinuedT
S250 mode – bias:
FHbF (%) Bias
TFTHbF (%) ctHb (g/dL)
0 10 3.3
0 15 5.5
0 20 5.6
S250 mode – imprecision:
FHbF (%)
TcTtHb (g/dL)
TsTOB2 B (%) S B0B S BDB S BABLB S BXB
0 10 100 4 4 5 8
0 15 100 2 3 7 8
0 20 100 2 2 10 11 NOTES: a, b.
C125 mode – bias:
FHbF (%) Bias
TFTHbF (%) ctHb (g/dL)
0 10 3.3
0 15 5.5
0 20 5.6
C125 mode – imprecision:
FHbF (%)
TcTtHb (g/dL)
TsTOB2 B (%) S B0B S BDB S BABLB S BXB
0 10 100 4 4 5 8
0 15 100 2 3 5 7
0 20 100 2 2 10 11 NOTES: a, b.
Continued on next page
FHbF – adult blood
5. Performance characteristics ABL800 FLEX reference manual
5-68
Performance test results – oximetry, TContinuedT
S250 mode – bias:
FHbF (%) Bias
TFTHbF (%) ctHb (g/dL)
0 10 5.9
0 15 3.3
0 20 2.6
S250 mode – imprecision:
FHbF (%)
TcTtHb (g/dL)
TsTOB2 B (%) S B0B S BDB S BABLB S BXB
0 10 100 4 5 5 9
0 15 100 3 3 6 8
0 20 100 2 3 6 7 NOTES: a, b.
C125 mode – bias:
FHbF (%) Bias
TFTHbF (%) ctHb (g/dL)
0 10 5.9
0 15 3.3
0 20 2.6
C125 mode – imprecision:
FHbF (%)
TcTtHb (g/dL)
TsTOB2 B (%) S B0B S BDB S BABLB S BXB
0 10 100 4 5 6 9
0 15 100 3 3 6 8
0 20 100 2 3 6 7 NOTES: a, b.
Continued on next page
FHbF – fetal blood
ABL800 FLEX reference manual 5. Performance characteristics
5-69
Performance test results – oximetry, TContinuedT
The following corrections should be geometrically added to SBInstB and SBXB for the analyzer's wavelength calibrated with the S7770:
Parameter Mode Level Correction (percentage
point)
Macromode All 0 TcTtHb
Micromode All 0
sOB2 B All sOB2 B (100 %) 0.23
FOB2 BHb All FOB2 BHb (100 %) 0.15
FCOHb All FCOHb (20 % and 0 %) 0.40
FHHb All FHHb (0 %) 0.23
NOTES:
a. pH = 7.4 ± 0.1. FHbF is adjusted with the pH sensitivity to a nominal pH = 7.4. For further details please refer to the TInterference Tests T section for oximetry parameters.
b. Specifications for imprecision are derived from worst-case values found during internal laboratory tests. 40 % relative is then added as a safety factor.
Contribution to Imprecision Specifications from S7770
5. Performance characteristics ABL800 FLEX reference manual
5-70
Performance test results – bilirubin
The standard bilirubin solution is traceable to Bechman C20 Standard Reference Material.
As the optical system is unchanged in the ABL8x7 FLEX analyzers compared to the ABL700 Series, the specifications for bilirubin have not been re-established – see the description of bilirubin in section TABL8x0/8x5 Performance characteristicsT earlier in this chapter.
S250 mode – bias:
ctBil (μmol/L) ctHb (g/dL) Bias
≈0 Plasma 0.7
≈0 15 6.6
≈0 20 9.9
≈400 Plasma 5.0
≈400 15 7.0
≈400 20 8.1
S250 mode – imprecision:
ctBil (μmol/L)
ctHb (g/dL)
S B0B S BDB S BABLB S BXB
≈0 Plasma 1.1 2.3 3.4 6.0
≈0 15 3.0 2.9 5.7 9.9
≈0 20 4.0 5.5 9.4 17.5
≈400 Plasma 2.5 2.5 11.6 16.0
≈400 15 6.0 5.0 14.6 21.7
≈400 20 9.0 10.0 17.7 29.2
Notes a, b, c.
Continued on next page
Reference method
TcTtBil
ABL800 FLEX reference manual 5. Performance characteristics
5-71
Performance test results – bilirubin, TContinuedT
C125 mode – bias:
ctBil (μmol/L) ctHb (g/dL) Bias
≈0 Plasma –0.1
≈0 15 1.6
≈0 20 2.7
≈400 Plasma –10.2
≈400 15 –0.4
≈400 20 –2.2
C125 mode – imprecision:
ctBil (μmol/L)
ctHb (g/dL)
S B0B S BDB S BABLB S BXB
≈0 Plasma 1.1 2.3 3.4 6.0
≈0 15 3.0 2.9 5.8 9.9
≈0 20 4.0 3.0 11.3 17.5
≈400 Plasma 5.0 2.5 11.6 19.0
≈400 15 9.0 5.0 14.6 22.3
≈400 20 11.0 6.5 19.6 33.9
Notes a, b, c.
NOTES:
a. Adult/fetal blood, pH = 7.4 ± 0.1, normal MCHC and albumin variation. Spiked with unconjugated bilirubin.
b. TcTtBil specification at level 200 μmol/L is interpolated from the measured specifications at 0 and 400 μmol/L.
c. The performance specifications apply to measurements performed using CLINITUBES capillary tubes with clot catchers and mixing wire from Radiometer.
TcTtBil T(continued)T
5. Performance characteristics ABL800 FLEX reference manual
5-72
Performance test conditions and results – TcTCrea
This section describes performance characteristics for the TcTCrea parameter and includes the following topics:
• Definitions
• Samples
• Test conditions
• Performance – whole blood samples
• Performance – plasma samples
• Performance – quality control
• Performance – NIST909b standards
• Effect of Hct concentration
• Linearity – ABL837 FLEX analyzer versus HPLC reference method
• Detection limit
• Comparison study (serum) – enzymatic method
• Comparison study (serum) – Jaffé method
• Comparison study – whole blood
• Imprecisions for the macro- and micromodes
The following terms are used to describe the performance specifications of the ABL8x7 FLEX analyzers:
Term Definition
Repeatability (within-run precision)
The term "within-run imprecision" is identical to the ISO term "repeatability", i.e. the closeness of the agreement between results of successive measurements of the same measurand carried out under the same conditions of measurement [11].
Reproducibili-ty (total imprecision)
The ISO term "reproducibility" describes the closeness of agreement of results of measurements under changed conditions. Reproducibility may include the following: “within-run imprecision”, “run-to-run imprecision,” “day-to-day imprecision” or/and "Instrument-to-instrument imprecision" [11].
S By/x B Standard Deviation about regression line
X Mean value
R Correlation coefficient (used in the second power: RP
2P)
N number of measurements on several analyzers used for the test
Continued on next page
Introduction
Definitions
ABL800 FLEX reference manual 5. Performance characteristics
5-73
Performance test conditions and results – TcTCrea, TContinuedT
Preparation: Different experiments were made in order to investigate the cCrea performance on the ABL837 FLEX analyzer under different conditions.
The concentration of creatinine in the samples was reached by spiking with a stock solution.
The samples used in these experiments were serum/plasma, whole-blood or aqueous samples. Aqueous and serum/plasma samples were spiked directly. Whole-blood samples were prepared by separating the erythrocyte and plasma/serum layer. The plasma/serum layer was then spiked and afterwards mixed with erythrocyte layer. In order to reach different hematocrit concentrations in the samples, different volumes of plasma/serum and erythrocyte layer were mixed. Equilibrium in the whole-blood samples was reached by gently mixing the samples for 2 hours at room temperature.
The whole blood samples were prepared the day of the analysis and not stored.
Storage of serum/plasma samples:
In order to prevent creatinine – creatine equilibrium in the serum/plasma samples, they were stored in the freezer at either –20 °C or –80 °C.
TUMaterials and methodsUT:
The performance on whole-blood samples was tested during three days. Five whole-blood samples with different creatinine concentrations were prepared every day. Each sample was measured five times in succession on nine ABL837 FLEX analyzers – see the table below.
Day TcTCrea(ABL837) (µmol/L)
Repeatability (CV%)
Reproducibility* (CV%)
N
1 87 1.3 3.6 45
1 320 1.4 2.5 45
1 566 1.2 2.5 45
1 1090 1.1 2.9 45
1 1575 1.0 3.3 45
2 55 1.1 4.6 45
2 252 1.0 2.5 45
2 538 1.1 2.7 45
2 1059 1.4 2.8 45
2 1575 1.4 3.4 45
3 59 1.5 3.7 45
Continued on next page
Samples
Performance – whole blood samples
5. Performance characteristics ABL800 FLEX reference manual
5-74
Performance test conditions and results – cCrea, TContinuedT
Day TcTCrea(ABL837) (µmol/L)
Repeatability (CV%)
Reproducibility* (CV%)
N
3 267 1.5 2.8 45
3 543 1.4 3.0 45
3 1046 1.1 3.0 45
3 1548 1.4 3.1 45
* Reproducibility includes within-run and instrument-to-instrument imprecisions.
TUMaterials and methods:
The precision performance on plasma samples was tested following the guidelines described in CLSI EP5-A [13]. The performance was evaluated during 20 test days where two separate runs (run = double determination) were performed each day on three different plasma samples with normal, medium and high concentration of creatinine, respectively. The study was performed on two ABL837 FLEX analyzers – see the table below.
Analyzer TcTCrea(ABL837) (μmol/L)
Repeatability (CV%)
Reproducibility* (CV%)
N
1 63 0.8 2.2 80
1 241 0.9 1.9 80
1 568 0.6 2.2 80
2 64 0.9 1.9 80
2 244 1.1 1.9 80
2 574 0.8 2.2 80
* Reproducibility includes within-run, run-to-run and day-to-day imprecisions.
TUMaterials and methods:
The precision performance on S7835, S7845 and S7855 AutoCheck6+ solutions was tested during a period of 24 days – see the table below.
TcTCrea(ABL837) (μmol/L)
Repeatability (CV%) Reproducibility* (CV%) N
238 1.4 2.5 490
30 1.8 3.1 494
460 1.1 2.3 486
* Reproducibility includes within-run, run-to-run, day-to-day and instrument-to-instrument imprecisions.
Continued on next page
Performance – whole blood samples T(continued)
Performance – plasma samples
Performance – quality control
ABL800 FLEX reference manual 5. Performance characteristics
5-75
Performance test conditions and results – cCrea, TContinuedT
TUMaterials and methods:
Bias against the NIST 909b was determined as follows: two levels of NIST 909b were tested during one day in order to check the bias of the creatinine parameter. The two levels of NIST 909b were prepared according to the recommended procedure; each level was measured twice on 10 ABL837 FLEX analyzers – see the table below.
NIST 909b (μmol/L)
TcTCrea(ABL8x7) (μmol/L)
Bias (%)
Repeatability (CV%)
Reproducibility* (CV%)
N
56.18 ± 0.55 57.3 2.1 0.3 1.4 20
467.4 ± 5.3 470.7 0.7 0.8 1.3 20
* Reproducibility includes within-run and instrument-to-instrument imprecisions.
The creatinine measurement on the ABL837 FLEX analyzer in whole blood has been standardized towards serum/plasma creatinine because it is an established international point of reference for predicate laboratory measurements of creatinine in the clinical setting.
This has been done in two steps:
• The relation between the used reference method, HPLC, and the ABL837 FLEX analyzer has been established on both serum and plasma samples.
• The relation between plasma and whole blood samples has been established on the ABL837 FLEX analyzer.
Traceability between whole-blood measurements on the ABL837 FLEX analyzer and serum/plasma measurements on the HPLC reference method has thereby been established.
The primary working standards are prepared from NIST SRM 914a (creatinine) and used to determine the creatinine concentration of seven serum pool standards and seven plasma standards. The measurements are performed using Reversed Phase HPLC (High Performance Liquid Chromatography). The method has been validated, using NIST SRM 909b (human serum).
TUMaterials and methodsUT:
A serum pool was dialyzed in order to remove all creatinine and creatine, and seven serum pool levels with different creatinine concentrations (25 to 1900 μmol/L) and zero creatine were prepared. The seven serum pools were measured on an HPLC reference method [12] and the mean values of the HPLC method were used to assign reference values to seven serum pool levels. The linear response of the TcTCrea(ABL8x7) was studied by performing measurements on the seven serum pool levels during three days. Three measurements were made on each level every day on nine ABL837 FLEX analyzers. A total of 535 measurements were made.
Continued on next page
Performance –NIST 909b standards
Linearity
Linearity – reference method
Linearity – serum samples
5. Performance characteristics ABL800 FLEX reference manual
5-76
Performance test conditions and results – cCrea, TContinuedT
The TcTCrea results from the HPLC reference method and the ABL837 FLEX analyzer show linear relationship.
Regression Plot
The equation is as follows:
TcTCrea(ABL8x7) = 0.993 × TcTCrea(HPLC) – 0.5
where S By/x B = 22.4 μmol/L, RP
2P = 0.999, N = 535
The Bias % Plot shows comparison of ABL837 and HPLC TcTCrea measurements.
(%)100Crea(HPLC)
Crea(HPLC)7)Crea(ABL8xBias ×−
=c
cc
Continued on next page
Linearity – serum samples T(continued)
TcTCrea(HPLC) (μmol/L)
TcTCrea(ABL837) (μmol/L)
ABL800 FLEX reference manual 5. Performance characteristics
5-77
Performance test conditions and results – cCrea, TContinuedT
Bias % Plot
TUMaterials and methodsUT:
A plasma pool was dialyzed in order to remove all creatinine and creatine, and seven plasma pool levels with different creatinine concentrations (25 to 2000 μmol/L) and zero creatine were prepared. The seven plasma pools were measured on an HPLC reference method [12] and the mean values of the HPLC method were used to assign reference values to seven plasma pool levels. The linear response of the TcTCrea(ABL8x7) was studied by performing measurements on the seven plasma pool levels. Three measurements were made on each level on nine ABL837 FLEX analyzers. A total of 184 measurements were made.
The TcTCrea results from the HPLC reference method and the ABL837 FLEX analyzer (see regression plot on the next page) show linear relationship.
Continued on next page
Linearity – serum samples T(continued)T
Linearity – plasma samples
Bias (%)
TcTCrea(HPLC) (μmol/L)
95 % prediction limits
5. Performance characteristics ABL800 FLEX reference manual
5-78
Performance test conditions and results – cCrea, TContinuedT
Regression Plot
The equation is as follows:
TcTCrea(ABL8x7) = 0.977 × TcTCrea(HPLC) + 1.7
where S By/x B = 16.6 μmol/L, RP
2P = 0.999, N = 184
The Bias % Plot shows comparison of ABL837 and HPLC TcTCrea measurements.
(%)100Crea(HPLC)
Crea(HPLC)7)Crea(ABL8xBias ×−
=c
cc
TUContinued on next page
Linearity – plasma samples T(continued) TcTCrea(ABL837)
(μmol/L)
TcTCrea(HPLC) (μmol/L)
ABL800 FLEX reference manual 5. Performance characteristics
5-79
Performance test conditions and results – cCrea, TContinuedT
Bias % PlotT
T
UMaterials and methods:
Whole-blood samples with different creatinine concentrations in the range of 41 to 1852µmol/L were used. Each sample was measured as duplicate or triplicate on ABL837 FLEX analyzers. The samples were then centrifuged. The plasma samples thus obtained were also measured as duplicates or triplicates on the same ABL837 FLEX analyzers. This experiment was performed both internally at Radiometer and externally at a hospital on patient samples.
Correlation between whole blood and plasma measurements on the ABL837 FLEX analyzer is shown on the regression plot.
Continued on next page
Linearity – plasma samples T(continued)T
Linearity – whole blood versus plasma
Bias (%)
TcTCrea(HPLC) (μmol/L)
95 % prediction limits
5. Performance characteristics ABL800 FLEX reference manual
5-80
Performance test conditions and results – cCrea, TContinuedT
Regression Plot
The plot shows a small difference between results on whole blood and plasma (linear relationship).
The ABL8x7 FLEX analyzers are designed to measure cCrea on whole blood. If plasma, serum or NIST SRM is to be measured on the ABL8x7 FLEX analyzers, cCrea can be corrected as follows:
TcTCrea(ABL8x7,plasma/serum) μmol/L = 0.950 × TcTCrea(ABL8x7,whole blood) – 0.4
TcTCrea(ABL8x7,plasma/serum) mg/dL = 0.950 × TcTCrea(ABL8x7,whole blood) – 0.005
where TcTCrea(ABL8x7,whole blood) are the results obtained on the analyzer.
UMaterials and methodsU:
Seven levels of creatinine serum pool were prepared in order to test the effect of creatine concentrations on the linearity of the TcTCrea(ABL8x7).
For each creatinine level three samples with 0, 100 and 250 μmol/L creatine, respectively, were prepared by spiking. The samples without creatine were prepared in order to obtain reference values (the same volume was added to these samples as for the 100 and 250 μmol/L creatine samples).
Each sample (three creatine levels and seven creatinine levels; 21 samples in total) was measured three times on eight ABL837 FLEX analyzers every day over three days. A total of 72 measurements were made on each sample.
Continued on next page
Linearity – whole blood versus plasma T(continued)T
Creatinine linearity at various creatine concentrations
TcTCrea(ABL837, whole blood), μmol/L
TcTCrea(ABL837, plasma), μmol/L
ABL800 FLEX reference manual 5. Performance characteristics
5-81
Performance test conditions and results – cCrea, TContinuedT
The cCrea (ABL837) results on samples with 100 μmol/L creatine show linear relationship.
Regression Plot
The equation is as follows:
TcTCrea(100 μmol/L creatine) = 1.000 × TcTCrea(0 μmol/L creatine) – 3.2
where S By/x B = 35.6 μmol/L, RP
2P = 0.999, N = 492
The cCrea(ABL837) results on samples with 250 μmol/L creatine show linear relationship (see the next page):
Continued on next page
Creatinine linearity at various creatine concentrations T(continued)T
Assigned mean value for TcTCrea(ABL837) (μmol/L)
5. Performance characteristics ABL800 FLEX reference manual
5-82
Performance test conditions and results – cCrea, TContinuedT
Regression Plot
The equation is as follows:
TcTCrea(250 μmol/L creatine) = 0.993 × TcTCrea(0 μmol/L creatine) + 0.9
where S By/x B = 35.6 μmol/L, RP
2P = 0.999, N = 492
UMaterials and methodsU:
Five samples with different hematocrit concentrations in the range of 20 to 70 % were prepared. Each sample was measured three times on 10 ABL837 FLEX analyzers – see the table below.
Hct (%) TcTCrea(ABL837) (µmol/L)
Δ TcTCrea (μmol/L)*
N
24 66 3 30
31 65 2 30
46 63 0 30
57 62 –1 30
69 59 –4 30
*Δ TcTCrea = TcTCrea (Hct) – TcTCrea (Hct = 46 %)
Continued on next page
Creatinine linearity at various creatine concentrations T(continued)T
Effect of Hct concentration
Assigned mean value for TcTCrea(ABL837) (μmol/L)
ABL800 FLEX reference manual 5. Performance characteristics
5-83
Performance test conditions and results – TcTCrea, TContinuedT
UMaterials and methods:
The detection limit of the TcTCrea(ABL8x7) parameter was determined following the guidelines in CLSI EP17-A [14].
Limit of Blank (LoB): In order to determine LoB, three samples with TcTCrea = 0 μmol/L and creatine concentrations of 0, 50 and 125 μmol/L were prepared. The samples were then measured on seven ABL837 FLEX analyzers over two days. Limit of Blank (LoB) was determined, using the following equation:
LoB = mean + 1.645 × Standard Deviation at TcTCrea = 0 μmol/L
Limit of Detection (LoD): The same procedure was followed to determine LoD on samples with TcTCrea = 10 μmol/L at 0, 50 and 125 μmol/L creatine concentrations. Limit of Detection (LoD) was determined, using the following equation:
LoD = LoB + 1.645 × Standard Deviation at TcTCrea = 10 μmol/L.
LoB (μmol/L creatinine)
N LoD (μmol/L creatinine)
N
3.2 70 5.3 70
The LoD of the cCrea was found to be 5.3 µmol/L on the ABL837 FLEX analyzer.
Limit of Quantitation (LoQ): To determine LoQ, samples with TcTCrea = 25μmol/L and creatine concentrations at 0, 50 and 125 μmol/L were measured in order to compare the total error of TcTCreaBABL837B at 10 and 25 μmol/L creatinine.
The LoQ (Limit of Quantitation) is the lowest actual amount of creatinine that can be reliably detected (the LoD) and at which the total error meets the requirements (acceptable for clinical use). So if the total error of the measurement used in the LoD study is acceptable, then LoQ = LoD.
The total analytical error is determined using the following equation:
Total Error = ± (|Bias| + 2 × Standard Deviation)
The table below shows the TcTCrea(ABL8x7) results for samples with creatinine at 10 and 25 µmol/L:
Level
(μmol/L)
Bias
(μmol/L)
SD
(μmol/L)
Total Error
(μmol/L)
N
10 1.7 1.3 4.4 70
25 1.7 1.5 4.7 67
Continued on next page
Detection limit
5. Performance characteristics ABL800 FLEX reference manual
5-84
Performance test conditions and results – cCrea, TContinuedT
The total Error of cCrea measurements on the ABL837 FLEX analyzer for the LoD study (TcTCrea = 10 μmol/L) is acceptable for clinical use and the limit of Quantitation therefore equals the Limit of Detection:
LoQ ( TcTCrea(ABL8x7)) = LoD ( TcTCrea(ABL8x7)) = 5.3 μmol/L creatinine
The lower detection limit of the ABL8x7 FLEX analyzer was set at 10 μmol/L creatinine; the values lower than 10µmol/L will not be reported by the analyzer.
UMaterials and methodsU: In order to compare repeatability and bias of the TcTCrea(ABL8x7) parameter with an enzymatic laboratory routine method, the guideline CLSI EP9-A [15] was followed. In the study performed at a university hospital in Denmark, the enzymatic method of COBAS Integra 700 from Roche (called Integra for short) was used as the comparative method. 104 patient (serum) samples (CLSI recommends a minimum of 40 samples) in a clinically significant range of 15-1263 μmol/L were collected. The samples were measured twice by both methods.
From duplicate 104 patient samples covering the range 15-1263 μmol/L creatinine, statistical calculation of repeatability was made for the Integra and the ABL837 FLEX analyzer:
Method Repeatability (CV%) N (number of duplicates)
Integra 1.5 104
ABL837 2.8 104
X100DeviationStandard%CV ×
=
where
( )
N2
meas2-meas1DeviationStandard
N
1
2∑=
where meas1 and meas2 are two measurements on the same sample, i.e. duplicates.
The regression plot (see the next page) describes the relationship between TcTCrea(ABL8x7) and TcTCrea(Integra).
Continued on next page
Detection limit T(continued)T
Comparison study (serum) – enzymatic method
ABL800 FLEX reference manual 5. Performance characteristics
5-85
Performance test conditions and results – cCrea, TContinuedT
Regression Plot
The relationship between TcTCrea(ABL8x7) and TcTCrea(Integra) is as follows:
TcTCrea(ABL8x7) = 1.014 × TcTCrea(Integra) + 1.8
where S By/x B = 13.4 μmol/L, RP
2P = 0.999, N = 208
The Bias % Plot (see the next page) shows comparison of the TcTCrea(ABL8x7) and TcTCrea(Integra).
( ) (%)100ra)Crea(Integ7)Crea(ABL8xMean
ra)Crea(Integ7)Crea(ABL8xBias ×+
−=
cccc
Continued on next page
Comparison study (serum) – enzymatic method T(continued)T
TcTCrea(ABL837) (μmol/L)
TcTCrea(Integra) (μmol/L)
5. Performance characteristics ABL800 FLEX reference manual
5-86
Performance test conditions and results – cCrea, TContinuedT
Bias % Plot Comparison study (serum) –enzymatic method T(continued)T
Comparison study (serum) –Jaffé method
Bias (%)
Note that the results outside thecorrespond to TcTCrea values of 20
UMaterials and methods: In order to compare repeatability Jaffé laboratory routine method, t In the study performed at a univemethod of Modular from Roche (– called Modular for short – was (serum) samples (CLSI recommesignificant range of 21-1263 μmomeasured twice by both methods.From duplicate 54 patient samplestatistical calculation of repeatabiABL837 FLEX analyzer:
Method Repeatab
Modular
ABL837
Mean (TcTCre
95 % prediction limits
prediction limits in the low concentration range -30 μmol/L.
and bias of the TcTCrea (ABL8x7) parameter with a he guideline CLSI EP9-A [15] was followed.
rsity hospital in Denmark, the Jaffé-rate blank Hitachi instrument with Boehringer Manheim kit) used as the comparative method. 54 patient nds a minimum of 40 samples) in a clinically l/L creatinine were collected. The samples were s covering the range 21-1263 μmol/L creatinine, lity was calculated for the Modular and the
ility (CV%) N (number of duplicates)
1.4 54
2.5 54
Continued on next page
a(ABL8x7) + TcTCrea(Integra)) (μmol/L)
ABL800 FLEX reference manual 5. Performance characteristics
5-87
Performance test conditions and results – cCrea, TContinuedT
X100DeviationStandard%CV ×
=
where
( )
N2
meas2-meas1DeviationStandard
N
1
2∑=
where meas1 and meas2 are measurements on the same sample, i.e. duplicates.
The Regression Plot below describes the relationship between TcTCrea(ABL8x7) and TcTCrea(Modular).
Regression Plot
Continued on next page
Comparison study (serum) – Jaffé method T(continued)T
TcTCrea(ABL837) (μmol/L)
TcTCrea(Modular) (μmol/L)
5. Performance characteristics ABL800 FLEX reference manual
5-88
Performance test conditions and results – TcTCrea, TContinuedT
The relationship between TcTCrea(ABL8x7) and TcTCrea(Modular)B Bis as follows:
TcTCrea(ABL8x7) = 1.154 × TcTCrea(Modular) – 33.2
where S By/x B = 15.8 μmol/L, RP
2P = 0.999, N = 108
The Bias % Plot below shows comparison of the TcTCrea(ABL8x7) and TcTCrea(Modular).
( ) (%)100ar)Crea(Modul7)Crea(ABL8xMean
ar)Crea(Modul7)Crea(ABL8xBias ×+
−=
cccc
Bias % Plot
UMaterials and methodsU: In order to compare repeatability and bias of the TcTCrea(ABL8x7) parameter with a whole blood creatinine method, the guideline CLSI EP9-A [15] was followed. In the study performed at a hospital in the USA, the i-STAT creatinine cartridges were used as the comparative method. 55 patient (heparinized whole blood) samples (CLSI recommends a minimum of 40 samples) in a clinically significant range of 44-1743 μmol/L were collected. The samples were measured twice by both methods.
Continued on next page
Comparison study (serum) – Jaffé method T(continued)T
Comparison study – whole blood
Bias (%)
Mean (TcTCrea(ABL8x7) + TcTCrea(Modular)) (μmol/L)
95 % prediction limits
ABL800 FLEX reference manual 5. Performance characteristics
Performance test conditions and results – cCrea, TContinuedT
From duplicate 55 patient samples covering the range 44-1743 μmol/L creatinine, statistical calculation of repeatability was calculated for the i-STAT and the ABL837 FLEX analyzer:
Method Repeatability (CV%) N (number of duplicates)
i-STAT 3.3 55
ABL837 2.1 55
X100DeviationStandard%CV ×
=
where
N
w
TTcTC
Comparison study – whole blood T(continued)
TcTCrea(AB(μmol/L)
5-89
( )
N2
meas2-meas1DeviationStandard 1
2∑=
here meas1 and meas2 are measurements on the same sample, i.e. duplicates.
he Regression Plot below describes the relationship between TcTCrea(ABL8x7) and rea(i-STAT).
Regression Plot
Continued on next page
L837)
TcTCrea(i-STAT) (μmol/L)
5. Performance characteristics ABL800 FLEX reference manual
5-90
Performance test conditions and results – cCrea, TContinuedT
The relationship between TcTCrea(ABL8x7) and TcTCrea(i-STAT)B Bis as follows:
TcTCrea(ABL8x7) = 0.990 × TcTCrea(i-STAT) – 18.5
where S By/x B = 38.2 μmol/L, RP
2P = 0.997, N = 110
The Bias % Plot below shows comparison of the TcTCrea(ABL8x7) and TcTCrea(i-STAT).
( ) (%)100STAT)-Crea(i7)Crea(ABL8xMean
STAT)-Crea(i7)Crea(ABL8xBias ×+
−=
cccc
Bias % Plot
Comparison study – whole blood T(continued)T
Imprecision
Bias (%)
95 % prediction limits
Test conditions were as follows:
Item Description
Primary reference method
Reverse Phase HPLC
Analyzers and test modes
10 ABL837 were tested over three days in the following modes:
• syringe S250
• capillary C125
Continued on next page
Mean (TcTCrea(ABL8x7) + TcTCrea(i-STAT)) (μmol/L)
ABL800 FLEX reference manual 5. Performance characteristics
5-91
Performance test conditions and results – cCrea, TContinuedT
Item Description
Blood samples Heparinized blood from healthy voluntary donors. Five blood pools were prepared to cover the test range.
Blood measurements
The measurements were performed by different operators
Calibration solution and gases
All calibration solutions and gases used for the tests are traceable to Primary Reference Standards.
Traceability certificates for the ABL800 FLEX calibration solutions and gases are found at the end of chapter 7: TSolutionsT.
Experimental conditions
Ambient temperature: 22-25 °C
Relative humidity: 30-50 %.
S250 mode:
cCrea (μmol/L) S B0B S BDB S BABLB S BXB
60 1.0 0.9 2.8 3.9
250 3.6 3.8 11 15
500 7.1 7.5 21 29
1000 14 15 43 59
1500 21 23 64 88
C125 mode:
cCrea (μmol/L) S B0B S BDB S BABLB S BXB
60 1.7 1.5 3.0 4.7
250 7.0 6.2 11 18
500 14 13 23 36
1000 28 25 45 73
1500 42 38 68 109
Imprecision T(continued)
5. Performance characteristics ABL800 FLEX reference manual
5-92
Interference tests
The following interference results are found for the pH and blood gas electrodes:
Substance Test conc. Interference on pOB2B electrode
Halothane 3 % 5 % increased sensitivity
Intralipid* (20 % solution) in a concentration less than 4 % (the final Intralipid level being 0.8 %) will not interfere on pH measurements.
If intralipid is found in higher concentrations it will interfere on the measurements, see table below: *Intralipid (Kabivitrum Inc., California and Stockholm) is a brand name for an aqueous suspension of lipid droplets that is sterile and suitable for intravenous feeding of patients
Final level of intralipid Interference on pH measurements
0.8 % 0 %
1.2 % 1.4 %
1.6 % 2.6 %
The following interference results are found for the electrolyte electrodes:
Interference on…
Substance Test conc. TcTKP
+P
(4 mmol/L level)
TcTNa P
+ P
(150 mmol/L level)
TcTCa P
2+P
(1.25 mmol/L level)
TcTCl P
−P
(110 mmol/L level)
LiP
+P 4 mmol/L 0 0 0
KP
+P 12 mmol/L −1 −0.01
NaP
+P 100-180
mmol/L 0.1 to −0.1
NHB4 PB
+P 1 mmol/L 0 0
CaP
2+P 5 mmol/L 0
Mg P
2+P 5 mmol/L 0 0 0.05
BrP
−P
10 mmol/L 41
F P
−P
1 mmol/L 0
IP
−P
3.0 mmol/L 30-90
ClOB4 PB
–P 1.5 mmol/L 8-30
HCOB3 PB
–P 25-50 mmol/L 0.1 mmol/L ClP
−P
per mmol/L HCOB3PB
–P
Lactate 10 mmol/L 0
Continued on next page
pH/blood gas
Electrolytes
ABL800 FLEX reference manual 5. Performance characteristics
5-93
Interference tests, TContinuedT
Interference on…
Substance Test conc. TcTKP
+P
(4 mmol/L level)
TcTNa P
+ P
(150 mmol/L level)
TcTCa P
2+P
(1.25 mmol/L level)
TcTCl P
−P
(110 mmol/L level)
Acetyl-salicylic acid
3.0 mmol/L
2
Salicylic acid
4.0 mmol/L
7
Ascorbic acid
1.0 mmol/L
0
pH £ 7.2 7.2 0 0 0.01 −1
pH ≥ 7.6 7.6 0 0 −0.01 1
Sulphide will give erroneously high TcTCl P
−P results.
The following interference results are found for the Glucose and Lactate electrodes:
Interference on …
Substance Test conc. (mmol/L)
Tc TGlucose (4.0 mmol/L level)
Tc TLactate (1.5 mmol/L level)
Acetoacetic acid 2 <⎜0.1⎜ <⎜0.1⎜
Acetylsalicylic acid 3 <⎜0.1⎜ <⎜0.1⎜
Ascorbic acid 2 <⎜0.1⎜ <⎜0.1⎜
Bilirubin (conjugated) 0.46 <⎜0.1⎜ <⎜0.1⎜
Bilirubin (unconjugated) 0.34 <⎜0.1⎜ <⎜0.1⎜
Chlorpromazine HCl 0.2 <⎜0.1⎜ <⎜0.1⎜
Citrate 50 −0.37 0.19
Creatinine 3 <⎜0.1⎜ <⎜0.1⎜
D-glucose 67 <⎜0.1⎜
Dopamine HCl 1.0 <⎜0.1⎜ <⎜0.1⎜
EDTA 3 <⎜0.1⎜ <⎜0.1⎜
Ethanol 79 <⎜0.1⎜ <⎜0.1⎜
Fluoride 50 −0.36 <⎜0.1⎜
Continued on next page
ElectrolytesT (continued)T
Glucose and Lactate
5. Performance characteristics ABL800 FLEX reference manual
5-94
Interference tests, TContinuedT
Interference on …
Substance Test conc. (mmol/L)
Tc TGlucose (4.0 mmol/L level)
Tc TLactate (1.5 mmol/L level)
Galactose 3.3 up to 1.88*
Glucosamine 2 up to 1.06*
Glycolic acid 1 <⎜0.1⎜ Interference
Heparin 8000 IU/dL <⎜0.1⎜ <⎜0.1⎜
Ibuprofen 2 <⎜0.1⎜ <⎜0.1⎜
Lactic acid 12 <⎜0.1⎜
Maltose 5 up to 0.4*
Mannose 1 up to 0.4*
Oxalate 90 −0.47 0.14
Paracetamol-4-acetamidophenol
2 <⎜0.1⎜ <⎜0.1⎜
Pralidoxime 0.045 <⎜0.1⎜ <⎜0.1⎜
Pyruvate 2 <⎜0.1⎜ <⎜0.1⎜
Salicylic acid 4 <⎜0.1⎜ <⎜0.1⎜
Thiocyanic acid 24 Interference Interference
Urea 84 <⎜0.1⎜ <⎜0.1⎜
Uric acid 1.5 <⎜0.1⎜ <⎜0.1⎜
Xylose 1 up to 0.34*
* Values determined at cGlu = 0 mmol/L. Interference at cGlu 4.0 mmol/L is expected to be the same.
ΔTcTLactate % at :
Hematocrit % 5 mmol/L level 15 mmol/L level
0 0.7 % 0.7 %
45 0.0 % 0.0 %
60 −0.5 % −2.0 %
75 −2.2 % −5.0 %
Continued on next page
Glucose and LactateT (continued)T
ABL800 FLEX reference manual 5. Performance characteristics
5-95
Interference tests, TContinuedT
The substances against which the oximetry parameters (TcTtHb, sOB2 B, FOB2 BHb, FCOHb, FMetHb, FHHb, FHbF) and TcTtBil were tested for interference are given in the table below:
(SAT100 blood reference test sample: ctHb = 15 g/dL, sOB2 B= 100 %, FCOHb = 0.7 %, FMetHb = 0.5 %, ctBil = 0, pH = 7.4. Parameter sensitivity from the influence on the absorbance spectrum from various substances.)
Change on … Substance Test conc. TcTtHb
(g/dL) TsTO B2 B (%)
TF TOB2 BHb (%)
TF TCOHb (%)
TF TMetHb (%)
TF THHb (%)
TF THbFT (%)T
TcTtBil T (μTmol/LT) T
Intralipid 4 Vol % P
e)P −0.5 0.1 −1.3 0.5 0.9 −0.1 11 0
4P
b)P
Intralipid 2 Vol % P
f)P −0.4 0.1 −0.3 0.3 0.1 −0.1 11 7
2P
b)P
HbF P
a), c)P 20 %P
P −0.02 1.17 0.04 0.73 0.37 −1.14 0 −14
TSTHb 10 % 0 −1.0 0.9 −0.1 0.1 −0.9 Not tested pH 7.1
7.9 −0.5 −0.6
−0.5 0.6
−0.2 −0.5
−0.4 1.0
0.1 0.1
0.5 −0.6
−19 13
0 −5
Cardio Green P
c)P 5 mg/L −0.16 0.29 1.14 0.07 −0.93 −0.29 −5 −20
Evans Blue P
c)P 5 mg/L −0.04 0.14 0.28 −0.20 −0.20 0.14 −5 5
Betacarotene in plasma P
c)P
3.7 μmol/L 0.0 −0.02 0.03 −0.01 −0.04 0.02 0.1 −0.2
Patent Blue V P
c)P 10 mg/L −0.16 0.39 0.86 −0.47 0.00 −0.38 −21 38
Methylene BlueP
c)P 30 mg/L −0.7 −3.4 5.6 −3.0 −6.2 3.6 −37 −25
HiCN P
c)P 0.11
mmol/L 0.26 −1.5 −3.0 −0.5 0.5 1.5 24 47
MCHC P
c), d)P
newborn range 320 g/L 350 g/L
No interference −12 17
Sedimentation rate 100 arb. Units
≤ ± 0.5 No interference Not tested
Notes: a) If function "Correction for HbF" is not activated, the change is 0 for all parameters.
b) Plasma sample. c) Calculated value from mathematical superposition of measured pure interference
spectrum on measured reference spectrum. d)
TcTtBil = 400 μmol/L. e) Intralipid (20 % solution) at 4 Vol % gives final test level of 0.8 %. f) Intralipid (20 % solution) at 2 Vol % gives final test level of 0.4 %. There is no interference from fetal hemoglobin (HbF) when the analyzer applies HbF
correction. There is no interference from bilirubin (conjugated/unconjugated) up to 1000 μmol/L.
Continued on next page
Oximetry parameters
5. Performance characteristics ABL800 FLEX reference manual
5-96
Interference tests, TContinued
The process of HbF correction introduces additional noise compared to measure-ment on adult samples. The following tables list the extra contribution which must be added geometrically to the imprecision specifications for adult samples in order to obtain the imprecision specifications for fetal samples (also for adult samples if function “Correction for HbF levels less than 20 %” is activated).
S S Sfetal adult HbF= +2 2 ; geometrical addition of imprecision
where S BfetalB is the calculated fetal imprecision; SBadultB is the corresponding adult imprecision; SBHbFB is the extra contribution from HbF correction which is listed in the following tables.
HbF correction contribution to 10 g/dL SAT100 fetal sample:
S B0B S BDB S BABLB S BXB
sOB2 B % 0.15 0.20 0.19 0.31
FHHb % 0.14 0.19 0.19 0.30
FOB2 BHb % 0.01 0.01 0.01 0.01
FCOHb % 0.09 0.13 0.12 0.20
FMetHb % 0.05 0.06 0.06 0.10
HbF correction contribution to 15 g/dL SAT100 fetal sample:
S B0B S BDB S BABLB S BXB
sOB2 B % 0.09 0.12 0.29 0.33
FHHb % 0.09 0.11 0.28 0.32
FOB2 BHb % 0.00 0.00 0.01 0.01
FCOHb % 0.06 0.07 0.18 0.21
FMetHb % 0.03 0.04 0.09 0.11
HbF correction contribution to 20 g/dL SAT100 fetal sample:
S B0B S BDB S BABLB S BXB
sOB2 B % 0.09 0.12 0.20 0.25
FHHb % 0.09 0.11 0.19 0.25
FOB2 BHb % 0.00 0.00 0.01 0.01
FCOHb % 0.06 0.07 0.13 0.16
FMetHb % 0.03 0.04 0.06 0.08
Continued on next page
Contribution to imprecision specifications from HbF correction
ABL800 FLEX reference manual 5. Performance characteristics
5-97
Interference tests, TContinuedT
FHbF is sensitive to pH deviations from the nominal value of pH = 7.4. If pH is converted into TcTHP
+P (hydrogen ion concentration), the relationship between the
changes in TcTHP
+P and FHbF is linear as seen from the following equation:
ΔFHbF = −0.48 %/(nmol/L) × (TcTHP
+P − 40 nmol/L)
where pH = 7.4 corresponds to TcTHP
+P = 40 nmol/L.
EXAMPLE: pH = 7.25 corresponds to TcTHP
+P = 56 nmol/L. Then:
ΔFHbF = −0.48 × (56 − 40) = −7.7 %.
MCHC (Mean Corpuscular Hemoglobin Concentration) is used to estimate hematocrit, Hct, which is used in the TcTtBil measurement. MCHC is an average Hb concentration in the red blood cell (RBC). If the RBC volume decreases, MCHC increases. If an RBC has iron deficit, MCHC decreases.
Hct is determined from ctHb as follows:
Hct tHbMCHC
=c
A standard value of 332 g/L is assumed for MCHC which gives
Hct = ctHb × 0.0301 if the unit for ctHb is g/dL.
MCHC can, however, deviate from this standard value as illustrated in the following table (see the next page).
Erythrocytometric values given for “apparently healthy” white and black subjects of different ages are taken from: “Geigy Scientific Tables, Physical Chemistry, Composition of Blood, Hematology, Somametric Data”, CIBA-GEIGY, 1984; 3, 207.
Continued on next page
FHbF sensitivity for pH changesT T
TcTtBil sensitivity for MCHC variations
5. Performance characteristics ABL800 FLEX reference manual
5-98
Interference tests, TContinuedT
Subjects Age Hct mean
Hct 95 % range
MCHC mean, g/L
MCHC 95 % range,
g/L
Men Adults 0.47 0.39-0.55 340 310-370
Women Adults 0.42 0.36-0.48 330 300-360
Boys Newborn
1 month
3 months
6 months
9 months
1 year
2 years
4 years
8 years
14 years
0.59
0.50
0.45
0.46
0.45
0.41
0.40
0.37
0.41
0.41
0.53-0.65
0.44-0.56
0.39-0.52
0.39-0.51
0.39-0.52
0.37-0.45
0.36-0.47
0.30-0.44
0.37-0.45
0.36-0.46
330
320
330
300
280
290
300
280
290
300
320-340
310-330
320-340
290-310
270-300
280-300
280-310
270-290
280-300
290-310
Girls Newborn
1 month
3 months
6 months
9 months
1 year
2 years
4 years
8 years
14 years
0.58
0.49
0.44
0.44
0.43
0.43
0.43
0.43
0.40
0.40
0.51-0.65
0.42-0.56
0.39-0.51
0.39-0.50
0.37-0.50
0.37-0.49
0.36-0.50
0.36-0.51
0.36-0.46
0.36-0.47
340
320
330
320
300
300
300
280
280
290
330-350
310-330
320-340
310-330
290-310
290-310
290-310
270-290
270-290
280-300
If ΔMCHC is defined as ΔMCHC = 332 g/L − MCHC, then the contribution to the relative error on the TcTtBil measurement is as follows:
Δ Δcc
tBiltBil
Hct1 Hct
MCHCMCHC
= −−
×
A worst-case example, using 95 % confidence values:
A newborn girl with Hct = 0.58, MCHC = 350 g/L and ctBil = 400 μmol/L. ctHb may be derived as Hct × MCHC = 0.58 x 350 g/L = 20.3 g/dL (reference range is 18.0 − 21.0 g/dL).
Continued on next page
TcTtBil sensitivity for MCHC variationsT (continued)T
ABL800 FLEX reference manual 5. Performance characteristics
5-99
Interference tests, TContinuedT
Δcc
tBiltBil
0.581 0 350
= −−
×−
= +.
.58
18 0 071 And ΔTcTtBil = 0.071 x 400 = 28 μmol/L.
If the reference value for Hct is known, it is possible to correct the displayed TcTtBil value, using the following equation:
c c ctBil(corrected) tBil(displayed tHb(displayed)Hct(reference)
= ×− ×
−) .1 0 0301
1
ctHb is measured in g/dL.
TcTtBil is slightly sensitive to pH deviations from the nominal value of pH = 7.4.
The following table shows the changes in Δ TcTtBil compared to the value at pH = 7.4.
Sample Type TcTtHb g/dL
Nominal TcTtBil μmol/L
ΔTcTtBil (7.4→7.1) μmol/L
ΔTcTtBil (7.4→7.9) μmol/L
Adult/fetal plasma 0 0 3 0
Adult blood, sOB2 B = 100 % 15 0 0 −5
Fetal blood, sOB2 B = 100 % 15 0 −13 4
Adult/fetal plasma spiked with unconjugated bilirubin
0 400 −2 −1
Adult/fetal plasma spiked with conjugated bilirubin
0 400 9 −11
Adult blood spiked with un-conjugated bilirubin, sOB2 B = 100 %
15 400 10 −26
Fetal blood spiked with un-conjugated bilirubin, sOB2 B = 100 %
15 400 −4 −16
Adult blood spiked with conjugated bilirubin, sOB2B = 100 %
15 400 14 −35
Fetal blood spiked with conjugated bilirubin, sOB2B = 100 %
15 400 0 −26
UMaterials and methodsU:
The interference testing of the TcTCrea(ABL8x7) parameter follows the guidelines described in CLSI EP7-A[16].
The interference test was performed on serum with normal creatinine concentration in the range of 50 to 92 μmol/L. A large serum pool was collected from a healthy person not undergoing any medical treatment.
Continued on next page
TcTtBil sensitivity for MCHC variationsT (continued)T
TcTtBil sensitivity for pH changes
Creatinine
5. Performance characteristics ABL800 FLEX reference manual
5-100
Interference tests, TContinuedT
Stock solutions of the tested interference were prepared in concentrations 20 times the test concentration on the day of the analysis.
Prior to measurement, one serum sample was spiked with the interference stock solution (sample T) and another serum sample was spiked with the same volume of the diluent used to prepare the stock solution (sample C).
The test of interferences was then performed as relative measurements. The sample with the tested interference (T) was compared to the control sample (C) measured at the same time: CB1 BTB1 BCB2BTB2 BCB3 BTB3B. Each sample was measured three times. The aim of the test has been to show interference above 8 μmol/L.
The test was performed at a university hospital in Sweden.
The following substances have been tested on samples with TcTCrea in the range 50 to 92 µmol/L.
Substance Test concentration Interference (μmol/L)
Ascorbic acid 227 μmol/L < |8|
Acetoacetate 10 mmol/L < |8|
Acetone 10 mmol/L < |8|
Acetylsalicylic acid 3.3 mmol/L < |8|
Ammonia 1 mmol/L < |8|
Ampicillin 152 µmol/L < |8|
Beta-hydroxybutyrate 10 mmol/L < |8|
Bilirubin 400 μmol/L < |8|
Bleomycin sulfate 45 mg/L < |8|
Calcium chloride 3 mmol/L < |8|
Cephalexin 337 μmol/L < |8|
Cephalothin 759 μmol/L < |8|
Cephotaxime 671 μmol/L < |8|
Cefoxitin 1.6 mmol/L < |8|
Chlorpromazine 6.3 μmol/L < |8|
Creatine 200 μmol/L < |8|
Citrate 50 mmol/L < |8|
Cyclosporin 12 μmol/L < |8|
D(+)-Glucose 60 mmol/L < |8|
Dipyrone 50 mg/L < |8|
Continued on next page
Creatinine T(continued)T
ABL800 FLEX reference manual 5. Performance characteristics
5-101
Interference tests, TContinuedT
Substance Test concentration Interference (μmol/L)
Dobutamine 10 μg/L < |8|
Dopamin 5.9 μmol/L < |8|
Doxycycline 67.5 μmol/L < |8|
Ethanol 86.8 mmol/L < |8|
Fluoride 50 mmol/L < |8|
Gentisic acid 117 μmol/L < |8|
Glutamate sodium salt 2 g/L < |8|
Gluthation – oxidized 10 mg/L < |8|
Gluthation – reduced 10 mg/L < |8|
HCOB3 PB
–P 40 mmol/L < |8|
Heparin (Li-salt) 80000 U/L < |8|
Hepes 20 mmol/L < |8|
Hemoglobin 10 % hemolysis < |8|
Hemoglobin 3.5 % hemolysis < |8|
Hydroxyurea 100 μmol/L < |8|
Ibuprofen 2.4 mmol/L < |8|
Intralipid 5 % < |8|
Intralipid 2.5 % < |8|
L-DOPA 20 μmol/L < |8|
L(+)-Glucose 60 mmol/L < |8|
Lithium nitrate 3.2 mmol/L < |8|
L-Lactic acid 30 mmol/L < |8|
L-Prolin 250 μmol/L < |8|
Lidocaine hydrochloride 100 μmol/L < |8|
Lidocaine powder 98 % 100 μmol/L < |8|
6-Mercaptopurine (13.1 mmol/L)* < |8|
Magnesium nitrate 3 mmol/L < |8|
Methotrexate 2.0 mmol/L < |8|
*Saturated stock solution
Continued on next page
Creatinine T(continued)
5. Performance characteristics ABL800 FLEX reference manual
5-102
Interference tests, TContinuedT
Substance Test concentration Interference (μmol/L)
Methyldopa 71 μmol/L < |8|
Oxo-(2)-butyric acid 5 mmol/L < |8|
Paracetamol (acetaminophen) 1.7 mmol/L < |8|
pH 8.0 – < |8|
pH 6.0 – < |8|
Phenylbutazone 325 μmol/L < |8|
Pyruvate 3 mmol/L < |8|
Rifampicin 78.1 μmol/L < |8|
Sarcosine 1 μmol/L < |8|
Sodium hydrogen carbonate 40 mmol/L < |8|
Sodium hydrogen phosphate 2 mmol/L < |8|
Theopyllin 222 μmol/L < |8|
Theophyllin acetic acid 200 μmol/L < |8|
Thiocyanate 24 mmol/L < |8|
Urea 50 mmol/L < |8|
Uric acid 3 mmol/L < |8|
Creatinine T(continued)
ABL800 FLEX reference manual 5. Performance characteristics
5-103
References
1. Kristensen HB, Salomon A, Kokholm G. International pH scales and certification of pH.
2. Definition of pH scales, standard reference values, measurement of pH and related terminology (Recommendations 1994). Pure and Appl Chem 1985; 57, 3: 531-42.
3. Burnett RW, Covington AK, Maas AHJ, Müller-Plathe O Tet alT. J Clin Chem Clin Biochem 1989; 27: 403-08.
4. IFCC reference methods and materials for measurement pH, gases and electrolytes in blood. Scand J Clin Lab Invest 1993; 53, Suppl 214: 84-94.
5. Glucose. CLSI/NCCLS Publication RS1-A. Clinical and Laboratory Standards Institute, 940 West Valley Road, Suite 1400, Wayne, PA 19087, 1989.
6. Reference and selected procedures for the quantitative determination of hemoglobin in blood. Approved Standard (3rd edition), CLSI/NCCLS Publication H15-2A. Clinical and Laboratory Standards Institute, 940 West Valley Road, Suite 1400, Wayne, PA 19087, 2000.
7. Evelyn K, Malloy H. Microdetermination of oxyhemoglobin, methemoglobin and sulfhemoglobin in a single sample of blood. Biological Chem 1938; 126: 655-62.
8. Kristoffersen K. An improved method for the estimation of small quantities of alkali-resistant hemoglobin in blood. Scand J Clin Lab Invest 1961; 13: 402.
9. Quantitative measurement of fetal hemoglobin using the alkali denaturation method. Approved Guideline. CLSI/NCCLS Publication H13-A. Clinical and Laboratory Standards Institute, 940 West Valley Road, Suite 1400, Wayne, PA 19087, 1989; 9, 18.
10. Begmeyer. Methods of enzymatic analysis. 3rd ed., Verlag Chemie Deerfield Beach 1984; 6: 582-88.
11. VIM93: ISO, International Vocabulary of Basic and General Terms in Metrology, Geneva: International Organization for Standardization; 1993.
12. Kristensen H.B. Traceability to the primary reference standards at Radiometer. Copenhagen: Radiometer Medical ApS, 2004. Code 918-541.
13. CLSI Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guidelines, EP5-A, Vol. 19, No. 2.
14. CLSI Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guidelines, EP17-A, Vol. 24, No. 34.
15. CLSI Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline – Second Edition, EP9-A2, Vol. 22, No. 17.
16. CLSI approved guideline for interference testing in clinical chemistry, EP7-A, Vol. 22, No. 27.
List of references
5. Performance characteristics ABL800 FLEX reference manual
5-104
6-1
6. 6. Parameters
Overview
The measured, input and derived parameters are described in this chapter.
This chapter contains the following topics.
General information ......................................................................................... X6-2X
Measured parameters ....................................................................................... X6-5X
Input parameters............................................................................................... X6-14X
Derived parameters .......................................................................................... X6-17X
Units of derived parameters ............................................................................. X6-22X
List of equations............................................................................................... X6-28X
Oxyhemoglobin dissociation curve (ODC)...................................................... X6-44X
Conversion of units .......................................................................................... X6-49X
Default values .................................................................................................. X6-51X
Altitude correction ........................................................................................... X6-52X
References ........................................................................................................ X6-53X
Introduction
Contents
6. Parameters ABL800 FLEX reference manual
6-2
General information
The Deep Picture developed by Radiometer [1], (visit our website Hwww.deep-picture.comH), expands traditional pH and blood gas analysis by evaluating the capability of arterial blood to carry sufficient oxygen to tissues and to release it. It simplifies interpretation by dividing the process into three stages:
Stage Description
Oxygen uptake
Oxygen uptake in the lungs indicates whether the pulmonary gas exchange is efficient enough to oxygenate arterial blood.
The uptake of oxygen in the lungs can be described by parameters in combination, primarily the arterial oxygen tension (pOB2B(a)), fraction of OB2B in dry inspired air (FOB2B(I)) and shunt fraction of
perfused blood (QBsB
· /Q· BtB)
However, other parameters may also be used, such as the difference in alveolar air and arterial blood oxygen tension (pOB2B(A-a)).
Oxygen transport
Oxygen transport reveals whether arterial blood contains sufficient oxygen.
The oxygen concentration of arterial blood (TcTtOB2B(a)) also termed oxygen content is determined by the concentration of total hemoglobin (ctHb(a)), the fraction of oxygenated hemoglobin (FOB2BHb(a)) and the arterial oxygen tension (pOB2B(a)). Other parameters which should be known are the oxygen saturation (sOB2 B(a)) and the fractions of dyshemoglobins (FCOHb(a) and FMetHb(a)).
Oxygen release
Oxygen release describes the ability of arterial blood to release oxygen to the tissues.
The release of oxygen from capillaries to tissues is determined by the oxygen tension gradient between the two. This release of oxygen is also influenced by the hemoglobin-oxygen affinity, which is indicated by the oxygen tension at 50 % saturation, p50.
The symbols for the parameters are based on the principles described by Wandrup [2]. Each symbol consists of three parts, described below:T
1. TPropertyT A symbol in italics describing the quantity
Tp T for pressure TcT for concentration TF T for fraction TV T for volume etc.
2. Component An abbreviation of the component name
OB2 B for oxygen COB2 B for carbon dioxide COHb for carboxyhemoglobin, etc.
Continued on next page
The Deep Picture
Symbols
ABL800 FLEX reference manual 6. Parameters
General information, TContinuedT
3. (System) Specification of the system B for blood P for plasma a for arterial blood v– for mixed venous blood A for alveolar air TT T for patient temperature
TExample:
T TpOB2B(a)
The parameters are
The following rang
TRangeT
Measuring Theanacorr"Int(VI
Reportable Is uCan
Reference "ReshoRef‘heapersvarystan
Refsectmar
Whlistea nucon
Critical limits are uchapter 3 T: Setup P
Symbols T(continued)
Ranges and limits
property
listed by s
es are used
measuringlyzer is phyesponds toernational vM).
ser-defined be selecte
ference ranuld not be rerence ranglthy’ indivons afflicte significandardization
. 10 has beeion. In somked by thei
en possibled. Referencmber of facdition.
ser-definedrograms (se
component
ymbol in t
:
range forsically cap the "rangeocabulary
; is equal td for all m
ges are valegarded ases should biduals ofted with disetly due to m" [10].
n the soure cases ther reference
, the referee ranges mtors, such
and can bction Ana
system
6-3
hree groups: measured, input and derived.
TDescription T
a parameter is the range within which the able of measuring. The measuring range of indication" as defined in the of basic and general terms in metrology"
o or narrower than the measuring range. easured and derived parameters.
uable guidelines for the clinician, but they absolute indicators of health and disease. e used with caution since values for
n overlap significantly with values for ase. In addition, laboratory values may ethodological differences and mode of
ce for the reference ranges given in this values are taken from other sources number.
nce ranges for arterial blood have been ust be used with caution as they depend on
as sex, age and normal physiological
e entered into the analyzer software – see lysis Setup) Tin the TOperator’s ManualT.
Continued on next page
6. Parameters ABL800 FLEX reference manual
6-4
General information, TContinuedTT
Derived parameters are calculated according to the equations stated.
If… Then… the required measured or input values are unknown
default values are used, unless a measured parameter does not have a value or is outside the measuring range.
all values are known the derived parameter is designated Tcalculated T and a ‘c’ is added to the result.
a default value is used the derived parameter is designated Testimated T and an ‘e’ is added to the result.
If one or more default values have been used in the calculation, the result may deviate significantly from the true value. The deviation on “estimated” oxygen status parameters may become particularly significant if default values are used instead of measured blood oximetry data. In some cases, however, the default value is not accepted as the input for the calculation. This is because the actual values of the missing parameter may deviate significantly from the default value, thus making the estimation clinically inappropriate. If sOB2B cannot be measured due to severe errors, it will be calculated.
Some of the listed parameters are measured, depending on the analyzer configuration. In these cases the equation given only applies if that parameter is TnotT directly measured by the analyzer.
Unless otherwise stated, a parameter will be calculated or estimated irrespective of the choice on the TPatient Identification Tscreen: "Arterial", "Capillary", "Venous", "Mixed venous", or "Not specified". Some parameters, however, are defined for arterial samples only; they will be calculated only for sample types entered as "Arterial" or "Capillary". The symbol for system (blood (B) or plasma (P)) is not stated in the equations unless it is important for the calculation.
The units given for each parameter refer to the units available on the analyzer for that parameter.
The default values are listed in TDefault Values Tat the end of this chapter.
Derived parameters
Measured parameters
Sample type
Units
Default values
ABL800 FLEX reference manual 6. Parameters
6-5
Measured parameters
The following is the used:
m = male
f = female
Reference range
for adult arterial blood
Reference
[10] Tietz NW, Logan NM. Reference ranges. In: Tietz NW, ed. Fundamentals of clinical chemistry. 3rd ed. Philadelphia: WB Saunders Company 1987: 944-75.
(unless otherwise specified)
Definition Indicates the acidity or alkalinity of the sample
Unit -
Measuring range 6.300-8.000
Reference range 7.35-7.45 (m, f)
Definition Concentration of hydrogen ions in blood
Unit nmol/L
Measuring range 10.0-501
Reference range 35.5-44.7 (m, f)
Is used both for blood and expired air samples.
Definition Partial pressure (or tension) of carbon dioxide in blood.
High and low pCOB2B values of arterial blood indicate blood hypercapnia and hypocapnia, respectively.
Unit mmHg; kPa; torr
Continued on next page
General information
pH
TcTHP
+P
pCOB2 B
6. Parameters ABL800 FLEX reference manual
6-6
Measured parameters, TContinuedT
Measuring range mmHg; torr: 5.0-250
kPa: 0.67-33.3
Reference range mmHg: 35-48 (m); 32-45 (f)
kPa: 4.67-6.40 (m); 4.27-6.00 (f)
Conversion of units p (kPa) = (mmHg)133322.0 p× = 0133322. × p (torr)
p (mmHg) = p (torr) = 7 500638. × p (kPa)
Is used for both blood and expired air samples.
Definition Partial pressure (or tension) of oxygen in blood.
High and low pOB2B values of arterial blood indicate blood hyperoxia and hypoxia, respectively.
Unit mmHg; kPa; torr
Measuring range mmHg; torr: 0.0-800
kPa: 0.00-107
Reference range mmHg: 83-108 (m, f)
kPa: 11.07-14.40 (m, f)
Conversion of units p (kPa) = (mmHg)133322.0 p× = 0133322. × p (torr)
p (mmHg) = p (torr) = 7 500638. × p (kPa)
Definition Ambient barometric pressure (Tp T(amb))
Unit mmHg; kPa; torr
Measuring range mmHg; torr: 450-800
kPa: 60.0-106.7
Reference range -
Conversion of units p (kPa) = (mmHg)133322.0 p× = 0133322. × p (torr)
p (mmHg) = p (torr) = 7 500638. × p (kPa)
Continued on next page
Tp TCOB2 B T(continued)
pOB2 B
TBaro
ABL800 FLEX reference manual 6. Parameters
6-7
Measured parameters, TContinuedT
Definition Concentration of total hemoglobin in blood.
Total hemoglobin includes all types of hemoglobin: deoxy-, oxy-, carboxy-, met-.
Unit g/dL; g/L; mmol/L
Measuring range g/dL: 0.00-27.7
g/L: 0.0-277
mmol/L: 0.00-17.2
Reference range g/dL: 13.5-17.5 (m); 12.0-16.0 (f)
g/L: 135-175 (m); 120-160 (f)
mmol/L: 8.4-10.9 (m); 7.4-9.9 (f)
Conversion of units TcTtHb (g/dL) = 1.61140 × ctHb (mmol/L);
ctHb (g/L) = 16.1140 × ctHb (mmol/L);
ctHb (mmol/L) = 0.62058 × ctHb (g/dL) = 0.062058 × ctHb (g/L)
Default value 9.3087 mmol/L, (15.0 g/dL or 150 g/L)
Can also be calculated.
Definition Oxygen saturation, the ratio between the concentrations of oxyhemoglobin and the hemoglobin minus the dyshemoglobins
Unit %; fraction
Measuring range %: 0.0-100.0
Fraction: 0.000-1.000
Reference range %: 95-99 (m, f)
Fraction: 0.95-0.99 (m, f)
Continued on next page
ctHb
sOB2 B
6. Parameters ABL800 FLEX reference manual
6-8
Measured parameters, TContinuedT
Reference Siggaard-Andersen O, Wimberley PD, Fogh-Andersen N, Gøthgen IH. Arterial oxygen status determined with routine pH/blood gas equipment and multi-wavelength hemoximetry: reference values, precision and accuracy. Scand J Clin Lab Invest 1990; 50, Suppl 203: 57-66. Available as AS106.
Equation The ODC is determined as described in equation for TOxyhemoglobin Dissociation CurveT (points I and III).
( )s
F FF F
OS 1 MetHb COHb
1- COHb - MetHb2 =× − −
where
S = ODC(P,A,TT T)
P p F
s F F= +
×× − −
pO O COHbO COHb MetHb2
2
2 ( )1
A = a
TT T = 37.0 P
oPC
Can also be calculated.
Definition Fraction of oxyhemoglobin in total hemoglobin in blood
Unit %; fraction
Measuring range %: 0.0-100.0
Fraction: 0.000-1.000
Reference range %: 94-98 (m, f)
Fraction: 0.94-0.98 (m, f)
Equation ( )F s F FO Hb O 1 COHb MetHb2 2= × − −
If TsTOB2 B is not measured, it will be calculated.
If dyshemoglobins (FCOHb, FMetHb) are not known, they are set to the default values.
Continued on next page
TsTOB2 B T(continued)
FOB2 BHb
ABL800 FLEX reference manual 6. Parameters
6-9
Measured parameters, TContinuedT
Definition Fraction of carboxyhemoglobin in total hemoglobin in blood
Unit %; fraction
Measuring range %: 0.0-100.0
Fraction: 0.000-1.000
Reference range %: 0.5-1.5 (m, f)
Fraction: 0.005-0.015 (m, f)
Default value 0.004 (0.4 %)
Definition Fraction of methemoglobin in total hemoglobin in blood
Unit %; fraction
Measuring range %: 0.0-100.0
Fraction: 0.000-1.000
Reference range %: 0.0-1.5 (m, f)
Fraction: 0.000-0.015 (m, f)
Default value 0.004 (0.4 %)
Can also be calculated.
Definition Fraction of deoxyhemoglobin in total hemoglobin in blood
Deoxyhemoglobin is the part of total hemoglobin which can bind oxygen forming oxyhemoglobin. It is also termed reduced hemoglobin, RHb.
Unit %; fraction
Measuring range %: 0.0-100.0
Fraction: 0.000-1.000
Continued on next page
FCOHb
FMetHb
FHHb
6. Parameters ABL800 FLEX reference manual
6-10
Measured parameters, TContinuedT
Equation FHHb = 1– sOB2 B × (1−FCOHb−FMetHb)−FCOHb− −FMetHb
If TsTOB2 B is not measured, it will be calculated from equation 39.
If dyshemoglobins (FCOHb, FMetHb) are not known, they are set to the default values.
Definition Fraction of fetal hemoglobin in total hemoglobin in blood
Unit %; fraction
Measuring range %: 0-100
Fraction: 0.00-1.00
Reference range (neonates)
%: ≈80 (m, f)
Fraction: ≈0.80 (m, f)
Definition Concentration of potassium ions in plasma
Unit mmol/L; meq/L
Measuring range mmol/L; meq/L: 0.5-25.0
Reference range m, f: 3.4-4.5 mmol/L
Conversion of units mmol/L = meq/L
Definition Concentration of sodium ions in plasma
Unit mmol/L; meq/L
Measuring range mmol/L; meq/L: 7-350
Reference range m, f; 136-146 mmol/L
Conversion of units mmol/L = meq/L
Continued on next page
TFTHHb T(continued)
FHbF
TcTKP
+P
TcTNa P
+P
ABL800 FLEX reference manual 6. Parameters
6-11
Measured parameters, TContinuedT
Definition Concentration of calcium ions in plasma
Unit mmol/L; meq/L; mg/dL
Measuring range mmol/L: 0.20-9.99 meq/L: 0.40-19.98 mg/dL: 0.80-40.04
Reference range m, f: 1.15-1.29 mmol/L; 2.30-2.58 meq/L
Conversion of units meq/L = 2 mmol/L
mg/dL = 4.008 mmol/L
Reference Siggaard-Andersen O, Thode J, Wandrup JH. The concentration of free calcium ions in the blood plasma ionized calcium. In: Siggaard-Andersen O, ed. Proceedings of the IFCC expert panel on pH and blood gases held at Herlev Hospital 1980, Copenhagen: Radiometer Medical A/S, 1981: 163-90. Available as AS79.
Definition Concentration of chloride ions in plasma
Unit mmol/L; meq/L
Measuring range mmol/L; meq/L: 7-350
Reference range 98-106 mmol/L (m, f)
Conversion of units mmol/L = meq/L
Definition Concentration of D-glucose in plasma
Unit mmol/L; mg/dL
Measuring range mmol/L: 0.0-60 mg/dL: 0-1081
Reference range m, f: 3.89-5.83 mmol/L; 70-105 mg/dL
Conversion of units TcTGlucose (mg/dL) = 18.016 × TcTGlucose (mmol/L)
cGlucose (mmol/L) = 0.055506 × TcTGlucose (mg/dL)
Continued on next page
TcTCa P
2+P
TcTCl-
TcTGlu
6. Parameters ABL800 FLEX reference manual
6-12
Measured parameters, TContinuedT
Definition Concentration of L-lactate in plasma
Unit mmol/L; meq/L; mg/dL
Measuring range mmol/L: 0.0-30 meq/L: 0.0-30 mg/dL: 0-270
Reference range m, f: 0.5-1.6 mmol/L; 4.5-14.4 mg/dL
Conversion of units cLactate (mg/dL) = 9.008 × TcTLactate (mmol/L)
cLactate (mmol/L) = 0.11101 × TcTLactate (mg/dL)
(conversion based on the molecular weight of lactic acid)
Definition Concentration of total bilirubin in plasma
Total bilirubin includes its two forms: conjugated and unconjugated.
Unit μmol/L; mg/dL; mg/L
Measuring range μmol/L: 0-1000 mg/dL: 0.0-58.5 mg/L: 0-585
Reference range See the table on the next page.
Conversion of units TcTtBil (μmol/L) = 17.1 × ctBil (mg/dL)
ctBil (μmol/L) = 1.71 × ctBil (mg/L)
ctBil (mg/dL) = 0.0585 × ctBil (μmol/L)
ctBil (mg/L) = 0.585 × ctBil (μmol/L)
Continued on next page
TcTLac
TcTtBil
ABL800 FLEX reference manual 6. Parameters
6-13
Measured parameters, TContinuedT
The reference ranges are as follows:
Age TcTtBil
≤24 hrs, premature 17-137 μmol/L 1.0-8.0 mg/dL 10-80 mg/L
≤24 hrs, full-term 34-103 μmol/L 2.0-6.0 mg/dL 20-60 mg/L
≤48 hrs, premature 103-205 μmol/L 6-12 mg/dL 60-120 mg/L
≤48 hrs 103-171 μmol/L 6-10 mg/dL 60-100 mg/L
3-5 days, premature 171-239 μmol/L 10-14 mg/dL 100-140 mg/L
3-5 days, full-term 68-137 μmol/L 4-8 mg/dL 40-80 mg/L
>1 month 3.4-17 μmol/L 0.2-1.0 mg/dL
2-10 mg/L
Definition Concentration of creatinine in blood
Unit μmol/L; mg/dL
Measuring range μmol/L: 10-1800 mg/dL: 0.11-20.4
Reference range m: 53-106 μmol/L, 0.6-1.2 mg/dL f: 44-97 μmol/L, 0.5-1.1 mg/dL
Conversion of units TcTCrea (μmol/L) = 88.40 × cCrea (mg/dL)
TcTtBil T(continued)T
TcTCrea
6. Parameters ABL800 FLEX reference manual
6-14
Input parameters
Input parameters are the parameters keyed in by the operator on the TPatient IdentificationT screen or transferred from an interfaced database.
All input parameters are given in this section.
Definition Patient temperature
Unit P
oPC; P
oPF
Measuring range P
oPC: 15.0-45.0
P
oPF: 59-113
Conversion TT T °F = 32
59 o +CT ; TT T°C = )32 (
95 o −FT
Definition Fraction of oxygen in dry inspired air
Unit %; fraction
Input range %: 0-100 fraction: 0.000-1.000
Reference range 35.5-44.7 (m, f)
Is used in the ABL800/05 FLEX analyzers.
Definition Concentration of total hemoglobin in blood
Input range /Unit g/dL: 0.0-33.0 g/L: 0-330 mmol/L: 0.0-20.5
Conversion TcTtHb (g/dL) = 1.61140 × ctHb (mmol/L);
ctHb (g/L) = 16.1140 × ctHb (mmol/L);
ctHb (mmol/L) = 0.62058 × ctHb (g/dL) = 0.062058 × ctHb (g/L)
Definition Respiratory quotient, ratio between the COB2B production and the OB2B consumption
Input range 0.00-2.00
Continued on next page
Definition
TT
FOB2B(I)
ctHb
RQ
ABL800 FLEX reference manual 6. Parameters
6-15
Input parameters, TContinuedT
Definition Oxygen tension of mixed venous blood
Input range/Unit mmHg; torr: 0.0-750.0 kPa: 0.00-100
Conversion Tp T(kPa) = 0.133322 × Tp T(mmHg) Tp T(mmHg) = (kPa)500638.7 p×
Definition Oxygen saturation of mixed venous blood
Input range/Unit %: 0.0-100.0 fraction: 0.000-1.000
Definition Cardiac output; volume of blood delivered from the left ventricle into the aorta per unit of time
Also termed CO or C.O.
Input range/Unit 0.0-1000.0 L/min
Definition Oxygen consumption; total amount of oxygen utilized by the whole organism per unit of time
Input range/Unit mL/min: 0-xxxx mmol/min: 0.0-xxx.x
Conversion (mmol/L)min = (mL/dLmin)/22.41
Definition Volume of carbon monoxide added to the patient for measurement and calculation of V(B) [5]
Input range/Unit 0.0-1000.0 mL
Continued on next page
pOB2B(v– )
sOB2B(v– )
Q· Bt B
V·OB2B
VCO
6. Parameters ABL800 FLEX reference manual
6-16
Input parameters, TContinuedT
Can also be a derived parameter.
Definition Partial pressure (or tension) of oxygen at half saturation (50 %) in blood at standard conditions: temperature = 37 P
oPC
pH = 7.40 pCOB2 B = 5.33 kPa FCOHb, FMetHb, FHbF set to 0
Tp T50(st) may, however, vary due to variations in 2,3-DPG concentration or to the presence of abnormal hemoglobins.
Input range/Unit mmHg; torr: 0.01-100.00 kPa: 0.001-13.332
Conversion Tp T(kPa) =0.133322 × Tp T(mmHg; torr) Tp T(mmHg; torr) = (kPa)500638.7 p×
Definition The fraction of COHb measured before the CO-injection
Input range/Unit %: 0.0-100.0 fraction: 0.000-1.000
Definition The fraction of COHb measured after the CO-injection
Input range/Unit %: 0.0 - 100.0 fraction: 0.000 - 1.000
p50(st)
FCOHb(1)
FCOHb(2)
ABL800 FLEX reference manual 6. Parameters
6-17
Derived parameters
In the TType Tcolumn the following symbols are used:
• ms for measured parameters
• dv for derived parameters
Symbol Definition Type Eq.
pH( TT T) pH of blood at patient temperature. dv T1 T
cHP
+P(TT T) Concentration of hydrogen ions in blood at
patient temperature. dv T2 T
pCOB2 B(TT T) Partial pressure (or tension) of carbon dioxide at patient temperature.
dv T3 T
cHCOB3 PB
–P(P) Concentration of hydrogen carbonate in plasma
(also termed actual bicarbonate). dv T4 T
TcTBase(B)
or ABE
Actual Base Excess, the concentration of titrable base when the blood is titrated with a strong base or acid to a plasma pH of 7.40, at pCOB2B of 5.33 kPa (40 mmHg) and 37 P
oPC, at the actual
oxygen saturation [4,5].
Positive values (base excess) indicate a relative deficit of non-carbonic acids; negative values (base deficit) indicate a relative excess of non-carbonic acids.
dv T5 T
TcTBase(B,ox) TcTBase(B) of fully oxygenated blood. dv T6 T
TcTBase(Ecf)
or SBE
Standard Base Excess, an in vivo expression of base excess [5, 6]. It refers to a model of the extracellular fluid (one part of blood is diluted by two parts of its own plasma) and is calculated using a standard value for the hemoglobin concentration of the total extracellular fluid.
dv T7 T
TcTBase(Ecf,ox) TcTBase(Ecf) of fully oxygenated blood. dv T8 T
cHCOB3 PB
–P(P,st) Standard Bicarbonate, the concentration of
hydrogen carbonate in the plasma from blood which is equilibrated with a gas mixture with pCOB2B = 5.33 kPa (40 mmHg) and pOB2B ≥ 13.33 kPa (100 mmHg) at 37 P
oPC [4,5].
dv T9 T
TcTtCOB2 B(P) Concentration of total carbon dioxide, (free COB2B + bound COB2B) in plasma.
dv T10T
Continued on next page
General information
Acid-base derived parameters
6. Parameters ABL800 FLEX reference manual
6-18
Derived parameters, TContinuedT
Symbol Definition Type Eq.
TcTtCOB2 B(B) Concentration of total carbon dioxide in whole blood (also termed COB2B content).
Calculated based on the total COB2B concentrations in the two phases: plasma and erythrocyte fluid [5].
dv T11T
pH(st) Standard pH (or eucapnic pH), defined as the pH of plasma of blood equilibrated to pCOB2B = 5.33 kPa (40 mmHg).
By ensuring the normal value of pCOB2B, the respiratory influence from pH is removed, and pH(P,st) therefore reflects the metabolic status of the blood plasma.
dv T12T
TV TCOB2 B/ TV T(dry air)
The volume fraction of carbon dioxide in dry air. dv T51T
Symbol Definition Type Eq.
FHHb Fraction of deoxyhemoglobin in total hemoglobin in blood.
Deoxyhemoglobin is the part of total hemoglobin which can bind oxygen forming oxyhemoglobin. It is also termed reduced hemoglobin, RHb.
ms/dv T41T
FOB2 BHb Fraction of oxyhemoglobin in total hemoglobin in blood.
ms/dv T40T
sOB2 B Oxygen saturation, the ratio between the concentrations of oxyhemoglobin and the hemoglobin minus the dyshemoglobins.
ms/dv T39T
Hct Hematocrit, the ratio between the volume of erythrocytes and the volume of whole blood.
dv T13T
Continued on next page
Acid-base derived parameters T(continued)
Oximetry derived parameters
ABL800 FLEX reference manual 6. Parameters
6-19
Derived parameters, TContinuedT
Symbol Definition Type Eq.
pOB2B(TT T) Partial pressure (or tension) of oxygen at patient temperature.
dv T14T
pOB2B(A) Partial pressure (or tension) of oxygen in alveolar air.
dv T15T
pOB2 B(A, TT T) Partial pressure (or tension) of oxygen in alveolar air at patient temperature.
dv T16T
pOB2B(a)/ TF TOB2 B(I)
Oxygen tension ratio of arterial blood and the fraction of oxygen in dry inspired air
dv T17T
pOB2B(a, TT T)/ TF TOB2 B(I)
Oxygen tension ratio of arterial blood at patient temperature and the fraction of of oxygen in dry inspired air
dv T18T
Tp T50 Partial pressure (or tension) of oxygen at half saturation (50 %) in blood.
High and low values indicate decreased and increased affinity of oxygen to hemoglobin, respectively.
dv T19T
Tp T50(TT T) Partial pressure (or tension) of oxygen at half saturation (50 %) in blood at patient temperature.
dv T20T
Tp T50(st) Partial pressure (or tension) of oxygen at half saturation (50 %) in blood at standard conditions: temperature = 37 P
oPC
pH = 7.40 pCOB2 B = 5.33 kPa FCOHb, FMetHb, FHbF set to 0
Tp T50(st) may however vary due to variations in 2,3-DPG concentration or to the presence of abnormal hemoglobins.
dv/in T21T
Tp TOB2B(A−a) Difference in the partial pressure (or tension) of oxygen in alveolar air and arterial blood.
Indicates the efficacy of the oxygenation process in the lungs.
dv T22T
pOB2B(A−a, TT T) Difference in the partial pressure (or tension) of oxygen in alveolar air and arterial blood at patient temperature.
dv T23T
Continued on next page
Oxygen derived parameters
6. Parameters ABL800 FLEX reference manual
6-20
Derived parameters, TContinuedT
Symbol Definition Type Eq.
pOB2B(a/A) Ratio of the partial pressure (or tension) of oxygen in arterial blood and alveolar air.
Indicates the efficacy of the oxygenation process in the lungs.
dv T24T
pOB2B(a/A, TT T) Ratio of the partial pressure (or tension) of oxygen in arterial blood and alveolar air at patient temperature.
dv T25T
pOB2B(x) or TpTBx B Oxygen extraction tension of arterial blood.
Reflects the integrated effects of changes in the arterial pOB2B(a), ctOB2B and Tp T50 on the ability of arterial blood to release OB2B to the tissues [8].
dv T26T
pOB2B(x, TT T) or Tp TBxB(TT T)
Oxygen extraction tension of arterial blood at patient temperature.
dv
ctOB2 B(B) Total oxygen concentration of blood.
Also termed OB2B content.
dv T27T
ctOB2 B(a−v– ) Oxygen concentration difference between arterial and mixed venous blood.
dv T28T
BOB2 B Hemoglobin oxygen capacity; the maximum concentration of oxygen bound to hemoglobin in blood saturated, so that all deoxyhemoglobin is converted to oxyhemoglobin.
dv T29T
ctOB2 B(x) Extractable oxygen concentration of arterial blood.
Defined as the amount of OB2B which can be extracted per liter of arterial blood at an oxygen tension of 5.0 kPa (38 mmHg), maintaining constant pH and pCOB2B [8].
dv T30T
D· OB2B
Oxygen delivery; the total amount of oxygen delivered to the whole organism per unit of time.
dv T31T
Q· BtB Cardiac output; volume of blood delivered from the left ventricle into the aorta per unit of time.
Also termed CO or C.O.
dv/in T32T
V·OB2B
Oxygen consumption; total amount of oxygen utilized by the whole organism per unit of time.
dv/in T33T
TF TOB2 B(I) Fraction of oxygen in dry inspired air. in
Continued on next page
Oxygen derived parameters T(continued)
ABL800 FLEX reference manual 6. Parameters
6-21
Derived parameters, TContinuedT
Symbol Definition Type Eq.
TF TShunt Relative physiological shunt or concentration-based shunt [5,8,9]. • Calculated from the pulmonary shunt
equation: &
&
( )
QQ tO (a v)
tO A tO (a)
s
t 2
2 2
=+
−−
1
1 cc c
if both arterial and mixed venous blood samples are used.
• May be estimated from one arterial sample by assuming a constant difference in the concentrations of total oxygen in arterial and mixed venous blood:
ctO (a v) 2.3 mmol / L (5.1 mL / dL)2 − =
dv T34T
TF TShunt (TT T) TF TShunt at patient temperature. dv T35T
RI Respiratory Index; ratio between the oxygen tension difference of alveolar air and arterial blood and the oxygen tension of arterial blood.
dv T36T
RI(TT T) Respiratory Index; ratio between the oxygen tension difference of alveolar air and arterial blood and the oxygen tension of arterial blood at patient temperature.
dv T37T
TV TOB2 B/ TV T(dry air)
Volume fraction of oxygen in dry air. dv T52T
QBx B Cardiac oxygen compensation factor of arterial blood defined as the factor by which the cardiac output should increase to allow release of 2.3 mmol/L (5.1 mL/dL) oxygen at a mixed venous pOB2B of 5.0 kPa (38 mmHg) [5,8].
dv T38T
TV T(B) Volume of blood, calculated when FCOHb and V(CO) values are keyed in [5].
dv T42T
GFR, if AA Glomerular filtration rate, if African American dv T53T
GFR, if non AA
Glomerular filtration rate, if non African American
dv T54T
Oxygen derived parameters T(continued)
6. Parameters ABL800 FLEX reference manual
6-22
Units of derived parameters
Derived parameters are calculated or estimated on the basis of measured and keyed-in data. Calculations are made using equations programmed into the analyzer. The accuracy of the calculations depends on the input parameters keyed into the analyzer’s computer.
If the calculation of a parameter requires input from the operator, but this input is not forthcoming, the analyzer will use certain default values (refer to the section TDefault ValuesT in this chapter).
Not all input parameters are stored as defaults. In these instances the dependent derived parameter will not be reported if the relevant input parameter(s) is/areT not T entered.
If the default values are used in the calculation of a parameter, then a parameter is considered Testimated T("e") rather than Tcalculated T ("c").
The ABL83x FLEX analyzer corresponds to ABL82x FLEX analyzer, but it can measure TcTtBil and FHbF.
Number of characters before "." is variable; number of decimals is constant.
Symbol Unit Num-ber of deci-mals
ABL805
FLEX
ABL 81x/ 820
FLEX
ABL825/827/83x FLEX
Input parameter
Sample type
pH( TT T) - .xxx c c c TT T
TcTHP
+P(TT T) nmol/L .x c c c TT T
pCOB2 B(TT T) mmHg; torr
.x c c c TT T
kPa .xx
cHCOB3 PB
–P(P) mmol/L .x c c c
TcTBase(B) mmol/L Range c c c ctHb
±30.0 e c c
TcTBase(B,ox) mmol/L .x e c c ctHb
e c c
TcTBase(Ecf) mmol/L Range ±30.0
c c c
TcTBase(Ecf,ox) mmol/L .x e c c
cHCOB3 PB
–P(P,st) mmol/L .x c c c ctHb
e c c
Continued on next page
Calculated versus estimated parameters
Acid-base parameters
ABL800 FLEX reference manual 6. Parameters
6-23
Units of derived parameters, TContinuedT
The ABL83x FLEX analyzer corresponds to ABL82x FLEX analyzer, but it can measure TcTtBil and FHbF.
Number of characters before "." is variable; number of decimals is constant.
Symbol Unit Num-ber of deci-mals
ABL805
FLEX
ABL 81x/ 820
FLEX
ABL825/827/83x
FLEX
Input parameter
Sample type
TcTtCOTB2 TB(P)
Vol %, mL/dL, mmol/L
.x c
c
c
TcTtCOTB2 TB(B)
Vol %, mL/dL, mmol/L
.x c c c ctHb
pH(st) - .xxx c c c
TV TCOB2 B/ TV T(dry % .x c c c
air) fraction .xxx
The ABL83X FLEX analyzer corresponds to ABL82X FLEX analyzer, but it can measure TcTtBil and FHbF).
Number of characters before "." is variable; number of decimals is constant.
Symbol Unit Num-ber of deci-mals
ABL805
FLEX
ABL 81x/ 820
FLEX
ABL825/827/83x
FLEX
Input parameter
Sample type
Hct % ctHb
fraction .xxx c c c
sOB2 B % .x
fraction .xxx e
TF TOB2 BHb % .x
fraction .xxx e e c
FHHb % .x
fraction .xxx e e c
Continued on next page
Acid-base parameters T(continued)
Oximetry parameters
6. Parameters ABL800 FLEX reference manual
6-24
Units of derived parameters, TContinuedT
The ABL83X analyzer corresponds to ABL82X analyzer, but it can measure TcTtBil and FHbF). Number of characters before "." is variable; number of decimals is constant.
Symbol Unit Num-ber of deci-mals
ABL805
FLEX
ABL 81x/ 820
FLEX
ABL825/827/83x
FLEX
Input parameter
Sample type
pOB2B(TT T) mmHg; torr
.x e e c TT T
kPa .xx
pOB2B(A) mmHg; torr
.x c c c TF TOB2 B(I)+RQ Arterial,
kPa .xx e e e capillary
pOB2B(A, TT T) mmHg; torr
.x c c c TF TOB2 B(I)+RQ+TT T Arterial, capillary
kPa .xx e e e
Tp T50 mmHg; torr
.xx e e e*
kPa .xx
Tp T50(TT T) mmHg; torr
.xx e e c* TT T
kPa .xx
Tp T50(st) mmHg; torr
.xx e e c*
kPa .xx
pOB2B(A−a) mmHg; torr
.x c c c TF TOB2 B(I)+ RQ Arterial,
kPa .xx e e e capillary
pOB2B(A−a, TT T) mmHg; torr
.x e e c TF TOB2 B(I)+RQ+TT T Arterial,
kPa .xx e e e capillary
pOB2B(a/A) % .x c c c TF TOB2 B(I)+RQ Arterial,
fraction .xxx e e e capillary
pOB2B(a/A, T T T) % .x c c c TF TOB2 B(I)+RQ+TT T Arterial,
fraction .xxx e e e capillary
pOB2B(a)/TF TOB2 B(I) mmHg; torr
.x c c c TF TOB2 B(I) Arterial,
kPa .xx capillary
Continued on next page
Oxygen parameters
ABL800 FLEX reference manual 6. Parameters
6-25
Units of derived parameters, TContinued
The ABL83x FLEX analyzer corresponds to ABL82x FLEX analyzer, but it can measure TcTtBil and FHbF.
Number of characters before "." is variable; number of decimals is constant.
Symbol Unit Num-ber of deci-mals
ABL 805
FLEX
ABL 81x/ 820
FLEX
ABL825/827/83x
FLEX
Input parameter
Sample type
pOB2B(a, TT T)/ % .x c c c TF TOB2 B(I)+TT T
TF TOB2 B(I) fraction .xx
pOB2B(x) mmHg; torr .x e e* c* ctHb+Tp T50(st) Arterial,
kPa .xx - e* c* capillary
pOB2B(x, TT T) mmHg; torr .x e e* c* ctHb+Tp T50(st)+TT T Arterial,
kPa .xx - e* c* capillary
ctOB2 B(B) Vol %, mL/dL, mmol/L
.x e
e
c
ctHb
ctOB2 B(a−v– ) Vol %, mL/dL, mmol/L
.x e e
c
ctHb Venous +
arterial
BOB2 B Vol %, mL/dL, mmol/L
.x e
e
c
ctHb
ctOB2 B(x) Vol %, mL/dL, mmol/L
.x e
e*
c*
ctHb P
P+Tp T50(st) P
P Arterial,
capillary
D· OB2B mL/min e e c QBt B
· Arterial,
mmol/min .x capillary
QBt B
· L/min .x e e c V· OB2B Venous + arterial
V· OB2B mL/min e e c QBt B
· Venous +
mmol/min .x arterial
TF TShunt % .x e e c* ctHb Venous
fraction .xxx + arterial
TF TShunt(TT T) % .x e e c* ctHb + TT T Venous +
fraction .xxx arterial
Continued on next page
Oxygen parametersT (continued)T
6. Parameters ABL800 FLEX reference manual
6-26
Units of derived parameters, TContinuedT
The ABL83x FLEX analyzer corresponds to ABL82x FLEX analyzer, but it can measure TcTtBil and FHbF.
Number of characters before "." is variable; number of decimals is constant.
Symbol Unit Num-ber of deci-mals
ABL 805
FLEX
ABL 81x/ 820
FLEX
ABL825/827/83x FLEX
Input parameter
Sample type
RI % c c c TF TOB2 B(I)+RQ Arterial,
fraction .xx e e e capillary
RI(TT T) % e e c TF TOB2 B(I)+RQ+T
T T
Arterial, capillary
fraction .xx e e e TT T
TV TOB2 B/ TV T(dry % .x c c c
air) fraction .xxx
QBx B - .x e e* c* ctHb P
1)
P+Tp T50(st) P
1)P
Arterial, capillary
e e* c*
TV T(B) L .x c c c ctHb+TV TCO+ FCOHb(1)+ FCOHb(2)
* If the sOB2B value for establishing the ODC is greater than 0.97, the calculation of the parameter is not performed unless the Tp T50(st) value is keyed in.
P
1) PIf not measured, e.g. ctHb (or derived by analyzer, e.g. Tp T50(st)).
Numerical format: number of characters before "." is variable; number of decimals is constant.
Symbol Unit Num-ber of deci-mals
ABL8x5/8x7 FLEX
Input parameter
Sample type
Anion Gap, KP
+P meq/L, mmol/L .x cP
2)P
Anion Gap meq/L, mmol/L .x cP
3)P
cCaP
2+P(7.4) meq/L, mg/dL, mmol/L .x cP
4)P
TmTOsm mmol/kg .x cP
5)P
2) If the analyzer includes K, Na and Cl electrodes.
3) If the analyzer includes Na and Cl electrodes.
4) If the analyzer includes Ca electrode.
5) If the analyzer includes Na and Glucose electrodes.
Continued on next page
Oxygen parametersT (continued)T
Electrolyte parameters
ABL800 FLEX reference manual 6. Parameters
6-27
Units of derived parameters, TContinuedT
Numerical format: number of characters before "." is variable; number of decimals is constant.
Symbol Unit Num-ber of deci-mals
ABL8x7 FLEX
Input parameter
Sample type
GFR, if AA µmol/L cP
6)P Crea, Sex,
Age Arterial
GFR, if non AA µmol/L cP
6)P Crea, Sex,
Age Arterial
6) For patients >18 years [24].
Metabolite parameters
6. Parameters ABL800 FLEX reference manual
6-28
List of equations
All definitions and equations are based on SI units. If "T" for patient temperature is not stated, the calculation is based on a temperature of 37.0 °C. The following SI units are used: concentration in mmol/L
temperature in °C
pressure in kPa
fractions (not %)
The following symbols are used in the equations:
log(x) = logB10B(x)
ln(x) = logBeB(x)
If ctHb is not measured or keyed in, the default value will be used.
If sOB2 B is not measured, it will be calculated from equation 39.
TEq. 1 T[13]:T
( )[ ][ ]pH( ) pH(37) pH(37)- -37T T= − + ×0 0146 0 0065 7 40. . .
TEq. 2 T: ( )c T TH ( )=10+ pH ( )9−
TEq. 3 T[4]:T
( )[ ]p T p TCO CO (37) 102 2- 37( ) .= × ×0 021
TEq. 4 T[5]:T
( )c pHCO (P) CO 103-
2pH - pKp= × ×0 23.
where ( )[ ]8.7pH
p 10+1log125.6pK −−=
cHCOB3PB
–P(P) includes ions of hydrogen carbonate, carbonate and carbamate in the
plasma.
Continued on next page
Units and symbols
pH( TT T)
TcTHP
+P( TT T)
Tp TCOB2 B(T)
TcTHCOB3 PB
–P(P)
ABL800 FLEX reference manual 6. Parameters
6-29
List of equations, TContinuedT
TEq. 5 T[4,14]:
c cBase(B) = 0.5 8a'-0.919a'
0.919 - 8a'a'
HCOa'
3-
× ⎛⎝⎜
⎞⎠⎟
+ × ⎛⎝⎜
⎞⎠⎟
− ×−0 5 4 24 47 5332
. . ( . )
where
Eq. Description
5.1 a'= 4.04 10 tHb-3× + × −4 25 10 4. c
5.2 ( )
cHCO 103-
pH(st)
( . ) . .533 0 23 5336.161
0.9524= × ×−⎡
⎣⎢
⎤
⎦⎥
5.3 pH(st) pH log 5.33
COpH(Hb) pH
log CO Hb log(7.5006 CO2 2 2
= +⎛⎝⎜
⎞⎠⎟ ×
−−
⎛⎝⎜
⎞⎠⎟
p p p( ) )
5.4 ( )pH(Hb) tHb +5.98- 10 tHb= × ×− −4 06 10 1922 0 16169. . .c c
5.5 ( )log CO (Hb) tHb+ 3.4046+ 2.12 102tHbp c c= − × ×− −17674 10 2 0.15158.
TEq. 6 T [4]:
c c c sBase(B,ox) Base(B) tHb O2= − × × −0 3062 1. ( )
TEq. 7 T [5]:
cBase(Ecf) = cBase(B) for ctHb = 3 mmol/L
TEq. 8 T:
cBase(Ecf,ox) = cBase(B,ox) for ctHb = 3 mmol/L
TEq. 9 T[4,14]:
( )cHCO P,st) 0.919 Z + Z a' Z-83- ( .= + × × ×24 47
where
Eq. Description
9.1 a'= 4.04 10 tHb-3× + × ×−4 25 10 4. c
9.2 ( )Z = × ×c c sBase(B)-0.3062 tHb 1- O2
TEq. 10 T[4,5]:
c p ctCO P CO HCO P2 2 3-( ) . ( )= × +0 23
Continued on next page
TcTBase(B)
TcTBase(B,ox)
TcTBase(Ecf)
TcTBase(Ecf,ox)
TcTHCOB3 PB
–P(P,st)
TcTtCOB2 B(P)
6. Parameters ABL800 FLEX reference manual
6-30
List of equations, TContinuedT
TEq. 11 T[5]: ( )[ ]
⎟⎠⎞
⎜⎝⎛ −×+
××××= −−
0.21tHb1)P(tCO
10+1tHbCO10286.9)B(tCO
2
KpH2
32
EryEry
cc
cpc p
where
Eq. Description
11.1 ( ) ( )pH pH -7.40 OEry 2= + × + × −7 19 0 77 0 035 1. . . s
11.2 ( )[ ]pK log 1+10ErypH OEry 2= − − − ×6125 7 84 0 06. . . s
TEq. 12T [14]:
pH(st): see equations 5.3-5.5.
Eq. 13 [15]:
Hct = 0.0485×TcTtHb + 8.3×10P
-3P
Hct cannot be calculated on the basis of a default ctHb value.
TEq. 14 T[16,17]:
The standard Oxygen Dissociation Curve (ODC) is used (i.e. p50(st) = 3.578 kPa) at actual values of pH, pCOB2B, FCOHb, FMetHb, FHbF (see equations 46-47 in the section TOxyhemoglobin Dissociation Curve Tfurther in this chapter).
pOB2B(TT T) is calculated by a numerical method using:
( )t ( ) tHb 1- COHb - MetHb O O Oi 2,i 2 2,iT c F F s T T p T= × × + ×( ) ( ) ( )α
where
Eq. Description See…
14.1 S = ODC(P,A,TT T) Eq. 47
14.2 ( )s T
F FO ( ) =
S 1- MetHb COHb1- COHb - MetHb2,i
× −
F F
Eq. 46.12
14.3
( )
p TF
s T F F
O ( ) = PCOHb
O ( ) COHb MetHb2,i
2
11
,i
+× − −
Eq. 46.10
Continued on next page
TcTtCOB2 B(B)
pH(st)
Hct
Tp TOB2 B( TT T)
ABL800 FLEX reference manual 6. Parameters
6-31
List of equations, TContinuedT
Eq. Description See…
14.4 ( ) ( )[ ]242 37.0-101.237.0-1015.132 1083.9O TTe ××+×−− −−
×=α
14.5 P is the variable during iteration.
14.6 ( )37.0-pH1.04-ac=A TT
××∂
∂
14.7 TT T= patient temperature in P
oPC (keyed-in).
14.8 ( )∂∂
pH( )
pH
When t then O O ( )i 2,i 2
TT p T p T
= − × − × × −
= =
− −146 10 6 5 10 37 7 40
37 0
2 3. . ( ) .
( ) ( . ), ( )ti
Eq. 15T T[5]:
( )( )[ ]
p F p
p F
O (A) O (I) (amb) -6.275
CO RQ O (I) RQ 1
2 2
21
21
= ×
− × − × −− −
If FOB2B(I) and RQ are not keyed in, they are set to the default values.
The calculation requires entering the sample type as “Arterial” or “Capillary”.
Eq. 16T T[4,5,18]:
[ ]( )[ ]1RQ(I)ORQ)(CO
)(OH-(amb)(I)O)A,(O1-
21
2
222
−×−×−
×=− FTp
TppFTp
( ) ( )[ ]p TT T
H O( ) = 6.275 102
×
× × − − × × −− −2 36 10 37 0 9 6 10 37 02 5 2. . . .
If FOB2B(I) and RQ are not keyed in, they are set to the default values.
The calculation requires entering the sample type as “Arterial” or “Capillary”.
Continued on next page
T pTOB2 B( TT T) T(continued)T
Tp TOB2 B(A)
Tp TOB2 B(A, TT T)
6. Parameters ABL800 FLEX reference manual
6-32
List of equations, TContinuedT
TEq. 17T:
(I)O(a)O
(I)O/(a)O2
222 F
pFp =
The calculation cannot be performed on the basis of the default TF TOB2B(I) value, and the calculation requires entering the sample as “Arterial” or “Capillary”.
TEq. 18T:
(I)O),(aO(I)O/)(a,O
2
222 F
TpFTp =
The calculation cannot be performed on the basis of the default TF TOB2B(I) value, and the calculation requires entering the sample as “Arterial” or “Capillary”.
TEq. 19T Refer to Eq. 46.10:
The ODC is determined as described in equations 46-47 in the section TOxyhemoglobin Dissociation Curve Tfurther in this chapter.
( )
p FF F
50= PCOHb
0.5 1- COHb- MetHb1+
×
where
Description See...
P = ODC(S,A,TT T) Eq. 47
( )S =
× − +0 5 1. F F FF
COHb - MetHb COHb1- MetHb
Eq. 46.11
A = a
TT T = 37.0 P
oPC Eq. 46.13
Eq. 20:
The ODC is determined as described in equations 46-47 in the section TOxyhemoglobin Dissociation Curve Tfurther in this chapter.
( )
p T FF F
50( )= PCOHb
0.5 1- COHb- MetHb1+
×
where
Continued on next page
pOB2 B(a)/TFTOB2 B(I)
pOB2 B(a,TT T)/ TFTOB2B(I)
Tp T50
Tp T50(TT T)
ABL800 FLEX reference manual 6. Parameters
6-33
List of equations, TContinuedT
Description See…
P = ODC(S,A,TT T) Eq. 47
( )S =
× − +0 5 1. F F FF
COHb - MetHb COHb1- MetHb
Eq. 46.11
( )0.37)(
pH04.1a −××−= TT
A∂∂
( )∂∂
pH pH( )
. . ( ) .T
= − × − × × −− −146 10 6 5 10 37 7 402 3
TT T = patient temperature in P
oPC (keyed-in)
TEq. 21T:
p50 is calculated for pH = 7.40, pCOB2B = 5.33 kPa, FCOHb = 0, FMetHb = 0, FHbF = 0.
The ODC is determined as described in equations 46-47 in the section TOxyhemoglobin Dissociation CurveT, see equation 47 further in this chapter.
p50(st) = ODC(S,A,TT T)
where
Description See…
S = 0.5 Eq. 46.11
A = a6 corresponds to pH = 7.40, pCOB2B = 5.33 kPa, FCOHb = 0, FMetHb = 0, FHbF = 0
Eq. 46.13
TT T = 37.0 P
oPC
TEq. 22T: T
p p pO (A a) = O (A) - O (a)2 2 2−
The calculation requires entering the sample type as “Arterial” or “Capillary”.
TEq. 23T: T
p T p T p TO (A a, ) = O (A, ) O (a, )2 2 2− −
The calculation requires entering the sample type as “Arterial” or “Capillary”.
Continued on next page
Tp T50(TT T) T (continued)T
Tp T50(st)
Tp TOB2 B(A-a)
Tp TOB2 B(A-a, TT T)
6. Parameters ABL800 FLEX reference manual
6-34
List of equations, TContinuedT
TEq. 24T: T
)A(O(a)O
(a/A)O2
22 p
pp =
The calculation requires entering the sample type as “Arterial” or “Capillary”.
TEq. 25T: T
)(A,O),(O
=)(a/A,O2
22 Tp
TapTp
The calculation requires entering the sample type as “Arterial” or “Capillary”.
TEq. 26 T[8]:T
The ODC is determined as described in equations 46-47 in the section TOxyhemoglobin Dissociation Curve Tfurther in this chapter.
pOB2B(x) is calculated by a numerical method, using:
Eq. Description See…
26.1 S = ODC(P,A,TT T) Eq. 47
26.2 ( )s
F FF F
OMetHb COHb
COHb MetHb2,i =× − −
− −
S 11
Eq. 46.12
26.3
( )
pF
s F F
O PCOHb
O COHb MetHb
2,i
2,i
=+
× − −1
1
Eq. 46.10
26.4 ( )t tHb 1 COHb MetHb O
+ 9.83 10 Oi 2,i
32,i
= × − − × +
× ×−
c F F s
p
26.5 A = a
26.6 TT T = 37 P
oPC
When tBi B= ctOB2 B– 2.3 mmol/L, then pOB2,i B= pOB2B(x), where ctOB2B is determined as described in equation 27.
pOB2B(x) cannot be calculated on the basis of a default ctHb value.
pOB2B(x) can only be calculated if the measured sOB2B(a) ≤ 0.97 (or p50(st) keyed in).
The calculation requires entering the sample type as "Arterial" or "Capillary".
Continued on next page
Tp TOB2 B(a/A)
Tp TOB2 B(a/A,TT T)
pOB2B(x)
(or Tp TBxB)
ABL800 FLEX reference manual 6. Parameters
6-35
List of equations, TContinuedT
TEq. 27T [5]:
( )c p s F F ctO O O O COHb MetHb tHb2 2 2 2= × + × − − ×α 1
αOB2B is the concentrational solubility coefficient for OB2B in blood (here set to 9.83 x 10P
−3P mmolLP
–1PkPaP
–1P at 37 P
oPC [5,19].
TcTtOB2 B cannot be calculated on the basis of a default TcTtHb value.
Eq. 28:
ctOB2B(a − v– ) = ctOB2B(a) – ctOB2B(v– )
where ctOB2B(a) and ctOB2B(v– ) are calculated from equation 27 for arterial and mixed
venous blood, respectively. The calculation requires two measurements.
TEq. 29T [7]:T
( )BO c F F2 1= × − −tHb COHb MetHb
BOB2B cannot be calculated on the basis of a default ctHb value.
TEq. 30T [8]:
The ODC is determined, as described in equations 46-47 in the section TOxyhemoglobin Dissociation Curve Tfurther in this chapter.
c ctO tO2 2( ) ( )x a t i= −
where
Eq. Description See…
30.1 ( )t tHbi = × − × +
+ × ×−
c F F s
p
1
9 83 10 3
COHb - MetHb O
O (5)2,i
2.
30.2 pOB2B(5) = 5.00 kPa
30.3 S = ODC(P,A,TT T) Eq. 47
30.4
( )P = × +
× − −
⎡
⎣⎢⎢
⎤
⎦⎥⎥
p Fs F F
O (5) COHbO COHb MetHb2
2,i
11
Eq. 46.9
30.5 ( )( )
sF F
F FO
MetHb COHbCOHb MetHb2,i =
× − −
− −
S 11
Eq. 46.12
30.6 A = a
30.7 TT T = 37.0 P
oPC
Continued on next page
TcTtOB2 B
TcTtOB2 B(a−v– )
TBTOB2 B
TcTtOB2 B(x) (or TcTBx B)
6. Parameters ABL800 FLEX reference manual
6-36
List of equations, TContinuedT
ctOB2B(a) is determined as described in equation 27.
ctOB2B(x) cannot be calculated on the basis of a default ctHb value.
ctOB2B(x) can only be calculated if the measured sOB2B(a) ≤ 0.97 (or if p50(st) is keyed in).
The calculation requires entering the sample type as “Arterial” or “Capillary”.
Eq. 31:
DO tO2 2
. .= ×c Qt
Q· BtB is the cardiac output and is an input parameter for calculation of D· OB2B.
If Q· BtB is not keyed in, D· OB2B will not be calculated.
The calculation requires entering the sample type as “Arterial” or “Capillary”.
Eq. 32:
Q VOtO a vt
. .=
−2
2c ( )
If V· OB2B is not keyed in, Q· BtB will not be calculated.
Eq. 33:
V O Q2 2
. .= × −t tO a vc ( )
If Q· BtB is not keyed in, V·OB2B will not be calculated.
Eq. 34T T[5]:
Fc cc c
Shunt =tO c tO atO c tO v
2 2
2 2
( ) ( )( ) ( )
−−
and
Eq. Description
34.1 F
c cc c
ShunttO A tO atO A tO v
2 2
2 2
≅−−
( ) ( )( ) ( )
Continued on next page
TcTtOB2 B(x) (or cBx B)T (continued)T
D· OB2 B
Q· Bt B
V·OB2 B
TFTShunt
ABL800 FLEX reference manual 6. Parameters
6-37
List of equations, TContinuedT
Eq. Description
34.2 F c c
c cShunt = tO a tO v
tO A tO a2 2
2 2
11
+−−
⎡
⎣⎢
⎤
⎦⎥
−( ) ( )( ) ( )
where
ctOB2B(c): total oxygen in pulmonary capillary blood
ctOB2B(a): total oxygen in arterial blood
ctOB2B(A): total oxygen in alveolar air. Oxygen tension = pOB2B(A).
ctOB2B(v– ): total oxygen in mixed venous blood
34.3 ( ) (a)OMetHbCOHb1tHb(a)O1083.9)a(tO 223
2 sFFcpc ×−−×+×= −
34.4
( ) (A)OMetHbCOHb1tHb(A)O1083.9)A(tO
2
23
2
sFFcpc
×−−×+×= −
34.5
( ) )v(OMetHbCOHb1tHb)v(O1083.9)v(tO
2
23
2
sFFcpc
×−−×+×= −
where:
pOB2B(a): oxygen tension in arterial blood; measured
pOB2B(A): oxygen tension in alveolar blood. See equation 15.
pOB2B(v– ): oxygen tension in mixed venous blood; measured and then entered
sOB2B(a): oxygen saturation in arterial blood; can be measured
sOB2B(A): oxygen saturation in (alveolar) blood calculated from equation 39 where P = pOB2B(A). If sOB2B(a) > 0.97, a keyed-in p50(st) will be used to determine the ODC. If sOB2B(a) > 0.97 and no p50(st) has been keyed in, the default value (3.578 kPa) will be used to determine the ODC.
sOB2B(v– ): oxygen saturation in mixed venous blood
If not keyed in, it will be calculated from equation 39 where P = pOB2B(v– ). If sOB2B(a) > 0.97, a keyed-in p50(st) will be used to determine the ODC.
The calculation requires entering the sample type as “Arterial” or “Capillary”. If sOB2B(a) > 0.97 and no p50(st) has been keyed in, the default value (3.578 kPa) will be used to estimate the ODC.
If no venous sample is measured, FShunt is estimated assuming:
ctOB2B(a) – ctOB2B(v– ) = 2.3 mmol/L in equation 34.2
Continued on next page
TFTShunt T(continued)T
6. Parameters ABL800 FLEX reference manual
6-38
List of equations, TContinuedT
TEq. 35T [5,16]:
F Tc T c Tc T c T
Shunt( ) =tO (a, ) - tO (v, )tO (A, ) - tO (a, )
2 2
2 2
11
+⎡
⎣⎢
⎤
⎦⎥
−
where
ctOB2B(a,TTT): total oxygen in arterial blood at patient temperature
ctOB2 B(A, TTT): total oxygen in alveolar blood at patient temperature ctOB2 B(v– , TTT): total oxygen in mixed venous blood at patient temperature
Eq. Description See…
35.1 ctOB2B(a,TTT) = ctOB2B calculated from equation 25 for arterial pOB2B and sOB2B values at 37 P
oPC
35.2
( )c T T pO T
c F F s TtO (A O A
tHb 1- COHb - MetHb O (A,2 2
2
, ) ( ) ( , ))
= ×
+ × ×
α 2
35.3 ( ) ( )[ ]αO23 2 1 10 37 0
9 83 104 2
( ) .. .
TT
= × − × × + × × −−
e-1.15 10 -37.0-2 T
35.4 pOB2B(A,TTT) is calculated from equation 15
35.5 sOB2B(A,TTT) = S
35.6 S = ODC(P,A,TTT) Eq. 47
35.7 P = pOB2B(A,TTT)
35.8 ( )A a pH= − × × −104 37 0.
( ).∂
∂ TT
35.9 TTT = patient temperature (keyed-in)
35.10 ( )∂∂
pH pH( )
. . ( ) .T
= × − × −− −146 10 6 5 10 37 7 402 3
If sOB2B(a) > 0.97, a keyed-in p50(st) will be used to determine the ODC. If sOB2B(a) > 0.97 and no p50(st) has been keyed in, the default value (3.578 kPa) will be used to determine the ODC.
35.11 ctOB2B(v– ,TTT) = ctOB2B(v– ) at 37 P
oPC is calculated from equation 27 for
mixed venous blood values of pOB2B and sOB2B. If sOB2B(v– ) > 0.97, a
keyed-in p50(st) will be used to determine the ODC.
If sOB2B(v– )>0.97 and no p50(st) has been keyed in, the default
value (3.578 kPa) will be used to estimate the ODC. If no mixed venous sample is measured, the FShunt(TTT) is estimated assuming ctOB2B(a, TTT) – ctOB2 B(v– , TTT) = 2.3 mmol/L in equation 35.
Continued on next page
TFTShunt( TT T)
ABL800 FLEX reference manual 6. Parameters
6-39
List of equations, TContinuedT
TEq. 36T:T
RI =O A O a
O a2 2
2
p pp
( ) ( )( )
−
The calculation requires entering the sample type as “Arterial” or “Capillary”.
TEq. 37T:T
RI( ) =O (A O (a
O (a2 2
2
Tp T
p T p T, ) , ), )
−
The calculation requires entering the sample type as “Arterial” or “Capillary”.
TEq. 38T [8]:
The ODC is determined as described in equations 46-47 in the section TOxyhemoglobin Dissociation Curve Tfurther in this chapter.
Qx =−
2 3.( )ctO a t2 i
Eq. Description See…
38.1 ( )t tHb 1- COHb - MetHb O Oi 2, i 2= × × + × −c F F s p9 83 10 53. ( )
38.2 pOB2B(5) = 5.00 kPa
38.3 S = ODC(P,A,TTT)
38.4
( )⎥⎥⎦⎤
⎢⎢⎣
⎡
−−×+×=
MetHbCOHb1OCOHb1(5)O
i2,2 FFs
FpP Eq. 46.9
38.5 ( )s
F FF F
OS 1 MetHb COHb
1- COHb - MetHb2,i =× − −
Eq. 46.12
38.6 A = a
38.7 TTT = 37.0 P
oPC
ctOB2B(a) is determined as described in equation 27.
Qx cannot be calculated on the basis of a default ctHb value.
Qx can only be calculated if the measured sOB2B(a)B B≤ 0.97 (or if p50(st) is keyed in).
The calculation requires entering the sample type as “Arterial” or “Capillary”.
Continued on next page
RI
RI( TT T)
QBx B
6. Parameters ABL800 FLEX reference manual
6-40
List of equations, TContinuedT
TEq. 39T:T
The ODC is determined as described in equation 46 (points I and III). See the section TOxyhemoglobin Dissociation Curve Tfurther in this chapter.
( )s
F FF F
OS 1 MetHb COHb
1- COHb - MetHb2 =× − −
where
Description See…
S = ODC(P,A,TTT)
P p Fs F F
= +×
× − −pO O COHb
O COHb MetHb22
2 ( )1
Eq. 46.9
A = a
TTT = 37.0 P
oPC
Eq. 40:
( )F s F FO Hb O 1 COHb MetHb2 2= × − −
If TsTOB2 B is not measured, it will be calculated from equation 39.
If dyshemoglobins (FCOHb, FMetHb) are not known, they are set to the default values.
Eq. 41:
( )F s F F F FHHb O 1 COHb MetHb COHb MetHb2= − × − − − −1
If TsTOB2 B is not measured, it will be calculated from equation 39.
If dyshemoglobins (FCOHb, FMetHb) are not known, they are set to the default values.
Eq. 42T T[5]:
( )V V
F F c( )
.B =
× ×× − × ×
1 10 (CO)COHb(2) COHb(1) tHb
3
24 0 91
Continued on next page
TsTOB2 B
TFTOB2 BHb
TFTHHb
TVT(B)
ABL800 FLEX reference manual 6. Parameters
6-41
List of equations, TContinuedT
Eq. Description
42.1 ( )
V B( ) =× × − ×
VF F c
(CO)2.184 10 COHb(2) COHb(1) tHb-2
42.2 V(CO) = volume (in mL) of carbon monoxide injected according to the procedure and the value keyed in
42.3 FCOHb(1) = fraction of COHb measured before the CO injection
42.4 FCOHb(2) = fraction of COHb measured after the CO injection
Eq. 43: −−++ −−+= 3
+ HCOClKNaK,GapAnion cccc (P)
TEq. 44:T
Anion Gap Na Cl HCO3= − −+ − −c c c (P)
Eq. 45 Ref. [12]:
( )[ ]c cCa Ca pH22 7 4 1 0 53 7 40+ += − × −( . ) . .
Due to biological variations this equation can only be used for a pH value in the range 7.2-7.6.
See TOxyhemoglobin dissociation curve (ODC), Tpage 6-44 in this chapterT. T
TEq. 48:
GluNa2Osm ccm += +
TEq. 49:T
An iterative method is used to calculate FHbF. The input parameters are sOB2 B, TcTeHb (effective hemoglobin concentration) and TcTOB2 BHbF (concentration of fetal oxyhe-moglobin).
In the calculations the following are assumed: pH = 7.4, pCOB2B = 5.33 kPa, FCOHb = 0, FMetHb = 0, TcTDPG = 5 mmol/L, and temp = 37 °C.
Step Description See…
1. An estimate of FHbF is made: FHbF BestB = 0.8
2. pOB2B, BestB = ODC (sOB2 B,A, TTT); Eq. 47
where the constant A depends on FHbF = FHbF BestB
Continued on next page
TVT(B) T(continued)T
Anion Gap,KP
+P
Anion Gap
TcTCa P
2+P(7.4)
Eq. 46-47
TmTOsm
FHbF
6. Parameters ABL800 FLEX reference manual
6-42
List of equations, TContinuedT
Step Description See…
3. sOB2 B (for fetal blood) = ODC (pOB2 B,BestB, A, TTT); Eq.47
where FHbF = 1
4. TcTOB2 BHbFBestB = sOB2 B (fetal blood) × TcTeHb × FHbF Best B
5. ΔF c c
cHbF O HbF O HbF
eHbest2 meas. 2 est=
−
6. If |ΔTF THbF BestB| ≥ 0.001, proceed to step T7T.
If |ΔTF THbF BestB| < 0.001, proceed to step T9T.
7. FHbF Best, new B= FHbF Best, old B+ Δ TF THbF Best B
8. Return to step T2 T.
9. End of iteration. The value for FHbF has converged.
TEq. 50 T[8]: T
The ODC is determined as described in equations 46-47 in TOxyhemoglobin Dissociation Curve Tfurther in this chapter.
pOB2B(x) is calculated by a numerical method, using:
Eq. Description See…
50.1 S = ODC(P,A,TT T) Eq. 47
50.2 ( )MetHbCOHb1
COHbMetHb1)(O i2, FFFFSTs
−−−−×
= Eq. 46.12
50.3
( )MetHbCOHb1)(OCOHb1
P)(O
i2,
i2,
FFTsF
Tp
−−×+
= Eq. 46.10
50.4 ( ))(O)(O+
)(OMetHbCOHb1tHb)(t
i2,2
i2,i
TpTTsFFcT
×
+×−−×=
α
50.5 A = a
50.6 TT T = patient temperature
50.7 [ ]25 )37(1021)37(115.02 00983.0)(αO −××+−×− −
×= TTeT
Continued on next page
TFTHbF T (continued)T
pOB2 B(x,TT T)
ABL800 FLEX reference manual 6. Parameters
6-43
List of equations, TContinuedT
Eq. Description
50.8 )(x,OO 2i2, Tpp =
when tBiB(TT T) = ctOB2 B(37 °C) − 2.3 mmol/L
pOB2B(x,TT T) is calculated in accordance with OSA V3.0.
pOB2B(x,TT T) can only be calculated if the measured sOB2B(a) ≤ 0.97 (or p50(st) keyed in).
pOB2B(x, TT T) is tagged with "?" if any of the following parameters: sOB2 B, FMetHb, FCOHb, pO B2 B, pCOB2B, pH or ctHb is tagged with "?".
The calculation requires entering the sample type as “Arterial” or “Capillary”.
TEq. 51T: T
275.6(amb)CO
air)(dry/CO 22 −
=p
pVV
TEq. 52T: T
275.6(amb)O
air)(dry/O 22 −
=p
pVV
TEq. 53 T[24, 25]T:
T
( ) ( ) ( ) ( )210.1
femaleif0.742Age/88.4S186m3mL/min/1.7GFR -0.203154.1
cr2
×
×××=−
TEq. 54 T[24, 25]T:
( ) ( ) ( ) ( )femaleif0.742Age/88.4S186m3mL/min/1.7GFR -0.203154.1
cr2 ×××=
−
Tp TOB2 B(x,TT T) T(continued)
TVTCOB2 B/ TVT(dry air)
TVTOB2 B/ TVT(dry air)
GFR, if AA
GFR, if non AA
6. Parameters ABL800 FLEX reference manual
6-44
Oxyhemoglobin dissociation curve (ODC)
These equations account for the effect of FCOHb on the shape of the Oxyhemoglobin Dissociation Curve (ODC) in accordance with the Haldane equation.
TEq. 46T [16,18]:T
( )[ ]y y x x h tanh k x xo o o o− = − + × −( )
where k P
oP = 0.5343
Eq. Description
46.1 x ln = p
46.2 y ln
1-=
ss
46.3 y ln
1-o
o
o=s
s where TsTP
oP = 0.867
46.4 x x a b ln( a bo oo oo= + + = + +p ) where Tp TP
ooP = 7 kPa
The actual position of the ODC in the coordinate system (ln(s/(1–s)) vs ln(p)) used in the mathematical model, is expressed by equations 46.3 and 46.4.
The symbols "a" and "b" reflect the ODC displacement from the reference position to its actual position in this coordinate system:
"a" describes the displacement at 37 °C. "b" the additional displacement due to the patient temperature difference from 37 °C.
The reference position of the ODC was chosen to be the one that corresponds to the default value for p50(st) = 3.578 kPa, which is traditionally considered the most likely value of p50 for adult humans under standard conditions, namely:
pH = 7.40
pCOB2B = 5.33 kPa
FCOHb, FMetHb, FHbF = 0
cDPG = 5 mmol/L
Continued on next page
ODC equations
The ODC reference position
ABL800 FLEX reference manual 6. Parameters
6-45
Oxyhemoglobin dissociation curve (ODC), TContinuedT
The ODC displacement which is described by "a" and "b" in the coordinate system (ln(s/(1–s))vsln(p)), is given by the change in p50 from the default to its actual value in a more common coordinate system (sOB2B, pOB2B).
Eq. Description
46.5 x x ln
7a bo− = − −
p
46.6 h = +h ao where hP
oP = 3.5
46.7 b = × −0 055. ( )T T o TT TP
oP = 37 P
oPC
46.8 p p p= + ×O CO2 M
where M × pCO is taken from the Haldane equation [20]:
pc
pc
OO Hb
COCOHb
2
2
= ×M , to give eq. 46.9
46.9 p p p
sF
- F - F= + × ⎡
⎣⎢⎤⎦⎥
O OO
COHb1 COHb MetHb2
2
2
or equation 46.10
46.10 ( )[ ]p
p s F F
FO
O COHb MetHb
COHb22=
× × − −
+
1
1
The ordinate, TsT, may loosely be termed the combined oxygen/carbon monoxide saturation of hemoglobin and is described by equation 46.11 below:
Eq. Description
46.11
( )
sc c
c c cs F F F
F
=+
+ +
=× +
−
O Hb COHbO Hb COHb HHbO 1- COHb - MetHb COHb
MetHb
2
2
2
1
or
46.12 ( )s
s F FF F
O1- MetHb COHb
COHb MetHb2 =× −
− −1
Continued on next page
The ODC displacement
6. Parameters ABL800 FLEX reference manual
6-46
Oxyhemoglobin dissociation curve (ODC), TContinuedT
The actual position of the ODC at 37 °C for a given sample is, in principle, determined in two steps:
1. The calculation of the combined effect on the ODC position at 37 °C of all known causes for displacement (= ac in equation 46.13), and based on this position.
2. The computation by a numerical method of the actual position of the ODC curve by shifting it to pass through the known set of coordinates (PB0B, SB0B).
Eq. Description
46.13 a = ac + a6
46.14 ac = a1 + a2 + a3 + a4 + a5
46.15 a1 = −0.88 × (pH − 7.40)
46.16 a2 ln
CO5.33
2= ×0 048.p
46.17 a3 MetHb= − ×0 7. F
46.18 ( ) ( )a4 HbF DPG= − × −0 06 0 02 5. . F c
46.19 a5 HbF= − ×0 25. F
Step Description
I:
pOB2B, sOB2 B can be used.
If TsTOB2 B > 0.97, the calculation is based on II or III – Tsee belowT.
Coordinates (P B0B, S B0B) are calculated from equations (46.9) and (46.11).
If FCOHb and FMetHb are not known, the default values are used.
The ODC is shifted from the reference position to a position that corresponds to the effect of all measured parameters according to step I.
The magnitude of the shift is “ac”.
The ODC is then further shifted to pass through the point (PB0B, SB0B).
The magnitude of the shift is "a6".
Continued on next page
The actual ODC position
Determining the actual displacement
ABL800 FLEX reference manual 6. Parameters
6-47
Oxyhemoglobin dissociation curve (ODC), TContinuedT
Step Description
II:
sOB2 B > 0.97 (or erroneous) and p50(st) is keyed in.
Coordinates (PB0B, SB0B) are calculated from (p50(st), 0.5) using equations 46.9 and 46.11.
Reference position of the ODC.
The ODC is shifted from the reference position to pass through the point (PB0B, SB0B). In this position, the ODC reflects the p50(st) of the patient, i.e., the particular patient but at standard conditions.
The ODC is further shifted, as determined by the effect of the measured parameters (“ac”), to its actual position. This position reflects the p50(act) of the patient.
(III):
sOB2 B> 0.97 (or erroneous) and no p50(st) has been keyed in.
Reference position of the ODC.
The position of the actual ODC can now be approximated from the reference position, using the actual values of pH, pCOB2B, FCOHb, FMetHb and FHbF to determine the shift "ac".
The curves are used only to illustrate the principles of the ODC determination
Continued on next page
Determining the actual displacement T (continued)T
TNOTE: T
6. Parameters ABL800 FLEX reference manual
6-48
Oxyhemoglobin dissociation curve (ODC), TContinuedT
Calculation of a set of coordinates on the ODC is symbolized by:
Eq. 47T:
S = ODC(P, A, TT T) or P = ODC(S, A, TT T)
These equations are symbolic representations of the relationship between saturation (S), tension (P), displacement (A) and temperature (TT T).
To calculate S or P and to further calculate sOB2B and pOB2B, the other variables should be specified. S and P are calculated using numerical methods.
P is input to equation 46.1.
S is input to equation 46.2.
A is input to equation 46.5.
TT T is input to equation 46.7.
Coordinates on the ODC
ABL800 FLEX reference manual 6. Parameters
6-49
Conversion of units
The equations stated above are based on the SI-unit system. If parameters are known in other units, they must be converted into a SI unit before entering the equations. The result will be in an SI unit.
After the calculation the result may be converted to the desired unit. Conversion of units may be performed, using the equations stated below:
TT T °F = 32) (59 o +CT
TT T °C = )32 (95 o −FT
TcTX (meq/L) = TcTX (mmol/L) where X is KP
+P, NaP
+P or ClP
−P.
TcTCaP
2+P (meq/L) = 2 × TcTCaP
2+P (mmol/L) or
TcTCaP
2+P (mg/dL) = 4.008 × TcTCaP
2+P (mmol/L)
TcTCaP
2+P (mmol/L) = 0.5 × TcTCaP
2+P (meq/L)or
TcTCaP
2+P (mmol/L) = 0.2495 × TcTCaP
2+P (mg/dL)
p (mmHg) = p (torr) = 7 500638. × p (kPa)
p (kPa) = (mmHg)133322.0 p×
= 0133322. × p (torr)
[4]
TcTtHb (g/dL) = 1.61140 × ctHb (mmol/L)
ctHb (g/L) = 16.1140 × ctHb (mmol/L) or
ctHb (mmol/L) = 0.62058 × ctHb (g/dL)
ctHb (mmol/L) = 0.062058 × ctHb (g/L)
Vol % = 2.241 × (mmol/L)
Vol % = mL/dL
mmol/L = 0.4462 × (mL/dL)
Continued on next page
SI units
Temperature
cKP
+P, cNa P
+P, cCl P
–P
cCa P
2+P
Pressure
ctHb
TcTtCOB2 B, ctOB2 B, TcTtOB2 B(a−v– ), TBTOB2 B
6. Parameters ABL800 FLEX reference manual
6-50
Conversion of units, TContinuedT
V·OB2B (mmol/L)/min = V·OB2B/22.41 (mL/dL)/min
[22]
cGlucose (mg/dL) = 18.016 × TcTGlucose (mmol/L) or
cGlucose (mmol/L) = 0.055506 × TcTGlucose (mg/dL)
[22]
cLactate (mg/dL) = 9.008 × TcTLactate (mmol/L) or
cLactate (mmol/L) = 0.11101 × TcTLactate (mg/dL)
cLactate (meq/L) = cLactate (mmol/L)
(conversion based on the molecular weight of lactic acid)
TcTtBil (μmol/L) = 17.1 × ctBil (mg/dL)
ctBil (μmol/L) = 1.71 × ctBil (mg/L) or
ctBil (mg/dL) = 0.0585 × ctBil (μmol/L)
ctBil (mg/L) = 0.585 × ctBil (μmol/L)
TcTCrea (μmol/L) = 88.40 × cCrea (mg/dL)
All conversions of units are made by the analyzer.
V·OB2 B
TcTGlucose
TcTLactate
TcTtBil
TcTCrea
TNOTE: T
ABL800 FLEX reference manual 6. Parameters
6-51
Default values
The following default values are used in the ABL800 FLEX analyzers, if other values are not keyed in.
TT T = 37.0 °C (99 °F)
FOB2B(I) = 0.21 (21.0 %)
RQ = 0.86
ctHb = 9.3087 mmol/L, (15.00 g/dL or 150 g/L)
FCOHb = 0.004 (0.4 %)
FMetHb = 0.004 (0.4 %)
p50(st) = 3.578 kPa (26.84 mmHg)
In addition to the above default values, the ABL8x7 FLEX analyzers use the following default:
Ambient temperature = 25.0 °C (77 °F).
Values
6. Parameters ABL800 FLEX reference manual
6-52
Altitude correction
The barometric pressure is measured by the analyzer's built-in barometer, and the effect of barometric pressure on blood samples is compensated by the analyzer’s software.
Quality control result for pOB2 B obtained on aqueous quality control solutions at low barometric pressure (at high altitudes) is affected as the properties of aqueous solutions differ from those of blood. The deviation from the pOB2 B value obtained at sea level can be expressed by an altitude correction that can be added to the control ranges.
The relationship between the altitude and barometric pressure can be expressed by the following equation:
( )actref
actref
BBBB
TA+
−×+×= 004.0116000
where:
TAT = altitude in m
TT T = temperature in °C
TBBrefTB = standard barometric pressure at sea level = 760 mmHg
TBBactTB = actual barometric pressure in mmHg.
Reference [23].
Equation for altitude correction
ABL800 FLEX reference manual 6. Parameters
6-53
References
1. The Deep PictureP
TMP, critical information from blood gas analysis. Copenhagen:
Radiometer Medical A/S, 1993: 1-14.
2. Wandrup JH. Physicochemical logic and simple symbol terminology of oxygen status. Blood Gas News 1993; 2,1: 9-11.
3. Siggaard-Andersen O, Durst RA, Maas AHJ. Approved recommendation (1984) on physicochemical quantities and units in clinical chemistry. J Clin Chem Clin Biochem 1987; 25: 369-91.
4. Siggaard-Andersen O. The acid-base status of the blood. 4th revised ed. Copenhagen: Munksgaard, 1976.
5. Siggaard-Andersen O, Wimberley PD, Fogh-Andersen N, Gøthgen IH. Measured and derived quantities with modern pH and blood gas equipment: calculation algorithms with 54 equations. Scand J Clin Lab Invest 1988; 48, Suppl 189: 7-15.
6. Burnett RW, Noonan DC. Calculations and correction factors used in determination of blood pH and blood gases. Clin Chem 1974; 20,12: 1499-1506.
7. Wimberley PD, Siggaard-Andersen O, Fogh-Andersen N, Zijlstra WG, Severinghaus JW. Hemoglobin oxygen saturation and related quantities: definitions, symbols and clinical use. Scand J Clin Lab Invest 1990; 50: 455-59. Available as AS104.
8. Siggaard-Andersen O, Gøthgen IH, Wimberley PD, Fogh-Andersen N. The oxygen status of the arterial blood revised: relevant oxygen parameters for monitoring the arterial oxygen availability. Scand J Clin Lab Invest 1990; 50, Suppl 203: 17-28. Available as AS108.
9. Wandrup JH. Oxygen uptake in the lungs. Blood Gas News 1992; 1,1: 3-5.
10. Tietz NW, Logan NM. Reference ranges. In: Tietz NW, ed. Fundamentals of clinical chemistry. 3rd ed. Philadelphia: WB Saunders Company, 1987: 944-75.
11. Siggaard-Andersen O, Wimberley PD, Fogh-Andersen N, Gøthgen IH. Arterial oxygen status determined with routine pH/blood gas equipment and multi-wavelength hemoximetry: reference values, precision and accuracy. Scand J Clin Lab Invest 1990; 50, Suppl 203: 57-66. Available as AS106.
12. Siggaard-Andersen O, Thode J, Wandrup JH. The concentration of free calcium ions in the blood plasma ionized calcium. In: Siggaard-Andersen O, ed. Proceedings of the IFCC expert panel on pH and blood gases held at Herlev Hospital 1980. Copenhagen: Radiometer Medical A/S, 1981: 163-90. Available as AS79.
13. Severinghaus JW. Blood gas calculator. J Appl Physiol 1966; 21,3: 1108-16. Available as ST36.
14. Christiansen TF. An algorithm for calculating the concentration of the base excess of blood. In: Siggaard-Andersen O, ed. Proceedings of the IFCC expert panel on pH and blood gases held at Herlev Hospital 1980. Copenhagen: Radiometer Medical A/S, 1981: 77-81.
Continued on next page
List of references
6. Parameters ABL800 FLEX reference manual
6-54
References, TContinuedT
15. Kokholm G. Simultaneous measurements of blood pH, pCOB2B, pOB2B and concentrations of hemoglobin and its derivatives – a multicenter study. Scand J Clin Lab Invest 1990; 50, Suppl 203: 75-86. Available as AS107.
16. Siggaard-Andersen O, Wimberley PD, Gøthgen IH, Siggaard-Andersen M. A mathematical model of the hemoglobin-oxygen dissociation curve of human blood and of the oxygen partial pressure as a function of temperature. Clin Chem 1984; 30: 1646-51.
17. Siggaard-Andersen O, Wimberley PD, Gøthgen IH, Fogh-Andersen N, Rasmussen JP. Variability of the temperature coefficients for pH, pCOB2B and pOB2B in blood. Scand J Clin Lab Invest 1988; 48, Suppl 189: 85-88.
18. Siggaard-Andersen O, Siggaard-Andersen M. The oxygen status algorithm: a computer program for calculating and displaying pH and blood gas data. Scand J Clin Lab Invest 1990; 50, Suppl 203: 29-45.
19. Bartels H, Christoforides C, Hedley-Whyte J, Laasberg L. Solubility coefficients of gases. In: Altman PL, Dittmer DS, eds. Respiration and circulation. Bethesda, Maryland: Fed Amer Soc Exper Biol, 1971: 16-18.
20. Roughton FJW, Darling RC. The effect of carbon monoxide on the oxyhemoglobin dissociation curve. Am J Physiol 1944; 141: 17-31.
21. Engquist A.. Fluids electrolytes nutrition. Copenhagen: Munksgaard, 1985: 56-68 and 118.
22. Olesen H et al. A proposal for an IUPAC/IFCC recommendation, quantities and units in clinical laboratory sciences. IUPAC/IFCC Stage 1, Draft 1, 1990: 1-361.
23. Kokholm G, Larsen E, Jensen ST, ChristiansenTF. 3rd ed. Blood gas measurements at high altitudes. Copenhagen: Radiometer Medical A/S, 1991. Available as AS109.
24. U.S Department of Health and Human Services, National Institutes of Health, National Institute of Diabetes and Digestive and Kidney Diseases: NKDEP National Kidney Disease Education Program. Rationale for Use and Reporting of Estimated GFR. NIH Publication No. 04-5509. Revised November 2005.
25. Myers GL, Miller WG, Coresh J, Fleming J, Greenberg N, Greene T, Hostetter T, Levey AS, Panteghini M, Welch M, and Eckfeldt JH for the National Kidney Disease Education Program Laboratory Working Group. Clin Chem, 52:5-18, 2006; First published December 6, 2005, 10.1373/clinchem.2005.0525144.
List of references T(continued)T
7-1
7. Solutions and gas mixtures
Overview
This chapter gives information about all the solutions and gases used with the ABL800 FLEX analyzer, their composition, use and consumption.
The Certificates of Traceability for the calibrating solutions are found at the end of the chapter.
This chapter contains the following topics.
General information ......................................................................................... X7-2X
Calibration solutions ........................................................................................ X7-3X
Rinse and Cleaning solutions ........................................................................... X7-6X
Electrolyte solutions......................................................................................... X7-8X
S5362 Hypochlorite Solution........................................................................... X7-10X
Gas mixtures (Gas 1 and Gas 2)....................................................................... X7-11X
Traceability certificates.................................................................................... X7-12X
Introduction
Contents
7. Solutions and gas mixtures ABL800 FLEX reference manual
7-2
General information
All the solutions described in this chapter are for Tin vitro T diagnostic use.
Each solution is identified with an "S" and is followed by 4 or 5 digits. The name of the solution comes after the number.
The two gas mixtures used by the analyzer are named Gas 1 and Gas 2.
The expiration date of a solution found on the label or on a sticker on the side of the container is stated as a year and month. Do not use a product after its expiration date.
Safety Data Sheets for all solutions are available from your Radiometer distributor.
Information for reordering solutions from Radiometer can be found in the TABL800 FLEX Operator’s Manual, Chapter 14T.
In Vitro Diagnostic Use
Solution numbers
Gas names
Expiration date
Safety Data Sheets
Reordering
ABL800 FLEX reference manual 7. Solutions and gas mixtures
7-3
Calibration solutions
Use: For calibration of the pH, electrolyte and metabolite electrodes in the ABL835/30/25/20/15/10/05 FLEX analyzers.
Quantity: 200 mL
Composition: Contains the following substances with the stated nominal concentrations:
Solution Substance Concentration (mmol/L)
S1820 KP
+
NaP
+
CaP
2+
Cl P
−
TcTGlu
cLac
buffer
4
145
1.25
102
10
4
Maintains a pH of 7.40
S1830 KP
+
NaP
+
CaP
2+
Cl P
−
buffer
40
20
5
50
Maintains a pH of 6.9
TThe exact values are included in the ba code. T
Additives: Preservatives and surfactants.
Storage: At 2-25 P
oPC (36-77 P
oPF).
Stability: Expiration date and Lot No. are printed on a label.
Stability in use: 4 weeks for S1820
8 weeks for S1830
Continued on next page
S1820 and S1830
7. Solutions and gas mixtures ABL800 FLEX reference manual
7-4
Calibration solutions, TContinuedT
Use: For calibration of the pH, electrolyte and metabolite electrodes in the ABL837/27/17 FLEX analyzers.
Quantity: 175 mL
Composition: Contains the following substances with the stated nominal concentrations:
TcTKP
+
TPcTNaP
+
TPcTCaP
2+
TPcTCl P
−
PcGlu cLac cCrea buffer
4 mmol/L 145 mmol/L 1.25 mmol/L 102 mmol/L 10 mmol/L 4 mmol/L 200 μmol/L Maintains a pH of 7.40
TThe exact values are included in the barcode. T
Additives: Preservatives and surfactants.
Storage: At 2-10 P
oPC (36-50 P
oPF).
Stability: Expiration date and Lot No. are printed on a label.
Stability in use: 2 weeks
Use: For calibration of the pH, electrolyte and metabolite electrodes in the ABL837/27/17 FLEX analyzers.
Quantity: 140 mL
Composition: Contains the following substances with the stated nominal concentrations:
TcTKP
+
TPcTNaP
+
TPcTCaP
2+
TPcTCl P
−
TPcTCreatine buffer
10 mmol/L 50 mmol/L 5 mmol/L 50 mmol/L 190 μmol/L Maintains a pH of 6.9
TThe exact values are included in the barcode. T
Additives: Preservatives and surfactants.
Storage: At 2-10 P
oPC (36-50 P
oPF).
Stability: Expiration date and Lot No. are printed on a label.
Stability in use: 2 weeks
Continued on next page
S1827 Calibration Solution 1
S1837 Calibration Solution 2
ABL800 FLEX reference manual 7. Solutions and gas mixtures
7-5
Calibration solutions, TContinuedT
Use: For calibration of the cuvette optical path length. The calibrated value can be ctHb, ctHb and TcTtBil, or TcTtBil depending on the analyzer version.
Quantity: 2 mL
Composition: Salts, a buffer, preservative and a coloring agent.
Storage: Keep in a dark place at 2-25 P
oPC (36-77 P
oPF).
After opening the solution must be used at once.
S7770 tHb Calibration Solution
7. Solutions and gas mixtures ABL800 FLEX reference manual
7-6
Rinse and Cleaning solutions
Use: For rinsing the liquid transport system after each measurement or calibration in the ABL835/30/25/20/15/10/05 FLEX analyzers.
Quantity: 600 mL
Composition: Contains salts, buffer, anticoagulant, preservatives, and surfactants.
Storage: At 2-32 P
oPC (36-90 P
oPF).
Stability: Expiration date and Lot no. are printed on a separate label.
When stored between 2-32 P
oPC (36-90 P
oPF), S4980 is stable for 25
months from the date of production, if unopened.
Use: For rinsing the liquid transport system after each measurement or calibration in the ABL837/27/17 FLEX analyzers.
Quantity: 600 mL
Composition: Contains salts, buffer, anticoagulants, preservatives and surfactants.
Storage: At 2-32 P
oPC (36-90 P
oPF).
Stability: Expiration date and Lot no. are printed on a separate label.
When stored between 2-32 P
oPC (36-90 P
oPF), S4987 is stable for 25
months from the date of production, if unopened.
Use: For cleaning the liquid transport system automatically or called by operator in the ABL835/30/25/20/15/10/05 FLEX analyzers.
Quantity: 200 mL
Composition: Contains salts, buffer, anticoagulant, preservatives and surfactants.
Storage: At 2-32 P
oPC (36-90 P
oPF).
Stability Expiration date and Lot no. are printed on a separate label.
Use: For adding to the S8370 Cleaning solution.
Composition: Contains powdered streptokinase.
Storage: At 2-10 °C (36-50 °F).
Stability: Expiration date and Lot no. are printed on a separate label.
The Cleaning Solution with the Cleaning Additive is stable for 2 months in use.
Continued on next page
S4980 Rinse Solution
S4987 Rinse Solution
S8370 Cleaning Solution
S5370 Cleaning Additive
ABL800 FLEX reference manual 7. Solutions and gas mixtures
7-7
Rinse and Cleaning solutions, TContinuedT
TWARNING/ CAUTION: T
TMay cause sensitization by inhalation and skin contact. Do not breathe dust. Avoid contact with skin.. Wear suitable gloves. In case of accident or if you feel unwell, seek medical advice immediately (show the label where possible). T
Use: For cleaning the measuring system of the ABL837/27/17 FLEX analyzers and for checking high creatinine level.
Quantity: 100 mL
Composition: Contains: T TcCrea = 1500 μmol/L, salts, buffer, anticoagulants, preservatives, surfactants and enzyme.
Storage: At 2-10 P
oPC (36-50 P
oPF).
Stability Expiration date and Lot no. are printed on a separate label.
The Cleaning Met II Solution is stable for 14 days in use.
WARNING/CAUTION
Contains streptokinase. May cause allergic reaction.
S5370 Cleaning Additive T(continued)
S8377 Cleaning Met II Solution
7. Solutions and gas mixtures ABL800 FLEX reference manual
7-8
Electrolyte solutions
The following electrolyte solutions contained in the electrode jackets of the Radiometer electrodes are used:
Electrolyte for… Quantity Composition
E1001 reference electrode
0.6 mL in four prefilled electrode jackets per D711 Membrane Box
Organic compounds and inorganic salts*.
E788 pCOB2B electrode
0.6 mL in four prefilled electrode jackets per D788 Membrane Box
Inorganic salts, buffer, hygroscopic compound, preservative and surfactant.
E799 pOB2B electrode
0.6 mL in four prefilled electrode jackets per D799 Membrane Box
Inorganic salts, organic compounds, buffer, preservative and surfactant.
E722 K electrode 0.6 mL in four prefilled electrode jackets per D722 Membrane Box
Organic compounds, inorganic salts, buffer, acid and preservative.
E755 Na electrode 0.6 mL in four prefilled electrode jackets per D755 Membrane Box
Inorganic salts, organic compounds, preservative and surfactant.
E733 Ca electrode 0.6 mL in four prefilled electrode jackets per D733 Membrane Box
Inorganic salts, organic compounds, buffer, preservative and surfactant.
E744 Cl electrode 0.6 mL in four prefilled electrode jackets per D744 Membrane Box
Inorganic salts, organic compounds, preservative, surfactant and hygroscopic products.
E7066 Glucose electrode
0.6 mL in five plastic capsules to fill the electrode jackets (four units) per D7066 Membrane Box
Buffer, inorganic salts, thickening agent, preservative and surfactant.
E7077 Lactate electrode
0.6 mL in five plastic capsules to fill the electrode jackets (four units) per D7077 Membrane Box
Buffer, inorganic salts, thickening agent, preservative and surfactant.
E8088 and E8089 Crea electrodes
0.6 mL in five plastic capsules to fill the electrode jackets (four untis) per D8088 and D8089 Membrane Box
Buffer, inorganic salts, thickening agent, preservative and surfactant.
* TWARNING/CAUTION TT: Irritating to eyes, respiratory system and skin. In case of contact with eyes, rinse immediately with plenty of water and seek medical advicTe.
Continued on next page
List of solutions
ABL800 FLEX reference manual 7. Solutions and gas mixtures
7-9
Electrolyte solutions, TContinuedT
TTemperature: T TElectrolyte solution:T
2-25 P
oPC (36-77 P
oPF) For glucose electrode
2-10 P
oPC (36-50 P
oPF) For lactate and crea
electrodes
2-32 P
oPC (36-90 P
oPF) For all other electrodes
Expiration date and Lot No. are printed on a label on the side of the membrane box.
Storage
Stability
7. Solutions and gas mixtures ABL800 FLEX reference manual
7-10
S5362 Hypochlorite Solution
Use: For protein removal and decontamination according to the procedures described in the TOperator's ManualT, chapter 4: TAnalyzer Menus and Programs T.
Quantity: 100 mL. Delivered with a 1-mL syringe.
Composition: Contains sodium hypochlorite (pH ≈12).
Storage: Keep in a dark place at 2-8 P
oPC (36-46 P
oPF). After use, keep the
bottle tightly capped to avoid contamination and decomposition.
Stability: Expiration date and Lot No. are printed on a separate label on the bottle.
S5362 Hypochlorite Solution
ABL800 FLEX reference manual 7. Solutions and gas mixtures
7-11
Gas mixtures (Gas 1 and Gas 2)
For calibration of the pCOB2B and pOB2 B electrodes.
The following types of Gas 1 cylinders are used depending on the geographical location of the analyzer:
T
Gas 1 T
T Gas 2 T
EU USA Japan
Cylinder volume 1 L 1 L 1 L 1 L
Gas volume 10 L 33 L 25 L 10 L
Fill pressure at 25 P
oPC
140 psi(10 bar)
500 psi(34 bar)
375 psi (26 bar)
140 psi (10 bar)
Composition 19.76 % O B2 B, 5.60 % CO B2B 74.64 % NB2 B
< 0.04 % O B2B, 11.22 % CO B2 B
88.78 % N B2 B
TPressurized container. Non-flammable compressed gas. Do not breathe gas. Gas mixtures containing less than 19.5 % oxygen may cause suffocation. Protect from sunlight and do not expose to temperatures exceeding 50 P
oPC (122 P
oPF). Store and
use with adequate ventilation. Keep away from oil and grease. Do not refill.T
TThe exact composition of each gas mixture is given in the barcode on the gas cylinder label. The barcode is entered by the barcode reader or manually. T
Gas 1 and Gas 2 are stable for 25 months from the date of filling.
The gas cylinders should be stored between 2-32 P
oPC (36-90 P
oPF).
Use
Cylinder types
WARNING/ CAUTION:
TNOTE: T
Stability
Storage
7. Solutions and gas mixtures ABL800 FLEX reference manual
7-12
Traceability certificates
ABL800 FLEX reference manual 7. Solutions and gas mixtures
7-13
7. Solutions and gas mixtures ABL800 FLEX reference manual
7-14
ABL800 FLEX reference manual 7. Solutions and gas mixtures
7-15
7. Solutions and gas mixtures ABL800 FLEX reference manual
7-16
ABL800 FLEX reference manual 7. Solutions and gas mixtures
7-17
7. Solutions and gas mixtures ABL800 FLEX reference manual
7-18
ABL800 FLEX Reference Manual Index
Index
A
ABL837/27/17 performance characteristics .................................................................... 5-72 Absorbance ........................................................................................................................ 3-4 Additional information about FLEXMODE .................................................................... 5-48 Altitude Correction .......................................................................................................... 6-52 Amperometric method ....................................................................................................... 2-2
B
Bias .................................................................................................................................... 5-2 BiasBABL B chart description ................................................................................................... 5-7
C
Calibration ..................................................................................................................1-3, 2-3 Calibration line .................................................................................................................. 1-4 Calibration materials...................................................................................................1-7, 2-3 Calibration Solutions ......................................................................................................... 7-3 Continuous spectrum ......................................................................................................... 3-5 Contribution to imprecision specifications from HbF correction .................................... 5-96 Conversion of Units ......................................................................................................... 6-49 Correcting for Interferences............................................................................................... 3-7 Correction Factors for Oximetry Parameters and Bilirubin ............................................... 4-4 Corrrection Factors for Electrolyte and Metabolite Parameters......................................... 4-7 Crea electrodes................................................................................................................. 2-23
Calibration material ..................................................................................................... 2-25 Description .................................................................................................................. 2-23 Drift ............................................................................................................................. 2-28 Measurements and corrections .................................................................................... 2-29 Sensitivity.................................................................................................................... 2-26 Sensitivity limits.......................................................................................................... 2-28 Zero current ................................................................................................................. 2-25
TcTtBil sensitivity for pH changes ....................................................................................... 5-99
D
Default Values ................................................................................................................. 6-51 Defintion of terms.............................................................................................................. 5-2 Derived Parameters.......................................................................................................... 6-17 Determining concentrations ............................................................................................... 3-6 Drift ................................................................................................................................... 1-6
E
Electrolyte electrodes Calibration solution values .......................................................................................... 1-26 Corrections .................................................................................................................. 1-30 Drift ............................................................................................................................. 1-28 Sensitivity.................................................................................................................... 1-27 Stability criteria ........................................................................................................... 1-35 Status ........................................................................................................................... 1-28
Electrolyte Electrodes ...................................................................................................... 1-23 Electrolyte Solutions.......................................................................................................... 7-8
F
TFTHbF sensitivity for pH changes ..................................................................................... 5-97
Index ABL800 FLEX Reference Manual
G
Gas Mixtures (Gas 1 and Gas 2) ...................................................................................... 7-11 Glucose and lactate electrodes ......................................................................................... 2-13
Corrections .................................................................................................................. 2-18 Drift ............................................................................................................................. 2-17 Sensitivity.................................................................................................................... 2-16 Stability criteria ........................................................................................................... 2-21 Zero current ................................................................................................................. 2-14
H
HbF versus HbA ................................................................................................................ 3-7 Hypochlorite Solution...................................................................................................... 7-10
I
Imprecision ........................................................................................................................ 5-3 Imprecision chart ............................................................................................................... 5-8 Input Parameters .............................................................................................................. 6-14 Interference tests
TcTCrea............................................................................................................................ 5-99 Electrolytes.................................................................................................................. 5-92 Metabolites .................................................................................................................. 5-93 Oximetry parameters ................................................................................................... 5-95 pH/blood gas................................................................................................................ 5-92
L
Lambert-Beer’s law ........................................................................................................... 3-4 List of Equations.............................................................................................................. 6-28
M
Matrix of constants ............................................................................................................ 3-6 Mean Corpuscular Hemoglobin Concentration
MCHC ......................................................................................................................... 5-97 Measured parameters ..................................................................................................3-2 ,6-5 Measuring time .................................................................................................................. 1-7
N
Nernst equation.........................................................................................................1-2, 1-25
O
Optical System................................................................................................................... 3-2 Oximetry and bilirubin
Measurement and Corrections ....................................................................................... 3-9 Oxyhemoglobin Dissociation Curve (ODC).................................................................... 6-44
P
Parameters Ranges and limits .......................................................................................................... 6-3 Symbols ......................................................................................................................... 6-2
TpTCO B2B electrode Corrections - blood samples ........................................................................................ 1-18 Corrections - expired air samples ................................................................................ 1-21 Drift ............................................................................................................................. 1-17 Sensitivity.................................................................................................................... 1-17 Stability criteria ........................................................................................................... 1-21 Status ........................................................................................................................... 1-17
TpTCO B2B Electrode ................................................................................................................ 1-15 Performance test results - bilirubin .................................................................................. 5-42 Performance test results – TcTCaP
2+P ...................................................................................... 5-24
Performance test results – TcTClP
–P ........................................................................................ 5-22
ABL800 FLEX Reference Manual Index
Performance test results - TcTCrea....................................................................................... 5-72 Performance test results – TcTGlu ....................................................................................... 5-26 Performance test results – TcTK .......................................................................................... 5-18 Performance test results – TcTLac ....................................................................................... 5-28 Performance test results – TcTNaP
+P ....................................................................................... 5-20
Performance test results – TcTtHb ....................................................................................... 5-30 Performance test results - oximetry ................................................................................. 5-32 Performance test results – TpTCOB2B ...................................................................................... 5-12 Performance test results - pH........................................................................................... 5-10 Performance test results – TpTOB2B ......................................................................................... 5-15 pH electrode
Corrections .................................................................................................................. 1-11 Drift ............................................................................................................................. 1-10 Sensitivity.................................................................................................................... 1-10 Stability criteria ........................................................................................................... 1-14 Status ........................................................................................................................... 1-10
pH Electrode ...................................................................................................................... 1-9 TpTO B2B electrode
Corrections - blood samples .......................................................................................... 2-8 Corrections - expired air samples ................................................................................ 2-11 Drift ............................................................................................................................... 2-6 Sensitivity...................................................................................................................... 2-5 Stability criteria ........................................................................................................... 2-11 Zero point ...................................................................................................................... 2-6
TpTO B2B Electrode..................................................................................................................... 2-4 Potentiometric method....................................................................................................... 1-2
R
Reference Electrode........................................................................................................... 1-8 Repeatability ...................................................................................................................... 5-2 Repeatability chart ............................................................................................................. 5-8 Residual spectrum.............................................................................................................. 3-8 Rinse and Cleaning Solutions ............................................................................................ 7-6
S
Sensitivity .......................................................................................................................... 1-5 Status ................................................................................................................................. 1-6
T
Test conditions – ABL8X5/8X0 ...............................................................................5-6, 5-50 The Deep Picture ............................................................................................................... 6-2 Total absorbance ................................................................................................................ 3-4 Traceability Certificates................................................................................................... 7-15
U
Units and Numerical Format of Derived Parameters ....................................................... 6-22 Updatings........................................................................................................................... 1-7 User-defined corrections.................................................................................................... 4-2
ABL800 FLEX Reference manual Date of issue
Date of issue
Radiometer representative: Manufacturer:
ABL800 FLEX reference manual - from software version 6.00 Publication: 200801 Edition: F
Code number: 989-963
Radiometer Medical ApS Åkandevej 21 2700 Brønshøj Denmark Uwww.radiometer.comU
If you have any questions or need assistance, please contact your local Radiometer representative.
Date of issue ABL800 FLEX Reference manual