1
References [1] Linder et al.; Bioanalysis (2015) 1(16) 2013-2039 [2] Kearney et al.; BMJ. (2006) 332(7553):1302-8 [3] Shah et al.; Nutrition Journal (2011) 10:4 Results - Pipeting blood Whole blood is a difficult liquid class of highly variable viscosity (defined by hematocrit value). To pipet blood reproducibly and accurately adjustment factors are needed. However, in the typ- ical hematocrit range (30 - 50%) an average correction is sufficient to correct for that variability. Automated Sample Prepraration Task and cleanining Procedure Whole blood consists of water, salt, proteins, and intact erythrocytes, which tend to coagulate on surfaces or in contact with organic solvents. Therefore, it is important to wash properly to avoid precipitation in the system as well as the deposition of biofilms. Different wash steps applying two different solvents were applied to reduce potential carryover. However, an additional step using a protein removal detergent (Decon™ Contrad™ 70) was used to remove remaining proteins. Results - Examples for Therapeutic Drug Monitoring (TDM) The LCMS-P 80 Tool was used to prepare samples for several typical drugs for TDM as well as compounds relevant in diagnostics. Spiked and blank whole blood samples were analyzed. Carbamazepine Carbamazepine is a representative of an anti-epileptic drug and very important to monitor in blood to set a proper and indi- vidual dosage. [1] Method: Eluent A: Water 0.1% Formic acid / Eluent B: Methanol 0.1% Formic acid // Flow: 0.45 ml/min // Runtime: 6 min Column: Waters Coretecs C18 2.1x100 mm 2.7 µm MS Parameter: Mode: positive MRM: 237>194 Blood Volume: 10 µL / Injection volume: 2 µL Diclophenac Diclophenac is a nonsteroidal anti-inflammatory drug and a representative compound for drug monitoring acording its molec- ulare properties [2] Method: Eluent A: Water 2 mM NaF / Eluent B: Acetonitril 2 mM NaF // Flow: 0.5 ml/min // Runtime: 7 min Column: Zorbax SB-C8 4.6x50 mm 1.8 µm MS Parameter: Mode: positive MRM: 296>249 Blood Volume: 10 µL / Injection volume: 2 µL 25 Hydroxyvitamin D3 25 hydroxy vitamn D3 is the important maker for the storage depot of vitamin D [3] Method: Eluent A: Water 0.1% Formic acid / Eluent B: Methanol 0.1% Formic acid // Flow: 0.45 ml/min // Runtime: 6 min Column: Waters Coretecs C18 2.1x100 mm 2.7 µm MS Parameter: Mode: positive MRM: 401>383 Blood Volume: 10 µL / Injection volume: 20 µL Materials and Methods PAL RTC autosampler with a 1 mL syringe, LC MS P80 injection tool, vortex mixer, centrifuge. Agilent LC 1200, Agilent MS 6460 TripleQ-MS Tested laboratory blood was supplied by blood donation in primary containers (Vacuette™, Vacutainer™, etc.). All chemicals wer purchased from Sigma-Aldrich. Hematocrit levels were prepared by mixing defined ratios of plasma and erythrocytes. Christian Berchtold 1 ,Guenter Boehm 2 , Renée Falsia 2 , Thomas Preiswerk 2 , Götz Schlotterbeck 1 1 University of Applied Sciences and Arts Northwestern Switzerland FHNW, Institute for Chemistry and Bioanalytics , Gründenstrasse 40, 4132 Muttenz, Switzerland 2 CTC Analytics AG; Industriestrasse 20, 4222 Zwingen, Switzerland A New Tool for the Automated Sample Preparation of Whole Blood Samples by LC-MS using a Commercial Autosampler Introduction Automated sample preparation reduces the costs per sample and minimizes sample handling errors. The use of robots is well established for therapeutic drug monitoring or diagnostics based on blood samples. Usually expensive and highly specialized pipetting robots are used. However, most of these systems are not designed with a direct interface for LC-MS applica- tions. In addition common pipetting systems are not optimized for smaller scale sample series. Here we present a new tool for liquid handling of whole blood samples and direct sample injec- tion. The tool features an optical sensor that monitors all liquid handling steps. It detects pres- ence or absence of sample, standards and reagents. The sensor is essential to ensure process safety for automated liquid handling steps. Poster reprint and info: [email protected] Conclusions - The new LCMS-P 80 Tool is capable of pipetting whole blood with normal hematocrit levels (30-50%) directly from primary tubes with good accuracy and precision, applying one average correction factor. For high hematocrit levels a different correction factor should be applied. - Wash procedures were established to minimize carryover of analytes. A procedure to mini- mize the deposition of blood on system surfaces proofed successful. - The system consisting of a commercial PAL RTC autosampler, connected to an Agilent LC- MS system can perform sample prep and LC-MS analysis for whole blood samples from primary tube. The performance of the system was tested with several typical compounds in therapeutically relevant concentrations. - Future work is aimed at evaluating the capability of the sensor to reliably detect certain error conditions. Blood of different hematocrit levels was pipetted and weighed. The volume was determined using the experimentally deter- mined density of the used blood. The red line corresponds to the nominal volume to be pipetted (50 µL). Each bar represents n=9 measurements. The tool pipettes 1 to 70 µL of blood with good precision (<5% CV, @ 40% hematocrit). The volume was determined gravimetrically using the experimental determined density of the used blood. The effect of the individual wash steps was followed with carbamazepine as marker. Each wash solution was collected and analyzed separatly after a transfer of 10 µL of a carbamazepinesolution of 0.5 mg/mL. The first wash by solvent 1 (H2O, 0.95 g/L NaN3) was defined as 100% peak area. Solvent 2 (H2O/MeOH/ACN/2-Prop 1:1:1:1, step 6.) wash reduces carryover further. Sample Transfer Clean Tool Wash with detergent Transfer crashing solvent Clean Tool Vortex, Centrifue Inject aliquote Clean tool injector Next Sample y = 0.9603x R² = 0.9996 0 10 20 30 40 50 60 70 80 0 20 40 60 80 µg measured µL set 45 46 47 48 49 50 Water Serum (hematocrit =0) hematocrit 30% hematocrit 40 % hematocrit 70 % Volume measured [µL] 13 mm Primary Tube 75 - 100 mm Correct Transfer No sample Clot Foam 2 sensor transitions 1 2 Sensor Sample Tube Rack Cold Storage Vortex Mixer 2 D Barcode Reader UHPLC Injection Valve Fast Wash Station Centrifuge, 8 x 2 mL max. RCF = 5000 LCMS-P80 Tool The PAL RTC (instrument view below) can perform additional sample prep steps, e.g. standard and reagent addition, centifugation, liquid / liquid extraction (LLE). The LCMS-P 80 Tool features a special needle that allows direct sampling from primary tubes. With special adapters it can transport primary tubes to and from modules on the system, e.g. for vortexing. Different adapt- ers allow the use of different formats of primary tubes. The optical sensor positioned on the sample loop (orange triangles in the sketch above) of the tool moitores the aspiration of sample, air gap or bubbles. Different sampling situations (injec- tion with air gaps, clogging, bubbles) generate distinct patterns of the sensor signal triggering defined actions of the system. 1 1000 1000000 1000000000 1000000000000 Peak Area 0. Carbamazepine 0.5 mg/mL 1. Solvent 1 - FastWash 1 2. Solvent 1 - NeedleWash 1 3. Solvent 1 - F astWash 2 4. Solvent 1 - NeedleWash 2 5. Solvent 2 - F astWash 1 6. Solvent 2 - NeedleWash 1 7. Solvent 2 - F astWash 2 8. Solvent 2 - NeedleWash 2 9. Solvent 1 - F astWash 3 10. Solvent 1 - NeedleWash 3 1. Solvent 1 2. Solvent 1 3. Solvent 1 4. Solvent 1 5. Solvent 2 6. Solvent 2 7. Solvent 2 8. Solvent 2 9. Solvent 1 10. Solvent 1 10 µg/mL carbamazepine 233 ng/mL diclophenac 80 ng/mL 25 hydroxy vitamin D3 LOD of the method approx. 2 ng/mL Needle of LCMS Tool Blank Blank Blank LCMS-P 80 Tool

A New Tool for the Automated Sample Preparation of Whole

  • Upload
    others

  • View
    4

  • Download
    0

Embed Size (px)

Citation preview

Page 1: A New Tool for the Automated Sample Preparation of Whole

References[1] Linder et al.; Bioanalysis (2015) 1(16) 2013-2039[2] Kearney et al.; BMJ. (2006) 332(7553):1302-8[3] Shah et al.; Nutrition Journal (2011) 10:4

Results - Pipeting bloodWhole blood is a difficult liquid class of highly variable viscosity (defined by hematocrit value). To pipet blood reproducibly and accurately adjustment factors are needed. However, in the typ-ical hematocrit range (30 - 50%) an average correction is sufficient to correct for that variability.

Automated Sample Prepraration Task and cleanining ProcedureWhole blood consists of water, salt, proteins, and intact erythrocytes, which tend to coagulate on surfaces or in contact with organic solvents. Therefore, it is important to wash properly to avoid precipitation in the system as well as the deposition of biofilms. Different wash steps applying two different solvents were applied to reduce potential carryover. However, an additional step using a protein removal detergent (Decon™ Contrad™ 70) was used to remove remaining proteins.

Results - Examples for Therapeutic Drug Monitoring (TDM)The LCMS-P 80 Tool was used to prepare samples for several typical drugs for TDM as well as compounds relevant in diagnostics. Spiked and blank whole blood samples were analyzed.

CarbamazepineCarbamazepine is a representative of an anti-epileptic drug and very important to monitor in blood to set a proper and indi-vidual dosage. [1]Method: Eluent A: Water 0.1% Formic acid / Eluent B: Methanol 0.1% Formic acid // Flow: 0.45 ml/min // Runtime: 6 min

Column: Waters Coretecs C18 2.1x100 mm 2.7 µm

MS Parameter: Mode: positive MRM: 237>194

Blood Volume: 10 µL / Injection volume: 2 µL

DiclophenacDiclophenac is a nonsteroidal anti-inflammatory drug and a representative compound for drug monitoring acording its molec-ulare properties [2]

Method: Eluent A: Water 2 mM NaF / Eluent B: Acetonitril 2 mM NaF // Flow: 0.5 ml/min // Runtime: 7 min

Column: Zorbax SB-C8 4.6x50 mm 1.8 µm

MS Parameter: Mode: positive MRM: 296>249

Blood Volume: 10 µL / Injection volume: 2 µL

25 Hydroxyvitamin D325 hydroxy vitamn D3 is the important maker for the storage depot of vitamin D [3]Method: Eluent A: Water 0.1% Formic acid / Eluent B: Methanol 0.1% Formic acid // Flow: 0.45 ml/min // Runtime: 6 min

Column: Waters Coretecs C18 2.1x100 mm 2.7 µm

MS Parameter: Mode: positive MRM: 401>383

Blood Volume: 10 µL / Injection volume: 20 µL

Materials and MethodsPAL RTC autosampler with a 1 mL syringe, LC MS P80 injection tool, vortex mixer, centrifuge.

Agilent LC 1200, Agilent MS 6460 TripleQ-MS

Tested laboratory blood was supplied by blood donation in primary containers (Vacuette™, Vacutainer™, etc.). All chemicals wer purchased from Sigma-Aldrich.

Hematocrit levels were prepared by mixing defined ratios of plasma and erythrocytes.

Christian Berchtold1,Guenter Boehm2, Renée Falsia2, Thomas Preiswerk2, Götz Schlotterbeck1

1 University of Applied Sciences and Arts Northwestern Switzerland FHNW, Institute for Chemistry and Bioanalytics , Gründenstrasse 40, 4132 Muttenz, Switzerland 2 CTC Analytics AG; Industriestrasse 20, 4222 Zwingen, Switzerland

A New Tool for the Automated Sample Preparation of Whole Blood Samples by LC-MS using a Commercial Autosampler

Introduction

Automated sample preparation reduces the costs per sample and minimizes sample handling errors. The use of robots is well established for therapeutic drug monitoring or diagnostics based on blood samples. Usually expensive and highly specialized pipetting robots are used. However, most of these systems are not designed with a direct interface for LC-MS applica-tions. In addition common pipetting systems are not optimized for smaller scale sample series.

Here we present a new tool for liquid handling of whole blood samples and direct sample injec-tion. The tool features an optical sensor that monitors all liquid handling steps. It detects pres-ence or absence of sample, standards and reagents. The sensor is essential to ensure process safety for automated liquid handling steps.

Poster reprint and info:[email protected]

Conclusions- The new LCMS-P 80 Tool is capable of pipetting whole blood with normal hematocrit levels

(30-50%) directly from primary tubes with good accuracy and precision, applying one average correction factor. For high hematocrit levels a different correction factor should be applied.

- Wash procedures were established to minimize carryover of analytes. A procedure to mini-mize the deposition of blood on system surfaces proofed successful.

- The system consisting of a commercial PAL RTC autosampler, connected to an Agilent LC-MS system can perform sample prep and LC-MS analysis for whole blood samples from primary tube. The performance of the system was tested with several typical compounds in therapeutically relevant concentrations.

- Future work is aimed at evaluating the capability of the sensor to reliably detect certain error conditions.

Blood of different hematocrit levels was pipetted and weighed. The volume was determined using the experimentally deter-mined density of the used blood. The red line corresponds to the nominal volume to be pipetted (50 µL). Each bar represents n=9 measurements.

The tool pipettes 1 to 70 µL of blood with good precision (<5% CV, @ 40% hematocrit). The volume was determined gravimetrically using the experimental determined density of the used blood.

The effect of the individual wash steps was followed with carbamazepine as marker. Each wash solution was collected and analyzed separatly after a transfer of 10 µL of a carbamazepinesolution of 0.5 mg/mL. The first wash by solvent 1 (H2O, 0.95 g/L NaN3) was defined as 100% peak area. Solvent 2 (H2O/MeOH/ACN/2-Prop 1:1:1:1, step 6.) wash reduces carryover further.

Sample Transfer Clean Tool Wash with

detergent

Transfer crashingsolvent

Clean ToolVortex,

CentrifueInject

aliquoteClean tool

injector Next Sample

y = 0.9603xR² = 0.9996

01020304050607080

0 20 40 60 80

µg

mea

sure

d

µL set 45

46

47

48

49

50

51

52

Water Serum(hematocrit =0)

hematocrit 30% hematocrit 40%

hematocrit 70%

Volu

me

mea

sure

d [µ

L]

13 mm

Prim

ary T

ube

75 -

100 m

m

Correct Transfer No sample Clot Foam

2 sensor transitions

1

2

Sensor

Sample Tube Rack

Cold Storage Vortex Mixer

2 D Barcode Reader

UHPLC Injection Valve

Fast Wash Station

Centrifuge, 8 x 2 mL max. RCF = 5000

LCMS-P80 Tool

The PAL RTC (instrument view below) can perform additional sample prep steps, e.g. standard and reagent addition, centifugation, liquid / liquid extraction (LLE). The LCMS-P 80 Tool features a special needle that allows direct sampling from primary tubes. With special adapters it can transport primary tubes to and from modules on the system, e.g. for vortexing. Different adapt-ers allow the use of different formats of primary tubes.

The optical sensor positioned on the sample loop (orange triangles in the sketch above) of the tool moitores the aspiration of sample, air gap or bubbles. Different sampling situations (injec-tion with air gaps, clogging, bubbles) generate distinct patterns of the sensor signal triggering defined actions of the system.

1

1000

1000000

1000000000

1000000000000

Peak

Are

a

0. Carb

amazepine 0.5 m

g/mL

1. Solvent 1

- FastW

ash 1

2. Solvent 1

- NeedleWash

1

3. Solvent 1

- FastW

ash 2

4. Solvent 1

- NeedleWash

2

5. Solvent 2

- FastW

ash 1

6. Solvent 2

- NeedleWash

1

7. Solvent 2

- FastW

ash 2

8. Solvent 2

- NeedleWash

2

9. Solvent 1

- FastW

ash 3

10. Solvent 1

- NeedleWash

3

1.Solvent 1

2.Solvent 1

3.Solvent 1

4.Solvent 1

5.Solvent 2

6.Solvent 2

7.Solvent 2

8.Solvent 2

9.Solvent 1

10.Solvent 1

10 µg/mL carbamazepine

233 ng/mL diclophenac

80 ng/mL 25 hydroxy vitamin D3 LOD of the method approx. 2 ng/mL

Needle of LCMS Tool

Blank

Blank

Blank

LCMS-P 80 Tool