A New Bone Conductor with Improved Performance Authors: Eric Burwood (NAL); George Raicevich (NAL); Harvey Dillon (NAL); Carit Laursen (Ortofon) Acknowledgement: This project was a collaboration between the National Acoustic Laboratories (NAL), a division of Australian Hearing and Ortofon, Denmark. Introduction: The most commonly used bone conductor in Australia is based on a design that dates back several decades; however, this device suffers from limitations of user comfort and speech distortion. Aim: To design a bone vibrator motor that can be used in a binaural bone conductor hearing aid with improved distortion and comfort; and without excessive decrease in performance; and that is safer than current models. Background Information: Bone conduction hearing aids have not been comfortable devices to wear and anecdotal reports indicate that continuous excessive pressure on the skin can produce skin ulceration. However, for people with middle-ear disorders, including those with active effusion, transmission of sound to the cochlea via bone conduction rather than air conduction via the ear canal can provide more effective stimulation, unaffected by fluctuation in conductive loss, and avoid exacerbating active middle ear infection. In Australia, the predominant need is for Indigenous people, especially children, because of their high incidence of middle ear infection, often with chronic perforation of the eardrum. Hearing aids that enable children with chronic middle ear disease to hear well in school are an important step to improve educational outcomes for these children, and long-term their living conditions and health. Method: • Produce design specifications. • Investigate the limitations of performance and comfort for existing bone conductor transducers. • Locate a commercial partner to design and manufacture the bone conductor. • Investigate different designs. • Investigate how the type and size of battery and headband force affects the output of a bone conductor. Conclusion: The outcome of this project will be a bone conductor that is comfortable to wear, cosmetically acceptable, has low distortion, enables binaural fittings, and has an effective output for practical use. The new bone conductor motor will enable children with chronic middle ear disease to hear better in school. Reference 1: Raicevich G, Burwood E, Dillon H, 2008. Taking the pressure off bone conduction hearing aid users. The Australian and New Zealand Journal of Audiology, Vol. 30 No.2, pp 113-118. Reference 2: Håkansson B, 2003. The balanced electromagnetic separation transducer: A new bone conduction transducer. J.Acoust. Soc. Am. 113(2), pp 818-825. Reference 3: ISO 389-3:1994, Acoustics - Reference zero for the calibration of audiometric equipment; Part 3: Reference equivalent threshold force levels for pure tones and bone vibrators. Reference 4: Byrne D, Dillon H, Tran K, et.al., 1994. An international comparison of long-term average speech spectra. J. Acoust. Soc. Am. 96(4), pp 2108-2120. A commercial partner, Ortofon, was located in Denmark that had design and manufacturing experience in small precision electromechanical transducers. The button size zinc air batteries can limit the force output of the bone conductor for some wire sizes used in a transducer. Results : Investigations at NAL found that the most used bone conductor was fitted with a headband pressure sufficient to close small capillaries in the skin (Ref 1). This caused discomfort to the user and long term use may damage the skin. Headband force must be selected to allow the maximum perceived signal level. The area of the footplate must be selected to provide safe operating pressure on the head resulting in Settle on BEST bone conductor design by Bo Håkansson (Ref 2). Ortofon optimised motor design to meet performance requirements. Display shows the bone conductor output for the new Ortofon device and for the existing BC461device when powered by a size 675 zinc air battery; Reference Equivalent Threshold Force Level (RETFL) (Ref 3) and speech (Ref 4) at 65 dB SPL with a 6 dB per octave filter between 250 and 2000 Hertz. BEST Bone conductoroutput fordifferentbatteries 40 60 80 100 120 140 100 1000 10000 Frequency (H ertz) F o rce (d B re 1 m icro -N ew ton Alkaline C ell Zinc A ir675 Zinc A ir13 B one C onductorPerform ance 20 40 60 80 100 120 140 100 1000 10000 Frequency (H ertz) Force Level(dB re 1 uN O rtofon B C 675 Zinc A irC ell C om m ercial B C 675 Zinc A irC ell S peech 65dB S P L -6dB /O ct250 to 2000 H z RETFL Volt-A m pere Pow er for D ifferent C ells w ith w ire at30 ºC 0.1 1.0 10.0 100.0 1000.0 0 50 100 150 N om inalW ire D iam eter(μm ) Volt-A m pere Pow er (m VA) Alkaline 675 Zinc Air 13 Zinc Air

A New Bone Conductor with Improved Performance Authors: Eric Burwood (NAL); George Raicevich (NAL); Harvey Dillon (NAL); Carit Laursen (Ortofon) Acknowledgement:

Embed Size (px)

Citation preview

Page 1: A New Bone Conductor with Improved Performance Authors: Eric Burwood (NAL); George Raicevich (NAL); Harvey Dillon (NAL); Carit Laursen (Ortofon) Acknowledgement:

A New Bone Conductor with Improved PerformanceAuthors: Eric Burwood (NAL); George Raicevich (NAL); Harvey Dillon (NAL); Carit Laursen (Ortofon)

Acknowledgement: This project was a collaboration between the National Acoustic Laboratories (NAL), a division of Australian Hearing and Ortofon, Denmark.

Introduction: The most commonly used bone conductor in Australia is based on a design that dates back several decades; however, this device suffers from limitations of user comfort and speech distortion.

Aim: To design a bone vibrator motor that can be used in a binaural bone conductor hearing aid with improved distortion and comfort; and without excessive decrease in performance; and that is safer than current models.

Background Information:Bone conduction hearing aids have not been comfortable devices to wear and anecdotal reports indicate that continuous excessive pressure on the skin can produce skin ulceration. However, for people with middle-ear disorders, including those with active effusion, transmission of sound to the cochlea via bone conduction rather than air conduction via the ear canal can provide more effective stimulation, unaffected by fluctuation in conductive loss, and avoid exacerbating active middle ear infection. In Australia, the predominant need is for Indigenous people, especially children, because of their high incidence of middle ear infection, often with chronic perforation of the eardrum. Hearing aids that enable children with chronic middle ear disease to hear well in school are an important step to improve educational outcomes for these children, and long-term their living conditions and health.

Method:• Produce design specifications.• Investigate the limitations of performance and comfort for existing bone conductor transducers.• Locate a commercial partner to design and manufacture the bone conductor.• Investigate different designs. • Investigate how the type and size of battery and headband force affects the output of a bone conductor.

Conclusion:The outcome of this project will be a bone conductor that is comfortable to wear, cosmetically acceptable, has low distortion, enables binaural fittings, and has an effective output for practical use. The new bone conductor motor will enable children with chronic middle ear disease to hear better in school.

Reference 1: Raicevich G, Burwood E, Dillon H, 2008. Taking the pressure off bone conduction hearing aid users. The Australian and New Zealand Journal of Audiology, Vol. 30 No.2, pp 113-118.

Reference 2: Håkansson B, 2003. The balanced electromagnetic separation transducer: A new bone conduction transducer. J.Acoust. Soc. Am. 113(2), pp 818-825.

Reference 3: ISO 389-3:1994, Acoustics - Reference zero for the calibration of audiometric equipment; Part 3: Reference equivalent threshold force levels for pure tones and bone vibrators.

Reference 4: Byrne D, Dillon H, Tran K, et.al., 1994. An international comparison of long-term average speech spectra. J. Acoust. Soc. Am. 96(4), pp 2108-2120.

A commercial partner, Ortofon, was located in Denmark that had design and manufacturing experience in small precision electromechanical transducers.

The button size zinc air batteries can limit the force output of the bone conductor for some wire sizes used in a transducer.

Results:

Investigations at NAL found that the most used bone conductor was fitted with a headband pressure sufficient to close small capillaries in the skin (Ref 1). This caused discomfort to the user and long term use may damage the skin. Headband force must be selected to allow the maximum perceived signal level. The area of the footplate must be selected to provide safe operating pressure on the head resulting in a comfortable bone conductor to wear.

Settle on BEST bone conductor design by Bo Håkansson (Ref 2). Ortofon optimised motor design to meet performance requirements.

Display shows the bone conductor output for the new Ortofon device and for the existing BC461device when powered by a size 675 zinc air battery; Reference Equivalent Threshold Force Level (RETFL) (Ref 3) and speech (Ref 4) at 65 dB SPL with a 6 dB per octave filter between 250 and 2000 Hertz.

BEST Bone conductor output for different batteries

40

60

80

100

120

140

100 1000 10000Frequency (Hertz)

Fo

rce (

dB

re 1

mic

ro -

New

ton

)

Alkaline CellZinc Air 675Zinc Air 13

Bone Conductor Performance

20

40

60

80

100

120

140

100 1000 10000

Frequency (Hertz)

Fo

rce

Le

ve

l (d

B r

e 1

uN

)

Ortofon BC 675 Zinc Air Cell

Commercial BC675 Zinc Air Cell

Speech 65dB SPL-6dB/Oct 250 to2000 HzRETFL Volt-Ampere Power for Different

Cells with wire at 30 ºC

0.1

1.0

10.0

100.0

1000.0

0 50 100 150Nominal Wire Diameter (μm)

Vo

lt-A

mp

ere

P

ow

er

(mV

A)

Alkaline

675 Zinc Air

13 Zinc Air