<GIFTS_T_Uncert&Cal.ppt>9 March 2004
GIFTSBlackbody SubsystemCritical Design Review
Temperature Uncertainty Budget and Calibration
Fred Best
9 March 2004
<GIFTS_T_Uncert&Cal.ppt>9 March 2004
Slide 2
Temperature Error BudgetTemperature Uncertainty 3 sigma error [K] RSS [K]
Temperature Calibration Standard* 0.005(Thermometrics SP60 Probe with Hart Scientific 2560 Thermistor Module)
0.005
Readout Electronics Uncertainty (at delivery)* 0.0140.014
Blackbody Thermistor Temperature Transfer UncertaintyGradient Between Temperature Standard and Cavity Thermistors** 0.020Calibration Fitting Equation Residual Error** 0.005
0.021
Cavity Temperature Uniformity UncertaintyCavity to Thermistor Gradient Uncertainty (1/3 of total max expected gradient) 0.030Thermistor Wire Heat Leak Temperature Bias Uncertainty* 0.008Paint Gradient (assumes nominal HBB Temp and conservative viewing geometry) 0.018
0.036
Long-term StabilityBlackbody Thermistor (10 years of drift assuming 105 C) 0.050Blackbody Controller Readout Electronics* 0.012
0.051
Effective Radiometric Temperature Weighting Factor UncertaintyMonte Carlo Ray Trace Model Uncertainty in Determining Teff 0.030(1/3 of total max expected gradient) 0.030
*Presented in earlier section 0.074**Based on experience with UW AERI Blackbodies
Blackbody Readout Electronics Uncertainty
<GIFTS_T_Uncert&Cal.ppt>9 March 2004
Slide 3
Blackbody Section – Side View
Enclosure Tube(Aluminum)
FrontCover(Alum)
Multi-layer Insulation(not shown)
Cavity Support/Insulator (Noryl)
ElectricalConnector
MountingBase (Noryl)
Circumferental Heater
Thermistors 4 Temperature Measurement 2 Redundant Temp Control 1 Overtemp
Blackbody Cavity(Aluminum)
ChemglazeZ306 Paint
<GIFTS_T_Uncert&Cal.ppt>9 March 2004
Slide 6
Blackbody Temperature Calibration
• The Blackbody and Controller will undergo end-to-end calibration.
• The temperature of the Blackbody and Controller can be independently controlled.
• Ambient Blackbody calibrated over the range from: -40 C to 0 C.
• Hot Blackbody calibrated over the range from: 0 C to +40 C.
• For calibration the Controller will be maintained at near 0 C. {The system will be calibrated at other Controller temperatures if a temperature dependence is discovered during EM-2 testing.
<GIFTS_T_Uncert&Cal.ppt>9 March 2004
Slide 7
Temperature Calibration Standard
AbsoluteTemperatureCalibration of end-to-end System at Hart has UncertaintyOf +/- 0.003 K (k=2)
ThermometricsSP-60 Probes (2)
• Hart 2563 Standard Thermistor Module• Hart 2564 Thermistor Scanner Module• Hart 3560 Extended Communications Module• Calibration Computer