4
[email protected] www.jenway.com Tel: 01785 810433 Fluorimeter 6285 Application note: A10-006A Determination of chlorophyll a using fluorimetry Introduction The photosynthetic pigment chlorophyll is present in most plants, algae and cyanobacteria. It can be measured both by spectrophotometry and fluorimetry as an indicator of the abundance of photosynthetic organisms in fresh and salt water. In fresh water, levels of chlorophyll are also an important factor in determining water quality. Chlorophyll pigments may be present in several forms in varying ratios, the most common being chlorophyll a. Here we demonstrate two methods of determining chlorophyll a using the Jenway 6285 fluorimeter based on EPO method 445.0 (1). We also show how the presence of chlorophyll b can affect the results. In addition we demonstrate that the model 6285 can detect as little as 0.01μg/l chlorophyll a. Methods Chlorophyll a and b from spinach were purchased from Sigma, part codes C5753 and C5878. The contents of the vials, containing 1mg of pure chlorophyll, were each dissolved in 50ml of 90% acetone (spectrophotometric grade, Sigma 154598) to give 20mg/l stock solutions. These were stored in aliquots in the dark at -20ºC until required. Stock solutions of chlorophylls a and b were diluted in 90% acetone to give solutions of 1mg/l which were determined spectrophotometrically using the formulae of Lichtenthaler and Wellburn (2). Briefly, the absorbance of each dilution was measured in a Jenway model 6505 spectrophotometer at the absorbance maxima for chlorophylls a and b (662.6nm and 645.6nm respectively). The formulae given below were used to calculate the chlorophyll concentrations. Ca = (11.75 x A662.6) – (2.35 x A645.6) Cb = (18.61 x A645.6) – (3.96 A662.6) where Ca and Cb are the concentrations of chlorophyll a and b respectively. The solutions were adjusted until they reached 1mg/l. Photosynthetic pigments were also extracted from a sample of slime algae isolated from a domestic marine fish tank. The algae were ground in a mortar containing ceramic beads with 90% acetone. The extract was filtered through Whatman No. 1 filter paper into a volumetric flask and made up to a volume of 50ml with 90% acetone. The extract was stored in aliquots in the dark at -20ºC until required. The two methods described in EPO method 445.0 are as follows: a. “Corrected chlorophyll adetermination, measuring the fluorescence before and after acidification of the sample. This uses broad excitation and emission filters. The filters used in the model 6285 for this assay were BG28 380- 500nm band pass filter (part code 627 124) for excitation and Kodak 29, 610nm cut-off filter (part code 627 127) for emission. b. “Uncorrected chlorophyll a” determination which uses narrow band pass filters to eliminate most of the spectral overlap with pheopyhtin a and chlorophyll b. The filters used in the model 6285 for this assay were 435nm interference filter (part code 627 165) for excitation and 680nm interference filter (part code 627 193) for emission. Results The detection limit of the model 6285 fluorimeter was first determined by serially diluting the stock chlorophyll a solution until it could no longer be detected above the blank (90% acetone). The fluorimeter was set up with the narrow band pass filters, 435nm and 680nm and the gain set to 100% for maximum sensitivity. Each sample was diluted in triplicate and 2.5ml placed in a glass fluorimeter cuvette (part code 060 254). The results are presented in Table 1. Concentration (μg/l) Average (RFU) SD (± RFU) 10 34.180 0.155 1.0 3.971 0.045 0.1 0.949 0.011 0.05 0.940 0.007 0.01 0.715 0.008 0 0.655 0.006 Table 1: Relative fluorescence unit (RFU) values for various dilutions of chlorophyll a. It can be seen from the results that 0.01μg/l chlorophyll a can be clearly distinguished from the blank by greater than 3 standard deviations from the mean, demonstrating the very high sensitivity of the instrument.

A10_006A_Chlorophyll.pdf

Embed Size (px)

Citation preview

  • [email protected] www.jenway.com

    Tel: 01785 810433

    Fluorimeter 6285 Application note: A10-006A

    Determination of chlorophyll a using fluorimetry

    Introduction The photosynthetic pigment chlorophyll is present in most plants, algae and cyanobacteria. It can be measured both by spectrophotometry and fluorimetry as an indicator of the abundance of photosynthetic organisms in fresh and salt water. In fresh water, levels of chlorophyll are also an important factor in determining water quality. Chlorophyll pigments may be present in several forms in varying ratios, the most common being chlorophyll a. Here we demonstrate two methods of determining chlorophyll a using the Jenway 6285 fluorimeter based on EPO method 445.0 (1). We also show how the presence of chlorophyll b can affect the results. In addition we demonstrate that the model 6285 can detect as little as 0.01g/l chlorophyll a.

    Methods Chlorophyll a and b from spinach were purchased from Sigma, part codes C5753 and C5878. The contents of the vials, containing 1mg of pure chlorophyll, were each dissolved in 50ml of 90% acetone (spectrophotometric grade, Sigma 154598) to give 20mg/l stock solutions. These were stored in aliquots in the dark at -20C until required.

    Stock solutions of chlorophylls a and b were diluted in 90% acetone to give solutions of 1mg/l which were determined spectrophotometrically using the formulae of Lichtenthaler and Wellburn (2). Briefly, the absorbance of each dilution was measured in a Jenway model 6505 spectrophotometer at the absorbance maxima for chlorophylls a and b (662.6nm and 645.6nm respectively). The formulae given below were used to calculate the chlorophyll concentrations.

    Ca = (11.75 x A662.6) (2.35 x A645.6)

    Cb = (18.61 x A645.6) (3.96 A662.6)

    where Ca and Cb are the concentrations of chlorophyll a and b respectively. The solutions were adjusted until they reached 1mg/l.

    Photosynthetic pigments were also extracted from a sample of slime algae isolated from a domestic marine fish tank. The algae were ground in a mortar containing ceramic beads with 90% acetone. The extract was filtered through Whatman No. 1 filter paper into a volumetric flask and made up to a volume of 50ml with 90% acetone. The extract was stored in aliquots in the dark at -20C until required.

    The two methods described in EPO method 445.0 are as follows:

    a. Corrected chlorophyll a determination, measuring the fluorescence before and after acidification of the sample. This uses broad excitation and emission filters. The filters used in the model 6285 for this assay were BG28 380-500nm band pass filter (part code 627 124) for excitation and Kodak 29, 610nm cut-off filter (part code 627 127) for emission.

    b. Uncorrected chlorophyll a determination which uses narrow band pass filters to eliminate most of the spectral overlap with pheopyhtin a and chlorophyll b. The filters used in the model 6285 for this assay were 435nm interference filter (part code 627 165) for excitation and 680nm interference filter (part code 627 193) for emission.

    Results The detection limit of the model 6285 fluorimeter was first determined by serially diluting the stock chlorophyll a solution until it could no longer be detected above the blank (90% acetone). The fluorimeter was set up with the narrow band pass filters, 435nm and 680nm and the gain set to 100% for maximum sensitivity. Each sample was diluted in triplicate and 2.5ml placed in a glass fluorimeter cuvette (part code 060 254). The results are presented in Table 1.

    Concentration (g/l)

    Average (RFU)

    SD ( RFU) 10 34.180 0.155 1.0 3.971 0.045 0.1 0.949 0.011 0.05 0.940 0.007 0.01 0.715 0.008

    0 0.655 0.006

    Table 1: Relative fluorescence unit (RFU) values for various dilutions of chlorophyll a.

    It can be seen from the results that 0.01g/l chlorophyll a can be clearly distinguished from the blank by greater than 3 standard deviations from the mean, demonstrating the very high sensitivity of the instrument.

  • [email protected] www.jenway.com

    Tel: 01785 810433

    Next, a set of chlorophyll a standards were prepared ranging from 0.2g/l to 200g/l to determine the linear dynamic range of the fluorimeter. Essentially this is to determine a suitable gain setting where all the standards fit on a linear calibration curve, such that the top standard is not more or less than 10% from the calculated concentration based on the linear regression equation of the line. The RFU values for each of the standards were read at the gain as set by the auto gain function and at three lower gain settings. This test was done using both sets of filters. The data was plotted for the four lowest standards and, using the equation of the line, the theoretical value for the 200g/l standard was calculated. The results are presented in Tables 2 and 3.

    Std conc. (g/l)

    RFU 52%

    RFU 50%

    RFU 45%

    RFU 40%

    0 0.038 0.027 0.012 0.005 0.2 0.174 0.134 0.064 0.027 2.0 1.295 0.991 0.472 0.205 5.0 3.245 2.487 1.185 0.516 20 13.22 10.16 4.852 2.126 200 112.1 90.67 47.31 21.33

    Calc. conc. (g/l)

    169.87 178.73 195.26 200.87

    R2 value 0.9997 0.9999 1 1

    Table 2: Test of linearity at different gain settings using the BG28 and Kodak 29 filters.

    Std conc. (g/l)

    RFU 81%

    RFU 75%

    RFU 70%

    RFU 65%

    0 0.145 0.085 0.050 0.027 0.2 0.271 0.159 0.094 0.052 2.0 1.413 0.818 0.506 0.299 5.0 3.350 1.947 1.204 0.722 20 13.26 7.757 4.803 2.870 200 110.6 71.25 46.14 28.30

    Calc. conc. (g/l)

    168.11 185.02 193.41 198.34

    R2 value 0.9997 0.9999 1 1

    Table 3: Test of linearity at different gain settings using the 435nm and 680nm interference filters.

    For the BG28 and Kodak 29 filters, the 45% gain setting gave a good linear relationship over the whole standard range with the calculated top standard fluorescence within 2.5% of the expected calculated value. This gain setting was therefore used in following experiments. Likewise, the 70% gain setting was used with the narrow band pass filters, with the top standard within 3.5% of the calculated value. The standard curves produced with both sets of filters are shown in Figure 1.

    From the gradient of these lines, the response factor described in EPO method 445.0 can be derived. The response factor, Fs, is equivalent to 1/gradient of the standard curve at the chosen gain setting, S.

    Once the appropriate gain settings had been determined, the effects of the presence of chlorophyll b in the solution were investigated using both the corrected and uncorrected chlorophyll a methods.

    Figure 1: Chlorophyll a standards curves measured using the BG28 and Kodak 29 filters (red) and 435nm and 680nm interference filters (blue).

    The conventional method of chlorophyll a determination described in EPO method 445.0 relies on acidification of the sample to correct for the contribution of any pheophytin a (the magnesium-free derivative of chlorophyll a) present in the samples. Spectral overlap of pheophytin a and chlorophyll b with chlorophyll a can lead to errors in the estimation of chlorophyll a.

    A number of mixtures were prepared containing different ratios of chlorophyll a and b; the total amount of chlorophyll remained the same in each case. Table 4 lists the composition of the mixtures. A 1 in 1000 dilution of the slime algae extract was also measured.

    Sample Chl a conc. (g/l)

    Chl b conc. (g/l)

    Vol. 20g/l Chl a (ml)

    Vol. 20g/l Chl b (ml)

    Vol. 90%

    acetone (ml)

    A 10 0 2 0 2.0 B 9 1 1.8 0.2 2.0 C 8 2 1.6 0.4 2.0 D 6 4 1.2 0.8 2.0 E 5 5 1.0 1.0 2.0 F 0 10 0 2.0 2.0

    Table 4: Mixtures of chlorophyll a and b tested.

    For the corrected method, which is based on acidification of the solution, the fluorescence of a set of standard dilutions plus a number of mixtures of chlorophyll a and b were measured before and 90 seconds after the addition of 0.1N HCl to a final concentration of 0.003N. Briefly, 2.5ml of each sample were placed in a glass fluorimeter cuvette and the fluorescence determined. 75l of 0.1N HCl was then added to the cuvette, mixed and incubated for 90s before reading the fluorescence again.

    y = 0.2304x + 0.076R2 = 1

    y = 0.2364x + 0.029R2 = 1

    05

    101520253035404550

    0 50 100 150 200

    Concentration (ug/l)

    RFU

  • [email protected] www.jenway.com

    Tel: 01785 810433

    Using a standard solution of chlorophyll a, a before-to-after acidification response ratio, r, can be determined:

    r = Rb/Ra where:

    Rb = Fluorescence of pure chlorophyll a standard solution before acidification

    Ra = Fluorescence of pure chlorophyll a standard solution after acidification

    The acidification response ratio for the mid-range chlorophyll a standards (2, 5 and 20g/l) averaged at 1.59. Table 5 gives the calculated results for uncorrected chlorophyll a, corrected chlorophyll a and pheophytin a, based on the following formulae:

    Uncorrected chlorophyll a:

    Cu = Rb x Fs where:

    Cu = Uncorrected chlorophyll a concentration (g/l)

    Rb = Fluorescence of the sample solution before acidification

    Fs = Fluorescence response factor at gain setting S

    Corrected chlorophyll a:

    Cc = Fs(r/r-1) (Rb Ra) where:

    Cc = Corrected chlorophyll a concentration (g/l)

    Fs = Fluorescence response factor at gain setting S

    r = The before-to-after acidification response ratio of a pure chlorophyll a solution

    Rb = Fluorescence of the sample solution before acidification

    Ra = Fluorescence of the sample solution after acidification

    Pheophytin a (P, g/l):

    P = Fs(r/r-1) (rRa Rb)

    Table 5 illustrates as expected that the uncorrected values (Cu) for the standard solutions are very close to the concentrations of the standard dilutions. For samples A to F where the total chlorophyll concentration remained at 10g/l, the values did drop slightly with increasing amount of chlorophyll b, since the fluorescence of chlorophyll b can be seen to be

    only about 44% that of the equivalent concentration of chlorophyll a.

    On studying the corrected values (Cc) it can be seen that the top standard (200 g/l) appears to contain a large proportion of pheophytin a; this could indicate some degradation of the sample. It is also clearer from these results the presence of the added chlorophyll b in the mixed samples A to F. Chlorophyll b is not degraded at such a rapid rate as chlorophyll a in the presence of acid, therefore the before-to-after ratio is not as great. In these experiments, the amount of chlorophyll b present is being measured as an overestimation of the pheophytin a concentration.

    Sample Rb (RFU)

    Ra (RFU)

    Cu (g/l)

    Cc (g/l)

    P (g/l)

    0 g/l 0.012 0.012 0.049 -0.005 0.086 0.2 g/l 0.066 0.046 0.270 0.222 0.077 2.0 g/l 0.474 0.301 1.945 1.907 0.060 5.0 g/l 1.192 0.750 4.892 4.884 0.012 20 g/l 4.881 3.051 20.027 20.237 -0.333

    200 g/l 47.49 31.69 194.86 174.72 32.028 A 2.441 1.531 10.02 10.06 -0.076 B 2.478 1.596 10.17 9.753 0.658 C 2.363 1.568 9.698 8.794 1.437 D 2.195 1.518 9.008 7.487 2.417 E 1.969 1.438 8.079 5.864 3.521 F 1.082 1.042 4.440 0.443 6.356

    Slime algae

    4.257 2.962 17.47 14.32 5.011

    Table 5: Fluorescence of standards and samples before and after acidification and the calculated uncorrected and corrected chlorophyll a and pheophytin a concentrations

    Figure 2 shows the correlation of chlorophyll a concentration in samples A to F compared to the corrected chlorophyll a value. Also in the graph, chlorophyll b is correlated to the calculated amount of pheophytin a.

    Figure 2: Correlation of diluted chlorophyll a and b solutions with the corrected values based on the fluorescence readings before and after acidification.

    Using these formulae, the sample extracted from the marine slime algae was calculated to have a ratio of approximately 1:0.35 chlorophyll a to pheophytin

    0

    2

    4

    6

    8

    10

    12

    0 2 4 6 8 10

    Chlorophyll concentration (ug/l)

    "Co

    rrec

    ted"

    co

    nce

    ntr

    atio

    n

    Chl a Chl b

  • [email protected] www.jenway.com

    Tel: 01785 810433

    a/chlorophyll b and an undiluted chlorophyll a concentration of 14.32mg/l.

    An alternative to the acidification assay is to measure chlorophyll a using narrow band pass interference filters. If this method is used, then only the uncorrected chlorophyll a is calculated. As for the acidification assay, a series of standards and mixtures of chlorophylls a and b were measured using these filters at the 70% gain setting as determined previously. The chlorophyll a concentrations were calculated using the formula for uncorrected chlorophyll a and the results are shown in Table 6.

    Sample RFU Cu (g/l) 0 g/l chl a 0.050 0.21

    0.2 g/l chl a 0.094 0.39 2.0 g/l chl a 0.506 2.12 5.0 g/l chl a 1.204 5.05 20 g/l chl a 4.803 20.14

    200 g/l chl a 46.14 193.5 10 g/l chl a 2.379 9.98

    10 g/l chl a + 2.5g/l chl b 2.411 10.11 10 g/l chl a + 5.0g/l chl b 2.418 10.14 10 g/l chl a + 10g/l chl b 2.870 12.03

    Slime algae 4.742 19.89

    Table 6: RFU values and derived chlorophyll a concentrations calculated from the formula for uncorrected chlorophyll a.

    Using this method, the presence of added chlorophyll b in the sample leads only to very little chlorophyll a overestimation. At a 2:1 ratio of chlorophyll a to b, the amount is overestimated by only 1.6%, however if the ratio is increased to 1:1 (the highest likely to be found naturally) then the overestimation is around 20%.

    Conclusions The Jenway model 6285 fluorimeter has a red-enhanced PMT detector which allows the detection of molecules which fluoresce at wavelengths up to 850nm. Using this instrument, we have demonstrated it is possible to detect chlorophyll a down to a concentration as low as 0.01g/l with a large linear dynamic range, making it several orders of magnitude more sensitive than a spectrophotometer.

    Based on methods described in EPO method 445.0, chlorophyll a can be determined using either narrow band pass interference filters allowing an uncorrected chlorophyll a calculation, or by using wider band pass filters and including an acidification step. In the former method, we have shown that chlorophyll b can be present up to as much as 30% of the total chlorophyll and have a minimal effect on chlorophyll a estimation.

    The acidification method has the advantage in that the amount of pheophytin a and/or chlorophyll b can also be estimated. With the narrow band pass filters, although a large amount of chlorophyll b was added, this could not be measured. With the acidification method the results calculated for pheophytin a correlated well with the amount of chlorophyll b spiked into the sample. Since there is a large degree of spectral overlap between chlorophyll b and pheophytin a, it would not possible to distinguish between them in natural samples.

    In summary, the model 6285 fluorimeter is an extremely sensitive instrument ideal for the estimation of chlorophyll a in the study of marine and freshwater algae as described in EPO method 445.0.

    References (1) EPA Method 445.0: In vitro determination of

    chlorophyll a and pheophytin a in marine and freshwater algae by fluorescence. Arar, E. J. and Collins, G. B. (1997).

    (2) Lichtenthaler, H.K., and Wellburn, A.R., Determination of total carotenoids and chlorophylls a and b of leaf in different solvents. Biol. Soc. Trans. 11: 591-592 (1983).