7
Biomaterials 27 (2006) 996–1002 Ability of zirconia double coated with titanium and hydroxyapatite to bond to bone under load-bearing conditions Takashi Suzuki a , Shunsuke Fujibayashi a , Yasuaki Nakagawa a , Iwao Noda b , Takashi Nakamura a, a Department of Orthopaedic Surgery, Kyoto University Graduate School of Medicine, Shogoin-Kawaharacho 54, Sakyo-ku, Kyoto 606-8507, Japan b Bioceram Division, Kyocera Corp., Gamocho-Kawai 10-1, Gamogun, Shiga 529-1595, Japan Received 3 January 2005; accepted 21 July 2005 Available online 22 August 2005 Abstract As a preclinical study, we evaluated the ability of hydroxyapatite and titanium on zirconia (HTOZ) to bond to bone under load- bearing conditions in animal experiments. HTOZ, HA, and Ti on Co–Cr alloy (HTOC) and Ti on Co–Cr alloy (TOC) were implanted into the weight-bearing portion of the femoral condyles of nine beagle dogs. Femurs were extracted 4, 12, and 52 weeks after implantation and examined mechanically by pullout testing, and histologically by toluidine blue staining, SEM, and calculation of the affinity index. The interfacial shear strengths (mean7SD) of the HTOZ, HTOC, and TOC groups were 4.4270.453, 3.9070.903, and 4.0870.790 MPa at 4 weeks; 6.8272.64, 6.0071.88, and 6.6371.63 MPa at 12 weeks; and 13.9871.94, 11.9571.51, and 10.7870.83 MPa at 52 weeks. There were no significant differences in the interfacial shear strengths between the three groups at any time. Affinity indices (mean7SD) obtained from SEM images of the HTOZ, HTOC, and TOC groups were 49.676.52%, 43.3710.43%, and 23.773.95% at 4 weeks; 55.076.72%, 51.573.07%, and 28.674.09% at 12 weeks; and 59.176.73%, 63.076.40%, and 34.376.72% at 52 weeks. HA-coated implants (HTOZ, HTOC) had significantly higher affinity indices than non-HA-coated implants (TOC) at all times. HTOZ has the ability to bond to bone equivalent to HTOC and TOC. HTOZ is an excellent material for components of cementless joint prostheses. r 2005 Elsevier Ltd. All rights reserved. Keywords: Zirconia; Cementless; Load-bearing; Total knee arthroplasty 1. Introduction Many ceramic materials have been used for the components of joint prostheses and, compared with metallic materials, using ceramic materials such as zirconia as the bearing parts of joint prostheses can reduce polyethylene wear volume [1–6]. Additionally, in comparison to alumina, zirconia has higher wear resistance [1,2,7], and higher bending strength and fracture toughness [8–10]. Because of their good combination of mechanical properties and excellent biocompatibility [11–13], zirconia ceramics are recog- nized as one of the best biomaterials for joint prostheses. However, metallic cobalt–chromium alloys are the standard materials for the femoral components of total knee arthroplasty, and ceramic femoral components have not been applied as widely in total knee arthro- plasty as in femoral heads for total hip arthroplasty. One of the reasons for this is that there are no ceramic femoral components of the cementless type for total knee replacement. In general, ceramic femoral compo- nents such as zirconia and alumina are secured to the femur using bone cement. In contrast, in cementless fixation, the metal implants are prepared with rough surfaces, obtained by using several coatings or by etching, to achieve the mechanical interlocking required for their anchoring. Moreover, to successfully obtain ARTICLE IN PRESS www.elsevier.com/locate/biomaterials 0142-9612/$ - see front matter r 2005 Elsevier Ltd. All rights reserved. doi:10.1016/j.biomaterials.2005.07.026 Corresponding author. E-mail address: [email protected] (T. Nakamura).

Ability of zirconia double coated with titanium and hydroxyapatite to bond to bone under load-bearing conditions

Embed Size (px)

Citation preview

ARTICLE IN PRESS

0142-9612/$ - se

doi:10.1016/j.bi

�CorrespondE-mail addr

Biomaterials 27 (2006) 996–1002

www.elsevier.com/locate/biomaterials

Ability of zirconia double coated with titanium and hydroxyapatite tobond to bone under load-bearing conditions

Takashi Suzukia, Shunsuke Fujibayashia, Yasuaki Nakagawaa,Iwao Nodab, Takashi Nakamuraa,�

aDepartment of Orthopaedic Surgery, Kyoto University Graduate School of Medicine, Shogoin-Kawaharacho 54, Sakyo-ku, Kyoto 606-8507, JapanbBioceram Division, Kyocera Corp., Gamocho-Kawai 10-1, Gamogun, Shiga 529-1595, Japan

Received 3 January 2005; accepted 21 July 2005

Available online 22 August 2005

Abstract

As a preclinical study, we evaluated the ability of hydroxyapatite and titanium on zirconia (HTOZ) to bond to bone under load-

bearing conditions in animal experiments. HTOZ, HA, and Ti on Co–Cr alloy (HTOC) and Ti on Co–Cr alloy (TOC) were

implanted into the weight-bearing portion of the femoral condyles of nine beagle dogs. Femurs were extracted 4, 12, and 52 weeks

after implantation and examined mechanically by pullout testing, and histologically by toluidine blue staining, SEM, and calculation

of the affinity index. The interfacial shear strengths (mean7SD) of the HTOZ, HTOC, and TOC groups were 4.4270.453,

3.9070.903, and 4.0870.790MPa at 4 weeks; 6.8272.64, 6.0071.88, and 6.6371.63MPa at 12 weeks; and 13.9871.94,

11.9571.51, and 10.7870.83MPa at 52 weeks. There were no significant differences in the interfacial shear strengths between the

three groups at any time. Affinity indices (mean7SD) obtained from SEM images of the HTOZ, HTOC, and TOC groups were

49.676.52%, 43.3710.43%, and 23.773.95% at 4 weeks; 55.076.72%, 51.573.07%, and 28.674.09% at 12 weeks; and

59.176.73%, 63.076.40%, and 34.376.72% at 52 weeks. HA-coated implants (HTOZ, HTOC) had significantly higher affinity

indices than non-HA-coated implants (TOC) at all times. HTOZ has the ability to bond to bone equivalent to HTOC and TOC.

HTOZ is an excellent material for components of cementless joint prostheses.

r 2005 Elsevier Ltd. All rights reserved.

Keywords: Zirconia; Cementless; Load-bearing; Total knee arthroplasty

1. Introduction

Many ceramic materials have been used for thecomponents of joint prostheses and, compared withmetallic materials, using ceramic materials such aszirconia as the bearing parts of joint prostheses canreduce polyethylene wear volume [1–6]. Additionally, incomparison to alumina, zirconia has higher wearresistance [1,2,7], and higher bending strength andfracture toughness [8–10]. Because of their goodcombination of mechanical properties and excellentbiocompatibility [11–13], zirconia ceramics are recog-

e front matter r 2005 Elsevier Ltd. All rights reserved.

omaterials.2005.07.026

ing author.

ess: [email protected] (T. Nakamura).

nized as one of the best biomaterials for joint prostheses.However, metallic cobalt–chromium alloys are thestandard materials for the femoral components of totalknee arthroplasty, and ceramic femoral componentshave not been applied as widely in total knee arthro-plasty as in femoral heads for total hip arthroplasty.One of the reasons for this is that there are no ceramicfemoral components of the cementless type for totalknee replacement. In general, ceramic femoral compo-nents such as zirconia and alumina are secured to thefemur using bone cement. In contrast, in cementlessfixation, the metal implants are prepared with roughsurfaces, obtained by using several coatings or byetching, to achieve the mechanical interlocking requiredfor their anchoring. Moreover, to successfully obtain

ARTICLE IN PRESS

Fig. 2. Cross-section of HTOZ, consisting of zirconia ceramic as

substrate, titanium as the deep coating layer, and hydroxyapatite as

the superficial coating layer.

T. Suzuki et al. / Biomaterials 27 (2006) 996–1002 997

early biological fixation, metal surfaces have beenprepared with HA or bioactive glass coatings. AlthoughHA coating is necessary to ensure secure and earlybiological fixation between the ceramic and thesurrounding bone, HA coating of ceramic implants istechnically difficult.To solve this problem, we have developed a new

composite material for cementless ceramic components:hydroxyapatite and titanium on zirconia (HTOZ),which consists of zirconia ceramic as substrate, titaniumas a deep coating layer, and hydroxyapatite as asuperficial coating layer. The basic mechanical proper-ties and biological safety of this material have beenalready reported [14,15]. As a preclinical study, here weevaluated the ability of HTOZ to bond to bone undersuch load-bearing conditions as occur in the clinicalsituation, using an animal model.

2. Materials and methods

2.1. Implant preparation

Three types of trapezoid-shaped implants (8.5� 8.5�

4.3mm) were prepared: HTOZ, HA and Ti on Co–Cr alloy

(HTOC), and Ti on Co–Cr alloy (TOC) (Fig. 1). HTOZ

consisted of zirconia ceramics stabilized with 3mol% Y2O3,

which satisfies the ISO standard criteria [16], as substrate;

titanium (Ti) as a deep coating layer; and hydroxyapatite (HA)

as a superficial coating layer (Fig. 2). HTOC consisted of

Co–Cr alloy (CoCrMo) as substrate, Ti as a deep coating

layer, and HA as a superficial coating layer. TOC consisted of

Co–Cr alloy as substrate and Ti alone as the coating layer.

The details of the methods of preparation of these implants

are described in Refs. [14,15]. Briefly, the initial roughness

(Rmax) of the zirconia was 1.17 mm, and that of the Co–Cy

alloy was 68.5mm. For the Ti surface treatments, HTOZ,

HTOC, and TOC coatings were performed using the inert gas-

shielded arc spray method. Melted titanium was blown onto

the surface of the substrate by the high-speed carrier gas in the

chamber into which argon gas was introduced. The peak

thickness of the Ti coating was determined to be 500 mm and

the surface roughness (Rmax) was approximately 360 mm. ForHTOZ and HTOC, HA coating of the Ti coating was

performed using the flame spray method, with an acetylene

Fig. 1. The three types of trapezoid-shaped implants. From left to

right: HTOZ, HTOC, and TOC.

and oxygen gas flame, and air used as the high-speed carrier

gas for the spray. The thickness of the HA coating was less

than 50 mm (mean 20 mm).

2.2. Surgical procedure

The implants were conventionally sterilized with ethylene

oxide gas and implanted into the weight-bearing portion of the

femoral condyles of nine adult beagle dogs weighing

9.0–11.0 kg. The surgical methods used have been described

previously [17,18]. Briefly, the dogs were anesthetized with

intramuscular administration of ketamine hydrochloride

(50mg/kg), diazepam (5mg), and atropine sulfate (0.5mg).

Just before the operation, pentobarbital sodium (0.5mL/kg)

was injected intravenously, with local administration of a

solution of 1% lidocaine.

The dogs were placed in the supine position and the right

knee exposed via a medial parapatellar approach in the usual

sterile manner. The implantation sites in the weight-bearing

portion of the medial and lateral femoral condyles were

prepared using a specially designed broach with cutting

surfaces, followed by a surgical electronic drill. During drilling,

the hole was continuously cooled with saline. Just before

insertion of the implants, the hole was irrigated with saline

containing isepamicin sulfate to remove any shards of bone.

An implant was then inserted into the hole by tapping in for

press fitting (Fig. 3a).

Two implants were inserted into each femur. The three types

of implants were inserted in a randomized manner to avoid

any position-related differences. Following irrigation, the

fascia and subcutaneous layers were closed with silk sutures

and the skin was closed with skin staples. The same surgical

procedure was repeated to insert further implants into the left

femoral condyles. No postoperative external immobilization

was applied. Thirty-six implants (12 HTOZ, 12 HTOC, and 12

TOC) were inserted into 18 knees of nine dogs. The animals

were housed individually in standard dog cages and fed

standard dog food and water ad lib. The animals were kept in

cages for 2 weeks after the operation and allowed to move

freely under observation. Both hind legs in each animal were

operated on during the same operation so that the animal

would be obliged to stand on the operated legs. Three dogs

were sacrificed at 4, 12, and 52 weeks after the operation with

an overdose of pentobarbital sodium, and both femurs

ARTICLE IN PRESS

Fig. 3. (a) Implants inserted into the load-bearing portion of both

medial and lateral femoral condyles. (b) AP soft X-ray image of the

femur with a HTOZ implant at 4 weeks after implantation.

T. Suzuki et al. / Biomaterials 27 (2006) 996–1002998

retrieved. Six femurs involving 12 implants (four HTOZ, four

HTOC, and four TOC) were prepared for mechanical tests and

histological examinations at each time point. The Kyoto

University guidelines for animal experiments were observed

throughout this study.

2.3. Mechanical testing

Following euthanasia, all the extracted femurs containing

implants were examined by soft X-ray. At each time point,

nine implants (three HTOZ, three HTOC, and three TOC)

were selected at random from 12 implants for pullout tests and

the other three (one HTOZ, one HTOC, and one TOC) were

used for histological examination only. All implant-containing

bone specimens were kept moist after harvesting. After the

specimens were fixed with resin into a cylindrical container, the

superficial bone and cartilage surrounding each implant were

removed minimally using an electronic surgical drill to allow

connection of the implant and a specially designed hook for

the pullout tests.

After the container was set on the base of an Instron-type

autograph (Instron 1123, Instron Japan Ltd., Tokyo, Japan),

the hook was connected to the implant. The implants were

pulled out from the condyles at a crosshead speed of 2.0mm/

min, taking care to ensure that the line of action of the pullout

force was parallel to the long axis of the implant.

The pullout failure load was measured when the implant

was dislodged from the bone. The failure load values were

divided by the bone–implant interface area to obtain the

interfacial shear strength. After the pullout tests, the extracted

implants with surrounding bone were preserved for histologi-

cal examination to examine the separation sites.

2.4. Histological examination

For the histological and morphological examinations, new

bone formation on the coated surface was evaluated by

toluidine blue staining and light microscopy, scanning electron

microscopy (SEM), and measurement of the affinity index. All

specimens from which the implants had not been pulled out,

and some specimens with the implants pulled out, from each

time point were prepared for histological examinations. These

specimens were fixed in 10% phosphate-buffered formalin for

7 d, embedded in polyester resin, then dehydrated in serial

concentrations of ethanol (50%, 70%, 90%, and 99.5% v/v)

changing every 3 d. Sections 500 mm thick were cut with a band

saw (BS-3000N, EXAKT cutting system, Norderstedt, Ger-

many) across the bone–implant interface. The sections were

then ground to a thickness of 100–150 mm using a grinding–-

sliding machine (Microgrinding MG-4000, EXAKT, Norder-

stedt, Germany). For examinations by light microscopy

(ECLIPSE E600, Nikon Ltd., Tokyo, Japan), sections were

stained with toluidine blue. The remaining sections were

coated with gold by sputtering and examined by SEM

(JSM5410LV, JEOL Ltd., Tokyo, Japan).

For quantitative morphological evaluation of the degree of

direct contact between the coating surface and the new bone,

the affinity indices of the specimens with non-pulled-out

implants at each time point were measured from the SEM

photographs using image analysis software (Winroof, Mitani

Ltd., Tokyo, Japan). To calculate the affinity index, the length

of bone in direct contact with the coated surface with no

intervening soft tissue was divided by the total length of the

coated surface, and this value was multiplied by 100 (Fig. 4).

2.5. Statistical analysis

Data were recorded as mean7standard deviation (SD) and

analyzed using a one-way factorial ANOVA with Fisher’s

ARTICLE IN PRESS

Fig. 4. Measurement of affinity indices: the length of bone in direct

contact with the coated surface was divided by the total length of the

coated surface.

T. Suzuki et al. / Biomaterials 27 (2006) 996–1002 999

PLSD testing as the post hoc test. Differences at po0:05 wereconsidered statistically significant.

Fig. 5. Results of mechanical tests. Interfacial shear strength (MPa)

between bone and implants in pullout tests at 4, 12, and 52 weeks.

There was no statistically significant difference in the interfacial shear

strength between the three groups at any time.

3. Results

3.1. Clinical observation

All animals were able to bear their body weight uponstanding within 1 week. When the animals were allowedto walk outside their cages 2 weeks after the operation,all were able to walk. Although there were no infectionsor postoperative complications in the operated animals,mild joint effusion was noted in some animals. Therange of motion of the knee joints was nearly normal inall animals.

3.2. Radiographic findings

Specimens taken from the animals immediately aftereuthanasia were examined by soft X-ray in bothanteroposterior and lateral views. Soft X-ray imagesshowed no breakage or sinking of the implants at anytime points (4, 12, and 52 weeks). No radiolucent zonearound the implant and no sclerotic changes in thesurrounding bone indicating loosening were observedfor any of the three types of implants at any time point.There was no obvious difference between the HA-coatedimplants (HTOZ, HTOC) and the non-HA-coatedimplants (TOC) in the soft X-ray examination (Fig. 3b).

3.3. Mechanical evaluation

The interfacial shear strengths (mean7SD) ofthe HTOZ, HTOC, and TOC implants, obtained bypullout testing, were 4.4270.453, 3.9070.903, and4.0870.790MPa, respectively, at 4 weeks after implan-tation; 6.8272.64, 6.0071.88, and 6.6371.63MPa at12 weeks; and 13.9871.94, 11.9571.51, and 10.7870.830MPa at 52 weeks. The interfacial shear strengthsincreased with time in all three types of implants, andthese values suggest that the implants were firmly

bonded to bone at each time point. Although thestrength of HTOZ bonding was equivalent to or a littlegreater than that of HTOC and TOC at 4, 12, and 52weeks, there was no significant difference in theinterfacial shear strengths between the three groups atany time points (Fig. 5).

3.4. Histological examination

Histological examinations after the pullout tests showedno separation of the coating layer, and all the separationsoccurred in the bone at the bone–implant interface.In the toluidine blue stained sections, at 4 weeks after

implantation, new bone with thin and irregular trabe-culae was observed on the coated surface of all threetypes of implants. At 12 weeks, the new trabecular bonewas somewhat thicker than that at 4 weeks. At 52 weeks,new bone formation with a lamellar pattern on thecoated surface was noted. The direct contact areabetween the new trabecular bone and the coated surface,and new bone formation adjacent to the implants,increased with time. Through all time points, moredirect contact of bone to the coated surface tended to beobserved with the HA-coated implants than the non-HA-coated implants.SEM images of HTOZ (Fig. 6a), HTOC (Fig. 6b),

and TOC (Fig. 6c) at 4 weeks after implantation showednew bone formation on and along the coated surface,similar to the optical microscopy findings. At 12 weeks,new bone appeared to be a little thicker than at 4 weeksfor all three types of implant. However, more directcontact bone on the coated surface was noted with theHA-coated implants than with the non-HA-coatedimplants at both 4 and 12 weeks (Fig. 6d–f). Althoughnew bone formation on the coated surface at 52 weekswas noted in all three types of implants, the new bonewith the non-HA-coated implants appeared to beslightly poorer than that with the HA-coated implants.

ARTICLE IN PRESS

Fig. 6. SEM images at 4 (a: HTOZ; b: HTOC; c: TOC) and 12 weeks (d: HTOZ; e: HTOC; f: TOC). More direct contact bone on the coated surface

was noted with the HA-coated implants than with the non-HA-coated implants at both 4 and 12 weeks.

T. Suzuki et al. / Biomaterials 27 (2006) 996–10021000

3.5. Morphological examination

Affinity indices (mean7SD) obtained from the SEMimages of the specimens involving HTOZ, HTOC, andTOC implants were 49.676.52%, 43.3710.43%,23.773.95%, respectively, at 4 weeks; 55.076.72%,51.573.07%, 28.674.09% at 12 weeks; and 59.176.73%, 63.076.40%, 34.376.72% at 52 weeks. At alltime points, the HA-coated implants had significantlyhigher affinity indices than the non-HA-coated implants(Fig. 7).

4. Discussion

Ceramic femoral components of the cemented typehave been used in our institute for total knee arthro-plasty since 1989. Akagi [19] reported that the clinical

results of surgery with alumina ceramic implants havebeen good, and no aseptic loosening or breakage of theceramic components was observed. Nakamura [20]documented a further study in which zirconia ceramicswere used instead of alumina ceramics, and showedgood clinical results equivalent to the former study.However, there are no ceramic femoral components ofthe cementless type and thus the components have beenfixed with cement in all cases at our institute. Becauseceramic femoral components of the cementless type wereneeded for several cases, HTOZ was developed and thepresent study was performed.At present, there is no established technique to make

porous coatings on ceramics by which the ceramics candirectory bond to bone. Therefore, ceramic jointprostheses fixed without cement have not been popular,except for the ball head for total hip arthroplasty.Bosetti et al. [21] developed zirconia coated with

ARTICLE IN PRESS

Fig. 7. Affinity indices obtained from SEM images of each implant at

4, 12, and 52 weeks. HA-coated implants (HTOZ and HTOC) have

significantly higher affinity indices than the non-HA-coated implant

(TOC) at all times.

T. Suzuki et al. / Biomaterials 27 (2006) 996–1002 1001

bioactive glass and described how the coating enhancedintegration with bone cells on the surface of the zirconiain vitro. Spector et al. [22] reported that oxidizedzirconium femoral components for total knee arthro-plasty reduced the wear volume of polyethylene in vitro.However, there are few studies investigating whetherporous-coated zirconia can directory bond to bone invivo. Therefore, our study is valuable in the realizationof ceramic femoral components for total knee arthro-plasty.The interfacial shear strengths of all three types of

implants, obtained from the pullout tests in this study,were 3.90–4.42MPa at 4 weeks, 6.00–6.82MPa at 12weeks, and 10.78–13.98MPa at 52 weeks. Compared tothe data from similar experiments in previous studies[17,18], the interfacial shear strengths of all three typesof implants in the present study were higher. Thissuggests that all three types of implants bonded firmly tobone. In addition, because the interfacial shear strengthof HTOZ in this study was higher than those in theprevious studies, and was equivalent to those of HTOCand TOC, HTOZ was considered to have sufficientability to bond to bone.With regard to HA coatings, Thomas et al. [23]

demonstrated that HA-coated implants exhibit signifi-cantly greater interfacial shear strengths than uncoatedimplants in pushout tests after 3, 5, 10, and 32 weeks,and they stated that HA coating provided an osteophilicsurface for bone deposition and allows for a more rapiddevelopment of implant–bone attachment. Conversely,Nakashima and Hayashi reported that HA-coatedimplants had higher interfacial shear strengths inpushout tests than non-HA-coated implants at 4 weeks,although there was no significant difference betweenHA-coated and non-HA-coated implants at 12 weeks.Therefore, they described HA coating as enhancingimplant fixation by direct chemical bonding to boneshortly after implantation due to the bioactivity of HA[24,25].

Although there were significant differences in affinityindices, there was no significant difference in theinterfacial shear strength in the pullout tests betweenHA-coated and non-HA-coated implants at any timepoints in this study. As for the mechanical examinations,it was supposed that the strength of interlocking betweenthe bone and the porous structure of the titaniumcoating might exceed the strength of the chemicalattachment between bone and HA, and thus the shearstrength in the pullout tests would partially depend onthe quality of the surrounding bone around the implants.With respect to the histological evaluation, affinity

indices showed significantly greater direct bone in-growth into the HA-coated implants than into thenon-HA-coated implants at all times. This result is asdescribed in previous studies [24,25], and suggests thatHA coating enhances new bone ingrowth into thesurface of the implants due to the good osteoconductionof HA. There were obvious differences in the histolo-gical evaluation despite there being no significantdifference in the mechanical tests. Therefore, we areconvinced that the HA coating is advantageous to earlydirect bonding between bone and implant.The number of samples in this study was limited for

several reasons. First, because this study was apreclinical study using a canine model, it was moreimportant to show the effectiveness of these implantsand illustrate some trends from their use in a total kneereplacement model than to show significance in statis-tical analysis. Second, because of animal welfareconcerns, it was difficult to increase the number ofexperimental animals. Third, in the quantitative mor-phological evaluation in this study, four sections weretaken of each sample and there was considerablevariation in the data obtained for each section, variationthat might have influenced the significance obtainedfrom statistical analysis.Nakashima and Hayashi [24] indicated that HA

coating should induce early bone ingrowth and thesubsequent fixation should depend upon the mechanicalanchoring strength achieved by the rough surface of theimplant. We believe that Ti coating has a macroporousstructure for interlocking with bone, and HA coatinghas good osteoconduction for early bone ingrowth, andthus, both coatings have important roles in the fixationof cementless prostheses to bone.The results of our study showed that HTOZ has

bonding-to-bone ability equivalent to HTOC and TOC,TOC being already in clinical use. It was important thatHTOZ could firmly bond to bone under load-bearingconditions similar to the clinical situation. We areconvinced that HTOZ is an excellent composite materialfor components of cementless joint prostheses. HTOZfemoral components for total knee arthroplasty areexpected to be used clinically and to contribute toimproved future surgical results.

ARTICLE IN PRESST. Suzuki et al. / Biomaterials 27 (2006) 996–10021002

References

[1] Kumar P, Oka M, Ikeuchi K, Yamamuro T, Okumura H,

Kotoura Y. Wear resistant properties of various prosthetic joint

materials. In: Heimke G, Soltesz U, Lee AJC, editors. Clinical

implant materials. Amsterdam: Elsevier; 1990. p. 373–8.

[2] Kumar P, Oka M, Ikeuchi K, Shimizu K, Yamamuro T,

Okumura H, Kotoura Y. Low wear rate of UHMWPE against

zirconia ceramic (Y-PSZ) in comparison to alumina ceramic and

SUS 316L alloy. J Biomed Mater Res 1991;25:813–28.

[3] Streicher RM, Weber H, Schon R, Semlitsch M. Wear behaviour

of different ceramic surfaces in comparison to TiN and OHD-

treated Ti–6Al–4V alloy paired with polyethylene. In: Vincenzini

P, editor. Ceramics in substitutive and reconstructive surgery.

Amsterdam, The Netherlands: Elsevier; 1991. p. 197–205.

[4] Willmann G, Fruh HJ, Pfaff HG. Wear characteristics of sliding

pairs of zirconia (Y-TZP) for hip endoprostheses. Biomaterials

1996;17:2157–62.

[5] Oonishi H, Ueno M, Okimatsu H, Amino H. Investigation on the

tribological behavior of ceramic on ceramic combination in total

hip prostheses. In: Kokubo T, Nakamura T, Miyaji F, editors.

Bioceramics, vol. 9. Amsterdam: Elsevier; 1996. p. 503–6.

[6] Ben Abdallah A, Treheux T. Friction and wear of ultra high

molecular weight polyethylene against various new ceramics.

Wear 1991;142:43–56.

[7] Oka M, Kumar P, Ikeuchi K, Yamamuro T, Nakamura T. Low

wear rate of UHMWPE against zirconia. In: Yamamuro T,

Kokubo T, Nakamura T, editors. Bioceramics, vol. 5. Kyoto,

Japan: Koubunnshi kankokai, Inc; 1992. p. 373–9.

[8] Covacci V, Bruzzese N, Maccauro G, Andreassi C, Ricci GA,

Piconi C, Marmo E, Burger W, Cittadini A. In vitro evaluation of

the mutagenic and carcinogenic power of high purity zirconia

ceramic. Biomaterials 1999;20:371–6.

[9] Richter HG, Burger W, Willmann G. Ceramic hip joint heads

made from alumina and zirconia—a comparison. Ceram Trans

1995;48:73–81.

[10] Christel P, Meunier A, Heller M, Torre JP, Peille CN. Mechanical

prosperities and short time in vivo evaluation of yttrium

oxide partially stabilized zirconia. J Biomed Mater Res 1989;23:

45–61.

[11] Li J, Liu Y, Hermansson L, So R. Evaluation of biocompatibility

of various ceramic powders with human fibroblasts in vitro. Clin

Mater 1993;12:197–201.

[12] Ito A, Tateishi T, Niwa S, Tange S. In vitro evaluation of the

cytocompatibility of wear particles generated by UHMWPE

zirconia friction. Clin Mater 1993;12:203–9.

[13] Dion I, Rouais F, Baquey C, Lahaye M, Salmon R, Trut L,

Cazorla JP, Huong PV, Monties JR, Havlik P. Physico-chemistry

and cytotoxicity of ceramics, Part II. J Mater Sci Mater Med

1994;5:18–24.

[14] Noda I, Masuda S, Kitano H, Ikeda J, Yoshihara Y, Mukai M,

Shimotoso T. New approach to a cementless ceramic component

for artificial knee joint. Key Eng Mater 2002;218–20:577–80.

[15] Fujisawa A, Noda I, Nishino Y, Okimatsu H. The development

of new titanium arc-sprayed artificial joints. Mater Sci Eng

1995;C2:151–7.

[16] ISO TC 150/SC 1. Implants for surgery-ceramic materials based

on yttria-stabilized tetragonal zirconia (Y-TZP). ISO/DIS 13356,

1995.

[17] Takagi H, Yamamuro T, Hyakuna K, Nakamura T. Bone

bonding behavior of bead-coated alumina ceramic under load-

bearing conditions. J Biomed Mater Res 1989;23:161–81.

[18] Li ZL, Kitsugi T, Yamamuro T, Chang YS, Senaha Y, Takagi H,

Nakamura T, Oka M. Bone-bonding behavior under load-bearing

conditions of an alumina ceramic implant incorporating beads

coated with glass-ceramic containing apatite and wollastonite.

J Biomed Mater Res 1995;29:1081–8.

[19] Akagi M, Nakamura T, Matsusue Y, Ueo T, Nishijyo K, Ohnishi

E. The Bisurface total knee replacement: a unique design for

flexion. Four-to-nine-year follow-up study. J Bone Joint Surg Am

2000;82:1626–33.

[20] Nakamura T, Akagi M, Yasuda T, Nakagawa Y, Shimizu M. A

new knee prosthesis with bisurface femoral component made of

zirconia-ceramics. Key Eng Mater 2002;218–20:563–6.

[21] Bosetti M, Verne E, Ferraris M, Ravaglioli A, Cannas M. In vitro

characterisation of zirconia coated by bioactive glass. Biomater-

ials 2001;22:987–94.

[22] Spector M, Ries MD, Bourne RB, Sauer WS, Long M, Hunter G.

Wear performance of ultra-high molecular weight polyethylene on

oxidized zirconium total knee femoral components. J Bone Joint

Surg Am 2001;83:S80–6.

[23] Thomas KA, Kay JF, Cook SD, Jarcho M. The effect of surface

macrotexture and hydroxylapatite coating on the mechanical

strengths and histologic profiles of titanium implant materials.

J Biomed Mater Res 1987;21:1395–414.

[24] Nakashima Y, Hayashi K, Inadome T, Uenoyama K, Hara T,

Kanemaru T, Sugioka Y, Noda I. Hydroxyapatite-coating on

titanium arc sprayed titanium implants. J Biomed Mater Res

1997;35:287–98.

[25] Hayashi K, Mashima T, Uenoyama K. The effect of hydro-

xyapatite coating on bony ingrowth into grooved titanium

implants. Biomaterials 1999;20:111–9.