A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY Mr Drago GROSELJ Dr Jérôme DUVERNOY Dr John GORMAN www. arso . sigov . si www. meteo . fr www. bom . gov .au Drago . Groselj @ gov . si Jerome. Duvernoy @ meteo . fr [email protected]

A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

Embed Size (px)

DESCRIPTION

A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY. Mr Drago GROSELJ Dr Jérôme DUVERNOY Dr John GORMAN www.arso.sigov.si www.meteo.fr www.bom.gov.au [email protected] [email protected] [email protected]. - PowerPoint PPT Presentation

Citation preview

Page 2: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

2

Specialization

PRESSURE

TEMPERATURE

HUMIDITY

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Page 3: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

3

PRESSURE

Primary Pressure Standard

M = Mass

P = Piston

C = Cylindre

Weight = Pressure

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Page 4: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

4

PRESSURE

Primary Pressure Standards

Dead weight gauge

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Page 5: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

5

PRESSURE

Advantages:

– Traceability to mass and length

– Long term stability

– Accuracy, reproducibility, reliability and stability

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Page 6: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

6

PRESSURE

Disadvantages:

– Time consuming

– Measurements corrections (T, g, U…)

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Page 7: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

7

PRESSURE

Secondary Pressure Standard

Quartz Barometer

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Page 8: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

8

PRESSURE

Advantages:

– Faster and easier to use

– Easy to adapt to automatic operation

– Generally less expensive

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Page 9: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

9

PRESSURE

Disadvantages:

– Not a basic measurement

– Higher measurement uncertainty

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Page 10: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

10

PRESSURE

Travelling pressure standard

An example of electronic barometer

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Page 11: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

11

PRESSURE

Calibration systems

example of a barometric chamber

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Page 12: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

12

PRESSURE

Calibration systems

example of a pressure generator

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Page 13: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

13

TEMPERATURE

EIT 90

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

METEOROLOGY

Page 14: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

14

TEMPERATURE

EIT 90

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Gallium

Mercury

Water

Page 15: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

15

TEMPERATURE

Standard Platinum Resistance Thermometer (SPRT)

Example of an SPRT

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Page 16: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

16

TEMPERATURE

Means of calibration

Stirring bath

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Page 17: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

17

TEMPERATURE

Means of calibration

Climatic chamber

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Page 18: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

18

HUMIDITY

Humidity Generators

Two Pressure Humidity Generator

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Page 19: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

19

HUMIDITY

Humidity Generators

Split stream system

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Page 20: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

20

HUMIDITY

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Climatic Chambers

Environmental Chamber

Page 21: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

21

HUMIDITY

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Climatic Chambers

Split stream chamber

Page 22: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

22

HUMIDITY

Salt Solutions Generator

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Salt/Temperature oC 5.0 10.0 15.0 20.0 25.0

Lithium chloride 11.3 11.3 11.3 11.3 11.3

Magnesium chloride 33.6 33.5 33.3 33.1 32.8

Potassium carbonate 43.1 43.1 43.1 43.2 43.2

Sodium bromide 63.5 62.2 60.7 59.1 57.6

Sodium chloride 75.7 75.7 75.6 75.7 75.3

Potassium chloride 87.7 86.8 85.9 85.1 84.3

Potassium sulphate 98.5 98.2 97.9 97.6 97.3

Page 23: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

23

HUMIDITY

Salt Solutions Generator

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

The salt solution system at Meteo France

Page 24: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

24

HUMIDITY

Approximate Uncertainties

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Page 25: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

25

HUMIDITY

Recommendations– The target uncertainty within a calibration laboratory = 2 % RH (0.3 oC dew point)

– Labs should have a humidity generator or chamber:

within the range 5 % RH at 15 oC to 90 % at 40 oC.

– The streams or parcels should be stable to within ± 0.2% RH per minute

– If the source of humidity is not a first principles device, then measurement of the humidity should be with a dew point hygrometer with an 95 % uncertainty of 0.2oC or better

– Calibration of field humidity probes should be done with a view to maintaining a low overall uncertainty. The whole of measurement cycle, including inspection frequency, must be considered and appropriate levels of calibration uncertainty applied

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Page 26: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

26

CONCLUSION

Some possibilities in terms of calibration. This first review is not exhaustive Every other remaining devices, standard, generator not presented

here are still acceptable Metrology is the science of measurement Expression of the uncertainty

ENVIRONMENTAL AGENCY OF THE REPUBLIC OF SLOVENIA

Page 27: A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

27

THE END