24
varLble-frequency sine-wave generator w ith a uniform pressure amplitude and waveform response at varying frequen cies is difficult [9,13,28] and requires the use of a variable-temperature pressure chamber and a temperature-compensated reference transducer. Although certain re searchers have experimentally validated the second-order system theoretical mod els for the resonant behavior of a few intra-arterial blood pressure measurement systems [21, 22] and validated a more complex electrical transmission-line m odel [9], other researchers have quanti tatively shown such theoretical models not to be adequate for their intra-arterial catheters [29-31], Additionally, the simi larity o f responses obtained with square- wave response techniques compared to those obtained from the sine-wave gen erator technique is shown to be both ade quate [32,33] and inadequate [12,31] for intra-arterial catheler-manomcter sys tems. Extrapolating the assumption or a second-order system (shown to be valid in certain fluid-filled calheter-manomelers) to fluid-filled catheters of different FG sizes, lengths, and material compositions, operated under different conditions and coupled to modern pressure transducers and domes, may be erroneous. Our results suggest a signi ficant devia tion of the fluid-filled PVC catheter- manometers’ SOFCs from the ORCs. Mathematical models, from which com monly used equations based on systems assumed to be second order are derived, state that the catheter-manometer acts as a transmission probe (based on electrical transmission line theory) in the manner of a distributed system [9, 15-17], These theories, and the equations based on them, incorporate several important catheter-manometer properties influenc ing the theoretical frequency responses: compliance (catheter-manometer walls and filling-fluid), inertance (axial and ra dial fluid mass, and radial catheter- manometer wall inertance), viscous re sistance (catheter-manometer Tiling fluid), and hysteresis (catneter- manometcr wall hysteresis and filling fluid inelasticity). One obvious differ ence between fluid-filled feeding cathe ters used to measure inlra-esophageal pressure and catheters used to measure intra-arterial pressure is the soft PVC catheter compound of the former versus the relatively rigid nylon or braided polyurethane of the latter. Whereas wall compliance and hysteresis are validly ig nored in the mathematical models ap plied to certain intra-arterial catheters (by assuming a tube o f constant dyi .cm. c compliance, with negligible hysteresis losses directly proportional to frequency [9]), PVC NGCs are palpably compliant and may exhibit significant dynamic en ergy losses through hysteresis. M oreo ver, in certain catheter-manometer systems compliance has been shown to be frequency dependent [21, 29], as in fluenced by catheter wall composition., PVC eatheter.compliance and hysteresis may not only be greater in magni tude, but also be more frequency and amplitude dependent than those o f relatively rigid intra-arterial lines assumed to be linear systems [9,17,21 ]. We have not assessed the contribution of such factors here, Theory predicts that a gas-filled catheter-manometer system may be treated as being linear if the catheter ra dius is large enough and the frequency low enough, despite the large compressi bility of the gas [17], In highly com pressible fluids, such as a gas, the catheter-manometer system’s com pli ance is principally affected by the gas compressibility in A-BCs, whereas catheter elasticity and transducer- diaphragm compliance is primarily re sponsible for the elastancc o f W-FCs. In our experiment, even though deviation from second-order systems was less in A-BCs than W-FCs, we did not find an adequate second-order system fit pre vailing within the range o f clinically used A-I3C FG sizes and lengths. The I 80.6 ° -8 i T I ▲ Uquld#od Catholets O Air-balloon Cathelora aJagS aPortO aVentO aPorS aVenS aPord aVen6 wVenS wVyg4 wPorlO wVenlO wPor8 wVenB wPorB wVenS wAdcS wPor4 Catheter 1 20 15 10 5 0 -5 -10 -15 -20 J.tlllgd Calholers alloon Calhotors ‘ I aJag8 aPorlO aVenlO aPorS aVenS aPor6 aVenS wVenS wVyg4 wPorlO wVenlO wPorB wVenB wPorB wVenB wAdcS wPor4 Catheter 3. Phase-shift difference (mean ±SD) between second-order fitted curve and original recorded curve (SOFC-ORC) for phase-frequoncy response trace variables. X-scale break separates commercial Jaeger adult from laboratory-made air-balloon cathe ter. SOFC-ORC of F as: SOFC-ORC phasc-shlfl difference at A = 1.05 (or/As); SOFC-ORC of Pfr: SOFC-ORC phase-shift difference at damped resonance fre quency. Catheter: catheter brand and French Gauge size; aPor or wPor: air-balloon or water-filled Portcx catheter. * Significant SOFC-ORC phase-shift difference. + P < 0.05 for SOFC-ORC phase-shift difference between A-BC and W-FC of same FG size and brand, Moy/Junc 1997 IEEE EHGIIIEERIIIG III EDIEIIIE AND 8101009 31

‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

  • Upload
    others

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

Page 1: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

va rL b le -fre q u e n cy sine-wave generator w ith a u n ifo rm pressure am plitude and w ave fo rm response at va ry ing frequen­c ies is d if f ic u lt [9 ,1 3 ,2 8 ] and requires the use o f a variab le-tem perature pressure cham ber and a temperature-compensated re ference transducer. A lthough certain re­searchers have experim enta lly validated the second-order system theoretical m od­e ls fo r the resonant behavior o f a few in tra -a rte ria l b lood pressure measurement system s [21, 22 ] and validated a more c o m p le x e le c tr ic a l tra n s m is s io n -lin e m o de l [9 ], o ther researchers have quanti­ta tiv e ly shown such theoretical models n o t to be adequate fo r the ir intra-arteria l catheters [29 -31], A d d itio n a lly , the s im i­la r ity o f responses obtained w ith square- w a ve response techniques compared to

those obta ined from the sine-wave gen­erator technique is shown to be both ade­quate [3 2 ,3 3 ] and inadequate [1 2 ,3 1 ] fo r in tra -a rte r ia l ca the le r-m anom cte r sys­tems. E xtrapo la ting the assumption o r a second-order system (shown to be va lid in certa in f lu id - f i l le d calheter-manomelers) to f lu id - f i l le d catheters o f d iffe ren t FG sizes, lengths, and m aterial com positions, operated under d iffe re n t conditions and coupled to m odern pressure transducers and domes, m ay be erroneous.

O ur results suggest a s ign i ficant devia­tio n o f the f lu id - f i l le d P V C catheter- m anom eters ’ SOFCs fro m the ORCs. M athem atica l models, from w hich com ­m on ly used equations based on systems assumed to be second order are derived, state that the catheter-m anom eter acts as a

transm ission probe (based on e le c tr ic a l transm ission line theory) in the m anner o f a d is trib u te d system [9 , 15-17], These theories, and the equa tions based on them , in c o rp o ra te severa l im p o r ta n t catheter-m anom eter p roperties in f lu e n c ­ing the theoretica l frequency responses: com pliance (ca the ter-m anom eter w a lls and f i l l in g - f lu id ) , inertance (ax ia l and ra­d ia l f lu id mass, and ra d ia l ca th e te r- m anom eter w a ll inertance), v iscous re ­sis tance (c a th e te r-m a n o m e te r T i l in g f l u id ) , a n d h y s te r e s is ( c a tn e te r - m anom etcr w a ll hysteresis and f i l l in g flu id ine las tic ity ). One obvious d if fe r ­ence between f lu id - f i l le d feed ing ca the­ters used to measure in lra -esophagea l pressure and catheters used to measure in tra -a rte ria l pressure is the so ft P V C catheter com pound o f the fo rm er versus the re la t iv e ly r ig id n y lo n o r b ra id e d polyurethane o f the latter. W hereas w a ll com pliance and hysteresis are v a lid ly ig ­nored in the m athem atical m odels ap ­plied to certa in in tra -a rte ria l catheters (by assum ing a tube o f constant d y i .cm. c com pliance, w ith n e g lig ib le hysteresis losses d ire c tly p ropo rtiona l to frequency[9 ]), P V C N G Cs are pa lpab ly c o m p lia n t and may e x h ib it s ig n ifica n t dynam ic en ­ergy losses th rough hysteresis. M o re o ­ve r, in c e r ta in c a th e te r -m a n o m e te r systems com pliance has been show n to be frequency dependent [21, 2 9 ], as in ­fluenced by catheter w a ll co m po s itio n ., PVC eatheter.com pliance and hysteresis may not on ly be greater in m agni tude, b u t also be m ore frequency and a m p litude dependent than those o f re la tiv e ly r ig id intra-arte ria l lines assumed to be lin e a r systems [9 ,1 7 ,2 1 ]. W e have not assessed the co n tribu tion o f such fac to rs here,

T h e o ry p re d ic ts th a t a g a s - f i l le d ca th e te r-m a n o m e te r sys te m m a y be treated as be ing linea r i f the ca the ter ra ­dius is large enough and the frequency low enough, despite the large com press i­b il i ty o f the gas [17 ], In h ig h ly c o m ­p ress ib le f lu id s , such as a gas, the ca the ter-m anom eter sys te m ’s c o m p li­ance is p r in c ip a lly a ffected by the gas c o m p re s s ib il i ty in A -B C s , w h e re a s c a th e te r e la s t ic i ty and tra n s d u c e r - diaphragm com pliance is p r im a r ily re ­sponsible fo r the elastancc o f W -F C s. In our experim ent, even though d e v ia tio n from second-order systems was less in A -B C s than W -FCs, we d id not f in d an adequate second-order system f i t p re ­va iling w ith in the range o f c lin ic a lly used A - I3 C FG s iz e s and le n g th s . T h e

I8 0 . 6

° -8

i T I

▲ Uquld#od Catholets O Air-balloon Cathelora

aJagS aPortO aVentO aPorS aVenS aPord aVen6 wVenS wVyg4 wPorlO wVenlO wPor8 wVenB wPorB wVenS wAdcS wPor4

Catheter

1

20

1510

50

-5-10-15-20

J.tlllgd Calholers alloon Calhotors

‘ IaJag8 aPorlO aVenlO aPorS aVenS aPor6 aVenS wVenS wVyg4

wPorlO wVenlO wPorB wVenB wPorB wVenB wAdcS wPor4

Catheter

3. Phase-shift difference (mean ±SD) between second-order fitted curve and original recorded curve (SOFC-ORC) for phase-frequoncy response trace variables. X-scale break separates commercial Jaeger adult from laboratory-made air-balloon cathe­ter. SOFC-ORC of F as: SOFC-ORC phasc-shlfl difference at A = 1.05 (or/As); SOFC-ORC of Pfr: SOFC-ORC phase-shift difference at damped resonance fre­quency. Catheter: catheter brand and French Gauge size; aPor or wPor: air-balloon or water-filled Portcx catheter. * Significant SOFC-ORC phase-shift difference. + P < 0.05 for SOFC-ORC phase-shift difference between A-BC and W -FC of same FG size and brand,

Moy/Junc 1997 IEEE EHGIIIEERIIIG III EDIEIIIE AND 8101009 31

Page 2: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

S O FC :O R C ratios o f our W -FCs d e v i­ated s ig n if ic a n t ly m ore than the A -B C s, even though the W -F C s f i l l in g f lu id was o f lo w e r c o m p re s s ib ility . T h is may be related to the presence o f gas m icrohub - b lcs in the W -F C s, w h ic h rem ained de­s p ite o u r e f fo r t s to re d u c e th e m . Trapped gas bubbles, w h ich lo w e r the ca the ie r-m anom ete r system elastance, rem ain an adverse fea ture o f W -FC s in the c l in ic a l s itu a tio n . T h e ir fo rm a tion is e n c o u ra g e d by th e n e g a tiv e in t r a - e s o p h a g e a l p re s s u re s e n c o u n te re d p h y s io lo g ic a l ly . W e s im u la te d these p h y s io lo g ic a lly neg a tive pressures in v i t r o in th is e x p e r im e n t u s in g the la b o ra to ry -b u ilt s ine -w ave generator.

O u r f in d in g s m ust be in te rp re te d W ith in the context o f c lin ic a l relevance. The energy contents o f m ost p hys io log i­cal pressure w ave fo rm s lie w e ll w ith in / r, b u t n o t n e c e s s a rily w ith in /a s o f a catheter-m anom eter system. S O F C O R C d ev ia ti ms and S O FC -O R C phase-shift d iffe rences o f / r, A r, and Pfr are not as c l in ic a lly re levant as the devia tions o f /a s . In o u r experim ent, the m agnitude o f S O F C :O R C a m p litu d e -fre q u e n c y re ­sponse dev ia tion is shown to be greater for/AS than f o r / . T h is fin d in g is un favor­able because devia tions observed f o r / in our expe rim e n t are c lin ic a lly acceptable, w h ile those observed a t/A s are c lin ic a lly re levan t (an overestim ation o f the u n i­f o r m i t y o f a c a th e te r ’ s a m p litu d e - frequency. response is unacceptable) as w e ll 3s s ta tis tic a lly s ign ifica n t. A d d i­t io n a lly , the observed m agnitude o f de­v ia tio n is greater fo r W -FCs than fo r A -B C s . T h is resu lt is also un favorab le fo r c l in ic ia n s because the in v ivo fast- flu sh test (square-w ave response tech­n ique ), w h ich is re lia n t on second-order system theory to p re d ic t /a s [1 2 ], can c u rre n tly be perfo rm ed on W -FC s o n ly and n o t on A -B C s i/i vivo. O u r results suggest th a t assum ing a second-order system is less v a lid fo r W -FCs than fo r A -B C s .

T h e observed S O F C -O R C phase- s h if t d iffe re n c e s are s ta tis tic a lly s ig ­n if ic a n t, bu t re la t iv e ly sm a ll in m a gn i­tu d e . In ira -c s o p h a g c a l p re s s u re is m easured to e s tim a te d y n a m ic lun g co m p lia n c e , the re fo re a s im u ltaneous d y n a m ic lu n g -v o lu m e m easurem ent is a lw a y s re c o rd e d . T l ic S O F C -O R C p h a s e -s h ift d if fe re n c e s may thus be c l in ic a l ly re levan t.

W e noted s ign ifica n tly increased mag­

nitudes o f S O F C :O R C ra tios w ith in ­c re as in g c a th e te r FG s ize . C a the te r elastance, w h ich decreases w ith larger ca the ter FG sizes (w a ll th ickness:ou te r d iam ete r ra tio [ 13) reduces w ith increas­ing FG size in o u r catheters), may have in fluenced th is . In v a lid a tin g the theo­re tica l assum ptions o f th e ir m a them ati­cal m odels, m o s t researchers have used o n ly o i\e sam ple o f each ca the te r fo r the e x p e r im e n ta lly o b ta in e d (s in e -w a v e generator) frequency response and com ­pared th is a ga ins t the th e o re tica l pre­d ic te d re s p o n s e . W e n o te d la rg e variances in S O F C :O R C ra tios w ith in both the A -B C s and the W -F C s samples fn = 8 each in o u r expe rim e n t); large m anu fac tu ring to lerances fo r any par­tic u la r ca the ter FG size m ay be an in f lu ­encing fa c to r here. N ote, how ever, that d if fe r in g ca the te r brands d id n o t s ig n if i­ca n tly a lte r the S O F C :O R C devia tions fo r any p a rt ic u la r FG -s ize catheter, and the c o m m e rc ia lly a v a ila b le A - B C s S O F C iO R C d e v ia tio n was n o t s ig n if i­ca n tly d if fe re n t fro m the co rrespond ing labo ra to ry-m ade A -B C s.

In c o n c lu s io n , A -B C and W -F C fre ­quency responses do no t adequate ly f i t s e c o n d -o rd e r sys te m s. T h is f in d in g casts doub t on the v a lid ity o f d e fin in g a nd a p p ly in g s e c o n d -o rd e r sys te m m a them a tica l m ode ls to p re d ic t such c a th e te r-m a n o m e te rs ' fre q u e n c y re ­sponses. S O F C :O R C d e v ia tio n s are s ig n if ic a n t ly la rg e r in W -F C s than in A -B C s , in la rg e F G -s ize d ca the ters, and at / a s than at / . T h e frequency- response s y s te m o rd e r o f d if fe r in g f l u i d - f i l l e d c a th e te r - m a n o m e te rs shou ld be v e r if ie d e x p e r im e n ta lly p r io r to p re d ic tin g th e ir fre q u en cy responses us ing the ty p ic a l equa tions a pp lied to square-w ave response techn iques. W e recom m end th a t the s in e -w a ve genera­to r, ra the r than the sq ua re -w ave re­sponse te ch n iq u e , boused to m easure m vitro frequency responses o f c o m p lia n t f lu id - f i l le d catheters.

AcknowledgmentG rants p ro v id e d fro m the South A f ­

rican F o u n da tio n fo r Research and D e­ve lopm ent, M e d ic a l Research C o u n c il, Levenberg B equest, U n iv e rs ity o f the W itw a te rs ra n d , and the M e d ic a l Fac­u lty Research C o u n c il. C a the te r dona­tio n s p rov id e d b y E rich -Jaeger, S e lw yn S lie r E le c tro M e d S .A . ; P o r te x , the

Boots Com pany S.A. and Portex U .K .; V en, V en Products S .A .; V y g o n , A r t M edica l Equipm ent S. A .; and Sabax, A d ­cock Ingram S.A.

C raig G. H a rtfo rd was b o rn in C ape T o w n . South A fr ic a . He re ­ceived h is M B BC h de­gree at the U n iv e rs ity o f the W itw a te rsrand in Johannesburg in 1987, a nd g a in e d f u r t h e r c lin ica l tra in in g in aca­

dem ic hosp ita l un its u n t il 1990. In 1993. he received h is M S c M ed by d isserta tion in card iovascular p hys io log y . He c u r­re n t ly lec tu re s in the d e p a rtm e n t o f P hys io logy and C hem ical E ng ineering at the U n iv e rs ity o f the W itw a te rs rand in Johannesburg. In 1996, he w on the IE E E E M B S S tudent W h ita ke r paper co m pe ti­tion . H e cu rren tly is a b io m e d ica l eng i­n e e r in g P h D s tu d e n t in te re s te d in p u lm o n a ry c o m p lia n ce and d y n a m ic pressure measurements.

Johan M . van Schalk- wyk acqu ired h is m e d i­cal degree in 1983 at the U n iv e rs ity o f the W itw a te rs ra n d in Jo ­hannesburg, South A f ­r ic a , a n d r e c e iv e d F e llow sh ip o f the C o l­lege o f P h ys ic ians o f

South A fr ic a in 1990. He was registered., in the sub-specia lty o f c r it ic a l care in 1995 and is cu rren tly a senio r consu ltan t in the Post-operative In tensive Care U n it under the departm ent o f Anaesthesia at Johannesburg H osp ita l. H is interests are com puter p rogram m ing (p a rt ic u la rly as­se m b ly language p ro g ra m m in g ) , o r ­thogonal language design, and character recogn ition .

G e o ff G. R ogers was born in South A fr ic a and com p le ted h is BSc m a jo r in g in p h y s io l­ogy at the U n iv e rs ity o f the W itw a te rs ran d , Jo ­h a n n e s b u rg . H e o b ­ta ined h is M S c at the U n iv e rs ity o f P re to ria

and h is P hD at the U n iv e rs ity o f the W it ­w atersrand in ca rd iovascu la r p h y s io l­o g y . H e is c u r r e n t ly an a s s o c ia te p rofessor in p hys io log y at the U n iv e r ­s ity o f the W itw a te rsrand , w ith p a rt ic u ­la r in te res t in ca rd iovascu la r p hy s io lo g y o f exercise.

32 IEEE EIIGIIIEERING IN MEDICINE AND OIOLOGY May/Juno 1997

Page 3: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

M a i/ in ./. Tu rn e r was born in South A fr ic a . H e re ce ive d h is B S c E lec Eng a l the U n iv e r­s ity o f the W itw a te rs - n m d in Johannesburg and there w en t on to com ple te his M S c Eng and PhD in in fan t resp i­

ra to ry gas-l'low measurements in 1991, w h ile le c tu r in g in the depa rtm en t, o f E le c tr ica l E ng ineering . He was attached to the research u n it o f the departm ent o f A naesthe tics at O x fo rd U n iv e rs ity in 1995. He is c u rre n tly an associate p rofes­sor in E lec trica l E ng ineering at the U n i­v e r s i t y o f th e W i tw a te r s r a n d , Johannesburg, w ith p articu la r interest in b io m e d ic a l c n g in e e iin g ca rd io re sp ira ­to ry concepts a pp lied to infants.

A d d re s s f o r C o rre s p o n d e n c e : C .G . H artfo rd , Departm ent o f Physio logy, U n i­v e rs ity o f the W itw a te rs ran d M e d ica l School, 7 Y o rk Road, Parktown, Johan­nesburg, 2193, South A frica .

References1. Coates A U Stocks J : Esophageal pressure ma­nometry in human infants. Peiliatr Pttbmmol 11: 350-360, 1991.2. CoiUcsAL, Stocks J,G erhardtT : Esophageal manometry, In: Stocks J, Sly PD, Tcppcr RS, Morgan WJ (EUs): Infant Respiratory Function Testing. John Wiley & Sons. Inc.. pp 241-258, 19%.,i. Muxted K,I,Shaw A, Macdonald TH : Choos­ing n catheter ..system lo r measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977,4. Kaplan M,(>nHoto I'M , Rultimanu LJE,ctal: Evaludtion of a gastric tube with esophageal bal­loon for neonatal use. Am J Perhuuolo# ? 1:355- 35V* 1984.5. Chavtiund l)A» Ye TH, Maarck ,IM, et al; Measurement of pleural pressure at low and high frequencies in normal rabbits. J A/;/)/ Physio! 63: 1142-1146,19^7,6. Charlrimd DA, Jvdoln Ct Couture J i Meas­urement of pleural pressure with oesophageal catheter-lip mleromanometerin anaesthetized liu- mans. Con J Anat'sih 38:518-21.1991.

7. iteardsniore CS, Wong YC, Stocks J, Silver- man M : Assessment o f the catheter tip pressure transducer for use in infant lung function tests. A W Biol Eng Compat 20: 657-660, 1982.8. Gardner RM: Accuracy and reliability o f dis­posable pressure transducers coupled with mod­ern pressure monitors. C rii Care Med 24: 379-882, 1996.9. Ycomattson CW, Kvans O il: The frequency response o f external transducer blood pressure measurement systems: a theoretical and experi­mental study. Clin Phvx Phvsiol A-ferr.v 4:435-449, 1983.10. Hartford CG, Rogers GG, Turner MJ: Suitability o f poly* myl chloride feeding catheters in catheter msrtumcicr systems. 18th International Conference o f the IEEE Engineering in Medicine and Biology Society, Amsterdam, The Nether­lands, Abstract 1.1.3-1, pp 37, 1996.

11. Hartford CG, Rogers GG, Turner M,1: Cor­rectly selecting a liquid-filled nasogastric infant feeding catheter to measure intra-esophagcal pressure. Pediatr Pulmmol In Press, 1997.12. HipkinsSF, Rutten AJ, Runcimnn WB; Ex­perimental analysis o f catheter-manometer sys­tems in vitro and in vivo. Anesthesiology 7U 893-906, 1989.13. Vanof HM , Rosen AL, McDonald NM, McDonald D At A critical study o f the responseof manometers to forced oscillations. Phys Med Biol 8(4);407-422, 1963.

14. Frank, O: Kritik dcr elaslischen manometer. ZBio l 44:445-613,1903.

15. Hansen AT, Warburg E: The theory for elas­tic liquid-containing membrane manometers: General part. Arm P lm M Scund 19:306-332, 1950.16. Hansen AT: The theory for elastic liquid- containing membrane manometers: Special part

“'Acta Physiol Scand 19:333-343,1950.

17. Fry DL: Physiologic recording o f modern in­struments with particular reference to pressure re­cording, Physio} Hev 40:753-788,1960.18. Thomson W T: Forced vibration with har­monic excursion, hi: Vibration Theory and Appli­cations. New Jersey, Prentice H all Inc., Engelwood Cliffs, pp 51-57, 1965.19. Kuo FF: The Laplace transform; amplitude, phase and delay, In; Network Analysis and Syn­thesis, John Wiley & Sons Inc. New York, pp 134-165,253-279. 1966.

20. M ilhorn H T : Transient analysis o f physical systems. In: The Application o f Control Theory to

Physiological Systems. WB Saunders Company. Philadelphia, pp 78-91, 1966.21. Hansen AT: Experimental examinations o f the theory for the elastic liquid-containing mem­brane manometer. In: Pressure Measurement in the Human Organism, Thesis. Copenhagen, Tek- nisk Forlag, Ac/n Physiol Scand 19:84-U5.1949.22. Parnell J, Beckman E L , Peterson LH ; Evaluation o f pressure transducer systems. Warminster report, National A ir Development Centre, NADC/MA5206 available through De­fense Technical Information Centre: .OTIC AD 00106V FT Belvoir, North Virginia. 1953.23. LeSouefPN, Lopes JM , England SJ, Bryan M H , Bryan AC: Influence o f chest wall distor­tion on oesophageal pressure. J Appl Physiol 55(2):353-358,1983.24. Beartismorc CS, Helms P, Stocks J, et al: Improved esophageal balloon technique for use in infants, P/i.vsiof 49:735-742.1930.25. Shapiro GG, Krovetz L.I: Damped and un­damped frequency responses o f underdamped catheter manometer systems. Am Heart J 80:226-236,1970.26. Luus HG, M u lle r FO, Meyer BH ; Statistical significance ver?,,'5 clinical relcvance; methods for calculating confidence intervals. South A fri­can Medical Journal 76(16):681-685. 1989.27. Zar JH: Biostatistical Analysis, Engelwood C liffs, New Jersey, Prentice-Hall. pp 206-252. 1974.28. Fry D L, Noble FW, Malles AJ: An evalua­tion o f modem pressure recording systems. 0 > - culation Research 5:40-46, 1957.29. Taylor M G : An experimental determination o f the propagation o f fluid oscillations in a tube with a visco-eUtstic wall; together with an analysis o f the characteristics required in an electrical ana­logue. Phys Med Biol 4:63-82, 1959.30. Latimer KE, Latim er RD: Measurements of pressure wave transmission in liquid-filled tubes used for intra^ascular blood pressure recording. Med Biol Eng 7:143-148, 1969.31. BlUict E, Colardyn F: Hazardous informa­tion from bedside rast-flush device test for fluid- filled pressure monitoring systems. Angioloyy 43(!2):988-995,1992.32. Scwhid HA: Frequency response evaluation o f radial artery catheter-manometer systems: si­nusoidal frequency analysis versus flush method. J CUn Monit 4 (3):18M 85,1983.33. Sheahan NFt Tuohy B, Kirkham R et al: The fast flush test; evaluation in radial arterx calhcler-manometer systems. Clin Phys Physbl Metis l2(3):247-252, 1991.

Moy/Juncl997 m ENGINEERING IN MEDICINE AND BIOLOGY 33

Page 4: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

Anesthesiology200T;92:550~8(p 2(KK) A m erican Society o f Anesthesiologists, Inc.Upplncott Williams & Wilkins, Inc,

Primate Pleuroesophageal Tissue Barrier Frequency Response and Esophageal Pressure Waveform Bandwidth in Health and Acute Lung InjuryCraig G. Hartford, M.D., M.Sc., Med.,* Johan M. van Schalkwyk, M.D., F.C.P.(SA).,1; Geoffrey G. Rogers, Ph.D.,' Martin J. Turner, Ph.D., Eng.*

Background: Dynamic Intraesophageal pressure (Pes) is used to estimate intrapleural pressure (Ppl) to calculate lung compli­ance and resistance. This study investigated the nonhuman primate Ppl-Pes tissue barrier frequency response and the dy­namic response requirements o f Pes manometers.

Methods: In healthy monkeys and monkeys with acute lung injury undergoing ventilation, simultaneous Ppl and Pes were measured directly to determine the Ppt-Pes tissue barrier am­plitude frequency response, using the swept-sine wave tech­nique, The bandw;ldths o f physiologic Pes waveforms acquired during conventional mechanical ventilation were calculated us­ing digital low-pass signal filtering.

Results} The Ppi-Pes tissue barrier is amplitude-uniform within the bandwidth o f conventional Pes waveforms in healthy and acute lung injury lungs, and does not significantly attenuate Ppl-Pes signal transmission between 1 and 40 Hz. At Pes frequencies higher than conventional clinical regions of interest the Ppl-Pes barrier resonates significantly, is pressure amplitude dependent at low-pressure offsets, and is signifi­cantly altered by acute lung injury.

• Lecturer, Associate Professor, Department o f Physiology, Univer­sity o f the Witwatcrsnmd Medical School and Johannesburg Hospital, Johannesburg, South Africa.

t Consultant, Department o f Anesthesiology, University o f the Wlt- watersrand Medical School and Johannesburg Hospital, Johannesburg, South Africa.

t Senior Researcher, Department o f Anesthesia, University o f Syd­ney, New South Wales, Australia.

Received from the Departments o f Physiology and Anesthetics, Univer­sity o f d ie Witwatersnnd Medical School and Johannesburg Hospital, Johannesburg, South Africa, and the Department o f Anesthesia, University o f Sydney, New South Wales, Australia. Submitted for publication March 19,1999. Accepted for publication September 30,1999. Supported by the South African Medical Research Council, Cape Town, South Africa; the Foundation for Research and Development, Cape Town, South Africa; the University o f the Wltwatersrand Research Council, Johannesburg, South A frin; and the Slcye Foundation, Joliann esburg, South Africa. Presented in p at the Institute for Electrical and Electronic Engineering (IEEE) Engl- mx/mg in Medicine and Biology Society annual conference, Hong Kong, China, October 29-November 1, 1998,

Address reprint requests to Professor Rogers: Department o f Physi­ology, University o f the Wltwatersrand Medical School, 7 York Road, Parktown, Johannesburg, 2193, South Africa. Address electronic mail to: [email protected]

Allowing for 5% or less Pes waveform error, the maximum Pes bandwidths during conventional ventilation were 1.9 Hz and 3.4 Hz for physiologic and extreme-cp.se waveforms in healthy lungs and 4.6 Hz and 8.5 Hz during acute lung injury.

Conclusions: In monkeys, the Ppl-Pes tissue barrier has a frequency response suitable for Ppl estimation dm isg low-fre- quency mechanical ventilation, and Pes manometers should have a minimum uniform frequency response up to 8.5 Hz. However, the Ppl-Pes tissue barrier adversely affects the accu­rate estimation of dynamic Ppl at lii^ i frequencies, with varied airway pressure amplitudes and offsets, such as the Ppl encoun­tered during liiglvfrequency oscillatory ventilation. (Key words: Lung clastance; manometry; respiratory mechanics.)

DYNAMIC intraesophageal pressure (Pes) is measure* t;o estimate intrapleural pressure (Ppl) for the calculation of dynamic transpulmonaty pressure and lung compli­ance.1 The accurate representation of dynamic Ppl using Pes measurements depends on the amplitude and phase frequency response of the intrapleural to intraesopha­geal tissue barrier (Ppl-Pes tissue barrier). This response is important for the accurate construction of dynamic lung compliance loops and resistance calculations and for the potential measurement of Pes during high-fre­quency oscillatory ventilation.2'3 One study directly com­pared Pes and Ppl in healthy adult dogs4; however, direct Ppl and Pes comparisons to determine the fre­quency response of the primate Ppl-Pes barrier have not been published.

Intraesophageal pressure manometers should have a uni form frequency response and linear phase shift over the dynamic Pes bandwidth. Frequency response characteris­tics of clinically used Pes manometers have been investi­gated5"7 but suggested minimum frequency response re­quirements for Pes manometers are based on dynamic airway pressure (Paw) waveform bandwidths.8 No data quantifying Pes bandwidth in primates have been pub­lished. Assumptions that Pes bandwidth is the same as Paw bandwidth may be invalid because P.’w frequency compo­nents transmitted to Pes may be amplified or attenuated.

Anesthesiology, V 92, No 2, Feb 2000

Page 5: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

PLEUBOESOPHAGEAL BARRIER FREQUENCY RESPONSE

We hypothesize that the Ppl-Pes tissue barrier has a Iform amplitude frequency response within the clini-

c.illy relevant Pes waveform bandwidth and is uniform, amplitude independent, and lung-condih n independent at low and high frequencies. We aim to determine the frequency response of the primate Ppl-Pes tissue barrier using simultaneous Ppl and Pes measurements in healthy mechanically ventilated monkeys, and those with acute lung injury (All), Furthermore, we aim m quantity the dynamic response requirements of Pes manometers by determining Pes bandwidth before and after A ll.

Method

InstrumentationTwelve female Vervet monkeys (C ercopithecus aepbi-

ops, 3.9 - 0.60 kg) were anesthetized (20 mg/kg ket- amine induction and 10 mg • h-1 • kg"*1 ketamine, > 30 jug • h ~1 • kg- 1 sufentanil [Janssen-Cilag, Johannesburg, South Africa], and 5 MS ’ h-1 • kg-1 adrenaline mainte­nance continuous infusions), paralyzed (10 MS ' h-1 • kg~1 vecuronium [Omnimed, Johannesburg, South Afri­ca]), intubated orally (cuffed 4.5 mm, Mallinkrodt, Ath- lone, Ireland), and mechanically ventilated in the supine position with warm humidified 100% 0 2 (fraction of inspired oxygen. fFioJ = 1.0) using a hgh-fiow-rate pres- sure-cycle pressure-limit ventilator (model 105; Preemi- care, San Antonio, Texas). University Animal Ethics Clearance (96/113/2B) was obtained. The right femoral vein was cannutated for injection of oleic acid to pro­duce ALL

Intrapleural Pressure and Intraesophageal Pressure Catheter Design and Pressure Measurements Intraesophageal pressure and Ppl catheters were

adapted for dynamic pressure measurements by sealing the tips of size 7 French gauge transducer-tip catheters (Millar Mikro-tip SPC 470; Millar Instruments, Inc., Hous­ton, Texas) into water-filled latex balloons 2 cm in length. The distal end of the Ppl transducer catheter was enclosed in a right-angled rigid perforated acrylic cylin­der, around which the balloon was tied. The cylinder was fashioned with a distal pointed end, such that the catheter would act as an introducer during transthoracic positioning, and, once inserted, the water-filled balloon rested parallel to the chest wall inside the pleural space. A second catheter attached to the Ppl catheter was 'inked to an underwater chest drain for pneumothorax deflation. The water volumes of the Ppl and Pes catheter

Anesthesiology, V 92, Mo 2, Feb 2001)

balloons were set to allow maximum balloon-catheter compliance. The frequency responses of the two cathe- teis were amplitude-linear and -equivalent.

With the aid of a transcutaneous cut-down to pleura and dilating trocars, the Ppl catheter was inserted trans- thoracically in either a cranial or a caudal direction into the left or right sides of the chest (n = 3 of each side per group) at the level of the seventh intercostal space in the mid axillary line. The in vivo Pes catheter site was chosen according to the best “paralyzed airway occlu­sion test," in which the airway is briefly occluded and abdominal pressure is applied to assess how faithfully APes reproduces APpl in an isovolumic chest of anes­thetized paralyzed subjects.9

Airway opening pressure was measured with a pi- ezoresistive differential pressure transducer (Micros­witch 170PC; Honeywell, Morristown, NJ) inserted into the proximal endotracheal tube perpendicular to the gas flow direction.

All pressure transducer signals were preamplified us­ing the same apparatus (Hellige Servomed, Freiburg, Germany) and digitized at 500 Hz (Biopac MP100 and AcqKnowif dge version 3.3.2; Biopac Systems Inc., Go- leta, CA).

Intrapleural-Intraesophageal Pressure TissueBarrier Frequency Response DeterminationIntrapleural pressure waveforms were generated by

applying sinusoidal Paw waveforms using the pneumatic driver unit of a high-frequency oscillatory ventilator (3100A; SensorMedics, Bilthoven, The Netherlands, and Manta Medical Systems, Johannes! urg, South Africa). To generate sine-wave rather than square-wave Paw out­puts, the square-wave frequency generator of the venti­lator was bypassed, and the DC-coupled amplifier was controlled by a separate sine-wave generator (Dynamic Signal Analyser 3562A; Hewlett Packard, Palo Alto, CA).

Amplitude frequency responses of the Ppl-Pes tissue barrier were calculated by comparing the Ppl and Pes waveform signals generated between 1 Hz (taken as the DC response to which the responses were normalized) and 40 Hz. Swept-sine waves10 were applied at 0.98 Hz intervals w:‘h a 20-s integration time and a 90% integra­tion threshold for each frequency interval (Dynamic Sig­nal Analyser 3562A). The dynamic signal analyzer per­forms a high-resolution Fourier transform at each measured frequency and extracts amplitude and phase information from the acquired Ppl and Pes waveform signals only at the frequency of interest, thereby ignor­ing any harmonics created by distortion. In each subject,

Page 6: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

, 552

HARTFORD ET AL.

10 swept-sine wave sequences (performed by discon­necting the standard ventilator and connecting the high- frequency oscillatory ventilator at 5-min intervals) were averaged at each of two mean Paw offsets: 1 cm HzO and 10 cm H20 above atmospheric pressure. These mean Paw pressure offsets were generated during the swept- sine measurements by injecting excess oxygen into die airway and allowing the excess to bleed off under water at the appropriate depth. Frequency response traces during which esophageal peristalsis occurred w e t: e-corded.

Frequency response measurements were recorded at two mean Paw offsets (1 or 10 cm H20 ) before and 2.5 h after intervention (oleic add (A ll group [AUG], n = 6) or saline (contro. group [CTRL], n = 6 injection). These measurements were recorded at a low applied Ppl am­plitude (low applied Ppl; APpl waveform mean ampli­tudes were set to be ± 1.0 cm H2Q at 1 Hz, reducing to ± 0.5 cm H20 at 40 Hz). At 7.5 h after intervention, the frequency response measurements were recorded at a mean Paw offset of 20 cm HzO at low applied Ppl amplitude, and again at Paw offset of 10 cm H20 , but at a larger applied Ppl amplitude (high Ppl amplitude; APpl waveform mean amplitudes were set to be ± 4.7 cm H20 at 1 Hz, reducing to ± 1.0 cm HaO at 40 Hz),

Intmesophageat Pressure Dynamic WaveformBandwidthIntraesophageal pressure and Paw measurements were

made in six subjects before (baseline, PesB, and PawB) and 5.5 h after (PesA and PawA) oleic acid injection in the AU group (n = 6). Respiratory rate (RR) and maxi­mum Paw (Paw pk) were adjusted to achieve an arterial carbon dioxide pressure (PaCO!) of 30-40 mmHg throughout the experiment (calibrated Stat Profile 3; Nova Biomedical, Waltham, MA), Physiologic PesB and PawB or physiologic PesA and PawA were acquired at a PaCOl of 30 - 40 mmHg. In addition, Pes and Paw traces were acquired during extreme conditions by raising peak Paw 50% above the physiologically required peak Paw and simultaneously doubling the respiratory rate (elevated PesB and PawB, or elevated PesA and PawA).

Trains of 10 PesB and PesA waveform sequences, each containing 10 mechanical breaths, were selected and low-pass filtered using a Blackman (AcqKnowledge ver­sion 3.3.2; Biopac Systems Inc.) finite impulse response linear phase filter with minimal phase shift, for which the response was tailored to be near —1 dB at chosen low-pass cut-off frequencies. Twelve low-pass cut-off fre­quencies were selected that reduced the areas under the

curve (%AUC) of the Pes continuous-power spectrum by predetermined proportions. Power spectra were deter­mined by ensemble-averaging of the power spectra of 10 PesB and PesA breaths.8 The mean value was subtracted from each ensemble and the data was padded with zeros before fast Fourier transform (AcqKnowledge version 3.3.2; Biopac Systems). One hundred percent of Pes waveform power was assumed to be contained between 0 and 40 Hz.

The average Pes amplitude between Pes at onset of expiration to Pes at onset of inspiration for each of 10 mechanical breaths was determined manually for the original trains of unfiltered Pes waveforms and com­pared with those of the incrementally low-pass-filtered Pes waveforms. The maximum waveform amplitude er­ror (% amplitude difference from original Pes waveform) averaged for 10 breaths was determined among the six subjects. The Pes bandwidth, up to which a Pes manom­eter should have a uniform amplitude frequency re­sponse to yield a 3% or less or 5% or less error when measuring the end-expiratory to end-inspiratory ampli­tude of physiologic or extreme-case Pes waveforms, was determined (Pes EI-EE bandwidth).

Statistical AnalysisAmplitude frequency responses in which the mean

Pes-Ppl ratio deviated by more than 10% from 1.0, plus the 95% confidence interval value excluded 1.0 (both conditions met), were considered to significantly deviate from that of uniformity,11 Significant witiiin-group (with­in the ALIG or within the CTRL) progressive changes were detected using Friedman analysis of variance (re­peated measures), followed by identification with the Wilcoxon signed rank test.12 The Mann-Whitney test was used to determine significant differences between the ALI and CTRL groups (Statistica; StatSoft, Tulsa, OK). Frequency response graphically displayed values are the mean ± SEM; all ventilation variables are the mean ± SD; and differences with P values < 0.05 are regarded as statistically significant.

Results

Intrapleural-Intraesophageal Tissue BanierFrequency' ResponseMean Pes-Ppl area under the curve ratios were more

than 0.90 in all groups for the anesthetized monkey Pes manometer occlusion tests (table 1). Figure 1 depicts the swept-sine amplitude frequency response at baseline in

Anesthesiology, V" 92, No 2, Feb 2000

Page 7: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

553

PLEUROESOPHAGEAL BARRIER FREQUENCY RESPONSE

Table 1. Anesthetized Monkey Occlusion Test Ratios

Baseline AUC (n = 12) Postlnterventlon AUC

CTRL (n = 6) AUG (n = 6)Paw Offset Pes/Ppl Paw/Ppl Pos/Ppl Paw/Ppl Pes/Ppl Paw/Ppl

1 cm H20 10 cm

H20

0.94 ± 0.067

0.90 + 0.080

0.95 ± 0.108

0.96 ± 0.065

0.90 ± 0.126 0.94 ± 0.061

0.90 ± 0.148 0.97 ± 0.163

0.95 ± 0.092

0.96 ± 0.059

0 30 ± 0.082

0.92 ± 0.077

AUC = area under the curve, ratio of two pressure waveform AUC values; Paw offset = airway pressure offset during occlusion test; Pes/Ppl = AUG ratio of esophageal to pleural pressure during occlusion test; Paw/Ppl = AUC ratio of airway to pleural pressure during occlusion tost; CTRL » control group receiving saline; ALIG = acute lung Injury group receiving oleic acid.

the combined (n = 12) ALIG and CTRL subjects (Paw offset = 1 cm HzO and 10 cm H20 ). The Ppl-Pes tissue barrier amplitude frequency response was uniform from 1 to 40 Hz when Paw offset was 10 cm H20 . However, the 95% confidence interval lower limit of the response was greater than unity and the mean response was greater than 1.10 at frequencies more than 14.7 Hz when Paw offset was 1 cm HzO. From 20.5-40 Hz, the ampli­tude frequency response when Paw offset was 1 cm H20 was significantly greater than when Paw offset was 10 cm H 20 (fig. 1).

After control saline injection the CTRL (n = 6) ampli­tude frequency response was uniform between 1 and 40 Hz at Paw offset of 10 cm H20 , and the amplitude gain was significantly greater at Paw offset of 1 cm HzO compared with Paw offset of 10 cm H20 between 19 and 35 Hz and 38 and 40 Hz (similar to the baseline values in fig. 1). However, unlike the CTRL, the ALIG

(n = 6) amplitude frequency response changed (fig. 2): At Paw offset of 1 cm H20 , the amplitude response was now uniform from 1 to 40 Hz (before lung injury, it was significantly raised; baseline, fig. 1) and, at Paw offset of 10 cm H20, the response was significantly resonant, from 13.7 to 15.7 Hz and from 26.4 to 32.2 Hz (before lung injury, it was uniform from 1 to 40 Hz; baseline, fig. 1). Within the ALIG, the amplitude response was signif­icantly different between Paw offset of 1 cm H20 versus

' Paw offset of 10 cm H20 , from 3.0 to 5.9 Hz and 8.9 to14.7 Hz and at 30.3 Hz (fig. 2).

At Paw offset of 20 cm H20, the CTPL amplitude response was uniform, from 1 to 40 Hz (similar to CTRL baseline at Paw offset of 10 cm H20 from fig. 1 and to CTRL after saline injection at Paw offset of 10 cm H20), whereas the ALIG amplitude response continued to de­viate from uniformity between 4.0 and 5 0 Hz and 30.3 and 36.1 Hz (resembling the AUG amplitude response at

Fig. 1. Baseline Intrapleural pressure (Pp!)-lntraesophageal pressure (Pes) tis­sue barrier amplitude frequency re­sponses at low and high mean airway pressure offsets. Hz = swept-slne fre­quency in cycles/s; Paw = airway open­ing pressure; Pes-Ppl = amplitude gain of esophageal pressure over intrapleural pressure. "P < 0.05 and " P < 0.01; ampli­tude gain significantly different for Paw offset o f 10 cm HzO versus 1 cm H20; +amplitude response deviates signifi­cantly from uniformity (see criteria in Methods).

2.0

- a - Pew offset = 1 cm H20 Paw offset -1 0 cm H20

+— +“ +“

i S "UJ 3 1.3 (0 3 < .15 1.?.

0.9

0.U

Swept-Sine Frequency (Hz)

Anestheslolog)\ V 92, No 2t Feb 2000

Page 8: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

554

HARTFORD ET AL.

2.0

- a - Paw offset ” 1 cm H20

Paw offset = 10 cm H,0% i.a %CL t.6

= 1.5o 1.4

< S uce ^® i.° g 0.9

U 0.6

0.7

Swcpt-Slne Frequency (Hz)

Fig. 2. Intrapleural pressure (PpO-iutrae- sophageal pressure (Pes) tissue barrier amplitude frequency responses at low and high mean airway pressure offsas after acute lung injury (AU; AH group). Hz = swept-stoe frequency in cycles/s; Paw = airway opening pressure; I'es- Ppl = amplitude gain o f esophageal pres, sure over intrapleural pressure. ’P <0.05; amplitude gain significantly differ­ent for Paw offset o f 10 cm HzO versus 1 cm H30; + amplitude response deviates significantly from uniformity (see crite­ria in Methods).

Paw offset of 10 cm H20 from fig. 2, which deviated from uniformity after AU).

At Paw offset of 10 cm H20 , but with high Ppl ampli­tude, CTRL was uniform between 1 and 40 Hz (similar to the baseline fig. 1 and postsaline CTRL responses at Paw offset of 10 cm H20 with low Ppl amplitude), whereas the ALIG significantly deviated from uniformity between23.5 and 31.3 Hz (similar to the AUG response at Paw offset of 10 cm H20 with low Ppl amplitude from fig. 2). Minimum to maximum phase differences were small: —11 to -1-26° among ail frequencies and lung conditions tested.

Intraesophageal Pressure Waveform EI-EEBandwidthThe physiologic Paw and Pes waveforms, as measured

before (physiologic PawB and PesB) and after lung injury (physiologic PawA and PesA) used for Pes EI-EE band­width determination are characterized in table 2. The effect of low-pass filtering of physiologic PesB wave­forms is shown in figure 3, Incrementally, low-pass til tering between the third and fourth (1,2 Hz) and the second and third (0.8 Hz) harmonic frequencies leads to errors in the end-inspiratory to end-expiratory waveform amplitudes.

The mean (n = 6 subjects) frequencies at which power in the Pes waves is 78-90% of total power are shown in figure 4. On average, among all the lung conditions, up to 90% and 78% of Pes waveform power is found to be less than 8.0 and 3.4 Hz, respec­tively,

Figure 5 shows the maximum Pes waveform error produced by low-pass filtering of Pes of healthy lungs (fig. 5A) and lungs with AU (fig. 5B), With increasing attenuation of Pes energy, the waveform error increases. The maximum waveform error showed a trend toward being larger for elevated PesB waveforms than for phys­iologic Pesr waveforms, The maximum waveform errors were larger i r PesA waveforms (after AU) than for PesB waveforms (at baseline; fig. SB, physiologic PesA to. fig. 5A, physiologic PesB).

Among Pes for all subjects, the highest cut-off frequen­cies (largest waveform BI-EE bandwidth) that yielded a more than 5% waveform error was 1.9, 3.4, 4,6, and 8,5 Hz for the physiologic PesB, elevated PesB, physiologic PesA, and elevated PesA waveforms, respectively, Corre-

Table 2, Physiologic Ventilation Characteristics for Pes EI-EE Bandwidth Determination

Before AU S.5 hr after AU

PawB PesB PawA PesA

RR 19 ± 4 .9 33 + 10.7Paw pis 12 ± 0.9 7 ± 2 .4 18 + 2,9 8.3 + 2.0PEEP 3 ± 1.0 1 a 1,9 5.5 + 1.3 3 ± 1 ,2Paw M 6 ± 0 .6 3 *2,2 10 + 1.3 5 ± 1.5

Values are those necessary for Pa002 of 30-40 mmHg.AU = oleic acld-4nduced acute lung injury; Paw pk.M <= peak and mean proximal airway pressure in cm H20; PawB, PawA = physiologic mean proximal airway pressure at baseline and In AU; PEEP «= positive end' expiratory pressure; PesB, PesA « physiologic mean intraesophageal pres­sure at baseline and In AU; RR = respiratory rate,

Anesthesiology, V 92, No 2, Feb 2000

Page 9: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

5 5 5

* PLEUROESOPHAGEAL BARRIER FREQUENCY RESPONSE

Fig. 3. Illustration of low-pass filtering effect on intraesophageal pressure (Pes) waveforms. Ovcrdamped Pes acquisi­tions result in gradual erosion of Pes waveform shape. Paw = original proxi­mal airway pressure waveform; Pes = original and two filtered (low-pass cut-off frequencies, 1.8 and 0.8 Hz, respectively) Intraesophageal physiologic pressure waveforms at baseline.

seconds10

spending values for a waveform error of more than 3% were 2.1, 4.9, 6.0, and 8.5 Hz.

Discussion

Intrapleural-Intraesopbageal Pressure Tissue Barrier Frequency Responses Concerns pertaining to Pes measurements are raised

through findings such as the lung volume dependency and Paw dependency of Pes changes in infants13 and adults,’4 and the frequency dependency and Paw de­pendency of lung compliance values in healthy adult

monkeys.15 The mechanical properties of the esopha­geal wall and surrounding structures are a potential cause of signal alteration when estimating dynamic Ppl from Pes.16

Our Ppl and Pes catheters are comparable to clinically used Pes air-balloon catheters but have a lower volume- dispiacement coefficient, are amplitude independent, and have a wider dynamic range suitable for assessing the Ppl-Pes tissue barrier at high frequencies. Using direct, simultaneous Ppl and Pes measurements, we have demonstrated that the amplitude frequency response of the Ppl-Pes tissue barrier is uniform within the EI-EE

Fig. 4. Frequency value o f power spec­trum o f Intraesophageal pressure (Pes) as a function o f Pes waveform energy con­tent, assuming 100% of waveform energy content lies between 0 and 40 Hz. AUC ™ % energy above 0 Hz remaining Under the power spectrum curve relative to that between 0 and 40 Hz (100%); Hz = aver­age frequency recorded for the particular area under the curve value. PesB, PcsA = mean Intraesophageal pressure wave­forms at baseline and after acute lung Injury, at physiologic (FaCOl of 30-40 mmllg) or artificially elevated airway pressures.

8.5

-e- Physiological PesB -o- Elevated PesB

Physiological PesA Elevated PesA7.0

4.0

3.5

3.0

Waveform Energy Content (%AUC)

Anusthcslology, V 92, No 2, Feb 2000

\

Page 10: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

556

HARTFORD ET AL

16

14* - Physiological PesB maximum orror i - Etevnted PesB maximum error

12

10

8

04

08 S 5 S S 6 S 8S S S

A Residua! W aveform Power (% AUC below cut-off frequency)

»- Physiological PesA maximum error i - Eievaled PesA maximum error&

max Hz 4.6

ICL

I3

B Residual Waveform Power (% AUC below cul-olf frequency)

Fig. 5. (A B) Intraesophageal pressure (Pcs) waveform error as a function of reducing Pcs waveform energies after low-pass filtering before t/t) and after (B) acute lung Injury. Waveform error (%) Is tlic maximum value for the end-lnsplratory onset to end-expiratory onset amplitude error after low-pass filtering at chosen cut-off frequencies defined from reducing waveform energy contents (reducing % areas under the curve). A ll = acute lung injury; max Hz = highest Pcs waveform frequency (largest waveform El-EE bandwidth) found at maximum wave­form errors o f 5% or more or 3% or more; PesB, PesA = mean Intraesophageal pressure waveforms at baseline (fop) and In Aid (bottom), at physiologic (Paco, o f 30-40 mmHg) or artifi­cially elevated airway pressures.

bandwidth (up to 8.5 Hz) of physiologic Pes waveforms measured in anesthetized mechanically ventilated pri­mates and with healtbv lungs and ALI. The barrier does not significantly attenuate Ppl-Pes signal transmission between I and 40 Hz. However, at higher frequencies in this range, the Ppl-Pes frequency response resonates

significantly, is amplitude dependent, and is significantly altered in the presence of lung disease.

Studies tiiat used direct Ppl measurements, but with markedly differing methodologies among the studies, have shown that the Ppl-Pes amplitude frequency re­sponse in dogs did not deviate by more than 10% from unity between 2 and 20 Hz, and no amplitude depen­dency was observed by varying the mean Paw offset.4 In rabbits there was slight attenuation of Pes measured at discrete frequencies of 30, 40, and 50 Hz.17

Alterations in the Ppl-Pes tissue barrier frequency re­sponse seen after induction of AJJ could be caused by changing lung conditions in A ll that affect Ppl-Pes trans­mission or by reduced amplitudes in the Ppl waveforms during the A ll swept-sine runs, i f the system is nonlin­ear. hi the current study, Ppl amplitude power spectra were similar before and after A ll; therefore, the alter­ations in the Ppl-Pes tissue barrier frequency response seen after induction of A ll probably are related to respi­ratory system A ll changes. Changes in the Ppl-Pes bar­rier could result from partial or complete isolation of the i Ppl or Pes catheters. This is also unlikely because we inserted the Ppl catheter among subjects into both the left or the right sides of the chest, in caudal or cranial directions, and the Ppl, Pes, and Paw occlusion test results before and after ALI were similar (table 1). The generation of negative pressures in the airways during oscillation, which occurred when mean Paw offset was 1 cm H20, may have altered the Ppl-Pes tissue barrier frequency responses. Inaccuracy of Pes measurements has been attributed to chest wall distortion producing uneven pleural pressure distribution,18 However, this is an unlikely contributing factor because the Ppl-Pes tis­sue barrier frequency responses at low frequencies did not depend on mean Paw (and thus on negative Paw) in our animals (fig. 1).

When Paw pressure offsets were switched from 1 to 10 cm H20, the Ppl-Pes tissue barrier frequency re­sponse changed, in both the CTRL and ALIG, although the 1- or 10-cm H20 Paw offsets at which the Ppl-Pes gain occurred is different in the ALIG vs, the CTRL. Such a change is not seen when the Paw pressure offsets were switched from 10 to 20 cm H20 . In addition, increasing applied Ppl amplitude at a Paw offset of 10 cm H20 did not significantly change the Ppl-Pes responses within the ALIG or the CTRL. To avoid the risk of Paw negative pressure trauma, we did not test the effect of a high Ppl amplitude when the Paw offset was 1 cm H20 , The amplitude dependency lies at lower Paw pressures (1-10 cm H20). These findings suggest that, in conditions in

Anesthesiology, V 92, No 2, Feb 2000

ean amm&aW W

Page 11: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

557

PLEUROESOPHAGEAL BARRIER FREQUENCY RESPONSE

which peak end-expiratory pressure or mean Paw are higher (such as occur during high-frequency osciUatory ventilation modes), it may be possible to correct the Pes waveform output using an esophageal transfer function to compensate for the Ppl-Pes tissue barrier gain (found during A ll at 10 and 20 cm H20 Paw offsets) because the Ppl-Pes frequency response exhibited a linear amplitude gain based on the 10- and 20-cm HzO Paw offsets.

Intraesqpbageal Pressure Waveform El-BEBandwidthThe bandwidth of pressure waveforms is influenced by

the type of transform used for frequency content analy­sis, characteristics of the chosen filters, and the criteria for waveform error.19 We based Pes waveform error on the Pm amplitude difference between end-inspiration and end-expiration because these are commonly used points of reference for lung compliance calculations in clinical settings.20 Using peak amplitude, or end-inspira­tory to end-expiratory amplitude, based on the first points of zero gas flow, as opposed to the method we chose, reduces the calculated Pes waveform error and, thus, plays down the potential distorting effect on dy­namic lung compliance loops. Our Pes EI-EE band- widths represent the maximum likely bandwidths of a range of Pes waveforms encountered in healthy mon­keys undergoing ventilation and monkeys with All.

The larger maximum EI-EE bandwidths noted in the supraphysiologic (elevated) PesB and PesA waveforms, compared with the physiologic PesB and PesA wave­forms, probably are caused by increased energy content found ut higher frequencies in some of the elevated Pes waveforms. H ie larger Pes EI-EE bandwidths noted after A ll (PesA) compared with healthy baseline lungs (PesB) may be caused by increased energy content at higher frequencies in some of the PeSA waveforms because of the higher Paw values necessary for A ll (table 2), or because of the A ll condition itself altering the Ppl-Pes transmission of waveform frequency components. The latter is unlikely because the frequency response of the Ppl-Pes tissue barrier was uniform up to the determined EI-EE bandwidths before and after A l l

Conclusion-iDirect simultaneous Ppl and Pes measurement reveals

that the Ppl-Pes tissue barrier has a uniform amplitude frequency response within the BI-EE bandwidth of con­ventional Pes waveforms in healthy lungs and A lt and

does not significantly attenuate Ppl-Pes signal transmis­sion between 1 and Hz. At Pes frequencies higher than conventional clinical regions of interest, the Ppl- Pes barrier resonates significantly, is pressure amplitude dependent at low pressure offsets, and is significantly altered by A l l

Allowing for 5% or less of Pes waveform error, the maximum Pes EI-EE bandwidths during conventional ventilation are 1.9 Hz and 3.4 Hz for physiologic and extreme-case waveforms in healthy lungs, and 4.6 Hz and 8.5 Hz during A l l For a 3% or less waveform error, the maximum Pes EI-EE bandwidth is 8.5 Hz.

In Vervet monkeys, the Ppl-Pes tissue barrier has a frequency response suitable for Ppl estimation during low-frequency mechanical ventilation, and Pes manom­eters should have a uniform frequency response up to8.5 Hz. However, the Ppl-Pes tissue barrier adversely affects the accurate estimation of dynamic Pp' at high frequencies, with varied airway pressure amplitudes and offsets, such as the Ppl encountered during high-fre­quency oscillatory ventilation. These findings may facil­itate the improvement of the accuracy of primate pul­monary function studies in hiyh-frequency respiratory mechanics.

The authors thank Ernest Somya and the University o f the Witwa- tersrand Central Animal Service, South Africa, for technical assistance.

References1. Coates AL, Stocks J: Esophageal pressure manometry In human

Infants. Pediatr Pulmonol 1991; 11:350-'602. Chartrand DA, Ye TH, Maarek JM, Chang HK: Measurement o f

pleural pressure at low and high frequencies In normal rabbits. J Appl Physiol 1987; 63:1142-6

3. P6siln R, Navnjas D, Rotger M, Varre IV Validity o f the esophageal balloon tedintque at high frequencies, J Appl Physiol 1993; 74(3): 1039--14

4. Nagels J, landser FJ, van dc Linden L, Clement J, van Woestijne KP: Mechanical properties o f lungs and chest wall during spontaneous breathing. J Appl Physiol 1980; 49:408-16

5. Hartford CG, van StimlkwykJ, Turner Ntf, Rogers GO: Application o f curve fitting to air-balloon and liquid-filled catheter frequency re­sponse traces, IEEE Eng Med Biol Mag 1997; 16:27-33

6. Hartford CG, Rogers GG, Turntiv M): Correctly selecting a fluid- filled nasogastric Infant feeding catheter to measure Intra-esophageal pressure. Pediatr Pulmonol 1997; 23:362-9

7. Hartford CG, Rogers GG, van Sclialkwyk J, Turner MJ: Fre­quency responses o f Infant air-balloon versus liquid-filled catheters fo r intra-esophagesl pressure measurement. Pediatr Pulmonol 1997; 24:353-9

8. Turner MJ, Davies VA, Ravel TJL: Bandwidths o f respiratory gas flow and pressure waveforms In mechanically ventilated infants, Physto! MeaS 1983; 14',419-3V

Anesthesiology, V 92, No 2, Feb 2000

Page 12: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

55 8

HARTFORD ET AL

9. Lanteri CJ, Kano S, Sly PD: Validation o f the esophageal pressure occlusion test after paralysis, Pediatr Pulmonol 1994; 17:56-62

10. Hipkins SF, Ruttcn AJ, Runciman WB: Experimental analysis o f catheter-manometer systems in vitro and in vivo. Anesthesiology 1989; 7 1 :8 9 3 -9 0 6

11. Luus HG, Muller FO, Meyer BK: Statistical significance versus clinical relevance; methods for calculating confidence intervals. S Afr Med J 1989; 76(l6);681-5

12. Glanrz SA: Confidence intervals, Alternatives to analysis o f vari­ance and the t-test based on ranks, Primer o f Biostatistics. New York, McGraw-Hill, 1992, pp 188-217, 320-63

13. Beardsmore CS, Stocks J, Silverman M: Esophageal pressure in infants at elevated lung volumes and positive airway pressure. J Appl Physiol 1983; 55:377-82

14. Baydur A, Cha E, Sassoon CSH: Validation o f esophageal balloon technique at different lung volumes and postures. J Appl Physiol 1987; 65:315-21

15. Wcgner CD, Jackson AC, Berry JD, Glllepsie JR: Dynamic respi­ratory mechanics in monkeys measured by forced oscillations. Respir Physiol 1984; 55:47-61

16. Dechman G, Sato J, Bates JHT: Factors affecting the accuracy o f esophageal balloon measurement o f pleural pressure in dogs. J Appl Physiol 1992; 72:383-8

17. Kaplan M, Galioto FM, Ruttimann UE, Hamosh P, Avery GB: Evaluation o f a gastric tube w ith esophageal balloon for neonatal use. Am J Perinatology 1984; 1:355-9

18. LeSouef PN, JM Lopes, SJ England, HM Bryan, CA Bryan: Influ­ence o f chest wall distortion on esophageal pressure. J Appl Physiol 1983; 55:353-8

19. Paulsen AW: Implications fo r clinical monitoring o f intra-arterial blood pressure based on the frcquenqr content o f worst-case pressure waveforms. Biomed Instrum Technol 1993; May;217-34

20. Ahrens TS: Effects o f mechanical ventilation on hemodynamic waveforms. Crit Care Nurse Clin North Am 1991; 3(4):629-39

Anesthesiology, V 92, No 2, Feb 2000

Page 13: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

THESIS ERRATA

Page viiAspects of the research performed in this thesis have also been presented by the candidate as a guest speaker at two international universities in 2000:(i) City University of New York, Biomedical Engineering Department, seminar.(ii) The University of Kent at Canterbury Electronics Engineering Department,

Medical Electronics Division, colloquium.

Page ixFINANCIAL SUPPORT Grants:The South African Foundation for Research and Development. Medical Research Council of South Africa. fris-EHen Iris Ellen Hodges grant. H.E. Griffin grant.

Chapter 3, page 85BOTTOM: Supplemental Figure S3.BAir-filled balloon-tip catheter ready for frequency response characterisation Size 10 French gauge polyvinyl chloride catheter

Chapter 5, page 161Supplemental Figure S5.BRespiratory instrumentation and monitoring during acute lung injury.I. Endotracheal tubeII. In-line inspiratory oxygen temperature and humidity controlIII. Pressure cycle pressure limit ventilator (PEEP, pkpeak Paw, RR)IV. Proximal airway pressure (differential pressure transducer, Paw)V. Flowmeter (screen pneumotachograph connected to Validyne , ressure

transducer, Faairway flow)VI. Animal End Tidal C02 monitor (sidestream, rapid response ETGOg)VII. Plethysmographic tongue monitor (arterial % haemoglobin oxygen

saturationSaGg)

Chapter 5, Page 156; Chapter 6, Pages 217, 220, 223Airway-opentng-pfesstife Proximal airway pressure (Paw)

Chapter 6, Page 203The right femoral vein was cannulated for injection of oteie-aeid saline vehicle or oleic acid (OA) to produce ALL

Page 14: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

THESIS ADDENDUM

Chapter 2, Page 6 7 and Chapter 3f page 111

Notes on the pressure offset and input pressures applied to the in

vitro model

In the in vitro model used in this thesis the term “pressure offset" referred to is

comparable to the term mean airway pressure due to mechanical ventilation in vivo,

and the term "input pressure" is comparable to the a Paw applied oscillating about

the mean Airway pressure in vivo. Dynamic in vivo Pes is dependent on the applied

Paw (mean and peak to peak), PEEP and CLdyn. Thus Pes may be negative in

spontaneously breathing subjects reflecting a negative intrathoracic pleural pressure

during inspiration or the negative intrathoracic pressure at end-expiration in non-

PEEP mechanical ventilation. Pes may also be positive during positive pressure

ventilation or forced expiration. In order to capture a broad rangs of Pes we applied

a positive offset pressure since most infants are ventilated using PEEP, and

oscillated the pressure in a sinusoidal fashion about this mean offset pressure to

include negative and positive Pes pressures.

The values chosen for mean pressure offset and peak to peak oscillating pressures

in vitro were initially determined by consulting a paediatric intensive care physician

from a large intensive care unit routinely managing respiratory distress syndrome,

prior to commencing the experiments of Chapters 2 and 3. Advice was obtained that

Page 15: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

a PEEP of 5 cmH20 and up to 40% transmission of peak Paw to Pes could be

expected. These values are confirmed in this thesis (Chapters 5 and 6 , see below)

and are similar to those reported in the literature where Ppl and Pes magnitude is

cited as being up to - 6 cmH20 in the negative direction^, which is similar to the

lower limit applied in Chapters 2 and 3, and anything up to + 30 cmH20 in the

positive direction, dependent on the applied Paw and lung compliance (a'b,c).

Chapters 5 and 6 of this thesis demonstrate in vivo that the expected magnitude of

Pes pressures may vary depending on the applied Paw, the disease state of the

lungs and the site of the Pes measurement. From Chapter 6 the mean % delivery of

dynamic Paw to Pes, defined by Pes AUC/Paw AUC for 10 respiratory cycle

ensembles was 64 ± 7% and 62 ± 8.7% for physiological PesB/PawB and elevated

PesB/PawB respectively. Following ALI, % delivery values were reduced to 41 ± 6.3

and 43 ± 7.7 for physiological PesA/PesA and elevated PesA/PawA respectively.

Thus ALI reduces the % delivery of Paw to Pes, but doubling the RR simultaneously

and raising the Paw above that required for physiological PaC02 did not significantly

affect the % delivery of Paw to Pes, before or after ALI (Chapter 6, Page 226).

Physiological (= mechanical ventilation parameters set to yield a normal PaC02)

peak Pes at baseline and in acute lung injury was 7 ± 2.4 and 8.3 ± 2.0 cmH20

respectively, and the corresponding mean Pes was 3 ± 2.2 and 5 ± 1.5 cmH20.

Thus the -5 to +15 cmH20 peak to peak oscillating pressure chosen for the in vitro

model of chapters 2- 4 comfortably covered the likely upper and lower levels (twice

the SD values) of the expected mean and peak Pes values during mechanical

Page 16: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

ventilation, and the in vitro sinusoidal waveforms produced using a loudspeaker

were suitably minimally distorted at this magnitude.

<a) Beardsmore CS, Helms P, Stocks J, Hatch DJ, Silverman M. Improved

esophageal balloon technique for use in infants, J Appl Physiol 1980;49:735-

42.

(b) Coates AL, Stocks J: Esophageal pressure manometry in human infants.

Pediatr Pulmonol 1991; 11:350-360

!c) Stocks J, Sly PD, Tepper RS, Morgan WJ: Infant Respiratory Function

Testing. New York, U.S.A., Wiley-Liss, Inc., 1996. Pages 242, 243, 248.

Chapter 4

Notes on the clinical relevance of an invalid second order system

assumption

Chapter 4‘s objective is clearly stated as being aimed at testing whether or not the

second order assumption significantly affects the determination of certain frequency

response variables, rather than whether or not the invalidity of this assumption will

impact on the frequency response regions of clinical interest. Regions of clinical

interest are highly variable with requirements differing between, for example,

laboratory based investigators and clinicians or the type of waveforms being

assessed (HFOV or conventional).

-iv-

Page 17: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

Perhaps the most useful variable measured in Chapter 4 indicative of whether or not

the invalidity of the second order assumption could be clinically relevant is the

SOFC:ORC at fA5 (the lowest frequency at which the mean amplitude frequency

response deviates by >= 5% from uniformity). This is a variable impacting on the

lower end of the frequency response spectrum of the catheters: most physiological

pressure waveforms' energy contents lie well within fr, but not necessarily within fA5

of a catheter-manometer system. Chapter 4 shows that by employing the second

order assumption essentially all the liquid-filled and air-balloon catheters’ resulting

fA5 values are statistically significantly incorrect (Chapter 4, Page 136). “Significant”

here was defined as SOFC:ORC fA5 ratios with confidence intervals which excluded

the ideal ratio of 1.0 and which additionally had mean ratios s 0.95 or ;> 1.05. Note

that (i) the significance definition used by the author here is a modest criterion only

and (ii) Figure 4.2 (Chapter 4, page 136) presents the SD values (indicated in the

legend) rather than the 95% confidence intervals or maxima. Purists might argue

that the upper limits of the ratios, rather than the means, should be presented to

determine the magnitude of the fA5 inaccuracy. Such an approach would have a

large effect on increasing the fA5 value’s reported inaccuracy and inter alia on

predicting the clinical relevance of the inaccurate fAg.

In summary, using only a modest statistical significance criterion rather than a

stringent one, and presenting the mean fA5 inaccuracies rather than the maximal

inaccuracies, the fA5 of both liquid and air-balloon catheters was shown to be

incorrectly estimated when using the second order assumption. The magnitude of

this inaccuracy was significantly different for liquid vs balloon catheters, being larger

Page 18: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

for liquid filled catheters. The latter catheters are those in which the second order

assumption techniques are more likely to be applied in clinical practice. The fA5 was

also overestimated rather than underestimated; overestimation is potentially of

greater clinical concern. However, the exact impact of these findings on clinical Pes

measurements is hard to predict since to quantify this the extremes rather than the

means of inaccuracy may need to be considered (worst-case observations from both

the catheters original fA5 values in Chapters 2 and 3, as well as the SOFC:ORC fA5

ratios from Chapter 4), and the nature of the Pes waveforms to be measured

clinically would need to be defined upfront. This is beyond the scope of Chapter 4's

objectives and in any event th6 iA5 inaccuracy problem can be avoided by using an

in vitro sine wave generator technique; this remains the concluding recommendation

based on Chapter 4's findings.

Chapter 5, Page 161, Figure S5.B.

Notes on the Sa02 monitoring

This is a supplemental figure included to show the scope of measurements and

monitoring required to maintain physiological stability in this experimental model of

acute respiratory distress syndrome. Sa02 is mentioned in the legend to the figure

and was monitored as a routine (late) warning device whilst an arterial line was

being established or re-established. Suitable in vivo calibration of Sa02

measurement apparatus remains an unsolved problem to date and this

unbenchmarked parameter should probably not be used as a valid scientific

endpoint in critical care illness laboratory-based studies. The measurement error in

Page 19: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

Sa02 readings increases with reducing Sa02, is site-dependent (we found that the

monkey tongue was the only position which yielded consistent signals), and the

algorithm generating the Sa02 reading may be species and age dependent. Our

Sa02 tracings (Nellcor pulse oximeter) were digitally captured, however the more

accurate Pa02 was presented as the variable indicative of arterial oxygenation

(measured using a fully calibrated Stat Profile 3, Nova biomedical, MA, USA) as per

chapter 5, page 167.

Chapter 5, page 174, Table 5.2 and Chapter 6, page 226, Table 6.2

Notes on the different % transmission of Paw to Pes values

presented in these tables

The minor differences in % transmission of Paw to Pes between Chapters 5 and 6

were not due to th~ measurement systems since in both chapters the same animals,

Paw transducer and Pes water-filled balloon-tipped transducer tip catheter were

used. The minor differences arise from the methodology used to determine % Paw

to Pes transmission. In Chapter 5 the Paw and Pes delta end-inspiration to end-

expiration value was measured, over 3 respiratory cycle ensembles (Chapter 5,

page 167, paragraph 2), at physiological ventilation parameters, whereas in Chapter

6 the Paw and Pes AUC was used, over ten respiratory cycle ensembles (Chapter 6,

page 226), at physiological and elevated ventilation parameters. Despite these

methodology differences the values between the two chapters are similar in

magnitude and direction.

Page 20: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

Chapters 5 and 6

Notes on the use of laboratory-modified micro-tip transducers to

measure Pes and Ppl.

Negative reports(a,b,c'd) currently outweigh positive reports in the literature concerning

the ease of use or the validity of using Pes micro-tip transducers to estimate Ppl for

lung compliance determination. However, in none of these validation reports is Ppl

measured directly and the validity of Pes microtip transducers is largely questioned

based on the relatively static occlusion test or based on comparisons with

conventional fluid-filled catheters.

The author cannot entirely exclude the possibility that in Chapter 5 the CLdyn

measurements could be relatively different to those which may have been obtained

had conventional fluid-filled catheters been used to measure Pes. However, such an

effect would be minimal for the following reasons:

The purpose of this experiment was to compare the Cudyn of control against

acute lung injury subjects. The same Pes measurement apparatus was used

in control and experimental subjects and the accuracy of the absolute values

is thus less important.

The laboratory-modified micro-tip transducer catheters were also validated in

vivo in the same fashion as that used to validate conventional fluid-filled

catheters in vivo using the occlusion test (Chapter 6, page 215, Table 6.1).

-v i i i -

Page 21: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

Additionally, any differences between the Pes microtip transducer used and

conventional Pes fluid-filled catheters would not have affected the CRgdyn

values reported in Chapter 5, pages 174 and 186.

Furthermore, the baseline CTRL and ALIG CLdyn values obtained are similar

to those of human infants<e) and CRSdyn or CLdyn values resulting from acute

lung injury fall within the ranges described for newborn infants with RDS<0 and

for low birth weight infants with severe RDS(9>.

In Chapter 6 Pes and Ppl was measured directly, using laboratory-modified microtip

catheters with similar catheters’ structural properties on both sides of the

transmission barrier (Ppl-Pes) undergoing evaluation. Therefore concerns raised

about the validity of these results based on the inherent structural properties of

micro-tip catheters' potential to interfere with the Ppl and Pes measurements are

minimised. One obvious problem with micro-tip transducer catheters in comparison

with conventional fluid-filled catheters is the reduced intra-esophageal volume over

which they measure Pes. This was not a factor in the author’s experiments since the

micro-tip transducer catheters were adapted in the laboratory to have a water-filled

balloon surrounding, unlike conventional micro-tip transducer catheters. The Pes

and Ppl manometers used were validated in vitro (Chapter 6, page 213) and equally

exhibited - i , exceptionally wide bandwidth, were linear over a large amplitude, and

yet measi/w j dynamic pressure over a large portion of the esophagus and pleura,

with which to characterise the Ppl-Pes tissue barrier.

Page 22: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

In Chapter 6 the "unconventional” laboratory-modified micro-tip transducer catheters

were deliberately used (for reasons outlined in Chapter 6, page 238) to measure

dynamic Pes and Ppl at low and high frequencies. These laboratory-modified

catheters were validated in vitro, but not in vivo against conventional catheters:

Although it may be possible to validate in vivo the laboratory-modified

catheters against conventional fluid-filled Pes catheters for respiratory

waveforms of low bandwidth, this would not be possible for waveforms with

higher frequency contents since conventional fluid-filled catheters do not

exhibit a uniform amplitude frequency response (Chapters 2 and 3).

Note too that conventional fluid-filled Pes catheters' dynamic Ppl estimations

have themselves not been validated against direct Ppl measurements

anyway.

As per above, the laboratory-modified micro-tip transducer catheters were

validated in vivo for Pes measurement in each subject in the same fashion as

that used to validate conventional fluid-filled catheters in vivo using the

occlusion test (Chapter 6, page 215, Table 6.1).

,a) Gappa M, Jackson E, Pilgrim L, Costeloe K, Stocks J. A new microtransducer

catheter for measuring esophageal pressure in infants. Pediatr Pulmonol

1996;22:117-24.

-x -

Page 23: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

Chartrand DA, Ye TH, Maarek JH, Chang HK. Measurement of pleural

pressure at low and high frequencies in normal rabbits. J Appl Physiol

1987;63:1142-114'’

Chartrand DA, Jodoin C, Couture J. Measurement of pleural pressure with

oesophageal catheter-tip micromanometer anaesthetized humans. Can J

Anaesth 1991;38:518-21.

Beardsmore CS, Stocks J, Silverman M. Esophageal pressure in infants at

elevated lung volumes and positive airway pressure, J Appl Physiol

1983;55:377-82.

Hjalmarson, 0 . 1974. Mechanics of breathing in newborn infants with

pulmonary disease. Acta Paediatrica Scandanavica 247:29,34.

Teague, W.G., A. Robert, A. Darnall. 1982. A noninvasive constant-flow

method for measuring respiratory compliance In newborn infants. Grit, Care

Med. 13:965-971,

Thomson, A., J. Elliott, M. Siiverman. 1983. Pulmonary compliance in sick low

birthweight infants. Arch. Dis. Child. 58:891-896.

Page 24: ‘ I...ing n catheter ..system lor measuring, imra*. esophageal pressure. Med Mol Eng Coniput 15: 398-401,1977, 4. Kaplan M,(>nHoto I'M, Rultimanu LJE,ctal: Evaludtion of a gastric

Author Hartford C G F

Name of thesis Accuracy Of Intra-Oesophageal Dynamic Pressure Probes, And Their Validity In Determining Dynamic

Intrapleural Pressure Hartford C G F 2000

PUBLISHER: University of the Witwatersrand, Johannesburg

©2013

LEGAL NOTICES:

Copyright Notice: All materials on the Un i ve r s i t y o f the Wi twa te r s rand , Johannesbu rg L ib ra ry website are protected by South African copyright law and may not be distributed, transmitted, displayed, or otherwise published in any format, without the prior written permission of the copyright owner.

Disclaimer and Terms of Use: Provided that you maintain all copyright and other notices contained therein, you may download material (one machine readable copy and one print copy per page) for your personal and/or educational non-commercial use only.

The University of the Witwatersrand, Johannesburg, is not responsible for any errors or omissions and excludes any and all liability for any errors in or omissions from the information on the Library website.